Abacus 5 Tubing Guide for Engineers
Abacus 5 Tubing Guide for Engineers
This guide was created to help service engineers to repair or refurbish five part analyzers by changing
used or dirty tubing of the unit to a new set. Tubing is an important and sensitive part of the
analyzer. The length, diameter and material of the tubing are very important to ensure the correct
function of the unit. It is highly recommended to use the defined tube types and solutions given in
this guide. Non-compliance in the use of the recommended tubing system could create overflow and
damage electical parts of the unit.
- Total tube system on the valve block (tube types with length and part number)
Tools needed:
- Cutter
Note: Sections of the tubing system can be also changed by using this guide
                                Warning
The original tubing system may contain human sample, clot and infective particles. Before removing
any tubing, it is highly recommended to prime cleaner solution through the unit by connecting
cleaner to all the inlets. Make stress or blank measurements to remove these particles to make
removal of the tubing more safe.
When re-tubing an analyzer always do this with completely drained system. Preparing for shipment is
highly recommended before removing the tubing (after cleaning is ready).
The removed tubing parts are highly biohazard material so place all removed tubing and parts into a
Biohazardous material container.
Preparing of the "T" tubes:
The system contains tubes connected to a plastic junction. These tubes are called "T" tubes and they
are pre-defined to make the tubing exchange faster. Creating these tubes is necessary before starting
the exchange of the tubing. The "T" tubes are indicated with letter T and the identification number.
small "T" junction part number (A5451) Not needed for valve block tubing
middle "T" junction part number (A556) in T0, T1, T2, T3, T5, T7, T12, T16
large "T" junction part number (A545) in T4, T6, T9, T10, T11,
"Y" junction part number (A546) in T2, T3, T8, T13, T14, T15, T16
During the procedure, the tube data is indicated between () signs. (order number, external/internal
diameter, tube material) Please use the images and the steps in combination.
     T0
     T0                                                                      T1
                     4                  8           Glass
         V 12/2                                                                   V 15/1        7                  12
                                                  Reservoir                                                                   V2/2
                                                   bottom
                           10
                                                                                                       2
V 1/3 V 12/3
           8                                         4                     22           Vacuum
                                                                                        Chamber
                     2,5
                                        2                      2
16
                2                7               2
  WBC                                                                Shear                                                               2
 chamber                                                             Valve
                            40               2                                                                            119                4
                                                                                        Wash head                                                        V14/1
V29/1 V11/2
                                                                                                  T6
 T5 (3 inlet wash head)                              V13/1
                                                                                                                     22              4
                                                         2                                        V36/2                                          V17/3
                                        114                     4                                                               3
                                                                                V14/1
V26/2
T7 T8
                4                                                                                                                   31
                                                                                                                                             V16/1
                            47              Sheath
                                            Reservoir                                   Dilu Reservoir           6
                                                                                              2.
                1,7
V32/2
 T9
                                        Dilu                      Dilu
                                     Reservoir 2.              Reservoir 1.
                                                 2                     2
               2,5               18,5                    8                      20              3,5            9,3
 V43/1                                                                                                                    V36/1
3,5 2 2
               2,5
 V41/1                                                                                  V42/1          V40/1
10
                     V37/1
T10                                                    T11
                         MIX                                                   RBC
                       chamber                                               chamber
2 3,5
             40                                                        40
V42/2                                                  V29/2
9 9
V9/2 V10/2
                                                                                Glass Reservoir         7
 Vacuum                 6                     3
                                                           V35/2                    middle
 Reservoir
T14 T15
Step1.1:
Locate the hole on the bottom right side of the valve door. Pass the two tubes (A542, 6/3, silicon)
connected to the pumps through the hole. Both tubes should have the converter (A547, 3.2/3.2) at
the end as shown on the image below. 20 cm tube is connected to pump 1's (outer pump) top, 50 cm
tube is connected to pump 2's (internal pump) top.
Step1.2:
Connect the 35 cm tube (A543, 4/1.8, tygon) to valve 13/3 and connect the other end to the
converter (A547, 3.2/3.2) on pump tube 50 cm.
