Simplygo Mini: S G M S & T I
Simplygo Mini: S G M S & T I
Page 1
1112893, Ver. 04
Page 2
1112893, Ver. 04
Page 3
1112893, Ver. 04
Page 4
1112893, Ver. 04
Page 5
1112893, Ver. 04
Version
Record of Revision Author
Number
Page 6
1112893, Ver. 04
AC Power Supply
DC Power Supply
Rechargeable Lithium Ion Battery
1.3 Contraindications
This device is not intended to be life-sustaining or life-supporting.
This device is not intended for newborn and infant use.
Page 7
1112893, Ver. 04
International
Phone: 1-724-387-4000
Fax: 1-724-387-5012
Page 8
1112893, Ver. 04
WARNING
Warnings indicate the possibility of harm to the operator or patient.
CAUTION
Cautions indicate the possibility of damage to the device.
NOTE
Notes are used to emphasize a characteristic or important consideration.
2.2 Warnings
The operator should read and understand this entire manual before servicing or repairing the
device.
The device is not intended for life support. Where the prescribing health care professional has
determined that an interruption in the supply of oxygen, for any reason, may have serious conse-
quences to the user, an alternate source of oxygen should be available for immediate use.
Geriatric or any other patient unable to communicate discomfort, or hear or see the alarms while
using this device, may require additional monitoring.
Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of
an open flame.
Do not use the device in the presence of a flammable anesthetic mixture in combination with ox-
ygen or air, or in the presence of nitrous oxide.
Do not use oil or grease on the concentrator or its components as these substances, when com-
bined with oxygen, can greatly increase the potential for a fire hazard and personal injury.
If you notice any of the following, discontinue use and contact your home care provider:
unexplained changes in the performance of this device
unusual or harsh sounds
dropped or mishandled device or the power supply
Page 9
1112893, Ver. 04
Page 10
1112893, Ver. 04
Page 11
1112893, Ver. 04
Philips Software
Description Version
SimplyGo Mini Service Tool 2.2.0.2
SimplyGo Mini Firmware 1.1.1.5
SimplyGo Mini O2 Balancing Tool 1.0.1.1
SimplyGo/SimplyGo Mini Service 01
Compressor Test
NOTE
This section is for service computers that are being set up for a first time installation. If your
computer has a previous version of the service tool installed, refer to Section 3.6.2 to upgrade
to the newest version of the service tool.
Page 12
1112893, Ver. 04
3. On the Service page, select the SimplyGo Mini option from the Service Software
Category list.
4. Select the Download button on the SimplyGo Mini Service Tool page.
5. Depending on your configuration, the buttons Run, Save, Cancel appear. Select Save or
Save As from the pull down list under Save. Download the SimplyGo Mini Service Tool
and follow the on screen instructions to the default locations on the computer.
Page 13
1112893, Ver. 04
7. Allow the software to download. If the system presents any Security Warning screens,
select the Run option as needed.
8. Depending on your configuration, you can either choose to launch the installer now, or wait
until later. In either case, when you launch the installer, the system self extracts and opens
the Self Extractor box.
9. When the SimplyGo Mini Service Tool Self Extractor box appears, select Install.
NOTE
It is strongly recommended to use the default
destination folder for extraction.
Page 14
1112893, Ver. 04
13. The License Agreement windows appears. Select the I accept the above 2 License
Agreements option and then select Next when it becomes active.
14. The Start Installation summary box appears. Review the details and select Next
15. The Command Prompt window may appear (XP users) during the installation. Do not close
this window, it will close automatically.
16. The Installation Complete box appears. Select Next.
17. The FTDI CDM Drivers box appears. This installs the drivers needed to configure the PC’s
com ports to enable interface with the device. Select Extract.
NOTE
If you have a previous version of the service tool installed, the FTDI driver installer
should not appear. If it does, let the installer step through the process. If prompted
to replace existing files, as in Step 10, select Yes.
18. The Device Driver Installation Wizard box appears. Select Next.
Page 15
1112893, Ver. 04
You can Save or Print the agreement for your records from this window.
20. The Completing the Device Driver Installation Wizard box appears. Select Finish
21. The Restart selection box appears. Before ANY restart, go to Step 22.
22. Attach the O2 RP Interface Kit to a USB port on the computer ONLY and then select
Restart.
IMPORTANT NOTE
Do not attach any unit to the interface at this time.
Page 16
1112893, Ver. 04
NOTE
You must have Administrator rights to your
computer to configure the com port. If you cannot
access the Control Panel or Properties for the com
port, contact your system administrator.
3. Right-click on My Computer icon, and select Manage. For Windows 7 users, click Start and
select Control Panel.
Page 17
1112893, Ver. 04
Page 18
1112893, Ver. 04
NOTE
Ensure that the Com Port settings are the same as
pictured above.
7. In the BM Options section of the Advanced Settings window, set the Latency Timer (msec)
field to ‘2’ and select OK.
8. Select OK on the Port Properties window and close the Device Manager. The FTDI box is
now ready to work with the SimplyGo Mini device.
Page 19
1112893, Ver. 04
NOTE
As a matter of habit, confirm the interface switch is in
the No-Programming position before attaching the
MiniDIN7 cable between the fixture and the SimplyGo
Mini device.
Page 20
1112893, Ver. 04
The release is found at the same location on my.respironics.com as the Service Tool. Download the ZIP file to
your desktop.
If you have already installed the Service Tool and configured the computer USB port to use the
O2 interface, there is no need to reconfigure.
If you have not installed the Service Tool, then refer to Section 3.2.2.1 and follow the instructions
for configuring the O2 Interface/computer com port.
Page 21
1112893, Ver. 04
Removal
Insert a small flat head screwdriver to the tab and gently pry the data port cover away from the
rear cover.
Installation
Align the data port cover and gently snap the tabs into place.
Page 22
1112893, Ver. 04
4. Turn on the Simplygo Mini. Double click the SimplyGo Mini Service Tool icon on the
desktop and switch the Interface kit to the Program Mode.
