Scope User Guide
Scope User Guide
5000/6000/7000
Series Oscilloscopes
User’s Guide
  Agilent Technologies
Notices
© Agilent Technologies, Inc. 2000-2010          Warranty                                         receive no greater than Restricted Rights as
                                                                                                 defined in FAR 52.227-19(c)(1-2) (June
No part of this manual may be reproduced in
                                                The material contained in this docu-             1987). U.S. Government users will receive
any form or by any means (including elec-
                                                ment is provided “as is,” and is sub-            no greater than Limited Rights as defined in
tronic storage and retrieval or translation
into a foreign language) without prior agree-
                                                ject to being changed, without notice,           FAR 52.227-14 (June 1987) or DFAR
ment and written consent from Agilent           in future editions. Further, to the max-         252.227-7015 (b)(2) (November 1995), as
Technologies, Inc. as governed by United        imum extent permitted by applicable              applicable in any technical data.
States and international copyright laws.        law, Agilent disclaims all warranties,           Product specifications, characteristics, and
                                                either express or implied, with regard           descriptions in this document are subject to
Manual Part Number                              to this manual and any information               change without notice.
                                                contained herein, including but not
54695-97024                                     limited to the implied warranties of             Safety Notices
                                                merchantability and fitness for a par-
Edition                                         ticular purpose. Agilent shall not be
Fourth Edition, June 2010                       liable for errors or for incidental or               CAUTION
                                                consequential damages in connec-
Print History                                   tion with the furnishing, use, or per-           A CAUTION notice denotes a haz-
                                                formance of this document or of any
54695-97015, June 2009                                                                           ard. It calls attention to an operat-
                                                information contained herein. Should
54695-97018, January 2010
                                                Agilent and the user have a separate             ing procedure, practice, or the like
54695-97022, February 2010                                                                       that, if not correctly performed or
                                                written agreement with warranty
54695-97024, June 2010
                                                terms covering the material in this              adhered to, could result in damage
                                                document that conflict with these                to the product or loss of important
Printed in Malaysia                             terms, the warranty terms in the sep-
                                                                                                 data. Do not proceed beyond a
Agilent Technologies, Inc.                      arate agreement shall control.
1900 Garden of the Gods Road
                                                                                                 CAUTION notice until the indicated
Colorado Springs, CO 80907 USA                  Technology Licenses                              conditions are fully understood and
                                                The hardware and/or software described in        met.
A newer version of this manual
                                                this document are furnished under a license
may be available at                             and may be used or copied only in accor-
www.agilent.com/find/5000manual                 dance with the terms of such license.
www.agilent.com/find/6000manual                                                                     WA R N I N G
www.agilent.com/find/7000manual                 Restricted Rights Legend
                                                                                                 A WARNING notice denotes a
                                                If software is for use in the performance of a
Software Revision                                                                                hazard. It calls attention to an
                                                U.S. Government prime contract or subcon-
                                                tract, Software is delivered and licensed as     operating procedure, practice, or
This guide was written for version 6.1 of the
Agilent InfiniiVision 5000/6000/7000 Series     “Commercial computer software” as                the like that, if not correctly per-
Oscilloscope software.                          defined in DFAR 252.227-7014 (June 1995),        formed or adhered to, could result
                                                or as a “commercial item” as defined in FAR
                                                2.101(a) or as “Restricted computer soft-
                                                                                                 in personal injury or death. Do not
Trademark Acknowledgments                       ware” as defined in FAR 52.227-19 (June          proceed beyond a WARNING
Java is a U.S. trademark of Sun Microsys-       1987) or any equivalent agency regulation or     notice until the indicated condi-
tems, Inc.                                      contract clause. Use, duplication or disclo-     tions are fully understood and
                                                sure of Software is subject to Agilent Tech-
Sun, Sun Microsystems, and the Sun Logo                                                          met.
                                                nologies’ standard commercial license
are trademarks or registered trademarks of
                                                terms, and non-DOD Departments and
Sun Microsystems, Inc. in the U.S. and other
                                                Agencies of the U.S. Government will
countries.
Navigate to a topic
Reference Index
                1    Introduction     21
                                     Models Covered in this Manual        22
                                     Specifications and Characteristics        23
                                     Licensed Options      24
                                     Notes     26
                2    Initial Setup    27
                                     Package Contents 28
                                        5000 Series Oscilloscope Package Contents 29
                                        6000A Series Oscilloscope Package Contents 30
                                        6000A Option BAT Oscilloscope Package Contents          31
                                        6000L Series Oscilloscope Package Contents 32
                                        7000A Series Oscilloscope Package Contents 33
                                     Accessories Available      34
                                     Tilt the oscilloscope up for easy viewing     36
                                         To tilt the 5000 Series oscilloscope up for easy viewing    36
                                         To tilt the 6000 Series oscilloscope up for easy viewing    37
                                         To tilt the 7000 Series oscilloscope up for easy viewing    38
                                     To install the optional front panel overlay    40
                                         5000 Series Overlay 40
                                         6000 Series Overlay 41
                                         7000 Series Overlay 42
Caution indicator. 50
                                     AutoProbe Interface        72
                                     Passive Probes       73
                                     Active Probes      74
                                     Active Probes for 6000 Series 100 MHz Bandwidth Models          75
                                     Connect the Probes to the Oscilloscope              75
                4    Triggering     81
                                     Triggering - General Information      83
                                     Trigger Mode and Coupling Menu            84
                                     Auto and Normal Trigger modes 84
                                        Choosing Auto Trigger Mode or Normal Trigger Mode          84
                                        Auto Mode 85
                                        Normal Mode 85
                                     Trigger Level Adjustment        86
                                     Trigger Coupling     87
                                     Trigger Noise Rejection 87
                                         HF Reject 88
                                         LF Reject 89
                                         Noise Rejection 89
           Trigger Holdoff 90
               Trigger Holdoff Operating Hints      90
               To set holdoff 90
           External Trigger Input 92
              2-Channel Oscilloscope External Trigger Input       92
           5   Displaying   173
                            Tips for Displaying Waveforms            174
                            Stabilizing the Display        174
                            Interpreting the Display        175
                            Graphic Symbols in Softkey Menus               176
                            AutoScale 177
                               How AutoScale works 177
                               Undo AutoScale 177
                               Enabling Fast Debug AutoScale 178
                               Specifying the Channels Displayed After AutoScale              178
                               Preserving the Acquisition Mode During AutoScale               178
                            Pan and Zoom       179
                            Waveform Intensity and Signal Detail             180
                            Grid Intensity    181
                            Infinite Persistence      182
                            Using Labels     183
                            Waveform Expansion Reference Point               187
                            Vectors (Connect the Dots)           187
                            Freeze Display     188
                            Antialiasing     189
                            XGA Video Output         189
           14   Utilities   391
                                  To restore the oscilloscope to its default configuration        392
                                  To perform service functions       393
                                  User Calibration     394
                                  Self Test   397
           15   Reference    399
                                  About Oscilloscope      400
                                  Software and Firmware Updates            400
                                  Installed Licenses    400
                                  Upgrading to an MSO        401
                                  Secure Environment Mode Option             402
Index 421
1
Introduction
Models Covered in this Manual 22
Specifications and Characteristics 23
Licensed Options 24
Notes 26
The Agilent InfiniiVision 5000, 6000, and 7000 Series oscilloscopes deliver
powerful features and high performance:
•   100 MHz, 300 MHz, 350 MHz, 500 MHz, and 1 GHz bandwidth models.
•   Up to 4 GSa/s sample rate.
•   6.3 inch or 12.1 inch XGA display.
•   Powerful triggering including analog HDTV, I2C, I2S, SPI, LIN, CAN,
    FlexRay, MIL-STD 1553, and USB.
•   USB and LAN ports make printing, saving and sharing data easy. GPIB
    ports on 5000 and 6000 Series models.
•   2-channel and 4-channel Digital Storage Oscilloscope (DSO) models.
•   2+16-channel and 4+16-channel Mixed Signal Oscilloscope (MSO)
    models.
•   An MSO lets you debug your mixed-signal designs using up to four
    analog signals and 16 tightly correlated digital signals simultaneously.
•   You can easily upgrade an InfiniiVision 6000 or 7000 Series
    oscilloscope from a DSO to an MSO and/or add features such as
    Segmented Memory and Mask Test.
                                                                        21
1    Introduction
Licensed Options
                    Many of the following licensed options can be easily installed without returning the
                    oscilloscope to a Service Center. Not all options can be installed on all models. See data
                    sheets for details. To see the list of options installed on your oscilloscope, press
                    [Utility]&Options&Licenses&Show license information.
UPGRADE OPTIONS
                    232     RS232/UART serial decode option          Order N5457A after purchase (Option 232 at
                            (for 4 channel or 4+16 channel           time of purchase). You can easily install this
                            models only)                             option yourself.
                    553     MIL-STD 1553 serial decode option        Order N5469A after purchase (Option 553 at
                            (for 4 channel or 4+16 channel           time of purchase). You can easily install this
                            models only)                             option yourself.
                    ALT     N5434A FPGA dynamic probe for            N5434A with Option 001 (Oscilloscope-locked
                            Altera (MSO recommended)                 license) or Option 002 (PC-locked license).
                                                                     Software is installed on an external PC.
                    AMS     CAN/LIN automotive triggering and        Order N5424A after purchase (Option AMS at
                            decode (for 4 channel or 4+16            time of purchase). You can easily install this
                            channel models only)                     option yourself.
                    FPG     N5406A FPGA dynamic probe for            N5406A with Option 001 (Oscilloscope-locked
                            Xilinx (MSO recommended)                 license) or Option 002 (PC-locked license).
                                                                     Software is installed on an external PC.
                    FLX     FlexRay trigger and decode (for 4        Order N5432C after purchase (Option FLX at
                            channel or 4+16 channel models           time of purchase). You can easily install this
                            only). Includes mask limit test option   option yourself. FlexRay triggering and decode
                            (LMT), segmented memory option           option.
                            (SGM), and FlexRay physical layer
                            conformance test application option
                            (FRC)
                    LMT     Mask Limit Test                          Order N5455A after purchase (Option LMT at
                                                                     time of purchase). You can easily install this
                                                                     option yourself.
                       LSS      I2C/SPI serial decode option (for 4  Order N5423A after purchase (Option LSS at
                                channel or 4+16 channel models only) time of purchase). You can easily install this
                                                                     option yourself.
                       MSO      Mixed Signal Oscilloscope (MSO).          Order N2735, N2736A, or N2737A. You can
                                Upgrade a DSO to an MSO.                  easily install this option yourself. The logic
                                                                          cable kit is supplied with the MSO license.
                       PWR      U1881A Power Application                  Order U1881A. Application runs on PC
                                                                          connected to InfiniiVision oscilloscope.
                                                                          Optional U1880A deskew fixture available.
                       SEC      Secure Environment Mode                   Nonvolatile memory is cleared of all setup and
                                                                          trace settings in compliance with National
                                                                          Industrial Security Program Operation Manual
                                                                          (NISPOM) Chapter 8 requirements. Available at
                                                                          time of purchase only. See also page 402.
                       SGM      Segmented Memory                          Order N5454A after purchase (Option SGM at
                                                                          time of purchase). You can easily install this
                                                                          option yourself.
                       SND      I2S serial decode option (for 4 channel   Order N5468A after purchase (Option SND at
                                or 4+16 channel models only)              time of purchase). You can easily install this
                                                                          option yourself.
CALIBRATION OPTION
Option Order
Notes
                    Digital Channels
                    Because all of the oscilloscopes in the Agilent InfiniiVision Series have analog channels, the
                    analog channel topics in this book apply to all instruments. Whenever a topic discusses the digital
                    channels, that information applies only to Mixed-Signal Oscilloscope (MSO) models or DSO
                    models that have been upgraded to an MSO (available on the 6000 and 7000 Series oscilloscopes).
2
Initial Setup
Package Contents 28
    5000 Series Oscilloscope Package Contents 29
    6000A Series Oscilloscope Package Contents 30
    6000A Option BAT Oscilloscope Package Contents 31
    6000L Series Oscilloscope Package Contents 32
    7000A Series Oscilloscope Package Contents 33
Accessories Available 34
Tilt the oscilloscope up for easy viewing 36
To install the optional front panel overlay 40
To mount the oscilloscope in a rack 43
Ventilation Requirements 47
Power Requirements 48
Power-on the 5000/6000/7000 Series Oscilloscope 49
Power-on the Battery-Powered 6000A Series Oscilloscope 49
Verifying Basic Oscilloscope Operation 54
Quick Help 55
Graphical User Interface Languages 56
To set the clock 57
To set up the screen saver 58
Keys, Softkeys, and the Entry Knob 53
Using the Analog Channels 59
To set up the Horizontal time base 64
This chapter shows package contents for each oscilloscope model, and
explains how to set up the oscilloscope. Information about using the
analog channels and setting up the horizontal timebase is given. For an
overview of front panel controls see “Controls and Connectors” on
page 351.
                                                                        27
2    Initial Setup
Package Contents
                     ✔ Inspect the shipping container for damage.
                        If your shipping container appears to be damaged, keep the shipping container or
                        cushioning material until you have inspected the contents of the shipment for
                        completeness and have checked the oscilloscope mechanically and electrically.
                     ✔ Verify that you received the following items and any optional accessories you may
                       have ordered:
                        • InfiniiVision Oscilloscope
                        • Front-panel cover (all models except 6000L Series)
                        • Power cord (country of origin determines specific type)
                        • Oscilloscope probes
                           • Two probes for 2-channel models
                           • Four probes for 4-channel models
                        • CD-ROM containing:
                           • User’s Guide
                           • Service Guide
                           • Programmer’s Guide
                        • Automation-Ready Software CD-ROM
                        • Accessory pouch (7000 Series only)
                        • Digital probe kit P/N 54620-68701 (MSO Models only)
                        • Power Supply P/N 0950-4866 (6000 Option BAT only)
                        • LAN Crossover Cable 5061-0701 (6000L Series only)
                        • GPIB cable extender P/N 5183-0803 (6000L Series only)
                        • 50 ohm feedthrough termination adapter, Qty. 4, P/N 0960-0301 (DSO6014L only)
                   Front-panel cover
                                                              InfiniiVision 5000 Series
                                                              Oscilloscope
                                                                                  Oscilloscope probes
                                                                                  N2863A or 10073C
                                                                                  (Qty 2 or 4)
Documentation CD
   Automation-Ready CD
                                       Power cord
                                       (Based on country of
                                       origin)
                                                                                          Oscilloscope probes
                                                                                          10073C or 10074C
                                                                                          (Qty 2 or 4)
Documentation CD
          Power cord
          (Based on country of
          origin)
Front-panel cover
                                                                             Oscilloscope probes
                           Ground wire                                        10073C or 10074C
                                                                                 (Qty 2 or 4)
                                     Power cord
                                  (see Power Cords
     AC/DC power adapter                table)
            50 ohm feedthrough
            termination adapter
            P/N 0960-0301, Qty. 4        GPIB cable extender
                                         P/N 5183-0803
                                                                          Oscilloscope Probes
                                                                          10073C or 10074C
                                                                          Qty. 4
Documentation CD
     Automation-Ready CD
                                                         Power cord
                                                         (Based on country of
                                                         origin)
                                                                                                   10073C or 1165A
                                                                                                   probes
                                                                                                   (Qty 2 or 4)
Documentation CD
             Power cord
             (Based on country
             of origin)
Accessories Available
                     Model         Description
                     N2918A        6000/7000 Series Oscilloscope Evaluation Kit
                     N2916B        5000/6000 Series Oscilloscope Rackmount Kit
                     N2732A        7000 Series Oscilloscope Rackmount Kit
                     54684-44101   5000/6000 Front Panel Cover
                     54913-44101   7000 Front-Panel Cover
                     various       Front Panel Overlays - see page 40, page 41, page 42.
                     N2760A        5000 Soft Carrying Case
                     N2717B        5000 Transit Case
                     N2733A        7000 Soft Carrying Case
                     N2734A        Transit Case
                     1180CZ        Testmobile Oscilloscope Cart for 6000 Series (requires N2919A adapter kit)
                     N2919A        Testmobile Adapter kit for 6000 Series oscilloscopes
                     N2605A-097    USB Cable
                     10833A        GPIB Cable (5000 and 6000 models only)
                     5061-0701     LAN Crossover Cable
                     54620-68701   Digital Probe Kit (standard with MSO models)
                     54684-42301   Digital Probe Cable Guide (cable tray) 6000 models only
                     01650-61607   Logic Cable and Terminator (use with 40-pin logic analyzer accessories)
                     0960-0301     50-Ohm Termination Feedthrough (only for 6000 Series 100 MHz models)
                     10070C        Passive Probe, 1:1 20 MHz, 1.5 m
                     10074C        Passive Probe, 10:1, 150 MHz, 1.5 m
                     10073C        Passive Probe, 10:1, 500 MHz, 1.5 m
                     1165A         Passive Probe, 10:1, 600 MHz, 1.5 m
                     10076A        Passive Probe, 100:1, 4 kV, 250 MHz
                     N2863A        Passive Probe, 10:1, 300 MHz, 1.2 m
                     N2771A        Passive Probe, 1000:1, 30 kV, 50 MHz
                     N2790A        High-Voltage Differential probe, 50:1 or 500:1 (switchable), 100 MHz
                     N2786A        2-legged Probe Positioner
                     N2784A        1-arm Probe Positioner
                     N2785A        2-arm Probe Positioner
                     N2880A        InfiniiMax In-line Attenuator kit, pair of 6 dB, 12 dB, and 20 dB attenuators in kit,
                                   to be used with InfiniiMax probe amplifiers and heads
                     N2881A        InfiniiMax DC blocking caps (qty 2), withstand up to 30 V dc, to be used with
                                   InfiniiMax probe amplifiers and heads
                     N2882A        75 ohm-to-50 ohm adapter, dc to 8 GHz bandwidth, 5.7 dB attenuation
                       Model            Description
                       W2637A           LPDDR BGA probe, x16, 404 MHz, 100 ohm input impedance
                       W2638A           LPDDR BGA probe, x32, 404 MHz , 100 ohm input impedance
                       W2639A           Oscilloscope Adapter Board, 1.5 GHz, 75 ohm input impedance
                       N5450A           InfiniiMax Extreme Temperature Extension Cables, to be used with InfiniiMax
                                        probe amplifiers and heads, 92 cm (36 in)
                       N2791A           Differential probe, 1 MOhm termination, 10:1 or 100:1 (switchable), 25 MHz
                       N2792A           Differential probe, 50 Ohm termination, 10:1, 200 MHz, compatible with all
                                        InfiniiVision models except 6000 Series 100 MHz models (MSO/DSO601x)
                       N2793A           Differential probe, 50 Ohm termination, 10:1, 800 MHz, compatible with all
                                        InfiniiVision models except 6000 Series 100 MHz models (MSO/DSO601x)
                       1156A            Active Probe, 1.5 GHz
                       1144A            Active Probe, 800 MHz (requires 1142A – power supply)
                       1145A†           Active Probe, 750 MHz 2-ch (requires 1142A – power supply)
                       1130A†           For active differential probes: 1.5 GHz InfiniiMax amplifier (requires one or more
                                        InfiniiMax probe head – E2675A, E2668A, E2669A).
