Stargas 1
Stargas 1
3
4
6
5
7 8
10
9
9001328
Operating instructions
Rel. 3 11/07
11
12
13
53
14 e
15
16
17
18 54
19 55
20
21 f
22
23
56 57
24
44 25
43 41
42 40 26 g
39 27
38 h
28
37
36
35 33 29 59
32 f 58
34 31
34 30
45 47
46
60
48
49
a
61
b
62
50 52
c d
51
63 64
65 66
67 68
69 70
71 72 73
__________________________________________________________________________ STARGAS BOX
Thank you for having chosen one of our instruments for your workshop. We are
certain that it will give the utmost satisfaction and be a notable help on the job.
Please become fully familiar with the instructions in this user’s manual. It should
be kept ready to hand for consultation whenever required.
The STARGAS BOX analyser ranks at the top of the last generation of gasoline engine
exhaust gas analysers and allows to provide the repairer with a reliable guide to the
analysis procedure.
STARGAS adds to the gas analyser functions - the analysis of CO, CO2, HC, O2, NO
(optional), of Lambda coefficient, of engine revolutions (RPM) and of temperature -
the Lambda probe operation and the relative probe-central unit system
control.
The analyser is nowadays absolutely necessary for a good engine tuning, and for the
diagnosis of possible operation drawbacks aimed at limiting the vehicle polluting
emissions.
If the emission values are not correctly adjusted it will be absolutely impossible to
obtain a good operation of engines, especially of electronic injection fuel ones.
OPERATING INSTRUCTIONS 1
• It is forbidden to even partially this handbook in any way unless prior written authorisation has
been obtained from the manufacturer.
• The data and characteristics indicated in this handbook are not binding. The manufacturer reserves
the right to make all those modifications as are considered necessary without being obliged to give
advance warning or make replacements.
• All the names of brands and products and the trade marks are the property of the respective
owners.
INDEX
OPERATING INSTRUCTIONS 3
STARGAS BOX ___________________________________________________________________________
4 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
Read carefully the instructions for the installation, use and maintenance contained in
the operative manual.
Do not allow unskilled personnel to use this equipment, in order to avoid accidents to
the people or damages to the equipment.
The work place must be dry, sufficiently bright and well aired.
Particularly, vehicle diagnosis operations involving engine ignition must be carried out in
a room provided with an exhaust fan.
We remind you that inhalation of carbon oxide (which is odourless) causes serious
health damages.
• Before going on with the test, to make sure that the car gear is in neutral position
(or in parking position if the vehicle is an automatic transmission one), to put on
the emergency or parking brake in the vehicle which is going to be tested and to
verify if the wheels are locked.
• To protect face, hands and feet and avoid any contact with hot surfaces such as
sparking plugs, exhaust pipes, radiators, cooling system connections.
• Not to smoke and not to light flames when working on the vehicle.
• Not to look directly and at a close distance into the carburettor suction pipe when
the engine is on.
• To keep one's hands and hair off the moving parts. Never to wear ties, large
clothes, wrist pieces of jewellery and watches when working on a vehicle,
especially if the engine is on.
• To keep off the fan; the cooling fan is controlled by a thermal switch which is
linked to the coolant temperature: thus disconnect the fan cable when working on
a still hot engine, in order to prevent the fan from starting to work suddenly even
if the engine is off.
• Not to pour fuel directly in to the carburettor in order to ease the engine starting.
OPERATING INSTRUCTIONS 5
STARGAS BOX ___________________________________________________________________________
• Not to unscrew the radiator plug before the engine temperature and consequently
the cooling system pressure have lowered.
• Not to touch the high voltage cables when the engine is on.
• To handle portable lamps carefully and use only the ones that are provided with
metallic protection.
• To wear protective glasses in order to protect the eyes from gasoline, dust or
metals.
• To remember that the catalytic muffler reaches very high temperatures that can
cause serious burns or fires.
• Hence check that near the muffler there are no oil stains, wipers, paper or other
easily inflammable materials.
Car batteries contain sulphuric acid and produce explosive gases; hence pay attention
to the following instructions:
• Never leave tools onto the battery since they could cause accidental contacts.
• Before carrying out the test or the recharge, cover with a wet wiper the battery
openings so as to choke the explosive gases.
• Avoid electrolyte sprinkles on the skin, on the eyes and on cloths, since it is a
corrosive and highly toxic compound.
6 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
Disposing of equipment
• The public administration and producers of electrical and electronic equipment are
involved in facilitating the processes of the re-use and recovery of waste electrical
and electronic equipment through the organisation of collection activities and the
use of appropriate planning arrangements.
Disposing of batteries
OPERATING INSTRUCTIONS 7
STARGAS BOX ___________________________________________________________________________
In order to use the STARGAS BOX gas analyser safely it is necessary to comply with
the following rules:
• The analyser must be used in dry places. Its exposure or use near sources of heat
or polluting emissions (stoves, ovens, etc...) must also be avoided.
• In order to connect the analyser to the feeding mains make sure that the mains
voltage/frequency corresponds to the ones which are suggested for the analyser.
• Replace the burnt fuses with fuses having the same features.
