Brachytherapy Applicator Guide
Brachytherapy Applicator Guide
The new ‘Safety through simplicity’ Flexitron family is designed for safety and better
team efficiency. With more logic and fewer variables the Flexitron afterloading
platform optimizes the treatment delivery process and offers you a safe investment
for today and in the future.
With the Oncentra Brachy treatment planning system, the creation of highly
conformal plans in an efficient workflow is within reach. The latest version of Oncentra
Brachy offers a series of useful tools that facilitate many repetitive tasks for you, such
as contouring and reconstruction. Oncentra Brachy buys you time to focus on what
counts: designing the optimal dose distribution.
When Oncentra Brachy treatment planning is used in combination with the new
Flexitron afterloading platform, you will optimally benefit from a number of smart and
standardized features, that allow an easy, standardized way of treatment planning –
while feeling confident about accurate and safe treatment delivery.
2 3
How to use this applicator guide Safety in the magnetic resonance environment
In this applicator guide, brachytherapy applicators are The symbols shown below are used to indicate safety of
presented by body site. When a particular applicator medical devices in the Magnetic Resonance environment.
can also be used for other body sites, a cross reference
is given on the index-page at the start of the relevant MR safe - an item that poses no known hazards
applicator section. in all MR environments (1.5T, 3T, 6T and others).
single use, delivered non-sterile • ot all products in this guide are available in
N Skin & surface 86
all markets. Please contact your local Elekta
multiple use, delivered non-sterile representative for further information. NEEDLES & FLEXIBLE IMPLANT TUBES
• Applicators as mentioned in this guide can 94
Sterilization advice be used with both Iridium-192 and Cobalt-60
Sterilization advice provided in this document is afterloaders, except for the Valencia and Leipzig QUALITY ASSURANCE TOOLS
applicators.
indicative and applies to newly purchased applicators. 104
For more information regarding sterilization of a • I f no information is provided under ‘part number’,
particular applicator, please consult the User Manual this item cannot be ordered separately. ACCESSORIES
or Sterilization Advice Chart. • The latest version of the applicator guide 110
is leading and can be downloaded from
www.elekta.com/brachytherapy/applicators INDEX
Find your information under the relevant bodysite.
116
• References to literature are not included
in this guide. APPENDIXES
Table for Flexitron X-ray, CT and MRI markers 118
Table for Flexitron transfer tubes 120
Table for microSelectron X-ray,
CT and MRI markers 122
Table for microSelectron transfer tubes 124
4 5
Gynecology
CERVIX ENDOMETRIUM
Advanced gynecological applicator – VeneziaTM 8 Rotte endometrial applicator set 48
Utrecht interstitial CT/MR applicator set 10
ALTERNATIVES FOR VAGINAL BRACHYTHERAPY
Fletcher CT/MR shielded applicator set 12
Fletcher CT/MR applicator set 14 Prostate stepper template set
Prostate, rectum & bladder 56
Standard CT/MR applicator set 16
MUPIT applicator set
Fletcher Williamson applicator set 18 Prostate, rectum & bladder 58
Cervix Rotterdam applicator set 20 Intracavitary mold applicator set
Henschke titanium applicator set 22 Prostate, rectum & bladder 60
VAGINA
Vaginal CT/MR multi channel applicator set 34
Vaginal CT/MR applicator set 36
Vaginal cylinder applicator set titanium 38
Shielded cylindrical applicator set 40
6 7
GYNECOLOGY
126.006 Interstitial lunar ovoid tube 30 mm 60° left • Ensure reproducible dosimetric
126.007 Vaginal cap 22 mm right results
126.008 Vaginal cap 22 mm left
various patient groups. The two lunar-shaped ovoids This guided needle placement ensures reproducible dosimetric 126.023 Intrauterine tube 70 mm 15°
form a ring when clicked together, allowing you to benefit results, making needle placement easier and more accurate. 126.024 Fixation element
from the advantages of a tandem and ovoid applicator in 126.025 Fixation clip “With Venezia, I expect
combination with the dose distribution of a ring - high dose Venezia is much easier to insert and assemble than
to treat advanced cervical
126.026 Perineal template dorsal 0°
to the target volume while sparing organs at risk. conventional gynecological applicators. Simply insert the 126.027 Perineal template dorsal 12°
first half of the applicator and attach the second half by
sliding and clicking the two parts together. The applicator
126.028 Perineal template ventral 0° cancer patients with more
Reproducible dosimetric results 126.029 Bracelet for guiding tubes (5 x 1)
The numbered hole positions on the Venezia ovoids and
contains no screws and requires no tools, making applicator
126.030 Number tag for guiding tubes (5 x 1)
ease. Venezia will allow me
assembly and insertion an easy, do-it-yourself task.
template facilitate consistent needle placement for every
fraction – every time. Additionally, the insertion tool helps
126.031 Guiding tubes for interstitial lunar ovoid (5 x 1) to reach tumors extended
Easily integrate into the workflow 126.032 Guiding tubes for vaginal cap (5 x 1)
to insert the needles at the exact prescribed depth.
Venezia is easy to insert and assemble and comes in various 126.033 Perineal bar to the parametrium
Venezia is included in the Oncentra® Brachy Applicator sizes to accommodate every patient’s pelvic anatomy and 101.012 Sterilizing caps CT/MR
because I can position
Modeling module, making reconstruction easy and fast. as such increases the patient’s comfort. And, depending on 110.268 ProGuide obturator bulk set, 6F x 294 mm (1 x 5)
Simply select Venezia, then import and modify to meet the patient’s indication, the physician can select vaginal and/ 189.714 Insertion tool needles at multiple angles,
clinical needs. Manual reconstruction becomes obsolete, or perineal needle insertion. The multi-purpose applicator
saving time and streamlining workflow. eliminates the need for a separate device for every indication.
189.040
189.043
Cleaning brush ø 2.5 mm x 250 mm (set of 3)
Venezia will always be clearly visible in the image dataset. 101.015 CT/MR sterilization box
Petch Alisanant, MD
Utrecht Interstitial CT/MR Applicator Set 189.043 Cleaning brush set (6.5 mm)
Accessories
A compelling workflow for image-guided adaptive Intended for:
Part number Description
brachytherapy (IGABT)
• Cervix 110.230* ProGuide round needle set, 6F x 294 mm
The Utrecht interstitial applicator is a state-of-the-art • Endometrium 110.231 ProGuide sharp needle set, 6F x 294 mm
treatment solution for combined intracavitary and 110.232* Guiding tube for ovoids
interstitial gynecological brachytherapy. Note: 189.769 Intrauterine tube, d =6 mm, 45º
10 11
GYNECOLOGY
110.957
Adjustable fixation mechanism
Fixation clip
your plans can be calculated with ACE (TG-186) 189.769 Intrauterine tube (6 mm), 45º
The Fletcher CT/MR shielded applicator is the first of in Oncentra Brachy, for both Iridium-192 and 189.566* Smit Sleeve CT/MR 40 mm (10 pcs)
its kind to overcome these issues. Moveable shielding Cobalt-60 sources. Oncentra also provides a shield 189.676*
189.676 and following Smit Sleeves, l = 20 mm
and longer, d = 6 mm (10 pcs)
incorporated in the ovoids allows you to lower the dose approximation that can be used for TG-43 calculations
101.031 Fixation mechanism for ovoid tubes
to organs at risk during gynecological brachytherapy for Iridium-192.
101.015 CT/MR sterilization box
treatments. High-quality CT and MR images for 3D dose
189.931 CT/MR applicator clamp with axle
planning can now be combined with protection of the Order information
189.932 CT/MR applicator clamp with base plate
organs at risk. Fletcher CT/MR shielded applicator set (4 mm)
Part # 110.980 110.700 Pins for adjustable fixation mechanism (5 pcs)
12 13
GYNECOLOGY
Order information
Why the Fletcher CT/MR
Fletcher CT/MR applicator set, part # 189.730
applicator set?
