Mercury: Cold Vapor Mercury Concentration Method Method 10065 0.1 To 2.5 G/L HG
Mercury: Cold Vapor Mercury Concentration Method Method 10065 0.1 To 2.5 G/L HG
01059
Test preparation
Instrument-specific information
Table 1 shows all of the instruments that have the program for this test. The table also
shows sample cell and orientation requirements for specific instruments.
To use the table, select an instrument, then read across to find the applicable information
for this test.
Table 1 Instrument-specific information
Instrument Sample cell orientation Sample cell
DR 6000 The fill line is to the right. 2495402
DR 3800
DR 2800
DR 2700
DR 1900
DR 5000 The fill line is toward the user.
DR 3900
Before starting
The test can release toxic chlorine or other gases. Do the test procedure in a fume hood.
Use dedicated digestion glassware and sample cells for this procedure.
Determine a reagent blank for each new lot of reagent: complete the procedure, including the digestion, with 1 liter of
deionized water instead of sample; add the same amount of potassium permanganate as required by the sample; subtract
the reagent blank value from the final results or complete a reagent blank adjust.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective
equipment.
Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal
information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.
Items to collect
Description Quantity
Refer to Consumables and replacement items on page 11 for a complete list of required
–
apparatus
Cold Vapor Mercury Apparatus Set 1
Cold Vapor Mercury Reagent Set (refer to Consumables and replacement items on page 10) 1
Digestion Reagents and Apparatus (refer to Consumables and replacement items
varies
on page 12)
Sample cells (For information about sample cells, adapters or light shields, refer to Instrument-
2
specific information on page 1.)
1
Refer to Consumables and replacement items on page 10 for order information.
Sample collection
• Collect 1000 mL of sample in an analytically clean, glass or polyethylene
terephthalate (PET) container.
• To preserve samples for later analysis, add 10 mL of concentrated hydrochloric acid
to the sample container before collection.
Note: Close the glass container with a ground glass stopper. Close a PET container with a
PET cap or a polypropylene cap (no liner).
• Fill the bottle completely full, then tighten the cap on the bottle.
• Keep the preserved samples at 2–6 °C (35.6–43 °F) for a maximum of 6 months.
• Correct the test result for the dilution caused by the volume additions.
System start-up
For more accurate results and system equilibration, complete several analyses on
mercury standards and blanks before the sample testing. This allows the system to
stabilize before processing samples.
Standard start-up
1. Follow the procedure Standard solution method on page 9. If the value is not
within the specified limits, continue to the next step.
2. Use a pipet to add 10.0 mL of the 0.1-mg/L mercury standard solution into the purged
solution in the Gas Washing Bottle. Immediately put the stopper in the Gas Washing
Bottle.
3. Use the test procedure to measure the concentration of the standard. Start at step 3
of phase 2.
4. Test the eluate as described in phase 3. The concentration must be 0.9–1.1 µg/L Hg.
Do steps 1 to 3 again if the value is not within these limits.
Blank start-up
After a satisfactory Standard start-up is complete, use the purged solution in the Gas
Washing Bottle to do a system Blank start-up.
1. Keep the purged solution in the Gas Washing Bottle. Do not add an aliquot of
mercury standard.
2. Use the test procedure to measure the concentration of the sample. Start at step 3 of
phase 2.
3. Test the eluate as described in phase 3. The concentration must be ≤ 0.2 µg/L Hg.
Do the Blank start-up procedure again until a reproducible value is shown.
WARNING
Gas inhalation hazard. Operate the instrument in a fume hood to prevent exposure to
hazardous gas.
CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to the
current safety data sheets (MSDS/SDS) for safety protocols.
5. Add 4.0 g of potassium 6. Add 7.5 g of potassium 7. Use a watch glass as the 8. Continue to stir and keep
persulfate to the sample. permanganate to the flask cover. After the the sample temperature at
Alternatively, add one sample. reagents dissolve, increase 90 °C for 2 hours.
5 gram measuring scoop of Alternatively, add a 10 gram the temperature of the The solution must stay dark
potassium persulfate to the measuring scoop of sample to 90 °C (194 °F). purple during the entire
sample. potassium permanganate to Do not boil. digestion. Some samples
the sample. Note: It is not necessary to (e.g., seawater, industrial
Stir until dissolved.
increase the temperature for effluents or samples that are
Stir until dissolved.
a mercury standard or high in organic matter or
reagent blank in distilled chloride) require additional
water. permanganate. It can be
difficult to see a dark purple
color if the sample contains
black/brown manganese
dioxide precipitate. Add
more potassium
permanganate if the solution
is not dark purple.
WARNING
Gas inhalation hazard. Operate the instrument in a fume hood to prevent exposure to
hazardous gas.
1. Pour the digested 2. Set the Gas Washing 3. Connect the 100-mL 4. Pipet 8 mL of HgEx
sample into the Cold Vapor Bottle in the support ring. Erlenmeyer flask to the Reagent B into the Mercury
Gas Washing Bottle. Place the top on the Gas mercury absorber column. Absorber column.
Note: The volume of the Washing Bottle. Wait until
digested sample must step 9 to connect the
contain 0.1 to 2.5 µg Hg. mercury absorber column to
the Gas Washing Bottle.
