VHP Decontamination Systems Comparison
VHP Decontamination Systems Comparison
Barrier Isolator
Decontamination Systems
Jim Fisher and Ross A. Caputo*
I
n recent years, the use of barrier isolators in a variety of ap-
plications has increased. In 1998, R. Friedman called isola-
tor technology a “promising technology [that] may repre-
sent a significant stride forward in aseptic processing” (1).
PHARMACEUTICAL SYSTEMS, INC.
Equipment operational differences to five hours to complete and must be done after approximately
The Clarus “C” and VHP 1000ED generators have many fea- 1000 g of hydrogen peroxide have been injected. The Clarus “C”
tures that are similar although not identical. Both store multi- unit uses refrigeration principles to withdraw the water from
ple cycles, provide printouts, have alarms and safeties, and can the isolator. This system does not require regular regeneration.
send and receive remote input and output. The most impor- Single versus dual air flow loops. The Clarus “C” has two airflow
tant differences in the operation of the two machines include loops. One loop includes the catalyst and refrigeration unit for
the methods for VHP and water removal; the use of single or removing moisture and VHP from the air, and is used for the
dual airflow loops; the availability of a parametric gassing op- conditioning and aeration phases. The second loop is for the
tion; and the use of a wet or dry cycle. ramping and decontamination phases. The second loop does
VHP and water removal. Both isolators use a heavy-metal cata- not include the catalyst or refrigeration units, so the VHP and
lytic converter to break the VHP down into water and oxygen. water content are not removed from the isolator. The genera-
The VHP 1000ED uses a desiccant dryer unit that absorbs and tor vaporizes additional VHP and injects it into the stream of
holds the water removed from the isolator until the unit is re- air that already has VHP entrained.
generated. When the unit goes through its regeneration cycle, The VHP 1000ED uses a single-loop system. As a result, the
it heats up the dryer and passes heated air through it to remove generator removes all VHP and a majority of the water content
all moisture from the desiccant material. This cycle takes four from the air stream that the generator draws from the isolator.
72 Pharmaceutical Technology NOVEMBER 2004 www.phar mtech.com
Wet versus dry decontamination cycle.
The major difference between the two
systems is that one uses a wet decon-
tamination cycle and the other uses a
dry cycle. The STERIS VHP 1000ED
operates on the principles that dry air
holds more VHP, and that the higher
the VHP concentration, the higher
the kill rate. Because the air inside the
isolator can only hold a finite amount
of VHP and water before condensa-
(a) (b) (c) tion begins, the VHP 1000ED dries
Figure 1: (a) The BioQuell Clarus “C” H2O2 gas generator, (b) the STERIS VHP 1000ED the air before gassing. This deconta-
biodecontamination system, and (c) the la Calhène isolator used in the study. mination method, known as the dry
cycle, allows the unit to produce a
higher VHP concentration without condensation.
The biodecontamination method of the BioQuell Clarus “C”
uses a wet-cycle with microcondensation. When hydrogen per-
oxide and water condense out of air they do not condense at
equal rates. In fact, the hydrogen peroxide condenses at a faster
rate than the water, creating a high hydrogen peroxide concen-
tration condensate. This condensate is believed to kill organ-
isms more quickly than VHP alone.
The microcondensation is purported to occur in extremely
small droplets that are invisible to the human eye, and is re-
quired over every surface of the isolator. A challenge in validat-
ing a wet cycle is showing that every surface has conditions suf-
ficient to yield microcondensation. Because of the high
saturation level of the air, and the fine line between micro- and
macro-condensation and temperature control of the room is
critical if a wet cycle is used.
Figure 2: Transfer isolator load.
Performance qualification (PQ)
The objective of PQ testing in an isolator decontamination ap-
plication is to demonstrate that the generator consistently biode-
contaminates the isolator using a specified cycle. During PQ,
decontamination is tested at the maximum and minimum iso-
lator load conditions.
The current aseptic processing guidance have left room for
interpretation in reference to the biological challenge during
validation, stating, “[Decontamination] cycles should be de-
veloped with an appropriate margin of extra kill to provide
confidence in robustness of the decontamination processes.