DESKTOP ICON
Page 23
1112893, Ver. 04
6. In the File menu, select Find Device and Connect. The tool searches for connected
devices that are running.
Page 24
1112893, Ver. 04
The software tool begins searching for connected, running devices. When a device is found the tool displays
the unit serial number, the software version currently installed on the unit and PC com port to which the
device is attached.
Page 25
1112893, Ver. 04
8. When the download of files and error codes to the destination file on the computer is
complete, you are prompted to Save the data.
9. Select OK to save the data. The Service Tool Home window appears.
Page 26
1112893, Ver. 04
NOTE
You can load the software update into the tool without
being connected to any device. Launch the tool and
Cancel the No Device Found warning.
The Service Tool Home Window has three menus File, Commands and About, and five action buttons. The
commands under the menu options are detailed later in this chapter. To load the Software/Firmware updates
into the service tool, do the following:
1. From the Service Tool Home window, select the Import Software Kit button. This option
lets you load SimplyGo Mini software/firmware that you downloaded to your desktop in
Section 5.3.1, into the service tool.
Page 27
1112893, Ver. 04
Page 28
1112893, Ver. 04
Import Software Kit lets you import updated software revisions into the tool.
Program Device lets you program selected devices with the latest version of the software/firm-
ware. This button is active only when the tool has imported a software kit for distribution.
View Live Data lets you view the live data from selected devices.
Delivery Valve Calibration lets you calibrate valve operation
Exit closes and exits the tool.
Page 29
1112893, Ver. 04
Connect Device option lets you select any connected device via the com port number.
Find Device and Connect lets you find and connect devices attached to the test PC.
Disconnect Device option lets you disconnect selected devices that are attached to the test PC.
Download Dataset option captures the data set from a selected device.
The Uninstall Software Kit option lets you remove a previous version of the software from the
service tool.
Page 30
1112893, Ver. 04
Reset User Hourmeter option lets you reset the device time-in-service meter.
Reset NVR for future release.
Reset O2 Factors for future release.
Set Valve Deltas for future release.
Set Motor Speed option lets you adjust the motor speed.
Set Demo Breath Rate option lets you adjust the Breath Rate in Demo mode.
Set Model & Serial Number option lets you enter the Model and Serial numbers from the de-
vice’s outer case. This option is used when the PCA is replaced.
Set Device Setting option lets you control the selected device’s setting via the service tool.
Set O2 Sensor State option lets you control the O2 sensor via that service tool.
Set Power On option.
Send Command (Hex) for future release.
Page 31
1112893, Ver. 04
To update the system software/firmware in the SimplyGo Mini device, after the software/firmware version has
been loaded into the tool, do the following:
1. From the Service Tool Home window, select the Program Device button. The Select
Software Kit to Program selection box opens. This button is available only when
software/firmware is loaded into the service tool.
2. From the pulldown menu, select the Release Number to be installed into the connected
device. Select Continue. The Software Tool updates the device software/firmware. The
programing is automatic and takes about 13 minutes to complete.
The following is an example of the status report of device passing/failing after the firmware is updated.
Page 32
1112893, Ver. 04
2. In the dialog box, set the Motor Speed to 1174. Ensure that the device is at Setting 2 in the
device interface display.
3. Select Continue.
4. The Command Successful message is displayed. Select OK.
Page 33
1112893, Ver. 04
3. Select Continue.
4. The Command Successful message is displayed. Select OK.
3. Select Continue.
4. The Command Successful message is displayed. Select OK.
Page 34
1112893, Ver. 04
3. Select Continue.
4. The Command Successful message is displayed. Select OK.
Page 35
1112893, Ver. 04
3. Select Continue.
4. The Command Successful message is displayed. Select OK.
Page 36
1112893, Ver. 04
View Menu
Device Parameters
Add/Subtract Parameters window.
List
2. From the View menu of the window, select the Number of Charts command. A selection
box appears.
3. Set the Number of Charts for example to 1. Typically, the chart area divides into the
number of charts you have selected. In this example, only one chart is visible.
Page 37
1112893, Ver. 04
5. Select Y-Scale Properties. The tool opens Edit Y-Scale Properties box. This box lets you
change the grid size on the Y-Axis and the line color.
6. In this example, use the default settings. Click on the Grid Color box to open the color
palette selection box.
Page 38
1112893, Ver. 04
10. Select OK and the chart reflects the changes you made.
11. Select Edit Background Color. The tool opens Edit Background Color box. This box lets you
change the background color of the chart.
12. Select the color and click OK and the chart reflects the changes you made.
Page 39
1112893, Ver. 04
NOTE
Previous versions of the tool do not save settings. When you exit the
tool all settings are lost and have to be reprogrammed.
From version 2.6.0.1 forward you can save the settings and apply them
to any new testing.
Page 40
1112893, Ver. 04
3. The device begins tracking the selected parameters in real time. You can change the color
of the tracking signal by clicking on the color block on the right of the selected parameter.
4. Remove parameters from the active list by highlighting the parameter and clicking the erase
icon.
5. Remove all parameters by clicking the erase all icon.
Page 41
1112893, Ver. 04
NOTE
Devices may display multiple failure modes.
Units with an improper power cord or power cord connection
can be diagnosed visually.
Page 42
1112893, Ver. 04
NOTE
To establish a baseline of the device’s mechanical performance, perform and record an initial
waveform review before proceeding to any other inspection or repair.
Refer to Chapter 3 for instructions on connecting the device to the test PC, and using the Service Tool
to generate and view waveform data.
Page 43
1112893, Ver. 04
Page 44
1112893, Ver. 04
Page 45
1112893, Ver. 04
NOTE
The preceding figures are only a sampling of what may occur.
