D 1125
D 1125
This standard has been approved for use by agencies of the Department of Defense.
Copyright © ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United States.
                                                                                        1
                                                                                         D 1125
                    TABLE 1 Electrical Conductivity Values Assigned to the Potassium Chloride in the Reference Solution A
                           Approximate                                                                                                      Electrical
     Reference                                                                                              Tempera-
                           Normality of                         Method of Preparation                                                      Conductivity,
      Solution                                                                                               ture, °C
                             Solution                                                                                                        µS/cm
          A                      1                   74.2460 g of KCl weighed in air per 1 L of                 0                            65 176
                                                                  solution at 20°C                             18                            97 838
                                                                                                               25                            111 342
          B                     0.1                   7.4365 g of KCl weighed in air per 1 L of                 0                             7 138
                                                                  solution at 20°C                             18                             11 167
                                                                                                               25                            12 856
          C                     0.01                  0.7440 g of KCl weighed in air per 1 L of                 0                              773.6
                                                                  solution at 20°C                             18                            1 220.5
                                                                                                               25                            1 408.8
          D                    0.001                 Dilute 100 mL of Solution C to 1 L at 20°C                 0                             77.69B
                                                                                                               18                            127.54B
                                                                                                               25                            146.93
  A
   Excluding the conductivity of the water used to prepare the solutions. (See 7.2 and Section 14.) These tabulated conductivity values are in international units. When
using measuring instruments calibrated in absolute units, multiply the tabular values by 0.999505.
  B
   From Glasstone (13).
  K2 5 conductivity of the water used to prepare the reference                                  5.3 If an unshielded cell is used to measure the resistivity/
solution, at the same temperature of measurement, µS/cm,                                     conductivity of high resistivity water there is a possibility of
  Q 5 temperature correction factor (see Section 11),                                        electrical pickup causing erroneous reading. For this reason it
  R 5 resistivity at 25°C, ohm · cm,                                                         is recommended that conductivity cells for this application be
  Rx 5 measured resistance, ohm.                                                             of coaxial shielded type or equivalent, and that the cables and
                                                                                             instrument also be shielded.
4. Significance and Use
  4.1 These test methods are applicable for such purposes as                                 6. Apparatus
impurity detection and, in some cases, the quantitative mea-
                                                                                                6.1 Measuring Circuit—The instrument may be a manually
surement of ionic constituents dissolved in waters. These
                                                                                             operated wheatstone bridge or the equivalent, or a direct
include dissolved electrolytes in natural and treated waters,
                                                                                             reading analog or digital meter. Instruments shall energize the
such as boiler water, boiler feedwater, cooling water, and saline
                                                                                             conductivity cell with alternating current and, together with the
and brackish water.
                                                                                             cell and any extension leadwire, shall be designed to reduce
  4.1.1 Their concentration may range from trace levels in
                                                                                             errors from the following sources:
pure waters (1)5 to significant levels in condensed steam (see
Test Methods D 2186 and D 4519, and Ref (2)), or pure salt                                      6.1.1 In highly conductive solutions—Uncompensated elec-
solutions.                                                                                   trode polarization due to excessive current density at the
  4.1.2 Where the principal interest in the use of conductivity                              electrode surfaces can cause negative conductivity errors.
methods is to determine steam purity, see Ref (3). These test                                Insufficient series capacitance at the electrode/solution inter-
methods may also be used for checking the correctness of                                     face can allow charging effects to distort the a-c measurement
water analyses (4).                                                                          and cause errors if not compensated. Leadwire resistance can
                                                                                             add significantly to the measured resistance.
5. Interferences                                                                                6.1.2 In low conductivity solutions—Excessive parallel ca-
   5.1 Exposure of a sample to the atmosphere may cause                                      pacitance in the cell and extension leadwire can shunt the
changes in conductivity/resistivity, due to loss or gain of                                  measurement and cause positive conductivity errors. Tempera-
dissolved gases. This is extremely important in the case of very                             ture compensation errors can be significant below 5 µS/cm if
pure waters with low concentrations of dissolved ionized                                     variable coefficient algorithms are not employed as described
materials. The carbon dioxide, normally present in the air, can                              in Test Method D 5391.