Step1.3:
Connect the other 35 cm tube (A543, 4/1.8, tygon) to valve 14/3 and connect the other end to the
converter (A547, 3.2/3.2) on pump tube 20 cm.
Step1.4:
Connect the 56 cm tube (A5401, 4/2, RCT-NOVA-971) to valve Diff 5 P puffer reservoir's bottom front
inlet. Thread the other end through the hole (to pumps), under the pumps through the tube holders
and to the Diff 5 P metal inlet.
Step1.5:
Connect the 73 cm tube (A543, 4/1.8, tygon) with green marker to valve 16/3. Thread the other end
through the hole (to pumps), under the pumps through the tube holders and to Diluent metal inlet.
Step1.6:
Attach tube "T0" to valve 1/3 (10 cm) and valve 12/2 (4 cm). Thread the unattached end through the
hole (To glass reservoir's bottom) and attach it to the bottom inlet of the glass reservoir.
Step1.7:
Connect the other 52 cm tube (A543, 4/1.8, tygon) with yellow marker to Lyse reservoir's bottom
front inlet. Thread the other end of the tube through the hole (to pumps) under the pumps through
the tube holders and to the Lyse metal inlet.
                                    Pressure sensor                       Vacuum         MIX                RBC             WBC
                                         board                            chamber      chamber            chamber         chamber
To glass
reservoir's
TOP
T0
                    Valve1    Valve2        Valve3     Valve4    Valve5     Valve6     Valve7         Valve8    Valve9        Valve10      Valve11
                   Valve12     Valve13      Valve14    Valve15   Valve16     Valve17      Valve18     Valve19   Valve20       Empty        Empty
To glass
reservoir's
BOTTOM
                                                                                                                                                   73
              T0   T0                                                                                                                 35
                                                                                                                     35
                                                                                                                                      C
                                                                                                                     C
                                                                                                                                      C
                   Valve23     Valve24      Valve25    Valve26   Empty      Valve28    Valve29        Valve30   EmptyC        Valve32      Valve33
Valve34 Valve35 Valve36 Valve37 Valve38 Valve39 Valve40 Valve41 Valve42 Valve43 Valve44
                                                                                                                                                                   20
                        Vacuum              Sheath           Diff 5 P                                Diluent         Diluent
                                                                                Lyse
                        reservoir          reservoir        reservoir                               reservoir       reservoir                           50
                                                                              reservoir
                                                                                                        1               2
                                                                                                                                                                   C
                                                                                                                                                        To pumps
                                                                                                                                                        CC         C
52
                                                                                                                         56
                                                                                                                                  C
                                                                                                                         C        C
                                                                                                                         C
Step2.1:
Connect a 22 cm tube to valve 3/2 and lead the other end through the hole (to glass reservoir's top)
and attach it to the glass reservoir's top inlet. Note that some glass reservoirs has two inlets on the
top. In this case "T13" should have a "Y" junction added with 2,5 cm and 7 cm on the two ends, so
both glass reservoir’s inlets can be covered with tygon. See the image below.
Step2.2:
Connect the "T" fitting middle of tube "T5" to valve 13/1 (2 cm) and 14/1 (4 cm). Pass the other end
of the tube (119 cm with the red mark) above the reservoirs, under the dilutors to the XY unit and
attach it to the washing head's top inlet. Use the tube holders attached to the frame to grip the tube.
The long tube of "T5" is 114 cm. Add an "Y" junction to this tube with 5 cm tube on both other ends.
Connect these 5 cm tubes to the top and bottom inlet of the washing head.
Step2.3:
Connect the 108 cm tube (A543, 4/1.8, tygon) to valve 18/2 and pass the other end of the tube above
the reservoirs, under the dilutors to the XY unit and attach it to the washing head's bottom inlet.
Connect the same tube to the middle inlet of the washing head.
Step2.4:
Connect the 25 cm tube (A543, 4/1.8, tygon) to valve 24/1. The other end of the tube should have a
converter (2.3/1.6). This end will be connected to the shear valve with 30 cm tube (A5433, 2.23/1.27,
tygon).
Step 2.5:
Connect valve 17/1 and valve 18/3 with an 5 cm tube (A543, 4/1.8, tygon).
