Designation: D2892 – 11a
Standard Test Method for
                      Distillation of Crude Petroleum (15-Theoretical Plate
                      Column)1
                      This standard is issued under the fixed designation D2892; the number immediately following the designation indicates the year of
                      original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
                      superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
  1. Scope*                                                                                          1.4.5 Annex A5—Test Method for Determination of the
     1.1 This test method covers the procedure for the distillation                               Temperature Response Time,
  of stabilized crude petroleum (see Note 1) to a final cut                                          1.4.6 Annex A6—Practice for the Calibration of Sensors,
  temperature of 400°C Atmospheric Equivalent Temperature                                            1.4.7 Annex A7—Test Method for the Verification of Reflux
  (AET). This test method employs a fractionating column                                          Dividing Valves,
  having an efficiency of 14 to 18 theoretical plates operated at a                                  1.4.8 Annex A8—Practice for Conversion of Observed
  reflux ratio of 5:1. Performance criteria for the necessary                                     Vapor Temperature to Atmospheric Equivalent Temperature
  equipment is specified. Some typical examples of acceptable                                     (AET),
  apparatus are presented in schematic form. This test method                                        1.4.9 Appendix X1—Test Method for Dehydration of a
  offers a compromise between efficiency and time in order to                                     Sample of Wet Crude Oil, and
  facilitate the comparison of distillation data between laborato-                                   1.4.10 Appendix X2—Practice for Performance Check.
  ries.                                                                                              1.5 The values stated in SI units are to be regarded as the
                                                                                                  standard. The values given in parentheses are for information
    NOTE 1—Defined as having a Reid vapor pressure less than 82.7 kPa                             only.
  (12 psi).
                                                                                                     1.6 This standard does not purport to address all of the
     1.2 This test method details procedures for the production of                                safety concerns, if any, associated with its use. It is the
  a liquefied gas, distillate fractions, and residuum of standard-                                responsibility of the user of this standard to establish appro-
  ized quality on which analytical data can be obtained, and the                                  priate safety and health practices and determine the applica-
  determination of yields of the above fractions by both mass and                                 bility of regulatory limitations prior to use. For specific
  volume. From the preceding information, a graph of tempera-                                     warning statements, see Section 10.
  ture versus mass % distilled can be produced. This distillation
  curve corresponds to a laboratory technique, which is defined                                   2. Referenced Documents
  at 15/5 (15 theoretical plate column, 5:1 reflux ratio) or TBP                                    2.1 ASTM Standards:2
  (true boiling point).                                                                              D941 Test Method for Density and Relative Density (Spe-
     1.3 This test method can also be applied to any petroleum                                        cific Gravity) of Liquids by Lipkin Bicapillary Pycnom-
  mixture except liquefied petroleum gases, very light naphthas,                                      eter3
  and fractions having initial boiling points above 400°C.                                           D1217 Test Method for Density and Relative Density (Spe-
     1.4 This test method contains the following annexes and                                          cific Gravity) of Liquids by Bingham Pycnometer
  appendixes:                                                                                        D1298 Test Method for Density, Relative Density (Specific
     1.4.1 Annex A1—Test Method for the Determination of the                                          Gravity), or API Gravity of Crude Petroleum and Liquid
  Efficiency of a Distillation Column,                                                                Petroleum Products by Hydrometer Method
     1.4.2 Annex A2—Test Method for the Determination of the                                         D2887 Test Method for Boiling Range Distribution of
  Dynamic Holdup of a Distillation Column,                                                            Petroleum Fractions by Gas Chromatography
     1.4.3 Annex A3—Test Method for the Determination of the                                         D3710 Test Method for Boiling Range Distribution of
  Heat Loss in a Distillation Column (Static Conditions),                                             Gasoline and Gasoline Fractions by Gas Chromatography
     1.4.4 Annex A4—Test Method for the Verification of Tem-                                         D4006 Test Method for Water in Crude Oil by Distillation
  perature Sensor Location,                                                                          D4052 Test Method for Density, Relative Density, and API
     1                                                                                               2
       This test method is under the jurisdiction of ASTM Committee D02 on                             For referenced ASTM standards, visit the ASTM website, www.astm.org, or
  Petroleum Products and Lubricants and is the direct responsibility of Subcommittee              contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
  D02.08 on Volatility.                                                                           Standards volume information, refer to the standard’s Document Summary page on
     Current edition approved Dec. 1, 2011. Published January 2012. Originally                    the ASTM website.
                                                                                                     3
  approved in 1970. Last previous edition approved in 2011 as D2892–11. DOI:                           Withdrawn. The last approved version of this historical standard is referenced
  10.1520/D2892-11A.                                                                              on www.astm.org.
                                                    *A Summary of Changes section appears at the end of this standard.
    Copyright (c) ASTM International. 100 Barr Harbor Dr., P.O. box C700, West Conshohocken Pennsylvania United States
Copyright by ASTM Int'l (all rights reserved);                                                1
                                                                 D2892 – 11a
       Gravity of Liquids by Digital Density Meter                           3.1.9 pressure drop, n—the difference between the pressure
      D4057 Practice for Manual Sampling of Petroleum and                 measured in the condenser and the pressure measured in the
       Petroleum Products                                                 distillation flask.
      D4177 Practice for Automatic Sampling of Petroleum and                 3.1.9.1 Discussion—It is expressed in kilopascals (mm Hg)
       Petroleum Products                                                 per metre of packed height for packed columns, or kilopascals
      D5134 Test Method for Detailed Analysis of Petroleum                (mm Hg) overall for real plate columns. It is higher for
       Naphthas through n-Nonane by Capillary Gas Chromatog-              aromatics than for paraffins, and for higher molecular weights
       raphy                                                              than for lighter molecules, at a given boilup rate.
      D6300 Practice for Determination of Precision and Bias                 3.1.10 reflux ratio, R, n—in distillation, the ratio of the
       Data for Use in Test Methods for Petroleum Products and            condensate at the head of the column that is returned to the
       Lubricants                                                         column (reflux) to that withdrawn as product.
      D6729 Test Method for Determination of Individual Com-                 3.1.11 static hold-up or wettage, n—the quantity of liquid
       ponents in Spark Ignition Engine Fuels by 100 Metre                retained in the column after draining at the end of a distillation.
       Capillary High Resolution Gas Chromatography                          3.1.11.1 Discussion—It is characteristic of the packing or
      D6730 Test Method for Determination of Individual Com-              the design of the plates, and depends on the composition of the
       ponents in Spark Ignition Engine Fuels by 100−Metre                material in the column at the final cut point and on the final
       Capillary (with Precolumn) High-Resolution Gas Chroma-             temperature.
       tography                                                              3.1.12 takeoff rate, n—in distillation, the volume of product
      D6733 Test Method for Determination of Individual Com-              withdrawn from the reflux divider over a specified period.
       ponents in Spark Ignition Engine Fuels by 50-Metre                    3.1.13 theoretical plate, n—the section of a column required
       Capillary High Resolution Gas Chromatography                       to achieve thermodynamic equilibrium between a liquid and its
                                                                          vapor.
   3. Terminology                                                            3.1.13.1 Discussion—The height equivalent to one theoreti-
      3.1 Definitions:                                                    cal plate (HETP) for packed columns is expressed in millime-
      3.1.1 adiabaticity, n—the condition in which there is no            tres. In the case of real plate columns, the efficiency is
   significant gain or loss of heat throughout the length of the          expressed as the percentage of one theoretical plate that is
   column.                                                                achieved on one real plate.
      3.1.1.1 Discussion—When distilling a mixture of com-
   pounds as is the case of crude petroleum, there will be a normal       4. Summary of Test Method
   increase in reflux ratio down the column. In the case where               4.1 A weighed sample of 1 to 30 L of stabilized crude
   heat losses occur in the column, the internal reflux is abnor-         petroleum is distilled to a maximum temperature of 400°C
   mally greater than the reflux in the head. The opposite is true        AET in a fractionating column having an efficiency at total
   when the column gains heat, as with an overheated mantle.              reflux of at least 14, but not greater than 18, theoretical plates.
      3.1.2 boilup rate, n—in distillation, the quantity of vapor            4.2 A reflux ratio of 5:1 is maintained at all operating
   entering the column per unit of time.                                  pressures, except that at the lowest operating pressures be-
      3.1.3 debutanization of crude petroleum, n—the removal of           tween 0.674 and 0.27 kPa (5 and 2 mm Hg), a reflux ratio of
   the light hydrocarbons up to and including n-butane, and               2:1 is optional. In cooperative testing or in cases of dispute, the
   retention of the heavier hydrocarbons.                                 stages of low pressure, the reflux ratios, and the temperatures
      3.1.3.1 Discussion—In practice, a crude petroleum is re-            of cut points must be mutually agreed upon by the interested
   garded as debutanized if the light hydrocarbon cut collected in        parties prior to beginning the distillation.
   the cold trap contains more than 95 % of the C2 to C4                     4.3 Observations of temperature, pressure, and other vari-
   hydrocarbons and less than 5 % of the C5 hydrocarbons                  ables are recorded at intervals and at the end of each cut or
   initially present in the sample.                                       fraction.
      3.1.4 distillation pressure, n—the pressure measured as                4.4 The mass and density of each cut or fraction are
   close as possible to the point where the vapor temperature is          obtained. Distillation yields by mass are calculated from the
   taken, normally at the top of the condenser.                           mass of all fractions, including liquefied gas cut and the
      3.1.5 distillation temperature, n—the temperature of the            residue. Distillation yields by volume of all fractions and the
   saturated vapor measured in the head just above the fraction-          residue at 15°C are calculated from mass and density.
   ating column.                                                             4.5 From these data the TBP curves in mass or volume %,
      3.1.5.1 Discussion—It is also known as the head tempera-            or both, versus AET are drawn.
   ture or the vapor temperature.
      3.1.6 dynamic hold-up, n—in column distillation, the quan-          5. Significance and Use
   tity of liquid held up in the column under normal operating              5.1 This test method is one of a number of tests conducted
   conditions.                                                            on a crude oil to determine its value. It provides an estimate of
      3.1.7 flood point, n—in distillation, the point at which the        the yields of fractions of various boiling ranges and is therefore
   velocity of the upflowing vapors obstructs the down-coming             valuable in technical discussions of a commercial nature.
   reflux and the column suddenly fills with liquid.                        5.2 This test method corresponds to the standard laboratory
      3.1.8 internal reflux, n—in distillation, the liquid normally       distillation efficiency referred to as 15/5. The fractions pro-
   running down inside the column.                                        duced can be analyzed as produced or combined to produce
Copyright by ASTM Int'l (all rights reserved);                        2
                                                                 D2892 – 11a
   samples for analytical studies, engineering, and product quality       matic devices can be employed provided they meet the same
   evaluations. The preparation and evaluation of such blends is          requirements. A typical apparatus is illustrated in Fig. 1.
   not part of this test method.                                             6.1.1 Distillation Flask—The distillation flask shall be of a
     5.3 This test method can be used as an analytical tool for           size that is at least 50 % larger than the volume of the charge.
   examination of other petroleum mixtures with the exception of          The size of the charge, between 1.0 and 30 L, is determined by
   LPG, very light naphthas, and mixtures with initial boiling            the holdup characteristics of the fractionating column, as
   points above 400°C.                                                    shown in Table 1 and described in Annex A2. The distillation
                                                                          flask shall have at least one sidearm.
   6. Apparatus                                                              6.1.1.1 The sidearm is used as a thermowell. It shall
     6.1 Distillation at Atmospheric Pressure—All components              terminate about 5 mm from the bottom of the flask to ensure its
   must conform to the requirements specified as follows. Auto-           immersion at the end of the distillation. When a second sidearm
                                                             FIG. 1 Apparatus
Copyright by ASTM Int'l (all rights reserved);                        3
                                                                                D2892 – 11a
             TABLE 1 Data for n-Heptane-Methylcyclohexane Test Mixture at 75 % of Maximum Boilup and 101.3 kPa (760 mm Hg)
                                                             PropakA,B,C,D,E                         HelipakF,G,H      Perforated PlatesE,I,J          Wire MeshE,K
   Column diameter, mm                           25           50               70            25           50         25          50             25           50
   Packing size, mm                              4            6                6             No. 2917     No. 2918   NAL         NAL            NAL          NAL
   Boilup, mL/h 3 cm2                            650          670              675           300          350        640         660            810          1050
   Dynamic holdup
     % of packed volume                          17           15.3             17.0          15           14.3       NAL         NAL            8.0          10.0
     mL/theoretical plate                        3.2          16               39            1.6          8.7        2.8         12.3           2.0          12.9
   Pressure drop
     kPa/m                                       1.2          1.05             0.94          1.53         1.41       NAL         NAL            0.97         0.75
     mm Hg/m                                     9.0          7.9              7.1           11.5         10.6       NAL         NAL            7.3          5.6
     kPa/theoretical plate                       0.045        0.056            0.06          0.03         0.045      0.15        0.16           0.05         0.05
     mm Hg/theoretical plate                     0.34         0.42             0.43          0.24         0.34       1.1         1.2            0.35         0.37
   HETP, mm (% of real plates)                   38           53               61            21           32         (60 %)      (65 %)         48           66
   For 15-plate Towers
     Packed height, cm (plates)                  57           80               91            31.5         48         (25)        (23)           72           99
     Packed volume, mL                           280          1570             3460          155          917        NAL         NAL            353          1940
     Dynamic holdup, mL                          47           240              590           23           131        42          184            28           194
   Pressure drop
     kPa                                         0.68         0.84             0.86          0.48         0.68       2.2         2.4            0.70         0.73
     mm Hg                                       5.1          6.3              6.5           3.6          5.1        16.5        18.0           5.3          5.5
   Charge volume, L
     Min (4 % Holdup)                            1.2          6.0              15            0.575        3.3        1.0         4.6            0.7          4.9
     Max (1 % Holdup)                            4.8          24.0             60            2.3          13.0       4.2         10.4           2.8          19.4
     A
       Cooke, G. M. and Jameson, B. G. Analytical Chemistry, Vol 27, 1955, p. 1798.
     B
       Struck, R. T. and Kinner, C. R. Industrial and Engineering Chemistry, Vol 42, 1950, p. 77.
     C
       Cannon, M. R. Industrial and Engineering Chemistry, Vol 41, No. 9, 1949, p. 1953.
     D
       Bulletin 23, Scientific Development Co. P.O. Box 795, State College, PA 16801.
     E
       Cooke, G. M. Analytical Chemistry, Vol 39, 1967, p. 286.
     F
       Bulletin of Podbielniak Div. of Reliance Glass Works, P.O. Box 825, Bensenville, IL 60106.
     G
        Feldman, J., et al, Industrial and Engineering Chemistry, Vol 45, January 1953, p. 214.
     H
       Helipak Performance Characteristics, Begemean, C. R. and Turkal, P. J. (Laboratory Report of Podbielniak Inc.), 1950.
     I
       Umholtz, C. L. and Van Winkle, M. Petroleum Refiner, Vol 34, 1955, p. 114 for NH:MCH. Pressure Drop Calculated from data obtained on o- and m-xylene binary.
     J
       Oldershaw, C. F. Industrial and Engineering Chemistry, Vol 13, 1941, p. 265.
     K
       Bragg, L. B. Industrial and Engineering Chemistry, Vol 49, 1957, p. 1062.
