Yp 3000

1. The document provides the operator's manual and warranty for an infant incubator. 2. The warranty covers defects in materials and workmanship for one year except for consumables and disposables which are only covered upon shipment. 3. The manual outlines important safety precautions including monitoring the infant's temperature regularly, not leaving the incubator in direct sunlight, and only allowing qualified personnel to service the equipment.

Uploaded by

Heidi Blue
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
2K views100 pages

Yp 3000

1. The document provides the operator's manual and warranty for an infant incubator. 2. The warranty covers defects in materials and workmanship for one year except for consumables and disposables which are only covered upon shipment. 3. The manual outlines important safety precautions including monitoring the infant's temperature regularly, not leaving the incubator in direct sunlight, and only allowing qualified personnel to service the equipment.

Uploaded by

Heidi Blue
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 100

YP-3000

INFANT INCUBATOR
Operator’s Manual

www.chiandavid.cn
OPERATOR'S MANUAL FOR INFANT INCUBATOR

WARRANTY

The product being described in this manual is warranted against defects in materials or

workmanship for one year from the date of shipment, with the following exceptions:

1. All consumable and disposable products are guaranteed to be free from defects upon

shipment only.

2. Calibrations are considered normal maintenance and are not included in the 1-year

warranty.

During the warranty period any defective parts other than those listed above will be replaced

at no charge to the customer.

This warranty is rendered void and our company cannot be held liable for conditions resultant

there from if:

1. Damage to the unit is incurred as a result of mishandling.

2. The customer fails to maintain the unit in a proper manner.

3. The customer uses any parts, accessories, or fittings not specified or sold by our company.

4. Sale or service is performed by the non-certified service/dealer agency.

This warranty is in lieu of all other warranties, expressed or implied, and our company shall in

no event be liable for incidental or consequential damages including loss of use, property damage,

or personal injury resulting from breach of warranty.

The Accreditation Manual for Hospitals requires each piece of equipment to be tested prior to

initial use and at least annually thereafter. To comply with this standard, we recommend that you

participate in our accreditation Testing compliance Program during the warranty period. This

service can be performed through our company and authorized dealers.

SERVICE

For optimal performance, product service should be performed only by qualified service

personnel who is authorized by manufacturer. Please contact the local agency or the After-Sales

of our company to get more technical information about maintenance.

i
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

OPERATING PRECAUTIONS
1. INFANT INCUBATOR (Hereinafter referred to as the equipment) belongs to high risk medical
device which can endanger infant's life. Therefore please use the device only in neonate nursing
room, children nursing room, pediatric intensive care unit or similar sickroom in hospital. Operators
for the device should be special trained and operate the device under the instruction of medical
practitioner. The device is intended for the infant whose weight is less than 10kg. (Not exceeding
8kg when using weighing scale).
2. The operator must keep observing the patient's condition while the equipment is working.
Supervise and record baby's temperature regularly to check whether the temperature of the
patient is over high/low or any other unusual conditions happen. Suggest monitoring the baby
temperature at least half hour a time.
3. Please stop using the equipment when it failure or disfunction. Turn off the power and transfer
the patient out from the equipment, then inform our company or our authorized agency for service.
DO NOT ask for service from person who's not been authorized by our company.
4. The equipment is only for intended use which is applied as specified in this manual.
5. Direct radiation from sunlight or other infrared source could cause overheating of the infant
without activating the Over Temperature Alarm. DO NOT leave the INCUBATOR in direct sunlight
or near other sources of radiant heat.
6. DO NOT leave the equipment in the presence of flammable anaesthetic gases or other
flammable materials, such as some types of cleaning fluids.
7. DO NOT leave the equipment in the presence of strong electromagnetic field. Portable and
mobile radio frequency communication equipment may have effect to this equipment.
8. Equipment which are easily interfered by magnetic field should not be used near the equipment
because they may interfered by the equipment.
9. The equipment does not equip the air cleaner, to make sure the good air quality inside hood, the
equipment should be used in the environment with clean air.
10. Please DO NOT use the equipment under working environment not stipulated in table 1.1, or
else, it may cause the failure or the equipment can not reach the requirements.
11. Make sure the patient, clothing, supervise wire, pipeline and other similar items inside the
range of bad when open or close the baffle and operation window. Before or after pull-out, rotate,
tilted the bed, must check all of the wire and pipeline which connect with patient. Because of the
bed moving may pull-out the wire or pipeline and cause the wire or pipeline break. This can make
some conditions occur like limited the flow of air or liquid, or the probe deviated the correct position,
and so on.

ii

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

12. The panel of the equipment is required to check periodically to make sure it is securely
installed to prevent the patient falling from bed and injured.
13. When the patient is inside the device, all the contact with patient should through baffle and
operate windows.
14. When the Access Panel and Access Doors are open, the temperature from the air temperature
indicator maybe not the real temperature inside incubator. Therefore, do not leave the Access
Panels and Access Doors open in a long period.
15. All access panel‘s latch should be firm plugged, in case accidental open.
16. For infant safety, DO NOT leave the infant unattended when the Front and Back Access
Panels are open.
17. Other accessories within the incubator which can alter the air flow pattern may affect
temperature uniformity and temperature variability.
18. Patient safety and incubator performance may be compromised if air flow passages are not
kept clear of obstructions (blankets, stuffed animals, etc.) during clinical usage.
19. Do not place surgical covers or blankets over the infant and warm air curtain or side vents
simultaneously. This may cause heat induced injury and burns.
20. Wearing clothes or blanket for patient will prevent their body receiving the infrared radiant heat
effectively. It is recommended to be exposed in the device.
21. The equipment should be moved by at least two personnel who have certain strength. Located
in the equipment back handle can be used as mobile handle (This handle can not carry the whole
machine). Please pull out all power cords before moving.
22. To prevent the harm on patient for accidental moving, please lock the wheels during usage.
23. To prevent accidental disconnection, secure all patient leads, infusion lines and ventilator
tubing to the mattress with sufficient excess length to allow for the full range of mattress height
adjustment.
24. To avoid water over flow from water chamber, and keeping the most stable position of the
equipment, before moving all of accessories should be fixed to the right position and let out all of
the water.
25. Using hands to touch the heater and protective cover is forbidden. It can prevent scalding.
26. Do not place any items on the box tent to avoid the falling items hurt operators when the box
tent is lifted. Do not place leads and lines or make them tied on the box tent. It can avoid the
disconnection of the patient with the leads and lines when lifted. Before lifting the box tent, you
must ensure there is no obstacles within the range of movement.
27. Do not switch on for longer time than essential, when the machine not connect to the power
supply, otherwise, power supply failure alarm will be actively and the internal battery current
waste.

iii
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

28. When using X-Ray tray via incubator hood, the shadow of hood will reflect on the X-ray
negative. That maybe influence the doctor’s diagnose.
29. Do not place any article higher than incubator’s caster under its VHA stand which may affect
the stabilization of VHA stand.
30. When operating the VHA stand, support the equipment with one hand on to prevent it from
unbalance.
31. In nursing operation, the operator can not touch the other charged equipment at the same time,
may bring shock hazard to patients.
32. Only the authorized and qualified maintenance personnel can replace the fuse according to
the specification. When replace the fuse, you should disconnect the power supply of the
equipment first, and can not touch the patients and metal parts at the same time.
33. The equipment must be fully cleaned and sterilized for the first time for initial use, after nursing
for one baby, after used it for one week or there's dirt in the equipment. Cleaning and sterilizing
methods please refer to Section 5.
34. Please use the neutral detergent/ disinfectant registered by nation. Other disinfectant (like
alcohol) will destroy some parts of the incubator. Please follow the instruction for detergent usage.
35. After cleaning the incubator by combustible cleaning solvent should airing the equipment
completely. The residual a handful of the flammable solvent (such as ethyl ether, ethanol or similar
cleaning solvent) in the equipment can cause a fire.
36. For the equipment safety, please only use our company skin temperature sensor, rechargeable
battery and other accessories.
37. The lifetime of equipment internal rechargeable battery is usually 3 years. Before using it every
time, we should inspect the status of internal rechargeable battery according to the requirement in
section 5.3. If it isn't getting through the inspection or the battery has been used more than three
years, it must be replaced. The internal rechargeable battery should be replaced by authorized
qualified service personnel.
38. The equipment should not be near or stack with other equipment to use. If have to, you should
observed in its use of configuration can run normally.
39. Damages will be easily caused if using the equipment after it reached its lifetime. Previous
capability guideline and requirement cannot be reached as well.
40. Must move the check lamp away while go up and down the box tent. Strictly prohibit to take the
check lamp under the radiation heat source that may cause the plastic enclose at the top of the
lamp distortion or damage by overheating. If the check lamp forced vibration or collision, or move
the lamp while is heating all may lead the life period of lamp greatly reduced. So that should
cooling the lamp 10 minutes at least before move the check lamp.

iv

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

41. While the radiation heater equipment is in used, other additional functions like bed heating or

treatment of jaundice used at the same time that may cause the patient overheating. Operator

must consider this factor before setting the radiation heating function.

42. Some parts of the patient's body will closer to the heater at the top of the box tent when the

bad is incline, that will cause the patient's body absorb more Infrared radiation. The user must

increase the check frequency for patient.

43. The life period of equipment is 8 years. The equipment, accessories and the packaging have

to be disposed of waste correctly at the end of the usage. Please follow Local Ordinances or

Regulations for disposal.

ELECTRICAL PRECAUTIONS
1. To avoid the electric shock, this equipment only can be connected with the supply mains which

have protective earth. If there is any doubt with the ground connection, please don’t operate the

equipment.

2. Because of the electric shock damage, while need to disconnect the equipment for

maintenance must submitted to the authorized and qualified service personnel.

3. Make sure the power supply conform to the requirement of VHA stand. Please take off the

power line for safety when stop using or servicing the equipment.

4. Use ancillary equipment not conform of the safety requirements of this equipment will reduce

the safety performance. When select the ancillary equipment must insure that the equipment had

tested according to the requirements in IEC60601-1 or other relevant standard, and acquire the

safety certificate.

5. Consider the safety factor, power plug and socket as the main insulated way with device power

supply. Please ensure the power plug simple operational.

6. Please pull out the power line when the equipment is maintenance or stop using. Must hold the

power plug in hand and don’t use the way like pull the wire while insert or pull out the power line.

Please don’t buckling the joint of power line.

7. Only can use the power line supplied with this equipment, otherwise will reduce the safety

performance.

8. DO NOT put any objects nearby the power socket, For fear that pull out the power socket

inconvenience in an emergency.

9. The applied part which is joint used with this equipment must achieve the require of BF type

applied part.

v
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

HUMIDITY PRECAUTIONS
1. The equipment is with the humidity controller, it can increase the Humidity of equipment
according to the clinical demands.
2. Higher relative humidity will, at any given time, decrease an infant’s water evaporation and may
cause an increase in infant temperature. Monitor the infant’s temperature as required.
3. Use only distilled water to fill or refill the Reservoir. Tap water may contain organisms that may
flourish in the heated water of the humidifier.
4. Make sure all hood access door gaskets and tubing ports are properly installed. Any open gaps
in the equipment hood will reduce the incubator’s internal relative humidity.
5. Fill the humidity chamber to the bottom of the MAXIUM LIMIT line. DO NOT OVERFILL. In order
to avoid water spillage.
6. At high humidity levels within the incubator hood (typically more than 80%RH), Because of the
higher humidity, it will appear spray on the inside walls of the hood.
7. When there is much more difference between the humidity setting value and the humidity of
environment, it is very difficult to reach the needed control value of the humidity inside of the
equipment.
8. Following the doctor’s advice when setting the relative humidity.
9. The life period of humidity sensor is 8 years.

OXYGEN PRECAUTIONS
1. This equipment is with oxygen control system, please use medical oxygen when feeding
oxygen.
2. Abusing of supplemental oxygen may result in serious aftereffects which include blindness,
brain damage, even death. Therefore, keeping to the main doctor's direction strictly and
monitoring the oxygen supplement condition for the patient in a regularly time.
3. If it is necessary to administer Oxygen in an emergency, notify the attending physician
immediately.
4. When supplementing the oxygen, calibrated oxygen analyzer must be turned on for monitoring
the oxygen concentration.
5. Oxygen feeding may increase the noise level inside the equipment.
6. As Oxygen use increases the danger of fire. To ensure the device safety, make all flammable
material far away from equipment, and auxiliary equipment producing sparks should not be placed
near equipment.
7. When the oil, grease, other fat substance and the compressed oxygen meet, it will self-ignite
seriously, therefore, try to avoid the oxygen pressure reducing valve/adjustment valve, valve for
oxygen cylinder, pipe, connector containing these substance.
8. Do not use combustible material like ether, alcohol etc. because once even a little ether, alcohol,
or other combustible material mixed with oxygen in the equipment, it’ll cause fire.

vi

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

9. When modulating the oxygen flux every time, please leave 30min. at least for the equipment
regaining the new oxygen concentration.
10. There’s a pressurize device filling with potassium hydroxide electrolyte installed inside the
oxygen sensor. If the sensor leaks, please stop using and chuck it. If the leaking electrolyte
touches skin or clothes, please wash with clean water immediately. If the leaking electrolyte
touches eyes, please wash eyes with clean water for 15 min. and keep them open, notify the
doctor immediately.
11. Seasonal check the gas and the oxygen transporting parts to see if they are eroded or broken.
12. Seasonal check the battery of oxygen sensor to if they are leaking or aging. Replace them if it
is necessary.
13. To operate the auxiliary oxygen equipment together with equipment, please refer to the
corresponding instruction manual.
14. Under 100% oxygen concentration, the oxygen sensor service life is up to 10000 hours.

WEIGHING PRECAUTIONS
1. The electrical scale installed in the bassinet must work after 30 minutes’ warm-up in the
equipment, that is to say the electrical scale can’t work until being put in the incubator which is set
the using temperature and begins working for at least 30min. Or else, the weighing accuracy will
surpass the regulated value.
2. Please keep the patient horizontal and in the center of the infant bed while weighing.
3. The maximum weighing weight of Infant scale is 8kg, please don’t over loading, or else the
scale will be damaged.
4. Lay the infant bed with infant scale gently when loading or unloading, don’t press surface of the
bassinet to avoid damage the weight sensor inside the infant bed.
5. The displayed weighing value is just for reference.

SEASONAL SAFETY CHECK


1. Please clean the plug of power cord at least once a year. Too much dust on plug may cause the
fire.
2. The life period of air sensor and air flow sensor is 8 years respectively. The air sensor should be
calibrated every half year, and only the authorized qualified service personnel can do that.
3. The life period of skin temperature sensor is 2 years. The skin temperature sensor should be
calibrated every half year, and only the authorized qualified service personnel can do that.
4. The following safety checks should be performed at least every 12 months by a qualified person
who has adequate training, knowledge, and practical experience to perform these tests. The data
should be recorded. If the equipment is not functioning properly or fails any of the above tests, the
equipment has to be repaired.

vii
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

①. Inspect the equipment and accessories for mechanical and functional damage.
②. Inspect the safety relevant labels for legibility.
③. Inspect the fuse to verify compliance with rated current and breaking characteristics.
④. Verify that the device functions properly as described in the instructions for use.
⑤. Test the protection earth resistance according IEC 60601-1:2005: Limit 0.1Ω.
⑥. Test the earth leakage current according IEC 60601-1:2005: Limit: NC 5mA, SFC: 10mA.
⑦. Test the enclosure leakage current according to IEC 60601-1:2005: Limit: NC 100μA,
SFC: 500μA.
⑧. Test the patient leakage current according IEC 60601-1:2005: Limit: for a.c.: 100μA (BF),
for d.c.: 10μA (BF).
⑨. Test the patient leakage current under single fault condition with mains voltage on the
applied part according IEC 60601-1:2005: Limit: for a.c.:500μA (BF), for d.c.: 50μA (BF).
⑩. According to the test methods of IEC 60601-1:2005, the patient leakage current (net
voltage should be added on the applied part) of the testing device must less than 5000μA.
⑪.Test the patient auxiliary leakage current according IEC 60601-1:2005: Limit: NC for a.c.:
100μA (BF), for d.c.: 10μA (BF).SFC for a.c.:500μA (BF), for d.c.: 50μA (BF).

TABLE OF DEFINITIONS AND SYMBOLS


TECHNICAL DEFINITIONS
SKIN TEMPERATURE SENSOR: A sensing device including the link with the equipment intended
to measure the infant’s skin temperature.
CONTROL TEMPERATURE: The temperature set at the temperature control.
TEMPERATURE ALARM CHECKOUT STATE: The difference between real temperature and
control temperature is within ±0.5℃ and such state lasts for over 10 minutes. The equipment must
stay in such state when check up the alarm about temperature.
LIFETIME OF PRODUCT: The period from sell-by date to the date of discarding as useless.
VHA STAND: Abbreviation of vertical height adjustment stand.

TECHNICAL DEFINITIONS UNDER THE INCUBATOR WORKING MODE


INCUBATOR TEMPERATURE: Air temperature at a point 10cm above and centered over the
mattress surface.
STEADY TEMPERATURE CONDITION: A condition which is reached when the temperature does
not vary by more than 1℃ over a period of 1 hour.
TEMPERATURE UNIFORMITY: The amount by which the average temperature at each of four
points 10cm above the mattress surface differs from the Average incubator Temperature at steady
Temperature Condition.

viii

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

TEMPERATURE RISING TIME: The time required for the Incubator Temperature to rise 11℃,
when the Air Control Temperature is at least 12℃ above ambient.
STEADY HUMIDITY CONDITION: A condition that the humidity error between displayed value
and controlling value in ±5%RH and keep this condition over than 1 minute.
AVERAGE INCUBATOR TEMPERATURE: The average of the maximum and minimum Incubator
temperatures achieved during Temperature Condition.

TECHNICAL DEFINITIONS UNDER THE WARMER WORKING MODE


PRE-WARM MODE: A mode of which can keep the mattress at a properly temperature. The
heater output will work automatically according to the set program.
MANUAL MODE: A mode of operation in which the heater output is either at a fixed level or a
proportion of its maximum output set by the operator.
SKIN TEMPERATURE MODE: A mode of operation in which the power output varies
automatically in response to the temperature of the baby, to achieve a temperature close to a
value set by the operator.
STEADY TEMPERATURE CONDITION: A condition which is reached when the temperature,
measured at the center of the TEST DEVICE positioned on the midpoint of the EQUIPMENT
mattress, does not vary by more than 1℃ over a period of 1 hour.
TEMPERATURE UNIFORMITY: The amount by which the average temperature at each of four
test devices above the mattress surface differs from the average temperature of test devices in the
middle of mattress at steady Temperature Condition.
APGAR TIMER: It offers the functions of three periods alarming indication: 1min, 5min, and 10min
for clinical treatment.

