Service Manual Surgi-Pro

This document provides a service manual for Surgic Pro+ and Surgic Pro surgical units. It contains: 1) A list of required tools for repairs and disassembly, including screwdrivers, pliers, and a wrist strap to prevent static damage. 2) Troubleshooting guidance with error codes and recommended remedies. 3) Step-by-step instructions for disassembling parts like the top case, motor cord harness, foot control harness, and USB memory module. 4) Additional information for addressing a blank screen, including removing the top case and a connector to access parts that may need replacement.

Uploaded by

dhcastano
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
2K views17 pages

Service Manual Surgi-Pro

This document provides a service manual for Surgic Pro+ and Surgic Pro surgical units. It contains: 1) A list of required tools for repairs and disassembly, including screwdrivers, pliers, and a wrist strap to prevent static damage. 2) Troubleshooting guidance with error codes and recommended remedies. 3) Step-by-step instructions for disassembling parts like the top case, motor cord harness, foot control harness, and USB memory module. 4) Additional information for addressing a blank screen, including removing the top case and a connector to access parts that may need replacement.

Uploaded by

dhcastano
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 17

SM-U0044 E

SERVICE MANUAL
For
Surgic Pro+ / Surgic Pro
Surgical Unit

Parts List No.


Surgic Pro+ (100V, 120V, 230V): G-229
Surgic Pro (100V, 120V, 230V): G-230

1 / 17
SM-U0044 E

CONTENTS

1. TOOLS REQUIRED………………………………………………….3

2. TROUBLESHOOTING…………….………………...………………4

3. DISASSEMBLY & ASSEMBLY PROCEDURE ………….……….9

4. ADDITIONAL INFORMATION FOR TROUBLESHOOTING……13

2 / 17
SM-U0044 E
1. TOOLS REQUIRED

2mm Hexagon Wrench


Pliers (For Surgic Pro+)
(For Surgic Pro+; to attach / detach
Screw Driver (+) USB Memory Module)

In addition, a tester and Grounded Static Control Wrist Strap are needed.

Caution: Please wear Grounded Static Control Wrist Strap to prevent any damage that static-
electricity might cause on the P.C.B.

Grounded Static Control Wrist Strap

3 / 17
SM-U0044 E
2. TROUBLESHOOTING
Please see below for Error code and remedy.

4 / 17
SM-U0044 E

5 / 17
SM-U0044 E

6 / 17
SM-U0044 E

7 / 17
SM-U0044 E

8 / 17
SM-U0044 E
3. DISASSEMBLY & ASSEMLY PROCEDURE

3-1. Removal of the Screws


Reverse the Control Unit.
Use Screw Driver (+) to remove the 4 Binding Head Screws from the Control Unit.

Bind Head Screw

Control Unit (reverse side)

3-2. Removal of the Top Case


Reverse the Control Unit to the right side.
Hold both sides of the Top Case and slightly lift the Top Case until the Connector Cord
(connecting the Display PCB and Main PCB) can be seen.

Top Case

Connector Cord

Take out the Connector Cord (connecting the Display PCB and Main PCB) from the Main PCB.

Connector Cord
Main PCB

9 / 17
SM-U0044 E
While holding the Top Case with one hand, take out the Connector Cord from the Top Case with
the other hand.

Connector Cord

Top Case Bottom Case Top Case

3-3. Removal of the Harness Assy for Motor Cord


Remove the Harness Assy for Motor Cord by pulling it straight up by hand.
When assembling, assemble the Harness Assy for Motor Cord to the position shown in the photo.
Push back the Harness Assy all the way in to position, since the Top Case may not be closed, if
not.

Harness Assy for Motor Cord

3-4. Removal of the Harness Assy for Foot Control


Remove the Harness Assy for Foot Control by pulling it straight up by hand.
When assembling, check that the Socket of the Harness Assy for Foot Control is assembled to
the direction shown in the below photo.

Harness Assy for Foot Control Socket of the Harness Assy


for Foot Control
10 / 17
SM-U0044 E
(For Surgic Pro+)
3-5. Removal of the USB Memory Module
Open the USB flash drive Socket by hand.
Use 2mm Hexagon Wrench to disassemble the Screw Bolt.

2mm
Hexagon
Wrench

Screw Bolt
USB flash drive
Socket

Use pliers to hold the Connector Cord (connecting USB Memory Module and Main PCB) at the
position shown in the photo. Pull the Connector Cord straight out from the USB Memory Module
(will need force to pull out). Be careful not to damage other parts when disassembling.

pliers Connector Cord

Connector Cord

In addition, take out 3 screws and Connector Cord to detach the USB Memory Module.

11 / 17
SM-U0044 E

Please refer to the Parts List for further information regarding the connection.

When assembling, reverse the procedure for disassembly.

12 / 17
SM-U0044 E
4. ADDITIONAL INFORMATION FOR TROUBLESHOOTING (Remedy for “Screen
is blank”)

4-1. Removal of the Top Case


Follow the steps written in the previous section 3-1 and 3-2 to remove Top Case.

4-2. Removal of the Connector (connecting Transformer to the Main PCB)


Please see the Parts List.
Take out the Connector “C”, attached to the Main PCB.

4-3. Remedy for “Screen is blank”


Follow the Steps 1 – 7, to determine which parts need to be replaced when “Screen is blank”.

13 / 17
SM-U0044 E

14 / 17
SM-U0044 E

15 / 17
SM-U0044 E

16 / 17
SM-U0044 E

17 / 17