Winct Manual
Winct Manual
40 06/10/2017
                                                                             for Windows 10 / 8.1 / Windows 7 / Windows Vista
                                                                                       Data communication software for windows
                                                                                     Copyright (c) 1998 A&D Company, Limited.
Instruction Manual
==================================================================================
    Contents
-------------------------------------------------------------------------------------------------------------------------------------------
1. Summary
2. License Statement and Limited Warranty
3. System Requirements
4. Interface Preparation
5. RsCom
6. RsKey
7. RsWeight
8. Troubleshooting
9. Examples
==================================================================================
1. Summary
-------------------------------------------------------------------------------------------------------------------------------------------
Windows Communication Tools (WinCT) is a program for transmitting the weighing data from an A&D balance to
a computer. WinCT consists of three applications: RsCom and RsKey and RsWeight.
RsCom:
This window can transmit the weighing data to the computer using a text file format, and can transmit "commands"
to control the balance.
RsKey:
The weighing data is directly pasted into an application such as Excel or Word, but this application cannot use
"commands" to control the balance.
RsWeight:
The weighing data can be retrieved from the balance and displayed in graph form on the monitor screen in real-time.
Maximum, minimum, average, standard deviation and coefficient of variation values of data can be calculated and
displayed.
                                                                    1
Caution:
- WinCT is a program for balances that use the A&D format. A&D does not guarantee performance or compatibility
 when other manufacturers' balances are used with WinCT.
- Back up your important data in the computer just in case.
- If you have problems or questions about compatibility with third party software, contact the manufactures for
 information.
==================================================================================
2. License Agreement and Limited Warranty
-------------------------------------------------------------------------------------------------------------------------------------------
- WinCT is subject to change without notice and does not represent a commitment by A&D to support earlier
 versions or to make changes to support non standard computers.
- WinCT is intended for the transmission of the weighing data from an A&D balance to a PC, and command
 transmission for controlling the balance.
- Under copyright laws, the WinCT program may not be copied, in whole or part, without the consent of A&D,
 except as specified in the license agreement.
- WinCT should only be installed on the hard disk or peripheral storage devices of a computer connected to an
 A&D balance.
- A&D bears no liability for direct, indirect, special, incidental, or consequential damages resulting from any defect
 in the WinCT program or the readme file, even when advised of the possibility of such damage.
- A&D is not responsible or liable for the loss of any programs and/or data stored in the computer on which WinCT
 was installed, including the costs of recovering said lost programs or data.
Microsoft, Windows, Excel and Word are trademarks or registered trademarks of Microsoft Corporation.
July 2, 2014
A&D Company, Limited
                                                                    2
==================================================================================
3. System Requirements
-------------------------------------------------------------------------------------------------------------------------------------------
3.1 A&D Balance (with RS-232C or USB port*1)
-------------------------------------------------------------------------------------------------------------------------------------------
The balance must be equipped with an RS-232C interface.
-------------------------------------------------------------------------------------------------------------------------------------------
3.2 Computer
-------------------------------------------------------------------------------------------------------------------------------------------
OS            Microsoft Windows 10 / 8.1 / Windows 7 / Windows Vista
             Port differs depending on the connection method.
                            Connection method                             Port
                               RS-232C cable                       RS-232C port
                                                                   (COM port)*2
                                 USB cable                              USB port
                                  Bluetooth                             USB port
                              (AD-8529PC-W)
                       *2     If the computer has no COM port, a Serial/USB converter is available as an
                              accessory to add a COM port to the computer. A cable is also attached for
                              connecting the balance.
                       Product, Accessory
                                                                     Balance has 25 pin                 Balance has 9 pin
                                                                            RS-232C                           RS-232C
                       Using Serial/USB converter                 AX-USB-25                          AX-USB-9P
                       Using Bluetooth converter                  AD-8529PC-W                        AD-8529PC-W
                                                                  AX-K05363-30
Caution:
- Use an English version of Windows to assure performance of WinCT.
- If WinCT is operated on Windows except for an English version, the computer may show incorrect characters of
 data which is already stored before installing WinCT.