• Do not lean objects on the supply cable and never bend it sharply.
8 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
11. Main switch: to turn on and off the tool if plugged to the mains supply.
12. Main switch: to turn on and off the tool if plugged to a battery.
13. Filter-holder with internal cooling air filter that can be washed.
15. Standard cartridge/filter: It allows to separate the impurities from the exhaust
gases.
16. Certification plate: All of the technical features can be found there.
18. Oxygen Sensor: It measures the oxygen percentage which is contained in the
exhaust gases.
19. COM1 Port: serial port for the connection to the autodiagnosis interface.
20. COM2 Port: serial port for the connection to the autodiagnosis interface.
21. PARALLEL port: Port for the connection to a 80-column standard printer.
22. USB Port: Port for the connection of a 80-column external printer and for
software update.
25. RPM Socket: Socket for the connection of the RPM clamp from battery or of
induction clamp (option).
26. TEMP Socket: Socket for the connection of the oil temperature probe.
27. Ambient sensors: Sensors to detect environment data (relative humidity and
temperature).
28. λ Socket: Socket for the connection of the lambda probe 1 Volt/5 Volt.
29. COM port: Port for the vehicle ground connection; battery negative.
30. Oscilloscope connector BNC: It allows the inlet of the measured signal through
the relevant wire connection.
OPERATING INSTRUCTIONS 9
STARGAS BOX ___________________________________________________________________________
31. VGA port: Port for the connection to a standard monitor or to a TV color
(PAL/NTSC systems).
33. SMOKEMETER port: Port for the connection to a partial flow smoke analysis cell
combined with external supplier.
34. Tightness Test Plugs: They verify the internal and external pneumatic circuit
tightness.
36. Condensate Separating Filter: It allows to separate the condensate from the
exhaust gases.
40. Mains Socket: Tool power supply socket by the connection of the relevant cable
to the power mains.
41. Battery Socket: Tool power socket by the connection of the relevant battery
cable (12 Volt).
42. Actived Carbon Filter: It separates the air that is sucked up for the internal
pneumatic circuit cleaning from unburnt hydrocarbons and impurities.
10 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
ATTENTION:
Remove the protection plugs which have been inserted in
the external connections.
It is necessary to keep the plugs because in case that a
new package is needed for the forwarding to the
laboratory (calibration, overhaul, repair), they assure a
suitable protection to the whole internal pneumatic
circuit.
• Insert the special pipe (49) into the condensate outlet connection
OUT WATER (44) so as to convey the condensate far from the analyser.
• Set up the gas pick-up system according to the following rules:
Insert the sampling gas probe (54) in the 6m piping (48-b).
Connect to the piping the external prefilter which is equipped with its
80cm pipe (48-a).
Insert the above-mentioned assemblage into the inlet gas connection (38).
• Connect the RPM battery cable (62) or in case of use the antenna system, connect
the relevant cables (69) o (70), to the RPM socket (25).
• Connect the ground cable (60) to the COM port (29). Such cable doesn’t have
to be connected, in case of use the aerial system for RPM surveying.
• Connect the temperature probe (61) to the TEMP. °C tap (26).
• Connect the extension (58) to the λ tap (28) to check the effectiveness of the
Lambda probe/central unit.
• Attach to the extension (58) the SHUNT cable (57) or the connecting cable of
the test prod clamp (56) with the prod (55).
• Connect the serial cable (71) to COM2 (20) or COM1 (19) port of the
STARGAS BOX and to one of the port presents on the PC. The configuration of the
communication port on the STARGAS BOX happens in automatic mode.
• Connect the sample gas bottle if required by the rules in force to the coupler (43).
• Attach the battery supply cable (59) to the battery socket (41) and to the tested
vehicle battery (in case the vehicle is not already supplied by the mains cable).
• Attach the mains cable (63) to the mains socket (40) and to the supply source
of the power mains (in case the vehicle is not already supplied by battery cable).
OPERATING INSTRUCTIONS 11
STARGAS BOX ___________________________________________________________________________
• For the connection to the feeding mains, make sure that the tap has the same
VOLTAGE/FREQUENCE features as those that are written on the plate which is
located on the analyser back side. The earthing is absolutely necessary.
ATTENTION:
The oxygen sensor (18) is one of chemical kind and it has to be replaced
when it is no longer possible to obtain the automatic span. To replace the
exhausted sensor it is necessary to ask for the intervention of the authorised
technical service. Not to alter the test results, let the gas outlet flow coupler
free (37).
Error messages are displayed on the monitor simultaneously to the emission of a sound
warning.
Once the connections have been performed, before starting the test check:
ATTENTION:
During the outdoor gas analysis tests be sure the gas sampling facility is not
directly exposed to sun rays. As a matter of fact, a temperature increase of
piping/filter can cause a steam accumulation inside the analyser thus
affecting measurements.
12 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
3.2 - Temperature Measurement
During the carrying out of the tests the engine temperature is displayed on the
TEMP. °C box, after beeing measured through the probe the analyser is equipped with.
The temperature (61) probe has to be introduced at the place of the oil dipstick.