drastically increase the conductivity of pure waters by approxi-                                6.1.3 These sources of error are minimized by an appropri-
mately 1 µS/cm. Contact with air should be avoided by using                                  ate combination of a-c drive voltage, wave shape, frequency,
flow-through or in-line cell where feasible. Chemically pure                                 phase correction, wave sampling technique and temperature
inert gases, such as nitrogen or helium, may be used to blanket                              compensation designed in by the instrument manufacturer. The
the surface of samples.                                                                      instrument manufacturer’s recommendations shall be followed
   5.2 Undissolved or slowly precipitating materials in the                                  in selecting the proper cell constant, leadwire size, and length
sample can form a coating on the electrodes of the conductivity                              and maintenance of the electrode surface condition for the
cell that may cause erroneous readings. For example, biofoul-                                range of measurement. Calibration may be in either conduc-
ing of the cell or a build-up of filming amines may cause poor                               tivity or resistivity units.
cell response. In most cases these problems can be eliminated                                   6.1.4 When an output signal is required from an on-line
by washing the cells with appropriate solvents.                                              instrument, it shall be electrically isolated from the cell drive
                                                                                             circuit to prevent interaction between a solution ground at the
   5
     The boldface numbers in parentheses refer to the list of references at the end of
                                                                                             cell and an external circuit ground.
these test methods.                                                                             6.2 Cells:
                                                                                         2
                                                                   D 1125
   6.2.1 Flow-through or in-line cells shall be used for mea-          rate to 0.1°C is acceptable for this application, when the
suring conductivities lower than 10 µS/cm (resistivities higher        instrument is not provided with manual or automatic tempera-
than 100 000 ohm · cm), to avoid contamination from the                ture compensation. (See Section 11).
atmosphere. However, samples with conductivity greater than
10 µS/cm may also be measured. In all other cases, pipet-type          7. Reagents
or dip cells can also be used. Pipet or dip cells may be used to          7.1 Purity of Reagents—Reagent grade chemicals shall be
measure samples in the range of 1 to 10 µS/cm if the sample is         used in all tests. Unless otherwise indicated, it is intended that
protected by an inert gaseous layer of nitrogen or helium.             all reagents shall conform to the specifications of the Commit-
   6.2.2 A cell constant shall be chosen which will give a             tee on Analytical Reagents of the American Chemical Society,
moderate cell resistance, matching the instrument manufactur-          where such specifications are available.6 Other grades may be
er’s requirements for the range of measurement. For laboratory         used, provided it is first ascertained that the reagent is of
bridges, Table 2 provides conservative guidelines.                     sufficiently high purity to permit its use without lessening the
   6.2.3 Flow-through and in-line cells shall be mounted so            accuracy of the determination.
that continuous flow of the sample through or past it is                  7.2 Purity of Water—Unless otherwise indicated, references
possible. Flow rate should be maintained at a constant rate            to water shall be understood to mean reagent water conforming
consistent with the manufacturer’s recommendations for the             to Specification D 1193, Type I. In making up the potassium
cell being used, particularly at conductivities below 10 µS/cm.        chloride solutions for cell constant determinations, use water of
The cell shall retain calibration under conditions of pressure,        conductivity not greater than 1.5 µS/cm. If necessary, stabilize
flow, and temperature change, and shall exclude the atmo-              to the laboratory atmosphere by aspirating air through the
sphere and be constructed of corrosion resistant, chemically           water from a fritted glass or stainless steel gas dispersion tube.
inert materials. The chamber or cell shall be equipped with            The equilibrium point is reached when the conductivity re-
means for accurate measurement of the temperature.                     mains constant but not greater than 1.5µ S/cm. The equilibrium
   6.2.4 Platinized cells shall not be used for measurement of         conductivity must be added to Table 1.
conductivities below 10 µS/cm, except that a trace or flash of            7.3 Alcohol—95 % ethyl alcohol. Alternatively, use isopro-
platinum black may be used on cells for measurements in the            pyl alcohol or methyl alcohol.
range of 0.1 to 10 µS/cm (see 9.4). Because of the cost and               7.4 Aqua Regia (3 + 1)—Mix 3 volumes of concentrated
fragility of platinum cells, it is common practice to use              hydrochloric acid (HCl, sp gr 1.19) with 1 volume of concen-
titanium, monel, and graphite electrodes for measurements              trated nitric acid (HNO3, sp gr 1.42). This reagent should be
with accuracies on the order of 1 %. Note that these electrodes        used immediately after its preparation.
