H2 PDF
H2 PDF
2a Screen-film mammography
2b Digital mammography
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   57
H2   20-09-2005   20:54    Pagina 58
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   58               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 59
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   Executive summary
                   A prerequisite for a successful screening project is that the mammograms contain sufficient
                   diagnostic information to be able to detect breast cancer, using as low a radiation dose as is
                   reasonably achievable (ALARA). This quality demand holds for every single mammogram. Quality
                   Control (QC) therefore must ascertain that the equipment performs at a constant high quality
                   level.
                   In the framework of ‘Europe Against Cancer’ (EAC), a European approach for mammography
                   screening is chosen to achieve comparable high quality results for all centres participating in the
                   mammography screening programme. Within this programme, Quality Assurance (QA) takes into
                   account the medical, organisational and technical aspects. This section is specifically concerned
                   with the quality control of physical and technical aspects of medical imaging in mammography
                   and the dosimetry.
                   The intention of this part of the guidelines is to indicate the basic test procedures, dose
                   measurements and their frequencies. The use of these tests and procedures is essential for
                   ensuring high quality mammography and enables comparison between centres. This document is
                   intended as a minimum standard for implementation throughout the EC Member States and does
                   not reduce more comprehensive and refined requirements for QC that are specified in local or
                   national QA Programmes. Therefore some screening programmes may implement additional
                   procedures.
                   QC of the physical and technical aspects in mammography screening starts with specification
                   and purchase of the appropriate equipment, meeting accepted standards of performance. Before
                   the system is put into clinical use, it must undergo acceptance testing to ensure that the
                   performance meets these standards. This holds for the mammography X-ray equipment, image
                   receptor, film processor, viewing device and QC test equipment. After acceptance, the
                   performance of all equipment must be maintained above the minimum level and at the highest
                   level possible.
The QC of the physical and technical aspects must guarantee that the following objectives are met:
                   1. The radiologist is provided with images that have the best possible diagnostic information
                      obtainable when the appropriate radiographic technique is employed. The images should at
                      least contain the defined acceptable level of information, necessary to detect the smaller
                      lesions (see CEC Document EUR 16260).
                   2. The image quality is stable with respect to information content and optical density and
                      consistent with that obtained by other participating screening centres.
                   3. The breast dose is As Low As Reasonably Achievable (ALARA) for the mammographic
                      information required.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   59
H2   20-09-2005   20:54     Pagina 60
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   Several measurements can be performed by the local staff. The more elaborate measurements
                   should be undertaken by medical physicists who are trained and experienced in diagnostic
                   radiology and specifically trained in mammography QC. Comparability and consistency of the
                   results from different centres is best achieved if data from all measurements, including those
                   performed by local technicians or radiographers are collected and analysed centrally.
                   Image quality and breast dose depend on the equipment used and the radiographic technique
                   employed. QC should be carried out by monitoring the physical and technical parameters of the
                   mammographic system and its components. The following components and system parameters
                   should be monitored:
                   The probability of change and the impact of a change on image quality and on breast dose
                   determine the frequencies at which the parameters should be measured. The protocol gives also
                   the acceptable and achievable limiting values for some QC parameters. The acceptable values
                   indicate the minimal performance limits. The achievable values indicate the limits that are
                   achievable. Limiting values are only indicated when consensus on the measurement method and
                   parameter values has been obtained. The equipment required for conducting QC tests are listed
                   together with the appropriate tolerances in Table II.
                   Methods of dosimetry are described in the ‘European Protocol on Dosimetry in Mammography’
                   (EUR16263). It provides accepted indicators for breast dose, from both measurements on a
                   group of women and test objects.
                   The first (1992) version of this document (REF: EUR 14821) was produced by a Study Group,
                   selected from the contractors of the CEC Radiation Protection Actions. In the second (1996) and
                   third (1999) version the test procedures and limiting values have been reviewed critically based
                   on literature, the experience gained by users of the document and comments from
                   manufacturers of equipment and screen-film systems. Due to the introduction of digital
                   mammography an addendum on digital mammography was made in 2003. The current version
                   incorporates both screen-film and digital mammography and is based on further practical
                   experience with the protocol, comments from manufacturers and the need to adapt to new
                   developments in equipment and in the literature.
                   60               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 61
                   Authors
                   R. van Engen
                   S. van Woudenberg
                   H. Bosmans
                   K. Young
                   M. Thijssen
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   61
H2   20-09-2005   20:54    Pagina 62
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                   EUROPEAN COMMISSION
                   Authors:
                   R. van Engen, Nijmegen, the Netherlands
                   S. van Woudenberg, Nijmegen, the Netherlands
                   H. Bosmans, Leuven, Belgium
                   K. Young, Guildford, United Kingdom
                   M. Thijssen, Nijmegen, the Netherlands
                   Contributors:
                   P. Baldelli, Ferrara, Italy
                   B. Beckers, Nijmegen, the Netherlands
                   R. Bijkerk, Nijmegen, the Netherlands
                   M. Borowski, Bremen, Germany
                   AK. Carton, Leuven, Belgium
                   D. Dance, London, United Kingdom
                   T. Deprez, Leuven, Belgium
                   D. Dierckx, Brussels, Belgium
                   A. Ferro de Carvalho, Lisbon, Portugal
                   M. Fitzgerald, London, United Kingdom
                   A. Flioni Vyza, Athens, Greece
                   M. Gambaccini, Ferrara, Italy
                   T. Geertse, Nijmegen, The Netherlands
                   G. Gennaro, Padua, Italy
                   N. Gerardy, Brussels, Belgium
                   A. de Hauwere, Gent, Belgium
                   P. Heid, Marseille, France
                   E. van der Kop, Nijmegen, the Netherlands
                   W. Leitz, Stockholm, Sweden
                   J. Lindeijer, Nijmegen, the Netherlands
                   R. van Loon, Brussels, Belgium
                   C. Maccia, Cachan, France
                   A. Maidment, Philadelphia, USA
                   H. Mol, Brussels, Belgium
                   B. Moores, Liverpool, United Kingdom
                   L. Oostveen, Nijmegen, the Netherlands
                   J. Pages, Brussels, Belgium
                   F. Rogge, Leuven, Belgium
                   M. Säbel, Erlangen, Germany
                   H. Schibilla, Brussels, EC
                   F. Shannoun, Luxembourg, Luxembourg
                   J. Shekdhar, London, United Kingdom
                   F. Stieve, Neuherberg, Germany
                   M. Swinkels, Nijmegen, the Netherlands
                   A. Taibi, Ferrara, Italy
                   D. Teunen, Luxembourg, EC
                   E. Vaño, Madrid, Spain
                   F. Verdun, Lausanne, Switserland
                   A. Watt, Edinburgh, United Kingdom
                   J. Zoetelief, Rijswijk, the Netherlands
                   Acknowledgements:
                   Comments have been received from the following manufacturers:
                   GE, Kodak
                   62               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 63
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                   a
                   Many measurements are performed using an exposure of a test object. All measurements are
                   performed under normal working conditions: no special adjustments of the equipment are
                   necessary.
                   Two standard types of exposures are specified:
                   • The reference exposure- which is intended to provide information on the system under defined
                     conditions, independent of the clinical settings
                   • The routine exposure- which is intended to provide information on the system under clinical
                     settings
                   For the production of the reference or routine exposure, an object is exposed using the machine
                   settings as follows (unless otherwise mentioned):
compression device in contact with test object in contact with test object
source-to-image distance matching with focused grid matching with focused grid
                   The optical density (OD) of the processed image is measured at the reference ROI, which lies 60
                   mm from the chest wall side and laterally centred. The reference optical density is preferably
                   1.60 ± 0.15 OD.
                   All measurements should be performed with the same cassette to rule out differences between
                   screens and cassettes except when testing individual cassettes (as in section 2a.2.2.2).
                   Limits of acceptable performance are given, but often a better result would be achievable. Both
                   the acceptable and achievable limits are summarised in section 2a.4, table 1. Occasionally no
                   limiting value is given, but only a typical value as an indication of what may normally be expected.
                   The measurement frequencies indicated in the protocol (summarised in section 2a.4) are the
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   63
H2   20-09-2005   20:54    Pagina 64
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                   minimum required. When the acceptable limiting value is exceeded the measurement should be
                   repeated. If necessary, additional measurements should be performed to determine the origin of
                   the observed problem and appropriate actions should be taken to solve the problem.
                   For guidance on the specific design and operating criteria of suitable test objects; see the
                   Proceedings of the CEC Workshop on Test Phantoms (see section 2a.5, Bibliography). Definition
                   of terms, such as the ‘reference ROI’ and the ‘reference density’ are given in section 2a.1.2. The
                   evaluation of the results of the QC measurements can be simplified by using the forms for QC
                   reporting provided in section 2a.6.
                   The staff conducting the daily/weekly QC-tests will need the following equipment2 at the
                   screening site:
                   The medical physics staff conducting the other QC-tests will need the following additional
                   equipment and may need duplicates of many of the above:
                   • Dosemeter                                                  • Stopwatch
                   • kVp-meter                                                  • Film/screen contact test device
                   • Exposure time meter                                        • Tape measure
                   • Light meter                                                • Compression force test device
                   • QC test objects                                            • Rubber foam
                   • Aluminium sheets                                           • Lead sheet
                   • Focal spot test device + stand                             • Aluminium stepwedge
                   Air kerma                                         Quotient of dEtr by dm where dEtr is the sum of initial kinetic
                                                                     energies of all the charged ionising particles liberated by
                                                                     uncharged ionising particles in a mass of air dm (adapted from
                                                                     ICRU 1980). The common unit for air kerma is milliGray (mGy).
                                                                     Air kerma measures, employing a ionization chamber or another
                                                                     dose detector calibrated in mammography energy range, can be used
                                                                     to evaluate the entrance dose (Entrance Surface Air Kerma – ESAK).
                   64               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 65
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                   a
                   Automatic Exposure                                           Operation mode of an X-ray machine by which the tube loading is
                   Control (AEC)                                                automatically controlled and terminated when a preset radiation
                                                                                exposure to a dose detector located under the image receptor is
                                                                                reached. Some more sophisticated equipment also allow the
                                                                                automatic selection of tube potential (kV), target and filter
                                                                                materials.
                   Average glandular dose                                       Reference term (ICRP 1987) for radiation dose estimation from X-
                   (AGD)                                                        ray mammography i.e. the average absorbed dose in the glandular
                                                                                tissue in a uniformly compressed breast. AGD value depends on X-
                                                                                ray beam quality (HVL), breast thickness and composition. If
                                                                                breast thickness and composition are not known, AGD can be
                                                                                referred to a standard breast.
                   Compression paddle                                           Thin device (few millimetres) rectangular shaped, made of plastic
                                                                                material (typical PMMA or polycarbonate) that can be positioned
                                                                                parallel to and above the breast table of a mammography
                                                                                apparatus.
                   Contrast threshold                                           Contrast level that produces a just visible difference between an
                                                                                object and the background.
                   Half Value Layer (HVL)                                       Thickness of absorber which attenuates the air kerma of non-
                                                                                monochromatic X-ray beams by half. The absorber normally used
                                                                                to evaluate HVL of low energy X-ray beams, such as mammography
                                                                                beams, is high purity aluminium (≥ 99.9%). It should be noticed
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   65
H2   20-09-2005   20:54    Pagina 66
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                   Mean gradient (MGrad)                             Parameter describing the film contrast in the exposure range,
                                                                     which contains most diagnostic information. MGrad is calculated
                                                                     as the slope of the line through the points D0.25 = (Dmin + 0.25) OD
                                                                     and D2 = (Dmin + 2.00) OD. Since the film curve is constructed from
                                                                     a limited number of points, D0.25 and D2 must be obtained by
                                                                     interpolation. Linear interpolation will result in sufficient accuracy.
                   Middle gradient (Grad1,2)                         Parameter describing the film contrast in the middle of the
                                                                     diagnostic range. Grad1,2 is calculated as the slope of the line
                                                                     through the points D1 = (Dmin+ 1.00) OD and D2 = (Dmin+ 2.00) OD.
                                                                     Since the film curve is constructed from a limited number of
                                                                     points, D1 and D2 must be obtained by interpolation. Linear
                                                                     interpolation will result in sufficient accuracy.
                   Net optical density                               Optical density excluding base and fog. Base+fog value is
                                                                     determined measuring the optical density into a non-exposed area
                                                                     of film (see Dmin).
                   Optical density (OD)                              Logarithm (base 10) of the ratio between light intensity produced
                                                                     by a visible light source and perpendicularly incident on a film (Io),
                                                                     and light intensity transmitted by the film (I): OD = log10 (Io/I)
                                                                     Optical density is measured by an instrument named
                                                                     densitometer, that measures transmitted light intensity into an
                                                                     area of the order of mm2.
                                                                     Variations in optical density should be measured along a direction
                                                                     parallel to the major axis of image receptor (perpendicular to
                                                                     cathode-anode direction), to avoid influences by the angular
                                                                     distribution of X-ray intensity (heel-effect).
                   Patient dose                                      Generic term for a variety of radiation dose quantities applied to a
                                                                     (group of) patient(s).
                   66               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 67
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                   a
                   Quality assurance                                            As defined by the WHO (1982): ‘All those planned and systematic
                                                                                actions necessary to provide adequate confidence that a
                                                                                structure, system or component will perform satisfactorily in
                                                                                service (ISO 6215-1980). Satisfactory performance in service
                                                                                implies the optimum quality of the entire diagnostic process i.e.,
                                                                                the consistent production of adequate diagnostic information with
                                                                                minimum exposure of both patients and personnel.’
                   QC test object                                               Object made of tissue simulating material (typically PMMA) for
                                                                                image quality evaluation; it generally includes objects simulating
                                                                                mammographic lesions (microcalcifications, fibers, masses)
                                                                                and/or resolution patterns and step wedges to measure
                                                                                parameters such as spatial resolution or contrast, related to
                                                                                image quality.
                   Reference cassette                                           Cassette, properly identified, used for QC tests. Using a single
                                                                                cassette permits to exclude variations in optical density caused by
                                                                                changes in absorption from different cassettes or individual
                                                                                screen efficiencies.
                   Reference exposure                                           Exposure of the standard test object with predetermined values of
                                                                                parameters to provide an image at reference conditions.
Reference optical density Optical density of (1.6 ± 0.1) OD, measured in the reference ROI.
                   Reference ROI                                                Considering an image obtained by the standard test block, the
                                                                                reference ROI is centred 60 mm perpendicular from the chest wall
                                                                                in the middle of the major film axis.
                   Region Of Interest (ROI)                                     Measurement area of optical density whose boundaries can be
                                                                                virtually defined on an image. ROI size can be around 1 cm2.
                   Routine exposure                                             Exposure of the standard test block under the conditions that
                                                                                would normally be used to produce a mammogram having the
                                                                                routine optical density into the reference ROI. The routine
                                                                                exposure is used to check optical density and dose stability under
                                                                                clinical conditions.
                   Routine optical density                                      Optical density measured in the reference ROI of a standard block
                                                                                image obtained by a routine exposure. This value is chosen by the
                                                                                site personnel as optimal value for mean clinical mammograms
                                                                                provided by a specific imaging chain. The routine net optical
                                                                                density should be included into the interval [1.4-1.9] OD.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   67
H2   20-09-2005   20:54    Pagina 68
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                   Standard test block                               PMMA test object to simulate the absorption of a standard breast.
