CST240 Service Manual
CST240 Service Manual
List .................................................................................................................................................... 1
Chapter 1 Instrument specification ................................................................................................... 2
Chapter 2 Installation ........................................................................................................................ 9
Chapter 3 Performance and test flow .............................................................................................. 12
Chapter 4 Module Instruction ……………………………………………………………………..21
Chapter 5 Instrument Fluid Pipeline………………………………………………………………33
Chapter 6 Instrument Hardware Circuit…………………………………………………………..41
Chapter 7 Maintenance……………………………………………………………………………56
Chapter 8 Analysis Method………………………………………………………………………..72
Chapter 9 Troubleshooting List……………………………………………………………………84
1
Chapter Ⅰ Instrument Specification
2
1.1.2 Front ① ②
① Model Logo
② Upper Cover
③ Front Cover
1.1.3 Rear
④ Detergent Inlet
⑦ Syringe Pump
⑨ R S232 Port
① ②③ ④ ⑤ ⑥ ⑦ ⑧ ⑨
3
1.1.4 Working Table
④ Stirring Unit
⑦ ⑧
1.1.5 Rightside
① Main Switch
② Power Socket
① ② ③ ④ ⑤ ⑥
4
1.2 Analysis Unit
The working speed of CS-T240 auto-chemistry analyzer is 240 tests / hour at constant speed
(single / double-reagent item), whose working period is 15 seconds. Instrument overall structure
adopts the project of "2-disks + 1-probe + 1-stirring rod"—one reaction disk, one sample &
reagent disk, one stirring rod, one sample & reagent probe used for adding reagent and sample,
one stirring rod used for mixing. "Grating + diode array" approach is adopted in optical
measurement unit for real-time optical collection of reaction cuvette. The 8-stop 12-step
automatically rinsing of the reaction cuvette is carried out during testing process.
1.2.1 Structure
5
1.2.3 Probe and Stirring Unit
Stirring Unit Probe Unit
Probe unit:1
Stirring Unit:1
6
① ② ③ ④
7
1.3 Function Overview
Main work flow:
1. All mechanical moving parts initialization.
2. 3 times water blank measure is implemented after six time automatic rinsing
3. Sample reagent probe assimilates quantitive reagent when it descents to reagent sample disk
after the sample reagent disk rotates to designated R1 reagent position. And then, sample reagent
probe assimilates quantitive sample when it descents to reagent sample disk after the sample
reagent disk rotates to designated sample position.
4. After 8-stop 12-step rinsing, reaction cuvette stops at the sampling position, and sample reagent
probe rotates to reaction disk and descends to reaction cuvette to discharge the mix liquid(reagent
and sample), the reagent 1 and sample adding is finished.
5. Reaction cuvette is stirred immediately when it rotates to R1stirring position.
6. Sample+R1 reagent react in reaction cuvette or temperatured.
7. If it is double reagent item test, sample reagent disk rotates to the designated R2 reagent
position and sample reagent probe descends to sample reagent disk to assimilate quantitive reagent
after a set period (6 mins).
8. The sample&reagent probe discharges R2 into reaction cuvette when it rotates to reaction disk.
9. Finishing R2 reagent adding, reaction cuvette is stirred after its one circle (R2 stirring position)
rotation.
10. Reaction cuvette carries out the collection of absorbance data when it passes the optical unit in
every period.
11.The reaction cuvette is rinsed automatically after reaction when passing the rinsing unit, and 15
minutes have been elapsed since sampling to rinsing.
8
Chapter 2 Installation
9
Please use air conditioner if the temperature or humidity can not meet the requirement above.
The heat generated during working process by the instrument will be emitted at the rear of the
instrument. Good ventilation should be kept well and ventilation equipment can be adopted if
necessary, but direct air current should be avoided, or inaccuracy of instrument test may be
caused.
Get through the pure water machine, computer host, display and analytical unit power supply, and
enter CS auto-chemistry analyzer systematic application software. Initial user name: 001, initial
password: 001.
After enter software, follow the steps below in “Maintenance” interface.
(a)Injection pump exhaust
Execute injection pump exhaust to expel air in pipeline.
(b)Cleaning liquid pipeline exhaust
Executing irrigation detergent pipeline exhaust is infusing detergent into pipeline to expel air in
pipeline.
(c)Reagent&sample probe horizontal check
Make sure sample&reagent probe is right above reaction cuvette, rinsing groove and reagent
bottle.
Place two standard cups at outer circle position 43, inner circle 44 respectively in the sample&
reagent disk, and put two blank bottle in inner circle 2 and outer circle 1.Make sure the sample&
reagent probe is above reaction cuvette, rinsing groove, standard cup by implementing reagent
sample probe horizontal check.
(d)Stirring rod horizontal check
In order to make sure the stirring rod is above the reaction cup, rinsing groove.
10
(e)Mechanical movement check
Execute 20 times mechanical movement checks to make sure whether the washing block of
rinsing mechanism nozzle abrases the reaction cuvette or not and each mechanism runs normally
or not.
(f)Rinse reaction cup+ ISE
Select rinsing reaction cuvette in “Maintenance” interface, and execute rinsing reaction cuvette +
ISE if ISE equipment is collocated.
(g)Light quantity check
Light quantity result should be attached to installation check report with its value no more than
18000.
(h)Cup blank test
No.1 cup blank value should be within 18000, and 2-120 reaction cup check value should be
within 18000 ±800.
11
Chapter 3 Performance and Test Flow
Wavelength
±2nm
precision
Reaction
37℃±0.1℃
temperature
Characteri
stics Simultaneously testing 60 colorimetric items
Test item
and 3 ISE items at most
Test
Rate assay ,end-point assay, 2-point assay.
method
12
Sample & reagent disk,sample Reagent & sample disk with refrigerator,
position, reagent position semiconductor cooling system.Total 67
positions(21 routine samples, 45 reagent
positions “CS-bacterial phosphate-free
detergent”, 1 detergent position)
13
type:code 128
Size:
Reaction cuvette
6mm
optical path
QC QC interval, monthly QC
14
power(VA) 650VA
15
3.2 Test Flow
3.2.1 典型测试流程
16
3.2.2 Test Flow Instruction
1. Periodic movement sequence of sample reagent probe
a. Switch the pump to zero, internal and external wall rinsing
b.Inhale 3ul air.
c.Rotate to the position above the sample reagent disk
d.Descend till the sample&reagent probe point into liquid level about 2mm
e. Assimilate quantitive volume + push back redundant sample
f.Move to above the rinsing pool from the reagent bottle to rinse the external wall.
g. Inhale 3ul air
h. Move to above the sample&reagent disk
i. Assimilate quantitive volume sample
j.Move to above the reaction disk from sample cup or tube
k.Add sample and reagent into the reaction cuvette
l. Rotate to above the rinsing pool from reaction cuvette
→ (next periodic movement sequence).
17
120 个反应杯位置号
反应杯清洗机构
加试剂与加
反应盘复位点
样位置
搅拌位置
光电检测位
参考位置编号
Outer circle figure: No. of reaction cuvette;inner circle figure: No. of mechanism position.
18
i. Reaction Cuvette Rinsing Movement Sequence
C D
A B E F G H
Reaction disk rotate direction Test cell blank for 3 times(1 stop,2 pass)
figure 3-3 Reaction cuvette rinsing probe position
Above figure shows that 8 steps are needed when rinsing reaction cuvette. (3 times cell blank test
is added) , therefore, to finish rinsing one reaction cuvette, 12 steps are needed:
ii. Optical measurement movement sequence
Photometry in the entire process is adopted. In 13-minute reaction time, the continuous
determination of the absorbance of reaction solution is carried out. Reaction disk rotates 1 plus 2
pitches, about 15 seconds, absorbance values are measured out when the 120 reaction cups passing
optical axis of the photometer one by one.