may require special surface preparation. Titanium and monel               7.5 Ethyl Ether.
electrodes are especially suitable for high resistance solutions          7.6 Hydrochloric Acid (sp gr 1.19)—Concentrated HCl.
such as ultrapure water, but may introduce a small surface                7.7 Hydrochloric Acid (1 + 1)—Mix 1 volume of concen-
resistance which limits their accuracy when the measured               trated HCl (sp gr 1.19) with 1 volume of water.
resistance is less than a few thousand ohms (1).                          7.8 Platinizing Solution—Dissolve 1.5 g of chloroplatinic
   6.2.5 It is recommended that cells intended for the measure-        acid (H2PtCl6· 6H2O) in 50 mL of water containing 0.0125 g of
ment of conductivities below 10 µS/cm be reserved exclusively          lead acetate (Pb(C2H3O2)2).
for such applications.                                                    7.9 Potassium Chloride (KCl)—The assay of the potassium
   6.3 Temperature Probes:                                             chloride must be 100.06 0.1 %. This standardization grade of
   6.3.1 For Temperature Control—The measurement of tem-               KCl is available from NIST and from commercial sources. Dry
perature is necessary for control of a temperature bath, manual        at 150°C for 2 h or until weight loss is less than 0.02 %; store
temperature compensation, or automatic temperature compen-             in desiccator.
sation, or all of these. Thermometers, thermistors, and resis-            7.10 Potassium Chloride Reference Solution A—Dissolve
tance temperature detectors with accuracies of 60.1°C or               74.2460 g of KCl (weighed in air) in water and dilute to 1 L at
better are acceptable for this application. An ASTM precision          20 6 2°C in a Class A volumetric flask.
thermometer, Number 63C, as defined in Specification E 1, is              7.11 Potassium Chloride Reference Solution B—Dissolve
recommended. The calibration of temperature probes should be           7.4365 g of KCl (weighed in air) in water and dilute to 1 L at
checked periodically by comparison to a reference temperature          20 6 2°C in a Class A volumetric flask.
probe whose calibration is traceable to the U.S. National                 7.12 Potassium Chloride Reference Solution C—Dissolve
Institute of Science and Technology (formerly NBS) or equiva-          0.7440 g of KCl (weighed in air) in water and to dilute 1 L at
lent.                                                                  20 6 2°C in a Class A volumetric flask.
   6.3.2 For Temperature Correction—A thermometer accu-                   7.13 Potassium Chloride Reference Solution D—Dilute 100
                                                                       mL of reference solution C to 1 L with water at 20 6 2°C in
TABLE 2 Recommended Cell Constants for Various Conductivity
                       Ranges
      Range of Conductivity, µS/cm        Cell Constant, cm −1             6
                                                                             Reagent Chemicals, American Chemical Society Specifications, American
         0.05 to 10                           0.01 to 0.1              Chemical Society, Washington, DC. For suggestions on the testing of reagents not
         10 to 200                              0.1 to 1               listed by the American Chemical Society, see Analar Standards for Laboratory
         200 to 5000                            1 to 10                Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia
         5000 to 1 000 000                    10 to 100                and National Formulary, U.S. Pharmaceutical Convention, Inc. (USPC), Rockville,
                                                                       MD.
                                                                   3
                                                                              D 1125
a Class A volumetric flask shortly before using. Store the                        using from 1.5 to 3 coulombs/cm2 of electrode area. For
solution in a glass-stoppered bottle of chemically resistant                      example, for an electrode having a total area (both sides) of 10
glass which has only been used for storage of this solution.                      cm2, the plating time at a current of 20 mA would be from 121⁄2
                                                                                  to 25 min. The current density may be from 1 to 4 mA/cm2 of
   NOTE 3—The electrical conductivity of each of the referenced solutions
is given in Table 1. The values for electrical conductivities for the             electrode area. Plate the electrodes one at a time with the aid of
solutions are those of G. Jones and B. C. Bradshaw (5), confirmed in 1987         an extra electrode. During the plating, agitate the solution
(6) and 1989 (7) by the National Institute of Standards and Technology            gently, or use ultrasonic bath. When not in use, platinized cells
(NIST). The data of T. Shedlovsky (8) are used for Solution D. Solutions          should be filled with water to prevent the drying out of
A, B, and C were prepared by Jones and Bradshaw using the molal or                electrodes while in storage.