Page 46
1112893, Ver. 04
Page 47
1112893, Ver. 04
Page 48
1112893, Ver. 04
Page 49
1112893, Ver. 04
Page 50
1112893, Ver. 04
Not charging
Complaint
Any complaint related to a dead battery
Not charging
Complaint
No flow
Low O2
Complaint
No flow
Clean/replace check
valves
Refer to Section 4.1.3 for examples of check
Debris, Leak, other valve malfunction and other wave form or *if issue persists with
failure in Core Running unit wave pressure anomalies new check valves
Assembly form (and after sieve tubes
Refer to Chapter 5 for instructions on replaced) replace
generating data waveforms. core
Page 51
1112893, Ver. 04
n/a - updated
High Breath Rate at No flow events at < 10 min run time
software (standard on
Startup Event log ACCOMPANIED by breath rates @/above
all repairs) will correct
20BrPM
Cracked screen
Complaint Damaged screen
Cracked/Distorted Broken screen Front cabinet
Screen assembly with UI
External physical Cracking or visible damage (unit on or off)
examination Touch screen not responsive
Noise
Complaint Low O2
No flow
Low O2
Complaint
No flow
Page 52
1112893, Ver. 04
Complaint No flow
Complaint Noise
Core
Noise generated by OR
fluttering diaphragm Unit emits a chirp, burp, bullfrog, squeak, or Airside assembly
Running unit noise other unusual noise accompanying the
EXHAUST (not delivery pulse)
Complaint Miscellaneous
Page 53
1112893, Ver. 04
Complaint No flow
Noise
Low O2
Complaint No flow
Noise
Low O2
Complaint No flow
Replace rubber
·Unit emits a grinding, rattling, or
Compressor Rubber Unit noise mounts and add
otherwise abnormal sounding noise
Mounts LOOSE Loctite
Low O2
Battery won’t charge
Complaint
Technical Fault
SMB BUS failure Battery PCA Wire
Harness
Physical
Loose battery PCA wire examination of Communication error between main
harness unit/accessories PCA and battery PCA
Page 54
1112893, Ver. 04
IMPORTANT NOTE
The sieve canisters and compressor inlet filters are to be replaced during ANY
service activity.
5.1 Overview
This section describes scheduled and routine maintenance procedures. Normal routine maintenance involves
periodic checking, cleaning, and or replacing the following items as necessary:
Case cleaning
Data log consistency
Pressure Waveforms
Check Valve Performance
Required replacement of Sieve Canisters during any service activity ONLY if the sieve canisters
are greater than 90 days old.
o 90-day definition is either from the date of manufacture if the device has not yet been
serviced or, from the last service date.
o If the Sieve Canisters are found to be faulty prior to the 90-day definition, replace the
canisters as per the Repair and Replacement Section of this manual.
Replacement of the Rear Cover with upgraded Rear Cover (Green Dot) during any service activ-
ity
Replacement of the Compressor Inlet Filters during any service activity ONLY if the Compressor
Inlet Filters are greater than 90 days old.
o 90-day definition is either from the date of manufacture if the device has not yet been
serviced or, from the last service date.
o If the Compressor Inlet Filters are found to be faulty prior to the 90-day definition, re-
place the filters as per the Repair and Replacement Section of this manual.
Inspect and replace loose Compressor Isolation Mounts
Upgrade device firmware if necessary (Section 3.6.2)
IMPORTANT NOTE
Page 55
1112893, Ver. 04
If the display screen is damaged or the hours are not accessible from the display screen, refer to Chapter 3
to retrieve hours using the service software tool or data log.
5.2 Cabinet
Commensurate with hospital or homecare policies, Respironics recommends cleaning the case and
inspecting for damage as necessary.
WARNING
To avoid electrical shock, do not remove the SimplyGo Mini covers. The covers should only be
removed by authorized service personnel. Do not apply liquid directly to the covers. Do not use
alcohol, solvents, polishes or any oily substances on the device as they are
flammable.
CAUTION
Do not allow liquids into any of the controls, the interior of the case, or the
oxygen tubing connector. If this occurs, contact your home care provider for
assistance.
Page 56
1112893, Ver. 04
LEFT ORIGINAL REAR COVER WITHOUT GUIDE FEATURE RIGHT NEW REAR COVER WITH GUIDE
FEATURE
Page 57
1112893, Ver. 04
NOTE
Regardless of the condition of the filters, all filters must be replaced as a
matter of routine maintenance after they are 90 days old. Be sure to replace
the filters with the upgraded filter
Debris contamination
DEBRIS CONTAMINATION
Page 58
1112893, Ver. 04
NOTE
The sieve canisters are to be replaced during ANY service activity.
NOTE
It is not necessary to remove the product tank tubes from the O2
butterfly plate
2. Stand the canisters on the air manifold base assembly with the O2 manifold plate up.
Air Manifold
Base Assembly
Page 59
1112893, Ver. 04
4. Remove and flip the O2 manifold plate over and lay it flat on the table.
O2 Manifold Plate
O-rings
NOTE
Remove and discard used O rings and filters. Replace with new rings
and filters each time.
Page 60
1112893, Ver. 04
Pan Head
Screws
Balance Valve
Butterfly
Plate Screws Butterfly Plate
6. Remove 2-T8 pan head screws securing the butterfly plate to the O2 manifold.
Page 61
1112893, Ver. 04
Butterfly Plate
O2 Manifold
NOTE
Remove and discard used O rings. Replace with new rings each time.
8. Remove and discard the old EPDM (white) check valve discs (2).
Check Valve
Check Valve Disc
Disc
2 CHECK VALVE
CAUTION
Gently tapping the O2 manifold on a hard surface should cause the
Discs to drop out.
If using a tool to dislodge the Discs, take care to not scratch or damage
the sealing surface at the bottom of the check valve pocket.
Page 62
1112893, Ver. 04
CAUTION
The unit cannot be reassembled with the O-rings in place. They MUST
be removed for reassembly.
The connections will NOT seal if you attempt to reassemble the unit
without removing the O-rings.