     L
       NA = not applicable.
   is present, it can be used for pressure drop detection with a                             6.1.2.2 The heat density in the flask heaters is approxi-
   nitrogen bleed or for mechanical stirring, or both.                                    mately equal to 0.5 to 0.6 W/cm2. This requires the use of
      6.1.1.2 If a magnetic stirrer is used with a spherical flask,                       nickel reinforced quartz fabric to ensure a reasonable service
   the flask shall have a slightly flattened or concave area at the                       life.
   bottom on which the magnetic stirrer can rotate without                                   6.1.2.3 Immersion heaters can be employed in a similar way
   grinding the glass. In this case, termination of the thermowell                        and have the advantage of faster response, but they are more
   shall be off center 40 6 5 mm to avoid the magnetic stirring                           fragile and require a specially designed flask to ensure that the
   bar. Boiling chips can be used as an alternative to a stirrer.                         heating elements remain immersed at the end of the run. When
      6.1.1.3 (Warning—While the advantage of visibility in                               used, their heat density should be approximately equal to 4
   glass distillation flasks is desirable, flasks of glass may become                     W/cm2.
   hazardous the larger the charge they contain. For this reason,                            6.1.2.4 The upper half of the flask shall be covered with a
   glass flasks of a volume greater than 10 L are not recom-                              mantle to avoid unnecessary heat losses from the upper surface
   mended.)                                                                               and shall have an electric heater supplying about 0.25 W/cm2
      6.1.2 Heating System—Heating of the flask shall be pro-                             at full-rated voltage.
   vided in such a way that full boilup can be maintained at a                               6.1.3 Fractionating Column—The fractionating column
   steady rate at all pressure levels. An electric heating mantle                         must contain either particulate packing or real plates similar to
   covering the lower half of the flask and having one third of the                       those whose performance characteristics are summarized in
   heat in an element located in the bottom central area and the                          Table 1 and meet the specifications stated in 6.1.3.1 through
   remaining two thirds in the rest of the hemisphere is recom-                           6.1.3.4. Table 2 lists current North American suppliers of
   mended. While proportioning controllers are preferred, heat                            suitable packings.
   input can be manually adjusted by use of a variable auto                                  6.1.3.1 The internal diameter shall be between 25 and 70
   transformer on each circuit, the smaller heater being automati-                        mm.
   cally controlled by an instrument sensing the pressure drop of                            6.1.3.2 The efficiency shall be between 14 and 18 theoreti-
   the column as registered in a differential pressure instrument or                      cal plates at total reflux when measured by the procedure
   alternatively by direct measurement of distillation rate.                              described in Annex A1.
      6.1.2.1 Minimum wattage required to provide full boilup of                             6.1.3.3 The fractionating column shall be comprised of a
   crude petroleum is approximately 0.125 W/mL of charge.                                 integral glass column and reflux divider totally enclosed in a
   Twice this amount is recommended for quick heat-up.                                    highly reflective vacuum jacket having a permanent vacuum of
Copyright by ASTM Int'l (all rights reserved);                                        4
                                                                                D2892 – 11a
     TABLE 2 North American Sources of Commercially Available                         with a calcium chloride drying tube between them to keep
                        Packing Materials                                             moisture from collecting in the traps. When analysis of the gas
         Name                  Size                       Source                      sample is required, the gas can be collected in an empty plastic
   Propak               6 by 6 mm         Scientific Development Co.                  balloon of suitable size either in place of the meter or following
                                          P.O. Box 795
                                          State College, PA 16801
                                                                                      it. The volume of its contents can be determined by calculation
   Helipak              2.5 by 4 mm       Reliance Glass Works Inc.                   from the rise in pressure after expanding the sample into an
                                          P.O. Box 825                                evacuated vessel of known volume.
                                          Bensenville, IL 60106
   Perforated plates    25 and 50 mm      Reliance Glass Works Inc.                       6.1.7 Fraction Collector—This part of the apparatus per-
                                          P.O. Box 825                                mits the collection of the distillate without interruption during
                                          Bensenville, IL 60106
                                          W.A. Sales Inc.
                                                                                      withdrawal of product from the receiver under atmospheric or
                                          419 Harvester Ct.                           reduced pressure. It also permits removal of product from the
                                          Wheeling, IL 60090                          vacuum system, without disturbing conditions in the column.
   Knitted wire mesh-                     Pegasus Industrial Specialties Ltd.
   Goodloe multiknit                      P.O. Box 319                                    6.1.8 Product Receivers—The receivers shall be of suitable
                                          Agincourt, Ontario MIS 3B9 Canada           size for the quantity of crude petroleum being distilled. The
                                          Packed Column Co.
                                          970 New Durham Rd.
                                                                                      recommended capacity is from 100 to 500 mL. They shall be
                                          Edison, NJ 08817                            calibrated and graduated to permit reading to the nearest 1 %.
                                                                                          6.2 Distillation Under Reduced Pressure—In addition to the
                                                                                      apparatus listed in 6.1, the apparatus for distillation under
   less than 0.1 mPa (;10−6 mm Hg). It shall be essentially                           reduced pressure shall include the following:
   adiabatic when tested in accordance with Annex A3.                                     6.2.1 Vacuum Pump—The vacuum system shall be capable
      NOTE 2—In the case of an adiabatic column when distilling a pure                of maintaining smooth pressure operation at all pressure levels.
   compound, the internal reflux is constant from top to bottom and is equal          It shall have the capacity to draw down the pressure in the
   to the reflux at the reflux divider. When distilling crude petroleum, the          receiver(s) from atmospheric to 0.25 kPa (2 mm Hg) in less
   fractionation occurring in the dynamic holdup will cause a temperature             than 30 s so as to avoid disturbance of the system during
   gradient to be established with attendant greater amount of internal reflux        emptying of receivers under vacuum. Alternatively, a separate
   at the bottom of the column.
                                                                                      pump can be employed for this purpose.
     6.1.3.4 The column shall be enclosed in a heat insulating                            6.2.2 Vacuum Gauge—The point of connection of the
   system, such as a glass-fabric mantle, capable of maintaining                      vacuum gauge to the system shall be as close as practical to the
   the temperature of the outer wall of the glass vacuum jacket                       reflux dividing head. The connecting tubing shall be of
   equal to that of the internal vapor temperature. To verify this,                   sufficient diameter to ensure that no measurable pressure drop
   the vacuum jacket shall have a temperature sensor, such as a                       occurs in the line. In no case shall the vacuum gauge
   thermocouple, soldered to about 6 cm2 of thin copper or brass                      connection be near the vacuum pump.
   sheet and fastened to the outer wall of the glass jacket at a level
                                                                                          6.2.2.1 All gauges shall be carefully protected from con-
   just below the reflux divider.
                                                                                      densable vapors, especially water vapor, by a cold trap main-
     NOTE 3—For certain types of columns there is no significant difference           tained at the temperature of dry ice.
   in yields and fraction qualities between an uncompensated and a heat-
                                                                                          6.2.3 Pressure Regulator—The regulator shall maintain the
   compensated column. In such a case, by mutual agreement between
   parties concerned, the application of a heated insulating system can be            pressure in the system essentially constant at all operating
   omitted.                                                                           pressures. Automatic regulation can be achieved by a device
                                                                                      that regulates the demand on the vacuum source. A satisfactory
      6.1.3.5 The adjustable reflux divider shall be located about
                                                                                      device is a solenoid valve positioned between the vacuum
   one column diameter above the top of the packing or topmost
                                                                                      source and a surge tank of at least 10-L capacity. Alternatively,
   plate. It must be capable of dividing the condensate with an
                                                                                      a manual bleed valve can be maintained by a trained operator
   accuracy of better than 90 % between the column and the
                                                                                      with a minimum of attention.
   takeoff line over a range of rates from 25 to 95 % of the
   maximum boilup rate of the column when determined in                                   6.3 Sensing and Recording Apparatus:
   accordance with Annex A7.                                                              6.3.1 Temperature Sensors—Only temperature measure-
      6.1.4 Condenser—The condenser shall have sufficient ca-                         ment systems meeting the requirements of 6.3.1.1 and 6.3.1.2
   pacity to condense essentially all the C4 and C5 vapors from the                   shall be used.
   crude at the specified rate, using a coolant temperature                               6.3.1.1 The vapor temperature sensor can be a platinum
   of −20°C.                                                                          resistance thermometer, a Type J thermocouple with the
      6.1.5 Cold Traps—Two efficient traps of adequate capacity                       junction head fused to the lower tip of the thermowell, or any
   cooled by dry ice and alcohol mixture shall be connected in                        other device that meets the requirements in this paragraph and
   series to the vent line of the condenser when light hydrocar-                      6.3.1.2. The tip of the sensor shall be located above the top of
   bons are present, as at the beginning of the distillation. For                     the packing or the topmost glass plate and in close proximity to
   vacuum distillation, a Dewar-style trap also cooled by dry ice                     the reflux divider but not in contact with the liquid reflux. The
   is used to protect the vacuum gauge from vapors.                                   location of the vapor temperature sensor shall be proved by the
      6.1.6 Gas Collector—If uncondensed gas is to be measured,                       test method described in Annex A4. The sensor shall have a
   a gas meter can be connected to the outlet of the cold trap but                    cooling time of not more than 175 s, as described in Annex A5.
Copyright by ASTM Int'l (all rights reserved);                                    5
                                                                              D2892 – 11a
                                                 FIG. 2 Approximate Pressure Drop-Fractionators Using Propak
      6.3.1.2 The vapor temperature measuring device shall have                         100-13.3          760 to 100         0.13           1.0
   an accuracy of 0.5°C or better and be measured with a                                13.3-1.33          99 to 10          0.013          0.1
                                                                                        1.33-0.266          9 to 2           0.006          0.06
   resolution of 0.1°C or better. The liquid temperature measuring
   device shall have an accuracy of 1.0°C or better and be                             6.3.2.2 Noncertified gauges shall be calibrated from a non-
   measured with a resolution of 0.5°C or better. Temperatures are                  tilting McLeod gauge or a secondary electronic standard
   recorded either manually or automatically.                                       traceable to a primary standard. A basic calibration procedure
      6.3.1.3 Temperature sensors shall be calibrated as described                  is described in Annex A6. Recalibrate when either the sensor or
   in Annex A6. Alternatively certified sensors may be used,                        the instrument is repaired or serviced. Verification of the
   provided the calibration of the sensor and its associated                        calibration of the electronic pressure sensors is to be made on
   recording instrument can be traced back to a primary tempera-                    a regular basis. A frequency of at least once a month is
   ture standard. Temperature sensors are calibrated over the full                  recommended. Verification of the calibration of the sensors can
   range of temperature (from 0 to 400°C) at the time of first use                  be accomplished using the procedures described in Annex A6
   of the sensor in combination with its associated instrument.                     or against a certified reference system.
   Recalibrate when either the sensor or the instrument is repaired                    6.3.3 Boilup Rate—The boilup rate is normally controlled
   or serviced. Verification of the calibration of the temperature                  by sensing the pressure drop in the column. The pressure drop
   sensors is to be made on a regular basis. For vapor temperature                  during operation is measured by means of a manometer or
   sensors, verification at least once a month is recommended and                   pressure transducer connected between the flask and the
   for liquid temperature sensors once every six months. Verifi-                    condenser. Prevention of condensation in the connecting tube
   cation of the calibration of the sensors can be accomplished                     can be accomplished by injecting a very small flow of nitrogen
   potentiometrically by the use of standard precision resistance
                                                                                    (8 cm3/s) between the pressure drop sensor manometer and the
   or by distilling a pure compound with accurately known
                                                                                    flask (see Fig. 1) or by placing a small water-cooled condenser
   boiling point.
                                                                                    between the flask and the pressure drop sensor. Alternatively,
      6.3.2 Vacuum Gauge—A nontilting McLeod gauge or a
                                                                                    the boilup rate can be controlled from the measurement of take
   mercury manometer are primary standards and can be used
                                                                                    off rate.
   without calibration when properly used and maintained. A
   mercury manometer, however, will only be of satisfactory
                                                                                    7. Verification of Apparatus Performance
   accuracy down to a pressure of about 1 kPa and then only when
   read with a good cathetometer (an instrument based on a                             7.1 Test Method D2892 provides a standard framework for
   telescope mounted on a vernier scale to determine levels very                    the laboratory distillation of crude oils in order to produce cuts
   accurately). Alternatively, a tensimeter or certified electronic                 of defined quality (for further testing) and the concurrent
   sensors may be used, provided the calibration of the sensor and                  production of a boiling point curve. As the quantity require-
   its associated recording instrument can be traced back to a                      ments and cut points might be widely different between
   primary pressure standard. Sensors of the diaphragm type have                    companies and application areas, this test method does not
   been found satisfactory. Vacuum gauges based on hot wires,                       standardize on equipment design but on equipment perfor-
   radiation, or electrical conductivity detectors are not recom-                   mance.
   mended.                                                                             7.2 The nature of the test method (the use of large sample
      6.3.2.1 The gauge for measuring subatmospheric pressures                      quantities and very time consuming) and the nature of the
   shall have an accuracy at least equal to that stated as follows:                 product being tested (highly volatile and unstable material),
              Distillation Pressure                        Accuracy                 precludes the use of standard statistical control techniques.
            kpa                  mm Hg               kPa              mm Hg         Moreover, this test method does not produce a single result, nor
Copyright by ASTM Int'l (all rights reserved);                                  6
                                                                   D2892 – 11a
   is the series of results (the boiling point curve) derived under         cluding both correctness of AET and column efficiency. It is the
   rigidly defined conditions (see 4.2).                                    responsibility of the laboratory to provide for sufficient quality
      7.3 Equipment performance in the context of Test Method               controls to guarantee conformance to the test method.
   D2892 consists of two elements; the efficiency of the column,
   defining cut quality, and the correctness of the cut point (AET),        8. Sampling
   defining the boiling point curve.                                           8.1 Obtain a sample for distillation in accordance with
      7.4 The correctness of the AET is mainly, but not only,               instructions given in Practice D4057 or D4177. The sample
   dependent on the accuracy of the (vapor) temperature and                 must be received in a sealed container and show no evidence of
   (operating) pressure sensors (Annex A6). Other factors affect-           leakage.
   ing the accuracy and precision of the boiling point curve are:              8.2 Cool the sample to between 0 and 5°C by placing it in
      7.4.1 The location of the temperature and pressure sensor             a refrigerator for several hours (preferably overnight) before
   (Annex A4).                                                              opening.
      7.4.2 The dynamic response of the sensors (Annex A5).                    8.3 If the sample appears waxy or too viscous, raise the
      7.4.3 The correct operation of the reflux divider (Annex              temperature to 5°C above its pour point.
   A7).                                                                        8.4 Agitate the sample by whatever means are appropriate
      7.4.4 The heat loss from the column (Annex A3).                       to its size to ensure that it is well-mixed.
      7.4.5 The efficiency of the column (Annex A1).                           8.5 Determine the water content of the sample by Test
      7.4.6 These factors are basically covered through the appro-          Method D4006 or any other suitable method. If the water
   priate annexes. However, it should be realized that this takes           content exceeds 0.3 % volume, the sample shall be dehydrated
   only care of individual components and does not cover the                prior to fractional distillation. A suitable practice for dehydra-
   combined effect of small deviations. Moreover, the aforemen-             tion of wet crude oil samples is described in Appendix X1.
   tioned tests are all done under more or less static conditions,            NOTE 4—Attempts to distill wet crude oil samples in glass columns
   not necessarily representative for the behavior of the system            might result in breakage of the glassware, which poses a potential fire
   under actual dynamic conditions.                                         hazard. Moreover, the presence of water will effect the accuracy of
      7.5 Cut quality is mainly defined by the efficiency of the            distillation yield in the naphtha region. These effects are more pronounced
   column (Annex A1), but is also affected by:                              for heavy crude oils, containing low amounts of hydrocarbons boiling
                                                                            below 100°C, than for light crudes where there is usually sufficient
      7.5.1 The correct operation of the reflux divider (reflux             hydrocarbon vapor generated to form an azeotrope and drive the water
   ratio) (Annex A7).                                                       vapors through the column without problems.