NOTE, IMPORTANT, CAUTION AND WARNING


NOTE: A note is inserted in text to point out procedures or conditions, which may otherwise be
misinterpreted or overlooked. A note may also be used to clarify apparently contradictory or
confusing situations.
IMPORTANT: Similar to a Note but be used where greater emphasis is required.
CAUTION: A caution is inserted in text to call attention of a procedure which, It not followed
exactly, can lead to damage or destruction of the equipment.
WARNING: A warning is inserted in text to call attention to dangerous or hazardous conditions
inherent to the operation, cleaning, and maintenance of the equipment which may result in
personal Injury or death of the operator or patient.

ix
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

SYMBOLS

General warning sign Loading prohibited

CLASS I Class I equipment Type BF Applied part

Main power on Main power off

on (only for a part of equipment) Off (only for a part of equipment)

Baby scale socket RS232 RS232 Connector

Sensor of air temperature, humidity,


Setting value indicator
oxygen

High water level Low water level

Caution, scald Date of manufacture

Canopy up key /height Canopy down key/ height


adjustment up key of whole unit adjustment down key of whole unit

110-120V~ /60Hz AC Power 110-120V/60Hz F 10AH/250V F Type Fuse 10AH/250V

SN Serial Number User-Defined key

Keyboard lock Silence /Reset Key

Keep away from the heat source Canopy operation indicator

Canopy operation indicator Refer to instruction manual

Assistant main power output,


Manufacturer 110-120V~ /60Hz MAX:1A
MAX:1A

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

TABLE OF CONTENTS
SECTION PAGE
1. GENERALINFORMATION…………..……………………………………………………..…..……1-1
1.1 INTRODUCTION……………………...…………………………………………….………….……1-1
1.2 PRODUCT CONTRAINDICATIONS………………………………………………………………1-1
1.3 INTENDED USE……………………………………………………………………………………1-1
1.4 COMPOSITION OF PRODUCTS…………………………………………………….……………1-1
1.5 DESCRIPTION………………………...………………………….…………………………………1-1
1.6 SPECIFICATIONS……………………………………………………...……………………………1-3

2. INSTALLATION……………………………………………………….…….…….……………..……2-1
2.1 GENERAL………………………………….………………………………..………………….……2-1
2.2 UNPACKING…………………………………………………………………………………………2-1
2.3 INSTALLATION………………………………………………………………………………………2-1

3. FUNCTIONAL DESCRIPTION ………………………………………………………...….…...…...3-1


3.1 GENERAL FUNCTIONAL DESCRIPTION………………..………………………………………3-1
3.2 TEMPERATURE CONTROL PRINCIPLE…………………………………………….....……..…3-2
3.3 DATA COMMUNICATION CONNECTOR………………………………………..………………..3-4
3.4 ALARMING AND SYSTEM INDICATION INFORMATION………………………………………3-4

4. OPERATION…………………………………..………………………………………………………4-1
4.1 GENERAL……….……………………………………………………………………………………4-1
4.2 POWER SUPPLY CONNECTION AND SWITCH CONTROL…………………………….….…4-1
4.3 THE CONTROLLER AND DISPLAY PANEL…………………………………...…………………4-2
4.4 OPERATIONAL CHECKOUT PROCEDURE………………………………………..…………4-9
4.5 GENERAL OPERATION PROCEDURE…………………………………………………………4-20

5. CLEANING AND MAINTENANCE………………………………….………………………………5-1


5.1 GENERAL……………………………………………………….……………………………………5-1
5.2 CLEANING……………………………...……………………………………………………………5-1
5.3 MAINTENANCE…………………….………………………………………………………………5-11
5.4 TROUBLE SHOOTING………………………………………………….………………...………5-11

6. OXYGEN CONCENTRATION CONTROL SYSTEM…………………………...…………………6-1


6.1 GENERAL………………………………………..…………………………………………..………6-1
6.2 REPLACE THE OXYGEN SENSOR………………………………………………..………...6-1

xi
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

6.3 FUNCTIONAL DESCRIPTION……………………………………………………………..………6-2


6.4 CONNECTION OF OXYGEN INPUT CONNECTOR……………………………..……………..6-2
6.5 OXYGEN SYSTEM OPERATIONAL CHECKOUT PROCEDURE……………………………..6-3
6.6 USING OF OXYGEN SUPPLY SYSTEM…………………………………………………..…..…6-4
6.7 MAINTENANCE…………………………………………………………………………….…….…6-7

7. HUMIDITY CONTROL SYSTEM……………………………………………………..….…7-1


7.1 GENERAL…………………………………………………………………………………………….7-1
7.2 FUNCTIONAL DESCRIPTION………………………….....……………………………………….7-1
7.3 OPERATION CALIBRATION PROCEDURES……………………………………..……………7-1
7.4 USING HUMIDITY CONTROL SYSTEM…………………………..……………………………7-2
7.5 MAINTENANCE………………………………………...…………………………………...………7-4

8. WEIGHT SYSTEM………………………………………………………………………………..…..8-1
8.1 GENERAL…………………………………………………………………………………………....8-1
8.2 INSTALLATION……………………………………………………………………………………....8-1
8.3 FUNCTION DESCRIPTION...……………………………………………………………………...8-2
8.4 OPERATION AND CALIBRATION PROCEDURES……………………….……………………8-2
8.5 OPERATION OF WEIGHING SYSTEM……………………………..…………………………….8-4
8.6 MAINTENANCE………………………………………..……………………………………………8-4

9. PARTS LIST……………………………………………..………………………………………….…9-1

NOTE: The product composition maybe different from this manual, but it does not affect
product function. Please understand.

xii

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

SECTION 1
GENERAL INFORMATION
1.1 INTRODUCTION
This manual provides instructions for installation, debugging, operation, cleaning and

maintenance of Infant Incubator (Refered as equipment). We are not responsible for the

malfunction which is caused due to not following the instruction on our manual, and we are also

not responsible for the running performance of the equipment which is not approved.

The operator should read and understand of the content of this manual.

This manual should be put together with the equipment so as to the client to check at any

moment.

Monitoring Tray and weighing system are the optional accessories for the product. If you

don’t purchase the optional part, please omit it.

1.2PRODUCT CONTRAINDICATIONS
It is not clear now.

1.3 INTENDED USE


The equipment is intended to provide a controlled environment for premature and invalidism
infants to grow up, also can provide an open type environment to nursing or saving the neonatal
infant’s life and adjust the infant’s temperature.

1.4 COMPOSITION OF PRODUCTS


The equipment consists of main body(includes infant compartment, crate, controller, I.V. Pole
and tray), sensor box,skin temperature sensor, box seat,RadiatorRadiation,Observation
Lamp,weighing device (Option available) and monitoring tray(Option available).

1.5 DESCRIPTION
The following diagram shows the main parts of theequipment.

1-1
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Controller Radiator Radiation

I.V. Pole
Observation Lamp

Tray
Infant Compartment

Monitoring Tray
Base

Mobile Cabinet

DESCRIPTION OF
EXPLANATION
PART

An important component at the top of infant compartment(In this


manual called CANOPY),comprise by heater and reflecting cover,
Radiator Radiation intended to provide the electromagnetic spectrum scope of infrared
radiation heat for patient on bed surface.
The lifetime of heater is about 10000 hours.

A kind of bearing part, it is used for hangingthe infusion bottle.


I.V. pole
The Max Load:2 Kg.The max HeightDuring Transport:142 cm.

The main part of the equipment, possesstemperature control, humidity


Controller control, oxygen concentration control system and Information display
function.

A kind of bearing part, it is used for putting some objects and auxiliary
Tray devices.
Max Load: 2 Kg.

Monitoring Tray A kind of bearing part, it is used for putting Infant monitor system.
(Option available) Max Load: 8Kg.

1-2
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

DESCRIPTION OF
EXPLANATION
PART

Aspace composed by 4 pieces of organic glassbaffle, canopy and


infant bed, it is used for placed the patient. Combine with the clinical
require, the bassinet can adjust the inclination angle and equipweighing
Infant Compartment
system for option part.
MaxLoad of infant bed: 10 Kg.
Max Load of weighing system mattress: 8kg.
An important part of the equipment, which composed of the heater,
Base radiator, humidifier, servo oxygen system, air circulation system, air
filter, and so on.

A illumination component of equipment, it is used for lighting during the


Observation Lamp
clinical operation.

A part which can support the main body of equipment. Cabinets are
Mobile Cabinet supplied in VHA stand which including 2 storage compartment.
Storage cover Max. load: 2 Kg.

The max. Load of tray and the other accessories is the value in the list, not
overloaded, so as not to damage to the accessories.

NOTE:Size of Infant Incubator: L1140mm×W660×H1515~L1140mm×W660×H2315.


Height of bed surface:880mm~1080mm.
Weight of Infant Incubator:160kg.

1.6 SPECIFICATIONS
This product’s classification as follows:

By the electric shock protection type classification: Type I equipment.

By the degree of shock proof classification:Type BF application part.

By the specified ofIEC60529 for liquid protection degree classification:IPX0.

By the manufacturers recommended disinfection and sterilization method classification:Use

neutral disinfection solvents or solution to clean, can use steam sterilization to clean the tank.

By the air mixer of flammable gas or with oxygen or nitrous oxide mixture of flammable

anaesthetic gas safety degree classification: It should not use in air mixer of flammablegas or

with oxygen or nitrous oxide mixture of flammable anaesthetic gas.

By operational mode classification:Continuous operation.

Specifications for the incubator are provided in table 1.1.

1-3
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

TABLE 1.1 SPECIFICATIONS

Power Requirements……………………………………………………….….…….AC110-120V, 60Hz

Power Input……………………………………………………….…………...…………………….1300VA

Auxiliary Mains Power Output……………………….…….….AC110-120V, 60Hz,Max. Current:1A

WorkingModes…………………………..………………………….…………………….Incubator Mode

Warmer Mode

Performance Index Under The Incubator Mode:

Temperature Control Modes……………….….….….…………..Air mode (Air temperature control)

Baby mode(Baby temperature control)

Air Temperature Control Range…………………………….…….….…………….…………25~37℃

37~39℃(overriding mode)

Skin Temperature Control Range………………………………………......…………………34~37℃

37~38℃(overriding mode)

Air Temperature and Skin Temperature Display Range………..………………….………5~65℃

Temperature Rise Time*(environment temperature is +22℃)………………......……...…≤40min

Difference Between Incubator Temperature and Average Incubator Temperature...........≤0.5℃

Difference Between Average Incubator Temperature andControl Temperature……...….≤1.0℃

Temperature Uniformity*(level mattress)…………………………….....................................≤0.8℃

Temperature Uniformity* (Tilt mattress)……………….………………………..…………….....≤1.0℃

Skin Temperature Sensor Precision……………………………...……………………….within±0.2℃

Humidity Control Precision……………..……………………………………………..…………±5%RH

Humidity DisplayPrecision…..…….…..…..…..…..…..…..…....…….…..….…………………±5%RH

Humidity Display Range…..………………………………………………...………………0%~99%RH

Humidity Control Range……………………………..……………………..………………0%~90%RH

Oxygen ConcentrationDisplay Range…………………………………………………0%O2~99%O2

Oxygen ConcentrationControlRange…………………………..……………...……20%O2~60%O2

Oxygen Concentration Display Precision………….…………±2%O2(The set value is below 25%)

±3%O2(The set value is above 25%)

Oxygen Concentration ControlPrecision…………………………..±4%O2(Volume concentration)


Weight Display Precision…………………………………………………………………...…within±1%

1-4
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

TABLE 1.1 SPECIFICATIONS(continued)

Performance Index Under The Warmer Mode:


Temperature Control Modes………………………………………….………………..Pre-warm mode*
Manual mode*
Baby mode* (Baby temperature control)
Temperature Control Range Under The Skin temp. Mode………………..………...34.5℃~37.5℃
Temperature Display Range Under The Skin temp. Mode…………..……………….……5℃~65℃
DifferenceBetween Skin Temperature Sensor and Control Temperature…………………≤0.5℃
Temperature Uniformity of Bed Surface…………………………………………….………..……≤2℃

ALARM(See 3.4 section)……………………..……………….……………………Power Failure Alarm

Fan Motor Alarm

Sensor Failure Alarm

Deviation Alarm

CanopyFailure Alarm

Over Temperature Alarm

Water tank placed false alarm

Water Shortage Alarm

Setting Alarm

Checking Alarm
System Alarm

ENVIRONMENT TEMP (Not to use in the environment exceed specified)


WorkingRange…………………………………..……………..……………………………...20℃~30℃
Transport and Storage Range……………………………………...……………..………-20℃~+55℃

RELATIVE HUMIDITY
WorkingRange…………………………………………………………………..………….….30%~75%
Transport and Storage Range……………………………………………………………….….…..≤93%

ATMOSPHERIC PRESSURE
Transport and Storage Atmospheric Pressure Range…………………….……500hPa~1060hPa
WorkingAtmospheric Pressure Range………………...…………………………800hPa~1060hPa

AIR FLOW RATE


Ambient Environment Air Movement Rate……………………………………………………<0.3m/s

1-5
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

TABLE 1.1 SPECIFICATIONS(continued)

Application environment altitude……………………………………..……………………≤2000m


Overvoltage category……………………………………..……………………………………………II
Pollution degree…………………………………………………………………………………………2

OTHER SPECIFICATION
Noise Inside The Hood Under Incubator Mode…………………………………………………………
Under the steady temperature condition≤50dB(A)
[Ambient noise: ≤40dB(A)]
Under other conditions≤55dB(A)
[Ambient noise: ≤45dB(A)]
Carbon dioxide(CO2) ConcentrationInside The Hood Under Incubator Mode……………….….
at a Point10cm above the center ofthe mattress
Less than 0.5% when a mixture CO2 is delivered at 750ml/min
Air Flow rate above MattressUnder Incubator Mode……………………………...………≤0.35m/s
APGAR Timer Under Warmer Mode*…………………………………..……Audible and visible alarm
when the device runs to 50″~1′, 4′50″~5′, 9′50″~10′

* Refer to Table of Definitions and Symbols

Note: Running under the incubator mode, if open the baffle or operation window, or using
infant bed, supplies or other equipment within incubator will alter the air flow pattern
that affect the temperature uniformity and variability. The temperature display on
incubator mutual relation with center mattress temperature and skin temperature.
Running under the warmer mode, if change the air flow rate inside the room or the
declination angle of bassinet, or useother heating device beside this equipment may
affect the temperature uniformity of mattress. If theambient environment air flow rate
over 0.3m/s, cause the patient thermal dissipation too quick and affectthermal
equilibrium.

1-6
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

SECTION 2
INSTALLATION
2.1 GENERAL
This section provides installation procedures about equipment.

2.2 UNPACKING
Generally, the equipment is usually packed into a wooden box. Take care when unpacking.
Do not damage the spare parts of the This section provides installation procedures about
equipment.

2.3 INSTALLATION
Required two professionals at least to installation the This section provides installation
procedures about equipment with spanners, the installation tools are wrench, internal hexagonal
wrench and cross screwdriver.
Installation steps:
A. Fixed the wheel

Restoration pedal 1
Wheel Lock

WHEEL UNLOCKED WHEEL LOCKED

NOTE: Depress the wheel lock catch to lock the wheel as the arrow 1, it can keep the
equipment in situ under this condition. Dial up the wheel lock catch or depress
restoration pedal to open the wheel lock, the equipment can move steadily under
this condition.

WARNING: To prevent the equipment from sliding when parked on an incline, the front
locking wheel for the VHA stand must be facing down the incline and locked.

FIGURE 2.1

B. Install the I.V. Pole


As the figure 2.2 shows,twisting the set screw at the fixed block of I.V. Pole towards the left,
and the dovetail groove is unlocked, put the buckle at the fixed block of I.V. Pole into the Card slot
of instrument stand (the two sides buckle should be put into the Card slot in sequence) and adjust
it at appropriate height, and then twist the set screw to make the dovetail groove locked.

2-1
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Card slot of instrument stand

Fixed block of I.V. Pole


Dovetail groove

Set screw

Dovetail groove unlocked Dovetail groove locked

NOTE: Don’t use the I.V. Pole as hand shank that may affect the firmness of I.V. Pole.

FIGURE 2.2

C. Install the Observation Lamp


As the figure 2.3 shows, turn left the set screw (on the fixed block of observation lamp) to
make the dovetail groove at the unlocked status, then put the buckle(on the fixed block of
observation lamp) into the Card slot of the instrument stand (the two sides buckle should be put
into the Card slot in sequence), and adjust it at appropriate height(the fixed block should be
installed above the mattress at least 15cm, the observation lamp must be kept at least 50cm with
patient), then twist the set screw to make the dovetail groove locked. The observation lamp
equipped with the flexible shank and axle, you can use the internal hexagonal wrench to twist the
screw on the axle and adjust the size of friction.

Card slot of instrument stand


Shank

Dovetail Groove
Axle

Screw

Fixed Block of Observation Lamp Set Screw

2-2

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

1. To prevent the eyes do not keep the patient under the lamplight for a long time.

2. The observation lamp will produce additional quantity of heat, so must keep the

safe distance between patient and lamp, and supervise the patient carefully.

3. Strictly prohibit place the lamp at the top of patient while adjust the lamp

support.

4. Strictly prohibit place the observation lamp under the radiant heater. That may

block the radiant energy for patient or destroy the observation lamp.

CAUSE: Move the observation lamp away while go up and down the canopy, avoid to
disturb the lifting of canopy and that may cause the canopy or observation lamp
damaged.

FIGURE 2.3

D. Install tray and monitoring tray (monitoring tray is optional, hereinafter referred to
as the tray)
As the figure 2.4.1 shows, twist the set screw on fixed block of tray towards the left to make
the dovetail groove at the unlocked state, then put the buckle into the card slot of instrument
stand (the two sides buckle should be put into the Card slot in sequence) , adjust it at appropriate
height, then twist the set screw to make the dovetail groove at the locked state. As the figure
2.4.2 shows, put the tray into the tray support.

Card slot of instrument stand Tray


Dovetail groove

Fixed block of tray Set screw Tray support

FIGURE 2.4.1 FIGUTR 2.4.2

CAUSE: Move the tray away while go up and go down the canopy, avoid to disturb the
lifting of canopy and cause the canopy damaged.

