0% found this document useful (0 votes)
577 views17 pages

Fluke 37

The Fluke 37 is a rugged, portable multi meter ideal for use on the bench or in the field. It combines the performance and accuracy of a digital meter with the speed and dynamic measurement capability of an analog meter. The 3200-count, digital display offers better resolution than a conventional 31f2-digit, 2000-count display. The Touch-Hold mode locks the measurement into the display for viewing and automatically updates the display when a new measurement is taken.

Uploaded by

bnewport1
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
577 views17 pages

Fluke 37

The Fluke 37 is a rugged, portable multi meter ideal for use on the bench or in the field. It combines the performance and accuracy of a digital meter with the speed and dynamic measurement capability of an analog meter. The 3200-count, digital display offers better resolution than a conventional 31f2-digit, 2000-count display. The Touch-Hold mode locks the measurement into the display for viewing and automatically updates the display when a new measurement is taken.

Uploaded by

bnewport1
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 17

37 MULT/METER

OPERATOR'S MANUAL MANUEL D'INSTRUCTION BEDIENUNGS-HANDBUCH MANUAL DE INSTRUCCIONES MANUALE D'IMPIEGO



lfXt&~~iI

FLUKE

n::an ucu

~

rn UL

SEE PAGE 2 VOIR PAGE 16 SIEHE SEITE 30 VEOI PAGINA 46 VER PAGINA 60 74 "'-;) ••

37 MULT/METER

English

Francais

Deutsch

Italiano

Espariot

ENGLISH

Operating Instructions



This meter has been designed and tested according to lEe Publication 348. Safety Requirements for Electronic Measuring Apparatus. This manual contains information and warnings which must be followed to ensure safe operation and retain the meter in safe condition.

INTRODUCTION

The Fluke 37 is a rugged, portable multi meter ideal for use on the bench or in the field. It provides unsurpassed performance and input protection in the most demanding working conditions.

The Fluke 37 combines the performance and accuracy of a digital meter with the speed and dynamic measurement capability of an analog meter. The 3200-count, digital display offers better resolution than a conventional 31f2-digit, 2000-count display, and the 31-segment analog bar graph display provides quick and easy dynamic measurement indications. In addition, a unique Touch-Hotd'vrnode allows you to watch the probes during critical measurements. The Touch-Hold mode locks the measurement into the display for viewing and automatically updates the display when a new measurement is taken. These and other useful features of the Fluke 37 are explained in detail in this manual.

MUL TIMETER SAFETY

Read this information before using the meter. WARNINGS denote hazards to the operator. CAUTIONS denote hazards to the meter. The following safe practices and proper operating procedures should be followed when using any multimeter:

• Inspectthe test leads for insulation damage or exposed metal. Damaged leads should be replaced.

• Check test lead continuity uslnqthe diode testt um ++-) mode.



Be certain the digital multimeter (DMM) itself is in good operating condition. During the continuity test, a meter reading that goes from overload (OL) to 0 generally means the meter is working properly.

-Touch-Hold is a trademark of the John Fluke Manufacturing Company



Select the proper function and range for your measurement.

WARNING

TO AVOID ELECTRICAL SHOCK, USE CAUTION WHEN WORKING ABOVE 60V DC OR 25V AC RMS. SUCH VOLTAGES POSE A SHOCK HAZARD.



Electrically disconnect the live, or hot, test lead before disconnecting the common test lead.



Follow all equipment safety procedures. Disconnect the input power and discharge all high-voltage capacitors through a protective impedance before testing in n and (til' ++- with the multimeter.



Avoid working alone.

0 OFF lpower) ~ DANGEROUS
SWITCH
POSITION VOLTAGE
I ON (power! ~
SWITCH GROUND
POSITION
AC- A SEE
-<:» ALTERNATING EXPLANATION
CURRENT IN MANUAL
-- DC- [Q] DOUBLE
--- D1RECI' CURRENT INSULATION
(Protection Class II)
~ EITHER a- FUSE
DCOR AC International Electrical Symbols

ENGLISH



When making a current measurement, turn the power off before connecting the multi meter in the circuit.

Overloading a current shunt will cause excessive heat.

When measuring transformer secondary or motor winding current, check the multi meter fuses first. (See Fuse Test in the Operator Maintenance Section.) An open fuse will allow high voltage build-up, which is potentially hazardous.

WARNING

TO AVOID ELECTRICAL SHOCK OR DAMAGE TO THE METER, DO NOT APPLY MORE THAN 1000V BETWEEN ANY TERMINAL AND EARTH GROUND.

OPERATING FEATURES



The following features are keyed by number to the illustration inside the front cover.

CD Digital Display:

3200 count, liquid crystal display with automatic decimal point pOSitioning. Updated two times per second. When the meter is first turned on, all

display segments appear while the instrument performs a brief power-up self-test.

