InBody 570 User's Manual
InBody 570 User's Manual
Please note the important information below before reading this manual.
NOTE DANGER
Warning
고 주의 참조 주의
Failure to comply with safety warnings and regulations can cause serious injury or death.
NOTE DANGER
Caution
고 주의 참조 주의
Failure to comply with safety cautions and regulations can cause injury or property damage.
For any problems with the InBody570 or any clinical questions, contact:
Biospace, Inc.
13850 Cerritos Corporate Dr., Unit C, Cerritos, CA 90703
TEL: +1-323-932-6503
FAX: +1-323-952-5009
Website: http://www.biospaceamerica.com
E-mail: usa@biospaceamerica.com
I. InBody570 Installation
A. Product Components 4
B. Operating Environment 5
C. Installation Instructions 5
D. Required Settings 8
E. Connecting Printer and Stadiometer 9
F. Maintenance 10
V. Others
A. Exterior and Functions 18
B. Safety Information 21
C. Classification 22
D. Specifications 22
I. InBody570 Installation
A. Product Components
The InBody570 consists of the following components. Please make sure all of the following components are
present.
* Please inspect each component of the InBody570 for defects prior to installation.
➊ InBody570 Keypad
➋ Adapter (DC 12V, 3.4A) 1 EA
➌ Power cord 1 EA
➍ User’s Manual 1 EA
Upper part
Hand electrodes
Foot electrodes
➊
Footplate
➊
재포장
➍
User’s M 570
anual
4
B. Operating Environment
Please make sure that the environment is adequate for the InBody570 installation. This equipment is designed
for indoor use. If installing outdoors, the following requirements must be fulfilled.
Temperature range 50 ~ 104℉ (10 ~ 40℃)
Relative humidity 30 ~ 75% RH
Atmospheric pressure 70 ~ 106 kPa
C. Installation Instructions
1. Open the packing box of the InBody570 and remove the packing pads. Then take the InBody570 out of the box.
Packing pads
Caution
참조 주의
재포장
참고 주의
• If you have any problems installing your InBody570, please contact BIOSPACE for assistance.
• Do not transport the equipment by holding the screen portion or the joints of the hand electrodes.
• Keep the packing materials provided for repacking the equipment in the future. Other wastes should be
disposed of according to relevant laws and regulations.
2. After raising the upper part of the InBody570 all the way, remove the polyethylene foam cover.
Joint frame
Caution
참고 주의 참조 주의
• Using the InBody570 on carpet may cause static electricity, which could damage the equipment. If installing
the InBody570 on carpet is unavoidable, please use an antistatic mat.
• Install the InBody570 on a leveled, non-vibrating surface. Installing the equipment on an uneven surface may
cause the examinee to fall down. Test results may also be inaccurate.
• Never clean the hand and foot electrodes with liquid spray or detergent directly. The equipment may corrode
and/or malfunction if the liquid or detergent leaks inside. Use the InBody Tissue when cleaning the InBody570.
5
3. Level the InBody570 by rotating the leveling screws under the footplate to the left and right so that the
air bubble is centered.
* Leveling the equipment is necessary for accurate measurement of weight. There are a total of 5 leveling screws.
Level Indicator
Caution
참고 주의 참조 주의
• Avoid injuring your hands when rotating the leveling screws under the footplate.
4. Connect the adapter (➋) to the power input port, which is located on the rear panel (➊). Connect the adapter (➋)
to the power cord (➌). Plug the power cord (➌) into a grounded 3-socket outlet.
* The InBody570 can be used in connection with other test equipment such as a stadiometer or data management software called
Lookin’ Body. For more information, please refer to ‘A. Exterior and Functions 3. Rear Panel’ in section ‘V. Others’ in this User’s Manual.
➊
Power input port
Rear panel
➋ ➌
Grounded outlet
UTION NOTE DANGER 3-socket outlet
Warning
참고 주의 참조 주의
• Do not place the InBody570 in a location making it difficult to disconnect the power cord.
• Do not plug in or pull out the power cord with wet hands. There is a risk of an electric shock.