      7.5.2 The heat loss from the column, that is, internal reflux
   (Annex A3).                                                              9. Preparation of Apparatus
      7.5.3 The dynamic hold up of the column (Annex A2).                     9.1 Clean and dry the distillation column and all the
      7.5.4 Column efficiency is covered in this test method                ancillary glass apparatus before the distillation begins.
   through Table 1 and Annex A1. However, Table 1 only                        9.2 Ensure that the system is leak-free and all heaters,
   provides an assumption on efficiency and is not a guarantee.             control devices, and instruments are on and in working order.
   Annex A1 only provides a check under static conditions,                  A clock or other timing device should be ready for use.
   infinite reflux ratio, rather low actual temperatures and a binary
   component system. Hence, although there is some safeguard on             10. Procedure
   standard performance, through conformance to Annex A1,                      10.1 Charging:
   Annex A2, Annex A3, and Annex A7, again it remains                          10.1.1 The charge size shall be such that the dynamic hold
   questionable whether this is truly representative for columns            up as determined in accordance with Annex A2 is between 1
   under actual operating conditions.                                       and 4 % of the charge when operating at 75 % of maximum
      7.6 Theoretically, an overall performance check, like the             boilup (see Table 1). Chill the flask to a temperature not lower
   one described in Appendix X2, is capable of verifying the                then 0°C.
   performance of a column and the correctness of the AET under                10.1.2 Insert the stirring device or place some pieces of
   actual operating conditions. Appendix X2, in principle, mea-             glass or porcelain into the flask to control bumping.
   sures the combined effect of all factors affecting the results of           10.1.3 Determine the density of the sample by Test Method
   Test Method D2892.                                                       D941, D1217, or D1298.
      7.6.1 The minimum tray number as defined in Appendix X2                  10.1.4 Calculate to within 65 % the mass of crude petro-
   is a measure of overall cut quality, and the difference between          leum corresponding to the desired volume of the charge. Weigh
   nominal cut point (AET) and effective cut point (ECP as                  to the nearest 1 % this quantity of sample into the flask.
   defined in Appendix X2) provides a measure for the correct-                 10.1.5 Attach the flask to the column and connect the
   ness of the AET. However, insufficient data are available right          pressure drop measuring device. Install the heating system,
   now to define the allowable tolerances in a rigid statistical way.       stirrer, and support device. (Warning—Poisonous H2S gas is
   Moreover, the test method described is very labor intensive and          frequently evolved from crude oil and precautions must be
   precludes its use on a regular, short time interval basis and,           taken either to absorb the gas that passes through the cold trap
   therefore, its use as a mandatory statistical control technique.         or to vent it to a safe place.)
      7.6.2 Appendix X2, therefore, provides only recommended                  10.2 Debutanization:
   guidelines for statistical control on column performance, in-               10.2.1 For necessary apparatus refer to 6.1.5 and 6.1.6.
Copyright by ASTM Int'l (all rights reserved);                          7
                                                                             D2892 – 11a
      10.2.2 Begin circulation of refrigerant at a temperature no
   higher than −20°C in the condenser, distillate cooler, and
   receiver, if so equipped.
      10.2.3 Record the barometric pressure at the beginning and
   periodically throughout the distillation.
      10.2.4 Apply heat to the flask at such a rate that vapors
   reach the top of the column between 20 and 50 min after
   startup. Adjust heat input so as to achieve a pressure drop of
   less than 0.13 kPa/m (1.0 mm Hg/m) in packed columns or less
   than 0.065 kPa (0.5 mm Hg) in real plate columns. Program
   automated equipment in accordance with the preceding direc-
   tions. Turn on the stirring device if used.
      10.2.5 Allow the column to operate at total reflux until the
   vapor temperature reaches equilibrium but not longer than 15
   min after the first drop of condensate appears in the reflux
   divider.
      10.2.6 Record the vapor temperature as the initial vapor
   temperature.
      10.2.7 Stop the circulation of the refrigerant and observe the
   vapor temperature. When the vapor temperature reaches 15°C,
   start the circulation of refrigerant again.
                                                                                              FIG. 3 Expected Takeoff Rates at 5:1 Reflux Ratio for
      10.2.8 If the vapor temperature drops below 15°C, continue                                          Fractionators Using Propak
   refluxing for at least 15 min. Repeat 10.2.7. If the vapor
   temperature remains at 15°C or rises, continue with the
   atmospheric distillation. (Warning—The following three steps
                                                                                       evaluation (see Annex A1) and is measured at the bottom of the column.
   should not be done until after the first naphtha cut has been                       The maximum boilup of the n-heptane-methylcyclohexane test mixture is
   removed to ensure that all the light gases have been recovered.)                    that which the column can handle under stable conditions without
      10.2.9 Remove and weigh the dry ice traps containing light                       flooding. In routine adiabatic operation, the boilup rate can be estimated
   hydrocarbon liquid after carefully wiping them dry.                                 roughly from the takeoff rate multiplied by the reflux ratio plus one.
      10.2.10 Sample the contents of the first dry ice trap using a                      10.3.3 Commence takeoff at a reflux ratio of 5:1 and total
   10 to 50 mL pressure vessel evacuated to no lower than 26.6                         cycle time of not over 30 s nor less than 18 s.
   kPa (200 mm Hg). Keep all containers at the temperature of
   dry ice to ensure no loss of volatiles. The first trap next to the                     NOTE 6—The vapor reaching the top of the column is totally condensed
   condenser should contain all of the sample. If condensate is                        and the resulting liquid is divided into two parts. One part L (reflux), is
                                                                                       returned to the column and the other part, D (distillate), is withdrawn as
   found in the second trap, sample both traps or combine the                          product. The reflux ratio (R = L/D), can vary from zero at total takeoff
   contents before sampling.                                                           (L = 0) to infinity at total reflux (D = 0).
      10.2.11 Submit the trap sample and gas balloon, if used, for
   analysis by a suitable gas chromatographic test method to be                           10.3.4 Take off distillate in separate and consecutive frac-
   reported on a fixed-gas free basis. Test Methods D6729,                             tions of suitable size. The recommended size of fraction is that
   D6730, and D6733, equipped with liquid or gas sampling                              corresponding to 5 or 10°C in vapor temperature. Collect
   valves, or both, for sample introduction equipment have been                        fractions boiling below 65°C in receivers cooled to 0°C or
   used successfully for this analysis.                                                below. When the vapor temperature reaches 65°C, refrigerant
      10.3 Distillation at Atmospheric Pressure:                                       in the condenser and related coolers can be discontinued and
                                                                                       water at ambient temperature substituted.
      10.3.1 Maintain a temperature below −20°C in the lines of
   the distillate cooler and receiver as well as in the condenser.                        10.3.5 At the end of each fraction and at each cut point,
   Turn on the column mantle heat controller and maintain the                          record the following observations:
   column jacket temperature 0 to 5°C below the vapor tempera-                            10.3.5.1 Time in hours and minutes,
   ture.                                                                                  10.3.5.2 Volume in millilitres,
      10.3.2 Regulate the heat input as necessary to establish and                        10.3.5.3 Vapor temperature in °C to the nearest 0.5°C,
   maintain a boilup rate approximately 75 % of maximum. Fig.                             10.3.5.4 Temperature of the boiling liquid in °C to the
   3 can be used as a guide for Propak. Rates for other sizes can                      nearest 1°C,
   be estimated by multiplying the boilup rate in Table 1 by the                          10.3.5.5 Atmospheric pressure in kPa (mm Hg), and
   cross-sectional area of the column and dividing by the sum of                          10.3.5.6 Pressure drop in the column in kPa (mm Hg).
   the reflux ratio + 1.                                                                  10.3.6 If signs of flooding are observed, reduce the heating
      NOTE 5—Boilup rate is expressed in millilitres of liquid per hour for a
                                                                                       rate while continuing takeoff until steady conditions are
   given column or in millilitres per hour per square centimetre of cross-             restored. If a cut point is encountered during this period, stop
   sectional area for comparative purposes. In the latter case, it refers to the       the distillation, cool the charge, and recombine the off-
   test mixture of n-heptane and methylcyclohexane in the efficiency                   condition cuts. Restart the distillation with a period at total
Copyright by ASTM Int'l (all rights reserved);                                     8
                                                                          D2892 – 11a
   reflux, not to exceed 15 min, to restore operating conditions                       10.4.7.2 Volume in millilitres observed at ambient tempera-
   before continuing takeoff. Do not make a cut within 5°C of                       ture,
   startup.                                                                            10.4.7.3 Vapor temperature in °C to the nearest 0.5°C with
      10.3.7 Continue taking cuts until the desired maximum                         correction, if any,
   vapor temperature is reached or until the charge shows signs of                     10.4.7.4 Temperature of the boiling liquid in °C to the
   cracking. Pronounced cracking is evidenced by a fog appearing                    nearest 1°C,
   in the flask and later at the reflux divider. Do not allow the                      10.4.7.5 Pressure drop in the column in kPa (mm Hg),
   vapor temperature to exceed 210°C nor the temperature of the                        10.4.7.6 Operating pressure measured at the top of the
   boiling liquid to exceed 310°C.                                                  column in kPa (mm Hg) absolute with correction, if any, and
      10.3.8 Shut off the reflux valve and the heating system.                         10.4.7.7 AET using the equations given in Annex A8.
   Allow the contents to cool to such a temperature that the                           10.4.8 Continue taking cuts until the desired maximum
   distillation can be commenced at 13.3 kPa (100 mm Hg)                            point is reached or until the charge shows signs of cracking.
   without flooding. This temperature can be estimated by adding                    Pronounced cracking is evidenced by the evolution of gases as
   the DT between the liquid and vapor temperatures found for the                   indicated by rising pressure as well as a fog appearing in the
   column during atmospheric operation to the expected initial                      flask (see Note 7). Do not allow the temperature of the boiling
   vapor temperature at the reduced pressure, or by subtracting                     liquid to exceed 310°C. (Warning—Automatic vacuum con-
   the DT from the last recorded liquid temperature.                                trollers could mask a slight rise in pressure due to cracking.
     NOTE 7—Cooling of the liquid in the flask can be accelerated by                Vigilance is required to avoid this.)
   blowing a gentle stream of compressed air onto the flask after its heating          10.4.9 Shut off the reflux valve and the heating system.
   mantle has been removed. Avoid strong jets of cold air. Alternately, turn        Allow the contents to cool to such a temperature that the
   on coolant in the quench coil of the flask, if used.                             distillation can be commenced at a lower pressure without
      10.3.9 Weigh all fractions and determine their densities.                     boiling. This temperature can be estimated by adding the DT
      10.3.10 Submit the first distillate fraction for analysis by gas              between the liquid and vapor temperatures found for the
   chromatography.                                                                  column during operation to the expected initial vapor tempera-
      10.4 Distillation at 13.3 kPa (100 mm Hg):                                    ture at the lower pressure, or by subtracting the DT from the
      10.4.1 If further cuts at higher temperatures are required,                   last recorded liquid temperature.
   distillation can be continued at reduced pressures, subject to                      10.4.10 Weigh all fractions and determine their densities at
   the maximum temperature that the boiling liquid will stand                       15°C.
   without significant cracking. This is about 310°C in most cases.                    10.5 Distillation at Lower Pressures:
   Notable exceptions are crude oils containing heat-sensitive                         10.5.1 If the final cut point has not been reached, distillation
   sulfur compounds. In any case, do not make a cut within 5°C                      can be continued at a lower pressure subject to the same
   of the temperature at startup because the column will not be at                  limitation as before (see 10.4.1). Only one pressure level
   equilibrium.                                                                     between 13.3 kPa (100 mm Hg) and 0.266 kPa (2 mm Hg) is
      10.4.2 Connect a vacuum pumping and control system to                         permitted. Where the maximum cut point is 400°C AET, the
   the apparatus as shown in Fig. 1.                                                minimum pressure is recommended.
      10.4.3 Start the vacuum pump and adjust the pressure                             10.5.2 Adjust the pressure to the desired level. If the liquid
   downward gradually to the value of 13.3 kPa (100 mm Hg) or                       boils before the pressure is reached, increase the pressure and
   set the pressure regulator at this value. The temperature of the                 cool further until the desired pressure can be achieved without
   liquid in the flask must be below that at which it will boil at                  boiling. Follow the procedure in 10.4.4.
   13.3 kPa (100 mm Hg). If the liquid boils before this pressure                      10.5.3 Circulate cooling water in the condenser and liquid
   is reached, increase the pressure and cool further until the                     cooler either at ambient temperature or warmed to a tempera-
   desired pressure can be achieved without boiling.                                ture that will ensure that wax does not crystallize in the
      10.4.4 Apply heat to the boiler and reestablish reflux at any                 condenser or takeoff lines. Alternatively, leave the cooling coils
   moderate rate in the reflux divider for about 15 min to reheat                   full of water but vented and not circulating, or else circulate a
   the column to operating temperature. Momentarily stop heat                       stream of air instead of water as a coolant.
   input and raise the pressure with N2 for 1 min to drop the                          10.5.4 Continue vacuum operation as in 10.4.5 through
   holdup into the distillation flask.                                              10.4.8. During this operation, a reflux ratio of 2:1 is allowed if
      10.4.5 Reapply heat to the distillation flask and adjust the                  mutually agreed upon in advance and noted in the report.
   rate of heating to maintain a constant pressure drop equivalent                  Correct observed and corrected vapor temperatures to AET
   to the boilup rate of approximately 75 % of the maximum rate                     using the equations given in Annex A8.
   for this pressure and begin takeoff without delay. The approxi-                     10.5.5 Check periodically that the condensate drips nor-
   mate pressure drops required for this purpose are indicated in                   mally in the condenser and that the distillate flows smoothly
   Fig. 3. Maintain a column insulation temperature 0 to 5°C                        into the takeoff line. If crystallization is observed, allow the
   below the vapor temperature throughout the operation.                            coolant in the condenser to warm as in 10.5.3.
      10.4.6 Remove separately, cuts of suitable size as in 10.3.4.                    10.5.6 When the final cut point has been reached, or when
      10.4.7 At the end of each distillate fraction and at each cut                 limits of boiling liquid temperature and column pressure
   point, record the following observations:                                        prevent further distillation, turn off the reflux valve and heating
      10.4.7.1 Time in hours and minutes,                                           system and allow to cool with the vacuum still applied.
Copyright by ASTM Int'l (all rights reserved);                                  9
                                                                  D2892 – 11a
      10.5.7 When the temperature of the residue in the flask has             10.5.9 Weigh all fractions and the residue in the flask and
   fallen below 230°C, shut off the vacuum pump. Vent the                   determine their densities at 15°C by Test Method D4052 or by
   fractionating unit with nitrogen or other inert gas. Do not use          another suitable method. Convert the density to 15°C, if
   air. (Warning—Air is suspected of initiating explosions in               necessary.
   fractionating units that are vented while too hot, such as at the
   end of a run.)                                                             NOTE 8—Heavier flasks, such as those for 50 and 70-mm diameter
      10.5.8 Stop circulation of coolant in the condenser and               columns, are not normally removed for weighing. In these cases the
                                                                            residue can be discharged at a temperature not over 200°C into a tared
   ancillary equipment. Disconnect the flask. Recover the static
                                                                            container for weighing. Nitrogen pressure of approximately 6.7 kPa (50
   holdup of the column (wettage) by distilling a small quantity of         mm Hg) will be sufficient for this. Wettage in these cases will include that
   solvent such as toluene in a separate flask to wash the column,          of the column and the flask together.
   condenser, and takeoff system. Evaporate the solvent from the
   collected residue at 10°C above the boiling point of the solvent,        11. Calculation
   using a small purge of nitrogen. For distillations not involving
   disagreement, or by mutual consent, the holdup can be esti-                11.1 Calculate the mass % of each distillate fraction and the
   mated using a graph similar to Fig. 4. The density of the holdup         residue to the 0.1 mass %, using Eq 1.
   is estimated by extrapolation of the density line for the                                           mass % 5 100~m/M!                             (1)
   preceding cuts. The static holdup can be treated as a separate
   small cut or blended into the bottoms before inspections are             where:
   made. The latter must be done if other analyses besides density          m = mass of fraction or residue, g, and
                                                                            M = mass of dry crude oil charged, g.
   are to be performed on the residue.