9. Use the glass elbow to 10. Shake an ampule of 11. Put the stopper on the 12. Use the quick
connect the Mercury HgEx Reagent A to suspend side neck of the Glass disconnect to connect the
Absorber column to the Gas undissolved reagent. Washing Bottle. vacuum pump to the
Washing Bottle. Open the ampule and Mercury Absorber Column
carefully pour the contents again. Apply the vacuum to
into the Gas Washing Bottle pull HgEx Reagent C
through the side neck. through the Mercury
Absorber Column packing
and into the 10-mL receiver.
Air bubbles should be
produced at the gas
dispersion tube in the Gas
Washing Bottle. Complete
the next two steps
immediately.
13. Start program 14. Start the instrument 15. After the timer expires, 16. Pipet 8 mL of HgEx
312 Mercury, Cold Vap. timer. A 5-minute reaction remove the glass elbow Reagent B into the Mercury
For information about time starts. Let the solution from the top of the Mercury Absorber Column to elute
sample cells, adapters or bubble for this period. Absorber Column. Keep the the captured mercury.
light shields, refer to The air flow rate through the vacuum pump power on. Continue to apply the
Instrument-specific Gas Washing Bottle should vacuum to pull the HgEx
information on page 1. be between 1-5 L/min. Let Reagent B into the Distilling
Note: Although the program the solution bubble for more Receiver.
name can be different time when the air flow rate is
between instruments, the low. For example, if the air
program number does not flow rate is 1 L/min., let the
change. solution bubble for
10 minutes.
17. Set the vacuum pump 18. Remove the distilling 19. Use a pipette to add
to off when the volume in Receiver from the Mercury 3 mL of HgEx Reagent B
the Distilling Receiver is at Absorber Column. Connect into the Mercury Absorber
the 10-mL mark. the 100-mL Erlenmeyer Column without applying
If necessary, adjust the flask to the column again. vacuum. This keeps the
volume in the Distilling absorber packing wet
Receiver a maximum of between tests.
10 mL with HgEx Reagent The Mercury Absorber
B. Column eluate in the
To prevent low volumes in Distilling Receiver is ready
the future, disconnect the for analysis.
vacuum sooner in step 6. Go to Phase 3: Colorimetric
This leaves more HgEx analysis on page 7.
Reagent B in the packing of
the Mercury Absorber
Column.
WARNING
Gas inhalation hazard. Operate the instrument in a fume hood to prevent exposure to
hazardous gas.
1. Use the supplied funnel 2. Put the stopper on the 3. Use the supplied funnel 4. Put the stopper on the
to add the contents of one receiver. Invert to dissolve to add the contents of one receiver. Invert to dissolve
HgEx Reagent 3 foil pillow the reagent. HgEx Reagent 4 foil pillow the reagent.
to the eluate in the Distilling to the Distilling Receiver.
Receiver.
5. Add 8 drops of HgEx 6. Put the stopper on the 7. Start the instrument 8. During the reaction
Reagent 5 to the Distilling receiver. Invert to mix the timer. A 2–minute reaction period, pour the solution into
Receiver. reagent. time starts. a sample cell.
Zero
9. Clean the prepared 10. Insert the sample cell 11. Push ZERO. The 12. Remove the cell from
sample cell. into the cell holder. display shows 0.1 µg/L Hg the cell holder. Add the
(this program uses a non- contents of one HgEx
zero intercept). Reagent 6 foil pillow to the
solution.
Note: Do not use the funnel
to add HgEx Reagent 6 to
the sample cell. HgEx
Reagent 6 contamination
from the funnel will make it
impossible to find mercury in
subsequent tests.
13. Swirl the cell until the 14. Clean the prepared 15. Insert the prepared 16. Push READ. Results
reagent is completely sample cell. sample into the cell holder. show in µg/L Hg. This is the
dissolved. Immediately concentration in the original
continue with next step . sample.
Interferences
Standards were used to prepare a single test solution with substances at the
concentrations shown in Table 2. A second test solution containing only mercury at the
same concentration was prepared as the control. The two solutions were digested, then
analyzed concurrently. There was no interference from the matrix of the test solution at
the concentrations listed.
In addition, no interference occurred with a test solution containing 1000 mg/L Na+,
1000 mg/L K+, 1000 mg/L Mg2+ and 400 mg/L Ca2+.
Table 2 Interfering substances
Interfering substance Interference level
Ag+ 7 mg/L
Al3+ 10 mg/L
Au3+ 500 µg/L
Cd2+ 10 mg/L
Co2+ 10 mg/L
Cr6+ 10 mg/L
Cu2+ 10 mg/L
F– 1.0 mg/L
Fe2+ 100 mg/L
Mo6+ 10 mg/L
Ni2+ 10 mg/L
NO3––N 50 mg/L
Pb2+ 10 mg/L
SiO2 100 mg/L
Zn2+ 10 mg/L
Accuracy check
Standard additions method
Use the standard additions method (for applicable instruments) to validate the test
procedure, reagents and instrument and to find if there is an interference in the sample.
Items to collect:
• 1000-mg/L Mercury Standard Solution
Required apparatus
Recommended standards
© Hach Company/Hach Lange GmbH, 2007, 2010, 2012, 2014. All rights reserved. 03/2014, Edition 8