-------------------------------------------------------------------------------------------------------------------------------------------
3.3 Third Party Software/Applications
-------------------------------------------------------------------------------------------------------------------------------------------
If RsKey is used, an application to write/edit the data is required (for example: Microsoft "Excel").
                                                                    3
-------------------------------------------------------------------------------------------------------------------------------------------
3.4 Cable
-------------------------------------------------------------------------------------------------------------------------------------------
Refer to each instruction manual for a cable to connect the balance and the computer.
These cables are included in the Serial/USB converters AX-USB-25P/9P.
Example 1
   Balance          GH, HR-i, HR-AZ/A, FZ/FX-i, EK-i, or BM, series (D-sub 9 pin)
   Computer          (D-sub 9 pin)
   Cable            Straight type, D-sub 9 pin (pin module), D-sub 9 pin (socket module)
Example 2
   Balance          GR, GX, GX-K, GF, GF-K, GP, MC series (D-sub 25 pin module)
   Computer          (D-sub 9 pin, socket module)
   Cable            Straight type, D-sub 25 pin (pin module), D-sub 9 pin (socket module)
Example 3
   Balance          GX-A(USB) *1
   Computer          (USB port)
   Cable            USB cable
                                                                    4
==================================================================================
4. Interface Preparation
-------------------------------------------------------------------------------------------------------------------------------------------
4.1 Connecting the RS-232C cable between an A&D balance and a computer.
-------------------------------------------------------------------------------------------------------------------------------------------
Step 1    Turn off the computer.
            Connect the cable to appropriate COM port of the computer.
            (e.g. COM1 of the computer)
-------------------------------------------------------------------------------------------------------------------------------------------
4.2 Specifying parameters of an A&D balance
-------------------------------------------------------------------------------------------------------------------------------------------
Refer to the balance's instruction manual for the "Data output mode", "Data format" and RS-232C interface
parameters. The following parameters are the factory settings for A&D's balances.
                                                                    5
==================================================================================
5. RsCom
-------------------------------------------------------------------------------------------------------------------------------------------
5.1 Use and Function of RsCom
-------------------------------------------------------------------------------------------------------------------------------------------
This window is used to transmit the weighing data to a computer, using a text file format.                    This window can also
be used to transmit "commands" to control an A&D balance. Software version 5.40.
-------------------------------------------------------------------------------------------------------------------------------------------
5.2 RsCom Features
-------------------------------------------------------------------------------------------------------------------------------------------
- This window is able to transmit "commands" to control an A&D balance.
- This window communicates data bi-directionally between the balance and the computer.
- Data is transmitted in ASCII text format and is used for display on a monitor.                 It can also be sent to a printer
 without modification.
- Multiple windows can be opened at the same time, depending on the available number of connecting COM ports,
 when multiple balances are connected.
- Other applications can be run at the same time as WinCT.
- This window is able to receive GLP data from an A&D balance.
- Interface parameters are stored in this window.
Caution:
- Use "A&D standard format" for the data communication with an A&D balance.
   Other formats can not be used with RsCom.
- Do not specify the same COM port for multiple windows.
                                                                    6
-------------------------------------------------------------------------------------------------------------------------------------------
5.3 RsCom Operation
-------------------------------------------------------------------------------------------------------------------------------------------
Below is an example of how to operate RsCom.
                                                                    7
Step 3   Starting Communication
          Select (click) the [Start] key in the RsCom window.
          The "o" mark starts blinking on the window and indicates that RsCom is ready to communicate.
Step 4   When the stability mark of a balance is displayed and the [Print] key of a balance is pressed, the current
         data is transmitted to a computer.
Step 5   Check the weighing data at [Received Data] in the RsCom window.
          If the data is not displayed on the computer monitor, refer to section "8. Troubleshooting".
                                                          8
Step 6     Controlling the Balance
            Select a command on [Command Data] in the RsCom window.
            Select (click) the [Command] key. Then the command is sent to the balance.
Caution:
- Data is in text format and can be edited on the monitor.
- Use "A&D standard format" for the data communication with an A&D balance.
  Other formats cannot be used with RsCom.
- Do not specify the same COM port for multiple windows.
