UCRL-92160
PREPRINT
AN IMPROVED APPROACH TO CHARACTERIZING AND
PRESENTING STREAK CAMERA PERFORMANCE
J . D. Wiedwald
B. A. Jones
Th is paper was prepared f o r submittal to
SPIE's 29th
th Annual International Technical Symposium,
San Diego, CA
August 18-23, 1985
July 15, 1985
This is a preprint of a paper intended for publication in a journal or proceedings. Since
changes may be made before publication, this preprint is made available with the un
derstanding that it will not be cited or reproduced without the permission of the author.
OtSWBUWti Of- THIS DOCUMENT iS UNLIMITED
W
An improved approach to characterizing and
presenting streak camera performance UCRL 92160
0. D. Wiedwald and B. A. Jones
DE85 015188
Lawrence Livermore National Laboratory
Laser Fusion Program
P. 0. Box 5508, L-479
Livermore, California 94550
(415) 422-1338
Abstract
Tne performance of a streak camera recording system is strongly linked to the technique used to amplify,
detect and quantify the streaked imaqe. At the Lawrence Livermore National Laboratory (LLNL) streak camera
images have been recorded both on film and by fiber-optically coupling to charge-coupled devices (CCD's).
During the development of a new process for recording these images (lens coupling the image onto a cooled
CCD) the definitions of important performance characteristics such as resolution and dynamic range were
re-examined. As a result of this development, these performance characteristics are now presented to the
streak camera user in a more useful format than in the past. This paper describes how these techniques are
used within the Laser Fusion Program at LLNL. The system resolution is presented as a modulation transfer
function, including the seldom reported effects that flare and light scattering have at low spatial
frequencies. Data are presented such that a user can adjust image intensifier gain and pixel averaging to
optimize the useful dynamic range in any particular application.
The LLNL streak camera system
The Laser Fusion Program at the Lawrence Livermore National Laboratory uses an internally designed streak
camera as a principal diagnostic instrument. This streak camera^ utilizes the RCA C73435 image converter
tube in suveral configurations as its streaking element. Various photocathodes are employed to detect
photons having energies ranging from 1.1 eV (1060 nm) to =»10 keV. To amDlify the image to a recordable
intensity an ITT F4113 intensifier tube with an MCP (micro-channel plate) and fiber optic input and output
windows is coupled to the output of the image converter tube. Until recently, the image was recorded either
directly onto film or fiber-optically coupled ta a two-dimensional charge-coupled device.
We have recently developed a streak camera readout system based on a thermo-electrically cooled CCD
camera^ manufactured by Photometries, Ltd. (Tucson, AZ). The output of the image intensifier is lens
coupled to the CCD camera by an fl.2 lens. Lens coupling was chosen over fiber optics to minimize
mechanical and thermal stress on the CCD. The linearity and large dynamic range of the CCD camera, coupled
with rapid data analysis capability, allow us to characterize the streak camera system more thoroughly than
in the past.
How streak cameras are used in the Laser Fusion Program
Streak cameras are used to record the temporal history of a radiation flux related to either the laser
target heating beam or to the effects of that beam on a target. In addition to the sweep direction, the
orthogonal axis is often used to display spatial or chromatic information by using a microscope or
spectrometer in front of the streak camera. In some cases there are multiple signal channels across tte
streak camera slit, provided by optical fibers carrying independent signals. These applications call for
high spatial resolution while temporal resolution requirements are typically not severe, 25-50 psec is
usually adequate. As a result, streak camera sweep speeds are usually in the 1.5 to 10 nsec range.
Laser fusion experiments occur relatively infrequently, once or twice a day is typical, and represent a
large investment per shot. Due to uncertainty in target performance, signal levels are somewhat
unpredictable. Therefore a large and well-defined measuring system dynamic range is particularly
beneficial. The experimenter wants to know where the limits of the dynamic range are, what happens if those
limits are exceeded and how to best adjust the system gain, both preceding and following the streak camera.
