08 Chapter3 PDF
08 Chapter3 PDF
CharacterizationTechniques
and
Instrumentation
0
SrFe12O19 hexaferrite samples synthesized by three different routes are
characterised by using different experimental techniques namely; Fourier Transform
Infrared (FTIR) spectroscopy, Differential Thermal Analysis (DTA), Differential
Scanning Calorimetry (DSC), Thermo Gravimetric Analysis (TGA), X-ray
Diffraction (XRD), Scanning Electron Microscopy (SEM), Vibrating Sample
Magnetometer (VSM) and Dielectric measurements. Details of these characterization
techniques are discussed in present chapter. The effect of Swift Heavy Ion (SHI)
irradiation on SrFe12O19 hexaferrites is studied. The details of SHI irradiation is
discussed in this chapter.
The IR region is commonly divided into three smaller areas: near - IR (400 -
10 cm-1), mid - IR (4000 - 400 cm-1), and far - IR (14000 – 4000 cm-1). Infrared
photons have enough energy to cause groups of atoms to vibrate with respect to the
bonds that connect them. Like electronic transitions, these vibrational transitions
correspond to distinct energies, and molecules absorb infrared radiation only at certain
wavelengths and frequencies. Chemical bonds vibrate at characteristic frequencies,
- 43 -
and when exposed to infrared radiation, they absorb the radiation at frequencies that
match their vibration modes. Measuring the radiation absorption
absorption as a function of
frequency produces a spectrum that can be used to identify functional groups and
compounds. Some impurities produce their own characteristic bands in infrared
region. Spectral measurements of these bands are used to determine concentration
conce of
the impurities and their bonding with the host materials. In order to make
identification, the measured interferogram signal can not be interpreted directly. A
means of “decoding” the individual frequencies is required. This can be accomplished
via a well-known
known mathematical technique called the Fourier transformation.
transformation This
transformation is performed by the computer which then presents the user with the
desired spectral information for analysis.
- 44 -
active components: a moving mirror, a fixed mirror, and a beam splitter (Figure 3.1).
The two mirrors are perpendicular to each other. The beam splitter is a semi-reflecting
device and is often made by depositing a thin film of germanium onto a flat KBr
substrate. Radiation from the broadband IR source is collimated and directed into the
interferometer, and impinges on the beam splitter.
IR spectra can be measured using liquid, solid, or gaseous samples that are
placed in the beam of infrared light. A drop of a liquid can be placed as a thin film
between two salt plates made of NaCl or KBr, which are transparent to infrared light
at most important frequencies. A solid can be ground with KBr and pressed into a disk
that is placed in the light beam. Alternatively, a solid sample can be ground into a
pasty mull with paraffin oil. As with a liquid, the mull is placed between two salt
plates. Solids can also be dissolved in common solvents such as CH2Cl2, CCl4, or CS2
that do not have absorptions in the areas of interest. Gases are placed in a longer cell
with polished salt windows. These gas cells often contain mirrors that reflect the beam
through the cell several times for stronger absorption.
The basic instrument design is quite simple. Figure 3.2 (a) illustrates the schematic
diagram and (b) a typical FTIR spectrometer.
a. The Source: Infrared energy is emitted from a glowing black-body source. This
beam passes through an aperture which controls the amount of energy presented to the
sample (and, ultimately, to the detector).
b. The Interferometer: The beam enters the interferometer where the ‘spectral
encoding’ takes place. The resulting interferogram signal then exits the
interferometer.
c. The Sample: The beam enters the sample compartment where it is transmitted
through or reflected off of the surface of the sample, depending on the type of analysis
- 45 -
being accomplished. This is where specific frequencies of energy, which are uniquely
characteristic of the sample, are absorbed.
(a) (b)
Figure 3.2 (a) Schematic diagram of FTIR and (b) FTIR instrument
d. The Detector: The beam finally passes to the detector for final measurement. The
detectors used are specially designed to measure the special interferogram signal.
e. The Computer: The measured signal is digitized and sent to the computer where
the Fourier transformation takes place. The final infrared spectrum is then presented
to the user for interpretation and any further manipulation.
- 46 -
individual compound. IR spectroscopy is mainly used in two ways: structural
elucidation and compound identification.
a. Structural Elucidation
Because of complex interactions of atoms within the molecule, IR absorption
of the functional groups may vary over a wide range. However, it has been found that
many functional groups give characteristic IR absorption at specific, narrow
frequency ranges regardless of their relationship with the rest of the molecule.