E. Install the handle


As the figure 2.5 shows, use the internal hexagonal wrench to screw off the pull rod and
screw on fixed block of handle , then insert the buckle into the card slot and adjust it at
appropriate height, twist the screw and fasten the handle.

2-3
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Fixed block of handle

Screw

Pull rod

NOTE: Make sure that two terminals of the handle at horizontal position after installed.
FIGURE 2.5

F. Install the tubing management arm


As the figure 2.6 shows, insert the tubing management arm into the D shape hole of bed
clapboard, the four holes can inserted arbitrarily.

Tubing Management Arm

D shape hole

NOTE: The tubing management arm can only be used for placing pipelines, shall not
placed other items over 200g.

FIGURE 2.6

G. Install the oxygen sensor


Refer to the section 6.

H. Install the weighing system


Refer to the section 8.

I. Insert the power cord


Connect the power cord into the socket of general power supply which located at the VHA
stand.
Check the equipment according to the instruction in section 4.3.
NOTE: The main power switch located at the VHA stand, please do not put the equipment
in the position which is difficult to operate the power switch.

2-4

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

SECTION 3
FUNCTIONAL DESCRIPTION
3.1 GENERAL FUNCTIONAL DESCRIPTION
The equipment combines the functions of incubator and warmer. It can transform the
incubator mode and warmer mode by controlling the up and down of canopy. It can be used as
the incubator when the canopy at the lowest location. It can be used as the warmer when the
canopy at the highest location.

3.1.1 FUNCTION OF INCUBATOR


While the equipment working as incubator: It has temperature control system (Air Mode

and Baby Mode), humidity control system, oxygen concentration control system and weighing

system.

Heat output control under the Air Mode: automatic control the output of heater

according to the temperature detected by the air temperature sensor, reference part A in

section 4.5.1.2.

Heat output control under the Baby Mode: automatic control the output of heater

according to the patient's skin temperature detected by the skin temperature sensor, reference

part B in section 4.5.1.2.

Oxygen input control: automatic control the switch of oxygen concentration valve

according to the oxygen concentration detected by the oxygen concentration sensor in

incubator, reference the section 6.


Tank humidifying output control: automatic control the humidifier according to the

humidity detected by the humidity sensor in incubator, reference the section 7.

Weighing system control: display the number on the Monitor according to the weight

detected by the infant scale sensor, and operate the functions, such as calibration and zero

setting, reference the section 8.

The controller provides measured value and set value of Air temperature, skin temperature,

humidity and oxygen concentration on the LCD Monitor. And heat proportion, weight measured

value and curve.

3.1.2 FUNCTION OF WARMER


While the equipment working as warmer: It has radiation temperature control system
(pre-warm mode, manual mode and baby mode), weighing system and timer function.
Heat output control under the Pre-warm Mode:The output of heater according to
warmer pre-warm procedure, reference part A in section 4.5.1.3.
3-1
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Heat output control under the Manual Mode: The output proportion of heater according
to the preset by operator and fixed heat proportion, uncontrolled by patient's skin temperature,
reference part B in section 4.5.1.3.
Heat output control under the Baby Mode:automatic control the output of heater
according to the skin temperature detected by skin temperature sensor, reference part C in
section 4.5.1.3.
Weighing system control: display the number on the Monitor according to the weight
detected by the infant scale sensor, and operate the functions like calibration and zero setting,
reference the section 8.
The controller provides measured value and set value of Air temperature, skin temperature,
humidity and oxygen concentration on the LCD Monitor. And heat proportion, weight measured
value and curve.
Except the component configuration in section 1.6. The equipment can provide the check
lamp, X-ray tray for clinical operation and four casters for moving the equipment (All wheels can
be locked).

3.1.3 TRANSFORM THE WORKING MODE BETWEEN INCUBATOR AND WARMER


While the equipment working as incubator, rise the canopy will automatic change into the
warmer mode. If the canopy is used as incubator under the baby mode before rising, the
canopy will keep the baby mode after rising. If the canopy is used as incubator and rise the
canopy while operating under the air mode, the canopy will be used as the warmer under the
manual mode.
While the equipment working as warmer, descend the canopy will automatic change into
the incubator mode. If the canopy is used as warmer under the baby mode before reducing, the
canopy will be used as the baby mode after reducing. If the canopy is used as warmer and
reduce the canopy while operating under the manual mode or pre-warm mode, the canopy will
be used as incubator under the air mode.

3.2 TEMPERATURE CONTROL PRINCIPLE


3.2.1 TEMPERATURE CONTROL PRINCIPLE UNDER INCUBATOR MODE
As the figure 3.1 shows, control the temperature, humidity and oxygen concentration of
inside box are achieved by the forced air circulation system.
When the oxygen control function is locked, after the outside air is filtered, the outside air
inhaled into the heated air circulation passage, heated by heater and flow through the
humidification device, then enter into the hood through the air outlet from two sides of bed shelf,
it can replenish the fresh air of incubator. Fan motor will form the hot air circulation system by
mixing and heating between air in hood and fresh air.
3-2
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

When turning on the oxygen control function, oxygen flow through the oxygen input
connector, then pass the electromagnetism valve, and inhaled into the heated air circulation
passage to replenish the oxygen.

FIGURE 3.1

3.2.2 TEMPERATURE CONTROL PRINCIPLE UNDER WARMER MODE


As the figure 3.2 shows the heat from the heater will radiate the surface of mattress
through the reflective cover in shape of parabola with high reflectivity. Then the heat could
reach to the patient.
Under the Manual Mode, the radiant heater outputs in the way of fixed proportion to make
the body temperature of patient resume; under the Baby Mode, the device will automatic adjust
the heat output proportion by comparing the skin temperature and the set temperature to keep
the heat balance.

NOTE: During the process of getting the radiation and heat, the heat loss in convection,
evaporation, radiation, conduction will affect the heat balance. Therefore, to
make the patient decrease the dissipation, it is necessary to use the device in the
environment of no rapid air flowing ,and cover waterproof velum (polyethylene
velum) the naked skin or to use the waterproof net to increase the humidity
around the skin to decrease the water evaporation.

3-3
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

FIGURE 3.2

3.3 DATA COMMUNICATION INTERFACE


The incubator is equipped with RS232 data communication interface which is used for data
terminal output.

1. If connect the auxiliary equipment on this interface, the assembly of ME SYSTEMS


and modifications during the actual service life require evaluation to the requirements
of IEC60601-1, clause 16.
2. Everyone should be responsible for the safety of the whole system requirements.
3. Only the equipment provided by our company can be connected with RS232 data
communication interface. must ensure the reliable connection when using.
4. The medical institutions maintenance department should be responsible for the
maintenance of data communication interface, and inspect the data communication
every year.
5. The connecting and using of the data communication must be performed by special
trained medical personnel, and the personnel should clear and definite the risk of data
communication.
6. Do not touch data communication interface and patient simultaneously.
7. If have any question, please contact with the agency or the after-sales service
department of our company.

3-4
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

3.4 ALARMING AND SYSTEM INDICATION INFORMATION


STATE
1. Alarm information
High priority: The most urgency information, red alarm light blinking, alarm sounds more than
65dB(A);
Five tones order alarm, ring twice, and every 2.5 seconds to repeat again. (Expect the power
failure alarm)

NOTE: The sound alarm indication of power failure alarm is different from the other high
priority alarm which sound source is a single buzzer.
Medium priority: Medium priorities information, yellow alarm light blinking, alarm sounds more
than 65dB (A);
Three tones order alarm, and every 7.5 seconds to repeat again.
Low priority: Low priority information, the yellow alarm light is continued on, alarm sounds
more than 65dB(A);
Two tones order alarm, and every 20 seconds to repeat again.
2. The alarm preference is arranged according to the alarm serial number, the bigger the serial
number is, the lower level is. When various failures appear, the alarm prompts according to the
priority, the sound is different too.

ALARM INTRODUCTION
Alarm
Alarm Alarm Alarm activation Control Heater Alarm
Alarm character delay
no. information conditions mode state level
time

Power failure Turn on the switch


Power failure High
1 alarm light is on, when without All Off <2s
alarm priority
sound alarm start power supply

Red alarm light


flashes, the
display screen Main engine
Internal System High
2 shows" Internal internal RAM or All Off <2s
of main MCU priority
System of main ALU failure
MCU ", sound
alarm start
Red alarm light
flashes, the
Communication
display screen
Communication failure between High
3 shows " All Off <14s
A Fault main engine and priority
Communication A
slave machine A
Fault ", sound
alarm start

3-5
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

ALARM INTRODUCTION(continued)
Alarm
Alarm Alarm Alarm activation Control Heater Alarm
Alarm character delay
no. information conditions mode state level
time
Red alarm light
flashes, the
Slave machine A
ROM of client display screen High
4 internal ROM All Off <2s
MCU A shows " ROM of priority
failure
client MCU A ",
sound alarm start
Red alarm light
flashes, the
display screen Slave machine A
Internal System High
5 shows" Internal internal RAM or All Off <2s
of client MCU A priority
System of client ALU failure
MCU A ", sound
alarm start
Red alarm light
flashes, the
Communication
display screen
Communication failure between High
6 shows" All Off <5s
B Fault slave machine A priority
Communication B
and B
Fault ", sound
alarm start
Red alarm light
flashes, the
Slave machine B
ROM of client display screen High
7 internal ROM All Off <2s
MCU B shows" ROM of priority
failure
client MCU B ",
sound alarm start
Red alarm light
flashes, the
display screen Slave machine B
Internal System High
8 shows" Internal internal RAM or All Off <2s
of client MCU B priority
System of client ALU failure
MCU B ", sound
alarm start
Red alarm light
flashes, the
Communication
display screen
Communication failure with Main High
9 shows" All Off <6s
C engine and Slave priority
Communication C
machine C
", sound alarm
start
Red alarm light
flashes, the
Main engine
display screen High
10 EEPROM Fault internal EEPROM All Off <2s
shows" EEPROM priority
failure
Fault ", sound
alarm start
3-6
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

ALARM INTRODUCTION(continued)
Alarm
Alarm Alarm Alarm activation Control Heater Alarm
Alarm character delay
no. information conditions mode state level
time
Red alarm light
flashes, the display
Controller internal High
11 SRAM Fault screen shows" All Off <9s
SRAM failure priority
SRAM Fault ", sound
alarm start
Red alarm light
flashes, the display AD converter in the
High
12 ADC 1 Fault screen shows" ADC sensor box All Off <34s
priority
1 Fault ", sound TLV2548 failure
alarm start
Red alarm light
Inside sensor box
flashes, the display
slave machine A High
13 ADC 2 Fault screen shows" ADC All Off <27s
(MS1210)internal priority
2 Fault ", sound
ADC failure
alarm start
Red alarm light
flashes, the display Inside the sensor
High
14 ADC 3 Fault screen shows" ADC box ADC TLV2544 All Off <33s
priority
3 Fault ", sound failure
alarm start
The location of
canopy is not in
place and not
Red alarm light
proceed go up and
flashes, the display
Canopy down less than 30 High
15 screen shows" All Off <4s
Fault seconds, or while priority
Canopy Fault ",
microswitch of
sound alarm start
canopy became
short circuit or
open circuit
The radiant door is
not complete High
Red alarm light Incubator Off <2s
closed under the priority
flashes, the display
Radiant incubator mode
16 screen shows"
Door Fault The radiant door is
Radiant Door Fault ",
not complete High
sound alarm start Warmer Off <2s
opened under the priority
warmer mode
Red alarm light
TEMP. flashes, the display The heater circuit
Heater screen shows" of for Incubator High
17 Incubator Off <27s
System TEMP. Heater temperature control priority
Fault System Fault ", system failure
sound alarm start

3-7
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

ALARM INTRODUCTION(continued)
Alarm
Alarm Alarm Alarm activation Control Heater Alarm
Alarm character delay
no. information conditions mode state level
time
Red alarm light
flashes, the The heater circuit of
Radiant
display screen for Warmer High
18 Heater Warmer Off <2s
shows" Radiant temperature control priority
System
Heater System ", system failure
sound alarm start
Red alarm light
Short-circuit, open
flashes, the
circuit or poor
Air Sensor display screen High
19 connection inside the Incubator Off <2s
Fault shows" Air Sensor priority
air temperature
Fault ", sound
sensor
alarm start
Red alarm light
Short-circuit, open
flashes, the
Isolated circuit or poor
display screen High
20 Sensor connection inside the Incubator Off <2s
shows" Isolated priority
Fault isolated
Sensor Fault ",
temperature sensor
sound alarm start
Red alarm light
The temperature
flashes, the
difference between
Air Sensor display screen High
21 air sensor and Incubator Off <117s
DIFF. Fault shows" Air Sensor priority
isolated sensor is
DIFF. Fault ",
above 0.8℃
sound alarm start
Red alarm light
Short-circuit, open
flashes, the
Air Flow circuit or poor
display screen High
22 Sensor connection inside the Incubator Off <3s
shows" Air Flow priority
Fault air flow temperature
Sensor Fault ",
sensor
sound alarm start
Red alarm light
Short-circuit, open
flashes, the
circuit or poor
Skin Sensor display screen Baby High
23 connection inside the Off <4s
1 Fault shows" Skin mode priority
skin temperature
Sensor 1 Fault ",
sensor 1
sound alarm start
Red alarm light
flashes, the
Internal isolated
Isolated display screen Warmer
probe of Skin Sensor High
24 Skin Sensor shows" Isolated baby Off <4s
become short-circuit priority
Fault Skin Sensor Fault mode
or open circuit
", sound alarm
start

Red alarm light


flashes, the
The difference
display screen Warmer
Skin Sensor between air sensor High
25 shows" Skin baby Off <60s
DIFF. Fault and isolated sensor priority
Sensor DIFF. mode
is above 0.8℃
Fault ", sound
alarm start

3-8
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

ALARM INTRODUCTION(continued)

Alarm
Alarm Alarm Alarm activation Control Heater Alarm
Alarm character delay
no. information conditions mode state level
time
Temperature of
Red alarm light incubator is not over
Air flashes, the display 38℃(set value is
High
Overheat screen shows" Air less than 37℃),or Incubator Off <6s
priority
Alarm Overheat Alarm ", not over 39.5℃(set
sound alarm start value is over than
26 37℃)
Red alarm light The temperature
flashes, the display measured by the
Skin Warmer
screen shows" skin temperature High
Overheat baby Off <6s
Skin Overheat sensor under the priority
Alarm mode
Alarm ", sound warmer mode is
alarm start over 38.5℃
Temperature
measured by airflow
sensor is over 57℃
Red alarm light High
(set value is less Incubator Off <15s
flashes, the display priority
Air flow than 37℃),or over
screen shows" Air
27 Overheat 60℃ (set value is
flow Overheat
Alarm over than 37℃)
Alarm ", sound
Temperature
alarm start
measured by air High
Warmer Off <15s
flow sensor is over priority
60℃
The temperature
measured by skin
Red alarm light sensor 1 is always
flashes, the display 2℃ lower than set
Skin Sensor Incubator
screen shows" temperature, the High
28 Position baby Off <110s
Skin Sensor temperature priority
Error mode
Position Error ", measured by air
sound alarm start sensor is lower than
set temperature
from 3.5°C to 4.5°C
Red alarm light
Motor stopped
flashes, the display
running or revolving High
29 Motor Fault screen shows" Incubator Off <11s
speed lower than priority
Motor Fault ",
800r/m
sound alarm start
Display temperature
Incubator High
is 3℃ higher than Off <2s
air mode priority
set temperature
Display temperature
Red alarm light Incubator High
is 3℃ lower than On <2s
flashes, the display air mode priority
Temp. set temperature
screen shows" Display temperature
30 Deviation
Temp. Deviation of sensor 1 is 1℃ High
Alarm baby Off <2s
Alarm ", sound higher than set priority
alarm start temperature
Display temperature
of sensor 1 is 1℃ High
baby On <2s
lower than set priority
temperature

3-9
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

ALARM INTRODUCTION(continued)

Alarm
Alarm Alarm Alarm activation Control Heater Alarm
Alarm character delay
no. information conditions mode state level
time
The temperature
Red alarm light
measured by skin
flashes, the display
sensor is lower than Warmer
Check Skin screen shows" High
31 set temperature baby Off <118s
Sensor A Check Skin Sensor priority
3.5℃ above, and mode
A ", sound alarm
the state will keep
start
for 2mins
When temperature
stable state, if the
temperature is out of
Red alarm light
the stable state over
flashes, the display
0.4℃by accident or Warmer
Check Skin screen shows" High
32 unreasonable action baby Off <131s
Sensor C Check Skin Sensor priority
and do not back to mode
C ", sound alarm
the stable state in
start
3mins and do not
occur the deviation
alarm.
Red alarm light
flashes, the display
Check Warmer
screen shows" Appears every 15 High
33 Manual manual Off <14’16”
Check Manual minutes priority
Mode mode
Mode ", sound
alarm start
Red alarm light
Short-circuit,
flashes, the display
O2 Sensor open-circuit inside High
34 screen shows" O2 Incubator On <4s
1 Fault the oxygen sensor 1 priority
Sensor 1 Fault ",
or own fault
sound alarm start
Red alarm light
Short-circuit,
flashes, the display
O2 Sensor open-circuit inside High
35 screen shows" O2 Incubator On <4s
2 Fault the oxygen sensor 2 priority
Sensor 2 Fault ",
or own fault
sound alarm start
Red alarm light
The difference
flashes, the display
O2 Sensor between oxygen 1 High
36 screen shows" O2 Incubator On <2s
DIFF. Fault and oxygen 2 is priority
Sensor DIFF. Fault
3%O2
", sound alarm start
Oxygen display
Red alarm light value is 5%O2 High
Incubator On <2s
flashes, the display higher than set priority
O2 Sensor
screen shows" O2 value
37 Deviation
Sensor Deviation
Alarm Oxygen display
Alarm ", sound High
value is 5%O2 lower Incubator On <2s
alarm start priority
than set value

3-10
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

ALARM INTRODUCTION(continued)

Alarm
Alarm Alarm Alarm activation Control Heater Alarm
Alarm character delay
no. information conditions mode state level
time
Red alarm light
flashes, the
Internal RTC of
display screen High
38 RTC Fault system failure or All Off <6s
shows" RTC Fault priority
accuracy is not good
", sound alarm
start
Yellow alarm light
flashes, the
HUM. The heater circuit
display screen
Heater inside the humidity Medium
39 shows" HUM. Incubator Off <42s
System control system is priority
Heater System
Fault fault
Fault ", sound
alarm start
Yellow alarm light
flashes, the Short-circuit,
HUM.
display screen open-circuit inside Medium
40 Sensor Incubator On <2s
shows" HUM. the humidity sensor priority
Fault
Sensor Fault ", or own fault
sound alarm start
Yellow alarm light Humidity display
flashes, the value is 15%RH Medium
HUM. Incubator On <2s
display screen higher than set priority
Sensor
41 shows" HUM. value
Deviation
Sensor Deviation Humidity display
Alarm Medium
Alarm ", sound value is 15%RH Incubator On <2s
priority
alarm start lower than set value
Yellow alarm light
Disconnection,
flashes, the
open-circuit,
Battery display screen Medium
short-circuit inside All Off <30s
Fault shows" Battery priority
42 the controller
Fault ", sound
rechargeable battery
alarm start
*NOTE: Within the first 5min after starting up, system doesn’t check the battery failure, only
charging the battery.
Yellow alarm light
flashes, the
Panel key
Keyboard display screen Medium
43 short-circuit fault All Off <10s
Fault shows" Keyboard priority
occurred
Fault ", sound
alarm start
Yellow alarm light
is continued on,
Tank the display screen Bad connection
Low
44 Position shows" Tank between water tank Incubator On <4’2”
priority
Error Position Error ", and base
sound alarm start
after 4mins
Yellow alarm light
is continued on, he
Water display screen Water tank is lack of Low
45 Incubator On <14s
Shortage shows" Water water priority
Shortage ", sound
alarm start
3-11
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

System fault alarm introduce


When system failure appears, the alarm light flashes, the display screen shows alarm
information and the sound alarm start. While system failure appears means the warmer can't
work as normal, please stop using the equipment at once and invite the authorized
maintenance personnel to maintain.

