Thermal Calibration Guide

This document is a calibration certificate from Tempsens Calibration Centre that: 1) Provides accreditation to ISO/IEC 17025:2005 standards and lists calibration capabilities for temperature measurement instruments between -196°C to 2900°C. 2) Details calibration methods for temperature sensors that include the use of liquid baths, dry block furnaces, and fixed point cells. 3) Specifies calibration measurement capabilities within ±0.01°C to ±4°C depending on the temperature range and instruments being calibrated.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
186 views5 pages

Thermal Calibration Guide

This document is a calibration certificate from Tempsens Calibration Centre that: 1) Provides accreditation to ISO/IEC 17025:2005 standards and lists calibration capabilities for temperature measurement instruments between -196°C to 2900°C. 2) Details calibration methods for temperature sensors that include the use of liquid baths, dry block furnaces, and fixed point cells. 3) Specifies calibration measurement capabilities within ±0.01°C to ±4°C depending on the temperature range and instruments being calibrated.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 5

Laboratory Tempsens Calibration Centre, A Division of Tempsens Instruments

(I) Pvt. Ltd., B-188 A, M.I.A., Road No. 5, Udaipur, Rajasthan


Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2840 (In lieu of C-0321) Page 1 of 5
Validity 25.10.2018 to 24.10.2020 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

THERMAL CALIBRATION

I. TEMPERATURE

1. Contact Thermometry Approx (-) 196 ºC 0.05 °C Using digital


RTD, Thermocouples, thermometer & SPRT
with or without with LN2 comparator
Temperature Indicator/
Recorder / Controller Using digital
& Digital (-)80 ºC to 0 ºC 0.03 °C thermometer & SPRT
$
Thermometer with Liquid Bath

0 ºC to 250 ºC 0.04 °C Using digital


thermometer & SPRT
with Liquid Bath

250 ºC to 650 ºC 0.12 °C Using digital


thermometer & SSPRT
with Dry Block Furnace

650 ºC to 1200 ºC 1.30 °C Using digital


thermometer &
Thermocouple with Dry
Block Furnace

1200 ºC to 1600 ºC 2.60 °C Using digital


thermometer &
Thermocouple with Dry
Block Furnace

Vishal Shukla Avijit Das


Convenor Program Manager
Laboratory Tempsens Calibration Centre, A Division of Tempsens Instruments
(I) Pvt. Ltd., B-188 A, M.I.A., Road No. 5, Udaipur, Rajasthan
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2840 (In lieu of C-0321) Page 2 of 5
Validity 25.10.2018 to 24.10.2020 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

2. Non Contact 0 ºC to 250 ºC 1.50 °C Using Digital


Thermometry Thermometer & PRT
& with BB Source
Infrared
Thermometers, 250 ºC to 500 ºC 2.40 °C Using Digital
Pyrometers, Thermal Thermometer & PRT
$
Imagers with BB Source

500 ºC to 1500 ºC 2.50 °C Using Pyrometer with


Black Body Source

1500 ºC to 1700 ºC 3.20 °C Using Pyrometer with


Black Body Source

1700 ºC to 2700 ºC 4.00 °C Using Pyrometer with


Black Body Source

2700 ºC to 2900 ºC 4.00 °C Using Pyrometer with


Black Body Source

Vishal Shukla Avijit Das


Convenor Program Manager
Laboratory Tempsens Calibration Centre, A Division of Tempsens Instruments
(I) Pvt. Ltd., B-188 A, M.I.A., Road No. 5, Udaipur, Rajasthan
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2840 (In lieu of C-0321) Page 3 of 5
Validity 25.10.2018 to 24.10.2020 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

3. Calibration of SPRT,
SSPRT, PRT,
Thermocouple at Fixed
$
Point of
Triple Point of Water 0.01°C 3.8 m°C Using Resistance
Bridge & SPRT with
TPW Cell &
Maintenance Apparatus