Each reaction Cup in 3-minute reaction time was measured 12 times (12 photometric points),
4-minute reaction time was measured 16 times (16 photometric points), 5-minute reaction time
was measured 20 times (20 photometric points), 10-minute reaction time was measured 40 times
(40 photometric points), 13-minute reaction time was measured 49 times (49 photometric
points).
Light starting from the light source was focused by the lens, and passed the reaction cup first, and
then was disparted by concave grating. After spectrophotometry, each wavelength were received
by 12 fixed photoelectric sensors simultaneously, and were amplified 12 amplifiers, after Log
transformation to derive the rate of change of absorbance or absorbance. When dual-wavelength
testing is used, the concentration value is calculated by the difference of the main and
sub-wavelength absorbance or that of absorbance change rate, and therefore dual-wavelength
testing can not only compensate the blood lipid, hemolytic, jaundice sample test, but also
compensate on the result impacted by voltage changes, so that measurement is more accurate,
more stable.
19
20
Chapter 4. Module Introduction
In addition, main function of reagent&sample probe component: liquid level detecting and
collision detection in movement process, probe block detecting function.
Probe drive mechanism plays a key role of reagents and samples adding. The way of probe are
only up-down and circular moving, so two step motors are necessary to drive.
21
4.1.2 Composition
Probe rotating arm
Probe rotating gear belt
position hole
Up-down motor
mounting hole
Lower rotating axie centor
Mandrel
22
Rotating probe drive ratio is 12:34, using 0.9° stepper motor and 8 segment controller. Control
accuracy can achieve 0.0398°.
Up-down main gear driver diameter is 19.1mm, the 60mm for the perimeter, also using 0.9°
stepper motor and 8 segment controller.Up-down control accuracy can achieve 0.01875mm.
2, The reaction disk turntable: reaction plate, the reaction cuvette, incubation bath, disk rotating
bracket, step motor-driven components
S/Driving seat
Sample&Reaction
Storehouse cover light sensor disk motor seat
23
Driven gear
Drive gear belt
disk and
light sensor
System transmission ratio is 10:1.Because of using 0.9 °step motor with 8 segment driver circuit,
Disk cover
45 teeth of sample&reagent switch light sensor
code disk
24
temperature maintains at 6 degrees -10 degrees, 21 sample positions and 46 reagent positions (the
45th fixed position for placing phosphor-free CS-anti-bacterial detergent)
Equipped
with
temperature-keeping
reagent
cooling storehouse
4-8 Reagent cooling storehouse and sample&reagent disk rack assembly configuration
Air inlet
Reagent kit
4-9 reagent cooling storehouse and sample&reagent disk rack assembly configuration
Gear drive adopted reaction disk due to the relatively high positioning precision.
25
Colorimetric cup Fixing pin
rotating rack
Driving gear
Driven gear
motor vibrator
cuvette
cuvette install pin
Locking screw of
colorimetric cup rack
26
stirring rod rinsing
bath
Spilling outlet of
Optical window incubation bath
Circular constant
Circular constant
temperature water
temperature water
outlet
inlet
4-13 Incubation bath components configuration
Phosphor-free
Incubation bath detergent inlet
Optical system
27
4.4 Stirring Unit
Rotating mechanism of stirring and rotating arm adopts the direct drive way of step motor output
axle, using 8 segment drive circuit of 0.9 ° motor, and the controlaccuracy can achieve 0.1125 °.
The largest angle is limited by the open angle of rotatation code disk mechanical limit. And
positioning is determined separately by the left and right light sensors.
Mechanism rotation
mechanical limit
Stirring mechanism
up-down motor
Stirring mechanism
up-down slider
Stirring mechanism
up-down 4-16 Stirring components configuration
light
sensor
Stirring mechanism
up-down motor
Up-down slider axle
Curve axle and curve handle drive is adopted by stirring up-down mechanism .
28
Up-down linear
sliding axle
Up-down
Up-down slider
guide axle
Because of mixing body movements are used by way of the crankshaft crank, so the
movements of the positioning accuracy at different locations different. Landing and taking-off
between the location of the speed of the highest, lowest accuracy, and precision at both ends of the
Electrical axis from the axis of the slider bearings the size is 16.5mm, crank in horizontal position,
Due to curve axle and curve handle drive is adopted by stirring up-down mechanism, the
up-down positioning precision are different at different position. The speed is the highest and
precision is lowest at the middle, however the precision is highest and speed is lowest at the two
ends.
The size of Motor axis from the axis of the slider bearings is 16.5mm, when curve handle locates
29
Up-down slider
Up-down slider
30
4.6 Colorimetric cup rinsing mechanism
Up-down
slider
Up-down step motor
4-20
4.7 Rack
Rack includes the main rack, electrical, gas liquid valve and water tanks and other racks.
Isolating
standing
board
Adjusting foot
31
Electrical control
refrigeration controller
box
Switch
Switch power
4-22
32
Chapter 5. Instrument Fluid Pipeline
The CS-T240 liquid line system can be divided into four parts: water inlet tank, 37 degrees
centigrade temperature, colorimetric cup cleaning, the inner and outer arms and stirring rod
cleaning.
1. CS-T240 liquid line system includes sampling subsystem and cleaning subsystem.
2. Sampling subsystem uses one probe plus one stirring rod and one injector. The injector use
500uL
3. The inner and outer wall cleaning of probe and stirring rod uses barotropic driving.
4.Cleaning bath: two cleaning bathes plus waste liquor in the reagent storehouse, together
with the flooding waste liquor in the reaction disk etc.
5. The reaction cuvette cleaning uses the way named 8-stop 12-step.
6.Water supply: uses the special outboard water-supply equipment and the special
water-supplying machine.
7. Waste liquor: reticulate pattern pipe as low concentrate waste pipe with inside diameter of
12mm, wall thickness 3mm, high concentrate waste pipe is also reticulate pattern pipe with inside
diameter is 8mm and wall thickness is 2mm
8. Source of power: the power of cleaning comes from the magnetic pump..
11. The inner and outer wall cleaning of probe uses independent solenoid valve while the
stirring rod use one solenoid valve.
33
5.2 Liquid Line Principle
5.2.1 CS-T240 Liquid line principle sketch map
34
5.2.1 liquid line of water inlet tank
Explanation:
Z1——heater; Z2——lamp coller; Z3——float; Z4——temperature sensor;
Z5——water tank; Z6——magnetic pump; Z7——water seven-pass;Z8——air discharger
Z9——flow controller;Z10——pressure sensor;Z11——5 valve plate assembly;A——detergent;
B——probe inner wall rinsing;C——Cuvette rinsing;D——stirring rod rinsing;E——probe external
wall rinsing ; F——water inlet ; G——incubation bath inlet ; H——air outlet ;
M——overflow outlet; N——cuvette rinsing
1, When the low water level float detects out the signal, open the inlet valve SV13, and water tank
begins to be infused water until the high water level float detects out signal, then turn off SV13 to stop
the water. Influent flow is as follows:
Stop
Alarm of low influent Alarm of high water influent
water level check water level check water
2, When the low water level float did not detect out signal, the water tank heater began to work,and
temperature control started. Magnetic pump began to work simultaneously.
3, Output water pressure of water tank is controlled by the magnetic pump and fixed damper regulator.
Magnetic pump head is 4.6 m(50Hz). Control water pressure is around 0.45kgf/cm2.
4,The tank water is pumped into the seven-pass through magnetic pump: the first pass to the
35
five-valve plate assembly, used for cuvette rinsing, stirring rinsing and external wall rinsing of the probe;
second pass through the degassing device for the internal wall rinsing of the probe; third pass with the
detergent used for the first probe rinsing of rinsing unit; fourth pass for the incubation bath; fifth pass
access to pressure sensor for detecting the tank pressure; sixth pass for regulating the water tank
pressure.