Normally, a four- to six-log reduction can be justified depend-
Figure 3: Flow trajectory plot. ing on the application” (5). Even though the type of biologic
challenge has not been addressed in the guidance, Geobacillus
During the ramping and decontamination phases, the unit in- stearothermophilus spores have become the industry-accepted
jects VHP into this stream of dry air to maintain the VHP con- organism for the decontamination challenge (6). A rather con-
centration within the isolator. servative approach was used in this biodecontamination vali-
Parametric gassing. The Clarus “C” has a parametric gassing dation. The cycles used in the validation are intended to pro-
option that allows the generator to monitor the decontamina- vide a minimum six-log reduction of a resistant biological
tion cycle and adjust parameters to meet decontamination ob- indicator (BI) when exposed to an overkill decontamination
jectives. BioQuell offers a condensation monitor that can au- cycle. G. stearothermophilus spores were used with a minimum
tomatically adjust the ramp gassing time based on the conditions initial concentration of 106 spores per indicator. The inactiva-
inside the isolator. This ensures the desired VHP and water con- tion of 106 spores at the three-quarter decontamination cycle
centration levels have been reached before the dwell-gassing time provides a theoretical total organism reduction of 108 or-
phase begins. ganisms for a full cycle.
74 Pharmaceutical Technology NOVEMBER 2004 www.phar mtech.com
of the cycle. In VHP decontamination, the surface
conditions of the load are the most critical factors
to consider because VHP is a surface-decontam-
ination process. The first surface condition to con-
sider is the compatibility of the surface material
with VHP. Ideally the load contents should not
react with, absorb, or adsorb the VHP. Secondly,
the more surface area the VHP has to interact with,
the longer it will take to decontaminate all of the
Figure 4: Velocity cut plots. (Left) 10.5 in. from rear wall. (Right) 40 in. from left wall. surfaces. Another concern is overloading the iso-
lator. If the load is packed too tightly, there may
be areas of low VHP concentration within the load resulting
from to reduced airflow through the load.
The final consideration for loading an isolator is to avoid sur-
face-to-surface contact between items or parts of the load it-
self, and between the load and the isolator. Areas where surface-
to-surface contact occurs are less likely to receive sufficient VHP
concentration, and therefore can pose a potential for contam-
ination. To help avoid areas of surface-to-surface contact, every-
thing should be placed on stainless steel wire racks.
The load used to test the VHP 1000ED consisted of forty-six
250-mL bottles of rinse fluid and samples, eight 100-mL vials
of media, two “Steritest kits,” scissors, forceps, pens, and a Dräger
pump with tubes. A similar load was used to test with the Clarus
Figure 5: Location of thermocouples, biological indicators, and “C” generator. The majority of the 250-mL bottles were placed
chemical indicators. on the bottom shelf of the two-shelf rack. The rest of the items
were placed on the top shelf. The loading configuration is shown
Table I: Settings used for each cycle during performance in Figure 2.
qualification.
Conducting CFD and smoke studies
Steris VHP 1000ED biodecontamination system
CFD is a finite element analysis for determining air patterns
Airflow 20 SCFM
and temperature distributions around or inside a given model.
Phase I: Dehumidification 15 min and RH ,4.6 mg/L The most recognizable use for this software is in the automo-
Phase II: Conditioning 2 min at 5.6 g/min bile and aerospace industries, where air patterns around a car
Phase III: Decontamination 34 min at 3.5 g/min or over the wing of an airplane can be modeled and analyzed.
Phase IV: Aeration 2 to 5 hours This technology can be used to analyze air patterns inside an
isolator. Once the patterns are established, the areas of worst
Bioquell Clarus “C” H2O2 gas generator flow or temperature can be found.
Airflow 500 L/min (~18 SCFM) There are four basic steps in conducting a CFD analysis on
Phase I: Conditioning 10 min and 40% RH an isolator. The first step is to create a three-dimensional solid
Phase II: Ramp gassing 15 min at 1.5 g/min model of the isolator, its components, and the placement of
Phase III: Dwell gassing 12 min at 1.1 g/min load inside the isolator. Once the system’s physical shapes and
Phase IV: Aeration 2.5 to 3 h dimensions are defined, the initial fluid conditions must be es-
The cycle times listed above are the full cycle times. During the tablished. These initial fluid conditions include properties such
PQ, the VHP 1000ED's Phase III and the Clarus's Phases II and III as temperature and flow rates of the inlet, the exhaust, any dis-
were run at three quarters of the times listed above. tribution fans, the isolator walls, and the air itself. When all the
initial conditions have been defined, the analysis is performed.
