Mgcplus PDF
Mgcplus PDF
English
MGCplus
Hottinger Baldwin Messtechnik GmbH
Im Tiefen See 45
D-64239 Darmstadt
Tel. +49 6151 803-0
Fax +49 6151 803-9100
info@hbm.com
www.hbm.com
Mat.: 7-2002.0612
DVS: A0534-30.0 HBM: public
03.2018
Subject to modifications.
All product descriptions are for general information only.
They are not to be understood as a guarantee of quality or
durability.
Table of contents
Table of contents
1 Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.1 Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2 Markings used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.1 Markings on the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.2 The markings used in this document . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.1 Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.2 Notes on documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.3 System description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.4 Layout of the MGCplus device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.5 MGCplus housing designs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.6 Possible amplifier/connection board combination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.7 Installation of the CP52 communication processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.8 Conditions at the place of installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.9 Maintenance and cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
4 Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.1 Connecting the MGCplus in a tabletop housing/rack frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.1.1 Mains connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.1.2 Synchronization of multiple CP52 devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.1.2.1 Synchronization of multiple CP52 devices via a synchronization jack . . . . . 32
4.1.3 Synchronization of CP52 with CP22/CP42 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.2 Shielding design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.3 Connecting the transducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.3.1 Connecting separate TEDS modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.3.2 SG full bridges, inductive full bridges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
4.3.3 Full bridge circuits on AP810i/AP815i . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5 Starting up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
5.1 Devices in the desktop housing and rack frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
7 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
7.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
7.2 General principles for adjusting a measurement channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
7.2.1 Adapting to the transducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
7.2.1.1 Extended functions of the ML38B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
7.2.2 TEDS transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
7.2.3 Signal conditioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
7.2.4 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
7.2.5 Analog outputs (single-channel modules only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
7.3 Adapting to the transducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
7.3.1 SG transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
7.3.1.1 Direct entry of transducer characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
7.3.1.2 Calibrating the characteristic curve of the transducer . . . . . . . . . . . . . . . . . . . 139
7.3.2 Strain gages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
7.3.2.1 Direct entry of transducer characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
7.3.3 Inductive transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
7.3.3.1 Direct entry of transducer characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
7.3.3.2 Calibrating the characteristic curve of the transducer . . . . . . . . . . . . . . . . . . . 148
7.3.4 Torque transducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
7.3.4.1 Direct entry of torque characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153
7.3.4.2 Calibration with shunt installed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
7.3.5 Adjusting the rotational speed channel, frequency measurement . . . . . . . . . . . . . . . . . . 159
7.3.6 Adjusting the rotational speed channel, power measurement . . . . . . . . . . . . . . . . . . . . . 162
7.3.7 Thermocouples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164
7.3.7.1 Direct entry of transducer characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165
7.3.8 Current and voltage measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166
7.3.8.1 Direct entry of transducer characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
7.3.9 Resistance temperature sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 169
7.3.9.1 Direct entry of transducer characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
9 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 213
9.1 Display format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 213
9.1.1 Select setup window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 214
9.1.2 Setup window Display format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 215
9.1.3 Setup window components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 216
9.1.3.1 Numeric value display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 217
9.1.3.2 Graphic display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 227
9.1.4 Limit value status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 230
9.1.5 Recording status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 231
9.2 F keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 232
9.2.1 F keys in measuring mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 232
9.2.2 F keys in Setup mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 234
9.3 Channel names . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 235
10 System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237
10.1 Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237
10.1.1 Define new user . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 238
10.1.2 Password protection activation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 239
10.1.3 Set access for operator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 240
10.1.4 Delete user . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 241
10.1.5 Change password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 242
10.2 Save/load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 243
10.3 Recording series of tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 248
10.3.1 Setting parameters of test series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 249
10.3.2 Format of the MGCplus measurement files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 267
10.3.2.1 Measured values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 267
10.3.2.2 Time channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 269
10.3.3 MEA format in detail (MGC binary format 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 270
1 Safety instructions
Intended use
The amplifier system is to be used exclusively for measurement tasks and
directly related control tasks. Use for any purpose other than the above is
deemed to be non-designated use.
In the interests of safety, the device should only be operated as described in
the operating manuals. It is also essential to comply with the legal and
safety requirements for the relevant application during use. The same ap
plies to the use of accessories.
Each time, before starting up the equipment, you must first run a project
planning and risk analysis that takes into account all the safety aspects of
automation technology. This particularly concerns personal and machine
protection.
Additional safety precautions must be taken in plants where malfunctions
could cause major damage, loss of data or even personal injury. In the
event of a fault, these precautions establish safe operating conditions.
This can be done, for example, by mechanical interlocking, error signaling,
limit switches, etc.
Residual dangers
The scope of supply and performance of the amplifier system covers only a
small area of measurement technology. In addition, equipment planners,
installers and operators should plan, implement and respond to the safety
engineering considerations of measurement technology in such a way as to
minimize residual dangers. On-site regulations must be complied with at all
times. There must be reference to the residual dangers connected with
measurement technology.
After making settings and carrying out activities that are password-
protected, you must make sure that any controls that may be connected re
main in a safe condition until the switching performance of the amplifier
system has been tested.
Working safely
The supply voltage connection, as well as the signal and sense leads, must
be installed in such a way that electromagnetic interference does not ad
versely affect device functionality (HBM recommendation: "Greenline
shielding design", can be downloaded from http://www.hbm.com/Greenline).
Automation equipment and devices must be designed in such a way that
adequate protection or locking against unintentional actuation is provided
(access checks, password protection, etc.).
When devices are working in a network, these networks must be designed
in such a way that malfunctions in individual nodes can be detected and
shut down.
Safety precautions must be taken both in terms of hardware and software,
so that a line break or other interruptions to signal transmission, such as via
the bus interfaces, do not cause undefined states or loss of data in the auto
mation device.
Error messages should only be acknowledged once the cause of the error is
removed and there is no further danger.
Qualified personnel
are persons entrusted with siting, mounting, starting up and operating the
product and who possess the appropriate qualifications for their function.
Safety rules
Before starting up, make sure that the mains voltage and type of current
stated on the type plate match the mains voltage and type of current at the
place of operation and that the circuit used is sufficiently protected.
The mains plug must only be inserted into a grounded socket with a protec
tion switch (protection class I).
Use only the mains cable included with delivery, which is fitted with a ferrite
core.
The device must be switched off and the mains plug disconnected from the
socket before opening the device.
Never pull the mains plug out of its socket by the supply lead.
Do not operate the device if the mains lead is damaged.
If an amplifier channel is removed, the module must be sealed with a blind
panel.
Only operate built-in devices once they are installed in the housing pro
vided.
The device complies with the safety requirements of DIN EN 61010 Part 1
(VDE 0411 Part 1); protection class I.
The insulation resistance of the connecting cables (v50V) must be at least
350V(AC).
The MGCplus device has been tested based on EMC product standard
EN 61326-1:2013. This standard includes definitions of limit values and test
levels for various electromagnetic environments.
Regarding emission (EME), requirements are included for class A (industrial
environments) and class B (residential, business and commercial environ
ments as well as small businesses). Laboratory applications also usually
require class B.
The product standard here references to EN 55011:2009+A1:2010.
Regarding immunity to interference, the product standard includes require
ments for controlled electro-magnetic environments (lowest requirements),
general environments and industrial environments (highest requirement).
MGCplus meet the following requirements:
- Emission (EME): Class B
- Immunity to interference: Industrial environment
The MGCplus series and the individual modules thus essentially meet the
highest requirements and are therefore suitable for use in all environments
described in the product standard.
2 Markings used
CE mark
By way of the CE mark the manufacturer guarantees that the product com
plies with the requirements of the relevant EC directives (the Declaration of
Conformity can be found at http://www.hbm.com/HBMdoc).
Any risk of residual dangers when working with the amplifier system are
pointed out in these instructions by means of the following symbols:
Symbol Meaning
This marking warns of an imminently threatening dangerous situation in
DANGER which failure to comply with safety requirements will result in death or
extremely serious physical injury.
This marking warns of a potentially dangerous situation in which failure to
WARNING comply with safety requirements could result in death or serious physical
injury.
Symbol Meaning
This marking warns of a potentially dangerous situation in which failure to
CAUTION comply with safety requirements could result in slight or moderate physi
cal injury.
This marking draws your attention to a situation in which failure to comply
Notice with safety requirements could lead to property damage.
This marking draws your attention to important information about the
product or about handling the product.
Important
This marking indicates tips for use or other information that is useful to
you.
Tip
This marking draws your attention to information about the product or
about handling the product.
Information
Emphasis Italics are used to emphasize and highlight text and identify references to
See … sections, diagrams, or external documents and files.
Device -> New Bold text indicates menu items, as well as dialog and window headings in
the program environment. Arrows between menu items indicate the se
quence in which the menus and sub-menus are called up
Sampling rate Bold text in italics indicates inputs and input fields in the user interfaces.
3 Introduction
The degree of protection given in the technical data indicates the suitability
of the housings for various ambient conditions and also the protection of
persons against potential risks when used. The letters IP (International Pro
tection) which are always present in the designation, are followed by two
digits. These indicate which degree of protection a housing offers against
contact or foreign objects (first digit) and moisture (second digit).
MGCplus devices are available with degree of protection IP20.
IP 2 0
The complete documentation for the MGCplus amplifier system includes the
following documents:
S The operating manual,
which explains manual operation of the device and how to perform mea
surements with it.
CD-ROMs containing the following documentation are included with every
system device:
S Operation with computer or terminal,
which contains commands for programming and measuring with com
puter or terminal.
S MGCplus Assistant,
Documentation of the program for parameterization and control of the
MGCplus measuring amplifier system.
This manual contains all the information required to operate the MGCplus.
Guides
Several guides are available to help you:
S The header shows you which section or sub-section you are currently
reading. For example:
Introduction
Notes on documentation
S See è section 6 „Functions and symbols of the AB22A“ for explanations
of the AB22A display and control unit.
S è Section 11 „Menu structure“ provides an overview of the drop-down
and setting menus of the display and control unit.
Additional MGCplus
PC interface
Profi-bus
switches
"10V
CAN
...
1 2 8
...
2400Hz
2400Hz
2400Hz
Digital signal
conditioning
Digital
Synchronization
signal conditioning ... Filtering
Scaling, zero ... CPU ... CPU Display
and
Communication
processor CPxx
balance, ...
Filtering, scaling, control
zero balance, ... panel
CPU CPU
8-channel Single
module channel Serial bus
module
Connection boards
(AP01i, AP815i, ...)
Power CPxx
supply Communication
processor
AB22A
display and
control unit
Amplifier
plug−in board
(ML30B, ML55B, ML801B...)
Fig. 3.2 Device layout with display and control unit AB22A
Desktop housing Rack frame Slots Supply voltage (V) Weight, approx. (kg)
TG/ER
TG001E − 6 230 (115) 5.9 1)
TG003E ER003E 16 230 (115) 8.3 / 5.5 1)
TG009E − 2 230 (115) 5.0 1)
1) With the NT030 power pack, the enclosures weigh about 150g less each
Single-channel amplifier
R
TEDS
AP01i
1,4,5,B1
1,4,5,B1
R
TEDS
AP03i
1,4,5,B1
1,4,5,B1
AP14
AP17
T3...T10
R Piezoresistive
SG full bridge circuit T3...T10
transducer
1,4,5,B1
T1, T4, T5, TB1
1) For the combination of ML55B with AP14, a one-time zero calibration must always be
performed after setting up the measurement chain.
Multi-channel amplifiers
AP402i
AP418i
AP455i
TEDS
AP455iS6
TEDS
AP460i
Thermocouples
Inductive full bridge
TEDS
AP815i
R Pulse counter, frequency
Piezoresistive
transducer
Potentiometer
AP835 200 −5000
Voltage
LVDT
TEDS
AP836i Current
Digital input
AP72 Serial I/O
Analog output
AP74 CANHEAD
ProfiBus
AP75
CAN CANBus
RS232, RS422,
Serial I/O
RS485 I/O
AP77
For type "D" housings (ER003D or TG001D, etc.) the existing communica
tion processor (CP22/CP42) can be replaced by the CP52 communication
processor.
► Loosen the screws on the old communication processor, the blind panel
(only with CP22) and the cover of the NT030 power supply unit.
► Remove the parts.
► Insert the new CP52 communication processor and screw it in place.
► Fit the power supply cover of the NT030 and screw the cover in place.
The process is similar when subsequently installing a CP52 communication
processor in an MGCplus housing (type "D" or type "E") that was initially
configured without a communication processor.
► Loosen the screws on the blind panels, if there are any, of the SY03 syn
chronization interface and the power supply cover.
► Remove the parts.
► Insert the new CP52 communication processor and screw it in place.
► Fit the power supply cover of the NT030 or NT040 and screw the cover
in place.
CP42
RS 232
CARDBUS
USB DEVICE
USB HOST
YE SLAVE
RD ERROR
ETHERNET GN MASTER
IN
SYN
CTRL I/O C
OU
T
24V 1 2 GND 1 2
OUT IN
Communication processor CP42, power supply Communication processor CP52, power supply
unit NT030 unit NT040
CP CP
yes (1) yes (1)
S3 S3
no (2) no (2)
Power S3 CP switch
supply p2 Interface switch
S1
CAUTION
The MGCplus system devices are maintenance-free. Please note the follow
ing points when cleaning the housing:
CAUTION
Disconnect the mains plug from the socket before cleaning.
S Clean the housing with a soft, slightly damp (not wet!) cloth. You should
never use solvents, since this may damage the labeling on the front
panel and the display field.
S When cleaning, ensure that no liquid gets into the device or connections.
4 Connection
Housing ground
CAUTION
The power supply fuse may only be replaced by the manufacturer's service
personnel!
Grounding switch
In the factory setting ( ), the grounding switch connects supply voltage
zero with the protective conductor. If external devices (transducer, com
puter) have already set up this connection resulting in ground loops (hum-
pickups), the grounding switch must be opened ( ).
Synchronization sockets
Fig. 4.1 Example of synchronizing two MGCplus systems equipped with CP52.
The overall length of the synchronization chain (total length of cable be
tween the sync master and the last sync slave) must be less than 150 m. A
termination resistor should be used if the line length is >15 m. We recom
mend that you attach a termination resistor connector to the Sync Out
socket (X2) of the last sync slave. This connector is available from HBM on
request. The maximum number of MGCplus units that can be synchronized
is 32.
CP42
RS 232
CARDBUS
USB DEVICE
USB HOST
YE SLAVE
RD ERROR
ETHERNET
GN MASTER
IN
SYNC
CTRL I/O
OUT
24V1 2 GN1 2
OUT D IN
Fig. 4.2 Example of synchronizing two MGCplus systems equipped with CP52
and CP42.
Notice
All parts of the measurement chain (including all cable connection points
such as plugs and couplings) must be surrounded by a closed EMC-tested
shield. Shield junctions must represent a full contact, closed and low-imped
ance connection. This is the case for original HBM plug connections.
AP01i AP03i
AP14
AP455i
1 9
A F
B E
G
15 C D
8
1 4
Cable shield Hsg. Hsg.
3
Sense lead (+) 13 F
Sense lead (-) 12 G
4
RB / 2 (on the transducer)
Hsg. = Housing
Measurement
signal (+)
Bridge excitation
4.3 Connecting the transducer
voltage (-)
2
Bridge excitation
voltage (+)
Measurement
1 4 signal (-) Important
Cable shield 1)
3 Transducers with four-wire configuration
Sense lead (+)
If you connect a transducer with a 4-wire cable, you must connect the sense
Feedback Sense lead (-) lead with the corresponding bridge excitation circuit in the transducer plug
bridges
(sense line (-) with bridge excitation voltage (-) and sense lead (+) with
bridge excitation voltage (+) 1). A cable extension may only be implemented
with 6-wire configuration.
1 9
Important
8 15
Single-channel amplifier MLxx (together with connection board AP01i) must
have at least hardware revision 1.32 or higher.
TEDS 4
TEDS data 9
1) For cable lengths >50m, a resistor of half the value of the bridge resistance (RB/2) must be activated on the transducer instead of each
feedback bridge. If the transducers are calibrated in a 6-wire configuration, resistors must be activated directly into the sense lead.
