Technical Publications: Direction 2305203-100 Rev. 004
Technical Publications: Direction 2305203-100 Rev. 004
Technical
Publications
Direction 2305203-100
Rev. 004
The Vivid 3 Systems meet the provisions of the European Council Directive 93/42/EEC concerning medical devices which apply to them.
Conformity assessment was performed according to Annex II by KEMA.
Any changes to accessories, peripheral units or any part of the system must be approved by the manufacturer, GE Medical Systems. Ignoring this advice may compromise
the regulatory approvals obtained for the product.
Requirements given by local laws/acts/ordinances must be followed by the user/installer. [For example, German: German Medical Device Act (Medizinproduktegesetz) and
German Medical Device Ordinance (Medizinprodukte-Betreiberverodnung).]
                                                                                      Standard/Directive              Scope
       0344                                                                           89/336/EEC                      EMC Directive
 GE ULTRASOUND ISRAEL PRODUCTS:                                                       93/42/EEC                       Medical Device Directive
 GE ULTRASOUND ISRAEL products are tested to meet all applicable requirements in
 relevant EU Directives and European/International standards (See “Standards used     IEC 801-2                       Electrostatic Discharge
 below”.)                                                                             IEC 801-3                       Radiated Electromagnetic Field
 Any changes to accessories, peripheral units or any part of the system must be
 approved by the manufacturer, GE Ultrasound Israel. Ignoring this advice may         IEC 801-4                       Electrical Fast Transient/Burst
 compromise the regulatory approvals obtained for the product.                        IEC 805-1                       Surge
 Please consult your local GE ULTRASOUND ISRAEL representative for further details.
                                                                                      EN 55011/CISPR 11               Electromagnetic Susceptibility
 GE ULTRASOUND EUROPE, Regulatory Affairs Dept. Beethoven Str. 239,Solingen
 Germany, tel +49-(0)212-2802-243, has been empowered to enter into commitments       EN 60601-1/IEC 601-1/UL 2601-1 Medical Electrical Equipment; General
 on our behalf in any matter concerning the provisions of the 93/42/EEC directive.                                   Requirements for Safety
 Standards used:                                                                      EN 61157/ IEC 61157             Requirements for the declaration of the acoustic
 To fulfill the requirements of relevant EC directives and/or European Harmonized/                                    output of medical diagnostic ultrasonic equipment
 International standards the following documents/standards have been used:
                                                                                      IEC 60601-2-27                  Medical electrical equipment - Part 2: Particular
                                                                                                                      requirements for the safety of electrocardiographic
                                                                                                                      monitoring equipment.
                                                                                      IEC 60601-1-4                   Medical electrical equipment: Part 1-4: General
                                                                                                                      requirements for collateral standard: Programmable
                                                                                                                      electrical medical systems
                                                                                      ISO 10993-1                     Biological evaluation of medical devices
                                                                                      ISO 13485                       Quality management standards for medical devices
GE Medical Systems
Revision History
       PAGE                      REVISION
      NUMBER                     HISTORY
       5 – 81                       0
         All                        1
         All                        2
         All                        3
         All                        4
       Please verify that you are using the latest revision of this document. Information pertaining to this document is maintained on GPC (GE Medical
       Systems Global Product Configuration). If you need to know the latest revision, contact your distributor, local GE Sales Representative or in the
       USA, call the GE Ultrasound Clinical Answer Center at 1-800-682-5327 or 262-524-5698.
Quick Guide                                                                     Direction 2305203-100 Rev. 004                                                                                        i
    Table of Contents
Quick Guide                                                                    Direction 2305203-100 Rev. 004                                                                                       ii
Contraindications
The GE Vivid 3 ultrasound unit is not intended for opthalmic use or any use
causing the ultrasound beam to pass through the eye. The acoustic levels at
which this system operates exceed the FDA limits for opthalmic use.
    Bold type       Denotes buttons, soft keys, soft key rotaries and
                    field names that are displayed on the screen.
                                                                                    For additional information, see “Safety” on page 88.
    Italic type     Denotes the names of windows, screens and dialog
                    boxes, for example, New Patient window.
Getting Started                                         10. Foot Brake                                          15. Floppy/MO Disk (optional)/CD-R Drive
                                                            Three-position brake. LEFT locks the wheels,        16. Air Filter Vent
For detailed information, see Chapter 2, Getting            MIDDLE unlocks the wheels, and RIGHT locks
                                                                                                                17. ECG Cable (Internal/External) Phono
Started in the Vivid 3 User’s Manual.                       the swivel action.
                                                                                                                    Connections
                                                        11. Footswitch (Not Shown)
                                                                                                                18. Footswitch Connection
                                                            Configurable footswitch that enables certain
Front and Side View                                         keyboard commands to be operated by foot.
1. Display Monitor                                      12. Side Bay                                            Rear View
   Swivels to the left and right, and tilts up and          The color printer or JVC VCR is contained in
                                                            the upper-left side bay of the unit.                1. Monitor cables, video and power cables
   down.
                                                        13. Peripherals Storage Area                            2. Power cable storage hook
    •     15" Monitor
                                                            All VCR recorders (excluding JVC VCRs) and          3. Circuit breaker On/Off switch
    •     17" Monitor
                                                            black & white printer are positioned on the         4. Ground Screw
2. Speakers                                                 peripheral storage area under the front panel.      5. Main power cable socket
3. Alphanumeric Keyboard                                14. Cable Hook
4. On/Off Switch
5. General Storage Areas
   Situated on either side of the front panel on the
   top surface, on either side of the monitor.                                                             1
    Either:                                                                                                                                                 3
    •     Two active probe ports, one park probe port                                                      10
          and one pencil probe port.                                                                                                                        4
    Or:                                                                                                                                                     5
                                                                        Getting Started
Vivid 3 User Guide                                          Direction 2305203-100 Rev. 004                                                                          5
                                                                      Getting Started
Vivid 3 User Guide                                          Direction 2305203-100 Rev. 004   6
Electromagnetic Interference
                                                                   Getting Started
Vivid 3 User Guide                                         Direction 2305203-100 Rev. 004                                                                     7
Connecting the Unit                                   Voltage Level Checks                                 Connecting to the Electrical Outlet
Before connecting the System, perform preliminary
                                                      Before connecting the System to the power source,    Before connecting the System to the electrical
checks of the power cord, voltage level, and
                                                      perform the following voltage level checks:          outlet, ensure that the wall outlet is of the
compliance with electrical safety requirements. Use
                                                                                                           appropriate type, and that the power switch is
only the power supply cords, cables, and plugs        1. Check the label near the mains input connector.
                                                                                                           turned off.
provided by, or designated by, GE Medical                A yellow label indicates the input connector is
Systems. Extension cords must not be used.               either 100 V, 120 V, 220 – 240 V.                 NOTE: To assure grounding reliability, connect to a
Ensure that the power cord and plug are intact and    2. Check the voltage indicated on the label:         "hospital grade" or "hospital only" grounded power
that the power plug is the proper hospital-grade                                                           outlet.
                                                      •   100 V 50 – 60 Hz 8A
type.
                                                      •   120 V 50 – 60 Hz 18A                             To connect the unit to the electrical outlet:
NOTE: A GE-qualified person should perform the
                                                      •   220 – 240 V 50 – 60 Hz 4A                        1. Uncoil the power cable, allowing sufficient slack
initial system installation.
                                                          Maximum power requirement = 1.2 KVa                 so the plug will not pull out of the power socket
Products equipped with a power source plug should                                                             if the unit is moved slightly.
be connected to the fixed power socket which has
                                                          Maximum allowed voltage deviation = + 10
                                                                                                           2. Secure the power plug in the mains power
the protective grounding conductor. Never use any                                                             socket.
                                                                   If the mains supply is not within the
adapter or converter to connect with a power
                                                                   specified range, DO NOT CONNECT
source plug (for example, three-prong to two-prong                                                                       The unit’s power must be supplied
                                                                   THE UNIT TO THE POWER
converter).                                                                                                              from a separate, properly rated outlet
                                                                   SOURCE. Contact the dealer to have
                                                                   the unit adjusted for the specific                    to avoid the risk of fire. Refer to the
             Failure to provide an adequate                                                                              Power Requirements section in
             ground circuit can cause electrical                   mains supply.
                                                                                                                         Chapter 2 of the Vivid 3 User’s
             shock, resulting in serious injury.                                                                         Manual for rating information. The
                                                      Electrical Safety Requirements                                     power plug should not, under any
                                                                                                                         circumstances, be altered to a
                                                      Ensure that all equipment connected to the System                  configuration rated less than that
                                                      complies with national safety requirements for                     specified for the current. DO NOT use
                                                      medical equipment. These safety requirements                       an extension cord or adapter plug.
                                                      include: IEC60601-1, CSA22.2, AS3200.1, and UL
                                                      2601-1.
                                                                       Getting Started
Vivid 3 User Guide                                                  Direction 2305203-100 Rev. 004                                                                          8
                                                                             Getting Started
Vivid 3 User Guide                                              Direction 2305203-100 Rev. 004                                                                         10
5. Place all probes in the probe holders and              To ensure safety while moving the unit on foot:         Transporting the Unit
   ensure that the probe cables do not protrude
                                                          •   Proceed cautiously when crossing door or
   from the unit or interfere with the wheels. Store                                                              Take extra care when transporting the unit by
                                                              elevator thresholds.
   all other probes in their original cases, soft cloth                                                           vehicle. In addition to following the moving
   or foam, to prevent damage.                            •   Grasp the front handle grips or the back handle
                                                                                                                  precautions listed above, make sure to follow the
                                                              bar and push or pull. DO NOT attempt to move
                                                                                                                  safety precautions below.
Important: Before moving to a new site, remove all            the unit using cables or probe connectors. Take
the probes and pack them in their original cases, soft        extra care while moving the unit long distances
                                                                                                                  Important: Save and reuse the original packaging if
cloth or foam, to prevent damage.                             and on inclines.
                                                                                                                  moving the unit to other sites. GE service
                                                          •   Ensure that the unit does not strike walls or       representatives or their assignees should perform the
6. Ensure that no loose items are left on the unit.           door frames.                                        move. A specially designed van should also be used.
   All loose items (such as: gel, ECG cables, CDs         •   Ensure that the pathway is clear.
   and optical disks), should be stored in the unit                                                               To ensure safety while transporting the unit by
                                                          •   Move the unit slowly and carefully.
   pockets.                                                                                                       vehicle:
                                                          •   Know where the foot brake is located and learn
7. Unlock the brake by pressing down on the                                                                       1. Disconnect all probes and secure them in their
                                                              how to use it. The foot brake must be set when
   pedal to return it to its middle position.                                                                        boxes.
                                                              moving of the unit stops.
To prepare the unit to be moved                                                                                   2. Use the control panel release handle, situated
                                                          •   One adult can usually move the unit along a
(Plug&Scan not installed) or for long term
                                                              level surface for short distances. Use two or          under the center front of the control panel, to
storage or long distance transportation:                                                                             lower the control panel to its minimum height.
                                                              more persons to move the unit on inclines or
1. Initialize a Full Shutdown, as described in the            over long distances.                                3. Ensure that all peripherals are secured.
   Full Shutdown section in Chapter 2, Getting
                                                                                                                  4. Ensure that the transporting vehicle is
   Started in the Vivid 3 User’s Manual.                                Avoid inclines that are steeper than
                                                                                                                     appropriate for the unit’s weight. The
2. Set the circuit breaker to OFF.                                      10 degrees, or bumps higher than
                                                                                                                     recommended load capacity is a minimum of
                                                                        2.5 cm (1 in), to prevent the unit from
3. Remove the plug from the mains power socket.                                                                      160 - 168 kg (353 - 370 lbs). The vehicle should
                                                                        tipping over. After reaching the
4. Follow steps 3 – 7 of the previous procedure.                                                                     have good shock absorbers and a door large
                                                                        destination, lock the front wheel
                                                                                                                     enough for the unit with its monitor and any
                                                                        brake.
                                                                                                                     peripherals to clear.
                                                                                                                  5. Park the vehicle on a level surface for loading
                                                                                                                     and unloading.
                                                                                                                  6. Secure the unit while it is on the lift, to prevent
                                                                                                                     rolling. Do not attempt to hold it in place by
                                                                                                                     hand. Cushion the unit and strap the lower part
                                                                                                                     so that it does not break loose.
                                                                                            Getting Started
Vivid 3 User Guide                                         Direction 2305203-100 Rev. 004                  12
                                      Monitor
                                      Adjustment
                                      Controls
                                                                                                     Basic Mode
                                        Archiving &                                    Scan Mode     Parameter
                             On/Off     Reporting      TGC Gain                        Selection     Adjustment
                             Button     Buttons        Sliders        Soft Keys        Buttons       Rotaries
        VCR
        Playback
                                                                                                                    Display
        Pre-examination                                                                                             Format
            Buttons                                                                                                 Buttons
                                                                                                                     Measurement
                                                                                                                     Buttons
                                                                                                                    VCR Control
                                                                                                                    and Print
                                                                                                                    Buttons
                                              Figure 3: Control Panel Layout Diagram               NOTE: For details on Soft Keys and Soft Menu
                                                                                                   Rocker functions, see “Soft Key and Soft Menu
                                                                                                   Functions” on page 21.
                                                                Getting Started
Vivid 3 User Guide                                              Direction 2305203-100 Rev. 004                                                                        14
Beginning an Examination                                  To enter a new patient:                                To select a preset from the top of the image screen:
                                                          1. Press Patient ID.                                   NOTE: A green triangle is located at the top of the
Entering Patient Details                                  2. TRACKBALL to Start New Patient in the Select        Image screen alongside the name of the
                                                             screen and press Select. The Main Details           currently-selected preset. Clicking on the triangle
To access patient information:                               screen is displayed.                                opens a pull-down menu in which all presets
                                                                                                                 available for the currently-selected probe are listed.
•     Press Patient ID, Date of Birth, or Last Name.
                                                          Important: Either an ID Number or Last Name is         NOTE: On the image screen, when the cursor is in
Important: The layout of the Patient List screen is       mandatory. All other patient details are optional.     the area around the preset name or the green
dependent on the screen layout template selected                                                                 triangle, the cursor changes to a hand-shaped
during system configuration. For details, refer to        3. TRACKBALL to the Patient ID or Last Name field      pointer. Clicking Select opens a drop-down menu;
Chapter 14, Presets in the Vivid 3 User’s Manual.            and press Select.                                   a checkmark is visible preceding the name of the
                                                          4. Enter the patient’s ID number or last name.         currently-active preset (Cardiac in the example
NOTE: The screen layout is displayed in accordance        5. Enter additional patient details in the remaining   shown below):
with the currently-selected Patient ID Page Type. (For       fields, using <Tab> or <Enter>, navigate to
details, refer to Vivid 3 Pro/Expert SW 2.2 Release          each field,
Notes). Any one of the following four types may have      6. Use the Select Page soft key rotary to navigate
been be selected:                                            through the data entry pages. Add patient
    • Type A:Full details - includes Patient ID, Weight      details as required.
              and Height                                  7. Press the Begin Exam soft key.
    • Type B:No Patient ID - includes Last Name,
              DOB, Weight and Height                      To select a probe and an application:
    • Type C:Patient ID on 2nd line; No Weight or         1. Press Probe.                                        1. Click the triangle at the top of the image screen (on
              Height                                      2. TRACKBALL to the required application preset in        the right of the currently-selected preset name).
    • Type D:No Patient ID; No Weight or Height              the frame of the appropriate probe and press        2. From the pull-down menu, trackball to the
                                                             Select. The name of the currently selected             required preset and press Select.
                                                             probe and application are displayed in the
                                                                                                                     The pull-down menu closes and the name of
                                                             information bar at the top of the 2D screen. The
                                                                                                                     the newly-selected preset is displayed at the
                                                             examination can now begin.
                                                                                                                     top of the screen.
                                                                                                                 NOTE: The names listed in the pull-down menu are
                                                                                                                 relative to the currently-selected main application.
                                                                                                                 User-modified preset names are preceeded by an
                                                                                                                 asterisk (*).