10. Use a lint free cotton swab LIGHTLY coated with alcohol to clean inside the check valve
pocket bottom and sides in the O2 manifold
11. Swab the flat surface of the check valve stop on the butterfly plate
Orifice Pocket
12. Let the alcohol fully evaporate before proceeding to step 14.
Page 63
1112893, Ver. 04
New Check
New Check Valve Disc
Valve Disc
CAUTION
The new discs must be free from dirt, debris and oil. Do not excessively
handle the discs with exposed fingertips. Use tweezers or a suitable
tool to handle the discs. If wearing latex gloves, be sure the gloves are
not powder coated.
14. Place 2 new O-rings (Red) on check valve posts and 1 new O-ring (Black) on orifice post
after LIGHTLY coating with Christo-Lube/Krytox and isoproply alcohol.
NOTE
Black Viton O-rings are to be coated in Christo-Lube/Krytox.
Red silicone O-rings are to be coated in isopropyl alcohol.
O-RING PLACEMENT
Page 64
1112893, Ver. 04
15. Position butterfly plate on corresponding O-ring glands of the O2 manifold and press straight
down with one finger on the round boss on top of the butterfly plate, holding it down while
installing 2 T8 pan head screws that secure the butterfly plate.
Butterfly Plate
O2 Manifold
Page 65
1112893, Ver. 04
Page 66
1112893, Ver. 04
18. Place 2 new O-rings (Red) on canister posts of the O2 plate assembly after LIGHTLY
coating with Christo-Lube/Krytox.
WARNING
Introducing petroleum based products into an oxygen rich environment can
be hazardous.
Applying a LIGHT coating of Christo-Lube/Krytox and then wiping the ring
completely with a dry cloth will leave a sufficient film of lubricant on the
ring.
Page 67
1112893, Ver. 04
NOTE
Make sure all 4 screws are started and bottomed before final tightening.
Page 68
1112893, Ver. 04
5.8 Compressor
In addition to inspecting and replacing the compressor inlet filters, inspect the device for loose isolation
mounts. Refer to section 6.3.8.3 to replace the compressor inlet filters. Additionally:
If it is suspected that the compressor has failed, refer to Section 7.6.1 for optional troubleshoot-
ing procedure to test the compressor for functionality.
If it has been confirmed that the compressor has failed, refer to Chapter 6 to remove and re-
place the compressor.
Page 69
1112893, Ver. 04
CAUTION
It is required to be ESD grounded during repair and replacement
procedures.
Loctite
262
Page 70
1112893, Ver. 04
CAUTION
Power drivers are not recommended for servicing or repairing the
device. Several of the Torx fasteners are easily stripped and
should only be removed or replaced manually.
Page 71
1112893, Ver. 04
CAUTION
Only remove the battery when the SimplyGo Mini is turned off. Never remove the battery while the
device is operating.
Release Button
Page 72
1112893, Ver. 04
NOTE
The SimplyGo Mini back cover and power connector have been
upgraded. These upgrades are to be made whenever the device is
in for service. Refer to the following sections to identify and
implement these upgrades.
Removal:
1. Remove the battery from the device if installed.
2. Remove the (8) T10 Torx screws securing the rear cover.
Page 73
1112893, Ver. 04
CAUTION
When removing the connector, avoid contact with chip U1.
Chip U1 is easily damaged. If the chip is damaged both the PCA and
O2 sensor boards will have to be replaced. When removing the
connector, be careful to not damage the chip.
Page 74
1112893, Ver. 04
NOTE
For devices with PCA Revision 5 and higher, the power connector
must be replaced with the two wire connector PN 132383. Even if
the back cover is the updated version, the power connector must
be replaced.
If the back cover AND the power connector are to be replaced use
replacement rear cover which has the two (2) wire plug already
included.
Page 75
1112893, Ver. 04
AC CONNECTOR
6. Install the power connector in the new rear cover. Do not overtighten the screws.
CAUTION
When replacing the two short screws at the top of the device, HAND
TIGHTEN ONLY. The threads in the cover are easily stripped.
7. Reconnect the AC power connector to the PCA. Avoid contact with chip U1.
Page 76
1112893, Ver. 04
Page 77
1112893, Ver. 04
CAUTION
When replacing the two short screws at the top of the device, HAND TIGHTEN
ONLY. The threads in the cover are easily stripped.
Stripping the threads will necessitate replacing the whole front cover.
5. Install a new service label as needed. See Section 6.4 for label part numbers.
Page 78
1112893, Ver. 04
Page 79
1112893, Ver. 04
Removal
1. Remove the Rear Cover. Refer to Rear Cover and AC Power Connector Removal.
2. Remove the (2) T8 Torx screws securing the Battery PCA.
Connector J4
Page 80
1112893, Ver. 04
Page 81
1112893, Ver. 04
J8
J14
J3
J12
J2 J10
J19
J6 J4
FRONT SIDE MAIN PCA CONNECTIONS
Page 82
1112893, Ver. 04
Page 83
1112893, Ver. 04
IMPORTANT NOTE
The Main PCA and the O2 Sensor Assembly PCA are electronically synchronized at the manufacturer
and must be replaced as a unit. If either the Main PCA or the O2 PCA fails, both must be replaced.
The Replacement Kit contains both boards.
IMPORTANT NOTE
DVG (delivery valve gain) calibration must be performed after replacing this component. Refer to the
Testing chapter for the calibration procedure.
CAUTION
Do not attempt to remove the tubes from the tube barbs on either the main PCA or the O2
sensor PCA. The tube barbs are easily damaged. Service the device by flipping the main
PCA up and out of the way of the front cover.
Page 84
1112893, Ver. 04
NOTE
It is not necessary to remove the main PCA when servicing most components, except the compressor
assembly, tilt alarm, battery PCA and display ribbon. This is to avoid any damage that may occur
when removing and reinstalling components from the device.
Page 85
1112893, Ver. 04
1. Remove the main PCA board as described in 6.3.5. Do not remove the tubes connecting the
O2 assembly to the main PCA.