Generalized tables of the positions and relative intensities of absorption bands (Table
3.1) have been established and used to determine the functional groups present or
absent in a molecule. Correlation charts are available in many data books [1, 2].
b. Compound Identification
Since the IR spectrum of every molecule is unique, one of the most positive
identification methods of an organic compound is to find a reference IR spectrum that
matches that of the unknown compound.
Transmittance (T) is the ratio of radiant power (I) transmitted by the sample to
the radiant power (I0) incident on the sample. Absorbance (A) is the logarithm to the
base 10 of the reciprocal of the transmittance.
= log
= − log = − log
(3.1)
= (3.2)
(i) Advantages
The FT-IR spectrometer has several major advantages over the dispersive
instrument. Its sensitivity is better because it measures all frequencies simultaneously
rather than scanning through the individual frequencies. Less energy is needed from
the source and less time (typically 1 to 2 seconds) is needed for a scan. Several scans
can be completed in a few seconds and averaged to improve the signal. Resolution
and accuracy are also improved because a laser beam is used alongside the IR beam to
control the speed of the moving mirror and to time the collection of data points. The
light beam is a precise frequency reference that keeps the spectrometer accurately
calibrated.
(ii) Limitations
Few limitations of FTIR spectrometer are listed below:
In DTA, the material under study and an inert reference (which does not go
through phase transition in the temperature range of interest) are heated (or cooled)
under identical conditions, while recording any temperature difference between
sample and reference. This differential temperature is then plotted against time, or
- 49 -
against temperature (DTA curve or thermo gram). Changes in the sample, either
exothermic or endothermic, can be detected relative to the inert reference. Thus, a
DTA curve provides data on the transformations that have occurred, such as glass
transitions, crystallization, melting and sublimation [4]. Differential temperatures can
arise between two inert samples when their response to the applied heat-treatment
may not identical. So DTA can also be used to study thermal properties and phase
changes which do not lead to a change in enthalpy.
3.2.2 Instrumentation
(i) Apparatus
2. Furnace
3. Temperature programmer
4. Recording system
The essential requirements of the furnace are that it should provide a stable
and sufficiently large hot-zone and must be able to respond rapidly to commands from
the temperature programmer. A temperature programmer is essential in order to
obtain constant heating rates. The recording system must have a low inertia to
faithfully reproduce variations in the experimental set up. Figure 3.3 shows the
experimental set up of a DTA cell.
The sample holder assembly consists of a thermocouple each for the sample
and reference, surrounded by a block to ensure an even heat distribution. The sample
is contained in a small crucible designed with an indentation on the base to ensure a
snug fit over the thermocouple bead. The crucible may be made of materials such as
Pyrex, silica, nickel or platinum, depending on the temperature and nature of the tests
involved. The thermocouples should not be placed in direct contact with the sample to
avoid contamination and degradation, although sensitivity may be compromised.
- 50 -
(a) (b)
Figure 3.3 (a) Concept sketch of a DTA cell and (b) DTA instrument
Metallic blocks are less prone to base-line drift when compared with ceramics
which contain porosity. On the other hand, their high thermal conductivity leads to
smaller DTA peaks. The sample assembly is isolated against electrical interference
from the furnace wiring with an earthed sheath, often made of platinum-coated
ceramic material. The sheath can also be used to contain the sample region within a
controlled atmosphere or a vacuum. During experiments at temperatures in the range
200 to 500°C, problems are encountered in transferring heat uniformly away from the
specimen. These may be mitigated by using thermocouples in the form of flat discs to
ensure optimum thermal contact with the now flat- bottomed sample container, made
of aluminium or platinum foil. To ensure reproducibility, it is then necessary to ensure
that the thermocouple and container are consistently located with respect to each
other.
- 51 -
reactions and can lead to large variations between apparently identical samples. In
some circumstances, the rate of heat evolution may be high enough to saturate the
response capability of the measuring system; it is better than to dilute the test sample
with inert material. For the measurement of phase transformation temperatures, it is
advisable to ensure that the peak temperature does not vary with sample size. The
shape of a DTA peak does depend on sample weight and the heating rate used.
Lowering the heating rate is roughly equivalent to reducing the sample weight; both
lead to sharper peaks with improved resolution, although this is only useful if the
signal to noise ratio is not compromised. The influence of heating rate on the peak
shape and disposition can be used to advantage in the study of decomposition
reactions, but for kinetic analysis it is important to minimize thermal gradients by
reducing specimen size or heating rate.
A simple DTA curve may consist of linear portions displaced from the
abscissa because the heat capacities and thermal conductivities of the test and
reference samples are not identical, and of peaks corresponding to the evolution or
absorption of heat following physical or chemical changes in the test sample. The test
results are a graph of the DTA signal (micro-volts) on the Y-axis plotted versus the
sample temperature in °C on the X-axis. The sample graph of enhanced output for
Ceramics-Structe Clay is shown in Figure 3.4.