In many cases the experimenter also wants to know how system spatial resolution will affect the diagnostic
measurement, which will result in channel crosstalk, spectral line broadening or degradation of spatial
resolution of an object source.
By using the new CCD camera we now generate a family of system transfer functions that describe the noise
and non-linearity characteristics of the streak camera system at various image intensifier gains. We also
measure the modulation transfer function (MTF) in the spatial direction at the streak camera input and both
spatial and temporal axes at the output of the image intensifier. With this information, the user can
determine the accuracy and limitations of the streak camera system as it is being applied.
-1-
System characterization - dynamic range
The upper end of the streak camera dynamic range is determined by one of two limits. There is a hard
limit to how intense a light pulse is obtainable from an HCP image intensifier when operated in a oulsed
mode. This is because the HCP behaves as a segmented capacitor, charged to several hundred volts with a
glass/vacuum dielectric. For excitation shorter than the recharge time (about 10 msec), the maximum charge
that can be extracted from the MCr is the charge stored in it. Phosphors used in streak tubes generally
have decay times much shorter than 10 msec. The charge stored in the HCP is typically about
4 x 10"' coulombs/cm^, or 50000 electrons/MCP channel. Therefore the maximum charge available from the
MCP is essentially independent of intensifier gain. This puts an upper limit on the number of photons/unit
area imaged onto the CCD or other recording media.
There is another limitation which may restrict the upper end of the dynamic range. This is a "soft"
limit caused by self-repulsion of the electron cloud in the streak camera^-S. When current density in the
streak tube is large due to high instantaneous illumination of the photo cathode, the electrons will
disperse both axially and radially. They spread over a larger area of the phoSphor screen" than they
should. The result is a measurement indicating lower peak intensity and longer duration tha«. the actual
input signals. The integral of the signal is, however, unchanged. This limitation is predominant for short
light pulses and fast streak speeds. For long pulses and slower speeds, MCP saturation is more important.
Streak camera users should be aware of which limit applies in their application. The effects are quite
different and will have different impact on signal recovery efforts.
There are also two factors which limit the lower end of streak camera dynamic range. The first is
quantum noise, due to detection of single photo-electrons emitted from the streak camera photo-cathode.
This depends on image intensifier gain and is dominant when the intensifier gain is high. The recording
system noise forms the other lower limit. This is independent of intensifier gain and in the case of a CCD
readout system is found by measuring the statistics of the electrical signal. The CCD camera we are now
using has an RMS noise of 5 analog-to-digital converter (ADC) counts, corresponding to 50 hole-electron
pairs or —100 incident photons per pixel. Since a background image must be subtracted during data reduction
the effective noise is increased by yT to 7 ADC counts.
We have found that the new CCD camera has sufficient dynamic range to both detect single streak camera
photo-electrons (at high image intensifier gain) and to measure Image intensifier saturation. At an image
intensifier luminous gain of 15000 a single photo-electron produces about 2000 hole-electron pairs in the
CC0. This results in a signal of about 200 analog-to-digital converter (ADC) counts, distributed over a
maximum of 25 pixels. The ADC which digitizes the CCD signal has 14-bit resolution for a maximum of 16383
counts per pi;.el. A totally discharged region of the MCP results In -1O000 ADC counts per pixel, which is
60S of the full.range of the CCD camera. This results in a system dynamic range of about 1250, even without
averaging. This permits us to characterize the dynamic range of a streak camera much more directly than in
the past.
Dynamic range measurement technique
A laser provides varying illumination to the streak camera by utilizing a 70* etalon with 500 psec
inter-pulse time. This provides a factor-of-2 reduction in the intensity of sequential input light pulses
(see figure 1 ) . Neutral density (ND) filters are used to provide a large input signal range. Typically six
100-psec-long pulses (amplitude range of 32:1) are recorded on each streak record. By using NO filters of
0, 1, and 2, the input signal cavers a dynamic range of 3200 in three shots. Careful alignment of the
etalon to prevent beam walking across the slit is essential. It is also important to assure that the CCD
columns are parallel to the streak camera sweep axis.
