Instruktioner
Dokumentnummer/Document Number Titel/Title Sida/Page
13241-2 Manual - BM800 Data Interface 1/16
Utfärdare/Originator Godkänd av/Approved by Gäller från/Effective date
Mats Rova Per Lofbom 2011-05-17
1. Introduction
BM800 Data Interface (BDI) is intended for transferring data from BM800 instrument to PC or as a sample data
interface between BM800 and LIS systems.
Features
- Transfer data from BM800 to PC via local COM port (RS232, USB) or Network (TCP Client, Server)
- Parse sample data from files
- Supports multiple channels / instruments (hardware dependent)
- Supports multiple connection types (RS232, USB, TCP/IP)
- Selectable parameter collections (Reduced, Standard, Extended)
- Multiple output formats (LIS, Text, Excel)
- Validation of transferred data
2. Minimum Requirements
Hardware
Pentium 4, 512 MB RAM
Operating System LIS Text Excel
Windows 2000 .Net 2.0* x x -
Windows XP SP3 .Net 2.0* x x -
Windows XP SP3 + .Net 2.0* + Microsoft Office XP x x x
*.NET is automatically installed by the BDI installer
3. Installation
See Installation Instructions (included in the zip-file)
NOTE 1 This manual refers to version 1.01 of BDI software. Future revision changes could make this information
obsolete. Only use the manual that was included in the installed software package.
NOTE 2 Always set Serial Output Setup - Send with Ack to enabled according to section 5 in order to ensure safe data
transfer between BM800 and PC (See section 4).
4. Windows Environment
Program Location: Setup will create a shortcut to the program in the start menu.
Default output Folder: Default folder for sample output is Program Files/Boule Medical/BM800_Data. Setup will also
create a shortcut to the output folder on the desktop.
Utskriftsdatum:2011-05-17 16:14
Original lagras och godkänns elektroniskt. Kontrollera att utskriften är giltig genom att jämföra med utgåvan i Centuri.
5. Typical Configurations
RS232 Configuration – One channel
RS232
RS232
A single BM800 is connected to PC via COM port (RS232)
1. Connect BM800 to an available COM port on PC
2. Configure BM800 communications settings as follows
3. Configure BDI channel settings, where “Available Port(s)” is the connected COM port.
13241-2 Manual - BM800 Data Interface Utskriftsdatum: 2011-05-17 16:14:00 2/16
Original lagras och godkänns elektroniskt. Kontrollera att utskriften är giltig genom att jämföra med utgåvan i Centuri
USB Configuration – One Channel
USB
USB
A single BM800 is connected to PC via USB port
1. Connect BM800 to USB device port on PC (see image above for position)
2. Install USB drivers (see Installation documentation, included in the zip file)
3. Configure both BM800 communications settings as follows (if changing USB VID&PID the
instrument needs to be restarted for the changes to take effect).
Note: If using firmware below 2.7.1 VID & PID selection will not be available and defaults to “1”
4. Configure BDI channel settings, where “Available Port(s)” is the virtual USB port
13241-2 Manual - BM800 Data Interface Utskriftsdatum: 2011-05-17 16:14:00 3/16
Original lagras och godkänns elektroniskt. Kontrollera att utskriften är giltig genom att jämföra med utgåvan i Centuri
TCP Server Configuration
COM
COM
TCP/IP
COM
COM
Typical setup (connecting multiple BM800 to single device server e.g. Moxa Nport 5610)
1. Connect BM800 to available port on device server (see your device server manual)
2. Configure your device server to act as client and setup IP address (the server IP address) and port number
3. Configure both BM800 communications settings as follows
4. Configure BDI connection as Server and configure IP address and Port to match the
the device server (Use the “Channels – Add” button to add a channel for next BM800)
13241-2 Manual - BM800 Data Interface Utskriftsdatum: 2011-05-17 16:14:00 4/16
Original lagras och godkänns elektroniskt. Kontrollera att utskriften är giltig genom att jämföra med utgåvan i Centuri
TCP Client Configuration
COM
COM
TCP/IP
COM
COM
Typical setup (connecting multiple BM800 to single device server)
1. Connect BM800 to available port on device server (see your device server manual)
2. Configure your device server to act as server and setup IP address (the server IP address) and port number
3. Configure both BM800 communications settings as following
4. Configure BDI connection as Client and configure IP address and Port to match the
device server (Use the “Channels – Add” button to add a channel for next BM800)
13241-2 Manual - BM800 Data Interface Utskriftsdatum: 2011-05-17 16:14:00 5/16
Original lagras och godkänns elektroniskt. Kontrollera att utskriften är giltig genom att jämföra med utgåvan i Centuri
6. Main Program Window
Connection / Data Log Displays configured connections and status. This text box can also be
configured to display real-time status of parsed sample data.