AP402i
View of the mating connector
4x (solder side)
5
6 4
1
3
2
4
Hsg.
TEDS 3
TEDS data 5
AP460i
7 6
8 5
9 10
1 4
2 3
10
1
2
4
Hsg.
5
1 9 A F 6 4
B E
G 1 3
C D
2
8 15
wh
Measurement signal (+) 8 A 1
bk Bridge excitation
voltage (-)
5 B 2
2 b
u
Bridge excitation
voltage (+)
6 C 3
r
d Measurement signal (-) 15 D 4
1 4
ye Hsg. Hsg.
Cable shield Hsg.
3
gn
Sense lead (+) 13 F 5
g
y
Sense lead (-) 12 G 6
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye= yellow; gn= green; gy= gray
AP810i/AP815i
Subchannel Subchannel Subchannel Subchannel
1
1/5 2/6 3/7 4/8
14
Subchannel 1...4
wh
Measurement signal (+) 2 5 8 11
bk
13 25 Bridge excitation
voltage (-)
1 4 7 10
2 bu Bridge excitation
voltage (+)
3 6 9 12
1 rd
14 Measurement signal (-) 15 18 21 24
1 4
ye Hsg. Hsg. Hsg.
Cable shield ) Hsg.
3
gn
Sense lead (+) 16 19 22 25
Subchannel 5...8
gy
13 25 Sense lead (-) 14 17 20 23
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye= yellow; gn= green; gy= gray
5
A F 6 4
1 9 B E
G 3
C D 1
2
8 15
wh
Measurement signal (+) 8 A 1
bk Bridge excitation
voltage (-)
5 B 2
2 b
u
Bridge excitation
voltage (+)
6 C 3
1
ye
Cable shield Hsg. Hsg. Hsg.
3
gn
Sense lead (+) 13 F 5
g
y
Sense lead (-) 12 G 6
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye= yellow; gn= green; gy= gray
AP455i AP455iS6
5
6 4
1 3
2
LVDT transducers
Measurement signal (+) 8 1
Bridge excitation voltage (-) 5 2
Bridge excitation voltage (+)
6 3
Measurement signal (-) 15 4
Cable shield ) Hsg. Hsg.
AP810i
Subchannel
1...4
25
rd Sense lead (-) 14 17 20 23
13
2
wh/rd Bridge excitation voltage (-)
1 4 7 10
AP815i
Subchannel
1...4
25
rd Sense lead (-) 14 17 20 23
13
2
wh/rd Bridge excitation voltage (-)
1 4 7 10
AP815i can measure decentralized half bridge circuits for which the active
SGs are separated by a line.
1) With decentralized half bridge circuits the measured value must be acquired at both ends of the connecting line between the active SGs.
With standard half bridge circuits a connector can also be bridged.
AP14
1 9
8 15
SG SG
Hsg.
AP814Bi
1
14
Subchannels
1 2 3 4 5 6 7 8
Subchannel 1...8
25
13
Measurement signal (+), 5
2 15 5 18 5
8 21 11 24
excitation voltage (+)
SG
Cable shield Hsg. Hsg. Hsg. Hsg. Hsg. Hsg. Hsg. Hsg.
Three-wire connection
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye= yellow; gn= green; gy= gray
AP815i
Subchannel
1...4
25
2 wh/rd Excitation voltage (-) 1 4 7 10
13
1
14
SG Cable shield Hsg. Hsg. Hsg. Hsg.
Subchannel
Measurement signal (+),
5...8 1 wh/gn sense lead (+)
2 5 8 11
25
13
4 gn Excitation voltage (+) 15 18 21 24
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye= yellow; gn= green; gy= gray
Subchannel
1
AP815i
2 Excitation voltage (-) 1
Subchannel
2'
1
14
Sense lead (-) 14 2/6
SG SG SG SG SG
Subchannel 5 4/8 3/7 2/6 1 Subchannel
Cable shield Hsg.
1...4 3/7
Measurement signal (+),
25
13 1 sense lead (+) 2 Subchannel
4 Excitation voltage (+) 15 4/8
1
14
Subchannel
Subchannel 5
5
18
5...8
13
25 8
21
11
24
2
15
You can operate a maximum of eight SGs at 120 ohms with a 5-V current
feed. Make certain that sensor point 2' of the SG chain is as close as possi
ble for the individual strain gages and the distances between the individual
strain gages are short.
If the distances between the individual strain gages cannot be kept small
(for example two 90 strain gage rosettes in different places), they must be
connected as follows:
Subchannel
1
AP815i 2 Excitation voltage (-) 1
Subchannel
2' 2
Sense lead (-) 14
1
14
SG2 SG1
Cable shield Hsg.
Subchannel
1...4 1
Measurement signal (+),
2
sense lead (+)
25
13
4 Excitation voltage (+) 15
1
14 5
18
Subchannel Subchannel
5
5...8
2 Excitation voltage (-) 1
13
25 Subchannel
2' 6
Sense lead (-) 14
SG4 SG3
Cable shield Hsg.
5
18
AP17
1 9
8 15
Plug 1
Md
bk
Power supply (0V) 5
bu
Supply voltage (18V ... 30 V) 6
r
2 3
4 d Torque measurement signal, frequency output (+) 12
w
1 h
Torque measurement signal, frequency output (-) 13
ye
7/5 6 Cable shield Hsg.
gn
Calibration signal trigger (approx. 5V) 14
g
y Ground 8
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye= yellow; gn= green; gy= gray
AP460i
7 6
8 5
9 10
1 4
2 3
Plug 1
Md
r
4 d Torque measurement signal, frequency output (+) 1
w
1 h
Torque measurement signal, frequency output (-) 2
ye
7/5 Cable shield Hsg.
gn
g
y Ground 10
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye= yellow; gn= green; gy= gray
Information
The torque flanges must be powered externally.
AP17
1 9
8 15
bk
Plug 2 and
n bu
Ground 8
rd
Rotational speed measurement signal, 0 (+) 12
wh
8 1
6 Rotational speed measurement signal, 0 (-) 13
gn
7 Rotational speed measurement signal, 90 (-) 14
3 Cable shield Hsg.
g
y
Rotational speed measurement signal, 90 (+) 15
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye= yellow; gn= green; gy= gray
AP460i
7 6
8 5
9 10
1 4
2 3
Plug 2
n ye
Ground 10
rd
Rotational speed measurement signal, 0 (+) 1
wh
8 1
6 Rotational speed measurement signal, 0 (-) 2
gn
7 Rotational speed measurement signal, 90 (-) 4
Cable shield Hsg.
3
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye= yellow; gn= green; gy= gray
AP17
1 9
8 15
Plug 2
n ye
Ground 8
rd
Rotational speed measurement signal, 0 (+) 12
wh
8 1
6 Rotational speed measurement signal, 0 (-) 13
gn
7 Rotational speed measurement signal, 90 (-) 14
Cable shield Hsg.
3 4 2
b
u
Reference signal (+) 2
bk
Reference signal (-) 3
g Rotational speed measurement signal, 90 (+) 15
y
5V (out) 11
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye= yellow; gn= green; gy= gray
AP460i
7 6
8 5
9 10
1 4
2 3
Plug 2
n ye
Ground 10
rd
Rotational speed measurement signal, 0 (+) 1
wh
8 1
6 Rotational speed measurement signal, 0 (-) 2
gn
7 Rotational speed measurement signal, 90 (-) 4
Cable shield Hsg.
3 4 2
b
u
Reference signal (+) 5
bk
Reference signal (-) 6
g Rotational speed measurement signal, 90 (+) 3
y
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye= yellow; gn= green; gy= gray
AP01i AP03i
AP142)
AP455i
A F
1 9
B E
G
C D
8 15
wh
Measurement signal (+) 8 A
bk Bridge excitation
voltage (-) 5 B
2 b
u
Bridge excitation
voltage (+) 6 C
r
1 4 d Measurement signal (-) 15 D
ye
Cable shield Hsg. Hsg.
3
Sense lead (+)
13 F
Feedback bridges for Sense lead (-)
12 G
transducer in a four wire
configuration *
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye= yellow; gn= green; gy= gray
AP460i
7 6
8 5
9 10
1 4
2 3
Rotational speed 1
measurement signal
(+) 1 k 5 k
Umax = 30V 5V
Rotational speed 2
measurement signal
Inductive (-)
tachometer
(T-R coil)
Please note the setting information for T-R coils on page 152.
4.3.12 Thermocouples
- +
Compensating
Thermocouple line
(-) −
(+) +
Compensating
line
Miniature thermo connector (see
table for matching type)
AP01i AP03i
1 9
Maximum input voltage against
A F
ground = 12V
B E
G
8 15 C D
(+) 8 A
U
Supply voltage zero1) 6 C
(-) 15 D
Cable shield Hsg. Hsg.
1) With a potential-free DC voltage source you must connect pin 15 with pin 6.
AP402i
5
6 4
4x
1 3
2
Maximum input voltage against
ground = " 100V
(+) 1
U
(-) 4
Cable shield
Hsg.
AP801 AP801S6
1 4
2 3
(+) (+) 1
U
U (-) 2
Hsg.
Power supply 0V 4
Hsg.
AP402i
5
4x
6 4
1 3
2
(+) 1
U (−) 4
Power supply 0V 3
Hsg.
AP836i
Subchannels
1/5 2/6 3/7 4/8
1
14
Subchannel 5...8
Sense lead (-) 14 17 20 23
Bridge excitation voltage (-) 1 4 7 10
Common signal and supply voltage zero, power line not corrected on one side.
Since the bridge excitation voltage used to supply the active transducer is symmetrical to GND/ground, the
design of the active transducer must without exception be potential-free!
AP836i
Subchannels
1/5 2/6 3/7 4/8
1
14
Subchannel 5...8
Sense lead (-) 14 17 20 23
Bridge excitation voltage (-) 1 4 7 10
Separate signal and supply voltage zero, power lines not corrected
Since the bridge excitation voltage used to supply the active transducer is symmetrical to GND/ground, the
design of the active transducer must without exception be potential-free!
AP836i
Subchannels
1/5 2/6 3/7 4/8
1
14
Subchannel 5...8
Sense lead (-) 14 17 20 23
Bridge excitation voltage (-) 1 4 7 10
Common signal and supply voltage zero, power lines fully corrected.
Since the bridge excitation voltage used to supply the active transducer is symmetrical to GND/ground,
the design of the active transducer must without exception be potential-free!
AP836i
Subchannels
1/5 2/6 3/7 4/8
1
14
Subchannel 1...4
Bridge excitation voltage (+) 3 6 9 12
Sense lead (+) 16 19 22 25
25
13 UB (+)
U_OUT Measurement signal (+) 2 5 8 11
0V Measurement signal (-) 15 18 21 24
UB (-) Cable shield 1) Hsg. Hsg. Hsg. Hsg.
Separate signal and supply voltage zero, power lines fully corrected.
Since the bridge excitation voltage used to supply the active transducer is symmetrical to GND/ground, the
design of the active transducer must without exception be potential-free!
AP01i AP03i
1 9
Maximum input voltage against A F
ground = 12V B
G
E
8 15 C D
I
(+) 5 B
Cable shield Hsg. Hsg.
AP402i
5
4x
6 4
Maximum input voltage against
ground = ±100V 1 3
2
(-) 4
I
(+) 2
Cable shield Hsg.
AP835
1 4
2 3
R
Cable shield Hsg.
AP01i AP03i
1 9
A F
B E
G
8 15 C D
Frequency
generator/
pulse
generator Cable shield Hsg. Hsg.
Rotational speed/pulse
signal 1 (frequency f1) 12 G
Deactivate the analysis of the f2 signal in this mode (factory setting: Off), see page 159.
AP01i AP03i
1 9
A F
B E
G
8 15 C D
Activate the analysis of the f_2 signal in this mode (factory setting: Off), see page 159
1 9 1 9
A F
B E
G
8 15 C D 8 15
8 A 10 8
f2
15 D 3 15
f1
Industrial 12 G 1 12
pulse generators
Zero index
6 C 5 2
Transducer error
5 B
AP17 AP460i
1 9
8 15
AP418i
Shield
Piezoelectric
transducer with
preamplifier
(with
T‐ID/TEDS for
AP418i)
Input *
Information
Information about AP418i connection boards:
When laying transducer cables outside of enclosed areas or with cable
lengths greater than 30 m between the connection board and transducer,
the sensor cables must be designed with an additional, separately grounded
shield to ensure overvoltage protection. This can be done for example by
laying the cable in a metallic pipe or using double-shielded cable, in which
case the outer shield must be connected to ground potential or protective
conductor potential where it is close to the connection board (for example
where it enters the switch cabinet). HBM recommends Triaxial cable for this
purpose.
AP01i AP03i
1 9
A F
B E
G
8 15 C D
AP01i AP03i
1 9
A F
B E
G
8 15 C D
1
Cable shield Hsg. Hsg.
3
Sense lead (+) 13 F
Sense lead (-) 12 G
AP836
Subchannels
1/5 2/6 3/7 4/8
1
14
1
Cable shield Hsg. Hsg. Hsg. Hsg.
3
Subchannel 5...8
Sense lead (+) 16 19 22 25
Sense lead (-) 14 17 20 23
Single SG; four-wire connection, AP815i only Strain gage half bridge on (AP810i) on
VT810/815i
RJ45 socket RJ45 socket
Hsg.
Cable shield 6
1 wh/gn Measurement signal (+) 3
3
wh/br Bridge excitation voltage (+) 7
4 gn Excitation voltage (+) 6 br Sense lead (+) 8
Connecting diagram
SG 1 D‐Sub25
1-Kab156-3 RJ45
D‐Sub25 1-Kab263-3
SG 2
1-Kab156-3
2
D‐Sub25
1-Kab263-3
1-AP810i
1-AP815i
SG 8
1-Kab156-3
8
10 measurement
channels
MGCplus
T-piece
10 10
measurement 10 measurement
channels measurement channels Termination resistor
channels
1) The AP74 connection board in the MGCplus contains a built-in termination resistor.
The CP52 communication processor has two digital inputs and outputs
each. The digital inputs and outputs must be operated with an external cur
rent feed (12 V ... 24 V).
Inputs (0 V ... 24 V)
You can assign the following functions to the digital inputs:
S Start to record measured data with an external trigger
The status of the inputs and outputs can also be queried with the MGCplus
terminal commands.
Outputs (0 V ... 24 V)
The following functions are assigned to the digital outputs:
S Disk is full
If the remaining storage space of the USB mass storage medium in the
CP52 is 1 Mbyte, the output is set to logical High. The status can be
changed using the MGCplus terminal commands.
S System is up and running
Inputs and outputs When the initialization is completed and measured values are acquired, the
output is set to logical High.
Notice
Functions can only be assigned to the digital inputs and outputs of the CP52
communication processor by using the MGCplus Assistant software from
HBM.
COM
NO NC
+ -
24V DC
Fig. 4.5 Wiring example for the "Start Trigger" function at the CP52 control inputs
On the front panel of a single-channel module there is a BNC socket for the
analog output signal VO1. (This socket is used for test purposes. Stationary
wiring should always be done with the connection boards, since there is no
noise voltage with this method).
Notice
Please note: The input resistance of the connected device must be greater
than 1MΩ.
When connecting a coaxial cable to the analog output of the ML60B ampli
fier module, a ferrite (available from Würth, art. no. 742 711 72, or similar)
must be placed on the cable for operation in environments of limit value
class B in accordance with EN55011 and EN55022 (residential applications,
business and commercial applications as well as small businesses).
Pin Function
1 14 Digital
1
2 Remote control 1 Input
3 Remote control 2 Input
4 Remote control 3 Input
13 25 5 Remote control 4 Input
6 Remote control 5 Input
Sct 2 7 Remote control 6 Input
Outputs 8 Remote control 7 Input
OUTPUT
9 Remote control 8 Input
10 no function -
11 no function -
12 VO2 (Ra>5k) Output
13 VO1 (Ra>5k) Output
16 Digital Input
17 Limit value output 1 Output
18 Limit value output 2 Output
19 Limit value output 3 Output
20 Limit value output 4 Output
21 Warning Output
22 no function -
23 no function -
24 to Pin 12
Analog
25 to Pin 13
Assignment of outputs
Analog outputs
S On pin 12 the analog output signal VO2 is present.