                                                                     Its thickness is (45.0 ± 0.5) mm and the remaining dimensions
                                                                     can be either rectangular ≥ 150 mm x 100 mm or semi-circular
                                                                     with a radius of ≥ 100 mm. The standard test block can be used to
                                                                     check the AEC behaviour or to evaluate a mean value of AGD.
                   Tube loading                                      Product of the X-ray tube current (milliampere, mA) and the
                                                                     exposure time (seconds, s). It is quantified in units of mAs.
                   Tube potential                                    The potential difference in units of kilovolt (kV) applied across the
                                                                     anode and cathode of an X-ray tube during a radiographic
                                                                     exposure.
                   68               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 69
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                   a
                   Tube yield                                                   Ratio between air kerma (mGy) measured without any test object
                                                                                and the tube loading (mAs), for a known distance between the X-
                                                                                ray source and the dosimeter and for preset exposure parameters.
                   Local basic safety tests should be followed. If no local basic safety tests are available, an
                   example of such tests can be found in appendix 1.
Fig. 2.1 Focal spot size measurement using the star pattern method
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   69
H2   20-09-2005   20:54    Pagina 70
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                   to obtain an optical density between 0.8 and 1.4 OD base and fog excluded (measured in the
                   central area of the image). The device should be imaged at the reference ROI of the image plane,
                   which is located at 60 mm from the chest wall side and laterally centred. Remove the
                   compression device and use the test stand to support the test device. Select about the same
                   focal spot charge (mAs) that is used to produce the standard image of 45 mm PMMA, which will
                   result in an optical density of the star pattern image in the range 0.8 to 1.4.
                   According the IEC/NEMA norm, an 0.3 nominal focal spot is limited to a width of 0.45 mm and a
                   length of 0.65 mm. A 0.4 nominal focal spot is limited to 0.60 and 0.85 mm respectively. No
                   specific limiting value is given here, since the measurement of imaging performance of the focal
                   spot is incorporated in the limits for spatial resolution at high contrast. (see 2a.2.5.2)
(2.1)
                   where q is the angle of the radiopaque spokes, and dblur is the diameter of the blur.
                   The magnification factor (mstar) is determined by measuring the diameter of the star pattern on
                   the acquired image (dimage) and the diameter of the device itself (dstar), directly on the star, and is
                   calculated by:
Note: mslit = mimage - 1, and the method requires a higher exposure than the star pattern method.
                   70               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 71
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                     a
Note: The method requires a higher exposure than the star pattern method.
                   The multi-pinhole device is used similarly. It allows an estimate of the focal spot size at any
                   position in the x-ray field. This method is not suitable for measuring the dimension of fine focus
                   because of the relatively large size of the pin-holes.
                                                                 f = source-to-image distance
                                                                       d
                                                                 d = distance between the object in position 1
                                                                      and 2
            breast support table           ob ject               a = size of the imaged object
            image receptor                                       p1 = size of the object on image 1 (object on
                                             p1                       the breast support table)
                                             p2                  p2 = size of the object on image 2 (object at
                                                                      a distance d above the breast support
                   Fig. 2.2 Source-to-image distance measurement      table)
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n     71
H2   20-09-2005   20:54    Pagina 72
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                   Note 1: The lateral edges of the X-ray field should at least expose the image receptor. A slight
                           extension beyond any edge of the image receptor is acceptable.
                   Note 2: If more than one field size or target is used, the measurement should be repeated for
                           each.
                   Limiting value                                     Not more than 1 mGy in 1 hour at 1 m from the focus at the
                                                                      maximum rating of the tube averaged over an area not exceeding
                                                                      100 cm2, and according to local regulations.
                   Frequency                                          At acceptance and after intervention on the tube housing.
                   Equipment                                          Dose meter and appropriate detector.
                   72               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 73
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                     a
                   focal spot and the reference ROI, in the absence of scatter material and attenuation (e.g. due to
                   the compression plate). A tube load (mAs) similar to that required for the reference exposure
                   should be used for the measurement. Correct for the distance from the focal spot to the detector
                   and calculate the specific output at 1 metre and the output rate at a distance equal to the focus-
                   to-film distance (FFD).
                   If the measurements are used for dosimetry, tube output measurements should be performed at
                   all relevant spectra with the compression plate in position.
                   Limiting values                                               Acceptable: > 30 µGy/mAs at 1 metre, achievable > 40 µGy/ mAs
                                                                                 at 1 metre > 70% of value at acceptance for all target-filter
                                                                                 combinations.
                   Frequency                                                     Every six months and when problems occur.
                   Equipment                                                     Dose-meter, exposure timer.
                   Note: A high output is desirable for a number of reasons e.g. it results in shorter exposure
                         times, minimising the effects of patient movement and ensures adequate penetration of
                         large/dense breasts within the setting of the guard timer. In addition any marked changes
                         in output require investigation.
Note: Consult the instruction manual of the kVp-meter for the correct positioning.
                   Limiting value                                                Accuracy for the range of clinically used tube voltages: < ± 1 kV,
                                                                                 reproducibility < ± 0.5 kV.
                   Frequency                                                     Every six months.
                   Equipment                                                     kVp-meter.
(2.6)
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n     73
H2   20-09-2005   20:54    Pagina 74
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                   The direct exposure reading is denoted as Y0; Y1 and Y2 are the exposure readings with added
                   aluminium thickness of X1 and X2 respectively.
                   Note 1: The purity of the aluminium ≥ 99.9% is required. The thickness of the aluminium sheets
                           should be measured with an accuracy of 1%.
                   Note 2: For this measurement the output of the X-ray machine needs to be stable.
                   Note 3: The HVL for other (clinical) tube voltages and other target materials and filters may also
                           be measured for assessment of the average glandular dose (see appendix 5 and the
                           European Protocol on Dosimetry in Mammography, ISBN 92-827-7289-6).
                   Note 4: Alternatively a digital HVL-meter can be used, but correct these readings under extra
                           filtration following the manufacturers’ manual.
                   Limiting value                                     For 28 kV Mo/Mo the HVL must be over 0.30 mm Al equivalent,
                                                                      and is typically < 0.40 mm Al. Typical values of HVL for relevant
                                                                      target-filter combinations and tube voltages, are shown in
                                                                      appendix 5, table A5.3.
                   Frequency                                          Yearly.
                   Equipment                                          Dosemeter, aluminium sheets: 0.30, 0.40, 0.50, 0.60 mm.
                   2a.2.1.3 AEC-system
                   The performance of the Automatic Exposure Control (AEC) system can be described by the
                   reproducibility and accuracy of the automatic optical density control under varying conditions, like
                   different object thickness and tube voltages. Essential prerequisites for these measurements
                   are a stable operating film-processor and the use of the reference cassette. If more than one
                   breast support table, with a different AEC detector attached, is used then each system must be
                   assessed separately.
                   2a.2.1.3.1 Optical density control setting: central value and difference per step
                   To compensate for the long term variations in mean density due to system variations the central
                   optical density setting and the difference per step of the selector are assessed. To verify the
                   adjustment of the optical density control, produce exposures in the clinically used AEC mode of
                   the standard test object with varying settings of the optical density control selector.
                   A target value for the mean optical density at the reference ROI should be established according
                   to local preference, in the range: 1.4 – 1.9 OD, base and fog included.
                   Limiting value                                     The optical density (base and fog included) of the step used
                                                                      clinically at the reference ROI should remain within ± 0.15 OD of
                                                                      the target value.
                                                                      The change produced by each step in the optical density control
                                                                      should be about 0.10 OD.
                                                                      Step-sizes within the range 0.05 to 0.20 OD are acceptable.
                                                                      The acceptable value for the range covered by full adjustment of
                                                                      the density control is > 1.0 OD.
                   Frequency                                          Step-size and adjustable range: every six months.
                                                                      Density and mAs-value for clinically used AEC setting: daily.
                   Equipment                                          Standard test block, densitometer.
                   Warning: An incorrect functioning of the back-up timer or security cut-off could damage the tube.
                            To avoid excessive tube load consult the manual for maximum permitted exposure time.
                   74               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 75
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                   a
                   Limiting value                                                The back-up timer and/or security cut-off should function properly.
                   Frequency                                                     Yearly.
                   Equipment                                                     PMMA plates or sheet of lead covering the detector.
                   Limiting value                                                Deviations from the mean value of exposures < ± 5%, achievable
                                                                                 < ± 2%.
                   Frequency                                                     Every six months.
                   Equipment                                                     Standard test block, dosemeter.
                   Note: For the assessment of the reproducibility, also compare these results from the short term
                         reproducibility with the results from the thickness and tube voltage compensation and
                         from the optical density control setting at 45 mm PMMA at identical settings. Any problem
                         will be indicated by a mismatch between those figures.
                   Limiting value                                                The variation from the target value must be within < ± 0.20 OD;
                                                                                 achievable < ± 0.15 OD.
                   Frequency                                                     Daily.
                   Equipment                                                     Standard test block or QC test object, densitometer.
                   Limiting value                                                All optical density variations from the chosen target optical
                                                                                 density must be within ± 0.15 OD. Achievable: ± 0.10 OD.
                                                                                 Typical spectra for each PMMA thickness can be found in
                                                                                 appendix 4. The value of the thickness indicator must be within
                                                                                 ± 0.5 cm of the thickness of the PMMA plates.
                   Frequency                                                     Every six months: full test.
                                                                                 Weekly: 20, 45, 65 mm PMMA exposed as for clinical setting.
                   Equipment                                                     PMMA: plates 10x180x240 mm3, densitometer.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   75
H2   20-09-2005   20:54    Pagina 76
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                   with each AEC sensor. Choose the sensors manually. The optical density at the position of the
                   AEC sensor, which was used for that particular image, should be measured.
                   To test whether the correct AEC sensor is chosen, extra attenuation material (for example: 2 or 3
                   aluminium sheets used for HVL measurements) should be positioned above one AEC sensor
                   position. The markers on the compression paddle can be used as guidance. The whole sensor
                   should be covered and adjacent sensors should not be covered. The sensor, above which the
                   extra attenuation has been placed, must be chosen automatically by the system. If another
                   sensor is chosen, increase the amount of attenuating material until the correct sensor is chosen
                   or until it is beyond any doubt that the sensor does not work properly. This procedure must be
                   repeated for all sensor positions.
                   Note: If the Heel effect is large, it may be necessary to add extra attenuating material for sensor
                         positions near nipple side. The marker on the compression paddle may not always
                         completely coincide with the real position of the sensor.
                   Limiting value                                     The variation in optical density between all AEC sensors should
                                                                      be within 0.20 OD. The correct AEC sensor must be chosen.
                   Frequency                                          Every six months: full test.
                   Equipment                                          Standard test block, densitometer.
                   2a.2.1.4 Compression
                   The compression of the breast tissue should be firm but tolerable. There is no optimal value
                   known for the force, but attention should be given to the applied compression and the accuracy
                   of the indication. All units must have motorised compression.
                   When compression force is indicated on the console, it should be verified whether the figure
                   corresponds with the measured value. It should also be verified whether the applied
                   compression force is maintained over a period of 1 minute. A loss of force over this time may be
                   explained, for example, by a leakage in the pneumatic system.
                   Limiting value                                     Maximum automatically applied force: 130 - 200 N. (~ 13-20 kg),
                                                                      and must be maintained unchanged for at least 1 minute. The
                                                                      indicated compression force should be within ± 20 N of the
                                                                      measured value.
                                                                      The compression device should not contain any cracks or sharp
                                                                      edges.
                   Frequency                                          Yearly.
                   Equipment                                          Compression force test device.
                   76               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 77
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                   a
bucky
foam rubber
Note: Not correcting the doses for the inverse square law will result in an over estimation of 5%.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   77
H2   20-09-2005   20:54    Pagina 78
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                   Remark: For some systems it is not possible to image the grid due to the minimum required
                           exposure time.
                   2a.2.2.2 Screen-film
                   The current image receptor in screen-film mammography consists of a cassette with one
                   intensifying screen in close contact with a single emulsion film. The performance of the stock of
                   cassettes is described by the inter cassette sensitivity variation and screen-film contact.
                   2a.2.2.2.1 Inter cassette sensitivity and attenuation variation and optical density range
                   The differences between cassettes can be assessed with the reference exposure (section 2a.1).
                   Select an AEC setting (should be the normal position and using a fixed tube voltage, target and
                   filter) to produce an image having about the clinically used mean optical density on the
                   processed film. Repeat for each cassette using films from the same box or batch. Make sure the
                   cassettes are identified properly. Measure the exposure (in terms of mGy or mAs) and the
                   corresponding optical densities on each film at the reference ROI. To ensure that the cassette
                   tests are valid the AEC system in the mammography unit needs to be sufficiently stable. It will be
                   sufficient if the variation in repeated exposures selected by the AEC for a single cassette is (in
                   terms of mGy and mAs) < ± 2%.
                   Limiting value                                     The exposure, in terms of mGy (or mAs), must be within ± 5% of
                                                                      the mean for all cassettes.
                                                                      The maximum difference in optical density between all
                                                                      cassettes: ± 0.10 OD is acceptable, ± 0.08 OD is achievable.
                   Frequency                                          Yearly, and after introducing new screens.
                   Equipment                                          Standard test object, dosemeter, densitometer.
                   Limiting value                                     No significant areas (i.e. > 1 cm2) of poor contact are allowed in
                                                                      the diagnostically relevant part of the film.
                   Frequency                                          Every six months and after introducing new screens.
                   Equipment                                          Mammography screen-film contact test device, densitometer and
                                                                      viewbox.
                   78               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 79
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                   a
                   2a.2.3.2.1 Sensitometry
                   Use a sensitometer to expose a film with light and insert the exposed side into the processor
                   first. Before measuring the optical densities of the step-wedge, a visual comparison can be made
                   with a reference strip to rule out a procedure fault, like exposure with a different colour of light or
                   exposure of the base instead of the emulsion side.
                   From the characteristic curve (the graph of measured optical density against the logarithm of
                   exposure by light) the values of base and fog, maximum density, speed and film gradients can be
                   derived. These parameters characterise the processing performance. A detailed description of
                   these ANSI-parameters and their clinical relevance can be found in appendix 2, film parameters.
                   Note: There is no clear evidence for the optimal value of film gradient; the ranges quoted are
                         based on what is typical of current practice and are dependent on the film, which is used.
                         At the top end of these ranges the high film gradient may lead to under- and over exposure
                         of parts of the image for some types of breast, thereby reducing the information content.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   79
H2   20-09-2005   20:54    Pagina 80
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                          A further complication of using a very high film contrast is that stable conditions with very
                          low variability of the parameters are required to achieve any benefit in terms of overall
                          image quality (See appendix 3).
                   The assessment of variations can be found in the use of the following table, where the values are
                   expressed as a range (Max value - Min value). Acceptable and achievable ranges are quoted in
                   the table below. For centres where computer facilities for calculating speed and film gradient
                   (Mgrad and Grad1,2) are not available, speed and contrast indices are given. However, this
                   approach is less satisfactory as these indices are not pure measures of speed and contrast.
mean gradient (Mgrad) < 10% of baseline value < 5% of baseline value
                   2a.2.3.2.3 Artefacts
                   An image of the standard test block obtained daily, using a routine exposure, should be
                   inspected. This should show a homogeneous density, without significant scratches, shades or
                   other marks indicating artefacts.
                   2a.2.3.3 Darkroom
                   Light tightness of the darkroom should be verified. It is reported, that about half of darkrooms
                   are found to be unacceptable. Extra fogging by the safelights must be within given limits.