File - Open File Opens a file (raw BM800 data captured via e.g. Hyper Terminal) to
parse data.
File - Exit Close BDI
Tools - Settings Open settings window
Tools - Restart Channel Restarts a single channel (only communication is restarted, not
BM800).
Tools - Restart All Channels Restarts all channels (only communication is restarted, not BM800).
Tools - Datalog / No Datalog Select if parse information for current sample shall be displayed in the
Connection / Data Log window.
Tools - Clear Screen Clears the Connection / Data Log screen.
Help Manual Opens this manual
Help About BDI version information
13241-2 Manual - BM800 Data Interface Utskriftsdatum: 2011-05-17 16:14:00 6/16
Original lagras och godkänns elektroniskt. Kontrollera att utskriften är giltig genom att jämföra med utgåvan i Centuri
7. Settings Window
7.1. Instrument Connection
The instrument connection panel displays all current active channels and settings.
Connection Select type of connection to be used for this channel (RS232/USB, TCP
Server, and TCP Client). Depending on selected type, COM Port, Server
Mode or Client Mode will be highlighted.
COM Port - Available Port(s) Display a list of available COM ports.
COM Port - Selected Port Displays current selected port (if selected port is a virtual USB port,
BM800 S/N will also be displayed).
COM Port - Baud Rate Selected baud rate (not changeable).
Server Mode - Local IP Address Device server IP Address.
Server Mode - Port Port number to be used.
Server Mode - Server Name Used only for reference if more than one device server is used.
13241-2 Manual - BM800 Data Interface Utskriftsdatum: 2011-05-17 16:14:00 7/16
Original lagras och godkänns elektroniskt. Kontrollera att utskriften är giltig genom att jämföra med utgåvan i Centuri
Client Mode - Server IP Address Display PC local IP address (if more than one network adapter is present
on the PC, only the first will be shown. To change to second adapter
manually enter IP Address)
Client Mode - Port Port number to be used.
Client Mode - Server Name Used only for reference if more than one device server is used.
Output Directory Path to this channels selected output folder.
7.2. File Format
Select which output format to be used. See section 7 File Format for detailed information. This is a global option
affecting all channels.
LIS Output is LIS format.
Text Output is Text format.
Excel Output is Excel format.
7.3. Sample Data
Select data to be parsed and stored on file. See section 8 Transferred Data for detailed information on which
parameters that are transferred. This is a global option affecting all channels.
Standard Collection Standard set of Instrument information, sample information and
parameters.
Reduced Collection Instrument information and sample information are reduced.
Extended Collection If check-boxes “Transfer all instrument data” and “Transfer all
sample information data” are checked, all BM800 parameter data is
transferred.
7.4. Channels
Add Add a channel to connect additional BM800 (default is one channel).
Remove Remove selected channel.
Default Set settings to default factory mode.
8. File Format
LIS - Format Description LIS files are intended to be used as interface between BM800 and installed
LIS system. All samples are stored in separate .txt files. Parameters are
stored as TAB separated ASCII values. LIS files are in a fixed collection
format i.e. it is not possible to change which parameters are transferred.
LIS - File name Sample ID_Date_Sequence.txt
Example: AUTOBACKGROUND_2010-03-23_s7979.txt
Text - Format Description Similar format as LIS but with the option to change collection size.