Several tests must be completed in an isolator PQ. The first The flow trajectories, the flow velocities, and the tempera-
step is to define the isolator load conditions for both minimum tures throughout the isolator are the three most important re-
and maximum loading. Next, the user must conduct a compu- sults of a CFD in an isolator application. VHP is distributed
tational fluid dynamics analysis (CFD) and/or smoke studies through the isolator by two methods: it can be directly car-
to determine the “worst case” airflow locations in the isolator. ried on the air stream, and it also can be spread by diffusion
The last step of the PQ is the biological challenge at both the from areas of high concentration to areas of low concentra-
minimum and the maximum loads to prove that the stated goals tion. It is preferable for the VHP to be carried on the air stream
of 106 spore reduction in a cycle time are met. to the decontamination site because it is distributed faster this
Defining the load. As with most other decontamination or ster- way.
ilization processes, load determination is critical to the efficacy The airflow trajectory plot generated from the CFD analy-
76 Pharmaceutical Technology NOVEMBER 2004 www.phar mtech.com
Total injected hydrogen peroxide vs time STERIS VHP1000ED and BioQuell Clarus "C" hydrogen
140 peroxide concentration vs time during the gassing phases
VHP 10000ED 2.0
Total injected peroxide (grams)
Figure 6: Total injected aqueous H2O2 versus time. Figure 7: VHP concentration versus time.
30
30
minimum-loaded conditions demonstrates that the load con-
28
tents had little effect on the VHP concentrations during the gassing
26
cycle.
24
The temperature profiles for the two generator systems are
22
shown in Figures 9 and 10.The temperature profiles for the two
20
generators are very similar. In other generators, the tempera-
18
0 15 30 45 60 75 90 105 120 135 150 165 180 195 210
ture increased continuously during the decontamination phase
Time (min)
and did not reach a steady state temperature until the aeration
phase.
Figure 10: Clarus “C” temperature profile.
Conclusion
After completing an installation qualification, operational qual-
Compared biodecontamination cycles ification, and performance qualification, and comparing the
The cycles that were programmed into each generator are listed operation and performance of the STERIS VHP 1000ED Bio-
in Table I. Aeration was continued until the isolator environ- decontamination System and BioQuell Clarus “C” H2O2 gas
ment was ,1 ppm VHP, as read by a Dräger pump and tube, generator, it can be concluded that both units can be validated,
to meet OSHA permissible exposure limit guidelines for accept- and are capable of effectively decontaminating an isolator. There
able hydrogen peroxide contact levels (14). Once the VHP level are differences between the two systems; however, these differ-
was below this limit, the cycle was considered complete and the ences do not affect the effectiveness of either unit to decontam-
BIs and CIs were harvested. inate an isolator system.
The first three phases (conditioning, ramping, and decont-
amination) of the VHP 1000ED totaled 62 min with an addi- References
tional 3 to 5h of aeration, for a total cycle time of 4 to 6h. The 1. R. Friedman, “Design of Barrier Isolators for Aseptic Processing: A
first three phases of the Clarus “C” totaled 37 min with 2.5–3h GMP Perspective,” Pharm. Eng. 18 (2), 28–33 (1998).
2. J.C. Lyda, “Regulatory Aspects of Isolator/Barrier Technology,” PDA
of aeration, for a total cycle time of 3 to 3.5. Because the Clarus Pharm. Sci. Tech., 49 (6), 200–304 (1995).
“C” has a shorter cycle time and lower hydrogen peroxide in- 3. C.M. Wagner and J. Raynor, “Industry Survey on Sterility Testing Iso-
jection rates, it uses less hydrogen peroxide during a cycle. Dur- lators: Current Status and Trends,” Pharm. Eng. 21 (2), 124–140 (2001).
ing the two three-quarter cycles, the Clarus “C” used 22.5 g of 4. J. Lysfjord and M. Porter, “Barrier Isolation History and Trends, A Mil-
hydrogen peroxide whereas the VHP 1000ED used 130.2 g. lennium Update,” Pharm. Eng. 21 (2), 142–145 (2001).
5. Food and Drug Administration, Guidance for Industry: Sterile Drug
Products Produced by Aseptic Processing–Current Good Manufacturing
Water and hydrogen peroxide concentrations and Practice (FDA, Rockville, MD, 2001).
temperatures 6. “Hydrogen Peroxide Vapour Biological Efficacy,” V2.5 (BioQuell
To monitor the concentration of VHP and water vapor in the Pharma, Andover Hampshire, United Kingdom, February 2003).
isolator, a guided wave vapor monitor was used. The adsorp- 7. “La Calhène Three Glove Transfer Isolator User’s Manual,” (La Cal-
Circle/eINFO 61
82 Pharmaceutical Technology NOVEMBER 2004 www.phar mtech.com