The connected load resistor must be greater than 5 kohms.
S On pin 13 the analog output signal of VO1 is present (and also on the
BNC socket on the front panel).
The connected load resistance must be greater than 5 kohms.
The following signals can be assigned to outputs VO1 and VO2:
S S1: Gross
S S2: Net
S S3: Peak value 1
S S4: Peak value 2
S S5: various differential, integration and mean value
S On pins 17 to 20 the switching states of limit switches 1...4 are present.
The switching states are indicated by two different HCMOS voltage lev
els:
Positive logic:
Level 0 V: limit switch OFF
Level 5 V: limit switch ON
S On pin 21 a level of 5 V (High level) is present, which can be used as a
warning signal. In case of a fault or broken transducer cable the output
signal is set to 0 V (Low). However, this signal is also set to zero during
the autocalibration cycle (every 5 minutes for about 1 s).
Remote controls
On pins 2 to 9 of socket 2, remote controls CTRL 1...8 are present for con
trol of some amplifier functions. These contacts are active if they have been
enabled with the AB22A display and control unit, i.e. in REMOTE operating
mode. The assignment of these remote controls is freely selectable. The
possible functions are described in
section 8 Additional functions .
Information
In the factory settings the contacts are not assigned.
External triggering
Remote control 7 is used as an external trigger input.
With AP01i and AP03i, jumpers can be used to adapt the filter settings of
the two analog outputs follows:
Filter
approx. 3 kHz Analog output VO1 Analog output VO2
approx. 2. order
OFF
ST1 ST6
for ML10B
On
ST1 ST6
for all others MLxx
Termination resistors must be connected for long lines (>100m) and high
frequencies (>200kHz). To do this the 3x DIP switch S2 on the motherboard
of the AP17 must be switched to "ON".
AP17
ON
3
2
1
Switch S2
Connection
Symmetrical input signals (RS 422): Input a/input b
Asymmetrical input signal, bipolar: Input a (signal ground on input b)
Asymmetrical input signal, unipolar: Input a (signal ground on pin 10, input b
must remain open)
Profibus
4.5.3.3 AP77
The pin assignment of the 9‐pin Sub‐D socket complies with Profibus stan
dards IEC 61158/61784.
1 6
Pin Function
1 -
2 -
5 9 3 RS485-B
4 RS485-RTS
5 GND
6 VCC
7 -
8 RS485-A
9 GND
Information
Further information can be found in the ML77B operating manual.
The AP75 connection board has 8 digital inputs, 8 digital outputs and 2 ana
log outputs. The digital outputs must be operated with an external current
feed (12 V...24 V). The AP75 connection board can be operated together
UD01 with the special function modules ML78B or ML70B.
24V
Please note that the measuring systems of the digital inputs and outputs are
separated from each other.
24V
Analog outputs V01 and V02 have a common ground system that is sepa
rated from the ground systems of the digital inputs and outputs.
UOUT
The AP78 connection board has 10 analog outputs. The analog outputs out
puts designated A03...A10 are electrically isolated, while outputs VO1 and
1 VO2 can be digitally filtered (together with ML78B). The AP78 connection
A03 14 GND A03
A04 board can be used together with the ML78B and the freely programmable
GND A04
GND A05 ML70B module (CoDeSys).
5 Starting up
This section shows you the necessary operating steps to place your mea
surement chain (measuring amplifier system and transducer) in operation.
This will enable you to perform a functional test of all the components. The
steps are deliberately described in very general terms so that there is no
need to go into the details of specific transducers or amplifier modules.
However the description can easily be applied to your measurement chain.
In some cases – especially when connecting transducers – reference is
made to the following sections. We also point out some typical errors that
may occur during start-up.
After the initial start-up is performed and the amplifier module is adapted to
your transducer, you will be ready to become familiar with the remaining
functions and possibilities of the MGCplus measuring amplifier system.
S Unpack the MGCplus.
S Check the MGCplus for damage.
S Is the delivery complete?
S Compare the contents of the package with the enclosed documentation
list. Is the documentation complete?
If you have not already received your measuring amplifier system complete,
note the following information as you put together your system:
S Plug in the measurement cards from the front and the corresponding
connection boards from the back.
The assignment is important in this process.
S If you are using the wide connection board or amplifiers (8 sub-units),
plug them into slots 1, 3, 5, etc. Slots 2, 4, 6, etc. on the front and back
must be left free or fitted with blind panels.
S For safety reasons all free slots (amplifiers or connection boards) must
be covered by blind panels.
S Check to make certain the amplifiers and connection boards are securely
plugged in.
S Connect the device to the mains with the mains cable provided.
S Connect your transducer to the applicable socket on the back of the
connection board (designation BU01). If you are using a cable pre-
assembled in-house, please note the pin assignments for your
transducer in section B.
Comply with the safety instructions on è page 11.
Switching on
S Turn the device on using the POWER button on the front of the device.
The AB22A is initialized (all the LEDs light up briefly) and detects the
components present.
MGCplus initialization
Password
Save/load
Recording
Interface
Print
Language
Time
LANGUAGE
Language: Deutsch
Deutsch
English
Francais
You can make all the settings on your device with the keys on the
AB22A/AB32 display and control unit. You can use the Switch key to
select the operating states "Measurement" or "Setting".
6.2 Display
After the mains voltage is turned on, the initialization of the device is in
dicated in the display by a horizontal progress bar. You are also informed of
the current software version.
After the opening display, the measured value of image type "1 measured
value" appears in the standard configuration (factory settings). Pressing the
Switch key takes you to Setup mode, where you can configure the sys
tem, display, amplifier and additional functions. We recommend first setting
the Language if you want to use a language other than German.
Gross Gross
Display with image type "6 measured values"
Gross Gross
To the next image type with
Gross Gross
Measure
Measure
Measure
LIMIT VALUE
1-LV 1 LV On Display with image type "Limit value status"
1-LV 2 LV Off
1-LV 3 LV On To the next image type with
1-LV 4 LV Off
Measure
Gross
Channel
Status of limit switches. If the set activation level of a limit switch is ex
ceeded, the switch number has a black background in the display.
Example: Activation level of limit switch 1 is exceeded
Local Remote control turned off
Remote Remote control turned on
System-related settings User-oriented settings of Amplifier settings for Additional settings for
that are usually only the display such as the each channel, for exam each individual amplifier.
made for initial operation preferred realization of ple to adapt to the ampli
or a new measurement measured values, assign fier module or signal con
task. ment of function keys or ditioning.
assigned channel names.
Press the Switch key to go to the Setting dialog. In the bottom section
of the display there is a selection bar with terms assigned to the function
keys below them (F1...F4) (selection level 1).
You can adjust your device in various menus that can be called with the
selection bar in the display. Depending on the function you can call up to 4
menu levels (selection and setting levels). In the first two levels you can se
lect terms. You are in the selection levels. In the next levels you will adjust
certain values or turn functions on/off. Then you will be in the setting levels.
Selection level 1
Selection level 2
Password
Password Pull‐up menu
Save/load Setting level 1
Display Recording PASSWORD
Interface
User: new delete
Print
Language Display Amp Password: chang
Setting level 2
Selection bar Add user
e
Access: set
User:
Setup window Display Amp Password:
Access: Operator
OK Cancel
Setup window
Press the Switch key . At first only the selection bar appears. Now if you
press the function keys F1...F4, the associated pull-up menu appears above
the corresponding term (in this example System). Move the light bar to the
desired term in the pull‐up menu with the cursor keys (in this case
Language) and press the confirmation key . Now you are in the setting
level for the selected term. The current setup window may lead to additional
setting levels.
Selection level 2
Password
Pull‐up menu Save/load
Recording
Interface
Print Light bar
Language
Time
Deutsch
Setting level 1 Deutsch O English
Selection window Francais
Measuring mode
Gross
Channel
Selection level 1
Selection level 2
Display Password
Password
Save/load
Recording Setting level 1
Interface
Print PASSWORD
Language User: .... new delete
Displa
Password: .... change
Access: .... set
Setting level 2
Add user
Display ....
User:
Password: ....
Access: Operator
OK Cancel
Before you exit a menu window and return to measuring mode, you always
Save settings? have the option of saving the settings you have made, deactivating them or
Yes No Cancel aborting the dialog window. The confirmation prompt shown on the side ap
pears for this purpose.
2. With direct input of the channel number using the alphanumeric keypad
(recommended with multi-channel modules).
Two examples of method 2:
► Confirm with .
(If you confirm now with , the first subchannel, in this case 3.1, is selected).
1. With the channel selection keys (select the channels one after the
other).
2. By entering the channel number in the channel number input field.
There is an option that allows you to jump directly from one channel to
any other.
► Press the cursor key to go to the channel input field of the header.
► Confirm with .
All the settings you make before the confirmation prompt are temporarily
saved in working memory (RAM) as soon as you have made a change and
confirmed with . The data is permanently saved as soon as you exit
Setup mode and confirm the confirmation prompt with Yes.
In the first two selection levels you can select terms. On the first level (menu
bar) by pressing the corresponding function key, on the second level by
making a selection in pop‐up menus.
Password S Select and confirm in the selection fields (pull‐up menu)
Save/load
Recording S The selected field appears in reverse video. Confirm the selection with
Interface
Print
Language
Time
Enter the parameters on the setting level with the setup window. The setup
windows contain Dialog boxes that include four different types.
SETUP WINDOW
Abs
S Switch on or off in activation fields
Gross
Net
The selected field appears in reverse video. Confirm the selection with .
A tick appears in selected boxes ("activated"). Pressing the activation button
again cancels activation.
Gross
Net
PV1
PV2
PV12
LV status
Multi-channel modules are identified by two dots under the channel number
(channel number 8 in the example). If you select the activation field of a
multi-channel module and activate it with , a new setup window appears
with activation fields for the individual subchannels.
Channel 1 2 3 4 5 6 7 ..
8 9 10 111213 141516
1 2 3 45 6 78
Gross All
Net
PV1
OK Cancel
PV2
PV12
LV status
Current feed: 5V
S Expand and make a selection in the selection fields
1V The selected field appears in reverse video. After you press the confirmation
Current feed: 2.5V key the selection field opens up. Select with the cursor keys and
5V
confirm your setting with . In the documentation (not in the display) fields
of this kind are identified by an arrow pointing down .
Information
The "Unit" selection field in the "Transducer" setup window can be edited by
double clicking with . Then you can enter up to four characters (for
The selected field appears in reverse video. Confirm your entries with .
In the example on the left you can see one edit field with no content and
another edit field with content.
An edit field together with content can be
S directly overwritten,
S partially edited
The minus sign of the key can also be used as a separator in text fields.
The key also has the superscript numbers 2, 3, and special character @
assigned to it.
Buttons
Cancel
The selected field appears in reverse video. Confirm with . If the button
or
label is followed by three dots, (change...), an additional setup window ap
Change... pears after confirmation.
7 Measuring
This section describes the steps you will need to follow to perform a mea
surement with MGCplus. At the beginning of each section an example ex
plains the transducer-specific details for adapting the amplifier modules. Af
ter you have adapted the amplifier modules you can start with the
measurements. Other functions such as limit switches and peak-value
memory are described in è section 8 "Additional functions", page 191 .
Basic unit
(mV/V, etc.) Analog output
Transducer Amplifier
(V)
Display
Inputs and outputs
(User unit: kg, etc.)
► Adjust amplifier
Once these steps are complete the most urgent settings and adaptations
have been made and simple measurements are possible. The following
steps are optional
DISPLAY CHANNEL 1
► Adjusting the display format
Unit kg
Coefficients/standardized
You can use the cubic polynomial characteristic curve to correct the trans
ducer signal (in mV/V). This adjusts the measured values to the true charac
teristic curve. To do this enter the coefficients A0, A1, A2 and A3 of the third
order polynomial, which is described by the following equation:
YPh = A3@Yel3 + A2@Yel2 + A1@Yel + A0
YPh: Display in physical unit with polynomial characteristic curve
Yel: Measured value in electrical unit
The values of the coefficients can be found on the transducer calibration
certificate.
The amplifier itself requires coefficients (C0 ...C3) standardized to the value
1. The coefficients from the calibration certificate (A0 ...A3) are automati
cally converted before they are forwarded to the amplifier (also in
MGCplus‐Assistant).
Notice
The actual amplifier will then be parameterized with the transducer settings
from the TEDS chip if it supports this functionality.
For an overview of which connection board/amplifier combinations are
TEDS-capable, see section 3.6. The amplifiers must be equipped with the
current firmware, which is available from www.hbm.com/downloads.
Zero offset: 0.0000 ... V −>0<− The amount by which the relative zero point is offset relative to absolute
Tare: 0.0000 ... V −>T<− zero.
Disable zero: 3 Disable taring: 3
100 Hz Bessel V
Example: A displacement transducer (nominal (rated) displacement 20mm)
Low pass
High pass Off
will be mounted at a height of 1m, measured from the machine foundation.
Absolute movement will be shown in the display.
Displacement transducer
Reference zero:
Zero offset:
Tare:
1000.00 mm
Tare
Activate the −>T<− button to initiate a taring. If the tare value is known, you
can also enter it directly in the edit field. Taring affects the net display.
A brief example will clarify the difference between a zero balance and taring:
A platform is standing on three load cells and will be used for tank weighing.
Tank
Platform
Two different components are poured into the tank and their individual
weight (net) will be displayed.
ÉÉÉÉÉÉÉÉÉÉ The table summarizes the states before and after zero balance or taring is
ÉÉÉÉÉÉÉÉÉÉ
Load cells
triggered.
Low pass
Step response Low-pass filters are used to suppress unwanted higher frequency
interference that lies above a specific cut-off frequency.
Amplitude response, runtime and step response are dependent on the filter
u
High pass
Bessel High-pass filters are used to suppress unwanted lower frequency
t
interference that lies below a specific cut-off frequency. With the high-pass
filter you can suppress slow fluctuations, such as those resulting from
Damping temperature influences and time drift.
The high-pass filter has no function when resistance thermometers and
thermocouples are connected.
u
Important
Butterworth f
To ensure that the high pass will function, the low-pass setting must not fall
below the following limits:
Damping
Bessel: .5Hz
Butterworth: .10Hz
u
The nominal (rated) values of the filters provided by the AB22A display and
control unit and by the MGCplus Assistant and catman software are
oriented towards the -1 dB values (see technical data).
Bessel f
7.2.4 Display
Unit1
Select the unit in the display. You can choose between the basic unit (mV/
System Display Amplifier Options V), the user unit (kg, etc.) and the unit of the analog output (V). Then the
corresponding measured value appears in the display.
You can achieve the same effect with the "Unit" function (factory settings:
DISPLAY CHANNEL1
F3/level 2; è see also section 9.2 "F keys", page 232).
Unit V 3 Abs
Decimal places 3 ... Abs
Display range from −10.000 V
to 10.000 V
If the absolute display is activated ( 3 ), the signal appears in the display at
Step 100
the amplifier input without signal conditioning (for example zero drift or tar
ing).
Decimal places
Number of places after the decimal.
Display range
Display range in the selected unit (displayed automatically).
Step
The step defines the size of the display jumps. It refers to the last decimal
place of the nominal (rated) value.
Example: Nominal (rated) value 20kg
Decimal places 1 (20.0kg)
Step 1 means display jumps every 100g
Step 5 means display jumps every 500g
Decimal places 3 (20.000kg)
Step 1 means display jumps every 1g
Step 5 means display jumps every 5g
Output VO1
Signal selection for analog output 1 (socket Bu2 on the back side of the de
System Display Amplifier Options vice and BNC socket on the front panel for test purposes).
Output VO2
Transducer Signal selection for analog output 2 (socket Bu2 on the back side of the de
Signal conditioning vice).
Display
Analog outputs
Switch
Output characteristic
The edit fields for characteristic curve points 1 and 2 are updated automati
cally as soon as you activate the Adjust amplifier button in the "Trans
ANALOG OUTPUTS CHANNEL1
ducer" setup window. However, you can also change the output characteris
Output Vo1: Gross Output Vo2: Net tics with direct entry.
Output characteristics ppm V
Pt.1: 0.0000 ... 0.0000 ...
Pt.2: 100.0000 ... 0.0000 ...
User unit
(ppm, etc.)
