1
Cathode Ray Oscilloscope(CRO)
Mr.P.Krishna, EEE Department, IIITN       19-Feb-20
                Cathode Ray Oscilloscope(CRO)
                                      2
 Very useful and versatile laboratory instrument (diagnostic
  lab equipment).
 Used for display, measurements and analysis of waveforms
  (signals) in electrical and electronic circuits at frequencies
  ranging from dc to GHz.
 Basic instrument for the study of all types of signals.
 Very fast in displaying/plotting (X-Y plotters).
 Capable of generating a graph of an input signal versus time
  (voltage-time mode) or a second variable (X-Y mode).
 Can be employed to measure quantities such as peak
  voltage, frequency, period, phase difference, pulse width etc.
Mr.P.Krishna, EEE Department, IIITN                      19-Feb-20
                                  Parts of a CRO:
                                           3
1) Cathode Ray Tube (CRT) and its associated controls.
                       An electron gunI.
                   II. Deflection plates
               III. Phosphorescent screen
2) Input circuitry (a time-base generator, vertical and
   horizontal amplifiers).
Mr.P.Krishna, EEE Department, IIITN                 19-Feb-20
            Simple CRO construction:
                                      4
Mr.P.Krishna, EEE Department, IIITN       19-Feb-20
                         CRO construction:
                                      5
Mr.P.Krishna, EEE Department, IIITN          19-Feb-20
   Block diagram of a basic CRO :
                                      6
Mr.P.Krishna, EEE Department, IIITN       19-Feb-20
  Operation: Vertical Deflection Plates(VDP)
                                      7
 Electron    beam generated from electron gun will be
  accelerated towards phosphorous screen.
 Beam passes through a set of parallel plates (vertical
  deflection plates) oriented horizontally.
 Voltage to be displayed is amplified by a vertical amplifier
  and produce an electric field which deflects the path of the
  electron vertically.
 The polarity of the signal determines whether the
  deflections will be up or down.
 Magnitude of the signal determines the vertical
  displacement of the electron.
Mr.P.Krishna, EEE Department, IIITN                     19-Feb-20
  Operation: Horizontal Deflection Plates(HDP)
                                      8
 After passing through vertical deflection plates, the beam
  passes through a set of similar plates (horizontal
  deflection plates) which are oriented vertically.
 The potential difference applied to these plates produce an
  electric field which deflects left or right.
 Under normal configuration these HDP are connected to a
  time-base circuitry.
 This circuit can control how fast the electron beam
  sweeps from the left to right
Mr.P.Krishna, EEE Department, IIITN                    19-Feb-20
                                      Operation:
                                          9
 If two deflection voltages were held constant, the electron beam would
  strike a point on the phosphorescent film and a stationary point would
  visible on the screen.
 Most voltages of interested are time-varying and so the voltage applied
  to the HDP is varied with time in such a way that the spot moves from
  left to right on the screen as time passes.
 The phosphorescent material has the property of emitting light for
  several milliseconds after the electrons have passed, the total effect is
  for the electrons to leave behind a visible trail –a time-varying signal.
 The horizontal deflection voltage is also varied in such a way that
  when the beam reaches the right-hand edge of the screen, it starts over
  at the left-right side.
Mr.P.Krishna, EEE Department, IIITN                                19-Feb-20
               Displaying a signal on the screen:
                                      10
Mr.P.Krishna, EEE Department, IIITN                 19-Feb-20
               Displaying a signal on the screen:
                                      11
Mr.P.Krishna, EEE Department, IIITN                 19-Feb-20
                                CRO in X-Y Mode:
                                       12
 Beside showing a time-varying signal, the oscilloscope
  lets you display an input signal rather than the time base
  on the horizontal axis. This is called X-Y mode as both
  the X and Y axis are tracing input voltage.
 The waveform resulting from XY arrangement of two
  periodic signals of different periods is called Lissajous
  Pattern.
 From the shape of Lissajous pattern we can determine
  relative phases of the signals as well as frequency ratio.
Mr.P.Krishna, EEE Department, IIITN                   19-Feb-20
                                      CRO controls:
                                            13
1. Y-Gain: Amplifies Y-deflection (small voltages are
   amplified by built-in amplifiers before applying Y-plate)
                    Y-Gain=0.5V/Div
2. Time-Base: A saw-tooth wave is applied internally
   across X-plates.
