Platelet Function Testing: Practice Among UK National External Quality Assessment Scheme For Blood Coagulation Participants, 2006
Platelet Function Testing: Practice Among UK National External Quality Assessment Scheme For Blood Coagulation Participants, 2006
com
Original article
Original article
Table 1 Number of centres performing each test, and approximate number of tests performed
Number of laboratories
performing tests Approximate number of tests per year
On
On all selected
patients patients Total* ,10 10–20 21–50 51–100 .100 .1000
centres indicated they did not exclude any patients from The source and range of reported reference values for the
investigation on these grounds. bleeding time are shown in table 2. Approximately 15% of
centres had established a locally derived reference range.
Bleeding time
A total of 130 centres reported that they performed bleeding PFA testing
times; these were requested exclusively by haematology The PFA-100 was widely used among UK laboratories in the
clinicians in 105 centres, and by haematology or other investigation of platelet function; 72 centres reported PFA-100
hospital-based clinicians in 23 centres; requests could also be measurement as part of their platelet function screening
generated by general practitioners in four centres. In two procedure. A total of 61% of centres performed PFA-100 analysis
centres bleeding time requests were generated by biomedical or on all patients investigated for platelet function defects, 39% of
clinical scientists, and in two centres by specialist nursing staff. centres performed the test on selected cases only. Fourteen
centres reported that they did not carry out further tests if the
The most widely used devices for measuring bleeding time
PFA-100 analysis was normal, 48 sometimes carried out further
were Simplate (Organon Teknika Corporation, Durham, North
analysis, and nine always proceeded with other investigations.
Carolina, US; n = 54), Surgicutt (ITC, Piscataway, New Jersey,
The source and range of reported reference values employed
US; n = 38) and Triplett (Helena Laboratories, Beaumont,
with the PFA-100 analyser are shown in table 3. Approximately
Texas, US; n = 21). Paediatric Simplate devices (n = 1) and
58% of centres had established a locally derived reference range.
Lancets (n = 2) were also employed. One centre reported use of
a Duke’s (ear-lobe) bleeding time.
A total of 106/111 (95%) of centres reported that venous Platelet aggregometry
pressure was maintained at 40 mm Hg when performing a A total of 88 centres performed platelet aggregometry investiga-
bleeding time, one centre reported 20 mm Hg, one reported an tions.
age-dependent range between 20 and 40 mm Hg, one reported
80 mm Hg, and one reported performing a bleeding time with Sample preparation
venous pressure at 140 mm Hg. All centres performed aggregometry on samples collected into
citrate anticoagulant. Of these, four reported using whole blood
aggregometry. Forty-seven centres reported use of samples
collected into 3.2% or 0.106–0.109 mM citrate, four centres
employed 3.8% citrate; the remainder did not state the citrate
concentration employed. Sixty-two (70%) applied maximum
and minimum time limits between which aggregation must be
performed. The minimum time limit before aggregometry could
be performed ranged from 10 to 90 min (median 30 min) and
the maximum limit applied ranged from 45 to 240 min (median
180 min).
Platelet-rich plasma (PRP) was prepared at speeds ranging
from 75 to 210 g, for between 5 and 15 min. The platelet count
was checked by 79/82 centres and adjusted by 73/79, of which
55 indicated that platelet-poor plasma (PPP) was used to dilute
Table 2 Source and range of reference ranges reported for the bleeding
time
Device Source n Median (min) Range (min)
Original article
Table 3 Source and range of reference ranges reported for the platelet function analyser
Test performed Test performed Normal range, Normal range Normal range
on all on selected Normal range lower value upper value upper value
patients (n) patients (n) source range (s) range (s) median (s)
the PRP. Eighteen centres did not indicate how the dilution was Epinephrine
effected. PRP was most frequently adjusted to a count of Fourteen different sources of epinephrine were reported; only
2006109/l, or to a range of 200–3006109/l. PPP was then the epinephrine supplied by Biodata (n = 17) was used by more
prepared by centrifugation between 1200 and 4000 g, for than 10 centres. Reported concentrations ranged between 1 and
between 5 and .20 min. 1000 mm, with 8/60 centres reporting concentrations in mg/ml.