2. Remove the (2) T10 Torx screws securing the O2 PCA bracket in place.
Page 86
1112893, Ver. 04
3. Pull the O2 PCA assembly with patient filter and patient delivery valve connected, up and out
of the slots on the front cover.
It is not necessary to remove the tubing from the cannula connector or the product canister. If you are
replacing the O2 assembly PCA, remove those tubes at the patient filter and O2 sensor housing.
Page 87
1112893, Ver. 04
2. If removed from sensor housing, connect tube from product tank to sensor housing.
Page 88
1112893, Ver. 04
Page 89
1112893, Ver. 04
O2 Assembly
PCA Orientation
top of J7 Blade
Connector
Page 90
1112893, Ver. 04
STANDOFF PEGS
2. Slide the board backward under the air manifold connector and insert the tang on the board,
just below component L3, into the slots on the air manifold.
3. Ensure the PCA aligns with the two standoff pegs.
4. Align and press connector J7 firmly onto the male blade connector of the O2 assembly
board. The standoff pegs guide this alignment.
5. Re-attach all connectors to the proper connection points. Refer to Section 6.3.4.
Page 91
1112893, Ver. 04
IMPORTANT NOTE
DVG (delivery valve gain) Calibration must be performed after replacing this component. Refer to
Testing for the calibration procedure
Remove
1. Remove the main PCA referring to section 4.3.5.
2. Remove the O2 assembly referring to section 4.3.5.
3. Remove the (2) screws securing the patient delivery valve to the O2 sensor housing.
4. It is not necessary to remove the sensor housing from the board to remove the valve.
Page 92
1112893, Ver. 04
1. Install the patient delivery valve to the O2 sensor assembly and secure with either (2) T6
Torx screws torqued to 1.5 INLBS or (2) pan head Philips screws torqued to 3.5 INLBS
depending on the model.
2. Install the O2 sensor assembly. Refer to Section 6.2.5.
3. Install the main PCA. Refer to Section 6.3.5.
4. Connect the power connector to J2.
5. Install the rear cover.
Removal
1. Remove the rear cover.
2. Remove the main PCA referring to section 6.3.5 .
3. Remove the (2) T10 Torx screws securing the O2 PCA bracket in place.
4. Remove the tubing between patient filter and cannula barb one of two ways:
Without removing the O2 assembly, use a thin blade screwdriver to pry the tubing
from the patient filter barb on the bottom of the filter. Removing the tubing from the
filter unit and not the patient cannula will facilitate installation.
Remove the O2 assembly referring to section 6.3.5 and use extended nose needle
nose pliers to remove the tubing from the cannula barb.
PATIENT FILTER
Page 93
1112893, Ver. 04
Removal
IMPORTANT NOTE
DVG (delivery valve gain) Calibration must be performed after replacing this component. Refer to
Testing for the calibration procedure
It is not necessary to remove the main PCA to remove and service the compressor.
Page 94
1112893, Ver. 04
Center Wall
COMPRESSOR CONNECTIONS
Page 95
1112893, Ver. 04
6. Remove the (4) grommets from the compressor bracket. Inspect the grommets for cracks or
damage.
Page 96
1112893, Ver. 04
Filter Removal
1. Follow the procedures for removing the compressor from the unit.
2. Remove the compressor inlet filter from the end of the compressor. The filter can be
removed by pulling the filter away from the compressor. The recommended method for
removing a filter is to use a long tool that can fit between the compressor and the filter inlet
tubing (e.g., screwdriver, needle-nose pliers) and, using the compressor body as a fulcrum
and the tool as a lever, apply a force on the tubing away from the compressor. Repeat this
step for both filters.
Apply a force
pushing the filter
away from the
compressor
Page 97
1112893, Ver. 04
Note that the filter tube points to the front of the compressor with a slightly upward attitude.
Page 98
1112893, Ver. 04
Page 99
1112893, Ver. 04
Stripping the threads will necessitate replacing the entire front cover.
Page 100
1112893, Ver. 04
6. Place the compressor back in its mounting bracket by pulling the mount tabs through the
bracket holes.
Page 101
1112893, Ver. 04
8. Follow the rest of the procedure in section 6.3.8 to install the compressor back in the device
Page 102
1112893, Ver. 04
Removal
It is not necessary to remove the Main PCA to replace the speaker.
Speaker Connector J5
Speaker
SPEAKER POSITION
Installation
4. Install the speaker to its housing by pushing down into the slot. Be sure to orient the
speaker’s grey mesh to the inside of the device.
SPEAKER
5. Reconnect J5.
6. Reconnect the power connector to J2.
7. Install the rear cover.
Page 103
1112893, Ver. 04
Removal
1. Remove the rear cover.
2. Remove the main PCA and flip it over the top of the device.
3. Remove patient filter and O2 assembly bracket and move aside.
If needed remove the patient filter to sensor tube.
Installation
5. Install the fan to it's housing. Ensure the fan label is facing the compressor compartment of
the device and the connector is to the top left facing the rear cover.
FAN WITH O2 ASSEMBLY REMOVED NOTE FAN LABEL AND WIRING ORIENTATION
Page 104
1112893, Ver. 04
Removal
1. Remove the rear cover.
2. Remove the display ribbon cable from the main PCA by gently pushing up on each side of
the ribbon cable locks. The locking mechanism will only move a very small distance. Slide
the wire harness out ot the connector.
3. Remove the display wire harness from the display the same way.
Installation
1. Be sure the ribbon connector lock is loose and out of the connector.
2. Install the ribbon to the display by sliding the ribbon gently into the connector slot. Confirm
that the conductive metal protion end of the ribbon is facing the metal portion of the
conductors in the connector.
CAUTION
When installing the ribbon into the connector, the blue band of the ribbon end faces AWAY
from the circuit boards on both ends.
Ribbon alignment is critical. When inserting the ribbon into the connection the blue band
should be perfectly parallel to the connector to avoid damage to the device.