Part number Description
189.750
Adjustable fixing mechanism
Fixation clip
the anatomy of the cervix and surrounding areas. • The separate Utrecht Interstitial applicator 189.735 Ovoid tube 1 110.296/.298/.300 Intrauterine tube (4 mm) 15°/30°/45°
complementary set (110.271) contains all components 189.740 Ovoid tube 2 189.566** Smit Sleeves CT/MR ø 9 mm (box of 10)
The applicator uses strong composite fiber tubing and to add interstitial functionality to a Fletcher CT/MR 110.296* Intrauterine tube (4 mm), 15º 189.676 and following Smit Sleeves, l = 20
189.676**
plastics to eliminate distortion on CT and MR imaging. applicator and thereby upgrade it to the state-of-the- mm and longer, d = 6 mm (10 pcs)
110.298* Intrauterine tube (4 mm), 30º
With the use of the Applicator Modeling module in art Utrecht Interstitial CT/MR applicator 189.769 Intrauterine tube 45° (6 mm)
110.300* Intrauterine tube (4 mm), 45º
Oncentra Brachy, the applicator is reconstructed with 101.031 Ovoid tube fixing mechanism
• Asymmetrical ovoid placement is possible 110.540 Ovoid pair 15 mm, R7.5 (micro ovoids)
just a few mouse clicks.
189.783 Ovoid pair 15 mm, R10 (half ovoids) 101.015 CT/MR sterilization box
• The applicator can be sterilized using steam
The flexible geometry of the applicator provides the 189.754 Ovoid pair 20 mm, R10
sterilization 189.900
Ovoid set with radius 12.5 mm
189.755 Ovoid pair 25 mm, R10 (25, 30, 35, 40 mm)
options to optimize the position of the ovoids to the cervix,
• User manual 777.00287 Ovoid set with radius 15 mm
and to adjust the intrauterine tube to ensure an optimal fit 189.722 Adjustable fixation mechanism 189.901
(30, 35, 40 mm)
with the patient’s anatomy. 189.750 Fixation clip
110.700 Pins for adjustable mechanism (5 pcs)
101.012 Sterilization caps (50 pcs.)
189.930 Spare fixation nuts (5 pcs)
Unique solution for narrow anatomy of vagina 110.283 Screwdriver titanium for ovoids and stopper
Utrecht interstitial applicator complimentary
008.033 Plastic screw M5 PPSU (10 x 008.033) 110.271 set. To expand the use of the Fletcher CT/MR
An extra small version of this applicator is available, the applicator to interstitial treatments
Fletcher CT/MR applicator set XS. This version has small Spare part set plastic screw M5 PPSU
189.894
ovoids that optimally fit women with a narrow vagina, e.g (10 pcs)
patients from the Asian region. The thin intrauterine tubes 110.315
Rectal retractor 30 mm with slot
(for intrauterine tubes 4 mm)
in this set facilitate easy insertion. 4 mm IU tube with new stopper
Rectal retractor 40 mm with slot
110.335
(for intrauterine tubes 4 mm)
189.741 Fixation nut for rectal retractor
14 15
GYNECOLOGY
16 17
GYNECOLOGY
Fletcher Williamson Asia Pacific set, part # 085.260 Why the Fletcher
Part number Description
Williamson applicator?
085.233 Ovoid tube 1
006.182 Screwdriver
MR Unsafe
Accessories
Part number Description
Fletcher Williamson Applicator Set 084.043 Ovoid pair, 25 mm, medium, unshielded
085.237
Ovoid cap pair, 25 mm, medium
J. Stitt, MD
University of Wisconsin, Madison
United States
18 19
GYNECOLOGY
Cervix Rotterdam applicator set IU-30°, part # 137.206 Why the Cervix Rotterdam
Part number Description applicator?
137.248 Right ovoid tube I, IU-30° (ti)
137.249 Left ovoid tube III, IU-30° (ti) • Dependable insertion through
137.191 Ovoid pair, Ø 20 mm, small fixed geometry
137.192 Ovoid pair, 25 mm, medium
137.256 Ovoid tube coupling for 30°, for ovoids only (ti) • Easy extension for radiation of
137.570 Closing caps (25/pack) vaginal wall with the optional
137.574 Screwdriver for ovoid screws vaginal cylinder
Accessories for 30° set
With the optional vaginal cylinder, the treatment
MR conditional can be easily extended to include the vaginal wall Part number Description
in the same treatment fraction.
137.245 Vaginal cylinder locking pin (ti)
20 21
GYNECOLOGY
Order information
Why the Henschke
Henschke titanium applicator set, part # 110.018
titanium applicator?
Part number Description
Accessories
MR Unsafe Part number Description
22 23
GYNECOLOGY
Interstitial Ring CT/MR Applicator Set Set, d = 26 mm, 6 mm, 60º (max. 7 needles),
part # 110.330 A01
101.051 Intrauterine tube, l = 20 mm, 60º (6 mm)
interstitial gynecological brachytherapy. Note: 101.053 Intrauterine tube, l = 60 mm, 60º (6 mm) 110.268 Obturators, set of 5
101.057 Rectal retractor, 60º 189.714 Insertion tool
• Can also be used as regular intracavitary
Targeted dose optimization ring applicator 189.930 Fixation nuts set (spare, 5 pcs) 189.040 Cleaning brush set (2.5 mm)
101.012 Sterilization caps CT/MR, (50 pcs) 189.043 Cleaning brush set (6.5 mm)
Besides the regular intracavitary treatment channels • Use of needles with rounded tip can reduce the risk
110.268 Obturators, (5 pcs)
of the applicator, the ring acts as a template for placing of perforation of critical structures, and is required for
189.714 Insertion tool Accessories for all sets
additional needles in the cervical area. The interstitial use with interstitial cervix applicators
189.040 Cleaning brush set
needles provide additional control of dose coverage to the Part number Description
• The applicator can be sterilized using 189.043 Cleaning brush set
tumor, while sparing the surrounding tissue. The insertion ProGuide round needle set, 6F x 294 mm
steam sterilization 110.230*
(set of 5, single packed, delivered sterile)
depth of the needles can be easily controlled using the Set, d = 30 mm, 6 mm, 60º (max. 9 needles),
• User manual 777.00083 part # 110.331 A01 110.231*
ProGuide sharp needle set, 6F x 294 mm
insertion tool. The needles are individually locked in place (set of 5, single packed, delivered sterile)
for treatment. Order information Part number Description
Guiding tube for ring set for interstitial ring
110.273*
(set of 5, single packed, delivered sterile)
Interstitial Ring CT/MR applicator sets (4 mm)* 110.351 Interstitial ring tube, d = 30 mm, 60º
Optimized for 3D imaging and planning 101.051 Intrauterine tube, l = 20 mm, 60º (6 mm)
Part number Description 110.350 Interstitial ring tube, d = 26 mm, 60º
The applicator uses strong composite fiber tubing 101.052 Intrauterine tube, l = 40 mm, 60º (6 mm) 110.351 Interstitial ring tube, d = 30 mm, 60º
110.674 Interstitial Ring applicator set, complete (4 mm)
and plastics to eliminate distortion on CT and MR 101.053 Intrauterine tube, l = 60 mm, 60º (6 mm) 110.352 Interstitial ring tube, d = 34 mm, 60º