                   80               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 81
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                     a
                   Open the cassette with pre-exposed film and position the film (emulsion up) on the (appropriate
                   part of the) workbench. Cover half the film and expose for two minutes. Position the cover
                   parallel to the tube axis to avoid the influence of the heel effect in the measurements. Measure
                   the optical density difference of the background (Dbg) and the fogged area (Dfogged). The extra fog
                   (ΔD) equals:
                   2a.2.3.3.2 Safelights
                   Perform a visual check that all safelights are in good working order (filters not cracked). To
                   measure the extra fog as a result of the safelights, repeat the procedure for light leakage but with
                   the safelights on. Make sure that the safelights were on for more than 5 minutes to avoid start-
                   up effects.
                   Since good viewing conditions are important for the correct interpretation of the diagnostic
                   images, they must be optimised. Although the need for relatively bright light boxes is generally
                   appreciated, the level of ambient lighting is also very important and should be kept low. In
                   addition it is imperative that glare is minimised by masking the film.
                   The procedures for photometric measurements and the values required for optimum
                   mammographic viewing are not well established. However there is general agreement on the
                   parameters that are important. The two main measurements in photometry are luminance and
                   illuminance. The luminance of viewing boxes is the amount of light emitted from a surface
                   measured in candela/m2. Illuminance is the amount of light falling on a surface and is measured
                   in lux (lumen/m2). The illuminance that is of concern here is the light falling on the viewing box,
                   i.e. the ambient light level. (An alternative approach is to measure the light falling on the film
                   reader’s eye by pointing the light detector at the viewing box from a suitable distance with the
                   viewing box off.) Whether one is measuring luminance or illuminance one requires a detector and
                   a photometric filter. This combination is designed to provide a spectral sensitivity similar to the
                   human eye. The collection geometry and calibration of the instrument is different for luminance
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n     81
H2   20-09-2005   20:54    Pagina 82
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                   and illuminance. To measure luminance a lens or fibre-optic probe is used, whereas a cosine
                   diffuser is fitted when measuring illuminance. Where the only instrument available is an
                   illuminance meter calibrated in lux it is common practice to measure luminance by placing the
                   light detector in contact facing the surface of the viewing box and converting from lux to cd/m2 by
                   dividing by π. Since this approach makes assumptions about the collection geometry, a correctly
                   calibrated luminance detector is preferred.
                   There is no clear consensus on what luminance is required for viewing boxes. It is generally
                   thought that viewing boxes for mammography need to be higher than for general radiography. In a
                   review of 20 viewing boxes used in mammographic screening in the UK, luminance averaged
                   4500 cd/m2 and ranged from 2300 to 6700 cd/m2. In the USA the ACR recommend a minimum
                   of 3500 cd/m2 for mammography. However some experts have suggested that the viewing box
                   luminance need not be very high provided the ambient light is sufficiently low and that the level
                   of ambient light is the most critical factor. The limiting values suggested here represent a
                   compromise position until clearer evidence is available.
                   2a.2.4.1.2 Homogeneity
                   The homogeneity of a single viewing box is measured by multiple readings of luminance over the
                   surface of the illuminator, compared with the luminance in the middle of the viewing panel.
                   Readings very near the edges (e.g. within 5 cm) of the viewing box should be avoided. Gross
                   mismatch between viewing boxes or between viewing conditions used by the radiologist and
                   those used by the radiographer should be avoided. If a colour mismatch exists, check to see that
                   all lamps are of the same brand, type and age. The local personnel has to make sure that all
                   tubes are changed at the same time. To avoid inhomogeneities as a result of dust, clean the light
                   boxes should be regularly cleaned inside and outside.
                   Limiting value                                     The luminance across each panel should be within 30% of the
                                                                      luminance in the centre of the panel.
                   Frequency                                          Yearly.
                   Equipment                                          Luminance meter.
                   82               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 83
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                           a
                   2a.2.5.1 Dosimetry
                   Image PMMA plates of 20 mm thickness in clinical settings. Record the entrance surface air
                   kerma and the exposure factors chosen by the AEC. Repeat this measurement for 30, 40, 45,
                   50, 60 and 70 mm PMMA thickness. Calculate the average glandular dose for a breast
                   equivalent to each PMMA thickness. A detailed description of the calculation of the average
                   glandular dose can be found in appendix 5.
Limiting value A maximum average glandular dose is set per PMMA thickness:
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n           83
H2   20-09-2005   20:54    Pagina 84
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                   Note: If the resolution is measured at different heights between 25 and 50 mm from the
                         tabletop it can differ by as much as 4 lp/mm. The distance from the chest wall edge is
                         critical, but the position parallel to the chest wall side is not critical within ± 5 cm from the
                         reference ROI. Resolution is generally worse parallel to the tube axis due to the
                         asymmetrical shape of the focal spot.
                   Remark: The value for image contrast is dependent on the whole imaging chain, therefore no
                           absolute limits are given. Ideally the object is part of, or placed on top of, the daily
                           quality control test object.
                   84               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 85
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                   a
                   • AEC reproducibility
                   • Tube output stability
                   • Reference optical density
                   • Spatial resolution
                   • Image contrast
                   • Threshold contrast visibility
                   • Homogeneity, artefacts
                   • Sensitometry (speed, contrast, fog)
                   Practical considerations:
                   • Ideally the sensitometric stepwedge should be on the same film as the image of the test
                     object, to be able to correct optimally for the processing conditions.
                   • To improve the accuracy of the daily measurement, the test object should be designed in such
                     a way that it can be positioned reproducibly on the bucky.
                   • The shape of the test object does not have to be breast-like. To be able to perform a good
                     homogeneity check, the test object should cover the normally imaged area on the image
                     receptor (180x240 mm).
                   • For testing the AEC reproducibility, the PMMA test object may comprise several layers of
                     PMMA, 10 or 20 mm thick. It is important to use the same PMMA blocks since variations in
                     thickness of the PMMA plates will influence the tube load (mAs) read-out. Sufficient blocks
                     are required to make up a thickness in the range 20-70 mm to adequately simulate the range
                     of breast thickness found clinically.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   85
H2   20-09-2005   20:54    Pagina 86
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
         2a.4 Tables
         Table 2a.4.1 Frequency of quality control, measured and limiting values
         2a.2.1 X-ray generation and control                         frequency                   typical                             limiting value                                                            unit
                                                                                                  value                      acceptable          achievable
         X-ray source
             - focal spot size                                               i                     0.3                       IEC/NEMA                                         -                                 -
             - source-to-image distance                                      i                    ≥ 600                          -                                            -                                mm
             - alignment of x-ray field/                                    12                       -                          <5                                         <5                                  mm
             - image receptor
             - film/bucky edge                                              12                         -                    ≤5                                             ≤5                            mm
             - radiation leakage                                             i                         -                    <1                                            <1                            mGy/hr
             * output                                                        6                         -                   > 30%                                          > 40                         µGy/mAs
                                                                                                                      > 70% of baseline
         tube voltage
             - reproducibility                                              6                          -                    < ± 0.5                                < ± 0.5                                      kV
             - accuracy (25 – 31 kV)                                        6                          -                    < ± 1.0                                < ± 1.0                                      kV
             - HVL                                                          12                         -                See appendix 5                         See appendix 5
                                                                                                                          table A5.3                             table A5.3
         AEC
               * central opt. dens                                           6                         -         < ± 0.15 of target value                                     -                                OD
               - control setting
               - target opt. dens. control setting                           6                 -                             1.4 -1.9                                                                          OD
               - opt. dens. control step                                     6                 -                            0.05 - 0.20                            0.05 - 0.10                                 OD
               - adjustable range                                            6                 -                             > 1.0z a                               > 1.0z a                                   OD
               * short term reproducibility                                  6                 -                             < ± 5%a                                < ± 2%zz                                   mGy
               * long term reproducibility                                   d                 -                             < ± 0.20                               < ± 0.15                                   OD
               - object thickness                                            w                 -                             < ± 0.15                               < ± 0.10                                   OD
               - and tube voltage compensation                               6                 -                             < ± 0.15                               < ± 0.10                                   OD
               - spectra                                                     6          See appendix 4
               - correspondence between AEC sensors                          6                 -                                 < 0.20                                                                        OD
         compression
            - compression force                                             12                         -                      130 - 200                                    -                                    N
            - maintain force for 1 minute                                   12                         -                         1                                        1                                    min
            - compression force indicator                                   12                         -                       < ± 20                                   < ± 20                                  N
            - compression plate alignment,                                  12                         -                         ≤5                                       ≤5                                   mm
            - symmetric
         2a.2.2 Bucky and image receptor                             frequency                  typical                              limiting value                                                            unit
                                                                                                 value                       acceptable          achievable
         screen-film
             * inter cassette sensitivity                                   12                         -                         < ± 5%                                 < ± 5%                                 mGy
             - - variation (mAs)
             * inter cassette sensitivity                                   12                         -                       < ± 0.10                               < ± 0.08                                 OD
             - - variation (OD range)
             - screen-film contact                                          12                         -            No significant areas                                      -                                  -
                                                                                                                      of poor contact
                   86               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005     20:54         Pagina 87
                     EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                     S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                       a
          processor
              - temperature                                                                    i                 34 - 36                                -                                       -         °C
              - processing time                                                                i                 90 - 120                               -                                       -          s
          film
                  - sensitometry: base and fog                                                d               0.15 - 0.251         -                 -                                                    OD
                                   speed                                                      d                    -               -                 -                                                     -
                                   Contrast Mgrad:                                            d                3.0 - 4.02          -                 -                                                     -
                                   Grad1,2                                                    d                3.5 - 5.0           -                 -                                                     -
                  - daily performance                                                         d                    -        See 2a.2.3.2        See 2a.2.3.2                                               -
                  - artefacts                                                                 d                    -    No disturbing artefacts      -                                                     -
          darkroom
              - light leakage (extra fog                                                    12                          -                       ≤ + 0.02                               ≤ + 0.02           OD
              - in 2 minutes)
              - safelights (extra fog                                                       12                          -                       ≤ + 0.10                               ≤ + 0.10           OD
              - in 2 minutes)
          viewing box
              - luminance                                                                   12                          -                  3000 - 6000                            3000 - 6000            cd/m2
              - homogeneity                                                                 12                          -                    < ± 30%                                < ± 30%              cd/m2
              - luminance difference                                                        12                          -                    < ± 15%                                < ± 15%              cd/m2
                between panels
          environment
              - ambient light level                                                         12                          -                          < 50                                   < 50            lux
          dosimetry
              * Glandular dose                                                                6
                - PMMA thickness (cm)
                  2.0                                                                                                                              < 1.0                                  < 0.6          mGy
                  3.0                                                                                                                              < 1.5                                  < 1.0          mGy
                  4.0                                                                                                                              < 2.0                                  < 1.6          mGy
                  4.5                                                                                                                              < 2.5                                  < 2.0          mGy
                  5.0                                                                                                                              < 3.0                                  < 2.4          mGy
                  6.0                                                                                                                              < 4.5                                  < 3.6          mGy
                  7.0                                                                                                                              < 6.5                                  < 5.1          mGy
          image quality                                                                       --
              * spatial resolution,                                                           w                         -                          > 12                                   > 15           lp/mm
              - - reference ROI
              * threshold contrast visibility                                               w                           -                        < 1.5%                                 < 1.5%             -
              * exposure time                                                               12                          -                          <2                                    < 1.5             s
          End of table 2a.4.1                         i = At acceptance; d = daily; w = weekly; 6 = every 6 months; 12 = every 12 months * standard measurement conditions
                                                      1. For standard blue based films only 2. Depend on the film which is used
                     E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n       87
H2   20-09-2005   20:54    Pagina 88
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
sensitometer - ± 2% OD
densitometer ± 2% at 1.0 ± 1% OD
dosemeter ± 5% ± 1% mGy
aluminium stepwedge
resolution pattern
stopwatch
tape measure
lead sheet
                   88               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54    Pagina 89
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y   a
                   2a.5 Bibliography
                   CEC-Reports
                   1. Technical and Physical Parameters for Quality Assurance in Medical Diagnostic Radiology;
                      Tolerances, Limiting Values and Appropriate Measuring Methods.
                      1989: British Institute of Radiology; BIR-Report 18, CEC-Report EUR 11620.
                   2. Optimisation of Image Quality and Patient Exposure in Diagnostic Radiology.
                      1989: British Institute of Radiology; BIR-Report 20, CEC-Report EUR 11842.
                   3. Dosimetry in Diagnostic Radiology.
                      Proceedings of a Seminar held in Luxembourg, March 19-21, 1991.
                      1992: Rad. Prot. Dosimetry vol 43, nr 1-4, CEC-Report EUR 14180.
                   4. Test Objects and Optimisation in Diagnostic Radiology and Nuclear Medicine.
                      Proceedings of a Discussion Workshop held in Würtzburg (FRG), June 15-17, 1992
                      1993: Rad. Prot. Dosimetry vol 49, nr 1-3; CEC-Report EUR 14767.
                   5. Quality Control and Radiation Protection of the Patient in Diagnostic Radiology and Nuclear
                      Medicine.
                      1995: Rad. Prot. Dosimetry vol 57, nr 1-4, CEC-Report EUR 15257.
                   6. European Guidelines on Quality Criteria for Diagnostic Radiographic Images.
                      1996: CEC-Report EUR 16260.
                   Protocols
                   1. The European Protocol for the Quality Control of the Technical Aspects of Mammography
                       Screening.
                       1993: CEC-Report EUR 14821.
                   2. European Protocol on Dosimetry in Mammography.
                       1996: CEC-Report EUR 16263.
                   3. Protocol acceptance inspection of screening units for breast cancer screening, version April
                       2002.(in Dutch)
                       National Expert and Training Centre for Breast Cancer Screening, University Hospital
                       Nijmegen (NL) 2002.
                   4. LNETI/DPSR: Protocol of quality control in mammography (in English) 1991.
                   5. ISS: Controllo di Qualità in Mammografia: aspetti technici e clinici.
                       Instituto superiore de sanità (in Italian),
                       1995: ISTASAN 95/12.
                   6. IPSM: Commissioning and Routine testing of Mammographic X-Ray Systems - second edition
                       The Institute of Physical Sciences in Medicine, York.
                       1994: Report no. 59/2.
                   7. American College of Radiology (ACR), Committee on Quality Assurance in Mammography:
                       Mammography quality control.
                       1994, revised edition.
                   8. American Association of Physicists in Medicine (AAPM): Equipment requirements and quality
                       control for mammography.
                       1990: report No. 29.
                   9. Quality Control in Mammography,
                       1995: Physics consulting group Ontario Breast Screening Programme.
                   10. QARAD/LUCK: Belgisch Protocol voor de kwaliteitszorg van de fysische en technishe
                       aspecten bij mammografische screening (in Dutch) 1999.
                   11. Protocollo italiano per il controllo di qualità degli aspetti fisici e tecnici in mammografia,
                       2004: AIFM report n. l, http://www.aifm.it/report/
                   Publications
                   1. Chakraborty D.P.: Quantitative versus subjective evaluation of mammography accreditation
                       test object images.
                       1995: Med. Phys. 22(2):133-143.
                   2. Wagner A.J.: Quantitative mammography contrast threshold test tool.
                       1995: Med. Phys. 22(2):127-132.
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                   90               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 91
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                   a
                   26. K.C. Young, M.L. Ramsdale, A. Rust: Mammographic dose and image quality in the UK
                       breast screening programme. 1998, NHSBSP report 35.
                   27. J. Zaers, S. van Woudenberg, G. Brix: Qualitätssicherung in der Röntgenmammographie
                       1997: Der Radiologe (37):617-620.