In order to introduce it deeply enough, adjust the rubber plug by using as a reference
the original dipstick length.
ATTENTION:
When introducing the temperature probe, do not bring the connecting cable
near sources of strong interferences (coil, spark plug cables, distributor).
Moreover do not twist the cables inside the engine compartment.
While carrying out the gas analysis test, it is also necessary to make sure that the
VEHICLE IS IDLING and at ACCELERATED SPEED (manufacturer’s data).
Battery mode
To measure the rpm you can use the internal revolution counter of the tool that
processes a signal from the battery positive terminal and detects the rpm.
That is set for 4-cylinder engines, but also 1, 2, 3, 5, 6, 8, 10, 12 cylinder can be set.
• Connect the RPM cable from battery (62) to the battery positive pole.
• Start the engine, the RPM case on the LCD will read the rpm.
OPERATING INSTRUCTIONS 13
STARGAS BOX ___________________________________________________________________________
NOTA BENE:
If the programme displays the RPM RANGE SELECTION page, usually when the engine
rpm number is lower than 700 RPM or when it exceeds 900 RPM, it is necessary to
select the RPM value being nearest to that of the tested engine. The programme
processes the data by setting as predefined value the measure it deems to be more
correct.
After carrying out the relevant checks the operator shall select the most reliable
measure (by using the cursor shifting keys) and confirm by ENTER.
If during the tests the rpm is unstable (anomalous values) the problems are due to the
generator charge pulses or to the charge current pulses interfering with the continuous
current of the electrical system. These currents cause a voltage drop on the electrical
system lines that can be measured by the tool revolution counter.
• Measuring the speed functions better with a partially discharged battery than with
a fully charged one. It is, therefore, worthwhile switching on several loads for a
few minutes before starting the engine.
• During measurement switch on several loads such as the lights, rear window
heating etc.
• Once the engine is on, press down fully on the accelerator, the generators
regulator can only be activated through this.
• Be sure that the V-belt used to drive the generator does not slip off, otherwise
faulty measurements could occur whilst the engine is accelerating.
• In some vehicles a switched on fan can distort measurement. In this case, switch
off the fan.
The rpm reading can be carried out by using the RPM cable from the electric lighter
(option) or the cable with induction clamp (option).
The latter is set to 4-time engines, but can be set to 2-time/DIS engines too.
ATTENTION:
At the end of measurement disconnect the RPM cable from the battery and
short circuit the clamps one another to avoid the display of a wrong rpm due
to the high sensitivity of the system. If a RPM cable from car lighter is used
(option) short circuit it through the special hook connected to it.
14 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
Induction clamp mode
If during the tests with induction clamp the revolution number is unstable (showing a
trend towards anomalous values), probably some parasitic signals or some other signals
coming from the ignition system of the other non concerned cylinders get into the
induction clamp.
• Spark plug cables too close one another and non perfectly insulated.
ATTENTION:
Check that no electrical discharges occur between the spark plug cable and
the clamp, which could seriously damage the analyser.
Antenna mode
An antenna positioned near by the vehicle, picks up from ignition peaks that are
opportunely processed.
• Position the antenna (69) or (70), according to that one that thinks more adapted
for the measure, near to the engine;
• start the engine in order to visualize into the RPM square, the RPM number.
OPERATING INSTRUCTIONS 15
STARGAS BOX ___________________________________________________________________________
This kind of antenna was designed to use the instrument with the four-cylinder engines
that are typical of road motorcycles. It is advisable to use it both in the case of ignition
coils separated from spark plugs, and in the case of ignition systems that have no spark
plug high voltage cable. This antenna is also suitable for gasoline engines of vehicles
with head coils and common rail diesel engines (for the correct positioning and fixation
see antenna mode with common rail engines).
Use the revolving antenna by positioning it near one cylinder only so as to enable the
instrument to receive the correct signal only.
NOTA BENE:
For this kind of engines the rpm measurement can often turn out to be easier by using
the induction clamp.
Use the revolving antenna and position it near the coil unit (ex. BMW motorcycles).
16 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
OPERATING INSTRUCTIONS 17
STARGAS BOX ___________________________________________________________________________
SENSIBILITY ADJUSTMENT
Selecting the Antenna mode, in the angle on the left the bar chart is visualized.
It indicates the power of the signal in entrance to the measure circuit.
If the indicator bar is positioned between the half and the maximum, the signal
received presents an optimal power for the measure. In contrary case approach mainly
the antenna at the source of the signal.
NOTA BENE:
The signal power decreases with the increase in the rpm number.
The cylinder number preset by the tool is 4; modifying the cylinders number, the
following chosen will be displayed:
• Position the selection level with suitable choice to the type of measure that must
be executed.
NOTA BENE:
The motorcycles type VESPA 125 PRIMAVERA ET3 are characterised by anomalous
signals such as 1S every driving shaft revolutions (2 time/DIS selection) when the
engine is idling and 2S every driving shaft revolutions (2 time-2 sparks selection) when
the engine is accelerated.
18 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
The sensitivity value preset by the tool is 200.
This value must not be modified if the tool shows the correct and stable rpm.
