Comparison Methods of Measurements: UNIT-3
Comparison Methods of Measurements: UNIT-3
D.C potentiometers, D.C (Wheat stone, Kelvin and Kelvin Double bridge) & A.C bridges (Maxwell,
Anderson and Schering bridges), transformer ratio bridges, self-balancing bridges. Interference &
screening – Multiple earth and earth loops – Electrostatic and electromagnetic Interference –
Grounding techniques.
UNIT-3
COMPARISON METHODS OF
MEASUREMENTS
DC Potentiometer
An instrument that precisely measures an electromotive force (emf) or a voltage by
opposing to it a known potential drop established by passing a definite current through a resistor
of known characteristics. (A three-terminal resistive voltage divider is sometimes also
called a potentiometer.) There are two ways of accomplishing this balance: the current I may be
held at a fixed value and the resistance R across which the IR drop is opposed to the unknown may
be varied, current may be varied across a fixed resistance to achieve the needed IR drop.
The essential features of a general-purpose constant-current instrument are shown in
the illustration. The value of the current is first fixed to match an IR drop to the emf of a reference
standard cell. With the standard-cell dial set to read the emf of the reference cell, and the
galvanometer (balance detector) in position G1, the resistance of the supply branch of the circuit
is adjusted until the IR drop in 10 steps of the coarse dial plus the set portion of the standard-cell
dial balances the known reference emf, indicated by a null reading of the galvanometer. This
adjustment permits the potentiometer to be read directly in volts. Then, with the galvanometer in
Position G2, the coarse, intermediate, and slide-wire dials are adjusted until the galvanometer
again reads null. If the potentiometer current has not changed, the emf of the unknown can be
read directly from the dial settings.
There is usually a switching arrangement so that the galvanometer can be quickly
shifted between positions 1 and 2 to check that the current.
terminals of the potentiometer. Here, of course, internal heating and consequent resistance change
of the current-carrying resistor (shunt) may be a critical factor in measurement accuracy; and the
shunt design may require attention to dissipation of heat resulting from its I2R power consumption.
Potentiometer techniques have been extended to alternating-voltage measurements,
but generally at a reduced accuracy level (usually 0.1% or so). Current is set on an ammeter
which must have the same response on ac as on dc, where it may be calibrated with a potentiometer
and shunt combination. Balance in opposing an unknown voltage is achieved in one of two ways:
a slide-wire and phase-adjustable supply; separate in-phase and quadrature adjustments on
Slide wires supplied from sources that have a 90° phase difference. Such potentiometers have
limited use in magnetic testing.
An instrument that precisely measures an electromotive force (emf) or a voltage by
opposing to it a known potential drop established by passing a definite current through a resistor
of known characteristics. (A three-terminal resistive voltage divider is sometimes also
called a potentiometer.) There are two ways of accomplishing this balance the current I may be
held at a fixed value and the resistance R across which the IR drop is opposed to the unknown may
be varied; current may be varied across a fixed resistance to achieve the needed IR drop.
The essential features of a general-purpose constant-current instrument are shown in the
illustration.
The value of the current is first fixed to match an IR drop to the emf of a reference
standard cell. With the standard-cell dial set to read the emf of the reference cell, and the
galvanometer (balance detector) in position G1, the resistance of the supply branch of the circuit
is adjusted until the IR drop in 10 steps of the coarse dial plus the set portion of the standard-cell
dial balances the known reference emf, indicated by a null reading of the galvanometer. This
adjustment permits the potentiometer to be read directly in volts. Then, with the galvanometer in
position G2, the coarse, intermediate, and slide-wire dials are adjusted until the galvanometer
again reads null. If the potentiometer current has not changed, the emf of the unknown can be
read directly from the dial settings. There is usually a switching arrangement so that the
galvanometer can be quickly shifted between positions 1 and 2 to check that the current has not
drifted from its set value.
Potentiometer techniques may also be used for current measurement, the unknown current
being sent through a known resistance and the IR drop opposed by balancing it at the voltage
terminals of the potentiometer. Here, of course, internal heating and consequent resistance change
of the current-carrying resistor (shunt) may be a critical factor in measurement accuracy; and the
shunt design may require attention to dissipation of heat resulting from its I2R power consumption.