Text - File name Sample ID_Date_Sequence.txt
Example: AUTOBACKGROUND_2010-03-23_s7979.txt
Excel - Format Description All samples are stored in a single Excel file. A new file is created each
day.
Excel - File name BM800_Instrument ID_Date.xls
Example: BM800_50559_2010-03-25.xls
13241-2 Manual - BM800 Data Interface Utskriftsdatum: 2011-05-17 16:14:00 8/16
Original lagras och godkänns elektroniskt. Kontrollera att utskriften är giltig genom att jämföra med utgåvan i Centuri
N.B Excel Important Information If you change sample collection size, you need to archive the old file and
delete the original (a new will be created automatically when first sample
is stored). This is to ensure sure that Excel column information
corresponds to current collection.
9. Transferred Data
For all collection types (standard, reduced and extended) possible future parameter data will be included
automatically.
Standard Collection SNO, IID, ID, SEQ, DATE, SORC, ASPM, ASPS, BLMD, BLNK,
ASWP, WDMA, CLVL, CEXP, APNA, RDPN, RDED, RLLI, RLPN,
RLED, RCLI, RCPN, RCED, RBC, MCV, HCT, MCH, MCHC, RDWR,
RDWA, PLT, MPV, PCT, PDW, LPCR, HGB, WBC, LA, MA, GA, LR,
MR, GR, EA, ER
Reduced Collection SNO, IID, ID, SEQ, DATE, SORC, ASPM, ASPS, BLMD, BLNK,
ASWP, WDMA, APNA, RBC, MCV, HCT, MCH, MCHC, RDWR,
RDWA, PLT, MPV PCT, PDW, LPCR, HGB, WBC, LA, MA, GA, LR,
MR, GR, EA, ER
Extended Instrument PRDI, SNO, BRND FIWV, IAPL, IID, ID, SEQ, DATE, SORC, ASPM,
ASPS, BLMD, BLNK, ASWP, WDMA, CLVL, CEXP, APNA, RDPN,
RDED, RLLI, RLPN, RLED, RCLI, RCPN, RCED, RBC, MCV, HCT,
MCH, MCHC, RDWR, RDWA, PLT, MPV, PCT, PDW, LPCR, HGB,
WBC, LA, MA, GA, LR, MR, GR, EA, ER
Extended Sample SNO, IID, ID, SEQ, DATE, SORC, STYP, ASPM, ASPS, BLMD, BLNK,
ASWP, RPD, RPDS, RPDL, RPDH, RPDF, WDMS, WDMA, WDFB,
WDLL, WDLH, WDCL, WDCH, LGL, WDIL, WDIH, WDDM, WDDP,
CAPL, CLVL, CEXP, CEXT, RGED, RGEL, RGEC, APNU, APNA,
XLT, RDLI, RDPN, RDED, RLLI, RLPN, RLED, RCLI, RCPN, RCED,
RBC, MCV, HCT, MCH, MCHC, RDWR, RDWA, PLT, MPV, PCT,
PDW, LPCR, HGB, WBC, LA, MA, GA, LR, MR, GR, EA, ER
13241-2 Manual - BM800 Data Interface Utskriftsdatum: 2011-05-17 16:14:00 9/16
Original lagras och godkänns elektroniskt. Kontrollera att utskriften är giltig genom att jämföra med utgåvan i Centuri
10. Parameters
10.1. Static Instrument Parameters
The static instrument parameters are parameters that are always the same during the instrument life cycle.
9.1.1 SNO - Serial Number
SNO is a 5-digit serial number. This number is unique for each instrument.
Value range: 00001 – 65534.
10.2. Instrument Configuration Parameters
The instrument configuration parameters are "semi-static" instrument parameters. The parameters are
probably the same during the instrument life cycle, but may change, for example after a software upgrade.
9.2.1 FIWV - Firmware Version
FIWV identifies the software used in the instrument. This parameter is always present with a value.
9.2.2 IID - User-Settable Instrument ID
IID identifies an individual BM800 instrument. It is a user settable string of 1-15 characters. This
parameter is always present, but the value is only present if the instrument ID is set.