Step 2.6:
Connect valve 14/2 and valve 15/3 with a 4 cm tube (A543, 4/1.8, tygon).
Step 2.7:
Connect valve 43/2 and valve 44/3 with a 4 cm tube (A543, 4/1.8, tygon).
                                     Pressure sensor                       Vacuum         MIX                 RBC             WBC
                                          board                            chamber      chamber             chamber         chamber
To glass
reservoir's        T13
TOP
T0
                   Valve1       Valve2       Valve3     Valve4    Valve5     Valve6        Valve7      Valve8    Valve9        Valve10   Valve11
                   Valve12      Valve13      Valve14    Valve15   Valve16    Valve17       Valve18     Valve19   Valve20       Empty     Empty
To glass
reservoir's
BOTTOM
                                                                                           5                                                     73                  108
                                                        4
              T0   T0                        T5                                                                                    35
                                                                                                          25           35
                                                                                                                                   C
                                                                                                                     C
                                                                                                                                   C
                   Valve23      Valve24      Valve25    Valve26   Empty       Valve28      Valve29     Valve30       C
                                                                                                                  Empty        Valve32   Valve33
Valve34 Valve35 Valve36 Valve37 Valve38 Valve39 Valve40 Valve41 Valve42 Valve43 Valve44
T5
                                                                                                                                                                20
                         Vacuum              Sheath           Diff 5 P                                Diluent          Diluent
                                                                                 Lyse
                         reservoir          reservoir        reservoir                               reservoir        reservoir                       50
                                                                               reservoir
                                                                                                         1                2                                     CC
                                                                                                                                                      To pumps
                                                                                                                                                      C
                                                                                                                                                      C
                                                                                                                                   52
                                                                                                                          56
                                                                                                                                   CC
                                                                                                                          C
                                                                                                                          C
                                                                                                                                                          Washing
                                                                                                                                                          head
Step 3.1:
Connect valve 24/3 to valve 25/1 with a 5 cm tube (A543, 4/1.8, tygon).
Step 3.2:
Connect valve 25/3 to valve 26/1 with an 5 cm tube (A543, 4/1.8, tygon).
Step 3.3:
Connect valve 1/2 to the tube transition 7th pin with a 33 cm tube (A543, 4/1.8, tygon).
Step3.4:
Connect valve 1/1 to valve 19/2 with a 25 cm tube (A543, 4/1.8, tygon).
Step 3.5:
Connect "T1" tube's 2 cm long tube to valve 12/3, the two "side" tubes are going to valve 15/1 (7 cm)
and valve 2/2 (12 cm) as shown on the image below.
Step 3.6:
Connect the 1,5 cm tube (A543, 4/1.8, tygon) with exausting went to valve 5/2.
                               Pressure sensor                        Vacuum         MIX                 RBC             WBC
                                    board                             chamber      chamber             chamber         chamber
  To glass
  reservoir's    T13
  TOP
                  T1
                                                                           33
                                                          25
Valve1 Valve2 Valve3 Valve4 Valve5 Valve6 Valve7 Valve8 Valve9 Valve10 Valve11
Valve12 Valve13 Valve14 Valve15 Valve16 Valve17 Valve18 Valve19 Valve20 Empty Empty
To glass
reservoir's
BOTTOM
                                                  4                                   5                                                                 108
                          T1
                                       5          5
                                                                                                     25
Valve23 Valve24 Valve25 Valve26 Empty Valve28 Valve29 Valve30 Empty Valve32 Valve33
Valve34 Valve35 Valve36 Valve37 Valve38 Valve39 Valve40 Valve41 Valve42 Valve43 Valve44
To pumps
                                                                                                                                              Washing
                                                                                                                                              head
Step 4.1:
Connect the Diff 5P reservoir's top inlet to valve 6/2 with a 25 cm tube (A5401, 4/2, RCT-NOVA-971).
Step 4.2:
Connect the 9.3 cm end of the "T9" tube to valve 36/1. Follow the tube to the nearest "T" junction
with 2 cm tube and connect it to valve 40/1. Connect the next "T" junction with 2 cm tube to valve
42/1.