® Function Selector Rotary Switch:

Turn to select any of 10 different functions, or OFF. Refer to the Specifications for available ranges and to Table 1 for input terminals and limits.

V

Voltsdc

mV

Millivolts dc

"-

V

Volts ac

"-

mV

Millivolts ac

{1

Ohms (resistance), also conductance (1/0) in nanosiemens (nS)

((III .....

Continuity or diode test

Table 1. Input Terminals and Umits

FUNCTION INPUT TERMINALS MIN DISPLAY MAX DISPLAY MAXIMUM
Red Lead Black Lead READING READING INPUT
--- ""
V V yO ..... COM 0.001 V 1oo0V 1000V
--- "- yO ..... COM
mY mY 0.1 mV 320.0 mV SOOV
0 yO ..... COM 0.10 32.00 MO SOOV
(nS) YO,*- COM 0.01 nS 32.00 nS SOOV
1(111 ..... YO,*- COM O.OOW 2.08V SOOV
A COM 0.01A 20.00A* 10A* 600V
--- "-
mAlA mAlA
mA COM 0.01 mA 320.0 mA 320mA 600V
pA
iiA "- mA
IIA pA COM 0.1 pA 3200 pA 320mA 600V
*10A continuous, 20A for 30 seconds maximum 2

ENGLISH

"-

mAlA

Milliamps or amperes ac

"-

pA

Microamps ac

ii.A/A

Milliamps or amperes dc

iiA

Microamps dc

® V 0 +r Volt,Ohms, Diode Testlnput Terminal:

Input terminal used in conjunction with the volts, mV (ac or de), ohms, or diode test position of the function selector rotary switch.

o COM Common Terminal:

Common or return terminal used for all measurements.

® mAlpA Miliiamp/Microamp Input Terminal:

Input terminal used for current measurements up to 320 mA (ac or dc) with the function selector rotary switch in the mA or pA position.

® A Amperes Input Terminal:

Input terminal used for current measurements up to 10A continuous (20A for 30 seconds) with the function selector rotary switch in the mAl A position (ac or dc).

o RANGE ®Manual Range Mode Pushbutton:

Press once to enter manual range mode, press again to increment range, press and hold for 2 seconds to return to autorange. Meter returns to auto range if the function selector is switched to any other position. There is no autorange annunciator; absence of the manual range annunciator indicates the meter is in autorange. If RANGE is depressed (>1 second) while the function switch is moved from OFF to any ON position, manual ranging will be selected in all functions.

® REL~ Relative Mode Pushbutton:

Press momentarily to enter the Relative mode and store the displayed reading. The display will read zero. Press again to update the stored digital reading. Press and hold for 2 seconds to exit the

Relative mode. The Relative mode stores a digital reading and displays the change (difference) between the stored reading and any following reading. For example, if the stored reading is 15.00V and the present reading is 14.10V, the display will indicate -0.90V. The analog bar graph continues to display the actual reading (14.1 OV).lf the difference exceeds 3999 counts (without overloading the input), OF (overflow) is displayed. The Relative mode selects manual ranging; changing ranges automatically exits the Relative mode.

® MINIMAX MINIMAX Mode Pushbutton:

Press momentarily to enter MIN/MAX mode, press again to toggle between MIN and MAX indications. Press and hold for 2 seconds to exit MIN/MAX mode. The meter stores the minimum and maximum digital readings, and will display either reading as selected by the operator. Press the HOLD/RESET button to reset the MIN/MAX readings to the present input. The MIN/MAX mode selects manual ranging; use a range that can record the maximum anticipated input. Range changes reset previously recorded MIN/MAX readings. Exiting the MIN/MAX mode does not reset the previously recorded readings unless the range or function is changed. The MIN/MAX mode overrides the Touch-Hold mode.

@ HOLD Cl Touch-Hold Mode Pushbutton:

Press momentarily to enter Touch-Hold mode. In Touch-Hold, the meter captures a stable measurement and holds it in the display. The operator can watch the probes while taking measurements in difficult or hazardous circuits, then look at the display when convenient. The meter beeps and the display is automatically updated each time a new, stable measurement is made. Press momentarily to manually update reading. Press and hold for 2 seconds to exit Touch-Hold mode. If HOLD is depressed (>1 second) while the function switch is moved from OFF to any ON position, the Touch-Hold mode will only update to a new reading when the HOLD button is pressed. (Automatic Touch-Hold updates are defeated.) This is useful when you want to take a reading at a specific time and hold it.

3

ENGLISH

@

MIN Minimum Annunciator:

• ..

Indicates that the meter is in the MIN/MAX recording mode, and the value displayed is the minimum digital reading taken since reset or since entering MINIMAX. Refer to item 9 for operation.

@ M A X Maximum Annunciator:

Indicates that the meter is in the MINIMAX recording mode, and the value displayed is the maximum digital reading taken since reset or since entering MINIMAX. Refer to item 9 for operation.

