CCT — 8300 Series
Conductivity/TDS/Resistivity/Temp.
Configuration transmitting controller
Instruction ManualI Function and application
e
ty
2
3)
4)
3)
6)
1)
2)
3)
7)
8)
»
10)
1)
12)
13)
14)
Applications and features
With 32 bit processor ARM as the core of the powerful kernel and operating system, make
products to IT new era.
Using advanced modulus hybrid technology, intelligent measure range automatically
measurement switch
Language man-machine interface, Guided operation menu, Humanized design process
128X64 matrix LCD display, Chinese/English language, multi parameter display in one
screen;
Grading permission password protect function, prevent noncertified person freely revise the
engineering parameter when log in.
LED background lighting can select long time lighten or delay sleep, fit low carbon
fashion,
Characteristic feature
New conductivity measurement technology, which integrates conductivity/ resistivity/
TDS/ temperature multi parameters measurement.
Meet ultra pure water(0,054,S/em)~ Condensed water (100mS/em) wide range
measuremes
Conductivity. resistivity, TDS can selected arbitrary, set and operate senior link, Simplified
setting process
Professional temperature measurement/temperature control, replace temperature
instruments directly
‘The virtual instrument technology compensate the nonlinearity relationship between
resistivity and temperature, Compensation of high precision and good repeatability
Double channel, full isolated current loop, conductivity/TDS/resistivity/temperature signal
configuration freely
double mode current loop output, Compatible with all types of signal disposal, set up
high-end system integration
Configurable photoelectric switch control (Choose to conductivity’ resistivity’
temperature/time),drive pulse metering pump or expand control directly;
For regular cleaning and maintenance, and water treatment technology of time associated
provides time function (cycle timing/appointed timing) ;
Power supply DC24V, fit high humidity site safety code, Port polarity internal identify
automatically
Calendar function, time setting, and making an appointment, provide time tags for
recording data;
SMT, AOLICT, computerized whole machine inspection, Strict quality control ;
Product batch electrical aging, Surface three protective treatment, Perfect electronic
production process management, high level quality;
Facing the quality supervision opening measure inspection function, Any authorized
verification institutions can use legal measure for verification;
Complete electromagnetic compatibility design, Good anti-disturbance performances, Can
satisfy most of electromagnetic environment operate safely
116) Electric meter and conductivity cell mark and inspect respectively, Package and storage
respectively; According to the purchasing to choose international order process.
3
4
© Main Technical
Instrument application
) High end water quality management, high automatic level selection:
) high pure water/ ultra pure water measure control, service to electron, electric power,
pharmaceutical, fine Chemicals, clinical medicine and Life Science Research ;
> Industry process solution salinity analysis,
‘concentrate management, cooling water circulation industrial coating and t
equipment
) used for inline mot
service to water treatment, seawater desalination,
water treatment
(oring in metallurgy, petrifaction and industry cleaning ete
dex
Model CCT~8301/CCT - 8302
: ‘Single-channel / double channel conductivity/TDS /resistivity/ temperature
Deseription :
|Configuration transmitting controller
Conductivity cell model | CON2126A-13 CON2124A-13. | CON3123A-13 | CONSI21A-13
Conductivity cell type 10,00cm" 1.000em 0.100cm"" 0.010cm
Conductivity cell constant
| Temperature compensation|
es (10.042)em" (1.080.2)em"" | (0.1020.02)em" | (0,01020.002)em
Measurement range
200pSicm~100mSlem | SpS/em~10mSfem | (0.5~200)pS/em | (0.05~18.25)MAem
Conductivity 0.5 p8/em~100.0mS%ern
[Measureme] Resistivity 0.05MQ-cm~18.25MQ-em
nbvange TDS 0.25ppm~SOppt
Temperature 0-s0'c
Conductivity 0.01nS/em
Resolution | Resistivity 0.01M2-em
‘TDs .01ppm
‘Temperature ore
Conductivity 15 level
Accuracy | Resistivity 15 level
TDs 15 level
temperature 05
Prio00
Work condition
temperature: 0~50°C Relative humidity; <85%RHCurrent output [Double channel, isolated ,configuration 4-20mA output, Maximum loop resistance 5002.