                   28. J. Law: Checking the consistency of sensitometers and film processors in a mammographic
                       screening.
                       programme. 1996 Brit. J. Of Radiology (69) 143-147.
                   29. K.J. Robson, C.J. Kotre, K. Faulkner : The use of a contrast-detail test object in the
                       optimization of optical density in mammography, 1995 Brit. J. Of Radiology (68) 277-282.
                   30. J.A. Terry, R.G. Waggener, M.A. Miller Blough: Half-value layer and intensity variations as a
                       function of position in the radiation field for film screen mammography 1999.
                       Med. Phys. 26 259-266.
                   31. S. Tang, G.T. Barnes, R.L. Tanner: Slit camera focal spot measurement errors in
                       mammography.
                       1995 Med. Phys. (22) 1803-1814.
                   32. J. Coletti et al.: Comparison of exposure standards in the mammography x-ray region, 1997.
                       Med. Phys (8) 1263-1267.
                   33. C. Kimme Smith et al. Mammography film processor replenishment rate: Bromide level
                       monitoring.
                       1997 Med. Phys. (3) 369-372.
                   35. M. Goodsitt, H. Chan, B. Liu: Investigation of the line-pair method for evaluating
                       mammographic focal spot performance, 1997. Med. Phys. (1) 11-15.
                   34. A. Krol et al. Scatter reduction in mammography with air gap.
                       1996 Med. Phys. (7) 1263-1270.
                   35. D. McLean, J. Gray: K-characteristic photon absorption from intensifying screens and other
                       materials:
                       Theoretical calculations and measurements.
                       1996 Med. Phys. (7) 1253-1261.
                   36. P. Rezentes, A de Almeida, G. Barnes: Mammographic Grid Performance.
                       1999 Radiology 210:227-232.
                   37. J. Hogge et al. Quality assurance in Mammography: Artifact Analysis.
                       1999 Radiographics 19:503-522.
                   38. J. Byng et al.: Analysis of Mammographic Density and Breast Cancer Risk from Digitized
                       Mammograms.
                       1998 Radiographics 18:1587-1598.
                   39. D.R. Dance et al.: Additional factors for the estimation of mean glandular breast dose using
                       the UK mammography dosimetry protocol, 2000 Phys. Med. Biol. (45) 3225-3240.
                   Other reports
                   1. International Electrotechnical Commission (IEC), Geneva, Switzerland: Characteristics of
                       focal spots in diagnostic X-ray tube assemblies for medical use.
                       1982: IEC-Publication 336.
                   2. Quality assurance in mammography - quality control of performance and constancy
                       1990: Series of Nordic Reports on radiation Safety No. 1, Denmark, Finland, Iceland, Norway
                       and Sweden.
                   3. Société française des physiciens d’hôpital, Nancy: Contrôle de qualité et mesure de dose
                       en mammographie - aspects théoriques et pratiques (in French).
                       1991.
                   4. Department Health & Social Security, Supplies Technology Division (DHSS): Guidance notes
                       for health authorities on mammographic equipment requirements for breast cancer
                       screening.
                       1987: STD
                   5. Department of Radiodiagnostic Radiology, University of Lund, Sweden: Quality Assurance in
                       Mammography.
                       1989.
                   6. Sicherung der Bildqualität in röntgendiagnostischen Betrieben - Filmverarbeitung.
                       1996: DIN 6868-2: Beuth Verlag GmbH, Berlin.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   91
H2   20-09-2005   20:54    Pagina 92
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                   92               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 93
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                   a
                   QC report
                   based on
Fourth edition
Date:
Contact:
Institute:
Address:
Telephone:
Conducted by:
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   93
H2   20-09-2005   20:54    Pagina 94
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                   * pinhole method
                         diameter pinhole                                                                                                                           µm
                         distance pinhole-to-film                                                            dpinhole-film                                          mm
                         distance focus-to-pinhole                                                           dfocus-pinhole                                         mm
                         diameter pinhole ⊥ AC axis                                                          f⊥                                                     mm
                         diameter pinhole // AC axis                                                         f //                                                   mm
f// = mm
Accepted: yes / no
                   94               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 95
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                        a
Accepted: yes / no
Accepted: yes / no
Accepted: yes / no
Accepted: yes / no
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n        95
H2   20-09-2005   20:54    Pagina 96
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
Accepted: yes / no
= mm Al
HVL: mm Al
Deviation exposure at 0 mm Al : %
                   96               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 97
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                    a
2a.2.1.3 AEC-system
                   2a.2.1.3.1 Optical density control setting: central value and difference per step
                      Target density value:     OD
Accepted: yes / no
Adjustable range: OD
Accepted: yes / no
Accepted: yes / no
2a.2.1.3.4 Long term reproducibility: Forms should be made to suit the local preferences.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n    97
H2   20-09-2005   20:54    Pagina 98
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
Accepted: yes / no
2a.2.1.4 Compression
Accepted: yes / no
                   98               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 99
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                   a
                      Symmetric load
                      Thickness indication:                                    cm
Accepted: yes / no
Accepted: yes / no
2. yes/no
3. yes/no
Accepted: yes / no
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   99
H2   20-09-2005   20:54    Pagina 100
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                   2a.2.2.2 Screen-film
                   2a.2.2.2.1 Inter cassette sensitivity and attenuation variation and optical density range
                    AEC setting:
Accepted: yes / no
Accepted: yes / no
                   100              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 101
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                    a
                                                                                     Developer                        Fixer
                      Reference/nominal:
                      Thermometer
                          Reference:
                          Local:
                          Console:
                   2a.2.3.3 Darkroom
                   2a.2.3.3.1 Light leakage
                    Fog (after 2 min.) of a pre-exposed film on the workbench:
                    point:                  1        2       3       4      5
                    D(fogged):                                                                                                                         OD
                    D(background):                                                                                                                     OD
                    Difference:                                                                                                                        OD
                    Average difference:          OD
Accepted: yes / no
2a.2.3.3.2 Safelights
Accepted: yes / no
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   101
H2   20-09-2005   20:54    Pagina 102
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
                   Viewing panel                         1                                 2                                3                                4                                5
                   Luminance (cd/m2)
                     Centre
                     Top left
                     Top right
                     Bottom left
                     Bottom right
                   Difference in
                   luminance
                   between the
                   centres (%)
                   Maximum deviation
                   in luminance compared
                   to the luminance in the
                   centre (%)
Accepted: yes / no
Accepted: yes / no
Accepted: yes / no
                   102              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 103
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y                                                                         a
image 1
image 2
image 3
image 4
Accepted: yes / no
Graph(s) attached
Accepted: yes / no
0,2
0,5
1,0
2,0
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n     103
H2   20-09-2005   20:54    Pagina 104
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   S   C    R   E    E    N    -   F    I   L    M        M     A   M     M    O    G   R   A   P    H    Y
Accepted: yes / no
                   104              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 105
                   Authors
                   R. van Engen
                   K. Young
                   H. Bosmans
                   M. Thijssen
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   105
H2   20-09-2005   20:54    Pagina 106
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   Authors:
                   R. van Engen, Nijmegen, the Netherlands
                   K. Young, Guildford, United Kingdom
                   H. Bosmans, Leuven, Belgium
                   M. Thijssen, Nijmegen, the Netherlands (project leader)
                   Co-authors:
                   R. Visser, Nijmegen, the Netherlands
                   L. Oostveen, Nijmegen, the Netherlands
                   T. Geertse, Nijmegen, the Netherlands
                   R. Bijkerk, Nijmegen, the Netherlands
                   P. Heid, Marseille, France
                   Contributors:
                   P. Baldelli, Ferrara, Italy                                                               A. Noel, Nancy, France
                   B. Beckers, Nijmegen, the Netherlands                                                     K. Pedersen, Oslo, Norway
                   A. Bloomquist, Toronto, Canada                                                            N. Phelan, Dublin, Ireland
                   M. Borowski, Bremen, Germany                                                              F. Rogge, Leuven, Belgium
                   AK. Carton, Leuven, Belgium                                                               M. Säbel, Erlangen, Germany
                   M. Chevalier, Madrid, Spain                                                               M. Schutten, Nijmegen, the Netherlands
                   M. Gambaccini, Ferrara, Italy                                                             F. Shannoun, Luxembourg, Luxembourg
                   S. von Gehlen, Oldenburg, Germany                                                         A. Stargardt, Aachen, Germany
                   G. Gennaro, Padua, Italy                                                                  M. Swinkels, Nijmegen, the Netherlands
                   A. de Hauwere, Gent, Belgium                                                              A. Taibi, Ferrara, Italy
                   B. Johnson, Guildford, United Kingdom                                                     J. Teubner, Heidelberg, Germany
                   B. Lazzari, Florence, Italy                                                               H. Thierens, Gent, Belgium
                   A. Lisbona, Nantes, France                                                                P. Torbica, Innsbruck, Austria
                   A. Maidment, Philadelphia, USA                                                            F. van der Meer, Rotterdam, the Netherlands
                   N. Marshall, London, United Kingdom                                                       F.R. Verdun, Lausanne, Switzerland
                   G. Mawdsley, Toronto, Canada                                                              A. Workman, Belfast, United Kingdom
                   P. Moran, Madrid, Spain
                   Acknowledgements:
                   Comments have been received from the following manufacturers:
                   Agfa, Barco, Eizo, Fischer, Fuji, GE, Hologic, IMS, Instrumentarium, Kodak, Mevis, Sectra,
                   Siemens, X-counter.
                   Furthermore the financial support of the European Commission through the European Breast
                   Cancer Network contracts 2000 (SI2.307923(2000CVG2-031)), 2001 (SI2.328176(2001CVG2-
                   013)) and 2002 (SPC2002482) is acknowledged.
                   106              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 107
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
                   Foreword
                   The ‘European guidelines for quality assurance in mammography screening’ (European
                   Guidelines, 2001) include as chapter 3 the ‘European protocol for the quality control of the
                   physical and technical aspects of mammography screening’. In this protocol the requirements for
                   an adequate screen-film imaging system are defined. In recent years, the image detection
                   technology used in mammography has extended to include the use of digital detector systems.
                   This technology is different in so many ways, that it is necessary to set new quality standards and
                   test procedures specifically for digital systems.
                   This document is based on the addendum to chapter 3 of the European guidelines (3rd edition),
                   which was released in November 2003 (Addendum, 2003). The approach to quality assessment
                   and control in this protocol is comparable in the sense, that the measurement and evaluation of
                   performance are in principle independent of the type and brand of the system used. The
                   measurements are generally based on parameters that are extracted from the images that are
                   produced when a phantom with known physical properties is exposed under defined conditions.
                   The limiting values are based upon the quality that is achieved by screen-film systems, which
                   fulfil the demands of the European guidelines.
                   To fulfil the European guidelines in mammography screening, the digital x-ray system must pass
                   all relevant tests at the acceptable level. The achievable level reflects the state of the art for the
                   individual parameter.
                   This protocol for digital mammography is work-in-progress and subject to improvements as more
                   experience in digital mammography is obtained and new types of digital mammography
                   equipment are developed. Changes in measuring techniques or limiting values will lead to a new
                   version number, changes in wording or added comments will change the sub-number. Updates on
                   the current version will be made available on the EUREF website (www.euref.org). It is
                   recommended that users check the website for updates before testing digital mammography
                   equipment.
In the text some lines are printed in parentheses [like these]. This text is a remark.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   107
H2   20-09-2005   20:54    Pagina 108
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   The general principles for testing the three main parts of the imaging chain, illustrated in figure 1,
                   are discussed below.
                   Processing system
                   • In future the processing system (fig 1, B) may be evaluated by the inspection and scoring of a
                     test set of images (either mammograms or phantom images), which have been processed in
                     the available standard processing algorithm.
                     - These images are to be inserted by the user as ‘unprocessed images’ (DICOM: ‘for
                       processing’) and processed by the software of the manufacturer before displaying.
                     - The manufacturer must provide information in general terms on the processing applied.
                   • The processing algorithms are built to enhance the visibility of specific image details. At this
                     moment little experience and literature on the effects is available. These algorithms therefore
                     are not addressed in the present protocol. The observer is urged to convince himself of the
                     value of the algorithms provided.
                   108              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 109
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
                   • Evaluation of processing algorithms and CAD (computer aided detection) will be addressed in
                     a future version of this protocol.
                   Display system
                   • The display system (fig 1, C) can be evaluated by the inspection on the display system (printer
                     or monitor) of synthetic test images, produced in DICOM format and independent of the
                     phantom images delivered by the manufacturer. The user must be able to insert these images as
                     ‘processed images’ (DICOM: ‘for presentation’). They are not processed further before displaying.
                     Evaluation of such images is necessary to confirm compliance to quality standards other than
                     those of the manufacturer. It must be possible to load and display these phantom images
                     using the imaging system under evaluation.
                   • For the evaluation of the display system this protocol follows the advice of AAPM Task Group
                     18 (Samei, 2004) and of preliminary results of the ACRIN Dmist trial.
                   The measurements in the protocol are in principle chosen and described to be generally
                   applicable. Where the tests are similar to those required for screen-film mammography, a
                   reference to the relevant part of the European guidelines is given. When necessary, different test
                   procedures are given for CR (computed radiology, i.e. photo-stimulable phosphor type) systems
                   and DR (direct radiology, i.e. solid state type, including scanning slot) systems separately.
                   Many measurements are performed by an exposure of a test object. All measurements are
                   performed under normal working conditions: no special adjustment of the equipment is
                   necessary. Since the available settings in the different systems vary in spectrum and X-ray
                   quantity for the different breast thicknesses, no common standard exposure can be indicated.
                   Therefore dose calculations for the comparison of systems are based on the AGD (average
                   glandular dose) to the breast (or simulated breast) rather than on entrance surface air kerma.
                   To evaluate the clinical use of a system, a standard type of exposure is specified: the routine
                   exposure, which is intended to provide information on the system under clinical settings.
                   For the production of the routine exposure, a test object is exposed using the machine settings
                   as follows (unless stated otherwise):
Routine exposure:
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   109
H2   20-09-2005   20:54    Pagina 110
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   Mean pixel values and their standard deviation are measured in a standard region of interest
                   (ROI), which has an area of 4 cm2 and is positioned 60 mm from the chest wall side and laterally
                   centred.
                   Limits of acceptable performance for image quality and dose are based on the limits of
                   acceptable performance of screen-film mammography systems. The relation between dose and
                   limits of visibility of details for a certain contrast are based on the performance of a large number
                   of screen-film systems in the UK, the Netherlands, Germany, Belgium and France. These
                   acceptable limits are given, but often a better result is achievable. When applicable the
                   achievable values are also given. Both the acceptable and achievable values are summarised in
                   Appendix 7. Occasionally no limiting value is given, but only a typical value as an indication of
                   what may normally be expected. The measurement frequencies indicated in the protocol
                   (appendix 6) are the minimum required. When the acceptable limiting value is exceeded the
                   measurement should be repeated. If necessary, additional measurements should be performed
                   to determine the origin of the observed problem and appropriate actions that should be taken to
                   solve the problem.
                   For some tests the limiting values are provisional, this means that the limiting value needs
                   further evaluation and may be changed in the future. Check the EUREF website for updates. In
                   some cases further remarks about the limiting values can be found in a box.
                   Guidance on the specific design and operating criteria of suitable test objects will be given by a
                   separate project group of the European Breast Cancer Network (EBCN). Definition of terms, such
                   as the reference ROI and signal-to-noise-ratio are given in section 2b.1.5. The evaluation of the
                   results of the QC measurements can be simplified by using the forms for QC reporting that are
                   provided on the EUREF homepage (www.euref.org).