IJ.. Relative Annunciator:

Indicates that the meter is in the Relative mode and that the value displayed is relative (the difference between the present measurement and the previously stored reading). Refer to item 8 for operation.

@ GI Touch-Hold Mode Annunciator:

Displayed when the Touch-Hold mode is in use. Refer to item 10 for operation.

M k 0 Resistance Annunciators:

The appropriate annunciator (0, k, or M) is displayed for the resistance range in use.

@

nS Conductance Range Annunciator (nS):

Top range ofthe resistance function is the conductance range. Displays conductance in nS (nanosiemens). 1000/nS converts to megohms. (Example: 2 nS converts to 500 MO.) Use for measuring resistance above 32 megohms. Select 0, open test leads, press RANGE button twice. (Refer to item 7 for manual range operation.)

@ __ _.__.. Analog Bar Graph Display:

Analog representation of input. Composed of 31 segments which illuminate starting from the left as the input increases. (See display inside rear cover.) A minus sign (-) is displayed for reversepolarity inputs. Updated 25 times per second.

4

.:. Decimal Point/Range Indicator:

Decimal point position and the digits (3, 30, 300) under the decimal point indicate the range in use.

@ ® 'Manual Range Annunciator:

Displayed in the Manual Range mode or if the selected function has only one range. Absence of the indicator implies autorange mode in use. The meter powers-up in autorange. In autorange, the meter automatically selects the measurement range. Refer to item 7 for operation.

@ ~ Low BaHery Annunciator:

At least 60 hours of battery life remain when first displayed. Battery voltage is tested each time the function switch is moved to a new position.

_ Negative Polarity Annunciator:

Automatically indicates negative inputs.

n. UL

Overload Indication:

These symbols indicate the input is too large for the input circuitry. (The location of the decimal point depends on the measurement range.)

@ D F Overflow Indication:

These symbols indicate the calculated difference in the Relative mode is too large to display (>3999 counts) and that the input is not overloaded.

@ Beeper (not Illustrated):

The beeper can produce beeps, clicks, or a continuous tone. It is used for audible indication in the diode test mode, when operating the push buttons, arid when a new reading is displayed in the Touch-Hold mode.

ACCESSORY COMPARTMENT

The multi meter provides an easy-access compartment for storing test leads and other accessories. The compartment is large enough to accommodate certain temperature probes, current probes, and rf probes. The accessory compartment also provides access for battery and fuse replacement

ENGLISH

BAIL ADJUSTMENT

The multi meter is shipped from the factory with a detachable bail stored in the accessory compartment The bail allows the multi meter to rest in four convenient viewing positions, as shown in Figure 1.

To install the bail, slide the bail's mounting tabs into the two sets of mounting brackets on the case bottom as shown in position I or II. While installing the bail, keep the mounting tabs parallel to the case bottom. To remove the bail, simply slide the bailout of the mounting brackets.

To rest the multimeter on a bench (view A), install the bail as shown in position I. To rest the multi meter on a shelf (view B) or the floor (view C), install the bail as shown in position II. The multimeter can also be stood on end (view 0).

APPLICATIONS

AC Measurement

The multimeter's ac ranges employ an average responding ac-coupled converter. This means that the multimeter measures the average value of the input and displays it as an equivalent rms value for a sine wave. As a result, measurement errors are introduced when the input waveform is non-sinusoidal. Further, any dc component of the input is blocked by the ac-coupled converter. Figure 2 shows some commonly encountered waveforms. If the waveform is known, multiply the reading displayed on the multi meter by the indicated factor for the desired conversion.

Voltage, AC/DC

The multi meter features five ac voltage ranges and five dc voltage ranges. All ranges present an input impedance of approximately 10 megohms in parallel with less than 100 pF. Measurement errors, due to circuit loading, can result when making either ac or dc voltage measurements on circuits with high source resistance. However, in most cases the error is negligible (0.1 % or less) if the measurement circuit source resistance is 10 kilohms or less. If circuit loading does present a problem, the percentage of error can be calculated using the appropriate formula from Figure 3.

When measuring voltages above 320V in Touch-Hold mode, use manual ranging to minimize readings of stray Voltages.

Current, AC/DC

WARNING

INSTRUMENT DAMAGE AND OPERATOR INJURY MAY RESULT IF THE FUSE BLOWS WHILE CURRENT IS BEING MEASURED IN A CIRCUIT WHICH EXHIBITS AN OPEN CIRCUIT VOLTAGE GREATER THAN 600V. DO NOT ATTEMPT AN IN-CIRCUIT CURRENT MEASUREMENT WHERE THE POTENTIAL IS GREATER THAN 600V.

The multi meter features five ac current ranges and five dc current ranges. All current ranges are fuse protected. If a fuse opens, refer to the fuse replacement procedures in the Operator Maintenance section of this manual.

