Instruction Manual                                                   Dear Customer,
Thank you for choosing a Hanna product. This manual will                       SPECIFICATIONS                                 DISPLAY CODE GUIDE
            HI 93727                                                             provide you with the necessary information for the correct
                                                                                                                                                                                                                  This indicates that the meter is in a
                                                                                 operation of the meter. Please read it carefully before using
       Color of Water ISM                                                        the meter. If you need additional technical information, do                                                                      ready state and zeroing can be per-
                                                                                 not hesitate to e-mail us at tech@hannainst.com.                                                                                 formed.
                                                                                 This instrument is in compliance with              directives                                                                    Sampling in Progress. This prompt ap-
                                                                                 EN 50081-1 and EN 50082-1.                                                                                                       pears each time the meter is performing
                                                                                                                                                                                                                  a measurement.
                                                                                      PRELIMINARY EXAMINATION                                                                                                     This indicates that the meter is in a
                                                                                                                                                                                                                  zeroed state and measurement can be
                                                                                 Remove the instrument from the packing material and                                                                              performed.
                                                                                 examine it carefully to make sure that no damage has                                                                             A zero reading was not taken. Insert a
                                                                                 occurred during shipment. If there is any damage, notify                                                                         sample before adding reagent and press
                                                                                 your Dealer.                                                                                                                     ZERO.
                                                                                 Each Ion Specific Meter is supplied complete with                                                                                Under range. A blinking "0.00" indicates
                                                                                           • 9V Battery                                                                                                           that the sample absorbs less light than
                                                                                           • Three Sample Cuvets and Caps                                                                                         the zero reference. Check the procedure
                                                                                           • One Transport Cap                                                                                                    and make sure you use the same cuvet
                                                                                 Note: Conserve all packing material until the instrument        SPECIFICATIONS                                                   for reference (zero) and measurement.
                                                                                        has been observed to function correctly. Any defective                                                                    Over range. A flashing value higher than
                                                                                                                                                 Range           0 to 500 PCU (Platinum Cobalt Units)
                                                   This Instrument is in                item must be returned in its original packing.                                                                            the maximum concentration readable (see
http://www.hannainst.com                     Compliance with the CE Directives                                                                   Resolution      10 PCU
                                                                                                                                                 Accuracy        ±10 PCU ± 5% of reading                          specifications) indicates that the sample
                                                                                                                                                 Typical EMC     ±10 PCU                                          absorbs too much light, meaning that
                                                                                          GENERAL DESCRIPTION                                    Deviation                                                        the concentration is too high. Dilute the
                                                                                                                                                 Light Source    Light Emitting Diode @ 470 nm                    sample.
                         WARRANTY
                                                                                 The HI 93727 meter measures the true and apparent color         Method          Adaptation of the Colorimetric Platinum          Light over range. The cuvet is not in-
HI 93727 is warranted for two years against defects in workman-                  in water and wastewater in the 0 to 500 PCU (Platinum                           Cobalt method                                    serted correctly and an excess ambient
ship and materials when used for its intended purpose and maintained             Cobalt Units) range.                                            Light Detector Silicon Photocell                                 light is reaching the detector. If the cover
according to instructions.                                                       The meter uses an exclusive positive-locking system to ensure   Environment 0 to 50°C (32 to 122°F);                             is properly installed, then contact your
This warranty is limited to repair or replacement free of charge.                that the cuvet is in the same place every time it is placed                     max 95% RH non-condensing                        dealer or the nearest Hanna Customer
Damages due to accident, misuse, tampering or lack of prescribed                                                                                 Battery Type/Life 1 x 9 volt/40 hours
maintenance are not covered.
                                                                                 into the measurement cell.                                                                                                       Service Center.
                                                                                 Display codes aid the user in routine operations.               Auto-Shut off After 10' of non-use                               Light under range. The zero sample is
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur-                                                                             Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8")                   too dark for proper zeroing. If this is not
                                                                                 The meters have an auto-shut off feature that will turn the     Weight          290 g (10 oz.).
chase, serial number and the nature of the failure. If the repair is not                                                                                                                                          the case, contact your dealer or the
covered by the warranty, you will be notified of the charges incurred. If        instrument off after 10 minutes of non-use.
the instrument is to be returned to Hanna Instruments, first obtain a                                                                            REQUIRED ACCESSORIES                                             nearest Hanna Customer Service Center.
Returned Goods Authorization Number from the Customer Service de-                                                                                0.45 µm membrane for true color measurement.              V      The "V" indicates that the battery volt-
partment and then send it with shipment costs prepaid. When shipping                                                                                                                                              age is getting low and the battery needs
any instrument, make sure it is properly packaged for complete protec-
tion.
                                                                                                                                                                                                                  to be replaced.
To validate your warranty, fill out and return the enclosed warranty card                                                                                                                                         This indicates that the battery is dead
within 14 days from the date of purchase.                                                                                                                                                                         and must be replaced.
All rights are reserved. Reproduction in whole or in part is prohibited                                                                                                                                           Note: once this indication is displayed,
without the written consent of the copyright owner, Hanna Instru-                                                                                                                                                 the meter will lockup. Change the bat-
ments Inc., Woonsocket, Rhode Island, 02895 , USA.