• Gradually lower sensitivity until reaching a correct rpm value, in case the displayed
rpm is higher than the true one.
• Gradually increase sensitivity or take the antenna near the signal source, in case
the displayed rpm is lower than the true one.
NOTA BENE:
Always check the rpm value also at the accelerated idle speed.
Approximately, the values of rpm number, at idling, are those that are listed in the
following table.
ATTENTION:
The cellular telephones can disturb the ANTENNA MODE measure.
OPERATING INSTRUCTIONS 19
STARGAS BOX ___________________________________________________________________________
The common rail engines revolution speed can be measured by an antenna system
thanks to the injectors electromagnetic emissions.
1
1
ATTENTION:
Do not attach the antenna grippers to the positive pole and/or
negative pole of the battery.
20 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
Tecnotest equipment is provided with different software for PCs with Windows OS.
An auto-install CD-ROM is provided for the installation of the VISA PLAnet software.
PC minimum requirements PC:
Pentium MMX 233 MHz
128 MB RAM
10 GB Hard Disk
Windows 98 Second Edition
1 Port USB
1 Port LPT
CD-ROM drive
Mouse
Keyboard
1 RS232 serial port for each connected device.
PC recommend requirements:
Pentium Celeron 850 MHz
128 MB RAM
40 GB Hard Disk
Windows 98 Second Edition
2 Port USB
1 Port LPT
CD-ROM drive
Mouse
Keyboard
1 RS232 serial port for each connected device.
Software prerequisites (if the following programmes are not installed on the PC, they
are automatically installed from the CD-ROM):
Java 2 Runtime Environment or higher.
Microsoft Internet Explorer 5.02 or higher.
.NET Framework 1.1.
.NET Framework 2.0.
OPERATING INSTRUCTIONS 21
STARGAS BOX ___________________________________________________________________________
The user interface is conforming to the Windows applications standard; once the
VISA PLAnet installation has started, the user shall press Next on each screen page.
After the prerequisites (if needed) have been installed, the VISA PLAnet installation
starts, displaying a page with the installation confirmation request.
The installation starts automatically and ends some seconds after a message for the
user is displayed
• Open the Control Panel, verify the installation (VISA PLAnet) and, if needed,
remove it, just like any other programme installed on the PC.
During the removal procedure, all the folders and files that have been copied on the PC
during the installation are cancelled. Files containing customized configurations will not
be removed. These files can be cancelled manually.
NOTA BENE:
The first time that each VISA PLAnet suite application is launched, the relevant license
contract is displayed. The contract shall be read and accepted before the software can
be used.
22 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
4.2 - Hardware key protection
• Make sure that the hardware key is inserted in way corrected in one door USB of
the PC.
• In order to verify the correct operation of the hardware key supplied with suite
VISA PLAnet, enter in menu the UTILITIES of the MAIN MENU and select
VERSIONS SOFTWARE.
The utilities program recalled, will control the presence of the hardware key and will
suggest to the customer like resolving eventual problems.
4.3 - Powering
• Press the (11) or (12) switch located on the back side of the analyser next to the
mains supply tap.
NOTA BENE:
To start the VISA2 central unit or the Personal Computer, please refer to the specific
operational manual.
To set the serial port on the VISA2, please refer to the specific operational manual of
the multioperational unit.
For the correct functioning of the gas analyser connected to a Personal Computer,
the serial port setting is required.
• Select the Protocol properties page and set the serial port to be used in the
analyser SERIAL PORT section.
• Once the setting has been completed, quit and restart the program.
OPERATING INSTRUCTIONS 23
STARGAS BOX ___________________________________________________________________________
• Select EXHAUST GASES function from MAIN MENU page and afterwards
GAS ANALYSIS.
The waiting message WAIT PLEASE is displayed and subsequently you enter the
GAS ANALYSIS programme.
Warming up Time
After to have selected one of the tests available in the programme, the WARMING UP
message is displayed; it signals the analyser warming up phase which can last till a
maximum of 60 seconds.
The analyser automatically starts the zero setting phase. The AUTOZERO IN
PROGRESS message is displayed.
At the end of this phase the analyser is ready to be used.
In case some calibration errors occur (for instance due to a defective or exhausted
sensor) the message REPLACE O2 SENSOR will be displayed.
24 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
• Press F4 to select the type of fuel. Make sure the chosen fuel is displayed in the
box.
• Perform two rapid idle accelerations and bring the engine back to the slow
running.
• Introduce the gas sampling probe into the exhaust pipe as deep as possible, and
anyway at a depth of at least 300mm.
• If the exhaust pipe extremity does not allow to introduce the probe completely, it
is necessary to add a special extension assuring the tightness in the joint area.
The analyser displays the values of the single gases, the Lambda coefficient, the
revolution number and the engine temperature value.
OPERATING INSTRUCTIONS 25
STARGAS BOX ___________________________________________________________________________
To carry out settings (in case they differ from those of default) follow the instructions
below:
• Select the type of revolution cable used to carry out the test by mouse.
• Select the engine times (Strokes N.) of the tested engine (2, 4 times or 2 times
2 sparks), only in case the induction clamp or the antenna system with a different
number of cylinders from that 4 cylinders preset, is used.