5, When abnormality occurs to the float, possibly water tank remains the status of inputting water. When
the water tank is full of water, SV13 not shut down, tank outflow water spills out from the overflow pipe
into the waste liquid pipe.
This system offers precise constant 37 degrees water to the incubation bath of reaction disk, and cools
the high temperature light source simultaneously. This system consists of magnetic pump, incubation
bath water inlet valve, incubation bath water outlet valve, liquid level detector and temperature
controller consists of the heater, temperature detector and the shell.
1, Open the inlet valve SV6 and turn off outlet valve SV10 to infuse water into incubation bath,
simultaneously with reagent&sample probe adding phosphor-free anti-bacterial rinsing liquid to the
incubation bath, liquid level detector determining whether to stop water.
2, Turn off outlet valve and inlet valve, and start the water circulation magnetic pump and temperature
36
controller. In order to improve the adjusting performance of the PID temperature controller in high
temperature environment, the system is added the cooling device through the water tank to get a
small amount of temperature cooled.
3, Turn off magnetic pump and temperature controller, open the drain valve SV10, time to turn off the
drain valve when the incubation bath is draining.
5.2.3 Probe internal and external wall rinsing
1,Open valve SV9 can get the inner wall of reagent sample probe rinsed. Probe position should be at
the top of the corresponding cleaning trough when cleaning so that waste liquid can get out of the
instrument.
2,Open the valves SV4 and SV5 to rinse the external wall of sample&reagent probe and stirring rod.
Fixing pressure adjusting piston is adopted to every external wall rinsing pipeline to avoid rinsing water
spilling out of rinsing bath.
improve cleaning speed, colorimetric cup adopts hydraulic valve switch and vacuum liquid exhaust.
1, warm water provided by heating tank of inlet water is about 34 ℃. Water pressure is produced by
37
magnetic pump from the water tank and adjusted by voltage regulator to the stable pressure, about
0.45kgf/cm2.
2, vacuum pump, vacuum tank and pressure detector composes vacuum source with pressure value about
3,Vacuum liquid discharging is high concentrate liquid – reaction liquid including concentrated patient
4, when the colorimetric cup cleaning mechanism descends, turn off valves open the valve SV11. It tarts
assimilating sample under the vacuum pressure, and the liquid of colorimetric cup will be discharged
after a short period of time when the rinsing mechanism arrives at the bottom of Colorimetric Cup.
5, cleaning liquid and ionized water adding is completed by the SV1、SV2、SV3、SV7 and SV8 which
are timed. Because the vacuum liquid discharging begins to work simultaneously when adding liquid to
discharge redundant liquid, the liquid will not spill outside colorimetric cup.
6, SV7 and SV8 valves are responsible for adding cleaning liquid.
SV8 valve is three links, NO port connect with vacuum tank, NC connect with water seven pass.Before
adding or after the previous adding, open the valve SV7, but valve SV8 is at COM and NO conduction
status (power off status). Because there is a one-way valve in the 3-way top pipeline, cleaning fluid
flows into the middle pipeline of the SV7 and SV8 valves, and time valve SV7, cleaning liquid will
remain in pipeline. Open valve SV8 when adding, cleaning liquid will be added into colorimetric cup
through one- way valve under the pressure. About 500ul cleaning liquid is consumed each time.
38
Explanation:Z1——cuvette rinsing unit;Z2——detergent component;Z3——five-valve plate unit;
39
40
Chapter 6 Instrument Hardware Circuit
PC
Sampling mechanism
Rinsing unit
Stirring unit
convertion
and
41
6.2 Security Note:
At working, touching hardware panel with hand or any other objects is forbidden.
In order to dismount panels, operation is only allowed when cut off power (220V, AC).
42
ISE 1. Communication with the main control board
2. ISE pump motor control
Circuit board
4. Magnetic valve control: ISV1、ISV2、ISV3、ISV4
Mother board 1. Providing reaction disk board, sample reagent board, rinsing
stirring board, main control board, ISE board with connection
and power supply.
1. Wiring
43
2. Function
As shown in above figure:
Power input: ~220V
L1、N1: AC power filter S1: power switch
G1: power filter connecting with the ground of bottom-board
P131: AC port of solid state relay board
J501: fan port of AC system J502: AC power port of cooling system
J503: power Supply Port of Halogen lamp J504: Zero line port
J505: power switch of circuit board box J506: power port of solid state relay board
N1:±12V, 5V switch power, power supply for the mother board of circuit board
44
6.4.2 Refrigeration system
1. Wiring
2. Function
Refrigeration system is composed of semiconductor refrigeration module, radiator, fan.
The main function of refrigeration board is controlling the semiconductor refrigeration module, to
keep the reagent disk temperature within 6-10℃.
As shown in the figure :
U1:refrigeration control board is the CPU of refrigeration system. It controls the working of
refrigeration system on the basis of the reagent disk temperature detected by temperature sensor.
Semiconductor refrigeration module and fan are controllable. The refrigeration control board starts
when the temperature beyond 6-10℃ to adjust the temperature. It also communicates with the main
control board.
DS18B20:the sensor used to detect the reagent disk and environment temperature
D1-D2:semiconductor cooling module(Peltier)
N3:12V switch power
N2:5V switch power
45
1.Wiring
46
2.Function
Communicate with PC through serial port to transmit data, command and alarm information.
Transmit data and command with reaction disk board, sample&reagent disk board, ISE board,
refrigeration board and AD board communication through mother board.
As shown in figure:
Solenoid Valve:
Refer to 5.2.1
47
6.4.4 Solid state relay board
1. Wiring
2. Function
Its main function is to control the solid state relay through main board to turn on and turn off the pump ,
motor, heater and lamp.
As shown in figure :
LAMP:Halogen lamp
48
6.4.5 Circuit board of reaction disk
1. Wiring
2. Function
The main function of circuit board of reaction disk is to receive the command from the main control
board, communicate with the main control board, transmit data and alarm command.
Controlling sample&reagent unit includes sample&reagent disk rotation, probe unit running.
Barcode scanning of sample&reagent disk
J702:motor drive port of reaction disk
J704:signal receiving port of reaction disk clear optocoupler, count optocoupler. J705:output port of
reaction disk count optocoupler
J706:output control signal port with AD board
49
6.4.6 Rinsing mixing circuit board
1. Wiring
50
2. Function
The main function of rinsing mixing circuit board is to receive the command from the main control
board, communicate with the main control board, transmit data and alarm command.
Controlling sample&reagent unit includes sample&reagent disk rotation, probe unit running.
Barcode scanning of sample&reagent disk
As shown in figure:
J041: signal port of rinsing mixing optocoupler
J042: drive port of solenoid valve
J045: stirring motor control port.
J048: communication port with cooling board
Solenoid valve:refer to 5.2.1
SV1:cuvette rinsing probe(the second one)control solenoid valve
SV2:cuvette rinsing probe(the third one)control solenoid valve
SV3:cuvette rinsing probe(the fourth one)control solenoid valve
SV4:stirring rod rinsing control solenoid valve
SV7,SV8:cuvette rinsing probe(the first one)control solenoid valve
SV11:cuvette vacuum control solenoid valve
51
6.4.7 Circuit board of sample&reagent disk
1. Wiring
52
2. Function
Receive the command from the main control board, communicate with the main control board, transmit
data and alarm command.
Controlling sample&reagent unit includes sample&reagent disk rotation, probe unit running.