Current Measurement Error Calculations

In an ac or dc current measurement, the voltage developed across the meter's terminals is called burden voltage. The burden voltage for a full-scale input is given for each range in the Specifications table. The burden voltage can affectthe accuracy of a current measurement if the current source is unregulated and the terminal resistance represents a significant portion (1/1000th or more) of the source resistance. If burden voltage does present a problem, the percentage of error can be calculated using the formula in Figure 4. Approximate terminal resistances for the current ranges are: 0.05 ohms for A, 5.5 ohms for mA, and 500 ohms for /lA.

Resistance Measurement

CAUTION

Turn test circuit power off and discharge a" capacitors before attempting in-circuit resistance measurements.

The multi meter features six resistance ranges and a conductance range. All ranges employ a two-wire measurement technique. As a result, test lead resistance may influence measurement accuracy on the 320-ohm range. To determine the error, short the test leads together and read the lead resistance. Correctthe measurement by subtracting the lead resistance from the measurement, or use the Relative (REL) mode to zero the display. The error is usually 0.1 to 0.2 ohms for a standard pair of test leads.

5

ENGLISH

A.

I.

B.

II.

c.

D.

6

Figure 1. Bail Adjustment

ENGLISH

DISPLAY MULTIPLIER FOR DISPLAY MULTIPLIER FOR
ACCOUPLED MEASUREMENT CONVERSION AC COUPLED MEASUREMENT CONVERSION
INPUT WAVEFORM RMS DC INPUT WAVEFORM DC
COMPONENT O-PK PK-PK RMS COMPONENT O-PK PK-PK
AC+OC ONLY AC+OC ONLY
SINE P~ru _i_ RECTIFIED SQUARE
PK-PK 1.000 0.000 1.414 2.828 PK _L
nn PK-PK 1.274 0.900 1.800 1.800
,-
RECTIFIED SINE\(FULL WAVE) o "T"
P:m _t_ 2.375 2.138 3.359 3.359 RECTANGULAR PULSE
PK-PK D=X/Y
-,- _l_ 0.450 ~ 0.450 0.450
RECTIFIED SINE (HALF WAVE) P:JUl PK-PK ro(l-D) (I-D) 0(1-0) 0(1-0)
PKf\..f\ __L_ 1.283 0.817 2.566 2.566 -I Y I- .-
PK-PK
0 -,- TRIANGLE SAWTOOTH
SQUARE __l_ P~'V _L.
P:~ PK-PK 1.040 0.000 1.800 3.600
PK-PK 0.900 0.000 0.900 1.800
.-- -,- Figure 2. Waveform Conversion

1. DC VOLTAGE MEASUREMENTS

L.oading Error in % = 100 x Rs:!: (Rs + 10')

Where: Rs = Source resistance in ohms of circuit being measured.

2. AC VOLTAGE MEASUREMENTS

First. determine input impedance. as follows:

Z in = 10'

y'1 + (2 .. F .Rin.C»

Where: Zin = effective input impedance Rin = 10' ohms

Cin = 100 x 10-" Farads F = frequency in Hz

Then. determine source loading error as follows: (Vector algebra required)

100 x Zs Loading Error in % = -Rs + Zin

Where: Zs = source impedance

Zin = input impedance (calculated) Rs = source resistance

Figure 3. Voltage Measurement Error Calculations

Some in-circuit resistance measurements can be made without removing diodes and transistors from the circuit. The full-scale measurement voltage produced on ranges below 32 megohms does not strongly forward bias silicon diodes or transistor junctions. Use the highest range you can (except 32 megohm) to minimize the possibility of turning on diodes or transistor junctions. Full scale measurement voltage in the 32-megohm range does strongly forward bias t diode or transistor.

Diode Test and Continuity

In diode test, there is only one range: 0 to +2.08 volts. Voltage is developed across the component(s) under test by a test current output from the multi meter. Voltages greater than 2.08V or open test leads produce an overload (Ol) condition. Negative inputs produce a negative indication (they are not suppressed). In the diode test function (11111 *), the beeper produces a continuous tone ifthe input is less than 0.1 V, and the beeper beeps once when the input descends through a 0.7V threshold.

Audible continuity testing is also performed with the function selector switch in the diode test/continuity position. A continuous tone sounds for test resistances below approximately 150 ohms. An intermittent connection produces erratic beeps, and can be a valuable troubleshooting aid.

7

ENGLISH

Erratic beeps can also occur, due to environmental noise, if a test value is very close to the threshold (150 ohms). Test resistances from approximately 150 ohms to 1000 ohms produce a short tone similar to a forward biased diode. Test resistances less than approximately 20 kilohms will produce an on-scale reading.

Conductance

Conductance measurement is performed with the function selector switch in the ohms (0) function. The conductance range can only be entered using manual range selection; autorange cannot enter the conductance range. The conductance range can be used both to measure conductance (1/0, the inverse of resistance) and to measure very high resistances (greater than 32 megohms).