10.3. Sample Identification Parameter Values
Sample identification parameter values identify the sample run.
9.3.1 ID - Blood Sample ID
ID is the identification of the blood sample. This parameter is always present, but the value is absent for
"blank" ID:s.
Value range: 1 - 15 characters
Note 1: Non-numeric characters are allowed. Leading or trailing white space will be filtered away.
Note 2: Non-ASCII characters may be used.
Note 3: All control blood sample ID:s are a number followed by a trailing "+". (The number is normally
the digits of the lot number of the control blood.)
Note 4: Leading zero digits in ID numbers are significant.
Note 5: The maximum number of characters in ID numbers may change.
9.3.2 SEQ - Sample Sequence Number
SEQ is the sequence number of the sample run. This parameter is always present with a value.
Value range: 1 - 9999
Note 1: SEQ numbers are not unique in samples from one instrument. The operator can set the sequence
number of the next sample, and sequence numbers wraps around.
Note 2: The upper limit of sequence numbers may change.
9.3.3 DATE - Sample Date and Time
DATE is the date and time of the sample run. This parameter is always present with a value except when
the date and time is not set in the instrument. The value is always formatted according to ISO 8601. See
example below. The date and time is always in the local time zone. No time zone info is provided.
Example: 2004-05-06T07:08:09
9.3.4 SORC – Sample or Control
SORC indicates if it is a normal sample run or a control sample run. This parameter is always present
with a value.
Value range: 0: normal sample
2: control sample with reference ranges
13241-2 Manual - BM800 Data Interface Utskriftsdatum: 2011-05-17 16:14:00 10/16
Original lagras och godkänns elektroniskt. Kontrollera att utskriften är giltig genom att jämföra med utgåvan i Centuri
Note 1: More sample or control types may be added. They will all be small positive numbers.
10.4. Aspiration Parameters
9.4.1 ASPM - Aspiration Mode
ASPM indicates the aspiration mode for the sample. This parameter is always present with a value.
Value Description: "OT" Open Tube
"PD" Pre-diluted sample
"CT" Closed Tube = Cap piercer
"MC" Micro capillary = MPA / MCI
"AS" Auto sampler
9.4.2 ASPS – Aspiration Status
ASPS indicates the aspiration status for the run. This parameter is always present with a value.
The value is a status flag indicating if blood (or some other sample type) was detected during aspiration.
Value range: 0: no blood (or other) detected
1: blood (or other) detected
2: indeterminate due to e.g. timeout during aspiration
9.4.3 BLMD - Background Mode Sample Run
BLMD indicates if the sample was run in background mode. Background mode is used for blank test runs.
This parameter is always present with a value.
Value range: 0: not background mode
1: background mode
Note: the parameter BLNK indicates if the sample actually was a blank sample.
9.4.4 BLNK - Blank Sample
BLNK indicates if the sample was a blank sample run or not, determined by the instrument. This
parameter is always present with a value.
Value range: 0: not a blank sample
1: a blank sample
Note: The definition of a blank is RBC < 0,50 and HGB < 2,0 (below approx 15% of a normal human blood
sample).
9.4.5 ASWP - Auto-sampler Wheel Position
ASWP indicates the wheel position for an auto-sampler run. This parameter is always present. The value is only
present for auto-sampler samples.
Value range: 1 - 20.
10.5. Differentiation – White Cells
The BM800 instrument has several different methods to separate the complete WBC population into three
subpopulations: LYM, MID , GRAN (in increasing size order). The instrument configuration allows or
disallows selection of the different WBC diff methods.
9.5.1 WDMA - WBC Diff Method Actually Used
WDMA indicates the floating WBC diff method actually used during the calculations. This parameter is
always present. The value is only present if the instrument actually tried a WBC diff.
Value range: 1: DM1 human floating
2: DM2 fixed
4: DM4 vet floating
Note 1: more diff methods may be added. They will all be small positive numbers.