• Always use an outlet connected to the rated power (AC 100~240 V). Using other power rated outlets may result
in fire or malfunction.
• When using a power surge protector, make sure that the outlet or the extension cable has adequate power capacity.
• Do not disassemble or modify the equipment including internal parts without written consent from the
manufacturer. This may cause electric shock or injury, product malfunction, inaccurate results, and will void the
manufacturers warranty.
• Do not directly contact the InBody570 with any other electronic device when the InBody570 is on.
This may result in an electric shock.
6
UTION NOTE DANGER
Caution
참고 주의 참조 주의
• If
the InBody570 is not plugged into a grounded outlet, it may cause damage through electric surges or
product malfunction. This may affect the test results.
• The test results may be inaccurate if the InBody570 is under electrical interference. Do not install the InBody570
near products that generate electrical interference such as fluorescent lights, large AC motor equipment
(treadmill, vibration plate, refrigerator, air-conditioner, compressor, etc.), high-frequency thermal therapy
equipments, or heating appliances. Do not share the power source of the InBody570 with other electrical
devices. This may affect the test results.
• When connecting the InBody570 with other test equipment, turn on the other equipment first. When
turning off other equipment, turn off the InBody570 first. This is necessary to minimize electrical surges on the
InBody570.
• Always use the specified adapter provided by BIOSPACE as it is a part of the InBody570. Using other adapters
may result in malfunction of the InBody570.
• Operation of the InBody570 6500 feet above sea level may affect the weight measurement.
Power switch
7
D. Required Settings
1. The InBody570 automatically starts booting when it is turned on. While booting, it performs a self weight
calibration.
* While booting (about 5 minutes), make sure there is nothing on top of the footplate. Please do not stand on the footplate, or
place objects on the footplate.
2. Press the Administrator Menu button on the screen, which appears when no one is on the footplate.
Administrator Menu
3. Input the password (default password: 0000) to have access the Administrator Menu.
4. The Administrator Menu will give you access to many different options such as basic settings, touch calibration,
Results Sheet management, printing options, and more.
8
E. Connecting Printer and Stadiometer
1. Printer
In order to print InBody Results Sheets, an InBody570 compatible printer is required.
* A list of printers compatible with the InBody570 can be found at http://www.inbodyservice.com.
2) Plug the USB cable provided with the printer into the USB HOST port on the rear panel of the InBody570
and plug the other end of the USB cable into the printer.
2. Stadiometer
If a stadiometer is connected to the InBody570, the height values measured by the stadiometer will be sent
directly to the InBody570.
* Always connect a stadiometer from BIOSPACE.
2) Plug the serial cable provided with the stadiometer to the 9-pin stadiometer serial port on the rear panel of
the InBody570.
9-pin stadiometer
serial port
9
F. Maintenance
ON NOTE DANGER
Caution
참고 주의 참조 주의
• Do not bend the handles of the hand electrodes or rotate them in the reverse direction beyond its limitation.
• Do not extend the handles of the hand electrodes beyond its limitation.
• Do not place any objects on the footplate.
• Do not apply excessive force on the equipment.
• Turn off the equipment if you are not using it for a day or longer.
• Donot allow any liquid substances to contact the equipment directly. Keep food and drinks away from the
equipment. Substances getting inside the equipment can cause critical damage to the electronic components.
• Use a lint-free cloth to gently wipe the external surface of the equipment about once every week. Be careful not
10
II. InBody Test
A. Precautionary Steps
ON NOTE DANGER
Warning
참고 주의 참조 주의
• Individuals with medical implant devices such as pacemakers, or essential support devices such as patient
monitoring systems, must not use this equipment. Safe, low-level currents will flow through the body during
the test, which may cause malfunctioning of the device or endanger lives.
• Bioelectrical Impedance Analysis (BIA) uses safe low level currents, which are not harmful to the body. However,
ON NOTE DANGER
Caution
참고 주의 참조 주의
• Stand upright for about 5 minutes before testing. Taking the test immediately after lying in bed or sitting for
a long period of time might result in a slight change in the test results. This is because body water tends to
move to the lower body as soon as the person stands or gets up.