High value resistance measurements are susceptible to induced noise, and may require careful shielding. Conductance measurements are displayed in nanosiemens (nS). Calculate megohms by dividing 1000 by the nanosiemens displayed (1000/nS is equivalentto megohms). Example: 2 nS converts to 500 megohms (1000/2).

Leakage Testing

The conductance range effectively extends the resistance measurement capability of the multimeterto the point where it can provide useful leakage measurements on passive

components. For example, the operator can detect leaky diodes, cables, connectors, printed circuit boards, etc. In all cases, the test voltage is less than 2V dc.

Leakage testing on purely resistive components such as cables and printed circuit boards is straightforward. Select the ohms function and manually increment the range to conductance (nS). Connect the test leads to the test points on the unit under test, and read the leakage in terms of conductance.

NOTE

There is normally a small residual reading with open test leads in the conductance range. To ensure accurate measurements, connect clean test leads to the mu/timeter and (with the leads open) read the residua/leakage in nanosiemens. Correct subsequent measurements by subtracting the residual from the readings. This can be done automatically using the Relative mode (REL) in the multimeter.

Diode leakage tests require that the diode junction be reverse biased when being measured. This is accomplished by connecting the anode of the diode to the COMMON input terminal and the cathode (ring) of the diode to the voltsl ohmsl diode test terminal. Leakage at the test voltage being applied can then be read in terms of conductance.

Es

1M)

J Ear

I A~METER SHCNTI

ES = Source voltage

Rl = load resistance + Source resistance

1M = Measured current (display reading in rnA)

EB = Burden voltage (calculated). i.e .. Display reading

READING .

expressed as a % of full-scale (100 x FUll-SCALE) times

full-scale burden voltage for selected range. See Table:

TYPICAL BURDEN VOLTAGE

RANGE

320 pA 3200 pA 32 mA 320mA

lOA

0.16V 1.6V .18V 1.8V 0.5V

Maximum cu,ent error due to Burden Voltage

• E

in % = 100 x B

ES - EB

_ EB x 'M - ES -EB

Example: ES = 15V, Rl = SOO.IM = 270 rnA.

in rnA

270

EB = 100 x 320 x 1.8 (from Table) =

84.4% x 1.8 = 1.519V

. 1~9 1~9

Error In % = 100 = 100 13.481 = 11.3%

15-1.519

Increase displayed current by 11.3% to obtain true current Error in rnA = 1.519 x 270 = ~ = 30.41 rnA

15 - 1.519 13.481

Increase displayed current by 30 rnA to obtain true current.

8

Figure 4. Current Measurement Error Calculations

ENGLISH

High-voltage stacked diode assemblies can usually be tested for forward and reverse resistance changes using conductance. These assemblies typically have such high forward voltage drops that the diode test or resistance modes cannot test them.

ANALOG BAR GRAPH APPLICATIONS

In looking at the analog bar graph, notice that it is composed of segments that simulate an analog needle. The bar graph performs the same function as an analog meter needle, but it eliminates the ~echanical overshoot inherent in needle

movements. \

A negative (-) annunciator is displayed at the left end of the bar graph when taking a reverse polarity dc measurement. Assume that a slowly varying dc voltage is the input signal. As the input goes more positive (from zero), a bar graph segment is displayed, and additional segments are displayed from left to right, to indicate the input level as it increases. Now, assume that the input level slowly decreases. Fewer bar graph segments are displayed as the signal decreases, then the - annunciator flashes as the signal level passes through O. As the signal goes more negative, the - annunciator is displayed, and additional bar graph segments are displayed from left to right, indicating a more negative input signal. The first segment is an indication greater than or equal to 20 counts.

Note that every fifth segment of the bar graph is slightly larger than those in between, and every tenth segment is larger yet. These larger segments provide a quick reference for bar graph indications. The largest segments (every 10th segment) divide the display into thirds. Thus, if the bar graph indicates 11 segments on the 32.00Vrange, the.inputvoltaqe is 1 0 to 11 volts; ifthe bar graph indicates 11 segments on the 320.0V range, the input voltage is 100 to 110 volts.lfthe input equals or exceeds 3000 counts on the range selected, the bar graph displays an arrow at the far right of the display. If the manual range annunciator ( ® ) is not displayed, the multi meter automatically switches to the next higher range if the input exceeds approximately 3260 counts. If the multimeter is in the manual range mode, the overrange arrow is displayed until the operator manually selects a range appropriate for the input value.

Using the Analog Bar Graph

The analog bar graph is most useful in making adjustments and performing limited diagnostics. Bar graph response is fast and precise, so it can be used to easily reach a setting within a few percent of the final adjustment. The bar graph can be used to make rough adjustments quickly; then the 3200-count digital display can be used for final adjustment.