Mr.P.Krishna, EEE Department, IIITN                   19-Feb-20
    CRO control:                              2.Time-Base
                                      14
 Controls the speed at which the spot sweeps across the
    screen horizontally from left to right.
Mr.P.Krishna, EEE Department, IIITN                   19-Feb-20
    CRO control:                           2.Time-Base
                                      15
Mr.P.Krishna, EEE Department, IIITN                19-Feb-20
    CRO control:                           2.Time-Base
                                      16
 Helps to display the actual waveform applied across Y-
  plates.
 Calibrated in s/cm or ms/cm or µs/cm.
 Gives the time required for the spot to sweep 1cm
  horizontally across the screen.
Mr.P.Krishna, EEE Department, IIITN                19-Feb-20
         Most Common Oscilloscope Controls:
                                      17
Mr.P.Krishna, EEE Department, IIITN           19-Feb-20
                      Measurement of frequency:
                                      18
Mr.P.Krishna, EEE Department, IIITN               19-Feb-20
           Measurement of phase difference:
                                      19
Mr.P.Krishna, EEE Department, IIITN           19-Feb-20
                                  Lissajous Figures:
                                          20
 Can    be displayed by applying two a.c signals
  simultaneously to X-plates and Y-plates of a CRO.
 As the frequency, amplitude and phase difference are
  altered, different patterns are seen on the screen of the
  CRO.
 Obtained when operated in X-Y mode.
 If the ratio of the two frequencies of the voltages is just
  equal to a rational number, standing figures appear on the
  oscilloscope (Lissajous’ figures); if the frequency ratio,
  however, deviates slightly from a rational number, these
  figures are moving.
Mr.P.Krishna, EEE Department, IIITN                    19-Feb-20
 Measurement of phase difference using Lissajous Pattern:
                                      21
Obtained by applying two sinusoidal signals to
 be compared at the vertical and horizontal plates.
The phase difference then measured from
 resulting Lissajous pattern.
The pattern may be either a straight line or a
 circle or an ellipse depending on the amount of
 phase difference.
Mr.P.Krishna, EEE Department, IIITN                 19-Feb-20
Measurement of phase difference using Lissajous Pattern:
                                      22
Mr.P.Krishna, EEE Department, IIITN                19-Feb-20
Measurement of phase difference using Lissajous Pattern:
                                      23
Mr.P.Krishna, EEE Department, IIITN                19-Feb-20
Measurement of phase difference using Lissajous Pattern:
                                      24
Mr.P.Krishna, EEE Department, IIITN                19-Feb-20
       Measurement of frequency using Lissajous Pattern:
                                      25
Mr.P.Krishna, EEE Department, IIITN                   19-Feb-20
       Measurement of frequency using Lissajous Pattern:
                                      26
Mr.P.Krishna, EEE Department, IIITN                   19-Feb-20
       Measurement of frequency using Lissajous Pattern:
                                      27
Mr.P.Krishna, EEE Department, IIITN                   19-Feb-20
                         Storage type oscilloscope:
                                      28
 Used for presentation of very slowly swept signals.
 Used in mechanical and biomedical fields.
 In conventional CRO the persistence of the screen is
  smaller than the rate at which the signal sweeps across the
  screen (the display will disappear within µs to seconds).
 The persistence of the screen can be adjusted to match
  with the sweep time. So the display stays for few seconds
  or even hours.
 A special CRT called storage tube is used which is same
  as conventional CRT but with additional number of
  special electrodes.
Mr.P.Krishna, EEE Department, IIITN                     19-Feb-20
                         Storage type oscilloscope:
                                      29
1) Two special guns called flood guns
2) Collimation electrodes
3) Collector mesh
4) Storage mesh or storage target
Mr.P.Krishna, EEE Department, IIITN                   19-Feb-20
                                      Example1:
                                          30
 Horizontal frequency=50Hz
 Vertical frequency=?
Mr.P.Krishna, EEE Department, IIITN               19-Feb-20
                                      Example2:
                                          31
 Phase difference=?
Mr.P.Krishna, EEE Department, IIITN               19-Feb-20