                  FIG. 4 Approximate Static Holdup for Average Crude Oil Using 4 mm Propak in a 25-mm ID 3 570-mm Column
Copyright by ASTM Int'l (all rights reserved);                         10
                                                                           D2892 – 11a
      11.1.1 The first fraction is the gas fraction collected in the                     12.1.5 The volume and mass % of each fraction to the
   balloon. If this fraction is less than 0.1 mass %, it can be                       nearest 0.1 %,
   ignored.                                                                              12.1.6 The cumulative volume and mass percentages, and
      11.1.2 The second fraction (or first, if no gas is collected) is                   12.1.7 The mass of water, if any.
   the condensate in the dry ice trap. With density at 15°C                              12.2 The gas, debutanized naphtha, and succeeding frac-
   calculated from the gas chromatographic data on a fixed gas                        tions are listed in order of ascending boiling point with residue
   free basis, its volume can be computed.                                            recorded last.
      11.1.3 The holdup is treated either as a separate cut or added                     12.3 The observations made in 10.3.5, 10.4.7, and 10.5.4 are
   to the residue fraction, in accordance with agreement. The                         included as a second sheet, which is normally attached to the
   amount of holdup is determined by actual recovery by solvent                       summary sheet.
   washing, as directed in 10.5.8, or estimated from Fig. 4.                             12.4 Make plots of the temperature in degrees Celsius AET
      11.2 Calculate the percent loss to the nearest 0.1 mass %,                      as the ordinate (y-axis) versus the percentage mass and volume
   using Eq 2.                                                                        distilled as the abscissa (x-axis). These are the final TBP
                                                                                      distillation curves.
                           Loss 5 100 2 ~ (100~m/M!!                       (2)
      The weight loss as calculated above must not be greater than                    13. Precision and Bias 4
   0.4 %, otherwise the distillation must be discarded. Losses less                      13.1 The precision of this test method, as determined by the
   than this should be allocated two thirds to the trap cut and one                   statistical examination of interlaboratory test results obtained
   third to the first naphtha cut. Where there is no trap cut, the                    from eight laboratories on five crude oil samples ranging in
   acceptable losses are to be normalized among all cuts.                             gravity from 20.0° to 41.0° API. The sample set contained
      11.3 Calculate the volume of the sample of crude oil in                         crudes with high and low sulfur content, high and low
   millilitres at 15°C, using Eq 3.                                                   naphthenic acid content, and asphaltic types. The results of the
                                    V 5 ~M/D!                              (3)        statistical examination follow, and the statistical evaluation and
                                                                                      validation can be found in the research report.
    where:                                                                               13.2 Repeatability—The difference between successive re-
    D = density of charge at 15°C, g/mL,                                              sults obtained by the same operator with the same apparatus
    M = mass of dry charge, g, and                                                    under constant operating conditions on identical test material
    V = volume of charge, mL.                                                         would, in the long run, in the normal and correct operation of
     11.4 Calculate the volume of each fraction and of the                            this test method, exceed the values indicated as follows in one
   residue in millilitres at 15°C, using Eq 4.                                        case in twenty.
                                      v 5 m/d                              (4)                                                      Mass %            Degrees of
                                                                                                                                                       Freedom
    where:                                                                            Atmospheric pressure                            0.6                 19
    d = density of the fraction or residue at 15°C, g/mL,                             13.33 kPa Vacuum pressure                       0.9                 19
                                                                                      1.33 kPa Vacuum pressure                        0.9                 19
    m = mass of fraction or residue corrected for loss, g, and
    v = volume of fraction, mL.                                                         13.3 Reproducibility—The difference between two single
     11.5 Calculate the volume % of each distillate fraction to the                   and independent results obtained by different operators work-
   nearest 0.1 volume %, using Eq 5.                                                  ing in different laboratories on identical test material would, in
                                 vol % 5 100~v/V!                          (5)
                                                                                      the normal and correct operation of this test method, exceed the
                                                                                      values indicated as follows in one case in twenty.
     11.6 Calculate the volume % gain or loss to the nearest 0.1                                                                    Mass %            Degrees of
   volume %, using Eq 6.                                                                                                                               Freedom
                                                                                      Atmospheric pressure                            1.3                 16
                            Loss 5 100 2 ~ (100~v/V!!                      (6)        13.33 kPa Vacuum pressure                       1.5                 25
                                                                                      1.33 kPa Vacuum pressure                        2.0                 15
     Usually, the above expression is negative due to volume
   expansion. Normalize any apparent expansion or contraction                           13.4 Bias:
   among fractions boiling below 150°C in proportion to their                           13.4.1 Absolute Bias—Since there is no accepted reference
   yields.                                                                            material suitable for determining the bias for the procedure in
                                                                                      Test Method D2892 in determining distillation properties of
     NOTE 9—In view of the foregoing rules for establishing yields, the ratio
                                                                                      crude petroleum, bias cannot be determined.
   of mass to volume is not precise enough to be used to calculate the density
   of any distillate fractions or residue.                                              13.4.2 Relative Bias—TBP is defined under the conditions
                                                                                      of this test method (see 1.2).
   12. Report                                                                           NOTE 10—The degrees of freedom associated with the repeatability/
     12.1 A summary sheet for the run must include:                                   reproducibility estimate from this round robin study ranged from 15 to 25.
                                                                                      Since the minimum requirement of 30 (in accordance with Practice
     12.1.1 The mass of the dry sample charged, g,
                                                                                      D6300) is not met, users are cautioned that the actual repeatability/
     12.1.2 The density of the sample at 15°C, g/mL,
     12.1.3 The volume of the sample at 15°C, mL,
     12.1.4 The gain or loss in mass and volume to the nearest                           4
                                                                                           Supporting data have been filed at ASTM International Headquarters and may
   0.1 %,                                                                             be obtained by requesting Research Report RR:D02-1705.
Copyright by ASTM Int'l (all rights reserved);                                   11
                                                                              D2892 – 11a
   reproducibility may be significantly different than these estimates. Given            14. Keywords
   the fact that it is very difficult to get laboratories to participate, coupled
   with the logistical difficulties of shipping precision study samples in the              14.1 boiling point distillation; crude oil distillation; distilla-
   quantities required, the members of Subcommittee D02.08 have decided                  tion; fractional distillation; TBP curves
   to publish all precision data, despite the lower degrees of freedom.
                                                                            ANNEXES
                                                                    (Mandatory Information)
               A1. TEST METHOD FOR THE DETERMINATION OF THE EFFICIENCY OF A DISTILLATION COLUMN
   A1.1 Scope                                                                            described. Cleaning can be accomplished by washing with a
      A1.1.1 This test method is for determining the efficiency of                       strong industrial detergent. Rinse thoroughly, dry, and reas-
   a distillation column, under total reflux conditions using the                        semble.
   test mixture n-heptane/methylcyclohexane at atmospheric                                  A1.5.2 Distill a small quantity of pure n-heptane at a high
   pressure.                                                                             boilup rate for at least 5 min. Take off several small quantities
      A1.1.2 The efficiency is not measured under vacuum con-                            through the overhead sampling system at intervals. Turn off the
   ditions because there is no satisfactory test mixture that has a                      heat, remove the flask, and dry the column with air while still
   constant relative volatility with pressure.                                           hot. To keep the apparatus completely dry, connect a moisture
                                                                                         trap at the vent of the overhead condenser.
   A1.2 Significance and Use
                                                                                         A1.6 Procedure
     A1.2.1 The efficiency of the distillation column must be
   between 14 and 18 theoretical plates to be used in Test Method                           A1.6.1 Introduce into the distillation flask a quantity of test
   D2892 (see 6.1.3.2).                                                                  mixture equal to the minimum volume permitted for the
     A1.2.2 The performance of particulate packings is well                              column (see Table 1) but not more than two thirds of the
   established in the literature. The data shown in Table 1 can be                       capacity of the flask. Add some pieces of glass or porcelain to
   used in place of this test method.                                                    promote even boiling.
                                                                                            A1.6.2 Connect the flask to the pressure drop manometer as
   A1.3 Apparatus                                                                        shown in Fig. A1.1.
      A1.3.1 An example of a suitable apparatus is shown in Fig.                            A1.6.3 Circulate water at ambient temperature in the con-
   A1.1. It consists of the following:                                                   denser.
      A1.3.1.1 Calibration Flask, of suitable size with a device                            A1.6.4 Apply heat to the flask until the test mixture boils,
   for heating. An example of a suitable calibration flask is shown                      then increase the heat progressively up to the flood point. This
   in Fig. A1.2.                                                                         will be noted by visible slugs of liquid in the packing or on the
      A1.3.1.2 Distillation Column and Condenser.                                        plates or by liquid filling the neck of the condenser and a
      A1.3.1.3 Manometer, or equivalent to measure the pressure                          sudden increase in pressure drop. Reduce heat to allow the
   drop in the column.                                                                   flooding to subside. Increase the heat gradually to just below
                                                                                         the flood point. Record boilup and DP measurements just
   A1.4 Reagents and Materials                                                           below the flood point. This is considered the maximum rate.
     A1.4.1 The test mixture is a 50/50 mixture by volume of                                NOTE A1.1—Under these conditions no vapor can reach the head and
   n-heptane and methylcyclohexane with refractive indexes of:                           the heat to the distillation flask must be reduced to establish normal
                    n-heptane                              nD20 = 1.38764                operations again. The flood point is normally determined during the
                    methylcyclohexane                      nD20 = 1.42312                efficiency evaluation of a column using the n-heptane-methylcyclohexane
                                                                                         test mixture (see Annex A1).
      A1.4.2 Chromatographic analysis of the components of the
   test mixture must show less than 0.01 % contamination with                               A1.6.5 Adjust the heat input until the column is refluxing at
   lighter compounds and greater than 99.75 % purity. They must                          a steady rate corresponding to about 200 mL/h times the
   be transparent to ultraviolet light at 260 to 270 nm to ensure                        cross-sectional area of the column in cm2.
   freedom from aromatics.                                                                  A1.6.6 Maintain the column under total reflux for 1 h.
      A1.4.3 If the components do not meet the above specifica-                          Record the pressure drop and determine the boilup rate by
   tion, they can be purified either by redistillation in a 30 to 50                     timing the filling of the calibrated bulb.
   plate column, or by percolation through 200-mesh silica gel,                             A1.6.7 Take rapidly, and almost simultaneously, a sample of
   discarding the first 10 % of the eluted liquid. Exercise care that                    the flask liquid and reflux in sufficient quantity for the
   the gel does not become overloaded.                                                   determination of the refractive indexes at 20°C but not more
                                                                                         than 0.5 mL each. The duration of withdrawal of the reflux
   A1.5 Preparation of Apparatus                                                         sample must not exceed 2 s.
     A1.5.1 The distillation column and all the glassware must                              A1.6.8 Measure refractive indexes of reflux and flask
   be clean and dry before proceeding with the test method                               samples.
Copyright by ASTM Int'l (all rights reserved);                                      12
                                                                             D2892 – 11a
      A1.6.9 Repeat A1.6.7 and A1.6.8 at intervals from 12 to 30                        ria have been determined by Adler et al.6 and relative volatility
   min until refractive index measurements indicate that steady                         is taken as 1.075 as recommended by IUPAC.7
   conditions of maximum efficiency have been attained.                                    A1.7.4 The efficiency of the column is numerically equal to
      A1.6.10 Check the ultraviolet absorption of the flask sample                      the difference between the plate numbers indicated on the
   at 260 and 270 nm. If an absorption is detected, aromatics                           curve for top and bottom samples. Note that one plate is
   contamination is present and the efficiency determination will                       subtracted for the contribution of the flask, or alternatively a
   be in error. The apparatus must be cleaned and the test mixture                      sample of liquid from the bottom of the column rather than the
   repercolated through fresh silica gel before starting over.                          flask can be taken to obtain the efficiency of the column alone.
      A1.6.11 Repeat successively the series of operations A1.6.5                          A1.7.5 For packed columns, draw curves of HETP in
   to A1.6.10, without preflooding, at four or more other boilup                        millimetres, DP in kPa and kPa/m, and number of theoretical
   rates approximately equally spaced between (200 mL/                                  plates as a function of boilup rate expressed in mL/h and
   h) 3 cm2 and the maximum rate. One of these should be at a                           absolute rate in (mL/h) 3 cm2. For real plate columns, the
   rate that is above 90 % of maximum.                                                  efficiency should be plotted as percent of a theoretical plate per
                                                                                        real plate and DP should be plotted both in kPa/real plate and
   A1.7 Calculation                                                                     in absolute units of kPa/theoretical plate using the same units
      A1.7.1 For each pair of samples taken, determine the molar                        for boilup rate as above for packed columns. The efficiency of
   composition of each sample by comparison of its refractive                           the column corresponds to the value of N determined from the
   index with a curve of refractive index versus molar composi-                         curve at 75 % of its maximum boilup rate.
   tion of n-heptane drawn from the following data:                                        A1.7.6 Typical curves for the popular packings and plate
         Molar fraction of n-heptane                 Refractive Index                   columns are shown in Figs. A1.4-A1.7. These columns are to
                     0.00                                1.4231                         be used in the unpreflooded condition. Data on other packings
                     0.10                                1.4191
                     0.20                                1.4151                         are inconclusive or incomplete in the literature and are not
                     0.30                                1.4113                         included.
                     0.40                                1.4076                            A1.7.7 Draw a vertical line on the graph at 75 % of
                     0.50                                1.4040
                     0.60                                1.4005                         maximum boilup and read the HETP or the plate efficiency
                     0.70                                1.3971                         corresponding to this rate.
                     0.80                                1.3939                            A1.7.8 Multiply the HETP in millimetres by 14 and by 18
                     0.90                                1.3907
                     1.00                                1.3876                         for packed columns to obtain the range of permissible heights
                                                                                        of packing for use in this test method. If the efficiency of a
     A1.7.2 Calculate the number of theoretical plates in the                           packed column is greater than 18 plates, as measured above,
   column by means of the Fenske equation5 (Eq A1.1):                                   the efficiency can be reduced by removing a suitable amount of
                                XD          XO                                          packing without detriment to performance.
                          log 1 2 X 2 log 1 2 X
                      N 5
                                   D            O
                                                  21                     (A1.1)
                                                                                           A1.7.9 For real plate columns, divide 14 and 18 by plate
                                   log a                                                efficiency to obtain the number of actual plates (rounded to the
    where:                                                                              nearest integer) for an acceptable column.