Interstitial Ring applicator set d = 26 mm,
110.330A02
imaging. With the use of transversal slices, and the 60˚(4 mm) 101.057 Rectal retractor, 60º Smit Sleeves CT/MR for 6 mm intrauterine
modules Applicator Modeling and Library Plans in Interstitial Ring applicator set d = 30 mm, 189.930 Fixation nuts set (spare), set of 5 189.566** tube, l = 40 mm, d =9 mm, open ended
110.331A02 (box of 10)
60˚ (4 mm)
Oncentra Brachy, the applicator is quickly reconstructed 101.012 Sterilization caps CT/MR, pack of 50
Interstitial Ring applicator set d = 34 mm, Smit Sleeves CT/MR for 4 mm intrauterine
189.676
during treatment planning. 110.332A02
60˚ (4 mm) 110.268 Obturators, set of 5 tube, l = 20 mm and longer, d = 6 mm
and following** (box of 10)
189.714 Insertion tool
101.054 Perineal bar
189.040 Cleaning brush set
Intrauterine tube, d = 4 mm, 60°,
189.043 Cleaning brush set 110.305/306/307
l =20/40/60 mm
24 25
GYNECOLOGY
189.799
Description
The Vienna Ring CT/MR applicator is based on the CT/MR Intended for: 101.051 Intrauterine tube, l = 20 mm, 60°
Ring applicator, enhanced with the addition of nine guide • Cervix 101.052 Intrauterine tube, l = 40 mm, 60°
holes in the ring tube. These guide holes allow placement 101.053 Intrauterine tube, l = 60 mm, 60°
• Endometrium
of interstitial titanium needles using the ring tube as a 101.057 Rectal retractor, 60°
needle template, while maintaining the treatment channel Note: 189.930 Set of 5 fixing nuts (spare)
of the ring tube. The addition of interstitial needles makes 101.012 Sterilization caps, CT/MR
• Can also be used as a standard intracavitary
it possible to achieve asymmetric alteration of the dose ring applicator 110.721 Needle fixation tool, d = 30 mm
distribution. The needles are inserted perpendicular to the 189.040 Cleaning brush set (2.5 mm)
• Use of needles with rounded tip can reduce the risk of
ring and in parallel to the tandem. The applicator’s plastic 189.043 Cleaning brush set (6.5 mm)
perforation of critical structures, and is recommended
and composite fiber materials minimize distortion on CT
for use with interstitial cervix applicators
or MR imaging. The Round Point needle is designed to Vienna Ring CT/MR applicator set 34 mm, 60°,
minimize tissue damage and for ease in implantation. This • Non-clinical testing has demonstrated that the part # 189.569 Accessories
applicator, developed by renowned clinical experts, offers titanium needles and the needle fixation tool are
Part number Description Part number Description
MR Conditional. These products can be scanned
the solution for treatment of advanced tumor stages with 189.567 Vienna Ring CT/MR tube, 60° (for 9 needles) 110.720 Needle fixation tool, d = 26 mm
safely under the following conditions: static magnetic
an optimised dose, minimizing dose to organs at risk. 101.051 Intrauterine tube, l =20 mm, 60° 110.721 Needle fixation tool, d = 30 mm
field up to 0.23 Tesla
101.052 Intrauterine tube, l = 40 mm, 60° 189.995 Needle fixation tool, d = 34 mm
• The applicator can be sterilized using
101.053 Intrauterine tube, l = 60 mm, 60° Round point titanium needle set, 1.9 mm x
steam sterilization 110.025
240 mm, 60°, 30 mm (set of 5 needles)
101.057 Rectal retractor, 60°
• User manual 090.720 Round point titanium needle set, 1.9 mm x
189.930 Set of 5 fixing nuts (spare) 110.027
240 mm, 60°, 40 mm (set of 5 needles)
101.012 Sterilization caps, CT/MR
Round point titanium needle set, 1.9 mm x
110.029
189.995 Needle fixation tool, d =34 mm 240 mm, 60°, 50 mm (set of 5 needles)
189.040 Cleaning brush set (2.5 mm) 083.252 Obiturator for needle, l = 100 mm
189.043 Cleaning brush set (6.5 mm) 089.636 Handle for obturator
26 27
GYNECOLOGY
Ring CT/MR applicator set, 6 mm, 45º, part # 101.061 Why the Ring CT/MR
Part number Description
applicator?
101.042 Ring tube, d = 26 mm, 45°
The Ring CT/MR applicator uses the ring to create a Part number Description 189.566* Smit Sleeve CT/MR (l = 40 mm, d = 9 mm)
pear-shaped dose to match the anatomy of the cervix 110.669 Ring CT/MR applicator set complete, 4 mm 101.054 Perineal bar (for part # 101.060/.061/.062)
and endometrium. 110.670 Ring CT/MR applicator set, 30°, 4 mm 101.038 Ring tube, d = 34 mm, 30° (for part # 101.060)
110.671 Ring CT/MR applicator set, 45°, 4 mm 101.044 Ring tube, d = 34 mm, 45° (for part # 101.061)
The applicator uses strong fiber composite tubing and
110.672 Ring CT/MR applicator set, 60°, 4 mm 101.050 Ring tube, d = 34 mm, 60° (for part # 101.062)
can be used in combination with CT and MR imaging for
110.303/.304 / Intrauterine tube, d =4 mm,
volume-based treatment planning. The fixed geometry of .305 /.306 /.307 l = 40 30°/45°/60°
the ring and the intrauterine tube provide a reproducible Ring CT/MR applicator set, 6 mm, part # 101.035
101.015 CT/MR sterilization box
applicator geometry. With the use of the Applicator Part number Description
110.290 MR line marker set
Modeling module in Oncentra Brachy, reconstruction of 101.060 Ring applicator set, 30°, CT/MR
the applicator takes only a few mouse clicks. 101.061 Ring applicator set, 45°, CT/MR
28 29
GYNECOLOGY
Ring applicator set 45°, part # 085.096 Why the Ring applicator?
Part number Description
085.097 Ring tube, 60° (for set # 085.097) Part number Description
Intended for: 084.051 Ring cap, d = 34 mm
084.072 Perineal bar
• Cervix 084.056 Ring tube, d = 34 mm, 30° 084.056 Ring tube, d = 34 mm, 30° (requires part # 085.095)
084.055 Ring tube, d = 34 mm, 45° 084.055 Ring tube, d = 34 mm, 45° (requires part # 085.096)
• Endometrium
084.054 Ring tube, d = 34 mm, 60° 084.054 Ring tube, d = 34 mm, 60° (requires part # 085.097)
Note: 084.051 Ring cap, d = 34 mm 084.072 Perineal bar (already included in part # 084.050)
• The applicator can be sterilized using steam 084.078 Intrauterine tube, l = 20 mm, 90°
Ring applicator set 30°, part # 085.095
sterilization and ethylene oxide 084.077 Intrauterine tube, l = 40 mm, 90°
Part number Description
084.076 Intrauterine tube, l = 60 mm, 90°
• User manual 090.610
084.062 Ring tube, d = 26 mm, 30°
084.338 Rectal retractor, 90°
084.059 Ring tube, d = 30 mm, 30°
101.015 CT/MR sterilization box
084.065 Intrauterine tube, l = 20 mm, 30°
189.676 Smit Sleeves CT/MR for 4 mm intrauterine tube,
084.064 Intrauterine tube, l = 40 mm, 30° and following* l = 20 mm and longer, d =6 mm (box of 10)
30 31
GYNECOLOGY
Order information
Adjustable cervix applicator set, part # 084.450
Why the Adjustable cervix
applicator set?