Barcode scanning of sample&reagent disk
As shown in figure :
J041: signal port of sample&reagent disk optocoupler
J042: signal inlet port of probe position limit optocoupler and liquid level detection optocoupler
J043: signal inlet port of injection pump optocoupler
J044: SV5,SV9 drive signal port of solenoid valve
J045: motor drive port of sample&reagent disk
J046: motor drive port of injection pump
J047: motor drive port of probe
J048: communication port of barcode scanning
53
6.4.8 ISE circuit board
1. Wiring
2. Function
54
The instrument add calibrator into the reaction cuvette.The SIP injection pump aspirate the calibrator,
and pump it into the Na、K、Cl electrode pipeline of the reaction tank. Test the electric potential on the
basis of reference electrode. The sample probe aspirates sample and diluent, and pump it into the
reaction cuvette. Test the electric potential, and the concentration can be calculated.
J2:optocoupler port of aspirate nozzle elevator mechanism
J5:motor port of proportioning pump
P13:optocoupler port of proportioning pump
J61:motor port of aspirate nozzle elevator mechanism
J62:solenoid valve port
J64:preamplifier board port
ISE preamplifier board:amplify electrode signal
55
Chapter 7 Maintenance
In order to ensure reliable system performance, excellent working status and span, please conduct
system operation and regular maintenance strictly in accordance with the requirements in the service
manual. Learning maintenance and overhaul of this chapter is also very important and in-depth study
will enable the instrument to achieve the best running status and exert the best performance.
Warning:
Do not touch any other parts except user self-operation and maintenance
which are clear recorded.
Do not splash water, reagent and other liquid onto the system's
mechanical or electrical parts.
56
7.1 Maintenace preparation
Tools, high concentrate detergent and alcohol maybe used in maintenance.
1.Tools
One set of hexagon wrench
Cruciform Screwdriver (large, medium and small)
Injection needle hose
Small tweezers
Clean gauze
Warning:
Acidic high concentrate detergent and alkaline high concentrate detergent
mixed generate poisonous gas. Do not mix them.
Caution:
Following high concentrate detergent designated by Dirui:
Acidic high concentrate detergent: 0.1mol / l hydrochloric acid; alkaline
high concentrate detergent: 0.5% (V / V) sodium hypochlorite.
Please use the high concentrate detergent designated by Dirui. If
undesignated types of high concentrate detergent are used, inappropriate
analysis results might be received.
Dirui recommends the use of alternating acidic and alkaline high
concentrate detergent, for example, use acidic high concentrate detergent
after power on, then use alkaline high concentrate detergent next time
after power on.
1 Make sure that Power of analysis part has been switched off.
57
2 The injection pump can be seen in figure.
If so, check the leakage causes, and check the pipeline and connector timely.
2 When cleaning, carefully observe whether the stirring rod works normally, If not
normal, check the corresponding liquid line channel, check whether water supply
of water tank and water pressure is normal.
58
2 When rinsing, carefully observe rinsing probe working and whether probe infusing
is normal and assimilating is completely.
If infusing abnormal, check pressure value of water infusing pressure gauge
If assimilating abnormal, check pressure value of assimilating vacuum
7.2.5 Check waste connection and discharging
1
Biological contamination danger:
During waste water operation, please put on gloves, put on work
clothes and if necessary, wear protective glasses.
Check whether liquid waste disposal system is normal every day, and maintain
waste liquid pipe is not bent and discharges smoothly and high and low
concentration waste liquid are disposed properly (refer to local standards of
dispose waste liquid).
Warning:
Please be careful to avoid hands from being scratched
Biological contamination danger:
In operation, please put on gloves, work cloths, and put on protective glasses
for the best.
Do not dispose the gauze used to clean sample probe at your own will, please
follow the relevant provisions for proper disposal.
59
1 Make sure the analysis part power supply is switched off.
2 Lift the rotating arms of samle&reagent probe by hands to the top position, and
rotate it to the top of sample&reagent storehouse for convenient operation.
3 Caution:
When cleaning, do not touch directly the probe surface to
prevent probe scratch; avoid too much hand force to prevent
deformation of the samle&reagent probe.
Note:
Acidic and alkaline detergent can be used alternatively, for
instance, acidic detergent is used at previous time
maintenance, use alkaline detergent at this time maintenance.
Wipe the external walls of sample&reagent probe lightly with cotton
stick moisturized with alcohol, especially the point of probe, until no
impurities left at all.
4 Wipe sample&reagent probe with the gauze dipped with deionized water.
5 After cleaning, lift the rotating arms of sample&reagent probe to the top
position, and rotate the rotating arm of sample&reagent probe to locate the
sample&reagent probe above the rinsing bath of sample&reagent probe.
Caution:
After cleaning the surface of a sample&reagent probe, please
make sure sample&reagent probe must be rotated to the top
of sample&reagent probe rinsing bath.
60
6 Switch on the power of analysis part and wait 30 seconds, enter the
"maintenance - routine maintenance" column to implement “instrument
resetting", the system will automatically reset the sample&reagent probe and
rinse them with deionized water.
2 Lift the stirring rod by hands to the top position, and rotate its rotating arm to a
position for convenient operation.
Caution:
When cleaning, do not touch the stirring rod surface directly to
prevent scratch; avoid too much hand force to prevent deformation
of the sample&reagent probe.
Note:
Acidic and alkaline detergent can be used alternatively, for instance,
acidic detergent is used at previous time maintenance, use alkaline
detergent at this time maintenance.
Wipe the surface of stirring rod lightly with cotton stick moisturized with
alcohol, especially the point of probe, until no impurities left at all.
4 Wipe stirring rod with the gauze dipped with deionized water
5 After cleaning, lift the rotating arm of stirring rod to the top position, and rotate
the rotating arm of stirring rod to locate the stirring rod to the top of the rinsing
bath.
61
6 Switch on the power of analysis part and wait 30 seconds, enter the "maintenance
- routine maintenance" column to implement “instrument resetting", the system
will automatically reset the sample&reagent probe and rinse them with deionized
water.
Caution:
Do not gaze scanning laser light, or it may cause eyes injury
Warning:
Please be careful to avoid being scratched by probe.
62
Biological contamination danger:
In operation, please put on gloves, work cloths, and put on
protective glasses for the best.
Do not dispose the gauze used to clean sample&reagent
probe.rinsing bath at your own will, please follow the relevant
provisions for proper disposal.
2 Lift the rotating arm of sample&reagent probe by hands to the its top position, and
rotate its rotating arm to keep reagent&sample probe away from rinsing bath for
convenient operation.
3 Clean the inside and appearance of sample&reagent probe rinsing bath with clean
cotton stick.
4 After cleaning, lift the rotating arm of sample&reagent probe to the top position,
and rotate the rotating arm of sample&reagent probe to locate the probe to the top
of the rinsing bath.
Caution:
After the work of sample&reagent probe surface rinsing,
please make sure to rotate reagent&sample probe to the top
of reagent&sample probe rinsing bath.
5 Switch on the power of analysis part and wait 30 seconds, enter the "maintenance -
routine maintenance" column to implement “instrument resetting", the system will
automatically reset the sample&reagent probe.
Warning:
Please be careful to avoid being scratched.
Biological contamination danger:
In operation, please put on gloves, work cloths, and put on
protective glasses for the best.
Do not dispose the gauze used to clean stirring rod at your own
will, please follow the relevant provisions for proper disposal.
63
7.5 Every 6 months maintenance
7.5.1 Check light source lamp
Lamp light source of optical system will gradually be aged in use, and will cause an increase in noise
during measurement. If the cuvette blank and light source intensity attenuation is out of range or the
working time of light source lamp accumulates over 2000 hours, the light source lamp should be
checked.
Caution:
Do not touch the light source lamp shell surface and lens in front of
the light source lamp by hand, because it may change the characteristics
of the light source. If you accidentally make noodle stained with filth,
absorbent cotton dipped by absolute alcohol can be used to clean it.
1 Turn off the system main power, so that the light source box and light source
lamp will be cooled for at least 15 minutes.