demal basis by dissolving 71.1352, 7.4191, and 0.7453 g, respectively, of            9.4 For measurement of conductivities in the range of 0.1 to
KCl (in vacuum) per 1000 g of solution (in vacuum). The method of
preparation given in Table 1 includes the corrections to weights of KCl (in
                                                                                  10µ S/cm, a trace or flash coating of platinum black may be
air against brass weights) per litre of solutions at 20°C and assumes the         used. For a flash coating, the cell is left in the platinic chloride
density of KCl 5 1.98, density of brass 5 8.4, and the density of                 solution for only 2 or 3 s at a current of about 20 mA. A flash
air 5 0.00118. The densities of 1.0 N, 0.10 N, and 0.010 N KCl at 20°C,           coating will leave the electrodes with their metallic appear-
1.04420, 1.00280, and 0.99871 g/mL, respectively, were interpolated from          ance, but with a faint blackish tint.
the data in the International Critical Tables (9).
                                                                                  10. Calibration
8. Sampling                                                                          10.1 Measuring Instrument—A calibrating resistor to be
   8.1 Samples shall be collected in accordance with Practice                     used in place of the conductivity cell may be furnished by the
D 1066, Specification D 1192, and Practices D 3370, as appli-                     manufacturer, together with information as to the correct scale
cable.                                                                            reading the instrument shall assume when this resistor is
   8.2 Avoid exposure of the sample to atmospheres containing                     connected in place of the conductivity cell. Follow the manu-
ammonia or acidic gases. Protect the sample to avoid gain or                      facturer’s instructions and periodically check the instrument.
loss of dissolved gases, particularly if there is some delay                      Alternatively, standard resistors with certified accuracy of
before the conductivity measurements are made. Preferably,                        60.05 % may be used with appropriate calculations adapted to
use a flow-type cell for sampling and measuring condensed                         the instrument scale. Some instruments may be factory cali-
steam or water having a conductivity of less than 10 µS/cm.                       brated, taking into account the resistance of the cable wire
For waters in the range of 5 to 10 µS/cm, a dip-type cell may                     attached to the conductivity cell; this may be indicated by a
be used if a layer of chemically pure nitrogen or helium is                       warning to avoid cutting or extending the cable length. When
maintained over the surface.                                                      lead wires between the instrument and the cell are long, check
                                                                                  the installation at least once by connecting the calibrating
9. Preparation of Electrodes                                                      resistor at the far end of the lead wire and noting the difference,
   9.1 If the cell constant as checked does not fall within                       if any, in reading with the long lead wire in the circuit. Check
reasonable limits of its nominal value, it is necessary to clean                  portable or manually operated instruments in a similar manner
or replatinize the electrodes or replace the cell. In general, no                 with one or several calibrating resistors. Note errors of signifi-
mechanical cleaning should be attempted. In high purity water                     cant magnitude and correct subsequent conductivity readings.
measurements, where the presence of finely divided platinum                       Calibration checks should be made at values as close as
is undesirable due to its long retention of impurities, platini-                  possible to the conductivity values expected in samples. This is
zation of electrodes should be omitted, especially for testing of                 especially important if the measurement is made at the extreme
water having a conductivity below 10 µS/cm (see 9.4). On the                      high or low end of an instrument’s range. Instruments sub-
other hand, clean and well-platinized electrodes are increas-                     jected to field use may require more frequent checks of
ingly important in testing water of higher conductivities,                        calibration. For direct reading instruments, the conductivity
particularly above 1000 µS/cm.                                                    check resistance in ohms equals the cell constant (cm −1)
   9.2 The cell manufacturer’s instructions may be followed                       divided by the conductivity desired (S/cm) while the resistivity
for cleaning the electrodes as well as other parts of the cell. A                 check resistance equals cell constant (cm −1) times the resis-
suitable cleaning solution consists of a mixture of 1 part by                     tivity desired (ohm · cm).
volume of isopropyl alcohol, 1 part of ethyl ether, (with                            10.2 Conductivity Cells—For field and routine laboratory
polymer cells, check compatibility) and 1 part of HCl (1 + 1).                    testing, the calibration of conductivity cells may be checked by
After cleaning, thoroughly flush the cell with water. If the old                  comparing instrument readings taken with the cell in question
platinum black coating is to be removed, judicious application                    against readings on the same sample or series of samples taken
of aqua regia to the electrodes, or electrolysis in HCl (sp gr                    with a conductivity cell of known or certified cell constant.