NOTE: 1. While system failure appears, must stop using the incubator at once and
invite authorized maintenance personnel to maintain.
2. All the above alarms except the deviation alarm and skin overheat alarm in
skin mode are physiological alarm status, the power failure alarm is other alarm
status; the others are all technology alarm status.
3. Except for the power failure alarm, the other alarms are all can silence by
pressing silence/reset key, the time for silence is 4mins(the silence time of
oxygen concentration alarm is about 115s). When the silence time is over, if the
alarm condition is still not solved, the alarm will activate again. If occur multiple
alarm at the same time, the device will give an alarm firstly for the higher grade.
Press the twice of silence /reset key can cancel the alarm state then the
equipment back to the set condition to monitor the alarm.
4. The power failure alarm lasts at least 10 minutes, if the power supply recover
before the alarm, the device will back to the alarm setting before the power
failure.
5. The alarm system will save all the alarm status logs automatically, and the
saved log will not change even the equipment is power failure.
6. While checking alarm appears, if don’t press silence /reset key, the heating
output proportion will volitional check to 30%. This kind of alarm doesn’t mean
the device get fault, only for remind the operator to check and supervise the
body temperature for patient.

WARNING: When using the equipment in any independent place, if use different
alarm preset, there will be the potential risk.

3-12
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

SECTION 4
OPERATION
4.1 GENERAL
This section provides the operating procedures for the equipment.

4.2 POWER SUPPLY CONNECTION AND SWITCH CONTROL


As shown in figure 4.1, the power supply connections and the power switch located under
the power supply box, RS232 connector is located on the back of the power supply box. The
control switch is located below the controller, the inside of the back panel.

Socket of power output Socket of power input Main power switch

Controller switch

Enlarge figure A Enlarge figure A

RS232 Connector

FIGURE 4.1

4-1
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

4.3 THE CONTROLLER AND DISPLAY PANEL


The key operation of this equipment is made of membrane button and rotate button.

4.3.1 CONTROLLER AND DISPLAY PANEL UNDER INCUBATOR MODE

FIRST INDICATION INTERFACE


AIR, SKIN, OXYGEN CONCENTRATION AND HUMIDITY
1 14 3 13 12 2 6 15 5 11

>37℃ Power

2015-03-24 10:10:00
SKIN

37.0 ℃ 45
SKIN(℃) TIMES(12h) AIR

SET --.- SKIN


40
HEATER 35 HUM.

30 O2
25
0 t
AIR
25.0 ℃ HUM. % O2 % WEIGHT g
3
2
.
0

SET
○ 22 ○ 21 1000
SET 0 SET 20

10 4 9 8 16 7

1. Display area of Air 5. Self-definite function key 9. Display area of weigh


temperature information On first indication interface information
The above display the designates air, skin, humidity, Display the weight on the
temperature of the air and oxygen concentration. baby scale.
temperature sensor probe, Press the air key to enter into
the below display the air the second interface A, Press 10. Display area of humidity
temperature setting value. the skin key to enter into the Information
When the equipment is second interface B, press the The above display the
running at baby mode, it humidity key to enter into the humidity of hood, the below
display“--.-”. second interface C, press the display the humidity setting
O2 function key to enter into value. The left icon display
2. Alarm indicator light the second interface D. As for humidity control condition.
Appear when the alarm the details, please refer to the As for the details. Refer the
equipment, the light is on. relevant instruction. section 7.
When this light 6. Display area of daily 11. Alarm silence/reset key
flashes, please stop information
using and consult This key has two kinds of
The above displays current functions, when occur
the description of
system’s date and time,>37℃ alarm, press once to stop
section 3.4.
indicate information, the the sound, press twice to
internal battery condition, and reset alarm condition and
alarm information. the equipment go back the
monitor condition again.

4-2

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

3. Display area skin temperature 7. Rotate button 12. Mode display icon
information This key has two kinds of It displays the current
The above display the functions. Press the down temperature control mode
temperature of skin temperature key to activate or lock the is air temperature or skin
sensor probe, when the skin keyboard, rotate right or temperature.
temperature sensor disconnect left to increase or
with the equipment, it decrease the setting 13. Power failure alarm
display“--.-”;the below display the value. indicator light
air temperature setting value. When the equipment
When the equipment is running at 8. Keyboard lock indication power failure, the light is
air mode, it display“--.-”. icon on.
When it is unlock, the
4. Display area of oxygen system can enter into the 14. >37℃ indicator light
concentration information second interface When the equipment is
The above display the oxygen condition, all the function running on >37℃
concentration in the hood, the has been activated, when condition, the light is on.
below display the oxygen it is locked, the system
concentration setting value, the cannot be setting and all 15. Curve display area
left icon display the oxygen the function has been
concentration control condition, locked, the equipment Display the current
As for the details, please refer to enter into control setting's curve and
the section6 condition. period.

FIGURE 4. 2A

SECOND INDICATION INTERFACE A


AIR TEMPERATURE INTERFACE

2015-03-24 10:10:00
SKIN

37.0 ℃
1
5

SKIN(℃) TIMES(12h) SET


2
0

45
>37℃
SET --.-
40
1
7

HEATER 35
TREND
1
6

30 HOME

25
0 t
AIR
25.0 ℃ HUM. % O2 % WEIGHT g
3
2
.
0

SET

23 ○
21 1000
SET 0 SET 20

4-3
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

SECOND INDICATION INTERFACE B


SKIN TEMPERATURE INTERFACE

2015-03-24 10:10:00
SKIN

37.0 ℃

2
1
SKIN(℃) TIMES(12h) SET

3
7
.
0

2
0
45
>37℃
SET
40

1
9
HEATER 35
TREND

1
8
30 HOME

25
0 t
AIR
25.0 ℃ HUM. %
○ 22
O2
○ 21
% WEIGHT g
1000
SET --.- SET SET 20
0

SECOND INDICATION INTERFACE C


HUMIDITY INTERFACE

2015-03-24 10:10:00
SKIN

37.0 ℃
2
5
SKIN(℃) TIMES(12h) SET
3
7
.
0

2
4
45
SET WEIGHT

40
2
3
HEATER 35
TREND
2
2

30 HOME

25
0 t
AIR
25.0 ℃ HUM. %
○ 22
O2
○ 21
% WEIGHT g
1000
SET --.- SET 0 SET 20

SECOND INDICATION INTERFACE D


OXYGEN CONCENTRATION INTERFACE

2015-03-24 10:10:00
SKIN

37.0 ℃
2
9

SKIN(℃) TIMES(12h) SET


3
7
.
0

2
8

45
SET ON/OFF

40
2
7

HEATER 35
TREND
2
6

30 HOME

25
0 t
AIR
25.0 ℃ HUM. %
○ 22
O2
○ 21
% WEIGHT g
1000
SET --.- SET 0 SET 20

4-4

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

20. Self-definite function key- >37℃ 22. Self-definite function 26. Self-definite function
indicator key key- setting exit key: key-exit indicator key:
In the second indicator interface A, In the second In the second indicator
the key is for >37℃ temperature indicator interface C, interface D, this key is as
override mode function key. this key is as exit the exit the oxygen
humidity interface concentration interface
17/19. Self-definite function key- function key. function key.
temperature control curve
display: 23. Self-definite function 27. Self-definite function
In the second indicator key-humidity control key-oxygen concentration
interface A, this key is as air curve display: curve display key:
temperature curve display key, In the second In the second indicator
in the second indicator indicator interface C, interface D, this key is as
interface B, this key is as skin this key is for oxygen concentration
temperature curve display key. humidity curve display key.
display key.
16/18. Self-definite function key - 28. Self-definite function
temperature mode enter and 24. Self-definite function key-oxygen concentration
exit return key: key-weigh display: switch indicator key:
In the second indicator In the second In the second indicator
interface A, this key is as air indicator interface C, interface D, this key is as
temperature mode to choice press this key to oxygen concentration
enter and exit, in the second enter weigh interface. switch key.
indicator interface B, this key is
as skin temperature to choice 25. Self-definite function 29. Self-definite function
enter and exit. key- humidity setting key-oxygen concentration
choice key: setting indicator Key:
15/21. Self-definite function key- In the second In the second indicator
temperature setting choice key: indicator interface C, interface D, this key is as
this key is as oxygen concentration
In the second indicator humidity setting setting key.
interface A, this key is as air choice function key.
temperature setting choice
function, In the second
indicator interface B, this key is
as skin temperature setting
choice function

FIGURE 4.2B

THIRD INDICATION INTERFACE A


WEIGHT INTERFACE

2015-03-24 10:10:00
SKIN

37.0 ℃
3
3

SKIN(℃) TIMES(12h) TREND


3
7
.
0

3
2

45
SET CAL.

40
3
1

HEATER 35
ZERO
3
0

30 HOME

25
0 t
AIR
25.0 ℃ HUM. %
○ 22
O2

% WEIGHT g
21 1000
SET --.- SET 0 SET 20

4-5
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

30. Self-definite function key-exit indicator key: 31. Self-definite function key-weight reset
In the third interface A, this key is as exit indicator key:
weigh interface function key. In the third interface A, this key is as weight
reset key.
32. Self-definite function key-calibration
indicator key: 33. Self-definite function key-weigh curve
In the third interface A, this key is as weight indicator key:
system calibration function key. In the third interface A, this key is as weigh
curve indicator key.

THIRD INDICATION INTERFACE B


CURVE SETTING INTERFACE

2015-03-24 10:10:00
SKIN

37.0 ℃

3
6
WEIGHT(kg) TIMES(1 week) times
3
7
.
0

3
5
45
SET STORE

40
HEATER 35

3
4
30 HOME

25
0 t
AIR
25.0 ℃ HUM. %
○ 21
O2
○ 21
% WEIGHT g
1000
SET --.- SET 0 SET 20

34. Self-definite function key- exit indicator key: 35. Self-definite function key-save indicator
In the third interface B, this key is as exit the key:
curve interface function key. In the third interface B, press this key can
save the curve trend data again according
36. Self-definite function key-cycle indicator to the current time cycle, and enter the
key: display enter interface:
Press this key continuous to choice time Press “YES” and back to the previous
interval of the curve display, the curve interface, the controller can eliminate
display area is 2,4,8,12 and 24 hours trend original storage record to start new storage.
figure. Press “No” and back to the previous
interface, the controller continues to store
record on the original recording.

FIGURE 4. 2C

4-6

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

4.3.2 CONTROLLER AND DISPLAY PANEL UNDER WARMER MODE

1 3 4 5
2015-03-24 10:10:00
SKIN

37.0 ℃ SKIN(℃)

45
TIMES(24h) PRE-WARM

MANUAL
SET --.-
40
HEATER 35
SKIN

30 APRAR

25
0 t
MODE MANUAL
1
:
1
6
APGAR WEIGHT g

1000

2 6 7
1. Display area of skin temperature 4. Curve display area: 6. Timer display area:
information: Display the current setting's Display the time under the
The above display the curve and period APGAR timer mode, the
temperature of skin temperature unit use second.
sensor probe, when the skin 5. Self-definite function:
temperature sensor disconnect At the first indication 7. Weigh display area:
with the equipment, it display interface, can select Display the weight on the
“--.-”;the below display the skin parameter display like baby scale.
temperature setting value. When pre-warm, manual, baby, or
the equipment is running at air timer according to the
mode, it display“--.-”. prompt on monitor. Press
pre-warm function keys
2. Mode display icon: select Pre-warm Mode.
Display the current temperature Press manual function keys
mode is pre-warm, manual select Manual Mode, can set
control, or baby mode. the ratio of heating output
Press baby function keys
3. Display area of normal enter into the second
information: indication interface A, Press
Display the system date and timer function keys enter into
time, the battery capacity the APGAR timer condition.
condition and the relevant alarm
information.

WARNING: Operate the controller must observe strictly the described in this Manual,
don’t press any key optionally on the panel.

FIGURE 4. 2D

4-7
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Servo control oxygen input port under the power box behind the equipment

FIGURE 4.3

Socket of weight sensor

Socket of skin temperature sensor

Skin temperature sensor socket and weight sensor socket located under the controller,
inside the back shield.

FIGURE 4.4A

Air temperature sensor Humidity sensor Oxygen concentration sensor

Sensor box located on the back shield

WARNING: To prevent accidental touch of the damage, while moving the patient, the
patient should maintain a certain distance from the sensor.

FIGURE 4.4B

4-8

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Down button Up button

The VHA stand control

NOTE: The VHA Stand is only for INTERMITTENT OPERATION, after finished a up or down,
it must wait 30 seconds before operation.

Canopy foot treadle Canopy Lifting control button

Canopy foot treadle and canopy control button located at the bottom of VHA stand. The
canopy lifting control button is near the stand column, beside the infant bed.
While need to rise the canopy, you should tread the foot treadle or press the up button, if
loosen the foot treadle or up button, the canopy will stop rising. While need to decline the canopy,
you should press the down button, if loosen the down button, the canopy will stop declining.

NOTE: The foot treadle only use for rise the canopy. If you need to decline the canopy, you
must press the down button.

FIGURE 4.5

4.4 OPERATIONAL CHECKOUT PROCEDURE

WARNING
1. Please stop using this equipment once some functions lost or the spare parts for
fixing the front panel are loose.
2. Set temperature must be 3℃ higher than ambient temperature. And then you can
proceed this checkout procedure. Do not perform calibration procedures before use,
when the patient is in the device.
3. Please do not damage the VHA stand during moving. And lower the cabinet to the
lowest position before moving so that the equipment is stable.

Equipment should be only operated by trained personnel who knows familiar about the
general risk of operating the incubator and under the instructions of medical practitioner.

4-9
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Please do the following checkout procedure each time before operation. If the does not meet
any standards of the verification procedures, you must stop using and need maintenance.
The operator should operate the equipment within 20cm in front of the device, and the
specific distance between the equipment and operator with the operation comfort level.

4.4.1 CHECK THE INTEGRITY OF INCUBATOR


 Make sure that the equipment has been disinfected;

 Make sure that the corner cover of canopy is accurate fit with sealing gasket;

 Make sure that there is no crack or sharp edge on the organic glass;

 Make sure that the I.V. Pole and tray are locked firmly;

 Make sure that the inclination angle of infant bed is appropriate and the location is locked;

 Make sure that the necessary accessories and other devices are available;

 Make sure that the power cord is connected and safety of installation, strictly prohibit place

the power cord on the organic glass.

 Make sure that the wheels are installed well.

Check whether the caster can drop when lifting the incubator 2cm above the ground. The

drop of the caster will cause danger when the incubator crossed the cable during transporting,

threshold or elevator threshold. Before replacing the loose casters, please do not use the

equipment.

4.4.2 CHECK THE INCUBATOR MODE

WARNING
1. To avoid the risk of electric shock, this equipment must only be connected to supply
mains with protective earth.
2. Make sure that the power supply is compatible with the electrical specifications label
on the incubator. Our company’s power cord is 3 meters long. For safety, the equipment
can’t use the extension power cord.

A. START-UP THE CONTROLLER

In the case of power has supplied, turn on the main power switch and controller switch the

can make a sound like “Di…” and appear the startup picture, then enter into system

self-inspection. Pass the self-inspection direct access the operation interface. If pass the

self-inspection failed, the controller need maintain.

4-10

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

B. CHECK THE POWER FAILURE ALARM

Disconnect the power supply of whole unit; the equipment should give a power failure alarm,
the power failure alarm light is on, red alarm light blink and the equipment gives continuous alarm
sound.
This operation is used for checking if the power failure is normal or not. Insert the power
cord again after finishing checking.
IMPORTANT: Make sure that the rechargeable battery is full charged before usage. If not
full, it may cause the power failure without the alarming indication. If full, and
there is no any indication after disconnecting the main power supply, please
refer to the qualified service personnel.

C. CHECK HEAT FUNCTION

Control the environment temperature at 21℃ ~ 26℃, and choose the Air Mode, and set the
temperature at 33.0℃, and all heat power indicator are on, and heater will output heat completely.
NOTE: When the equipment works under the set state, and continue the following
operation procedure.

D. CHECK THE ACCURACY OF TEMPERATURE CONTROL

Choice the Air Mode, and set the air temperature at 36℃, after the air temperature enters
into the STEADY TEMPERATURE CONDITION, put the calibrated temperature measuring
equipment on the position above 10cm from the center of mattress to measure the air
temperature, compared with the indicated air temperature to check whether the deviation
between them is within 0.8℃.

E. CHECK SKIN TEMPERATURE MODE

Insert the skin temperature sensor into the socket of skin sensor, the arrow sign on the plug
should aim to the opening on the sensor socket so that the sensor is inserted correctly.

Arrow Gap

Socket of skin temperature sensor

1. Insert or pull out the skin sensor, you must hold the plug of skin sensor,
pulling the leads is forbidden.
2. Please do not bend the connection of sensor.