Thermo Gravimetric analysis (TGA) technique uses heat to force reactions and
physical changes in materials. TGA provides quantitative measurement of mass
change in materials associated with transition and thermal degradation. TGA records
change in mass from dehydration, decomposition, and oxidation of a sample with time
and temperature. Characteristic thermo-gravimetric curves are given for specific
materials and chemical compounds due to unique sequence from physicochemical
reactions occurring over specific temperature ranges and heating rates. These unique
characteristics are related to the molecular structure of the sample [5]. When used in
combination with FTIR, TGA/FTIR is capable of detailed FTIR analysis of evolved
gases produced from the TGA.
3.3.2 Instrumentation
- 53 -
(a) (b)
Figure 3.5 (a) An instrument for TGA and (b) Typical Characteristics curve of TGA
- 54 -
As many weight loss curves look similar, the weight loss curve may require
transformation before results may be interpreted. A derivative weight loss curve can
be used to tell the point at which weight loss is most apparent. Again, interpretation is
limited without further modifications and deconvolution of the overlapping peaks may
be required.
= 2 sin (3.3)
= Angle between incident X-rays and plane of the crystal. The diffracted
beam makes an angle 2θ.
- 55 -
In case of fine particles, with reduction in the size of the particles, the XRD
lines get broadened, which indicates clearly that particle size has been reduced.
Information of the particle size is obtained from the full width at half maximum
(FWHMs) of the diffraction peaks.
After recording the X-ray diffraction pattern, first step involves the indexing
of XRD peaks. The indexing means assigning the correct Miller indices to each peak
of the diffraction pattern.
There are three main methods for indexing of X-ray diffraction pattern:
(i) Comparing the measured XRD pattern with the standard data base (JCPDS-cards)
- 57 -
Figure 3.8 Typical XRD instrument
X-ray diffraction provides us the lattice cell parameters of crystal unit cell and
therefore its volume.
% = / , ()ℎ+,+ = . (3.6)
√& /
%= (3.7)
/
Where a, b, and c are the unit cell dimensions called unit cell or lattice parameters.
- 58 -
b. X-ray density
G HJI .MMN/
F= ≡ Å& (3.9)
E E
c. Porosity
R
P = 1 −
× 100 % (3.10)
S
d. Particle size
When the particle size of the individual crystal is less than 100 Å3, the term
particle size is used. Crystals of this size range cause broadening of the Bragg peak,
the extent of the broadening is given by
.XY
W= (3.11)
Where, B = broadening of the diffraction line (Bragg peak) measured at full width at
half of its maximum intensity (FWHM) (in radian)
λ = wavelength of target
- 59 -
3.5 Scanning Electron Microscope (SEM)
A deflection coil system in front of the last lens scans the electron probe in a
raster across the specimen and in synchronism with the electron beam of a separate
cathode ray tube (CRT). The intensity of the CRT is modulated by one of the signals
- 60 -
recorded to form an image. The magnification can be increased simply by decreasing
the scan-coil current and keeping image size on CRT constant.
Here an electron beam scans the object (the specimen) and because of
synchronized scans of electron beam and the CRT screen (nowadays, monitor), there
is one-to-one correspondence between the spot on the specimen and the spot on the
screen. Unlike Optical microscopy, SEM requires the vacuum environment and
specimen surface to be electrically conductive. Figure 3.9 (a) shows the general
schematic diagram of SEM.
(a) (b)
Figure 3.9 (a) Schematic diagram of Scanning Electron Microscope (SEM) BSE=back
scattered electrons, SE=secondary electrons, SC=specimen current, EBIC=electron-beam-
induced current, X=x-rays, CRT=cathode-ray tube and (b) Typical SEM instrument
- 62 -
elementary atomic interaction processes takes place when electron beam interacts
with the specimen. The incident electrons interact with specimen atoms and are
significantly scattered by them.
Electron beam -specimen interaction gives various signals which can be used
to form images and also can be used for other important information.
Electron images are obtained by rastering the electron beam across the
specimen surface using the deflection coils inside the objective lens and
synchronously rastering the output signal of the detector on a cathode-ray tube (CRT).
The ratio of the area rastered on the specimen to that of the CRT gives the
magnification. For example, a rastered area of 200 µm2 (200 x 10-6 m2) displayed on
CRT with an area of 20 cm (200,000 x 10-6 m2) yields a magnification of 1000x. This
is a very different process than the production of an image by an optical microscope.