Preset
                                                               Getting Started
Vivid 3 User Guide                                            Direction 2305203-100 Rev. 004                                                                        16
Basic Scanning Operations                               3. Use the alphanumeric keyboard to type the            Zoom
                                                           required text. To add additional annotations,
The System enables the user to perform various             press Menu.                                          Magnifies the display in both frozen and live 2D,
supplementary operations to improve both the
                                                        To move a line of text:                                 M-Mode and combined mode images.
examination process and quality of the data
archived for future use.                                If required, a line of text may be moved                To zoom an image:
                                                        (up/down/left/right)
For detailed information regarding scanning mode                                                                1. Rotate the zoom rotary to magnify the image.
functions, refer to Chapter 3, Modes in the Vivid 3     1. Trackball to the line and press Select.              2. Press the Zoom rotary to toggle between the
User’s Manual.                                          2. While holding down the Select button on the             actual-sized image and the magnified image.
                                                           control console, drag the selected line of text to
Storing Images and Cineloops                               the required position, using the trackball.          Performing Measurements
                                                        3. Release the Select button. The line of text will
Images and cineloops can be stored at any time                                                                  To perform measurements:
                                                           remain fixed in the new position.
during the scanning session, by using the Store
soft key.                                                                                                       •   Press Caliper or Measurement. See Chapter
                                                        Deleting Text Annotations                                   6, Cardiac Measurement and Analysis and
NOTE: The Store soft key should be pressed twice.                                                                   Chapter 7, Vascular Measurement and Analysis
The first time to preview the image, the second to      To delete a specific text annotation from the               in the User’s Manual for details.
accept and store the image.                             screen:
For further information, refer to Chapter 10, Patient   1. TRACKBALL to the annotation to be deleted and
Management in the Vivid 3 User’s Manual.                   press Select to highlight.
                                                        2. Press <Line Erase>. The selected annotation
Storing Images on VCR                                      is erased.
                                                        NOTE: If you did not select the annotation (Step 1),
Images can be recorded on a video tape. Press
                                                        the last line typed will be erased.
VCR Record to start recording. For detailed
information regarding the VCR recorder, refer to
Chapter 12, Peripherals in the Vivid 3 User’s           To erase all the annotations from the screen:
Manual.
                                                        •   Press <Page Erase>.
Text Annotations
                                                        Depth Adjustment
To insert a text annotation:
                                                        To adjust the image depth:
1. Press <Text>. A text cursor (I) is displayed in
   the upper-left corner of the display screen, or in   •   Rotate the Depth rotary.
   the user-defined home position.
2. TRACKBALL the text cursor to the position at
   which text is to be inserted.
For additional information, see Chapter 3, Modes in    To manually select a heartbeat:                         To split the screen:
the Vivid 3 User’s Manual.                             1. While in any scanning mode, press Freeze.            •   Press 1/2/4 once to split the screen into two
                                                          The scan mode freezes and the First Marker               windows.
                                                          and End Marker are displayed on either side of       •   Press the button again to split the screen into
2D-Mode Imaging                                           the last detected heartbeat on the ECG trace.            four windows
                                                       2. Press the Cine Scroll soft key rotary to start the   •   Press the button a third time to revert to a single
Accessing 2D-Mode                                         cineloop playback.                                       image display.
                                                       3. Use the Previous/Next soft key rotary to select      To toggle between windows:
To access 2D-Mode:
                                                          the heart cycle to be played back.
•   Press 2D.                                                                                                  •   Press Window Select. The active window is
                                                       4. Use the First Marker and/or End Marker soft
                                                                                                                   marked with a solid light-blue circle near the top
                                                          key rotaries to trim or expand the cineloop
Adjusting the Gain in a 2D Image                                                                                   of the image and a yellow frame. Non-active
                                                          boundaries.
                                                                                                                   windows are marked with a hollow circle near
                                                       NOTE: By default, one cycle is stored in one cine.          the top of the image and no frame.
Use the 2D Gain rotary to adjust the overall gain on   However, up to 8 cycles may be stored. Select the
both live and frozen images, as well as on images                                                              To enlarge the display area:
                                                       required number of cycles, using the Cycles for
retrieved from the image archive.                      Store soft key:                                         •   Press Image Size. The images will increase in
                                                                                                                   size, and the parameter display column will not
Using Zoom in 2D-Mode                                                                                              be displayed.
•   Rotate the Zoom rotary counterclockwise. The       •   Rotate the Cine Speed soft key rotary to adjust
    ROI border increases in size, producing a              the frame-rate of the cineloop.
    smaller enlargement (zoom factor) of the larger        The default is Nominal speed.
    area. Continued counterclockwise rotation          To view a cineloop frame by frame:
    reaches a point at which the zoom is
    automatically switched off and the reference       1. Press Stop Cine.
    image is no longer displayed.                      2. Use the Cine Scroll soft key rotary.
To exit the zoom function:
•   Press the Zoom rotary once after the ROI has
    been sized and positioned as required.
                                                                       Getting Started
Vivid 3 User Guide                                           Direction 2305203-100 Rev. 004                                                                    18
                                                                       Getting Started
Vivid 3 User Guide                                         Direction 2305203-100 Rev. 004                                                                       19
Resizing/Repositioning the M-Mode                    Doppler Mode Imaging                                   5. While the spectrum is scrolling, use the
Display                                                                                                        TRACKBALL to reposition the Doppler cursor
                                                                                                               and/or gate position.
                                                     Regular Doppler Display
To resize and/or reposition the image display:                                                              6. Press Active Image to toggle between the
                                                     To access PW Doppler Mode or CW Doppler                   active soft key menus.
1. Press the Image Size soft key to toggle
   between the M scroll sizes: either half, two      Mode:                                                  7. Press Select or 2D to freeze the Doppler
   thirds, or full display.                                                                                    spectrum. Then press 2D Update to update the
                                                     1. Perform a 2D-Mode scan.
                                                                                                               2D image.
2. Press Layout to toggle between a top/bottom       2. Press either PW or CW. The Doppler cursor
   or side-by-side split screen display.                                                                    To access PW Doppler Mode or CW Doppler
                                                        (and the Doppler gate, when in PW Doppler
                                                                                                            Mode (Triplex Doppler display):
                                                        Mode) is displayed on the image.
Enlarging the 2D Image                               3. Use the TRACKBALL to position the Doppler           1. Perform a 2D-Mode scan.
                                                        cursor (and Doppler gate, when using                2. Press either PW or CW on the control panel.
To display a full-sized 2D image:                       PW-Mode) over the appropriate anatomy.                 The Doppler cursor (and the Doppler gate,
1. In M-Mode, press Freeze. The M-Mode display       4. When positioned, press PW or CW to run the             when in PW Doppler Mode) is displayed on the
   is frozen.                                           PW or CW spectrum.                                     image.
2. Press M. The M scroll will temporarily            5. While the spectrum is scrolling, use the            3. Use the TRACKBALL to position the Doppler
   disappear, displaying a full-sized 2D image.         TRACKBALL to reposition the Doppler cursor             cursor (and Doppler gate when in PW-Mode)
                                                        and/or gate position.                                  over the appropriate anatomy.
3. Press M again to re-display the original M-Mode
   layout.                                           6. Press Select or 2D Update to freeze the             4. Once positioned, press PW to run the PW
                                                        Doppler spectrum and update the 2D image.              spectrum.
Exiting M-Mode                                                                                              NOTE: Only in specific velocities will Interleave
                                                     Duplex Doppler Display                                 Triplex be activated automatically.
To exit M-Mode:                                                                                             5. While the spectrum is scrolling, use the
•   Press M. The image reverts to 2D-Mode. The M     To access PW Doppler Mode or CW Doppler                   TRACKBALL to reposition the Doppler cursor
                                                     Mode:                                                     and/or gate position.
    cursor remains on the image; pressing M again
    will re-enter M-Mode.                            1. Follow steps 1 – 4 above.                           6. Press Active Mode to toggle between the
To access Anatomical M-Mode:                         2. Press one of the arrows on the Soft Menu               active soft key menus.
                                                        Rocker. A pop-up menu is displayed on the           7. Press Select or 2D to freeze the Doppler
1. Perform a 2D-Mode scan and press Freeze.
                                                        screen.                                                spectrum, then press 2D Update to update the
2. Press M to generate the Anatomical M-Mode                                                                   2D image.
                                                     3. Use the vertical arrows to highlight the Enable
   sweep.
                                                        Triplex option from the soft menu.                  8. Press 2D Freeze to freeze the 2D display, if
To position the M cursor:                                                                                      required.
                                                     4. Use the left or right arrows to change the status
1. Rotate the Line Move soft key rotary clockwise       to 1 (enabled). The display changes to Duplex
   to move the M Cursor to the right.                   Doppler, and both the 2D image and Doppler
2. Rotate the Line Rotate soft key rotary               Spectrum are active.
   clockwise to rotate the M cursor to the right.
                                                                    Getting Started
Vivid 3 User Guide                                           Direction 2305203-100 Rev. 004                                                                        20
Adjusting the Gain in Doppler Mode                     LVO Imaging (optional)                                    •   Tissue harmonics
                                                                                                                     Tissue harmonics give rise to contrast-like
                                                       Use the Left Ventricular Opacification Imaging (LVO           signals independent of the presence of the
In both CW Doppler and PW Doppler Modes, the
                                                       Imaging) application to help identify LV thrombus             contrast agent.
Active Gain rotary enables the overall spectral gain
                                                       and evaluate wall motion.
to be adjusted in either active or freeze modes and                                                              The LVO Imaging application default settings
on images retrieved from the image archive.            Support of LVO Imaging is an optional feature of the      minimize the occurrences of artifacts. Nevertheless,
                                                       Vivid 3 system.                                           the operator has to take these into account when
The 2D Gain rotary enables the overall gain of the
2D image to be adjusted, without affecting the         The LVO Imaging application is preset to:                 analyzing the data and has to confirm the results
Doppler spectrum.                                                                                                using other techniques.
                                                       •   optimize resolution and enhance delineation of
                                                           endocardial borders, and
Enlarging the 2D Image
                                                       •   optimize assessment of wall motion and wall
                                                                                                                 Optimizing LV Contrast
                                                           thickening.
To temporarily hide Doppler display:
                                                                   WARNING: Read and carefully follow            Usually, the default settings for the LVO Imaging
1. With PW/CW image frozen, press PW or CW. A
                                                                   the manufacturer’s instructions on the        application are the optimal settings. However,
   full-sized 2D image is displayed.
                                                                   contrast agent’s label.                       depending on the acoustic window of the patient or
2. Press PW or CW again to re-display original                                                                   the specific contrast agent being used, there may
   Doppler layout.                                                                                               be need for adjustment.
Steering on Vascular Applications                      Misdiagnosis Based on Image Artifacts                               C CAUTION: Too high a power setting will
                                                                                                                             destroy the contrast agent in the LV
To steer the Doppler interrogation beam:               Misdiagnosis in ultrasound contrast images might                      cavity. Destruction of the contrast
                                                       be caused by several artifacts. These include:                        agent is shown as a swirling pattern.
1. Press PW.
2. Use the Steer rocker key to steer the Doppler       •   Motion artifacts                                      If, before injection of the contrast agent, image
   interrogation beam to the right, left or                Motion artifacts give rise to contrast-like signals   quality is not acceptable, increase the power level.
   perpendicular to the probe surface.                     independent of the presence of contrast.              Then, when the contrast agent is observed in the
3. Use the Doppler Angle soft key rotary to align          Motion artifacts may be caused by patient             left ventricle, reduce the power to a level giving
   the Doppler angle cursor with the measured              movement, including respiration, or by probe          homogenous opacification.
   vessel.                                                 movement influenced by the operator.
                                                       •   Regional drop-outs                                    If a swirling pattern—that is, destruction of the
To exit PW Doppler Mode or CW Doppler Mode:                                                                      contrast agent—is observed and persists after the
                                                           Regional drop-outs can be caused by the
1. If the system is in PW-Mode, press PW.                  unintentional destruction of the contrast agent,      LV cavity has been filled with the contrast agent,
   OR                                                      too low a concentration of the contrast agent,        continue to reduce the power until homogenous
   If the system is in CW-Mode, press CW. Full             poor acoustic penetration due to rib and/or lung      opacification is obtained.
   2D-Mode or CFM-Mode will be displayed.                  shadows, or by the system failing to detect the
                                                                                                                             WARNING: Read and carefully follow
2. Press Cursor on the control panel to hide the           contrast agent. The last cause might be the
                                                                                                                             the manufacturer’s instructions on the
   Doppler cursor.                                         result of incorrect settings introduced by the
                                                                                                                             contrast agent’s label.
                                                           operator.
                                                                        Getting Started
Vivid 3 User Guide                                          Direction 2305203-100 Rev. 004                                                                    21
Angio Mode
         TM                                          Soft Key and Soft Menu Functions                        pressed and/or turned. The lit (yellow) label
                                                                                                             indicates the currently active function.
AngioTM Mode (Power Doppler) is application          The soft keys are unlabeled grey buttons and
dependent and is only used in vascular and shared    rotaries located at the top of the control panel.       For more detailed information, refer to Chapter 3,
services applications.                                                                                       Modes in the Vivid 3 User’s Manual.
                                                     A set of corresponding soft key icons is displayed at
NOTE: AngioTM Mode is only available on selected     the bottom of the screen.
probes and applications.
                                                     The functions of the soft keys vary according to the
                                                     mode in which the user is working.
Accessing AngioTM Mode
                                                     In each mode, a label above and/or below each soft
To activate AngioTM Mode (Power Doppler):            key icon on the screen indicates the function of the
                                                     corresponding soft key on the control panel, when
•   Press Mode 2.
    As shown in the example below, PDI (Power
    Doppler Image) parameters are displayed on
    the right of the screen, indicating that Power
    Doppler mode is currently activated:
                                                      Soft keys on the screen
                                                                          Getting Started
Vivid 3 User Guide                                                         Direction 2305203-100 Rev. 004                                                                                                 22
The Soft Key functions are as follows:                              •   Compression M – increases the compression level and                 •   Focus – changes the location of the focal zone marker by
                                                                        reduces the M-Mode image contrast when rotated                          displaying a triangular focus marker that indicates the depth
•   %DR – measures two diameters of one blood vessel                    clockwise, to soften the image. When rotated counter-                   of the optimized focal point. In Color Mode, the focus
    (minimum and maximum).                                              clockwise, increases the M image contrast.                              location tracks the ROI position.
•   %AR – measures two areas of one blood vessel (minimum           •   Contour M – controls image processing related to the                •   Foci Number - enables the user to choose the required
    and maximum).                                                       extent of edge enhancement applied to an image.                         number of focal zone markers. The number of markers
•   2D Maps – displays a 2D Map menu. The menu enables an           •   Curve – activates Curved Anatomical M-Mode, enabling                    available for selection will depend upon the scanning depth
    option from a list of non-linear grey-curves or different 2D-       the user to generate a synthetic M-Mode display from a                  of the scan currently in progress.
    colorized curves to be selected.                                    curved line drawn on a 2D image.                                        NOTE:Changing the number of focal zone markers affects the
•   2D Caliper – measures distance on the 2D image.                 •   D Caliper – measures maximum velocity, time, and slope                  Frame Rate.
•   2D Area – measures area on the 2D image.                            on the spectral Doppler.                                            •   Foci Position - sets the position of the focal zone markers
•   2D Volume - measures volume of 2D area.                         •   Directional - shows direction of the Power Doppler (only in             (relative to each other).
•   Angle – controls the size or angular width of the 2D image          Angio [Power Doppler] Mode).                                            NOTE:Toggles between Foci Number/Foci Position.
    sector (in degrees). A smaller angle generally produces a       •   DDP Persistence - enables adjustment of 2D image in
                                                                                                                                            •   Foci Spread - sets the spread of the focal zone markers.
    scan with a higher frame rate.                                      time. The current frame retains some 2D information from                Turning the rotary to the left sets markers closer together;
•   ECG Position – the user can position the ECG according              previous frames, to prevent noise.                                      turning it to the right, set them further apart.
    to preference.                                                  •   Doppler Angle – enables the user to correct the Doppler             •   Frame Rate – adjusts the frame rate (in frames/sec).
