Minimally Invasive Therapies Group System Risk Management PB500 Series Product Risk Assessment / Risk Analysis Chart - General Information
Minimally Invasive Therapies Group System Risk Management PB500 Series Product Risk Assessment / Risk Analysis Chart - General Information
PB500 Series Product Risk Assessment / Risk Analysis Chart - General Information
This document covers the patient, all intended users, and the device itself during manufacture, use, servicing, transportation, and storage of the medical device.
2 Assumptions:
This hazard analysis assumes that the PB560, PB540, PB520 and PowerPack shall only be operated by qualified, properly trained personnel (i.e. used by prescription
only). The users of thePB560, PB540, PB520 and PowerPack include: clinicians, physicians, nurses, homecare providers, patients and patient’s families.
Per the Risk Management Plan, 10035479, risk management activities are retrospective to comply with EN ISO 14971:2012, and Remediation Plan for Risk
Management Documents 10130032, and shall be carried out in accordance with MITG Risk Management Method GS10004411. Because this is retrospective, the
columns in the Product Risk Analysis table below that call for the 'Pre-mitigation risk assessment' will not be populated. The 'Post mitigation' (risidual) risk assessment
columns will be the only ones used because the mitigations have already been applied. The PB500 Risk Assessment Team considered the probability of both P1 and
P2 occurring together without defining each other separately. The residual probability of harm was arrived by combining engineering design expertise with data from a
3 years period of PB500 customer complaints.
Within this file cells that have been shaded brown represent the start of the hazard. All causes, harms and mitigations listed below this line are appropriate for the
hazard. P1, P2 and risk zones are correct for the row associated with the 'harm x severity' found on the same row.
Grey shading has been used to block out unused cells and are considered equivalent to 'not applicable' for documentation purposes.
3.2 The intended use environments for the above personnel include:
The PB500 ventilator is intended for use in institutional, home, and portable settings. The ventilator is not for use as an emergency transport ventilator.
The PB500 series ventilators are indicated for the continuous or intermittent mechanical ventilatory support of patients weighing at least 11 lbs (5 kg)
who require mechanical ventilation delivered invasively or non-invasively.
4 System Definition/Exclusions:
N/A
6 Standard Revisions:
The standards referenced in the Risk Anaysis Chart correspond with the following standard revisions:
Medical Electrical Equipment - Part 1: General Requirements for Safety - Collateral Standard - Safety Requirements for Medical
EN 60601-1:2006 + A12:2014 Electrical Systems
EN ISO 10993-1:2009 Biological Evaluation of medical devices Part 1: Evaluation and testing within a risk management process
Lung Ventilators for Medical Use- Particular Requirements for Basic Safety and Essential Performance Part 2: Home Care
EN 10651-2:2009 Ventilators for Ventilator-Dependent Patient
IEC 60601-1-11:2015 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance – Collateral Standard:
IEC 60601-1-11:2015 Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
EN 60601-1-6:2010 + A1:2014
IEC 60601-1-6: 2010 +A1:2013 Medical Electrical Equipment - Part 1: General Requirements for Safety -2- Collateral standard: Usability
EN 60601-1-8:2007 + A11:2017
IEC 60601-1-8:2006 + A1:2012 General
Medical Requirements, tests and
electrical equipment Partguidance for alarm
2-55: Particular systems in medical
requirements for basicelectrical
safety andequipment
essentialand medical electrical
performance systems
of respiratory gas
EN ISO 80601-2-55:2011 monitor
Medical Electrical Equipment - Part 2-72:Particular requirements for basic safety and essential performance of home healthcare
EN ISO 80601-2-72:2015 environments ventilator-dependent patients
EN 60601-1-2:2015 Medical Electrical Equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard:
IEC 60601-1-2:2014 Electromagnetic Disturbances - requirements and tests
GS10004411, Attachment D
10037183 Rev W Revision 202 Page 2 of 14
Minimally Invasive Therapies Group System Risk Management
PB500 Series Product Risk Assessment / Risk Analysis Chart - Annex C Questions
Response Potential
Question Factors to Consider Potential Hazards(s) Notes / Comments
(N/A = Not Relevant) Hazardous situation(s)
What is the intended use and how What is the medical device's role The Puritan Bennett™ 560 Ventilator is Refer to RE00076775 N/A N/A
is the medical device to be used? relative to: indicated for the continuous or Ventilation Hazards
• diagnosis, prevention, intermittent mechanical
monitoring, treatment, or ventilatory support of patients weighing
alleviation of disease, at least 11 lb (5 kg) who require
mechanical ventilation.
