GS 12A01A03-01EN
n General
The model FLXA 202 2-Wire Analyzer, one model
TM
FLEXA, FLXA, SENCOM are trademarks or registered trademarks of Yokogawa Electric Corporation.
All other company and product names mentioned in this document are trademarks or registered trademarks of their respective companies.
We do not use TM or ® mark to indicate those trademarks or registered trademarks in this document.
All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 12A01A03-01EN Mar. 23, 2018-00
3
2-3. Inductive Conductivity (ISC) ■ Output Range
■ Input Specification DO concentration:
Compatible with the Yokogawa inductive conductivity mg/l (ppm):
ISC40 series with integrated temperature sensor: min.: 1 mg/l (ppm)
NTC30k or Pt1000. max.: 50 mg/l (ppm)
ppb:
■ Input Range min.: 1 ppb
Conductivity: 0 to 2000 mS/cm at 25 ºC reference max.: 9999 ppb
temperature. % saturation:
Temperature: -20 to 140 ºC min.: 10 %
Cable length: max.: 600 %
max. 60 meters total length of fixed sensor Temperature:
cable + WF10(J) extension cable. min. span 25 ºC
Influence of cable can be adjusted max. span 170 ºC
by doing an AIR CAL with the cable
connected to a dry cell. ■ Performance (Accuracy)
(The specifications are expressed with simulated
■ Output Range inputs.)
Conductivity: Performance in ppm mode:
min. span: 100 µS/cm Linearity: ±0.05 ppm or ±0.8% F.S., whichever is
max. span: 2000 mS/cm (max 90% zero greater
suppression) Repeatability: ±0.05 ppm or ±0.8% F.S., whichever
Temperature: is greater
min. span 25 ºC Accuracy: ±0.05 ppm or ±0.8% F.S., whichever is
max. span 160 ºC greater
■ Performance (Accuracy) Performance in ppb mode:
(The specifications are expressed with simulated Linearity: ±1 ppb or ±0.8% F.S., whichever is
inputs.) greater
(Output span is 0-100 µS/cm or more) Repeatability: ±1 ppb or ±0.8% F.S., whichever is
Conductivity: greater
Linearity: ±(0.4 %F.S. + 0.3 µS/cm) Accuracy: ±1 ppb or ±0.8% F.S., whichever is
Repeatability: ±(0.4 %F.S. + 0.3 µS/cm) greater
Temperature: ±0.3 ºC Temperature
Step response: 90 % (< 2 decades) in 8 seconds Linearity: ±0.3 ºC
Note: "F.S." means maximum setting value of analyzer Repeatability: ±0.1 ºC
output. Accuracy: ±0.3 ºC
2-4. Dissolved Oxygen (DO) Note: "F.S." means maximum setting value of analyzer
output.
■ Input Specification
The FLXA202 accepts output from membrane 2-5. pH/Oxidation-reduction Potential (pH/ORP)
covered Dissolved Oxygen sensors. These sensors with digital sensor, FU20F pH/ORP SENCOM
can be Galvanic type, where the sensor generates Sensor
its own driving voltage or Polarographic type, where ■ Input Specification
the sensor uses external driving voltage from the Bi-directional digital communication (RS-485)
converter. between FU20F and FLXA202
The input range is 0 to 50 µA for Galvanic sensors ■ Input Range (depending on FU20F)
and 0 to 1 µA for Polarographic sensors. pH: 0 to 14 pH
For temperature compensation, the FLXA202 ORP: -1500 to 1500 mV
accepts Pt1000 (DO30 sensor) and NTC22k rH: 0 to 100 rH
elements (OXYFERM and OXYGOLD sensors). Temperature: -10 to 105 ºC
■ Input Range ■ Output Range
Dissolved Oxygen: 0 to 50 mg/l (ppm) pH: min. span 1 pH
Temperature: -20 to 150 ºC max. span 20 pH
DO30G sensor: ORP: min. span 100 mV
Measurement range: 0 to 20 mg/l (ppm) max. span 3000 mV
Temperature: 0 to 40 ºC rH: min. span 2 rH
Hamilton sensors: max. span 100 rH
Oxyferm: Temperature: min. span 25 ºC
Measurement range: 10 ppb to 40 ppm max. span 170 ºC
Temperature range: 0 to 130 ºC
Oxygold G:
Measurement range: 2 ppb to 40 ppm
Temperature range: 0 to 130 ºC
Oxygold B:
Measurement range: 8 ppb to 40 ppm
Temperature range: 0 to 100 ºC
All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 12A01A03-01EN July 30, 2019-03
4
3. Electrical ● Maximum Load Resistance
pH/ORP (analog sensor), SC and DO:
■ Output Signal
Refer to the Figure 1.