                       N2772A           Active Differential Probe, 20 MHz, 1.2 kVDC + peak AC max (requires N2773A
                                        power supply)
                       1141A            Active Differential Probe, 200 MHz, 200 VDC + peak AC max (requires 1142A
                                        power supply)
                       1146A            Current Probe, 100 kHz, 100 A, AC/DC
                       1147A†           Current Probe, 50 MHz, 30 A, AC/DC with AutoProbe interface
                       N2780A           Current Probe, 2 MHz, 500 A, AC/DC (use with N2779A power supply)
                       N2781A           Current Probe, 10 MHz, 150 A, AC/DC (use with N2779A power supply)
                       N2782A           Current Probe, 50 MHz, 30 A, AC/DC (use with N2779A power supply)
                       N2783A           Current Probe, 100 MHz, 30 A, AC/DC (use with N2779A power supply)
                       10072A           Fine-Pitch Probe Kit
                       10075A           0.5 mm IC Clip Kit
                       10076A           100:1, 4 kV 250 MHz Probe
                       E2613B           0.5 mm Wedge Probe Adapter, 3-signal, qty 2
                       E2614A           0.5 mm Wedge Probe Adapter, 8-signal, qty 1
                       E2615B           0.65 mm Wedge Probe Adapter, 3-signal, qty 2
                       E2616A           0.65 mm Wedge Probe Adapter, 8-signal, qty 1
                       E2643A           0.5 mm Wedge Probe Adapter, 16-signal, qty 1
                       E2644A           0.65 mm Wedge Probe Adapter, 16-signal, qty 1
                       †
                        Indicates a maximum of two of this model probe can be connected to each oscilloscope due to
                       AutoProbe interface current supply limitation. See also “Passive Probes” on page 73 and
                       “Active Probes” on page 74.
                         1 To rotate the handle, grasp the handle hubs on each side of the instrument and pull the
                           hubs out until they stop.
                         2 Without releasing the hubs, rotate the handle to the desired position. Then release the
                           hubs. Continue rotating the handle until it clicks into a set position.
                       The tilt tabs (underneath the oscilloscope) can be positioned as shown in the center
                       picture below. The handle can be used as a stand when placing the oscilloscope on a
                       floor, as shown in the picture on the right.
                         2 Without releasing the hubs, rotate the handle to the desired position. Then release the
                           hubs. Continue rotating the handle until it clicks into a set position.
                                                                                           Removal
                                                                                           Tab
                         To remove the label, carefully pull it away using the tab at the lower right corner.
                         The overlays will be available at www.parts.agilent.com using the following part
                         numbers:
                                                                                      Removal
                                                                                      Tab
                       To remove the label, carefully pull it away using the tab at the lower right corner.
                       The overlays will be available at www.parts.agilent.com using the following part
                       numbers:
                                                                                         Removal
                                                                                         Tab
                         To remove the label, carefully pull it away using the tab at the lower right corner.
                         The overlays can also be ordered separately at www.parts.agilent.com using the
                         following part numbers:
Step 4
If needed
                                                                                           Step 1,
                                                                                           step 5
Step 2
                     Step 3
                                                  If needed
                      1 Loosely attach the Front Extender Supports to the Rear Extender Supports with four
                        (4) clip-nuts and four (4) of the 10-32 x 0.375 Rail Screws. (The screws require a Torx
                        T20 driver.) Choose the correct set of slots in the supports such that their overall
                        length is approximately correct for the depth of your cabinet.
                      2 Fasten the Rack Mount Extenders to the oscilloscope chassis with the four (4) M3 x 6
                        mm screws, using a Torx T10 driver as follows:
                       The sets of holes in the Rack Mount Extenders are slightly offset. This was done to
       NOTE            ensure that the Rack Mount Extenders are attached to the oscilloscope at the correct
                       points so that the oscilloscope’s ventilation area is not obscured. The holes in the Rack
                       Mount Extenders will align with the correct holes in the oscilloscope and the screws will
                       go in easily. Do not force the screws into the wrong holes.
                          a Attach a Rack Mount Extender to the left side of the oscilloscope using two (2) of
                            the M3 x 6 mm screws in the inner set of holes on the Rack Mount Extender.
                          Use inner
                          holes in
                          extender
                                                                                     Use outer
                                                                                     holes in
                                                                                     extender
                          b Attach the other Rack Mount Extender to the right side of the oscilloscope using
                            two (2) of the M3 x 6 mm screws in the outer set of holes on the rack mount
                            extender.
                       3 Place the instrument in the rack. Install the four (4)
                         10-32 x 0.625 Dress Screws in the chassis front ears to secure the front of the
                         instrument to the rack. Use the Phillips screwdriver.
                       4 Align the ears in the Rear Mount Extenders with the correct set of holes in the rear of
                         the rack and secure the Rack Mount Extenders to the rack using the four (4) remaining
                         10-32 x 0.375 Rail Screws. Use the Torx T20 driver.
                       5 Securely attach the Rear Extender Supports to the Front Extender Supports by
                         tightening the four (4) 10-32 x 0.375 Rail Screws screws that you loosely attached in
                         step 1.
Ventilation Requirements
                       The air intake and exhaust areas must be free from obstructions. Unrestricted air flow is
                       required for proper cooling. Always ensure that the air intake and exhaust areas are free
                       from obstructions.
Power Requirements
                     5000, 6000A (without Option BAT), and 7000 Line voltage, frequency, and power
                     ~Line 100-120 Vac, 50/60/400 Hz
                     100-240 Vac, 50/60 Hz
                     120 W max
                       Always use a grounded power cord. Do not defeat the power cord ground.
    WA R N I N G
                                     54684A         1 GHz
                                     Oscilloscope   4 GSa/s
                                                                                                                                                     Mode
                                                                                               Quick           Acquire      Display        Edge     Coupling
                                                                                  Cursors      Meas
Digital Analog
Select
                                                               D15
                                                               Thru
                                                                D0
                                                                              1        Math                2             Label         3                         4
                                                                         AC                   AC                                 AC                   AC
                                                                         BW                   BW                                 BW                   BW
                                                                         50                   50                                 50                   50
Position
                                                                              1                            2       1M ~ ~ 14pF         3                         4
                                                                                                                   300 V RMS
                                                                         X                    Y                        CAT I                         Z
50 5 V RMS
54684e82
                       The 6000A Series Option BAT oscilloscopes are battery powered. They have additional
                       LED indicators on the front panel:
                     The battery will last approximately 1.75 hours before requiring recharging. Operating time
                     depends on the oscilloscope configuration.
Ground Post
You can use the oscilloscope while the battery is being charged.
                     Battery Replacement
                     The battery (actually two batteries) can be replaced by a qualified technician. Refer to the
                     service guide for replacement instructions. The service guide is available on the CD that
                     was included with the oscilloscope at time of purchase, and at
                     www.agilent.com/find/6000manual.
                     Battery Warranty
                     The battery is warranted for a period of one year from date of purchase.
                       The simplest way to set up the oscilloscope is to connect it to the signals of interest and
       NOTE            press the [AutoScale] key.
To find out how to use a key or connector, see “Controls and Connectors” on page 351.
                     If you see the waveform, but the square wave is not shaped correctly as shown above,
                     perform the procedure “Compensating Passive Probes” on page 77.
                     If you do not see the waveform, ensure your power source is adequate, the oscilloscope
                     is powered-on, the power cord is firmly inserted into the oscilloscope receptacle, and the
                     probe is connected securely to the front-panel oscilloscope channel input BNC and to the
                     Probe Comp terminal.
Quick Help
                       By default, Quick Help remains on the screen until another key is pressed or a knob is
                       turned. You can set Quick Help to close when you release the key. Press the [Utility] key,
                       then press the Language softkey, then press the Help Close on Release/Remain on
                       Screen softkey ([Utility]&Language&Help).
                            Year set       Month set      Day set       Hour set      Minute set    Return to
                                                                                                    previous menu
                       2 Press the Year, Month, Day, Hour or Minute softkey and rotate the Entry knob to set to
                         the desired number.
                       The hours are shown in the 24-hour format. So 1:00 PM is hour 13.
                       The real-time clock only allows selection of valid dates. If a day is selected and the month
                       or year is changed so the day is invalid, the day is automatically adjusted.
                     3 Turn the Entry knob to select the number of minutes to wait before the selected screen
                       saver activates.
                        When you turn the Entry knob, the number of minutes is displayed in the Wait softkey.
                        The default time is 180 minutes (3 hours).
                       4 Press the Preview softkey to preview the screen saver you have selected with the
                         Saver softkey.
                       5 To view the normal display after the screen saver has started, press any key or turn any
                         knob.
 Channel 1
 ground level
 Channel 2
 ground level
                        Pressing an analog channel key displays the channel’s menu and turns the display of
                        the channel on or off. The channel is displayed when the key is illuminated.
                     Vertical sensitivity     Turn the large knob above the channel key marked         to set the
                     sensitivity (volts/division) for the channel. The vertical sensitivity knob changes the
                     analog channel sensitivity in a 1-2-5 step sequence (with a 1:1 probe attached). The
                     analog channel Volts/Div value is displayed in the status line.
                     Fine Press the Fine softkey to turn on fine control V/div adjustment (vernier) for the
                     selected channel. When Fine is selected, you can change the channel’s vertical
                     sensitivity in smaller increments. The channel sensitivity remains fully calibrated when
                     Fine is on. The V/div value is displayed in the status line at the top of the display.
                     When Fine is turned off, turning the volts/division knob changes the channel sensitivity
                     in a 1-2-5 step sequence.
                     Vertical expansion      The default mode for expanding the signal when you turn the
                     volts/division knob is vertical expansion about the ground level of the channel. To set the
                     expansion mode to expand about the center of the screen instead, press Expand in the
                     User Preferences Menu ([Utility]&Options&Preferences) and select Center.
                     Ground level The ground level of the signal for each displayed analog channel is
                     identified by the position of the icon at the far-left side of the display.
                     Vertical position Turn the small vertical position knob ( ) to move the channel’s
                     waveform up or down on the display. The voltage value momentarily displayed in the
                     upper right portion of the display represents the voltage difference between the vertical
                     center of the display and the ground level ( ) icon. It also represents the voltage at the
                     vertical center of the display if vertical expansion is set to expand about ground.
                       Measurement Hints
                       If the channel is DC coupled, you can quickly measure the DC component of the signal by simply
                       noting its distance from the ground symbol.
                       If the channel is AC coupled, the DC component of the signal is removed, allowing you to use
                       greater sensitivity to display the AC component of the signal.
                       3 Press the channel’s on/off key, then press the Coupling softkey to select the input
                         channel coupling.
                          Coupling changes the channel's input coupling to either AC (alternating current) or DC
                          (direct current). AC coupling places a 3.5 Hz high-pass filter in series with the input
                          waveform that removes any DC offset voltage from the waveform. When AC is
                          selected, “AC” is illuminated on the front panel next to the channel position knob ( ).
                          • DC coupling is useful for viewing waveforms as low as 0 Hz that do not have large
                            DC offsets.
                          • AC coupling is useful for viewing waveforms with large DC offsets. When AC
                            coupling is chosen, you cannot select 50 mode. This is done to prevent damage
                            to the oscilloscope.
                       Note that Channel Coupling is independent of Trigger Coupling. To change trigger
                       coupling see page 87.
                       4 Press the Imped (impedance) softkey.
                          The analog channel input impedance can be set to either 1M Ohm or 50 Ohm by
                          pressing the Imped softkey.
                          • 50 Ohm mode matches 50-ohm cables commonly used in making high frequency
                            measurements, and 50-ohm active probes. This impedance matching gives you the
                            most accurate measurements since reflections are minimized along the signal
                            path. When 50 Ohm is selected, “50” is illuminated on the front panel next to the
                            channel position knob. The oscilloscope automatically switches to 1 M Ohm mode
                            to prevent possible damage if AC coupling is selected.
                        • 1M Ohm mode is for use with many passive probes and for general-purpose
                          measurements. The higher impedance minimizes the loading effect of the
                          oscilloscope on the device under test.
                     5 Press the BW Limit softkey to turn on bandwidth limiting.
                        Pressing the BW Limit softkey turns the bandwidth limit on or off for the selected
                        channel. When bandwidth limit is on, the maximum bandwidth for the channel is
                        approximately 25 MHz. For waveforms with frequencies below this, turning bandwidth
                        limit on removes unwanted high frequency noise from the waveform. The bandwidth
                        limit also limits the trigger signal path of any channel that has BW Limit turned on.
                        When BW Limit is selected, “BW” is illuminated on the front panel next to the
                        channel position knob ( ).
                     6 Press the Invert softkey to invert the selected channel.
                        When Invert is selected, the voltage values of the displayed waveform are inverted.
                        Invert affects how a channel is displayed, but does not affect triggering. If the
                        oscilloscope is set to trigger on a rising edge, it remains set to trigger on a same edge
                        (triggers at the same point on the waveform) after the channel is inverted.
                        Inverting a channel will also change the result of any function selected in the
                        Waveform Math Menu or any measurement.
                     7 Press the Probe softkey to display the Channel Probe Menu.
                        This menu lets you select additional probe parameters such as attenuation factor and
                        units of measurement for the connected probe.
                                         Probe calibration not needed, not available when this softkey is ghosted.
                        • Probe Units — Press the Units softkey to select the proper measurement unit for
                          the connected probe. Select Volts for a voltage probe and select Amps for a current
                          probe. Channel sensitivity, trigger level, measurement results, and math functions
                          will reflect the measurement units you have selected.
           Normal
           sweep mode
           XY or Roll
           mode
                                                            Sample rate
                                             Zoom sweep                      Horiz. pos   Time base      Time
                                             mode                            knob mode    fine           reference
                                                                                          adjustment
                        The Horizontal Menu lets you select the horizontal mode (Normal, Zoom, XY, or Roll), and
                        set the time base fine adjustment and time reference.
                        The current sample rate is displayed above the Fine and Time Ref softkeys.
                        Normal mode
                        1 Press the Normal softkey to select Normal horizontal mode.
                           Normal horizontal mode is the normal viewing mode for the oscilloscope. When the
                           oscilloscope is stopped, you can use the Horizontal knobs to pan and zoom the
                           waveform.
                        2 Turn the large knob (time/division) in the Horizontal section and notice the change it
                          makes to the status line.
                           When the oscilloscope is running in Normal mode, use the large Horizontal knob
                           marked           to change sweep speed and use the small knob marked        to set the
                           delay time (horizontal position). When the oscilloscope is stopped, use these knobs to
                           pan and zoom your waveform. The sweep speed (seconds/division) value is displayed
                           in the status line.
                       3 Press the Fine softkey to turn on the time base fine adjustment.
                          The Fine softkey lets you change the sweep speed in smaller increments with the
                          time/division knob. The sweep speed remains fully calibrated when Fine is on. The
                          value is displayed in the status line at the top of the display.
                          When Fine is turned off, the Horizontal sweep speed knob changes the time base
                          sweep speed in a 1-2-5 step sequence.
                       4 Note the setting of the Time Ref (time reference) softkey.
                          Time reference is the reference point on the display for delay time (horizontal
                          position). The time reference can be set to one major division from the left or right
                          edge, or to the center of the display.
                          A small hollow triangle ( at the top of the display grid marks the position of the time
                          reference. When delay time is set to zero, the trigger point indicator () overlays the
                          time reference indicator.
                          Turning the Horizontal sweep speed knob will expand or contract the waveform about
                          the time reference point (). Turning the Horizontal position ( ) knob in Normal
                          mode will move the trigger point indicator () to the left or right of the time reference
                          point ().
                          The time reference position sets the initial position of the trigger event within
                          acquisition memory and on the display, with delay set to 0. The delay setting sets the
                          specific location of the trigger event with respect to the time reference position. The
                          time reference setting affects the Zoom sweep as described in the following ways:
                          • When the horizontal mode is set to Normal, the delay knob positions the normal
                            sweep relative to the trigger. This delay is a fixed number. Changing this delay value
                            does not affect the sweep speed.
                          • When the horizontal mode is set to Zoom, the delay knob controls the position of
                            the zoom sweep window inside the normal sweep display. This delay value is
                            independent of sampling interval and sweep speed. Changing this delay value does
                            not effect the position of the normal window.