                                                             9
-------------------------------------------------------------------------------------------------------------------------------------------
5.4 RsCom Functions
-------------------------------------------------------------------------------------------------------------------------------------------
[Start / Command] key
             [Start] key        Opens the RS-232C port.
             [Command] key          Transmits commands to the balance.
[Clear] key
             Deletes all data displayed in the RsCom window.
[Save] key
             Stores data displayed in the RsCom window to a file.
[Printer] key
             Sends data to the attached printer.
[RS-232C]
             - Specifies the following parameter of RS-232C interface so as to
               adjust to a balance's parameters.
               [Port:], [Baud Rate], [Parity], [Length], [Stop Bit], [Terminator]
             - The "*" is the default parameter settings.
               (The default settings for RsCom communication parameters are the same as the factory settings for
               A&D's balances.)
   [Port:]
             COM port connecting a cable on the computer.
             COM ports recognized by the computer are listed.
             COM ports not on the list have not been recognized correctly by the computer.
   [Baud Rate]
             600, 1200, *2400, 4800, 9600, 14400, 19200, 28800, 38400 bps
   [Parity]
             *E (Even), O (Odd), N (None)
                                                                   10
  [Length]
         Data length
         *7, 8 bits
  [Stop Bit]
         *1, 1.5, 2 bits
  [Terminator]
         *CR/LF, CR
         CR: ASCII code 0Dh
         LF: ASCII code 0Ah
[Manual / Repeat]
         Toggles between manual and automatic operations
         - When the Repeat mark is checked, commands are transmitted periodically.
               Periodic time can be set in seconds. Use values of 1 second or more.
               Caution:
               -Due to internal processing, data cannot be acquired correctly at intervals less than 1 second.
                To acquire data at intervals less than 1 second, we recommend setting the data output setting of the
                balance to Stream Mode and acquiring data as the display is refreshed.
         - When the Repeat mark is not checked, commands are transmitted each time the Command key is
          clicked.
[Data Format]
         Selects items to output with weighing data.
         [Time]
         [Date]
         [Seq. No.]        Sequence number.
         [Command]
         [PU]              Utility software for A&D printers.
         [Beep]             Beep when receiving data.
                           Depending on your computer there may not be sound.
[Received Data]
         Displays current data transmitted from the balance.
                                                            11
[Command Data]
        Selects a command from popup list.
        Refer to the balance manual for command selections.
                                                    12
==================================================================================
6. RsKey
-------------------------------------------------------------------------------------------------------------------------------------------
6.1 Use and Function of RsKey
-------------------------------------------------------------------------------------------------------------------------------------------
The weighing data is directly pasted into an application, such as Excel or Word. (NOTE: this mode can not be
used to send "commands" to the balance) Software version 5.40.
-------------------------------------------------------------------------------------------------------------------------------------------
6.2 RsKey Features
-------------------------------------------------------------------------------------------------------------------------------------------
- The weighing data is directly pasted into an application such as Excel or Word.
- The RsKey mode can paste data transmitted from a balance, as if data is entered by a keyboard.
- The RsKey mode can be used with applications such as Excel, Word and Text Editor, etc.
- This window is able to receive GLP data from a balance.
- Interface parameters are stored in this mode.
Caution
- Turn the Caps Lock off. If the Caps Lock is on, upper-case and lower-case letters of the alphabet will toggle.
                                                                   13
-------------------------------------------------------------------------------------------------------------------------------------------
6.3 RsKey Operation
-------------------------------------------------------------------------------------------------------------------------------------------
This is an example of an RsKey operation.
                                                                   14
Step 3 Test
          Select (click) the [Test] key.
Step 4   When the balance displays the stability indicator, press the balance's [Print] key.
          Check weighing data in the window.
          If the data is not displayed on the computer monitor, refer to section "8. Troubleshooting".
Step 5 Select (click) the [ x ] icon in the window to close the test mode.
Step 7 Place the cursor in the importing application where data is to be entered.
                                                          15
Step 9     When the balance displays the stability indicator and the [Print] key of a balance is pressed, the data is
           entered at the cursor's position.
Step 10 Select (click) the [RsKey Com#] on the task bar to stop communication. The display will return to the
           RsKey window.