General: One output of 4-20 mA DC
Note: Tolerance: ±0.02 mA ISC and pH/ORP SENCOM sensor:
Bi-directional HART digital communication, Refer to the Figure 2.
superimposed on mA (4-20mA) signal ■ Display
Output function: LCD with a touch screen:
Linear or Non-linear (21-step table) Black/White: 213 x 160 pixels
Burn out function: (NAMUR 43 except ISC) Contrast adjustment available on the touch screen
Without HART/PH201G: Message language:
Down: 3.6 mA 12 (English, Chinese, Czech, French,
(signal: 3.8 to 20.5 mA for pH/ORP, SC German, Italian, Japanese, Korean,
and DO) Polish, Portuguese, Russian and Spanish)
(signal: 3.9 to 20.5 mA for ISC) One analyzer has all 12 languages.
Up: 22mA Note: Description for a selection of language and
With HART/PH201G: language names are written in English.
Down: 3.6 mA for pH/ORP, SC and DO Note: Only English alphabet and numeric are available
for a tag number, an additional description for each
Down: 3.9 mA for ISC value on the display screen and passwords.
(signal: 3.8 to 20.5 mA for pH/ORP, SC Note: Only for message language on the screen, 12
and DO) languages are provided.
(signal: 3.9 to 20.5 mA for ISC)
Up: 22mA 4. Mechanical and others
■ Power Supply ■ Housing
Nominal 24 V DC loop powered system Case, Cover:
One (1) Sensor module (1 input): • Aluminum alloy cast + epoxy coating
16 to 40V DC (for pH/ORP (analog • Aluminum alloy cast + urethane coating
sensor), SC and DO) • Aluminum alloy cast + high anti-corrosion
17 to 40V DC (for ISC) coating
21 to 40V DC (for pH/ORP SENCOM Color: Silver gray
sensor) Protection: IP66 (except Canada), NEMA Type 4X
Two (2) Sensor modules (2 inputs): (USA), CSA Type 3S/4X (Canada)
22.8 to 40V DC (for pH/ORP (analog sensor), ■ Cable and Terminal
SC and DO) Cable size:
Note: When the FLXA202 is used in the multi-drop mode Outer diameter:
of HART communication, the output signal is 6 to 12 mm (suitable for M20 cable gland)
changed from 12.5 mA DC to 4 mA DC just after the Terminal screw size: M4
power is turned on. Enough power supply for the torque of screw up: 1.2 N•m
instruments is to be provided. Wire terminal:
Pin terminal, ring terminal and spade
2-sensor measurement terminal can be used for analyzer’s power
1295
supply terminals and sensor terminals.
Load resistance (Ω)
V - 11.5
R=
0.022 Grounding terminal:
1000 Ring terminal should be used.
Pin terminal: pin diameter: max. 1.9 mm
Ring and spade terminal: width: max. 7.8 mm
600
■ Cable Entry
Digital Communication
Range (HART) 3 holes,
250 M20 cable gland x 3 pcs
Close up plug x 1 pc
0 1617 18 22.8 24.7 40 Note: Cable gland and plug are delivered with an analyzer,
Voltage (V) but not assembled into the analyzer.
Figure 1 Supply Voltage and Load Resistance ■ Mounting
for pH/ORP (analog sensor), SC and DO Mounting hardware (option):
• Universal mounting kit (Note)
• Pipe and wall mounting hardware
1295
• Panel mounting hardware
Load resistance (Ω)
V - 11.5
R= Note: This kit contains the pipe and wall mounting
0.022
1000 hardware and the panel mounting hardware.
Except SENCOM
Hood (option):
• Stainless steel
600 • Stainless steel with urethane coating
516
Digital Communication
• Stainless steel with epoxy coating
304 Range (HART) ■ Stainless Steel Tag Plate
250
When the additional code “/SCT” with a tag number
is specified, the tag plate on which the tag number is
0 17 18 22.86 24.7 40 inscribed is delivered with the analyzer.