In order to obtain the system transfer function, peak intensity measurements are made at several values
of image intensifier gain from 500 to 15000. After correcting for sweep non-linearity and non-uniform light
collection a relative system transfer function curve is obtained for each gain. Sweep linearity correction
is necessary because signal amplitude varies inversely with sweep velocity. Non-uniform light collection is
currently corrected by averaging the profiles of several shots that are well within the linear range of the
system and applying a compensation to the measured amplitudes. Both of these steps have been done manually
thus far. The rapid and accurate image digitization and computer access to the data will allow this to
become more automated in the future.
The system noise, which is used to define the lower end of the dynamic range, is obtained by histograming
a dark field. The result is a distribution which describes the noise of the system. For relatively low
image intensifier gains, this is simply the noise of the CCD camera. As the intensifier gain is increased,
single photo-electrons produce detectable CCD sinnal, raising the noise level. We are still evaluating how
to best quantify this latter component of the noise.
-2-
(a) (b)
10000
• ! I !
1000
-MAfttar
n ;
i 1 1
100
0 60 120 180 240 300 360 420 480 540 60 120 180 240 300 360 420 480 540
Time (pixels) Time (pixels)
Figure 1 A wide signal dynamic range is obtained by using a pulsed laser and a 70% etalon. Two cases show
operation in the linear range (a) and when the image intensifier is saturated (b).
Characterization of resolution
We have not attempted to change the standard techniques for determining the temporal resolution of streak
7
cameras . We do, however, make measurements that allow us to report the modulation transfer functions
(both axes) of the CCD camera-lens system, as well as along the slit at the input to the streak camera.
This is done by installing an opaque edge either on the intensifier output or at a plane imaged onto the
slit (Figure 2 ) . By differentiating the CCD signal along a column or row, then performing a Fourier
transform, we obtain the MTF of the system being measured. A significant advantage of this approach is the
ability to observe large scale effects such as lens flare, fiber optic cross talk, and electron scattering
in the streak tube. These effects are typically masked when using resolution charts to measure resolution.
They show up on a MTF curve as poor modulation at low spatial frequencies, below two lp/mm at the streak
camera input. This reduces the ability to accurately measure weak signals in the vicinity of intense ones.
The limiting visual resolution of the system, generally defined as the spatial frequency where the MTF has a
value of 0.04, is also available from the MTF.
10000
Dark Light_
1000
Ps
100 \
u
Q
\
< / \,
M
10
-^*
40
J ttf (1*if
A/i
80 120 160 200 240 280 320 360
Distance across slit (pixels)
Figure 2 - The edge response (logarithmic vertical scale) of the streak camera system. The opaque edge is
located at a slit which is imaged onto the photocathode. The image is produced by a 100-psec
long light pulse and the camera is streaked at lp.mm/nsec.
-3-
Presentation of performance data
The data obtained from the characterization described above is presented to the streak camera user in the
form of two plots (see figures 3 and 4 ) . Figure 3 gives the system transfer function for ."our image
intensifier gain settings. The input intensity is in relative intensity, although in the future we plan to
make this an absolute measurement. The output is in ADC counts per pixel. Image intensifier saturation is
seen primarily at high intensifier gains. As the intensifier gain decreases, the transfer function rolls
off more gradually due to pulse broadening of the electron cloud in the streak tube.
The signal levels where 10% pulse compression (horizontal arrows) and 10% pulse broadening (vertical
arrows) occur are indicated. Ten percent pulse compression will occur whenever the output of the
8
intensifier exceeds 40* of HCP saturation. This agrees with one model of an HCP becoming non-linear .
The non-linearity may occur at a lower output level if it is due to electron cloud self-repulsion in the
streak tube. Ten percent pulse broadening will occur at a fixed input intensity if it is due to
self-repulsion of electrons in the streak tube. If it is due to HCP non-linearity it will occur at a lower
intensity. The user should determine whether pulse compression or broadening is more detrimental in a
particular application, and choose the intensifier gain accordingly.