Part number Description
MR Unsafe
32 33
GYNECOLOGY
Accessories
Applicator catheters following the
MR safe Part number Description
curve of the tip to reach close to
the vaginal vault
Catheter set with collar 6F x 293 mm
110.800*
(delivered sterile and for single use only)
unique design incorporating channels which follow the 101.069 IU3 x-ray marker for central tandem allows me to get the
Intended for:
curve of the applicator tip. This brings the dosimetry
exact target coverage
measurably closer to the vaginal vault, allowing you to • Vagina
reliably and more effectively cover the planning target
• Cervix
I planned and gives me
volume.
• Endometrium confidence in conformity
Unique treatment precision
Note: in the treatment of my
The channels, located immediately underneath the • The center of the surface channels is located 5 mm gynecologic cancer
surface of the applicator, provide high accuracy and underneath the cylinder surface
optimized treatment precision, while tailoring the patients.”
• The applicator can be sterilized using steam
dose to the target and saving the organs at risk around
sterilization
e.g. bowel and bladder.
• User manual for applicator 777.00171 B.R. Pieters, MD, PHD
Head of Brachytherapy Department
• User manual for catheters 777.00172 Academic Medical Centre (AMC), Amsterdam
The Netherlands
34 35
GYNECOLOGY
MR safe Accessories
Part number Description
36 37
GYNECOLOGY
Accessories
38 39
GYNECOLOGY
Accessories
Part number Description
40 41
GYNECOLOGY
006.182 Screwdriver
42 43
GYNECOLOGY
Accessories
MR Unsafe Part number Description
084.009
Fixing nut
Cervical stopper
The Vaginal applicator set for intracavitary treatment is Intended for: 084.355 4 cylinder parts, d = 40 mm with marker
very user-friendly. The set is available with three different • Vagina 084.875 4 cylinder parts, d = 15 mm with marker
lengths and curvatures of the intrauterine tube to provide 101.149 Knob PSU
• Cervix
additional treatment options. The cylinders are available in 189.676 and Smit Sleeves CT/MR for 4 mm intrauterine tube,
• Endometrium following* l = 20 mm and longer, d = 6 mm (box of 10)
different diameters to optimize the dose distribution and
101.015 CT/MR sterilization box
reduce the mucosal dose. • Rectum
44 45
GYNECOLOGY
MR Unsafe
46 47
GYNECOLOGY
MR Unsafe
48 49
Prostate, Rectum
& Bladder
Content
BLADDER
Luneray 52
PROSTATE
Martinez prostate template set 54
Prostate stepper template set 56
RECTUM
MUPIT applicator set 58
Intracavitary mold applicator set 60
ALTERNATIVES FOR
RECTAL BRACHYTHERAPY
Prostate stepper template set
Prostate, Rectum & Bladder 56
50 51
PROSTATE, RECTUM & BLADDER
types of bladder cancer, with good clinical outcome. A key Luneray™ catheter, 6F, 50 cm, needle l = 47 mm,
Interstitial brachytherapy of soft tissue that allows the 110.565
radius 45 mm (5 pcs)
advantage of brachytherapy is the potential to spare the
insertion of this flexible applicator, e.g.:
bladder, which improves the quality of life for the patient. A 110.566
Luneray™ catheter, 6F, 50 cm, needle l = 57 mm,
radius 60 mm (5 pcs)
hurdle for the adoption of bladder brachytherapy, however,
• Abdominal structures / organs (e.g. urinary bladder) Luneray™ catheter, 6F, 50 cm, needle l = 67 mm,
was the invasiveness and complexity of the procedure. 110.567
radius 75 mm (5 pcs)
• Head and neck (e.g. ear, lip, nose, tongue, oral mucosa)
Luneray overcomes this barrier by offering a unique
• Skin (e.g. keloids) Accessories
system of an integrated catheter and needle, designed
Part number Description
to enable laparoscopic, minimally invasive bladder • Soft tissue sarcoma
189.367 Cutter for flexibles (1 pcs)
brachytherapy procedures.
Note: 189.282 Radio-opaque button white, 6F (72 pcs)
Unique catheter/needle system • Each package holds five, identical, pre-assembled 189.284 Radio-opaque button black, 6F (72 pcs) “The Luneray catheter
needle/catheter products, single packed in a double
creates new opportunities
189.286 Radio-opaque button red, 6F (72 pcs)
Luneray combines a flexible catheter with a curved needle
pouch 189.288 Radio-opaque button blue, 6F (72 pcs)
which enables a minimally invasive catheter placement.
• Radio-opaque buttons to fixate the catheter after 189.290 Radio-opaque button green, 6F (72 pcs) to treat select bladder
insertion. Available in 7 different colors 189.292 Radio-opaque button yellow, 6F (72 pcs)
Precise catheter placement
189.294 Radio-opaque button brown, 6F (72 pcs)
cancers in a minimally
Luneray’s state-of-the-art curved needle design allows for • Imaging for treatment planning and verification
optimal maneuverability and precise catheter placement by way of X-ray, CT and MR invasive way.”
for accurately targeted bladder brachytherapy procedures.
• Centimeter marks: every 5 cm (line and number)
The catheter is made of exclusive material and compiled
with fill-wire, offering unsurpassed kink-resistance. • Detailed marking in the center of the catheter to G. Smits, MD, PhD
assist accurate catheter placement: 1 cm increments Urologist
Enables organ-sparing, laparoscopic procedures between 20-22 cm and 28-30 cm Rijnstate hospital, Arnhem, The Netherlands
Luneray is designed to enable minimally invasive access to • Manual number 777.00293
the bladder. Minimally invasive procedures are associated E. van der Steen – Banasik, MD
with quick recovery and fewer complications, supporting
Radiation oncologist
a patient friendly treatment.
Arnhem radiotherapy institute, The Netherlands
* The Luneray catheter is MR safe when the needle is removed
as described in the user manual.
• Luneray: MR unsafe
• Luneray-needle: MR unsafe
• Luneray-catheter: MR safe
52 53
PROSTATE, RECTUM & BLADDER
137.285 Template fixation pin set • Individual needle locks for various
needle sizes
Accessories
Part number Description • Quick implant removal
137.286 Lock inserts set for 5F plastic needles
*Non-clinical testing demonstrated that the Martinez prostate template is MR Conditional. A patient can be scanned safely,
immediately after placement under the following conditions:
• Static magnetic field of 1.5 Tesla or 3 Tesla
• Maximum spatial gradient magnetic field of 720 gauss/cm
• M
aximum MR system reported, whole body averaged specific absorption rate (SAR) of 2 W/kg for 15 minutes of scanning (i.e. per pulse sequence)
54 55
PROSTATE, RECTUM & BLADDER
Fixation plates
Part number Description
for applying interstitial techniques in HDR prostate uniform prostate brachytherapy treatment plans, users OncoSmart ProGuide sharp needle set,
treatments. It is placed in a stepper holder, which is can be sure that clinical objectives are met in the most
2 189.602
6F x 200 mm clinical situations, Prostate
available for various ultrasound stepper devices. efficient and dedicated way. Prostate stepper template
is best used in combination with ProGuide needles and
Prostate stepper template set, 6F part # 189.067
stepper template gives you
Quantity Part number Description
Convenient design supporting a variety of needles OncoSelect stepper.
1 189.158 Base plate stepper 6F
exactly what you need.”