13241-2 Manual - BM800 Data Interface Utskriftsdatum: 2011-05-17 16:14:00 11/16
Original lagras och godkänns elektroniskt. Kontrollera att utskriften är giltig genom att jämföra med utgåvan i Centuri
10.6. Control Blood with Reference Ranges
The operator inputs control blood assay information to specify reference ranges and other information about
a control blood. All control blood lots have a unique ID number that is used to separate different control
blood lots.
Note 1: The unique ID number is normally the digits from the control blood lot number followed by a plus
(+) sign. Example: lot number “0606-123” gets the ID number “0606123+”.
Note 2: Leading zero digits in control blood ID numbers are significant.
9.6.1 CLVL – Control Blood Level
CLVL identifies the level (normal/low/high/calibrator) of a control blood with reference ranges. This
parameter is always present. The value is only present if the blood was a control blood with reference ranges.
The value is a one-character string: "N" for normal level, "L" for low level, "H" for high level, and "C" for
calibrator. Other control blood levels may be added. All control blood level codes will be one uppercase letter.
Note 1: This information comes from the control blood reference range assay information.
Note 2: The control blood level is also present as a name in the parameter APNA (see 9.9.2)
9.6.2 CEXP - Expiry Date of a Control Blood with Reference Ranges
CEXP is the expiry date of a control blood with reference ranges. This parameter is always present. The
value is only present if the blood was a control blood with reference ranges.
The value is a date, always formatted according to ISO 8601. See example below.
Note: This information comes from the control blood reference range assay information.
Example: 2005-03-09
10.7. Settings
9.7.1 APNA – Analysis Profile Name / control blood level name
APNA is the name (if defined) of the Analysis Profile (species in vet instruments) selected for the sample. This
parameter is always present. The value is present if a name is defined for the selected Analysis Profile (species).
A qualified operator can set the value of this parameter.
The value is a string of 1 to 15 characters.
Note 1: Non-ASCII characters may be used.
Note 2: The initial instrument setup uses "BLOOD" for the “blood” analysis profile, and
"BACKGROUND" for “background”. The control bloods with reference ranges have "LOW",
"NORMAL", "HIGH", and "CALIBRATOR".
Note 3: The predefined names in the previous note may be translated to other languages.
Note 4: A qualified operator can set the names of analysis profiles.
10.8. Reagent statistics
The BM800 instrument has an optional reagent statistics system that keeps track of the reagents used in the
instrument.
Note: The diluent and hemolyzing regent information is exactly the same if a reagent “combo” pack is used.
9.8.1 RDLI Reagent Statistics Diluent Lot identifier
RDLI is the lot identifier of the diluent reagent. This parameter is always present. The value is only present if the
reagent statistics system is enabled.
The value is always formatted as “YYMM-NNN”, where YY, MM and NNN are digits. (Normally YY =
production year, MM =production month and NNN = lot sequence number.) See example below.
Note 1: This information comes from the reagent package information.
Note 2: leading zero digits in reagent lot identifiers are significant.
Example: 0606-123
13241-2 Manual - BM800 Data Interface Utskriftsdatum: 2011-05-17 16:14:00 12/16
Original lagras och godkänns elektroniskt. Kontrollera att utskriften är giltig genom att jämföra med utgåvan i Centuri
9.8.2 RDED Reagent Statistics Diluent Expiry date
RDED is the expiry date of the diluent reagent. This parameter is always present. The value is only present if the
reagent statistics system is enabled.
The value is a date, always formatted according to ISO 8601. See example below.
Note: This information comes from the reagent package information.
Example: 2006-06-29
9.8.3 RLLI Reagent Statistics Hemolyzer Lot identifier
RLLI is the lot identifier of the hemolyzing reagent. This parameter is always present. The value is only
present if the reagent statistics system is enabled.
The value is always formatted as “YYMM-NNN”, where YY, MM and NNN are digits. (Normally YY =
production year, MM =production month and NNN = lot sequence number.) See example below.
Note 1: This information comes from the reagent package information.
Note 2: leading zero digits in reagent lot identifiers are significant.