The ventilator is a restricted medical
• compensation for injury or
device intended for use by qualified,
handicap or
trained personnel under the direction of
a doctor.
of ventilation
• what are the indications for use
• Breath types including Volume Control,
(e.g., patient population)?
Pressure Control, and Pressure Support
Is the medical device intended to Consider: the location of N/A Refer to RE00076775 N/A N/A
be implanted? implantation, the characteristics Ventilation Hazards
of the patient population, age,
weight, physical activity, the
C.2.2
GS10004411, Attachment D
10037183 Rev W Revision 202 Page 3 of 14
Ventilation Hazards
Is the medical device intended to Nature of the intended contact, N/A N/A N/A
be in contact with the patient or i.e. surface contact,
other persons? invasive contact, or implantation None of the parts of the ventilator is in
and, for each, the period and direct patient contact. A breathing circuit
frequency of contact. is connected to the ventilator. The
• surface contact points; breathing circuit does not come into
• invasive contact / contact contact with the patient directly. The
C.2.3
What materials or components Consider: Clinicians may use a variety of N/A N/A
are utilized in the medical device • compatibility with relevant interfaces to connect patients to the
or are used with, or are in contact substances; ventilator: nasal masks or full face
with, the medical device? • compatibility with tissues or masks; endotracheal or tracheotomy
body fluids; tubes.
• whether characteristics relevant
to safety are known;
• is the device manufactured
utilizing materials of animal
C.2.4
Are substances delivered to or Consider: The ventilator delivers regulated air and/or
Refer to RE00076775 Ventilation Hazards N/A N/A
extracted from the patient? • whether the substance is delivered oxygen to the patient via a medical pipeline
or extracted; system or from gas cylinders. It is supplied
to the patient circuit through check valves
• whether it is a single substance or and exits to the patient circuit close to
range of substances; patient pressure.
C.2.6
• the maximum and minimum transfer This ventilator is not for use with anesthetic
rates and control thereof.
gases.
GS10004411, Attachment D
10037183 Rev W Revision 202 Page 4 of 14
Minimally Invasive Therapies Group System Risk Management
PB500 Series Product Risk Assessment / Risk Analysis Chart - Annex C Questions
• Simplicity of disassembling /
reassembling and any USE errors
connected to these?
Is the medical device intended to Consider: All external panels and surfaces should be Refer to RE00076775 N/A N/A
be routinely cleaned and • the types of cleaning or disinfecting cleaned before and after each patient use Ventilation Hazards
disinfected by the user? agents to be used and any limitations and as often as necessary to keep the
on the number of cleaning cycles. ventilator clean. Additionally the ventilator
should be cleaned the ventilator
periodically, whenever it is soiled or dirty,
before any maintenance operation, and
before storing the ventilator.
Approved Cleaning Solutions for Exterior
Ventilator Surfaces:
Mild dishwashing detergent
C.2.9
Is the medical device intended to Consider: The ventilator modifies the patient's Refer to RE00076775 N/A N/A
modify the patient environment? • temperature; breathing environment. User-selectable Ventilation Hazards
• humidity; ventilation modes are:
• atmospheric gas composition; • Assisted Controlled Volume (V A/C)
• pressure; • Assisted Controlled Pressure (P A/C)
• light. • Volume Synchronized Intermittent
Mandatory Ventilation (V SIMV)
• Pressure Synchronized Intermittent
C.2.10
Are there unwanted outputs of • Energy-related Consider: noise and There are no unwanted outputs of energy Refer to RE00076775 N/A N/A
energy or substances? vibration, heat, radiation (including or substances under normal operating Ventilation Hazards
ionizing, non-ionizing, and conditions.
ultraviolet /visible/infrared radiation),
contact temperatures, leakage
currents, and electric or magnetic
fields.
C.2.14
GS10004411, Attachment D
10037183 Rev W Revision 202 Page 6 of 14
Are there unwanted outputs of There are no unwanted outputs of energy Refer to RE00076775 N/A N/A
energy or substances? Minimally
or substances Invasive
under normal operatingTherapies Group
Ventilation System Risk Management
Hazards
PB500 Series Product Risk Assessment / Risk Analysis Chart - Annex C Questions
conditions.
• Substance-related Consider:
substances used in manufacturing,
cleaning or testing having unwanted
physiological effects if they remain in
the product.