The analog bar graph is useful for performing limited diagnostics in applications where rapidly fluctuating signal levels cause the flashing digits of a digital display to be useless. Like the traditional VOM needle, the analog bar graph excels at displaying trends, or slowly changing signals. In addition, autoranging on the multimeter allows monitoring the signal change through changing ranges.

Many diagnostic routines using the bar graph require practice. The operator is looking for good or bad signal patterns that occur over some span of time. Capacitance checks and noisy resistance measurements create such patterns. Therefore, familiarity with analog bar graph response and movement is necessary to accurately interpret a signal pattern. Compare the bar graph response when making measurements on a known-gOOd unit to the bar graph response when making measurements on a faulty unit.

Specific Applications-Nulling

The multimeter bar graph is ideal for nulling adjustments. As an adjustment approaches zero, fewer bar graph segments are displayed, then no bar graph segments are displayed. The - annunciator flickers when the input level is within 10 counts of zero. The flickering null indication is displayed every time the input approaches zero or swings from one polarity to the other. The operator merely watches for the - annunciator indication, then reverses the direction of the adjustment when the polarity sign is displayed. In one or two passes, a near-zero input level is possible, then the digital display can be used for exact zero adjustment.

When using an analog VOM without a center scale, the operator must manually switch the polarity between each adjustment. Also, with the traditional analog VOM, there is no digital display to use for fine adjustment after the analog needle is at zero.

Specific Applications-Contact Bounce

When subject to vibration, relay contacts may begin to bounce open. Checking for this intermittent problem is a routine troubleshooting measure associated with many types of equipment, including computers. Since the bounce problem will worsen as the relay fatigues, early diagnosis is important.

When the contact bounces open, its resistance value changes momentarily from zero to infinity and back. Ordinary hand-held DMMs take more than 300 milliseconds to update their displays-much too long to detect a brief contact bounce. A traditional VOM needle will move slightly at the instant of contact bounce, but the inertia of the needle movement dampens the response.

9

ENGLISH

Table 2. Capacitance Vs. Time to Full Scale

The analog bar graph, however, will display at Ilast one segment the moment the contact opens. The bar graph can detect contact bounce as brief as 0.2 milliseconds, while most analog needle movements require a 3 millisecond opening before they will respond.

Since the analog bar graph is ten times more sensitive to erratic signals than most analog needle movements, the bar graph can detect faulty contacts earlier than ever before. The severity of the problem is indicated by the number of segments displayed.

Specific Applications-Checking Capacitors

Volt-ohm meters are often used as simple capacitor checkers. In the capacitor kick test, the needle of the VOM in the resistance mode moves quickly from open (infinite ohms) toward short (zero ohms) as the capacitor is placed across the VOM input. The VOM battery charges the capacitor and the needle slowly moves back to the open (infinite ohms) position. The higher resistance ranges offer increased sensitivity for checking smaller capacitors.

The analog bar graph can make similar checks in the resistance function, even in the autoranging mode. As a capacitor is placed across the inputs, the analog bar graph quickly shortens, then rapidly down-ranges, depending on the size of the capacitor. As the capacitor charges, the bar graph slowly extends back to its full 31-segment length, up-ranging if necessary. For capacitors as small as 0.02 pF, only the 30-megohm range is involved, the lastfew segments blink off, then back on.

In a fixed range (using manual range mode), the time it takes for the bar graph to extend from zero to full scale indicates the approximate capacitance value. Table 2 gives typical capacitance values for various charge times on different resistance ranges. For very small capacitors, use the conductance (nS) mode.

Specific Applications-Noisy Resistance Measurements Most digital multimeters are so sensitive they cannottolerate as much as 50 mV of line noise while making resistance measurements; their digital displays become unreadable due to the line noise. On the other hand, because of the mechanical inertia of the analog needle, the noise alternately pulls the needle to the left and then to the right, averaging out any movement and leaving a fairly stable resistance reading.

10

Resistance 3200 3.2k0 32k0 320kO 3.2MO 32M 0
Range
Capacitance
Value
10,0001lF 4 sec 33 sec Smin ext ext ext
1,000 pF blink 4 sec 30 sec ext ext ext
100pF nil blink 4 sec 32 sec ext ext
10llF nil nil blink 4 sec 30 sec ext
11lF nil nil nil blink 3 sec 19 sec
0.11lF nil nil nil nil blink 2 sec
0.021lF nil nil nil nil nil blink
ext = extended time. nil = no indication The multimeter's resistance measurement circuit is designed to tolerate ac noise far better than the usual DMM. Readable 2-kilohm readings can be obtained even in the presence of 1 V ac noise. Readings of 1 megohm may be obtained with up to 2V ac noise. The noise appears as about 50 counts of change and an oscillating bar graph.