Gross
Net
PV1 Pt.2
PV2
OFF
Comb. PV
OCC
Pt.1
Transducer 1. Use the channel selection keys or a direct entry (è see page 114) to
Signal conditioning
Display select the desired channel.
Analog outputs
Switch
2. Use the Switch key to switch to Setup mode.
3. Press the function key, use the cursor keys to select the
TRANSDUCER CHANNEL1
Transducer and confirm with .
Type SG full bridge circuit
Current feed 5V
Unit: kg mV/V 4. Use the cursor keys to select SG full bridge circuit and confirm
Zero pt.: 0.0000 ... 0.0000 ... measure
Nom. val.: 50.0000 ... 2.0000 ...
with .
calibrate... Adjust amplifier
5. Use to switch to the Current feed selection field, press and
Gage factor: 0.0000 ...
select 5 V.
6. Confirm with .
8. Use to switch to the Zero point edit field and enter the value 0 in
the left edit field. Confirm with .
10. Use to switch to the measure... button and confirm with (the mea
sured value appears in the right edit field of Zero point)
System Display Amplifier Options
11. Use to switch to the Nominal (rated) value edit field and enter the
value 50. Confirm with .
Transducer
Signal conditioning 12. Use to switch to the right edit field (mV/V) and enter the value 2.
Display
Analog outputs Confirm with .
Switch
13. Use to select the Adjust amplifier button and confirm with .
14. Use the Switch key to go to measuring mode and confirm the con
TRANSDUCER CHANNEL1
7.3.1 SG transducers
SG transducers (load cells and force transducer from HBM) are passive
transducers that have the following features:
S They must be supplied with an excitation voltage (carrier frequency or
direct voltage)
S They are SG full bridge circuits
S The transducer is identified by the following data:
- Maximum capacity (for example 20 kg)
- Nominal (rated) sensitivity (for example 2 mV/V)
These data items must be taken into consideration when adapting the
amplifier.
The standard setting for the excitation voltage UB with SG full bridge circuits
is 5 V. If higher measuring ranges are required or multiple transducers are
connected in parallel, a lower excitation voltage must be selected. The
individual values depend on which module is used.
Important
The specified full scale values are maximum values that cannot be affected
by settings of the analog output!
7. Confirm with .
9. Use to switch to the Zero point edit field and enter the value 0 in
the left edit field. Confirm with .
11. Use to switch to the measure button and confirm with (the
measured value appears in the right edit field of Zero point)
System Display Amplifier Options
12. In the left edit field Nominal (rated) value enter the value 20.
13. In the right edit field Nominal (rated) value enter the value 2 (under the
unit mV/V).
TransducerTransducer 14. Use to select the Adjust amplifier button and confirm with . If
Signal conditioning
Display you do not want to make any more changes to the display or output
Analog outputs
Switch characteristics, you can now continue with step 24.
Unit kg 16. In the pull-up menu select Display and confirm with .
Decimal places 2 ...
Display range from −20.000 kg 17. In the Decimal places edit field enter the desired number of decimal
to 20.000 kg
places and confirm with .
Step 1
18. In the Step selection field enter the value 2 and confirm with .
Information
The step refers to the last decimal place of the display value.
Example:
Entry 10.0 kg
Step 1 means display jumps every 100g
Step 5 means display jumps every 500g
Entry 10.000 kg
Step 1 means display jumps every 1g
Step 5 means display jumps every 5g
System Display Amplifier Options 20. In the pull-up menu select Analog outputs and confirm with .
21. In the Output Vo1 selection field select the desired signal and confirm
with .
Transducer
Signal conditioning
Display 22. In the Output Vo2 selection field select the desired signal and confirm
Analog outputs
Switch
with .
23. Use to select the Output characteristics pt.2 edit field and enter
the desired value (left for the display, right for the analog output). Con
ANALOG OUTPUTS CHANNEL1 firm with .
Output Vo1: Gross Output Vo2: Net
Output characteristics ppm V
24. Use the Switch key to go to measuring mode and confirm the con
Pt.1: 0.0000 ... 0.0000 ...
Pt.2: 100.0000 ... 0.0000 ... firmation prompt with .
Information
Transducer
Signal conditioning If the zero point and nominal (rated) value are not changed (for example in a
Display
Analog outputs recalibration), you can skip steps 1-10.
Switch
1. Use the channel selection keys or a direct entry to choose the de
calibrate... Adjust amplifier 4. In the pull-up menu select Transducer and confirm with .
Gage factor: 0.0000 ...
9. Use to switch to the Zero point edit field and enter the value 0 in
the left edit field. Confirm with .
10. Use to switch to the Nominal (rated) value edit field and enter the
value 10 in the left edit field. Confirm with .
11. Use to select the Adjust amplifier button and confirm with .
calibrate...
12. Use the cursor keys to select the button and confirm
with .
15. Use the cursor key to select the measure button in line Point 1 and
confirm with .
16. Use the cursor key to select the left edit field in line Point 2, enter
the value 4 and confirm with .
18. Use the cursor keys to select the measure button symbol in line
Point 2. Now if you press , a measurement starts and the current
measured value appears in mv/V on the left next to the measure button.
19. Use the cursor keys to select the OK button and confirm with
(the amplifier converts the nominal (rated) value to 10 kg. The calibra
tion data for 4 kg remains intact).
20. Use the Switch key to go to measuring mode and confirm the con
The resistance of the SG changes due to strain. Since this change lies in
the mΩ and μΩ range, a "Wheatstone bridge" must be used with great
accuracy.
UA
+ k @
UB 4
1 4
UB
UB = Bridge excitation voltage
2 3 UA = bridge output voltage
UA
ML10B + relativechangeoftheexcitationvoltage
UB
ML30B
UA
ML55B =
Total strain
k = strain gage sensitivity (gage factor)
You can use the MGCplus measuring system to measure the strain of one
SG or the total strain of multiple SGs. The table below summarizes the pos
sible bridge circuits and the necessary device configuration:
e2 e3
1) For the combination of ML55B with AP14, a one-time zero calibration must always be performed after setting up the measurement chain.
2) Only in combination with AP01i and AP03i
The measurement system always measures the total strain () of the indi
vidual SGs active in the Wheatstone bridge. Individual SGs are usually used
for mechanical stress analysis in practical applications.
The half bridge circuit and full bridge circuit are used on account of tempera
ture compensation, the larger measurement signal, better cable resistance
compensation and for some stress states (measuring on a bending beam,
etc.). The most important characteristic value of a SG is the gage factor.
The MGCplus measurement system automatically switches to strain mea
surements if a gage factor >0 is entered. The zero point and measuring
range are scaled in m/m. By changing the unit and the display value you
can derive other physical quantity quantities (such as mechanical stress in
N/mm2). If zero is entered as a gage factor, the system switches to mV/V.
1. Use the channel selection keys or a direct entry to choose the de
3. Press .
TRANSDUCER CHANNEL1 4. In the pull-up menu select Transducer and confirm with .
Type SG 1/4 bridge circuit 120ohms 4−wr
5. In the pull‐up menu select Type.
Current feed 2.5V
Unit: % m/m 6. Select type in selection field: SG‐1/4 bridge 120 ohms 4‐wr. In the Cur
Zero pt.: 0.0000 ... 0.0000 ... measure
rent feed selection field choose 2.5 V.
Nom. val.: 100.000 ... 1000.000 ...
calibrate... Adjust amplifier 7. Use to select the Gage factor edit field and enter the value 2.05.
Gage factor: 3 2.0500 ... Press the confirmation button .
Since the amplifier automatically changes the basic unit from mV/V to
m/m after the gage factor is entered, the physical quantity plays no role
in this example. The menu window shown here is based on the display in
the factory settings.
8. Use to switch to the right edit field Zero point and enter the value 0.
Confirm with .
System Display Amplifier Options
9. Use to switch to the right edit field Nominal (rated) value and enter
the value 1000. Confirm with .
Transducer
Signal conditioning Adjust amplifier
Display 10. Use to select the button and confirm with .
Analog outputs
Switch
11. Press .
16. In the Step selection field select the desired step and confirm with .
Information
The step refers to the last decimal place of the display value.
18. Perform a zero balance in unloaded state ( function key in the fac
tory settings).
Inductive transducers (for example force transducer from HBM) are passive
transducers that have the following features:
S They must be supplied with an excitation voltage (carrier frequency)
S They are inductive half or full bridge circuits
S The transducer is identified by the following data:
- Nominal (rated) displacement (for example 20 mm)
- Nominal (rated) sensitivity (for example 10 mV/V)
The standard setting for the excitation voltage UB with inductive transducers
is 2.5 V. If higher measuring ranges are required or multiple transducers are
connected in parallel, a lower excitation voltage must be selected. The indi
vidual values depend on which module is used.
Important
The specified full scale values are maximum values that cannot be affected
by settings of the analog output!
1. Use the channel selection keys or a direct entry to choose the de
Transducer
Signal conditioning
sired channel (for example 3.2 )..
Display
Analog outputs
Switch 2. Use to switch to Setup mode.
TRANSDUCER CHANNEL1 4. In the pull-up menu select Transducer and confirm with .
Gage factor: ✓
0.0000 ... 7. Confirm with .
9. Use to switch to the Zero point edit field and enter the value 0 in
the left edit field. Confirm with .
12. In the left edit field Nominal (rated) value enter the value 20.
13. In the right edit field Nominal (rated) value enter the value 10 (under
System Display Amplifier Options the unit mV/V).
14. Use to select the Adjust amplifier button and confirm with . If
you do not want to make any more changes to the display, you can now
Transducer continue with step 20.
Signal conditioning
Display
Analog outputs 15. Go back to the pull-up menu with .
Switch
DISPLAY CHANNEL1
17. In the Decimal places edit field enter the desired number of decimal
V
places and confirm with .
Unit 3 Abs
Decimal places 3 ...
Display range from −10.000 V 18. In the Step selection field select the desired step and confirm with .
to 10.000 V
Step 100
Information
The step refers to the last decimal place of the display value.
20. Use the Switch key to go to measuring mode and confirm the con
Transducer
Signal conditioning
Display Information
Analog outputs
Switch If the zero point and nominal (rated) sensitivity are not changed (for exam
ple in a recalibration), you can skip steps 1-12.
OK Cancel 8. Use to switch to the Unit selection field and press . Select mm
as the unit and confirm with .
9. Use to switch to the Zero point edit field and enter the value 0 in
the left edit field. Confirm with .
10. Use to switch to the measure button and confirm with (the mea
sured value appears in the right edit field of Zero point)
System Display Amplifier Options
11. In the left edit field Nominal (rated) value enter the value 20 (under the
unit mm) and confirm with .
18. Position the gaging block under the probe tip of the displacement trans
Characteristic curve points ducer.
Unit mm mV/V Shunt off
19. Use the cursor keys to select the measure button symbol in line
Point 1: 0.0000 0.0021 measure
Point 2 10.000 5.0018 measure Point 2. Now if you press , a measurement starts and the current
OK Cancel measured value appears in mV/V on the left next to the button measure
20. Use to select the OK button and confirm with (the amplifier con
verts the nominal (rated) value to 20 mm. The calibration data for 10
System Display Amplifier Options mm remains intact).
Transducer 22. In the pull-up menu select Analog outputs and confirm with .
Signal conditioning
Display
Analog outputs 23. In the Output Vo1 selection field select the gross signal and confirm
Switch
with .
24. In the Output Vo2 selection field select the desired signal and confirm
ANALOG OUTPUTS CHANNEL1 with .
Output Vo1: Gross Output Vo2: Net
Output characteristics ppm V
25. Use to select the Output characteristics pt.2 edit field and enter
Pt.1: 0.0000 ... 0.0000 ... the desired value (left 15 for the display, right 10 for the analog output).
Pt.2: 100.0000 ... 0.0000 ...
Confirm with .
26. Use the Switch key to go to measuring mode and confirm the con
Freq.0..2kHz
Freq.0..20kHz
Freq.0..200kHz
Freq.0..500kHz
Pulse counter
PWM
Duration
Duration
Only the absolute pulse duration (THigh) is evaluated for the pulse duration.
The measuring range is 0 ... 2500 ms.
The following example explains the operating steps for adjusting the torque,
rotational speed and power channel. Explanation of the extended functions
can be found on è page 158.
System Display Amplifier Options
Nominal (rated) torque of the shaft: 50Nm
Nominal (rated) sensitivity: 5kHz
Calibration value: 24.22Nm
Transducer Nominal (rated) torque: 3000rpm
Signal conditioning
Display Rotational speed range to be displayed
Analog outputs
Switch for the application: 2000rpm
Display range for power: 10kW
6. Use to switch to the Level selection field, press and select 5V.
Confirm with .
7. Use to switch to the Unit selection field and press . Select NVm
as the unit and confirm with .
Decimal places 3 ... 15. In the Decimal places edit field enter the desired number of decimal
Display range from −50.000 Nm
to 50.000 Nm places and confirm with .
Step 1
16. In the Step selection field select the desired step and confirm with .
17. Use the Switch key to go to measuring mode and confirm the con
Information
The step refers to the last decimal place of the display value.
Example:
Entry 50.0 Nm
Step 1 means display jumps every 0.1Nm
Step 5 means display jumps every 0.5Nm
Entry 50.000 Nm
Step 1 means display jumps every 0.001Nm
Step 5 means display jumps every 0.005Nm
Information
System Display Amplifier Options
If the zero point and nominal (rated) torque are not changed (for example in
a recalibration), you can skip steps 1.-9.
Transducer
Signal conditioning 1. Use or a direct entry to choose the desired channel (for example 3.2
Display
Analog outputs
Switch ).
Characteristic curve points 7. Use to switch to the Unit selection field and press . Select NVm
Unit: Nm kHz Shunt on
as the unit and confirm with .
Point 1: 0.0000 10.056 measure
Point 2 24.220 2.4210 measure 8. In the left edit field Nominal (rated) value enter the value 50.
OK Cancel
Transducer 12. In the left edit field for characteristic curve point 1 enter the value 0 and
Signal conditioning
Display
confirm with .
Analog outputs
Switch
13. Use the cursor key to select the measure button in line "Point 1" and
confirm with .
TRANSDUCER CHANNEL1 14. Use to switch to the Shunt off selection field, press and select
Type Freq.0..20kHz Shunt on. Confirm with .
Level: 5V Ext.fct...
Unit: Nm kHz
0.0000 ... 10.0000 ... measure
15. Use to select the measure button symbol in line Point 2. Now if you
Zero pt.:
Nom. val.: 50.0000 ... 5.0000 ...
press a measurement starts and the current measured value ap
calibrate... Adjust amplifier
pears in kHz on the left next to the measure button symbol.
16. Use to select the left edit field in line Point 2. Enter the calibration
value 24.22 and confirm with
Characteristic curve points 17. Use to select the OK button symbol and confirm with (the am
Unit: Nm kHz Shunt on plifier converts the nominal (rated) value to 50 Nm. The calibration data
Point 1: 0.0000 0.0000 measure for 25 Nm remains intact).
Point 2 24.220 2.4210 measure
18. In the pull-up menu select Display and confirm with .
OK Cancel
19. In the Decimal places edit field enter the desired number of decimal
places and confirm with .
20. In the Step selection field select the desired step and confirm with .
Information
The step refers to the last decimal place of the display value.
Example:
Entry 50.0 Nm
Step 1 means display jumps every 0.1Nm
Step 5 means display jumps every 0.5Nm
Entry 50.000 Nm
Step 1 means display jumps every 0.001Nm
Step 5 means display jumps every 0.005Nm
21. Use the Switch key to go to measuring mode and confirm the con
Transducer
Signal conditioning Switch output LV1 (ML60B only)
Display
Analog outputs The frequency signal F1 or the counting signal can be applied to the switch
Switch output of limit switch 1 (see illustration).
Type: Freq.0..20kHz
Level: 5V Ext.fct...
Unit: Nm kHz
F1
Zero pt.: 0.0000 ... 10.0000 ... measure
Nom. val.: 50.0000 ... 5.0000 ...
1.6s
calibrate... Adjust amplifier
Enhanced functions
Glitch filter (min. pulse width 1.6s)
Frequency quadrupling
Analyze F2 signal Counting signal
Zero index input active on LV1
Transducer error input active
From
F2 to LV2
direction of rotation to LV2
From
F1 to LV1
F1 counting signal to LV1
1. Use or a direct entry to choose the desired channel (for example 3.2
TRANSDUCER CHANNEL1
6. Use to switch to the Level selection field, press and select 5V.
Type: Freq.0..200kHz
Level: 5V Ext.fct... Confirm with .