Step 11 Select (click) the [End] key in the RsKey window to exit (close) RsKey.
Caution:
- Use "A&D standard format" for data communication with an A&D balance. Other formats cannot be used with
 "RsKey."
- Do not move the cursor during communication while in RsKey mode. If the cursor is moved during
 communication, an error may occur.
- If you have questions about the importing application (Word, Excel, etc.), contact the manufacture.
- You do not need to start up RsKey before starting the importing application (Word, Excel, etc.).
- Perform RsKey minimization after placing the cursor where data is pasted.
- When the mouse or key freezes, turn off the balance, stop receiving data and refresh the [Interval].
- The status of the Caps Lock key and Num Lock key may be changed by the "RsKey".
- When inputting data and plotting a graph in Excel in real time, the software may not operate correctly because
 Excel requires time to process large information.
                                                            16
-------------------------------------------------------------------------------------------------------------------------------------------
6.4 RsKey Functions
-------------------------------------------------------------------------------------------------------------------------------------------
The "*" is default settings.
[Start] key
            - When the [Start] key is selected (clicked), the RsKey window is minimized and is ready to receive data.
               When the data is received, it is entered into an application, such as Excel.
            - To stop the RsKey mode, click the [RsKey Com#] on the task bar and the RsKey window returns to the
              display.
[End] key
            The RsKey mode is closed and exited.
[Test] key
            - Verifies readiness to receive.
               The right condition is that when data is received from a balance it is displayed on the computer.
            - The display format changes according to header type from the balance.
               The "o" (stability) mark is displayed with a stable header (ex. "ST").
               The "o" (stability) mark is not displayed with an unstable header (ex. "US").
            - If the balance is set for "stream mode," clicking this [Test] key will display the data (i.e. on the
              computer monitor).
[RS-232C]
            - Set the computer's RS-232C interface parameter as to adjust to the balance's parameters.
               [Port:], [Baud Rate], [Parity], [Length], [Stop Bit], [Terminator]
            - The "*" is default parameter settings.
               (The default settings for RsKey communication parameters are the same as the factory settings for
               A&D's balances.)
    [Port:]
      COM port to which the cable is connected on the computer.
      COM ports recognized by the computer are listed.
      COM ports not on the list have not been recognized correctly by the computer.
    [Baud Rate]
            600, 1200, *2400, 4800, 9600, 14400, 19200, 28800, 38400 bps
                                                                   17
   [Length]
            Data length
            *7, 8 bits
   [Parity]
            *E (Even), O (Odd), N (None)
   [Stop Bit]
            *1, 1.5, 2 bits
   [Terminator]
            *CR/LF, CR
            CR : ASCII code 0Dh
            LF : ASCII code 0Ah
[Data]
            Selecting the way to paste data with the [Cell] and [Type].
   [Cell]
            Selects an action (operation) after each data is entered.
            *Enter (same operation as the enter key of a computer), Right, Down, Auto (Right or Down)
   [Type]
            Selects the type of input data.
            *All (All data), Number (only numerical data)
[Separator]
            Selects a separator to place in the data, when the [Type] of [Data] is "All".   Set1 and Set2 can set
            arbitrary characters.
            [Comma / Space] "Comma" or "Space" are selected as separator.
            [Tab]
            [Set1]
            [Set2]
[Data Format]
            Selects items with weighing data to display on the monitor window.
            [Time]
            [Date]
            [Seq. No.]              Sequence number.
                                                            18
[Interval]
             Interval is used for periodic data transmission. Set the balance to Stream Mode.
             This setting is used when computer processing is slow or data is periodically required.
             The interval can be set in seconds. The default is 2 seconds. Use values of 2 seconds or more.
             If the interval is set to zero seconds, all data is transmitted from the balance to the application.
             (Please use 0 seconds only in TEST mode)
               Caution:
               -Due to internal processing, data cannot be acquired correctly at intervals less than 2 seconds.
[Beep]
         Beep when receiving data.
         Depending on your computer there may not be sound.
