Impulse 7000DP Upgrade
Instruction Sheet
Introduction
The Impulse 7010 Selectable Defibrillator Load
Upgrading Analyzer Firmware
will only work with an Impulse 7000DP If the firmware version of your Analyzer is lower
Defibrillator Analyzer (herafter the Analyzer) than 1.04, then it will need to be upgraded to
running firmware version 1.04 or later. This operate with an Impulse 7010 Selectable Load.
instruction sheet describes how to check the The necessary firmware is included on the
Firmware version of the Analyzer and, if accompanying CD (PN 3351841).
necessary, how to upgrade it with the latest For further instructions on updating the Analyzer
firmware. firmware, insert the accompanying CD into a PC
and open the CD file labeled “Impulse
Checking the Firmware Version 6000D7000DP Firmware Update.pdf”. These
instructions require running the AnsurMLC utility.
There are two ways to check for the Firmware AnsurMLC is a separate program installed with
Version number of the Analyzer: at power up and Ansur. It should be on your Start Menu at
through the menu of the Analyzer. Programs > Fluke > Ansur > AnsurMLC. If the
If the Analyzer is off, press the power button (O) Ansur software is not already installed on your
on the top panel. During power up, the screen PC, refer to the CD that came with the Impulse
shown in Figure 1 will be displayed. The 7000DP. If the Ansur version is not 2.5.0 or later,
firmware version is displayed along the bottom of then download the Ansur software from the web
the display. site at www.flukebiomedical.com.
fak150.eps
Figure 1. Analyzer Power-Up Screen
If the Analyzer is already on, press Qto open
the setup main menu. Press the softkey labeled
More to reveal additional setup selections. Press
the softkey labeled Instrument Info to display
the screen shown in Figure 2.
fak24.eps
Figure 2. Analyzer Information Screen