•   Audio Volume – enables the adjustment of the volume                 velocity scale by defining the angle between the Doppler
                                                                                                                                            •   Frequency 2D – enables adjustment of the transmit and
    level in Doppler Mode.                                              beam and the interrogated blood vessel (in degrees).
                                                                                                                                                receive operating frequency (in MHz).
•   Auto Update – activates the 2D image for a short time           •   Doppler Maps – displays a drop-down menu of different
                                                                                                                                            •   Horizontal Sweep – enables modification of the horizontal
    whenever the gate is moved, allowing it to refresh for a            Doppler colorization/grey maps when rotated. Select the
                                                                        required non-linear grey curves or "colorized Doppler"                  sweep speed (in m/sec).
    short distance before the Doppler process is continued.                                                                                     In PW-Mode: focus tracks the Doppler gate position.
                                                                        curves.
•   Baseline (color) – modifies the color distribution map and                                                                                  In CW-Mode: focus tracks the small horizontal marker
    changes the color contents of the image accordingly.            •   Dual Focus - provides another focal point in near field (only
                                                                                                                                                located on the Doppler cursor.
•   Baseline (Doppler) – enables the Doppler spectrum to be             in Cardiac Mode).                                                       In M Mode: focus tracks time (in msecs)
    shifted up and down.                                            •   ECG - toggles between ON/OFF to show/hide ECG trace                     In ECG Mode:focus changes the trace speed.
•   B/C - controls the priority of black & white image (B)          •   ECG Internal/External - enables the user to select the              •   HR – measures heart rate.
    vs. color (C).                                                      required signal source (either from the ECG cables
                                                                                                                                            •   Invert
                                                                        connected to the system, or from an external ECG source.
•   Cine Scroll – enables manual scrolling of the cineloop, in                                                                                  •    Left/Right Invert: Enables a mirror image of the 2D
    any of the modes, as an alternative to scrolling with the       •   ECG/Phono Gain - controls the gain of the ECG/Phono
                                                                                                                                                     image to be created.
    TRACKBALL.                                                          signal.
                                                                                                                                                •    Up/Down Invert: Enables the 2D image to be rotated
•   Cine Speed – controls the cineloop playback speed. When         •   End Marker – moves the red vertical line that is displayed                   180 degrees.
    Nominal is displayed on the rotary on the screen, the loop is       over the ECG trace to indicate the end of a cineloop. Move
                                                                                                                                                •    Invert Color: Enables the color scheme assigned to
    running at its original speed.                                      left or right to define the section of the cineloop that is to be
                                                                                                                                                     positive and negative velocities to be inverted.
•   CO – acquires two measurements; one on the Doppler                  displayed.
                                                                                                                                                •    Invert (Doppler): Enables the spectrum to be flipped
    trace and one on the 2D valve diameter.                         •   Envelope Adjustment – corrects and adjusts the envelope
                                                                        for the Doppler spectrum image auto trace, assuring better                   180 degrees, so that negative velocities are displayed
•   Color Maps – displays a menu of color map options. Rotate                                                                                        above the baseline and positive velocities below the
                                                                        accuracy.
    the rotary to select a color map.                                                                                                                baseline. If the Doppler Mode is combined with Color
                                                                    •   First Marker – moves the green vertical line that is
•   Colorize – enables the user to colorize black and white                                                                                          Mode, the color map will also be inverted.
    images. With the use of the 2D Maps soft key, the user              displayed over the ECG trace to indicate the beginning of a
                                                                        cineloop. Move left or right to define the section of the           •   Layout – enables toggling between side-by-side and top-
    selects the color map required.                                                                                                             bottom layouts when viewing two modes in parallel.
                                                                        cineloop that is to be displayed.
•   Compression 2D – increases compression when rotated                                                                                     •   Line Move – enables the M Cursor displayed over 2D
    clockwise, resulting in reduced contrast of the 2D image.                                                                                   images in Anatomical M-Mode to be moved from left to right
•   Cycles for Store - enables the user to select the number of                                                                                 across the scanning screen.
    cycles to be stored in one cine.
                                                                                         Getting Started
Vivid 3 User Guide                                                        Direction 2305203-100 Rev. 004                                                                                             23
•   Line Rotate – enables the M Cursor displayed over 2D            •   Sample Volume – enables the TRACKBALL to be assigned                3D Option:
    images in Anatomical M-mode to be rotated 360 degrees.              to control the size of the sample volume gate (in mm), in
                                                                                                                                        •   Layout - 3D - enables the user to select the required 3D
•   Low PRF – enables toggling between High and Low Pulse               PW-Mode.
                                                                                                                                            View Layout. Options available for selection:
    Repetition Frequency (PRF) functions, only in PW-Mode.          •   Scroll Results – enables scrolling within the list of
                                                                                                                                            •    3D + Cube
•   Low Vel. Reject (Color) – enables the extent of low                 measurements taken and displayed in the measurement
                                                                        result list window.                                                 •    Render
    velocity removal to be adjusted (in cm/s).
                                                                    •   Sensitivity - active in CW, enables the sensitivity of the          •    Main Plane
•   Low Vel. Reject (Doppler) – enables the low velocity
    portions of the spectrum to be filtered (in cm/s), since the        Doppler signal to be increased/decreased.                           •    Landscape
    Doppler spectrum and audio can contain strong,                  •   Simultaneous – synchronizes all the cineloops to run            •   Volumes - enables the user to select the required 3D
    undesirable wall-motion signals.                                    simultaneously on the screen when working in multiple               Viewing option. Choose from the following:
•   LV – measures left ventricle parameters (IVS, LVID, LVPW).          view. When turned off, only the image in the active segment         •    Gray
                                                                        of the screen will run. When running cineloops in split             •    Color
•   Manual Trace/Auto Trace – enables Doppler signal
    envelope tracing.                                                   screen mode, enables cineloops to be run in only one
                                                                                                                                            •    Gray + Color
                                                                        window or in all windows.
•   M Caliper – measures distance, time, and slope on the M-                                                                                NOTE:Volumes selection only available in Color/Angio Mode.
                                                                    •   Stop Cine – stops the continuous playback of a cineloop
    Mode display.
                                                                        when pressed. Active when cineloop is running.                  •   Auto Movie - enables the user to record the rendering of a
•   M Height – measures distance on the M-Mode display.                                                                                     3D image.
                                                                    •   Tilt – enables the sector of the 2D image to be tilted to the
•   M Maps - displays a drop-down menu of different M-Mode                                                                              •   3D Reset - returns the display to the initial default 3D Cube.
                                                                        left or right. By default, the axis of symmetry of a 2D image
    grey or colorized maps.
                                                                        is vertical.                                                    •   Scan Distance - in post-processing, enables the user to
•   Previous/Next – enables the user to manually move to the                                                                                manipulate the incremental distance between each
                                                                    •   Triplex Live - simultaneously presents 2D, Color, & PW - all
    next or previous heart cycle. Active when cineloop is               live.                                                               neighboring scan plan.
    running. (Toggles between Previous/Next cycles.)
                                                                    •   Variance – enables the velocity variance calculation and        •   Save As - when enabled, used to save a 3D image to a
•   PS + ED – measures two velocity points on the spectral                                                                                  hard drive, removable drive, floppy disk, etc.
                                                                        display velocity variable as an indication of the amount of
    Doppler window.
                                                                        turbulent flow.
•   Reject 2D – enables 2D (grey) reject in CFM-Mode. Setting                                                                               Hip Joint Measurement:
                                                                    •   Velocity – measures one velocity point on the Doppler
    is differential to the value set in 2D only. Returning to 2D-       window.                                                         •   Hip Joint - enables the user to superimpose lines on an
    Mode disables Reject 2D.                                                                                                                image and align them with anatomical features, for the
                                                                    •   Velocity Scale – controls the PRF (Pulse Repetition
•   Reject M – rejects the low-level echoes in the M-Mode                                                                                   calculation of Hip Joint Angles measurement.
                                                                        Frequency) values. It enables maximal detected velocity to
    image when rotated clockwise, producing a clearer image.                                                                            •   Rotate Line - used to rotate any of the three lines
                                                                        be modified. In CFM, it also controls the distribution of
    Turn the rotary counterclockwise to decrease the low-level          shades in the color map over the different velocity values in       (superimposed on the image) to the required angle.
    echo rejection.                                                                                                                     •   2D Caliper - used to perform calculation of the angle
                                                                        the flow.
•   Resize Results – enables shortening and/or enlarging of                                                                                 between two lines.
                                                                    •   VF – measures velocity time integral (VTI) and heart rate
    the measurement result list shown in the window on the                                                                              •   Ellipse - provides an alternative method of measuring two
                                                                        (HR) on the spectral Doppler window, and blood vessel
    display screen.                                                     diameter on the 2D window.                                          diameters (within an elliptical shape).
•   ROI Size – enables the TRACKBALL to control the                 •   Volumes -                                                       •   2 Diameters - assigns a ratio to a hip joint measurement.
    dimensions of the color ROI (Region of Interest, when
    pressed. Use the TRACKBALL right, left, up, or down to
                                                                    •   Width – controls the sector width in the case of a linear           Biopsy Guide Option:
                                                                        probe, and angular width for all other probes.
    control ROI height/width.                                                                                                           •   Biopsy Marker - used to adjust and rotate the biopsy
•   RUN Cine – activates an endless playback of a cineloop                                                                                  “depth marker” to the required position (estimates the
    when pressed. The time frame of the played back cineloop                                                                                penetration depth of the biopsy needle).
    is defined by the colored markers displayed on the ECG                                                                              •   Biopsy Setup - used to select one of three angles when
    trace. In 2D-Mode, displays a scrolling image. In Doppler                                                                               changing the angle of the biopsy bracket attachment.
    Mode, displays a scrolling ECG pointer. Active when                                                                                     NOTE:Only applicable to variable-angle adjustable bracket
    cineloop is stopped.                                                                                                                    attachments.
                                                                                       Getting Started
Vivid 3 User Guide                                            Direction 2305203-100 Rev. 004                                                                                              24
Soft Menu Rocker Functions                              The Soft Menu Rocker functions are as follows:                        •   Frequency – enables the transmission frequency in
                                                                                                                                  Doppler and Color Modes (in MHz) to be modified
The Soft Menu rocker enables access to                  •   Contour 2D – enables the sharpness of images to be                    to control sensitivity or level of penetration.
                                                            increased in the radial direction in 2D or M-Mode.
mode-specific pop-up menus that provide display                                                                               •   Persistence – enables adjustment of color images, so
                                                        •   Diff (Difficult) – reduces these reverberation artifacts,             the current frame retains some color information from
adjustment options. For additional information, refer       however frame rate will be reduced as a result. When set to           previous frames, to avoid noise.
to Chapter 3, Modes in the Vivid 3 User’s Manual.           0 (zero), this function is active. The factory default for this   •   Power – controls the amount of acoustic power applied in
                                                            function is 1.
                                                                                                                                  all modes (in dB).
                                                        •   Doppler DR – adjusts the range of color scale used to
                                                                                                                              •   Reject 2D – enables 2D (grey) reject in CFM-Mode. Setting
                                                            represent the span of received signal strength from the
                                                                                                                                  is differential to the value set in 2D only. Returning to
                                                            blood. It makes the spectrum softer or harder.                        2D-Mode disables Reject 2D.
                                                        •   Doppler Post Processing – controls the amount of                  •   Reject (Doppler) – enables undesirable background noise
                                                            contrast of the Doppler display.
                                                                                                                                  to be removed from the Doppler spectrum.
                                                        •   2D Dynamic Range – enables control of the dynamic range
                                                                                                                              •   Sample Volume – controls the size of the CFM sample
                                                            or contrast of an image in 2D and Doppler Modes (in dB).
                                                                                                                                  volume.
                                                        •   Flash Removal – enables an algorithm that removes
                                                                                                                              •   Steer Angle – controls the angular direction of the
                                                            undesired color flashes due to probe motion or strong
                                                                                                                                  ultrasound beam.
                                                            tissue motion. (Note: High values can cause a reduction of
                                                            color sensitivity.)                                               •   Lateral and Axial Smooth – enables radial and lateral
                                                                                                                                  smoothing of the color pixels.
                                                        •   Frame Rate – enables the frame rate (in Hz) to be
                                                            increased when examining fast moving tissue. When
                                                            examining non-moving tissue, the frame rate can be
                                                            decreased to increase resolution.
                                                                            Getting Started
Vivid 3 User Guide                                          Direction 2305203-100 Rev. 004                                                                   25
                                                                      Getting Started
Vivid 3 User Guide                                          Direction 2305203-100 Rev. 004                                                                         26
To rotate the 3D VOI:                                                                                       NOTE: The yellow hand appears only when the
                                                                                                            pointer is on an edge of the 3D VOI. If the hand is
1. Position the white trackball pointer (in the                                                             not visible, trackball closer to the edge. Only the
   shape of a pointing finger) outside of the                                                               edge of the plane closest to the hand is highlighted
   rectangle and press Select.                                                                              in yellow.
   The white hand will close, as shown in Figure 7.
                                                                                                            To manipulate the corner of the 3D VOI:
                                                                                                            1. Position the green trackball pointer (in the
                                                         Figure 8: Moving through the 3D VOI (Red Hand)
                                                                                                               shape of a pointing finger) on one of the
                                                      NOTE: Any plane in the volume can be made                corners of the VOI and press Select.
                                                      active (highlighted by a red rectangle) by placing       The trackball cursor shape changes to a green
                                                      the red hand over it and clicking on it.                 closed hand and a small green dot appears on
                                                      To manipulate the edge of the 3D VOI:                    the selected corner.
                                                                                                            2. Use the Trackball to move the green hand and
                                                      1. Position the trackball pointer over any visible       pull back the corner of the plane accordingly, as
                                                         edge. The hand pointer turns yellow.                  shown in Figure 8.
      Figure 7: Rotating the 3D VOI (White Hand)      2. Position the yellow pointer (in the shape of a
                                                         pointing finger) on the required edge and press
2. Trackball to rotate the 3D VOI in any direction,
                                                         Select.
   as required. You can also tilt it backwards or
                                                         The trackball cursor shape changes to a yellow
   forwards, if necessary.
                                                         closed hand and the edge of the selected plane
3. Press Select again to release the trackball from      (closest to the hand) becomes highlighted in
   rotating the VOI. The white hand will open in         yellow.
   the shape of a pointing finger.
                                                      3. Use the Trackball to move the yellow hand (and
                                                         the edge of the plane) accordingly - to view
                                                         specific anatomical portions, as shown in
                                                         Figure 9.                                          Figure 10: Manipulating the Corner of a 3D VOI (Green Hand)
                                                                      Getting Started
Vivid 3 User Guide                                          Direction 2305203-100 Rev. 004                                                                          27
NOTE: A bracket attachment is supplied in a                          • Follow the probe cleaning and disinfection procedures and precautions to properly
"Starter Kit" together with some accessories in a                      prepare the probe.
separate Biopsy Kit. For details and a list of                       • Follow the manufacturer's instructions for the cleaning of biopsy devices and accessories.
contents of the kit, refer to theVivid 3 Release                     • Use protective barriers such as gloves and probe sheaths.
Notes SW 3.1.
                                                                     • After use, follow proper procedures for decontamination, cleaning, and waste disposal.
                                                                 C CAUTION: Improper cleaning methods and the use of certain cleaning and
                                                                   disinfecting agents can cause damage to the plastic components that will degrade
                                                                   imaging performance, or increase the risk of electric shock.
                                                                     For more information, refer to the safety instructions supplied with each probe.
                                                                     Getting Started
Vivid 3 User Guide                                            Direction 2305203-100 Rev. 004                                                                         28
Displaying the Biopsy Guide Zone Screen                     The following message is displayed: "Be sure to
                                                            select the proper biopsy kit and set up the
The Biopsy Guide Zone is an interactive display             proper bracket angle".
drawn on the screen, and associated with the                For details on selecting the proper Biopsy Kit,
currently-selected bracket angle of the probe               refer to the information provided in the Vivid 3
attachment.                                                 Release Notes SW 3.1.