                       5 Turn the delay knob (     ) and notice that its value is displayed in the status line.
                          The delay knob moves the normal sweep horizontally, and it pauses at 0.00 s,
                          mimicking a mechanical detent. Changing the delay time moves the sweep
                          horizontally and indicates how far the trigger point (solid inverted triangle) is from the
                          time reference point (hollow inverted triangle ). These reference points are indicated
                          along the top of the display grid. The previous figure shows the trigger point with the
                        delay time set to 400 s. The delay time number tells you how far the time reference
                        point is located from the trigger point. When delay time is set to zero, the delay time
                        indicator overlays the time reference indicator.
                        All events displayed left of the trigger point happened before the trigger occurred, and
                        these events are called pre-trigger information. You will find this feature very useful
                        because you can now see the events that led up to the trigger point. Everything to the
                        right of the trigger point is called post-trigger information. The amount of delay range
                        (pre-trigger and post-trigger information) available depends on the sweep speed
                        selected and memory depth.
                     Zoom mode
                     Zoom, formerly called Delayed sweep mode, is a horizontally expanded version of normal
                     display. When Zoom is selected, the display divides in half and the Zoom mode          icon is
                     displayed in the middle of the line at the top of the display. The top half of the display
                     shows the normal sweep and the bottom half displays the Zoom sweep.
                     The Zoom window is a magnified portion of the normal sweep. You can use Zoom to
                     locate and horizontally expand part of the normal sweep for a more detailed
                     (higher-resolution) analysis of signals.
                     The following steps show you how to use Zoom.
                     1 Connect a signal to the oscilloscope and obtain a stable display.
                     2 Press the [Menu/Zoom] key (or [Main/Delayed] key on some oscilloscopes).
                     3 Press the Zoom softkey.
                     To change the sweep speed for the Zoom window, turn the horizontal sweep speed
                     control knob. As you turn the knob, the sweep speed is highlighted in the status line
                     above the waveform display area.
                     The area of the normal display that is expanded is outlined with a box and the rest of the
                     normal display is lightened. The box shows the portion of the normal sweep that is
                     expanded in the lower half. The Horizontal Sweep Speed controls the size of the box and
                     the Horizontal Position (delay time) control sets the left-to-right position of the zoom
                     sweep. The delay value, which is the time being displayed relative to the trigger point) is
                     momentarily displayed in the upper-right portion of the display when the delay time ( )
                     knob is turned. Negative delay values indicate you’re looking at a portion of the waveform
                     before the trigger event, and positive values indicate you’re looking at the waveform after
                     the trigger event.
                     To change the sweep speed for the normal sweep window, press the Normal softkey,
                     then turn the sweep speed knob.
 Normal
 sweep
 window
      Signal
      anomaly
      expanded
      in zoom
      window
 Zoom
 window
                               Select Normal
                               sweep or Zoom
                       When the horizontal mode is set to Normal, the horizontal position control (delay control)
                       positions the normal sweep relative to the trigger. This delay is a fixed number. Changing
                       this delay value does not affect the sweep speed. When the horizontal mode is set to
                       Zoom, the delay controls the position of the zoom sweep window inside the normal
                       sweep display. This delay value is independent of sampling interval and sweep speed.
                       To change the sweep speed for the normal sweep window, press the Normal softkey,
                       then turn the sweep speed knob.
                     To specify the measurement window for Zoom mode When the zoomed time base
                     is displayed, you can specify whether the upper, Main window or the lower, Zoom
                     window is used as the measurement window.
                     1 Press [Utility] > Options > Preferences > More.
                     2 Press Window to select from these measurement window options:
                        • Auto Select – The measurement is attempted in the lower, Zoom window; if it
                          cannot be made there, the upper, Main window is used.
                        • Main – The measurement window is the upper, Main window.
                        • Zoom – The measurement window is the lower, Zoom window.
                     For information about using zoom mode for measurements, refer to Chapter 6,
                     “Measurements and Math Functions,” starting on page 191.
                     Roll mode
                     • Press the [Menu/Zoom] key (or [Main/Delayed] key on some oscilloscopes), then
                       press the XY or Roll softkey to select “Roll”.
                     • Roll mode causes the waveform to move slowly across the screen from right to left. It
                       only operates on time base settings of 500 ms/div and slower. If the current time base
                       setting is faster than the 500 ms/div limit, it will be set to 500 ms/div when Roll mode
                       is entered.
                     • In Normal horizontal mode, signal events occurring before the trigger are plotted to
                       the left of the trigger point (t) and signal events after the trigger plotted to the right of
                       the trigger point.
                     • In Roll mode there is no trigger. The fixed reference point on the screen is the right
                       edge of the screen and refers to the current moment in time. Events that have
                       occurred are scrolled to the left of the reference point. Since there is no trigger, no
                       pre-trigger information is available.
                     If you would like to pause the display in Roll mode press the [Single] key. To clear the
                     display and restart an acquisition in Roll mode, press the [Single] key again.
                     Use Roll mode on low-frequency waveforms to yield a display much like a strip chart
                     recorder. It allows the waveform to roll across the display.
                       XY mode
                       XY mode changes the display from a volts-versus-time display to a volts-versus-volts
                       display. The time base is turned off. Channel 1 amplitude is plotted on the X-axis and
                       Channel 2 amplitude is plotted on the Y-axis.
                       You can use XY mode to compare frequency and phase relationships between two
                       signals. XY mode can also be used with transducers to display strain versus
                       displacement, flow versus pressure, volts versus current, or voltage versus frequency.
                       Use the cursors to make measurements on XY mode waveforms.
                       For more information about using XY mode for measurements, refer to “XY Horizontal
                       Mode” on page 220.
                       Knob softkey
                       The “Knob “softkey is used in conjunction with Segmented Memory. You can set the
                       Horizontal Position knob to adjust the horizontal position or move between segments.
                       See “Segmented Memory” on page 271.
3
Connecting to the Device Under Test
Analog Input Impedance (50 Ohm or 1 MOhm) 72
AutoProbe Interface 72
Passive Probes 73
Active Probes 74
Active Probes for 6000 Series 100 MHz Bandwidth Models 75
Connect the Probes to the Oscilloscope 75
Compensating Passive Probes 77
Calibrating Probes 78
Manually Setting the Probe Attenuation Factor 80
Digital Probes 80
                                                                     71
3    Connecting to the Device Under Test
AutoProbe Interface
                      The AutoProbe interface uses a series of contacts directly below the channel’s BNC
                      connector to transfer information between the oscilloscope and the probe. When you
                      connect a compatible probe to the oscilloscope, the AutoProbe interface determines the
                      type of probe and sets the oscilloscope’s parameters (units, offset, attenuation, coupling,
                      and impedance) accordingly.
Passive Probes
                       All InfiniiVision oscilloscopes recognize passive probes such as the 10073C, 10074C and
                       1165A. These probes have a pin on their connector that connects to the ring around the
                       oscilloscope’s BNC connector. Therefore, the oscilloscope will automatically set the
                       attenuation factor for recognized Agilent passive probes.
                       Passive probes that do not have a pin that connects to the ring around the BNC connector
                       will not be recognized by the oscilloscope, and you must set the probe attenuation factor
                       manually. See page 80.
                       The following passive probes can be used with the InfiniiVision oscilloscopes. Any
                       combination of passive probes can be used.
PASSIVE PROBES
Active Probes
                      Most Agilent active probes are compatible with the AutoProbe interface. Active probes
                      that do not have their own external power supply require substantial power from the
                      AutoProbe interface. “Quantity Supported” indicates the maximum number of each type
                      of active probe that can be connected to the oscilloscope. If too much current is drawn
                      from the AutoProbe interface, an error message will be displayed, indicating that you
                      must momentarily disconnect all probes to reset the AutoProbe interface, then connect
                      only the supported quantity of active probes.
ACTIVE PROBES
                      Do not float the oscilloscope chassis: Defeating the ground connection and
     CAUTION
                      “floating” the oscilloscope chassis will probably result in inaccurate measurements
                      and may also cause equipment damage. The probe ground lead is connected to the
                      oscilloscope chassis and the ground wire in the power cord. If you need to measure
                      between two live points, use a differential probe with sufficient dynamic range.
                      Do not negate the protective action of the ground connection to the oscilloscope.
     WA R N I N G
                      The oscilloscope must remain grounded through its power cord. Defeating the
                      ground creates an electric shock hazard.
Perfectly compensated
Over compensated
                          Under compensated
                                                                                          comp.cdr
Calibrating Probes
                      Some probes, such as the 10073C, 10074C, and 1165A passive probes, do not require
                      calibration. When one of these probes is connected, the Calibrate Probe softkey in the
                      Channel Probe Menu is grayed-out (displayed in faint text).
                      However, for certain active probes, such as InfiniiMax probes, the oscilloscope can
                      accurately calibrate its analog channels for the probe. When you connect a probe that
                      can be calibrated, the Calibrate Probe softkey in the Channel Probe Menu becomes
                      active. To calibrate one of these probes:
                      1 First, plug your probe into one of the oscilloscope channels.
                         This could be, for example, an InfiniiMax probe amplifier/probe head with attenuators
                         attached.
                      2 Connect the probe to the Probe Comp terminal, and the probe ground to the Probe
                        Comp ground terminal.
                      When calibrating a differential probe, connect the positive lead to the Probe Comp
       NOTE           terminal and the negative lead to the Probe Comp ground terminal. You may need to
                      connect an alligator clip to the ground lug to allow a differential probe to span between
                      the Probe Comp test point and ground. A good ground connection ensures the most
                      accurate probe calibration.
                      3 Press the Channel on/off key to turn the channel on (if the channel is off).
                      4 In the Channel Menu, press the Probe softkey.
                      5 In the Channel Probe Menu, the second softkey from the left is for specifying your
                        probe head (and attenuation). Repeatedly press this softkey until the probe head
                        selection matches the attenuator you are using.
                         The choices are:
                         • 10:1 single-ended browser (no attenuator)
                         • 10:1 differential browser (no attenuator)
                         • 10:1 (+6 dB Atten) single-ended browser
                         • 10:1 (+6 dB Atten) differential browser
                         • 10:1 (+12 dB Atten) single-ended browser
                         • 10:1 (+12 dB Atten) differential browser
Digital Probes
                      Please see “To connect the digital probes to the device under test” on page 372 for
                      information about digital probing.
4
Triggering
Triggering - General Information 83
Trigger Mode and Coupling Menu 84
Auto and Normal Trigger modes 84
Trigger Level Adjustment 86
Trigger Coupling 87
Trigger Noise Rejection 87
Trigger Holdoff 90
External Trigger Input 92
Trigger Output 95
Trigger Types 96
    CAN Trigger 98
    Duration Trigger 102
    Edge Trigger 105
    FlexRay Setup and Triggering 107
    Glitch or Narrow Pulse Capture 112
    I2C Trigger 115
    I2S Trigger 121
    LIN Trigger 128
    MIL-STD 1553 Setup and Triggering 131
    Nth Edge Burst Trigger 135
    Pattern Trigger 137
    Pulse Width Trigger 140
    Sequence Trigger 143
    SPI Trigger 150
    TV Trigger 156
    UART/RS232 Trigger 166
    USB Trigger 170
                                                         81
4    Triggering
                  This chapter provides instructions for setting up triggers for various signal types. Setups
                  can be saved if desired (see “Saving Oscilloscope Data” on page 247).
                  The easiest way to trigger on a waveform is to use AutoScale. Simply press the
                  [AutoScale] key and the oscilloscope will attempt to trigger on the waveform using a
                  simple Edge Trigger type. For AutoScale details see “AutoScale” on page 177.
                  Complex waveforms may require specific trigger setups as described in this chapter.
Trigger Event
Acquisition Memory
                       Agilent InfiniiVision oscilloscopes provide a full set of features to help automate your
                       measurement tasks. MegaZoom technology lets you capture and examine triggered or
                       untriggered waveforms. You can:
                       • Modify the way the oscilloscope acquires data.
                       • Set up simple or complex trigger conditions as needed, to capture only the sequence
                         of events you want to examine.
                       Adding options such as the Low Speed Serial (LSS) or Automotive Serial (AMS) serial
       NOTE            decode feature to the oscilloscope will add more trigger features to the relevant trigger
                       types (I2C and SPI, or CAN and LIN). Adding option FRS (FlexRay serial decode) will add
                       the FlexRay triggers.
                                                                                                2-channel
                                                                                                models only
                 Auto Mode
                       Use the auto trigger modes for signals other than low-repetitive-rate signals and for
                       unknown signal levels. To display a DC signal, you must use auto trigger mode since there
                       are no edges on which to trigger.
                       When you select [Run], the oscilloscope operates by first filling the pre-trigger buffer. It
                       starts searching for a trigger after the pre-trigger buffer is filled, and continues to flow
                       data through this buffer while it searches for the trigger. While searching for the trigger,
                       the oscilloscope overflows the pre-trigger buffer; the first data put into the buffer is the
                       first pushed out (FIFO). When a trigger is found, the pre-trigger buffer will contain the
                       events that occurred just before the trigger. If no trigger is found, the oscilloscope
                       generates a trigger and displays the data as though a trigger had occurred. In this case,
                       the background of the Auto indicator at the top of the display will flash, and the Auto
                       indicator will change to Auto? indicating that the oscilloscope is forcing triggers.
                       When you press the [Single] key, the oscilloscope will fill pre-trigger buffer memory, and
                       continue flowing data through the pre-trigger buffer until the auto trigger overrides the
                       searching and forces a trigger. At the end of the trace, the oscilloscope will stop and
                       display the results.
                 Normal Mode
                       Use Normal trigger mode for low repetitive-rate signals or when Auto trigger is not
                       required.
                       In Normal mode the oscilloscope must fill the pre-trigger buffer with data before it will
                       begin searching for a trigger event. The trigger mode indicator on the status line flashes
                       Trig’d? to indicate the oscilloscope is filling the pre-trigger buffer. While searching for the
                       trigger, the oscilloscope overflows the pre-trigger buffer; the first data put into the buffer
                       is the first pushed out (FIFO).
                       When the trigger event is found, the oscilloscope will fill the post-trigger buffer and
                       display the acquisition memory. The trigger mode indicator on the status line indicates
                       Trig’d (not flashing). If the acquisition was initiated by [Run/Stop], the process repeats.
                       If the acquisition was initiated by pressing [Single], then the acquisition stops and you
                       can Pan and Zoom the waveform.
                       In either Auto or Normal mode, the trigger may be missed completely under certain
                       conditions. This is because the oscilloscope will not recognize a trigger event until the
                       pre-trigger buffer is full. Suppose you set the Time/Div knob to a slow sweep speed, such
                       as 500 ms/div. If the trigger condition occurs before the oscilloscope has filled the
                  pre-trigger buffer, the trigger will not be found. If you use Normal mode and wait for the
                  trigger condition indicator to flash before causing the action in the circuit, the
                  oscilloscope will always find the trigger condition.
                  Some measurements you want to make will require you to take some action in the device
                  under test to cause the trigger event. Usually, these are single-shot acquisitions, where
                  you will use the [Single] key.
Trigger Coupling
                       1 Press the [Mode/Coupling] key.
                       2 Press the Coupling softkey, then select DC, AC, or LF Reject coupling.
                          • DC coupling allows DC and AC signals into the trigger path.
                          • AC coupling places a 10 Hz high-pass filter in the trigger path removing any DC
                            offset voltage from the trigger waveform. The high-pass filter in the External Trigger
                            input path is 3.5 Hz for all models. Use AC coupling to get a stable edge trigger
                            when your waveform has a large DC offset.
                          • LF (low frequency) Reject coupling places a 50 kHz high-pass filter in series with
                            the trigger waveform. Low frequency reject removes any unwanted low frequency
                            components from a trigger waveform, such as power line frequencies, that can
                            interfere with proper triggering. Use this coupling to get a stable edge trigger when
                            your waveform has low frequency noise.
                          • TV coupling is normally grayed-out, but is automatically selected when TV trigger is
                            enabled in the Trigger More Menu.
                       Note that Trigger Coupling is independent of Channel Coupling. To change Channel
                       Coupling see page 61.
                  HF Reject
                     HF Reject adds a 50 kHz low-pass filter in the trigger path to remove high frequency
                     components from the trigger waveform. You can use HF Reject to remove high-frequency
                     noise, such as AM or FM broadcast stations or noise from fast system clocks, from the
                     trigger path.
                     1 Press [Mode/Coupling]&HF Reject.
0 dB
                        -3 dB                          Pass
                                                       Band
DC 50 kHz
                 LF Reject
                       Low-frequency reject (LF reject) adds a high-pass filter with the 3-dB point at 50 kHz. LF
                       reject removes low-frequency signals, such as power line noise, from the trigger path.
                       1 Press [Mode/Coupling]&Coupling&LF Reject.
0 dB
                         3 dB down point
                                                                                      Pass
                                                                                      Band
DC 50 kHz
                 Noise Rejection
                       Noise Rej adds additional hysteresis to the trigger circuitry. By increasing the trigger
                       hysteresis band, you reduce the possibility of triggering on noise. However, this also
                       decreases the trigger sensitivity so that a slightly larger signal is required to trigger the
                       oscilloscope.
                       1 Press [Mode/Coupling]&HF Reject.
Trigger Holdoff
                  To set holdoff
                      1 Press the [Mode/Coupling] key; then press the Holdoff softkey.
                      2 Turn the Entry knob        to increase or decrease the trigger holdoff time shown.
                      Holdoff sets the amount of time that the oscilloscope waits before re-arming the trigger
                      circuitry. Use Holdoff to stabilize the display of complex waveforms.
                      To get a stable trigger on the pulse burst shown below, set the holdoff time to be >200 ns
                      but <600 ns.
200 ns 600 ns
                       By setting the Holdoff, you can synchronize triggers. The oscilloscope will trigger on one
                       edge of the waveform, and ignore further edges until the holdoff time expires. The
                       oscilloscope will then re-arm the trigger circuit to search for the next edge trigger. This
                       allows the oscilloscope to trigger on a repeating pattern in a waveform.