The prostate stepper template is designed for guidance 1 189.161 Front plate stepper 6F
and fixation of OncoSmart ProGuide plastic needles in Intended for:
1 189.166 Fixation plate support assembly Alberto Bossi, MD
sizes 5F and 6F. This square light-weight template has • Prostate
1 189.165 Disposable fixation plates 6F (10 pcs)
Gustave Roussy
a universal 5 mm grid array and a single-action locking • Vagina 1 189.167 Eccentric axle
Genito Urinary Oncology
system that locks all implanted needles in a single
2 189.170 Assembly bolt template
• Perineum Head of Prostate Brachytherapy Unit
movement. The single-action locking system accelerates
1 189.205 Assembly screwdriver
the procedure and is designed to eliminate inadvertent • Rectum Villejuif Cedex - France
1 189.173 Fixation screwdriver
movement of individual needles. Optional grid face plates
1 189.041 Cleaning brush set ø3 x 100 mm
Note:
for the template correspond with the grid of the ultrasound
1 189.043 Cleaning brush set ø6.5 x 100 mm
system used, such as GE/ Siemens and B&K ultrasound • Holder and grid face plates need to be
systems. Grid face plates help determine the correct ordered separately Fixation plates Grid Face plates
needle position with use of the ultrasound image, and • User manual 090.699 Part number Description Part number Description
can be mounted on the template itself or on the 189.165 Set of 10 disposable fixation plates, 6F (5x2) 189.073 Grid face plate set, GE/Siemens
58 59
PROSTATE, RECTUM & BLADDER
089.786 Screwdriver
Accessories
MR Unsafe
Part number Description
189.538
Central catheter set (box of 5)
60 61
Breast
Content
BREAST TEMPLATES
Breast CT/MR template set 64
Kuske breast applicator set 66
Rowland adjustable breast template (RABIT) 68
ONCOSMART CATHETERS
CT/MR OncoSmart Catheter system 70
62 63
BREAST
Accessories
MR safe
Part number Set components
64 65
BREAST
1 189.012 Kuske breast template set • Designed for optimal fit with
6 189.307
Leader insertion needle, 1.5 x 200 mm patient anatomy
(box of 36)
The Kuske breast applicator set was developed in close Note: 1 189.015 Kuske breast template holder volume breast implants.
cooperation with Dr. Robert Kuske. The adjustable • The Kuske breast template set can be sterilized using
1 189.026 Left template clamp
The adjustable template
1 189.027 Right template clamp
template with integrated guiding plate ensures exact ethylene oxide
placement of needles and OncoSmart catheters or flexible 1 189.046 Implant template has an integrated guiding
• User manual 090.706
implant tubes. The single large template pair with special
• Sterilization Advice Chart 094.446
1
1
189.047
189.048
Target template
Guiding template
template which ensures
hole geometry ensures optimal coverage of the complete
target volume. Markers are integrated in the template to 2* 189.222 Implant template fixation nut an exact placement of
needles and flexible
aid the imaging procedures and provide the opportunity to 2 189.223 Bridge fixation nut
R. Kuske MD
Arizona Oncology Services - Scottsdale, AZ,
United states
66 67
BREAST
083.040 or
Interstitial needle set
083.900
MR Unsafe 1 189.400 or
Flexible implant tube set
(Either) 189.401
189.007 or
Optional accessories
The Rowland adjustable breast template (RABIT) holder
Quantity Part number Description
was developed to ensure parallel positioning of the
1 045.022 Double plane, 12 mm template pairs (5)
template pairs. The breast template set provides a dose
1 045.023 Double plane, 14 mm template pairs (5)
distribution according to the Paris dosimetry system.
1 045.025 Double plane, 18 mm template pairs (5)
It includes subset of single, double and triple plane
1 045.037 Double plane, 20 mm template pairs (5)
template pairs with different basal distances. Hence all
1 045.027 Triple plane, 12 mm template pairs (3)
target areas can be covered with this template.
1 045.028 Triple plane, 14 mm template pairs (3)
Note:
• The template set can be sterilized using
steam sterilization
68 69
BREAST
CT/MR OncoSmart Catheter System OncoSmart flange molding system, part # 189.016
Quantity Part number Description
For outpatient breast brachytherapy Note:
1 189.320 Flange molding tool
The CT/MR OncoSmart catheter system is an elegant • The catheter set contents are for single use only, 2 417.018 Li-ion battery Pack
treatment solution for breast brachytherapy. The system and delivered sterile, with the exception of the
1 417.006 Battery charge station
permits breast brachytherapy treatments on an outpatient 4F ProGuide
1 189.316 Flange molding cutter
basis, enabling the patient to resume daily life between • Insertion needles and the stop tags are delivered 1 189.367 Cutter flexible implants
fractions. It is ideally suited for treatment regimens such unsterile 1 189.369 Spare part set flange molding tool
as Accelerated Partial Breast Irradiation (APBI ) or the
• Set contents are packed per 6 items, OncoSmart Upgrade set, part #103.110 |
delivery of a brachytherapy boost dose as part of Whole except for # 189.802 for microSelectron
Breast Irradiation (WBI).
• User manual 090.709 Part number Description
In addition to breast brachytherapy, the CT/MR OncoSmart CT/MR OncoSmart catheter set
Order information 189.410 (contains 189.418 (= sterile, single use)
catheter system also can be used for brachytherapy in head
and 189.016 (non-sterile, re-usable)
& neck, tongue and soft tissue cases. CT/MR OncoSmart catheter system
189.036 CT-marker set 240 mm black
Part number Description
030.100 Push fit disconnection tool
Optimizing clinical workflow CT/MR OncoSmart catheter system, l =240 mm
(contains 30 flexible implant tubes (5 x 6 tubes), 4F Transfer tube set for microSelectron for 4F flexibles
The CT/MR OncoSmart catheters are delivered sterile and 189.416 (see transfer tube appendixes for Flexitron and
ProGuide needles (240 mm), stop tags, buttons, and
insertion needles (200 mm)) microSelectron)
are ready for immediate use. The treatment catheters are
inserted only once, at the start of the actual treatment, CT/MR OncoSmart catheter system, l =240 mm OncoSmart Upgrade set for Flexitron, aggregate
(contains 18 flexible implant tubes (3 x 6 tubes), 4F
189.418
and remain in place in between fractions. Small buttons ProGuide needles (240 mm), stop tags, buttons, and Part number Description
insertion needles (200 mm))
with corresponding channel numbers fix the catheters in
CT/MR OncoSmart catheter set
CT/MR OncoSmart catheter system, l =240 mm,
place and are used for accurate and quick reconnection single package (contains 6 flexible implant tubes, 4F
189.410 (contains 189.418 (= sterile, single use)
189.802 and 189.016 (non-sterile, re-usable)
for sequential fractions. Tailored material characteristics ProGuide needles (240 mm), stop tags, buttons, and
insertion needles (200 mm))
189.036 CT-marker set 240 mm black
make OncoSmart catheter sets optimally visible on either
CT/MR OncoSmart catheter system, l =200 mm, 030.100 Push fit disconnection tool
CT or MR images. for Head & Neck (contains 6 flexible implant tubes,
189.801
4F ProGuide needles (200 mm), stop tags, buttons, 111.850 Transfer tube set for 4F, flexibles (1-10)
and insertion needles (160 mm))
Intended for: 111.861 Transfer tube set for 4F, flexibles (11-20)
CT/MR OncoSmart catheter system, l =294 mm
(contains 30 flexible implant tubes (5 x 6 tubes), 4F
• Breast 189.840
ProGuide needles (294 mm), stop tags, buttons, and
insertion needles (280 mm))
• Head & Neck
CT/MR OncoSmart catheter system, l =294 mm
• Sarcomas 189.846
(contains 30 fexible implant tubes (5 x 6 tubes), 4F
ProGuide needles (294 mm), stop tags, buttons, and
insertion needles (200 mm))
• Tongue
70 71
Bronchus & Esophagus
Content
BRONCHUS
Lumencare Azure 74
ESOPHAGUS
Bonvoisin-Gérard esophageal applicator set 76
Esophageal applicator set 78
72 73
BRONCHUS & ESOPHAGUS
The LumenCare Azure offers clinicians an excellent Advanced solution for the most complex procedures Lumencare Azure 6F x 150 cm,
part# 110.683
solution for lung brachytherapy applications. It is State-of-the-art material characteristics make hard-to-
specifically designed to improve the delivery of lung reach places more easily accessible. In the event that the
Quantity Part number Description
cancer brachytherapy treatments, expanding the catheter gets obstructed e.g. through coughing, guide
2 110.682 LumenCare Azure 6F x 150 cm (5 pcs)
treatable area. wires are supplied with the LumenCare Azure, allowing
1 089.511 X-ray marker channel 1 - 150 cm
The smoothness of the catheter enables easy insertion Janusz Skowronek, MD, PhD. Ass. Prof.
and smooth source travel into the bronchus or any other Head of Brachytherapy Department
lumen, facilitating fast and simple lung brachytherapy Greater Poland Cancer Centre, Poland
procedures.