NOTE: For a full description and detailed                   For “Bracket Attachment Angle Set-up” see the
explanation of this feature, refer to the Vivid 3           instructions, below.
Release Notes SW 3.1.                                       NOTE: If the current probe does not support
NOTE: The user is able to change the way in which           use of a biopsy bracket, the message "Biopsy       Figure 12: Bracket Attachment Angle Selection Set-up
the Biopsy Guide Zone is displayed, in accordance           Guide is not supported" is displayed.
                                                                                                               2. Select the required bracket angle (positions 1,
with individual preferences. For example, change        2. Adjust the Biopsy Marker rotary and rotate it to       2, or 3) by clicking on one of the following:
the distance between dots; hide/display the center         move the "depth marker" to the required                MBX-1, MBX-2, or MBX-3.
line, outer lines etc. For instructions, refer to the      position. Use this tool to estimate the needle’s        Once the required bracket angle is selected,
Vivid 3 Release Notes SW 3.1.                              penetration depth.                                      the pop-up menu closes and the Guide Zone
To activate display of the Biopsy Guide Zone:                                                                      display accordingly reflects the new selection.
1. While in 2D or Color Mode, press Biopsy on                                                                  NOTE: If you wish to exit the pop-up menu without
   the alphanumeric keyboard.                                                                                  making changes to the bracket angle, simply click
                                                                                                               on Exit.
    If the current probe supports use of a biopsy           The changed position of the depth marker is
    bracket, the Biopsy Guide Zone is displayed on          reflected on the screen.                           3. Check the display at the
    the screen, as shown in Figure 11 below.                                                                      top of the screen to make
                                                        Bracket Attachment Angle Set-up                           sure that the selected
                                                                                                                  angle is appropriate. If not, repeat Step 2 to
                                                                                                                  change the selection.
                                                        NOTE: This procedure is intended for variable-angle
                                                        adjustable bracket attachment only.                                 DANGER: Failure to match the
                                                                                                                            screen-displayed Guide Zone to the
                                                        To change the angle of the bracket attachment:
                                                                                                                            physical angle of the bracket’s needle
                                                        1. Press the Biopsy Setup soft rotary key.                          guide may cause the needle to track a
                                                                                                                            path outside the Guide Zone.
                                                            A pop-up menu is displayed, showing pictures
                                                                                                                            It is extremely important that when using the
                                                            of the bracket attachment at three different
                                                                                                                            adjustable angle biopsy guides, the angle
                                                            angles, as shown in Figure 12.
                                                                                                                            displayed on the screen matches the angle
                                                                                                                            set on the guide, otherwise the needle will
                                                                                                                            not follow the displayed guide zone. This
                                                                                                                            could result in repeated biopsies, or even
                                                                                                                            patient injury. For more details on Biopsy
   Figure 11: Biopsy Guide Zone displayed on screen                                                                         Needle Deviation, see the Vivid 3 Release
                                                                                                                            Notes SW 3.1
                                                                        Getting Started
Vivid 3 User Guide                                              Direction 2305203-100 Rev. 004                                                                                  29
NOTE: Whenever the system is turned OFF and ON
again, the bracket-hole number - as well as the                Needle clip attachment                                               C CAUTION: When pushing the pin to
location of the depth marker - is maintained as before.                          Orientation mark                                     secure the angle position of the
                                                                                                                                      needle guide attachment, hold the
Preparing the Multi-Angle Guide                                                                                                       bracket firmly in place on the probe.
                                                                                                                                      Excessive force may cause the
Attachment                                                                                                                            bracket to release from the probe.
             WARNING: DO NOT attempt to use
                                                                                                                         9. Apply an adequate amount of coupling gel to
             the multi-angle biopsy bracket and
                                                                        Bracket            Probe                            the face of the probe.
             needle guide until the
             manufacturer's instructions                                                                                 10. Place the proper sanitary sheath tightly over the
                                                           Figure 14: Aligning the Orientation Mark
             (provided with the biopsy bracket                                                                               probe and biopsy bracket. To hold the sheath
                                                          6. Attach the biopsy bracket to the probe by                       firmly in place, use the rubber bands supplied,
             and needle guide in the Biopsy Kit),            sliding the bracket over the end of the probe
             have been read and thoroughly                                                                                   as shown in Figure 16.
                                                             until it clicks or locks in place.
             understood.
                                                          7. Select the required needle guide attachment
To assemble the multi-angle biopsy guide:                    angle by pulling the metal locking pin upwards
                                                             to freely move the needle guide attachment and
1. Scan the patient and identify the target for the          slide it out of the slotted opening on the plastic
   biopsy.                                                   bracket.
2. Move the probe to locate the target to the
   center of the image.                                    Metal
                                                           locking                                                       Figure 16: Applying the Sanitary Sheath to the Probe
3. Enable the system Biopsy Guide Zone and try
   the guide zone angles MBX1, 2 or 3 to decide            pin                                                           11. Place the needle clip attachment onto the
   the best angle setting for the needle path. See                                                                           biopsy guide bracket and snap into position, as
   “Displaying the Biopsy Guide Zone Screen” on                                                                              shown in Figure 17.
   page 28.
4. Identify the appropriate Biopsy Guide bracket              Locking pin in Position 1 Locking pin in new position              Needle clip attachment
   by matching the label on the bracket with that
                                                          Figure 15: Changing Attachment Angle & Locking into Position
   on the probe to be used.
                                                          8. Align the pin with the currently-selected angle
                                                             of the needle guide attachment in order to
                                                             match that of the Guide Zone displayed on the
                                                             Vivid 3 screen: MBX1, MBX2, or MBX3. For
                                                             example, if MBX2 is selected, align the pin with
   Figure 13: Multi-Angle Biopsy Guide Bracket               position 2 (as shown above) and push it into
5. Orient the bracket so that the needle clip                position in the slot. Press the pin downwards to
                                                             secure it firmly in place (see picture on right, in          Figure 17: Fitting the Needle Clip Attachment onto the
   attachment will be on the same side as the                                                                                              Biopsy Guide Bracket
   orientation mark (ridge) on the side of the               Figure 15 above).
   probe, as shown in Figure 14.
                                                                            Getting Started
Vivid 3 User Guide                                                    Direction 2305203-100 Rev. 004                                                                       30
12. Push the locking mechanism towards the                      NOTE: When inserting the needle barrel into the         7. Push the locking mechanism towards the
    bracket to secure the lock. Make sure the                   needle clip, make sure the two embossed arrow              bracket to secure the lock. Make sure the
    needle guide is firmly attached to the bracket -            points are facing each other; in this way correct          needle guide is firmly attached to the bracket -
    see Figure 18.                                              positioning is assured. The gauge marked on one            see Figure 18 on page 30.
                                                                side of the needle barrel will be visible, since this   8. Choose a needle barrel of the required gauge
                   Lock                                         will be facing towards you (as shown in Figure 20).        (size) and twist it back and forth to remove it
                                                                                                                           from the plastic holder (tree) - refer to Figure 19
                                                                Preparing the Fixed-Angle Guide                            on page 30.
                                                                Attachment                                              9. Place the selected needle barrel into the needle
                                                                                                                           clip (see note below) and push to snap into
                                                                            WARNING: DO NOT attempt to use                 place - see Figure 20 on page 30.
                                                                            the fixed-angle biopsy bracket and
 Figure 18: Locking the Needle Clip Attachment into Position                                                            NOTE: When inserting the needle barrel into the
                                                                            needle guide until the
                                                                                                                        needle clip, make sure the two embossed arrow
13. Choose a needle barrel of the required gauge                            manufacturer's instructions
                                                                                                                        points are facing each other; in this way correct
    (size) and twist it back and forth to remove it                         (provided with the biopsy bracket
                                                                                                                        positioning is assured. The gauge marked on one
    from the plastic holder (tree) - see below.                             and needle guide in the Biopsy Kit),
                                                                                                                        side of the needle barrel will be visible, since this
                                                                            have been read and thoroughly
                                    Plastic holder                                                                      will be facing towards you (as shown in Figure 20).
                                                                            understood.
                                                                                Getting Started
Vivid 3 User Guide                                          Direction 2305203-100 Rev. 004                 31
Performing the Biopsy Procedure                        To complete the Biopsy procedure:
                                                       1. When the biopsy is complete, remove the
            WARNING: Biopsy procedures must               needle barrel, needle clip and probe sheath.
            only be performed on live images.          2. Properly dispose of these items in accordance
                                                          with current facility guidelines.
                                                       NOTE: The biopsy bracket can be sterilized in a
                                                       recommended disinfecting agent (refer to the
          C CAUTION: Before performing a               manufacturer’s instructions - see also the Biopsy
            biopsy, ensure that all Biopsy Guide       Bracket Cleaning and Sterilization instructions
            parts are seated properly.                 provided in the Vivid 3 Release Notes SW 3.1),
            Refer to the following instructions, as    after which it may be re-used, as required.
            appropriate: “Preparing the Multi-Angle
            Guide Attachment” on page 29;
            “Preparing the Fixed-Angle Guide
            Attachment” on page 30.
To perform the Biopsy procedure:
1. Place coupling gel on the scanning surface of
   the probe/sheath/biopsy guide assembly.
2. Activate the Biopsy Guide Zone on the Vivid 3
   system by pressing Biopsy on the
   alphanumeric keyboard. (Refer to “Displaying
   the Biopsy Guide Zone Screen” on page 28).
    When using multi-angle guides, ensure that the
    proper Guide Zone angle is displayed.
    (Refer to “Bracket Attachment Angle Set-up” on
    page 28).
3. Scan to locate the target. Center the target in
   the electronic Guide Zone path.
4. Carefully place the needle in the guide between
   the needle barrel and needle clip. Direct it into
   the area of interest for specimen retrieval.
                                                                      Getting Started
Vivid 3 User Guide                                       Direction 2305203-100 Rev. 004                                                                   32
Physiological Traces Changing the ECG Trace Amplitude Hiding the ECG Signal Trace
For additional information, see Chapter 4,        To change the ECG trace amplitude:                      To hide the ECG signal trace:
Physiological Traces in the User’s Manual.        1. Press Physio.                                        1. Press Physio .
                                                  2. Turn Gain ECG rotary.                                2. Press the ECG soft key to hide the ECG signal
Connecting the Electrodes                         To change the ECG trace position:                          trace. The on-screen LED will dim. Press the
                                                                                                             ECG soft key again to display the ECG signal
                                                  1. Press Physio.                                           trace.
                                                  2. Turn ECG position rotary.
                                                                                                          Using the Heart Sound Microphone
                                                  Selecting an External ECG Input                         Device
 1                                      3
 2                                      4         To select external ECG signal input:                    To use the heart sound microphone device:
                                                  1. Press Physio.                                        1. Connect a microphone cable in the Rear
                                                  2. Press the ECG Internal/External soft key so             Connector Panel.
                                                     that External is lit on screen. To return to the     2. Press Physio.
                                                     internal ECG unit, press the ECG Internal/           3. Press the Phono soft key to display the Phono
                                                     External soft key again so that Internal is lit on      trace.
1. Footswitch (black)
                                                     screen.
2. Phono (blue)
                                                  NOTE: ECG Internal/External can also be set by          Changing the Phono Trace Position
3. Internal ECG (yellow)                          accessing the System Configuration and changing
4. ECG (green)                                    to the required settings under the ECG/VCR tab.         To change the Phono trace position:
                                                  NOTE: If any ECG trace problems occur, use the          1. Press Physio.
                                                  Reset ECG soft key.
Adjusting ECG & Phono Display                                                                             2. Press the ECG position / Phono Position soft
                                                                                                             key rotary to activate the Phono position
Changing the Trace Sweep Speed in                                                                            function.
2D-Mode                                                                                                   3. Turn the Phono position soft key rotary.
For additional information, see Chapter 5, Stress      Selecting an Image                                    To delete an acquired image:
Echo in the User’s Manual.                                                                                   1. TRACKBALL to the projection/level image cell
                                                       To begin acquisition:                                    and press Select.
Accessing Stress Echo Mode                                                                                   2. Press Menu. The Delete option is displayed.
                                                       1. Toggle Freeze to off.
                                                                                                                TRACKBALL to the on-screen Delete button and
To access Stress Echo Mode and select a stress         2. Perform a scan in 2D-Mode.
                                                                                                                press Select. A pop-up dialog box, requesting
protocol:                                              3. Press Freeze to stop and replay the last              confirmation of the deletion is displayed.
1. Press Protocol. A list of available stress             complete heart-loop.
                                                                                                             3. TRACKBALL to Yes and press Select.
   protocol templates is displayed on the              4. After selecting the preferred loop, press Store.
                                                                                                             NOTE: To delete a level, the selected level must
   clipboard.                                          5. Repeat previous steps until all required           not contain any images.
NOTE: Stress protocol templates can be edited or          projections have been made and completed.
created, as described in Chapter 5,Stress Echo in                                                            To replace an acquired image:
                                                       NOTE: You may also press Store twice (to accept
the User’s Manual.                                     loops) without performing Steps 1 -3 above.           1. TRACKBALL to the projection/level image cell
2. Use the Select Template soft key rotary to                                                                   that is to be replaced in the clipboard’s matrix
                                                       To select a loop on a Quad View display:
   highlight the required template.                                                                             and press Select.
                                                       1. Follow steps 1 – 3 of the previous procedure to    2. Press Store.
3. After selecting the required protocol, press           acquire an image.
   Freeze to exit Freeze Mode and begin                                                                      3. TRACKBALL to Yes and press Select.
                                                       2. Press the Quad View soft key so the relevant
   acquisition.The name of the selected template
                                                          on-screen icon is selected.
   is displayed at the top of the clipboard.
                                                       3. TRACKBALL to the required quad window and
                                                          press Select.
                                                       4. Manually select the required image using the
                                                          soft keys and soft key rotaries.
                                                       5. Press Store.
                                                       6. Repeat the previous steps until all required
                                                          projections have been acquired and completed.
                                         Stress Echo
Vivid 3 User Guide                                             Direction 2305203-100 Rev. 004                                                                        34
                                           Stress Echo
Vivid 3 User Guide                                            Direction 2305203-100 Rev. 004                                                                    35
Selecting and Storing Heart Cycles                       Displaying, Hiding, or Resetting the                  Assigning New Labels to Levels or
                                                         Capture Buffer                                        Projections
After the buffer is filled with continuously captured
cycles, the user can begin to review the contents of     To display, hide, or reset the capture buffer:        To assign new labels to projections:
the buffer and store images to the scoring matrix.
                                                         1. TRACKBALL to the template name and press           1. TRACKBALL to the projection label that is to be
The procedure options available for reviewing the           Select.                                               changed and press Menu. A pop-up menu is
contents of the buffer are described below.                                                                       displayed with a list of available labels.
                                                         2. TRACKBALL to the required option and press
To review and select the buffer bar’s captures in           Select.                                            2. TRACKBALL to the required projection label and
Full View:                                                                                                        press Select.
                                                         NOTE: For details on viewing scoring and
•   Use the Previous/Next soft key rotary to scroll      measurements during the Stress exam, refer to         NOTE: Once a level has an image stored, the label
    through the buffer. A blue line indicates the        “Reporting” on page 52.                               of that level can no longer be changed.
    location of the heart cycle in the buffer bar, and
    the heart cycle is displayed on the screen,          Editing and/or Creating Stress                        Defining New Groups
                                                         Templates                                             To assign a name to a group of image cells:
To enter, review, and select the captures in the
                                                                                                               1. TRACKBALL to the group name that is to be
buffer bar in quad view:                                 Selecting a Base Template                                changed in the group list and press Select.
1. Press the Quad View soft key. The first four          To select a base template:                            NOTE: If cells are assigned while the group name
   captured heart cycles are displayed.                                                                        is highlighted, the cells associated with the group
                                                         1. Press Protocol.
NOTE: If the Quad View soft key is not available,                                                              will be identified by yellow frames.
press Active Mode. Scroll using Active Mode              2. Use the Select Template soft key rotary to
                                                            highlight the one to use as the basis of the new   NOTE: If the vertical cursor is not displayed in the
through the different soft key menu options, while in
                                                            stress template.                                   desired location, move the cursor with the TRACKBALL
Freeze Mode, until the Quad View soft key is
                                                                                                               and press Select again to reassign cursor location.
displayed.