Connect the 10 cm end of "T9" and connect it to valve 37/1. Follow the tube to nearest "T" junction
with 2.5 cm tube and connect it to valve 41/1. Connect the next "T" junction with 2.5 cm tube to
valve 43/1.
Connect the last two "T" fitting with 2 cm tubes to Diluent reservoir's bottom rear inlets. Settle the
tube as shown on the image below.
Step 4.3:
Connect "T14" "Y" junction with the 4 cm tubes to the top inlets of the two Diluent reservoirs.
Connect other end of "T14" to valve 8/2 (23 cm) as shown on the image.
Step 4.4:
Connect lyse reservoir's top inlet to valve 7/2 with a 25 cm tube (A543, 4/1.8, tygon).
Step 4.5:
Connect "T8" "Y" junction with 6 cm tube to Diluent reservoir 2 bottom front inlet, connect the 9 cm
tube to Diluent reservoir 1 bottom front inlet. Connect the 31 cm tube of "T8" to valve 16/1 as shown
on the image.
Step 4.6:
Connect "T4" "T" fitting (with 2 cm tube facing up) to bottom of WBC chamber. Find the second "T"
fitting (with 2 cm tube facing down) and connect it to valve 11/2. Connect the 40 cm tube to valve
29/1 with a curve shown on the image below.
                              Pressure sensor                     Vacuum             MIX                RBC             WBC
                                   board                          chamber          chamber            chamber         chamber
 To glass
 reservoir's
 TOP
                 T1                                                           25
                                                                       33                                                         T4
                                                     25     25
                                                                                                  T14
Valve1 Valve2 Valve3 Valve4 Valve5 Valve6 Valve7 Valve8 Valve9 Valve10 Valve11
Valve12 Valve13 Valve14 Valve15 Valve16 Valve17 Valve18 Valve19 Valve20 Empty Empty
To glass
reservoir's
BOTTOM
                         T1
                                      5          5           T8
                                                                                                                T4
Valve23 Valve24 Valve25 Valve26 Empty Valve28 Valve29 Valve30 Empty Valve32 Valve33
Valve34 Valve35 Valve36 Valve37 Valve38 Valve39 Valve40 Valve41 Valve42 Valve43 Valve44
                                                                                                                     T9
                                                                                                 T14
T8
                                                                                                 T9
Step 5.1:
Find "T11" "T" junction and connect the 2 cm long tube to RBC chamber's bottom. Connect the other,
9 cm long tube on "T" junction to valve 10/2. Lead the 40 cm long tube of "T11" as shown on the
image and connect it to valve 29/2.
Step 5.2:
Find "T10" "T" junction and connect the 2 cm long tube to MIX chamber's bottom. Connect the other,
9 cm long tube on "T" junction to valve 9/2. Lead the 40 cm long tube of "T10" as shown on the
image and connect it to valve 42/2.
Step 5.3:
Connect a 6 cm tube (A543, 4/1.8, tygon) with 1.6/2.3 converter to valve 2/1.
Step 5.4:
Connect "T7" "T" fitting with 4 cm tube to valve 20/2 and 1.7 cm long tube to valve 32/2. Connect the
middle, 47 cm long tube of the "T" fitting to the bottom of sheath reservoir's rear inlet as shown on
the image.
Step 5.5:
Connect a 12 cm tube (A543, 4/1.8, tygon) with 1.6/2.3 converter to valve 18/1.
Step 5.6:
Connect a 6 cm tube (A543, 4/1.8, tygon) with 1.6/2.3 converter to valve 19/1.