1.19) is frequently successful.                                                   Exercise care to ensure that both working and reference cells
   9.3 Platinize the electrodes of the cell with H2PtCl6 solu-                    are at the same temperature or, alternatively, at different but
tion. A suitable plating apparatus consists of a 6 volt a-c supply,               known temperatures so that a correction as later described can
a variable resistor, milliammeter, and an electrode. The deposit                  be applied. Resistivity-reading instruments will indicate in
should present a black, velvety appearance and should adhere                      direct proportion to the cell constant, while conductivity
well to the electrode surface. The procedure for platinizing is                   reading instruments will indicate in inverse proportion to cell
not critical. Follow the manufacturer’s instructions or the                       constant. Conductivity cells may be calibrated with reference
following guidelines. Good platinized coatings are obtained                       solutions in accordance with Section 14.
                                                                              4
                                                                                D 1125
11. Temperature Coefficient of Conductivity/Resistivity                             temperature T, and this ratio or correction factor, Q, taken from
   11.1 The conductivity/resistivity of water and aqueous so-                       the smoothed curve.
lutions depends strongly upon the temperature. (See Table 3.)                         NOTE 4—Depending on the type of compensation used, uncompensated
The normal practice is to report conductivity and resistivity                       readings may be obtained by setting temperature to 25°C, by putting the
values referenced to 25.0°C. The coefficient varies depending                       temperature probe in a 25°C bath, or by substituting an electrical
upon the nature and composition of the dissolved electrolytes,                      resistance equivalent to 25°C.
and upon the concentration. The dissociation of water contrib-                        11.5 When using an instrument provided with a manual or
utes significantly to conductivities at 5 µS/cm or less and                         automatic temperature compensator, follow the manufacturer’s
increases the temperature coefficient from near 2 % per °C at                       instructions to calibrate the compensator or check its accuracy
above 5 µS/cm to near 5 % per °C at 0.055 µS/cm. To avoid                           and applicability to the sample being tested.
making a correction, it is necessary to hold the temperature of
the sample to 25 6 0.1°C. If this cannot be done, the                                   TEST METHOD A—FIELD AND ROUTINE
temperature coefficient must be determined and a correction                          LABORATORY MEASUREMENT OF STATIC (NON-
applied. This requires a series of conductivity and temperature                                FLOWING) SAMPLES
measurements on the sample over the required temperature
range. Where automatic temperature compensation is used, the                        12. Scope
temperature compensation algorithm should be chosen that                              12.1 This test method is applicable to field and routine
best simulates the composition of the samples to be tested. In                      laboratory measurements of the electrical conductivity of water
high purity water, 5µ S/cm or less, the variable coefficient shall                  using static samples.
be automatically determined and applied across the range of
measurement for both the dissociation of water and its inter-                       13. Summary of Test Method
action with salt or other contaminations. (See Test Method                            13.1 This test method utilizes dip-type or pipet-type con-
D 5391 and Refs (10), (11), and (12) for more information.)                         ductivity cells for testing static samples having conductivities
   11.2 In static systems, exercise care to avoid change of                         greater than 10µ S/cm. Temperature control and correction
composition caused by loss of volatile constituents or by                           methods are also provided.
pick-up of contaminants from the air to the containing vessel
during the series of measurements.                                                  14. Determination of Cell Constant
   11.3 In flowing systems, provide means for variable heating                        14.1 For the purposes of this test method, the cell constant
or cooling so that the desired range of temperature will be                         of the conductivity cell used shall be known within 61 %. The
covered. Regulate the rate of flow through each cell so as to                       manufacturer’s certification of the cell constant within this
keep the cell adequately flushed.                                                   accuracy is generally considered satisfactory but the user is
   11.4 From the data obtained, plot conductivity against                           advised that damage could occur in shipment and it is best to
temperature. Make sure that the conductivity readings are                           recheck the cell constant when received. If the conductivity
uncompensated. From the curve a table of temperature correc-                        cell has been in service for a period subsequent to this
tion factors may be prepared, or the ratio of conductivity at                       certification, it shall be rechecked by the manufacturer, or in
temperature T to conductivity at 25°C may be plotted against                        the laboratory.