Unit: rpm kHz
Zero pt.: 0.0000 ... 0.0000 ... measure
Nom. val.: 3000.0 ... 18,000 ...
11. Use to select the Adjust amplifier button and confirm with . If
you do not want to make any more changes to the display, you can now
continue with step 16..
14. In the Decimal places edit field enter the desired number of decimal
places and confirm with .
15. In the Step selection field select the desired step and confirm with .
Information
The step refers to the last decimal place of the display value.
Example:
Entry 1000.0 rpm
Step 1 means display jumps every 0.1rpm
Step 5 means display jumps every 0.5rpm
Entry 1000.000 rpm
Step 1 means display jumps every 0.001rpm
Step 5 means display jumps every 0.005rpm
16. Use the Switch key to go to measuring mode and confirm the con
1. Use or a direct entry to choose the desired channel (for example 3.2
TRANSDUCER CHANNEL1
6. Use to switch to the Unit selection field and press . Select kW
Type 10V
Current feed None as the unit and confirm with .
Unit: kW Vi
Zero pt.: 0.0000 ... 0.0000 ... measure 7. In the left edit field Nominal (rated) value enter the value 10.472.
Nom. val.: 10.4720 ... 10.0000 ...
8. In the right edit field Nominal (rated) value enter the value 10.
calibrate... Adjust amplifier
12. In the Decimal places edit field enter the desired number of decimal
places and confirm with .
System Display Amplifier Options
13. In the Step selection field select the desired step and confirm with .
Transducer
Signal conditioning Information
Display
Analog outputs The step refers to the last decimal place of the display value.
Switch
Example:
Entry 10.0 kW
ANALOG OUTPUTS CHANNEL1 Step 1 means display jumps every 0.1 kW
Step 5 means display jumps every 0.5 kW
Output Vo1: Gross Output Vo2: Net
Output characteristics ppm V Entry 10.000 kW
Pt.1: 0.0000 ... 0.0000 ... Step 1 means display jumps every 0.001 kW
Pt.2: 100.0000 ... 0.0000 ...
Step 5 means display jumps every 0.005 kW
15. In the pull-up menu select Analog outputs and confirm with .
16. In the Output Vo1 selection field select the desired signal and confirm
with .
17. In the Output Vo2 selection field select the desired signal and confirm
with .
18. Use to select the Output characteristics pt.2 edit field and enter
the desired value 5 (left for the display, right for the analog output). Con
firm with .
19. Use the Switch key to go to measuring mode and confirm the con
7.3.7 Thermocouples
1. Use or a direct entry to choose the desired channel (for example 3.2
TRANSDUCER CHANNEL1
6. Use to select the Adjust amplifier button and confirm with . If
Type Thermocpl.K
you do not want to make any more changes to the display, you can now
Current feed None
Unit: C C
continue with step 11.
Zero pt.: 50.0000 ... 50.0000 ... measure
Nom. val.: 20.0000 ... 20.0000 ... 7. Go back to the pull-up menu with .
calibrate... Adjust amplifier
8. In the pull-up menu select Display and confirm with .
9. In the Decimal places edit field enter the desired number of decimal
places and confirm with .
10. In the Step selection field select the desired step and confirm with .
Information
The step refers to the last decimal place of the display value.
Example:
Entry 10.0C
Step 1 means display jumps every 0.1C
Step 5 means display jumps every 0.5C
Entry 50.000 C
Step 1 means display jumps every 0.001C
Step 5 means display jumps every 0.005C
11. Use the Switch key to go to measuring mode and confirm the con
To measure current and voltage signals you will need amplifier module
ML01B or ML801B and AP402i. For simple voltage measurements you can
also use the ML801B multi-channel module with connection board AP801 or
AP801S6.
The settings are explained with the following example:
A torque transducer with integrated amplifier returns a maximum output
signal of 3V, corresponding to a nominal (rated) torque of 20 Nm. The
display range will be set to 20.000 Nm. An output signal of 10V is required
for control.
1. Use or a direct entry to choose the desired channel (for example 3.2
TRANSDUCER CHANNEL1
6. Use to switch to the Unit selection field and press . Select Nm
Type 10V
Current feed None
as the unit and confirm with .
Unit: Nm Vi
Zero pt.: 0.00000 ... 0.0000 ... measure 7. In the left edit field Zero point enter the value 0.
Nom. val.: 20.0000 ... 3.0000 ...
8. In the right edit field Zero point enter the value 0.
calibrate... Adjust amplifier
9. In the left edit field Nominal (rated) value enter the value 20.
10. In the right edit field Nominal (rated) value enter the value 3.
11. Use to select the Adjust amplifier button and confirm with . If
you do not want to make any more changes to the display, you can now
continue with step 16.
14. In the Decimal places edit field enter the desired number of decimal
places and confirm with .
15. In the Step selection field select the desired step and confirm with .
Information
The step refers to the last decimal place of the display value.
Transducer
Signal conditioning
Display Example:
Analog outputs
Switch Entry 10.0 Nm
Step 1 means display jumps every 0.1 N@m
Step 5 means display jumps every 0.5 N@m
Entry 10.000 Nm
DISPLAY CHANNEL1 Step 1 means display jumps every 0.001 N@m
Unit V 3 Abs Step 5 means display jumps every 0.005 N@m
Decimal places 3 ...
Display range from −10.000 V 16. Use the Switch key to go to measuring mode and confirm the con
to 10.000 V
1) The permitted ranges depend on the bridge excitation voltage. If you set a measuring range that lies outside of of the adjustment range of
the amplifier, the maximum or minimum value will be applied.
1. Use or a direct entry to choose the desired channel (for example 3.2
TRANSDUCER CHANNEL1 6. In the Connector selection field choose the desired connection type.1)
Type: Pt100
Connection: 4−wire
7. Use to switch to the Unit selection field and press . Select °C as
Unit: C C
50.0000 ... 50.0000 ... measure
the unit and confirm with .
Zero pt.:
Nom. val.: 20.0000 ... 20.0000 ...
8. In the left edit field Zero point enter the value 50.
calibrate... Adjust amplifier
9. In the left edit field Nominal (rated) value enter the value 20
(70°C-50°C=20°C).
10. Use to select the Adjust amplifier button and confirm with . If
you do not want to make any more changes to the display, you can now
continue with step 15.
13. In the Decimal places edit field enter the desired number of decimal
places and confirm with .
System Display Amplifier Options
14. In the Step selection field select the desired step and confirm with .
Transducer
Signal conditioning Information
Display
Analog outputs The step refers to the last decimal place of the display value.
Switch
Example:
Entry 10.0 C
DISPLAY CHANNEL1 Step 1 means display jumps every 0.1 C
Step 5 means display jumps every 0.5 C
Unit V 3 Abs
Decimal places 3 ... Entry 50.000 C
Display range from −10.000 V Step 1 means display jumps every 0.001 C
to 10.000 V
100
Step 5 means display jumps every 0.005 C
Step
15. Use the Switch key to go to measuring mode and confirm the con
7.3.10 Resistors
Resistors are passive transducers. For these transducers you will need the
ML801B amplifier module with AP835, which will provide you with a coarse
measuring range (0...500 Ω.
The settings are explained with the following example:
Resistance value 400 ΩDisplay full scale 400.00 Ω
1. Use or a direct entry to choose the desired channel (for example 3.2
5. Use to select 500ohms (this is the input range of the amplifier) and
TRANSDUCER CHANNEL1
confirm with .
Type: 500 ohms 6. In the Connector selection field choose the desired connection type.1)
Connection: 4−wire
Unit: ohms ohms
7. Use to switch to the Unit selection field and press . Select
Zero pt.: 0.00000 ... 0.00000 ... measure
Nom. val.: 400.0000 ... 400.0000 ... ohms as the unit and confirm with .
calibrate... Adjust amplifier
8. In both the Zero point edit fields enter the value 0.
9. In the left edit field Nominal (rated) value enter the value 400.
10. Use to select the Adjust amplifier button and confirm with . If
you do not want to make any more changes to the display, you can now
continue with step 15.
13. In the Decimal places edit field enter the desired number of decimal
System Display Amplifier Options
places and confirm with .
14. In the Step selection field select the desired step and confirm with .
Transducer
Signal conditioning
Display
Analog outputs
Switch Information
The step refers to the last decimal place of the display value.
Example:
DISPLAY CHANNEL1
V
Entry 10.0
Unit 3 Abs
Step 1 means display jumps every 0.1
Decimal places 3 ...
Display range from −10.000 V
Step 5 means display jumps every 0.5
to 10.000 V
100
Entry 10.000
Step
Step 1 means display jumps every 0.001
Step 5 means display jumps every 0.005
with .
To count pulses you will need the ML60B or ML460 amplifier. This amplifier
is able to process a maximum pulse string frequency of 1 MHz. Further in
formation about possible settings of this amplifier can be found in è section
7.3.4 "Torque transducer", page 158.
Incremental transmitters (used for example to measure angles) return two
square wave signals which are phase-shifted by 90. The duty ratio between
the High and Low of the two signals must be about 1:1.
The settings are explained with the following example:
An angle of rotation transmitter returns two signals, each with 180 pulses/
revolution. They will be realized in the display as 360. Due to the pulse
number per revolution, a value of 1 is specified as the step, since a higher
revolution is not practical. The level of the square wave signal is 10 V.
In addition to the measured value, a frequency signal with four times the
count pulses is required at the output.
1. Use or a direct entry to choose the desired channel (for example 3.2
Type: Freq.0..20kHz 6. Use to switch to the Level selection field, press and select 5V.
Level: 5V Ext.fct...
Unit: deg kImp
Confirm with .
Zero pt.: 0.0000 ... 0.0000 ... measure
Nom. val.: 360.0000 ... 0.1800 ... Ext. fct. ...
7. Use to select the button and confirm with (for
calibrate... Adjust amplifier
explanations of extended functions).è see page 180)
Enhanced functions
8. Use to select the activation fields Frequency quadrupling and An
Glitch filter (min. pulse width 1.6s)
Frequency quadrupling
Analyze F2 signal alyze F2 signal and activate them with .
Zero index input active
Transducer error input active
9. Use to switch to the Switch output LV1 selection field, select F1
Switch output LV1: F1 counting signal to LV1
Counting signal to LV1 and confirm with .
Switch output LV2: Off
OK Cancel OK
10. Use to select the button and confirm with .
11. Use to switch to the Unit selection field and press . Select deg
as the unit and confirm with .
12. In both the Zero point edit fields enter the value 0.
13. In the left edit field Nominal (rated) value enter the value 360.
14. In the right edit field Nominal (rated) value enter the value 0.180.
15. Use to select the Adjust amplifier button and confirm with . If
you do not want to make any more changes to the display, you can now
continue with step 20.
System Display Amplifier Options 17. In the pull-up menu select Display and confirm with .
18. In the Decimal places edit field enter the desired number of decimal
places and confirm with .
Transducer
Signal conditioning
Display
19. In the Step selection field choose 2 (1 pulse is equivalent to 2°; 180
Analog outputs pulses/revolution is equivalent to 360°) and confirm with .
Switch
with .
DISPLAY CHANNEL1
Unit V 3 Abs Assign the Zero balance function to a free F-key (factory settings level
Decimal places 3 ... 1/F1). Then return to measuring mode. If you press the corresponding
Display range from −10.000 V F‐key, the pulse counter is set to "0". è See also section "Assigning
to 10.000 V
100
F‐keys"; page 232.
Step
Transducer
Signal conditioning Frequency quadrupling
Display
Analog outputs Frequency quadrupling is equivalent to increasing the signal resolution.
Switch Both the positive and negative signal edges are counted. If both frequency
inputs (F1 and F2) are connected, the measurement frequency is quadru
pled. If F2 is not connected the measurement frequency is doubled.
TRANSDUCER CHANNEL1
Analyze F2 signal (direction detection)
Type: Freq.0..20kHz
Level: 5V Ext.fct... The phase-offset signal F2 is analyzed, which makes it possible to display
Unit: deg kImp the direction of rotation or the direction of motion.
Zero pt.: 0.0000 ... 0.0000 ... measure
Nom. val.: 360.0000 ... 0.1800 ...
Zero index input active
calibrate... Adjust amplifier
For incremental encoders this input is used in counting mode to reset the
Enhanced functions
counter.
Glitch filter (min. pulse width 1.6s)
Frequency quadrupling
Analyze F2 signal
Zero index input active
Transducer error input active (only with AP01i)
Transducer error input active
If a signal is present (level 0V), the measured value is interpreted by the
Switch output LV1: Off amplifier as an error (for example when the light source fails in optical sys
Switch output LV2: Off tems).
OK Cancel
Switch output LV1 (only with ML60B)
From
F2 to LV2 The frequency signal F1 or the counting signal can be applied to the switch
direction of rotation to LV2
output of limit switch 1.
From
F1 to LV1
F1 counting signal to LV1
Switch output LV2 (only with ML60B)
The frequency signal F2 or the direction of rotation signal can be applied to
the switch output of limit switch 2.
F1
F2
Counting signal to
LV1 with quadruple
analysis activated
Piezoelectric transducers with a built-in preamplifier are often used for ac
celeration and force measurements. These transducers require a constant
current for supply. The measurement signal corresponds to the modulated
voltage on the power line. The DeltaTronT acceleration transducer from
Brüel&Kjaer is a typical representative of this product family.
To operate current-fed piezoelectric transducers you will need the AP418i
connection board and the ML810B amplifier. You can select coarse and fine
adjustment for the measuring range. 3 input ranges are available for coarse
adjustment: "0.05 V; "0.5 V; "5 V.
Example: If the required measuring range is 4 V, the next larger input range
"5 V is set.
The zero point defines the 0 V voltage generated at the output of the ampli
fier. The nominal (rated) value is always defined relative to the zero point.
The settings are explained with the following example:
A DeltaTronT transducer will be used to measure an acceleration of up to
300m/s 2. The nominal (rated) value of the transducer is 1mV/ m/s2.
Measuring range (fine adjustment) 300 m/s2 x 1 mV/m/s2 = 0.3 V
Input range (coarse adjustment): "0.5 V ( > 0.3 V)
1. Use or a direct entry to choose the desired channel (for example 3.2
Type Deltatron 1V 6. Use to switch to the Unit selection field and press . Select m/s2
Current feed: 2−20mA
Unit: m/s2 Vi as the unit and confirm with .
Zero pt.: 0.00000 ... 0.00000 ... measure
Nom. val.: 300.0000 ... 0.30000 ... 7. In both the Zero point edit fields enter the value 0.
calibrate... Adjust amplifier 8. In the left edit field Nominal (rated) value enter the value 300.
9. In the right edit field Nominal (rated) value enter the value 0.3.
10. Use to select the Adjust amplifier button and confirm with . If
you do not want to make any more changes to the display or output
characteristics, you can now continue with step 15.
13. In the Decimal places edit field enter the desired number of decimal
System Display Amplifier Options
places and confirm with .
14. In the Step selection field select the desired step and confirm with .
Transducer
Signal conditioning
Display
Analog outputs
Switch Information
The step refers to the last decimal place of the display value.
Example:
DISPLAY CHANNEL1
V
Entry 10.0m/s2
Unit 3 Abs
Step 1 means display jumps every 0.1 m/s2
Decimal places 3 ...
Display range from −10.000 V
Step 5 means display jumps every 0.5 m/s2
to 10.000 V
100
Entry 10.000m/s2
Step
Step 1 means display jumps every 0.001 m/s2
Step 5 means display jumps every 0.005 m/s2
with .
1. Use or a direct entry to choose the desired channel (for example 3.2
5. Use to select transducer type Full bridge circuit high level and
confirm with .
TRANSDUCER CHANNEL1
10. In the right edit field Nominal (rated) value enter the value 200.
11. Use to select the Adjust amplifier button and confirm with . If
you do not want to make any more changes to the display or output
characteristics, you can now continue with step 16.
System Display Amplifier Options 13. In the pull-up menu select Display and confirm with .
14. In the Decimal places edit field enter the desired number of decimal
places and confirm with .