                                  Pressure sensor                          Vacuum             MIX                RBC             WBC
                                       board                               chamber          chamber            chamber         chamber
     To glass
     reservoir's
     TOP
                   T11                                                                 25
                                                                                                                                            T4
                                        6                            25              T10
                                                                                                           T14
                   Valve1    Valve2         Valve3        Valve4    Valve5    Valve6        Valve7       Valve8      Valve9         Valve10      Valve11
                   Valve12   Valve13        Valve14       Valve15   Valve16   Valve17       Valve18     Valve19     Valve20         Empty        Empty
To glass
reservoir's
BOTTOM                                                                                                                                 6
                                                                                                                                                           12
                                                                      T8
                                                                        T11                                              T4
                                                                                                                               T7
Valve23 Valve24 Valve25 Valve26 Empty Valve28 Valve29 Valve30 Empty Valve32 Valve33
Valve34 Valve35 Valve36 Valve37 Valve38 Valve39 Valve40 Valve41 Valve42 Valve43 Valve44
T10
                                                                                                                              T9
                                                                                                          T14
To pumps
                                                                                                                    T8
                                                     T7
                                                                                                          T9
Step 6.1:
Connect the 4 cm tube (A543, 4/1.8, tygon) with 1.6/2.3 converter to valve 20/1.
Step 6.2:
Connect valve 38/1 to the bottom of the Diff 5P reservoir's rear inlet with a 40 cm tube (A5401, 4/2,
RCT-NOVA-971) as shown on the image below.
Step 6.3:
Connect "T6" "T" junction with 4 cm tube to valve 17/3 and the 3 cm tube to valve 26/2. Connect the
22 cm long tube of "T6" valve 36/2.
Step 6.4:
Connect valve 39/1 to the bottom of the Lyse reservoir's rear inlet with a 34 cm long tube (A543,
4/1.8, tygon).
Step 6.5:
Connect "T12" "T" junction with the 3 cm tube to valve 35/2 and the 2 cm tube on the middle of the
"T" fitting to valve 34/2. Connect the 6 cm tube to the top of the Vacuum reservoir's front inlet.
Step 6.6:
Connect a 5 cm tube (A543, 4/1.8, tygon) with 1.6/2.3 converter to valve 28/1 .
                                  Pressure sensor                            Vacuum          MIX                  RBC             WBC
                                       board                                 chamber       chamber              chamber         chamber
     To glass
     reservoir's
     TOP
T11
6 T10
                   Valve1     Valve2        Valve3        Valve4    Valve5       Valve6       Valve7         Valve8   Valve9         Valve10   Valve11
                   Valve12   Valve13        Valve14       Valve15   Valve16      Valve17      Valve18     Valve19     Valve20        Empty         Empty
To glass
reservoir's
BOTTOM                                                                                                                                  6
                                                                                                                                                           12
                                               T6
                                                                                                         5
                                                                           T11                                                                 4
                                                                                                                                T7
Valve23 Valve24 Valve25 Valve26 Empty Valve28 Valve29 Valve30 Empty Valve32 Valve33
Valve34 Valve35 Valve36 Valve37 Valve38 Valve39 Valve40 Valve41 Valve42 Valve43 Valve44
T10
                    T12
                                                                                              34
To pumps
                                                     T7
                                                                                                         40
Step 7.1:
Connect "T2" "Y" junction with 10 cm tube to pressure sensor board's sheath sensor (right side). Find
the 3 cm tube to the first "T" junction with 2 cm tube in the middle and connect the tube to valve
3/1. Find the next "T" junction with 2 cm tube in the middle and connect the tube to valve 5/1. Find
the 3rd "T" junction with 2 cm tube in the middle and connect the tube to valve 7/1.
Connect the 2cm tubes on the "T" junctions on the opposite side of "T2" to valve 4/2, valve 6/1 and
valve 8/1.
Find the "Y" junction connecting the two sides of "T2". Lead the 25 cm tube from "Y" junction under
valve 13.
Connect the first "T" junction's middle 2,5 cm tube to valve 13/2. Find the last "T" fitting of "T2" and
connect the 2,5 cm long tube on the junction to valve 12/1 and the 2 cm tube to 23/2. See the image
below.
Step 7.2:
Find "T3" "Y" junction and connect the 16 cm tube on the middle of the junction to 15/2. Find the 8
cm tube on the other side of the "Y" junction and connect it to valve 10/1.
Follow the 2,5 cm tube to "T" fitting, find the 2 cm tube on the middle of the junction and connect it
to valve 9/1. Find the second "T" fitting with the 2 cm tube on the middle and connect it to valve
11/1.
Find the 22 cm tube on the "T" fitting and connect it to the bottom of the Vacuum chamber.