                      Probe Attenuation Turn the Entry knob to set the attenuation factor displayed in the
                      Probe softkey for the connected probe. The attenuation factor can be set from 0.1:1 to
                      1000:1 in a 1-2-5 sequence.
                      When you connect an AutoProbe self-sensing probe, the oscilloscope will automatically
                      configure your probe to the correct attenuation factor.
                      The probe correction factor must be set properly for measurements to be made correctly.
                      Range     The input voltage range can be set to 1.0 Volts or 8.0 Volts. When in current
                      mode, the range is fixed at 1.0 Amps. Range is automatically scaled according to the
                      probe’s attenuation factor.
                       Input Impedance    The 2-channel oscilloscopes have selectable external trigger input
                       impedance. The impedance can be set to either 1M Ohm or 50 Ohm by pressing the
                       Imped softkey.
                       • 50 Ohm — matches 50-ohm cables commonly used in making high frequency
                         measurements. This impedance matching gives you the most accurate measurements
                         since reflections are minimized along the signal path.
                       • 1M Ohm — is for use with many passive probes and for general-purpose
                         measurements. The higher impedance minimizes the loading effect of the
                         oscilloscope on the device under test.
                       Probe Units Press the Units softkey to select the proper measurement unit for the
                       connected probe. Select Volts for a voltage probe and select Amps for a current probe.
                       Measurement results, channel sensitivity, and trigger level will reflect the measurement
                       units you have selected.
                      Input Voltage The input voltage sensitivity is 500 mV, from DC to 500 MHz. The input
                      voltage range is ±15 V.
                      There are no range or units settings for the External Trigger input of the 4-channel
                      oscilloscope.
Trigger Output
                       You can select one of the following signals to be output at the TRIG OUT connector on the
                       rear panel of the oscilloscope:
                       • Triggers
                       • Source frequency
                       • Source frequency/8
                       • Mask test pass/fail (see “Mask Test Trigger Output” on page 342.)
                 Triggers
                       This is the default selection. In this mode, a rising edge is output each time the
                       oscilloscope triggers. This rising edge is delayed 17 ns from the oscilloscope’s trigger
                       point. The output level is 0-5 V into an open circuit, and 0-2.5 V into 50 .
                 Source Frequency
                       This mode is only available when the oscilloscope is set to positive or negative edge
                       trigger on a front-panel source (analog channel or the external input of a 2-channel
                       oscilloscope). In this mode, the TRIG OUT BNC is connected to the output of the trigger
                       comparator. The output level is 0-580 mV into an open circuit, and 0-290 mV into 50 .
                       The maximum frequency output is 350 MHz due to the bandwidth limitation of the TRIG
                       OUT BNC amplifier. This selection is useful for driving an external frequency counter.
                 Source Frequency/8
                       This selection produces the same signal as described in Source frequency except the
                       output frequency is one-eighth of the trigger comparator output frequency. This mode is
                       useful when your input signal is faster than 350 MHz.
                       The Trigger Out connector also provides the User Cal signal. See “User Calibration” on
                       page 394.
Trigger Types
                  The oscilloscope lets you synchronize the display to the actions of the device under test
                  by defining a trigger condition. You can use any input channel or the External Trigger
                  Input BNC as the source for most trigger types.
                       Changes to the trigger specification are applied when you make them. If the oscilloscope
                       is stopped when you change a trigger specification, the oscilloscope will use the new
                       specification when you press [Run/Stop] or [Single]. If the oscilloscope is running when
                       you change a triggering specification, it uses the new trigger definition when it starts the
                       next acquisition.
CAN Trigger
                  The Controller Area Network (CAN) trigger allows triggering on CAN version 2.0A and
                  2.0B signals. Setup consists of connecting the oscilloscope to a CAN signal, using the
                  Signals Menu to specify the signal source, baud rate, and sample point, and using the
                  Settings Menu to specify the event to trigger upon.
                  If the AMS license is not installed on the oscilloscope, you can still trigger on the Start of
                  Frame (SOF) bit of a CAN data frame. If the N5424A CAN/LIN Automotive Triggering and
                  Decode option (license AMS) is installed on your oscilloscope, the following additional
                  CAN trigger types will also be available: Remote Frame ID (RTR), Data Frame ID (~RTR),
                  Remote or Data Frame ID, Data Frame ID and Data, Error Frame, All Errors, Acknowledge
                  Error, and Overload Frame.
                  A CAN message frame in CAN_L signal type is shown below:
SOF edge
                       6 Repeatedly press and release the Trigger: softkey or rotate the Entry knob to select the
                         trigger condition.
                          • SOF - Start of Frame — The oscilloscope triggers at the start of a frame.
                          • Remote Frame ID (RTR) — The oscilloscope triggers on remote frames with the
                            specified ID. Press the Bits softkey to select the ID.
                       For details about using the CAN Bits Menu softkeys, press and hold the softkey in
       NOTE            question to display the built-in help.
                          • Data Frame ID (~RTR) - The oscilloscope will trigger on data frames matching the
                            specified ID. Press the Bits softkey to select the ID.
                          • Remote or Data Frame ID - The oscilloscope will trigger on remote or data frames
                            matching the specified ID. Press the Bits softkey to select the ID.
                          • Data Frame ID and Data - The oscilloscope will trigger on data frames matching
                            the specified ID and data. Press the Bits softkey to select the ID and set up the
                            number of data bytes and values.
                          • Error Frame - The oscilloscope will trigger on CAN active error frames.
                          • All Errors - The oscilloscope will trigger when any form error or active error is
                            encountered.
                          • Acknowledge Error - The oscilloscope will trigger when the acknowledge bit is
                            recessive (high).
                          • Overload Frame - The oscilloscope will trigger on CAN overload frames.
                   8 Press the Source softkey to select the channel connected to the CAN signal line.
                      As you repeatedly press the Source softkey (or rotate the Entry knob), the CAN label
                      for the source channel is automatically set and the channel you select is shown in the
                      upper-right corner of the display next to “CAN”.
                      If you have connected one of the oscilloscope’s analog source channels to the CAN
                      signal: Adjust the trigger level for the selected analog channel to the middle of the
                      waveform by turning the Trigger Level knob.
                      If you have connected one of the oscilloscope’s digital source channels to the CAN
                      signal (this applies to MSO model oscilloscopes only): Press the [D15-D0] key and
                      select Thresholds to set an appropriate threshold level for digital channels.
                      The value of the trigger level or digital threshold is displayed in the upper-right corner
                      of the display.
                   9 Press the Signal softkey and select the type and polarity of the CAN signal. This also
                     automatically sets the channel label for the source channel.
                      • CAN_H — The actual CAN_H differential bus
                      Dominant low signals:
                      • CAN_L — The actual CAN_L differential bus signal
                      • Rx — The Receive signal from the CAN bus transceiver
                      • Tx — The Transmit signal from the CAN bus transceiver
                      • Differential — The CAN differential bus signals connected to an analog source
                        channel using a differential probe. Connect the probe’s positive lead to the
                        dominant-low CAN signal (CAN_L) and connect the negative lead to the
                        dominant-high CAN signal (CAN_H).
                   10 Repeatedly press and release the Baud softkey to set the CAN signal baud rate to
                      match your CAN bus signal.
                                               
                                                 
                                  	
 
                                                    
                       You can use the Zoom mode for easier navigation of the decoded data.
                       If the setup does not produce a stable trigger, the CAN signal may be slow enough that
       NOTE            the oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode
                       softkey to set the trigger mode from Auto to Normal.
Duration Trigger
                   Duration trigger lets you define a pattern, then trigger on a specified time duration of this
                   logical AND combination of the channels.
                   1 Press the [More] key in the Trigger section of the front panel, rotate the Entry knob
                     until Duration is displayed in the Trigger softkey, then press the Settings softkey to
                     display the Duration Trigger Menu.
                                                                            Duration    Selected    Trigger level
                                                                            trigger     channel     or threshold
                   2 For each analog or digital channel you want to include in the desired pattern, press the
                     Channel softkey to select the channel.
                      This is the channel source for the H, L, or X condition. As you press the Channel
                      softkey (or rotate the Entry knob), the channel you select is highlighted in the
                      Pattern = line directly above the softkeys and in the upper-right corner of the
                      display next to “Dur”. External trigger may also be specified as a channel in the
                      pattern when using the 2-channel and 4-channel oscilloscopes.
                      Adjust the trigger level for the selected analog channel by turning the Trigger Level
                      knob. Press the [D15-D0] key and select Thresholds to set the threshold level for
                      digital channels. The value of the trigger level or digital threshold is displayed in the
                      upper-right corner of the display.
                   3 For each channel you select, press the logic level softkey to set the condition for that
                     channel in the pattern.
                      • 1 sets the pattern to 1 (high) on the selected channel. A high is a voltage level that
                        is greater than the channel's trigger level or threshold level.
                      • 0 sets the pattern to 0 (low) on the selected channel. A low is a voltage level that is
                        less than the channel's trigger level or threshold level.
                          • X sets the pattern to don't care on the selected channel. Any channel set to don't
                            care is ignored and is not used as part of the pattern. If all channels in the pattern
                            are set to don't care, the oscilloscope will not trigger.
                       4 Press the Qualifier softkey to set the time duration qualifier for the pattern.
                          The time qualifier can set the oscilloscope to trigger on a channel pattern whose time
                          duration is:
                          • less than a time value (<)
                          • greater than a time value (>)
                          • greater than a time value, but with timeout (Timeout). A trigger will be forced at the
                            timeout value, rather than occurring when the pattern is exited.
                          • within a range of time values (><)
                          • outside a range of time values (<>)
                          The time values for the selected qualifier are set using the qualifier time set softkeys
                          (< and >) and the Entry knob.
                       5 Select a qualifier time set softkey (< or >), then rotate the Entry knob to set the
                         duration qualifier time.
Edge Trigger
                       The Edge trigger type identifies a trigger by looking for a specified edge (slope) and
                       voltage level on a waveform. You can define the trigger source and slope in this menu.
                       The slope can be set to rising edge or falling edge, and can be set to alternating edges or
                       either edge on all sources except Line. The trigger type, source, and level are displayed in
                       the upper-right corner of the display.
                       1 Press the [Edge] key in the Trigger section of the front panel to display the Edge
                         Trigger Menu.
                                                                                Edge      Trigger     Trigger level
                                                                                          source      or threshold
                             Trigger         Slope
                             source
                       2 Press the Slope softkey and select rising edge, falling edge, alternating edges, or
                         either edge. The selected slope is displayed in the upper-right corner of the display.
                       Alternating edge mode is useful when you want to trigger on both edges of a clock (for
       NOTE            example, DDR signals). Either edge mode is useful when you want to trigger on any
                       activity of a selected source. All modes operate up to the bandwidth of the oscilloscope
                       except Either edge mode, which has a limitation. Either edge mode will trigger on
                       Constant Wave signals up to 100 MHz, but can trigger on isolated pulses down to
                       1/(2*oscilloscope's bandwidth).
                   You can select analog channel 1 or 2, Ext, or Line as the trigger source on any Agilent
                   InfiniiVision oscilloscope. The trigger source can also be set to channel 3 and 4 on the
                   4-channel oscilloscopes, or digital channels D15 through D0 on the mixed-signal
                   oscilloscopes. You can choose a channel that is turned off (not displayed) as the
                   source for the edge trigger.
                   The selected trigger source is indicated in the upper-right corner of the display next to
                   the slope symbol:
                      1 through 4 = analog channels
                      D0 through D15 = digital channels
                      E = External trigger
                      L = Line trigger
                       3 Press Settings.
                       4 In the FlexRay Trigger Menu, press Signals.
                       5 In the FlexRay Signals Menu, press Source and select the analog channel that is
                         probing the FlexRay signal.
                       6 Press Baud and select the baud rate of the FlexRay signal being probed.
                       7 Press Bus and select the bus type of the FlexRay signal being probed.
                         It is important to specify the correct bus because this setting affects CRC error
                         detection.
                      8 Press Auto Setup to perform the following actions:
                         • Set the selected source channel’s impedance to 50 Ohms.
                         • Set the selected source channel’s probe attenuation to 10:1.
                         • Set the trigger level (on the selected source channel) to -300 mV.
                         • Turns on trigger Noise Reject.
                         • Turns on Serial Decode.
                         • Sets the trigger type to FlexRay.
                   FlexRay Triggering
                      To set up the oscilloscope to capture a FlexRay signal, see “Setup for MIL-STD 1553
                      Signals” on page 132.
                      After you have set up the oscilloscope to capture a FlexRay signal, you can then set up
                      triggers on frames, errors, or events.
                      To display FlexRay serial decode, see “FlexRay Serial Decode” on page 302.
        NOTE
                      3 Press Settings.
                      4 In the FlexRay Trigger Menu, press Trigger; then, select Frame.
5 Press the Frames softkey to access the FlexRay Frame Trigger Menu.
                       6 Press the Frame ID softkey, and use the Entry knob to select the frame ID value from
                         All or 1 to 2047.
                       7 Press the Frame Type softkey to select the frame type:
                          • All Frames
                          • Startup Frames
                          • NULL Frames
                          • Sync Frames
                          • Normal Frames
                          • NOT Startup Frames
                          • NOT NULL Frames
                          • NOT Sync Frames
                       8 Press the Cyc Ct Rep softkey, and use the Entry knob to select the cycle count
                         repetition factor (2, 4, 8, 16, 32, or 64, or All).
                       9 Press the Cyc Ct Bas softkey, and use the Entry knob to select the cycle count base
                         factor from 0 through the Cyc Ct Rep factor minus 1.
                          For example, with a base factor of 1 and a repetition factor of 16, the oscilloscope
                          triggers on cycles 1, 17, 33, 49, and 65.
                          To trigger on a particular cycle, set the Cycle Repetition factor to 64 and use the cycle
                          base factor to choose a cycle.
                          To trigger all (any) cycles, set the Cycle Repetition factor to All. The scope will trigger
                          on any and all cycles.
                       Because specific FlexRay frames may occur infrequently, it may be helpful to set the
       NOTE            oscilloscope to Normal trigger mode instead of Auto trigger mode. This prevents the
                       oscilloscope from Auto triggering while waiting for a particular frame and cycle
                       combination.
                   3 Press Settings.
                   4 In the FlexRay Trigger Menu, press Trigger; then, select Error.
                   Since FlexRay errors occur infrequently it may be helpful to set the oscilloscope to
        NOTE       Normal trigger mode instead of Auto trigger mode. This prevents the oscilloscope from
                   Auto triggering while waiting for an error to occur. You may need to adjust trigger holdoff
                   to see a particular error when multiple errors exist.
3 Press Settings.
I2C Trigger
                       An I2C (Inter-IC bus) trigger setup consists of connecting the oscilloscope to the serial
                       data (SDA) line and the serial clock (SCL) line, then triggering on a stop/start condition, a
                       restart, a missing acknowledge, an EEPROM data read, or on a read/write frame with a
                       specific device address and data value.
                       1 Press [Save/Recall]&Default Setup.
                       2 Press the [Label] key to switch labels on.
                       3 Turn on any analog or digital channels that you will be using for the I2C signals.
                       4 Press the [More] key in the Trigger section of the front panel, rotate the Entry knob
                         until I2C is displayed in the Trigger softkey, then press the Settings softkey to display
                         the I2C Trigger Menu.
                                                                             Currently selected Clock      Trigger level
                                                                             or Data channel               or threshold
                                                                     I2C trigger
                       6 Connect an oscilloscope channel to the SCL (serial clock) line in the device under test,
                         then set the SCL clock channel softkey to that channel.
                      As you press the SCL softkey (or rotate the Entry knob), the SCL label for the source
                      channel is automatically set and the channel you select is shown in the upper-right
                      corner of the display next to “I2C”.
                   7 Set the trigger levels for the SCL and SDA signals to the middle of the signals:
                      • If your I2C signals are connected to analog channels, press the SCL softkey and
                        rotate the Trigger Level knob, then press the SDA softkey and rotate the Trigger
                        Level knob.
                      • If your I2C signals are connected to digital channels (this applies to MSO model
                        oscilloscopes only), press the [D15-D0] key and the Thresholds softkey to access
                        the threshold level setting softkeys and set the thresholds to the approximate
                        middle of the signals.
                   8 Choose 7-bit or 8-bit address size. Use 8-bit address size to include the R/W bit as part
                     of the address value, or choose 7-bit address size to exclude the R/W bit from the
                     address value.
                   9 Connect an oscilloscope channel to the SDA (serial data) line in the device under test,
                     then set the SDA data channel softkey to that channel.
                   10 As you press the SDA softkey (or rotate the Entry knob), the SDA label for the source
                      channel is automatically set and the channel you select is shown in the upper-right
                      corner of the display next to “I2C”.
                      Adjust the trigger level for the selected analog channel by turning the Trigger Level
                      knob. Press the [D15-D0] key and select Thresholds to set the threshold level for
                      digital channels. The value of the trigger level or digital threshold is displayed in the
                      upper-right corner of the display.
                      Data needs to be stable during the whole high clock cycle or it will be interpreted as a
                      start or stop condition (data transitioning while the clock is high).
                   11 Return to the previous menu; then, press the Trigger: softkey to select one of the
                      following I2C trigger conditions:
                      • Start Condition — The oscilloscope triggers when SDA data transitions from high
                        to low while the SCL clock is high. For triggering purposes (including frame
                        triggers), a restart is treated as a start condition.
                          • Stop Condition — The oscilloscope triggers when data (SDA) transitions from low
                            to high while the clock (SCL) is high.
                             SDA
                              SCL
                                    Start Address          R/        Ack         Data            Ack Stop
                                    Condition                                                        Condition
                          • Missing Acknowledge — The oscilloscope triggers when SDA data is high during
                            any Ack SCL clock bit.