18.2 21 Voltage (V) Tag plate is hanging type.
Figure 2 Supply Voltage and Load Resistance
for ISC and pH/ORP SENCOM sensor
All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 12A01A03-01EN Mar. 23, 2018-00
5
■ Conduit Adapter 5. Digital Communication
Using optional adapter
■ Kind of Digital Communication
• G1/2 (quantity: 3)
• HART (HART 5) or PH201G dedicated distributor
• 1/2NPT (quantity: 3) Note: Only one kind of digital communication is available
• M20 x 1.5 (quantity: 3) for one analyzer.
These conduit adapters are delivered with an
analyzer, but not assembled into the analyzer. ■ Output Value Parameter (HART)
Four value parameters (measured values) are
■ Size of Housing Case available for one digital communication.
165 x 165 x 155 mm (W x H x D) (without cable gland) • For 1-sensor measurement, these parameters are
■ Weight measured values.
Approx. 2.5 kg • For 2-sensor measurement, refer to the next item.
■ Ambient Operating Temperature ■ Digital Communication of 2-Sensor
-20 to +55 ºC Measurement (HART)
■ Storage Temperature Even when two sensor modules are installed, only
-30 to +70 ºC one digital communication is available for 2-sensor
measurement.
■ Humidity Four value parameters can be selected from the
10 to 90% RH at 40ºC (Non-condensing) followings;
■ Document Measured values of two sensors
Following documents are delivered with an analyzer; Calculated data of 2-sensor measurement
Paper copy: Redundant system output
Start-up Manual ■ Specific Contact Output with dedicated
written in English distributor, model PH201G (Style B)
Safety Precautions The distributor, model PH201G, is designed to
written in English connect with the 2-Wire Analyzer.
CD-ROM: This distributor supplies drive power to the analyzer
Start-up Manual and receives simultaneously 4-20 mA DC signal from
written in English the analyzer.
User's Manual This signal is converted to 1-5 V DC signal in the
written in English distributor.
Safety Regulations Manual This distributor also receives digital signals
for European region superimposed on the 4-20 mA DC signal, and
written in 25 languages provides contact outputs
General Specifications Input/Output signal:
written in English Number of available drive/signal point: 1
Technical Information Output signal: 1-5 V DC (2 points) (Note)
for HART Communication Load resistance: 2 kΩ or less (1-5 V DC output)
written in English Isolation system: Loop isolation type
User Setting Table Note: Two output signals for one analyzer’s analog output
of 5 kinds of measurement/sensor type are provided. Two 1-5 V DC output signals are same.
written in English Contact output:
Contact rating:
250 V AC, maximum 100 VA
220 V DC, maximum 50 VA
Hold contact output:
NC contact, normally energized
Contact closes when power is off or
during Hold situation.
Fail contact output:
NC contact, normally energized
Contact closes when power is off or
during Fail/Warning conditions.
Wash contact output:
NO contact
Contact closes during wash cycles.
All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 12A01A03-01EN Mar. 23, 2018-00
6
n Regulatory Compliance (FLXA202) ■ Explosion Protected Type Compliance
■ Safety, EMC and RoHS Compliance ‘Type’
Item Description in MS
Safety: UL 61010-1 code
UL 61010-2-030 Europe [Intrinsic safety “ia”] -CB
CAN/CSA-C22.2 No.61010-1 (ATEX) Applicable Standard:
CAN/CSA-C22.2 No.61010-2-030 EN 60079-0: 2012 + A11: 2013,
EN 61010-1 EN 60079-11: 2012
Certificate No:
EN 61010-2-030
DEKRA 11ATEX0109X
EMC: EN 61326-1 Class A,Table 2 (For use in Marking/Rating:
industrial locations) II 1 G Ex ia IIC T4 Ga
EN 61326-2-3 Ambient Temperature: -20 to 55°C
RCM: EN 61326-1 Class A,Table 2 Control Drawing: Refer to (1)
Korea Electromagnetic Conformity International [Intrinsic safety “ia”]
(IECEx) Applicable Standard:
Standard Class A 한국 전자파적합성 기준 IEC 60079-0: 2011, IEC 60079-
Russian: TR CU 020/2011 11: 2011
RoHS: EN 50581: 2012 (Style 1.02 or newer) Certificate No: IECEx DEK 11.0044X