4-11
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

When the Air Mode works stability change it into the Skin temperature Mode, set temperature
at 37℃. Keep the temperature detected by skin temperature sensor is 2℃ lower than the set
temperature while making the displayed air temperature 4℃ lower than the set skin temperature.
After a while the equipment can give a high priority alarm sound, “Skin Sensor Position Error”
information on the screen, the alarm character should be consistent with the description of
section 3.4.
Disconnect the connection of skin temperature sensor. The equipment should give a high
priority alarm sound, “Skin Sensor 1 Fault” information on the screen, and the alarm character
should be consistent with the description of section 3.4.

F. CHECK PRECISION OF SKIN TEMPERATURE SENSOR

Put the skin temperature sensor with the mercury thermometer for the accuracy with ±0.1℃
into the water cup with warm water. Make the probe of the skin temperature sensor and the
mercury ball as closely as possible and stir enough, then read the value of mercury thermometer.
Compare the value of the skin temperature sensor and the mercury thermometer, and the
deviation must be within± 0.2℃.

NOTE: Please check again if the accuracy of the skin temperature sensor exceeds the
permissible deviation. Please let the professional maintenance man service
machine if the accuracy the skin temperature sensor exceeds the permissible of
deviation again.

G. CHECK FAN MOTOR ALARM

Refer to section 4.4.4 step B to turn down the organic glass, then refer to the steps in
section 5.2.1 to take out the mattress, subplate, infant bed and tray. As the follow figure, hold the
Motor Fan on hand carefully, turn on the switch of the controller power, the equipment should give
a high priority alarm sound, “Motor Fault” information will indication on the screen, the alarm
character should be consistent with the description of section 3.4. Press the Silence/Reset key,
alarm is over and the Controller return to operation status.

NOTE: Must take at least 45 minutes after the equipment stopped working to perform this
operation.

4-12

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

H. CHECK OVERHEAT ALARM

While the equipment under the working state, press the air temperature self-definite function

key and turn the rotate key to the right at the same time as figure 4.2. All heating indicators are on,

if the set temperature is no indication means the controller has entered into the over-temp test

state. After a while, the equipment should make a high priority alarm sound, “Air Overheat

Alarm” information will indication on the display screen and the alarm character should be

consistent with the description in section 3.4.

Equipment under the over-heating state and the sitting temperature is over 37℃, press the

air temperature self-definite function key and turn the rotate key to the right at the same time. All

heating indicators are on, if the set temperature is no indication means the controller has entered

into the over-temp test state. After a while, the equipment should make a high priority alarm

sound, “Air Overheat Alarm” information will indication on the display screen and the alarm

character should be consistent with the description in section 3.4.

Choice the skin temperature mode, press the air temperature self-definite function key and

turn the rotate key to the right at the same time as figure 4.2. All heating indicators are on, if the

set temperature is no indication means the controller has entered into the over-temp test state.

After a while, the equipment should make a high priority alarm sound, “Air Overheat Alarm”

information will indication on the display screen and the alarm character should be consistent with

the description in section 3.4.

I. CHECK DEVIATION ALARM

In Air Mode, close all doors and windows, set the temperature at 32℃. Enter

TEMPERATURE ALARM CHECKOUT STATE(refer to descriptions of terms and symbols

concerned), fan into the hot air inside the hood. When the air temperature indicates 35.1℃, the

equipment can give a high priority alarm sound, “Temp. Deviation Alarm” information will

indication on the screen, and the alarm character should be consistent with the description of

section 3.4. Set the temperature at 35℃, after the equipment enters into the temperature

alarming checkout state, open the front access panel, when the air temperature indicates 31.9℃,

the equipment can give a high priority alarm sound, “Temp. Deviation Alarm” information will

indicator on the screen, and the alarm character should be consistent with the description of

section 3.4.

NOTE: If the system does not enter into the TEMPERATURE ALARM CHECKOUT
CONDITION or the air Temperature does vary within ±3℃ than the setting
temperature, the deviation alarm can’t occur.
4-13
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

In Baby Mode, set the temperature at 35℃. Enter TEMPERATURE ALARM CHECKOUT

STATE, put the skin sensor into the 37℃ water cup. When the skin temperature indicates 36.1℃,

the equipment can give a high priority alarm sound, “Temp. Deviation Alarm” information will

indication on the screen, and the alarm character should be consistent with the description of

section 3.4; Set the temperature at 35℃, after the equipment enters into the temperature

alarming checkout state, put the skin sensor into the 33℃ water cup, when the skin temperature

indicates 33.9℃, the equipment can give a high priority alarm sound, “Temp. Deviation Alarm”

information will indication on the screen, and the alarm character should be consistent with the

description of section 3.4.


NOTE: If the system does not enter into TEMPERATURE ALARM CHECKOUT STATE or the
skin temperature does vary within ±1℃ than the setting temperature, the
deviation alarm can’t occur.

J. CHECK THE WATER SHORTAGE ALARM

Equipment working under humidifying system, pour out the water in the water tank. The

equipment can give a low priority alarm sound, “Water Shortage” information indicates on the

screen, and the alarm character should be consistent with the description of section 3.4.

K. CHECK THE TANK POSITION WRONG ALARM

When the equipment is in the working state, pulling out the water tank, the equipment should

give a low priority alarm sound, the yellow alarm light continue on, “Tank position error”

information will indicate on the screen, and the alarm character should be consistent with the

description of section 3.4. The water tank is still not in position after 4 minutes, the alarm sound

start immediately.

L. CHECK THE TIME

When the controller is normally working, the message shall display the current real time. If

the displayed time is not correct, please reset the time. The setting method is as following:

Press the first self-definite function key (from top to button) under the controller closed

condition, and switch on the controller, and the controller can enter into the interface in follow

figure to set the recording time of controller.The time can be indicated year (four figures, the

former two are the century value, the rest are the year), month, data, week, hour, minute, second.

4-14

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

SELECT

REAL TIME CLOCK:


STORE

SET
2015-3-24,TUE,10:13:39
EXIT

Press set self-definite function key, the underlined century value can indicate, and then press

the select self-definite function key, the underline can move with the time indication, press the left

key or the right key can decrease or increase the value on underlined, after correction, press the

store key to change the time and record the time, press Exit key, the controller can enter into the

first interface.

4.4.3 CHECK UNDER WARMER MODE

A. START-UP THE CONTROLLER

In the case of power has supplied, turn on the main power switch and controller switch, the

equipment can make a sound like “Di…” and appear the startup picture, then enter into system

self-inspection. Pass the self-inspection direct access the operation interface. If pass the

self-inspection failed, the controller need maintain.

B. CHECK HEAT FUNCTION

Control the environment temperature at 21℃ ~ 25℃, and set the skin temperature setting

value at 35.0℃, all the heat power indicator lamp are on and heater should power output

completely.

NOTE: Running the equipment under the setting status, and continue the following
operation procedure.

C. CHECK APGAR TIMER

When the controller under the main interface and the keyboard is unlocked, press the

Self-definite function key, then choice APGAR. the top right corner of display screen will indicate

the timer from 0s, when the time indicates 50”~1’, 4’50”~5’, 9’50”~10’, the equipment can

sound like “Du…”, the time of APGAR will blink in display screen at the same time.

4-15
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

D. CHECK SENSOR ALARM FUNCTION

Pull off the skin temperature sensor in Skin Temperature Mode, The equipment should make

a high priority alarm sound and the normal information display area will show “Skin Sensor 1

Fault”. The alarm character should be consistent with the description in section 3.4.

Gap

Sensor Socket Arrow


When insert the sensor, make sure the arrow sign align with the gap place.

E. CHECK SETTING ALARM FUNCTION

In Skin Temperature Mode, put the skin temperature sensor into the water cup at 3.5℃ for

2mins which temperature is lower than the skin set temperature, the equipment should give a

high priority alarm sound. Normal information display area shows " Check Skin Sensor A". The

alarm character should be consistent with the description of section 3.4.

Set the temperature at 36℃, after the equipment enters into the TEMPERATURE STEADY

STATE, put the skin sensor into the water cup at 35.3℃±0.1℃ for 3min, the equipment should

give a high priority alarm sound. Normal information display area shows " Check Skin Sensor C".

The alarm character should be consistent with the description of section 3.4.

F. DEVIATION ALARM FUNCTION

In Skin temperature Mode, set the temperature at 35.0℃. Put the skin sensor into the water

cup at 37℃, when the temperature rise to 36.1℃, the equipment should make a high priority

alarm sound and the normal information display area will show “Temp. Deviation Alarm”. The

alarm character should be consistent with the description in section 3.4.

Set the temperature at 35.0℃, Put the skin sensor into the water cup at 33℃, when the

temperature at 33.9℃, the equipment should make a high priority alarm sound and the normal

information display area will show “Temp. Deviation Alarm”. The alarm character should be

consistent with the description in section 3.4.

4-16

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

G. CHECK OVERHEAT ALARM

In Skin temperature Mode, put the skin sensor into the water cup at 39.5℃±0.5℃, the
equipment should make a high priority alarm sound and the normal information display area will
show “Skin Overheat Alarm”. The alarm character should be consistent with the description in
section 3.4.

H. CHECK “CHECK” ALARM

In Manual Mode, set the heating proportion at 50%, after 15min, the equipment should make
a high priority alarm sound and the normal information display area will show “Check Manual
Mode”. The alarm character should be consistent with the description in section 3.4.

4.4.4 MECHANICAL CHECK

A. CHECK CANOPY LIFTING FUNCTION

Press on the down control button, decline Press on the up control button, rise the
the canopy at the lowest condition and the canopy at the highest condition and the
radiation door will complete closed at the radiation door will complete opened at the
same time, the device is at incubator mode. same time, the device is at warmer mode.

WARNING
1. The canopy only can locate at the highest or lowest condition, otherwise
the equipment will send out the error signal.
2. To avoid the scald, please don’t touch the hearer before it’s cooling.

CAUTION: To avoid disturb the canopy go up and down, please disconnect all of the
installed auxiliary equipment before rise or decline the canopy. Make sure
there is no barrier in the moving range of canopy.

B. CHECK SIDE DOOR AND FRONT DOOR PANEL

CAUTION: Make sure disconnect all of the installed auxiliary equipment before open the
panel. To avoid disturb the operation.

4-17
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Lock Seat Unlocked

Side Door Panel

Lock Seat Locked


Press the lock seat at the unlocked condition, then open the left side and right side panel at
maximum (turn down). Close the left side and right side panel, make sure every lock seat is
complete locked, prevent the panel unexpected open.

Front Door Panel

While the canopy raised, perpendicularly lift the front door panel according to the direction of
arrow 1, then open the panel at maximum (turn down) according to the direction of arrow 2. When
close the panel, should upturning the panel at the vertical location then press downward, make
sure the panel install strong and accurate.

C. CHECK THE LATCH AND GASKET OF ACCESS DOOR

Latch Open Latch Close

4-18

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Push away the latch of each access door as the arrow shows, the operation window can
automatic bounce off. Then close the access door, you can determine whether the latch cover
securely closed and sealed tightly with the access door gasket.

D. CHECK THE INNER WALL

Inner wall unlocked Inner wall locked Bottom of inner wall locked
Open the left side and right side panel, the top of inner wall should be locked condition, the
bottom of inner wall should fixed in the inner wall hold down groove.

E. CHECK TILT MECHANISM

Tilt Button

Press on the tilt button and tilt angle crib should be able to be adjusted. Release the tilt
button when the infant bed is in any position inclined path, the infant bed should be locked in this
position.

F. CHECK THE INFANT BED

4-19
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Put the infant bed in a horizontal position. Open the left and right sides of the panel, as the
direction of the arrow to pull-out the infant bed until it cannot move (infant bed should be pulled
out from two sides), try to press infant bed, to see whether there is enough support in order to
provide baby a smooth platform, then the rotate bassinet as the direction of the arrow (bassinet
must first be pulled out or keep side door open to free rotating bassinet), see if it can be rotated
360 °. Return infant bed and close the panel, then lock the seat.

G. CHECK THE AIR INTAKE FILTER

Take off the tank and you will see the air filter device, screw out the hand screw to remove
the air intake filter cover and check the air filter material, if dirty, it should be replaced.

Hand screw Air filter cover Air filter material

WARNING: A dirty air filtering material will increase the concentration of carbon dioxide
inside the hood, according to the actual condition, must check the air filter
materials regularly. If the air filters materials look dirty or used more than 2
months, please replace it.

H. CHECK THE X-RAY TRAY

X-Ray Tray

Open the left side or right side panel, both of the side all can take out the X-ray tray.

4-20

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

I. CHECK THE SEALING GASKET AND CORNER COVER

Canopy
Corner cover
Sealing gasket

Check the sealing gasket and corner cover on canopy, make sure both of them install in
place and accurate fit.

J.CHECK THE HIGH ADJUSTMENT FUNCTION

Step the up and down button with feet as the figure 4.5 shows, the height of equipment can
be adjusted.
CAUTION: Must decline the equipment at the lowest position before moving the device.

K.CHECK THE OBSERVATION LAMP

Connect the power line of observation lamp to the equipment power supply output socket,
then press the observation lamp button, the lamp can be turned on or turned off.

L. CHECK I.V. POLE AND TRAY

Check the I.V. Pole and tray are tightening or not.

4.5 GENERAL OPERATION PROCEDURE

WARNING
1. Please read this operator’s manual carefully before use.
2. You should not use the equipment without the checkout procedure, and please refer
to the qualified person.
3. For the equipment’s normal function, the setting temperature must 3℃ higher than
environment temperature.
4. The display module of the controller is susceptible with the interference of
electromagnetic, so don’t use the controller under the high electromagnetic field. If a
equipment which sends or receive weak signal installed in the equipment nearby, it may
be affected by the electromagnetic wave sent by this equipment. Before using, please
check if the equipment has been affected.

4-21
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

4.5.1 PREPARATION
4.5.1.1 Connect the power supply, skin temperature sensor, weight sensor.
4.5.1.2 Air Mode and Baby Mode under the incubator mode:
Pre-warm the incubator, then put the patient on the infant bed after the air temperature
reaches the stable temperature. To avoid the body temperature of the patient decreasing in the
non-pre-warm incubator. The specific pre-warm method is as follows:
Choice air mode then set a proper control temperature and make the equipment continuous
operation until the air temperature reaches the set value.

IMPORTANT: 1.Because of the air outlet at two sides of the bed, while the infant placed
into the equipment from the side door should avoid the slotted hole which
provide for warm air flow.
2. Please ask the physician to decide the set temperature.
A. Use The Air Mode
Air Mode control the baby's body temperature by controling the inside temperature of
incubator. It will automatically control the inside temperature of incubator at the set temperature.
Under the Air Mode, the controller make a comparison between the temperature value (which
is detected by air sensor) and air setting value, And control the heat proportion according to the
comparative result, make the inside temperature reach the set value.

CHOICE AIR TEMPERATURE CONTROL AND SET POINT

On first indication interface, the keyboard is under the open condition and the keyboard

indicator light is on(you can press the rotate knob key to open or close), press the Air self-definite

function key to enter into the second interface A (Air temperature setting interface). Refer to

Figure 4.6, after that press setting self-definite function key, the light flashes and set the Air

Temperature to the prescribed temperature. Once stabilized, the air temperature will be

maintained within± 0.2℃ of the set temperature.

The mode means the equipment is operating by air temperature mode. Air temperature and

set temperature are display in air temperature and set position at same time. If skin temperature

sensor connect with socket, the skin temperature sensor probe temperature is display skin

temperature position, if not connected, the display is“--.-”.

4-22

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

If need to change the set temperature (lower or higher) during the equipment operation

process, press the keyboard locking key at first, after unluck the keyboard, choice the air mode

and press the setting button, then turn the rotary knob to adjust the setting temperature. If set the

override mode, first set the air temperature at 37℃, then press the>37℃ button in first indication

interace A. The indicator light of >37℃ will turn on means the equipment is under the override

mode, turn the rotary knob at this time can set the temperature >37℃. After selected the

appropriate temperature, press the keyboard locking key or wait for 10 seconds without doing any

thing can enter into the Control Mode. The settings area of air temperature information shows the

setting value. The display area of air temperature information shows the air temperature

measured with air temperature sensor. The display area of skin temperature information shows

the skin temperature measured with skin temperature sensor.

2015-03-24 10:10:00
SKIN

AIR
37.0 ℃ SKIN(℃) TIMES(12h)

45
SET

>37℃
SET --.-
40
SKIN
HEATER 35
TREND

HUM. 30 HOME

O2 25
0 t
AIR
25.0 ℃ HUM. % O2 % WEIGHT g
3
2
.
0

SET

23 ○
21 1000
SET 0 SET 20

FIGURE 4.6

NOTE: 1. All settings only can be operated under the unlocked condition.
2. In spite of any operation, if don’t do anything within 10 seconds, the display will
auto return to the first indication interace and save the latest operating result.

1 AIR

Press this key to enter into the Air Temperature setting interface.

2 SET

Press this key to enter the setstate of Air Temperature set value, the Set Value blink and turn the
rotary knob to adjust theset value.

4-23
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Rotate this rotate knob right or left to change the setting value increase or decrease. Rotate right

rotate knob key to increase setting value, raise set temperature to 37℃. In Temperature Override

Mode (Step 4), please press this key to raise set temperature from 37℃ to 39℃. Rotate one

case, set temperature raise 0.1℃. Rotate left rotate knob key to decrease setting value, lower set

temperature to 25℃. Rotate one case, set temperature lower 0.1℃.

4 >37℃

Press the>37℃ button, the equipment will enter into temperature override mode. >37℃ will

show on the normal information display area, the controller can proceed the >37℃ temperature

setting.
NOTE: While the set value is not displayed at 37℃, this button is inoperative.

5 CURVE
Press this key to enter the curve interface. Refer to section 4.5.1.4 for details.

6 HOME

Press this key to confirm the new set temperature and return to first indication interface.

Press this rotate knob, the keyboard lock will enter controller set state (Display screen

display ). After setting back to first interface, press this knob to enter keyboard closed

condition (Display screen display ), or don’t press any key, it will automatically enter keyboard

closed condition after 10 seconds.

8
This key has two functions, when occur alarm indication, press once can cancel the alarm sound

and the top right corner of the display screen shows , press twice can reset the alarm state,
then the equipment back to the set condition to monitor the alarm and top right corner of the

display screen shows .

4-24

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

B. Use the Baby Mode

Baby Mode is a temperature control method which means the controller will control the skin

temperature automatically to keep the baby temperature close to set temperature.