                          • Address with no Ack — The oscilloscope triggers when the acknowledge of the
                            selected address field is false. The R/W bit is ignored.
                          • Restart — The oscilloscope triggers when another start condition occurs before a
                            stop condition.
                          • EEPROM Data Read — The trigger looks for EEPROM control byte value 1010xxx
                            on the SDA line, followed by a Read bit and an Ack bit. It then looks for the data
                            value and qualifier set by the Data softkey and the Data is softkey. When this event
                            occurs, the oscilloscope will trigger on the clock edge for the Ack bit after the data
                            byte. This data byte does not need to occur directly after the control byte.
                                                           Read
                              SDA
                              SCL
                                            Control byte        R/         Ack    Data          Ack
                              Start or
                              Restart Condition                                          Trigger point
                    • Frame (Start: Addr7: Read: Ack: Data) or Frame (Start: Addr7: Write: Ack: Data)
                      — The oscilloscope triggers on a read or write frame in 7-bit addressing mode on
                      the 17th clock edge if all bits in the pattern match. For triggering purposes, a restart
                      is treated as a start condition.
                                                  Read Write
                         SDA
                         SCL
                               Start or Address        R/     Ack      Data                  Ack Stop
                               Restart                                         Trigger point     Condition
                               Condition                                       17th clock edge
                    • Frame (Start: Addr7: Read: Ack: Data: Ack: Data2) or Frame (Start: Addr7: Write:
                      Ack: Data: Ack: Data2) — The oscilloscope triggers on a read or write frame in
                      7-bit addressing mode on the 26th clock edge if all bits in the pattern match. For
                      triggering purposes, a restart is treated as a start condition.
                                        Read
                                               Write
                   SDA
SCL
                    • 10-bit Write — The oscilloscope triggers on a 10-bit write frame on the 26th clock
                      edge if all bits in the pattern match. The frame is in the format:
                         Frame (Start: Address byte 1: Write: Address byte 2: Ack: Data)
                           SCL
                                      Address     R/      Ack1 Address Ack2        Data             Ack Stop
                          Start or                             2nd byte
                                      1st byte                                        Trigger point     Condition
                          Restart
                          Condition                                                   26th clock edge
                       12 If you have set the oscilloscope to trigger on an EEPROM Data Read condition:
                          Press the Data is softkey to set the oscilloscope to trigger when data is = (equal to),
                           (not equal to), < (less than), or > (greater than) the data value set in the Data
                          softkey.
                          The oscilloscope will trigger on the clock edge for the Ack bit after the trigger event is
                          found. This data byte does not need to occur directly after the control byte. The
                          oscilloscope will trigger on any data byte that meets the criteria defined by the Data is
                          and Data softkeys during a current address read or a random read or a sequential read
                          cycle.
                       13 If you have set the oscilloscope to trigger on a 7-bit address read or write frame
                          condition or a 10-bit write frame condition:
                          a Press the Address softkey and turn the Entry knob to select the 7-bit or 10-bit
                            device address.
                             You can select from an address range of 0x00 to 0x7F (7-bit) or 0x3FF (10-bit)
                             hexadecimal. When triggering on a read/write frame, the oscilloscope will trigger
                             after the start, address, read/write, acknowledge, and data events occur.
                             If don't care is selected (0xXX or 0xXXX) for the address, the address will be
                             ignored. The trigger will always occur on the 17th clock for 7-bit addressing or 26th
                             clock for 10-bit addressing.
                          b Press the Data value softkey and turn the Entry knob to select the 8-bit data pattern
                            on which to trigger.
                             You can select a data value in the range of 0x00 to 0xFF (hexadecimal). The
                             oscilloscope will trigger after the start, address, read/write, acknowledge, and data
                             events occur.
                         If don't care (0xXX) is selected for data, the data will be ignored. The trigger will
                         always occur on the 17th clock for 7-bit addressing or 26th clock for 10-bit
                         addressing.
                      c If you have selected a three-byte trigger, press the Data2 value softkey and turn the
                        Entry knob to select the 8-bit data pattern on which to trigger.
I2S Trigger
                       The N5468A (Option SND) I²S trigger and serial decode option adds the ability to decode
                       serial data to 4-channel or 4+16 channel oscilloscopes. To control serial decode press the
                       [Acquire] key.
                       An I2S (Inter-IC Sound or Integrated Interchip Sound) trigger setup consists of connecting
                       the oscilloscope to the serial clock, word select, and serial data lines and then triggering
                       on a data value.
                       1 Press [Save/Recall]&Default Setup.
                       2 Press the [Label] key to switch labels on.
                       3 Turn on any analog or digital channels that you will be using for the I2S signals.
                       4 Press the [More] key in the Trigger section of the front panel. Rotate the Entry knob
                         until I2S is displayed in the Trigger softkey.
                       5 Press the Settings softkey to display the I2S Trigger Menu.
                                                                         Currently selected channel    Trigger level
                                                                                                       or threshold
                                                                     I2S trigger
                   6 Press the Signals softkey to display the I2S Signals Menu. A diagram appears showing
                     WS, SCLK, and SDATA signals for the currently specified bus configuration.
                   7 Connect an oscilloscope channel to the SCLK (serial clock) line in the device under
                     test. (If the channel is not already switched on, switch it on now, and return to this
                     menu by pressing [More]&Settings& Signals.)
                      Rotate the Entry knob to set the SCLK clock channel softkey to the SCLK channel.
                      As you rotate the Entry knob the SCLK label for the source channel is automatically
                      set and the channel you select is shown in the upper-right corner of the display next to
                      “I2S”.
                   8 Connect an oscilloscope channel to the WS (word select) line in the device under test,
                     switch the channel on, then set the WS channel softkey to that channel.
                   9 Connect an oscilloscope channel to the SDATA (serial data) line in the device under
                     test, switch the channel on, then set the SDATA channel softkey to that channel.
                       10 Set the trigger levels for the SCLK, WS, and SDATA signals to the middle of the signals:
                          • If your I2S signals are connected to analog channels, press the SCLK softkey and
                            rotate the Trigger Level knob (not the Entry knob). Repeat for the WS and SDATA
                            softkeys.
                          • If your I2S signals are connected to digital channels (on MSO model oscilloscopes
                            only), press the [D15-D0] key and the Thresholds softkey to access the threshold
                            level setting softkeys and set the thresholds to the approximate middle of the
                            signals. For details see “To change the logic threshold for digital channels” on
                            page 380.
                       11 Press [More]&Settings&Bus Config. The I2S Bus Configuration Menu is displayed,
                          along with a diagram showing WS, SCLK, and SDATA signals for the currently
                          specified bus configuration.
                       12 Press the Word Size softkey. Rotate the Entry knob to match the transmitter word size
                          of the device under test (from 4 to 32 bits).
                       13 Press the Receiver softkey. Rotate the Entry knob to match the receiver word size of
                          the device under test (from 4 to 32 bits).
                   14 Press the Alignment softkey and rotate the Entry knob to select the desired alignment
                      of the data signal (SDATA). The on-screen diagram changes with your selection.
                      Standard Alignment: MSB of data for each sample is sent first, LSB is sent last. The
                      MSB appears on the SDATA line one bit clock after the edge of the WS transition.
                                               A:;I8=6CC:A                                    G><=I8=6CC:A
LH
                        H8A@
                                     DC:H8A@8N8A:
                      H96I6           %    &    '            c"' c"&                   %   &   '             c"' c"&
                                     BH7                        AH7                  BH7                        AH7
                      Left-Justified: Data transmission (MSB first) begins at the edge of the WS transition
                      (without the one-bit delay that Standard format employs).
                                               A:;I8=6CC:A                                    G><=I8=6CC:A
LH
H8A@
LH
H8A@
                       15 Press the WS Low softkey to select whether WS Low indicates Left or Right channel
                          data. The on-screen diagram changes with your selection.
                          WS Low = Left Channel: Left-channel data corresponds to WS=low; right-channel
                          data corresponds to WS=high. WS Low=Left is the oscilloscope’s default WS setting.
                                                   A:;I8=6CC:A                   G><=I8=6CC:A
LH
                            H8A@
                                        DC:H8A@8N8A:
                           H96I6         %    &     '        c"' c"&     %    &   '        c"' c"&
                                        BH7                       AH7   BH7                    AH7
LH
                            H8A@
                                        DC:H8A@8N8A:
                           H96I6         %     &    '        c"' c"&     %    &    '        c"' c"&
                                        BH7                       AH7   BH7                       AH7
                       16 Press the SCLK Slope softkey to select the SCLK edge on which data is clocked in
                          your device under test: either rising or falling. The on-screen diagram changes with
                          your selection.
                       17 Press the up arrow softkey.
                       18 Press the Trigger Setup softkey.
                       19 Press the Audio softkey and rotate the Entry knob to choose to trigger on Left channel
                          events, Right channel events, or events that occur on Either channel.
"Trigger" value
"Armed" value
                          • Decreasing value — similar to the description above except the trigger occurs on a
                            descreasing data word value, and the “Armed” value is the value to which the data
                            must rise in order to re-arm the trigger.
                       21 Press the Base softkey and select a number base for entering data values:
                          • Binary (2’s complement).
                             When Binary is selected, the Bits softkey appears. This softkey opens the I2S Bits
                             Menu for entering data values.
                             When the trigger qualifier requires a pair of values (as with In Range, Out of Range,
                             Increasing value, or Decreasing value), the first softkey in the I2S Bits Menu lets
                             you select which value of the pair.
                             In the I2S Bits Menu, press the Bit softkey and rotate the Entry knob to select each
                             bit; then, use the 0 1 X softkey to set each bit value to zero, one, or don’t care. You
                             can use the Set all Bits softkey to set all bits to the value chosen on the 0 1 X
                             softkey.
                          • Signed decimal.
                             When Decimal is selected, the softkey(s) to the right let you enter decimal values
                             with the Entry knob. These softkeys can be Data, <, >, or Threshold depending on
                             the selected trigger qualifier.
                       22 If the setup does not produce a stable trigger, the I²S signal may be slow enough that
                          the oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the
                          Mode softkey to set the trigger mode from Auto to Normal.
LIN Trigger
                   LIN (Local Interconnect Network) trigger setup consists of connecting the oscilloscope to
                   a serial LIN signal.
                   LIN triggering will trigger on the rising edge at the Sync Break exit of the LIN single-wire
                   bus signal that marks the beginning of the message frame. If the N5424A CAN/LIN
                   Automotive Triggering and Decode option is installed on your oscilloscope, the Frame ID
                   and Frame ID and Data trigger types are also available.
                   A LIN signal message frame is shown below:
5 Press the Signals softkey. The LIN Signals Menu will be displayed.
                       6 Press the Source softkey to select the channel connected to the LIN signal line.
                          If you have connected one of the oscilloscope’s analog source channels to the LIN
                          signal: Adjust the trigger level for the selected analog channel to the middle of the
                          waveform by turning the Trigger Level knob.
                          If you have connected one of the oscilloscope’s digital source channels to the LIN
                          signal (this applies to MSO model oscilloscopes only): Press the [D15-D0] key and
                          select Thresholds to set an appropriate threshold level for digital channels.
                          The value of the trigger level or digital threshold is displayed in the upper-right corner
                          of the display.
                       7 Press the Baud softkey to set the LIN signal baud rate to match your LIN bus signal.
                          The LIN baud rate can be set to 2400 b/s, 9600 b/s, 10.4 kb/s, 19.2 kb/s, 115.2 kb/s,
                          625 kb/s, or User Defined. The default baud rate is 2400 b/s.
                          If the desired baud rate is not shown in the list, select User Defined. You can set the
                          LIN baud rate from 2.4 kb/s to 625 kb/s in increments of 100 b/s. Press the User Baud
                          softkey and rotate the Entry knob to make your selection.
                       8 Press the Smpl Pt softkey to select the sample point at which the oscilloscope will
                         sample the bit value.
                                                
                                                  
                                   	
 
                                                     
                       9 Press the Standard softkey to select the LIN standard your are measuring (LIN 1.3 or
                         LIN 2.0).
                   10 Press the Sync Break softkey and select the minimum number of clocks that define a
                      sync break in your LIN signal.
                   11 Press the up-arrow softkey to return to the LIN Trigger Menu.
                   12 Press the Trigger: softkey and choose the type of trigger:
                      • Sync (Sync Break) — The oscilloscope triggers on the rising edge at the Sync
                        Break exit of the LIN single-wire bus signal that marks the beginning the message
                        frame.
                      • ID (Frame ID) — The oscilloscope triggers when a frame with an ID equal to the
                        selected value is detected. Use the Entry knob to select the value for the Frame ID.
                      • ID & Data (Frame ID and Data) — The oscilloscope triggers when a frame with an
                        ID and data equal to the selected values is detected. When triggering on a frame ID
                        and data:
                         • To select the frame ID value, press the Frame ID softkey, and use the Entry
                           knob.
                            Note that you can enter a “don’t care” value for the frame ID and trigger on data
                            values only.
                         • To set up the number of data bytes and enter their values (in hexadecimal or
                           binary), press the Bits softkey to open the LIN Bits Menu.
                       Hex or Binary No. of bytes   Nibble or Bit   Nibble or Bit   Set all        Return to
                       entry         (up to 8)      position        value           nibbles/bits   previous menu
                   For details about using the LIN Bits Menu softkeys, press and hold the softkey in
        NOTE       question to display the built-in help.
                       This N2791A differential probe works well because it is designed for 1 MOhm impedance
                       oscilloscope inputs.
                       If you use a different differential probe that is designed for 50 Ohm impedance
                       oscilloscope inputs, be sure to set only one of the oscilloscope channel input impedances
                       to 50 Ohms (because the two input channel’s impedances are seen in parallel).
                      3 Press Settings.
                      4 In the MIL-STD-1553 Trigger Settings Menu:
                      • Upper Ch — selects the channel used for the upper threshold. Because defined pairs
                        of channels are used for upper and lower thresholds (either channels 1 and 2, or
                        channels 3 and 4), this softkey also selects the channel for the lower threshhold.
                      • Lower Ch — shows the channel selected for lower thresholds.
                      • Auto Setup — performs these actions:
                         • Copies the Upper Threshold Channel position and volts/div settings to the Lower
                           Threshold Channel.
                         • Sets both the Upper Threshold Channel and the Lower Threshold Channel probe
                           attenuation to 10:1.
                         • Sets the Upper Threshold Channel trigger level to 500 mV and the Lower Threshold
                           Channel trigger level to -500 mV.
                         • Turns off trigger noise reject.
                         • Turns on Serial Decode.
                         • Sets the trigger type to MIL-1553.
                         These sofkeys appear in both the Trigger Menu and the Serial Decode Menu.
                       3 Press Settings.
                       4 In the MIL-STD-1553 Trigger Settings Menu, press Trigger; then, select from these
                         options:
                          • Data Word Start – triggers on the start of a Data word (at the end of a valid Data
                            Sync pulse).
                          • Data Word Stop – triggers on the end of a Data word.
                          • Command/Status Word Start – triggers on the start of Command/Status word (at
                            the end of a valid C/S Sync pulse).
                          • Command/Status Word Stop – triggers on the end of a Command/Status word.
                          • Remote Terminal Address – triggers if the RTA of the Command/Status word
                            matches the specified value.
                             When this option is selected, the RTA softkey appears and lets you select the hex
                             Remote Terminal Address value to trigger on. If you select 0xXX (don’t cares), the
                             oscilloscope will trigger on any RTA.
                          • Remote Terminal Address + 11 Bits – triggers if the RTA and the remaining 11 bits
                            match the specified criteria.
                             When this option is selected, the RTA + Bits softkey appears and opens the
                             MIL-STD-1553 RTA + 11 Bits Menu. In this menu:
                             • The RTA softkey lets you select the hex Remote Terminal Address value.
                        • The Bit Time softkey lets you select the bit time position.
                        • The 0 1 X softkey lets you set the bit time position value as a 1, 0, or X (don't
                          care).
                     • Parity Error – triggers if the (odd) parity bit is incorrect for the data in the word.
                     • Manchester Error – triggers if a Manchester encoding error is detected.
                     • Sync Error – triggers if an invalid Sync pulse is found.
                   To display MIL-STD-1553 serial decode, see “MIL-STD 1553 Serial Decode” on page 320.
        NOTE
                       Nth Edge Burst trigger set up consists of selecting the source, the slope of the edge, the
                       idle time, and the number of the edge:
                       1 Press the [More] key in the Trigger section of the front panel, rotate the Entry knob
                         until Nth Edge Burst is displayed in the Trigger softkey.
                       2 Press the Settings softkey to display the Nth Edge Burst Trigger Menu.
Pattern Trigger
                       The Pattern trigger identifies a trigger condition by looking for a specified pattern. This
                       pattern is a logical AND combination of the channels. Each channel can have a value of 0
                       (low), 1 (high), or don't care (X). A rising or falling edge can be specified for one channel
                       included in the pattern. You can also trigger on a hex bus value as described on page 139.
                       1 Press the [Pattern] key in the Trigger section of the front panel to display the
                         PatternTrigger Menu.
                                                                                 Pattern     Selected     Trigger level
                                                                                 trigger     channel      or threshold
                                    Analog                Digital
                                    Channels              Channels                         Bus 1      Bus 2
                       2 For each analog or digital channel you want to include in the desired pattern, press the
                         Channel softkey to select the channel.
                          This is the channel source for the 0, 1, X, or edge condition. As you press the Channel
                          softkey (or rotate the Entry knob), the channel you select is highlighted in the
                          Pattern = line directly above the softkeys and in the upper-right corner of the
                          display next to “Pat”. External trigger may also be specified as a channel in the
                          pattern when using the 2-channel and 4-channel oscilloscopes.