• In case the battery cable or antenna system is used, press many times the F2 key
until displaying the number of cylinders of the tested vehicle.
To calibrate the rpm (only in case the battery cable is used) follow instructions below:
• Select the RPM COUNTER CONFIGURATION page and connect the revolution cable
from the battery (see chapter 3.3).
• The light on the RPM COUNTER CONFIGURATION page is usually red
(no calibration performed), while when the tool starts calibration it turns yellow.
• At the end of calibration the light becomes green and the RPM box indicates the
revolution number.
• If at the end of automatic calibration the value is not correct, select F1 key to
carry out a second calibration.
NOTA BENE:
In the case an adapter model 2033 is used, select the INDUCTION CLAMP mode and
lower sensibility from 200 (default value) to 120.
To carry out a correct measure of the rpm number refer to the use of the rpm counter
(see chapter 3.3).
26 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
• Press the F3 key to store the new settings and quit the RPM COUNTER
CONFIGURATION page.
• Select the ESC function to quit the RPM COUNTER CONFIGURATION page without
storing the new settings.
5.1 - Print
At the end of the printing the GAS ANALYSIS page is displayed again.
• Compare the retrieved values with the data that have been specified by the car
maker and if it is necessary correct them by means of the special adjustment
systems (screw on the carburettor, screw on the flow meter, screw on the central
unit etc...).
The autozero function allows the machinery to zero the values. The autozero message
is automatically displayed either according to the preset parameters or every time the
user selects the function.
OPERATING INSTRUCTIONS 27
STARGAS BOX ___________________________________________________________________________
The Lambda coefficient shows the stoichiometric ratio, that is the ratio between the air
and petrol quantities in the combustion chamber.
When this value is very close to 1 it means that the combustion inside the chamber is
as near perfect as can be expected.
⎡ ⎛ ⎞ ⎤
⎢ ⎜∗ ⎟ ⎥
CO Hcv 3,5
21× ⎢CO 2 + + O2 + ⎜ × − 0,0087 ⎟ × (CO 2 + CO)⎥
⎢ 2 ⎜ 4 CO ⎟ ⎥
⎢ ⎜ 3,5 + ⎟ ⎥
⎣ ⎝ CO 2 ⎠ ⎦
λ =
⎡ CO ⎤
⎢ CO 2 ⎥ ⎡ ∗
Hcv 0,01754 ⎤
⎢21 + 0,5628 × ⎥ × ⎢1 + − ⎥ × [CO 2 + CO+ HC× 6 × 10 ]
−4
⎢ CO ⎥ ⎣ 4 2 ⎦
3,5 +
⎢ CO ⎥
⎣ 2 ⎦
28 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
5.5 - Values exhaust gases
The value of the single gases (CO, CO2, O2) is expressed in percentage-volume.
The unburnt hydrocarbons (HC) and Azote oxide values (NOX) are expressed in
PPM VOL (Parts Per Million-Volume).
The corrected CO indicates the CO when the following conditions are satisfied
simultaneously:
(CO × 15∗ )
The applied formula is the following one: CO corr =
(CO+CO2)
The corrected CO is always printed (if CO2 ≥ 6%) together with CO. The percentage
value has to be taken into consideration only when the corrected CO value turns out to
be different from the CO value.
The corrected CO indicates the value of the CO which is emitted by the engine even
when the vehicle exhaust system is not tight.
As a matter of fact the gas values are diluted by the air suction hence they are not
reliable.
The corrected CO is printed with larger fonts, as provided for by the law in force.
The corrected CO value must be taken into consideration only when the measured gas
values are stabilized.
OPERATING INSTRUCTIONS 29
STARGAS BOX ___________________________________________________________________________
DIAGRAM OF COMPARISON
AMONG THE 4 ANALYSED GASES (CO, CO2, HC, O2)
ON A SAMPLE OF 100 FUEL INJECTION VEHICLES
ALIMENTATI AD INIEZIONE
IN GOOD MAINTENANCE CONDITIONS
In order to read the diagram correctly, consider the CO (Carbon Monoxide) as the
reference gas.
30 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
Possible causes for CO - CO2 - HC and O2 wrong values
• IGNITION LEAKAGES:
Defective reducer contacts
Defective spark plug cables
Wrong advance
Defective spark plugs
HC • INCOMPLETE COMBUSTION:
Lean mixture
Defective suction circuit tightness
• MECHANICAL DEFECTS
Insufficient compression
Valves that are not tight
O2 • Rich mixture
• Defective Lambda Probe
OPERATING INSTRUCTIONS 31
STARGAS BOX ___________________________________________________________________________
6.0 - CURVES
• Select CURVES from the page GAS ANALYSIS in order to display the exhaust gas
variation graphically.
1 2 3
The graph which is visualised on the display indicates the measured gases progress.
The values correspondents to every single gas (1), are displayed in their respective
boxes.
• Press with the mouse on the key (2) in order to interrupt the values of the
selected gas acquisition.
• With the mouse carry the cursor (3) in the point of the acquired curve that is
wanted to be inspected, while the relevant gas values are displayed.