The system noise level is also provided on this graph. The noise is due to the CCO camera except at high
intensifier gains. At high gains the photo-electron ouantum noise becomes significant. The diagonal line,
shown here at the lower left, was determined by histograming faint images at high gains. We are not
satisfied with this approach and plan to find a better one.
It is important to note that the noise is based on single pixel statistics. The user may reduce the
effective system noise by averaging pixels, thus increasing dynamic range. Averaging over n rows and m
columns will decrease the effective noise by yn x m. The increased dynamic range is obtained at the expense
of reduced resolution.
In addition to the system transfer function, several modulation transfer functions (see for example
fiqure 4) are also provided to the user. These are obtained as described above, with some manipulation of
the data to make the lower intensity side of the edge response dominate the computation, minimizing the
effects of image non-uniformity.
10OOO
I.U V I I I I I I
0.8
S 0.6 -
0.4 \ -
Quantum
noise-' io 0.2
I I I iT—-i- -
Line' pairs/mm
at streak camera photocathode
-3 -2 -1
Log intensity (relative)
Figure 3 - The system transfer function provides Figure 4 - The modulation transfer function calculated
information required to properly apply a from the data of figure 2. The dip in the
streak camera and interpret data fro'n i t . v i c i n i t y of 0.5 lp/mrn is a t t r i b u t e d to
electron scattering from the MCP input face.
Summary
We are trying to present data to the streak camera user in a form that makes 1t easier to properly set up
the camera and interpret the data from it. The recent acquisition of an accurate, all electronic, computer
interfaced camera has made this possible. Future work will include a better means to measure the effects of
photo-electron noise.
Acknowledgments
The authors would like to thank John Severyn, Nick Selchow, Roger Griffith, Sue Bremer, and Don Browning
for contributing to the CCD camera evaluation and measurements described here. This work was performed
under the auspices of the U.S. Department of Energy by the Lawrence Livermore Laboratory under contract
number W-7405-ENG-48.
References
1. Thomas, S.W., et al, "The LLL compact 10 psec streak camera - 1974 update," in Proc. 11th Int'l. Cong, on
High Speed Photography, London, 1974, p. 101.
2. Jones, B.A., "A lens coupled streak camera readout system utilizing a thermoelectrically cooled CCD,"
Conf. on High Speed Photography, Videography and Photonics III, SPIE Proc. 569, August, 1985.
3. Bradley, D.J. et al, "Optimized electron-optical computations of image tube 'designs for high temporal
resolution chronoscopy," Proc. of the 13th Cong, on High Speed Photography and photonics, 1978, p. 451.
4. Kalibjian, R., "Space-charge temporal broadening effects in streak-camera tubei," Proc. of the 13th Cong,
on High Speed Photography and Photonics, 1978, pp. 452-455. f
5. Degtyareva, V.P. and Platonov, V.N., "Computer evaluation of Coulomb repuj'sion effects in picosecond
streak image camera," Proc. 15th Int'l. Cong, on High Speed Photography and'Photonics, (1982) SPIE Vol.
348, pp. 1001-1004.
6. Kalibjian, R. and Peterson, G.G., "Analysis of space charge in photonic tubes," J. Appl. Phys. 54(8) Aug.
1983, pp. 4295-4301.
7. Fleurot, N., "Workshop on picosecond streak cameras," Proc. 16th Int'l. Cong, on High Speed Photography
and Photonics, (1984) SPIE Vol. 491, pp. 374-375.
Attendees at this workshop concluded:
Temporal resolution should be specified by one of the two following criteria:
- F.W.H.M. of the pulse response to a delta intensity input function
- or by Rayleigh limit iT such that the contrast
KAT-H
C-$J.ZM
The following complementary parameters should be simultaneously specified:
- sample width in spatial direction
- sweep speed
- flux density level
- wavelength
- spatial resolution corresponding to this temporal resolution.
Information concerning degradation of temporal resolution off center should be given.
Csorba, I.P., "Current gain parameters of microchannel plates," Applied Optics, Vol. 19, No. 22, Nov.
1980, p. 3863.
-5-