Macrotest HT5030

This document provides specifications for the HT ITALIA HV005030 continuity tester and insulation resistance meter. It can measure continuity, line impedance, insulation resistance, short circuit current, earth resistance, RCD tripping time, contact voltage, and frequency. Measurement ranges and accuracies are provided for each test function. The device meets relevant safety and EMC standards, runs on batteries, has a graphic display, and memory for test results.
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (1 vote)
2K views2 pages

Macrotest HT5030

This document provides specifications for the HT ITALIA HV005030 continuity tester and insulation resistance meter. It can measure continuity, line impedance, insulation resistance, short circuit current, earth resistance, RCD tripping time, contact voltage, and frequency. Measurement ranges and accuracies are provided for each test function. The device meets relevant safety and EMC standards, runs on batteries, has a graphic display, and memory for test results.
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

HT ITALIA – DATA SHEET Rel 1.

00 of 27/07/00

MACROTEST5030. Metel: HV005030

• CONTINUITY TEST OF EARTH, PROTECTIVE AND • LINE IMPEDANCE MEASUREMENT (phase-phase, phase-
EQUALISING POTENTIAL CONDUCTORS neutral) or FAULT LOOP IMPEDANCE (phase-earth)

Range Resolution Accuracy Ω)


Range (Ω Ω)
Resolution (Ω Accuracy
Ω)
(Ω Ω)
(Ω (*) 0.01 – 99.99 0.01
0.01 – 19,99 0.01 ±(2% Reading + 2 dgt) 100.0 – 999.9 0.1 ±(2%Reading + 2 dgt)
(*) Incluse the resistance of the calibration. 1000 - 1999 1
Test current > 200mA DC Test current 10A ± 10%
Open circuit voltage > 4,5V Test duration 10ms

• INSULATION RESISTANCE MEASUREMENT • CALCULATION OF SHORT CIRCUIT CURRENT


Resolution ( Accuracy Ik max 40kA
Ω)
Range (MΩ
kΩΩ) (*)
0.01 – 99.99 10 • EARTH RESISTANCE MEASUREMENT
±(2% Reading + 2 dgt)
100.0 – 999.9 100
Range RE (ΩΩ) Resolution (Ω) Accuracy
Automatic selection of the measuring range.
DC test voltage 100V, 250V, 500V, 1000V 0.01 – 99.99 0.01
Open circuit voltage 1.3 x Tensione di Prova. 100.0 – 999.9 0.1 ±(2%Reading + 2 dgt)
Short circuit current 1.4mA (max).
Rated mesuring current 1mA.
100 - 1999 1
Test frequency 125Hz ± 1Hz
Test current ≤10mA
• MEASUREMENT OF TRIPPING TIME, CONTACT VOLTAGE, Open circuit voltage ≤80V
EARTH RESISTANCE (RE) IN TEST OF GENERAL AND Test voltage waveform Sine
SELECTIVE RCD, TYPE AC AND A.
Tripping time t∆N • SAFETY STANDARDS
Range Res The instrument complies with the: EN 61010-1 and EN 61557
Accuracy
(ms) (ms) Insulation class 2, double insulation
Pollution level 2
½ I∆N, I∆N 0 – 999 Inside use; max height: 2000m
2 I∆N (***) 0 – 999 Overvoltage category CAT III 300V (phase - earth)
5 I∆N RCD s (*) 0 – 150 (**) 1 ±(2% Reading + 2 dgt) • MECHANICAL FEATURES
5 I∆N RCD (*) 0 – 50 (**) Dimensions 230(L)x220(La)x90(H)mm
Weight (batteries incluse) approx. 1500g
Jaws opening HT99 21mm
(*) Not for 500mA and 1000mA.
Max conductor size 20mm
(**) With contact voltage monitoring.
(***) Not for 1000mA. • POWER SUPPLY
Contact voltage Ut Battery type 4 batteries 1.5-LR14-E93
Range (V) Resolution (V) Accuracy Battery life approx. 50 ore
Fusible F315mA-250V
+10% - 0% of UBLIM
0 – 2 UBLIMITE 0.1
UBLIM: 25V o 50V • DISPLAY
Grafic
Earth resistance measurement RE without RCD tripping • MEMORY
Range Ω)
Resolution (Ω accuracy I∆N 128kByte
1Ω - 2kΩ 1 +10% - 0%±0.2 V/ I∆N • ENVIRONMENTAL WORKING CONDITIONS
Test current 0.5 I∆N of RDC Reference temperature 23° ± 5°C
Operating temperature -5°C ÷ 50°C
Allowed relative temperature < 80%
• TRIPPING CURRENT FOR GENERAL RCD TYPE A AND AC. Storage temperature -10 ÷ 60°C
Storage humidity < 70%
Range I∆N Resolution Accuracy I∆N • STANDARDS
0.5 – 1.1 I∆N 0.1 I∆N ±0.15 I∆N CEI 64-8 612.2 Test continuity
Test current 0.5 I∆N of RDC CEI 64-8 612.3 Insulation resistance
CEI 64-8 612.6.2 Earth resistance
CEI 64-8 612.6.3 Fault loop impedance
• TEST OF RCD. CEI 64-8 612.9 and app.D RCD tripping test
CEI EN 60947-2 point B 4.2.4.1. (CEI 17-5)
Rcd type AC, A
Minimum Phase – Earth voltage 127V(RCD type AC), 230(RCD type A) • EMC
Rated tripping current (I∆N) 10mA, 30mA, 100mA, 300mA, 500mA, This instrument was designed in compliance with the EMC standards in force and its
1000mA compatibility was tested relating to:

• FREQUENCY MEASUREMENT. EN 55011 radiated emission


EN 50140, EN 61000 immunity
Range (Hz) Resolution (Hz) Accuracy EN 61000-4-2 elettrostatic discharges
EN 50140 R.F. field
15.3 – 99.9 0.1
±(0.1%Reading + 1 dgt) EN 61000-4-4 Fast transient
100 - 499 1
This instrument was designed in compliance with the 72/23/CEE, CEM 89/336/CEE, and the
• VOLTAGE MEASUREMENT 93/68/CEE.

Range (V) Resolution (V) Accuracy • NOTE ABAUT ACCURACY


500 1 ±(2%Reading + 2 dgt) Accuracy is indicated as [% reading + digit number]. It refers to the following atmospheric
conditions: temperature 23°C ± 5°C with a relative humidity of <75%.