Four centres stored platelets at 37uC prior to testing, and all Only 11/60 used more than one dilution of epinephrine.
other centres stored platelets at ambient temperature (18–
25uC). Arachidonic acid
Fewer sources of arachidonic acid were reported (7), of which
Agonists employed in platelet aggregometry two are used by .10 centres (Biodata, n = 28; Helena
Table 4 shows the most frequently employed agonists and the Laboratories, n = 18). Greater conformity in concentrations
patterns of use amongst participants. Other agonists employed was also noted, with 31/59 centres using a single concentration
were thrombin (n = 1), tartrate-resistant acid phosphatase of 0.5 mM in their investigation. A further 12 centres used a
(n = 1) and cryoprecipitate (n = 1). Spontaneous aggregation single concentration of 1 mM.
was assessed by two centres.
The range of sources and concentrations of each agonist Collagen
reported by participants was substantial. These are summarised Nine different sources of collagen were reported. Seventeen
below. reported use of bovine source of collagen, 23 reported an equine
source, and four reported a human source. However, some
ADP confusion was evident in the source of collagen, as some centres
Ten different sources of ADP were reported; the most widely reported collagen obtained from different sources but from the
employed were from BioData (Alpha Laboratories, Eastleigh, same manufacturer. Final concentrations varied markedly, but
UK) (n = 23), Helena Laboratories (n = 18), and Sigma, Poole, were related to the source of reagent—centres using equine
UK (n = 14). Within these groups, up to 12 different combina- collagen from Helena Laboratories employed concentrations
tions of ADP concentrations were reported, from a range of from 0.2 to 10 mg/ml; centres using Biodata bovine collagen
0.16–5 mm employed by one participant to 20–40 mm employed reported concentrations between 190 and 1900 mg/ml.
by another. A total of 28 of 76 centres used a single
concentration of ADP, the most frequent of which was 20 mM. Measurement of aggregation response
Participants were asked for the method by which aggregation
Ristocetin responses were reported; fig 2 shows the distribution of
Seventeen different sources of ristocetin were reported; the responses. Normal ranges employed by centres for quantitative
most widely employed were from Biodata (n = 10), Helena aggregometry were established from manufacturers’ informa-
Laboratories (n = 22), and ABP, Marlton, New Jersy, USA tion (n = 14 centres), peer-reviewed literature (n = 8), or in-
(n = 12). Within these groups, up to 15 different combinations house evaluation (n = 23), of which 11 centres reported their
of ristocetin concentrations were reported. A single concentra- observed range of responses and 12 employed the mean (SD) of
tion of ristocetin was used by 40/77 centres, and the most percentage aggregation to establish reference ranges. Fifty-nine
frequent concentration was 1.5 mg/ml. The lowest concentra- centres employed internal quality controls when performing
aggregometry; these controls were obtained predominantly
tion employed was 0.15 mg/ml, and the highest was 3 mg/ml.
from laboratory staff, and sometimes age- and sex-matched to
Twenty-seven centres employed two or more dilutions of
the patient under investigation. Thirteen centres indicated they
ristocetin, the most frequent being a range between 0.5 and
tested a control with every patient, and two reported they only
1.5 mg/ml.
did so if a patient result was abnormal.
Original article
Original article
aggregation response (compared with only 1/47 in the Nascola Competing interests: None.
study), which is of necessity subjective, and may fail to detect mild
but significant reductions in aggregation response. Fifty-six per REFERENCES
cent of centres determined maximal aggregation, of which almost 1. Poller L. The bleeding time, current practice in the UK. Clin Lab Haematol
1984;6:369–73.