Transducer
Signal conditioning
Display 15. In the Step selection field select the desired step and confirm with .
Analog outputs
Switch
16. Use to go to measuring mode and confirm the confirmation prompt
with .
DISPLAY CHANNEL1
Step 100
Example:
Entry 10.0bar
Step 1 means display jumps every 0.1 bar
Step 5 means display jumps every 0.5 bar
Entry 10.000bar
Step 1 means display jumps every 0.001 bar
Step 5 means display jumps every 0.005 bar
Information
If the zero point and nominal (rated) value are not changed (for example in a
Transducer
Signal conditioning
recalibration), you can skip steps 1-10.
Display
Analog outputs
Switch
1. Use or a direct entry to choose the desired channel (for example 3.2
).
TRANSDUCER CHANNEL1
2. Use to switch to Setup mode.
Type Full bridge circuit or high level
Current feed 5V
bar
3. Press .
Unit: mV/V
Zero pt.:
0.0000 ... 0.0000 ... measure
Nom. val.: 300.0000 ... 200.000 ... 4. In the pull-up menu select Transducer and confirm with .
calibrate... Adjust amplifier
0.0000 ... 5. Use to switch to the Current feed selection field, press and
Gage factor:
select 10V.
6. Confirm with .
Characteristic curve points 7. Use to switch to the Unit selection field and press . Select bar
Unit: bar mV/V Shunt off
as the unit and confirm with .
Point 1: 0.0000 0.0021 measure
9. Use to switch to the Nominal (rated) value edit field and enter the
value 300 in the left edit field. Confirm with .
10. Use to select the Adjust amplifier button and confirm with .
calibrate...
System Display Amplifier Options 11. Use the cursor keys to select the button and confirm
with .
Transducer
12. Unload the transducer.
Signal conditioning
Display 13. In the left edit field for characteristic curve point 1 enter the value 0 and
Analog outputs
Switch confirm with .
14. Use to select the measure button in line Point 1 and confirm with
.
TRANSDUCER CHANNEL1
15. Use to select the left edit field in line Point 2, enter the value 250
Type Full bridge circuit or high level
Current feed 5V and confirm with .
Unit: bar mV/V
Zero pt.: 0.0000 ... 0.0000 ... measure 16. Apply the test pressure to the transducer.
Nom. val.: 10.0000 ... 2.0000 ...
calibrate... Adjust amplifier 17. Use the cursor keys to select the measure button in line Point 2. Now
Gage factor: 0.0000 ... if you press , a measurement starts and the current measured value
appears in mv/V on the left next to the measure button.
18. Use to select the OK button and confirm with (the amplifier
Characteristic curve points converts the nominal (rated) value to 300 bar. The calibration data for
Unit: bar mV/V Shunt off 250 bar remains intact).
Point 1: 0.0000 0.0021 measure 19. Use to go to measuring mode and confirm the confirmation prompt
Point 2 250.000 166.67 measure
OK Cancel
with .
1. Use or a direct entry to choose the desired channel (for example 3.2
5. Use to select transducer type Half bridge circuit high level and
confirm with .
TRANSDUCER CHANNEL1
11. Use to select the Adjust amplifier button and confirm with . If
you do not want to make any more changes to the display or output
characteristics, you can now continue with step 16.
14. In the Decimal places edit field enter the desired number of decimal
places and confirm with .
15. In the Step selection field select the desired step and confirm with .
with .
Information
The step refers to the last decimal place of the display value.
Example:
Entry 10.0mm
Step 1 means display jumps every 0.1 mm
Step 5 means display jumps every 0.5 mm
Entry 10.000mm
Step 1 means display jumps every 0.001 mm
Step 5 means display jumps every 0.005 mm
Information
If the zero point and nominal (rated) value are not changed (for example in a
Transducer
Signal conditioning
recalibration), you can skip steps 1-10.
Display
Analog outputs
Switch
1. Use or a direct entry to choose the desired channel (for example 3.2
).
TRANSDUCER CHANNEL1
2. Use to switch to Setup mode.
Type Half bridge circuit or high level
Current feed 2.5V
Unit: mm mV/V
3. Press .
Zero pt.: 0.0000 ... 0.0000 ... measure
Nom. val.: 10.0000 ... 1000.000 ... 4. In the pull-up menu select Transducer and confirm with .
calibrate... Adjust amplifier
Gage factor: 0.0000 ... 5. Use to switch to the Current feed selection field, press and
select 2.5V.
6. Confirm with .
Characteristic curve points 7. Use to switch to the Unit selection field and press . Select mm
Unit: mm mV/V Shunt off
as the unit and confirm with .
Point 1: 0.0000 0.0021 measure
9. Use to switch to the Nominal (rated) value edit field and enter the
value 10 in the left edit field. Confirm with .
10. Use to switch to the right edit field Nominal (rated) value and enter
the value 1000. Confirm with .
System Display Amplifier Options
calibrate...
11. Use to select the button and confirm with .
Transducer
12. Slide the core into the transducer until the display indicates zero.
Signal conditioning
Display 13. In the left edit field for characteristic curve point 1 enter the value 0 and
Analog outputs
Switch confirm with .
14. Use to select the measure button in line Point 1 and confirm with
.
TRANSDUCER CHANNEL1
15. Use the cursor key to select the left edit field in line Point 2, enter
Type Half bridge circuit or high level
Current feed 2.5V the value 7 and confirm with .
Unit: mm mV/V
Zero pt.: 0.0000 ... 0.0000 ... measure 16. Position the gaging block under the probe tip of the displacement trans
Nom. val.: 10.0000 ... 1000.000 ...
ducer.
calibrate... Adjust amplifier
17. Use to select the measure button in line Point 2. Now if you press
Gage factor: 0.0000 ...
8 Additional functions
Control inputs The remote control contacts are used to control important functions of the
amplifiers via digital inputs. These contacts are only active if remote control
is turned on
Control inputs Channel 1 There are three ways to turn remote control on or off.
On
Remote control On
Input 1 Not assigned
Off S In measuring mode: Press the function key twice and then the
Input 2 Not assigned function key (factory settings).
Input 3 Not assigned
Not assigned
Input 4
Not assigned
S In setup mode: Press the function key. You can turn remote control
Input 5
Input 6 Not assigned on or off in the "Switch" setup window.
Input 7 Not assigned
Input 8 Not assigned S Via the remote control contacts (with the REMT function).
Not assigned
ACAL-autocal
TARA-tare
Delete CPV1-PV1
HLD1-hold memory1
Delete CPV2-PV2
HLD2-hold memory2
ZERO-Zero balance
SHNT-Shunt On/Off
CAL input: Calibration signal
ZERO input Zero signal
INV sign reversal
PSEL1-P-set coding line 1
PSEL2-P-set coding line 2
PSEL4-P-set coding line 4
REMT-Remote control On/Off
INT-Start/stop integration
The assignment of control contacts is freely selectable. They are not as
signed in the factory settings. The table below shows the possible functions
with a description of the function.
Tab. 8.1
The limit switches are not enabled in the factory settings (Enable "Off").
2. Print
Limit switches 3. In the pop-up menu select Limit switches and confirm with .
Input Free
Message when On 1-LV1 ON... Normal
Enabled
Disabled Positive logic
Negative logic
Exceed
Undershoot
Normal
Reverse video
Gross
Net
PV1
PV2
PV12
No
Yes
Level
Limit value delay time
12 3 4 Enter the response level in display units (kg, etc.).
Limit value On Delay time
Limit value Off 0ms Hysteresis
The Hysteresis is the modified onset of the switching effect between the
OK
OK Cancel
OK states "On" and "Off". The hysteresis prevents the limit switch from "flicker
ing" when the switching level is reached.
Direction
Enter the switching direction or the working direction of the limit switch.
Switching level
Hysteresis
Hysteresis
Switching level
30ms
Switching level
Hysteresis
1
0
60ms
Delay time
The limit value should only switch if the signal is persistently above the
switching level for an extended time (in this case 60ms). The limit switch
should not engage if the signal is only present above the switching level
briefly (in this case 30ms).
Output logic
You can change the output logic of the remote controls as required.
Input
Enable or disable entry of the limit value level via function keys.
Message when On
Edit field for a message in the display value when switching on (for example
"Below 20 kg", è see also illustration on page 201). You can also select the
type of realization (normal=black print and white background; Reverse
video=Light print and dark background).
Information
Limit switch 1 Channel 1
Limit switches 1 .... Using the horizontal cursor keys for a function enable is more efficient, for
Name 1-LV1 ... example. As soon as you are in the selection field (edit field) you want (in
Enable No
Input signal Gross
this case Enable) and you have made your setting with , press .
Activation level 10.000 ... V Then you will remain in the selection field (edit field) you want, but you will
Deactivation level 9.9000 ... V
move to the next limit switch.
Input Disabled
Hysteresis Variable
(edit field)
You can use this function to logically combine selected limit switches with a
limit value output.
System Display Amplifier Options
limit value output
Entry for the desired limit value output (---, 1, 2, 3, 4).
3
Input signal 1...4
Limit value output
AND
Combination: Select the inputs (LV1...LV4 or control input) to be combined.
Limit value 1
Input signal 1:
Limit value 2
Input signal 2:
−−−
Input signal 3:
−−−
Input signal 4:
----
Limit value 4
Limit value 4 inverted
Control input 4
Control input 4 inverted
AND
OR
EXOR
NAND
NOR
NEXOR
−−−
1
2
3
4
Input Free 1
Below 20kN ... Normal
LV1
Message when On 0
Reverse
Message when Off Over 20kN... video
1
LV2
Limit switch 3 Channel 1 0
Limit switches 3
1
Name 1−LV3 LV3
Yes
0
Enable
Input signal Gross
Level 0.000 ... kN
Hysteresis 0.010 ... kN
Direction Exceed
Output logic Positive logic
Delay define...
Input Free
Message when On IO Normal
Reverse
Message when Off NOK video
Measured peak
value 8.4.1 Peak-value memory
The "Peak values" function can be used to record and save isolated signal
peaks and minimum/maximum signal amplitudes. Each amplifier contains
two peak-value memories.
You can use these to save:
26 μsec
S Maxima
S Minima or
S Peak-to-peak amplitudes
Gross
Peak value memory Minimum (envelope curve) Display
Maximum peak-peak
With fast dynamic signals, you must take into account that the peak values
are determined in a fixed time grid. The grid is sampled at set filter frequen
cies > 5 Hz Bessel 10 Hz Butterworth 38,400 times per second, equivalent
to 26 μsec.
Due to the abbreviated notation, the peak-value memories in the setup win
dows are referred to as PV1 and PV2.
Subtraction
Off
On Minimum, gross PV2
Maximum gross
---
Maximum net
Difference PV1,PV2
Minimum gross Fig. 8.3 Combining peak values with subtraction
Mean value
Minimum net
Integrate gross
Integrate net
Mean-value calculation
( PV1)PV2 )
2
System Display Amplifier Options
F
PV1
Peak-value memory
Mean value
PV2
Peak-value memory Channel 1
Integrate gross
G/n g = Gross signal
Off
n = number of points
On
Maximum gross
---
Difference PV1,PV2
Maximum net Integrate net
Minimum gross
Mean value
Integrate gross
Minimum net N/n N = Net signal
Integrate net n = number of points
The mean value of the gross/net signal is calculated over a freely selectable
time interval. The values are totalized at a sampling rate of 1200 Hz (for fil
ter frequencies >5 Hz: Bessel; >10 Hz: Butterworth).
The beginning/end of integration can be set by assigning an F‐key ("Start/
Stop integration function"; è see page 234) or with a remote control contact
(INT).
Due to the abbreviated notation, the peak-value memories in the setup win
dows are referred to as PV1 and PV2.
System Display Amplifier Options
2. Use the remote control contacts CPV1/CPV2 (in pull‐up menu Ampli
fier, Setup window Switch to set selection field Remote control ON).
Peak-value memory Channel 1
Memory 1 Function Maximum gross 3. Via a computer using the CPV computer.
Memory 1 Envelope curve Off 0 ... ms
Memory 2 Function Maximum gross
Memory 2 Envelope curve Off 0 ... ms
PV 12-combination −−−
Off
On
Maximum gross
---
Maximum net
Difference PV1,PV2
Minimum gross
Mean value
Minimum net
Integrate gross
Integrate net
8.5 Version
You can use the channel selection keys to show information about the ver
sion of your device components one after the other. The name of the device
component (AB, CP, channel1, etc.) whose version will be displayed ap
pears at the right of the header. In the field underneath, the first line shows
System Display Amplifier Options
details on device identification while the second line is available for user
comments.
The Serial/revision number button is used to open a new information win
dow.
Version
Version CHANNEL 1
Serial/revision number
Serial.no. Rev.no.
ML801B xxxx xxxx
8.6 Switching
The Switching setup window contains first of all three buttons you can use to
->T<-
Tare
System Display Amplifier Options
->0<-
Perform a zero balance
->T<- ->0<-
Autocal
Autocal Off Remote control Off
Param.set 1-internal
Switches automatic autocalibration on and off. You can use AUTOCAL On
Ampl. input Zero LED display Status
to improve the temperature response of the zero point and the long-term
consistency of the amplifiers. If you need the analog output signal for contin
uous monitoring, you must turn off autocalibration. This is because no mea
sured values are acquired during calibration. Therefore there is a "gap"
(lasting for approximately 1 second at intervals of about 5 minutes, depend
ing on the filter setting) in the output of measured values.
Parameter set
Selection of saved parameter sets (è see also page 243).
Ampl. input
Zero: Zero signal; input internally at zero potential
Cal: Calibration signal
Measure: The current measurement signal is present at the input
Shunt: Shunt resistor turned on (shunt calibration)
SWITCH CHANNEL 1
Remote control
->T<- ->0<-
Turning remote control on/off (activates remote control contacts)
Autocal Off Remote control Off
Param.set 1-internal
Ampl. input Zero LED display Status
LED display
Switches the function of the LEDs on the front panel of the amplifier.
Status
Off
Zero 1-internal Status display (amplifier active, error, limit values)
5 Min
Cal 2-internal
one time
Measure 3-internal Status Level
Level Degree of control exerted by the amplifier
4-internal
5-internal
6-internal Off
7-internal On
8-internal
9 Display
1 measured value
Gross 9.1 Display format
Channel
Measure The settings affect how the selectable signals in the display are realized.
Measure
Normally you can select four different signals (gross, net, limit values and
3 measured values peak values) for each channel.
Gross
Gross
The measured values can be realized as numeric values or as a graph. A
Gross maximum of six measured values can appear at the same time in the nu
Measure
meric value realization.
6 measured values
Gross Gross The display states shown on the left, referred to as Image types, can be se
Gross Gross lected in the setting menu.
Gross Gross
x-y realization
Realization as a graph
S y‐t realization
S x‐y realization
Limit value status
Display of four limit value states
LIMIT VALUE
1−ML30
1−LV1
Gross
Below 10kN
8.483 kN
S Limit value status
1−LV2 Below 20kN
1−LV3 NOK
Measure Display with image type “Recording"
Recording status S Pre‐trigger status
Test series 3 SAMPLING RATE: 50Hz
TIME: 00: 00: 00 PERIODS: 100 S Post‐trigger status
1
−4.0S 16.0S
FILE NAME:MGCP0000.MEA 120 MB FREE
Load AP1 Acal ... ...
Display format
You will then be in the Display format setup window.
F−keys
Channel names
Image type
With this image type you define the number of measurement signals that
can be realized simultaneously (for numeric values only) or the type of real
ization (graph only). You can also display the status of the four selected limit
switches.
Absolute
The measured value and the signal of the selected channel are displayed
independently of the basic value. The number that is entered corresponds to
the actual channel number. Switching channels in measuring mode has no
effect on this display value.
Important
This values does not correspond to the actual channel number!
Setup window
DISPLAY FORMAT
Image no.: 0 ...
CHANNEL/SIGNAL SELECTION
Image type: 3 measured values
OK Cancel All channels All signals
Display value 1 (basic value):
C1 C2 C3 C4 C5 C6
MGCplus device
(−) (+)
Display −1
+2
Channel Value 1 Basic value
Channel Gross Value 2
Channel Gross
Value 3
Display
Basic value K3 Value 1 K4 Value 1 K5 Value 1
Value 2: Channel+2 relative K5 Value 2 K6 Value 2 K1 Value 2
Value 3: Channel-1 relative K2 Value 3 K3 Value 3 K4 Value 3
When the cursor keys (SIGNAL) are activated, all signals of the basic
channel set in the Channel/signal selection menu are displayed.