• Press with the mouse again to resume the acquisition of the value.
32 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
7.0 - HISTOGRAMS
This kind of visualisation proves to be very useful because it allows both to adjust the
CO value rapidly and without errors, and to check immediately if also CO2, HC and O2
are within the limits that have been specified by the Car Makers.
During these two graphic tests the diagrams will display the gases only if they do not
exceed the most significant preset limit values.
This is done in order to obtain a good resolution thus enabling to single out even very
little variations of the measured gases.
The values are always precisely displayed in their respective boxes, although they are
not graphically visualised.
OPERATING INSTRUCTIONS 33
STARGAS BOX ___________________________________________________________________________
This test allows to check the Lambda probe efficiency and the correct functioning of the
injection unit being connected to the probe.
• ANALYSIS
• SIMULATION
8.1 - Analysis
Two different types of lambda probe are currently being used: heated probe
(usually with 3 or 4 wires) and non-heated probe (usually with 1 or 2 wires).
In case of heated probe, before any operation, check the resistance and its power
supply using an ordinary multimeter, whereas in case of probe without heating
resistance, check the signal.
Once the type of probe has been identified, prepare the suitable connection in relation
to the type of vehicle being testes, as indicated on the following page.
• Connect the analyser to the Lambda probe, by means of the Lambda probe
extension cable (58) and the Shunt cable (57), or using other specific cables, if
needed.
Connections can be established both when the tool is turned off and on; in any case,
start the analyser for the preparation of the test.
Now warm up the engine as you would normally do by covering at least 3Km at a speed
such as to allow for the probe to attain the correct temperature, if non-heated, and
especially for both types of probes, to eliminate possible residues deposits due to
regular use; in some case, it is advisable to keep the engine accelerated, for
some minutes, between 2500 and 3000rpm during the test as well.
After warming up the engine as required, in the garage, immediately check the
efficiency of the lambda probe signal, through the connection previously prepared, and
insert the gas sampling probe into the vehicle’s exhaust pipe to check (alternatively) the
exhaust gas reading as well.
If the probe status is DEPLETED, immediately check the wiring through a simulation
test.
34 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
If the status is immediately signalled as EFFICIENT, you make sure this status is
maintained by carrying out other tests.
If efficiency decreases up to the DEPLETED condition, the possible causes can be the
following:
• Engine with relatively high unburnt hydrocarbons emissions (HC) which, often,
cannot be detected at the exhaust pipe due to the additional reduction by the
catalytic converter and/or the emission of carbon residues.
• Non-heated Lambda probe not attaining (at idling) a working temperature
sufficient to issue a correct signal to the electronic central unit.
• Lambda probe (heated) not working properly; using a multimeter, measure and
verify the voltage (12V) in parallel to the heater, as well as the correct resistance
value (2÷10Ω values when engine is cool) according to the type of probe.
NOTA BENE:
The lambda probe diagnosis is very accurate thanks to parameters developed through
tests and studies carried out for the German market during the AU type approval.
After the 10 seconds required to display the complete graph, F1 can be selected to
repeat the test.
OPERATING INSTRUCTIONS 35
STARGAS BOX ___________________________________________________________________________
8.2 - Diagnosis
The DIAGNOSIS function within the ANALYSIS test allows for the assessment of the
probe efficiency through a graphic indication.
• Select the DIAGNOSIS function, after the 10 seconds of the previous graph
(wave-shaped) display.
A bar indicating the probe condition is displayed, starting from the DEPLETED status up
and stopping at the point indicating that condition (DEPLETED –MEDIUM –EFFICIENT).
The space ranging between the depleted area and the efficient area is
regarded as medium.
8.3 - Simulation
Through this test, the analyser simulates the functioning of the Lambda probe,
replacing it.
The maximum duration of each test is 60 seconds, to ensure that the
ecu-catalytic converter system of the vehicle being tested is not damaged.
Through the simulation, the analyser checks the functioning of the ecu and of the
relevant wirings in the connection to the Lambda probe.
ATTENTION:
Improper use of the analyser can cause damage; carefully follow the
instructions.
• Select F2 SIM. λ < 1 to enrich the mixture after starting the test.
• Select F3 SIM. λ > 1 to weaken the mixture after starting the test.
36 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
8.4 - Rich mixture
On the LAMBDA PROBE SIMULATION page, in the box on the lower left-hand side, the
CO values and the lambda factor correspond to those measured by means of the gas
sampling probe.
The box indicating the lambda factor will change colour when the value is comprised
between 0,97 and 1,03 (ideal carburation).
The function λ > 1 allows for the simulation of a weakening of the mixture; in the
WEAK MIXTURE box, the CO values and the lambda factor are displayed.
The graph displays the curve transforming into a straight line tending to a value close
to 1 Volt. After about 15 seconds, by selecting the function λ > 1 once again, check the
correct variation of the CO values and lambda factor.
In the box on the lower left-hand side, make sure the normal working values of the
system have been restored, starting from the saved data in the WEAK MIXTURE box.
OPERATING INSTRUCTIONS 37
STARGAS BOX ___________________________________________________________________________
9.0 - CONFIGURATION
• Enter the garage data (the fields for the entry of the Garage Data are enabled only
if set on local mode in the Configuration Type) and set the language to be used for
printing and for the test.