Melting Point of 29.7646°C 6.5 m°C Using Resistance


Gallium Bridge & SPRT with Ga
Cell & Maintenance
Apparatus

Melting Point of Tin 231.928°C 6.5 m°C Using Resistance


Bridge & SPRT with Sn
Cell & Maintenance
Apparatus

Melting Point of Zinc 419.527°C 7.1 m°C Using Resistance


Bridge & SPRT with Zn
Cell & Maintenance
Apparatus

Melting Point of 660.323°C 7.5 m°C Using Resistance Bridge


Aluminum & SPRT with Al Cell &
Maintenance Apparatus

Calibration of
Thermocouple at
Secondary Fixed Point
$
of Melting Point of 1064.18 °C 0.72°C Using Pure Gold Wire,
Gold Furnace & 6.5 DMM
( Wire Bridge Method )

Melting Point of 1554.8 °C 0.83°C Using Pure Gold Wire,


Palladium Furnace & 6.5 DMM
(Wire Bridge Method )

Vishal Shukla Avijit Das


Convenor Program Manager
Laboratory Tempsens Calibration Centre, A Division of Tempsens Instruments
(I) Pvt. Ltd., B-188 A, M.I.A., Road No. 5, Udaipur, Rajasthan
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2840 (In lieu of C-0321) Page 4 of 5
Validity 25.10.2018 to 24.10.2020 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

4. Temperature Of (-) 25°C to 0 °C 0.07 °C Using digital


Contact Thermometry thermometer & SSPRT

Type Sensor with Dry Block
( RTD’s,
Thermocouples, With 0°C to 140°C 0.04°C Using digital
Or Without Indicator, thermometer & SSPRT
Recorder, Controller & with Dry Block
Digital Thermometer )
140°C to 250°C 0.09 °C Using digital
thermometer & SSPRT
with Liquid Bath

250°C to 650°C 0.12 °C Using digital


thermometer & SSPRT
with Dry Block Furnace

650°C to 1200°C 1.30 °C Using digital


thermometer &
Thermocouple with Dry
Block Furnace

5. Non Contact 0°C to 250°C 1.50 °C Using PRT with BB


Thermometry Infrared Furnace
Thermometers,

Pyrometers 250°C to 500°C 2.40 °C Using PRT with BB
Furnace

500°C to 1200°C 2.5 °C Using Pyrometer with


Tempsens BB Furnace

6. Calibration of Oven, (-)80°C to 200°C 2.8 °C Using 9 nos. of PRT



Furnaces, Chambers with Multichannel
Datalogger

200°C to 1200°C 4.1 °C Using 9 nos. of T/C with


Multichannel Datalogger

Vishal Shukla Avijit Das


Convenor Program Manager
Laboratory Tempsens Calibration Centre, A Division of Tempsens Instruments
(I) Pvt. Ltd., B-188 A, M.I.A., Road No. 5, Udaipur, Rajasthan
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2840 (In lieu of C-0321) Page 5 of 5
Validity 25.10.2018 to 24.10.2020 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

7. Temperature (-)80°C to 650°C 0.20 ºC Using digital


Indicators with sensor thermometer & SSPRT
of temperature Baths,
#
Furnaces, Ovens 650°C to 1200°C 1.70 ºC Using digital
thermometer &
Thermocouple
1200°C to 1600°C 3.0 ºC
Using digital
thermometer &
Thermocouple

8. Temperature 0°C to 1700°C 3.0 ºC Using master Pyrometer


Indicators with sensor
of Black Bodies
#
Furnaces

9. Calibration of Black 0°C to 500°C 2.3 ºC Using Master Pyrometer


#
Body Furnace 500°C to 1700°C 5.2 ºC

* Measurement Capability is expressed as an uncertainty () at a confidence probability of 95%


$
Only in Permanent Laboratory

Only for Site Calibration
#
The laboratory is also capable for site calibration however, the uncertainty at site depends on the
prevailing actual environmental conditions and master equipment used.

Vishal Shukla Avijit Das


Convenor Program Manager