Potentiometer techniques have been extended to alternating-voltage measurements, but generally
at a reduced accuracy level (usually 0.1% or so). Current is set on an ammeter which must have
the same response on ac as on dc, where it may be calibrated with a potentiometer and shunt
combination.
Balance in opposing an unknown voltage is achieved in one of two ways: a slide-wire
and phase-adjustable supply; separate in-phase and quadrature adjustments on slide wires
supplied from sources that have a 90° phase difference. Such potentiometers have limited use in
magnetic testing an electrical measuring device used in determining the electromotive force (emf)
or voltage by means of the compensation method. When used with calibrated standard
resistors, a potentiometer can be employed to measure current, power, and other electrical
quantities; when used with the appropriate measuring transducer, it can be used to gauge various
nonelectrical quantities, such as temperature, pressure, and the composition of gases.
A distinction is made between DC and AC potentiometers. In DC potentiometers, the
voltage being measured is compared to the emf of a standard cell. Since at the instant of
compensation the current in the circuit of the voltage being measured equals zero, measurements
can be made without reductions in this voltage. For this type of potentiometer, accuracy can
exceed 0.01 percent. DC potentiometers are categorized as either high-resistance, with a slide-wire
resistance ranging from 104 to 105 ohms (Ω) and a current ranging from 10-1 to 10-9 amperes (A),
or low-resistance, with a slide-wire resistance below 2 × 103 ohms and a current ranging from 10-1 to
10-3 A.
The higher resistance class can measure up to 2 volts (V) and is used in testing highly
accurate apparatus. The low-resistance class is used in measuring voltage up to 100 mV. To
measure higher voltages, up to 600 V, and to test voltmeters, voltage dividers are connected to
potentiometers. Here the voltage drop across one of the resistances of the voltage divider is
compensated; this constitutes a known fraction of the total voltage being measured.
BRIDGES
Bridge circuits are mainly used to measure unknown quantities such as resistance,
inductance, capacitance, Impedance and admittance. Bridge circuit consists of 4 resistance arms
forming a closed circuit with dc source of current applied to two opposite junctions and a current
detector connected to the other two junctions.
Types of bridges
A.C Bridges
D.C Bridges
DC Bridges
Wheatstone bridge
A Wheatstone bridge is an electrical circuit invented by Samuel Hunter Christie in 1833
and improved and popularized by Sir Charles Wheatstone in 1843. It is used to measure an
unknown electrical resistance by balancing two legs of a bridge circuit, one leg of which includes
the unknown component. Its operation is similar to the original potentiometer.
Operation
resistance is related to measuring the impact of some physical phenomenon - such as force,
temperature, pressure, etc. which thereby allows the use of Wheatstone bridge in measuring
those elements indirectly.
KELVIN BRIDGE
A Kelvin bridge (also called a Kelvin double bridge and some countries Thomson bridge)
is a measuring instrument invented by William Thomson, 1st Baron Kelvin. It is used to measure
an unknown electrical resistance below 1 Ω. Its operation is similar to the Wheatstone
bridge except for presence of additional resistors. These additional low value resistors and the
internal configuration of the bridge are arranged to substantially reduce measurement errors
introduced by voltage drops in the high current (low resistance) arm of the bridge
Accuracy
There are some commercial the devices reaching accuracies of 2% for resistance
ranges from 0.000001 to 25 Ω. Often, ohmmeters include Kelvin bridges, amongst other
measuring instruments, in order to obtain large measure ranges, for example, the Valhalla 4100
ATC Low-Range Ohmmeter.
The instruments for measuring sub-ohm values are often referred to as low-
resistance ohmmeters, milli-ohmmeters, micro-ohmmeters, etc
Principle of operation
The measurement is made by adjusting some resistors in the bridge, and the balance
is achieved when: in 1865 and further improved by Alan Bulletin in about 1926.
Resistance R should be as low as possible (much lower than the measured value)
and for that reason is usually made as a short thick rod of solid copper.