When the cursor keys (SIGNAL) are activated, all signals of the basic
channel set in the first display line in the Channel/signal selection menu
are displayed in measuring mode.
affects
2. Press .
7. Press .
11. Use to select the define... button symbol and confirm with .
12. Use to select the Channel 1/Net control field and confirm with
(a n appears in the control field).
15. Use to select edit field Absolute/relative, select Absolute and con
firm with .
16. Use to select edit field Signal, select Net and confirm with .
19. Use to select edit field Signal, select Net and confirm with .
To return to measuring mode, press the Switch key and confirm the
confirmation prompt with .
Displ. PV 1 Displ. P2 PV 1
Force P1 PV 1 Force P2 PV 1
Force P1 PV 2 Force P2 PV 2
CHANNEL NAMES
7. In edit field Channel 4 enter DISPLACEMENT P2 and confirm with .
Channel 1: Force P1
Channel 2: Displacement P1 8. Press .
Channel 3: Force P1
Channel 4: Displacement P2
9. In the pull-up menu select Display format and confirm with .
DISPLAY FORMAT 3 measured values
Image no.: 1 10. In edit field Image no. enter 1 and confirm with .
Image type: 3 measured values
Display value 1 (basic value): 11. Use in the Image type selection field to select 3 MEASURED VAL
Channels/signals: Selection define...
UES and confirm with .
Display value 2: Channel 2 Absolute
Signal Net
12. Use in the Display value 1: Channels/signals selection field to
Display value 3: Channel 3 Absolute
Signal Net choose SELECTION and confirm with .
CHANNEL/SIGNAL SELECTION
OK Cancel All channels All signals
13. Use to select the define... button symbol and confirm with .
1 2 3 4 5 6 7 8 910 111213 14 1516
Channel 14. Use to select the Channel 2/PV1 control field and confirm with
Gross
Net
PV1
(a n appears in the control field).
PV2 4 4
PV12 15. Use to select the Channel 4/PV1 control field and confirm with
LV1
LV2
LV3
(a n appears in the control field).
LV4
16. Use to select the OK button symbol and confirm with .
22. Use in the Signal selection field to select PV2 and confirm with .
23. Press the Switch key and confirm the confirmation prompt with .
Sampling rate
Time interval of signal sampling.
Channels/signals
Setup window
You can define here which channels will be realized with which signals in
DISPLAY FORMAT YT realization
the display. You can define your settings for all channels or only for certain
Image no.: 0 ...
YT realization
ones (Selection) (button symbol define... ). You can call a maximum of
Image type:
1s
four signals one after the other per channel.
Sampling rate
ymax
Setup window
Maximum value of the display relative to the current measuring range (as a %).
DISPLAY FORMAT YT realization
Zero point
y grid
1 subunit YMIN
Channels/signals
You can define here which channels will be realized with which signals in
the display. You can define your settings for all channels or only for certain
ones (Selection) (button symbol define... ). You can call a maximum of
four signals one after the other per channel.
Setting menu
DISPLAY FORMAT XY realization define...
Image no.: 0 ...
Image type: XY realization Button symbol opens new window Channel/signal selection.
Sampling rate: 1s
Display value 1 (basic value):
Channels/signals: All define...
XMax 100... %
XMin -100... %
Setting menu
ymax
DISPLAY FORMAT XY realization
Image no.: 0 ... Maximum value of the vertical axis relative to the current measuring range
Image type: XY realization (%).
Sampling rate: 1s
Display value 1 (basic value): ymin
Channels/signals: All define...
100...
Minimum value of the vertical axis relative to the current measuring range
XMax %
(%).
XMin -100... %
xmin
Minimum value of the horizontal axis relative to the current measuring range
(%).
Current
y grid 1 subunit
measured value (x)
y grid
Scaling
YMIN x grid
1 subunit Current
measured value (y)
XMIN XMAX
1−LV2 Below 20kN You can define here which channels will be displayed with which signals in
1−LV3 NOK the first status line. You can define your settings for all channels or only for
Measure certain ones (Selection) (button symbol define... ). You can call a maxi
mum of four signals one after the other per channel.
Setup window
define...
DISPLAY FORMAT Limit value status
0 ...
Button symbol opens new menu Channel/signal selection.
Image no.:
Image type: Limit value status
Absolute/relative to basic value
Title Limit value
Display value 1 (basic value): You can use this setting to determine whether a reference to the basic
Channels/signals: All define... channel should be set up.
Relative to
Status line 2: Channel 2 basic value Absolute
LV 1
The measured value and the signal of the selected channel are displayed
Status line 3: Channel 2 Absolute
independently of the basic value. The number that is entered corresponds to
LV 1
the actual channel number. Switching channels has no effect.
Status line 4: Channel 2 Absolute
LV 1 Relative to basic value
The number entered here refers to the basic channel. The basic channel
number is 0. Channels located to the left of the basic channel are entered
with a negative sign, channels to the right with a positive sign.
Important
This values does not correspond to the actual channel number
Information
Enter the designation of the limit values in the
Options Limit switch
menu.
Measurement time
Current measurement period
Time elapsed since the beginning of the measurement period.
Periods
Number of measurement periods
File name
Name of the saved recording file. The file will not be overwritten when the
measurement program is restarted. Instead the counter (last four places)
will be incremented.
xxx MB free
Indicates the free storage space on the hard disk.
Information
The MGCplus computer detects when the storage capacity of the hard disk
falls below 1 MB. In this case recording is canceled and the measurement
file is closed.
9.2 F keys
F keys level 1
9.2.1 F keys in measuring mode
Gross
Channel
Function keys F1...F4 are active both in measuring mode and in Setup
mode.
Measure
−> 0 <− −> T <− −II− You can switch a total of 9 functions on three levels in measuring mode:
Level 1
S F1 Zero balance
F keys level 2
S F2 Tare
Acal TEDS Unit ... S F3 Delete peak-value memory
Level 2
S F1 Autocalibration
Function Effect
Tare Taring is triggered.
Zero balance A zero balance is being performed.
Delete memory Peak-value memory is deleted.
Zero/Cal/Measure Switching between zero signal, calibration signal and measurement signal.
Autocal Autocalibration is turned on/off.
LED Status/Level Switching between LED display "Status" and "Level".
Limit value level The "Limit value level" menu is called.
Remote/Local Turn remote control on/off.
Shunt On/Off Turn shunt (transducer, AP14) on/off.
Start/Stop display The current measured value in the display is "frozen" (pause function).
Load next P-set The next parameter set of the amplifier(s) is loaded (1...8).
Load previous P-set The previous parameter set of the amplifier(s) is loaded (1...8).
Switch unit Switches between the basic unit (mV/V), user unit (for example kg) and analog
output (V).
Start/Stop integration Integration of peak values is turned on/off.
Start/Stop recording Recording of measured values is turned on/off.
Load rec. Parameters Parameter sets are loaded from RAM (parameter set 0) or the hard disk param
eter sets 1...16).
Load rec. Comments Recording comments are loaded from the hard disk.
Linearization On/Off Turn linearization of the characteristic curve of the transducer on/off.
TEDS The transducer data is read into the amplifier
Setup window
Select setup window
F−key assignment Level 1
Not assigned
Tare One channel
Zero balance
Clear memory All channels
Zero/Cal/Measure
Autocal
LED Status/Level
Limit value level
Remote/Local
Shunt On/Off
Start/Stop display
Print image
Print measured values
Print settings
Page feed
Load XM001
Load next P-set
Load previous next P-set
Switch unit
Start/Stop integration
Start/Stop recording
Load rec. Parameters
Load rec. Comments
Linearization On/Off
TEDS
9.2.2 F keys in Setup mode
Password
Save/load In setup mode, use the function keys to open the pull‐up menus belonging
Recording
Interface to the menu bar.
Print
Language
All channels present in the setup window have the channel names assigned
in the factory at first.
Empty edit fields identify free slots in your device.
Information
If you want to rewrite an edit field that is already written, use the delete key
to delete the entire entry.
10 System
10.1 Password
You can protect all the settings of your device with a password. This
password protection is turned off in the factory settings. As soon as the
password protection is activated, the password must be entered every time
Setup mode is called (and every time the device is turned on again). Only
then can the settings be edited. The password does not need to be entered
for measuring mode alone.
Access authorization is connected with the password:
S System (all settings can be changed)
S Operator (only enabled settings can be changed)
You can define a password and access authorization for a maximum of nine
users.
Important
In the factory settings, password protection cannot be turned on until a new
user has been defined with "System" access authorization.
Access: set... 8. Use to select the OK button symbol and confirm with .
Add user
User:
Password:
Access: Operator
OK Cancel
Operator
System
Password: change...
1. Use the Switch key to switch to Setup mode.
Access: set...
2. Press .
5. Press .
Access rights for operator
OK Cancel
Password: change...
1. Use the Switch key to switch to Setup mode
Access: set...
2. Press .
5. Use to choose the desired No/Yes selection field and confirm with
Access rights for operator
OK Cancel
If you are still in the Password setup window, continue with item 4.
2. Press .
Password
Save/load 3. In the pull-up menu select Password and confirm with .
Recording
Interface You will then be in the Passwordsetup window.
Print
Language 4. Use to select the delete... button symbol and confirm with .
Time
5. Use to select the delete... button symbol after the user and confirm
PASSWORD
with .
Access: set...
Delete user
User3 delete
OK Cancel
If you are still in the Password setup window, continue with item 4.
2. Print
Password
Save/load
3. In the pull-up menu select Password and confirm with .
Recording
Interface
You will then be in the Passwordsetup window.
Print
Language
4. Use to select the change... button symbol and confirm with .
Time
5. Enter the password in the New password edit field and confirm with .
PASSWORD 6. Press (so you will go immediately to the OK key) and confirm with .
User: new... delete...
Password: change...
Access: set...
Change password
New password:
OK Cancel
10.2 Save/load
You can use the Save/load function to permanently save the current
settings of the AB22A, the CP42/CP52 or the amplifier modules (up to 8
System Display Amplifier Options parameter sets per channel) or to load previously saved settings. You can
load the settings in effect at the time of delivery with "Factory settings".
You can also copy the settings from one amplifier to another.
Password
-
Save/load
Recording
Interface
Print
Language
Time
Save/Load setup
Copy…
OK
Copy settings
Channel 1 2 3 4 5 6 7 8 910111213141516
Param. set 1 1 1
CP 1 Automatically from disk
OK Cancel
Save parameters
Channel 1 2 3 4 5 6 7 8 910111213141516
Param. set 1 1 1
CP 1
OK Cancel
Save parameters
2. Print
5. In the Param.set activation field, under the channel number, enter the
parameter set for which you want to save the settings and confirm with
.
To return to measuring mode, press the Switch key and confirm the
confirmation prompt with .
Load parameters
OK
Automatically from disk 3
Copy settings
Channel 1 2 3 4 5 6 7 8 910111213141516
Settings of amplifiers that are saved on the storage medium are
Param. set 1 1 1 automatically loaded when the MGCplus is turned on or when the storage
CP 1 Automatically from disk medium is connected to the device after the device is turned on.
OK Cancel
Factory settings
A new window opens under Factory settings where you can define
whether all or only certain amplifiers should be set to factory settings. You
System Display Amplifier Options can also load the factory settings of the display and control unit (AB) and the
communication processor (CP).
Save… Load… Factory settings… fields 1...16 and confirm with (a check mark appears in the
selection field). You can use the All button to mark all control fields at
Copy…
once.
6. For multi-channel modules, two dots appear under the channel number.
When you select this type of module a window opens in which you can
OK select individual subchannels or all of them.
Copy amplifier
All or only certain settings can be transferred from one channel to another
(or to multiple channels) as defined by the user. Use the Copy button to do
System Display Amplifier Options this.
Password 2. Print
Save/load
Recording 3. In the pull‐up menu select Save/load and confirm with .
Interface
Print You will then be in the Save/load settings setup window.
Language
Time
4. Use to select the Copy button and confirm with .
5. Use to select the channel numbers for which you want to copy the
Save/load settings settings in the Copy from channel selection field and confirm with .
Save Load Factory settings
6. Use to select the channels for which you want to transfer the
Copy
settings in control fields 1...16 and confirm with (a check mark
appears in the selection field). To transfer all channels select All and
OK
confirm with .
Copy settings
Copy from channel: 1
7. Use to select which settings will be transferred in the vertical control
to the channels 1 2 3 4 5 6 7 8 910111213141516 fields and confirm with (a check mark appears in the selection field).
All
All Transducer
Control input Signal conditioning 8. Use to select the OK button symbol and confirm with .
LV switch Display
Peak values Analog outputs
OK Cancel
To return to measuring mode, press the Switch key and confirm the
confirmation prompt with .
CAUTION
The parameters for these series of tests can be adjusted either with the
AB22A display and control panel or with HBM's "MGCplus Assistant" PC
software.
Recording
Recording and confirm with .
Measurement periods
Number of measurement periods If zero is selected here, the measurement
period will be repeated infinitely after the measurement is started with an F
key.
Example:
50 measured values
1sec
20sec 20sec
Measurement period 1 Measurement period 2
Time channel
RECORDING PARAMETERS
Trigger function
RECORDING PARAMETERS
To use the trigger function you must first define the trigger conditions.
Measuring interval/sampling rate: 300Hz
Number: 1200 ... Duration: 4.0 s...
There are three trigger types available:
Measurement periods: 1 ... 1. Start trigger (with maximum four trigger conditions)
Time
Channels/signals: define... channel: 2. Stop trigger (with maximum four trigger conditions)
Trigger: define...
3. Sampling rate trigger (with start and stop)
PreTrigger: 20.00 % ...
One start condition and one stop condition (trigger condition) can be:
Start condition: Immediately
Example
Pre-trigger measure
Time
Recording of a
Load AP
measurement period
(F−key) Start AP
Start Trigger 1
Start Trigger 2
In the example on è page 252 the following could have been defined as a
Start Trigger 3 start condition:
Start Trigger 4
Stop Trigger 1 S Start measurement when force greater than 5 kN
Stop Trigger 2
Stop Trigger 3
Stop Trigger 4 or
Start sampling rate trg.
Stop sampling rate trg. S when pressure is less than 5 bar
or
S when temperature is higher than 22 °C
Define trigger Start Trigger 1
Mode
Start Trigger 1
Start Trigger 2 Configure start/stop conditions.
Start Trigger 3
Start Trigger 4 Off: Trigger inactive
Stop Trigger 1
Stop Trigger 2 Measurement level: The trigger condition is triggered when the
Stop Trigger 3
Stop Trigger 4 measurement signal exceeds/undershoots the level value.
Start sampling rate trg.
Stop sampling rate trg. Example:
Example:
Off
Measured value level
Measured value band
LV1 Level
LV2
LV3
LV4
External trigger Time
Start Trigger 1
Start Trigger 2 Measured value band: The trigger condition is triggered when the
Start Trigger 3 measurement signal exceeds/undershoots the band limits. The band lies
Start Trigger 4
Stop Trigger 1 between "Level 1" and "Level 2".
Stop Trigger 2
Stop Trigger 3 Example:
Stop Trigger 4
Start sampling rate trg.
Stop sampling rate trg.
Measured value band: outside → inside
Level 2
Define trigger Start Trigger 1
OK Cancel
Example:
Time
Start Trigger 1
Start Trigger 2 LV1-LV4: The trigger is in response to a limit value (High: LED on, recording
Start Trigger 3 program is started; Low: LED off)
Start Trigger 4
Stop Trigger 1
Stop Trigger 2
External trigger: The trigger is in response to an external signal (remote
Stop Trigger 3 control 7 on the connection board).
Stop Trigger 4
Start sampling rate trg.
Stop sampling rate trg. Channel
Source channel for the trigger definitions.
OK Cancel
Off
Measured value level
Measured value band
LV1
LV2
LV3
LV4
External trigger
Mode
Gross
Net You can use this mode to define when a trigger condition is met.