Warning:
2 Loosen the fixing screws of rinsing mechanism; remove the rinsing mechanism
of reaction cuvette. Loosen the setscrews of reaction disk and remove the
reaction disk. Place the reaction cuvette at dry and clean position. Loosen the
two fixing wire connecting poles of halogen and remove down-lead.
64
3 Loosen the two screws fixing light source seat to remove halogen lamp.
4 Mount a new halogen lamp according to the above opposite steps; pay attention to
tighten the screws. The cooling rubber hose in the lamp room can not be twisted and
down-lead can not be loosed or cocked.
5 Remount the reaction disk, the reaction cuvette and rinsing mechanism; switch on
the power supply of analysis part. After standby mode, single-click “Next” in
“System maintenance” window; infuse purified water into reaction groove. After
instrument standby mode, execute light quantity check function. Check the back of
halogen lamp if the light quantity conforms to the requirement to start test.
Warning:
Please be careful to avoid being scratched.
65
1 Make sure the analysis part power is switched off.
3 Lift the rotating arm of sample&reagent probe by hands to its top position, and
rotate its rotating arm to keep sample&reagent probe away from rinsing bath for
convenient operation.
4 Hold shell claw of probe rotating arm with fingers and lift to remove.
5
6 Loosen pipeline interface
7 warning:
Carefully place dismounted sample reagent&probe and prevent it
scratching human body and sample&reagent probe damage.
Note:
Take out sample reagent&probe from the rotating arm and be
careful to operate to avoid the damage of probe point caused by
touching rotating arm.
66
Note:
Sample&reagent probe is precisely processed to ensure the sample and
reagent adding precision. If the probe point is damaged or bent, checking
sample&reagent probe is a must, or no guarantee can be made for test
precision.
Warning:
Please be careful to avoid being scratched by probe.
Biological contamination danger:
In operation, please put on gloves, work cloths, and put on protective glasses
for the best.
Do not dispose the gauze used to clean sample&reagent probe at your own
will, please follow the relevant provisions for proper disposal.
1 Put a stainless steel wire through the sample&reagent probe point to clean the
impurity in the probe.
Caution:
Sample&reagent probe is precisely processed to ensure the sample and
reagent adding precision. If the probe point is damaged or bent, checking
reagent&sample probe is a must, or no guarantee can be made for test
precision.
Warning:
Please be careful to avoid being scratched by probe.
67
Caution:
Sample&reagent probe is precisely processed to ensure the sample and
reagent adding precision. If the probe point is damaged or bent, checking
sample reagent probe is a must, or no guarantee can be made for test
precision.
Warning:
In operation, please put on gloves, work cloths, and put on protective glasses
for the best
2 Lift the rotating arm of sample&reagent probe by hands to its top position,
and rotate its rotating arm to keep sample&reagent probe away from rinsing
bath for convenient operation.
5 Lift the rotating arm of sample&reagent probe by hands to its top position,
and rotate its rotating arm to keep sample&reagent probe above the rinsing
bath of sample&reagent probe.
Caution:
Please rotate the sample&reagent probe to the top of sample&
reagent probe rinsing bath after clean rinsing bath of sample
&reagent probe.
68
If the stirring rod is damaged, please check stirring rod in accordance with following steps strictly.
Warning:
5 Prepare a new stirring rod, and wipe the front of the stirring rod with gauze
dipped with 2% -antibacterial detergent.
69
6 When mounting new stirring rod, insert stirring rod till the bottom of motor axis
and tighten it with M2 screw.
Caution:
When inserting stirring rod, make sure the direction of force is
vertical to that of axis of rotating arm. Lateral force may damage the
axis or stirring rod.
Pushing the stirring rod completely
7 After stirring rod check, visually check whether stirring rod and its rotating arm
are vertical with each other.
If not vertical, return to step 5 and remount the stirring rod.
If vertical, continue to next.
8 Lift the rotating arm of stirring rod by hands to its top position, and rotate its
rotating arm to the top of its rinsing bath.
Caution:
Please make sure to rotate stirring rod to its rinsing bath top after
mount it.
9 Switch on the power of analysis part and wait 30 seconds, enter the
"maintenance - routine maintenance" column to implement “instrument
resetting", the system will automatically reset the sample&reagent probe and
rinse it with deionized water. Observe the outflow of reagent&sample probe.
7.8.6 Check reaction cuvette
Warning:
Please be careful to avoid being scratched by sample&reagent probe.
Place each probe and pole into proper position for convenience.
Biological contamination danger:
In operation, please put on gloves, work cloths, and put on protective glasses
for the best
Please deal with removed reaction cuvette properly which is broken.
Caution:
Please use consumables recommended by DIRUI company, using other
consumables may cause system performance degradation.
70
2
4 Rinse the new reaction cuvette dipped in with water; rinse inside and outside
of reaction cuvette and no scratch is allowed.
5 Rotate reaction disk by hands, and mount new reaction cuvette on reaction
disk and check the six sets reaction cuvette simultaneously.
6 Mount reaction disk with the opposite steps and make sure the fixing screw of
reaction disk is tight.
7 Switch on the power of analysis part
71
Select and execute “cuvette blank test” after click “System maintenance”, and
observe execution result and reaction status.
The analysis principle of auto-chemistry analyzer is based on the Beer-Lambert law that material
absorpts light selectively.
The main principle is: When monochromatic light with specific wavelength passes through the cuvette
with sample, the monochromatic light absorbency and sample liquid concentration vary in positivw
proportion as the distance of the light possing through the liquid.
I0
A = lg(1/T)= lg( )= ε b c
It
A -Absorbency of the light when passes through liquid
I0 - Incident intensity
It - Transmitted intensity
Solution layer thickness (b): Optical path, which is fixed by instrument. Molar absorption
coefficient (ε) is the correlation coefficient of the wavelength, solution and solution temperature.
Linear relationship is displayed between solution thickness and absorbency when in stable temperature
and single wavelength(ε value is given on the reagent bottle by factory)
72
If the sample liquid adequate distribution, interaction between liquid and incidence
monochromatic light only happens during absorbing process. No fluorescence, disperse and
photochemical appear. No interaction between substances in the solution while absorbing process. The
absorbency possess conducts nature, and this condition conforms to the Beer-Lambert law
1-point
L– 0– 0– 0 B1 + B 2 + B 3 AL + AL −1
End point
analysis 1<L≤49 3 2
method
2-points L– M– 0- 0 B1 + B 2 + B 3 ( AM + AM −1 ) − k ( AL + AL −1 )
analysis
1<L<M≤49 3 2
method
Time
2-points AM + AM −1 AL + AL −1 (minutes)
B1 + B 2 + B 3 −
Rate analysis L– M– 0– 0 3
2 2 between
method t metering
1<L<M≤49
points L,M
L– M – 0 - 0
B1 + B 2 + B 3
Rate method A 1<L<M≤49 △A(M-L)
3
L +2<M
Fist half
L– 0 – 0 – 0 B1 + B 2 + B 3 AL + AL −1
3 2
1 < M < N≤L < P <
1-point rate Q≤49
double items
Second half
analysis
method M – N – P – Q
B1 + B 2 + B 3
1 < M < N≤L < P < △(AQ- P)-k△(AN -M)
3
Q≤49
M+2<N ,P+2<Q
73
Fist half
B1 + B 2 + B 3 AL + AL −1
L– 0 – 0 – 0
3 2
3-points 1<L≤M<N≤49
double items Second half
B1 + B 2 + B 3 ( AN + AN −1 ) − k ( AM + AM −1 )
M – N – 0 – 0
3 2
1<L≤M<N≤49
Fist half
L– M – 0 - 0 B1 + B 2 + B 3
△A(M-L)
3≤L<M<N<P≤49 3
Rate B double
items L +2<M
analysis
When it is different from the
method Assay Second half first half part wavelength of
(mode 1 ) the project △ A (P-N)
N – P – 0 – 0 B1 + B 2 + B 3 When it is the same with the Two conditions
3≤L<M<N<P≤49 3 second part of wavelength of
the project: △A (P-N)–k
N+2<P △A(M-L)
L– M – 0 – 0 B1 + B 2 + B 3
△A(M-L)
3≤L<M<N<P<Q< 3
R≤49
Rate B double
items L +2<M
analysis
method Assay Second half part of
(mode 2 ) the project
N – P –Q – R B1 + B 2 + B 3
△A(R-Q)–k△A(P-N)
3≤L<M<N<P<Q< 3
R≤49
N+2<P ,Q+2<R
74
L,m,n,p,q,r : Photometric points
(B1,B2,B3 )/3 :Average value of three times passing through cell blanks
S :Sample volume
Rj 、Ri: a is the volumn of reagents without correction
Note 1: After adding reagent 2, the 5th metering point does not immediately stir. But after the reaction
disk rotates one circle around plus 2 reaction cuvettes and then pause, it need to rotate another 22
reaction cuvettes, after which will pause and then stir.