                                                              19
============================================================================
7. RsWeight
-------------------------------------------------------------------------------------------------------------------------------------------
7.1 Use and Function of RsWeight
-------------------------------------------------------------------------------------------------------------------------------------------
This application can produce a graph in real-time of the data transmitted from an A&D balance via RS-232C.
Software version 5.40.
-------------------------------------------------------------------------------------------------------------------------------------------
7.2 RsWeight Features
-------------------------------------------------------------------------------------------------------------------------------------------
- The weighing data transmitted from the balance can be displayed in graph form on the monitor in real-time.
- The axis scale of a graph automatically alters according to the data. It can be also changed manually.
- The graph can be printed out using a printer.
- The recorded data can be saved in CSV format.
- Maximum, minimum, average, standard deviation and coefficient of variation values can be calculated and
 displayed on the monitor.
                                                                   20
-------------------------------------------------------------------------------------------------------------------------------------------
7.3 RsWeight Operation
-------------------------------------------------------------------------------------------------------------------------------------------
Step 1    Start the RsWeight Window
            Run the "RsWeight" menu.
            Default menu address is [Start] - [Program] - [A&D WinCT] - [RsWeight].
                                                                   21
Step 3   Starting Transmission
          Select the [Start] key. The "o" mark starts blinking on the window and indicates that RsWeight is ready to
          start data transmission.
                                                             22
Step 5    Stopping a Measurement
            Select the [Stop] key to terminate the measurement.
Note : Do not change units during a measurement since statistical operation error may occur.
-------------------------------------------------------------------------------------------------------------------------------------------
7.4. RsWeight Functions
-------------------------------------------------------------------------------------------------------------------------------------------
[Table]
   Records transmitted data.
   [Time]     Time of measurement
   [Data] Double-click the table to change the title and value.
[Graph]
- Creates a graph showing transmitted data
- If X Auto, Y Auto are checked, the scales of the axes are automatically changed to suit the scale of the data.
- When the values for X scale (X mini, X max), Y scale (Y mini, Y max) are clicked, the minimum or maximum
 values for each axis can be changed.
- The title of the graph (New.csv) can be changed by the “File Name” option in the menu.
- Remarks can be edited by clicking "Remarks:_____".
   (It is advisable to type a sample name, measurement condition, etc. here.)
- Final value is shown on the right side of the graph.
       [X auto], [Y auto]
       The scales of checked axes (X Auto, Y Auto) are automatically changed
       to match the data(auto scaling function)
       X scale
       [X mini]       The minimum value on the X-axis (time).
       [X max]         The maximum value on the X-axis (time).
       Y scale
       [Y mini]       The minimum value on the Y-axis (weighing value).
       [Y max]         The maximum value on the Y-axis (weighing value).
                                                                   23
[Repeat/Auto Stop]
  When the Repeat mark is checked, commands are transmitted periodically.
  Periodic time can be set in seconds. Use values of 1 second or more.
  Caution:
  -Due to internal processing, data cannot be acquired correctly at intervals less than 1 second.
   To acquire data at intervals less than 1 second, we recommend setting the data output setting of the balance to
   Stream Mode and acquiring data as the display is refreshed.
  When the Repeat mark is not checked, commands are transmitted each time the Command key is clicked.
  When both "Repeat" and "Auto Stop" are checked, the measurement terminates automatically in a set period
  (set in units of minute). (For one set period, data should not exceed 10,000 in total.)
[Start] key
  Starts a measurement(Opens the RS-232C port).
[Stop/End] key
   [End] key          Closes the RsCom Window.
   [Stop] key         Stops the communication data flow.
[Command] key
  Transmits a command.
[Received Data]
  Displays received data on the monitor.
[Command Data]
  Sets a command to transmit.
  For a data query, select "Q."
【Auto Scroll】
  When【Auto Scroll】is checked, the data grid display automatically scrolls to the bottom so the most recent data
  can be viewed at all times.
                                                         24
   [Save]
     [Data Save]
                     Table data is saved in a file. File type can be chosen from CSV or TXT.
                     Choosing the sample material or conditions as the file name is the simplest method.Please
                     note that when the file name is entered here it is automatically populated in the graph title at
                     the top of the graph.
     [Chart Save]
                    The graph can be saved as an image file.