    N = number of theoretical plates in the column,                                     A1.8 Precision and Bias
    a    = relative volatility of n-heptane to methylcyclo-
            hexane,                                                                        A1.8.1 No statement is made concerning either the preci-
    XD = molar fraction of n-heptane in the reflux liquid, and                          sion or bias of Annex A1 for measuring the efficiency of a
    XO = molar fraction of n-heptane in the flask liquid.                               distillation column because the result is used to determine
      A1.7.3 Fig. A1.3 is a graphical solution to the equation for                      whether there is conformance to the criteria stated in Test
   the n-heptane-methylcyclohexane binary. Vapor-liquid equilib-                        Method D2892.
     5                                                                                    6
       A more convenient form of the equation from Fenske, M. R., Industrial and              Adler, et al., American Institute of Chemical Engineers , Vol 12, 1966, p. 629.
                                                                                          7
   Engineering Chemistry, Vol 24, 1932, p. 482.                                               Seig, Chemie-Ingenieur-Tecknik, Vol 22, 1950, p. 322.
Copyright by ASTM Int'l (all rights reserved);                                     13
                                                               D2892 – 11a
                                                 FIG. A1.1 Determination of Efficiency
                                                     FIG. A1.2 Boiling Rate Timer
Copyright by ASTM Int'l (all rights reserved);                    14
                                                                       D2892 – 11a
                                                               FIG. A1.3 Refractive Index
                                                 FIG. A1.4 4-mm Propak in 25-mm Inside Diameter Towers
Copyright by ASTM Int'l (all rights reserved);                            15
                                                                       D2892 – 11a
                                                 FIG. A1.5 4-mm Propak in 50-mm Inside Diameter Towers
                                                        FIG. A1.6 6-mm Propak in 50-mm Towers
Copyright by ASTM Int'l (all rights reserved);                            16
                                                                               D2892 – 11a
                                  NOTE—(1) Analytical Chemistry, ANCHA, Vol 39, 1967, p. 286. (2) Exxon unpublished data.
                                                      FIG. A1.7 Perforated Plates 50-mm Inside Diameter
          A2. TEST METHOD FOR THE DETERMINATION OF THE DYNAMIC HOLDUP OF A DISTILLATION COLUMN
   A2.1 Scope                                                                             A2.4 Apparatus
     A2.1.1 This test method is for determining the dynamic                                 A2.4.1 The apparatus is identical to that described in A1.3,
   holdup of a distillation column using a test mixture of stearic                        Fig. A1.1.
   acid in n-heptane.
                                                                                          A2.5 Reagents and Materials
     NOTE A2.1—Dynamic hold-up is expressed as a percentage of the
   packed volume for packed columns so that the data can be compared. For                   A2.5.1 The test mixture consists of 20 mass % stearic acid
   real plate columns, it is expressed in millilitres per plate. The data can only        in n-heptane. The n-heptane shall have a refractive index at
   be compared with others of the same diameter because of different tray                 20°C of 1.3878 6 0.0002. The stearic acid shall be greater than
   spacing. Data for packed columns cannot be compared with those of real                 95 % pure and have a melting point of 68 to 70°C.
   plate columns except in absolute units of millilitres per theoretical plate
   (see Table 1). Dynamic hold-up increases with increasing distillation rate
   up to the flood point and varies from one kind of fractionator to another.             A2.6 Preparation of Apparatus
                                                                                           A2.6.1 See A1.5.
   A2.2 Summary of Test Method
      A2.2.1 A test mixture, composed of stearic acid in                                  A2.7 Procedure
   n-heptane, is distilled under total reflux conditions. From the                          A2.7.1 Measure the concentration of stearic acid in the test
   difference in concentration of stearic acid in the initial mixture                     mixture by a convenient means. For example, titrate with 0.1 N
   and in the mixture during refluxing, the dynamic holdup of the                         NaOH solution to a pH of 9.0 potentiometrically. Record the
   column is calculated.                                                                  results as mass % stearic acid (PO).
                                                                                            A2.7.2 Introduce pieces of glass or porcelain into the flask
   A2.3 Significance and Use                                                              or use a good stirrer to promote even boiling.
      A2.3.1 The amount of sample charged to a particular                                   A2.7.3 Add 1 L of the test mixture for a 25-mm inside
   distillation column must be of such a size that the dynamic                            diameter column or 4 L for a 50-mm column to the flask.
   holdup of that column is between 2 and 4 % of that charge size                         Weigh the flask to the nearest 1 g.
   at 75 % boilup rate (see 9.1.1).                                                         A2.7.4 Attach the flask to the distillation column and to the
      A2.3.2 The performance of the particle packings is well                             DP measuring system.
   enough defined in the literature that the data in Table 1 can be                         A2.7.5 Circulate water at ambient temperature through the
   used instead of this test method.                                                      condenser.
Copyright by ASTM Int'l (all rights reserved);                                       17
                                                                    D2892 – 11a
      A2.7.6 Apply heat to the flask and bring the test mixture to            d     = density of n-heptane, 0.688 g/mL at 15°C.
   a boil. Adjust the boilup rate to approximately 200 mL/h times               A2.8.2 Calculate the dynamic holdup per theoretical plate at
   the cross-sectional area of the column in centimetres squared              each rate of boilup at which the determination was made, using
   measured by timing the filling of the calibrated bulb. When the            Eq A2.2:
   desired rate has been established, hold for 30 min noting the
                                                                                                               H
   pressure drop.                                                                                           h 5N                           (A2.2)
      A2.7.7 Sample the reflux and the flask liquids, taking
   samples of not more than the amount necessary for a determi-                where:
   nation, and immediately observe and record the boilup rate and              N = efficiency in theoretical plates of the column under
   DP.                                                                                 total reflux at this boilup rate (see Annex A1).
      A2.7.8 Analyze the reflux and flask samples to determine                   A2.8.3 Convert the boilup measurement in litres per hour to
   the concentration of stearic acid in mass % (P).                           rates per hour in (mL/h) 3 cm2. Convert also the DP measure-
      A2.7.9 Repeat A2.7.7 and A2.7.8 at 15 min intervals until               ments to kPa/m (mm Hg/m) and the holdup measurement to
   the concentration of stearic acid in the flask sample is steady.           millilitres per theoretical plate.
      A2.7.10 Raise the boilup rate by an additional 200 (mL/                    A2.8.4 Plot all data as ordinates versus boilup rate in litres
   h) 3 cm2 and repeat A2.7.7-A2.7.9.                                         per hour and (mL/h) 3 cm2 as abscissa. Draw smooth curves
      A2.7.11 Continue making measurements as above at incre-                 through the points for holdup in millilitres and in millilitres per
   ments of about 200 (mL/h) 3 cm2 until near the flood point. At             theoretical plate. The boilup versus DP measurements should
   least four sets of measurements should be obtained including               be compared with those made in the efficiency measurements if
   one near the maximum operable rate.                                        available, to ensure that they are in reasonable agreement.
   A2.8 Calculation                                                              A2.8.5 Draw a vertical line at 75 % of maximum boilup
     A2.8.1 Calculate the dynamic holdup of the column for                    rate. The dynamic holdup for the column will be that read from
   each observation, using Eq A2.1:                                           the intersection of the holdup curve and the line for 75 % of
                                                                              maximum boilup rate. The charge size, ranging from a holdup
                                      P 2 PO M                                ratio from 2 to 4 % will thus be from 50 to 25 times the above
                                H 5     P 3 d                   (A2.1)
                                                                              figure.
    where:
    H =      dynamic holdup in column, mL at 15°C,                            A2.9 Precision and Bias
    M =      mass of test mixture in the flask, g,                               A2.9.1 No statement is made concerning either the preci-
    PO =     stearic acid in the test mixture initially, mass %,              sion or bias of Annex A2 for measuring dynamic holdup
    P =      stearic acid in the test mixture after distillation, mass        because the result is used to determine whether there is
             %, and                                                           conformance to the criteria stated in Test Method D2892.
                A3. TEST METHOD FOR THE DETERMINATION OF THE HEAT LOSS IN A DISTILLATION COLUMN
                                             (STATIC CONDITIONS)
   A3.1 Scope                                                                   A3.3.2 The test should be performed on all new glass
      A3.1.1 This test method is for determining the heat loss of             vacuum jackets before use, and checked at least once per year
   a distillation column under static conditions when a tempera-              thereafter.
   ture differential exists between the inner and outer walls of a              A3.3.3 The heat loss as determined by this test method must
   distillation column.                                                       be less than 30 % for the column to be acceptable for use in
                                                                              Test Method D2892.
   A3.2 Summary of Test Method
      A3.2.1 The outer wall of the column vacuum jacket is                    A3.4 Apparatus
   maintained at an elevated constant temperature. The tempera-
   ture increase inside the column, as recorded by the sensor in                A3.4.1 The column is enclosed in its heat compensating
   the reflux divider, is a measure of the heat gained and thus heat          mantle with the thermocouple in place on the column wall.
   lost by the column.                                                          A3.4.2 A twin pen chart recorder to monitor the column
                                                                              outer wall and heat sensor temperatures and an automatic
   A3.3 Significance and Use
                                                                              proportioning controller for the heat input are recommended.
      A3.3.1 It is important to have an effective silvered glass
   vacuum jacket surrounding the column and reflux divider. This              A3.5 Preparation of Apparatus
   reduces the effects of ambient air temperature near the distil-
   lation apparatus and promotes easier control at the maximum                   A3.5.1 The heat sensor location, response time, and calibra-
   distillation temperatures. The use of a heat compensating                  tion must be checked as specified in Annex A4 to Annex A6.
   mantle further reduces losses by reducing the temperature                     A3.5.2 To reduce chimney effects inside the column during
   gradient between inside of the column and the ambient air.                 the test, the top of the condenser must be closed.
Copyright by ASTM Int'l (all rights reserved);                           18
                                                                 D2892 – 11a
   A3.6 Procedure                                                                                     B2A
                                                                                                  Q 5 C 2 A 3 100                     (A3.1)
      A3.6.1 Record the time and ambient temperature.
      A3.6.2 Apply heat to the outer wall of the column vacuum             where:
   jacket. Adjust the heat progressively until the temperature             A = ambient temperature, °C,
   sensor located on the column wall records a temperature 100°C           B = temperature inside column, °C, and
   above ambient. This condition provides a suitable temperature           C = temperature of outer wall, °C.
   difference to measure without placing unnecessary thermal                or, when the differential temperature is 100°C, use Eq A3.2:
   strain on the glassware. Attain the 100°C differential tempera-
   ture within 30 min and maintain the column at this temperature                                     Q5B2A                           (A3.2)
   for 1 h.
      A3.6.3 Record the time, the temperature of column outer              A3.8 Precision and Bias
   wall, and the temperature inside the column.                               A3.8.1 No statement is made concerning either the preci-
   A3.7 Calculation                                                        sion or bias of Annex A3 for measuring heat loss because the
     A3.7.1 Calculate the heat gained and thus lost by the                 result is used to determine whether there is conformance to the
   column using Eq A3.1:                                                   stated criteria (see A3.3.3).
                         A4. TEST METHOD FOR THE VERIFICATION OF TEMPERATURE SENSOR LOCATION
   A4.1 Scope                                                                 A4.4.1.4 If the observed boiling point is not 253.5 6 0.5°C,
     A4.1.1 This test method is for determining whether the                the location of the sensor is suspect and must be corrected
   temperature sensor is in the proper position for optimum                before use in this test method.
   performance.                                                               A4.4.2 Vacuum Distillation:
   A4.2 Summary of Test Method                                                A4.4.2.1 Assemble the apparatus for vacuum distillation.
                                                                           Charge 0.5 to 1 L of pure, dry n-hexadecane containing less
     A4.2.1 The vapor temperature of a pure compound mea-
                                                                           than 0.1 % light contaminants as determined by gas chroma-
   sured by the sensor and its recording instrument is compared to
                                                                           tography. Calibrate the temperature and vacuum instruments as
   the accepted boiling point for the compound. The test is
                                                                           prescribed in Annex A6.
   conducted both at atmospheric pressure and under vacuum of
   0.133 kPa (1 mm Hg).                                                       A4.4.2.2 Reduce the pressure in the system to 0.133 kPa (1
                                                                           mm Hg). Circulate air in the condenser to avoid crystallization.
   A4.3 Significance and Use
                                                                              A4.4.2.3 Apply heat and establish equilibrium at a boilup
      A4.3.1 A poorly positioned sensor can give temperatures              rate between 25 and 50 (mL/h) 3 cm2. This corresponds to a
   that are in error due to inadequate heat supply from the vapors.        DP of about 0.133 kPa/m (1 mm Hg/m) or less for particle
   It is especially important under vacuum when heat content of            packing or 0.08 kPa/plate (0.06 mm Hg/plate) for real plate
   the vapor is at a minimum.                                              columns to 50-mm diameter. Hold at these conditions for 15
      A4.3.2 The procedure is normally performed once only for             min.
   the approval of a design and need not be repeated thereafter.
                                                                              A4.4.2.4 Remove 2 % overhead at 20:1 reflux ratio and then
   A4.4 Procedure                                                          again hold at total reflux for 5 min. Record the boiling point to
      A4.4.1 Atmospheric Distillation:                                     the nearest 0.5°C and the operating pressure to the nearest 0.05
      A4.4.1.1 Assemble the apparatus for atmospheric distilla-            kPa (0.3 mm Hg), using instruments calibrated as described in
   tion. Use cooling water at ambient temperature. Charge 0.5 to           Annex A6.
   1 L of pure, dry n-tetradecane containing less than 0.1 % light            A4.4.2.5 If the observed boiling point is not steady with
   contaminants as determined by gas chromatography. Calibrate             0.2°C at 0.133 kPa (1 mm Hg), remove 2 % cuts at 5:1 reflux
   the temperature sensor as prescribed in Annex A6.                       ratio and then hold at total reflux for 15 min, repeating the
      A4.4.1.2 Apply heat and establish equilibrium at a boilup            above until total reflux conditions produce a steady tempera-
   rate of about 400 (mL/h) 3 cm2. This corresponds to a DP of             ture. Not more than three trials should be needed.
   about 0.4 kPa/m (3 mm Hg/m) for particle packing or 0.09                   A4.4.2.6 If the observed boiling point is not 105.2 6 0.5°C
   kPa/plate (0.07 mm Hg/plate) for real plate columns. Hold at            at 0.133 kPa (1 mm Hg), the location of the sensor is suspect
   these conditions for 15 min. If the vapor temperature drops             and must be corrected before use in this test method.
   more than 0.2°C, remove 2 % overhead at 5:1 reflux ratio and
   then again hold at total reflux for 15 min. Continue taking 2 %
                                                                           A4.5 Precision and Bias
   cuts as above until the vapor temperature remains steady within
   0.2°C for 15 min at total reflux.                                          A4.5.1 No statement is made concerning either the preci-
      A4.4.1.3 Record the boiling point to the nearest 0.5°C and           sion or bias of Annex A4 for checking the location of the
   the atmospheric pressure to the nearest 0.1 kPa (1 mm Hg),              temperature sensor because the result is used to determine
   using instruments calibrated in accordance with Annex A6.               whether there is conformance to the criteria stated in A4.4.2.6.
Copyright by ASTM Int'l (all rights reserved);                        19
                                                                  D2892 – 11a
                            A5. TEST METHOD FOR DETERMINATION OF TEMPERATURE RESPONSE TIME
   A5.1 Scope                                                               connect the sensor to a strip chart recorder of suitable range
      A5.1.1 This test method is for the determination of tempera-          allowing interpolation to 0.1°C. Set the chart speed at 30 cm/h
   ture response time based upon the rate of cooling of the sensor          for readability.
   under prescribed conditions.                                                A5.3.3 Insert the sensor into a hole in the center of one side
                                                                            of a closed cardboard box about 30 cm on a side. Hold the
   A5.2 Significance and Use                                                sensor in place by a friction fit on the joint. Allow the sensor to
                                                                            reach equilibrium temperature. Record the temperature when it
     A5.2.1 This test method is performed to ensure that the
                                                                            becomes stable.
   sensor is able to respond to changes in temperature fast enough
                                                                               A5.3.4 Remove the sensor and insert it into the heated
   that no error due to lag is introduced in a rapidly rising
                                                                            thermowell in the beaker of water. After the sensor has reached
   temperature curve.