Example: 0606-124
9.8.4 RLED Reagent Statistics Hemolyzer Expiry date
RLED is the expiry date of the hemolyzing reagent. This parameter is always present. The value is only
present if the reagent statistics system is enabled.
The value is a date, always formatted according to ISO 8601. See example below.
Note: This information comes from the reagent package information.
Example: 2006-06-29
9.8.5 RCLI Reagent Statistics Cleaner Lot identifier
RCLI is the lot identifier of the cleaning reagent. This parameter is optional, and may be present in vet
instruments. The value is only present if the reagent statistics system is enabled.
The value is always formatted as “YYMM-NNN”, where YY, MM and NNN are digits. (Normally YY =
production year, MM =production month and NNN = lot sequence number.) See example below.
Note 1: This information comes from the reagent package information.
Note 2: leading zero digits in reagent lot identifiers are significant.
Example: 0606-125
9.8.6 RCED Reagent Statistics Cleaner Expiry date
RLED is the expiry date of the cleaning reagent. This parameter is optional, and may be present in vet
instruments. The value is only present if the reagent statistics system is enabled.
The value is a date, always formatted according to ISO 8601. See example below.
Note: This information comes from the reagent package information.
Example: 2006-06-29
13241-2 Manual - BM800 Data Interface Utskriftsdatum: 2011-05-17 16:14:00 13/16
Original lagras och godkänns elektroniskt. Kontrollera att utskriften är giltig genom att jämföra med utgåvan i Centuri
10.9. Sample Result Parameter Values
Sample result parameter values are the measured and calculated hematology parameters from a sample run.
The parameter order below is not fixed. The parameters might be in another order. Parameter definitions may be
added or removed with new software releases.
9.9.1 Result Parameter Values
Figure 3: Parameter values
9.9.2 Result Parameter Format
For each parameter a set of 5 values are present
Parameter Name Value
ParamaterName Actual parameter value
ParameterName_L Low range value
ParameterName_H High range value
ParameterName_F Flag value
ParameterName_O Out of range flag
A result parameter always has a Parameter name and a normal / reference range (ParamaterName_L,
ParamaterName_H) . It could also have a value (ParamaterName) or an out-of range (ParamaterName_O)
13241-2 Manual - BM800 Data Interface Utskriftsdatum: 2011-05-17 16:14:00 14/16
Original lagras och godkänns elektroniskt. Kontrollera att utskriften är giltig genom att jämföra med utgåvan i Centuri
flag, and an error flag (ParamaterName_F).
Example (PLT, Background within range, no error flag present)
PLT 3
PLT_L 0
PLT_H 10
PLT_F
PLT_O
9.9.3 Parameter Value
The data type of the result parameters is either floating point or scaled integer.
For example, if the instrument counts RBC as 0.00, then there are no cells to calculate an MCV value from. The
instrument reports the RBC value as 0.00, since it did actually not count any cells, but the MCV value is absent.
The value range and the number of decimals depend of the parameter.
The parameter value and out of range value are mutually exclusive. A value is either not calculated (neither
"value" nor "out of range value", calculated ("value" only), or “out-of-range " only.
9.9.4 Parameter Out-of-Range
The " ParameterName_O " tag is only present if the parameter value is either completely out-of-range, or could
not be determined for some unusual reason. For example, if the MCV value is absent due to a blank run, then
there is no " ParameterName_O " tag. But if the RBC value is absent due to a ridiculously high count (above 14
in human models), then BM800 outputs a " ParameterName_O: H".
The " ParameterName_O " flag can take the values "H" (value too high) or "L" (value too low).
The " ParameterName_O " and " ParameterName" values are mutually exclusive. A value is either not calculated
(neither " ParameterName " nor " ParameterName_O "), calculated ("ParameterName" only), or out-of-range
("ParameterName_O " only).
9.9.5 Parameter Error Flag
The " ParameterName_F" value is an optional parameter error flag. Some errors block the calculation of the
corresponding parameter value, while others provide additional information.
The error flag is always two upper-case letters, or one upper-case letter followed by a digit. The list of
possible flags are not included here.
Note: Only these parameters can have an error flag: RBC, MCV, PLT, HGB, WBC, LYM, MID, GRAN.