ATTENTION:
The CONFIGURATION TYPE, MODE, LOGO and TEST PROCEDURE
functions are protected by a password and can thus be used
exclusively by the Technical Assistance personnel.
• Select the serial port used for the connection of the gas analyser.
38 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
• Select the RPM meter properties page to display the RPM COUNTER
CONFIGURATION page. To carry out a correct measure of the rpm number refer to
the use of the rpm counter (see chapter 3.3).
On this page, the selection of the type of gas analyser being used can be displayed.
ATTENTION:
The ANALYZER MODEL function, is protected by a password and can
thus be used exclusively by the Technical Assistance personnel.
• Select the License property in order to visualize and eventually to print the
relative contract.
ATTENTION:
The PROTECT SERVICES function, is protected by a password and can
thus be used exclusively by the Technical Assistance personnel.
OPERATING INSTRUCTIONS 39
STARGAS BOX ___________________________________________________________________________
The machinery automatically checks the state of the parameters being necessary for a
good operation.
The pump turning off is visualised on the display by the respective symbol.
The machinery automatically controls the gas flow by checking the vacuum value inside
the suction circuit.
Insert the probe into the analyser so as to guarantee that the flow control is
carried out correctly.
VACUUM LOW
VACUUM HIGH
NOTA BENE:
For a correct reading of the flow, never remove the tube (6 meters).
40 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
10.3 - Supply Voltage
The supply voltage is very important for the correct analyser operation.
In case the voltage exceeds the allowed limits (±15% of the supply voltage value), the
error is signalled by the displaying of two messages:
In case one of these messages is displayed while the test is being carried out, it is
necessary to repeat the test when the machinery supply voltage is regular again.
NOTA BENE:
If the supply voltage should keep being unstable or exceeding the allowed thresholds, it
is suggested to use a mains voltage stabiliser (provided with earthing), suitable for the
specific features that are written on the approval plate which is located on the back side
of the analyser.
OPERATING INSTRUCTIONS 41
STARGAS BOX ___________________________________________________________________________
11.0 - CONTROLS
F1-CALIBRATION
F2-LEAK TEST
F3-RESIDUAL HC
F4- SERIAL NUMBER
The values of the sample gases contained in the gas bottle must range between the
following concentrations:
Example:
NOTA BENE:
The P.E.F. value for this bench can vary. The bench is highly accurate since it complies
with the BAR 97 California standard and is OIML CLASS 0 type-approved.
42 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
11.2 - Gas bottle connection
The gas bottle must be connected to the coupler (43) which is located in the back side
of the analyser.
NOTA BENE:
It is compulsory to adjust the sample gas flow, by means of a reducer/adjuster, at a
value of 3,5 litres per minute in free air before connecting the pipe to the
connection (43), in order to prevent too a high pressure from damaging the analyser.
11.3 - Calibration
The calibration operation becomes necessary after ascertaining (through standard gas)
that the analyser does not carry out correct measurements.
During the test, bottle sample gases values are controlled. They are necessary for one
corrected calibration.
The leak test is necessary to make sure that in the whole analyser suction circuit there
are no air infiltrations, which could alter the gas analysis result.
1st TEST
• Introduce the sampling gas probe extremity (54) set up like gas pick-up system
(see chapter 2.0), in the leak test plug (34) which is located on the analyser back
side.
• Select LEAK TEST function from the CONTROLS page and press F1 to proceed.
The pump is on for 30 seconds, after which the vacuum electronic circuit checks the
tightness of the internal pneumatic system.
OPERATING INSTRUCTIONS 43
STARGAS BOX ___________________________________________________________________________
On the contrary, if a leakage is detected, the LEAK TEST NOT CORRECT message is
displayed.
2 st TEST
• Insert directly the leak test plug (34) directly into the gas inlet port (38).
In such a case check the gas sampling device conditions in order to eliminate the
external leakage.
On the contrary, if a leakage is detected, the LEAK TEST NOT CORRECT message is
displayed.
In case even after the second leakage test the LEAK TEST NOT CORRECT message is
displayed, go on according to the following instruction.
3 st TEST
• Carry out the leak test again according to the second test procedure.
• Introduce the second tightness plug in the out water (44) connection when the
pump stops operating.
Only if in this last case too the LEAK TEST NOT CORRECT message is displayed,
make sure that:
• The ring nuts of the condensate trap and of the standard cartridge/filter are
screwed tight
• The tightness gaskets of the filter plugs are fit in their slots and in good conditions.
44 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
11.5 - Residual HC
The RESIDUAL HC test allows to check if in the analyser pneumatic circuit there are
hydrocarbons which could alter the reading.
ATTENTION:
The gas sampling probe must not be introduced in the exhaust pipe.
After the AUTOZERO phase, the unburnt hydrocarbons value (HC in ppm) is measured.
If the analyser detects any impurities in the pneumatic circuit, the HC-TEST NOT
CORRECT or RESIDUAL HC-TEST NOT CORRECT message is displayed.