Note 2: During photometry , the reaction liquid should be more than or equal to 150μL, and less than
or equal to 450μL.
Note 3: Be sure to enter "0(zero)" when there will be no photomitric point.
75
bs
A
Cell blank
Time
76
conversion
Reference Adult;5.1~19umol/L(0.3~1.1mg/dL)
value Newly born:20~200umol/L(1.2~12mg/dL)
Two points analysis method (fixed time assay) is also named as first class
dynamics analysis method, which means during a certain period time of reaction,
reaction speed is in direct proportion to the simple power of substrate
concentration in specified time, namely v=k[S]. Due to the reduction of
substrate, the whole reaction speed is decreasing gradually, which shows the
increase (decrease) of absorbance. Because it takes a very long time to reach
balance, it can be monitored at any time theoretically, but due to the
complexity of serum ingredient, it must take a certain period of time to enter
into stable reaction phase.
Absorbance
Cell blank
Time
Time
77
Two points end point analysis two points rate analysis
In fact, because substrate concentration is not high enough, with the reaction
proceeds, the reaction is no longer zero class when substrate is consumed to
a certain extent, Therefore, zero-class dynamics analysisn is targeted at a
certain period of time; Because reaction time to reach balance is very long,
it can be monitored at any time theoretically, but because of the complexity
of serum ingredient and much reaction, therefore, it takes a certain period
of time to enter in stable reaction phase. So all reagent manufactures have
strict requirements to thesse two time periods
Metering point in accordance with the input form, dynamics method can be
divided into single-band and dual-band dynamics analysis.
Cell blank
Time
78
Example 3:ALT/GPT - Alanine aminotransferase (IFCC)
Calculating method:
�
A/min * TV *1000
ALT(U/L)=
6.22 * SV * P
SV=sample volume(mL)
79
8.6 Principle of electrolyte measurement
8.6.1 Principle
RT
E =E 0 + 2.303 × × l og(ai )
nF ………………………………………………………(1)
ai= f × Ci ……………………………………………………………………………(2)
80
T: absolute temperature (t ℃ +273.15) (K)
f: activity coefficient
Ci: concentration
Measure low concentration slope of liquid (S1) and high concentration slope
liquid (S2), and determine slope value (sensitivity) of K, Na, Cl the
electrode.
E ( H ) − E ( L)
SL =
C(H )
log
C ( L) …………………………………………………… (3)
E ( IS ) − E ( L )
C (=
IS ) C ( L) ×10 SL
…………………………………………… (4)
81
C (IS): the concentration of internal standard solution
E ( S ) − E ( IS )
=
C ( S ) C ( IS ) ×10 SL
………………………………………… (5)
=
C (VALUE ) C (C ) − C ( X ) ………………………………………… (6)
C ' ( S ) IF {C ( S ) + C (VALUE )}
=
.. ... ... ... ... ... ... ... ... ... ... ..
. ... ... (7)
Item specification
82
Diluent volume 600ul
Processing
100samples/h(only measuring electrolyte)
ability
Na + 20 ~ 200mmol / L (when only serum )
10 ~ 400mmol / L (when measuring urine)
K + 1.0 ~ 15.0mmol / L (when only serum)
Measuring
1 ~ 200mmol / L (when measuring urine)
range
Cl-20 ~ 200mmol / L (when only serum)
10 ~ 400mmol / L (when measuring urine)
Note:
Internal standard solution will be added one time in order to activate the
electrode if there is no electrolyte analysis for more than 10 minutes.
83
Chapter 9 Troubleshooting List
Alarm Detailed
Description Solution
Code description
Solution:
84
2、No change on multimeter reading
solution:Mechanical repair
85
mechanical repair is required if resistance is big.
Solution:
1) Check conductors
1) Check conductors
86
In CS-T240 software, enter into the "system
maintenance" interface after on-line, implement
"mechanical movement check", and observe the
stirring rod running status
Malfunction 1:stirring mechanism doesn’t move
Solution:
87
When
resetting,
Stirring Stirring
1-6 mechanism mechanism The solution is the same as 1-4
abnormal failed to
reach rinsing
bath.
Stirring
mechanism
Stirring
failed to If stirring mechanism failed to reach the top when
1-8 mechanism
reach the top rotating,the solution is the same as 1-1
abnormal
when
rotating.
Alarm Detailed
Description Solution
Code description
Solution:
Rinsing
Rinsing 1, Check rinsing mechanism sway movements,
mechanism of
mechanism of mechanical repair is required if resistance is big.
3-1 reaction cup
reaction cup
fails to reach the
abnormal 2, check whether both ends of connector of the motor are
top
connected well.
88
Malfunction 2:Rinsing mechanism can reach the top,
but can not check the zero position.
1) Check conductors
Check 1,2,3 P02 core plug is connected well and good
connections at both ends is connected
solution:Mechanical repair
89
Malfunction 2:Rinsing mechanism of reaction cup
cannot reach the top
90
9.3 Reaction disk malfunction analysis
Alarm Detailed
Description Solution
Code description
91
9.4 Sample probe mechanism malfunction analysis
Alarm Detailed
Description Solution
Code description
Alarm Detailed
Description Solution
Code description
92
can run, but the photo sensor cannot detect any signal
Solution: Manually make the probe body go up to the
peak repeatedly, transfer the multimeter to DC, negative
points the first 6-core plugs P042, P042 positive pole core
plugs 5.
1, multimeter readings between the positive changes from
5V to 0, indicating photo sensor and lead are normal
(1) Check conductors
Check whether the sample probe’s photo sensor J202 to
the reagent disc circuit board (P042) is in good
connection, and whether the both ends of the connectors
are good.
(2) Check the input of the sample reagent disc circuit
board’s photo sensor.
2, No change on multimeter reading
(1) Check conductors
Check P042 plugs 4,5,6 core is connected well,
connection of both ends is good.
(2) Check the photo sensor
Check voltage between P042 pin plugs 4,6, if the voltage
is not 5v, check the connection plug is connected well. If
the voltage is 5v, check the J042 socket 11 feet’s potential.
Rising to the peak when the potential is high, if not, the
potential is low. If the potential is not normal, check the
photo sensor.