                                                                            a temperature of 80°C, remove it and immediately insert it into
     A5.2.2 The importance of this test method is greatest under
                                                                            the hole in the box. Note with a stopwatch, or record on the
   vacuum conditions when the heat content of the vapors is
                                                                            strip chart, the time interval while the sensor cools from 30°C
   minimal.
                                                                            above to 5°C above the temperature recorded in A5.3.3.
                                                                               A5.3.5 A time interval in excess of 175 s is unacceptable.
   A5.3 Procedure
     A5.3.1 Arrange a 1-L beaker of water on a hot plate with a             A5.4 Precision and Bias
   glass thermowell supported vertically in the water. Maintain                A5.4.1 No statement is made concerning either the preci-
   the temperature of the water at 90 6 5°C.                                sion or bias of Annex A5 for measuring temperature response
     A5.3.2 Correct the sensor to an instrument, preferably with            because the result is used to determine whether there is
   a digital readout, with readability to 0.1°C. Alternatively,             conformance to the criteria stated in A5.3.5.
                                                 A6. PRACTICE FOR CALIBRATION OF SENSORS
   A6.1 Principle
     A6.1.1 This practice deals with the basic calibration of
   temperature sensors and vacuum sensors and their associated
   recording instruments.
     A6.1.2 The temperature sensor and its associated instrument
   are calibrated by observing and recording the temperatures of
   the melting point and boiling point of pure compounds or
   eutectic mixtures.
     A6.1.3 The vacuum sensor and its associated instrument are
   calibrated against a McLeod gauge.
   A6.2 Temperature Sensors
     A6.2.1 Apparatus—A suitable apparatus is shown in Fig.
   A6.1. For the freezing point of water, a Dewar flask filled with
   crushed ice and water can be substituted. For the boiling point
   of water, use an equilibrium still or ebulliometer, a tensimeter,
   or other apparatus for measuring vapor-liquid equilibrium.
     A6.2.2 Procedure:
     A6.2.2.1 Ensure that approximately 0.5 mL of silicone oil or
   other inert liquid is in the bottom of the thermowell and insert
   one or more thermocouples or other sensors connected to their
   respective measuring instruments.
     A6.2.2.2 Heat the melting point bath to a temperature 10°C
   above the melting point of the metal inside and hold this
   temperature at least 5 min to ensure that all of the metal is
   melted.
     A6.2.2.3 Discontinue heat input to the melting point bath
   and observe and record the cooling curve. When the curve
                                                                                 FIG. A6.1 Melting Point Bath for Temperature Standards
   exhibits a plateau of constant temperature for longer than 1
   min, the temperature of the recorded plateau is accepted as the
Copyright by ASTM Int'l (all rights reserved);                         20
                                                                               D2892 – 11a
   calibration temperature. If the freezing plateau is too short, it                  nometer is the primary reference standard. Fig. A6.3 shows a
   can be prolonged by employing some heat during the cooling                         convenient design in which any internal air above the mercury
   cycle. Alternatively, the melt bath may have become contami-                       is trapped in the fine tip at the top where it can be seen at a
   nated or excessively oxidized. In this case, replace the metal.                    glance. For a suitably evacuated gauge, there is no air in the tip
     A6.2.2.4 Record the calibration temperature at each of the                       visible to the unaided eye when vented. The glass shall be
   following points to the nearest 0.1°C:                                             chemically clean inside.
                  Material                             Temperature, °C
   Ice                                    melting point                  0.0
                                                                                         A6.3.1.3 If air can be seen in the tip, clamp the gauge in a
   Water                                  boiling point                100.0          horizontal position with the vacuum connection facing upward
   Tin: Lead: Cadmium (50:32:18)          melting point                145.0          and the top of the gauge lower than the bottom so as to expose
   Sn                                     melting point                231.9
   Pb                                     melting point                327.4
                                                                                      the hole in the bottom of the central tube. Apply a vacuum of
                                                                                      less than 0.0133 kPa (0.1 mm Hg) and heat the end of the
      A6.2.2.5 Set up a correction table by listing the correction to                 gauge containing the mercury with an infrared heat lamp or a
   be added algebraically to the observed temperature to give the                     hot air gun until the mercury comes to its boiling point.
   true temperature at each calibration point. A graphical plot of
                                                                                      Continue heating slowly until the mercury in the central tube
   the above corrections connected by a smooth curve may be
                                                                                      has partly distilled out into the outer tube. Condensed mercury
   helpful in routine use.
                                                                                      on the wall will be evidence of this. Return the manometer to
   A6.3 Vacuum Sensors                                                                the vertical position and slowly release the vacuum. Verify that
      A6.3.1 Apparatus:                                                               there is no air visible at the tip of the central tube when fully
      A6.3.1.1 Assemble a vacuum manifold such as that shown                          vented. (Warning—Mercury vapor is poisonous. Harmful or
   in Fig. A6.2. It shall be capable of maintaining steady pressures                  fatal if inhaled or ingested.)
   within 1% at all desired levels.                                                      A6.3.1.4 For pressures below 13.3 kPa, the nontilting
      A6.3.1.2 For pressures between atmospheric down to 13.3                         McLeod gauge is the primary standard and shall be carefully
   kPa (100 mm Hg), an evacuated mercury barometer or ma-                             maintained.
                                                           FIG. A6.2 Calibration of Vacuum Gauges
Copyright by ASTM Int'l (all rights reserved);                                   21
                                                                D2892 – 11a
                                                                          pressure below 10 Pa (0.075 mm Hg). Thereafter, carefully
                                                                          protect the reference gauge from exposure to moisture such as
                                                                          that from atmospheric air. The use of two reference McLeod
                                                                          gauges of different pressure ranges is recommended as a
                                                                          precaution. If they agree at the test pressure, it is an indication
                                                                          that the system is free of moisture and other condensables.
                                                                            NOTE A6.1—The general principles of construction of McLeod gauges
                                                                          are well-established. The dimensions and tolerances of such a gauge are
                                                                          beyond the scope of this test method.
                                                                             A6.3.1.6 Alternatively, certified secondary gauges, elec-
                                                                          tronic or otherwise, can be used, provided the output can be
                                                                          traced back to a primary standard. Secondary gauges shall be
                                                                          recertified at a regular basis, but at least once a year.
                                                                             A6.3.2 Procedure:
                                                                             A6.3.2.1 Set up the test manifold such as that shown in Fig.
                                                                          A6.2. Ensure that the test manifold is leak-free and can be
                                                                          maintained at a steady pressure at the required level. A suitable
                                                                          leak test is to pump down to a pressure below 0.1 kPa and
                                                                          isolate the pump. Observe the pressure inside the unit for at
                                                                          least 1 min. If the pressure rises no more than 0.01 kPa in that
                                                                          period, the apparatus is considered acceptable.
                                                                             A6.3.2.2 Connect the reference (primary) vacuum gauge(s)
                                                                          and the gauge(s) to be calibrated to the manifold. The gauges
                                                                          shall have such a range that the desired calibration pressure
                                                                          falls between 10 and 90 % of the scale. Insert a dry ice trap
                                                                          between the manifold and the vacuum pump. Adjust the
                                                                          pressure to the required level for the test and run a final leak
                                                                          test as above.
                                                                             A6.3.2.3 After steady conditions have been maintained for
                                                                          at least 3 min, readings are made of all gauges and compared
                        FIG. A6.3 Vacuum Manometer
                                                                          with the reference gauge.
                                                                             A6.3.2.4 Repeat the above procedure at the other required
      A6.3.1.5 Choose a McLeod gauge with a range such that the           pressure levels.
   desired calibration pressure falls between 10 and 90 % of the             A6.3.2.5 Make up a chart of corrections to be added at each
   scale. Before refilling with clean mercury, heat the empty             pressure level for each gauge tested. This can be used for
   reference McLeod gauge at 250°C for at least 30 min at a               interpolation when necessary.
                                  A7. TEST METHOD FOR VERIFICATION OF REFLUX DIVIDING VALVES
   A7.1 Scope                                                             A7.4 Apparatus
      A7.1.1 This test method is for determining whether a liquid            A7.4.1 The apparatus consists of the column and reflux
   reflux dividing valve produces the prescribed reflux ratio.            divider with its controller and the condenser if necessary for
                                                                          convenient entry of liquid to the divider.
   A7.2 Summary of Test Method                                               A7.4.2 A means must be provided for introducing the test
                                                                          liquid at steady rates over a range that includes 10 to 90 % of
      A7.2.1 A hydrocarbon distillate of medium density is intro-
                                                                          the maximum capacity of the column under test. This apparatus
   duced to the reflux dividing valve while operating in the
                                                                          can be a pumping system, but a simple gravity flow system can
   normal way. The reflux ratio is determined by the ratio of the
                                                                          be used if its reservoir is of a capacity that will ensure
   two streams so obtained. The test is conducted over a range of
                                                                          relatively uniform rates for each test. A supply of 400-mL
   rates normally encountered in use to ensure that performance is
                                                                          beakers tared to the nearest gram and a stopwatch are also
   acceptable at all levels.
                                                                          required.
   A7.3 Significance and Use                                              A7.5 Reagents and Materials
      A7.3.1 This test method is intended to ensure that the valve          A7.5.1 The test liquid can be any hydrocarbon fraction in
   in operation actually divides the reflux in the desired ratio.         the kerosine to diesel oil range (from 150 to 300°C).
Copyright by ASTM Int'l (all rights reserved);                       22
                                                                             D2892 – 11a
   A7.6 Preparation of Apparatus                                                           A7.7.9 Repeat sequence A7.7.1-A7.7.7 for a valve opera-
      A7.6.1 Assemble the column in the normal way with the                             tion in which the closed portion of the cycle is equal to two
   valve in place. Attach the condenser in its normal position if                       times the open period (2:1 ratio).
   this will facilitate the introduction of the test liquid to the                         A7.7.10 When the actual reflux ratio differs by more than
   divider. Connect the control device to the liquid dividing valve.                    10 % from the desired ratio, the valve is not acceptable and
                                                                                        must be corrected. Fig. A7.1 illustrates graphically the test
      A7.6.2 Set an untared beaker of suitable size under the                           results of a typical valve of acceptable performance.
   bottom of the column and another under the takeoff point.
      A7.6.3 Mount the liquid flow system so that the necessary                         A7.8 Precision and Bias
   range of flow rates can be provided.                                                    A7.8.1 No statement is made concerning either the preci-
                                                                                        sion or bias of Annex A7 for checking the location of the
   A7.7 Procedure                                                                       temperature sensor because the result is used to determine
                                                                                        whether there is conformance to the criteria stated in A7.7.10.
      A7.7.1 Start the valve and set the control device in such a
   way that the valve is open for not less than 4 s and not more
   than 6 s, and closed for a period that is five times as long (5:1
   ratio).
      A7.7.2 Commence introduction of the test liquid to the
   valve at a rate equal to about 10 % of the maximum boilup rate
   for the column under test.
      A7.7.3 After 2 min have elapsed, simultaneously replace the
   beakers under the column and under the takeoff point with
   tared beakers.
      A7.7.4 A precise 5 min intervals, simultaneously replace
   those beakers with another pair of tared beakers. After three
   sets of tared beakers have been collected, return the untared
   beakers and shut off the flow of liquid.
      A7.7.5 Weigh each of the tared beakers and obtain the mass
   of liquid in each to the nearest gram.
      A7.7.6 Calculate the ratio of the mass of test liquid recov-
   ered in the beaker at the bottom of the column to that at the top
   of the column in each of the three 5-min tests.
      A7.7.7 If the ratios so obtained are not consistent to within
   65 %, repeat A7.7.1 through A7.7.6.
      A7.7.8 Repeat the sequence A7.7.1-A7.7.7 at flow rates
   equal to about 30, 60, and 90 % of the maximum boilup rates                          FIG. A7.1 Graphical Illustration of Actual Reflux Ratio of a Typical
   for the column under test.                                                                                      Reflux Divider
          A8. PRACTICE FOR CONVERSION OF OBSERVED VAPOR TEMPERATURE TO ATMOSPHERIC EQUIVALENT
                                            TEMPERATURE (AET)
   A8.1 Scope                                                                           A8.3 Calculation
      A8.1.1 This practice is for conversion of the actual distilla-                      A8.3.1 Convert observed vapor temperature to atmospheric
   tion temperature obtained at sub-ambient pressure to AET                             equivalent temperature using Eq A8.1:
   corresponding to the equivalent boiling point at atmospheric
                                                                                                                       748.1A
   pressure, 101.3 kPa (760 mm Hg), by means of equations                                     AET 5                                          2 273.1
                                                                                                      @1/~T 1 273.1!# 1 0.3861A 2 0.00051606
   derived by Maxwell and Bonnell.8                                                                                                                    (A8.1)
   A8.2 Significance and Use                                                            where:
     A8.2.1 Final data on atmospheric equivalent temperatures                           AET = atmospheric equivalent temperature, °C, and
   are to be obtained by computation.                                                   T     = observed vapor temperature, °C.
                                                                                          A8.3.1.1 Calculate A using Eq A8.2 or Eq A8.3:
      8
        Maxwell and Bonnell, Industrial Engineering Chemistry , Vol 49, 1957, p.                              5.143222 2 0.972546 log10 P
                                                                                                         A5     2579.329 2 95.76 log10 P               (A8.2)
   1187.
Copyright by ASTM Int'l (all rights reserved);                                     23
                                                                        D2892 – 11a
    where:                                                                         where:
    P = operating pressure, kPa, (operating pressure $0.266                        B = mean average boiling point, °C, and
         kPa), or                                                                  D = relative density at 15.6/15.6C.
                      5.994295 2 0.972546 log10 P                                    A8.3.3.1 By custom, either the mid vapor temperature of the
                  A 5 2663.129 2 95.76 log P          (A8.3)                      fraction or the midpoint of a gas chromatographic distillation
                                             10
                                                                                  of the fraction can be used for the mean average boiling point.
    where:                                                                        In either case the method must be specified.
    P = operating pressure, mm Hg (operating pressure $2 mm                          A8.3.3.2 An estimate of the K-factor can be made using Fig.
          Hg).                                                                    A8.1.
     A8.3.2 The equations are correct only for fractions that have                   A8.3.4 Calculate the correction to be applied to the AET
   a Watson K-factor of 12.0 6 0.2. The K-factor shall be                         using Eq A8.5:
   assumed to be 12 and any effect of K-factor ignored unless
   there is mutual agreement to the contrary.                                                                      F S DG Pa
                                                                                                   t 5 21.4[K 2 12] log10 P
                                                                                                                            o
                                                                                                                                           (A8.5)
     A8.3.3 If correction is required, calculate the K-factor using
                                                                                   where:
   Eq A8.4:                                                                        t = correction, °C
                                   3
                                   = 1.8~B 1 273.1!                                Pa = atmospheric pressure, kPa (mm Hg), and
                             K5            D                        (A8.4)         Po = observed pressure, kPa (mm Hg).
                                                                                    A8.3.4.1 An estimate of the correction can be made using
                                                                                  Fig. A8.2.