                          Adjust the trigger level for the selected analog channel by turning the Trigger Level
                          knob. Press the [D15-D0] key and select Thresholds to set the threshold level for
                          digital channels. The value of the trigger level or digital threshold is displayed in the
                          upper-right corner of the display.
                       3 For each channel you select, press one of the condition softkeys to set the condition
                         for that channel in the pattern.
                     • 0 sets the pattern to zero (low) on the selected channel. A low is a voltage level
                       that is less than the channel's trigger level or threshold level.
                     • 1 sets the pattern to 1 (high) on the selected channel. A high is a voltage level that
                       is greater than the channel's trigger level or threshold level.
                     • X sets the pattern to don't care on the selected channel. Any channel set to don't
                       care is ignored and is not used as part of the pattern. However, if all channels in the
                       pattern are set to don't care, the oscilloscope will not trigger.
                     • The rising edge ( ) or falling edge ( ) softkey sets the pattern to an edge on the
                       selected channel. Only one rising or falling edge can be specified in the pattern.
                       When an edge is specified, the oscilloscope will trigger at the edge specified if the
                       pattern set for the other channels is true.
                        If no edge is specified, the oscilloscope will trigger on the last edge that makes the
                        pattern true.
                       If a digit is made up of less than four bits, then the value of the digit will be limited to the
       NOTE            value that can be created by the selected bits.
                       5 You can use the Set all Digits softkey to set all digits to a particular value.
                       When a hex bus digit contains one or more don’t care (X) bits and one or more bit with a
                       value or 0 or 1, the “$” sign will be displayed for the digit.
                       For information regarding digital bus display when Pattern triggering see “Bus values are
                       displayed when using Pattern trigger” on page 384.
                   2 Press the Source softkey (or rotate the Entry knob) to select a channel source for the
                     trigger.
                      The channel you select is shown in the upper-right corner of the display next to the
                      polarity symbol.
                      The source can be any analog or digital channel available on your oscilloscope.
                      External trigger may also be specified as a source when using a 2-channel
                      oscilloscope.
                      Adjust the trigger level for the selected analog channel by turning the Trigger Level
                      knob. Press the [D15-D0] key and select Thresholds to set the threshold level for
                      digital channels. The value of the trigger level or digital threshold is displayed in the
                      upper-right corner of the display.
                   3 Press the pulse polarity softkey to select positive (    )or negative (    ) polarity for the
                     pulse width you want to capture.
                      The selected pulse polarity is displayed in the upper-right corner of the display. A
                      positive pulse is higher than the current trigger level or threshold and a negative pulse
                      is lower than the current trigger level or threshold.
                          When triggering on a positive pulse, the trigger will occur on the high to low transition
                          of the pulse if the qualifying condition is true. When triggering on a negative pulse, the
                          trigger will occur on the low to high transition of the pulse if the qualifying condition is
                          true.
                       4 Press the qualifier softkey (< > ><) to select the time qualifier.
                          The Qualifier softkey can set the oscilloscope to trigger on a pulse width that is:
                          • less than a time value (<).
                             For example, for a positive pulse, if you set t<10 ns:
                                 10 ns                                                10 ns     Trigger
10 ns 15 ns 12 ns Trigger
                       5 Select the qualifier time set softkey (< or >), then rotate the Entry knob to set the
                         pulse width qualifier time.
                          The qualifiers can be set as follows:
                          • 2 ns to 10 s for > or < qualifier (5 ns to 10 s for 350 MHz bandwidth models)
                          • 10 ns to 10 s for >< qualifier, with minimum difference of 5 ns between upper and
                            lower settings
                       • When the time range (><) qualifier is selected, the Entry knob sets the upper time
                         range value.
Sequence Trigger
                         Sequence trigger lets you trigger the oscilloscope after finding a sequence of events.
                         Defining a sequence trigger requires three steps:
                         1 Define an event to find before you search for the trigger.
                            The “find” event can be a pattern, an edge from a single channel, or the combination
                            of a pattern and a channel edge.
                         2 Define the trigger event.
                            The “trigger on” event can be a pattern, an edge from a single channel, the
                            combination of a pattern and a channel edge, or the nth occurrence of an edge from a
                            single channel.
                         3 Set an optional reset event.
                            If you choose to define a “reset” event, the event can be a pattern, an edge from a
                            single channel, the combination of a pattern and a channel edge, or a timeout value.
                                                            Yes                                 Yes
                 Start                          Find:?                          Trigger on:?
                                                                               No                          Scope
                                           No                                                              triggers
                                                                    No
                                                                                Reset on:?
                                                                                                          Trigger
                                                                                                          holdoff
                                                                              Yes
                         To access the sequence trigger settings, press the [More] key in the Trigger section of
                         the front panel, rotate the Entry knob until Sequence is displayed in the Trigger softkey,
                         then press the Settings softkey to display Sequence Trigger Menu.
    Sequence stage
    definitions
Term definitions
                       As you set stage, term, and channel definitions for the sequence trigger, these settings
                       will be shown in the waveform area of the display.
                       5 If you select an edge term, one channel must be set to a rising edge or a falling edge.
                         All other channel edges will be set to don’t care (X).
                          a Press the Channel softkey (or rotate the Entry knob) to select the channel.
                             As you select a channel, the channel is highlighted in the selected pattern list
                             shown in the waveform area.
                          b Then press the       X softkey to select rising edge or falling edge. All other
                             channels will be defaulted to don’t care (X).
                             If you want to reassign an edge to a different channel, repeat the above step. The
                             value of the original channel edge will be defaulted to X (don't care).
                        If the term(s) used in the “Find:” Sequence Stage condition are set to “don’t care,” the
                        oscilloscope will not trigger. At least one term in the stage must be set to a value other than X
                        (don’t care).
                       4 If you select a pattern term, each channel in the pattern must be set to a 1 (high),
                         0 (low), or X (don't care).
                          a Press the Channel softkey (or rotate the Entry knob) to select the channel.
                          b Press the 0 1 X softkey to set a level for the channel.
                          c Repeat for all channels in the pattern.
                       5 If you select an edge term, one channel must be set to a rising edge or a falling edge.
                         All other channel edges will be set to don’t care (X).
                          a Press the Channel softkey (or rotate the Entry knob) to select the channel.
                             The channel you select is shown in the upper-right corner of the display next to
                             “Seq”.
                          b Then press the      X softkey to select rising edge or falling edge. All other
                             channel edges will be defaulted to don’t care.
                       6 When you set a Trigger on: condition to trigger on Edge 2, you can also select which
                         occurrence of Edge 2 to trigger on.
                          a Make sure Nth Edge 2 or Nth Edge 2 (no re-find) is selected in the Trigger: softkey.
                             When Nth Edge 2 is selected, if the Find event occurs again before the Count (N)
                             event is satisfied, Count (N) will be reset to zero.
                             When Nth Edge 2 (no re-find) is selected, if the Find event occurs again before the
                             Count (N) event is satisfied, the Count (N) is not reset to zero.
                          b Press the Term softkey and select Count (N).
                          c Press the N softkey, then turn the Entry knob to set the number of edges to wait
                            before triggering.
                             N can be set from 1 to 10,000.
                        If the term(s) used in the “Trigger on:” Sequence Stage condition are set to “don’t care,” the
                        oscilloscope will not trigger. At least one term in the stage must be set to a value other than X
                        (don’t care).
SPI Trigger
                   Serial Peripheral Interface (SPI) trigger setup consists of connecting the oscilloscope to a
                   clock, data, and framing signal. You can then trigger on a data pattern that occurs at the
                   start of a frame. The serial data string can be specified to be from 4 to 32 bits long.
                   When you press the Settings softkey, a graphic will be displayed showing the current
                   state of the frame signal, clock slope, number of data bits, and data bit values. Press the
                   Signals softkey in the Settings Menu to see the current source channels for the clock,
                   data, and frame signals.
                   1 Press [Save/Recall]&Default Setup.
                   2 Press the [Label] key to switch labels on.
                   3 Turn on any analog or digital channels that you will be using for the SPI signals.
                   4 Press the [More] key in the Trigger section of the front panel, rotate the Entry knob
                     until SPI is displayed in the Trigger softkey, then press the Settings softkey to display
                     the SPI Trigger Menu.
 Graphic showing
 current state of the
 SPI trigger setup
            Data string
            values
                                Assign        # data bits   Data bit          Data bit    Set all data      Return to
                                channels      in string     select            value       bits to value     previous menu
                   6 Press the Clock softkey or turn the Entry knob to select the channel connected to the
                     SPI serial clock line.
                      As you press the Clock softkey (or rotate the Entry knob), the CLK label for the source
                      channel is automatically set and the channel you select is shown in the upper-right
                      corner of the display next to “SPI”.
                      If you have connected one of the oscilloscope’s analog source channels to the Clock
                      signal: Adjust the trigger level for the selected analog channel to the middle of the
                      waveform by turning the Trigger Level knob.
                      If you have connected one of the oscilloscope’s digital source channels to the Clock
                      signal (this applies to MSO model oscilloscopes only): Press the [D15-D0] key and
                      select Thresholds to set an appropriate threshold level for digital channels.
                   7 Press the slope softkey (      ) to select rising edge or falling edge for the selected
                     Clock source.
                      This determines which clock edge the oscilloscope will use to latch the serial data.
                      When you press the slope softkey, the graphic shown on the display changes to show
                      the current state of the clock signal.
  Labels
  automatically
  set for clock,
  data, and chip
  select signals
  Graphic showing
  current state of SPI
  trigger clock slope and
  chip select polarity or
  timeout selection
                            8 Press the Data softkey or turn the Entry knob to select the channel that is connected
                              to the SPI serial data line. (If the channel you selected is off, switch it on.)
                               As you press the Data softkey (or rotate the Entry knob), the DATA label for the source
                               channel is automatically set and the channel you select is shown in the upper-right
                               corner of the display next to “SPI”.
                               If you have connected one of the oscilloscope’s analog source channels to the Data
                               signal: Adjust the trigger level for the selected analog channel to the middle of the
                               waveform by turning the Trigger Level knob.
                               If you have connected one of the oscilloscope’s digital source channels to the Data
                               signal (this applies to MSO model oscilloscopes only): Press the [D15-D0] key and
                               select Thresholds to set an appropriate threshold level for digital channels.
                            9 Press the Frame by softkey to select a framing signal that the oscilloscope will use for
                              determining which clock edge is the first clock edge in the serial stream.
                               You can set the oscilloscope to trigger during a high chip select (CS), a low chip select
                               (~CS), or after a Timeout period during which the clock signal has been idle.
                      • If the framing signal is set to CS (or ~CS), the first clock edge as defined, rising or
                        falling, seen after the CS (or ~CS) signal transitions from low to high (or high to
                        low) is the first clock in the serial stream.
                         Chip Select — Press the CS or ~CS softkey or turn the Entry knob to select the
                         channel that is connected to the SPI frame line. The label (~CS or CS) for the
                         source channel is automatically set. The data pattern and the clock transition must
                         occur during the time when the framing signal is valid. The framing signal must be
                         valid for the entire data pattern.
                         As you press the CS or ~CS softkey (or rotate the Entry knob), the CS or ~CS label
                         for the source channel is automatically set and the channel you select is shown in
                         the upper-right corner of the display next to “SPI”. When you press the Frame by
                         softkey, the graphic shown on the previous page changes to show timeout
                         selection or the current state of the chip select signal.
                         If you have connected one of the oscilloscope’s analog source channels to the Chip
                         Select signal: Adjust the trigger level for the selected analog channel to the middle
                         of the waveform by turning the Trigger Level knob.
                         If you have connected one of the oscilloscope’s digital source channels to the Chip
                         Select signal (this applies to MSO model oscilloscopes only): Press the [D15-D0]
                         key and select Thresholds to set an appropriate threshold level for digital channels.
                      • If the framing signal is set to Timeout, the oscilloscope generates it’s own internal
                        framing signal after it sees inactivity on the serial clock line.
                         Clock Timeout — Select Clock Timeout in the Frame by softkey, then select the
                         Timeout softkey and turn the Entry knob to set the minimum time that the Clock
                         signal must be idle (not transitioning) before the oscilloscope will search for the
                         Data pattern on which to trigger. When you press the Frame by softkey, the graphic
                         shown on the previous page changes to show timeout selection or the current
                         state of the chip select signal.
                         The Timeout value can be set anywhere from 100 ns to 10 s.
                   10 Press the up-arrow softkey to return to the previous menu.
TV Trigger
                    TV triggering can be used to capture the complicated waveforms of most standard and
                    high-definition analog video signals. The trigger circuitry detects the vertical and
                    horizontal interval of the waveform and produces triggers based on the TV trigger settings
                    you have selected.
                    The oscilloscope’s MegaZoom III technology gives you bright, easily viewed displays of
                    any part of the video waveform. Analysis of video waveforms is simplified by the
                    oscilloscope’s ability to trigger on any selected line of the video signal.
                   It is important, when using a 10:1 passive probe, that the probe is correctly compensated.
      NOTE         The oscilloscope is sensitive to this and will not trigger if the probe is not properly
                   compensated, especially for progressive formats.
                    1 Press the [More] key in the Trigger section of the front panel. If TV is not selected,
                      rotate the Entry knob until TV is displayed in the Trigger softkey, then press the
                      Settings softkey to display TV Trigger Menu.
                                                                            TV             Selected
                                                                            trigger        channel       TV Mode
                       2 Press the Source softkey and select any analog channel as the TV trigger source.
                          The selected trigger source is displayed in the upper-right corner of the display.
                          Turning the Trigger Level knob does not change the trigger level because the trigger
                          level is automatically set to the sync pulse. Trigger coupling is automatically set to TV
                          in the Trigger Mode and Coupling Menu.
                       3 Press the sync polarity softkey to set the TV trigger to either positive (       ) or
                         negative ( ) sync polarity.
                   5 Press the Mode softkey to select the portion of the video signal that you would like to
                     trigger on.
                      The TV trigger modes available are:
                      • Field1 and Field2 — Trigger on the rising edge of the first serration pulse of field 1
                        or field 2 (interlaced standards only).
                      • All Fields — Trigger on the rising edge of the first pulse in the vertical sync interval
                        (not available in Generic mode).
                      • All Lines — Trigger on all horizontal sync pulses.
                      • Line — Trigger on the selected line # (EDTV and HDTV standards only).
                      • Line: Field1 and Line:Field2 — Trigger on the selected line # in field 1 or field 2
                        (interlaced standards only except 1080i).
                      • Line: Alternate — Alternately trigger on the selected line # in field 1 and field 2
                        (NTSC, PAL, PAL-M, and SECAM only).
                      • Vertical — Trigger on the rising edge of the first serration pulse or approximately
                        70 s after the start of vertical sync, whichever occurs first (only available in
                        Generic mode).
                      • Count: Vertical — Counts falling edges of sync pulses; triggers on the selected
                        count number (only available in Generic mode).
                       6 If you select a line # mode, press the Line # softkey, then rotate the Entry knob to
                         select the line number on which you want to trigger.
                       7 When using the Generic standard and you select a line # mode or Count:Vertical,
                         press the Count # softkey and rotate the Entry knob to select the desired count
                         number.
                          Listed below are the line (or count) numbers per field for each non-HDTV/EDTV video
                          standard.
                          Video standard        Field 1               Field 2               Alt Field
                          NTSC                  1 to 263              1 to 262              1 to 262
                          PAL                   1 to 313              314 to 625            1 to 312
                          PAL-M                 1 to 263              264 to 525            1 to 262
                          SECAM                 1 to 313              314 to 625            1 to 312
                          Generic               1 to 1024             1 to 1024             1 to 1024 (vertical)
                 TV Triggering Examples
                       The following are exercises to familiarize you with TV triggering. These exercises use the
                       NTSC video standard.
                        Alternate Triggering
                        If Line:Alternate is selected, the oscilloscope will alternately trigger on the selected line number
                        in Field 1 and Field 2. This is a quick way to compare the Field 1 VITS and Field 2 VITS or to check
                        for the correct insertion of the half line at the end of Field 1.
                   If a more detailed analysis is required, then only one color field should be selected to be
                   the trigger. You can do this by using the TV Holdoff softkey in the trigger More Trigger
                   Menu when the trigger type is set to TV. Press the TV Holdoff softkey and use the Entry
                   knob to adjust the holdoff in half-field increments until the oscilloscope triggers on only
                   one phase of the color burst.
                   A quick way to synchronize to the other phase is to briefly disconnect the signal and then
                   reconnect it. Repeat until the correct phase is displayed.
                   When holdoff is adjusted using the TV Holdoff softkey and the Entry knob, the
                   corresponding holdoff time will be displayed in the Trigger Mode and Coupling Menu.
UART/RS232 Trigger
                   To trigger on a UART (Universal Asynchronous Receiver/Transmitter) signal connect the
                   oscilloscope to the Rx and Tx lines and set up a trigger condition. RS232 (Recommended
                   Standard 232) is one example of a UART protocol.
                   1 Press [Save/Recall]&Default Setup.
                   2 Press the [Label] key to switch labels on.
                   3 Turn on any analog or digital channels that you will be using for the UART/RS232
                     signals.
                   4 Press the [More] key in the Trigger section of the front panel. Rotate the Entry knob
                     until UART/RS232 is displayed in the Trigger softkey
                   5 Press the Settings softkey to display the UART/RS232 Trigger Menu.
                                                                              Currently         Trigger level
                                                                              selected Rx or    or threshold
                                                                              Tx channel
                                                                              UART/RS232
                                                                              trigger mode
                             Rx             Tx                                                       Return to
                             channel        channel                                                  previous menu
                       7 Connect an oscilloscope channel to the Rx signal in the device under test, then press
                         the Rx softkey and rotate the Entry knob to select the channel.
                          As you press the Rx softkey (or rotate the Entry knob), the channel you select is
                          shown in the upper-right corner of the display next to “URT”.