Installation altitude: 2000 m or less Marking/Rating: Ex ia IIC T4 Ga
Category based on IEC 61010: I (Note 1) Ambient Temperature: -20 to 55°C
Pollution degree based on IEC 61010: 2 (Note 2) Control Drawing: Refer to (1)
Note 1: Installation category, called over-voltage category, United [Intrinsically safe / Nonincendive] -CD
specifies impulse withstand voltage. States (FM) Applicable Standard:
Equipment with “Category I” (ex. two wire Class 3600: 2011, Class 3610:
transmitter) is used for connection to circuits in 2010, Class 3611: 2004, Class
which measures are taken to limit transient over- 3810:2005, NEMA 250: 2014,
voltages to an appropriately low level. ANSI/ISA 60079-0: 2013, ANSI/
Note 2: Pollution degree indicates the degree of existence ISA 60079-11: 2014
of solid, liquid, gas or other inclusions which may Certificate No: 3039632
reduce dielectric strength. Degree 2 is the normal Marking/Rating:
indoor environment. IS CL I, DIV 1, GP ABCD CL I, ZN
Information of the WEEE Directive 0, AEx ia IIC
This product is purposely designed to be NI CL I, DIV 2, GP ABCD CL I,
ZN 2 IIC
used in a large scale fixed installations only
T4: for ambient temperature:
and, therefore, is out of scope of the WEEE -20 to 55°C
Directive. The WEEE Directive does not apply. Enclosure: Type 4X
The WEEE Directive is only valid in the EU. Control Drawing: Refer to (3)
Canada [Intrinsically safe / Nonincendive]
(CSA) Applicable Standard:
C22.2 No.0-10 (R2015), CAN/
CSA-C22.2 No.94-M91 (R2011),
C22.2 No.213-M1987 (R2013),
CAN/CSA-C22.2 No.60079-0:11,
CAN/CSA-C22.2 No.60079-
11:14, CAN/CSA-C22.2
No.61010-1-12, CAN/CSA-C22.2
No.61010-2-030-12
Certificate No: 2425510
Marking/Rating: Ex ia IIC T4 Ga
Intrinsicaly safe for Class I,
Division 1, Groups A, B, C, D, T4
Nonincendive for Class I, Division
2, Groups A, B, C, D, T4
Ambient Temperature: -20 to 55°C
Ambient Humidity:
0 – 100% (No Condensation)
Enclosure: IP66, NEMA 4X
Control Drawing: Refer to (2)
All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 12A01A03-01EN Mar. 23, 2018-00
7
‘Type’ ‘Type’
Item Description in MS Item Description in MS
code code
United [Nonincendive] -DD Europe [Type of protection ‘n’] -DB
States (FM) Applicable Standard: (ATEX) Certificate:
Class 3600: 2011, Class 3611: Not Applicable as per Annex VIII
2004, Class 3810: 2005, NEMA to ATEX 2014/34/EU
250: 2014 Applicable Standard:
Certificate No: 3039632 EN 60079-0: 2012 + A11:2013,
Marking/Rating: EN 60079-11: 2012, EN
NI CL I, DIV 2, GP ABCD ZN 2 IIC 60079-15: 2010, EN 60529:
T4: for ambient temperature: 1991+A2:2013
-20 to 55°C Marking/Rating:
Enclosure: Type 4X II 3 G Ex nA ic IIC T4 Gc
Control Drawing: Refer to (3) Ambient Temperature: -20 to 55°C
Canada Applicable Standard: Enclosure: IP66
(CSA) C22.2 No.0-10 (R2015), CAN/ Control Drawing: Refer to (4)
CSA-C22.2 No.94-M91 (R2011), International [Type of protection ‘n’]
C22.2 No.213-M1987 (R2013), (IECEx) Applicable Standard:
CAN/CSA-C22.2 No.61010-1-12, IEC 60079-0: 2011, IEC 60079-
CAN/CSA-C22.2 No.61010-2- 11:2011, IEC 60079-15: 2010
030-12 Certificate No: IECEx DEK 16.0034X
Certificate No: 2425510 Marking/Rating: Ex nA ic IIC T4 Gc
Marking/Rating: Ambient Temperature: -20 to 55°C
Nonincendive for Class I, Division Enclosure: IP66
2, Groups A, B, C, D, T4 Control Drawing: Refer to (5)
Ambient Temperature: -20 to 55°C China [Intrinsic safety “ia”] -CH
Ambient Humidity: (NEPSI) Applicable Standard:
0 – 100% (No Condensation) GB3836.1-2010, GB3836.4-2010,
Enclosure: IP66, NEMA 4X GB 3836.20-2010
Control Drawing: Refer to (2) Certificate No: GYJ18.1051X
Canada [Type of protection ‘n’ / Nonincendive] -DE Marking/Rating: Ex ia IIC T4 Ga
(CSA) Applicable Standard: Ambient Temperature: -20 to 55°C
C22.2 No.0-10 (R2015), CAN/ Control Drawing: Refer to (6)
CSA-C22.2 No.94-M91 (R2011), Korea [Intrinsic safety “ia”] -CG
C22.2 No.213-M1987 (R2013), (KOSHA) Applicable Standard:
CAN/CSA-C22.2 No.60079-0:11, Notice of Ministry of Labor No.