If the condition R3R4’ =R3’R 4 met (and value of R is low), then the last component
in the equation can be neglected and it can be assumed that:
In the Schering Bridge above, the resistance values of resistors R1 and R2 are known,
while the resistance value of resistor R3 is unknown. The capacitance values of C1 and C2
are also known, while the capacitance of C3 is the value being measured. To measure R3 and
C3, the values of C2 and R2 are fixed, while the values of R1 and C1 are adjusted until the
current through the ammeter between points A and B becomes zero. This happens when
the voltages at points A and B are equal, in which case the bridge is said to
be 'balanced'.
When the bridge is balanced, Z1/C2 = R2/Z3, where Z1 is the impedance of R1 in
parallel with C1 and Z3 is the impedance of R3 in series with C3. In an AC circuit that has
a capacitor, the capacitor contributes a capacitive reactance to the impedance. The
capacitive reactance of a capacitor C is 1/2πfC.
As such, Z1 = R1/[2πfC1((1/2πfC1) + R1)] = R1/(1 + 2πfC1R1) while Z3 = 1/2πfC3 + R3.
Prepared by :U.NAGABALAN., AP/EEE ., M.VENKATESWARAN., AP/EEE ., R.VANITHA., AP/EEE
EE8403 UNIT-3 COMPARATIVE METHODS OF MEASUREMENTS
Maxwell's Bridge
second inductor to react with in the Maxwell bridge, this problem is eliminated.
Anderson Bridge
It is a modified version of Maxwell’s inductance capacitance bridge. In this method, the
self inductance is measured in terms of a standard capacitance. It is applicable for precise
measurement of self-inductance over a wide range of values.
r,𝑅 2 ,𝑅3 ,𝑅4 be the known non-inductive resistances,and C be the fixed standard
capacitance
At balance
I1 = I3 and I2 = IC
1
𝐼1 𝑅3 = 𝐼𝐶 × 𝑗𝜔𝑐
𝐼𝐶 = 𝐼1 j𝜔c𝑅3
𝐼1 (𝑟1 + 𝑅𝑋 + 𝑗𝜔𝐿𝑋 ) = 𝐼2 𝑅2 + 𝐼𝑒
𝑅2𝑅3
From the above equations,𝑅𝑋 = −𝑟1
𝑅4
𝑅
𝐿𝑋 = C𝑅3 [𝑟(𝑅4 + 𝑅2 ) + 𝑅2 𝑅4 ]
4
Advantages
A fixed capacitor can be used instead of a variable capacitors in the case of
Maxwell’s bridge.
This bridge may be used for accurate determination of capacitance in terms of
inductance.
It is much easier to obtain balance in case of Anderson’s bridge than in Maxwell’s
bridge for low Q coils.
Disadvantages
The Anderson’s bridge is more complex than its prototype Maxwell’s bridge.
An additional junction point increases the difficulty of shielding the bridge.
Different values of V1 and V2 may be had by changing the sliding contact on the tapings.
The construction of ideal transform we is impossible. But the ideals of zero winding
resistance, zero core loss and perfect coupling can be closely achieved by using special design
similar to those for instrument transformers.
Measurement of resistance
In the unknown resistance R is in comparison with a standard resistance R s. The position
of the wiper is adjusted till the detector D shows null position. Current through the unknown
resistance
I1=V1/R= K1 N1/R
Current through the standard resistance I 2=V2/Rs= K1 N2/Rs
Under balance conditions, the current through the detector D is zero ieI 1=I2
Hence K1 N1/R = K1 N2/Rs
Or R= N1/ N2 Rs
Measurement of Capacitance
The measurement of capacitance by using ratio transformer. Here, unknown capacitance
C is measured in comparison with a standard capacitance Cs. A resistance R is connected across
the unknown capacitor C. This resistance represents the loss of the capacitor. Under balance
position, the magnitude and phase of the currents passing through detector should be the same; a
variable standard resistance is connected in parallel with the standard capacitor.