9.9.6 Parameter Normal/Reference Range Low
The "ParameterName_L" value is always present. Its value is the low end of the normal / reference(*) range. If
the parameter value is exactly equal to "ParameterName_L", then it is within the normal / reference range. (* It is
called the normal range for normal blood, and reference range for control blood with reference ranges.)
9.9.7 Parameter Normal/Reference Range High
The "ParameterName_H" tag is always present. Its value is the high end of the normal / reference range. If the
parameter value is exactly equal to " ParameterName_H", then it is within the normal / reference range.
11. Limitations and Requirements
BDI only supports Excel output for PC with Microsoft Office XP and above installed.
BM800 ‘Send with Ack’ must be set to active for proper data transfer (NOTE! If “Send with Ack” is not set
sample data could be lost without any warnings.
If a sample have been lost and “Send with Ack” is activated, the error will be indicated by BM800 (A beep and a
Serial Output Alarm, no further samples will be sent until the issue is acknowledged manually).
All BM800 sample data must be deleted manually by user input on the instrument. BDI does not change any data
or settings on the instrument.
13241-2 Manual - BM800 Data Interface Utskriftsdatum: 2011-05-17 16:14:00 15/16
Original lagras och godkänns elektroniskt. Kontrollera att utskriften är giltig genom att jämföra med utgåvan i Centuri
12. FAQ
Samples are lost when transferred from BM800 to computer
1. Make sure that BM800 is set to Send with Ack under Advanced -> Setup - >Serial.
BDI validates received data. All data that fails validation will be discarded. BDI signals BM800 if a sample has
been transferred successfully by transmitting an acknowledge message. If Send with Ack is not set on BM800
there is no possibility for BM800 to acknowledge that a sample needs to be re-transmitted.
2. Ensure that the computer meets stated HW requirements / Ensure that no unnecessary programs/services are
running on the computer.
BM800 will timeout within 5 seconds after a sample is sent. If the computer is too slow, it may not allow for
validation and parsing within this time frame.
How do I know which COM port to use?
Connected via USB: Instrument serial number will be shown in Settings -> Selected Port.
Connected via RS232: You need to know which COM port connects to a physical port. One way to resolve this is
to:
1. Use Settings -> Selected Port to select a COM port and click Save.
2. If the program successfully connects to an instrument the serial number will be shown in the connection
log. If the connection fails serial number will be shown as N/A in the connection log. Retry with another
COM port.
I disconnected the USB cable and now the program does not work
1. Click Tools -> Restart All Channels.
2. Remove and re insert USB cable at PC end.
3. Click Tools -> Restart All Channels.
4. Check the “Connection / Data Log” text box in main window to see if connection is on-line
If above 4 steps does not work close the application and restart.
Local IP Address is wrong when I setup the program in Server Mode
This is set automatically to the address of the first detected network adapter in your PC. If you have two or more
network adapters you need to change the address manually to the one you want. To find the address of available
network adapters: start Command Prompt (Start -> Programs -> Accessories -> Command Prompt) and type
ipconfig.exe, then hit Enter key. A list of available adapters with corresponding IP Addresses will be shown.
I changed the Sample Data collection size and now Excel tables are not matching with previous data?
When changing collection size it is recommended to archive current excel file and then delete it (a new will be
created automatically as the first sample is parsed). BDI creates a new excel file each day. Within that day the
collection size must be fixed.
Where is the sample data?
Default location can be found by opening the shortcut BM800 Sample Data on your desktop
I sent the same sample twice but only one was transferred?
BDI will ignore the second sample if two consecutive samples have the same sequence number and Sample ID.
Why can I not change sample collection size when saving as LIS?
LIS output is the same as Text but with fixed collection size (Extended)
Boule Medical (2010)
http://www.boule.se
13241-2 Manual - BM800 Data Interface Utskriftsdatum: 2011-05-17 16:14:00 16/16
Original lagras och godkänns elektroniskt. Kontrollera att utskriften är giltig genom att jämföra med utgåvan i Centuri