CAN/CSA-C22.2 No.60079- 2016-54
11:14, CAN/CSA-C22.2 Certificate No: 15-AV4BO-0160X
No.60079-15:12, CAN/ Marking/Rating: Ex ia IIC T4 Ga
CSA-C22.2 No.61010-1-12, CAN/ Ambient Temperature: -20 to 55°C
CSA-C22.2 No.61010-2-030-12 Control Drawing: Refer to (6)
Certificate No: 2425510
Marking/Rating: Ex nA ic IIC T4 Gc
Nonincendive for Class I, Division
2, Groups A, B, C, D, T4
Ambient Temperature: -20 to 55°C
Ambient Humidity:
0 – 100% (No Condensation)
Enclosure: IP66, NEMA 4X
Control Drawing: Refer to (2)
All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 12A01A03-01EN Mar. 23, 2018-00
8
Lo 1.7 mH 8 mH 0.45 mH
Notes:
1. The associated apparatus must be a linear source.
2. Measuring Module 2 is not necessarily installed. As for ISC module and SENCOM module, only
one module is permitted to be installed at a time.
3. Sensor 1 and Sensor 2 may be simple apparatus or intrinsically safe apparatus.
4. WARNING – POTENTIAL ELECTROSTATIC CHARGING HAZARD – SEE USER’S
MANUAL
■ Control Drawings
Rev.2: July. 25, 2016 Doc. No.: IKE039-A12 P.1-1
Rev.4: July. 25, 2016 Doc. No.: IKE039-A12 P.1
Yokogawa Electric Corporation
(1)
Yokogawa Electric Corporation
9
Associated Apparatus
Sensor 1 Measuring Module 1 Supply + +
(Note 2)
(Note 6) Control Equipment
Supply – – Sensor 1 Measuring Module 1 (Note 7, 8)
(Note 6)
Supply – –
Sensor 2 Measuring Module 2 (Note 5)
(Note 6) Sensor 2
(Note 6) Measuring Module 2 (Note 5)
Supply +, Supply – (Note 2):
Ui: 30 V
Ii: 100 mA Supply +, Supply – (Note 7):
Pi: 0.75 W Ui: 30 V
Ci: 13 nF Ci: 13 nF
Li: 0 mH Li: 0 mH
All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 12A01A03-01EN Mar. 23, 2018-00
16
6: When a 2nd input is selected, only the same kind of the 1st input is available.
For example, when a 1st input is “-P1”, the 2nd input must be the same “-P1”.
The combination of ISC and ISC is not available. And, the combination of SENCOM sensor and SENCOM sensor is not
available, either.
7: These languages are message languages on the analyzer’s display.
One analyzer has English and 11 languages.
All languages are as follows; English, Chinese, Czech, French, German, Italian, Japanese, Korean, Polish, Portuguese,
Russian and Spanish.
8: When an analyzer is used in Japan, it must meet the Japanese Measurement Law, please select the “-J”.
Only SI units must be used on the analyzer and its documents in Japan.
9: The universal mounting kit contains the pipe and wall mounting hardware (/U) and the panel mounting hardware (/PM).
10: “/CB5”, “/CD5”, “/CF5” are exclusively for type “-DB” or “-DE”. “/CB4”, “/CD4”, “/CF4” cannot be used with type “-DB” or “-DE”.