Page 105
1112893, Ver. 04
Page 106
1112893, Ver. 04
IMPORTANT NOTE
Before proceeding with the steps defined in this section, carefully review the procedures in
Section 5.5 Periodic Check Valve Cleaning for proper installation of O rings.
ESD SHIELD
The sieve canister assembly is more easily removed and intalled when the battery PCA connector is removed
from the battery PCA.
Page 107
1112893, Ver. 04
7. Remove the product tank from the bottom side of the SCA by removing the tubing from the
top of the assembly.
NOTE
An RP kit for the sieve canister assembly (top/bottom manifolds and sieve canisters), is not
available. The SimplyGo Mini sieve canisters must be replaced individually.
Page 108
1112893, Ver. 04
Page 109
1112893, Ver. 04
Sieve Canister
Assembly
replacement label.
Enter Date/Hours
NOTE
The purpose of this label is to inform the repair technician when / if the sieve canisters have
been replaced before. This is useful documentation to track repair history.
The sieve canister replacement label has no impact on the device or sieve tubes warranties.
Reference the User Manual for warranty information.
Page 110
1112893, Ver. 04
Air Manifold
Base Assembly
Page 111
1112893, Ver. 04
4. Remove the 2 Phillips head screws securing the balance valve and set the valve aside.
Phillips
Head
Balance Valve
Screws
Butterfly
Plate Screws Butterfly Plate
NOTE
Replace the balance valve O rings with new rings when replacing
the valve.
CAUTION
The unit cannot be reassembled with the canister O-rings in
place. They MUST be removed for reassembly.
The connections will NOT seal if you attempt to reassemble the
unit without removing the O-rings from the canisters.
Page 112
1112893, Ver. 04
Page 113
1112893, Ver. 04
Place the sieve air side assembly on the sieve canisters and secure using (4) T15 torx screws.
torque to 8 IN LB. Ensure the O rings are attached to the air assembly prior to fitting the assem-
bly to the canisters.
CAUTION
The unit cannot be reassembled with the canister O-rings in
place. They MUST be removed for reassembly.
The connections will NOT seal if you attempt to reassemble the
unit without removing the O-rings from the canisters.
Page 114
1112893, Ver. 04
Page 115
1112893, Ver. 04
Remove the sieve O2 side assembly. Refer to sieve O2 side assembly removal procedure.
Remove the sieve air side assembly. Refer to sieve air side assembly removal procedure.
6.3.12.10 Sieve Canister Assembly Installation
1. Arrange the replacement sieve canisters so the air side and O2 aides are aligned. (air side to
air side and O2 side to O2 side)
CAUTION
The unit cannot be reassembled with the canister O-rings in place.
They MUST be removed for reassembly.
The connections will NOT seal if you attempt to reassemble the unit
without removing the O-rings from the canisters.
CAUTION
The canisters in the parts kit are a matched pair and must be replaced
as such. Ensure that new and used canisters are not mixed together.
Page 116
1112893, Ver. 04
2. Install the sieve O2 side assembly. Refer to sieve O2 side assembly installation procedure.
3. Install the sieve air side assembly. Refer to sieve air side assembly installation procedure.
IMPORTANT NOTE
If the sieve canister assembly has been replaced, a Sieve Assembly
Replacement Date/Hours label must be applied. Refer to section
4.3.12.2
Page 117
1112893, Ver. 04
Removal
1. Remove the rear cover.
2. Remove the battery PCA, battery PCA wire harness and battery PCA backing plate.
3. Remove the main PCA and O2 sensor assembly.
4. Remove the compressor.
5. Remove the fan.
6. Remove the speaker.
7. Remove the display wire harness.
8. Remove the sieve canister assembly.
Installation.
Install the new front cover by reversing the removal procedure.
Devices that are currently out in the field and manufactured before September 24th, 2016 that are returned for
repair, return to stock, conversion, or otherwise do not need to be relabeled with UDI compliant serial/model
labeling. However, the device packaging will need to be updated to reflect UDI information.
Devices that are manufactured after September 24th, 2016 with a UDI compliant serial/model label and is
converted to R/U from the base part number will not need new UDI serial/model labeling applied. However, a
new UDI number will be assigned to the corresponding R/U part number for the packaging labeling.
Serial/model labels need to be replaced with UDI compliant labeling if the label has become illegible or the
repair part where the serial label is applied requires replacement.
Page 118
1112893, Ver. 04
Depending on which option is chosen, the printer used needs to ensure the barcode is legible and can be
easily scanned. When reprinting a 2D barcode, a higher resolution, 300 DPI or higher, printer may be
necessary.
Label printer
monochrome
direct thermal / thermal transfer
Roll (4.25 in)
300 dpi
up to 240.9 inch/min
USB
LAN
serial
RECOMMENDED SOFTWARE:
ACTION TAKEN
Page 119
1112893, Ver. 04
PROCEDURE:
1. Print a UDI compliant serial/model number label in accordance with the specifications detailed above.
2. Apply the new label to the product in the same location as the original label.
3. Place a clear overlay over the label, if necessary.
Page 120
1112893, Ver. 04
Page 121
1112893, Ver. 04
Page 122
1112893, Ver. 04
PRESS AND HOLD THE AREA TO THE RIGHT OF THE “SIMPLYGO MINI” TEXT
Page 123
1112893, Ver. 04
DEMO
DEMO FLAG IN UPPER LEFT CORNER INDICATES THE DEVICE IS IN DEMO MODE
Page 124
1112893, Ver. 04
NOTE
Without a “T” fitting, some analyzers can indicate an
erroneous low oxygen concentration value.
IMPORTANT NOTE
DVG (delivery valve gain) Calibration and Adjustment must be performed after replacing the
compressor, Main PCA or delivery valve component.
Page 125
1112893, Ver. 04
5. Refer to the Pulse Volume chart to determine the bolus volume for each pulse.
Page 126
1112893, Ver. 04
6. Use the adjustment arrow buttons in the Delivery Valve Gain box to increase or decrease
the volume of the pulse.