[Accuracy: 20.1mA
‘Triple OC output
Control output
Load current 25mA (DCSV) load voltage DC24V (max)
‘Communication function S485 communication (Modbus agreement)
Power supply pemVve4v
Basic power consumption 3.9W
Max power consumption 48w
Protection level IP6S (with rear cover)
Installation mode Panel
Dimension 96 mm x96 mm *10Smm (H«W»D)
Slot dimension 91mm*9tmm
IL. Dimension and wire connection
i Outline dimension-Terminals
@ Wire connection
24VA, Power input interface, connect DC 24V
Internal automatic polarity recognition
24VB Power input interface, connect DC 24V
RS485A, Communication interface
RS485B Communication interface
14 Instrument mode output, instrument internal power supply.
Ti+/T1-__| Transmitter mode output, conditioning module external power supply
THI2- Instrument mode output, instrument internal power supply
T2412 ‘Transmitter mode output, conditioning module external power supply
ocioci’ The first channel photoelectronic Switch controlled contact,
oc2/0C2" The second channel photoelectronic Switch controlled contact
ocsiocs” - The third channel photoelectric Switch controlled contact
Instrument wire connection Conductivity cell wire connection
CCTS301 CCTS302 (CON2126A-13/CON?2124A-13/CON3123A-13/ CONSI21A-13
CELLI' CELLI' | CELLI White
CELL2' ceLt2' | CELL2 EMPTY
CELLS ceLt3' | CELL3 Green
CELLA CELL’ | CELL4 Yellow
CELLS" ceiis' | CELLS Red
CELLO CELL6' | CELL EMPTY
Connect with EARTH port of AC/DC Power conversion module
EARTH EARTH | EARTH 5
- (for removing the interference )-mA transfer wire connection
“| Left: External connection under
meter mode
tenis
Right: External connection under
: | 7
transmitter mode a Hegel
i]
1 Z| t-be --
|
NOTE: The second channel mA transfer wire connection is the same as above.
iv Electricity wire connection diagram
ee
ental eadpnant 2
fy.
Diagram: 0-1 Please use middle relay to drive and expand high pressure or
Dynamic load.EE Metering Metering
| pump 3 ] pump 2
Photoelectronie switch controls metering pump
Y .Conductivity high limit and window control
uS/em(ppm)
Action high limit alarm
Action low limit release
High limit control application methodE.G. 1, Water quality below high limit.
In the flelds of pharmaceutical, food, beverage, water purification, precision cleaning,
electronics and processing industries, water quality will be limited at high-water limit "Hi",
the solenoid valve to switch the direction of flow, in order to ensure the purity of process
water; When the water quality returns to "LO * system back to the original process. The
period between " Hi “and " LO "is the delay period. The delay function effectively avoid the
solenoid valve shock
pS/cm(ppm)
Hi JAction high limit alarm
|Action low limit release
ts
Water quality Control area between high and low
E.G. 2, Window- Control area between high and low
In the circulating cooling water, cutting coolants, cleaning fluids and other industrial
applications, when the water quality degradation to the high limit "Hit wil discharge or
regeneration, when the water restored to limit "LO", stops to discharge. This is the section
type window-control
vi. Resistivity low limit and window-control diagram
MQ,cm (ppm)
ai High iit lam release
Lo
Low limit alarm aston
ts
Jow limit control application method
Water quality should be controlled above the low limitUnder High pure / ultrapure water measurement control, water quality control request above
17MQ + cm ormore, when the water quality below the Lo setting, the solenoid valve will
switch the direction of flow, in order to ensure high purity process water; When the water
return to "HI" , system will back to the original process. The period between "LO" and “HI” is
the delay period, the delay function effectively avoid solenoid valve shocks.