At balance position,
C = N2/ N1 Cs
R = N1/ N2 Rs
Dissipation factor D
D = 1/ωCR=1/ ωCsRs
Measurement of phase angle
Figure 3.11 shows the measurement of phase angle by using ratio transformer.
amplifier. Bolometer measurements are based on the dissipation of the RF power in a small
temperature sensitive resistive element, called a Bolometer.
remedies only aim at reducing the effect of interference to such a degree, that it is neither audible
nor disturbing. This is mainly achieved by different ways of screening.
This paper will explain the technical background of interference and provides some
common rules and hints which may help you to reduce the problems.
Types of Interference
Theoretically, the effects and mechanism of a single interference can well be calculated.
But in practice, the complex coupling systems between pieces of equipment prevent precise
prediction of interference. The following picture shows the different types of interference
coupling.
If we consider all possible coupling paths in the diagram above we will find 10 different
paths. This means a variety of 1024 different combinations. It should be noted, that not only the
number of paths, but also their intensity is important.
Symmetrical and Asymmetrical Interference
Having a closer look at the interference of cable, we find that high frequency-interference
currents cause measurable levels on signal (audio) lines and on supply lines.
Through interference, asymmetrical signals are produced in respect to the ground. The
asymmetrical interference current flows along the two wires of the symmetrical line to the sink
and via the ground back to the source. These interference signals are cancelled at the
symmetrical input.
which means voltages of some 10kV in respect to ground. Dry air, artificial fabrics and friction
favour these conditions. When touching grounded equipment, an instantaneous discharge produces
arcing with short, high current pulses and associated strong changes of the e.m. field.
Screening
When considering the effect of electrical and magnetic fields, we have to distinguish
between low and high frequencies. At high frequencies the skin effect plays an important roll for
the screening. The penetration describes the depth from the surface of the conductor, where the
current density has decayed to 37% compared to the surface of the conductor.
The interference and never fully prevent it. This means there will never be a system which
is 100% safe from interference. Because the efforts and the cost will rise with the degree of
reduction of interference, a compromise has to be found between the effort and the result. The
requirement for the reduction of interference will depend on:
A lapped screen with wires laid parallel to the central signal conductor. The screening
converge is not perfect, and can be badly degraded as it tends to open up on the outside of
cable bends.
A braided screen around the central signal wires. This is more expensive, but opens up
less on bends. Screening is not 100%, but certainly better than screen.
An overlapping foil screen, with the ground wire running down the inside of the foil and
in electrical contact with it. This is usually the most effective as the foil cannot open up
on the outside of bends, and should give perfect electrostatic screening, However, the
higher resistance of aluminium foil compared with Copper braid means that RF screening
may be worse.
Advantages of Balancing
It discriminates against noise and crosstalk.
Balanced interconnect aloes 6 dB more signal level on the line.
Electrical Noise
Noise gets into signal cables in three major ways:
Electrostatic coupling
An interfering signal with significant voltage amplitude couples directly to the inner
signal line, through stray capacitance.
In critical situations the equality of these voltage is maximised by minimising the loop area
between the two signal wires, usually by twisting them tightly together. In practice most audio
Prepared by :U.NAGABALAN., AP/EEE ., M.VENKATESWARAN., AP/EEE ., R.VANITHA., AP/EEE
EE8403 UNIT-3 COMPARATIVE METHODS OF MEASUREMENTS
cables have parallel rather than twisted signal conductors, and this seems adequate most of the
time. Magnetic coupling falls off with the square of distance, so rearranging the cable-run away
from the source of magnetic field is usually all that is required. It is unusual for it to present
serious difficulties in a domestic environment.
Ground voltages coupled in through the common ground impedance
This is the root of most ground loop problems. In the equipment safety ground
causes a loop ABCD; the more existence of a loop in itself does no harm, but it is
invariably immersed in a 50 Hz magnetic field that will induce mains frequency current
plus odd harmonics into it.
This current produces a voltage drop down the non- negligible ground-wire
resistance, and this once again effectively appears as a voltage source in each of the to
signal lines. Since the CMRR is finite , a proportion of this voltage will appear to be
differential signal, and will be reproduced as such.
GROUNDING TECHNIQUES
Grounding (or)Earthling
This is one of the simplest but most efficient methods to reduce interference.