                                                         Adding/Deleting Levels                                2. Use the alphanumeric keyboard to enter a new
2. Use the Previous/Next soft key rotary to scroll
                                                         To add levels:                                           name.
   through the buffer to display the four previous
   or next heart cycles. A blue line indicates the                                                             To define cell options:
                                                         •   Rotate the Set Levels or (Set Peak Level) soft
   location of the first window’s heart cycle in the         key rotary clockwise until the required number    Cell options, such as whether or not the T2 timer is
   buffer bar.                                               of levels is reached.                             displayed, or when scoring is entered, can be
To store heart cycles selected from the buffer           To delete levels:                                     configured for each cell.
bar:
                                                         •   Rotate the Set Levels (or Set Peak Level) soft    1. TRACKBALL to the required image cell and press
1. Press Store.                                              key rotary counter-clockwise to remove the           Select.
2. Follow the procedures for analysis, described in          unrequired level/s                                2. Press Menu.
   Chapter 5, Stress Echo of the User’s Manual.          NOTE: Be sure not to remove the last and soliday         A pop-up menu with cell editing options is
NOTE: Press Exit to return to continuous capture         Peak level. Once no Peak level remains in the            displayed (e.g., Delete, Enter Quad on Freeze,
acquire mode.                                            template, the user will not be able to add a Peak        Unfreeze after Store, Timer 2 Reset or Off,
                                                         level to the template.                                   Enter Scoring).
                                           Stress Echo
Vivid 3 User Guide                                      Direction 2305203-100 Rev. 004   36
Deleting Groups
To delete a group:
1. TRACKBALL to the name of the group that is to
   be deleted along the right side of the screen
   and press Menu.
2. TRACKBALL to Empty and press Select. The
   group is removed.
                                          Stress Echo
Vivid 3 User Guide                                         Direction 2305203-100 Rev. 004                                                                    37
                                                                  Cardiac Measurement
                                                                  and Analysis
Vivid 3 User Guide                                        Direction 2305203-100 Rev. 004                                                                  38
                                                                 Cardiac Measurement
                                                                 and Analysis
Vivid 3 User Guide                                         Direction 2305203-100 Rev. 004                                                               39
Cardiac Output                                        Modifying Measurements                              Deleting Assigned Measurements
1. Measure the LVOT diameter in 2D/M Mode.
                                                      To modify a measurement:                            To delete assigned measurements:
2. Generate the spectrum to be measured and
   press Freeze.                                      1. Highlight the required measurements.             1. Highlight the required measurement.
3. Press Caliper.                                     2. Press Select twice.                              2. Press Menu.
4. Press the CO soft key.                             3. TRACKBALL the marker to the required position    3. TRACKBALL to Delete and press Select.
                                                         and press Select to anchor the marker.
5. TRACKBALL to the start time point of the trace
                                                      4. Perform the required modifications.                  OR
   and press Select.
6. Use the TRACKBALL to trace the Doppler             Assigning a Parameter Label After                   1. Highlight the required measurement.
   envelope from left to right on the required part   Measurement
   of the spectrum and press Select.                                                                      2. Press <Delete>. The highlighted measurement
                                                                                                             is deleted from the image and from the
7. TRACKBALL to the bar to indicate the end of the    •   A measurement can be taken more than once.
                                                                                                             database.
   heart cycle and press Select.                          Only the last measurement for a specific
                                                                                                             OR
8. To obtain the Cardiac Output (CO) estimation,          parameter label can be active and can be
                                                                                                             Press <Del Meas>.
   assign the result to SystCo (systolic cardiac          modified.
                                                                                                          To remove/clear all the measurements from the
   output), or to one of the heart valves.            •   The last six measurements for a parameter can
                                                                                                          screen:
9. The corresponding cross-sectional diameter of          be saved in the database and viewed in the
   the valve measured in Step 8 should be                 Worksheet.                                      •   Press Exit.
   measured and assigned using a corresponding        •   A measurement that has not been assigned to         OR
   2D image.                                              a parameter will not be retained in the             Press Caliper.
NOTE: The CO will be calculated, and will be              database.                                       NOTE: When measurements are removed using
displayed on the screen and in the Worksheet.         To assign a predefined parameter label to           the above procedure, measurements that have
                                                      a measurement:                                      been assigned are retained in the database.
General Controls                                      1. Highlight the required measurement.              Controlling the Result Table Size
                                                      2. Press Menu.
Highlighting Measurements                                                                                 To modify the size of the Result table:
                                                      3. TRACKBALL to the required label abbreviation.
To highlight a measurement:                           4. Press Select.                                    •   Rotate the Resize Results/Scroll Results soft
                                                                                                              key rotary.
•   TRACKBALL to the required measurement             To assign a user-defined parameter label to a
    marker on the image,                              measurement:                                        To scroll through the result table:
    OR                                                1. Highlight the required measurement.              1. Press the Resize Results/Scroll Results soft
    TRACKBALL to the required measurement in the                                                             key rotary to enable the Scroll Results
    result table in the upper-left corner of the      2. Press Menu.
                                                                                                             function.
    screen. The measurement is highlighted in         3. TRACKBALL to User and press Select.
                                                                                                          2. Rotate the Resize Results/Scroll Results soft
    green and can be modified.                        4. Use the alphanumeric keyboard to enter the          key rotary to scroll up or down through the
NOTE: There is an option to jump automatically to        required label.                                     results displayed in the Result table.
the next measurement.                                 5. TRACKBALL to OK and press Select.
                                                                   Cardiac Measurement
                                                                   and Analysis
Vivid 3 User Guide                                           Direction 2305203-100 Rev. 004                                                                          40
The Measure button activates this convention.          All the measurements and calculations taken during         •   Use the Scroll Content soft key rotary to
                                                       the examination can be viewed at any time using                navigate.
A list of studies (protocols) is displayed on the
                                                       the Worksheet. The Worksheet is the only source            To adjust the method by which a calculation is
clipboard area on the left of the screen. Each study
                                                       that retains all the measurement information.              made:
consists of a list of measurement labels.
Measurements are taken according to the order of       The Worksheet enables the user to view, edit, delete       1. TRACKBALL to the relevant cell in the Method
the listed measurement labels for the selected         or print data independently of a report. Measurements         column and press Select.
study.                                                 and values can be changed or deleted.                         The following options are available:
                                                       Worksheet formats provided are:                                •   Min
See the Reference Manual for a full list of all
available studies and parameters.                      •   Normal: Displays the parameters and                        •   Max
                                                           calculations for which measurements have
NOTE: The user can reconfigure the individual                                                                         •   Last
                                                           been assigned.
studies and their parameters. For further                                                                             •   Aver
                                                       •   Compact: Displays only the value of the
information, refer to Chapter 14, Presets and                                                                     2. TRACKBALL to the required option and press
                                                           measurements and calculations, and not the
System Setup in the User’s Manual.                                                                                   Select. The values are recalculated.
                                                           measurements themselves.
To select a study (protocol) and                       •   Expanded: Displays all available parameters
perform a measurement:                                     and calculations, regardless of whether                Including or Excluding Values in a
1. Press Measure.                                          measurements have been assigned to them.               Calculation
2. TRACKBALL to the required study and press           Accessing a Worksheet                                      •   TRACKBALL to the measurement that is to be
   Select.                                             To access a Worksheet:                                         included/excluded from the calculation and
3. Perform the measurement and press Select.           •   Press Worksheet.                                           toggle Select.
   The system will automatically jump to the next          On the left side of the screen, the list of mode-      NOTE: An excluded measurement changes color
   measurement to be performed, unless the                 specific pages for the Worksheet is displayed in the   to signify that its value has not been included in the
   Autojump option is enabled.                             clipboard area. The current page is highlighted (by    calculation. The value in the Value field is adjusted
To select a different study:                               default, this is according to the currently-selected   accordingly.
                                                           application).
•   TRACKBALL to the required study and press
    Select.                                            •   You are able to scroll through different worksheets,   Manually Changing a Value
                                                           then highlight the required worksheet and select it
                                                           by pressing the Select Worksheet button. The           1. TRACKBALL to the value that is to be changed
                                                           selected Worksheet list is displayed in the upper         and press Select.
                                                           left side of the screen.                               2. Use the alphanumeric keyboard to enter the
                                                       To scroll through the pages:                                  required value. An asterisk (*) indicates that the
                                                       •   Use the Scroll Pages soft key rotary to                   value has been manually altered.
                                                           navigate.                                              NOTE: To restore automatic calculation, position
                                                           OR                                                     the TRACKBALL over the asterisk and press Select.
                                                           TRACKBALL to the required page and press               The manually entered value is replaced by an
                                                           Select.                                                automatically calculated value.
                                                                     Cardiac Measurement
                                                                     and Analysis
Vivid 3 User Guide                                          Direction 2305203-100 Rev. 004                                                                 41
Manual Trace
                                                                                                    Vascular Measurements
                                                                                                    and Analysis
Vivid 3 User Guide                                         Direction 2305203-100 Rev. 004                                                                     42
Real-time Spectral Tracing and                       The IMT measurement is based on automatic               General Workflow
                                                     tracing of the leading edge of the intima surface and
Measurement Function                                 the leading edge of the adventitia surface. After       Scanning
1. Perform a scan in one of the Doppler Modes.       tracing the two surfaces, multiple measurements
2. Press Trackball, and position the TRACKBALL       are made between pairs of pixels located on both        1. Generate a longitudinal scan of the carotid
   over the spectrum.                                traces, as shown in the following figure:                  artery using normal scanning methods.
3. Press Menu.                                                                                               NOTE: In order to locate the ED time segment
                                                                                                             easily, we recommend using the ECG when doing
4. TRACKBALL to Show and press Select. To
                                                                                                             the IMT measurement.
   disable the function, press Select again.
5. Select Positive, Negative, or Both.                                                                       2. Optimize the posterior wall of the carotid. The
                                                                                                                measurement will be made from there.
NOTE: The accuracy of the tracing may be
adjusted by using Envelope Adjustment enabled        If some parts of the intima or adventitia are not       3. When the image is optimized, press Freeze.
by pressing the Soft Menu Rocker. Using the          sufficiently clear, the tracing algorithm omits them.
vertical arrows, select Envelope Adjustment. Use
                                                                                                             Measuring
                                                     As a result, the trace may appear discontinuous. To
the horizontal arrows for adjustment.                ensure accuracy, measurements are only taken            4. Use TRACKBALL to browse the cine-loop and
                                                     from parts of the trace which have been drawn.             locate a frame that was acquired during end-
Important: The same functions also can be                                                                       diastolic (ED). Activate ZOOM to improve
performed on frozen Doppler or archived images.      As soon as the trace is drawn and the multiple
                                                     measurements are made, the result window (shown            accuracy.
                                                     in Figure 1:) displays:                                 5. Press Measure.
Automatic IMT Carotid Artery                         •   The average of all of the measurement points        6. Select IMT.
Measurement                                              for the present trace (RTCCA IMT).                  7. Select either Left IMT or Right IMT, as shown in
                                                     •   The standard deviation calculated for all of the       the following figure:
Use this feature to measure the IMT (Intima-Media
                                                         measured points from the current trace
Thickness) of the posterior wall of the Carotid
                                                         (RTCCA IMT Sd).
artery, as shown in the following figure:
                                                     •   The maximum value (RTCCA IMT Max).
                                                     NOTE:
                                                     Due to the physics of ultrasound imaging, the IMT
                                                     of the anterior (near) wall of the carotid might be
                                                     incorrect. This is due to ultrasound artifacts. We
                                                     therefore recommend measuring IMT from the
                                                                                                             8. Place a vertical arrow anywhere inside the
                                                     posterior wall only. The algorithm in this package is
                                                                                                                carotid. This marks the area in the carotid
Measure the IMT using any good-quality 2D image      set to measure the IMT from the posterior wall only.
                                                                                                                where IMT tracing starts.
of the carotid. Standard scan optimization is        NOTE:                                                   9. Press SET. A second vertical arrow appears.
required, such as is required to produce any good-   Variations in blood pressure during the heart-cycle
quality image of the vessel.                                                                                 10. Position the second vertical arrow to mark the
                                                     cause the IMT to change. Therefore, we
                                                                                                                 end of the segment where IMT measurement
                                                     recommend measuring the IMT at ED (end-
                                                                                                                 should take place.
                                                     diastolic).
                                                                                                            Vascular Measurements
                                                                                                            and Analysis
Vivid 3 User Guide                                          Direction 2305203-100 Rev. 004                                                                   44
The time that has passed since the scan began            •   While in the single view FlexiView, press the LV   1. Insert the Convex C358 probe into an available
(“Time from Start”) is displayed in the upper-left           Monitoring soft key.                                  port.
corner of the live window. All cineloops stored in the   To exit FlexiView:                                     2. Press Probe.
archive retain the “Time from Start” stamp.              •   Select a preset other than OR in the Select        3. TRACKBALL to the appropriate OB application
NOTE: The timer of the live scan can be reset at             Probe and Application screen.                         preset and press Select.
any time.                                                                                                       It is now possible to use either the Caliper
To reset the “Time from Start” timer:                                                                           (Measure, or Measure and Assign methods) or
                                                         OB Package                                             Measure (Assign and Measure method) buttons to
1. TRACKBALL to the timer displayed in the upper-        NOTE: The OB applications are only available           perform measurements.
   left corner of the live image window.                 when the Convex C358 probe is connected.
                                                                                                                Measurements are accumulated in the database.
2. Press Menu.
                                                                                                                These measurements can be viewed and edited in
3. TRACKBALL to the Reset Time option and press          Operating the OB Package                               a Worksheet, and displayed in the on-screen Mini-
   Select. The next automatically stored loop will                                                              Report.
   reflect the new “Time from Start”.                    The general operation is similar to that of the
                                                         Cardiac Package. Refer to Chapter 6, Cardiac           On-Screen OB Mini-Report
FlexiView for Single Screen                              Measurement and Analysis in the User’s Manual for
                                                         detailed operating instructions.                       Each measurement made and assigned to a
A full single screen option enables the user to view                                                            parameter will appear in a condensed on-screen
full screen, live scanned images in all modes.           Patient Details                                        Mini-Report in the upper-right corner of the image.
Automatic storage of cineloops to the clipboard                                                                 During scanning, this on-screen Mini-Report is not
                                                         When an OB application is selected, the Main
according to user configured, predefined time                                                                   displayed. The Mini-Report is redisplayed when the
                                                         Details page of the Patient Details screen is
intervals takes place, without interrupting the                                                                 Caliper or Measure buttons are pressed.
                                                         modified to include:
procedure.
                                                         •   LMP
To activate full single screen FlexiView:
                                                         •   EGA
1. Select an OR preset.
                                                         NOTE: If the Application Type field in the Patient
2. Press the LV Monitoring soft key. The screen          Details screen is not OB, due to a different, prior
   automatically changes to a quad screen                procedure or application, TRACKBALL to OB in the
   display.                                              dropdown list and press Select.
3. Press the LV Monitoring soft key again. The
   system will display the live scan.
      Shared Services
Vivid 3 User Guide                                         Direction 2305203-100 Rev. 004                                                                   46
2D Caliper                                          1. Generate the image to be measured and press         1. Select an image to be measured and press
                                                       Freeze.                                                Freeze, or retrieve a cineloop from the archive.
1. Generate the image to be measured and press      2. Press Caliper.                                      2. Press Caliper.
   Freeze.                                          3. Press the Ellipse soft key.                         3. Press the HR soft key.
2. Press Caliper.                                   4. TRACKBALL to the start point for the                4. Press the Cine Scroll/Run Cine soft key rotary
3. Press the 2D Caliper soft key.                      circumference measurement and press Select             to activate and highlight the corresponding on-
4. TRACKBALL to the start point of the                 to anchor the caliper.                                 screen Run Cine soft key icon.
   measurement and press Select to anchor the       5. TRACKBALL to adjust the major axis until an         5. Press Active Mode twice to display the Cine
   caliper.                                            ellipse of the required shape and size is              soft key menu.