Step 7.3:
Connect a 5 cm tube (A543, 4/1.8, tygon) with 1.6/2.3 converter to valve 30/1.
Step 7.4:
Connect a 8 cm tube (A543, 4/1.8, tygon) with 1.6/2.3 converter to valve 33/1.
Step 7.5:
Connect a 3 cm tube (A543, 4/1.8, tygon) with 1.6/2.3 converter to valve 33/2.
Step 7.6:
Connect a 11 cm tube (A543, 4/1.8, tygon) with 1.6/2.3 converter to valve 44/1. Set the tube as
shown on the image below.
                                        Pressure sensor                     Vacuum         MIX                  RBC             WBC
                                            board                           chamber      chamber              chamber         chamber
    To glass
    reservoir's
    TOP
                  T2
                                                                                                                                                              T3
                   Valve1          Valve2       Valve3     Valve4    Valve5    Valve6       Valve7         Valve8   Valve9      Valve10   Valve11
                   Valve12         Valve13      Valve14    Valve15   Valve16   Valve17      Valve18     Valve19     Valve20     Empty         Empty
To glass
reservoir's
BOTTOM                                                                                                                              11                3
                       T2                          T6
                                                                                                       5
                                                                                                                                          4               8
                                                                                                                        5
Valve23 Valve24 Valve25 Valve26 Empty Valve28 Valve29 Valve30 Empty Valve32 Valve33
Valve34 Valve35 Valve36 Valve37 Valve38 Valve39 Valve40 Valve41 Valve42 Valve43 Valve44
                       T12
                                                                                            34
To pumps
                                                                                                       40
Step 8.1:
Connect the Drain sensor's right side on the middle of the pressure sensor board to the top inlet of
the Vacuum chamber with a 12 cm tube (A543, 4/1.8, tygon).
Step 8.2:
Connect the Chamber sensor's right side on the top of the pressure sensor board to the top rear inlet
of the Vacuum reservoir with a 33 cm tube (A543, 4/1.8, tygon).
Step 8.3:
Connect valve 17/2 to the RBC chamber's lower inlet on the side of the chamber with a 19 cm tube
(A543, 4/1.8, tygon).
Step 8.4:
Connect valve 30/2 to the WBC chamber's lower inlet on the side of the chamber with a 19 cm tube
(A543, 4/1.8, tygon).
Step 8.5:
Connect valve 4/3 to Sheath reservoir's top inlet with a 28 cm tube (A543, 4/1.8, tygon).
Step 8.6:
Connect valve 16/2 to the bottom of Sheath reservoir's front inlet with a 40 cm tube (A543, 4/1.8,
tygon).
                                                   Pressure sensor                     Vacuum         MIX                 RBC            WBC
                                                        board              12          chamber      chamber             chamber        chamber
     To glass
     reservoir's
     TOP
                             T2
                                                                                                            19                                                       T3
                              Valve1          Valve2        Valve3    Valve4    Valve5    Valve6       Valve7       Valve8   Valve9      Valve10   Valve11
                              Valve12         Valve13      Valve14    Valve15   Valve16   Valve17      Valve18     Valve19   Valve20      Empty    Empty
To glass
reservoir's
BOTTOM                                                                                                                                        11             3
                                                                                                                              19
                                  T2
                   33                                                                                                                                            8
                                                                                                                                   5
Valve23 Valve24 Valve25 Valve26 Empty Valve28 Valve29 Valve30 Empty Valve32 Valve33
Valve34 Valve35 Valve36 Valve37 Valve38 Valve39 Valve40 Valve41 Valve42 Valve43 Valve44
28
40
                                                                                                                                                                     To pumps
Step 9.1:
Connect valve 34/1 to RBC chamber's top inlet on the side of the chamber with a 36 cm tube (A543,
4/1.8, tygon).
Step 9.2:
Connect valve 35/1 to WBC chamber's top inlet on the side of the chamber with a 42 cm tube (A543,
4/1.8, tygon).
Step 9.3:
Connect valve 37/2 to valve 26/3 with a 14 cm tube (A543, 4/1.8, tygon) as shown on the image
below.