Is the medical device susceptible Consider: the operational, transport Yes – temperature, pressure, humidity, Refer to RE00076775
to environmental influences? and storage environments. These vibration, electrical/magnetic field. Ventilation Hazards
include light, temperature, humidity,
vibrations, spillage, susceptibility to
C.2.15
Does the medical device Consider:: Yes – alarm noises/notifications. Refer to RE00076775 N/A N/A
influence the environment? • the effects on power and cooling Ventilation Hazards
supplies;
C.2.16
GS10004411, Attachment D
10037183 Rev W Revision 202 Page 7 of 14
Minimally Invasive Therapies Group System Risk Management
PB500 Series Product Risk Assessment / Risk Analysis Chart - Annex C Questions
Are there essential consumables Consider: specifications for such Consumables and Replacement Intervals Refer to RE00076775 N/A N/A
or accessories associated with consumables or accessories and any Ventilation Hazards
the medical device? restrictions placed upon users in their • Air Inlet Filter - (Foam + Fine Particle)
selection of these. Once a month or more often, depending on
the extent of soiling
• Inspirited Bacteria Filter See
manufacturer’s recommendation
• Patient Circuit - See manufacturer’s
recommendation - Single use Single patient
• FIO2 Sensor - 14 to 18 months or more
often in case of persistent calibration failure
• Exhalation Block 4 months(*) (and for
each new patient) humidifiers, water traps
• Oxygen solenoid valve - Every 15,000
hours of use
•Turbine - Every 15,000 hours of use
•Exhalation solenoid valve - Every 15,000
hours of use
•Cooling fan - Every 15,000 hours of use
C.2.17
Is maintenance or calibration Consider: Each time the exhalation block or circuit is Refer to RE00076775 N/A N/A
necessary? • whether maintenance or calibration removed and reinstalled or after installing a Ventilation Hazards
are to be carried out by the operator new exhalation block, the exhalation flow
or user or by a specialist; sensor must be recalibrated before the
using the ventilator. This process is
automatic and does not require the use of a
measurement device.
Each time the FIO2 sensor is removed and
C.2.18
Does the medical device contain Consider: whether software is Software is installed by a qualified service N/A N/A N/A
software? intended to be installed, verified, engineer
C.2.19
GS10004411, Attachment D
10037183 Rev W Revision 202 Page 8 of 14
Minimally Invasive Therapies Group System Risk Management
PB500 Series Product Risk Assessment / Risk Analysis Chart - Annex C Questions
Does the medical device have a Consider: labeling or indicators and N/A N/A N/A N/A
C.2.20
Are there any delayed or long- Consider: ergonomic and cumulative The patient can become dependent on the Refer to RE00076775 N/A N/A
term use effects? effects. ventilator to provide respiration. Ventilation Hazards
Examples could include pumps for
saline that corrode over time, The ventilator will remain in good service
mechanical fatigue, loosening of straps condition when the preventative
and attachments, vibration effects, maintenance schedule is adhered to
C.2.21
To what mechanical forces will Consider: whether the forces to which Shipping – ISTA 2A; shock and vibration Refer to RE00076775 N/A N/A
the medical device be subjected? the medical device will be subjected The ventilator is suitable for use in Ventilation Hazards
are under the control of the user or institutional, home, and portable settings. It
controlled by interaction with other is not intended for use as an emergency
C.2.22
What determines the lifetime of Consider: aging and battery depletion. The lifetime of the device is determined by Refer to RE00076775 N/A N/A
the medical device? a number of factors: Ventilation Hazards
• Life of product line is from inception to
withdraw from market
C.2.23
Is the medical device intended for Consider: does the medical device self- No N/A N/A N/A
C.2.24
Is safe decommissioning or Consider: the waste products that are The ventilator is marked to ensure that it Refer to RE00076775 N/A N/A
disposal of the medical device generated during the disposal of the enters the recycled waste system. 'Recycle Ventilation Hazards
necessary? medical device itself. For example, electrical and electronic items with this
does it contain toxic or hazardous marking according to local regulations
C.2.25
GS10004411, Attachment D
10037183 Rev W Revision 202 Page 9 of 14
Minimally Invasive Therapies Group System Risk Management
PB500 Series Product Risk Assessment / Risk Analysis Chart - Annex C Questions
Does installation or use of the Consider: the novelty of the medical The ventilator is for use as prescribed by a Refer to RE00076775 N/A N/A
medical device require special device and the likely skill and training doctor. Ventilation Hazards
training or special skills? of the person installing the device.