                       8 Make sure the trigger or threshold levels are set to the middle of the signal:
                          • For analog channels, turn the Trigger Level knob.
                          • For digital channels, press the [D15-D0] key and the Thresholds softkey to access
                            the threshold level setting softkeys.
                          The value of the trigger level or digital threshold is displayed in the upper-right corner
                          of the display.
                       9 Repeat these steps for the Tx signal.
                       10 Press the [More] key, then the Settings softkey to return to the UART/RS232 Trigger
                          Menu.
                       11 Press the Bus Config softkey and set the following parameters.
                          a #Bits: Set the number of bits in the UART/RS232 words to match your device
                            under test (selectable from 5-9 bits).
                          b Parity: Choose odd, even, or none, based on your device under test.
                          c Baud: Press the Baud Rate softkey, then press the Baud softkey and select a baud
                            rate to match the signal in your device under test. If the desired baud rate is not
                            listed, select User Defined on the Baud softkey, then select the desired baud rate
                            using the User Baud softkey. You can set the UART baud rate from 1.2 kb/s to
                            3.0000 Mb/s in increments of 100 b/s.
                          d Polarity: Select idle low or idle high to match your device under test’s state when
                            at idle. For RS232 select idle low.
                          e Bit Order: Select whether the most significant bit (MSB) or the least significant bit
                            (LSB) is presented after the start bit in the signal from your device under test. For
                            RS232 select LSB.
USB Trigger
                   USB trigger will trigger on a Start of Packet (SOP), End of Packet (EOP) signal, Reset
                   Complete (RC), Enter Suspend (Suspend), or Exit Suspend (Exit Sus) on the differential
                   USB data lines (D+ and D-). USB Low Speed and Full Speed are supported by this trigger.
                                                                                    End of packet trigger
                   D+,–
                             Bus Idle
                   D–,+
                                        SOP                                        SE0 portion of EOP
                   5 Press the Speed softkey to select the speed of the transaction being probed.
                      You may select Low Speed (1.5 Mb/s) or Full Speed (12 Mb/s).
                   6 Press the D+ and D– softkeys to select the channel connected to the USB signal D+
                     and D– lines. The D+ and D– labels for the source channels are automatically set.
                          As you press the D+ or D– softkey (or rotate the Entry knob), the D+ and D– label for
                          the source channel is automatically set and the channel you select is shown in the
                          upper-right corner of the display next to “USB”.
                          If you have connected the oscilloscope’s analog source channels to the D+ and D-
                          signals: Adjust the trigger level for each connected analog channel to the middle of
                          the waveform by pressing the D+ or D- softkey, then turning the Trigger Level knob.
                          If you have connected the oscilloscope’s digital source channels to the D+ and D-
                          signals (this applies to MSO model oscilloscopes only): Press the [D15-D0] key and
                          select Thresholds to set an appropriate threshold level for the digital channels.
                          The value of the trigger level or digital threshold is displayed in the upper-right corner
                          of the display.
                       7 Press the Trigger: softkey to select where the USB trigger will occur:
                          • SOP (Start of Packet) — triggers at the Sync bit at the beginning of the packet.
                          • EOP (End of Packet) — triggers at the end of the SE0 portion of the EOP.
                          • RC (Reset Complete) — triggers when SE0 is > 10 ms.
                          • Suspend (Enter Suspend) — triggers when the bus is idle > 3 ms.
                          • Exit Sus (Exit Suspend) — triggers when exiting an idle state > 10 ms. This is used
                            to see the suspend/resume transition.
                                                                                            End of packet trigger
                           D+,–
                                    Bus Idle
                           D–,+
                                               SOP                                        SE0 portion of EOP
           5
           Displaying
           Tips for Displaying Waveforms 174
           Stabilizing the Display 174
           Interpreting the Display 175
           Graphic Symbols in Softkey Menus 176
           AutoScale 177
           Pan and Zoom 179
           Waveform Intensity and Signal Detail 180
           Grid Intensity 181
           Infinite Persistence 182
           Using Labels 183
           Waveform Expansion Reference Point 187
           Vectors (Connect the Dots) 187
           Freeze Display 188
           Antialiasing 189
           XGA Video Output 189
                                                                                  173
5     Displaying
                                                                                                      Measurement
                                                                                                      statistics
 Measurement
 line
 Softkeys
                       Status line     The top line of the display contains vertical, horizontal, and trigger setup
                       information.
                       Display area The display area contains the waveform acquisitions, channel identifiers,
                       and analog trigger and ground level indicators. Each analog channel’s information
                       appears in a different color.
                       Measurement line        This line normally contains automatic measurement and cursor
                       results, but can also display advanced trigger setup data and menu information.
                       Softkeys      The softkeys let you set up additional parameters for the selected mode or
                       menu.
AutoScale
                       Press the [AutoScale] key to quickly display the input signals.
                 Undo AutoScale
                       Press the Undo AutoScale softkey to return the oscilloscope to the settings that existed
                       before you pressed the [AutoScale] key.
                       This is useful if you have unintentionally pressed the [AutoScale] key or do not like the
                       settings AutoScale has selected and want to return to your previous settings.
                        Zoom
                        The screen will still contain a relatively good display if you zoom in horizontally by a factor of 1000
                        and zoom in vertically by a factor of 10 to display the information from where it was acquired.
                        Remember that you can only make automatic measurements on displayed data.
                   3 Turn the Horizontal Position (Delay Time) knob ( ) to pan horizontally and turn the
                     channel’s vertical position knob ( ) to pan vertically.
                   The stopped display may contain several triggers worth of information, but only the last
                   trigger acquisition is available for pan and zoom.
Grid Intensity
                       To adjust the grid (graticule) intensity, press [Display]&Grid and use the Entry knob
                       to adjust the intensity.
Infinite Persistence
                   With infinite persistence, the oscilloscope updates the display with new acquisitions, but
                   does not erase the results of previous acquisitions. All previous acquisitions are
                   displayed in gray with reduced intensity. New acquisitions are shown in their normal
                   color with normal intensity. Waveform persistence is kept only for the current display
                   area; you cannot pan and zoom the infinite persistence display.
                   Use infinite persistence to measure noise and jitter, to see the worst-case extremes of
                   varying waveforms, to look for timing violations, or to capture events that occur
                   infrequently.
                   3 Press  Persist to turn on infinite persistence. The display will begin accumulating
                     multiple acquisitions.
                   4 Press the Clear Display softkey to erase previous acquisitions. The oscilloscope will
                     start to accumulate acquisitions again.
                   5 Turn off infinite persistence, then press the Clear Display softkey to return the
                     oscilloscope to the normal display mode.
Using Labels
                       You can define labels and assign them to each analog input channel, or you can turn
                       labels off to increase the waveform display area. Labels can also be applied to digital
                       channels on MSO models.
2 To turn the labels off, press the [Label] key until it is not illuminated.
                      The figure above shows the list of channels and their default labels. The channel does
                      not have to be turned on to have a label assigned to it.
                   3 Press the Library softkey, then turn the Entry knob or successively press the Library
                     softkey to select a predefined label from the library.
                   4 Press the Apply New Label softkey to assign the label to your selected channel.
                   5 Repeat the above procedure for each predefined label you want to assign to a channel.
                          The channel does not have to be turned on to have a label assigned to it. If the channel
                          is turned on, its current label will be highlighted.
                       3 Press the Spell softkey, then turn the Entry knob to select the first character in the
                         new label.
                          Turning the Entry knob selects a character to enter into the highlighted position shown
                          in the “New label =” line above the softkeys and in the Spell softkey. Labels can
                          be up to ten characters in length.
                       4 Press the Enter softkey to enter the selected character and to go to the next character
                         position.
                          You may position the highlight on any character in the label name by successively
                          pressing the Enter softkey.
                       5 To delete a character from the label, press the Enter softkey until the letter you want to
                         delete is highlighted, then press the Delete Character softkey.
                       6 When you are done entering characters for the label, press the Apply New Label
                         softkey to assign the label to the selected channel.
                       When you define a new label, it is added to the nonvolatile label list.
                   Pressing the Default Library softkey will remove all user-defined labels from the
        NOTE
                   library and set the labels back to the factory default. Once deleted, these
                   user-defined labels cannot be recovered.
                          This will delete all user-defined labels from the library and set the labels in the library
                          back to the factory default. However, this does not default the labels currently
                          assigned to the channels (those labels that appear in the waveform area).
                       Expand About Ground The displayed waveform will expand about the position of the
                       channel's ground. This is the default setting. The ground level of the signal is identified by
                       the position of the ground level ( ) icon at the far-left side of the display. The ground
                       level will not move when you adjust the vertical sensitivity (volts/division) control.
                       If the ground level is off screen, the waveform will expand about the top or bottom edge of
                       the screen based on where the ground is off screen.
                       Expand About Center The displayed waveform will expand about the center of the
                       display.
                   The oscilloscope turns vectors on whenever the acquisition system stops. Digital
                   channels on the mixed-signal oscilloscope are not affected by the Display Menu. They are
                   always displayed with vectors on. They also only contain one trigger worth of information.
Freeze Display
                   To freeze the display without stopping running acquisitions:
                   1 Press the [Display] key.
                   2 Press the Freeze Display softkey.
                   To un-freeze the display, press Freeze Display again.
                   Many activities, such as adjusting the trigger level, adjusting vertical or horizontal
                   settings, or saving data will un-freeze the display.
                   Manual cursors can be used on the frozen display.
Antialiasing
                       At slower sweep speeds, the sample rate is reduced and a proprietary display algorithm is
                       used to minimize the likelihood of aliasing.
                       By default, Antialiasing is enabled. You should leave Antialiasing enabled unless there is
                       a specific reason to switch it off.
                       If you need to switch Antialiasing off, press [Utility]&Options&Preferences and press
                       the Antialiasing softkey to switch the feature off. The displayed waveforms will be more
                       susceptible to aliasing.
           6
           Measurements and Math Functions
           List of Automatic Measurements 192
           Making Automatic Measurements 193
              Measurement Statistics 194
              Measurement Thresholds 197
              Measurement Definitions 199
              Time Measurements 199
              Delay and Phase Measurements 203
              Voltage Measurements 205
              Overshoot and Preshoot Measurements 211
           Cursor Measurements 214
           XY Horizontal Mode 220
           Math Functions 224
              Multiply 227
              Add or Subtract 228
              Differentiate 230
              Integrate 232
              Square Root 234
              FFT Measurement 236
           Enabling Presision Measurements and Math 242
See Also “To specify the measurement window for Zoom mode” on page 68.
                                                                         191
6     Measurements and Math Functions
                     Time Measurements
                     •   “Counter” on page 199.
                     •   “Duty Cycle” on page 200.
                     •   “Frequency” on page 200.
                     •   “Period” on page 201.
                     •   “Rise Time” on page 201.
                     •   “Fall Time” on page 201.
                     •   “+ Width” on page 202.
                     •   “– Width” on page 202.
                     •   “X at Max Y” on page 202.
                     •   “X at Min Y” on page 202.
                     Voltage Measurements
                     •   “Average” on page 207.
                     •   “Amplitude” on page 206.
                     •   “Base” on page 207.
                     •   “Extinction Ratio” on page 207.
                     •   “Maximum” on page 207.
                     •   “Minimum” on page 207.
                     •   “Peak-Peak” on page 208.
                     •   “Ratio” on page 208.
                     •   “RMS” on page 208.
                     •   “Std Deviation” on page 208.
                     •   “Top” on page 210.
                       2 Press the Source softkey to select the channel or running math function to be
                         measured.
                          Only channels or math functions that are displayed are available for measurements. If
                          you choose an invalid source channel for a measurement, the measurement will
                          default to the nearest in the list that makes the source valid.
                          If a portion of the waveform required for a measurement is not displayed or does not
                          display enough resolution to make the measurement, the result will display “No
                          Edges”, “Clipped”, “Low Signal”, “< value”, or “> value”, or a similar message to
                          indicate that the measurement may not be reliable.
                     3 Press the Select softkey then rotate the Entry knob to select a measurement to be
                       made.
                        After all measurements have been cleared, when [Quick Meas] is pressed again, the
                        default measurements will be Frequency and Peak-Peak.
                Measurement Statistics
                     Press the [Quick Meas] key to enter the Measurement Menu. By default, statistics are
                     displayed, and frequency and voltage are measured on channel 1.
                     Select the measurements you desire for the channels you are using (see page 192).
                     From the Measurement Menu, press the Statistics softkey to enter the Statistics Menu.
Statistics
     Source channel
     (color coded)
                       The following statistics are shown: Name of the measurement, current measured value,
                       mean, minimum measured value, maximum measured value, standard deviation, and the
                       number of times the measurement has been made (count). Statistics are based on the
                       total number of captured waveforms (count).
                       The standard deviation shown in Statistics is calculated using the same formula used for
                       calculating the standard deviation Quick Measurement. The formula is shown in the
                       section titled “Std Deviation” on page 208.
                       The source channel of the measurement is shown in parenthesis after the measurement
                       name. For example: "Freq(1)" indicates a frequency measurement on channel 1.
                       You can switch the statistics Display On or Display Off. Statistics continue to accumulate
                       even when the statistics display is off.
                       When you navigate away from the Measurements Menu, the statistics display will no
                       longer be displayed, but statistical data collection will continue. Return to the
                       Measurements Menu to see the data again.
                     To reset the statistics measurements, press the Reset Statistics softkey. This resets all
                     statistics and begins recording statistical data again.
                     Each time a new measurement (for example: frequency, period, or amplitude) is added the
                     statistics are reset and accumulation of statistical data begins again.
                     When the [Single] key is pressed, statistics are reset and a single measurement is done
                     (count = 1). Successive [Single] acquisitions accumulate statistical data ( and the count
                     is incremented).
                     Press the Transparent softkey to disable the Transparent mode. This shows statistics
                     with a gray background. Press the Transparent softkey again to enable Transparent mode.
                     This writes measurement values, statistics, and cursor values on the screen without a
                     background. This controls the display of statistics in both Quick Meas and the optional
                     Mask test feature.
                     The Increment Statistics softkey only appears when the acquisition is stopped and the
                     optional segmented memory feature is off. Press the [Single] or [Run/Stop] key to stop
                     the acquisition. You can use the horizontal position control (in the Horizontal control
                     section of the front panel) to pan through the waveform. Active measurements will stay
                     on screen, allowing you to measure various aspects of the captured waveforms. Press
                     Increment Statistics to add the currently measured waveform to the collected statistical
                     data.
                     The Analyze Segments softkey only appears when the acquisition is stopped and the
                     optional segmented memory feature is on. After an acquisition has completed (and the
                     oscilloscope is stopped), you can press the Analyze Segments softkey to accumulate
                     measurement statistics for the acquired segments.
                     You can also turn on infinite persistence (in the Display Menu) and press the Analyze
                     Segments softkey to create an infinite persistence display.
                 Measurement Thresholds
                       Setting measurement thresholds defines the vertical levels where measurements will be
                       taken on an analog channel or math waveform.
                       1 From the Quick Meas Menu, press the Settings softkey, then press the Thresholds
                         softkey to set analog channel measurement thresholds.
                       2 Press the Source softkey to select the analog channel or math waveform source for
                         which you want to change measurement thresholds. Each analog channel and the
                         math waveform can be assigned unique threshold values.
                       3 Press the Type softkey to set the measurement threshold to % (percentage of Top and
                         Base value) or to Absolute (absolute value.)
                          • Percentage thresholds can be set from 5% to 95%.
                          • The units for absolute threshold for each channel is set in the channel probe menu.
                          • When the Source is set to Math: f(t), the threshold Type can only be set to
                            Percent.
                       4 Press the Lower softkey, then turn the Entry knob to set the lower measurement
                         threshold value.
                        Increasing the lower value beyond the set middle value will automatically increase the
                        middle value to be more than the lower value. The default lower threshold is 10% or
                        800 mV.
                        If threshold Type is set to %, the lower threshold value can be set from 5% to 93%.
                     5 Press the Middle softkey, then turn the Entry knob to set the middle measurement
                       threshold value.
                        The middle value is bounded by the values set for lower and upper thresholds. The
                        default middle threshold is 50% or 1.20 V.
                        • If threshold Type is set to %, the middle threshold value can be set from 6% to 94%.
                     6 Press the Upper softkey, then turn the Entry knob to set the upper measurement
                       threshold value.
                        Decreasing the upper value below the set middle value will automatically decrease the
                        middle value to be less than the upper value. The default upper threshold is 90% or
                        1.50 V.
                        • If threshold Type is set to %, the upper threshold value can be set from 7% to 95%.
Measurement Definitions
Time Measurements
                        FFT measurements
                        When you make an X at Max Y or X at Min Y measurement on a math FFT function, the resultant
                        units will be in Hertz. No other time related automatic measurement can be made on the FFT math
                        function. Use the cursors to make other measurements on FFT.
                       The default lower, middle, and upper measurement thresholds are 10%, 50%, and 90%
                       between Top and Base values. See “Measurement Thresholds” on page 197 for other
                       percentage threshold and absolute value threshold settings.
                       The following figure shows time measurement points.
                             Rise Time              Fall Time
                                                                                                        Thresholds
                                                                                                           Upper
                                                                                                           Middle
                                                                                                           Lower
                                         + Width                       – Width
                                                            Period
                       Counter
                       The InfiniiVision 7000 Series oscilloscopes have an integrated hardware frequency
                       counter which counts the number of cycles that occur within a period of time (known as
                       the gate time) to measure the frequency of a signal.
                       The gate time for the Counter measurement is automatically adjusted to be 100 ms or
                       twice the current time window, whichever is longer, up to 1 second.
                       The Counter can measure frequencies up to the bandwidth of the oscilloscope. The
                       minimum frequency supported is 1/(2 X gate time).