74 75
BRONCHUS & ESOPHAGUS
The Bonvoisin-Gérard Esophageal applicator is designed Intended for: Order information for microSelectron
to assist in the treatment of an endoscopically obstructed Esophagus
Bonvoisin-Gérard esophageal applicator set,
esophagus. By using this special dilation applicator with
part # 137.xxx (will be available soon)
tapered flexible tip the treatment area can be reached by Note:
Part number Description
bouginage. The bougie tube will also assist in centering
• No transfer tube is required
137.xxx (will be
the treatment catheter internally to minimize the mucosal The applicator set connects directly to the afterloader available soon)
Treatment tube
dose, while ensuring optimal depth dose penetration of the 137.261 Bonvoisin Gerard esophageal appl., 18F (Ø 6 mm)
• eparate applicator sets are available for Flexitron and
S
esophageal wall. Radiographic markers on the bougie tube 137.263 Bonvoisin Gerard esophageal appl., 24F (Ø 8 mm)
microSelectron afterloaders
aid in fluoroscopic placement and localization of 137.265 Bonvoisin Gerard esophageal appl., 30F (Ø 10 mm)
the applicator. • Flexitron needs an applicator adapter, see accessories
137.276 Applicator fixation mask
• User manual 090.628 for #085.035 137.277 Applicator cleaning accessories
• 094.353 Sterilization Advice Chart for #085.035 137.279 Applicator mask strap (pack of 5)
137.xxx (will be
Bougie tube, d = 10 mm
available soon)
137.xxx (will be
Bougie tube, d = 15 mm
available soon)
76 77
BRONCHUS & ESOPHAGUS
• This applicator set can only be used on 189.004 Esophageal fixing mask set
microSelectron afterloaders
78 79
Head & Neck
Content
NASOPHARYNX
Rotterdam nasopharyngeal mold 82
TONGUE
Pipard tongue template set 84
ALTERNATIVE FOR
TONGUE BRACHYTHERAPY
CT/MR OncoSmart catheter system 70
Breast
80 81
HEAD & NECK
MR safe
The Rotterdam nasopharyngeal mold is manufactured • The mold applicator and catheter set
from a flexible silicone material to easily shape the are deliverd unsterile and are for single use only.
radiation dose in the nasopharyngeal cavity. It consists of
two channels on the upper surface to optimize the dose • User manual 090.678
82 83
HEAD & NECK
MR Unsafe
84 85
Skin & Surface
Content
RIGID APPLICATORS
Valencia skin applicator set 88
Leipzig applicator set 90
FLEXIBLE APPLICATOR
Freiburg Flap applicator set 92
86 87
SKIN & SURFACE
Brachytherapy offers an excellent non-invasive solution • Skin and surface treatments 1 189.788 Valencia skin applicator, d = 20 mm
to complement or replace surgery for skin cancer. With our 1 189.760 Valencia skin applicator, d = 30 mm
Note:
specialized solutions, in a few fractions, a high radiation 2 085.115 Cap with marker ring, d = 20 mm
dose can be delivered to the tumor with a low dose to • Applicator provided with two types of treatment caps:
2 085.116 Cap with marker ring, d = 30 mm
the surrounding area, providing a short and effective 1) cap with marker and 2) cap for treatment. The cap
2 085.118 Cap, d = 20 mm
with the marker ring is used to determine the correct
treatment. 2 085.119 Cap, d = 30 mm
position of the applicator on the surface
1 084.158 Sterilization caps (pack of 50)
Controlled dosimetry for skin and • Applicator must be used with treatment cap in place
surface treatments Advanced Skin applicator set, part # 103.118
• The applicator set can be sterilized using
Part number Description
The Valencia skin applicator is a tungsten-shielded steam sterilization
applicator for contact treatments. The applicator’s unique 085.041 Leipzig applicator d = 10 mm
• Dose distribution charts are available for Iridium-192,
fixed flattening filter provides a flat and homogeneous 085.042 Leipzig applicator d = 20 mm
not for Cobalt-60. The charts are not included in
isodose pattern. The flat dose can be applied very 085.043 Leipzig applicator d = 30 mm
Oncentra Brachy treatment planning system.
precisely to skin and surface targets. This unique 085.114 Cap, d = 10 mm with marking ring
• Manual 777.00237 085.115 Cap, d = 20 mm with marking ring
applicator was designed in cooperation with the
La Fe University Hospital in Valencia, Spain. 085.116 Cap, d = 30 mm with marking ring
085.117 Cap, d = 10 mm
085.119 Cap, d = 30 mm
The Valencia skin applicator also contains a source
089.095 Freiburg Flap applicator set
stopper to secure a distal reference point for the source
189.701 Valencia applicator set
in the applicator, resulting in a reliable and reproducible
treatment. Plastic caps on the applicator help to avoid Accessories
over-dosage and assist with correct applicator positioning.