Refer to the step E in section 4.4.2, Connect the Skin sensor to the socket on the sensor

box correctly. On first indication interface, press skin self-definite function key to select Baby

Mode. About temperature setting, please refer to the figure 4.7 and skin temperature setting part.

Once stabilized, the sensor probe temperature will be automatically controlled within ±0.2℃and it

means that the incubator is working under Baby Mode


IMPORTANT: The Infant incubator controls the temperature by sensor. Therefore, the
sensor must be connected to sensor socket when the controller controls the
temperature by skin temperature.
In Baby Mode, Skin Temperature, Air Temperature and Set Temperature are displayed on

display screen at the same. If the skin temperature sensor is connected with sensor socket, the

temperature from the Skin Temperature Sensor Probe will also be displayed there.

The skin temperature sensor should be closed to baby skin surface under the skin mode.

The incubator internal controller compare temperature value (which is detected from skin

temperature sensor) with the set skin temperature, and control the heat proportion of heater

according to comparative result, eventually make the skin temperature sensor detects the

temperature stability near the set value.

If need to change the set temperature (lower or higher) during the equipment operation

process, press the keyboard locking key at first, after unluck the kepad, choice the skin mode and

turn the rotary knob to adjust the setting temperature. If set the override mode, first set the skin

temperature at 37℃, then press the >37℃ key in second indication interace A. After enter into

the override mode, turn the rotary knob to set the control temperature >37℃. After choose the

appropriate temperature, press the keyboard locking key or wait for 10 seconds without doing any

thing,it will automatically enter into the Control Mode. The settings area of skin temperature

information shows the setting value. The display area of air temperature information shows the air

temperature measured from air temperature sensor. The display area of skin temperature

information shows the skin temperature measured from skin temperature sensor.

In order to setting the appropriate temperature and correctly control the infant's skin

temperature, the incubator has following protection functions: When the temperature detected by

the skin temperature sensor is always below the set temperature for more than 2℃, the incubator

4-25
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

will rise the air temperature in the incubator in the speed of 1℃ per hour to make the infant's

temperature slowly rise, and avoid the damage during rapid rising of the temperature. If the skin

temperature is always below the set temperature for more than 2℃ in 1 hour, the system will

occur " Skin Sensor Position Error ". At that time, the incubator still rise the air temperature in

the hood slowly until the air temperature reaches the steady temperature condition which

temperature is equal to “ skin temperature set value -4℃”. For example, the set temperature is

36.5℃ under the skin mode, the temperature in incubator will finally stable at 32.5℃.

In order to make the temperature unnecessary lower because of infant fever, the incubator

has following protection functions: If the baby skin temperature surpasses the set temperature for

less than 0.5℃, the air temperature in the incubator will not lower 5℃ than the baby skin

temperature. If the baby skin temperature surpasses the set temperature for over than 0.5℃, the

air temperature in the incubator will not below 25℃. For example, if the set temperature is 36℃

while the baby temperature rising to 36.3℃, the air temperature in incubator will not lower than

31.3℃. While the baby temperature rising to 36.7℃ which is higher 0.5℃ than the set

temperature, the air temperature in the incubator will not lower than 25℃.

WARNING
1. Make sure that the probe of skin sensor is attached closely with the skin of patient. If
the probe falls off, the measured temperature from sensor is not the real skin
temperature, maybe the air temperature or the mattress surface temperature, and it may
cause the patient receive more heat or lose heat, even scald or death.
2. Please do not cover the blanket or diaper on the probe of skin sensor that may affect
the accuracy of temperature.
3. The equipment can‘t differentiate the different between skin temperature is low but
body temperature is high (fever) and both skin temperature and body temperature are
all low. Skin temperature sensor only measure the skin temperature of patient, not the
real body temperature, Therefore, measure the body temperature regularly, and check
whether the patient has a fever or not, whether the temperature of patient decreases.

CHOICE BABY MODE AND SET POINT

Press the skin temperature user-defined function key to enter the indication interace as the
figure 4.7 shows, when the keypod lock unlocked.

4-26

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

2015-03-24 10:10:00
SKIN

37.0 ℃ SKIN(℃) TIMES(12h) SET

3
7
.
0
45
>37℃
SET
AIR 40
HEATER 35
TREND

SKIN
30 HOME
HUM.
25
0 t
O2
AIR
25.0 ℃ HUM. %
○ 22
O2

% WEIGHT g
21 1000
SET --.- SET SET 20
0

FIGURE 4.7

NOTE: 1. All settings only can be operated under the unlocked condition.
2. In spite of any operation, if don’t do anything within 10 seconds, the display will
auto return to the first indication interace and save the latest operating result.

1 SKIN

Press this key to enter the Skin Temperature control mode interface.

2 SET

Press this key to enter the Skin Temperature Setting, the Indicator “Set” blink.

Rotate this rotate knob right or left to change the setting value increase or decrease. Rotate right
rotate knob key to increase setting value, raise set temperature to 37℃. In Temperature Override
Mode (Step 4), please press this key to raise set temperature from 37℃ to 38℃. Rotate one
case, set temperature raise 0.1℃. Rotate left rotate knob key to decrease setting value, lower set
temperature to 25℃. Rotate one case, set temperature lower 0.1℃.

4 >37℃

Press the >37℃ button, the equipment will enter into temperature override mode. >37℃ will
show behind the mode, the controller can proceed the >37℃ temperature setting.
NOTE: While the set value is not displayed at 37℃, this key is inoperative.

5 CURVE

Press this key to enter the curve set interface of the third display interface B, refer to the section
4.5.1.4.

4-27
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

6 HOME

Press this key to confirm the new set temperature and return to first display interface.

Press this rotate knob, the keyboard lock will enter controller set state ( Display screen
display ) After setting back to first interface, press this knob to enter keyboard closed
condition(Display screen display ), or don’t press any key, it will automatically enter keyboard
closed condition after 10 seconds.

8
This key has two functions, when occur alarm indication, press once can cancel the alarm sound
and the top right corner of the display screen shows , press twice can reset the alarm state,
then the equipment back to the set condition to monitor the alarm and top right corner of the
display screen shows .

Skin temperature sensor probe connect with patient skin surface:


In Baby Mode, must make sure the skin temperature sensor probe connect with patient skin
reliably. Before put the probe on the infant skin, you must use ethyl alcohol or warm water to
clean the grease and dirt on the skin which intended to place the sensor probe, and dry it. you
can use medical proof fabric or analogue to fix the sensor probe.
If the patient is supine, you should place the probe between abdominal sword cartilage and
navel, keep away from the liver part.
If the patient is grovel, should place the probe at the backside of patient, the best location is
the kidney part.
If the patient lie on the side, the location place the sensor probe should ask the advice from
attending doctor.

NOTE: 1. Skin temperature sensor must be disinfected before use.


2. Don’t place the skin temperature sensor probe under the patient.
3. Skin temperature sensor probe can’t be used as the rectal thermometer.

4.5.1.3 Pre-warm Mode, Manual Mode and Baby Mode Under Warmer Mode
A. Pre-warm Mode
Pre-warm the warmer, put the patient on the infant bed after the mattress surface
temperature reaches the stable temperature, to avoid the body temperature of the naked patient
decreasing quickly. Please transfer to manual mode or skin mode immediately when put the
patient into the equipment.
4-28

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

This equipment has Pre-warm Mode, after the starting up self-inspection, the equipment will

enter into the interface in figure 4.8 without pressing any key:

2015-03-24 10:10:00
SKIN

37.0 ℃ SKIN(℃)

45
TIMES(24h) PRE-WARM

SET --.- MANUAL

40
HEATER 35 SKIN

30 APRAR
25
0 t
MODE Pre-warm

0
:
4
4
APGAR WEIGHT g

1000

FIGURE 4.8

The system will output the heat according to the set procedure, after a while, the system will

output the heat according to 30% heat proportion(Radiation heating power output ≤10mW/cm2)

to keep the mattress temperature until the mode changes. During the pre-warm process, the Set

Temperature window shows “--.-”, Skin Temperature display window shows the real-time

temperature. Under the pre-warm mode (the environment temperature is 23℃±2℃), after the

equipment working 30 minutes since the cooling status, the surface temperature of mattress can

rise 4℃ at least. So suggest the operator should keep the warmer working 30 minutes at least

under the Pre-warm Mode.

B. Manual Mode

Manual Mode, the equipment output the heat according to the set heating proportion. This

mode only use for provide short time or emergency treatment the patient, or warm the

hypothermic patient. While the equipment transform from incubator air mode to warmer mode,

direct enter into the interface in figure 4.9.

NOTE: The operator should choice and adjust the heating output proportion according to
the clinical operation requirement.

WARNING
1. Operator should not leave for fear that the patient under the unattended condition.
2. Must regular check and measure the body temperature of patient.

4-29
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

2015-03-24 10:10:00
SKIN

37.0 ℃ SKIN(℃)

45
TIMES(24h) PRE-WARM

MANUAL
SET --.-
40
HEATER 35
SKIN

30 APRAR

25
0 t
MODE MANUAL

APGAR

:
WEIGHT g

1000

FIGURE 4.9
The heat output of system is 30% (heating power indication lamp with three grades). Make

the keyboard unlock during manual mode, the operator can operate the left button or right button

to adjust the heating output proportion in the range of 0%-100%. when the heating output

proportion>30%, the warmer give alarming every 15mins, the alarm lamp on and the status bar

of control instrument will show “Check Manual Mode”, then press the Silence/Reset key one

time can stop alarming, press twice can cancel the alarming and equipment keep working

according to the current mode and heating output proportion. If the operator doesn’t press

Silence/Reset key, for the safety of patient, system will automatically limit the heating output

proportion at 30% after 1 minute.

NOTE: 1. In Manual Mode, the heating output proportion is fixed, uncontrolled the
temperature measured by skin temperature sensor. Therefore, pay attention to the
change of patient's body temperature.
2. While the heating output proportion over 30%, it will send out voice and light
alarm every 15 minutes, to remind the operator that the patient is under the
dangerous environment. For the safety, please check and measure the body
temperature of patient.

C. Baby Mode
The baby mode is a kind of working mode which can self-maintenance the patient‘s skin
temperatuer according to the setting temperature value. Under this mode, the system according
to the difference between the temperature measured by skin temperature sensor and controlling
temperature, automatically adjust the heat output of Infrared radiation to keep the heat balance
for patient. This mode is intended to hold the body temperature of patient. Skin temperature
sensor is the applied part. While the equipment transform from incubator baby mode to warmer
mode, direct enter into the interface in figure 4.10.
4-30

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

2015-03-24 10:10:00
SKIN

37.0 ℃ SKIN(℃) TIMES(24h) PRE-WARM

3
7
.
0
45
MANUAL
SET
40
HEATER 35
SKIN

30 APRAR

25
0 t
MODE SKIN

:
APGAR WEIGHT g

1000

FIGURE 4.10
Manual mode output the infared rediation heat according to the heating proportion, please
do not self-regulation by the practical situation of patient. If the equipment continuous operation
and continuous output infrared radiation thermal energy may cause the patient overheating,
therefore, consider the safety of patient, suggest the operator use the Baby Mode.
NOTE: Patient under the state of shock or fever, please don’t use this mode.
Please do not use the Baby Mode when the patient is under the state of shock, because the
baby temperature is much lower than the normal temperature during shock, or else, the patient
will be over-heat.
Please do not use the Baby Mode when the patient has a fever, because the baby
temperature is much higher than the normal temperature during fever, if use the self-regulation
function to control, it will cause the body temperature of the patient decrease.
For the patient’s temperature, usually, the operators think it means the rectum temperature,
which is measured by putting the thermometer into the rectum of patient. To measure the core
temperature of patient accurately, the operator needs to insert the thermometer in to the rectum
at depth of 5cm, so it will be dangerous (the struggle of patient will cause the crack of
thermometer), if the insert depth is not enough or time is not long, the operator can’t get the real
body temperature of patient, therefore, it is improper to control the output of heater via rectum
temperature.
The connection method of skin temperature sensor probe and patient skin surface are same
as the section "use Baby Mode" under the incubator mode.

CHOICE THE SKIN TEMPERATURE CONTROL VALUE:


Operator can adjust the baby temperature by turn left or turn right the rotary knob under the
setting state.
NOTE: Operator should according to the clinical condition and the suggestion of attending
doctor to choice the skin temperature control value.

4-31
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

WARNING
1. Skin temperature of patient is servo control under the baby mode, but the operator
also should not leave. Avoid the patient is caused the danger under the non-cared
condition.
2. The patient maybe not feel the water loss under the continuing heat radiation, please
add some water for patient to avoid dehydration.
3. See the warning in section “Baby Mode” under the incubator mode.

4.5.1.4 CURVE DISPLAY


Press the curve self-definite function key to choice the menu of the following curve display.

SELECT CURVE TO ENTRE THE CURVE SETTING INFTERFACE(FIGURE 4.11)

2015-03-24 10:10:00
SKIN

37.0 ℃ WEIGHT(kg) TIMES(1 week) times


3
7
.
0

05
SET SET STORE

04
>37℃ HEATER 03
02 HOME

01
HOME 0 t
AIR
25.0 ℃ HUM. %
○ 21
O2

% WEIGHT g
21 1000
SET --.- SET 0 SET 20

FIGURE 4.11

NOTE: 1. All settings should be operated under the keyboard unlocked condition.
2. In the state of air temperature control, press the curve self-definite function key,
and then the choice is the air temperature curve display, while in the state of skin
temperature control, press the curve self-definite function key, and then the choice
is the skin temperature curve display.
3. In spite of any operation, if don’t do anything within 10 seconds, the display will
automatically return to the first indication interace and save the latest operation
result. ( Except inquiry innterface and return to the last interface in the step2).

1 TIMES

Press this key continuously to choice the time intervals of curve trend, the curve display area
successively shows the trend display diagram at 2 hrs, 4 hrs, 8 hrs, 12 hrs, and 24hrs.

2 STORE

Press this key to store the curve trend from the current state again, the following interface will be
displayed at the same time, and inquiry the operator YES or NO.

4-32

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Note:Delete?

SKIN(℃ ) TIMES(12h) YES


45
40
35
30
NO
25
0 t

Press " Yes " key to confirm and return to the last interface, the controller will store again by
canceling the original information.
Press " NO " key to deny and return to the last interface, the controller will continue to store the
information on basis of the original information.

3 HOME

Press this key to return to the first interface, the curve display window will shows the temperature
curve of control mode.

Press this rotate knob, the keyboard lock will enter controller set state (Display screen
display ). After setting back to first interface, press this knob to enter keyboard closed
condition (Display screen display ), or don’t press any key, it will automatically enter keyboard
closed condition after 10 seconds.

4.5.2 OTHER OPERATION


A. RAISE THE HEAD OR FOOT OF PATIENT
Raise the head or foot of patient according to the step E in section 4.3.4.
CAUTION: 1. Please do not add the over load on the infant bed.
2. The bassinet tilt will affect the temperature uniformity of bed surface, the
horizontal position uniformity is best.

B. OPERATION THE X-RAY TRAY


1. Open the side door baffle and pull-out the X-ray Tray under the infant bed.
2. Place the X-ray box into the X-ray Tray.
3. Place the Infant at the center of infant bed. Close the side door baffle. When the X-Ray is
completed, remove the X-ray cassette from the tray and push the tray back.

4-33
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

C. OPERATION TO THE OXYGEN CONCENTRATION CONTROL SYSTEM.


Please refer to the section 6.

D. OPERATION TO THE HUMIDITY CONTROL SYSTEM


Please refer to the section 7.

E. OPERATION TO THE WEIGHTING SYSTEM


Please refer to the section 8.

4.5.3 SHUTDOWN
After finish the operation, please turn off the power switch of controller and main power
switch, and disconnect the power line.

4-34

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

SECTION 5
CLEANING AND MAINTENANCE
5.1 GENERAL
This section provides cleaning and maintenance instructions about the equipment.

WARNING: Disconnect all the connections with oxygen feeding device before cleaning
and maintenance. Cleaning or maintenance in the environment full with
oxygen will cause fire or explosion.

5.2 CLEANING
This equipment must be thoroughly clean and disinfect for the first time, or finish the

salvage/nursing/cultivating for one patient, or after used it for one week.

Date of manufacture: See the product electrical nameplate.

5.2.1 DISASSEMBLY BEFORE CLEANING

1. Rise the Canopy, turn off the Power Switch on the VHA stand and remove the Power Cord.

Refer to Paragraph 4.4.4, Step B and disconnect the cables from the Sensor Module and slowly

turn down the baffle.

CAUTION: Before rising the Canopy or turn down the baffle, ensure that all mounted

accessories have been removed, to prevent interference.

2. Refer to Paragraph 4.4.4, Step H, and remove the X-ray Tray from the infant bed.

3. Refer to Figure 5.1 and remove the mattress and Mattress Tray. (For the infant bed with

the baby scale, please see the figure 8.1, disconnect the weight sensor and take the baby scale

out)

Mattress

Mattress Tray

FIGURE 5.1

5-1
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

4. After remove the mattress and mattress tray, take out the baby scale as the figure 5.2
shows. (optional part)

Baby scale

FIGURE 5.2
5. The figure 5.3 shows the disassembly steps of bed body.

Step 1: Put the mobile platform and rotating bed at the middle of the main deck, and raise
them up.

Rotating bed
Buckle

Mobile platform

Step 2: Separate the mobile platform and rotating bed by move the rotating bed and
squeeze the two buckles at the middle.
5-2

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Tilt button

Connect button

Step 3: Squeeze the tilt button to make the tilt platform at the lowest position, and move the
connect button as the arrow direction to take out the tilt screw.

Enlarged Figure A Enlarged Figure A

Fixed Button

Center handle

Step 4: Put the handle vertically, and as the enlarged figure A shows, press the fixed button
at the sides to take out the tile platform.

FIGURE 5.3
6. Refer to the figure 5.4 to remove the bed tray.

The heater maybe cause burns because of overheating. Please remove


the bed tray or touching the Heater until the unit has been switched off
for at least 45 minutes.

FIGURE 5.4

5-3
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

7. Refer to the figure 5.5 to unlock the lock catch at the top of inner container, then break
away the bottom of inner container from the hold down groove (show in Enlarged Figure A), can
take out the inner container.

Inner Container

Enlarged Figure A

Enlarged Figure A

FIGURE 5.5

8. Refer to the figure 5.6 shows the steps to take out the water tank:

Step 1: Handing the groove of tank bottom Step 2: Use two hands hold the button of tank
and pull out. and pull down, will take out the tank.