The ventilator may be operated by:
• Respiratory therapists
• Doctors
• Nurses
• Homecare providers
• Patient and patient’s families
C.2.26
How will information for safe use Consider: Training and manuals Refer to RE00076775 N/A N/A
be provided? • whether information will be Ventilation Hazards
provided directly to the end user by
the manufacturer or will it involve the
participation of third parties such as
installers, care providers, health care
professionals, or pharmacists, and
whether this will have implications for
training;
C.2.27
Will new manufacturing Consider: new technology or a new N/A N/A N/A N/A
C.2.28
dependent on human factors operation, menus for software driven devices, visibility of warnings, and audibility of alarms, standardized color coding.
such as the user interface?
GS10004411, Attachment D
10037183 Rev W Revision 202 Page 10 of 14
Minimally Invasive Therapies Group System Risk Management
PB500 Series Product Risk Assessment / Risk Analysis Chart - Annex C Questions
Can the user interface design Factors that should be considered are Yes Refer to RE00076775 N/A N/A
features contribute to use error? user interface design features that can Ventilation Hazards
contribute to use error.
Examples of interface design features
include: control and indicators,
symbols used, ergonomic features,
physical design and layout, hierarchy
of operation, menus for software
C.2.29.1
Is the medical device used in an Consider: Yes Refer to RE00076775 N/A N/A
environment where distractions • the consequence of use error; Ventilation Hazards
can cause use error?
C.2.29.2
Does the medical device have a Consider: Yes Refer to RE00076775 N/A N/A
control interface? spacing, coding, grouping, mapping, Ventilation Hazards
modes of feedback,
blunders, slips, control differentiation,
visibility, direction of activation or
C.2.29.4
Does the medical device display Consider: Yes Refer to RE00076775 N/A N/A
information? visibility in various environments, Ventilation Hazards
orientation, the visual capabilities of
the user, populations and
C.2.29.5
GS10004411, Attachment D
10037183 Rev W Revision 202 Page 11 of 14
Minimally Invasive Therapies Group System Risk Management
PB500 Series Product Risk Assessment / Risk Analysis Chart - Annex C Questions
Is the medical device controlled Consider: complexity and number of Yes Refer to RE00076775 N/A N/A
by a menu? layers, awareness of state, location of Ventilation Hazards
settings, navigation method, number
of steps per action, sequence clarity
and memorization problems, and
C.2.29.6
Will the medical device be used Consider: the user, their mental and Only if the user is differently abled Refer to RE00076775 N/A N/A
by persons with special needs? physical abilities, skill, and training, Ventilation Hazards
ergonomic aspects, the use
environment, installation
requirements, and the patient's
capability to control
or influence the use of the medical
device. Special attention should be
paid to users with special needs, such
as handicapped persons, the elderly,
and children. Their special needs might
C.2.29.7
Can the user interface be used to Consider: the possibility of initiating a yes N/A N/A N/A
initiate user actions? deliberate action for the user to enter
a controlled operation mode, which
C.2.29.8
Does the medical device use an Consider: Yes N/A N/A N/A
alarm system? are the risk of false alarms, missing
alarms, disconnected alarm systems,
unreliable remote alarm systems, and
the medical staff’s possibility of
understanding how the
C.2.30
GS10004411, Attachment D
10037183 Rev W Revision 202 Page 12 of 14
Minimally Invasive Therapies Group System Risk Management
PB500 Series Product Risk Assessment / Risk Analysis Chart - Annex C Questions
In what way(s) might the medical Consider: are incorrect use of User error inputting the prescribed settings Refer to RE00076775 N/A N/A
device be deliberately misused? connectors, disabling safety features User Error - ventilator settings Ventilation Hazards
or alarms, inappropriate for patient
neglect of manufacturer's User Error - not adhering to the PM
recommended maintenance. schedule
User Error - O2 not connected or
obstructed
User Error - cables not connected correctly
User Error - Physical damage from impact
or vibration, or transport bag fault
User Error - environmental temperature or
C.2.31
Does the medical device hold Consider: the consequence of the data Yes Refer to RE00076775 N/A N/A
C.2.32
Does the use of the medical Consider: for example, the Yes N/A N/A N/A
device depend on essential characteristics of the output of life-
performance? supporting
devices or the operation of an alarm.
C.2.34
GS10004411, Attachment D
10037183 Rev W Revision 202 Page 13 of 14
Minimally Invasive Therapies Group System Risk Management
PB500 Series Product Risk Assessment / Risk Analysis Chart - Risk Analysis Chart
RISK ASSESSMENT
Req. ID
PRD296
GS10004411, Attachment D
10037183 Rev. W Revision 202 Page 14 of 14