Tn
TN
                                   Part Number: RAN456D
                                               Pentra C200
                               Software V1.11 Release Note
QUAL-QA-TEMP-0220_rev15
                                                   Date: 29/1/14
                                                     ECR: 5393
Software V1.11 Release Note                                                                                  TN: RAN456D
      Objective Bug correction
  Intervention
                   Intervention priority   Mandatory on next visit - Immediate and mandatory
   information
                   Tracking number                 TN-RAN456D           Intervention estimated duration:            0h30
                   User process impact          YES            NO       Potential patient result impact       YES      NO
                   User training needed         YES            NO       User documentation update             YES      NO
                                           on instrument serial #
                   Applied in factory
                                           since instrument serial #
 Note history      Note index    Reason for change
                   A to B        New software version
                   B to C        New software version
                   C to D        New software version
       Software    Note index           Previous       New          Mini. requirements
        version    A to B               V1.06          V1.07        V1.04
                   B to C               V1.07          V1.10        V1.07
                   C to D               V1.10          V1.11        V1.10
      New parts    Parts designation        Previous     New                          Reason for change
        New        Adjustments              Previous                 New                     Reason for change
 adjustments
        New        Documentation            Previous                 New                     Reason for change
   documents
  Additionnal      Designation                  Part number                 Link
requirements       Pentra C200 Technical        RAA042DEN (Hard copy)
                                                                            RAS610: Software version installation
                                                                            RAS619: Regional settings
                   Manual                       RAX073D (CD-ROM)
                                                                            RAS595: Check up after intervention
2/8
Software V1.11 Release Note                                                                                     TN: RAN456D
   Part details                                           Component
                                                                                                                    availa-
                     kit or Assy    Designation                           Component designation               Qty   ble for
                                                          p/n
                                                                                                                    order
         1. Required software version
         The V1.11 software version is now available for the Pentra C200.
         This software version fixes the bug of the V1.10 software version concerning the automatic transmission of the results
         to the host in online batch mode.
                     The V1.11 software version must be installed on Pentra C200 that already have the V1.10 software version
                     and the V1.54 ISE software version (refer to the "RAN456C: Software V1.10 Release Note" technical note).
         ■ Check   the software version on your instrument in the Version screen (Services > Customer Services > Version).
                     Please note that you can install the V1.11 software version without installing the V1.10 software version
                     before, on condition that your instrument is not equipped with an ISE module and it already has the V1.07
                     software version.
                     However, you must update the customer documentation and train the customer for the V1.10 software
                     version (refer to the "RAN456C: Software V1.10 Release Note" technical note).
         ■ Check    the OS version and its compatibility with the language to be installed.
             - For EN and FR: the V1.11 software version can be installed on an instrument equipped with OS 2550414106 (DOM
             1GB) or OS 2550498107 (DOM 1GB) or OS 2550584108 (DOM 2GB).
             - For other languages, the V1.11 software version can only be installed on an instrument equipped with OS
             2550584108 (DOM 2GB).
         ■   If the OS version is not compatible with the language to be installed, update your instrument with the DOM Ref.
             F10B40431.
                                                                                                                           3/8
Software V1.11 Release Note                                                                                  TN: RAN456D
         2. V1.11 software version download
         ■ Download   the V1.11 software version in the language to be installed.
           The file contains:     - the user-interface software (UI Main)
                                  - the mechanical control software (Unit Main)
          Language                     Language code (xx)           UI Main               Unit Main           OS
                                                                                                         2550414106 or
          English                              EN                  2550497111            2550494109      2550498107 or
                                                                                                          2550584108
                                                                                                         2550414106 or
          French                               FR                  2550530111            2550494109      2550498107 or
                                                                                                          2550584108
          Spanish                              ES                  2550568111            2550494109     2550584108 only
          Italian                              IT                  2550572111            2550494109     2550584108 only
          Portuguese                           PT                  2550570111            2550494109     2550584108 only
          German                               DE                  2550532111            2550494109     2550584108 only
          Polish                               PL                  2550574111            2550494109     2550584108 only
          Russian                              RU                  2550576111            2550494109     2550584108 only
         ■ Unzip    the file to the root directory of a USB key.
         3. V1.11 software version update
         Here is a summary of the steps to follow in order to update your instrument with the V1.11 software version.
         For more information about software version updates, refer to the "RAS610: Software version installation" procedure
         of your technical manual.
         ■ Insert   your USB key.
         ■ Back     up the database.
         ■ Press    OS Reboot.
         ■ From     Windows, right-click the Start button, click Explore and select Removable disk E.
         ■ Open     the UI Main folder and double-click setup90.bat.
         ■ Open     the Unit Main folder and double-click update.bat.
         ■ Check     the regional settings according to the language to be installed.
         ■ Click    the Start button, then click UI to switch off and on the analyzer.
         ■ Restore    the database (do not shut down the analyzer at this stage of the procedure).
         ■ Register  the ABX Pentra Clean-Chem CP in the Reagent screen (Services > System Configuration > Reagent)
           as follows:
           - Select the reagent code #850 (reagent name: CLEA).
           - Press Delete.
           - Press Save.
           - Press OK.
           - Enter the data below and press Save to register the ABX Pentra Clean-Chem CP with the reagent code #815
           (reagent name: Clean).
4/8
Software V1.11 Release Note                                                                             TN: RAN456D
         ■ Reconfigure the ABX Pentra Clean-Chem CP in both DET.W bottles (probe and mixer) in the Reagent Tray screen
           (Inventory > Reagent Tray).
                                                                                                                  5/8
Software V1.11 Release Note                                                                                      TN: RAN456D
         ■ (Optional)  If your instrument is equipped with an ISE module, configure the chloride electrode as follows:
             - Go to Services > Customer Services > Sequence.
             - Select ON or OFF for Cl Electrode (depending on individual customer needs) and press Exec.
             - A pop-up is displayed: "Please Shut Down the system. OK?"
             - Press OK.
         ■ Switch   off then switch on the instrument.
         4. Checks after intervention
         ■ Check    the software version on your instrument in the Version screen (Services > Customer Services > Version).
         ■   Make sure that you have access to the Pentra C200 User Manual on the customer USB key when you press Help
             from the generic toolbar.
         ■ Check    the pipetting accuracy and precision of your instrument. Refer to the "RAS595: Check up after intervention"
             of your technical manual.
6/8
Software V1.11 Release Note                                                                                    TN: RAN456D
         5. Information about the V1.11 software version
5.1.     Corrected bugs
         The bug identified in the V1.10 software version has been corrected. Refer to the "RAN731A: V1.10 software bugs list"
         service information letter.
                                                                                                                          7/8
Software V1.11 Release Note   TN: RAN456D
8/8