To avoid burns or damage to the modules, please remove the tank


after sufficient cooling.
FIGURE 5.6
9. As the figure 5.7 shows, take out the fan.

Hold the fan and draw the fan upward gently as the arrow direction.
FIGURE 5.7

5-4

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

10. As the figure 5.8 shows, take out the oxygen input pipe.

Clamping Ring

Move the clamping ring as the arrow direction to take out the oxygen input pipe.
FIGURE 5.8
11. As the figure 5.9 shows, the disassembly step about the sealing gasket of Canopy as
follows:

Corner Cover

Step 1: Slightly to raise the cover and push it up, then take out the corner cover.

Sealing Gasket

Step 2: According to the direction of arrow, slide the two sides long gasket can take off them
from the guiderail.

Sealing Gasket

Step 3: According to the direction of arrow, slide the front short gasket can take off them
from the guiderail.
5-5
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Groove Sealing Gasket

Step 4: Pull according to the direction of arrow, then take out the groove sealing gasket.

FIGURE 5.9

12. Take down all of the operating window gaskets from the panel.
13. Take down all of the Infusion soft mat from the panel.
14. Take down the Air filter by loosening the hand screw. Refer to Step G in section 4.3.3.

5.2.2 CLEANING PROCEDURE


CAUTION: Some chemical cleaning agents/disinfectant may be conductive or leave a
residue which may permit a build-up of dust or dirt which may be conductive.
Do not permit cleaning agents to contact electrical components. Do not spray
cleaning solutions onto any of those surfaces.
A. Clean the skin temperature sensor
Use a state resisted neutral disinfectant-detergent to thoroughly clean the skin temperature
(including probe), then disinfect it with neutral disinfectant or ultraviolet.
CAUTION: 1. Prohibit put the sensor into the disinfectant-detergent or sterilized by high
temperature and high pressure.
2. Skin temperature sensor should be changed every two years, to avoid the
damage of sensor surface and strong impact for long-time use and disinfection.
B. Clean the I.V. soft port, Access door gasket, Canopy sealing gasket
Allow parts to soak as national registered neutral disinfectant-detergent, then remove them
and dry completely with a clean cloth or paper towel.
C. Clean the panel, box cover, inner wall, mattress tray and sensor box
Use the neutral disinfectant-detergent to thoroughly clean all surfaces, including sensor box,
box cover, inner wall, mattress tray and all contacted doors and windows, ensure clean all holes
and sag places, then dry with a clean cloth and paper towel.
CAUTION: 1. To avoid the panel, box cover and inner wall appeared crazing. Do not use
alcohol, acetone, or any organic solvents for cleaning. Do not expose the panel
assembly to direct ultraviolet radiation.
2. Please avoid the liquid flowing into the sensor box.
5-6

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

D. Clean the controller

Use neutral disinfectant-detergent to clean the all surfaces. Then dry it with clean cloth.

E. Clean the heater and fan

Remove any lint build-up on the Heater and Fan. Use neutral disinfectant-detergent to clean

the all surfaces. Then dry it with clean cloth.

WARNING: Not clean the heater and fan will result in accumulate of fibre dust, impede the

air flow. That will influence the temperature control and result in high CO2

concentration.

F. Clean the mattress of infant bed, baby scale (optional part), mobile platform, tilt

platform, rotating bed, tray, X-tray tray, storage compartment and bed body tray and corner

cover.

Use a neutral disinfectant-detergent to thoroughly clean all surfaces and dry with clean cloth

or paper towel.

NOTE: There are high precision of weight sensor in the baby scale must handle with care

while clean it. Strictly prohibit to cleanout and steep the electronic scale in

cleanser or disinfectant.

G. Clean the air filter

Use a neutral disinfectant-detergent to thoroughly clean all surfaces, including all holes and

cupped place, then dry them with a clean cloth or paper towel. If the air filter is dirt or is used for

more than two months, please replace it.

H. Clean observation lamp

Use a neutral disinfectant-detergent to thoroughly clean all surfaces, the observation lamp

and camera lens can use common glass cleanser to clean.

CAUTION: Strictly prohibit to steep the observation lamp in water to clean.

I. Clean the water tank

Take out the cover plate of humidifying chamber from the water tank as figure 5.10 shows.

Use the neutral disinfectant-detergent to thoroughly clean the all surfaces and cupped place.

Clean it again with clean water and then dry it with clean cloth or paper towel.

Please pay special attention that clean the inner and outer surface of the probe with

disinfection soft cloth, it can control the microorganism breed in the water tank, and ensure the

water shortage detection without error.

5-7
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Reek outlet Detecting probe Cover board

Hold the cover board press to the right and left direction as the arrow shows, level up the

cover board to take out it.


FIGURE 5.10
NOTE: 1. Please clean the water tank after the radiating cover inside the water tank totally

cooled down.

2. Make sure the water inlet hole on the bottom and the reek outlet hole of tank

clear unobstructed after cleaning. Two gaps on the water proof must placed

downwards.
3. If the water tank is unclean, especially there are dusts near the probe, there will

be a possibility that water shortage alarm failure.

4. Please use soft cloth to wipe the surface of the probe when it is covered by noble

metal. Do not use the hard material to scrape the probe so as not to damage it.

The water tank can use steam sterilization. The way of steam sterilization: it should clean

and dry the water tank thoroughly before steam sterilization.

The duration of steam sterilization is: 132℃ for 3 to 5 minutes; 121℃ for 15 to 20 minutes.

Repeating sterilize will lead to crack for some parts, it will damage the water tank and eventually

need to replace the water tank.


J. Clean the oxygen input pipe

Use clean cloth to clean the surface of oxygen input pipe.

5-8

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

K. Clean the surface of equipment


Use a neutral disinfectant-detergent to thoroughly clean all surfaces, including all holes and
cupped place, then dry them with clean cloth or paper towel.

5.2.3 INSTALL AFTER CLEANING


NOTE: Before installing parts into Incubator, please check carefully whether there is any
crack or damage. Replace the parts in time if damaged.
A. Install the water tank and air filter
install the water tank to the crate through the opposite sequence, then put into the air filter
cover and tighten the fixed screw.
B. Install the fan as the figure 5.11 shows

Connect Ring

FIGURE 5.11

NOTE: Before take back of the fan, please make sure the fan is always fixed on the
middleware and the connect ring is visible. After ensure the fan can free to rotate,
you can install other modules back.
C. Install the bed body
Refer to the figure 5.3 and 5.4 opposite sequence to install back the bed body. Refer to the
figure 5.1, put the mattress and base plate back. After that, refer to the step H in section 4.3.3 to
put back the X-ray Tray.
D. Install the inner wall
Refer to the opposite sequence of figure 5.5 to install back the inner container.
E. As the figure 5.12 shows, install the gasket of access door

gasket of access door


FIGURE 5.12

5-9
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

F. As the figure 5.13 shows, install the I.V. soft port

NOTE: The gap of I.V. soft port must be outward.

Gap

FIGURE 5.13
G. Install the gasket of Canopy
Refer to the figure 5.14 to install back the gasket of Canopy.

Groove Sealing Gasket

Step1: Insert the gasket into the groove according to the direction of arrow, hen slide up the snap
fastener in place.

Align at Guiderail Big Sealing Lip


Sealing Gasket Guiderail

Step 2: Adown the big sealing lip of gasket, glide the gasket along the guiderail till fixed, the
gasket should align at the end of guiderail.

5-10

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Enlarge figure A

Corner Seat Corner cover

Step 3: Lock the corner cover on the corner seat according to the direction of arrow, and make
sure it’s locked in place.

WARNING: Must make sure every sealing gaskets of Canopy are matched in place
after finish installing.

FIGURE 5.14

5.3 MAINTENANCE

WARING: Do not refit the incubator to avoid effect the safety of performance.

Please check the condition of the internal rechargeable battery before the first use or the
using interval period.
A. Continuous operate the device for 12 to 24 hours.
B. Enablement the power failure alarm by disconnecting the power supply cord.
C. The power failure alarm should continuous alarm for 10 minutes at least.
D. Insert the power supply cord of device and operate it, for recharge the internal battery.
If the power failure alarm can’t last over than 10 minutes, please replace the internal
rechargeable battery. The battery should be replaced by the authorized and qualified personnel.
NOTE: 1. While the device is unused, use the above method to check the internal
rechargeable battery for every three months and make sure the battery is fully
charged.
2. The replaced rechargeable battery will affect the environment if discard, must
according the local regulations to dispose.
If the heater is over the useful life period, for insure the effect of infrared irradiation. The
heater must be replaced, even though the heater can work as normal. This is because:
The infrared irradiation effect of heater will recede with the using time, that will cause the
device cannot reach the performance of temperature asserted in table 1.1 within this Operator’s
Manual, and may result device lose the intended effect while keep warm for the patient.

5-11
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

The heater should be replaced by the authorized and qualified personnel.


The bulb should be replaced in time if damaged. It must use the bulb with same wattage and
voltage, and it need to replace the bulb after the check lamp cooling.

5.4 TROUBLE SHOOTING


Troubleshooting of the temperature control system is presented in the following table. If the

fault cannot be found out from the table or the fault is not appeared in the following table, the

equipment should be removed from service and servicing should be referred to our company or

serviceman with authorized and qualified.

TROUBLESHOOTING TABLE

SYMPTOM POSSIBLE CAUSE REMEDY

No display and No alarm Power Switch is not on Open the Power Switch

The color of screen is High temperature for switch off the temperature
controller, restart it after
abnormal controller
cooling.

Power failure alarm light is Power failure Turn off the Power Switch
on, red alarm light blink, the
display screen display
"power failure alarm". Power supply cord is not Connect the power supply
connected cord

Skin temperature sensor


Connect the Skin temperature
disconnect or connect not sensor correctly.
Red alarm light blink, the
well
display screen display “Skin
Sensor 1 Fault” Skin temperature sensor Replace the skin temperature
damaged sensor

Install the O2 sensor 1


O2 sensor 1 not installed
Red alarm light blink, the correctly
display screen display “O2
Sensor 1 Fault”
O2 sensor 1 damaged Replace the O2 Sensor 1

Install the O2 sensor 2


O2 sensor 2 not installed
Red alarm light blink, the correctly
display screen display “O2
Sensor 2 Fault”
O2 sensor 2 damaged Replace the O2 Sensor 2

5-12

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

TROUBLESHOOTING TABLE(Continued)

SYMPTOM POSSIBLE CAUSE REMEDY

Far away from the heat


Ambient temperature is too
source or decrease the
Red alarm light blink, the high
ambient temperature
display screen display “Air
Overheat Alarm”
Decrease the humidity
High humidity
inside hood

Door or window of panel Close the door or window


is open of panel

Red alarm light blink, the Ambient temperature Check the Ambient
display screen display changes a lot temperature
“Temp. Deviation Alarm” Far away from the heat
Heat source nearby
source
Ambient temperature Check the Ambient
changes a lot temperature
Power cord is not connected Connect the power cord
VHA stand can’t go up
and go down
Power switch is not on Turn on the power switch

5-13
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

SECTION 6
OXYGEN CONCENTRATION CONTROL SYSTEM
6.1 GENERAL
This Section provides operation checkout procedure and operation instruction for the
Oxygen Control System (only use under the incubator mode).

6.2 REPLACE THE OXYGEN SENSOR


Replace the oxygen concentration sensor as the figure 6.1 shows.

Sensor Cover plate

Screwdriver

Unscrew the screw of the sensor cover plate and open the cover plate.

Oxygen Sensor

Connecting Line

Disconnect the connecting line of oxygen sensor then screw out the oxygen sensor with
counter-clockwise direction to replace it. After the replacement, Twist the oxygen sensor into the
sensor cover plate with clockwise direction and connect the connecting line correctly then close
the sensor cover plate and tighten the screw.

6-1
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

CAUTION: 1. Please pull out and/or install into the sensor box in correct way.
2. Sensor box is an important part for controlling the temperature
of incubator, please operate it carefully.

FIGURE 6.1

6.3 FUNCTIONAL DESCRIPTION


The incubator has function of oxygen concentration control. The range of oxygen
concentration display is 0%~99%,setting range is 20%~60% (every time adjust 1%).
The incubator has alarm function of O2 Sensor 1 Fault, O2 Sensor 2 Fault, O2 Sensor DIFF. Fault,
O2 Sensor Deviation Alarm, Please see the table 3.4 to get details of alarm message. If the above
alarms are active, press the Silence /Reset key to stop alarming for about 115s. If the power of
the heater is not cut off, it will cut off the oxygen supplying automatically (Except for O2 lower
deviation alarm).

6.4 CONNECTION OF OXYGEN INPUT CONNECTOR

WARNING
1. Please read the operation manual carefully before operating.
2. Oxygen concentration control system must use oxygen analyzer to supply oxygen
through the oxygen input connector, the relative operation must refer to the operation
manual of oxygen analyzer or other documents similarly.
3. Only given decompress valve or pressure adjusting valve can be used on the
oxygen cylinder.
4. According to the theory that the oxygen concentration in-breathed by the patient
cannot judge his artery PaO2. Accurately, so other acceptable clinical measures
should be taken to testify the patient’s artery PaO2.

Connect the oxygen transferring tube of servo control oxygen interface (see figure 4.3) and
the oxygen-offering system equipped with the oxygen monitoring device, and make sure the seal
of connecting part.

NOTE: 1. For safety, when offering the oxygen, must be a calibrated oxygen analyzer that
can monitor the oxygen concentration inside of hood for long time.
2. Oxygen transferring tube offered by our company only can connect with servo
control oxygen input interface.

WARNING: Over high oxygen concentration can cause the great hurt on the baby, so please
follow the advice of doctor when offering the oxygen, and use the relevant
oxygen monitoring device to make it reach the setting state set by physician.
Please read the indication values, the oxygen concentration table is just
regarded as a reference.

6-2

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

6.5 OXYGEN SYSTEM OPERATIONAL CHECKOUT PROCEDURE


Before the first using and reuse after maintenance, must perform the oxygen concentration
calibration system operating process.
The system support two kinds of oxygen concentration calibration: 21% oxygen
concentration and 100% oxygen concentration. When the oxygen concentration of environment is
below 21%, suggest using 100% oxygen concentration calibration.

21% OXYGEN CONCENTRATION CALIBRATION

Keep the incubator in the environment which oxygen concentration is 21%.


a. After press the second self-definition function key (from top to bottom), turning the
controller to enter the interface of oxygen concentration calibration. Press “21% Cal.” Key, see the
figure 6.2. The controller can calibrate 21% oxygen automatically and the indicator display the
“O2 calibration ”After calibration, the oxygen concentration display the calibration value. If
calibration work fails, the indicator displays the failure hint, and then users can solve problems
according to the hint. If these failures cannot be canceled, please ask the authorized and qualified
personnel for repair.

21%

O2 CALIBRATION PAGE
100%
EXIT
Press and open the equipment
to enter the oxygen concentration
calibration interface
FIGURE 6.2

100% OXYGEN CONCENTRATION CALIBRATION

a. Turn off the controller as figure 6.3 shows, connect the oxygen calibration device which
equipped with oxygen connecting pipe to the sensor box, supplying the device with 100% oxygen
for more than 2min.
b. After press the second self-definition function key, turning on the controller to enter the
oxygen concentration calibration interface as the figure 6.2 shows. After that, press “100% Cal.”
Key, the controller can calibrate oxygen sensor automatically. After calibration, the oxygen
concentration display the calibration value. If calibration work fails, the indicator displays the
failure hint, and then users can solve problems according to the hint. If these failures cannot be
canceled, please ask the authorized and qualified personnel for repair.

6-3
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Sensor box

Oxygen concentration calibration device


FIGURE 6.3

CHECK THE OXYGEN CONCENTRATION MONITOR FUNCTION

1. Place a calibrated oxygen concentration analyzer on the middle of the bassinet.


2. Input oxygen to the device, wait until the oxygen concentration displays steady, the value
deviation displayed between the value displays in the Oxygen Concentration indicator and the
oxygen concentration analyzer should in the range of ±3%.

CHECK THE OXYGEN CONCENTRATION DEVIATION ALARMING FUNCTION

Connect the oxygen supply system and servo control oxygen input interface, set the oxygen
concentration as 30%, when the oxygen concentration reaches the setting value, increase the
oxygen flow immediately, when the display value is different from the setting value +5%, the
device should give a high priority alarm sound, “O2 sensor deviation alarm” information will
indication on the screen, the alarm character should be consistent with the description of
section3.4; set the oxygen concentration as 30%, when the oxygen concentration reaches the
setting value, decrease the oxygen flow immediately, when the display value is different from the
setting value -5%, the device should give a high priority alarm sound, “O2 sensor deviation
alarm” Information indication on the screen, the alarm character should be consistent with the
description of section 3.4. When alarm occurs, press silence/reset key once can cancel the
alarm sound, but the alarm light still flashes until the display value is within ±5% of setting value.

6.6 USE OF OXYGEN SUPPLY SYSTEM


6.6.1 PREPARATION
Connecting the oxygen flow-meter and oxygen input valve with one PU-10×6.5 medical pipe.
6.6.2 OPERATION
6.6.2.1 According to the requirements of physician, on the first screen, press oxygen
self-definition function key, please refer to the figure 6.4 to set the oxygen concentration
monitoring datum mark.
6-4

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

NOTE: During operation, if there is information like “O2 Module Switched off” displays on
the screen, there must be something wrong with the oxygen module: not been
installed or fails to work.

WARNING: 1. If the artery blood oxygen pressure of patient (PaO2) cannot maintain at
65%O2 saturation need to change another oxygen transfusion method.
2. Strictly prohibit provide humidified oxygen to the oxygen supply port.
Supply humidified oxygen bottles or other utensils will damage humidified
oxygen supply port inside the incubator.

OXYGEN MODULE OPEN/CLOSE SETTINGS

Under the keyboard unlocked condition, enter into second display interface D (see figure 6.4)
of oxygen concentration operation interface.
Press On / Off key to open or close the oxygen module. Under the opening state, oxygen
concentration display window can display the actual oxygen concentration inside of the incubator.
After entering into the normal condition, the oxygen concentration display window can indicate the
circular marks on the middle of the oxygen indicating area.

2015-03-24 10:10:00
SKIN

37.0 ℃ SKIN(℃) TIMES(12h) SET


3
7
.
0

45
SET ON/OFF
40
HEATER 35
TREND

30 HOME

25
0 t
AIR
25.0 ℃ HUM. %
○ 22
O2

%
21
WEIGHT g

1000
SET --.- SET 0 SET 20

O2 % O2 %
21 ○ 21
SET 20 SET 20

The start-up mark of oxygen input valve under The closed mark of oxygen input valve under
the oxygen concentration servo control the oxygen concentration servo control
system. system.