                   The staff conducting the daily/weekly QC-tests will need the following equipment 8 at the
                   screening site:
                   The medical physics staff conducting the other QC-tests will need the following additional
                   equipment and may need duplicates of some of the above3:
                   110              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 111
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
                   AEC
                   The As Low As Reasonably Achievable (ALARA) principle on dose administered to the patient
                   necessitates the use of an automated exposure control (AEC) system to ensure the optimal
                   exposure of the image receptor compensating for breast thickness and composition. The use of
                   a look up table, only based on the measured thickness of the compressed breast, increases the
                   mean dose to the patients. This is due to the necessary margin in exposure to avoid increased
                   noise by underexposure in dense breasts and to compensate for the incorrect reading of the
                   thickness.
                   Image receptor
                   The required physical size of the image receptor and the amount of missed tissue at the short
                   sides and especially at the chest wall side are important for an optimal imaging of the breast
                   tissues. An upper limit is given for the amount of missed tissue at chest wall side, but the
                   acceptance of other margins remains the responsibility of the radiologist.
                   Display system
                   Optimal transfer of the information in digital mammograms will be reached, when every pixel in
                   the matrix is projected to at least one pixel on the display system and when the pixel size on the
                   display system is sufficiently small to show details that coincide with the maximum sensitivity of
                   the eye of the observer (1-3 lp/mm at a viewing distance of 30 cm). In screening the monitor
                   should allow for the inspection of the image at full size in full resolution, since the number of
                   images read does not allow time consuming procedures like roaming or zooming. Normally two
                   images are viewed at the same time, and with the current technology it is therefore
                   recommended that diagnostic workstations with two large (45-50 cm diagonal (19-21II)) high
                   quality, 5 megapixel monitors are used.
                   On the acquisition unit it may be acceptable to use a monitor with lower specification, depending
                   on the tasks of the radiographer.
                   Further research is needed to demonstrate whether cheaper solutions (e.g. 3 megapixel
                   monitors) may be sufficient in clinical situations.
                   Viewing conditions
                   Since the maximum intensity on the monitor (300-800 cd/m2) is much lower than that of a
                   viewing box with unexposed and developed film (3000-6000 cd/m2) and due to the reflection
                   characteristics of the monitor, the amount of ambient light might seriously diminish the visible
                   dynamic range and the visibility of low contrast lesions. The ambient light level therefore should
                   be low (less than 10 lux) to allow maximal extension to the lower part of the range. Although this
                   level has proven to be acceptable, a short time to adapt to this level might be necessary.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   111
H2   20-09-2005   20:54     Pagina 112
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   Printer
                   The pixel size of the printer should be in the same order of magnitude as (or less than) the pixel
                   size of the image and should be < 100 micron.
Fig.1
Fig.1A
                   112              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 113
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
Fig.1C (monitor)
Fig.1C (printer)
                   2b.1.4 Philosophy
                   Introduction
                   The primary scope of this document is setting standards for mammography screening, however
                   similar standards are expected for diagnostic mammography.
                   The imaging chain in digital mammography can be divided into three independent parts:
                   1. Image acquisition, which includes the X-ray generation, the image receptor and (for some
                       systems) image receptor corrections.
                   2. Image processing, which includes the image processing software.
                   3. Image presentation, including monitor, image presentation software, printer and viewing box.
                   In the European protocol for the quality control of the physical and technical aspects of
                   mammography screening these parts of the imaging chain are evaluated separately. This is a
                   practical approach because each requires very different evaluation techniques and it allows the
                   use of equipment and software from different vendors. In the present version of the protocol
                   (version 4) only image acquisition and image presentation are considered. Due to the large
                   number of processing techniques and the shortcomings of classical test objects with regard to
                   the evaluation of post processing as histogram and texture based processing, evaluation of the
                   image processing part of the imaging chain has not been addressed (yet). However, manufacturers
                   have to specify in general terms, which image processing techniques are applied and it is
                   advised to evaluate image processing by comparing mammograms to images from the previous
                   screening round by experienced readers.
                   The digital section of this chapter of the European guidelines should not be considered as a
                   guide for the optimal working point of a particular system or as a guide to optimise image quality.
                   Different research groups are studying these issues and manufacturers are still working on the
                   optimizing of current systems and the development of new techniques. We urge the reader of this
                   document to keep track of all new developments in this rapidly changing technology. Updates of
                   this protocol will be available at www.euref.org.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   113
H2   20-09-2005   20:54    Pagina 114
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   Before publication the test methods have been evaluated using a number of different types of
                   digital systems. For some types of systems only a small number of evaluation tests have been
                   performed due to limited accessibility. Due to the rapidly changing technologies, new methods of
                   testing may be necessary in the future. Check for updates on the EUREF website.
                   To remain up-to-date with the latest insights, the protocol will be updated continuously. Latest
                   versions will be made available on the EUREF website (www.euref.org).
                   The philosophy of important QC tests and remarks are explained in the following paragraphs.
                   Therefore, the authors have stated that systems used for mammography screening should
                   incorporate an AEC. Manufacturers of equipment without Automatic Exposure Control (AEC), are
                   114              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 115
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
                   urged to implement such a device in their systems before January 2006. For the time being
                   systems which incorporate an exposure table in the software that account only for compressed
                   breast thickness, will be allowed. We advise against systems in which both spectrum and dose
                   have to be set completely manually.
                   To increase reliability at least six phantom images are required. To reduce the (in-) visibility of
                   small disks due to the accidental relative position of the disks and the dels of the detector the
                   phantom has to be repositioned slightly after acquisition of each image. Extensive window
                   levelling and zooming must be performed to optimize the visibility of the dots in each section of
                   the phantom image. This prevents the monitor from being the limiting factor for the threshold
                   contrast evaluation instead of the quality of the unprocessed image. At least three readers
                   should score two different images each.
                   A problem with scoring CD images is the inter- and intra-reader variability. Therefore CDMAM
                   images with scores are available on the EUREF website for reference purposes. In future, the
                   threshold contrast visibility test may be performed using computer readout of the phantom
                   images. This will solve problems with inter- and intra-observer variability Allowance may need to
                   be made for differences between human and machine measurements of threshold contrast.
                   At this moment image quality is evaluated using a total attenuation equivalent to 50 mm PMMA
                   thickness. This has been chosen because image quality information was available for this
                   thickness. In the future, the image quality evaluation may be performed at the thickness of 45
                   mm PMMA, which has been chosen as the standard thickness for other tests in the European
                   Guidelines.
                   Two kinds of limiting values have been set: acceptable and achievable limiting values. The
                   acceptable limiting values have been derived from screen-film mammography, the achievable
                   limiting values have been derived from current full-field digital mammography systems.
                   The acceptable limiting values have been derived by stating that image quality of digital
                   mammography must be (at least) comparable to screen-film mammography (Young, 2004). For
                   this purpose the image quality of a large number of screen-film mammography systems in
                   different screening programmes has been determined using CD analysis. The CDMAM phantom
                   has been used for these measurements. It was chosen that the image quality limiting values for
                   digital mammography should be such that 97.5% of the screen-film systems in the UK would
                   comply. This means that the image quality limits are not very demanding and it must be realized
                   that a system just complying to the acceptable limits would probably not be considered
                   equivalent to top quality screen-film systems.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   115
H2   20-09-2005   20:54    Pagina 116
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   The resulting limiting values have been checked with the image quality levels found in the Dutch
                   screening and in some of the German screening projects (of which data was available) and were
                   found to be realistic.
                   Furthermore the limiting values have been checked with the CD curves from some hospitals in
                   which it was established (by radiologists) that image quality of the digital system was too low for
                   mammography. (The visibility of microcalcifications was regarded insufficient). Threshold
                   contrast visibility for small diameters did not meet the acceptable limiting values in these
                   hospitals. The image quality of a system is only acceptable if contrast threshold values for all
                   diameters comply with the limiting values.
                   The achievable limiting values have been derived as averages from a number of established
                   digital mammography systems. At the EUREF website CDMAM images with scores can be found
                   for reference purposes.
                   In the view of the authors CNR would be the right measure to quantify image quality at
                   thicknesses other than standard thickness. CNR should not necessarily be equal across the
                   whole range of breast thicknesses. However, problems arise when setting the CNR value at
                   standard thickness as reference for other thicknesses using the method described in version 1.0
                   of the Protocol for digital mammography. If the CNR at standard thickness is high, CNR at other
                   thicknesses may fail, not because image quality is too low, but because image quality at
                   standard thickness is relatively high. So the method of testing and limiting values needed to be
                   revised.
                   In this fourth edition of the guidelines the value of CNR at standard thickness is estimated which
                   would be obtained on a system if this system just complied with the acceptable limiting values of
                   threshold contrast visibility. In the calculation of this minimum CNR level it is assumed that
                   quantum noise is the main source of noise in the system. The calculation is based on the Rose
                   theory, from which can be derived that threshold contrast visibility is inversely related to CNR.
                   The calculated CNR at the acceptable limiting value of threshold contrast is the lower limit of
                   CNR at standard thickness. Lower limits of CNR at other thickness are related to this value
                   providing sufficient CNR for the whole range of breast thickness.
                   116              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 117
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
                   In the third edition of the protocol for screen-film systems, limiting dose values were measured
                   using a standard spectrum. This requirement of the third edition cannot be fulfilled by some
                   digital mammography systems due to the available spectra. For example: scanning slot systems
                   use tungsten instead of molybdenum targets due to the required tube loading. Furthermore using
                   clinical spectra in dose measurements is closer to clinical practice.
                   In the third edition, Entrance Surface Air Kerma (ESAK) limits at standard thickness have to be
                   measured for a given optical density. Practical ESAK values are found to be far below the limiting
                   value, even at the clinically used optical density. In digital mammography the link between
                   limiting dose values and OD is non-existent. Therefore a choice had to be made what limiting
                   dose value would be appropriate for digital mammography. In the view of the authors, inspired by
                   the ALARA principle, dose should not increase substantially when changing to digital
                   mammography. Data from the Dutch (Beckers 2003), Swedish (Leitz 2001), Norwegian (Pedersen
                   2000) and UK (NHSBSP 2000, 2003) screening programmes show that average glandular dose
                   levels in screen-film mammography systems are between 0.8 and 2.5 mGy for 4.5 cm PMMA in
                   clinical settings (corrected for difference in standard PMMA thickness in the UK and the
                   Netherlands). Therefore an average glandular dose limit of 2.5 mGy at standard thickness in
                   clinical settings has been chosen to ensure that dose levels in digital mammography will not
                   exceed those of screen-film mammography. This limiting value is comparable to the objective of
                   the NHSBSP in the UK to have average glandular dose levels of 2 mGy or less (for 4.0 cm PMMA)
                   and the limiting average glandular dose value for the Dutch screening programme (3 mGy for 5.0
                   cm PMMA). In the present version of the protocol limiting dose values for a range of PMMA
                   thickness have been introduced. This has been done because in some non-AEC systems it was
                   noticed that manufacturers decreased dose at standard thicknesses to comply with the limiting
                   value at standard thickness while dose levels at other thickness were found to be (much) higher
                   than found in screen-film mammography. Besides this it has been found that some systems did
                   use much lower tube voltages than in screen-film mammography (thus increasing patient dose
                   substantially). In measurements performed by some of the authors, these very low values proved
                   unnecessary for image quality, therefore the use of these tube voltages does not comply with the
                   ALARA principle. Setting limiting dose levels per PMMA thickness prevents this situation. The
                   limiting values for PMMA thicknesses other than standard thickness have been obtained by
                   averaging all measured glandular dose levels per PMMA thickness from all X-ray units of the
                   Dutch screening programme and some German screening trials. The resulting average glandular
                   dose against PMMA thickness curve has been scaled to the limiting value at standard thickness.
                   The results have been compared with the dose values per PMMA thickness found in the UK and
                   some of the German screening projects (screen-film mammography). The limiting values were
                   found to be reasonable.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   117
H2   20-09-2005   20:54    Pagina 118
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   118              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 119
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
                   performed. For some systems, the latter correction has to be performed by the user. If the user
                   has to perform this calibration, sufficient time must have passed since the last images were
                   made to prevent the influence of possible ghost images on the calibration. These corrections can
                   be checked in a homogeneity test.
                   Strictly speaking the correction for differences in sensitivity is only valid at the spectrum and
                   simulated breast thickness at which calibration is performed. Therefore it is advised to perform
                   the homogeneity test at several clinically relevant spectra and thickness at acceptance. At
                   acceptance, a baseline is established for the homogeneity test. For some types of DR detectors
                   homogeneity changes relatively quickly over time. Therefore it is advised to check image
                   homogeneity regularly (weekly) and check the results with the baseline. The frequency may
                   change in future and may be dependent on the type of digital system. For CR systems the
                   usefulness of the homogeneity test needs to be established. For the moment it is also
                   recommended to perform this test on CR systems.
                   Problems may occur if the Heel effect and geometric effects are relatively large. These effects
                   might influence the results of the image receptor homogeneity measurement. If a specific
                   system does not comply with the provisional limiting values it is advised to check whether
                   geometry or the Heel effect causes this deviation or any malfunction in the system. It is
                   recommended that the images are checked visually for artefacts.
                   2b.1.4.7.7 Ghosting
                   Several reports on ghosting in DR systems have been published (for example: Siewerdsen,
                   1999). In CR systems ghosts may occur if the erasure of the screens is not performed optimally.
                   This ghosting is quantified by comparing the pixel value of an induced ghost image to a known
                   contrast in the image (contrast of an aluminium sheet). After the ghosting measurement it is
                   advised to make some additional images with a homogeneous block of PMMA covering the whole
                   detector to make sure that ghosts will not appear on clinical images.
                   For scanning slot systems ghosts will not show with the proposed method of testing, but any
                   ghosting is included in MTF measurements.
2b.1.4.8.1 Monitors
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   119
H2   20-09-2005   20:54    Pagina 120
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
2b.1.4.3.2 Printers
                   Active display area                               The part of the display used for displaying images, applications
                                                                     and the desktop.
                   Bad pixel map                                     A map (either an image or a table) which defines the position of all
                                                                     pixels of which the pixel value is not based on its own del reading.
                   Computer Aided Detection                          Software to aid the radiologists’ detection of suspect areas in the
                   (CAD)                                             breast image.
                   Contrast to Noise Ratio                           The CNR is calculated as follows for a specific test object (e.g. 0.2
                   (CNR)                                             mm Al thickness on 45 mm PMMA).
                   Detective Quantum                                 Function which describes the transfer of SNR as function of spatial
                   Efficiency (DQE)                                  frequency when recording an X-ray image. The DQE gives the
                                                                     efficiency with which the device uses the available quanta.
                   120              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 121
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
                   Direct Radiography (DR)                                      Digital radiology technology using sealed units mounted on a
                                                                                radiography system, which captures X-rays and produces a digital
                                                                                image by sampling the X-ray image.
Digital Driving Level (DDL) Digital value which is the input for a display system.
                   Exposure time                                                The time between the first and last moment that primary X-rays
                                                                                reach an individual part of an imaged object.
                   Modulation Transfer                                          Function, which describes how the contrast of image components
                   Function (MTF)                                               is transmitted as a function of their spatial frequency content.
                   Noise power spectrum                                         Function which describes image noise as a function of spatial
                   (NPS)                                                        frequency.
                   Pixel pitch                                                  Physical distance between the centres of adjacent pixels. In the
                                                                                DICOM tags pixel pitch is called imager pixel spacing and is
                                                                                generally equal to detector element spacing.