“/CB4”, “/CD4”, “/CF4” can be used with other types of the analyzer except for “-DB” or “-DE”.
11: The analyzer with Japanese Measurement Law certificate is available only for the following model;
FLXA202-D-[Housing code]-D-AB-P1-NN-A-N-LA-J-NN/[option code except /K]/K
Only one pH measurement with an analog sensor is certified. The digital sensor (FU20F) is not certified.
The output signal of 4 - 20 mA is certified. HART communication is not certified
12: Urethane coating is for acid resistance, and epoxy coating is for alkali resistance. For high anti-corrosion coating, both
urethane coating and epoxy coating are applied.
13: Type “-CF” is anti-reflection coated. Other types are anti-glare coated.
14: Product registration is done by Yokogawa Taiwan Corporation as an importer in Taiwan.
15: For detailed information refer to Japanese GS 12A01A03-01JA.
All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 12A01A03-01EN Dec. 13, 2018-01
17
n Dimensions and Mounting
146 9 Unit: mm
165 40 80
165
137
80
For sensor 1 cable 4-M6 depth 7
For sensor 2 cable
137 FB4_01.ai
Approx. Packing
55(2.2") 49
(1.93")
G1/2 screw (/CB4), 1/2 NPT screw (/CD4)
Adapter M20x1.5 screw (/CF4)
F0204.ai
Approx. Packing
64(2.52")
All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 12A01A03-01EN Mar. 23, 2018-00
18
(Note) The universal mounting kit (/UM) contains the pipe and wall mounting hardware (/U) and the panel mounting
hardware (/PM).
Unit: mm
Panel thickness
1 to 12
121
2-M5 length 35 +1
138 0
+1
138 0
4-M6 *
195
100
185
Unit: mm
165 168 4-M6 *
13
165
234.5
200
50
3-ø10 holes
70
100
*: Tighten the four screws to a torque of 2 N•m.
FB4-202_03.ai
Note: The wall on which the analyzer is mounted should be strong enough to bear the weight of more than 8 kg.
All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 12A01A03-01EN Mar. 23, 2018-00
19
Pipe mounting hardware (Option code: /U, /UM)
Unit: mm
165
(209)
155 Pipe 50A
(ø60.5)
4-M6 *
184.5
200
50
M8 U-bolt
Pipe mounting (Horizontal) Pipe mounting (Vertical)
100
*: Tighten the four screws to a torque of 2 N•m. FB4-202_04.ai
165
(50)
(70)
(100)
FB4-202_05.ai
All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 12A01A03-01EN Mar. 23, 2018-00
20
n Wiring Diagrams
WTB10 or BA10 Case of Distributor
Terminal box *6 FLXA202 2-Wire Analyzer PH201G (Style B)
– CMN CMN
n n n n n n n n n *4 + B D H
A C F
*1 L N
Sensor
+ a c e Power supply
*3 b d f
– 20 to 130V DC
or
80 to 138V AC, 47 to 63Hz
*2 HOLD WASH
WTB10 or BA10 FAIL
Terminal box *6 Case of Distributor
SDBT
– –
+ B + H
A F
L N
Power supply
+
1 – 20 to 130V DC
2 or
(*3) 80 to 138V AC, 47 to 63Hz
*1: Use a 2-wire shielded cable with an outside diameter of 6 to 12 mm.
*2: Connect the analyzer to ground. (Class D ground: 100 ohm or less)
*3: This line is connected to a distributor or 24V DC power supply.
*4: Terminal numbers for each sensor module are shown below.
*5: Two modules of the same kind of measurement/sensor type can be installed. When measuring inductive
conductivity or pH/ORP with the SENCOM sensor, only one module can be installed.
*6: The terminal box may be necessary depending on the sensor cable length and the distance between the
analyzer and the sensor.
The SENCOM sensor is to be connected directly to the analyzer without a terminal box.
*7: Two outputs, output 1 and output2, of PH201G or SDBT are same signals.
11 12 14 18 13 17 19 15 16
PH Module PH
11 12 13 14 15 16 NC
SC Module SC
11 12 13 17 14 16 15
ISC Module ISC
11 12 16 15 13 14 17 18 NC
DO Module DO
NC 82 83 84 NC 86 87
SENCOM Module SENCOM
All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 12A01A03-01EN Mar. 23, 2018-00
21
n Inquiry Specifications Sheet for FLXA202 2-Wire Analyzer
Make inquiries by placing checkmarks () in the pertinent boxes and filling in the blanks.