7. When the desired volume is achieved, select the Save button. The adjustment is saved in
the device and the Delivery Valve Gain Confirmation box appears.
Page 127
1112893, Ver. 04
8. Manually adjust the gain for the remaining settings by highlighting and editing the gain
volume. Ignore setting 6.
9. Select the Continue button. The gain calibration is loaded to the device.
10. Select Cancel to exit the process. The original settings are loaded to the device.
Page 128
1112893, Ver. 04
Page 129
1112893, Ver. 04
1. 1.
2. 2.
3. 3.
4. 4.
5. 5.
6. 6.
2. ml 2. ml
3. Divide the SUM by 6 and 3. Divide the SUM by 6
record the average below and record the average
below
4. ml 4. ml
18.7 to 25.3 ml 42.5 to 57.5 ml
Page 130
1112893, Ver. 04
7.5.1 Interfacing the device with the SimplyGo Mini O2 Balance Tool
NOTE
Valve Timing Optimization takes from four to twelve hours.
The SimplyGo Mini MUST be powered with the AC power
supply.
NOTE
Set the test computer power settings to Never Sleep in the
Power Options of the Control Panel. If the computer sleeps
during optimization, you will have to restart the procedure.
NOTE
After a PCA replacement (see Section 6.3.5), the PCA must
have the model and serial number re-programed into the
PCA before using the Balance Tool with the device.
NOTE
The software version numbers represented in these
samples are for example only. The version numbers you
encounter depend on the most recently released version.
Page 131
1112893, Ver. 04
The Valve Timing Optimization and Run In window presents four options:
Add Device option lets you search for and add a connected device to the tool.
Remove Device option lets you remove the device from the tool.
Stop Device option lets you turn off any connected device.
Exit lets you close the tool.
Page 132
1112893, Ver. 04
Page 133
1112893, Ver. 04
9. Select Ok.
Page 134
1112893, Ver. 04
The optimization and run in process begins. This process can take anywhere from 4 to 12 hours to
complete. Do not disconnect or unplug either the device or test computer during that time. Ensure that the
test computer and its monitor are set to Never Sleep in the system’s configuration settings.
Page 135
1112893, Ver. 04
If the unit fails the process, the screen displays the failed parameters in red.
Page 136
1112893, Ver. 04
PASS FAIL:
Signature: Date:
Page 137
1112893, Ver. 04
NOTE
Devices that will have the compressor replaced due to
physical damage will not need to complete the test.
Description
Core i5 2.70 GHz
250GB Hard Drive
4 GB RAM
Monitor
Windows 7 Pro (64 Bit) Operating System
LabVIEW Run-Time 2014
NI VISA 16.0
NI MAX 16.0 (Measurement & Automation Explorer)
FTDI USB to RS232 driver
1 RS232 COM port
1 USB ports on PC
1 Network Connection
1 O2 Interface Kit (no program mode)
Page 138
1112893, Ver. 04
Mc Master Carr_52245K531 Compressor Fitting 1/8’ NPT x ¼ Tube OD (Stainless 1 Test fitting
Steel)
Mc Master Carr_5779K372 Barb fitting (stainless steel) 2 Test fitting
Mc Master Carr_9579K62 Bag of Snap grip clamps (one is used for the test station) 1 Test fitting
Page 139
1112893, Ver. 04
1. National Instrument- Measurement & Automation Explorer (NI-MAX for short) COM port baud rates
and Aliases will need to be setup/configured before the compressor test application can be used.
2. Double click on the NI-MAX shortcut.
3. Click on the “Devices and Interfaces” directories in the left hand pane.
a. You will need to figure out which COM port the Alicat is connected to and select it. In the
center pane toward the bottom click on the “General” tab, type “Alicat” w/o quotes in the field
labeled VISA Alias on My System.
Page 140
1112893, Ver. 04
4. You will need to do the same thing for the COM port the UUT is connected to. Instead of typing
“Alicat” it will be “UUT-ComPort” and for “Port Settings” change the baud rate to 115200.
a. Open the Windows Device Manager and select the COM port the UUT is connected to.
b. Click on the “Port Settings” tab, then click the “Advanced” button and change the “Latency
Timer (msec)” from 16 to 1.
c. Click OK to exit.
Refer to the SimplyGo Mini Service manual 1112893 for more information on setting COM
ports.
The service manual for SimplyGo 1097626 also explains on how to set COM ports.
The PC’s Com Ports must be configured as described in the table below.
Data Bits 8 8
Stop Bits 1 1
SLPM: Press the button shown below and verify the center display reads +0.00
SLPM
AIR
.
Main Menu
PSIA °C TARE V
14.30 24.75
+0.00 SLPM
AIR
+0.00 +0.00
LPM SLPM MENU
Press
Software Rev: Press the buttons shown below to navigate to the MODEL INFO screen and
verify the SW REV displays “5v13.0-xx”. Note SW REV: 6v is not compatible with the
Alicat’s LabVIEW device drivers. Press the MAIN button to return the main menu.
Page 141
1112893, Ver. 04
Press Press
PSIA °C TARE V MFG RS232 MODEL
MICS
14.30 24.75 DATA COMM INFO
SERIAL NO: XXXXXXXX
DATE MFG: XXXXXXXX
ALICAT SCIENTIFIC
DATE CAL: XXXXXXXX
+0.00 SLPM
AIR
www.alicat.com
Ph 520-290-6060
Fax 520-290-0109
CAL BY:
SW REV:
XX
5v13.0-XX
Press Press
Pressure Averaging: Press the buttons shown below to navigate to the PRESS AVG screen
and change the Avg Time Value to 200 msec. Use the SELECT DIGIT , UP arrow, and
DOWN arrow buttons to change the indicated value. Press the SET button when done then
press the MAIN button to return the main menu.