                                                                                                                                                                                                                  tery to restart.
 Hanna Instruments reserves the right to modify the design, construction
 and appearance of its products without advance notice.
                                                       • Remove the blank, insert the
           OPERATIONAL GUIDE                             apparent color cuvet into the                                           TIPS FOR AN ACCURATE                                                            ACCESSORIES
                                                         instrument and ensure that                                                   MEASUREMENT                                     OTHER ACCESSORIES
MEASUREMENT PROCEDURE                           ON       the notch on the cap is posi-
                                                                                                                       The instruction listed below should be carefully followed      HI 710009             Blue rubber boot
• Turn the meter on by pressing ON/OFF.        OFF       tioned securely into the
                                                                                                                       during testing to ensure best accuracy.                        HI 710010             Orange rubber boot
                                                         groove.
                                                                                                                       • Do not touch the cuvet walls with hands.                     HI 721310             9V battery (10 pcs)
• When the LCD displays "- - -", it is                 • Press READ and "SIP" will appear on the display.
  ready.                                                                                                               • In order to maintain the same conditions during the          HI 731318             Tissue for wiping cuvets (4 pcs)
                                                                                                                           zeroing and the measuring phases, it is necessary to       HI 731321             Glass cuvets (4 pcs)
                                             1.5 cm                                                                        close the cuvet to prevent any contamination.              HI 731325             Caps for cuvets (4 pcs)
• Fill one cuvet up to 1.5 cm (¾")
                                                                                                                       • Do not let the test sample stand too long after reagent      HI 93703-50           Cuvets cleaning solution (230 mL).
  below the rim with deionized water
                                                                                                                           is added or accuracy will be lost.
  and replace the cap. This is the
  blank.                                               • The meter directly displays the value of apparent color.      • Whenever the cuvet is placed into the measurement cell,
                                                       • Remove the cuvet, insert the                                      it must be completely free of fingerprints, oil or dirt.
• Fill another cuvet up to 1.5 cm (¾")
                                             1.5 cm
                                                         true color cuvet into the in-                                     Wipe it thoroughly with HI 731318 or a lint-free cloth
  below the rim with unfiltered sample
                                                         strument and ensure that                                          prior to insertion.
  and replace the cap. This is the
  apparent color.                                        the notch on the cap is posi-                                 • It is important that the sample does not contain any
                                                         tioned securely into the                                          debris. This would corrupt the readings.
                                                         groove.                                                       • It is possible to take multiple readings in a row, but it                          CE DECLARATION
• Filter 50 mL of sample through a                     • Press READ and "SIP" will appear on the display.                  is recommended that a zero reading be taken for each                             OF CONFORMITY
                                                                                                                           sample and that the same cuvet is used for zeroing and
  filter with a 0.45 µm membrane.                                                                                          measurement.
                                                                                                                       • It is important to discard the sample immediately after
                                                                                                                           the reading is taken because the glass might become
• Fill the third cuvet up to 1.5 cm           1.5 cm
                                                                                                                           permanently stained.
  (¾") below the rim with the filtered                                                                                 • Shaking the cuvet can generate bubbles in the sample,
                                                       • The meter directly displays concentration in color units of
  sample and replace the cap. This the                                                                                     causing higher readings. To obtain accurate measure-
                                                         true color on the Liquid Crystal Display.
  true color.                                                                                                              ments, remove such bubbles by swirling or by gently
                                                       INTERFERENCES                                                       tapping the vial.
• Place the blank into the holder                      There are no known interferences for this method.               • All the reaction times reported in this manual are
  and ensure that the notch on                                                                                             referred to 20°C (68°F). As a general rule of thumb,
  the cap is positioned securely                                                                                           they should be doubled at 10°C (50°F) and halved at
  into the groove.                                                                                                         30°C (86°F).
• Press ZERO and "SIP" will appear on the display.                                                                              BATTERY REPLACEMENT
                 ZERO                                                                                                  Battery replacement must only take place in a non-hazard-
                                                                                                                       ous area using a 9V alkaline battery.
                                                                                                                       Simply slide off the battery cover on the back of the meter.   Recommendations for Users
                                                                                                                                                                                      Before using these products, make sure that they are entirely suitable for the environ-
• Wait for a few seconds and the dis-                                                                                  Detach the battery from the terminals and attach a fresh       ment in which they are used.
  play will show "-0.0-". Now the meter                                                                                9V battery while paying attention to the correct polarity.     Operation of these instruments in residential area could cause unacceptable interferences
  is zeroed and ready for measure-                                                                                     Replace the battery and the cover.                             to radio and TV equipments, requiring the operator to take all necessary steps to correct
                                                                                                                                                                                      interferences.
  ment.                                                                                                                                                                               Any variation introduced by the user to the supplied equipment may degrade the
                                                                                                                                                                                      instruments' EMC performance.
                                                                                                                                                    9V
                                                                                                                                                                                      To avoid damages or burns, do not perform any measurement in microwave ovens.
                                                                                                                                                             SLIDE
                                                                                                                                                              OFF
                                                                                                                                                                                       PRINTED IN PORTUGAL                        IST93727R2                       01/00