PV1
PV2 You can set different modes depending on which trigger type is set:
Type Mode
Define trigger Start Trigger 1 Measured value greater; less; less → greater; greater → less
Start Trigger 1
level
Trigger:
Type: Off Measured value Inside; outside; outside → inside; inside → outside
Channel: 1 Signal: Gross band
Mode: greater than Limit value high, low; low → high; high → low
Level: 0.000000% ... 0.000000% ...
External trigger high, low; low → high; high → low
OK Cancel
Example: greater
The trigger condition is only met when the measurement signal is greater
greater than Level 1 than the trigger level.
less than
less than greater than Level 2
PreTrigger
You can define in the Pre‐trigger input box how many measured values to
System Display Amplifier Options record before the trigger event (measurement acquisition is divided into the
pre‐trigger range and the post‐trigger range ). The definition refers to the
"Number" entry field in the settings dialog.
Password
Example:
Save/load
Recording Number: 1.200
Interface
Print Pre-trigger: 20%
Language
240 measured values are acquired in the pre‐trigger range and 1200-240
Time
= 960 measured values are acquired in the post‐trigger range.
Stop condition
Number of measured values: The measurement ends when the number of
System Display Amplifier Options measured values specified (entered under
"Number") is reached.
AND‐combined: All stop conditions must be met to stop
Password recording data.
Save/load
Recording
OR‐combined: Data logging ends when one of the stop
Interface conditions is met.
Print
Language Recording to
Time
A file: Recording to a file on the hard disk of the
MGCplus. Enter the name under "Recording
file".
RECORDING PARAMETERS
Measuring interval/sampling rate: 300Hz
Number:1200 ... Duration: 4.0 s... Information
Measurement periods: 1 ... The file will not be overwritten when the measurement program is restarted.
Time
Channels/signals: define... channel: Instead the counter will be incremented (last first four letters are retained
Trigger: define...
and the last four are incremented).
PreTrigger: 20.00 % ...
Multiple files: Recording to multiple files on the hard disk of
Start condition: Immediately
the MGCplus (max. 999)
Stop condition: Number of measured values
Recording to: a file
Recording information file: MGCP0000.MEA ...
Information
Rec. Comments: change
If you activate compression, only one file will be created.
Recording format: 4−byte integer LSB.. MSB
Compression factor: 480
Internal RAM: Recording to the internal RAM of the
Autostart: No
communication processor
Number of measuredsave
Param.set: values load
1
Trigger (AND−combined)
Trigger (OR−combined)
Recording information file
Immediately
Trigger (AND−combined) If you selected "Recording to a file", assign the name of the file here.
Trigger (OR−combined)
Information
Measured value files larger than 2 Gbyte cannot be processed on
computers with Windows R operating systems. The communication
System Display Amplifier Options processor will therefore close the file if it has a size of 2 Gbyte and will
continue recording in a new file.
Recd. comments
Password
Save/load You can assign any comment you choose to the recording. The comments
Recording can be seen at the top left of the display with image type Recording.
Interface
Print
Change recording comments
Language
Time You can change an existing comment, i.e. re-enter it or add to it.
Compression factor
. . . Recording format
Reading a recorded file
To be able to read and interpret the data in the file, you must be familiar with
the internal format of the file. The data is recorded in a binary file structured
as follows:
The following header data items come before the actual data area:
S File ID (4‐byte LONG) // currently = 6001
S Number of channels (4‐byte LONG)
S Length of a data line (i.e. one measured value for all channels/signals) in
bytes (4‐byte LONG)
S Number of data lines (i.e. measured values/channel) in the file (4‐byte
LONG)
S The data format used to save measured values (4‐byte LONG)
S Sampling rate used to acquire measured values (4‐byte LONG)
S Size of the header area in bytes (4‐byte LONG) // currently = 512
S Reserved (4‐byte LONG)
The signal mask indicates how many values will appear in the data line per
channel:
S Bit 0 set: GROSS signal
S Bit 1 set: NET signal
S Bit 2 set: Peak value 1
S Bit 3 set: Peak value 2
This means that a maximum of 4 consecutive values can be displayed for
each individual channel.
This channel block is followed by information about the date and time:
S TimeDate (30‐byte CHARACTER)
The bytes that follow to the end of the header are reserved for future use
and may be left out. They are followed by the values, data line for data line
(the example shown here is based on bit 0 (Gross) and bit 2 (peak value 1)
being set):
S C1, 1 (Gross) C1: Channel 1
S C1, 1 (PV1)
S C2, 1 (Gross)
S C2, 1 (PV 1)
__________________________
Information
The 1253 LONG format contains status information in the least significant
byte. To access this status, you should check this byte first (!) Scale, since
this will cause the least significant byte to be lost due to division by 256.
Autostart
Save
The recording parameters are saved in the selected parameter set.
System Display Amplifier Options
You can save up to 16 measured value recording programs on the hard disk
of the MGCplus.
Password Load
Save/load
Recording A recording parameter set - saved under a ParaSetNo. - is loaded into
Interface MGCplus. Now you can view this recording parameter set or if necessary
Print change it and save it again.
Language
Time
RECORDING PARAMETERS
Measuring interval/sampling rate: 300Hz
Number: 1200 ... Duration: 4.0 s...
LSB MSB
Byte 0 Byte 1 Byte 2 Byte 3
Status information Measured value (24 bits)
Bit no.
0 1 2 3 4 5 6 7
Gross value overflow
Net value overflow This status information is displayed as an integer in the range
sŮ{0 ... 255}.
Calibration error
When measured values are imported from an MEA file (4-byte
Change flag
LV4 status
LV1 status
LV2 status
LV3 status
1) The notation bxc here describes the floor function, which rounds the number down to the next smallest integer: b12.2c = 12; b-12.2c = -13
The "MGCplus Device Time" time channel (called the "HBM" time channel in
the AB22A) is used to count 76.8 kHz periods. 46 bits are available for the
time count, so it will take 29 years for the counter to overflow.
Counter 1 Counter 2
Byte 0 Byte 1 Byte 2 Byte 3 Byte 0 Byte 1 Byte 2 Byte 3
8-bit status 23-bit counter + 1 bit sign 8-bit status 23-bit counter + 1 bit sign
Bit no.
0 7 8 3 0 7 8
1
In this format the time begins to be counted after the MGCplus system starts
up with t=0.
The format of the NTP time is an absolute format that contains the system
time of the communication processor. The time information is split between
2 counters of 32 bits each, with one counter containing the number of
seconds since January 1, 1970 00:00:00 UTC and the second counter
counting the s in the current second.
After February 7, 2106 (that is in about 100 years) there will be problems
with the NPT time.
Header
Date-Time String
Offset Type Content
Offset+0 char[26] Date and Time of file creation (MGC System Time)
e. g. Wed Sep 03 14:43:06 1997
Comments
Offset Type Content
Offset+0 char[80] comment of measurement
Offset+80 char[80] comment of active parameter set
10.4 Interface
You can use the Interface function to adjust the characteristics of the
interfaces:
S Device address
S Subnet mask
S DHCP Yes or No
S Baud rate
S Format (word length, parity, stop bit)
The following interfaces are available with the CP42 and CP52
communication processor:
CP52:
- 2 x independent Ethernet (DHCP capability), upper interface (X4)
additionally with APIPA capability
- RS232 (optionally via USB/RS232 converter)
CP42:
- Ethernet (permanent IP address)
- USB
- RS232
Password
Direct Ethernet connection between PC and MGCplus with
Save/load DHCP/APIPA
Recording
The CP52 communication processor has two independent Ethernet inter
Interface
Print
faces. The upper interface (X4) has DCHP and APIPA capability, the lower
Language interface (X5) has only DHCP capability. The CP52 is delivered from the
Time factory with DHCP adjusted on both interfaces.
To connect your PC with a CP52 in a network with a DHCP server (as is
typical in most companies), both the PC and the CP52 must be set to
Interface settings DHCP. The DHCP administers the IP addresses of all the devices in the
network and assigns IP addresses to both.
Ethernet 1… Info Ethernet 2 ... Info
If there is a direct connection between the PC and CP52 (without network
RS232… connection with a DHCP server) addresses are also promptly assigned to
3 AB disabled in computer mode the PC and CP52 by means of automatic addressing (APIPA) and the
connection can be made. This only applies to the upper Ethernet interface
Ethernet 1 settings
(X4).
Address: 172.20.14.182 ...
Subnet mask: 255.255.0.0 ...
OK Cancel
Router
IP address
Subnet mask
Router A address
Computer Computer
Ethernet Router A
Worldwide
Ethernet Router B
If data packets are sent via node points, for example from the company
computer to the Internet or a WAN / LAN, routers are needed to distribute
the data packets in different directions based on their address.
Routers are computers designed especially for this purpose that are
equipped with their own operating system. They have the task of analyzing
millions of IP packets, reading their headers and then forwarding them on in
the right direction.
Notice
But every client can effect changes in the parameterization of every
amplifier. For example, should a client change the zero offset or the filter
setting of an amplifier, this will immediately have repercussions for the
recorded data of another client process. In this case, the recorded
measurement traceability information will not match the effective amplifier
parameters.
10.5 Language
You can use this function to select the language in the display, menu and
texts.
Password
Save/load
Recording
Interface
Print
Language
Time
LANGUAGE
Language: Deutsch
Deutsch
English
Francais
10.6 Time
You can use this function to set the date, day of the week and time.
Password
Save/load
Recording
Interface
Print
Language
Time
DATE/TIME SET−UP
Day: Friday
11 Menu structure
The menu structure shown below will help you find the required setting
menus more rapidly. The necessary key sequence is also shown here.
Structural elements
Selection bar
Setup window
Selection field
Symbols
mV/V
Selection field
0.000000...
Edit field
measure
Button (triggers action)
measure ...
Button (opens new setup window)
Activation fields
Cursor keys
or
Password
PASSWORD
Password: change...
Access: set...
OK Cancel
or or
Save/load B
Save/load settings
Save... Load ... Factory settings ...
Copy
OK
Load parameters
Channel 1 2 3 4 5 6 7 8 910111213141516
Param.set 111
CP 1 Automatically from disk
OK Cancel
Save parameters
Channel 1 2 3 4 5 6 7 8 910111213141516
Param.set 111
CP 1
OK Cancel
Measure
or or
B Recording C
or or
C Recording
D
RECORDING PARAMETERS
Measuring interval/sampling rate:
1: 2400 Hz 2: 2400 Hz 3: 100 Hz
CP52
or or
D Interface E
OK Interface settings
OK
Ethernet 1… Info Ethernet 2 ... Info
RS232…
Ethernet 1 settings Information on Ethernet 2 settings 1)
Address: 172.20.14.182 ... 3 AB disabled in computer mode Ipv4 act:
Subnet mask: 255.255.0.0 ... Mask act:
OK Cancel
OK Cancel
RS232 settings
Baud rate: 9600 baud
Format: 8E1
OK Cancel
CP42
or or
D Interface
E
OK
OK
Ethernet settings
OK IEEE488USB settings
Address: 172.20.14.182 ...
Address: CP42p 400000 ...
Subnet mask: 255.255.0.0 ...
Format: 8E1
OK Cancel
300 baud
8N1 600 baud
801 1200 baud
8E1 2400 baud
8N2 4800 baud
802 9600 baud
8E2 19200 baud
38400 baud
57600 baud
CP52
oder
E Synchronization
F
Synchronization
Status
Status: MASTER
NTP off
PTP off
OK
NTP
Server:
OK
CP42
oder
E Synchronization
F
Synchronization
Status
Status: MASTER
NTP off
OK
NTP
Server:
OK
or or
E Language Time
Day: Friday
Deutsch
English
Francais
or or
Display format F−keys Channel names
or or
Transducer Signal conditioning Display
F
TRANSDUCER CHANNEL1 SIGNAL CONDITIONING CHANNEL 1 DISPLAY CHANNEL1
or
F Analog outputs Switching
Off
Gross
Net 5 Min Zero
PV1 one time Cal Off
PV2 Status
Measure On
Comb. PV 1−internal Level
2−internal
3−internal
4−internal
5−internal
6−internal
7−internal
8−internal
or or
Control inputs Limit switch Limit value combination G
OK
OK Cancel
OK
or or
G Peak−value memory Version
Off
On
12 Keyword index
A B
AB22A, Control elements, 103 Button, 116, 119, 284
AB22A/AB32
Display, Measuring mode, 105
Drop-down menus, 116 C
Menus, 112
Call menus, 112
Save setting, 115
Setting mode, 110 CANHEAD, connect, 82
Absolute/relative to basic value, 217 Channel selection in measuring mode, 114
Activation fields, 116, 284 Channel selection in setting mode, 115
Adapting to the transducer Communication card ML74, 83
Current and voltage measurement, 166 - 169 Compression factor, 261
Inductive transducers, 145
Piezoresistive transducers, 182 - 185 Conditions at the place of installation, 29
Potentiometric transducers, 186 - 189 Confirmation prompt, 115
Power channel, 165 - 179
Connect transducer, 39, 192
Resistance temperature sensor, 169 - 171
DC power sources, 70
Resistors, 172 - 175
DC voltage sources, 63
Rotational speed, 159 - 166
Distributor board VT810/815, 80
Strain gage transducers, 135
Frequency generators, 73, 74
Strain gages, 140
Inductive full bridge, 42, 43, 46, 47
adjust, Basic parameters, 121 Inductive half bridge, 44
Adjust limit values, 196 LVDT, 45
Piezoelectric transducers, 77
Amplifier Piezoresistive transducers, 78
adjust, 122 Potentiometric transducers, 79
First measurement, 133 Pulse generators, 75, 76
Aufzeichnungsparameter, 252, 258, 259, 260, 261, Resistors, Pt10,100, 1000, 72
265, 266 SG full bridge circuit, 42, 43, 46, 47
SG half bridge circuit, 44
Aufzeichnungsprogramm, 252, 254
Thermocouples, 62
AUTOCAL, 210 Torque flange, T10F-SF1, T10F-SU2, 53
Autostart, 265 Torque shaft, T1A, T4A/WA-S3, T5, TB1A, 59
Connection board
AP460, Connector pin assignment, 92
AP77, 93
Connection board AP74, 84
E
Edit field, 116, 284
I
Edit fields, 118 Image no., 216
F
Factory settings, of remote controls, AP01...AP14, 89
L
Feedback lines, 59 Language, Desktop device, 101
delete, 208
M Turn off, 208
Mains connection, Desktop housing, 31, 191, 232 Periods, 231
Maintenance and cleaning, 30 Piezoelectric transducers, connect, 77
Measured value band, 255 Piezoresistive transducers, 179, 182
Measuring connect, 78
pulses/frequencies, 174 - 177 Pin assignment, AP01i...AP14i, 88
Torque transducers, 151
Polynomial characteristic curve, Cubic, 125
Measuring with
Inductive transducers, 145 - 149 Potentiometric transducers, 186
piezoresistive transducers, 179 - 182 connect, 79
Strain gages, 140 - 144 Pre-trigger, 258
Menu, Pull-up menu, 112 Pulse counting, 75, 76
Menu structure, 283 Pulse generators, connect, 75, 76
Menus, exit, 113
Messperiode, 249
Messraten-Trigger, 252
R
Recording, 106, 213, 231
Messratentrigger, 253
Recording comments, 231, 260 - 264
Messreihen aufzeichnen, 248
Recording file, 259
Mode, 254, 257
Recording format, 260 - 264
Recording parameter set, 265
O Recording parameters, 287, 288
Output stage module, AP01i and AP03i, 217, 227 Remote controls, 85
Assignment AP01i...AP14i, 89
Factory settings, 89
P Function, 191
Parameter set, 210 Replacing the communication processor, 26
Peak-value memory, 202 Resistors, 172
"Envelope curve" operating mode, 207 PT10, 100, 1000, connect, 72
"Instantaneous value" operating mode, 206 Rotational speed measurement
"Peak value" operating mode, 205 T10F-SF1, T10F-SU2 (symmetrical signals), 55
adjust, 202 T10FS-SF1, T10FS-SU2 (symmetrical signals), 57
combine, 203
Rotational speed measurement with inductive
control, 205
transducers, 61
T
TEDS transducer, 126
Thermocouples, connect, 62
Torque flange T10F-SF1, T10F-SU2, 53
Torque measurement
T10F-SF1, T10F-SU2, 53
T1A, T4A/WA-S3, T5, TB1A (slip rings or direct
cable connection), 59