Pixel value offset Constant value that is added to the values of all pixels.
                   Presentation value                                           Pixel value after Value Of Interest Look-Up-Table (VOI LUT) or
                                                                                window width and window level settings have been applied.
                   Primary class display                                        A display device used for the interpretation of medical images
                   device                                                       (also referred to in the text as ‘diagnostic display device’).
                   Processed image                                              The image after image processing, ready for presentation on the
                                                                                monitor or print-out. In the DICOM file the value of tag Pixel
                                                                                Intensity Relationship (0028,1040) is ‘for presentation’.
                   Secondary class                                              A display device used for viewing the images, but not for
                   display device                                               diagnosis.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   121
H2   20-09-2005   20:54    Pagina 122
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   Signal to Noise Ratio                             The SNR is calculated as follows for a specific ROI:
                   (SNR)
                                                                                            mean pixel value - pixel value offset
                                                                     SNR =
                                                                                             standard deviation in pixel value
                   Standard test block                               PMMA test object to represent approximately the average breast
                                                                     (although not an exact tissue-substitute) so that the X-ray machine
                                                                     operates correctly under automatic exposure control and the dose
                                                                     meter readings may be converted into dose to glandular tissue.
                                                                     The thickness is 45 ± 0.5 mm. The standard test block covers the
                                                                     whole detector.
                   Threshold contrast                                The smallest detectable contrast for a given detail size that can be
                                                                     shown by the imaging system with different intensity (density) over
                                                                     the whole dynamic range. The threshold contrast is a measure for
                                                                     imaging of low-contrast structures.
                   Uncorrected image                                 The image in a DR system before any image processing, including
                                                                     detector corrections and flat-fielding, is performed.
                   Window centre                                     Setting defining (together with window width) a linear relationship
                                                                     between modality pixel values and pixel values meaningful for
                                                                     presentation (presentation values).
                   Window width                                      Setting defining (together with window centre) a linear relationship
                                                                     between modality pixel values and pixel values meaningful for
                                                                     presentation (presentation values).
                   122              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 123
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
                         To prevent ghosting artefacts, it is advised to cover the detector with a lead sheet during
                         all tests for which no image is required and use the non-imaging mode (if available) on the
                         X-ray unit.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   123
H2   20-09-2005   20:54    Pagina 124
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   2b.2.1.3 AEC-system
                   It is generally recommended that systems used for mammography screening incorporate an AEC.
                   The performance of the AEC system should be tested in terms of reproducibility and accuracy
                   under varying conditions (object thickness and beam quality). The AEC system should adjust
                   target, filter and tube voltage such that image quality is sufficient and dose is within an
                   acceptable range. Semi-automated systems that start from a user defined target, filter and tube
                   voltage but adapt dose according to breast transparency, are also acceptable.
                   The use of a look-up-table (LUT) for the determination of target, filter, tube voltage and dose
                   based on compressed breast thickness can only be allowed if this LUT is programmed into the X-
                   ray unit. However, it must be realized that these systems do not take breast composition into
                   account and therefore cannot be fully optimized with respect to image quality and dose. For this
                   kind of system some guidance for QC measurements is given in appendix 8.
                   For dose measurements it is essential that the dosimeter is positioned outside the region in
                   which the exposure settings are determined. Alternatively, dose can be calculated using tube
                   loading (mAs) and tube output.
                   2b.2.1.3.1 Exposure control steps: central value and difference per step (if applicable)
                   This test item only applies to mammography units with exposure control steps. Image the
                   standard test block at the different exposure control steps (or a relevant subset). Record
                   entrance dose (or tube loading). Calculate exposure steps in entrance dose (or tube loading).
                   Remark: If it is noticed that the system switches between two spectra, release the compression
                           paddle and compress again or use another PMMA thickness (add for example 0.5 cm
                           PMMA) to force the choice of one single spectrum and repeat the measurement.
                   The central setting is the standard setting. In this setting image quality must be sufficient, this is
                   determined by contrast threshold visibility measurements, see section 2b.2.4.1.
                   124              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 125
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
                   Warning: An incorrect functioning of the back-up timer could damage the tube. To avoid excessive
                            tube load, consult the manual for maximum permitted exposure time.
                   Remark: If it is noticed that the system switches between two spectra, release the compression
                           paddle and compress again or use another PMMA thickness (add for example 0.5 cm
                           PMMA) to force the choice of one single spectrum and repeat the measurement.
                   Limiting value                                                The variation of SNR in the reference ROI and dose < ± 10%.
                   Frequency                                                     Weekly.
                   Equipment                                                     Standard test block.
                   Image PMMA plates of 20 mm thickness, with an aluminium object of 0.2 mm thickness on top,
                   if necessary in manual mode and with settings as close as possible to the clinical AEC settings
                   (if manual mode is used, substract the pre-exposure from the settings). Position the aluminium
                   object as shown in figure 2.1. Measure the mean pixel value and standard deviation in a ROI (4
                   cm2) with (position 2) and without (position 1) aluminium object. Calculate CNR. Repeat this
                   measurement for 30, 40, 45, 50, 60 and 70 mm PMMA thickness.
Fig. 2.1 Position of the aluminium filter for the CNR measurement
                   Image quality is evaluated for one thickness (at the equivalent of 5.0 cm PMMA) using contrast
                   threshold measurements (section 2.4.1). At other PMMA thicknesses CNRlimiting value is related to
                   the CNRlimiting value at 5.0 cm PMMA to ensure image quality at other thicknesses14.
The following formula is used to calculate the limiting value of CNR at standard thickness:
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   125
H2   20-09-2005   20:54    Pagina 126
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   The value of CNR at 5.0 cm thickness is related to the measured threshold contrast visibility in
                   section 2b.2.4.1. Using the formula above the limiting value of CNR at standard thickness can
                   be estimated using the measured threshold contrast in section 2b.2.4.1 and the (acceptable)
                   limiting value of value of the 0.1 mm diameter disc. The calculated CNRlimiting value should be used
                   as the 100% level mentioned in the limiting values below.
                   Limiting value                                     CNR per PMMA thickness, see table for provisional limiting
                                                                      values; Compare CNR values with results at acceptance
                                                                                   PMMA                                                           CNR15
                                                                                 Thickness                                             (relative to 5.0 cm PMMA)
[cm] [%]
6.0 > 95
7.0 > 90
                   2b.2.1.4 Compression
                   Use the method and limiting values described in section 2a.2.1.4 of the screen-film part of the
                   European guidelines.
                   126              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 127
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
                   For systems with a non-linear response, such as currently available CR systems, plot mean pixel
                   value against . log relative entrance surface air kerma. Refer to the information provided by the
                   manufacturer whether pixel value should be linear or logarithmic versus entrance surface air
                   kerma at the applied screen processing. Post processing should be turned off. The screen
                   processing should be turned off as much as possible (see appendix 7). Determine linearity by
                   plotting a best fit through all measured points. Calculate the square of the correlation coefficient
                   (R2). Compare the results to previous measurements.
                   Appendix 7 provides information about the relation between entrance surface air kerma and
                   exposure indicator for some CR systems and screen processing modes.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   127
H2   20-09-2005   20:54    Pagina 128
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   For CR systems: No post processing should be applied, the screen processing should be turned
                   off as much as possible (see appendix 7).
                   Perform this measurement at acceptance also at other PMMA thickness (for example with PMMA
                   blocks of 20 and 70 mm thickness). For all measurements clinical settings should be used.
                   For CR systems: No post processing should be applied, the screen processing should be turned
                   off as much as possible (see appendix 7).
                   128              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 129
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
                         It is acknowledged that the Heel effect and geometry effects influences the results of the
                         homogeneity measurement. If a specific system does not comply with the provisional
                         limiting values it is advised to check whether geometry or the Heel effect causes this
                         deviation or some malfunction in the system. For CR systems an additional homogeneity
                         image can be obtained by exposing a cassette using half dose under normal conditions
                         and half dose with the cassette rotated 180° in the bucky to minimize the Heel effect and
                         geometric effects.
                   Limiting value                                                (provisional) Maximum deviation in mean pixel value < ± 15% of
                                                                                 mean pixel value in whole image, maximum deviation in SNR
                                                                                 < ± 15% of mean SNR in all ROI’s, maximum variation of the
                                                                                 mean SNR between weekly images < ± 10%, entrance surface air
                                                                                 kerma (or tube loading) between weekly images < ± 10%.
                   Frequency                                                     Weekly and after maintenance, at acceptance also at 20 and
                                                                                 70 mm PMMA thickness.
                   Equipment                                                     Standard test block covering the complete detector, at acceptance
                                                                                 also PMMA blocks of 20 and 70 mm thickness covering the complete
                                                                                 detector, software for determining detector homogeneity.
                   Limiting value                                                No limits have been set yet on the number of uncorrected
                                                                                 defective detector elements.
                   Frequency                                                     Weekly.
                   Equipment                                                     Standard test block covering the complete detector, at acceptance
                                                                                 also PMMA blocks of 20 and 70 mm thickness covering the
                                                                                 complete detector.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   129
H2   20-09-2005   20:54    Pagina 130
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   Limiting values                         SNR variation in the reference ROI between all imaging plates
                                                           < ± 15%, variation in entrance surface air kerma (or tube loading)
                                                           < ± 10%, no major inhomogeneities on the images.
                   Frequency                               Yearly and after introducing new imaging plates.
                   Equipment                               Standard test block.
                   2b.2.3 Dosimetry
                   Use the method and limiting values described in paragraph 2.5.1 of the screen-film part of the
                   European guidelines. The PMMA plates should cover the whole detector. For dose measurements
                   it is essential that the dose probe is positioned outside the region in which the exposure settings
                   are determined. Alternatively, dose can be calculated using tube loading (mAs) and tube output.
                   130                    European guidelines for quality assurance in mammography screening Four th edition
H2   20-09-2005   20:54    Pagina 131
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y            b
                   images with different relative position of the details and the detector elements. Three
                   experienced observers should determine the minimal contrast visible on two images. Every
                   observer must score two different images. The whole detail diameter range specified in the table
                   below must be covered. In this range minimal contrast visible for a large number of detail
                   diameter must be determined at acceptance and at least 5 detail diameters in subsequent tests.
                   This evaluation should be done on unprocessed images. The window width and level and zoom
                   facilities must be adjusted to maximise the visibility of the details on the displayed images.
                        It is acknowledged that at present it is not possible to get unprocessed images from some
                        systems. For these systems threshold contrast visibility evaluation should be done on
                        processed images. The image processing may introduce artefacts on phantom images and
                        may be different from image processing for mammograms due to histogram or local
                        texture based processing techniques. Therefore care needs to be taken in interpretation
                        of these processed images.
                   The threshold contrast performance specified here relates to the nominal contrast calculated for
                   the details for a 28 kV tube voltage with molybdenum target and filter materials as explained in
                   appendix 6. This nominal contrast depends on the thickness and materials used to manufacture
                   the test object, and is independent of the actual spectrum used to form the image, which should
                   be that used clinically. It does not include the effects of scatter. The average nominal threshold
                   contrasts should be compared with the limiting values below.
                   For CR systems: No post processing should be applied, the screen processing should be turned
                   off as much as possible (see appendix 7). If the screens comply with the limiting values of
                   section 2b.2.2.4 inter plate sensitivity variations, it is not necessary to use the same screen in
                   the threshold contrast visibility measurement.
                                                                               Threshold contrast
                                                      Acceptable value                                       Achievable value
                   Frequency                               Yearly.
                   Equipment                               Contrast detail phantom.
                   The threshold contrast standards defined in the table above are chosen to ensure that digital
                   mammography systems perform at least as well as screen-film systems (Young, 2004). They
                   have been derived from measurements on screen-film and digital mammography systems using
                   the Nijmegen CDMAM contrast detail phantom version 3.4 (see section 2b.1.4). However it is
                   intended that they are sufficiently flexible to allow testing by other designs and makes of test
                   European guidelines for quality assurance in mammography screening Four th edition                                 131
H2   20-09-2005   20:54    Pagina 132
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   objects. The values quoted form a smooth curve and may be interpolated for other detail
                   diameters. It is expected that a new design of test object will be developed that will simplify the
                   testing against these standards on a routine basis.
                   On the EUREF website (www.euref.org) CDMAM images and scores are available for reference
                   purposes.
                   Remark: For most systems exposure time increases rapidly with breast thickness and content.
                           Depending on the screen-film combination and the clinically used spectra this range
                           may vary from 0.2 to 3 seconds. For some scanning slot systems however, scanning
                           time and exposure time are fairly constant for the whole range of breast thickness and
                           content. Due to this design, these systems may not comply with the limiting value of 2
                           seconds at standard thickness. Ideally exposure time should be below a certain limiting
                           value even for very thick and dense breasts, so the limiting value at standard thickness
                           may not be the right measure to prevent motion unsharpness for all breasts. Because
                           this worst case liming value has not been determined yet, the value of 2 seconds at
                           standard thickness is maintained, with the exception that scanning slot systems for
                           which exposure time is only slightly dependent on breast thickness and content do not
                           have to comply. For these systems clinical results will have to show that motion
                           unsharpness is not a problem.
                   Limiting value                          Exposure time: acceptable: < 2 s19; achievable: <1.5 s; scanning
                                                           time: values at acceptance are used as reference, typical value:
                                                           5 - 8 s.
                   Frequency                               Yearly.
                   Equipment                               Exposure time meter, standard test block.
                   132                    European guidelines for quality assurance in mammography screening Four th edition
H2   20-09-2005   20:54    Pagina 133
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y    b
                   For the different digital systems, different types of artefacts can occur. Inspect all test images for
                   artefacts.
                   In manual mode an image of the standard test block is made using clinical settings. The block is
                   positioned such that half of the detector is covered and half of the detector is not covered. For
                   the second image (at clinical settings) the standard test block covers the whole detector and the
                   aluminium object is placed exactly centred on top of the standard block (see figure 2.2). The time
                   between both images should be approximately one minute.
                   For CR systems: No post processing should be applied, the screen processing should be turned
                   off as much as possible (see appendix 7).
                   Measure the mean pixel value (PV) in the ROI (area: 4 cm2) on the locations shown in the figure
                   above (on the second image) and calculate the ‘ghost image’-factor.
                    If the system fails to meet the limiting value, check the homogeneity of the image. If the Heel
                   effect is large regions 1 to 3 should be chosen on a line parallel to chest wall side.
                       If the ghost image test is performed last, it is advised to make a number of images of a
                       homogeneous block PMMA covering the whole detector afterwards to get rid of possible
                       ghosts.
                   European guidelines for quality assurance in mammography screening Four th edition                         133
H2   20-09-2005   20:54    Pagina 134
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   2b.4.1 Monitors
                   2b.4.1.1 Ambient light
                   Most of the quality tests in this chapter are highly sensitive to ambient light, therefore all of them
                   should be performed under clinical conditions (room lights, light boxes and other display devices
                   should be at the same luminance level as under clinical conditions). The ambient light should be
                   measured at the centre of the display with the light detector facing outwards and the display
                   switched off.
                   Limiting value                                     Ambient light should be less than 10 lux for primary display
                                                                      devices. [The maximum ambient light actually depends on the
                                                                      reflection characteristics and minimum luminance of the monitor,
                                                                      but for reasons of simplicity this is ignored here.]
                   Frequency                                          Every six months. (Every time the system is used, it has to be
                                                                      made sure that ambient light conditions have not changed.)
                   Equipment                                          Illuminance meter.