5. TRACKBALL to the measurement end point.             obtained, then press Select to anchor the           6. Use the First Marker and/or End Marker soft
6. Press Select to anchor the caliper.                 second caliper. The third and fourth calipers are      key rotaries to adjust the markers.
                                                       displayed along the minor axis, enabling the
7. Repeat steps 4 – 6 for additional                                                                       7. After positioning the markers correctly,
                                                       user to further adjust the shape and size of the
   measurements.                                                                                              TRACKBALL to the green checkmark next to the
                                                       ellipse.
8. Assign labels to the measurements as                                                                       heart rate in the upper-left corner of the screen
                                                    6. To adjust the size of the ellipse, TRACKBALL to        and press Select.
   described in the Assigning a Parameter Label        the appropriate caliper and press Select.
   After Measurement on page 39.                                                                              OR
                                                       TRACKBALL the caliper to the required location         To reject and cancel the measurement,
                                                       and press Select to anchor the position.               TRACKBALL to the red “X” and press Select.
                                                    7. Repeat steps 4 – 6 for additional
                                                       measurements.
                                                    8. Assign labels to the measurements.
      Shared Services
Vivid 3 User Guide                                           Direction 2305203-100 Rev. 004                                                                        48
OB Twins Measurement Tool                              3. From the list of pre-defined Fetal Growth_B          OB Graphs Tool
                                                          measurements displayed, select the required
The OB Twins Preset enables the measurement of            item. For example, BPD_B.                            The OB Graphs option is used to estimate fetal
twins. When performing an OB examination, all                                                                  growth and development against pre-defined
parameters for each fetus are clearly identified and                                                           standards - according to gestational age. For more
labelled (and saved) as "A" and "B".                                                                           details, refer to the Vivid 3 Release Notes SW 3.1
For more details, refer to the Vivid 3 Release                                                                 You may select the appropriate chart against which
Notes SW 3.1                                                                                                   to make a comparison (for example, Hadlock,
To activate the OB Twins measurement option:                                                                   Jeanty, etc.) from a collection of charts.
                                                                                                               To activate the OB graphs option:
1. Select the OB Twins preset (as shown below):
                                                                                                               1. Press Worksheet on the control console.
                                                                                                               2. Position the cursor on
                                                       4. Perform the selected measurement, using the             EGA and press Select.
                                                          cursor displayed on the screen.                          A chart is displayed (as
                                                       NOTE: In the worksheet, displayed measurements              shown below), depicting the fetal measurements
                                                       and parameters for Twin A and Twin B are clearly            in comparison to standard measurement values
                                                       separated by a thick horizontal line, as shown in the       at the same gestational age, according to the
                                                       example below.                                              selected author (Jeanty in this example).
                                                           Twin
                                                            A
2. Trackball to the pre-defined list of measurement
   options (left side of the screen) and make the
   appropriate selection in order to perform the
   required measurement. For example,                      Twin
   Fetal_Growth_B (as shown below):                         B
        Shared Services
Vivid 3 User Guide                                            Direction 2305203-100 Rev. 004                                                                     50
Hip Joint Measurement and                              4. Move the second line (Labrum Acetabular)
                                                          and drag it into the appropriate position, using
Analysis                                                  the rotary key to rotate it as necessary. Press
The Hip Joint Measurements package is a useful            Select to fix the line in position; the angle
diagnostic aid for assessing the development of the       between the two lines (α) is displayed.
infant hip. This provides two measurement tools for    5. Move the third line (Roof Line) and drag it into
performing the following measurements:                    position, rotating it as necessary. Press Select to
    •   Hip Joint Angles                                  fix the line in position; the ß (Beta) angle is
                                                          displayed.
    •   Diameter Ratio
                                                       NOTE: By default, the line names are displayed. If
Note: All hip joint measurements automatically
                                                       preferred the names may be hidden - for details,
appear in the worksheet and in the Hip Joint Report.
                                                       see Vivid 3 Release Notes SW 3.1.
For additional information, see the Vivid 3 Release
                                                       The measurement of the angles
Notes SW 3.1.
                                                       (degrees) between the lines is
                                                       automatically calculated and
                                                       displayed in a results box at the top of the screen,      NOTE: Pressing Cancel on the drop-down menu,
To perform Hip joint measurements:                                                                               cancels the currently-selected line; Delete removes
                                                       together with the measurement number - for
                                                       example, 1 as shown here.                                 the entire measurement (all three lines).
1. From the Presets menu (Small Parts), select
   Hip Joint.                                          NOTE: Measurements may be assigned to either              8. Repeat Step 3 through Step 7 for any additional
2. Freeze the image and press Caliper on the           the Left or Right hip, as described below.                   measurements, as required. These will be
   control console.                                                                                                 displayed on the screen accordingly.
                                                       6. If you wish to assign the measurements to the
   The Hip Joint soft rotary is displayed below the       Left or Right hip, trackball to the results box (the
   scanning screen:
                                                          cursor becomes hand-shaped) and right-click.
        Shared Services
Vivid 3 User Guide                                            Direction 2305203-100 Rev. 004                                                                      52
Reporting                                               5. Press the Save soft key.                            Deleting Images From Reports
                                                        6. Use the alphanumeric keyboard to enter a
For additional information, see Chapter 9,                 name for the new report in the dialog box.          The option to erase images from a report and
Reporting in the User’s Manual.                         NOTE: The system will automatically add the exam       "populate" it with new images, is provided.
                                                        date (DDMMMYY) as an extension of the saved            To delete images from a report:
Creating Reports                                        report name given by the user to assure easy and
To generate a report:                                   improved identification.                               1. Trackball to the image area corresponding to
                                                                                                                  the image you wish to erase from the report.
•   Press Report.                                       7. TRACKBALL to OK and press Select to save the
                                                                                                               2. Press Menu on the control console.
                                                           report.
Editing Reports                                         8. Press Report on the control panel to exit the
                                                                                                               3. A pop-up menu is displayed, as shown below:
To edit a report:                                          report and return to the current patient’s
                                                           scanning screen.
1. Press Report.
2. TRACKBALL to the report on the screen and press      Including Images in Reports
   Menu. An asterisk is displayed to the left of the
                                                        To add/replace images in a report:
   presently selected and displayed template.
                                                        1. Press Report.
NOTE: Instead of pressing Menu, you may use the
Template soft key, for easy access to the complete      2. TRACKBALL to the report on the screen and
list of report templates available.                        press Menu.
                                                        3. TRACKBALL to an alternative template layout that
NOTE: For details about configuring alternative
                                                           includes image fields and press Select to
report templates, refer to the Stress Tab and Report
                                                           highlight the template.
Tab sections in Chapter 14, Presets and System
Setup of the User’s Manual.                             4. TRACKBALL to the required image in the
                                                           clipboard and press Select. A yellow frame
3. TRACKBALL to an alternative template, if required,      indicates that the image has been selected.         NOTE: The illustration above shows only one
   and press Select to highlight the template.                                                                 image from a collection of images in the Report.
                                                        5. Simultaneously press Select and TRACKBALL
4. The report headings, such as Name, ID, Birth            the yellow framed image to the required             4. Select Clear Image to erase the
   Date, Indications, and Comments are                     location in the report template. When Select is        currently-selected image from the report.
   highlighted in green, and hyperlinked to                released, the image is placed and displayed in         Alternatively, select Clear All Images to
   facilitate editing.                                     the selected location.                                 remove all images from the report.
    •   TRACKBALL to such a heading and press                                                                  5. If necessary, use the drag-and-drop method to
                                                        Important: The user must save the report in order to
        Select.                                                                                                   "re-populate" the report with the required
                                                        be able to view the report at a later time with the
    •   Press OK or Cancel to return to the report      images as placed. Once a report is saved, no data         images.
        after editing is complete.                      or image can be added to the same report.
NOTE: In any reports, all light green fields are
hyperlinked. Double-clicking such a field will take
you to the original source field, where modifications
may be made if necessary.
                                          Reporting
Vivid 3 User Guide                                       Direction 2305203-100 Rev. 004                    53
Printing Reports
To print a report:
NOTE: The report is printed only to an externally
connected printer.
•   Press the PC Print soft key.
                                      Reporting
Vivid 3 User Guide                                                  Direction 2305203-100 Rev. 004                                                                                      54
Patient Management and                                      5. After locating and highlighting the required                     Annotations or measurements that are marked on
                                                               patient, press Select.                                           the screen are also stored with the image. When a
Archiving                                                   6. Edit the patient information as required.                        split screen image is stored, it is stored exactly as
                                                                                                                                displayed on the screen.
For additional information, see Chapter 10, Patient         7. Press the OK soft key.
Management and Archiving in the User’s Manual.              8. If the patient record cannot be located                          The stored images and cineloops are displayed as
                                                               (indicating that no previous record exists):                     thumbnails on the clipboard as shown below.
                                                                •    Press the Cancel soft key.                                 •    A filmstrip icon is displayed in the bottom left
Archiving Package (option)                                      •    Press the Begin Exam soft key.                                  corner of each cineloop thumbnail.
The Archiving Package enables the user to                                                                                       •    A split screen icon is displayed in the bottom
permanently store images and cineloops, as well as          Archiving/Storing Images and                                             left corner of each side-by-side thumbnail.
patient measurements and reports.                           Cineloops                                                           NOTE: To enable measurements to be made on
                                                                                                                                images recorded on VCR, the M&A on VCR option
Important: Without the Archiving Package, the data          When an image is stored, all the additional                         must be selected in the configuration setup. For
is only stored for the duration of the current              information that is displayed, such as the probe                    further information, refer to Chapter 14, Presets and
examination.                                                application and image settings, is saved with it.                   System Setup of the User’s Manual. In addition to the
                                                                                                                                archive, images and cineloops can also be backed up
A system message informs the user if the archived                                                                               on MOD (Magnetic Optical Disks) or CD-Rs.
patient information is currently in use.
                                                                                               Clipboard area
NOTE: In addition, the Archiving Package provides
the ability to generate and export a list of patients, in
Microsoft Excel format (see page 57). The exported
list provides a summary of demographic details
pertaining to patient exams performed on the system
for a specific period, as selected by the user.
                                                                                 Stored
                                                                                 cine
Beginning an Exam on an Archived Patient                                         loop
                                                                   Patient Management
                                                                   and Archiving
Vivid 3 User Guide                                          Direction 2305203-100 Rev. 004                                                                     56
To retrieve and/or edit an archived examination:      To retrieve a stress examination:                      Deleting Archived Information
1. Press Archive.                                     1. Press Archive.
                                                                                                             To delete a patient record:
2. Search the Patient List screen by scrolling        2. Using the Select Patient soft key rotary, select
   through the list using the Select Patient soft        the required patient record that includes the       1. Press Archive.
   key rotary.                                           stored stress examinations.                         2. Use the Select Patient soft key rotary to
3. Press the Exam List soft key.                      NOTE: Protocol is enabled and its label is lit only       highlight the patient record to be deleted.
NOTE: The images on the most recent                   for patient records that include stored stress         3. Press the Delete soft key. A pop-up dialog box
examination are displayed on the clipboard.           examinations.                                             is displayed.
4. Scroll through the exam list using the Select      3. Press Protocol.                                     4. TRACKBALL to Yes and press Select.
   Exam soft key rotary.                              4. Perform the stress analysis again, as required.     NOTE: The current patient record and examination
5. Press the Edit soft key.                              Refer to Chapter 5, Stress Echo in the User’s       cannot be deleted.
                                                         Manual for further information.                     To delete a group of patient records:
6. Edit the test as required, using the Select Page
   and Select Field soft key rotaries.                5. Press Protocol again to exit the examination,
                                                         saving all scoring and analysis, and to return to   1. Press Archive.
7. Press the OK soft key to save all changes                                                                 2. Press the Select Patients soft key rotary to
                                                         the current patient’s scanning screen.
   OR
                                                      To retrieve and/or edit a Worksheet:                      select the first patient record in the group of
   Press Cancel to ignore all changes.
                                                                                                                records that is to be deleted, and continue
To retrieve and/or edit an archived image:            1. Press Archive.                                         scrolling with the rotary to highlight the group.
1. Enter the archive and select the required          2. Using the Select Patient soft key rotary, select    3. Press the Select Patients soft key rotary again
   patient and examination.                              the required patient record.                           to indicate that the group is complete.
2. TRACKBALL to the required image on the             3. Press Worksheet.                                    NOTE: The selection process can be cancelled by
   clipboard and press Select.                        4. Using the alphanumeric keyboard, edit the           centering the TRACKBALL on the list and pressing
                                                         values as required.                                 Select.
Important: If the selected image displays             5. Press Worksheet again to return to the current      4. Press the Delete soft key. A pop-up dialog box
measurements and annotations, press the Remove           patient screen.                                        is displayed.
Overlay soft key to remove all annotations and
measurements                                                                                                 5. TRACKBALL to Yes and press Select.
                                                      Redisplaying the Clipboard
                                                                                                             NOTE: The current patient record and examination
3. Press Measure or Caliper and make the                                                                     cannot be deleted.
                                                      To redisplay the clipboard and its images:
   required measurements and annotations.
4. Press Store.                                       •   TRACKBALL over the clipboard area. The
                                                          thumbnail images are redisplayed.
                                                                  Patient Management
                                                                  and Archiving
Vivid 3 User Guide                                         Direction 2305203-100 Rev. 004                                                                  58
                                                                    Patient Management
                                                                    and Archiving
Vivid 3 User Guide                                                 Direction 2305203-100 Rev. 004                                                                                                      60
                                                                                                                                               Connectivity
Vivid 3 User Guide                                         Direction 2305203-100 Rev. 004                                                                         62
Supported Services                                  Exporting Data to DICOM Image                             The default system generated file is:
                                                                                                              Last Name_First Name_Birthdate
Services are the prime elements of a dataflow.      Storage, DICOM Print, Media                               (yyyy,mmm,dd)_Patient ID number.xls.
NOTE: The services associated with the System’s     Storage, Excel, or MPEGVue                                To activate the Export Shortcut:
dataflows cannot be configured by the user. A       Data may be exported from the System to various
meeting of both the GE Service representative and                                                             NOTE: Short-cut keys are assigned through the
                                                    configured services, including:
the local IT personnel must be arranged for                                                                   Archive Links tab in the System Configuration
configuration.                                      •    DICOM Image Storage                                  window - for details refer to the Vivid 3 Release
                                                    •    DICOM Print (see page 63)                            Notes SW 3.1.
The following services are supported by the
System:                                             •    Media Storage                                        1. When the current exam is finished and you wish
                                                    •    Excel                                                   to export it to another device, press the Patient
•   Echo Database 3.0                                                                                            ID button on the control console.
                                                    •    MPEGVue (see page 63)
•   DICOM Media Creator                                                                                          The Select window is displayed:
                                                    NOTE: Multiple patient exams may be exported
•   DICOM Storage
                                                    simultaneously to DICOM Image Storage or Media
•   DICOM Storage Commitment                        Storage services. However, when exporting to
•   DICOM Verification                              Excel, each exam must be exported individually.
•   DICOM Modality Worklist                         To export one or more patient exams:
•   DICOM Modality Perform Procedure Step
    (MPPS)                                          1. Press Archive.
                                                    2. Use the Select Patient/Select Patients soft
                                                       key rotary and select the patient, or patients
                                                       whose images are to be exported.
                                                    3. Press Menu.
                                                    4. TRACKBALL to the required option: Export to
                                                       DICOM Image Storage, Export to Media
                                                       Storage, or Export to Excel.
                                                    NOTE: Alternatively, if an Export short-cut key has       2. If the required Export function is located on one
                                                    been assigned, you may use this key - see the                of the soft-menu buttons (for example, Excel
                                                    instructions described below.                                Export as seen above), simply click the button.
                                                    5. Press Select.                                             The system will prompt for the next steps,
                                                                                                                 accordingly.
