Calibration of UV visible spectroscopy
A Control of absorbance
60 mg (+/-3) mg potassium di chromate in 1000 ml 0.005 M H2So4
Measuring configuration in the UV key in parameter in the
spectrum mode as followed: -
Measuring mode absorbance
Recording range0 to 3
Wave length range 400 to 200 nM
Slit width (nm) 2
Scan speed Slow
Ensure the H2So4 is placed in the sample compartment to
obstruct the optical path way. Perform base line correction click
on the base line button, after that click on the auto zero button to
take the spectroscopy.
Place K2Cr2O7 solution in the sample holder and the spectrum
with .005 M. H2So4 acid. Reference measured the absorbance
235, 257, 313 and 350 nm.
Calculate the extraction coefficient E as Below: -
%E (`1%, 1CM) = measured absorbance* 10000/weight of
K2Cr2O7
% E value should be 122.9 – 126.2, 142.4 – 145.7, 47.00 – 50.0,
104.2 – 108.2 respectively wavelength.
B Limit of stray light
1.2% w/v KCL solution
Measurement configuration
Measuring mode abs
Recording range 0–3
Wave length 250 – 190 nm
Slit width 2 nm
Scan speed Slow
Place distil water in both cubtte and do base line place KCL
solution in sample holder and measure the spectrum with water
reference.
Specification
The absorbance value should be more than 2 @ 200 nm.
C Resolution power
.02 % V/V solution of toluene in hexane.
Measurement configuration
Measuring mode abs
Recording range 0 – 1.0
Wave length 275 – 255 nm
Slit width 2 nm
Scan speed Slow
Ensure that nothing is placed in sample compartment to obstruct the
optical path way.
Put hexane in both cubtte and perform base line correction. Place .02
v/v sample of toluene in hexane in the sample holder and measure the
spectrum with hexane reference.
Specification
The ratio of absorbance about max 269 nm – min 266 nm. Should be at
least 1.2