93
4, check the sample&reagent motor drive circuit board
94
2. No chang on multimeter readings
95
5, Check if the up down and back to zero of photo sensor
are working normally
96
Solution:same as to 8-5
When
Sample&reage
Reagent probe nt probe
8-7 Probe couldn’t rise to the peak, solution same as to 8-1.
abnormal rotates, it
deciates from
the peak
Solution:
it couldn’t 1, Check the liquid level detection reagent with the sample
Reagent probe detect liquid plate to connect the circuit board is the connection of a
8-8
abnormal surface good guide, the connector ends are good
existence
2, check the liquid level detection circuit board, test the
sensitivity of the sample reagent is normal needle
Alarm Detailed
Description Solution
Code description
97
location well
3, Check the conductivity of the electrical conductor lead
is good
4, Check sample&reagent circuit board’s motor drive
circuit board kit part
Malfunction 2: When sample&reagent disc rotates to the
specified location, the sample&reagent disk does not stop
at the specified position, or when turns to specified
location, the photo sensor couldn’t detect it, or the signal
of the photo sensor is not correct
Solution:
98
well of this mechanism
3, Check whether the conductivity of the conductor’s lead
are good
4, Check the circuit board of the sample reagent disk
Sample&reagen Couldn’t find 1, Check whether the barcode reader is well connected to
t disk barcode sample the instrument
10-4
reader reagent
abnormal barcode reader 2, Check whether the barcode reader is broken
Please reset the barcode reader, if it still doesn’t work,
check barcode reader. If it starts to work normally,
indicates that there is no problem on the barcode reader. It
might be the problem of digital wire or circuit board.
99
9.12 Injection pump malfunction analysis
Alarm Detailed
Description Solution
Code description
Solution:
100
9.14 Incubation bath malfunction analysis
Alarm Detailed
Description Solution
Code description
The
temperature of
the incubation (1) Make sure the room temperature is in the scope of
bath is out of 15-32
the scope of
Incubation bath (2) Make sure the cooling fan of the instrument is rotating
20-2 temperature 37℃ ±5℃
narmally
abnormal
(Only check
(3) Make sure the water is cycling in the incubation bath
when the
instrument (4) Replace main control board or the temperature sensor
under
operation)
ISE QC is
ISE is over over (1)Please check whether the Control
0-32 Note measuring measuring volume, reagent volume is adequate, and
upper limit upper limit, whether the position is correct
the current
QC has no
101
result.
ISE
Calibration
is over (1)Please check whether the Calibrator
ISE is over measuring volume, reagent volume is adequate, and
0-33 Note measuring lower limit, whether the position is correct
lower limit the current
Calibration
has no
result.
ISE
Calibration
is over
ISE is over measuring (1)Please check whether the Calibrator
0-34 Note measuring upper limit, volume, reagent volume is adequate, and
upper limit the current whether the position is correct
Calibration
has no
result.
ISE sample
is over
ISE is over measuring (1)Please check whether the sample
0-35 Note measuring upper or volume, reagent volume is adequate, and
range lower limit, whether the position is correct
no test
result.
Electrolyte
system
stops
working
due to the (1)Enter into system maintenance, and
issued execute "Resetting”, and then execute
Electrolyte
19-1 Note function stops
alarm. “Mechanical movement check”
Note: the
stop status
of adding
sample
means to
restart.
102
volume)
remaining
volume of
internal (1)add ISE internal standard solution
ISE internal standard
(2)Enter into system maintenance, and
standard solution is
38-1 Note liquid is inadequate execute "ISE internal standard solution
inadequate (less then reagent pipeline rinsing”
the user
(3)Execute ISE calibration
designed
volume)
remaining
volume of (1)add ISE diluent
diluent is
ISE diluent is (2)Enter into system maintenance, and
inadequate
39-1 Note (less then
execute "ISE diluent reagent pipeline
inadequate rinsing”
the user
designed
(3)Execute ISE calibration
volume)
In the
5-measurin
g point
potential of
internal
standard (1) Enter into system maintenance
solution, the window, and execute "ISE check."
average
value
ISE LEVEL
60-1 Note (2) please refer to " electrolyte device
error (EAV)of maintenance " in the "user manual" for
three detail.
potentials is
over the
following
range
(internal
standard
solution)K:
-90.0mv≤E
AV≤-10mv
103
potential of (2) please refer to " electrolyte device
internal maintenance " in the "user manual" for
standard detail.
solution, the
average
value
(EAV)of
three
potentials is
over the
following
range
(internal
standard
solution)K:
-90.0mv≤E
AV≤-10mv
In the
5-measurin
g point
potential of
internal
standard
solution, the
average
value (1) Enter into system maintenance, and
execute "ISE check."
(EAV)of
ISE LEVEL three
60-3 Note potentials is (2) please refer to " electrolyte device
error maintenance " in the "user manual" for
over the
following detail.
range
(internal
standard
solution)
Cl:
80.0mv≤EA
V≤160mv
In the
(1) Enter into system maintenance, and
5-measurin
execute "ISE check."
ISE Noise g point
61-1 Note
error
potential of
(2) please refer to " electrolyte device
internal
maintenance " in the "user manual" for
standard
104
solution, the detail.
difference
(FI) of
the
maximum
and
minimum is
within
following
range
Na:
0.7mv≤|FIV
(2)-FIV
(4)|
In the
5-measurin
g point
potential of
internal
standard
solution, the
difference
(FIV) of (1) Enter into system maintenance, and
the execute "ISE check."
maximum
ISE Noise
61-2 Note and
error minimum is (2) please refer to " electrolyte device
within maintenance " in the "user manual" for
following detail.
range
(internal
standard,
sample)
K:
1.0mv≤|FIV
(2)-FIV
(4)
In the
5-measurin (1) Enter into system maintenance, and
g point execute "ISE check."
105
maximum
and
minimum is
within
following
range
(internal
standard,
sample)
Cl:
0.8mv≤|FIV
(2)-FIV
(4)|
The slope
value of
calibration
result is
within
following
range or the
impact of
electrode is
low (1) Enter into system maintenance, and
(cross-conta execute "ISE check."
ISE mination
62-1 Note
Prepare rate (A) is (2) please refer to " electrolyte device
as maintenance " in the "user manual" for
abnormal following detail.
Na:(1)
32.0mv≤SL
OPE≤37mv
or
68.1mv≤SL
OPE
The slope
value of
calibration (1) Enter into system maintenance, and
result is execute "ISE check."
ISE within
following
62-2 Note Prepare range or the
(2) please refer to " electrolyte device
maintenance " in the "user manual" for
abnormal impact of
detail.
electrode is
low
(cross-conta
mination
rate (A) is
106
as
following
K:(1)
32.0mv≤SL
OPE≤37mv
or68.1mv≤S
LOPE
The slope
value of
calibration
result is
within
following
range or the
impact of
electrode is (1) Enter into system maintenance, and
execute "ISE check."
ISE low
(cross-conta
62-3 Note Prepare mination (2) please refer to " electrolyte device
abnormal rate (A) is maintenance " in the "user manual" for
as detail.
following
Cl:(1)
-30mv≤SL
OPE≤-25m
v
or-68.1mv≥
SLOPE
(1)The
slope value
of
calibration
result is
within
following (1) Enter into system maintenance, and
range or the execute "ISE check."
ISE slope impact of
value electrode is
(2) please refer to " electrolyte device
63-1 Note low
(SLOPE) (cross-conta
maintenance " in the "user manual" for
abnormal detail.
mination
rate (A) is
as (3) Please re-execute ISE calibration.
following:
Na:(1)
SLOPE<
32.0mv
(2)current
107
ISE
calibration
result is not
updated
(1)The
slope value
of
calibration
result is
within
following
range or the
impact of (1) Enter into system maintenance, and
electrode is execute "ISE check."
ISE slope low
value (cross-conta
(2) please refer to " electrolyte device
63-2 Note mination
(SLOPE) rate (A) is
maintenance " in the "user manual" for
abnormal detail.
as
following:K
:(1) (3) Please re-execute ISE calibration.