• Do not eat before testing. In cases where the examinee has already eaten, the test should be put off for at least
two hours after the meal. This is because food mass is included in the examinee’s weight and thus, may result
in measurement errors.
• Use the bathroom before testing. Waste is not included in the body’s compositional elements, but the volume
of urine and excrement is included in the weight measurement affecting accuracy of the test results.
• Do not exercise before testing. Strenuous exercise or sharp movements can cause temporary changes in body
composition. Even light exercise can change your body composition temporarily.
• Take the test in the morning, if possible. Body water tends to gravitate towards the lower body throughout
the day, affecting accuracy of the test results.
• Thoroughly wipe the palms and soles with the InBody Tissue before testing. Testing may be difficult if the
examinee’s palms and soles are too dry or if the examinee has too many calluses.
• Avoid contact with the examinee during testing. Contact may lead to interference affecting test results.
11
B. Test Instructions
1. Step on the footplate when the screen below is shown.
* The screens vary according to the Settings of the Administrator Menu ‘05. Self Mode/Professional Mode’.
Professional Mode: An examiner is present and guiding the examinee through the InBody Test.
Self Mode: The examinee takes the InBody Test following the instructions that are displayed on screen.
12
6. When the test is completed, the results will be shown on screen.
C. Test Posture
The examinee must maintain proper posture to have accurate test results.
* The test will proceed when there is good electrical contact.
Place the heels on the rear sole Step on the footplate barefoot.
electrodes.
13
III. Transportation and Storage
A. Cautions during Transportation
When transporting, have two people keep the InBody570 parallel to the ground.
B. Repacking Instructions
Once the InBody570 is installed, avoid transporting the equipment. If it must be transported, repack it in
the following sequence.
3. Put the InBody into the packing box. Place the packing pad over the equipment and tape up the
packing box.
재포장
Caution
참고 주의 참조 주의
• Always use the protective packing materials provided by BIOSPACE when repacking.
14
IV. Frequently Asked Questions (FAQ)
Even if no problems arise from the equipment, users may still have many questions, especially regarding
clinical procedures. Few common questions and answers are listed below. If your questions are not answered
here, please contact BIOSPACE.
* Customer contact information can be found under Settings of the Administrator Menu ‘24. Customer Service Information’.
Question Answer
• My InBody570 does not • Insert the power plug completely into a grounded 3-socket outlet.
turn on.
User’s M 570
anual
제품보증
서
• When using a power surge protector, the equipment may not power on if
the power switch on the power surge protector is turned off. Check the
power surge protector which the power plug is connected to.
• The User’s M
problem may occur if the 570
anpower
ual cord is not completely plugged into
제품보증
the adapter. Insert the power cord completely into 서the adapter.
• The problem may occur if you are using the power adapter that is not provide
by BIOSPACE. Always connect a power adapter (DC 12V, 3.4 A) provided
by BIOSPACE.
15
Question Answer
• My touchscreen • Calibrate
the touchscreen under Settings of the Administrator Menu
is inaccurate or not ‘23. Touchscreen Alignment’.
responsive. * Press firmly to optimize touchscreen response.
• If you cannot enter the Administrator Menu due to touchscreen problems,
please restart the InBody. The InBody stores the last touchscreen alignment
and will automatically recall the previous touchscreen settings. The InBody
can also recognize if its touchscreen alignment is off screen and will
automatically take the user to the calibration screen after restarting.
•Iwould like to connect • Pleaserefer to ‘E. Connecting Printer and Stadiometer’ in section
other equipment to the ‘I. InBody570 Installation’ in this User’s Manual.
InBody570.
16
B. Regarding the InBody Test
Some of the more common clinical questions are answered below. If additional questions or more clarification is
desired, please contact BIOSPACE.
Question Answer
• Mustsocks or stockings • Bare skin contact is essential in the analysis using the BIA method. Socks
be removed for the or stockings may cause a varying degree of distortion in the results. Socks
InBody Test? or stockings must be removed to obtain accurate data.
