Multi-Purpose, Multi-Tissue

Ultrasound Phantom

The Standard for Ultrasound


Quality Assurance
• Unique dual attenuation design allows testing on low
frequency abdominal probes up to 5 MHz and high
frequency probes to 15 MHz and higher
• Detachable water wells allow for testing curvilinear
and endocavity probes
• Only general purpose QA phantom on market with
elasticity targets
• Ensure over ten years of reliable use through rein-
spection and repair services
Includes best in industry four-year warranty

The CIRS Model 040GSE Multi-Purpose, Multi-Tissue Key Tests with Model 040GSE
Ultrasound Phantom is the most complete solution avail-
able for performance and quality assurance testing. Its
• Uniformity
dual frequency design and detachable water wells allows
testing of most transducer shapes – including curvilinear • Depth of Penetration
and endocavity – and frequencies. It is also the only QA • Beam Profile/ Focal Zone/ Lateral Response
phantom on the market that provides both elasticity tar- Width
gets and all the standard B-mode imaging test objects.
• Vertical Distance Measurement
The phantom is made of CIRS' proprietary Zerdine®
hydrogel polymer, which has been formulated to provide • Horizontal Distance Measurement
tissue mimicking properties including compatibility with • Axial and Lateral Resolution
harmonic imaging. To maximize phantom lifetime, this
• Elevational Resolution
gel is contained in a rugged ABS plastic housing with a
Saran-based laminate membrane. • Contrast Resolution

CIRS ultrasound QA phantoms come standard with a • Grayscale Contrast Sensitivity


robust housing, carry case, 48-month warranty, and user • Elasticity Sensitivity
guide.
• Dead Zone Assessment

900 Asbury Ave • Norfolk, Virginia 23513 • USA • Tel: 757-855-2765 • WWW.CIRSINC.COM
MULTI-PURPOSE MULTI- TISSUE ULTRASOUND PHANTOM
MULTI-PURPOSE MULTI- TISSUE ULTRASOUND PHANTOM Model 040GSE

Overall Ultrasound image of Model 040GSE

Phantom comes with detachable


scanning wells to accommodate large
sector probes and small endocavity
probes. It is packaged in a foam lined
carry case.

SPECIFICATIONS
MODEL 040GSE INCLUDES
DIMENSIONS 17.8 cm x 12.7 cm x 20.3 cm QTY COMPONENT DESCRIPTION
(7" x 5" x 8”)
1 Multi-Purpose, Multi-Tissue Ultrasound Phantom
PHANTOM WEIGHT 11 lbs. (4.1 kg) 1 Detachable Protective Cover
HOUSING MATERIALS ABS Plastic 1 Detachable Water Well (1cm deep)
MEMBRANE Saran-based laminate 1 Detachable Endocavity Well
TISSUE-MIMICKING Zerdine® solid elastic hydrogel 1 Carry Case
MATERIAL
- 48-Month Warranty
- User Guide
-

ZERDINE® PROPERTIES AXIAL-LATERAL RESOLUTION GROUPS


Freezing point: 0° C Wire diameter: 80 microns
Melting point: Above 100° C
Speed of Sound: 1540 m/s Group 1&2 Depths: 3 & 6.5 cm
Other: Compatible with harmonic imaging Axial separation: 4, 3, 2, 1, 0.5 & 0.25 mm
Lateral separation: 4, 3, 2, 1, 0.5 & 0.25 mm
VERTICAL DISTANCE GROUP
Number of targets: 16 Group 3 Depths: 10.5 cm
Wire diameter: 100-micron, nylon monofilament Axial separation: 5, 4, 3, 2 & 1 mm
Depth range: 1 to 16 cm Lateral separation: 5, 4, 3, 2 & 1 mm
Spacing: 10 mm
ANECHOIC STEPPED CYLINDERS
HORIZONTAL DISTANCE GROUP Number of targets: 12
Number of groups: 2 Diameter of targets: 1.3, 2.0, 3.0, 4.5, 6.7 & 10.0 mm
Wire: 100-micron, nylon monofilament Depth of Targets: 1.5, 4.5, 7.0, 10.0, 13.0, 16.0 cm
Depths: 4 & 9 cm Contrast: Anechoic, Cyst-like, 1mm
Number of Targets: 6 & 7 respectively
Spacing: 10 & 20 mm respectively GRAY SCALE TARGETS
Group 1: 3 cm deep, Ø 8 mm
NEAR FIELD GROUP Contrast: -9, -6, -3, +3, +6 & > +15 dB with respect to back
Number of targets: 5 ground
Wire Diameter: 100 microns Group 2: 11.5 cm deep, Ø 10mm
Depth range: 1 to 5 mm Contrast: -6, -3, +3, +6 & > +15 dB with respect to back
Distance b/w Targets: 1 mm
ground
ELASTICITY TARGETS
Group 1: 1.5 cm deep, Ø 6 mm
Group 2: 5 cm deep, Ø 8 mm
Elasticity: 10, 40 & 60 kPa
Grayscale Contrast: 0 db with respect to background

©
2013 Computerized Imaging Reference Systems Inc. All rights reserved. Computerized Imaging Reference Systems, Inc. has been certified by UL
Specifications subject to change without notice. DQS Inc. to (ISO) 13485:2016. Certificate Registration No.10000905-
Publication: 040GSE DS 120418 MP2016.