FIGURE 6.4

NOTE: 1. All settings only can be operated under the keyboard unlocked condition.
2. If not do anything within 10 seconds during the value setting procedure, the
indicator will automatically revert to first indication interface and save the result of
latest operation.

6-5
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

1 SET

Under the oxygen module opening condition, press this key to enter the oxygen concentration
setting mode, you can set the datum mark of oxygen concentration, and the setting value in
indicator will blink.

2
SET
20
Rotate this knob to left or right can increase or decrease the fiducial value.
Rotate right to increase the fiducial value, raise to 60%O2 from current value. Each rotation of one
case to rise 1%O2 fiducial value.
Rotate left to decrease the fiducial value, reduce 20%O2 from current value. Each rotation of one
case to reduce 1%O2 fiducial value.

3 ON/OFF

Press this key to close or open the oxygen module. Under the close condition, all of the relevant
oxygen function is in vain, including the oxygen failure alarm.

2015-03-24 10:10:00
SKIN

37.0 ℃ SKIN(℃) TIMES(12h)

45
SET

SET --.-
ON/OFF
40
HEATER 35
TREND

30 HOME

25
0 t
AIR
25.0 ℃ HUM. % O2
MODULE
% WEIGHT g
3
9
.
0

SET
○ 18 OFF 1000
SET 0 SET

4 CURVE

Press this key to enter the curve setting interface. Refer to Paragraph 4.4 for detail.

5 HOME
Press this key to confirm the new humidity set value, and return to first indication interface.

6-6

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

Press this rotate knob, the keyboard lock will enter controller set state (Display screen
display ). back to first interface after setting, press this knob to enter keyboard close
condition(Display screen display ), or don’t press any key , the keyboard lock will be closed
automatically after 10 seconds.

7
This key has two functions, when occur alarm indication, press once can cancel the alarm sound
and the top right corner of the display screen shows , press twice can reset the alarm state
then the equipment back to the set condition to monitor the alarm and top right corner of the
display screen shows .

NOTE: 1. All settings only can be operated under the keyboard unlocked condition.
2. If no keys are pressed within 10 seconds of selecting this Mode or setting
Humidity, it will automatically revert to first indication interface and save the result
of latest operation. (But for the step 2 inquiry screen, the indicator displays the
previous screen instead).

6.7 MAINTENANCE
A. Maintenance of oxygen concentration sensor
When the oxygen concentration lifetime is in due, replace it. Even it is workable for ensure
the veracity of the detecting value. That is because:
The detecting accuracy of oxygen concentration sensor is influenced by the electrolyte inside
the sensor modular, and the electrolyte will be consumed along with the working time. So using
the over lifetime oxygen concentration sensor will result in the wrong detected value.
B. Common fault
Troubleshooting of the oxygen concentration sensor is presented in the following table. If the
fault cannot be found out from the table, the equipment should be serviced by authorized and
qualified service personnel.

SYMPTOM POSSIBLE CAUSE REMEDY

No indication of oxygen
Not install the oxygen sensor install the oxygen sensor
concentration value
Please calibrate the oxygen
Oxygen concentration displayed sensor again
too high
Adjust the proper oxygen flow
The red alarm light will
blink, and the display Please calibrate the oxygen
screen will display "O2 Oxygen concentration displayed sensor again
Sensor Deviation Alarm” too low
Adjust the proper oxygen flow

Open the door and window Close the door and window

6-7
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

SECTION 7
HUMIDITY CONTROL SYSTEM
7.1 GENERAL
This section provides operation and checkout procedures and operation manual about

humidity control system. (This function only used in incubator mode)

7.2 FUNCTION DESCRIPTION


The humidity display range of the incubator is 0%RH~99%RH, setting range is 0%~

90%RH(Note:if the environment humidity is too high, that may cannot reach the lower humidity

control).The deviation alarm value is ±15%RH, under the steady condition, the incubator will

alarming if the humidity is ±15%RH higher or lower than the set value.

Water vapor will produced while the water in the humidity chamber flow by the heat pipe on

the bottom of the chamber, enters into the hood by vapor outlet port. The humidity control system

enables the humidity increasing inside the chamber. The evaporating speed decided by the

power of the heater. There’s a humidity sensor inside the sensor modular of the hood which

adjust the output power of the evaporator.

Please see the alarm introduce of section 3.4 for more details of the humidity control function.

If the alarms about humidity are active, press silence/reset key, cancel the alarm sound for 4mins,

the power of the temperature heater is not cut off, the power of the humidity heater will automatic

cut off when the high humidity alarm is active.

7.3 OPERATION CALIBRATION PROCEDURES


Operation the calibration procedures while the incubator is used for the first time or reuse

after cleaning and installation. Please refer to section 7.4 before calibrating the device to familiar

with the humidity setting of incubator.

A. Fulfill the water tank at the highest level with water, then place a hygrometer in the middle

of the box cover.

B. Pre-warm the incubator at 36℃, then set the humidity value at 50%.

C. Wait till the value of hygrometer rise to 50%RH, the humidity indicator should display

50%RH±5%RH.

D. Check the humidity deviation alarm

7-1
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

On humidity set mode, set the value at 40%RH, when the humidity reaches the setting value,
increase the humidity display value, make the display value +15%RH high than the set value, the
device should give a medium priority alarm, the “HUM. Sensor Deviation Alarm” information will
indication on the screen, the alarm character should be consistent with description in section 3.4.
Set the value at 40%RH, when the humidity reaches the setting value, decrease the humidity
display value, let the display value -15%RH lower than set value, the device should give a
medium priority alarm, the “HUM. Sensor Deviation Alarm” information will indication on the
screen, the alarm character should be consistent with description of section 3.4. Press
silence/reset key can cancel the alarm sound, but the alarm light still flashes until the display
value is within ±15%RH. If 4 mins later, the alarm condition is still not cancel, alarm sound will
start again.

NOTE:If the displayed value won’t rise or decline the deviation rate with set value over
±15% anymore, the alarm won’t occur. When checking the low-deviation
alarm, please make sure the environmental humidity is lower than the
humidity in the hood, with that the alarm can be active.

7.4 USING HUMIDITY CONTROL SYSTEM

WARNING: In any temperature, the relatively high humidity in the incubator will
decrease the heat vapor from the patient and result in increasing the
patient’s temperature. This is especially clear in the neonate. Therefore,
temperature control mode, temperature value and humidity value should
certified by the chief doctor. Monitor the patient’s recta and armpit
temperature according to the doctor’s instruction.

7.4.1 Preparation
Pull the water tank out as following figure shows, adding proper distilled water into the
humidifying chamber. Cannot over the high water level. Set the incubator in the Air mode. It can
pre-warm the air in the hood according to the set temperature which set by the main doctor or
chief nurse.
Affusion meatus humidity heater handle

Water level display window humidifying chamber


7-2

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

1. Strictly prohibit use the humidifying function under the water shortage
condition, that will cause the humidifying system damaged.
2. Must push the water tank in position while the incubator is operating, if not
that will affect the normal temperature control.
3. To extend the service life of humidity device, only can use the distilled water
to clean, the sterilized water cannot take place of distilled water.
4. The water tank should be thoroughly cleaned every 24 hours and replace
cleaned distilled water, to prevent the microbial propagation and the pollution of
humidifying chamber. The water changing should be performed after the water
tank is cooling sufficient, so as not to scald.
5. Strictly prohibit use the water tank as the handle to push the device, prevent
the water tank fall off or scald the operator.

IMPORTANT: 1. Only can use the distilled water to clean the water tank for lengthening the
humidity device service life, the sterilized water cannot take place of distilled
water.
2. Setting the humidity value at 0%RH without using the humidity function.

Under the condition of high temperature and high humidity (temp. ≥37℃,
humidity ≥85%RH), some disadvantageous results may caused to the
humidifier just like it may burn itself etc. Therefore, DO NOT touch the top of
heater, and wait no less than 45minutes when taking out the water tank or
restart the humidifier.

7.4.2 Operation
7.4.2.1 According to the physician's requirements to choose the control value of the humidity. on
the first screen, press humidity self-definite function key, please refer to the figure 7.1 to set the
humidity value.
NOTE: The actual humidity achievable inside the system is depend on the Incubator Set
temperature and environment conditions.

SELECTING HUMIDITY CONTROL AND SET POINT

When the keyboard is unlocked, press Humidity user-defined key to enter Figure 7.1 on first
indication interface.

7-3
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

2015-03-24 10:10:00
SKIN

37.0 ℃ SKIN(℃) TIMES(12h) SET

3
7
.
0
45
SET WEIGHT
40
AIR
HEATER 35
TREND

SKIN 30 HOME

25
HUM.
0 t
O2
AIR
25.0 ℃ HUM. %
○ 22
O2

%
21
WEIGHT g

1000
SET --.- SET 0 SET 20

Humidity control system is Humidity control system stops working


in the state of humidification
FIGURE 7.1

NOTE: 1. All settings should be operating under the keyboard unlocked condition.
2. If no keys are pressed within 10 seconds of selecting this Mode or setting
Humidity, it will automatically revert to first indication interface and save the result
of latest operation.

1 SET

Press this key to enter the humidity setting, the set value in humidity window will blink.

Rotate this rotate knob to right or left can increase or decrease the setting value.
Rotary knob to the right to increase the set value, set the value of the humidity rises from 0% RH
to 90% RH. Each rotation of one cell rises the value 1% RH.
Rotary knob to the left to reduce the set value, set the value of the humidity decline from 90% RH
to 0% RH. Each rotation of one cell decline the value 1% RH.

3 CURVE

Press this key to enter the curve interface, please refer to section 4.4 for details.

4 HOME

Press this key to confirm the new humidity set value, and return to first indication interface.

7-4

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

5
Press this rotate knob, the keyboard lock will enter controller set state (Display screen display ).
back to first interface after setting, press this knob to enter keyboard close condition(Display
screen display ), or don’t press any key , the keyboard lock will be closed automatically after
10 seconds.

NOTE: 1. All settings can be proceed under the keyboard unlocked condition.
2. If no keys are pressed within 10 seconds of selecting this Mode or setting
Humidity, it will automatically revert to first indication interface and save the result
of latest operation.(But for the step 2 inquiry screen, the indicator displays the
previous screen instead).

7.5 MAINTENANCE
Troubleshooting of the humidity control system is presented in the following table. If the fault
cannot be found out from the table, the device should be serviced by authorized and qualified
service personnel.

SYMPTOM POSSIBLE CAUSE REMEDY

Yellow alarm indicator light Environmental humidity Check the surrounding humidity
blinked, the display screen change greatly condition
display " HUM. Sensor
Deviation Alarm" No distilled water in the tank Add distilled water into the tank

Yellow alarm indicator light The distilled water in tank is


continuous on, the display Add distilled water into the tank
reduced to the lowest water
screen display " Water (Wash it when there's a need)
Shortage " level

Push the water tank to the correct


Yellow alarm indicator light Bad connection between
continuous on, the display position to fully connecting with
water tank and the main body
screen display " Tank the
Position Error " of the device
main body of the device.
Excessive microorganism
Thoroughly clean the water tank
Alarm not be active when breed in the water tank or the
water tank haven't been and clean around the detect point
water shortage
cleaned for many days. with soft cloth in disinfector.

Over-heat relay works Clean the water tank thoroughly


Heating circuit failure after it cool down, use the soft
because the dusts near the
alarm cloth with disinfector to wipe
probe. around the detect point.
Fill the water tank with
water after water Humidifier over-heat
Waiting for about 10 minutes
shortage, "humidify" can't protection not recovered
be started immediately
7-5
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

SECTION 8
WEIGHING SYSTEM
8.1 GENERAL
This section provides operation checkout procedures and operation manual about weighing
system.
Weighing system is optional part.

8.2 INSTALLATION
Install the weighing system on the equipment as the figure 8.1 shows.

Open the panel, make the rotating bed at the horizontal position, put the bassinet with baby
scale at the rotating bed and make sure it can push, pull or rotate 360 degrees as the arrow
direction.

Socket of weighing sensor Arrow Gap

Make the weighing sensor can be inserted into the gap of socket correctly.

WARNING: 1. When plug in or out the power supply cord, the plug must be taken hold
and the wring should not be pulled.
2. Do not bend the connection of sensor.

Connect the plug of sensor on the baby scale with the socket of sensor box correctly.

FIGURE 8.1

8-1
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

8.3 FUNCTION DESCRIPTION


The weighing system is equipped with the operation of calibration, and the balance must be
able to zero. User can do this operation by pressing the self-definite function key on the panel of
temperature controller. The display range of weighing system is 100~8000g, and its display
distinguish rate is 1g, the display precision is ±1%, and the reset range is 0~2000g.

8.4 OPERATION CALIBRATION PROCEDURES


NOTE: Calibration the weight sensor after the incubator first install or maintenance.

CALIBRATION OF WEIGHT SENSOR

As the figure 8.2 shows, when the keyboard is unlocked, press the weighing self-definite
function key to enter the interface to calibrate the weight sensor in humidity interface of the
second interface C.

SET
AIR

SKIN WEIGHT

HUM. TREND

HOME
O2

2015-03-24 10:10:00
SKIN

37.0 ℃ SKIN(℃) TIMES(12h) TREND


3
7
.
0

45
SET CAL.

40
HEATER 35
ZERO

30 HOME

25
0 t
AIR
25.0 ℃ HUM. %
○ 21
O2

% WEIGHT g
21 1000
SET --.- SET 0 SET 20

FIGURE 8.2

NOTE: 1. All settings should be done under the keyboard unlocked condition.
2. Whichever operations, If not proceed the next step after pressing the key for 10s,
and then the indication can come back to the first interface and restore the
operation result of last time.

8-2

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

A. Take out all the things of the infant bed and make sure the bed surface clean.
B. Press the reset self-definite function key, weight indicator should display 0000g±5g, if
exceeds this range, press reset key again.
C. Put the 5000g standard weight on the baby scale, then press Calibration self-definite
function key, the system will enter the calibration condition and display “wait for calibrating”.
The indication window can indicate the weight within 5000g±5g. while exceeds the range, repress
Calibration self-definite function key until the weight is in the right range. If failed, the baby scale
should be repaired by qualified personnel.

CHECK THE CONNECTING OF WEIGHT SENSOR

Disconnect the connection between weight sensor and sensor box. At the first interface will
display the sensor error information as the figure 8.3 shows. The controller will auto cancel the
failure indication information and indicate the weight of sensor after reconnecting.
2015-03-24 10:10:00
SKIN

37.0 ℃ SKIN(℃) TIMES(24h)

45
TREND

SET --.-
CAL.

40
HEATER 35
ZERO

30 HOME

25
0 t
AIR
25.0 ℃ HUM. % O2 % WEIGHT g
3
6
.
0

SET
○ 31 ○ 21 SENSOR
FAULT
SET 0 SET 20

FIGURE 8.3

OPERATE ALL KEYS OF WEIGHT INTERFACE

NOTE: All of the keys can be operated while the keyboard lock indicator lamp is on.

1 CURVE

Press this key to enter the curve setup interface. Refer to the section 4.4 for details.

2 CALIBRATION

This key is just used for calibrating the weight sensor; do not press this key in normal operation.

3 RESET

Press this key to make the weight value back to zero, the range is 0~2000g.

HOME
4
Press this key to confirm the new setting value and back to the first indication interface.

8-3
I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

8.5 OPERATION OF WEIGHING SYSTEM


After calibrating and reset the weighing system then place in the patient, the Weight indicator
will display the patient’s weight measured by weight sensor.

8.6 MAINTENANCE
Troubleshooting of the weight system is presented in the following table after long time using.
If the fault cannot be localized from the table, please call the authorized and qualified service
personnel to repair.

SYMPTOM POSSIBLE CAUSE SOLUTION

No display on the Turn on the power switch of the


No power supply
Weighing indicator controller
Take out all articles on the infant
The reset range is over
bed and make sure the bed
0~2000g
Weight indicator displays surface clean
“Weighing Sensor Fault”
Weighing sensor is not Correct connect the weighing
connected sensor

The infant bed is not placed Adjust the infant bed into
in horizontal position horizontal position

Infant is not placed in the center Lay the infant in the center of
Weighing inaccurate and of the mattress the mattress
the weight value is not
stable Environmental temperature Check the environmental
varies greatly temperature

Weighing sensor have been Please calibrate it again with a


incorrect calibrated 5000g standard weight

8-4

I12-1
OPERATOR'S MANUAL FOR INFANT INCUBATOR

SECTION 9
PARTS LIST
This section provides the lists of accessories and removable parts of the incubator. The
users only can use the materials provided by our company, or it will reduce the safety of
equipment.

Replacement
NO. Name of Parts Serial Number
Period

1 Skin temperature sensor 88.04.001.00 2 years

Clearly dirty or
damaged or
2 Air-purifying filter 16.01.036.00
used more than
2 months

Window seals 16.01.093.00


Access door
3
gaskets
Window sealsⅡ 17.01.328.00

4 I.V. soft port 16.01.099.00


If damaged
5 Power cord 86.46.013.00

6 Oxygen input pipe 13.16.036.02

7 Oxygen calibration tool box components 16.01.354.00S

Use 10000hrs
at most under
8 Oxygen sensor 86.60.093.00
100% oxygen
concentration

9-1
I12-1
SPECIAL STATEMENT: All of the content in the manual is checked carefully, if there is any
error or content of printing misunderstanding, our company retains
finally explanation of this card-usage.
NOTE: The product’s appearances maybe differ from the one in this manual, but it does not
affect the capability of product. Please understand if it brings you troubles.

NINGBO DAVID MEDICAL DEVICE CO., LTD.


ADD: NO. 35, JINXING ROAD, BINHAI INDUSTRIAL PARK, XIANGSHAN ECONOMIC DEVELOPMENT

ZONE, NINGBO, ZHEJIANG PROVINCE, P.R.CHINA

POST CODE: 315712 FAX: +86-574-65962111

MARKETING CENTER: NO. 100, JINGHUA ROAD, HI-TECH INDUSTRIAL DEVELOPMENT ZONE,

NINGBO, CHINA

POST CODE: 315040

TEL: +86-574-87800008, 87800007 FAX: +86-574-87801111, 87803714

E-mail: sales@nbdavid.com

http://www.chinadavid.cn

EDITION/REVISION A/0