• Is it okay to wear accessories • Theideal condition for the analysis is simply standing with no clothes and
(jewelry, watch, rings, etc) or wearing no accessories. However, this may not always be possible.
metal objects while taking Therefore, we recommend that the examinee remove as many clothing items
the InBody Test? and accessories that may affect the weight as possible.
• Who cannot take the • Individualswith medical implant devices such as pacemakers, or essential
InBody Test or will have support devices such as patient monitoring systems, must not use this
difficulties taking the equipment. The currents will flow through the body during the test, which
InBody Test? may cause malfunctioning of the device or endanger lives.
• Can a person with metal • Theideal test methodology is where the examinee does not wear anything
implants in their body metallic. Individuals with metallic implants may have skewed test results
take the InBody Test? due to the conductivity of the metal affecting the results.
• Asthe weight of clothes and other wear affects the results of the body
composition analysis, it is strongly recommended to take off any heavy
clothing or metallic wear. Except for the weight, jewelry does not effect the
body composition analysis, as the contact point with the InBody570 are
the hands and feet.
• Is
the electric current • The physiological electric impedance method uses safe low level currents
harmful to the body? that is not harmful to the body. The safety of the InBody has been tested
and proven. The InBody products have been approved for medical use by
the CE and all over the world. Many medical institutions around the world
are actively using the InBody.
• How often should I take • Individualswho are undergoing any programs that may affect their body
the InBody Test? composition are strongly recommended to have the InBody Test done every two
to four weeks.
• Consistent testing will allow individuals to track and monitor their progress
over time.
• What are the precautionary • Please refer to ‘A. Precautionary Steps’ in section ‘II. InBody Test’ in this
steps to ensure accuracy of User’s Manual.
the InBody Test?
17
V. Others
* The InBody570 is manufactured according to the quality management procedure of BIOSPACE. BIOSPACE complies with the ISO9001
and ISO13485 which are international quality management systems.
* This equipment satisfies the IEC60601-1 (EN60601-1), an international safety standard for electronic medical equipment. This equipment
also satisfies the IEC60601-1-2 (EN60601-1-2), an international standard for electromagnetic conformity.
1. Upper Part
➊ Status LED: The LED emits a blue light when the InBody570 is turned on.
➋ Hand electrode: Examinee holds the hand electrode so that the 4 fingers wrap the surface of the
bottom hand electrode while the thumb is placed on the oval electrode.
➌ Hand electrode joint and handle: Supports the hand electrode and encloses the wiring for the electrode.
➍ Body: Connects the upper part of the equipment to the lower part.
➎ LCD screen: Shows each stage of the test, instructions, test results, etc. You can touch the screen to
input the data required for the test, configure settings, or view test results.
➏ Number keypad: Used for inputting age, height, and other number-based data.
➐ Text button: Shows the virtual keyboard on the screen. Used when inputting ID.
➒ Decimal point button: Used for inputting the decimal point in ID, height, age, or weight.
Enter button: Used to finish inputting data or to save changes in Administrator Menu.
Delete button: Used for deleting inputted data.
➊
➌ ➎
➏
➐ ➓
➑
➒
➋
➍
18
2. Footplate
➊ Front sole electrode: The examinee makes contact with this electrode by stepping with the front part of their foot.
➋ Rear sole electrode: The examinee makes contact with this electrode by stepping with the heel of their foot.
➌ Footplate: This is connected to the scale, which measures the examinee’s weight.
➍ Level Indicator: Indicates the current horizontal level of the InBody570.
➎ Joint: Joins the upper part and lower part of the equipment together.
➏ Leveling screws: Used for adjusting the horizontal level of the equipment.
➊
➋ ➎
19
3. Rear Panel
➊ Speaker: Provides audible indication for test in progress, test complete, and successfully saved setting changes.
➋ Rear cover: Designed to be opened only by BIOSPACE service personnel for checking internal circuits, etc.
➌ 9-pin PC serial port (Female, RS-232C): Used for connecting the InBody570 to Lookin’ Body installed on the computer.