                                                                                      14.2 Rinse the conductivity cell several times with water,
 TABLE 3 Conductivity Values of Pure Water and Increases Due
                                                                                    then at least twice with the KCl reference solution that has a
                    to Sodium Chloride A                                            conductivity nearest to that of the sample under test (Table 1).
                   Conductivity of Pure Water, Conductivity Increase Due to 1
                                                                                    Control the solution temperature to 25 6 0.1°C. Measure the
 Temperature, °C                                                                    resistance of the cell. Repeat the measurement on additional
                            µ S/cm                  mg/L NaCl, µS/cm
        0                  0.011649                       1.1463                    portions of the KCl reference solution until the value obtained
        5                  0.016607                       1.3311                    remains constant to within the limits of precision in accordance
       10                  0.02310                        1.5261                    with Section 18.
       15                  0.03143                        1.7297
       20                  0.04194                        1.9435                      14.3 For instruments reading measured resistance in ohms,
       25                  0.05501                        2.1642                    calculate the cell constant:
       30                  0.07101                        2.3935
       35                  0.09037                        2.6296                                            J 5 10 26 · Rx ~K1 1 K2!
       40                  0.11351                        2.8760
       45                  0.14081                        3.1257                      14.4 For instruments reading measured conductance in Si-
       50                  0.17268                        3.3841                    emens, calculate the cell constant:
       55                  0.2095                         3.6476
       60                  0.2514                         3.9179                                            J 5 10 26 · ~K1 1 K2!/Kx
       65                  0.2987                         4.1882
       70                  0.3516                         4.4773                      NOTE 5—Since the conductivities of a mixture of two solutions are not
       75                  0.4102                         4.7648                    exactly additive, the use of K1 + K2 is only an approximation and requires
       80                  0.4744                         5.0556                    that K2 be much smaller than K1.
       85                  0.5444                         5.3550
       90                  0.6205                         5.6527                    15. Procedure
       95                  0.7030                         5.9264
      100                  0.7930                         6.1933                      15.1 Precision Method Using Temperature Control—Use a
 A
  From Thornton (1).                                                                dip-type or pipet-type cell. Rinse the cell, container, and
                                                                                5
                                                                     D 1125
thermometer thoroughly several times with water and then two             18. Precision and Bias 7
or more times with the sample. Adjust the temperature to 25 6               18.1 This test method was tested by nine laboratories, at
0.1°C as indicated by a thermometer as described in 6.3.1.               four concentration levels, with each operator analyzing each
Allow sufficient time for equalization of temperatures. Read             sample on three different days. These collaborative test data
the conductance or resistance. Calculate conductivity or resis-          were obtained on reagent water. For other matrices, these data
tivity according to Section 16 using Q 5 1, since no tempera-            may not apply. These data were developed using the routine
ture correction is required.                                             method (temperature correction) described in 15.2. The actual
   15.2 Routine Method Using Temperature Correction—Use                  temperature of samples tested by the participants ranged from
a dip-type or pipet-type cell. Rinse the conductivity cell               18.5 to 26.0°C.
thoroughly several times with water and then two or more                    18.1.1 Precision—The precision of this test method within
times with the sample. Measure the resistance or the conduc-             its designated range appears in Table 4.
tance, and the temperature (to the nearest 0.1°C), on successive            18.1.2 Bias—Recoveries of known amounts of conductivity
portions of the sample until a constant value is obtained. If the        values in a series of prepared standards appears in Table 5.
measuring instrument is provided with a manual temperature                  18.2 This test method meets requirements for precision and
compensator, adjust this to the sample temperature value                 bias specified in Practice D 2777 – 86.
before reading the instrument. If an automatic temperature
compensator is provided, no adjustment is necessary, but                       TEST METHOD B—CONTINUOUS, IN-LINE
sufficient time must be allowed to permit equalization of                                MEASUREMENT
temperature. If the instrument has no means of temperature
compensation, determine a temperature correction factor in               19. Scope
accordance with the instructions in Section 11 to convert                  19.1 This test method is applicable to the continuous, in-line
readings to 25°C. If instrument temperature compensation is              measurement of the electrical conductivity of water.
used, calculate conductivity or resistivity according to Section
16 using Q 5 1, otherwise use Q as determined in Section 11.             20. Summary of Test Method
                                                                           20.1 This test method utilizes a flow-type conductivity cell
16. Calculation
                                                                         to sample a continuous stream of the water under test.