Grounding can be used for three different purposes:
Protection Ground
Provides protection for the operators from dangerous voltages. Widely used on
mains-operated equipment.
Function Ground
The ground is used as a conductive path for signals.
Example: in asymmetrical cables screen, which is one conductor for the signal, is
connected to the ground.
Screening Ground
Used to provide a neutral electrical path for the interference, to prevent that the
interfering voltages or currents from entering the circuit.
In this chapter we will only consider the third aspect. Grounding of equipment is often
required for the cases 1 or 2 anyhow, so that the screening ground is available "free
of charge".
Sometimes the grounding potential, provided by the mains connection, is very
"polluted". This means that the ground potential itself already carries an interfering signal.
Prepared by :U.NAGABALAN., AP/EEE ., M.VENKATESWARAN., AP/EEE ., R.VANITHA., AP/EEE
EE8403 UNIT-3 COMPARATIVE METHODS OF MEASUREMENTS
This is especially likely if there are big power consumers in the neighbourhood or even in the
same building. Using such a ground might do more harm than good. The quality of the ground
line can be tested by measuring it with a storage scope against some other ground connection,
e.g. a metal water pipe or some metal parts of the construction. Never use the Neutral (N) of
the mains as ground. It might contain strong interference, Because it carries the load current of
all electrical consumers.
The grounding can be done by single-point grounding or by multi-point grounding. Each
method has advantages which depend on the frequency range of the signal frequencies. All parts
to be grounded are connected to one central point. This results in no "ground loops" being
produced. This means the grounding conductors do not form any closed conductive path in
which magnetic interference could induce currents. Furthermore, conductive lines between
the equipment are avoided, which could produce galvanic coupling of interference. Central
grounding requires consistent arrangement of the grounding circuit and requires
insulation of the individual parts of the circuit. This is sometimes very difficult to achieve. A
system using the single-point grounding.
Multi-Point Grounding
In multi-point grounding all parts are connected to ground at as many points as possible.
This requires that the ground potential itself is as widely spread as possible.
In practice, all conductive parts of the chassis, the cases, the shielding, the room and the
installation are included in the network. The interconnection of these parts should be done at as
many point possible.
Screening
When considering the effect of electrical and magnetic fields, we have to distinguish
between low and high frequencies. At high frequencies the skin effect plays an important roll for
the screening. The penetration describes the depth from the surface of the conductor, where the
current density has decayed to 37% compared to the surface of the conductor. the interference and
never fully prevent it. This means there will never be a system which is 100% safe from
interference. Because the efforts and the cost will rise with the degree of reduction of
interference, a compromise has to be found between the effort and the result.
The requirement for the reduction of interference will depend on:
the strength of the interference source
Prepared by :U.NAGABALAN., AP/EEE ., M.VENKATESWARAN., AP/EEE ., R.VANITHA., AP/EEE
EE8403 UNIT-3 COMPARATIVE METHODS OF MEASUREMENTS
Earthling
Proper earthling bus bar has been provided.
Handling
Lifting lugs has been provided on top of four corners of the housing for lifting the DPS
by crane and chains as a single unit, otherwise this can be done by a forklift of sufficient
capacity, but the lifting fork must be positioned under the transformer portion. Schering Bridge is
independent of frequency.
PART – B
1. a) Explain in detail about the laboratory type DC potentiometer.( 1 2 )
b) Give the applications of AC potentiometers. (4)
2. a)Describe about the multiple earth and earth loops.(8)
b) Explain the different techniques of grounding. (8)
3. Explain voltage sensitive self balancing bridge, and derive the bridge sensitivity of
voltage sensitive bridge with fundamentals.( 1 6 )
4. a) With fundamentals distinguish between DC and AC Potentiometers, and give any
two specific applications for each.(8)
b) Discuss the advantages and limitations of electromagnetic Interference in measurements.(8)
5. a) Explain Kelvin’s double bridge method for the measurement of low resistance.(8)
b) Explain how inductance in measured by using Maxwell’s bridge. (8)
6. a)Explain the working principle of Anderson’s bridge and also derive its balance
equations. (8)
b) Explain the working principle of Schering bridge and also derive its balance equations. (8)