* The InBody570 can be connected to Lookin’ Body installed on a computer using one of the ports ➌, ➎, or ➏.
➍ 9-pin stadiometer serial port (Female, RS-232C): Used for connecting the InBody570 to a stadiometer.
* Only compatible with a BIOSPACE stadiometer.
➎ LAN port (10T Base): Used for connecting the InBody570 to Lookin’ Body installed on a computer
* The InBody570 can be connected to Lookin’ Body installed on a computer using one of the ports ➌, ➎, or ➏.
➏ USB SLAVE port: Used for connecting the InBody570 to Lookin’ Body installed on a computer.
* The InBody570 can be connected to Lookin’ Body installed on a computer using one of the ports ➌, ➎, or ➏.
➐ USB HOST port: Used for connecting to a printer or a USB Thumb Drive.
➋
➍
➒
➑
20
B. Safety Information
Indicators
USB port
Safety Symbols
Warning, Caution
4
BF Type Equipment
4
Adapter
4
4 Power On
4
4
Power Off
4
WARNING
Electric shock hazard – do not dismantle.
Dismantling will void the warranty.
AVERTISSEMENT
Risque de choc électrique - ne pas démonter.
Le démontage annulera la garantie.
DANGER
Do not use this equipment with electrical medical device such as a pacemaker.
Ne pas utiliser cet équipement avec des appareils médicaux électriques comme un stimulateur cardiaque.
CAUTION
Do not spray any liquid substance directly onto the device.
Ne pulverisez aucune substances liquids directement sur l’appareil.
CAUTION
No excessive force on shoulder joint
Ne pas appliquer de force excessive sur les bars articulés.
21
C. Classification
Body Composition Analyzer of Direct Segmental Multi-frequency Bioelectrical Impedance Analysis Method
• Type of protection against electric shock: Class I
• Type of the applied parts: BF Type
• Degree of protection against water infiltration: IPX0
• EMC Immunity: Level A
• EMC Emission: Level A
• Equipment is not suitable for use in the presence of flammable mixtures.
D. Specifications
Bioelectrical 15 Impedance Measurements by Using 3 Different Frequencies
Bioelectrical
Impedance Analysis (BIA) (5kHz, 50kHz, 500kHz) at Each of 5 Segments (Right Arm, Left Arm,
Impedance (Z)
Measurement Items Trunk, Right Leg and Left Leg)
Measurement Method Direct Segmental Measurement Bioelectrical Impedance Analysis Method, DSM-BIA
Body Composition
No Empirical Estimation
Calculation Method
Outputs Intercellular Water, Extracellular Water, Total Body Water, Dry Lean Mass, Lean Body Mass,
Body Fat Mass, Weight, Skeletal Muscle Mass, Body Mass Index, Percent Body Fat,
Segmental Lean Analysis (Right Arm, Left Arm, Trunk, Right Leg, Left Leg),
Segmental Fat Analysis (Right Arm, Left Arm, Trunk, Right Leg, Left Leg), ECW/TBW,
Body Composition History (Weight, Skeletal Muscle Mass, Percent Body Fat, ECW/TBW),
Basal Metabolic Rate, Visceral Fat Level, Impedance at each segment/frequency
Logo Name, Address, and Contact Information can be shown on the InBody Results Sheet.
Provides audible indication for test in progress, test complete, and successfully saved
Voice Guidance
settings changes.
Test results can be saved if the member ID is utilized. The InBody can save up to 100,000
Database
results.
Copy, backup, or restore the InBody570 data (data can be viewed on Excel or Lookin'Body
USB Thumb Drive
data management software)
22
Applied Rating Current 400µA(± 40µA)
Model JMW140KA1240F02
External Interface RS-232C 4EA, USB HOST 2EA, USB SLAVE 1EA, LAN (Ethernet, 10T) 1EA,
Bluetooth 1EA, Wi-Fi 1EA
Storage Environment -4 ~ 158℉ (-20 ~ 70℃), 10 ~ 95% RH, 50 ~ 106kPa (No Condensation)
23