                                                    6. TRACKBALL to OK and press Select.
                                                                                                              3. If the requested Export function does not
                                                    Important: When exporting to Excel, the destination          appear on any soft-key (has not been defined
                                                    path to which the data will be sent must be assigned         as a short-cut), just press Cancel and proceed
                                                    in the Archive Tab. When exporting to Excel, a               with the normal export procedure (as explained
                                                    dialog box will be displayed enabling the user to            on page 62).
                                                    enter a name for the Excel file being created.
                                                                                                               Connectivity
Vivid 3 User Guide                                        Direction 2305203-100 Rev. 004                                                                     64
4. Select the required media - for example,        6. Name the media by typing a suitable name in              The system exports the entire exam data from
   MPEGvueCDR (or any other available                 the Media Label window.                                  each of the currently-selected patient records to
   destination media type, such as MPEGvueNET,     NOTE: You may use a maximum of 8 characters                 the selected media, together with an embedded
   MPEGvueUSB, etc.).                              and type in upper or lower case letters.                    MPEG viewer.
   The following dialog box opens (if              Do not use any special characters (for                      During this process, the exam review progress
   MPEGvueCDR was selected) and the standard       example, #, *, etc.)                                        is visible in the displayed sequence of preview
   procedure for writing to CDR will commence:                                                                 screens; the time required for export will
                                                   7. Trackball to Format and press Select.
                                                                                                               depend upon the amount of data; longer
                                                        The following warning message is displayed:            examinations will obviously require longer time.
                                                                                                               The progress status showing the percentage of
                                                                                                               completed exams (for example, done 60%) is
                                                                                                               displayed during the export process.
                                                                                                           NOTE: While export is in progress, do not
                                                                                                           attempt to use the system!
                                                   8. Trackball to Yes and press Select.                       When export is complete, the following
                                                        The media is formatted and the following               message is displayed:
                                                        message is displayed:
                                                                                                             Connectivity
Vivid 3 User Guide                                               Direction 2305203-100 Rev. 004                                                        66
Peripherals
For additional information, see Chapter 12,
Peripherals in the User’s Manual.                       Link                           Type
The following peripherals can operate with the          VGA Sockets (2)
System.
                                                        Socket                         Video Out
•   Sony Black & White Thermal Video Printer
                                                        Composite B&W                  CCIR or RS-130
•   Sony Color Thermal Video Printer
                                                        Composite Color                PAL or NTSC
•   HP InkJet Printer
                                                        S-Video                        Y/C, PAL or NTSC
•   Video Cassette Recorder (VCR)
•   JVC VCR                                             RGB                            PAL or NTSC
NOTE: The ultrasound system may also be used            Composite color PAL or         NTSC Video In
                                                        S-Video                        Y/C, PAL or NTSC
safely while connected to devices other than those
recommended, if the devices and their                   Audio Stereo Left-Right Out    Audio Out (to VCR)
specifications, installation and interconnection with
                                                        Audio Stereo Left-Right In     Audio In (from VCR)
the system conform to the requirements of IEC/EN
60601-1-1. The added device/s must have                 External Microphone
appropriate safety standard conformance and CE          Remote Print Command           Make-Break by Relay
Marking.                                                                               TTL Levels (+5V to Ground)
Printing                                              Video Cassette Recorder                                          The soft keys can be used to perform the following
                                                                                                                       actions:
The System supports printing for both black &         The VCR is connected to the unit and operated
white, and color thermal (on board) printers.         using the control panel buttons and soft keys.                   •   Play and Stop: Starts and stops the playback.
These two printers are controlled from the control                                                                     •   Fast Forward: Fast forwards the recording.
                                                      Important: Manual calibration and all measurement                •   Rewind: Rewinds the recording.
panel using the Print A and Print B buttons.
                                                      capabilities of the system can be performed on
To print on the Black & White thermal printer:        VCR-captured images.                                             •   Play speed: Adjusts the speed of all play
                                                                                                                           modes (regular playback, fast forward and
•   Press Print A button.                                                                                                  rewind).
                                                      VCR Operation
To print on the Color thermal printer:                                                                                 •   Freeze: Freezes the playback. The soft key
•   Press Print B button.                             The soft key map and table below describe the                        controls enable cineloop review.
                                                      function of each soft key in the VCR operation.                  •   Search: Searches for a specific counter
HP InkJet Printer Configuration                                                                                            number location while the VCR is in either Play
                                                      Normal      1       3          5           7      9        11        or Freeze status.
                                                      Pressed     2       4          6           8      10       12    •   Search Blank: Searches for the start of a blank
To activate the HP DeskJet Toolbox:
                                                                                                                           (unrecorded) section. (Operates in forwards
1. While scanning in 2D-Mode, press Freeze to
                                                                                                                           direction only).
   freeze the image.
                                                                                                                       •   Eject: Ejects the VCR tape.
2. TRACKBALL to the image and press Menu.
3. TRACKBALL to the Explorer option on the pop-        Soft Key       Status/Function                                  Important: Once connected to the unit, only the
   up menu and press Select.                                                                                           Eject button is active on the VCR machine.
                                                                      Play               Freeze         Stop
4. On drive U:, TRACKBALL to the appropriate
                                                       1                                 Cine Scroll
   printer file (For example, HP DeskJet 970c
   Series V.2.0 Toolbox, HP DeskJet 990,               2                                 RUN Cine
   HP Deskjet 6122) and press Select.
                                                       3              Stop               Stop           Play
5. Configure the printer settings according to the
                                                       4              Search             Search         Stop
   HP instructions.
NOTE: Other printer setting utilities are available    5
12 Freeze/Play Freeze/Play
        Peripherals
Vivid 3 User Guide                                       Direction 2305203-100 Rev. 004                                                                       68
•      Tape name
•      Counter number                                               Shuttle - Jog
•      VCR status
The following VCR status icons, displayed on the                    VCR is inactive, although a video
                                                                    tape is in the VCR.
screen in the VCR Control area, indicate the
current VCR function.
                                                   Important: Before starting to record, ensure that the
                                                   tape label has been entered into the system, and
                                                   that the VCR counter is set correctly.
To adjust the counter:                             To search for a specific counter number:             To search for the start of a blank (unrecorded)
                                                                                                        section:
1. TRACKBALL to the counter in the VCR Control     1. Press the Search soft key.
   area on the screen.                             2. TRACKBALL to the Counter Number field and         1. Press the Search soft key.
2. Enter an identifying name in the Tape name         use the alphanumeric keyboard to manually         2. TRACKBALL to Search Blank and press Select:
   field, if required.                                adjust the counter number to the required
3. Enter the counter number in the counter field      number.
   OR                                              3. TRACKBALL to the Go To button and press
   TRACKBALL to Reset Counter and press Select        Select. The VCR will begin the search,
   to reset the counter to 0:00:00.                   displaying a Search icon in the VCR Control
4. TRACKBALL to Done and press Select to save         area of the screen. Once the required counter
   the changes.                                       number is found, the VCR status icon will
                                                      switch to the Play icon, and the VCR will begin
NOTE: Selecting Cancel will undo the current                                                               The search commences and the message
                                                      playing back the tape from the user-selected
changes to the counter.                                                                                    "Searching blank space, please wait" is
                                                      location.
                                                                                                           displayed.
                                                                                                           On reaching the start of the unrecorded section,
                                                                                                           the search process stops and the message
                                                                                                           window closes.
       Peripherals
Vivid 3 User Guide                                                 Direction 2305203-100 Rev. 004                                                          70
For additional detailed information, see Chapter 13,                                                     Each probe is labeled with the following
Probes in the User’s Manual.                                                                             information:
                                                                                                         •   Name of distributor and manufacturer
Environmental Requirements                                                                               •   Operating frequency
                                                                                                         •   GE model number
Probes should be operated, stored, and transported
within the parameters outlined below.                                                                    •   Probe serial number
                                                                                                         •   Month and year of manufacture
                                                                                                         The probe name, displayed on the probe housing
                                                                                                         and the connector, can be read when the probe is
 Property     Operational      Storage          Transport
                                                                                                         connected to the unit.
 Temp.        10o to 40o C     -10o to 60o C    - 40o to 60o C
              50o to 104o F    14o to 140o F    - 40o to 140o F
Probe Orientation
Probe Name
                                                     Probes
Vivid 3 User Guide                                                  Direction 2305203-100 Rev. 004                                                                           71
The probe name, together with a picture of the                 •   Vivid 3 Systems with 3 ports:                              Probe Integration
probes currently connected to the unit, is displayed               The Select Probe and Application screen on the
on the Select Probe Application screen.                            Vivid 3 ultrasound unit will display four possible                      The transesophageal probes require
•   Vivid 3 Systems with 2 ports:                                  probe connections, three active probes and one                          special handling. Refer to the user
                                                                   pencil probe, as shown below:                                           documentation enclosed with these
    The Select Probe Application screen on the
                                                                                                                                           probes.
    Vivid 3 ultrasound unit will display three
    possible probe connections, two active probes
    and one pencil probe as shown below:                                                                                      Connecting the Probe
                                                  Probes
Vivid 3 User Guide                                           Direction 2305203-100 Rev. 004                                                                    72
                                          Probes
Vivid 3 User Guide                                                  Direction 2305203-100 Rev. 004                                                                                  73
Lens
                                                Probes
Vivid 3 User Guide                                         Direction 2305203-100 Rev. 004                                                                       74
                                         Probes
Vivid 3 User Guide                                          Direction 2305203-100 Rev. 004                                                      75
                                         Probes
Vivid 3 User Guide                                               Direction 2305203-100 Rev. 004                                                             76
•   Linear Array Probes: Linear Array probes are        i or T              Intraoperative                Phased Array Sector Probes
    designated by the prefix/suffix "L".
                                                        T or P              TEE (Single or Multiplane)
•   Intraoperative Probes: Intraoperative probes                                                          3S Probe
                                                        P or D              CW Doppler
    are designated by the prefix "i" or "T".
                                                                                                           Catalog No.         H4550SZ
                                                       Non-Imaging Doppler Pencil Probes
                                                                                                           Intended Uses       •     General Cardiology
                                                                                                                               •     Pediatric Cardiology
                                                       P2D Probe                                                               •     Transcranial
                                                                                                                               •     Vascular
                                          Probes
Vivid 3 User Guide                                                Direction 2305203-100 Rev. 004                                                              77
7S Probe                                                   9T Probe                                          Linear Array Probes
Doppler Frequency 4.0 - 5.0 Image Frequency 4.0 - 10.0 Catalog No. H40212LH
                                                  Probes
Vivid 3 User Guide                                          Direction 2305203-100 Rev. 004                            78
T739 Probe
                                                     E721 Probe
 Catalog No.         H40212LM
                                                      Catalog No.         H40206LN
 Intended Uses       •     General
                     •     Intraoperative             Intended Uses       •     Transvaginal-
                                                                                OB/GYN
 Image Frequency     5.7 - 10.0                                           •     Transrectal-
                                                                                Urology
 Doppler Frequency   4.4 - 6.7
                                                      Image Frequency     4.0 - 10.0
i8L Probe
                                                     PAMTE Adapter
 Catalog No.         H45511NW
Transesophageal Probe
                                            Probes
Vivid 3 User Guide                                               Direction 2305203-100 Rev. 004                                                                79
Presets and System Setup Selecting an Application Preset User-created Application Presets
For additional information, see Chapter 14, Presets        To select a probe application preset:                To create a new user-created application preset:
and System Setup in the User’s Manual.                     1. Press Probe.                                      1. Select a probe and application preset.
                                                           2. TRACKBALL to the required application preset      2. Adjust the scanning parameters as required.
Application Presets                                           and press Select.                                 3. While the unit is in scanning mode,
                                                                                                                   simultaneously press <Ctrl> + <Preset>.
Application presets, which determine the settings a        Modifying Factory-defined Presets
probe will use for the study, are configured for each                                                           4. TRACKBALL to the Create new field and use the
probe type. The presets are selected in the Select                                                                 alphanumeric keyboard to type a name for the
                                                           To modify a factory preset:                             new application preset.
Probe and Application screen. The name of the
current application preset is displayed in the title bar   1. Select the probe and application preset.          5. TRACKBALL to Save with New Name and press
at the top of the screen when in any scanning              2. Adjust the scanning parameters in any of the         Select.
mode.                                                         scanning modes, as required.
                                                                                                                Important: Ensure that the preset is saved under a
There are three different types of presets:                3. While the unit is in scanning mode,
                                                                                                                unique name. If a factory-defined or user-modified
                                                              simultaneously press <Ctrl> + <Preset>.
•   Factory-defined presets: Supplied as part of                                                                preset name is used, the previous preset will be
                                                           4. TRACKBALL to Save Current and press Select        replaced and lost.
    the system.
                                                              to save the user-modified preset under the
•   User-modified presets: Modified by the user               factory-defined preset name.
    and saved under the factory-defined name.                                                                   To delete a user-created application preset:
•   User-created presets: New presets created              Restoring Factory-defined Presets                    1. Press Probe.
    and defined by the user, and saved under a                                                                  2. TRACKBALL to the name of the user-created
    unique name.                                           Important: Only user-modified presets can be            preset that is to be deleted.
                                                           restored to their original settings.                 3. Press Menu.
Important: New preset settings are valid only for the
specific combination of probe, application icon, and                                                            4. TRACKBALL to Delete and press Select.
                                                           To restore factory presets:
patient type.
                                                           1. Press Probe.                                      Important: Only user-created presets can be deleted
                                                           2. TRACKBALL to the application preset for which     from the system. Factory-defined presets and user-
                                                              the factory preset settings are to be restored,   modified presets cannot be deleted.
                                                              then press Select.
                                                           3. Simultaneously press <Ctrl> + <Preset>.
                                                           4. TRACKBALL to Restore Factory and press
                                                              Select.
Enables the user to assign short-cut keys for     Enables the user to manage stress templates,              Enables the user to set parameters for the video
easy archiving of an exam. In this way, the       including:                                                recorder and ECG unit connected to the System.
linking of source (From) and destination (To)     •   Stress template selection.                            Configures the printer assignment of both Print A
directories saves time, by eliminating the need   •   Default stress template designation.                  and Print B control panel buttons.
to scroll through many windows.
                                                  •   Definition of the order of Quad View stress
Export functions may be assigned as                   exams during Analyze Mode.                            Annotation Settings Tab
short-cuts to any one of the four assignable      •   Templates available in the System, listed in the
                                                                                                            Enables the user to define the primary and
soft-menu buttons:                                    left area.
                                                                                                            secondary fonts used for annotations. In this tab,
   •   DICOM Print                                •   Templates activated for the Stress Package,           the user can define the fonts, as well as the default
                                                      listed in the right area.                             position (home location) of the text cursor on the
   •   DICOM Storage Server
                                                                                                            screen when the Text key on the alphanumeric
   •   DICOMVueCDR                                Important: The Stress tab will only be displayed if the   keyboard is pressed.
   •   DICOMVueMOD525                             Stress Package option is installed.
   •   Excel Export                                                                                         System Options Tab
                                                  Report Tab
   •   MPEGVueCDR                                                                                           Enables the user to install optional features to the
   •   MPEGVueNET                                 Enables the user to set general parameters related        system, such as: Patient Archive, 3D Option,
                                                  to reports, including:                                    MPEGvue, and DICOM Print.
1. Press Config.                                  •   Report template selection for specific                         If an incorrect password is entered in the
2. TRACKBALL to the Archive Links tab and             applications.                                                  Key field and the Apply button is
   press Select.                                  •   User-defined report template creation.                         selected, the option, even if it has already
   The Archive Links tab is displayed.            •   Default report template designation.                           been correctly installed in the System, will
3. TRACKBALL to the left side of the window       •   Insertion of user-supplied logo into report                    be removed from the System.
   and select the required action; select the         template(s).                                          To access the System Options Tab:
   required export function to be associated      •   All available factory-defined templates, any
   with the shortcut (refer to the above list).                                                             1. Press Config.
                                                      user-created templates for the selected
                                                      application, or any user templates designed on        2. Trackball to System Options tab; press Select.