Til. Front panel and key functions
i Front panela and main interface
is
LEFT'S n00n
12.30 wom
ne OMA
ConductivityrResistviy
\Xel|
CCT ~8301 front panel CCT-8302 front panel
And main interface And main interface
ii, Manual operation
Keyboard | description Funetion
1. Return to previous menu after parameter preserve.
[e} Backspace | 2.Cancel the current setting
3. Switchover the main and inquiry interface.
1. Cursor Movement key
Dp | see 2. Under the parameter setting mode, select the items shift left to fight
elect : :
3. Switehover the main and temperature, current output
interface(CCT8302)
1. Digital 0-9 cycle operation keys
2. Under the parameter setting mode, select the items shift up to down
Add :
A 3. Switchover the main and temperature, current output
interface(CCT8302)
J | enter 1. Enter the password setting menu |
2. Save the parameter and enter next mentsSoft keyboard
(suggestion)
Mark meaning
Cancel the current setting or return to the previous menu
T Up shift menu
2. Digital 0-9 cycle operation
1 down shift menu
2.after enter parameter query, down shift to check the parameter setting
1.left shift menu
2. after enter parameter query , enter the parameter display interface
2. after enter parameter query , enter the parameter display interface
save the parameter and enter next menu
[Note]: the soft keyboard
is the current role paraphrase of hard keyboard corresponding which
display on LCD screen,
TV.settings and operation
i. Main monitoring interface
Curent Conroe time
07-12-03 17: 13:00]
ff row
12°35 ee =]
pen OO isi
(Cont ate
[ cusrent ame
07-12-03,
cere NEG eng. Saha
a | 5 0
alec
cones] Cot tae
8 eo :
Ea" i lo 00 a} 1
cH
B= 20 OO
= ae[ote] CCT-8301 conductivity data and temperature data can transmitting output at the same time,
‘main interface altemate display double channel transmitting data.
‘CCT-8302 when any channel high alarm or low alarm, the channel indicator and alarm
indicator recycling flash to indicate,
ii, Menu fuction introduction
Press” under Main monitoring interface enter the log in menu, select engineer Permission and
‘input password 2000 enter the menu setting,
Menu Pee :
Number ce Menu Function introduction
description
Blectrode constant, temperature compensation, measuring u
; Measurement | Filter coefficient, data migration, measurement high limit,
parameter resistivity display precision and TDS Transform coefficient setting
menu showing automatic link with the setting
Control Triple channel photoelectric switch parameter setting, measured
2 - pt i,
parameter parameters any configuration control output
3 Transmitting, Double channel transmitting parameter setting which can configure
| parameter and control output the parameter
4 Communication | RS485 baud rate and address code setting, communication
parameter protocol, see Appendix A
A Password Engineering users, normal users two level permission settings
setting
6 Time setting __| Set and modify the clock in the time zone
Backlighting : :
7 fae constant lighting and delay close backlight can be select
L setting
8 Alarm setting | Setting the buzzer ON ot OFF |
Language
9 Under engineering permission setting Chinese and English displey
selection
Restore factory setting, save present seiting, restore previous
10 parameter
setting (CCT8301)
Manufacturer | Inquire manufacturer information and website, obtain service
n
information | information
a). Measure parameter
[No [item select Function Introduce
I select conductivity cell type O.0lem™, O.lem™ . 10cm",
10.0c!
1 cell constant
2 check the mark on matched conductivity cell, input the accurate
cell constant
10Iselect measure parameter to conductivity/TDS , temperature
3 | Temperature | coefficient compensation default 0.020
compensation | 2, select measure parameter to resistivity, temperature compensation
select ON
Select conductivity’TDS/resistivity corresponding unit uS/em .