OPERATOR MAINTENANCE

WARNING

TO AVOID ELECTRICAL SHOCK, REMOVE THE TEST LEADS AND ANY INPUT SIGNALS BEFORE REPLACING THE BATTERY OR FUSES.

Battery Installation or Replacement

The multi meter is powered by a single 9V battery (NEDA 1604, 6F22, or 006P). Referring to Figure 5, use the following procedure to replace the battery:

1. Disconnect test leads from any live source, turn the rotary switch to OFF, and remove the test leads from the front terminals.

2. Open the hinged lid of the accessory compartment by pressing back and up on the button. Remove any stored accessories.

3. Using a slot-head screwdriver, car key, or other appropriate tool, pry open the top of the battery cover and lift the cover out.

4. Remove the battery from the battery cover and disconnect the battery connector leads.

ENGLISH

5. Snap the battery connector leads to the terminals on a new battery and slide the battery into the battery cover.

6. Reinsert the battery cover in the multi meter.

Fuse Test

1. Turn the function selector switch to the a position.

2. Connect a test lead from the va ++- input terminal to the A input terminal.

3. The display should indicate Qetween 0.1 ohm and 0.3 ohm. This tests F3 (15A, 600V fast).

4. Move one end of the test lead from the A input terminal to the mAl pA input terminal.

5. The display should indicate between 5.3 ohms and 6.C ohms. This tests Fl (2A, 600V fast) and F2 (630 mA, 250V fast).

6. If either of the above display indications is OL (overload), replace the appropriate fuse.

Fuse Replacement

WARNING

TO PREVENT DAMAGE OR INJURY,INSTALL QUICK ACTING FUSES WITH THE AMP IVOLT RATINGS SHOWN IN FIGURE 5.

Referring to Figure 5, use the following procedure to check or replace the multi meter's fuses:

1. Perform steps 1 through 3 of the battery replacement procedure.

2. Remove the battery from the battery cover and set the cover aside.

3. Remove the defective fuse by prying the fuse loose on one end and sliding the fuse out of the fuse bracket.

4. Install a new fuse of the same size and rating. (Note that there is a spare 630 mA, 250V fuse stored below the battery inside the battery cover.)

5. Reinstall the battery in the cover and insert the cover into the multimeter.

F3 (15A, 600V)

!~~~-- Fl (2A,600V)

F2 (630 mA, 250V)

9VBATIERY

(006P, 6FZZ, NEDA 1604)

Figure 5. BaDery and Fuse Replacement

11

ENGLISH

General Maintenance

Clean the case with a damp cloth and detergent; do not use abrasives or solvents.

Have the meter calibrated by a qualified technician once a year to ensure specified performance. Contact the nearest Fluke Service Center or refer to the Fluke Service Manual (part number 785311) for calibration or repair. Refer to the parts list at the end of this manual for operator replaceable parts.

BATTERY ELIMINATOR ACCESSORY

Table 3 lists the battery eliminators available to provide power for operation of the Fluke 37 from 100V ac, 115V ac, 220V ac, or 240V ac line voltages.

WARNING

TO PREVENT ELECTRICAL SHOCK OR DAMAGE, USE ONLY FLUKE MODEL AS1 BATTERY ELIMINATORS.

Table 3.

SERVICE

If the instrument fails, forward it, postage paid, to the nearest Fluke Service Center (refer to the list at the back of this manual). Include a description of the difficulty, and pack the instrument securely; Fluke shall assume NO responsibility for damage in transit.

IN WARRANTY: Instruments covered by the limited warranty will be promptly repaired or replaced, at Fluke's option, and returned, all at no charge. See the registration card for warranty terms.

OUT OF WARRANTY (USA AND CANADA): The instrument will be repaired and returned for a fixed fee. (Repairs needed because of abuse or accidental damage will be quoted.) Contact the nearest Service Center for current prices. Include a check, money order, or purchase order with the instrument.

OUT OF WARRANTY (OUTSIDE USA AND CANADA):

Service programs may vary by country. Contact the nearest Service Center for information.

MODEL INPUT CONNECTOR SAFETY
A81-100 100Vac NEMA 1-15P Designed
48-66 Hz to UL 1244
A81-115 120V ac NEMA 1-1SP UL Listed
60 Hz CSA Cert
A81-220B 220V ac CEE 7 VDE
50 Hz Pending
A81-230 220V ac CEE 7 IEC 348
48-66 Hz (Outside Europe)
A81-230-3 240V ac BS-1363 IEC 348
SO Hz
OUTPUT
All models 8.2V dc @ 2 mA, S.8V dc @ 0.1SA 12