Electronic images are sequentially "constructed" during the rastering of the beam,
whereas in optical systems all parts of the sample are imaged simultaneously.
Brightness can be increased /decreased by amplifying /decreasing the signal from the
sample and it can be affected by topography, composition, electrical conductivity, and
other properties of the sample. Contrast reflects the variation in the signal from point
to point. Contrast can also be enhanced by electronically increasing the difference
between small and large signals. Working distance (WD), probe current and probe
diameter, incorrect electron gun alignment, astigmatism, diffused scattering of
electrons from the edges (edge effect), electrostatic charging of the sample, external
disturbances(magnetic fields, flooring, improper grounding) are factors affecting the
quality of the image. The SEM image looks three dimensional Because of SEM's
ability to focus up to certain depth and hence the images can be easily interpreted.
- 63 -
3.5.6 Detectors
There are number of detectors around the specimen in the specimen chamber
which collects different signals for the required information. The most common
imaging mode collects low-energy (< 50 eV) secondary electrons that are ejected
from the k-orbitals of the specimen atoms by inelastic scattering interactions with
beam electrons. Due to their low energy, these electrons originate within a few
nanometers from the sample surface [9]. The electrons are detected by a type of
scintillator-photomultiplier system known as an Everhart-Thornley detector [10]. The
secondary electrons are first collected by attracting them towards an electrically-
biased grid at about + 400 V, then further accelerated towards a phosphor or
scintillator positively biased to about + 2,000 V. The accelerated secondary electrons
are now sufficiently energetic to cause the scintillator to emit flashes of light
(cathode-luminescence) which are conducted to a photomultiplier outside the SEM
column via a light pipe and a window in the wall of the specimen chamber. The
amplified electrical signal output by the photomultiplier is displayed as a two-
dimensional intensity distribution that can be viewed and photographed on an
analogue video display, or subjected to analog-to-digital conversion and displayed
and saved as a digital image. This process relies on a raster-scanned primary beam.
The brightness of the signal depends on the number of secondary electrons reaching
the detector. If the beam enters the sample perpendicular to the surface, then the
activated region is uniform about the axis of the beam and a certain number of
electrons ‘escape’ from within the sample. As the angle of incidence increases, the
‘escape’ distance of one side of the beam will decrease, and more secondary electrons
will be emitted. Thus steep surfaces and edges tend to be brighter than flat surfaces,
which results in images with a well-defined, three-dimensional appearance. Image
resolution less than 0.5 nm is possible using SEM technique.
- 64 -
ions. In a bulk ferrite, there exist domains called “Wiess Domains” in which all these
molecular magnets are aligned in one direction. Domain walls separates different
domains aligned in random directions and in the presence of an external magnetic
field these moments can be forced to align in one direction as shown in Figure 3.10.
Some energy has to be spent for this process and the magnetization always lags
behind the magnetizing field and results in a magnetization loop called as B-H Loop.
(a) (b)
Figure 3.11 (a) Theoretical M-H curve (b) Variation of the coercivity (Hc) of magnetic
nanoparticles with size
- 65 -
Figure 3.11 (a) shows a theoretical magnetization, (M) versus magnetic field,
(H) curve for superparamagnetic (SPM) and ferromagnetic nanoparticles (FM); where
the coercive field (Hc), the saturation magnetization (MS) and the remanent
magnetization (MR) parameters are indicated. Figure 3.11 (b) indicates the relation
between particle size and coercivity.
3.6.2 Principle
(a) (b)
Figure 3.12 (a) Schematic diagram of a VSM instrument and (b) Photograph of a VSM
instrument
- 66 -
The magnetic (dipole) moment is measured in the unit [Am2]. If the sample
weight is known then one can calculate the magnetic moment per unit weight in
[Am2/kg]. If also the density or the volume of the sample is known, it is possible to
calculate the magnetic moment per unit volume which corresponds to the
magnetization in [A/m].
The disc shaped specimens are used for Dielectric measurements. The flat
faces of disc are lapped, a record of the thickness and the diameter is done and then
- 67 -
the parallel faces are painted with silver paste. Silver paste is applied to both the
surfaces (front and back) to provide electrical contact.
The sample is first inserted in the cell and the electrodes closed with the
micrometer until they just touch the sample. The electrodes should not be forced
against the sample. The micrometer is turned with a light finger touch and the
electrometer setting recorded. Experimental set-up is shown schematically in Figure
3.14. The capacitance and the dielectric loss tangent values of the ceramics can be
determined simultaneously. The dielectric constants of samples can be determined
using various equations.