3 | Measure unit :
mSlem. ppm. ppt, MQ-em KQem
4 _| Fiterconstants | Change filter constant, optimize Measure stability
= | When measured value and standard value error bigger, modify data
3 | Data transfer ee as cl
migration achieve its accuracy
Only aimed at resistivity, measurement high limit set_as
Measurement
ee 20MSem/our,
igh limi
. ‘Transmitting output 0~20MQ-cm corresponding 4~20mA,
Resistivity
7 | sete secicion | Measure display data select one or two decimal places
display precision
g | TDSconversion | Measurement unt elect TDS, can modify the Conversion relation of
coefficient
ppm and S/em
b).Control parameter
No | Item select Funetion Introduce
1 configuration mode select (conductivity/TDS/resistivity/temperature )
1 OC Isetting | 2.contoal mode select (Normally on /Normally off /pulse)
3. Action value set_ (on/off)
| configuration select (conductivity/TDS/resistivity/Aemperature)
2 | OCIIsetting | 2.control mode select ( NO/NC /pulse )
3. Action value set_ (ON/OFF)
3 [ocuttseting | ! telat cleaning cleaning period/leaning tine) |
2.Time cleaning(start time! finish time)
¢). Transmitting parameter
No | Item select Function Introduce
1 select channel 72
1 | Tansmitting! | 2.select measured parameter ( conductivity/ TDS/ resistivity/
setting | temperature)
3.sct4mA~20mA corresponding data
~ | Tselect channel 1/2
| Transmitting I | 2.select measured parameter ( conductivity’ TDS/_resistivity/
setting
temperature)
3.set 4mA~20mA corresponding data
4), Password setting
No Item select Function Introduce
1 | Engineer | Under engineer user permission, can change both engineer user
user___| and normal user password, ay
2 Normal user | Under normal user permission, only can change the normal user
nal
password.
¢).Parameter management (CCT-8301)
No | lem select
Funetion Introduce
1 _| Restore factory
Restore to the factory default setting, this function used to quality
is setting supervise, measure and inspection,
2 | Save present | Revise the present parameter and save present every
setting _| parameter setting
3 | Restorelast | This function used to unaware operation influence the
setting _| measure control, restore to the last setting.
Under main monitoring status, press
i Parameter inquiry interface
interface through instrument panel
enter parameter seting inquiry
Teton Ta ae eam ae
EET ee en aeons toon
Tae] _CETST [3] =
ST aaa aa era ae
conteciviny Eo | OSE) lo: me
SS ess CiasT T=)
|e
clenise
[ao
Crate pea) C=
fear cles 2 1 00 [CED rnin ett: 12 + BO
SO eee
ing Lom aa
TEESE, tepecatre |_|
a eT)
a sien
20st: 0000.00
mars} = mest) SS Tes tS
la
ne cat ae Bry
eo CEST Cs)
jap
Career Sarre
tom | tet ew!
foes) es re)
=
1.00 tvan
ost
a5[Note}: Under parameter ing terface or parameter set interface, it would
back to main interface if without operation over 3 minutes.
V.Sensor installation _ nes
.Sensor constant selection
Maem 10 1 ot 001 1kOom 0.4 0.01
50.02] 00s
2 DE [0205 | 200 50 [28
wm Ot 1 10 100 tmsiem 10 100
Industrial Concentrated
‘Ultra pure
water water solution
water
NOTE. Please choose the suitable sensor to get accurate reading, Such as, it’s
better to choose K=0.1 sensor for RO two stage outlet, if choose K=1.0, the
reading will not be so good.
ii installation and maintenance
(1) The electrode should be installed in a place in the pipeline where the stream is steady and
air bubbles are hard to generate.
(2) No matter the conductance cell is horizontally or vertically installed, it should be deeply
inserted into the moving water and against the flow direction
(3) The conductivity signal is weak electronic signal and its collecting cable should be
separately installed, When threading cable joint or connecting terminal board is used, to
avoid wetting interference or breakdown of measurement unit circuit, they should not be
connected to the same group of cable joint or terminal board with the power line or
control line.
B(4) When the measurement cable needs to be lengthened, it’s recommended to make an
agreement with the factory before placing an order.
(3) Please keep the measuring part of electrode clean, and do not directly contact the surface
by hands or contact with the oil stain objects
(6) Electrode is a kind of precision components, so please do not change any part of the
electrode. The accuracy will be incorrect if the electrode was destroyed by the strong
acid, strong alkali, scrape from machine and etc.