Part number Description
Reinforced shielding at the back of the applicator adds to
Well chamber adapter for HDR 1000 PLUS
189.786*
patient safety during treatment. (Standard Imaging, Middleton, Wisconsin)
085.117 Cap, d = 10 mm
085.118 Cap, d = 20 mm
085.119 Cap, d = 30 mm
MR Unsafe
084.158 Sterilizing caps, 3.2 mm
085.119 Cap, d = 30 mm
90 91
SKIN & SURFACE
Dr. S. Stange
Albert Ludwigs Universität,
Freiburg, Germany
92 93
Needles & Flexible
Implant Tubes
Content
PROGUIDE NEEDLES,
OBTURATORS AND MARKERS
96
94 95
NEEDLES & FLEXIBLE IMPLANT TUBES
The OncoSmart ProGuide CT Markers (see appendixes) Part number Description 110.584 ProGuide sharp needle set, 6F x 200 mm (3x6) 089.013 Obturator for 1.5 mm needle x 200 mm
are designed to indicate the first source positions on ProGuide needle set, 5F, sharp, l =240 mm,
110.585 ProGuide sharp needle set, 6F x 200 mm (6x6)
083.250 Obturator for 1.9 mm needle x 80 mm
189.604 18 needles (set contains 3 pouches of 6 needles each,
CT images of ProGuide interstitial needle implants. The and 6 obturators)
110.586 ProGuide sharp needle set, 6F x 200 mm (12x6)
083.252 Obturator for 1.9 mm needle x 100 mm
markers fit needle diameters from 5F to 6F and are kept 110.587 ProGuide sharp needle set, 6F x 200 mm (10x1)
ProGuide needle bulk set, 5F, sharp, l =240 mm, 083.254 Obturator for 1.9 mm needle x 120 mm
in place during imaging by means of a built-in locking 189.619 55 needles (set contains 8 pouches of 6 needles each,
and 7 pouches with single-packed needles)
mechanism. The markers are reusable and can be 110.588 ProGuide sharp needle set, 5F x 240 mm (3x6) 083.256 Obturator for 1.9 mm needle x 140 mm
autoclaved. Each marker length has a unique color- 110.589 ProGuide sharp needle set, 5F x 240 mm (6x6) 083.258 Obturator for 1.9 mm needle x 160 mm
coded knob. 110.590 ProGuide sharp needle set, 5F x 240 mm (12x6)
083.260 Obturator for 1.9 mm needle x 180 mm
110.591 ProGuide sharp needle set, 5F x 240 mm (10x1)
083.262 Obturator for 1.9 mm needle x 200 mm
96 97
NEEDLES & FLEXIBLE IMPLANT TUBES
Accessories
Part number Description
MR conditional
083.252 Obturator, l =100 mm (for all titanium needle lenghts)
045.004
Titanium needle stopper set (set of 10)
Interstitial needles are used for treatment of carcinoma Note: 083.147 Sterilizing caps, 1.9 mm (50 pcs)
where no lumen or cavity is available, or where additional • Needles are described by their outside diameter
and very precise target area coverage is required. The and length
needles are designed for optimal needle rigidity and
• Needles should be inspected prior to use. Any needle
sharpness throughout multiple interstitial treatments, in
that is blunt, corroded or bent should not be used
combination with the appropriate templates. Titanium
needles have a needle tip that is either round or trocar • Metal needles can be sterilized using steam
sterilization and ethylene oxide
shaped. The round tip is designed to minimize tissue
damage and optimize ease in implantation. The trocar tip • The use of titanium needles and related titanium
facilitates insertion in the tissue. needle stoppers is MR Conditional up to 0.23 Tesla
98 99
NEEDLES & FLEXIBLE IMPLANT TUBES
microSelectron *When ordered separately the minimum order quantity is 10 pcs. 10* 083.924 Trocar needle, 1.5 mm x 180 mm 083.147 Sterilizing caps, 1.9 mm (50 pcs)
10* 083.926 Trocar needle, 1.5 mm x 200 mm
• Needles should be inspected prior to use. Any needle
that is blunted, corroded, or bent should be discarded
100 101
NEEDLES & FLEXIBLE IMPLANT TUBES
Order information
Flexible implant tube set, 6F, 30 cm, part # 189.400* Why Flexible implant tubes?
Quantity Part number Description
• Easy to use for interstitial
2 189.300
Flexible implant tube, 6F, single leader, treatments, i.e. tongue, breast
30 cm (box of 36)
and soft tissue
Flexible implant tube, 6F, double leader, 30
2 189.298
cm (box of 36)
Flexible Implant Tubes CT/MR flexible implant tube set, 6F, 30 cm,
Accessories
Part number Description
part # 189.401
The flexible implant tubes are developed for treatments Note: 189.445
CT/MR flexible implant tube, 6F, single leader,
Quantity Part number Description 30 cm (box of 36)
that require a moldable source path. The flexible implant • Flexible implant tubes are delivered sterile, and are
CT/MR flexible implant tube, 6F, Flexible implant tube, 6F, single leader, 50 cm
tubes are equipped with ‘leaders’ to run the implant tube for single use only 2 189.445
single leader, 30 cm (box of 36)
189.504**
(box of 36)
through a channel in the tissue that is preformed by a Flexible implant tube, 6F, double leader, Flexible implant tube, 6F, double leader, 50 cm
• Flexible implant tubes are packed in pouches of 2 189.298 189.502
leader insertion needle. The flexible implant tubes are 30 cm (box of 36) (box of 36)
6, and with 6 pouches per box
available in various configurations: single leader with Flexible implant tube, 6F, blind end, Flexible implant tube, 6F, blind end, 50 cm
2 189.296 189.500
30 cm (box of 36) (box of 36)
button end; single leader with blind end, or double leader. • Buttons are packed in pouches of 6, and with
12 pouches per box 1 189.458 Radio-lucent button white, 6F (box of 72) Flexible implant tube, 5F, single leader, 30 cm
189.510**
They arrive sterilized and ready for immediate clinical (box of 36)
1 189.460 Radio-lucent button black, 6F (box of 72)
use. To improve visibility of the flexible implant tubes in • For Flexitron afterloaders, only Flexible Implant Flexible implant tube, 5F, double leader, 30 cm
1 189.462 Radio-lucent button red, 6F (box of 72) 189.508
(box of 36)
combination with state of the art CT and MR imaging tubes with length ≤ 400 mm can be used
1 189.464 Radio-lucent button blue, 6F (box of 72) Flexible implant tube, 5F, blind end, 30 cm
modalities, dedicated CT/MR flexible implant sets and • User manual 090.710 189.506
(box of 36)
1 189.466 Radio-lucent button green, 6F (box of 72)
buttons are available. 189.292** Radio-opaque button yellow, 6F (box of 72)
• The insertion needles are not MR safe 1 189.367 Cutter flexible implants
189.294** Radio-opaque button brown, 6F (box of 72)
Leader Insertion needles, 1.5 x 200 mm,
1 189.437
(box of 72) 189.472** Radio-opaque button white, 5F (box of 72)
Content
SOURCE POSITION CHECK RULER /
SOURCE POSITION SIMULATOR
106 / 107
104 105
QUALITY ASSURANCE TOOLS
Source Position Check Ruler Flexitron adaptor overview Source Position simulator Note:
The Source Position Check Ruler is designed for
Part number
Source position check ruler Transfer tube The Source Position simulator is designed to determine • User Manual for Flexitron: 777.00264
adapter set to connect part number
verification of the correct treatment position of the
the distal length of an applicator. • User Manual for microSelectron: 090.682ENG
sources and check cable, and of the length of each of the Gynecological titanium trans-
137.040 137.063
fer tubes
transfer tubes. It can be used with either the mechanical Order information
5F catheter transfer tubes
indicator or GAFCHROMIC* filmstrip, a method that (old versions - compatible to 137.090 / 137.091
Source position simulator for Flexitron:
enables instant logging of the result of the measurement. transfer tube sets delivered 137.092 /137.093
before May 2014)
Part number Description
137.041
The Source Position Check Ruler has embedded markers 6F catheter transfer tubes
137.094 /137.095 138.200 Source position simulator set for Flexitron
(old versions - compatible to
that allow multiple measurements and serve as reference transfer tube sets delivered 137.096 /137.097 Source position simulator set for Flexitron with
before May 2014) 138.202
x-ray markers
to validate correct positioning of the GAFCHROMIC®
filmstrip. The films are transparent for easy digitizing or 137.090 / 137.091
Accessories for Flexitron
137.042 5F catheter transfer tubes
scanning, and are self-developing. 137.092 /137.093