Figure 3.13 Schematic representation of the set-up for dielectric property investigations
Figure 3.14 Precision LCR meter with sample holder and sample pallets
H
Z = [ [′ (3.12)
;
3.8.1 Pelletron
The main building at IUAC houses the Pelletron tower, beam hall and the
laboratory complex. The 15UD Pelletron [14], a versatile heavy ion tandem
electrostatic accelerator is installed in a vertical configuration in an insulating tank of
26.5 meter height and 5.5 meter diameter. The accelerator tank is filled with an
insulating, gas sulphur hexafluoride (SF6) maintained at a pressure of 6-7 atmosphere.
A schematic diagram of 15UD Pelletron is shown in Figure 3.15. In this machine the
negative ions are produced and preaccelerated to ~ 300 keV by Cesium sputter ion
source known as SNICS (Source of Negative Ions by Cesium Sputtering). Now it has
been replaced by name MCSNICS (Multi Cathode SNICS).
The ions are mass analyzed by dipole magnet called injector magnet and are
turned vertically downward direction. The ions then enter the strong electric field
inside the accelerator. A terminal shell is located at the centre of the tank which can
be charged to a high voltage (~ 15 MV) by a pellet charging system. The negative
ions on traversing through the accelerating tubes from the column top of the tank to
the positive terminal get accelerated. On reaching the terminal they pass through a
- 69 -
stripper (foil or gas) that removes some electrons from the negative ions thus
transforming the negative ions into the positive
positive ions. For very heavy ions (i.e.
( atomic
number A >50), the lifetimes of the carbon foils used in stripper are limited to a few
hours due to radiation damage. Therefore a gas filled canal or a combination of the
gas stripper followed by a foil stripper is
is used for heavy ions. The transformed
positive ions are then repelled away from the positively charged terminal and are
accelerated to ground potential to the tank.
In this way, same terminal potential is used twice to accelerate the ions.
Therefore this accelerator is called tandem Pelletron accelerator. The final energy of
the emerging ions from the accelerator is given by
_8 = _R7C`<; + (1 + a8 . % (3.13)
- 70 -
Where, Ei is the energy of the ion having a charge state qi after stripping; V is
the terminal potential in MV and Edecpot is the deck potential of the SNICS source. On
exiting from the tank, the ions are bent into horizontal plane by analyzing magnet.
Figure 3.16 A schematic diagram represents the principle of acceleration of ions in Pelletron
- 71 -
Figure 3.17 Schematic of a Fast closing system
The accelerated beam of desired ions from the Pelletron is brought to the beam
hall and switched to any one of the seven beam lines using the switching magnet. The
Material Science beam line is at 15º to the right with respect to the zero degree beam
line. This beam line houses three chamber-high vacuum chamber, ultra high vacuum
chamber and goniometer chamber. The high vacuum chamber is a cylindrical shaped
multiport stainless steel chamber.
After the observation of the beam on the quartz, the samples to be irradiated
are brought to the same position as that of the quartz by moving the ladder in the
desirable position. A CCD camera is attached to one of the ports of the chamber for
viewing the sample and the quartz position. The positions can be monitored using
close circuit television (CCTV) in the data acquisition room. The magnetic scanner
that can sweep the beam (25 mm in y-direction and 10 mm in x-direction) ensures the
sample ladder, which is kept at a negative potential of 120 V. This enclosure
suppresses the secondary electrons coming out of sample during irradiation. An
opening in the suppressor allows the ion beam to fall on the sample. The total number
of particles/charges falling on the sample can be estimated by a combination of the
current integrator and the pulse counter (Faraday cup) from which the irradiation
fluence can be measured.
The second chamber, UHV chamber contains facilities like in-situ scanning
tunnelling microscopy (STM) for in-situ surface studies and residual gas analyzer
(RGA) for ion beam induced desorption. Third chamber is the goniometer chamber,
which incorporates the in-situ X-ray reflectivity (XRR), large area position sensitive
detector for Elastic recoil detector analysis (ERDA) and other channelling facilities.
- 73 -
irradiated were mounted on the four sides of the target ladder (on copper block),
which were separated from each other by a distance of about 1 cm.
The counts calculated for the fluence for each sample can be calculated by
using the following relation
or
- 74 -
References
[1] D. R. Lide, ed., CRC Handbook of Chemistry and Physics, 75th ed., Boca Raton,
FL: CRC Press, 1994, 9–79.
[12] M. N. Afsar, J. R Birch, R. N. Clarke, IEEE Trans. Microwave Theory Tech, MTT- 25,
1977.
[15] J. F. Ziegler, J. P. Biersach, U. Littmark, The Stopping and Range of Ions in Matter,
Pergamon Press, New York, 2 – 6, 1985.
- 75 -