ENGLISH

Fluke 37 Specifications

FUNCTION RANGE RESOLUTION ACCURACY"
3.200V O.OOW ±(0.1%+1)
--- 32.00V 0.01V ±(0.1%+1)
V 320.0V 0.1V ±(0.1%+1)
1000V 1V ±(0.1%+1)
--- 320.0 mv 0.1 mV ±(0.1%+1)
mV
320.00 0.10 ±(0.3%+2)
3.200 kn 0.001 kn ±(0.2%+1)
32.00 kn 0.01 kn ±(0.2%+1)
n 320.0 kn 0.1 kn ;1:(0.2%+1)
(nS) 3.200 MO 0.001 MO ±(0.2%+1)
32.00 MO 0.01 MO ±(1%+1)
32.00 nS 0.01 nS ±(2%+10)
(It,, ..... 2.080V O.OOW ±(1%+1) typical
40 Hz-2 kHz 2 kHz-10 kHz 10kHz-3D kHz
"-' 3.200V 0.001V ±(0.5%+3) ±(2%+3) ±(4%+W)
V 32.00V 0.01V ±(0.5%+3) ±(2%+3) ±(4%+10)
320.0V 0.1V ±(0.5%+3) ±(2%+3) ±(4%+10)
1000V 1V ±(1%+3) ±(3%+3) Not Specified
...;.... 320.0 mV 0.1 mV ±(0.5% +3) ±(2%+3) ±(4%+10)
mV FUNCTION RANGE RESOLUTION ACCURACY· TYPICAL BURDEN VOLTAGE
32.00 mA 0.01 mA ±(0.75%+2) 5.6 mV/mA
mAlA 320.0 mA 0.1 mA ±(0.75%+2) 5.6 mV/mA
10.00A 0.01A ±(0.75%+2) 50 mV/A
iii>. 320.0 pA 0.1 pA ±(0.75%+2) 0.5 mV/pA
3200pA 1 pA ±(0.75%+2) 0.5 mV/pA
"-' 32.00 mA 0.01 mA ±(1.5%+2) 5.6 mV/mA
mAlA 320.0 mA 0.1 mA ±(1.5%+2) 5.6 mV/mA
40-1000 Hz 10.00A 0.01A ±(1.5%+2) 50 mV/A
"-' 320.0 pA 0.1 pA ±(1.5%+2) 0.5 mV/pA
pA
40-1000 Hz 3200 pA 1 pA ±(1.5%+2) 0.5 mV/pA
. . . . * Accuracy IS specltled as ±([% of reading] + [number of least significant digits]) .

Basic electrical accuracy is specified from 18°C to 28°C with relative humidity up to 90%, for a period of one year after calibration. All ac conversions are ac coupled, average responding, and calibrated to read the true rms value of a sine wave input.

Ranging is either automatic or manual in all functions with more than one range. Test resistance below approximately 1500 in the 1111' ..... function produces a continuous audible tone.

13

ENGLISH

Fluke 37 Specifications (cont)

OVERLOAD INPUT COMMON MODE NORMAL MODE
FUNCTION PROTECTION IMPEDANCE REJECTION RATIO REJECTION
(nominal) (1 kn unbalance)
--- 1000V rms** 10 MO in I I with <100 pF >120 dB at dc, 50 Hz, or 60 Hz >60 dB at 50 Hz or 60 Hz
V
--- 500V rms** 10 MO in II with <100 pF >120 dB at dc, 50 Hz, or 60 Hz >60 dB at 50 Hz or 60 Hz
mV
""' 1000V rms 10 MO in II with <100 pF >60 dB, dc to 60 Hz
V (107 V-Hz max) (ac coupled)
""' 500V rms 10 MO in II with <100 pF >60 dB, dcto 60 Hz
mV (107 V-Hz max) (accoupled)
OPEN CIRCUIT FULL SCALE VOLTAGE
n 500V rms TEST VOLTAGE Up to 3.2 MO 32 MO or nS
<2.8V dc <420 mV de <1.3V dc ** 107 V- Hz max

FUNCTION FUSE PROTECTION
rnA or jJA 630 rnA 250V FAST, 2A SOOV FAST
A 15A 600V FAST MAXIMUM VOLTAGE BETWEEN ANY TERMINAL AND EARTH GROUND

1000V

Digital Display 3200 counts, updates 2/sec

Analog Display 31 segments, updates 25/sec

Operating Temperature -15°C to 55°C, to -40°C for 20 minutes when taken from 20°C

Storage Temperature --40°C to +60°C

Temperature Coefflclent 0.1 x (specified accuracvj/vc

«18°C or >28°C)

Relative Humidity 0% to 90% (O°C to 35°C) 0% to 70% (35°C to 55°C)

BaHery Type 9V, NEDA 1604 or 6F22 or oo6P

BaHery Life 1000 hrs typical

Shock and Vibration Per MIL- T -28800

Size (HxWxL) 3.5 in x 8.5 in x 8.9 in (8.9 cm x 21.6 cm x 22.6 em)

Weight 2.5 pounds (0.94 kg)

Safety Protection Class II per IEC 348 and ANSI C39.5

14

You might also like