(1) The meter is made by precision integrated circuit and electronic components, so it
needs to place in case or dry environment.
(8) Ifthe strong acid , strong alkali, or mechanical rubs may change the sensor constant.
(9) The sensor cable can’t be cut or extended by clients, or the
i. Model and outline diemension
-—_____1imn_
a = pier]
8 Fp
12271
CONS121A-13 (C=0.01em") sensor outline dimension
®
@
0
©
©
o
®CON2124A-13 (C=1.0cm") outline dimension
133m
mm __. VUZNET Pipeline threed
|
o | [os
*CON2126A-13 (C=10.0cm"') sensor outline dimension
0127
a12V DC No
Transmitter mode THT No DC24Vv
Common Fault Analysis
Phenomenon Possible factors
Elimination method
A. check instrument power input terminal between 24V A and
ing | Ax Power not connected ieee
No reading ao a 24Y B, if there is 24V power
| B. instrument fault :
B._please contact with professional maintenance person
‘A. Sensor connection
Display error + adjust according to the manual
instability | B. bubble in pipelin
instability
C. Water quality
A
© | B. Rectification of pipeline or choose measuring point
C. use stable source of water exclusion of instrument reason
‘A. Constant setting error
B. Electrode constant
reset electrode constant
: change
Reading, . change new sensor or re-calibrate sensor constant
C. velocity measuring 5 fl
Yarget | aint unsuliable install conductivity cell in proper position of velocity
ion | D» Follow the sensor i
D. Sensor installation | D» Follow the sensor installation instructions
error _|
Low | A. velocity is not enough | A. Improve sensor installation position
resistivity | B. Sensor pollution
B. Remove the sensor and clean the body it carefully!
A. Instrument or se
snsor | A Put the sensor off-line inspection, Proof and exclude
Big error on fault installation reasons, re-setpoint to make corrections
reading B. Instrument setting | B.If off-line measure result still differ greatly, focus on
error
inspecting instrument parameter setting
red‘A. Instrument or sensor
Reading low
when under | fault A. Find out the cause of instrument and sensor, take
high pure |B. Sensor settinghave | countermeasures
water dead angle B. Clean the sensor and improve installation position.
measurement
A estou Sl ADC current meter directly connected in series measure loop
‘Transmission |B. PLC Intemat | SUE compare with Meter value, Exclude instrument's
data different | engineering quantity | "°*S°"
setting is incorrect
B. Inspect mA loop resistance, reset receiving modules of
migration
Note}:
4) High pure water, ultra pure water can’t choose opening or sampling measurement
‘comparison method, when high pure water exposed in the moment of the air, immediately
large amounts of carbon dioxide quickly to dissolved into the water, meanwhile vessel wall
not clean and dust in the air will be to dissolve in water, lead to double error, high pure
‘water only allow the sealed, liquidity, side stream circulation slot verification, laboratory
‘apparatus for opening measuring to measure high pure water is a wrong cognition, can not
comparable.
‘Water regeneration easy to cause the sensor pollution, after cleaning with alcohol cotton
ball can be recovered.
© Indicator I set (contain one pair of fast clamp)
© Operation manual L copy
© Conductivity Sensor | piece (standard cable 5 meter)
© Resistivity Sensor | set (mateh/standard cable 10 meter)
© = Waterproof rear cover | piece
‘Note}: CCT8301 match with 1 piece sensor, CCT8302 match 2 pieces sensor
1) Appoint the conductivity cable length when order the product, it would according to the
2
3)
4a
standard allocation leave factory if there is no special request;
select the suitable conductivity cell according the measured media before leave the factory.
obiain the accurate measure data and resolution ratios
allocate the conductivity cell ’s constant 0.01em -I directly when measuring the high pure
water,: Allocate the conductivity cell's constant 0.lem-1 when measuring KQ-em level
water
‘The meter’s power supply should be appointed the stander of AC power supply when select
24V DC power.
“Without the influence on the operation, any small change or improvement on the products by the
manufacturer will not be notified separately, Please make the object as the standard