half also quantified further parameters (slope, lag phase, area under
2. Moffat KA, Ledford-Kraemer MR, Nichols WL, et al. Variability in clinical laboratory
curve). Fifty per cent of centres determined a local reference range practice in testing for disorders of platelet function: results of two surveys of the
for maximal aggregation; several authors have recommended this North American Specialized Coagulation Laboratory Association. Thromb Haemost
approach.2 7 10 However, the merits of using single concentrations 2005;93:549–53
3. Breddin HK. Can platelet aggregometry be standardized? Platelets 2005;16:151–8.
of agonists and comparing maximum aggregation to a reference 4. Zhou L, Schmaier AH. Platelet aggregation testing in platelet-rich plasma. Description
range compared with determining agonist dose responses requires of procedures with the aim to develop standards in the field. Am J Clin Pathol
further consideration. Hayward and Eikelboom10 have emphasised 2005;123:172–83.
5. The Clinical and Laboratory Standards Institute. http://www.clsi.org (accessed
the difficulty in adopting reference ranges based on mean (SD) 27 June 2008).
(used by over half of UK NEQAS participants) as responses may 6. The British Society for Haematology BCSH Haemostasis and Thrombosis
not be normally distributed. Task Force. Guidelines on platelet function testing. J Clin Pathol 1988;41:1322–30.
7. Bolton-Maggs PH, Chalmers EA, Collins PW, et al. UKHCDO. A review of inherited
Finally, less than a quarter of centres reported performing platelet disorders with guidelines for their management on behalf of the UKHCDO.
nucleotide measurement, despite reports that aggregation Br J Haematol 2006;135:603–33.
studies may not detect these defects.16 Our survey has indicated 8. Duke WW. The relation of blood platelets to hemorrhagic disease. Description of a
method for determining the bleeding time and coagulation time and report of three
the wide variability of current practice and the need for updated cases of hemorrhagic disease relieved by transfusion JAMA 1910;55:1185.
recommendations and guidelines in this difficult area of 9. Podda GM, Bucciarelli P, Lussana F, et al. Usefulness of PFA-100 testing in the
haemostasis testing. diagnostic screening of patients with suspected abnormalities of hemostasis:
comparison with the bleeding time. J Thromb Haemost 2007;5:2393–8.
10. Hayward CPM, Eikelboom J. Platelet function testing, quality assurance. Semin
Thromb Hemost 2006;33:273–82.
11. Hayward CP, Harrison P, Cattaneo M, et al; The Platelet Physiology Subcommittee of
Take-home messages the Scientific and Standardization Committee of the International Society on Thrombosis
and Haemostasis. Platelet function analyzer (PFA)-100 closure time in the evaluation of
platelet disorders and platelet function. J Thromb Haemost 2006;4:312–9.
c External quality assessment is not available for tests of 12. Haubelt H, Anders C, Vogt A, et al. Variables influencing Platelet Function Analyzer-
platelet function, and internal quality control is not widely 100 closure times in healthy individuals. Br J Haematol 2005;130:759–67.
employed. 13. Born GVR, Cross MJ. The aggregation of blood platelets. J Physiol 1963;168:178–95.
14. Cattaneo M, Lecchi A, Zighetti ML, et al. Platelet aggregation studies: autologous
c Many centres employed screening tests to exclude further platelet-poor plasma inhibits platelet aggregation when added to platelet-rich plasma
investigation despite reports indicating that these methods are to normalize platelet count. Haematologica 2007;92:694–7.
not sensitive to all platelet function defects. 15. Linnemann B, Schwonberg J, Mani H, et al. Standardization of light transmittance
aggregometry for monitoring antiplatelet therapy: an adjustment for platelet count is
c Updated guidelines and recommendations are required to not necessary. J Thromb Haemost 2008;6:677–83.
improve standardisation and between-centre comparability. 16. Nieuwenhuis HK, Akkerman JW, Sixma JJ. Patients with a prolonged bleeding time
and normal aggregation tests may have storage pool deficiency: studies on one
hundred six patients. Blood 1987;70:620–3.
These include:
References This article cites 15 articles, 3 of which you can access for free at:
http://jcp.bmj.com/content/61/8/950#BIBL
Email alerting Receive free email alerts when new articles cite this article. Sign up in the
service box at the top right corner of the online article.
Notes