The change you need
KEY FEATURES
Fully Automatic, Random Access Clinical Chemistry Analyser with direct reading system.
System having throughput of 100 Tests/hour.
Reusable cuvettes with efficient on-board laundry.
STAT facility available for all sample positions.
Excellent On-board refrigeration for reagents.
Efficient water consumption.
TECHNICAL SPECIFICATIONS
Assay Model Random access with direct reading system
Method End point, Kinetic, Fixed time, Immunoturbidimetric, 1-2 reagent,
Multistandard, reagent/serum blank.
Assay Item 36 assay
Throughput 90 Tests/hr
Sample Position 50
Including standard, control and stat; either serum cup or primary tube
available
Sample Volume 1‐100 µl, 0.1 µl/ step
Reagent Position 36
Reagent Volume R1: 1‐400 µl, 1 µl/ step, R2: 1-400 µl, 1 µl/step
Reagent/ Sample Probe With liquid level sensor, collision sensor, Teflon coating and automatic
washing station
Reaction Time 0‐999 seconds
Cuvette 60 reaction cuvettes (High quality UV- transmitted plastic cuvette)
Cuvette Washing 2 Times washing
Probe and Mixer Washing Independent washing station for each sample/reagent probe and mixer
Alarm Automatic alarm for malfunction (eg: collision, reagent/sample
insufficiency, full waste container, etc)
Calibration Linear, Non-linear, Multi standard, K factor
Sample Dilution/ Retest Sample can be diluted and retested if results are out of range or sample
in insufficient. Also with calibrator auto diluting function
Wavelength 340‐800nm
Light Source Halogen lamp
Absorbance Resolution 0.0001A
Absorbance Accuracy ±0.0003A (0-2.5A)
Repeatability (CV %) ! 2%
Quality Control Levy Jennings quality control program at 3 levels
Bar Code Optional
Printer Multiple report formats available
System Windows XP or Vista; Windows 7
Power Supply 220V/ 50Hz or 110V/ 60hz (optional)
Dimension 65x53x52 (cm)
Weight 45 kgs
Cat No. T19953
LIST OF COMPONENTS SUPPLIED ALONG WITH THE INSTRUMENT
Sr. No. Components Quantity Sr. No. Components Quantity
1. Power Cord 1 No. 10. Reagent Bottle R2 10 Nos
2. Fuse 4 Nos. 11. Sample Cups 400 Nos.
3. Wash Water Inlet 1 No. 12. Reagent Bottle Caps 40+10 Nos.
4. Water & Waste Sensor 1 No. 13. Halogen Lamp 1 No.
5. Tube Kit 1 No. 14. Allen Key Set 1 No.
6. Reaction Cuvettes 90+10 Nos. 15. Precision Screw Driver 1 No.
7. 3 Pin Top 1 No. 16. Sample Probe 1 No.
8. Serial Cable RS232 1 No. 17. Sensor 1 No.
9. Reagent Bottle R1 40 Nos 18. Ground Wire 1 No.
ACCESSORIES/CONSUMABLES FOR USE WITH CORALYZER MINI
Accessories/Consumables Cat No. Pack Size
Serial Cable T19429 1 No.
Sample cups T19266 200 Nos
Reaction Cuvettes (90+10) T19462 90 + 10 Nos
Tube kit T14537 1 No.
Wash water inlet T19431 1 No.
Water sensor T19428 1 No.
Waste sensor T19461 1 No.
Halogen Lamps T19423 1 No.
Sample probe T19240 1 No.
Ground wire T19430 1 No.
R2 Adapters T19845 10 Nos.