                   134              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 135
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
                   Remark: It should be kept in mind that the luminance of LCD monitors depends on the viewing
                           angle. When large viewing angles are used, contrast visibility may not comply with the
                           limiting values.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   135
H2   20-09-2005   20:54    Pagina 136
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   Limiting value                                     All corner patches should be visible, the 5% and 95% pixel value
                                                                      squares should be clearly visible.
                   Frequency                                          Daily.
                   Equipment                                          TG18-QC test pattern.
                   2b.4.1.4 Resolution
                   Evaluate horizontal and vertical line patterns to check display resolution visually.
                   AAPM Task Group 18 provides 6 line patterns at different background luminance levels.
                   (Horizontal line patterns TG18-LPH10, -LPH50 and -LPH89; Vertical line patterns TG18-LPV10,
                   -LPV50 and -LPV89.)
                   136              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 137
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
                   Remark: It should be kept in mind that the luminance of LCD monitors depends on the viewing
                           angle. When large viewing angles are used, the luminance range may not comply with
                           the limiting values.
                   The greyscale display function (GDF) can be determined by measuring the luminance of the 18
                   AAPM luminance test patterns (TG18-LN12-01 through TG18-LN12-18). The test patterns should
                   be displayed full screen and the luminance has to be measured at the centre of the screen. The
                   shape of the GDF depends on the ambient light in the room. Therefore room lights, light boxes
                   and other display devices should be at the same luminance level as when the system is used
                   clinically. A telescopic luminance meter should be used to include the influence of ambient light.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   137
H2   20-09-2005   20:54    Pagina 138
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   The measured values can be inserted into a spreadsheet (available on the EUREF website:
                   www.euref.org) to automatically determine GSDF conformance.
                   After doing this measurement, the amount of ambient light may not be increased anymore,
                   otherwise the contrast response has to be measured again!
                   Remark: This test only applies to primary and secondary display systems. The acquisition
                           workstation monitor is excluded from this test. Due to the required ambient light levels
                           in the mammography room the acquisition workstation monitor will not comply with the
                           limiting values of primary and secondary displays. Therefore this monitor should only be
                           used to check positioning techniques, not for diagnosis and image quality checks.
                       It is acknowledged that some displaying systems do not comply with the DICOM Greyscale
                       Standard Function. Manufacturers are urged to comply with this standard.
                   Remark: It should be kept in mind that the luminance of LCD monitors depends on the viewing
                           angle. When large viewing angles are used, the display on a monitor may not comply with
                           the GSDF.
                   Limiting value                                     The calculated contrast response should fall within ± 10% of the
                                                                      GSDF contrast response for primary class displays (± 20% for
                                                                      secondary class displays).
                   Frequency                                          Every six months and when contrast visibility has changed.
                   Equipment                                          Telescopic luminance meter, TG18-LN12-01 through TG18-LN12-
                                                                      18 test patterns.
                   138              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 139
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
                   2b.4.2 Printers
                   2b.4.2.1 Geometrical distortion
                   Print the TG18-QC test pattern (fig. 4.1) and check visually if the image is printed without
                   geometrical distortion. Only the lines and borders of the test pattern are used to do this.
                   Limiting value                                                All corner patches should be visible, the 5% and 95% pixel value
                                                                                 squares should be clearly visible.
                   Frequency                                                     Daily.
                   Equipment                                                     TG18-QC test pattern.
                   2b.4.2.3 Resolution
                                                                                                                  Evaluate horizontal and vertical line patterns to
                                                                                                                  check the resolution of a print-out.
                                                                                                                  The fine detail horizontal and vertical line
                                                                                                                  patterns in the TG18-PQC test pattern (fig 4.5)
                                                                                                                  can be used.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   139
H2   20-09-2005   20:54    Pagina 140
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   Limiting value                                     Dmin < 0.25 OD, Dmax > 3.40 OD21 (provisional).
                   Frequency                                          Every six months.
                   Equipment                                          Densitometer, TG18-QC test pattern.
                   The greyscale display function (GDF) can be determined by printing the TG18-PQC test pattern
                   and measuring the optical density of marked regions of the 18 bars. The GDF is determined by
                   the luminance corresponding with the optical density. The relationship between the luminance (L)
                   and the optical density (D) of the printed bars is:
                   L = La + L0 * 10-D
                   where: La is the luminance contribution due to ambient illuminance reflected off the film, and
                           L0 is the luminance of the light box with no film present
                   Printed mammograms may be viewed on different viewing boxes and under a variety of viewing
                   conditions. It is not desirable to repeat this measurement for each viewing box. Assuming each
                   viewing box, on which printed mammograms will be diagnosed, complies with the limiting values,
                   a standard viewing box is defined. For this standard viewing box La is 1 cd/m2 and L0 is 4000
                   cd/m2.
                   The measured values can be inserted into an spreadsheet (available on the EUREF website:
                   www.euref.org) to automatically determine GSDF conformance.
                   Limiting value                                     The calculated contrast response should fall within ± 10% of the
                                                                      GSDF contrast response.
                   Frequency                                          Every six months and when contrast visibility has changed.
                   Equipment                                          Densitometer, TG18-PQC test pattern.
                   Limiting value                                     Maximum optical density deviation should be less than 10%
                                                                      ((Dmax-Dmin)/Dcentre < 0.1).
                   Frequency                                          Every six months and when contrast visibility has changed.
                   Equipment                                          Densitometer, TG18-UNL10 and TG18-UNL80 test patterns.
                   140              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 141
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   141
H2   20-09-2005   20:54    Pagina 142
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   2b.6.2 Bibliography
                   Protocols
                   1. European Guidelines for Quality Assurance in Mammography Screening, third edition,
                       European Communities, 2001.
                   2. Quality Control Procedures for Full-field Digital Mammography, 2002: ACRIN # 6652, Digital
                       Mammography Imaging Screening Trial, rev. 2.06, April 2002.
                   3. Meetprotocol Digitale mammografie: Acceptatietest van Screeningseenheden voor
                       Bevolkingsonderzoek op Borstkanker, versie 2002, National Expert and Training Centre for
                       Breast Cancer Screening, University Medical Centre Nijmegen.
                   4. Recommandations pour un programme d’assurance de qualité en mammographie
                       numérique. A. Noel, J. Stinès (red.). J. Radiol 2003; 84: 723-9.
                   5. European protocol on dosimetry in mammography, European Communities, 1996.
                   6. IPSM: Commissioning and routine testing of mammography X-ray systems – second edition,
                       The Institute of Physical Sciences in Medicine, York, 1994 (report no. 59/2).
                   7. Guidelines on quality assurance visits, NHSBSP, Second edition October 2000 (NHSBSP
                       Publication No 40).
                   8. Addendum on digital mammography to chapter 3 of the: European Guidelines for Quality
                       Assurance in Mammography Screening, version 1.0, November 2003.
                   9. Assessment of display performance for medical imaging systems, pre-final draft (version
                       8.1), AAPM, Task Group 18, E. Samei (chairman) et al.
                       2002: AAPM TG 18.
                   11. Performance of mammographic equipment in the UK breast cancer screening programme in
                       1998/99, 2000 (NHSBSP report no 45).
                   12. Review of radiation risk in breast screening, 2003 (NHSBSP report no 54).
                   13. Results of technical quality control in the Dutch breast cancer screening programme (2001-
                       2002), S.W. Beckers et al., Nijmegen 2003.
                   14. Patientdoser från röntgenundersökningar i Sverige, W. Leitz et al. Statens strålskyddsinstitut
                       2001.
                   15. Mammografiscreening, teknisk kvalitetskontroll – resultater og evaluering etter fire års
                       prøveprosjekt, K. Pedersen et al., Statens strålevern 2000.
                   Publications
                   1. E. Samei, J.A. Seibert, C.E. Willis, M.J. Flynn, E. Mah, K.L. Junck: Performance evaluation of
                       computed radiography systems.
                       2001: Med. Phys 28(3): 361-371.
                   2. C. Kimme-Smith, C. Lewis, M. Beifuss, L.W.Bassett: Establishing minimum performance
                       standards, calibration intervals and optimal exposure values for a whole breast digital
                       mammography unit.
                       1998: Med. Phys. 25 (12): December .
                   3. H. Fujita, D.Tsai, T. Itoh, K. Doi, J. Morishita, K. Ueda, A. Ohtsuka: A simple method for
                       determining the modulation transfer function in digital radiography.
                       1992: IEEE Transactions on Medical Imaging 11(1): 34-39.
                   4. E. Samei, M.J. Flynn, D.A. Reimann: A method for measuring the presampled MTF of digital
                       radiographic systems using an edge test device.
                       1998: Med. Phys. 25(1): 102-113.
                   5. A. Noël, J. Stines: Contrôle de qualité: du conventionnel au numérique.
                       2002: J. Le Sein 12(1-2): 17-21.
                   142              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 143
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                    b
                   6.       K.G. Lisk: SMPTE test pattern for certification of medical diagnostic display devices
                            1984: Proc. of SPIE 486: 79-82.
                   7.       W.Huda, A.M. Sajewicz, K.M. Ogden, E.M. Scalzetti, D.R. Dance: How good is the ACR
                            accreditation phantom for assessing image quality in digital mammography.
                            2002: Acad. Radiol. 9: 764-772.
                   8.       J.T. Dobbins: Effects of undersampling on the proper interpretation of modulation transfer
                            function, noise power spectra and noise equivalent quanta of digital imaging systems
                            Med. Phys. 22: 171-181.
                   9.       J.P. Moy, B. Bosset: How does real offset and gain corrections affect the DQE in images from
                            X-ray Flat detectors.
                            1999: proc. of SPIE 3659.
                   10.      C.K. Ly: Softcopy Display Quality Assurance Program at Texas Children’s Hospital.
                            2002: Journal Of Digital Imaging, online publication: March.
                   11.      D.R. Dance et al., Monte Carlo calculation of conversion factors for the estimation of mean
                            glandular breast dose, Phys.Med.Biol., 1990, Vol. 35, 1211-1219.
                   12.      D.R. Dance et al., Additional factors for the estimation of mean glandular dose using the UK
                            mammography protocol, Phys.Med.Biol., 2000, Vol. 45, 3225-3240.
                   13.      Z.F. Lu et al., Monthly monitoring program on DryviewTM laser imager: One year experience
                            on five Imation units, Med. Phys. 26 (9), 1817-1821.
                   14.      H. Jung et al., Assessment of flat panel LCD primary class display performance based on
                            AAPM TG18 acceptance protocol, Med. Phys. 31 (7), 2155-2164.
                   15.      J. H. Siewerdsen, D. A. Jaffray, A ghost story: Spatio-temporal response characteristics of
                            an indirect-detection flat-panel imager Med. Phys. 26 (8), 1624-1641.
                   16.      Development of minimum standards for image quality and dose in digital mammography,
                            K.C. Young, B. Johnson, H. Bosmans, R.E. van Engen (to be published in IWDM 2004
                            proceedings).
                   17.      M. Thijssen, W. Veldkamp, R. van Engen, M. Swinkels, N. Karssemeijer, J. Hendriks,
                            Comparison of the detectability of small details in a film-screen and a digital mammography
                            system by the imageing of a new CDMAM phantom, in M.J. Yaffe (ed). Digital mammography
                            IWDM 2000, Medical Physics Publishing 2001, 666-672.
                   18.      AK. Carton, Development and application of methods for the assessment of image quality
                            and detector performance in digital mammography. PhD thesis at the KU Leuven, July 2004.
                   Other reports
                   1. Samei E, Badano A, Chakraborty D, Compton K, Cornelius C, Corrigan K, Flynn MJ,
                       Hemminger B, Hangiandreou N, Johnson J, Moxley M, Pavlicek W, Roehrig H, Rutz L, Shepard
                       J, Uzenoff R, Wang J, and Willis C. Assessment of Display Performance for Medical Imaging
                       Systems. Draft Report of the American Association of Physicists in Medicine (AAPM) Task
                       Group 18, Version 10.0, August 2004.
                       2004: AAPM TG 18
                   2. Mammography – recent technical developments and their clinical potential, B. Hemdal et al.
                       2002: Statens strålskyddinstitut, Swedish Radiation Protection Authority.
                   3. Quality Assurance, meeting the challenge in the Digital Medical Enterprise, B.I. Reiner, E.L.
                       Siegel, J.A. Carrino (ed.).
                       2002: Society for Computer Applications in Radiology.
                   4. Qualitätssicherung an Bildwiedergabegeräten, ed. D. Richter.
                       2002: ZVEI-Fachverband Elektromedizinische Technik.
                   5. DIN 6868-13:2003-02 Sicherung der Bildqualität in röntgendiagnostischen Betrieben – Teil
                       13: Konstanzprüfungbei Projektionsradiographie mit digitalen Bildempfänger-Systemen.
                   6. DIN V 6868-58:2001-01 Sicherung der Bildqualität in röntgendiagnostischen Betrieben
                       – Teil 58: Abnahmeprüfung an medizinischen Röntgeneinrichtungen der Projektions-
                       radiographie mit digitalen Bildempfängersystemen.
                   7. International Electrotechnical Commission (IEC), Geneva, Switzerland: Evaluation and
                       routine testing in medical imaging departments, part 3-2: Acceptance tests – Imaging
                       performance of mammographic X-ray equipment.
                       2004: IEC 61223-3-2 Ed. 2.
                   8. Handbook of Medical Imaging vol 1, J. Beutel, H.L. Kundel, R.L. Van Metter
                       2000: SPIE Press.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   143
H2   20-09-2005   20:54    Pagina 144
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
                   9.  Breast dose surveys in the NHSBSP, software and instruction manual, K.C. Young., NHS
                       Cancer Screening Programmes, 2001 (NHSBSP report no 01/10).
                   10. IPEM report 78 Catalogue of diagnostic X-ray spectra & other data, K. Cranley et al., 1997.
                   Internet
                   1. EUREF website: www.euref.org
                   2. AAPM Task group 18 (test patterns monitor QC): http://deckard.mc.duke.edu/~samei/tg18
                   3. DICOM standard: http://medical.nema.org
                   144              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 145
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                      b
2b.2.1.3 AEC-system
O: optional test, X: required test => This table is continued on the next page
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n     145
H2   20-09-2005   20:54    Pagina 146
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
2b.2.1.4 Compression X X
                       2b.2.2.3.3 Uncorrected                                               X                                                                              X
                                  defective DELs (DR)
2b.2.3 Dosimetry X X
                   146               E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 147
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                        b
2b.4.1 Monitors
                         2b.4.1.2                 Geometrical                                         X                                                                                        X
                                                  distortion (CRT)
2b.4.1.4 Resolution X X
2b.4.2 Printers
2b.4.2.3 Resolution X
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n     147
H2   20-09-2005   20:54    Pagina 148
                   EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
         X-ray source
            See European Guidelines, part A, table 4.1.
         tube voltage
            See European Guidelines, part A, table 4.1.