Step 9.4:
Connect valve 39/2 to valve 28/3 with a 14 cm tube (A543, 4/1.8, tygon) as shown on the image
below.
Step 9.5:
Connect a 17 cm tube (A5401, 4/2, RCT-NOVA-971) with 1.6/2.3 converter to valve 38/2. Set the tube
as shown on the image below. (The converter will be connected to the tube transition 4th pin with 20
cm tube (A5433, 2.23/1.27, tygon).
Step 9.6:
Connect valve 44/2 to valve 33/3 with a 14 cm tube (A543, 4/1.8, tygon) as shown on the image
below.
                                               Pressure sensor                        Vacuum          MIX                RBC             WBC
                                                    board              12             chamber       chamber            chamber         chamber
     To glass
     resreservoir's
     TOP
                                                                           36                               19
                                                                                                                        42
                                Valve1    Valve2        Valve3    Valve4        Valve5    Valve6       Valve7       Valve8   Valve9      Valve10    Valve11
                                Valve12   Valve13      Valve14    Valve15       Valve16   Valve17      Valve18     Valve19   Valve20     Empty          Empty
To glass
reservoir's
BOTTOM
                                                                      17                                                      19
                                                                                                                                                                           20
                      33
Valve23 Valve24 Valve25 Valve26 Empty Valve28 Valve29 Valve30 Empty Valve32 Valve33
Valve34 Valve35 Valve36 Valve37 Valve38 Valve39 Valve40 Valve41 Valve42 Valve43 Valve44
14 14 14
28
40
                                                                                                                                                                To pumps
Step 10.1:
Connect a 15 cm tube (A543, 4/1.8, tygon) with 1.6/2.3 converter to valve 24/2.
Step 10.2:
Connect a 12 cm tube (A543, 4/1.8, tygon) with 1.6/2.3 converter to valve 25/2.
Step 10.3:
Connect valve 41/2 to valve 30/3 with a 14 cm tube (A543, 4/1.8, tygon) as shown on the image
below.
Step 10.4:
Connect valve 40/2 to valve 29/3 with a 14 cm tube (A543, 4/1.8, tygon) as shown on the image
below.
Step 10.5:
Connect a 5 cm tube (A543, 4/1.8, tygon) with 1.6/2.3 converter to valve 28/2. The converter will be
connected to the tube transition 5th pin with 20 cm tube (A5433, 2.23/1.27, tygon).
Step 10.6:
Connect "T15" 23 cm long tube (single end) to valve 23/1. Connect both 10 cm tubes on the other
side of the "Y" junction to the vacuum reservoir's bottom inlets.
Step 10.7:
Connect a 2 cm tube (A543, 4/1.8, tygon) with 1.6/2.3 converter to valve 32/1. The converter will be
connected to the tube transition 6th pin with 14 cm tube (A5433, 2.23/1.27, tygon).
                                        Pressure sensor                        Vacuum           MIX                 RBC            WBC
                                             board                             chamber        chamber             chamber        chamber
     To glass
     reservoir's
     TOP
14
                                                                    36
                                                                                                                  42
                         Valve1    Valve2        Valve3    Valve4        Valve5     Valve6       Valve7       Valve8   Valve9      Valve10    Valve11
                         Valve12   Valve13      Valve14    Valve15       Valve16    Valve17      Valve18     Valve19   Valve20      Empty         Empty
To glass
reservoir's
BOTTOM
                                                                                                                                                             20
                                                                                     5
                                                               17                                                                                                      20
                                   15            12                                                                                     2
                                                                               14
                         Valve23   Valve24      Valve25    Valve26       Empty      Valve28      Valve29     Valve30    Empty      Valve32    Valve33
Valve34 Valve35 Valve36 Valve37 Valve38 Valve39 Valve40 Valve41 Valve42 Valve43 Valve44
14 14 14 14
T15
                                                                                                                                                          To pumps
                    Step 11:
                    Connect dilutor pistons with the appropriate valves according to the image. Use same tube (A543,
                    4/1.8, tygon) as in the valves. Valve 38 to dilutor piston is a different type (A5401, 4/2, RCT-NOVA-
                    971).
37
37
37
37
35
35
35
17