1. General Information
Company name
Contact Person; Department;
Plant name;
Measurement location;
Purpose of use; Indication, Recording, Alarm, Control
2. Measurement Conditions
(1) Process temperature; to Normally [°C]
(2) Process pressure; to Normally [kPa]
(3) Flow rate; to Normally [l/min]
(4) Flow speed; to Normally [m/s]
(5) Slurry or contaminants; No, Yes
(6) Name of process fluid;
(7) Components of process fluid;
(8) Others;
3. Installation Site
(1) Ambient temperature; to [°C]
(2) Location; Outdoors, Indoors
(3) Others;
4. Requirements
1st Input; pH/ORP (analog sensor) Conductivity (SC) Inductive conductivity (ISC)
Dissolved oxygen (DO) pH/ORP (digital sensor, FU20F)
2nd Input; With (same as 1st Input) Without
All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 12A01A03-01EN Mar. 23, 2018-00
22
4.2 Conductivity
1st Input
(1) Measuring range;
(2) Transmission output; 4 to 20 mA DC
-1 -1
(3) Detector/sensor; SC4AJ Two electrode system (0.02 cm ) Two electrode system (0.1 cm )
-1 -1
SC8SG Two electrode system (0.01 cm ) Two electrode system (10 cm ),
-1
Four electrode system (10 cm )
-1 -1
SC210G Two electrode system (0.05 cm ) Two electrode system (5 cm )
(4) Detector/sensor mounting method;
SC4AJ Adapter mounting, Welding socket, Welding clamp
SC8SG Screw-in, Flow-through
SC210G Screw-in, Flange, Flow-through, Screw-in with gate valve
(5) Electrode cable length; SC4AJ 3m, 5m, 10m, 20m
SC8SG 5.5m, 10m, 20m
SC210G 3m, 5m, 10m, 15m, 20m
(6) Others;
2nd Input
(1) Measuring range;
(2) Transmission output; 4 to 20 mA DC
-1 -1
(3) Detector/sensor; SC4AJ Two electrode system (0.02 cm ) Two electrode system (0.1 cm )
-1 -1
SC8SG Two electrode system (0.01 cm ) Two electrode system (10 cm ),
-1
Four electrode system (10 cm )
-1 -1
SC210G Two electrode system (0.05 cm ) Two electrode system (5 cm )
(4) Detector/sensor mounting method;
SC4AJ Adapter mounting, Welding socket, Welding clamp
SC8SG Screw-in, Flow-through
SC210G Screw-in, Flange, Flow-through, Screw-in with gate valve
(5) Electrode cable length; SC4AJ 3m, 5m, 10m, 20m
SC8SG 5.5m, 10m, 20m
SC210G 3m, 5m, 10m, 15m, 20m
(6) Others;
All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 12A01A03-01EN Mar. 23, 2018-00
23
4.4 Dissolved oxygen
1st Input
(1) Measuring range; 0 to 50 mg/L
(2) Transmission output; 4 to 20 mA DC
(3) System configuration selection; Electrode, Holder, Converter, Cleaning system,
Terminal box, Maintenance parts set, Calibration set
(4) Electrode cable length; 3m, 5m, 10m, 15m, 20m
(5) Type of holder; Guide pipe, Submersion, Flow-through, Suspension,
Angled floating ball, Vertical floating ball
(6) Cleaning method; No cleaning, Jet cleaning
(7) Others;
2nd Input
(1) Measuring range; 0 to 50 mg/L
(2) Transmission output; 4 to 20 mA DC
(3) System configuration selection; Electrode, Holder, Converter, Cleaning system,
Terminal box, Maintenance parts set, Calibration set
(4) Electrode cable length; 3m, 5m, 10m, 15m, 20m
(5) Type of holder; Guide pipe, Submersion, Flow-through, Suspension,
Angled floating ball, Vertical floating ball
(6) Cleaning method; No cleaning, Jet cleaning
(7) Others;
All Rights Reserved. Copyright © 2015, Yokogawa Electric Corporation GS 12A01A03-01EN Mar. 23, 2018-00
Subject to change without notice.