4. Click OK.
                                                      the Report Designer, listed in the left area.         3. Trackball to the required option and activate the
                                                  •   All templates that the user has selected which           appropriate radio button:
                                                      are currently active, listed in the right area.           •   Patient Archive
                                                                                                                •   3D Option
                                                                                                                •   MPEGvue
                                                                                                                •   DICOM Print
                                                                                                            4. Click OK.
                                                                                                              User Maintenance
Vivid 3 User Guide                                             Direction 2305203-100 Rev. 004                                                                      86
Troubleshooting                                          Activity Log and Problem Reports                         2. Enter a description of the problem or
                                                                                                                     occurrence.
The system is engineered for reliable operation and
                                                         The System automatically logs any error messages         3. Select the Attach screen snapshot checkbox
consistent, high-quality performance. Automatic
                                                         that occur during system operation, on a daily              to attach the current screen capture, if required.
self-testing facilities are provided to monitor system
                                                         basis, for technical support analysis purposes (if
operation and to detect faulty operation as soon as
                                                         required). These daily log reports are retained in the
possible, thereby eliminating unnecessary
                                                         system and can be recalled for 90 days.
downtime. The detection of any serious malfunction
results in immediate interruption of scanning, and is    A reporting tool exists which enables the user or GE
indicated by an error message on the screen.             service personnel to download the log report to a
                                                         floppy disk.
Monitor Alignment
                                                                                                               User Maintenance
Vivid 3 User Guide                                            Direction 2305203-100 Rev. 004                                                                    88
              The concerns listed below can             The images and calculations provided by the             Damaged probes or improper use and manipulation
              seriously affect the safety of            System are intended for use by competent users as       of the transesophageal probe can result in injury or
              patients undergoing a diagnostic          a diagnostic tool. They are explicitly not to be        increased risk of infection. Inspect probes
              ultrasound examination.                   regarded as the sole, irrefutable basis for clinical    frequently for sharp, pointed or rough surface
                                                        diagnosis. Users are encouraged to study the            damage that could cause injury or tear protective
                                                        literature and reach their own professional             barriers (gloves and sheaths).
Patient Identification                                  conclusions regarding the clinical utility of the
                                                        system.                                                               Transesophageal probe safety:
Always include proper identification with all patient                                                                         Never use excessive force when
data and verify the accuracy of the patient’s name      The user should be aware of the product                               manipulating the transesophageal
and/or identity number when entering such data.         specifications and of the system accuracy and                         probe. The detailed user manual
Ensure that the correct patient ID is provided on all   stability limitations. These limitations must be                      enclosed with the transesophageal
recorded data and hard copy prints. Identification      considered before making any decision based on                        probe should be read carefully.
errors could result in an incorrect diagnosis.          quantitative values. If in doubt, consult the nearest
                                                        GE Ultrasound Service Office should be consulted.                Electrical
                                                        Equipment malfunction or incorrect settings can                  Hazard
                                                        result in measurement errors or failure to detect
                                                        details in the image. The user must become              A damaged probe can increase the risk of electric
                                                        thoroughly familiar with the operation of the unit in   shock if conductive solutions come in contact with
                                                        order to optimize its performance and to recognize      internal live pads. Inspect probes often for cracks or
                                                        possible malfunctions. Application training is          openings in the housing and holes in and around
                                                        available through the sales representative.             the acoustic lens, or other damage that could allow
                                                                                                                moisture to enter. Become familiar with the use and
                                                                                                                care precautions described in the Care and
                                                                                                                Maintenance section, in Chapter 13 in the User’s
                                                                                                                Manual.
        Safety
Vivid 3 User Guide                                           Direction 2305203-100 Rev. 004                                                                       90
        Safety
Vivid 3 User Guide                                                 Direction 2305203-100 Rev. 004                                                                            92
Owner Responsibility                                        For information about specific requirements and               Safety of Ultrasound
                                                            regulations applicable to the use of electronic
It is the responsibility of the owner to ensure that        medical equipment, consult the appropriate local,
anyone operating the system reads and                       state and federal agencies.
                                                                                                                          ALARA Statement
understands this section of the manual. However,
there is no representation that the act of reading                           For USA Only: Federal law restricts          Ultrasound procedures should be performed using
this manual renders the reader qualified to operate,                        this device to use by, or on the orders       output levels and exposure times As Low As
inspect, test, align, calibrate, troubleshoot, repair, or                   of, a physician.                              Reasonably Achievable (ALARA) while acquiring
modify the system. The owner should make certain                                                                          clinical information.
that only properly trained, fully qualified service
                                                                                                                          The System employs the principle of ALARA in
personnel undertake the installation, maintenance,
troubleshooting, calibration, and repair of the
                                                            Possible Biological Effects                                   configuring factory defaults.This default level takes
                                                                                                                          effect:
equipment.                                                  For a detailed explanation of possible biological
                                                            effects, refer to the Reference Manual, which                 •   When the system is first turned on.
The System owners should ensure that only
                                                            includes a copy of Medical Ultrasound Safety,                 •   When a new patient is entered.
properly trained, fully qualified personnel are
                                                            American Institute of Ultrasound and Medicine,
authorized to operate the system. Before                                                                                  •   When changing from one exam category to
                                                            1994.
authorizing anyone to operate the system, it should                                                                           another.
be verified that the person has read, and fully                                                                           •   When changing from one application to
understands, the operating instructions contained in        Power Information on Screen                                       another.
this manual. It is advisable to maintain a list of
                                                            In the scan parameter area, two fields are allocated          Refer to the Reference Manual, which includes a
authorized users.
                                                            for the display of power values as follows:                   copy of Medical Ultrasound Safety, American
Should the system fail to operate correctly, or if the                                                                    Institute of Ultrasound and Medicine, 1994, for
unit does not respond to the commands described                                                                           further details.
in this manual, the user should contact the nearest          MI =
Field Service Office.                                        TIX =                                                        Training
        Safety
Vivid 3 User Guide                                               Direction 2305203-100 Rev. 004                                                               94
Electromagnetic Compatibility                           There is no guarantee that interference will not           To comply with the regulations on electromagnetic
                                                        occur in a particular installation. If this equipment is   interference for a Group 2, Class A Medical Device,
(EMC)                                                   found to cause or respond to interference, which           all interconnecting cables to peripheral devices
AII types of electronic equipment may                   may be determined by turning equipment on and              must be shielded and properly grounded. Use of
characteristically cause electromagnetic                off, qualified service personnel should attempt to         cables not properly shielded and grounded may
interference with other equipment, transmitted          correct the problem by one or more of the following        result in the equipment causing or responding to
either through air or connecting cables. The term       measures:                                                  radio frequency interference, in violation of the
Electromagnetic Compatibility (EMC) indicates the                                                                  European Union Medical Device Directive and FCC
                                                        •   Reorient or relocate the affected device.
capability of the equipment to curb electromagnetic                                                                regulations.
influence from other equipment, while at the same       •   Increase the separation between the unit and
                                                            the affected device.                                   Do not use devices which intentionally transmit RF
time not affecting other equipment with similar                                                                    signals, for example, cellular phones, transceivers
electromagnetic radiation.                              •   Power the equipment from a source other than
                                                                                                                   or radio controlled products, in the vicinity of this
                                                            that of the affected device.
Radiated or conducted EMC can cause distortion,                                                                    equipment, as they may cause performance outside
degradation or artifacts in the ultrasound image that   •   Consult the service representative for further         the published specifications. Keep the power to
could potentially obscure diagnostic information.           suggestions.                                           these types of devices turned off when near this
                                                        The manufacturer is not responsible for any                equipment.
Important: This unit carries the CE mark. The           interference or responses caused by the use of
System complies with regulatory requirements of         interconnecting cables other than those
the European Directive, 93/42/EEC, concerning           recommended, or by unauthorized changes or
medical devices. It also complies with emission         modifications to the unit. Unauthorized changes or
limits for a Group 2, Class A Medical Device as         modifications could void the user’s authority to
stated in EN 60601-1-2 (IEC 60101-1).                   operate the equipment.
        Safety
Vivid 3 User Guide                                                  Direction 2305203-100 Rev. 004                                                                 96
                                             Index
Vivid 3 User Guide                                             Direction 2305203-100 Rev. 004                                                               97
ellipse tool                       hide doppler display 20                         OB Graphs 48                       owner responsibility 92
      OB package 46                Hip joint measurement and analysis 50           OB Twins 48                        P
enter new patient 14               hospital tab 80                                 perform 16
environmental requirements 5       HP InkJet printer configuration 67          mechanical hazards                     packages
                                                                                                                            archiving 54
      probes 70                    I                                               probes 75
examination                                                                    miscellaneous tab 82                         GYN 49
      begin 14                     icons                                       M-Mode 18                                    OB 45
      record 68                           alert 2                                  2D image 18                              operating room 44
examinations                       iLinq 84                                        resize 19                          patient details
      exercise stress 34           image                                           total gain 18                            OB package 45
export                                    delete 57                                zoom 18                            patient identification 89
      image/loop 59                       export 59                            modes                                  patient information
      shortcuts 62                 image acquisition                               2D 17                                    access 14
      to DICOM Print 63                   stress echo 33                           angio 21                           patient record
      to MPEGvue 63                images and cineloops                            CFM-Mode 18                              delete 56
                                          store 16                                                                    patient safety 89
F                                                                                  combined 94
                                   Initialization 8                                M-Mode 18                          perform
factory presets                    insert                                          regular doppler 19                       backup 58
      modify 79                           text annotation 16                       split screen 17                          measurements 16
      restore 79                   L                                               standby 8                          periodic maintenance 84
fetal heart evaluation 49                                                          stress echo 33                     peripherals 66
FlexiView 44                       label                                                                                    connect 8
                                        probe 70                               modify
      single screen 45                                                             factory preset 79                        external connection 91
foot brake 4                       linear array probes 77                                                                   internally connected 91
                                   log on                                          measurements 39
footswitch 4                                                                   monitor                                      specifications 66
      connect 8                         users 82                                                                      personnel and equipment safety 90
                                   LVO imaging 20                                  adjust 11
format                                                                             alignment 86                       phased array sector probes 76
      disk 59                      M                                                                                  phono filter 32
                                                                               moving unit 9
front and side view                MA options tab 80                           MPEGvue 63                             power
      unit 4                       maintenance                                                                              display values 92
                                                                               N
G                                      periodic 84                                                                    printer configuration 67
                                       probes 73                               new group                              printers tab 82
GYN package                                                                         define 35                         probe labeling 70
   abdominal evaluation 49             user 84
                                   measure and assign convention 37            new patient                            probe orientation 70
   fetal heart evaluation 49                                                        enter 14                          probes
                                   measurement tools
H                                                                              non-imaging doppler pencil probes 76         biological hazards 75
                                       hip joint 50
handling probes 85                     OB package 46                           O                                            connect 71
hazards                                vascular 41                                                                          curved array (convex) 77
                                                                               OB package                                   definitions 76
     biological 91                 measurements                                     OB Graphs tool 48
     electrical 90                     2D-Mode 37                                                                           disconnect 9, 72
                                                                                    OB Twins measurement tool 48            disinfect 74
     explosion 90                      assign parameter label 39                    patient details 45
     implosion 90                      carotid artery 42                                                                    electrical hazards 75
                                                                               obstetric evaluation 45                      environment requirements 70
     mechanical 89                     delete 39                               operating environment 5
     moving 90                         doppler mode 38                                                                      handling 85
                                                                               option                                       holders 4
     pacemaker 91                      highlighting 39                              3D 25
heart cycles                           hip joint 50                                                                         linear array 77
                                                                                    biopsy guide 27                         maintenance 73
     store 35                          M-Mode 37                               orientation
heart sound microphone device 32       modifying 39                                                                         mechanical hazards 75
                                                                                    probe 70                                non-imaging doppler pencil 76
                                       Index
Vivid 3 User Guide                                                    Direction 2305203-100 Rev. 004                                                                   98
     phased array sector 76                         biological hazard 91                  steering                                            VCR/ECG 81
     ports 4                                        changing imaging modes 93                   vascular applications 20                      vocabulary 82
     safety 75                                      diagnostic information 89             storage 4                                     templates
     select 14                                      electrical 91                         store                                               add/delete levels 35
     transesophageal 78                             electrical hazard 90                        images and cineloops 16                       editing 35
problem report 86                                   explosion hazard 90                         images on VCR 16                              report 53
purge backup data 59                                implosion hazard 90                   store heart cycles 35                         text annotation 16
R                                                   mechanical hazards 89                 stress echo                                         delete 16
                                                    moving hazard 90                            analyze 34                              time from start stamp 45
record                                              pacemaker hazard 91                         image aquisition 33                     timers
      examination 68                                patient 89                                  score 34                                      T1 34
reference cineloop 44                               patient identification 89             stress echo mode 33                                 T2 34
regular doppler mode 19                             personnel and equipment 90            stress tab 81                                 total gain
removable media utility screen 59                   probe 75                              supported services 62                               M-Mode 18
report tab 81                                       probe selection 93                          DICOM media creator 62                  traces
reports                                             transmit power 94                           DICOM modality perform procedure step         ECG signal 32
      activity log 86                               ultrasound 92                                          62                                 ECG trace amplitude 32
      create 52                                     user modification 88                        DICOM modality worklist 62                    phono 32
      edit 52                                 scanning                                          DICOM print 62                                sweep speed 32
      images 52                                     operations 16                               DICOM storage 62                        training 92
      installing a new report template 53           screen 15                                   DICOM storage commitment 62             transesophageal probe 78
      problem 86                              schedule backup 58                                DICOM verification 62                   transporting unit 10
      retrieve 53                             score stress echo data 34                         echo database 3.0 62                    troubleshooting 86
      view 53                                 screens                                     switch off 8                                        monitor alignment 86
requirements                                        removable utility screen 59           switch on 8                                   twins
      electrical safety 7                           scanning 15                           system                                              graphs 48
      environmental 5                               select probe and application 9, 79          add user 82                                   measurement 48
      operating environment 5                 select                                            edit user 83                            typographic conventions 2
      power 5                                       application preset 79                 system configuration 80                       U
      site 5                                  select probe and application 14                   perform 80
resize M-Mode 19                              select probe and application screen 9, 79   system malfunction 87                         ultrasound safety 92
restarting unit 87                            shutdown                                    system options tab 81                         unit
restore                                             full 8                                system tab 80                                       cleaning 85
      factory preset 79                                                                                                                       connect 7
                                              soft keys 21                                T
result table                                        functions 22                                                                              front and side view 4
      size 39                                 soft menu                                   T1 timer 34                                         moving 9
retrieve                                            VCR 68                                T2 timer 34                                         rear view 4
      report 53                               soft menu rocker 24                         tabs                                                restart 87
ROI                                                 functions 24                               annotation settings 81                         switch off 8
      size 23                                 sonographers 5                                   archive 80, 81                                 switch on 8
S                                             speakers 4                                       connectivity 82                                transporting 10
                                              split screen mode 17                             hospital 80                              unit acclimation time 11
Safety                                                                                         MA options 80                            user groups 82
     ALARA 92                                 standards
                                                    IEC 601-1 91                               miscellaneous 82                         user log on 82
     application selection 93                                                                  printers 82                              user maintenance 84
     electromagnetic compatibility (EMC) 95         possible biological effects 92
                                                    UL-2601-1 91                               report 81                                user modification
safety 1, 88                                                                                   stress 81                                      safety 88
     alert icons 2                            standby mode 8
                                              static electricity interference 85               system 80
     application selection 93                                                                  system options 81
                                               Index
Vivid 3 User Guide                                  Direction 2305203-100 Rev. 004   99
V
vascular applications
     steering 20
vascular measurement
     2D-Mode tools 41
vascular measurement tools 41
VCR
     counter 69
     operation 67
     search for start of blank section 69
     soft menu 68
     store images 16
VCR/ECG tab 81
view
     report 53
     stored image 55
vocabulary tab 82
voltage
     checks 7
     level 7
W
warnings 5
WEEE 91
worksheets 40, 49
Z
zoom 16
    2D-Mode 17
    CFM-Mode 18
    M-Mode 18
Index