Main Menu
+0.00 SLPM
AIR
Press
200
LCD BACK/
BACK CONSRAST CLEAR
MIAN CANCEL SET
Press Press
Flow Averaging: Press the buttons shown below to navigate to the FLOW AVG screen and
change the Avg Time Value to 200 msec. Use the SELECT DIGIT , UP arrow, and DOWN
arrow buttons to change the indicated value. Press the SET button when done then press the
MAIN button to return the main menu.
Page 142
1112893, Ver. 04
+0.00 SLPM
AIR
Press
200
LCD BACK/
BACK CONSRAST CLEAR
MIAN CANCEL SET
Press Press
RS232 COMM: Press the buttons shown below to navigate to the RS232 COMM screen and
verify the UNIT ID is set to A and the BAUD is set to 19200. Press the MAIN button when
done to return to the main menu.
Main Menu
Press
PSIA °C TARE V MFG RS232 UNIT ID BAUD
MICS
14.30 24.75 DATA COMM A 19200
+0.00 SLPM
AIR
Press Press
SELECT GAS: Press the buttons shown below to navigate to the SELECT GAS screen and
change/verify the Alicat is set to cense AIR. Press the MAIN button when done to return to
the main menu.
Page 143
1112893, Ver. 04
Press
PSIA °C TARE V MFG RS232
MICS UP DOWN PAGE UP DOWN PAGE
14.30 24.75 DATA COMM >Resent Resent
Standard >Standard
COMPOSER User Mixes COMPOSER User Mixes
+0.00 SLPM
AIR
Bioreactor
Breathing
Chromatography
Bioreactor
Breathing
Chromatography
Fuel Fuel
+0.00 +0.00 GAS CONTROL Laser Laser
CANCEL SELECT CANCEL SELECT
LPM SLPM MENU SELECT SETUP MIAN
UP DOWN PAGE
>Air
Argon
CH4 Methane
CO Carbon Monoxide
CO2 Carbon Dioxide
C2H6 Ethane
H2 Hydrogen
HE Helim
CANCEL SELECT
Press
CONTROL SETUP: Press the buttons shown below to navigate to the CONTROL SETUP
screen and change the Alicat’s control mode to SERIAL. Serial refers to a remote digital
RS-232 set-point applied via a serial connection to a PC. Press the MAIN button when done
to return to the main menu.
Main Menu
Press
PSIA °C TARE V MFG RS232 SETPT LOOP SETPT
MICS UP DOWN
14.30 24.75 DATA COMM SOURCE VAR +0.00
>SERIAL
FRONT PANEL
+0.00 SLPM
AIR
ANALOG
Press
Refer to the Alicat’s operating manual TABLE OF CONTENTS for further details on displays and menu
settings.
Page 144
1112893, Ver. 04
Page 145
1112893, Ver. 04
EDIT:
Reinitialize values to default (software reboot to default settings)
OPERATE:
Print at completion
Log at completion
Connect to remote panel: not enabled for use outside of Philips Engineering
TOOLS:
Security: Login and domain account manager are not enabled for outside of Philips Engineering
WINDOW:
Full screen option
ARROW FUNCTION: Program running and brings the test screen up.
ROTATING ARROWS FUNCTION: Refresh the connection to the Alicat
RED CIRCLE FUNCTION: Abort Execution of the test
Page 146
1112893, Ver. 04
Page 147
1112893, Ver. 04
Per Section 7.5, SimplyGo Mini devices are adequately tested for leakage regardless of the implementation of
a core assembly leak test. However, the core assembly leak test can identify leakage earlier in the service
repair process; and thus, aid in rapid troubleshooting.
This test utilizes the SimplyGo Mini Service Tool software. Refer to the install instructions in Section 3.2.
The PC and the SimplyGo Mini Service Tool communicates with the SimplyGo Mini device to control the core
assembly’s process valves.
Page 148
1112893, Ver. 04
NOTE
The argon gas tank, tubes, and tube fixtures are not shown in the recommended test station equipment table
above and should be obtained at the discretion of the test rig preparer (e.g., tubes should be sized to work with
the flow meter of choice, tube fittings should be chosen based on the selected tubes, etc.).
Additionally, ensure that the flowmeter purchased for testing is compatible with your their test set up (e.g.,
configure for Argon gas when using Argon to test for leaks).
Page 149
1112893, Ver. 04
o The primary pressure gauge on the argon gas cylinder should always remain above 100
PSI during testing. The secondary pressure gauge should be set to 40 - 60 PSI.
o The precision regulator downstream of the gas cylinder should be collocated with the
test setup and be set to 12 PSI. It should always read approximately 12 PSI during test-
ing.
Example Regulator
Page 150
1112893, Ver. 04
Example Valves
o The flow protection orifice protects the flow meter from damage that could be caused
from excessive flow.
Example Orifice
o The flow meter measures the mass flow of gas into the core assembly. If the core is as-
sembled correctly, this flow should be approximately zero. An example flow meter is
Page 151
1112893, Ver. 04
o A T-connector fitting reconnects the flows paths from the 3-way bypass valve.
o A plug is used to occlude the product tank outlet.
Page 152
1112893, Ver. 04
o The core should be removed from the unit, placed on an electrostatic discharge surface
close enough to the SimplyGo Mini device to plug the process valve connector into the
device.
o The SimplyGo Mini device should have its power brick plugged in.
COM port procedure:
o The SimplyGo Mini Service Tool consists of a PC running a test application developed in
LabVIEW. The test PC should be configured to have at least one (1) dedicated COM
port to communicate to the SimplyGo Mini device, any available COM port is acceptable.
o The PC’s COM Ports must be configured as follows:
Com Port User’s Discretion
Baud 115200
Data Bits 8
Parity None
Stop Bits 1
Flow None
Page 153
1112893, Ver. 04
Page 154
1112893, Ver. 04
Page 155
1112893, Ver. 04
PASS FAIL
NOTES:
Page 156
1112893, Ver. 04