                    File type can be chosen from JPEG or BMP.
                                                          25
[Option] Various settings
   [Line]       Sets the line thickness on the graph.
                 (1: Standard, *2: Heavy line)
   [Marker]     Sets the marker on the graph (*0: No, 1: Yes)
   [Decimal     Point] Sets decimal point. (*Dot, Comma)
   [X-Axis]     Sets time axis unit. (Second, Minute, Hour, *Number)
   [Y-Axis]     Sets the item of Y-Axis.
                 (Header, *Data, Unit)
   [Beep]       Beep when receiving data.
                 Depending on your computer there may not be sound.
    ※When measuring in stream mode on the balance, all data may not be able to be imported, depending on
       the baud rate. Also, when there are many measurement data entries, the graph display may take some
       time to update (about 1 second each time the display is changed).
       When you wish to import all measurement data in stream mode, please use RsCom.
                                                        26
==================================================================================
8. Troubleshooting
-------------------------------------------------------------------------------------------------------------------------------------------
If the setup and procedure is correct, when the balance's [Print] key is pressed, data is displayed on the computer
monitor.     If data is not transmitted from the balance to the computer using RsCom or RsKey or RsWeight, check
the following:
Check 2 Is the cable connected to the right COM port on the computer?
            Sometimes COM port is misunderstood as printer port.
Check 3 Are the RS-232C interface parameters, "Data format" and "Data output mode" for a balance correct?
Check 6 When the balance's [Print] key is pressed, is the stability indicator on the balance displayed?
Check 7 Try a different COM port. (Some computer may not be able to use COM1 if an internal modem is used).
Check 9 If you disconnect the cable during data transmission and transmit data again, an error may be displayed.
Close the software and reinsert the cable. Reload the software and start data transmission again.
                                                                   27
==================================================================================
9. Examples
-------------------------------------------------------------------------------------------------------------------------------------------
The following example assumes the use of an A&D balance, such as the GH , GR, GX, GF, GP, FX-i, EK-i series,
WinCT and Microsoft Excel. Refer to the appropriate manual for operating instructions.
-------------------------------------------------------------------------------------------------------------------------------------------
9.1 Basic Operation
-------------------------------------------------------------------------------------------------------------------------------------------
9.1.1 Collecting data
Data is transmitted using the balance's [Print] key and is entered into "Excel".                   The functions: average, standard
deviation, maximum and minimum can be used in "Excel".
      Balance status             Key mode (factory settings)
      Mode of WinCT              RsKey
      Application                Excel
                                                                   28
-------------------------------------------------------------------------------------------------------------------------------------------
9.2 Using the Balance's Functions
-------------------------------------------------------------------------------------------------------------------------------------------
If the GLP data output function of the balance is used, this GLP data can be sent to a printer connected to a
computer.     It is possible to print weighing data and printing/recording GLP data.
      Balance status             GLP output mode
      Mode of WinCT              RsCom
      Application                Unnecessary
                                                                   29
9.2.2 Using "data memory function"
Some A&D balances, such as the GR Balance, can store multiple weighing data into the balance's memory.
The stored data is then transmitted in a bundle to the computer.
     Balance status          Data memory function
     Mode of WinCT            RsCom,RsWeight
     Application             Unnecessary
If the balance's "auto-print mode" is used, each tablet can be automatically weighed and the data is transmitted to
the computer.
     Balance status          Auto-print mode, Auto-rezero mode
     Mode of WinCT          RsKey
     Application             Excel
When using the "test mode" of RsKey, the stream mode data is displayed on a computer monitor, which works as
an external indicator for the balance.
     Balance status          Stream mode
     Mode of WinCT            RsKey (with interval)
     Application             Unnecessary
                                                         30
9.2.5 Counting parts
When using the balance's "counting mode", the parts are counted.
The counting data is transmitted to a computer and totaled.
     Balance status         Counting mode
     Mode of WinCT         RsKey
     Application            Excel
When using the balance's "animal mode", animals are automatically weighed.
The data is transmitted to a computer.
     Balance status         Animal mode
     Mode of WinCT           RsKey
     Application            Excel
==================================================================================
End of Readme file
==================================================================================
31