  16.1 For instruments reading measured resistance in ohms,              Temperature control and correction methods are also provided.
calculate the conductivity of the sample:
                         K 5 10 6 · J/RxQ
                                                                         21. Procedure
  16.2 For instruments reading measured resistance in ohms,                 21.1 Precision Method Using Temperature Control—Use a
calculate the resistivity of the sample:                                 flow-type conductivity cell. Adjust the sample stream, known
                                                                         to be free of corrosion products and other particulate contami-
                            R 5 RxQ/J                                    nation, to a proper flow rate and bring the temperature to 25 6
  16.3 For instruments reading measured conductance in Si-               0.1°C as indicated by a thermometer as described in 6.3. Allow
emens, calculate the conductivity of the sample:                         sufficient time to reach equalization of temperatures. Read the
                         K 5 10 6 · JKx/Q
                                                                         conductance or resistance. Calculate the conductivity or resis-
                                                                         tivity according to Section 16 using Q 5 1, since no tempera-
  16.4 For instruments reading measured conductance in Si-               ture correction is required.
emens, calculate the resistivity of the sample:                             21.2 Routine Method Using Temperature Correction—Use
                            R 5 Q/JKx                                    a flow-type conductivity cell. Adjust the sample stream, known
   16.5 Automatic recorders and indicators provided with tem-            to be free of corrosion products and other particulate contami-
perature compensators, when used with conductivity cells of              nation, to a proper flow rate and bring the temperature to a
the required cell constant, usually read directly in terms of            steady value as near 25°C as possible. Read the temperature to
siemens per centimetre orµ S/cm referred to 25°C. No calcu-              the nearest 0.1°C. If the measuring instrument is provided with
lations are necessary if the compensator is corrected for the            a manual temperature compensator, adjust this to the sample
solution in the cell.                                                    temperature value. If an automatic temperature compensator is
                                                                         provided, no adjustment is necessary but sufficient time must
17. Report
   17.1 Report the value of the conductivity at 25°C in terms of
microsiemens per centimetre to the nearest 1 % of the deter-                7
                                                                              Supporting data are available from ASTM Headquarters. Request RR:D-19-
mined conductivity if measurements were made at 25 6 0.1°C,              1139.
otherwise report to the nearest 3 % of the determined conduc-
tivity.                                                                                  TABLE 4 Precision of Test Method A
   17.2 Alternatively, report the value of the resistivity at 25°C         Mean Concentration,       Overall Precision, St,   Single-Operator Precision
in terms of ohm-centimetres to the nearest 1 % of the deter-                     µS/cm                      µS/cm                Pooled, So, µS/cm
mined resistivity if measurements were made at 25 6 0.1°C,                           25.6                      3.4                       0.83
otherwise report to the nearest 3 % of the determined resistiv-                     162.0                      6.1                       3.8
                                                                                   1378.8                     60.9                      14.1
ity.                                                                             108169                     6600                      2918
                                                                     6
                                                                                      D 1125
                    TABLE 5 Bias of Test Method A                                            22.2 Alternatively, report the value of the resistivity at 25°C
     Amounts
                             Mean
                                                                      Statistically
                                                                                          in terms of ohm-centimetres to the nearest 1 % of the deter-
                             Recovery,x,         Bias      % Bias
   Added, µS/cm
                             µS/cm
                                                                      Significant         mined resistivity if measurements were made at 25 6 0.1°C,
                                                                                          otherwise report to the nearest 3 % of the determined resistiv-
        27.8                  25.6              −2.2         − 7.9        yes
       167.6                 162.0              −5.6         − 3.3        yes             ity.