Part number Description
106 107
QUALITY ASSURANCE TOOLS
(PBC)*
Coupling transfer tube set 6F
1 111.480
flexibles (1-18)
111.420
Quick release coupling set
108 109
Accessories
Torque screwdriver Push fit disconnection tool Torque screwdriver, 2.5 mm
Part number Description Part number Description Part number Description
Perineal bar for vaginal Cervical stopper for vaginal Ovoid pair, 20 mm, small
084.008 084.009 084.042
applicator applicator unshielded
Perineal bar for the ring 084.158 Sterilizing caps 3.2 mm 086.011 Fixing mechanism_retractor
084.072
applicator
Kink protection tube, 089.636 Handle for obturator Leader insertion needle,
089.345 089.700
for transfer tube 1.5 x 120 mm (1 pcs)
110 111
ACCESSORIES
Note:
• Delivered non-sterile and not filled with fluid
• User manual 777.00081
Intrauterine tube, ovoids, 4 mm,
Intrauterine tube, ring, 4 mm Rectal retractor, 4 mm, 30 mm Rectal retractor, 4 mm, 40 mm various degrees Interstitial lunar ovoid tubes Vaginal caps
Part number Description Part number Description Part number Description Part number Description Part number Description Part number Description
Intrauterine tube 4 mm Rectal retractor, 30 mm with Rectal retractor, 40 mm with Interstitial lunar ovoid tube 126.007 Vaginal cap 22 mm right
110.303 until Intrauterine tube, 4 mm, 126.001
(available in various lengths 110.315 slot for 4 mm intrauterine 110.335 slot for 4 mm intrauterine 110.971 22 mm 60° right
110.307 various degrees 126.008 Vaginal cap 22 mm left
and degrees) tubes tubes
Interstitial lunar ovoid tube
126.002 126.009 Vaginal cap 26 mm right
22 mm 60° left
126.010 Vaginal cap 26 mm left
Interstitial lunar ovoid tube
126.003
26 mm 60° right 126.011 Vaginal cap 30 mm right
Part number Description Part number Description Part number Description Interstitial lunar ovoid tube
126.006
30 mm 60° left
Cervical stopper for 4 mm
110.395 110.396 Cervical stopper tool 110.540 Ovoid pair,15 mm,micro ovoids
intrauterine tubes
Perineal bar set Applicator clamp adapter set Catheter with collar 126.020 Intrauterine tube 40 mm 15°
Part number Description Part number Description Part number Description 126.021 Intrauterine tube 50 mm 15°
112 113
ACCESSORIES
Intrauterine tube, 6 mm,
Perineal template dorsal Perineal template ventral Bracelet for guiding tubes Insertion tool Internal fixation mechanism various degrees
Part number Description Part number Description Part number Description Part number Description Part number Description Part number Description
Number tag for guiding tubes Guiding tubes Perineal bar Fixation nut for rectal retractor Rectal retractor, 6 mm, 30 mm Rectal retractor, 6 mm, 40 mm
Part number Description Part number Description Part number Description Part number Description Part number Description Part number Description
Guiding tubes for interstitial Rectal retractor, 30 mm with Rectal retractor, 40 mm with
126.030 Number tag for guiding tubes 126.031 126.033 Perineal bar Fixation nut for rectal
lunar ovoid 189.741 189.742 slot for 6 mm intrauterine 189.751 slot for 6 mm intrauterine
retractor tubes
tubes
126.032 Guiding tubes for vaginal cap
gynecological applicator 8
116 117
X-ray, CT and MRI Markers for Flexitron
X-ray marker set CT Marker set 200 mm CT Marker set 240 mm CT Marker set 294 mm CT marker set 5F and 6F catheters (3 pcs)
The appropriate X-ray and CT-markers for use of a specific applicator with X-ray
MR Line
catheter
your afterloader can be selected with this table. If no crossmatch is indicated in X-ray catheter set 30 cm X-ray catheter set 50 cm Marker
set
this table, it is not possible to connect that applicator and afterloader. set
40 cm metal, metal, yellow, yellow,
CT/MR SS Ti metal, 6 black, 6 black, 12 black, 18 yellow, 6
12 mark- 18 mark- 12 mark- 18 mark- 100 mm 200 mm 300 mm 400 mm
GYN GYN GYN markers markers markers markers markers
ers ers ers ers
For channel: 1-3-5 1-9 10-18 1-18 19-30 3-11 1-9 10-18 19-30
Part number: 110.400 110.570 137.070 089.082 089.083 089.420 189.850 137.089 089.089 089.090 189.851 110.285 189.691 189.034 110.286 189.692 189.036 110.287 189.693 189.039 110.610 110.611 110.612 110.613 110.290
Breast 110.518 Breast CT/MR applicator set Depending on use of OncoSmart catheters or type of needles
Needles various Titanium needle sets (Trocar point & round point) x x x x x x x x x x
Note: markers for bronchus & esophagus applicators are not included on this page as they are included in the applicator set
118 119
Transfer tubes for Flexitron
For channel: 1-3-5 1-3-5 1-2-3 5 1-10 11-20 21-30 31-40 1-10 11-20 21-30 31-40 1-10 11-20 21-30 31-40 1-10 11-20 21-30 31-40
Part number: 111.650 111.680 137.063 111.683 111.805 111.816 111.827 111.838 111.850 111.861 111.872 111.883 137.090 137.091 137.092 137.093 137.094 137.095 137.096 137.097
Breast 110.518 Breast CT/MR applicator set Depending on use of OncoSmart catheters or type of needles
Head & Neck 045.410 Pipard tongue template set Depending on use of flexible implants
Note: bronchus & esophagus applicator sets include adapters to insert the applicator directly in the indexer
120 121
X-ray, CT and MRI Markers for microSelectron
X-ray catheter set CT marker set 200 mm CT marker set 240 mm CT marker set 294 mm
GYN X-ray
The appropriate X-ray and CT-markers for use of a specific applicator with X-ray
X-ray catheter X-ray MR line
catheter
your afterloader can be selected with this table. If no crossmatch is indicated in marker set, X-ray catheter set 30 cm X-ray catheter set 50 cm catheter marker
set 40
this table, it is not possible to connect that applicator and afterloader. set, 300 Norman 100 cm set
cm metal, 6 metal, 12 metal, 18 black, 6 black, 12 black, 18 yellow, 6 yellow, 12 yellow, 18
standard CT/MR mm Simon
USA GYN
GYN GYN markers markers markers markers markers markers markers markers markers
For channel: 1-3-5 1-9 1-9 10-18 1-18 19-30 3-11 1-9 10-18 19-30 1
Part number: 085.075 084.880 101.065 137.070 089.096 089.082 089.083 089.420 189.850 137.089 089.089 089.090 189.851 089.401 110.285 189.691 189.034 110.286 189.692 189.036 110.287 189.693 189.039 110.290
Breast 110.518 Breast CT/MR applicator set Depending on use of OncoSmart catheters or type of needles
Note: markers for bronchus & esophagus applicators are not included on this page as they are included in the applicator set
122 123
Transfer tubes for microSelectron
For channel: 1-3 1-3 1-2-3 3 1-9 10-18 19-30 1-9 10-18 19-30 1-9 10-18 19-30 1-9 10-18 19-30
Part number: 111.001 111.006 111.100 111.004 111.011 111.021 111.140 111.071 111.081 111.600 111.051 111.061 111.614 111.031 111.041 111.628
Breast 110.518 Breast CT/MR applicator set Depending on use of OncoSmart or type of needles
Prostate, rectum & bladder 189.069 Prostate stepper template set 5F Depending on use of type of needles
Head & Neck 045.410 Pipard tongue template set Depending on use of flexible implants
Note: bronchus & esophagus applicator sets include adapters to insert the applicator directly in the indexer
124 125
BRACHYTHERAPY APPLICATOR AND ACCESSORY GUIDE
ABOUT ELEKTA
Elekta’s purpose is to invent and develop effective solutions for the treatment of cancer and brain disorders. Our goal
is to help our customers deliver the best care for every patient. Our oncology and neurosurgery tools and treatment
planning systems are used in more than 6,000 hospitals worldwide. They help treat over 100,000 patients every day.
The company was founded in 1974 by Professor Lars Leksell, a physician. Today, with its headquarters in Stockholm,
Sweden, Elekta employs around 4,000 people in more than 30 offices across 24 countries. The company is listed on
NASDAQ OMX Stockholm..
Art. nr. 308880 MKT [13] ©2016 Elekta AB (publ). All mentioned trademarks and registered trademarks are the property of the Elekta Group. All
rights reserved. No part of this document may be reproduced in any form without written permission from the copyright holder.