In case of negative result, please refer to the tool’s MAINTENANCE illustrative pages, in
the paragraphs concerning the cleaning operations on the pipe and gas sampling probe
through the replacement of the external filters.
OPERATING INSTRUCTIONS 45
STARGAS BOX ___________________________________________________________________________
12.0 - MAINTENANCE
The white cartridge/filter (15) cannot be cleaned but it must be replaced any time
it looks blackened or when the VACUUM HIGH message is displayed.
The condensate separating filter (36) has to be cleaned on average every two
replacements of the standard cartridge/filter (15) or whenever it is obstructed. In order
to clean it wash it with suds, hence rinse and dry it with compressed air. On the
contrary the separating filter has to be replaced if after the cleaning it appears to be
equally deteriorated.
NOTA BENE:
The wire gauze filter is provided with 2 different colour gaskets (blue/black) according
to the different filter sections where they have to be inserted.
When the filter is no longer in good conditions it is necessary to replace it with a new
one provided with new gaskets.
The actived carbon filter has to be replaced on average every two years.
This operation is easily carried out by extracting it out of the pipe to which it is
connected.
In the cases in which during function ZERO SET it did not aspire perfectly, it is
concurred the use of a small screwdriver in order to free the dirt plug that has been
deposited in the income of the filter.
ATTENTION:
Mount the new filter with the arrow towards the ZERO GAS IN connection.
46 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
12.4 - Replacement of the O2 sensor
The external transparent precleaner (52) must be cleaned or even replaced if the
cleaning is no longer enough. To clean the precleaner wash it with suds, hence rinse
and dry with compressed air.
The cleaning of the pick-up pipe has to be periodically carried out to clear the pipe from
possible carbon residues or condensate which have deposited inside of it.
Before blowing down in the pick-up pipe disconnect it from the connection.
ATTENTION:
Do not blow compressed air in the analyser.
The filtering baffle (51) allows to the analyser to work in dusty environments
preventing this working condition from causing troubles to the analyser correct
operation in the long term.
One cleaning periodic is advised to be carried out with water and soap. In case that
also after the washing is not succeeded to clean up the filter is necessary to replace it.
In order to replace the filtering baffle carry out the following procedure:
• Remove the filter-holder (13) on the back of the STARGAS BOX, to the inside of
which the filter (51) is positioned.
• Place again the filter to the inside of the filter-holder and subsequently insert the
all in the appropriate lodging making pressure with the hands.
OPERATING INSTRUCTIONS 47
STARGAS BOX ___________________________________________________________________________
CONTROL UNIT
Power supply:
Supply options:
Mains 90 ÷ 270V, 50 ÷ 60Hz
Battery 10 ÷ 16 Volt (5 [A] F fuse).
Mains supply voltage (110/220/240 Volt), automatic indication with print out of the
error code for variations > -15% +10%.
Max power 70W.
Serial port:
COM1 - COM2 for connection to the autodiagnosis interface
SMOKEMETER for connection to partial flow smoke cell combined with external
supplier
standard serial port RS232/PC for connection to a PC or to other diagnosis
stations
RS485 standard industrial serial connection
Printer port:
PARALLEL for connection to 80-column standard printer (colour too)
USB for connection to 80-column standard printer
Socket video:
VGA for connection to standard monitor or to a TV color (PAL - NTSC system)
Socket COM:
grounding of the tested vehicle (battery negative pole) for revolution counter
signals from battery, lambda prove and oscilloscope
48 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
Operation temperature:
+5°C ÷ +40°C
Stocking temperature:
Min. -25 Max. +70°C
Dimensions:
400 × 180 × 450mm
Weight:
8,6kg (approximately)
GAS ANALYSER
Measuring range:
CO 0 ÷ 15,000 % Vol. res. 0,001
CO2 0 ÷ 20,00 % Vol. res. 0,01
HC 0 ÷ 30000 ppm Vol. res. 1
O2 0 ÷ 25,00 % Vol. res. 0,01
NO 0 ÷ 5000 ppm Vol. res. 1 (optional)
Lambda 0,5 ÷ 2,000 [-] res. 0,001
Thermometer:
5 ÷ 200°C res. 1
Flow control:
Internal and automatical
Leak test:
Automatical
OPERATING INSTRUCTIONS 49
STARGAS BOX ___________________________________________________________________________
Condensate drainage:
Continual and automatical
Response time:
<10 sec. (Probe length 3m.)
Heating time:
Max. 60 seconds
Zero setting:
Electronic and automatical
Calibration:
Electronic and automatical (only with the gas bottle the analyser is equipped)
Socket λ:
socket allowing:
Lambda probes analysis (1 Volt/5 Volt) with performance automatic calculation
Simulation to evaluate the correct operation of the control unit wiring
50 OPERATING INSTRUCTIONS
__________________________________________________________________________ STARGAS BOX
OPERATING INSTRUCTIONS 51
STARGAS BOX ___________________________________________________________________________
The optional cables are equipped with wiring for the 12V supply voltage and Lambda
probe heating resistance control.
To carry out these two measures (Volt-Ohm) use any multimeter.
52 OPERATING INSTRUCTIONS