                                             FIG. A8.1 Watson Characterization Factor of Petroleum Fractions
Copyright by ASTM Int'l (all rights reserved);                               24
                                                                    D2892 – 11a
                                                 FIG. A8.2 Boiling Point Corrections for K-Factor
                                                                APPENDIXES
                                                          (Nonmandatory Information)
                                   X1. PRACTICE FOR DEHYDRATION OF A SAMPLE OF WET CRUDE OIL
   X1.1 Scope                                                                X1.4 Apparatus
     X1.1.1 This practice is for dehydrating a sample of wet                    X1.4.1 The dehydration of a sample of wet crude oil
   crude oil prior to fractional distillation.                               requires apparatus, such as that shown in Fig. 1, composed of:
   X1.2 Summary of Practice                                                     X1.4.1.1 Distillation Flask, with two side arms. In place of
                                                                             the differential pressure manometer in the second sidearm, a
      X1.2.1 A sufficient quantity of the sample is distilled in a
   low efficiency column under atmospheric pressure at zero                  capillary is fitted for the passage of nitrogen into the liquid.
   reflux ratio (total takeoff) to 130°C, the water decanted, and dry        When a sample is suspected of containing emulsified water or
   components recombined.                                                    significant amounts of clay or sediment, or both, additional risk
                                                                             to glass apparatus is involved. In this case, remove gross water
   X1.3 Significance and Use                                                 and sediment and use a metal flask for dewatering.
      X1.3.1 Dehydration is important in order to achieve accu-                 X1.4.1.2 Distillation Column, shall be of the type described
   rate yields in the light naphtha region.                                  in 6.1.3.
Copyright by ASTM Int'l (all rights reserved);                          25
                                                                            D2892 – 11a
     X1.4.1.3 The rest of the apparatus is identical to that                              X1.6.5 To separate the water from each distillate fraction,
   described in 6.1.                                                                   cool to − 5°C and decant the hydrocarbon liquid. Weigh the
                                                                                       water.
   X1.5 Preparation of Apparatus                                                          X1.6.6 Remove the condenser and rinse it with alcohol or
     X1.5.1 Clean the distillation column and all the glassware                        acetone to remove adhering drops of water. Dry with air and
   before starting the test.                                                           replace it.
                                                                                          X1.6.7 Recombine the cooled decanted fractions with the
   X1.6 Procedure                                                                      distillation residue observing the usual precautions against
      X1.6.1 Cool the charge to a temperature not lower than 0°C.                      losses. Do not recombine the trap fraction.
   Decant any bulk water that may be present. Weigh by differ-                            X1.6.8 Record the quantity of dry oil recovered. If reblend-
   ence to the nearest gram, into a chilled distillation flask                         ing of the dried fractions was done in the original flask, this
   containing some pieces of glass or porcelain, a given volume of                     flask can be used for subsequent distillation.
   wet crude oil.
      X1.6.2 Attach the flask to the column and pass a slow                            X1.7 Calculation
   stream of nitrogen (8 cm3/s) through the capillary. Vent the                         X1.7.1 Calculate the mass % of water using Eq X1.1.
   condenser through two traps in series maintained at the
                                                                                                                 W 5 100~A/B!                   (X1.1)
   temperature of dry ice. Circulate coolant at a temperature
   of − 20°C in the condensers.                                                        where:
      X1.6.3 Apply heat to the flask, regulating it to attain a                        A        =   mass of water recovered, g,
   moderate reflux rate as specified in 10.2.4 to 10.2.8 for                           B        =   mass of charge, g,
   debutanization. Remove distillate slowly at total takeoff (reflux                   W        =   mass % of water, and
   ratio = 0) until a vapor temperature of 130°C is reached.                           100      =   percentage constant.
   Collect the fraction distilling below 65°C in a receiver cooled
   to − 20°C or lower.                                                                 X1.8 Precision and Bias
      X1.6.4 Shut off the reflux valve and the heating system.                            X1.8.1 No statement is made concerning either the preci-
   Cool the flask and contents to ambient temperature. Maintain                        sion or bias of Appendix X1 for mass % water because the test
   the traps at the temperature of dry ice. Weigh the distillate                       method is used primarily for sample preparation for Test
   fractions obtained.                                                                 Method D2892.
                                                    X2. PRACTICE FOR PERFORMANCE CHECK
   X2.1 Scope                                                                          scribed and is recommended. For convenience sake, a simple
      X2.1.1 This practice covers a procedure for calculating                          graphical solution based on the same method is also included.
   column performance from GC boiling point distributions on                              X2.1.4 Overall column performance is assessed in terms of
   fractions and residues, obtained by distilling an average (30 to                    column efficiency (minimum tray number) and in terms of the
   40 API-gravity) crude oil under actual Test Method D2892                            differential between the nominal cut point (AET) and the
   distillation conditions.                                                            calculated effective cut point (ECP). Criteria are given for
     NOTE X2.1—There are no theoretical reasons to limit the API-gravity               acceptance of column performance. Possible corrective ac-
   range from 30 to 40. However, the use of a crude oil in the quoted range            tion(s), if required, are also indicated.
   will, more or less, ensure that sufficient product is available to assess
   performance, both at the upper and lower end of the temperature scale.              X2.2 Significance and Use
   The use of heavier crudes may not yield sufficient quantities at the low
   end, while the reverse is true for lighter crudes.                                     X2.2.1 Good agreement in yield of fractions can be
      X2.1.2 The assessment of column performance can be made                          achieved under a variety of conditions because efficiency has
   at any cut point where samples of two adjacent fractions or a                       no measurable effect on fraction yields except for very low
   residue can be analyzed by gas chromatography. Either fresh or                      efficiencies and at the beginning and end of a distillation.
   stored samples can be analyzed, as long as they have been                           However, fractions produced at high efficiency will have a
   protected from loss by evaporation. Recommendations are                             narrower boiling range and hence some different properties
   given for the number and spacing of cut points to be analyzed                       than the same fraction made at low efficiency. To arrive at a
   for performance.                                                                    standard level of efficiency and standard fraction quality, an
      X2.1.3 A precise mathematical method9 for the calculation                        overall check is recommended to assess the actual performance
   of distillation efficiency of multi-component mixtures is de-                       of the system. If the results of this test (Appendix X2) are
                                                                                       unacceptable, then the basic performance determinations as
      9
        Butler and Pasternak, The Canadian Journal of Chemical Engineering, Vol        outlined in Annex A1-Annex A5 inclusive and Annex A7
   42, 1964, p. 47.                                                                    should be considered to determine the cause. The results of the
Copyright by ASTM Int'l (all rights reserved);                                    26
                                                                                D2892 – 11a
   overall performance check as described in this appendix                                 provided that they are based on the same principles of boiling
   provides clues for possible causes and corrective actions in                            point separation and calibration, that is, employ a methylsili-
   such a case.                                                                            cone column and retention times calibrated for boiling point by
                                                                                           paraffins with boiling points assigned as given in Test Method
   X2.3 Summary of Practice                                                                D2887.
      X2.3.1 The distribution of (pseudo) components between
                                                                                             NOTE X2.4—A high temperature SIMDIST method is still under
   the overlapping tail and front of two adjacent fractions (or a                          development by ASTM D02.04.H and in its present draft is only
   residue) is calculated from GC analysis. The linearized trans-                          applicable to petroleum distillates. However, the methodology is widely
   form of the distribution coefficient is plotted against the boiling                     used in the industry to measure the boiling range distribution of residues
   point of the (pseudo) component. By linear regression, the best                         as well. Hence, it is envisaged that at a later date residues will be included
   possible straight line is obtained. The efficiency is calculated                        in the scope of this proposed test method.
   from the slope of the line by means of a modified Fenske                                   X2.4.2 Using the results of the GC analyses of two con-
   equation. The ECP is defined as the temperature at which the                            secutive distillate fractions, formulate groups defined by
   overlaps are equal on the basis of mass-percent of the charge,                          equally spaced temperature ranges (pseudo components). The
   or in mathematical terms, the temperature at which the distri-                          groups shall be formed such that the temperature range of the
   bution coefficient equals 0.5. This form of the efficiency                              first group and that of the last group bracket the overlapping
   calculation is mathematically identical to the graphical proce-                         portions of the distillate fractions. For each group (pseudo
   dure described by Butler and Pasternak.                                                 component) list the noncumulative yields (% M/M on whole
                                                                                           fraction, referred to as Yield o.f.) contained within that tem-
   X2.4 Procedure
                                                                                           perature range in each fraction. The temperature range used
      X2.4.1 Obtain samples of contiguous fractions or a residue                           should have a width not smaller than 5°C nor wider then 15°C.
   for each level of pressure and analyze them according to an                             The GC slices of adjacent distillate cuts by Test Method D2892
   appropriate boiling point distribution method. The fractions                            shall have the same start and end boiling point temperature. To
   shall be wide enough so that the overlap by GC analysis does                            ensure sufficient accuracy, within the constraint of fraction
   not extend beyond the fraction. Uniform 25°C wide fractions                             width, at least four non-outlying components should be iden-
   are recommended, but wider fractions can be employed with-                              tifiable with a distribution factor $0.05 and #0.95 (see
   out deterioration of the reliability of the assessment. The                             X2.5.1).
   recommended number of fractions to be obtained and the                                     X2.4.3 Calculate the yield of the (pseudo) components in %
   spacing of the cut points is discussed in X2.6. The range of                            M/M on the original (Test Method D2892) charge mass:
   application of the available boiling point distribution GC test
                                                                                                                                       mass fraction, g
   methods, together with the recommended application to cuts by                                            Yield o.c. 5 Yield o. f. 3 mass charge, g             (X2.1)
   Test Method D2892 is listed in X2.4.1.1-X2.4.1.3. Although
   adjacent fractions may be analyzed by different test methods, it                          X2.4.3.1 Assign a boiling point to each (pseudo) component
   is highly recommended that adjacent fractions be analyzed by                            exactly halfway the start and end boiling point temperature of
   the same test method.                                                                   the (pseudo) component.
      NOTE X2.2—Different GC boiling range test methods might show                            NOTE X2.5—If actual components as obtained by Test Method D5134
   relative bias towards each other. If such a bias exists it will affect the slope        are used in the calculations, assign a boiling point as obtained from
   of the regression line and, therefore, could lead to misleading efficiency              literature for that component.
   results.
     X2.4.1.1 Fractions–Cut Point # 150°C—Analyze by Test                                  X2.5 Calculation
   Method D3710. The final boiling point of the fraction >150°C                              X2.5.1 Calculate the distribution coefficient (D) for each
   should not exceed 260°C. Alternatively Test Method D5134                                (pseudo) component by means of the following equation:
   can be used. In the latter case it can only be applied to cuts                                                                   Y1i
   #100°C. If Test Method D5134 is used, the distribution                                                                Di 5                                     (X2.2)
                                                                                                                                ~Y1i 1 Y2i!
   coefficients are calculated for well defined and separated
   individual compounds. Any number of compounds can be                                    where:
   used, with a minimum of four. Aromatic compounds shall not                              Y1i = yield (% M/M on charge) of (pseudo) component i in
   be used for efficiency calculations.                                                           Fraction 1, and
                                                                                           Y2i = yield (% M/M on charge) of (pseudo) component i in
      NOTE X2.3—Due to azeotropic effects, aromatic compounds will not                            Fraction 2.
   fall on the same straight line as aliphatic compounds with the same boiling
   point.
                                                                                            X2.5.2 Calculate the coefficients of the following equation:
                                                                                                                           T 5 aX 1 b                             (X2.3)
      X2.4.1.2 Fractions—Cut Point >100°C—Analyze by Test
   Method D2887.                                                                             from:
      X2.4.1.3 Residues (FBP >538°C)—Presently there are no                                                             (Ti Xi 2 ~ ( Ti ( Xi/n!
   standardized methods for the characterization of the boiling                                                    a5                                             (X2.4)
                                                                                                                          ( Ti2 2 ~~ ( Ti!2/n!
   point distribution of residues on the books. However, methods
   for residue containing materials are under development (see                               and:
   Note X2.3). For the time being, in-house methods can be used                                                      b 5 ( Xi/n 2 a~ ( Ti/n!                      (X2.5)
Copyright by ASTM Int'l (all rights reserved);                                        27
                                                                         D2892 – 11a
    where:                                                                          Nominal cut point (AET)                    = 300°C
    n = number of (pseudo) components, and                                          Actual cut temperature                     = 221.4°C (from Annex A8)
    Ti = boiling point of (pseudo) component, °C.                                   Yield % M/M on charge; Fraction 1 (<200°C) = 10.3
     and:                                                                           Yield % M/M on charge; Fraction 2 (>200°C) = 9.8
                                            S
                                            Di
                               Xi 5 log10 1 2D
                                               i
                                                   D                                  X2.5.6.1 Linear regression of Xi versus Ti (X2.5.2):
                                                                                              T = − 14.77X + 300.1 thus: ECP = 300.1°C
     X2.5.3 By definition, the ECP is the temperature where Di =                      X2.5.6.2 Taking as a reference component a component
   0.5, hence Xi = 0 thus ECP = b.                                                 with a boiling point T = 280°C ( Tb = 280°C) and substituting
     X2.5.4 The efficiency is calculated from:                                     this in the derived equation with the corresponding X = 1.3605.
                                    X 3 ~Tt 1 273!                                    X2.5.6.3 For the actual efficiency Tt = 300°C. For the
                               N5                                    (X2.6)        minimum efficiency (corrected for pressure) Tt = 221.4°C.
                                    4.6 3 ~Tc 2 Tb!
                                                                                   Substituting these numbers in Eq X2.5 will yield an actual
    where:                                                                         efficiency Nactual = 8.4 and a minimum efficiency of Nminimum =
    N = efficiency,                                                                7.3.
    X =
                   S
                   Dx
                              D
           log10 1 2 D of reference component,
                      x
                                                                                      X2.5.7 Graphical Solution:
                                                                                      X2.5.7.1 A graphical solution to Eq X2.2 is shown in Fig.
                                                                                   X2.1. On Chart A percent to bottoms is plotted against the
    Tt = average tower temperature, °C,
    Tc = cut point temperature, °C, and                                            boiling point (BP) of the (pseudo) component. The best
    Tb = boiling point of reference component, °C.                                 possible straight line is drawn through the points and extended
      X2.5.4.1 As a reference compound any (pseudo) component                      down through Chart B.
   can be chosen in the boiling range overlap of the two adjacent                     X2.5.7.2 From Chart A, the boiling point corresponding
   fractions, except a component whose boiling point (Tb) is                       with the intersection of the line with 50 % to bottoms line is
   equivalent to the ECP. Subsequently, the corresponding X is                     read. This is the ECP.
   calculated from Eq X2.1.                                                           X2.5.7.3 From the intersection of the line on Chart B with
      X2.5.4.2 For cuts obtained under atmospheric conditions                      the curved line for temperature denoting the ECP, the efficiency
   and to obtain the actual tray number use the following                          Nactual can be read from the vertical axis. For atmospheric
   temperatures:                                                                   distillations the minimum efficiency, Nminimum, is equivalent to
                      Tc = ECP = AET, °C, and                                      the actual efficiency, Nactual. For distillations at subatmospheric
                 Tt = actual (nominal) cut point, °C.                              pressure, the minimum efficiency can be calculated from the
      X2.5.4.3 For cuts obtained under reduced pressure and for                    following equation:
   obtaining the column efficiency (Nminimum) for comparative                                                                 Tt 1 273
                                                                                                         Nminimum 5 Nactual 5 T 1 273                 (X2.7)
   purposes, the actual column temperature shall be used. In this                                                              n
   case the tower temperature Tt shall be considered equivalent to
                                                                                    where:
   the actual vapor temperature at the nominal cutpoint (AET)
                                                                                    Tt = actual cut point, °C, and
   under the pressure condition prevailing at the nominal cut                       Tn = nominal cut point, °C.
   point.