      1408.8                1378.8             −30.0         − 2.1        yes
     111342                108169             − 3173         − 2.8        yes
                                                                                          23. Precision and Bias
                                                                                             23.1 Since this test method involves continuous sampling, a
be allowed to permit equalization of temperatures. Read the                               general statement of precision and bias is not applicable.
conductance or resistance. If the instrument has no means of                                 23.2 Experience has shown that errors of 1 to 30 % may be
temperature compensation, determine a temperature correction                              encountered, depending on the equipment and techniques used.
factor in accordance with Section 11 to convert readings to                               Errors in temperature compensation are especially troublesome
25°C. If instrument temperature compensation is used, calcu-                              at conductivities below 10 µS/cm. Additional errors may be
late conductivity or resistivity according to Section 16 using Q                          encountered with low constant cells because of the difficulties
 5 1, otherwise use Q as determined in Section 11.                                        involved in verifying the cell constant at low conductivity
22. Report                                                                                levels.
   22.1 Report the value of the conductivity at 25°C in terms of
microsiemens per centimetre to the nearest 1 % of the deter-                              24. Keywords
mined conductivity if measurements were made at 25 6 0.1°C,                                  24.1 cell constant; conductivity; resistivity; temperature co-
otherwise report to the nearest 3 % of the determined conduc-                             efficient
tivity.
REFERENCES
(1) Thornton, R. D., Light, T. S., “A New Approach to Accurate Resis-                         Specific Conductance of Primary Standard KCl Solutions,” Journal of
    tivity Measurement of High Purity Water,” Ultrapure Water, Vol 6, No.                     Solution Chemistry, Vol 18, No. 6, 1989, pp. 515–528.
    5, 1989, pp. 14–26.                                                                   (8) Shedlovsky, T.,“ The Electrolytic Conductivity of Some Univalent
(2) Symposium on Power Plant Instrumentation for Measurement of                               Electrolytes in Water at 25°C,” Journal of American Chemical Society,
    High-Purity Water Quality, ASTM STP 742, ASTM, 1981.                                      Vol 54, 1932, p. 1411.
(3) “Methods for Determination of Quality and Purity of Steam,” ASME                      (9) International Critical Tables, Vol 3, 1928, p. 87.
    Power Test Code, Supplement on Instruments and Apparatus, Part                        (10) Light, T. S., Licht, L. L. “Conductivity and Resistivity of Water from
    19.11.                                                                                     the Melting to Critical Points,” Analytical Chemistry, Vol 59, 1987,
(4) Rossum, J. R., “Conductance Method for Checking Accuracy of Water                          pp. 2327–2330.
    Analyses,” Analytical Chemistry, Vol 21, 1949, p. 631.                                (11) Harned, H. S., Owen, B. B., The Physical Chemistry of Electrolytic
(5) Jones, G., Bradshaw, B. C., “The Measurement of the Conductance of                         Solutions, Third Ed., Reinhold Publishing Corp., New York, 1958, p.
    Electrolytes, V. A. Redetermination of the Conductance of Standard                         234.
    KCl Solutions in Absolute Units,” Journal of American Chemical                        (12) Gray, D. M., Tenney, A. S., “Improved Conductivity/Resistivity
    Society, Vol 55, 1933, p. 1780.                                                            Temperature Compensation for High Purity Water,” Ultrapure Water,
(6) Wu, Y. C., Koch, W. F., Hamer, W. J., Kay, R. L.,“ Review of                               July/August, 1986.
    Electrolytic Conductance Standards,” Journal of Solution Chemistry,                   (13) Glasstone, S., An Introduction to Electrochemistry, D. Van Nostrand,
    Vol 16, No. 12, 1987, pp. 985–997.                                                         New York, 1942, pp. 50, 56, 61.
(7) Wu, Y. C., Pratt, K. W., Koch, K. F.,“ Determination of the Absolute
                     The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted in connection
                  with any item mentioned in this standard. Users of this standard are expressly advised that determination of the validity of any such
                  patent rights, and the risk of infringement of such rights, are entirely their own responsibility.
                      This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
                  if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
                  and should be addressed to ASTM Headquarters. Your comments will receive careful consideration at a meeting of the responsible
                  technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should make your
                  views known to the ASTM Committee on Standards, at the address shown below.
                     This standard is copyrighted by ASTM, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
                  Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above address or at
                  610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website (www.astm.org).