         AEC
           - exposure contol steps                                                               5 - 15%                                                                                             mGy or mAs
           - back-up timer and security cut-off                                                      -                 function properly
           - short-term reproducibility                                                              -                      < ± 5%                                      < ± 2%                                 mGy
           - long-term reproducibility
             variation in SNR                                                                          -                       < ± 10%                                                                         mGy
             variation in dose                                                                         -                       < ± 10%                                                                         mGy
           - object thickness and tube voltage compensation
             CNR per PMMA thickness
                                    2.0 cm                                                             -                        > 115%
                                    3.0 cm                                                             -                        > 110%
                                    4.0 cm                                                             -                        > 105%
                                    4.5 cm                                                             -                        > 103%
                                    5.0 cm                                                             -                        > 100%
                                    6.0 cm                                                             -                        > 95%
                                    7.0 cm                                                             -                        > 90%
         compression
           See European Guidelines, part A, table 4.1.
         response function
            - linearity                                                                                -                      R 2 > 0.99                                      -                                 -
            - noise evaluation                                                                         -                           -                                          -                                 -
                   148              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005    20:54          Pagina 149
                     EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                     D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y                                                                        b
            uncorrected dels
               - number of uncorrected defective dels                                                                      -             not yet established not yet established                           -
               - position of uncorrected defective dels                                                                    -             not yet established not yet established                           -
scanning time 5 to 8 s
                     E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n       149
H2   20-09-2005       20:54      Pagina 150
                        EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                        D    I     G    I     T    A      L         M      A     M     M     O     G     R    A    P     H     Y
         2b.4.1 monitors
            - ambient light                                                                                   -                  < 10                                                -                                lux
            - geometrical distortion                                                                          -              straight lines
            - contrast visibility                                                                             -         corner patches visible                                       -
                                                                                                                            squares visible                                          -
               - resolution                                                                                   -        line pattern discernible                                      -
               - display artefacts                                                                            -        no disturbing artefacts                                       -
               - luminance range
                 * ratio maximum/minimum luminance                                                            -                     250                                              -
                 * difference in luminance left and right monitor                                             -                      5%                                              -                           Cd/m2
               - DICOM greyscale standard display function                                                    -                 ± 10% of GSDF                                        -
               - luminance uniformity
                 * deviation in luminance (CRT display)                                                       -                           30%                                        -                           Cd/m2
         2b.4.2 printers
            - geometrical distortion                                                                          -               straight lines                                         -
            - contrast visibility                                                                             -          corner patches visible                                      -
                                                                                                                             squares visible
               - resolution                                                                                   -         line pattern discernible                                     -
               - printer artefacts                                                                            -         no disturbing artefacts                                      -
               - optical density range (optional)                                                             -        Dmin < 0.251, Dmax > 3.41                                     -                                OD
               - DICOM greyscale standard display function                                                    -             ± 10% of GSDF                                            -
               - density uniformity
                 * deviation in optical density                                                               -                         < 10%                                        -                                OD
         1
             Provisional limiting values
                        150                E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 151
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   European protocol for the quality control of the physical and technical aspects
                   of mammography screening
                   Appendices
                   Appendix 1: Mechanical and electrical safety checks
                   Introduction
                   Basic mechanical and electrical safety tests should be performed according to local regulations.
                   If such regulations do not exist this appendix gives an example of such tests based on the UK
                   protocol.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   151
H2   20-09-2005   20:54    Pagina 152
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   Radiation safety
                   The following checks relate to the safe operation of the x-ray unit:
                   • A mains isolator, accessible from the normal operating position should be provided.
                   • A visible indication must be provided on the control panel to show that the mains are switched
                     on.
                   • The visible exposure warning indication must function correctly.
                   • The total filtration must be equivalent to at least 0.5 mm Al or 0.03 mm Mo.
                   • If the added filtration is removable or interchangeable, an interlock must be provided to
                     prevent exposure if the filter is removed or incorrectly inserted.
                   • If the field-limiting diaphragm can be removed, an interlock should be provided to prevent
                     exposure unless the diaphragm is properly aligned.
                   • The exposure must terminate if the exposure control is released prematurely.
                   • The location of the exposure control should confine the operator to the protected area during
                     exposure.
                   • The exposure control should be designed to prevent inadvertent production of x-rays.
                   • The design of the exposure control should prevent further exposure unless pressure on the
                     control is first released.
                   X-ray room
                   • Room warning lights should be provided at all entrances to the x-ray room. These should
                     indicate when x-rays are being or are about to be generated.
                   • A check on the room shielding, either visually, against the local requirements at the planning
                     stage, or by transmission measurements, should be undertaken at or prior to installation.
                   152              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 153
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   Appendix 2: Film-parameters
                   The film curve can be characterised by a few parameters. Most important items are contrast,
                   sensitivity and base and fog. There are different methods to calculate the film parameters.
                   Existing normalisation’s differ so much that the following method is suggested, derived from the
                   Dutch protocol (1991), which is based on the ANSI (1983) norm.
                   Very high contrast can be a problem because of an associated reduction in dynamic range which
                   may result in dense breast tissue being imaged in relatively low film densities where the film
                   performance is relatively poor. To some extent this can be compensated for by setting relatively
                   high average film densities, but even then a lower film contrast may better image local areas of
                   dense tissue. Conversely a very low overall film contrast may indicate an inadequately processed
                   film and subtle details may be missed by the radiologist.
                   Research has shown that film gradient measured by light sensitometry correlates well with film
                   gradient measured by x-ray sensitometry using a fixed kV and target filter combination. One must
                   bear in mind that film emulsions may respond slightly differently to the light from a sensitometer
                   as opposed to the light from the screen used for imaging.
                   Dmin                                                          Base and fog; the optical density of a non exposed film after
                                                                                 developing. The minimum optical density can be visualised by
                                                                                 fixation only of an unexposed film. The extra fog is a result of
                                                                                 developing the (unexposed) emulsion.
                   Dmax                                                          The maximum density achievable with an exposed film; i.e. the
                                                                                 highest density step.
                   Speed                                                         Sensitivity; the property of the film emulsion directly related to the
                                                                                 dose. The Speed is calculated as the x-axis cut-off at optical
                                                                                 density 1.00+Dmin, also called ‘Speedpoint’. The higher the figure
                                                                                 for Speed, the more dose is needed to obtain the right optical
                                                                                 density. Since the film curve is constructed from a limited number
                                                                                 of points, the Speed must be interpolated. Linear interpolation
                                                                                 will result in sufficient accuracy.
                                                                                 Since these parameters are derived from the characteristic curve by
                                                                                 interpolation they are not very practical if a computer is not available.
                                                                                 A simpler procedure is to use the parameters below which are based
                                                                                 on density measurements of particular sensitometric steps.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   153
H2   20-09-2005   20:54    Pagina 154
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   Speed Index                                        The density of the step near to the speedpoint density 1.0 OD,
                                                                      base and fog excluded. Usually this is the density of step 11 of
                                                                      the sensitometric stepwedge.
                   Contrast Index 1                                   The difference in density found between the step nearest to the
                                                                      speedpoint density (1.0 OD, base and fog excluded) and the one
                                                                      with a 0.6 log E (factor 4) higher light exposure (normally
                                                                      4 density steps) (ACR).
                   Contrast Index 2                                   The difference in density steps found between the step nearest
                                                                      to the speedpoint and the step nearest to a density at 2.0 OD,
                                                                      base and fog excluded (IPSM, see bibliography).
                   154              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 155
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   When the optical density of several images, taken under identical conditions, are measured,
                   there will be a range of optical densities. This can either be the result of a change in exposure or
                   a change in developing conditions. By calculating the relative figure log (I’) we are able to
                   distinguish between processor faults and tube malfunctions.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   155
H2   20-09-2005   20:54    Pagina 156
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
Spectrum
2 25, 26 kV
3 25-27 kV
4 26-28 kV 26, 27 kV
5 27-29 kV 26, 27 kV
                   156              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 157
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
D = Kgcs (A5.1)
                   where K is the entrance surface air kerma (without backscatter) calculated at the upper surface
                   of the PMMA. The factor g, corresponds to a glandularity of 50%, and is derived from the values
                   calculated by Dance et al 2000 and is shown in table A5.1 for a range of HVL. The c-factor
                   corrects for the difference in composition of typical breasts from 50% glandularity [Dance et al
                   2000] and is given here for typical breasts in the age range 50 to 64 in table A5.2. Note that the
                   c and g-factors applied are those for the corresponding thickness of typical breast rather than the
                   thickness of PMMA block used. Where necessary interpolation may be made for different values
                   of HVL. Typical values of HVL for various spectra are given in table A5.3. The factor s shown in
                   table A5.4 corrects for differences due to the choice of X-ray spectrum (Dance et al 2000).
                   The dose should be determined using the usual clinically selected exposure factors including any
                   automatic selection of kV and target/filter combination.
D = Kgcs (A5.2)
                   where K is the entrance surface air kerma calculated (in the absence of scatter) at the upper
                   surface of the breast. The factor g, corresponds to a glandularity of 50%, and is shown in table
                   A5.5 (Dance et al 2000). The factor c corrects for any difference in breast composition from 50%
                   glandularity. C-factors for typical breast compositions in the age range 50 to 64 and 40 to 49 are
                   shown in tables A5.6 and A5.7. The factor s corrects for differences due to the choice of X-ray
                   spectrum as noted earlier. Measurement of compressed breast thickness for this purpose is
                   performed by the radiographer, by reading the displayed compressed thickness on the X-ray set.
                   The accuracy of the displayed thickness should be verified by applying a typical force (e.g. 100
                   N) to rigid material of known thickness. It may be necessary to apply correction factors if the
                   displayed values are in error. An accuracy of better than ± 2 mm is required. Software for making
                   such dose calculations has been published by the UK Breast Screening Programme (Young,
                   2001).
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n     157
H2   20-09-2005   20:54    Pagina 158
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   158              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 159
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   Table A5.3: Typical HVL measurements for different tube voltage and target filter
                               combinations. (Data includes the effect on measured HVL of attenuation by a
                               PMMA compression plate*.)
25 0.33 ± .02 0.40 ± .02 0.38 ± .02 0.52 ± .03 0.31 ± .03
28 0.36 ± .02 0.42 ± .02 0.43 ± .02 0.54 ± .03 0.37 ± .03
31 0.39 ± .02 0.44 ± .02 0.48 ± .02 0.56 ± .03 0.42 ± .03
                   * Some compression paddles are made of Lexan, the HVL values with this type of compression
                     plate are 0.01 mm Al lower compared with the values in the table.
Table A5.4: s-factors for clinically used spectra [Dance et al. 2000]
Spectrum s-factor
Mo/Mo 1.000
Mo/Rh 1.017
Rh/Rh 1.061
Rh/Al 1.044
W/Rh 1.042
W/Al 1.05*
                   * This value is not given in the paper of Dance et al. The value in the table has been estimated
                     using the S-values of other spectra.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n       159
H2   20-09-2005   20:54    Pagina 160
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   Table A5.5: g-factors (mGy/mGy) for breast thicknesses of 2-11 cm and the HVL range 0.30-
                               0.60 mm Al. The g-factors for breast thicknesses of 2-8 cm are taken from Dance
                               (1990), and for 9-11 cm from Dance et al. (2000)
                   160              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 161
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
Table A5.6: c-factors for average breasts for women in age group 50 to 64 (Dance et al. 2000)
                   Breast                                                                                             c-factors
                   Thickness
                   (cm)                                                                                            HVL (mm Al)
Table A5.7: c-factors for average breasts for women in age group 40 to 49 (Dance et al. 2000)
                   Breast                                                                                             c-factors
                   Thickness
                   (cm)                                                                                            HVL (mm Al)
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n      161
H2   20-09-2005   20:54    Pagina 162
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   Table A6.1: Calculated radiation contrast for various gold thickness on the standard test
                               object
                   162              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 163
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   Remark: For all measurements on the Fuji system an L-value of 2 is advised (which resembles
                           the L-value in clinical practice). If clipping occurs with an S-value of 120, another S-value
                           should be chosen.
Semi EDR screen processing Plot sensitivity index versus inverse entrance surface air kerma
                   Kodak systems:
                   Pattern screen processing                                     Plot the mean pixel value in the reference ROI versus log
                                                                                 entrance surface air kerma
                   Agfa systems:
                   System diagnostics/flat                                       Plot the mean pixel value in the reference ROI versus log
                   field screen processing                                       entrance surface air kerma
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   163
H2   20-09-2005   20:54    Pagina 164
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   Fuji systems                                       Use FIXED EDR screen processing. The S and L value must be
                                                                      chosen such that they are typical for the clinical situation. These
                                                                      values may differ from site to site. Typical values (according to
                                                                      Fuji): S = 40 to 100, L = 1.8 to 2.6.
                   Fuji systems                                       Use FIXED EDR screen processing. The S and L value must be
                                                                      chosen such that they are typical for the clinical situation. These
                                                                      values may differ from site to site. Typical values (according to
                                                                      Fuji): S = 40 to 100, L = 1.8 to 2.6.
                   164              E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n
H2   20-09-2005   20:54         Pagina 165
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
                   Notes
                   1.       This is the PMMA thickness most commonly used, but others may be specified in parts of this
                            protocol.
                   2.       The specifications of the listed equipment are given, where appropriate, in section 4, table 2.
                   3.       The standard test block may be composed of several PMMA plates.
                   4.       PMMA (polymethylmethacrylate) is commercially available under several brandnames, e.g.
                            Lucite, Plexiglas and perspex.
                   5.       150 X 100 mm or semi-circular with a radius of ≥µ100 mm, and covering a total thickness
                            range from 20 to 70 mm PMMA.
                   6.       In future the PMMA thickness may change to the ‘standard thickness’ of 45 mm with the
                            details positioned at a height of 40 to 45 mm above the breast support table. This may mean
                            that the limiting values need slight adjustment.
                   7.       If the exposure-to-read-time other than one minute is more relevant for practical reasons, that
                            other time should be chosen.
                   8.       The specifications of the listed equipment are given, where appropriate, in chapter 3.5, table
                            2 of the European Guidelines, third edition.
                   9.       The standard test block, covering the whole imaging area, may be composed of several PMMA
                            plates.
                   10.      PMMA (polymethylmethacrylate) is commercially available under several brand names, e.g.
                            Lucite, Plexiglas and Perspex.
                   11.      Covering the whole imaging area, and covering a total thickness range from 20 to 70 mm
                            PMMA (Normally PMMA of 180 X 240mm is available).
                   12.      These films have been reported as suitable for use in collimation assessment by Beideck and
                            Gingold at the AAPM 2004 annual meeting.
                   13.      These values are derived from screen-film mammography. At this moment no limiting values
                            on exposure increase per step for digital mammography have been set, but they should be
                            approximately uniform.
                   14.      In future the contrast threshold visibility may be determined at the standard PMMA thickness
                            of 45 mm, so CNR limits will also be relative to 45 mm in future.
                   15.      These values are provisional, it is advised to check the EUREF website for alterations
                   16.      For some scanning slot systems only a limited range of mA or mAs settings are available, for
                            these systems images should be made at all settings.
                   17.      In future the PMMA thickness may change to the ‘standard thickness’ of 45 mm with the
                            details positioned at a height of 40 to 45 mm above the breast support table. This may mean
                            that the limiting values need slight adjustment.
                   18.      CDMAM phantom with a 4 cm thickness of PMMA, see appendix 6.
                   19.      For some scanning slot systems, see the remark above.
                   20.      Aliasing problems may occur due to the difference in pixel size of the printer and test pattern.
                   21.      Further research is necessary to investigate whether the Dmin and Dmax limiting values are
                            appropriate.
                   22.      Data partly based on: Bengt Hemdal, Lars Herrnsdorf, Ingvar Andersson, Gert Bengtsson, Boel
                            Heddson and Magnus Olsson, Average glandular dose in routine mammography screening with
                            Sectra MicroDose Mammography, MDM, poster at: Medicinska Riksstämman, Göteborg,
                            Sweden 2004.
                   E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n   165
H2   20-09-2005   20:54    Pagina 166
EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING
166 E u r o p e a n g u i d e l i n e s f o r q u a l i t y a s s u r a n c e i n b r e a s t c a n c e r s c r e e n i n g a n d d i a g n o s i s Fo u r t h e d i t i o n