SLOPE<
32mv
(2)current
ISE
calibration
result is not
updated
(1)The
slope value
of
calibration
result is
within
following (1) Enter into system maintenance, and
range or the execute "ISE check."
impact of
electrode is
ISE SLOPE
63-3 Note low (2) please refer to " electrolyte device
abnormal (cross-conta maintenance " in the "user manual" for
mination detail.
rate (A) is
as (3) Please re-execute ISE calibration.
following:C
l:(1)
SLOPE>
-25.0mv
(2)current
ISE
108
calibration
result is not
updated
(1)the
concentratio
n of
internal
solution(C
(IS))is
within (1) Enter into system maintenance, and
following execute "ISE check."
ISE the
range :Na:
concentration (2) please refer to " electrolyte device
64-1 Note of internal C(IS)<
120.0mmol/ maintenance " in the "user manual" for
solution detail.
abnormity l or
190.0mmol/
(3) Please re-execute ISE calibration.
<C(IS)
(2)current
ISE
calibration
result is not
updated
(1)the
concentratio
n of
internal
solution(C
(IS))is (1) Enter into system maintenance, and
within execute "ISE check."
following
ISE the range .K:C (2) please refer to " electrolyte device
concentration (IS)< maintenance " in the "user manual" for
64-2 Note of internal 3.0mmol/l detail.
solution or 8.0mmol/
abnormity <C(IS)
(3) Please re-execute ISE calibration.
(2)current
IS
calibration
result is not
updated
109
abnormity (IS))is maintenance " in the "user manual" for
within detail.
following
range Cl:C
(3) Please re-execute ISE calibration.
(IS)<
80.0mmol/l
or
140.0mmol/
<C(IS)
(2)current
ISE
calibration
result is not
updated
Because of
the
implementa
tion of the
ISE
maintenanc
e (ISE
ISE required
cleaning, Executive ISE calibration
calibration
complete
cleaning), it
is necessary
to
implement
ISE
calibration
110
Alarm Detailed
Description Solution
Code description
solution:
111
Alarm Detailed
Description Solution
Code description
solution:
112
Alar
Detailed
m Description Solution
description
Code
(4) Inspect whether the rise and fall zero light sensor
is working well.
113
Check voltage between P142A plug pin 1,3, if the
voltage is not 5v, check the connection of plug. If the
voltage is 5v, check potential of P142 plug pin 2. The
potential is high when reaching to the top position,
otherwise the potential is low. If not normal, check
the light sensor
Alar
Detailed
m Description Solution
description
Code
16-2 (4) Inspect whether the rise and fall zero light sensor
The nozzle of SIP nozzle is working well.
electrolyte SIP fails to reach
Malfunction 2:The motor swing to the required
abnormal the top
position, but the signal is not detected by the light
sensor.
114
(2)Check the light sensor input circuit of sample
reagent disk circuit board.
Alar
m Detailed
Code Description Solution
description
Time
synchronizatio Sending time
143 Check the communications of control board and
n synchronization
-1 sub-boards respectively.
order failure
failure
115
failure.
Reaction disk
detects
143 Reaction disk malfunction Lower machine monitoring software monitors error
-4 reset failure when resetting, information, and analyze the reaction disk
Reaction disk board-related failure to eliminate the error
reset failure
Sample&
reagent disk
detects Use lower machine monitoring software to monitor
Sample& error information, and refer to the related
143 malfunction
reagent disk information of sample&reagent disk board to debug.
-5 when resetting,
reset failure
sample&
reagent disk
fails to reset
fails to detect
liquid level
143 liquid level
detection Refer to liquid detection error analysis to debug
-10 when adding
failure
detergent
116
Check whether the liquid level detector of reaction
bath is clean, whether there is water in reaction bath,
Reaction bath Reaction bath whether water level reaches the position where the
143
liquid level liquid level detector can detect, and check liquid level detector
-13
malfunction malfunction in reaction bath is working normally; check whether
relevant electrical units of liquid level detection and
control board are normal.
Bar code Bar code Use lower machine monitoring software to monitor
143
scanning scanning error information, and refer to the related error
-14
overtime error overtime error information of reagent&sample disks to debug.
143 Previous Previous stirring Use lower machine monitoring software to monitor
-24 stirringfailure failure error information, and refer to the related error
117
information of reaction disk to debug.
Sample
Sample barcode
barcode Use lower machine monitoring software to monitor
143 scanning
error information, and refer to the related error
scanning
-32 overtime in
information of sample board to debug.
overtime in
testing
testing
ISE detects
Observe whether there is ISE alarm information;
malfunction in
143 ISE reset
resetting;
Use lower machine monitoring software to monitor
-33 failure
resetting
error information, and refer to the related error
information of ISE board to debug.
failed.
ISE pipeline ISE pipeline Observe whether there is ISE alarm information;
143 rinsing is over Use lower machine monitoring software to monitor
rinsing
-35 specified time error information, and refer to the related error
overtime
information of ISE board to debug.
118
ISE malfunction
ISE Observe whether there is ISE alarm information;
143 when testing,
Use lower machine monitoring software to monitor
malfunction in
-36 follow-up ISE
error information, and refer to the related error
testing
stops
information of ISE board to debug.
Sending reagent
Sending
mapping
reagent
143 information
Refer to instrument module error analysis
mapping
-42 failure, adding
information
reagent may
failure
fail.
119
Check whether light source, water quality of
Continuous incubation bath, reaction cup and counting light
Continuous
143 emergence of 5 sensor of reaction disk are normal; test data
emergence of
-45 dirty cups collecting board lines by using the testing program
dirty cups
of data collecting board.
Alar
m Detailed
Code Description Solution
description
120
(1)observe whether current value displayed on
digital pipe of refrigeration system is normal.
refrigeration refrigeration
144-2 system current
(2)Make sure which current is abnormal, and cope
abnormal abnormal
with it after check the corresponding abnormal Peltier
and circuit board.
The 1st line The 1st line Check the 1st line refrigeration Peltier and cooling
refrigeration refrigeration chip
145-1
chip current <5A
malfunction
The 2nd line The 2nd Check the 2nd line refrigeration Peltier and
refrigeration line refrigeration chip
145-2
chip refrigeration
malfunction current <5A
9.18 AD Collector
Alarm Detailed
Description Solution
Code description
The 1st line The Check the AD collection board and preamp board
146-1 AD collector measuring
value of the
121
malfunction 1st line AD
collector is
over normal
range
The 2nd line The Check the AD collection board and preamp board
AD collector measuring
malfunction value of the
146-2 2nd line AD
collector is
over normal
range
The 3rd line The Check the AD collection board and preamp board
AD collector measuring
malfunction value of the
146-3 3rd line AD
collector is
over normal
range
The 4th line The Check the AD collection board and preamp board
AD collector measuring
malfunction value of the
146-4 4th line AD
collector is
over normal
range
The fifth line The Check the AD collection board and preamp board
AD collector measuring
malfunction value of the
146-5 fifth line AD
collector is
over normal
range
The sixth line The Check the AD collection board and preamp board
AD collector measuring
malfunction value of the
146-6 sixth line AD
collector is
over normal
range
122
The seventh The Check the AD collection board and preamp board
line AD measuring
collector value of the
146-7 malfunction seventh line
AD collector
is over
normal range
The eighth The Check the AD collection board and preamp board
line AD measuring
collector value of the
146-8 malfunction eighth line
AD collector
is over
normal range
The ninth The Check the AD collection board and preamp board
line AD measuring
collector value of the
146-9 malfunction ninth line AD
collector is
over normal
range
The tenth The Check the AD collection board and preamp board
line AD measuring
collector value of the
146-10 malfunction tenth line
AD collector
is over
normal range
The 11th line The Check the AD collection board and preamp board
AD collector measuring
malfunction value of the
146-11 11th line AD
collector is
over normal
range
The 12th line The Check the AD collection board and preamp board
AD collector measuring
146-12 malfunction value of the
12th line AD
collector is
123
over normal
range
124