                                                                                      X2.5.7.4 Examples—The same basic data as for the ex-
      X2.5.5 Example of efficiency calculation for cuts obtained
                                                                                   amples in X2.5.5 and X2.5.6 are used. Percent to bottoms is
   under atmospheric conditions.
                                                                                   just another way of expressing the distribution coefficient and
   Excerpt of Test Method D2892 Analysis:
    Operating pressure                                 = 101.3 kPa
                                                                                   is related to the distribution coefficient by:
                                                                                                 Percent to bottoms = (1 − Di) 3 100
    Nominal cut point (AET)                            = 200°C                          (1) For convenience sake the relevant values for the
    Actual cut point                                   = 200°C
                                                                                   graphical solution are given in Table X2.3.
    Yield % M/M on charge; Fraction 1 (<200°C)         = 8.5                          X2.5.8 Outliers—Because of inaccuracies in the GC data,
    Yield % M/M on charge; Fraction 2 (>200°C)         = 9.0
                                                                                   especially at the beginning and end of the (GC) distillation
      X2.5.5.1 Linear regression of Xi versus Ti (X2.5.2):                         curve, it is not unusual to observe outlying points for distribu-
             T = −17.32X + 199.9 thus ECP = 199.9°C                                tion coefficients <0.05 and >0.95, therefore, it is recommended
      X2.5.5.2 Taking as a reference component a component                         to inspect the linear plot of T versus X for outliers. The same
   with a boiling point T = 180°C (Tb = 180°C) and substituting                    holds for the graphical solution, but there outliers will be
   this in the derived equation the corresponding X = 1.1503.                      readily observed. If outliers occur they should be removed,
   Furthermore, Tt = 200°C. Substituting these numbers in Eq                       because of their very pronounced effect on the slope of the line
   X2.5 will yield an efficiency of N = 5.9; or Nactual= Nminimum=                 and hence the value for N. Another cause for significant scatter
   5.9.                                                                            in data points can be incorrect sample size (too small) with
      X2.5.6 Example of efficiency calculation for cuts obtained                   respect to column diameter or dynamic hold-up, or both. Check
   under subatmospheric conditions.                                                Table 1 or Annex A2, or both, for correct sample size.
   Excerpt of Test Method D2892 analysis:                                          Inaccuracies in the GC data can be especially pronounced in
    Operating pressure                           = 13.3 kPa                        cuts with very narrow boiling ranges. This is especially evident
                                                                                   in lighter boiling cuts (<100 °C). Appropriate attention to skew
Copyright by ASTM Int'l (all rights reserved);                                28
                                                                          D2892 – 11a
                                                              FIG. X2.1 Efficiency Calculation
                                            TABLE X2.1 Relevant Values for Calculation Method of Efficiency
     NOTE—Excerpts from GC analysis (X2.4.2), calculation of distribution coefficient (X2.5.1) and Xi (X2.5.2) where Yield o.f. is % M/M of the
   component on the fraction (GC result from analysis of the fraction).
                                                 Fraction 1                      Fraction 2
                                                   Yield          Yield            Yield          Yield
        Component               BP                  o.f.           o.c.             o.f.          o.c.
                                Ti°C                               Y1i                             Y2i            Di                 Xi
         160-170                 165               19.75          1.68              0.21          0.02           0.989             1.9485
         170–180                 175               19.40          1.65              0.62          0.06           0.967             1.4706
         180–190                 185               17.95          1.53              2.03          0.18           0.893             0.9217
         190–200                 195               13.04          1.11              6.39          0.58           0.658             0.2850
         200–210                 205                6.96          0.59             12.52          1.13           0.344            −0.2798
         210–220                 215                2.05          0.17             16.67          1.50           0.104            −0.9350
         220–230                 225                0.77          0.07             18.45          1.66           0.038            −1.4043
         230–240                 235                0.18          0.02             18.72          1.68           0.009            −2.0419
                                            TABLE X2.2 Relevant Values for Calculation Method of Efficiency
     NOTE—Excerpts from GC analysis (X2.4.2), calculation of distribution coefficient (X2.5.1), and Xi (X2.5.2) where Yield o.f. is % M/M of the
   component on the fraction (GC result from analysis of the fraction).
                                                 Fraction 1                      Fraction 2
                                                   Yield          Yield            Yield          Yield
        Component               BP                  o.f.           o.c.             o.f.          o.c.
                                Ti°C                               Y1i                             Y2i            Di                 Xi
         260–270                 265               16.43          1.69              0.08          0.01           0.995             2.3342
         270–280                 275               16.26          1.67              0.27          0.03           0.984             1.8014
         280–290                 285               15.00          1.55              1.58          0.15           0.909             0.9991
         290–300                 295               11.07          1.14              5.92          0.58           0.663             0.2934
         300–310                 305                5.23          0.54             11.85          1.16           0.317            −0.3336
         310–320                 315                1.50          0.15             15.77          1.55           0.091            −1.0001
         320–330                 325                0.30          0.03             16.98          1.66           0.018            −1.7312
         330–340                 335                0.08          0.01             17.27          1.69           0.005            −2.3126
   (column overloading) and linearity of signal (detector over-                  methods mentioned in X2.1.4. Ensure that correct units are
   loading) is advisable if unusual GC results are detected. If                  compared (that is, obtain Test Method D3710 results in mass
   using Test Method D3710, note that results are typically                      percent).
   calculated in liquid volume %, not mass % as in the other                       X2.5.9 Final Calculation:
Copyright by ASTM Int'l (all rights reserved);                              29
                                                                                      D2892 – 11a
            TABLE X2.3 Relevant Values for Graphical Solution                                     X2.6.2.1 Only three pressure levels are permitted in Test
     NOTE—From Chart A for Example 1, an ECP of 200°C is read and for                          Method D2892. Distillation at 0.266 kPa provides an alterna-
   Example 2 an ECP of 300°C. From Chart B for Example 1, the efficiency                       tive to distillation under 1.33 kPa, but both pressure levels shall
   Nactual = 5.9 and for Example 2 Nactual= 8.4. For Example 1, atmospheric                    not be applied consecutively in the same distillation run. The
   distillation, Nactual = Nminimum = 5.9. For Example 2, subatmospheric                       reflux ratio applied shall be 5:1 in all cases, including distilla-
   distillation, Nminimum is derived from Eq X2.7, and amounts to 7.2.                         tion at 0.266 kPa.
                  Example 1                   Example 2                                           X2.6.3 Performance Verification—Determine the ECP and
      BP            Di          % to bottom      BP               Di        % to bottom
                                                                                               the efficiency Nminimum at three appropriate pressure levels, but
      165          0.989            1.0          265           0.995            0.5            only at the highest cut point usually attained at that pressure
      175          0.967            3.0          275           0.984            1.6
      185          0.893           10.7          285           0.909            9.1            level.
      195          0.658           34.2          295           0.663           33.7               X2.6.4 Data Interpretation–Effıciency:
      205          0.344           65.6          305           0.317           68.3
      215          0.104           89.6          315           0.091           90.9
                                                                                                  X2.6.4.1 Regression of the efficiency (Nminimum) obtained
      225          0.038           96.2          325           0.018           98.2            against the AET should indicate a best fit for linear regression
      235          0.009           99.1          335           0.005           99.5            and produce a straight line more or less parallel to the lines
                                                                                               given in Fig. X2.1. The regression line shall not exceed the
                                                                                               upper or lower limits shown in Fig. X2.1. Individual points
     X2.5.9.1 Calculate the efficiency Nminimum and ECP (°C) at                                should be randomly spread along the regression line and be
   each point. Calculate the difference between ECP and AET for                                within 0.7 theoretical plates from the regression line. If one
   each cut point.                                                                             point is outside these limits, rerun the appropriate GCs, redo
     X2.5.9.2 Table X2.4 defines the standard efficiency perfor-                               the efficiency calculation, and check again. If this check fails or
   mance of a Test Method D2892 TBP (15/5) distillation,                                       more then one point fails this criterion, the performance of the
   including the standard spread for individual data points. Fig.                              column shall be considered suspect. Consult X2.6.4.2-X2.6.4.4
   X2.2 is a graphical presentation of the standard efficiency and                             for possible causes. Take corrective action and repeat the whole
   the acceptable upper and lower limits of Test Method D2892                                  procedure, including the distillation (see Note X2.6).
   column efficiency performance (14 to 18 theoretical plates). A                                 X2.6.4.2 If the regression line is curved, yielding relatively
   discussion on the acceptability of deviations, probable causes                              high efficiency (Nminimum) values at the lowest pressure level, it
   and corrective actions is given in the next chapter.                                        is an indication for unacceptable heat loss from the column.
                                                                                               Check Annex A3. The reverse is true for an overhead column
   X2.6 Frequency of Test and Data Interpretation                                              (see 6.1.3.4).
      X2.6.1 It is recommended to carry out a full performance                                    X2.6.4.3 If the regression line is not parallel and yielding a
   check at the first time of commissioning or any other point of                              steeper slope, it is an indication of unacceptable heat loss in the
   time when there are significant changes to the equipment. After                             column (check Annex A3). The reverse is true for an over-
   that, a shortened evaluation may be satisfactory, if carried out                            heated column (see 6.1.3.4). Nonparallel regression lines can
   at regular intervals, for example, once or twice a year, to verify                          also be an indication of the application of a non-approved and
   that the equipment is still performing satisfactory.                                        non suitable packing (see 6.1.3 and Table 1).
      X2.6.2 Full Performance Check—Determine the ECP and                                         X2.6.4.4 If the (parallel) regression line is located outside
   the efficiency Nminimum for at least two, but if possibly three,                            the upper or lower Nminimum limits, or both, the following
   cut points at each of the three most frequently used pressure                               causes might be applicable.
   levels. One of the cut points should be the maximum cut point                                    (1) Efficiency of the packing too high or too low. Check
   (AET) usually obtained at the appropriate pressure level and                                Table 1 or Annex A1, or both. If appropriate, correct the
   one of the cut points should be obtained at the lower end of the                            efficiency by adding or subtracting packing material from the
   temperature range usually covered at the appropriate pressure                               column.
   level. Table X2.5 gives an example of such a scheme.                                             (2) Reflux ratio too high or too low. Check Annex A7.
                                                                                                    (3) Incorrect distillation rate. Consult 10.3.2 and 10.4.5.
                                                                                                  X2.6.4.5 For a (shortened) periodic verification of effi-
     TABLE X2.4 Standard Efficiency for 15/5, Test Method D2892
                             Columns                                                           ciency, the three values for efficiency (Nminimum) obtained shall
                                                                                               be located inside the band for the regression line obtained from
     NOTE—The data from Table X2.4 were estimated from the efficiency                          the full performance check and more or less parallel to that
   results obtained as a spin off of a round robin carried out in 1978 on a
                                                                                               regression line. If one or both criteria are not met, it is an
   crude oil with an API-gravity of 35. However, the lack of detailed
   documentation of these results prevents rigid statistical treatment.                        indication that the performance of the column has changed. For
                                                                                               possible causes refer to X2.6.4.1-X2.6.4.4.
                                                       Standard Efficiency,
                Cut Point, °C
                                                            Nminimum                             NOTE X2.6—It is recognized that the setting of acceptable tolerance
                     50                                     4.1   6   0.5                      limits to trigger corrective action is the responsibility of the laboratory.
                    100                                     4.7   6   0.6                      However, it is recommended that they do not exceed the limits as
                    150                                     5.3   6   0.6                      indicated.
                    200                                     5.9   6   0.7
                    250                                     6.6   6   0.8                        X2.6.5 Data Interpretation–Cut Point:
                    300                                     7.2   6   0.9
                    350                                     7.8   6   0.9                        X2.6.5.1 When using an average crude oil (30 to 40 API)
                                                                                               distilled under TBP (15/5) conditions, it is expected that the
Copyright by ASTM Int'l (all rights reserved);                                            30
                                                                              D2892 – 11a
                                                             FIG. X2.2 Efficiency Nomograph
           TABLE X2.5 Performance Check; Cut Point Scheme                                 (2) If the difference between ECP and AET is significantly
    Pressure,                Range for Recommended Cut Point, °C                     larger at one or more reduced pressure level(s), it is an
      kPa                                                                            indication for incorrect pressure measurement. Check Annex
                       Low                  Middle                 High
   101.3             60 − 70              130 − 140           190   −   200          A6.
   13.3             230 −240              260 − 270           290   −   300               (3) Incorrect efficiency (too low). Refer to X2.6.4.
   1.33             330 − 340                 −               360   −   370               (4) Incorrect sample size (too small) with respect to column
   0.266            330 − 340             360 − 370           390   −   400
                                                                                     diameter or dynamic hold-up, or both. Check Table 1 or Annex
                                                                                     A2, or both.
                                                                                          (5) Excessive heat loss. Evidence for the latter can also be
                                                                                     obtained by careful examination of the T versus X plot
   difference between ECP and AET will not exceed 0.7 times the
                                                                                     (X2.5.8). In case of excessive heat loss the distribution will be
   reproducibility of the test method. The reproducibility of the
                                                                                     asymmetrical, evidenced by significantly more outlying points
   test method in degrees Celsius is tentatively estimated for
                                                                                     at the high end of the temperature scale than at the low end. An
   average crude oils to be approximately 8°C at any pressure
                                                                                     overheated mantle (6.1.3.4) will lead to a too low ECP in
   level. Therefore, deviations in excess of 6°C shall be consid-
                                                                                     comparison to the targeted AET.
   ered suspect (see Note X2.7).
      X2.6.5.2 The following causes for excess deviations of ECP                       NOTE X2.7—The setting of acceptable tolerance limits to trigger
                                                                                     corrective action is the responsibility of the laboratory. However, the limit
   from AET can be identified.
                                                                                     given has been proven indicative of incorrect performance in many years
       (1) If the ECP is consistently and significantly higher or                    of application of Test Method D2892.
   lower than the AET, it is an indication of incorrect temperature                    NOTE X2.8—Too high efficiency will have no measurable effect on the
   measurement, Check Annex A4-Annex A6.                                             ECP-AET differential.
                                                             SUMMARY OF CHANGES
                  Subcommittee D02.08 has identified the location of selected changes to this standard since the last issue
                (D2892–11) that may impact the use of this standard. (Approved Dec. 1, 2011.)
   (1) Revised 3.1.7 to be consistent with one definition and
   Form and Style.
                  Subcommittee D02.08 has identified the location of selected changes to this standard since the last issue
                (D2892–10) that may impact the use of this standard. (Approved July 1, 2011.)
   (1) Updated terms in Section 3.                                                   (2) Made some of the original discussions into notes.
Copyright by ASTM Int'l (all rights reserved);                                  31
                                                                                   D2892 – 11a
                Subcommittee D02.08 has identified the location of selected changes to this standard since the last issue
              (D2892–05) that may impact the use of this standard. (Approved Oct. 1, 2010.)
   (1) Revised Section 13 to reflect the actual precision state-                            the research report containing the data and statistical evaluation
   ment from the D2892 precision study, including reference to                              and validation.
                        ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
                    in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
                    of infringement of such rights, are entirely their own responsibility.
                        This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
                    if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
                    and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
                    responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
                    make your views known to the ASTM Committee on Standards, at the address shown below.
                      This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
                    United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
                    address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
                    (www.astm.org). Permission rights to photocopy the standard may also be secured from the ASTM website (www.astm.org/
                    COPYRIGHT/).
Copyright by ASTM Int'l (all rights reserved);                                         32