C CB 003, Venous Pressure
C CB 003 000 CASE ERROR
Transient error
General description of conditions for occurring:
- Internal software error
C CB 006, Blood Pump Control
C CB 006 000 CASE ERROR
Transient error
General description of conditions for occurring:
- Internal software error
C CB 006 001 ERROR DURING REGISTER LOAD/READ
Unconditional error, blood pump not available
General description of conditions for occurring:
- The blood pump timer is not working properly.
Technical description of conditions for occurring:
- The timer 9513 has returned error message(s).
Some possible explanations why the error has occurred:
- The timer IC 34 on the control driver board is broken.
C CB 006 002 ERROR DURING SET DISCRETE PORT
Unconditional error, blood pump not available
General description of conditions for occurring:
- Internal software error
C CB 007, Art And Ven Clamp Control
C CB 007 000 CASE ERROR
Transient error
General description of conditions for occurring:
- Internal software error
HCEN12371 Rev. 10.2011 AK 96® - Error Codes 1 (115)
Program version 3.50