                     The measured frequency is normally displayed in 5 digits, but can be displayed in 8 digits
                     when an external 10 MHz frequency reference is provided at the 10 MHz REF rear panel
                     BNC and the gate time is 1 second (50 ms/div sweep speed or greater). See (“To supply
                     a sample clock to the oscilloscope” on page 406.)
                     The hardware counter uses the trigger comparator output. Therefore, the counted
                     channel’s trigger level (or threshold for digital channels) must be set correctly. The Y
                     cursor shows the threshold level used in the measurement.
                     Any channel except Math can be selected as the source.
                     Only one Counter measurement can be displayed at a time.
                     Duty Cycle
                     The duty cycle of a repetitive pulse train is the ratio of the positive pulse width to the
                     period, expressed as a percentage. The X cursors show the time period being measured.
                     The Y cursor shows the middle threshold point.
                                   + Width
                     Duty cycle = ------------------- X 100
                                      Period
                     Frequency
                     Frequency is defined as 1/Period. Period is defined as the time between the middle
                     threshold crossings of two consecutive, like-polarity edges. A middle threshold crossing
                     must also travel through the lower and upper threshold levels which eliminates runt
                     pulses. The X cursors show what portion of the waveform is being measured. The Y
                     cursor shows the middle threshold point.
                     To isolate an event for frequency measurement The following figure shows how to
                     use Zoom mode to isolate an event for a frequency measurement. If the measurement is
                     not possible in the Zoom mode, then the normal time base is used. If the waveform is
                     clipped, it may not be possible to make the measurement.
                       Period
                       Period is the time period of the complete waveform cycle. The time is measured between
                       the middle threshold points of two consecutive, like-polarity edges. A middle threshold
                       crossing must also travel through the lower and upper threshold levels which eliminates
                       runt pulses. The X cursors show what portion of the waveform is being measured. The Y
                       cursor shows the middle threshold point.
                       Fall Time
                       The fall time of a signal is the time difference between the crossing of the upper
                       threshold and the crossing of the lower threshold for a negative-going edge. The X cursor
                       shows the edge being measured. For maximum measurement accuracy, set the sweep
                       speed as fast as possible while leaving the complete falling edge of the waveform on the
                       display. The Y cursors show the lower and upper threshold points.
                       Rise Time
                       The rise time of a signal is the time difference between the crossing of the lower
                       threshold and the crossing of the upper threshold for a positive-going edge. The X cursor
                       shows the edge being measured. For maximum measurement accuracy, set the sweep
                       speed as fast as possible while leaving the complete rising edge of the waveform on the
                       display. The Y cursors show the lower and upper threshold points.
                     + Width
                     + Width is the time from the middle threshold of the rising edge to the middle threshold of
                     the next falling edge. The X cursors show the pulse being measured. The Y cursor shows
                     the middle threshold point.
                     – Width
                     – Width is the time from the middle threshold of the falling edge to the middle threshold
                     of the next rising edge. The X cursors show the pulse being measured. The Y cursor
                     shows the middle threshold point.
                     X at Max Y
                     X at Max Y is the X axis value (usually time) at the first displayed occurrence of the
                     waveform Maximum, starting from the left-side of the display. For periodic signals, the
                     position of the maximum may vary throughout the waveform. The X cursor shows where
                     the current X at Max Y value is being measured.
                     To measure the peak of an FFT:
                     1 Select FFT as the Operator in the Waveform Math Menu.
                     2 Choose Math: f(t) as the source in the Quick Meas Menu.
                     3 Choose Maximum and X at Max Y measurements.
                        Maximum units are in dB and X at Max Y units are in Hertz for FFT.
                     X at Min Y
                     X at Min Y is the X axis value (usually time) at the first displayed occurrence of the
                     waveform Minimum, starting from the left-side of the display. For periodic signals, the
                     position of the minimum may vary throughout the waveform. The X cursor shows where
                     the current X at Min Y value is being measured.
                       Delay
                       Delay measures the time difference from the selected edge on source 1 and the selected
                       edge on source 2 closest to the trigger reference point at the middle threshold points on
                       the waveforms. Negative delay values indicate that the selected edge of source 1
                       occurred after the selected edge of source 2.
                       Source 1
                                                                 Delay
                       Source 2
                       1 Press the [Quick Meas]&Select and select Delay. Press the Settings softkey to
                         select the source channels and slope for the delay measurement.
                          The default Delay settings measure from the rising edge of channel 1 to the rising
                          edge of channel 2.
                       2 Press the Measure Delay softkey to make the measurement.
                     The example below shows a delay measurement between the rising edge of channel 1
                     and the rising edge of channel 2.
                     Phase
                     Phase is the calculated phase shift from source 1 to source 2, expressed in degrees.
                     Negative phase shift values indicate that the rising edge of source 1 occurred after the
                     rising edge of source 2.
                                          Delay
                     Phase = ----------------------------------------- X 360
                                Source 1 Period
                                                                                       Period
                     Source 1
                                                                               Delay
                     Source 2
                     1 Press the Settings softkey to select the source 1 and source 2 channels for the phase
                       measurement.
                         The default Phase settings measure from channel 1 to channel 2.
                       The example below shows a phase measurement between the channel 1 and the math
                       d/dt function on channel 1.
                 Voltage Measurements
                       Measurement units for each input channel can be set to Volts or Amps using the channel
                       Probe Units softkey. A scale unit of U (undefined) will be displayed for math function 1-2
                       and for d/dt, and  dt when 1-2 or 1+2 is the selected source if channel 1 and channel 2
                       are set to dissimilar units in the channel Probe Units softkey.
Maximum Top
Amplitude Peak-Peak
                                                                                                             Base
                                                                                              Minimum
                     Amplitude
                     The Amplitude of a waveform is the difference between its Top and Base values. The Y
                     cursors show the values being measured.
                       Average
                       Average is the sum of the waveform samples divided by the number of samples.
                       Base
                       The Base of a waveform is the mode (most common value) of the lower part of the
                       waveform, or if the mode is not well defined, the base is the same as Minimum. The Y
                       cursor shows the value being measured.
                       Extinction Ratio
                       The extinction ratio is the ratio of the "logic one" level amplitude (V top) to the "logic
                       zero" level amplitude (V base), in dB. That is, the extinction ratio is: 10 log(Vtop/Vbase).
                       Maximum
                       Maximum is the highest value in the waveform display. The Y cursor shows the value
                       being measured.
                       Minimum
                       Minimum is the lowest value in the waveform display. The Y cursor shows the value being
                       measured.
                     Peak-Peak
                     The peak-to-peak value is the difference between Maximum and Minimum values. The Y
                     cursors show the values being measured.
                     Ratio
                     The Ratio measurement displays the ratio of the AC RMS voltages of two sources,
                     expressed in dB. Press the Settings softkey to select the source channels for the
                     measurement.
                     RMS
                     RMS (DC) is the root-mean-square value of the waveform.
                                        n
                                       xi         2
                                                        where xi = value at ith point being measured
                                     i=1                n = number of points in measurement interval
                                     ----------------
                      RMS (dc) =            n
                     Std Deviation
                     The Std Deviation measurement shows the standard deviation of the displayed voltage
                     values. It is an RMS measurement across the full screen with the DC component
                     removed. It is useful, for example, for measuring power supply noise.
                     The standard deviation of a measurement is the amount that a measurement varies from
                     the mean value. The Mean value of a measurement is the statistical average of the
                     measurement.
                       The following figure graphically shows the mean and standard deviation. Standard
                       deviation is represented by the Greek letter sigma: . For a Gaussian distribution, two
                       sigma (± 1) from the mean, is where 68.3 percent of the measurement results reside. Six
                       sigma (± 3) from is where 99.7 percent of the measurement results reside.
mean
                                    xi
                                  =1 -
                          x = i-------------
                                    N
                       where:
                          x = the mean.
                          N = the number of measurements taken.
                          xi = the ith measurement result.
                       The standard deviation is calculated as follows:
                                        N
                                                               2
                                          xi – x 
                           =        i=1
                                     -----------------------------
                                                                 -
                                              N–1
                     where:
                         = the standard deviation.
                        N = the number of measurements taken.
                        xi = the ith measurement result.
                        x = the mean.
                     Top
                     The Top of a waveform is the mode (most common value) of the upper part of the
                     waveform, or if the mode is not well defined, the top is the same as Maximum. The Y
                     cursor shows the value being measured.
                     To isolate a pulse for Top measurement The following figure shows how to use
                     Zoom mode to isolate a pulse for a Top measurement.
                     ISOLATING AREA FOR TOP MEASUREMENT
                       Preshoot
                       Preshoot is distortion that precedes a major edge transition expressed as a percentage of
                       Amplitude. The X cursors show which edge is being measured (edge closest to the trigger
                       reference point).
                                              Base - D local Minimum
                       Rising edge preshoot = ----------------------------------------------------------- X 100
                                                               Amplitude
Base
                     Overshoot
                     Overshoot is distortion that follows a major edge transition expressed as a percentage of
                     Amplitude. The X cursors show which edge is being measured (edge closest to the trigger
                     reference point).
                                             local Maximum - D Top
                     Rising edge overshoot = ---------------------------------------------------------- X 100
                                                             Amplitude
                                              Base - D local Minimum
                     Falling edge overshoot = ----------------------------------------------------------- X 100
                                                               Amplitude
                                                  Overshoot
                                                                                   local Maximum
                                                                                                                     Top
Base
Cursor Measurements
                     Cursors are horizontal and vertical markers that indicate X-axis values (usually time) and
                     Y-axis values (usually voltage) on a selected waveform source. You can use cursors to
                     make custom voltage or time measurements on oscilloscope signals, and timing
                     measurements on digital channels. Cursor information is displayed on the line above the
                     softkeys.
                     Cursors are not always limited to the visible display. If you set a cursor, then pan and
                     zoom the waveform until the cursor is off screen, its value will not be changed. It will still
                     be there when you return to its original location.
                     X cursors are vertical dashed lines that adjust horizontally and normally indicate time
                     relative to the trigger point. When used with the FFT math function as a source, the X
                     cursors indicate frequency.
                     The X1 cursor (short-dashed vertical line) and X2 cursor (long-dashed-vertical line) adjust
                     horizontally and indicate time relative to the trigger point for all sources except math FFT
                     (frequency is indicated).
                     In XY horizontal mode, the X cursors display channel 1 values (Volts or Amps).
                     The cursor values for the selected waveform source are displayed within the X1 and X2
                     softkeys.
                     The difference between X1 and X2 (X) and 1/X are displayed on the dedicated line
                     above the softkeys or in the display area when some menus are selected.
                     Y cursors are horizontal dashed lines that adjust vertically and normally indicate Volts or
                     Amps, dependent on the channel Probe Units setting. When math functions are used as
                     a source, the measurement units correspond to that math function.
                     The Y1 cursor (short-dashed horizontal line) and Y2 cursor (long-dashed horizontal line)
                     adjust vertically and indicate values relative to the waveform's ground point, except math
                     FFT where the values are relative to 0 dB.
                     In XY horizontal mode, the Y cursors display channel 2 values (Volts or Amps).
                     The cursor values for the selected waveform source are displayed within the Y1 and Y2
                     softkeys.
                     The difference between Y1 and Y2 (Y) is displayed on the dedicated line above the
                     softkeys or in the display area when some menus are selected.
                       If you intend to make cursor measurements on a trace that you recall from memory, be
       NOTE            sure to recall both setup and trace. See “To recall waveform trace and/or oscilloscope
                       setup” on page 256.
                        Manual and Track Waveform modes can be used on waveforms that are displayed on
                        the analog input channels (including math functions).
                        Binary and Hex modes apply to digital signals (of MSO oscilloscope models).
                        In Hex and Binary modes, a level can be displayed as 1 (higher than trigger level), 0
                        (lower than trigger level), indeterminate state ( ), or X (don't care).
                        In Binary mode, X is displayed if the channel is turned off.
                        In Hex mode, the channel is interpreted as a 0 if turned off.
                     4 Press Source (or X1 Source, X2 Source in the Track Waveform mode); then, select
                       the input source for cursor values.
                     5 In the Manual mode, press X Y to select between X and Y cursors for adjustment.
                     6 Adjust the cursor(s) by pressing one of these softkeys:
                        • X1 — adjust horizontally and normally measure time.
                        • X2 — adjust horizontally and normally measure time.
                        • X1 X2 — adjusts both cursors at the same time, while the delta value remains the
                          same. This can be useful, for example, for checking pulse width variations in a
                          pulse train.
                        • Y1 — adjust vertically and normally measure voltage.
                        • Y2 — adjust vertically and normally measure voltage.
                        • Y1 Y2 — adjusts both cursors at the same time, while the delta value remains the
                          same.
                        Then, turn the Entry knob.
                        The cursor currently assigned to the Entry knob displays brighter than the other
                        cursors.
                     After adjusting the X1, X2, X1 X2, Y1, Y2, or Y1 Y2 controls, you can press the Entry knob
        NOTE         to adjust the next control.
                 Cursor Examples
                       CURSORS USED TO MEASURE PULSE WIDTHS OTHER THAN MIDDLE THRESHOLD POINTS
                     Expand the display with Zoom mode, then characterize the event of interest with the
                     cursors.
                     CURSORS TRACK ZOOM SWEEP
                     Put the X1 cursor on one side of a pulse and the X2 cursor on the other side of the pulse.
                     MEASURING PULSE WIDTH WITH CURSORS
                       Press the X1 X2 softkey and move the cursors together to check for pulse width
                       variations in a pulse train.
                       MOVING THE CURSORS TOGETHER TO CHECK PULSE WIDTH VARIATIONS
XY Horizontal Mode
                     The XY horizontal mode converts the oscilloscope from a volts-versus-time display to a
                     volts-versus-volts display using two input channels. Channel 1 is the X-axis input, channel
                     2 is the Y-axis input. You can use various transducers so the display could show strain
                     versus displacement, flow versus pressure, volts versus current, or voltage versus
                     frequency. This exercise shows a common use of the XY display mode by measuring the
                     phase difference between two signals of the same frequency with the Lissajous method.
                     1 Connect a sine wave signal to channel 1, and a sine wave signal of the same
                       frequency but out of phase to channel 2.
                     2 Press the [AutoScale] key, press the [Menu/Zoom] key (or [Main/Delayed] key on
                       some oscilloscopes), then press the XY or Roll softkey to select “XY”.
                     3 Center the signal on the display with the channel 1 and 2 position ( ) knobs. Use the
                       channel 1 and 2 volts/div knobs and the channel 1 and 2 Fine softkeys to expand the
                       signal for convenient viewing.
                        The phase difference angle () can be calculated using the following formula
                        (assuming the amplitude is the same on both channels):
                                 A      C
                         sin  = --- or ----
                                 B D
                                  Signal must
                                  be centered in
                                  “X”
D A B
                     6 Move the Y1 and Y2 cursors to the intersection of the signal and the Y axis. Again,
                       note the Y value.
                     CURSORS SET TO CENTER OF SIGNAL
Math Functions
                     Math functions can be performed on analog channels. The resulting math waveform is
                     displayed in light purple.
                     You can use a math function on a channel even if you choose not to display the channel
                     on-screen.
                     You can:
                     • Perform an arithmetic operation (add, subtract, or multiply) on channels 1 and 2 or on
                       channels 3 and 4.
                     • Perform a function (differentiate, integrate, FFT, or square root) on the signal acquired
                       on an analog channel.
                     • Perform a function on the result of an arithmetic operation.
                     2 If f(t) is not already shown on the Function softkey, press the Function sofkey and
                       select f(t): Displayed.
                     3 Use the Operator softkey to select an operator.
                     4 Use the Source 1 softkey to select the analog channel on which to perform math. You
                       can rotate the Entry knob or repetitively press the Source 1 softkey to make your
                       selection. If you choose a function (differentiate, integrate, FFT, or square root) the
                       result is displayed.
                     5 If you select an arithmetic operator, use the Source 2 softkey to select the second
                       source for the arithmetic operation. The result is displayed.
                     6 Use the Scale and Offset softkeys to re-size and re-position the math waveform.
                       1 In the Waveform Math Menu, press the Scale or Offset sofkeys to set your own scale
                         factors (units/division) or offset (units) for the selected math function. The Function
                         softkey must be set to f(t): Displayed in order to set scale and offset.
                       2 Press the Scale or Offset softkey, then turn the Entry knob to rescale or change the
                         offset value for your math function.
                Units
                     Units for each input channel can be set to Volts or Amps using the Units softkey in the
                     channel’s Probe Menu. Scale and offset units are:
add or subtract V or A
 dt Vs or As (V-seconds or A-seconds)
                     * When the FFT source is channel 1, 2, 3 or 4, FFT units will be displayed in dBV when channel
                     units is set to Volts and channel impedance is set to 1 M. FFT units will be displayed in dBm
                     when channel units is set to Volts and channel impedance is set to 50. FFT units will be
                     displayed as dB for all other FFT sources or when a source channel’s units has been set to Amps.
                     A scale unit of U (undefined) will be displayed for math functions when two source
                     channels are used and they are set to dissimilar units and the combination of units
                     cannot be resolved.
Multiply
                       When you select the multiply math function, the voltage values of the channel chosen in
                       Source 1 are multiplied point by point by the voltage values of the channel chosen in
                       Source 2, and the result is displayed. Multiply is useful for seeing power relationships
                       when one of the channels is proportional to the current.
                       Example: Multiply channel 1 by channel 2
                          • Scale — lets you set your own vertical scale factors for multiply expressed as
                            V2/div (Volts-squared/division), A2/div (Amps-squared/division), or W/div
                            (Watts/division or Volt-Amps/division). Units are set in the channel Probe Menu.
                            Press the Scale softkey, then turn the Entry knob to rescale the math waveform.
                          • Offset — lets you set your own offset for the multiply math function. The offset
                            value is in V2 (Volts-squared), A2 (Amps-squared), or W (Watts) and is represented
                            by the center horizontal grid line of the display. Press the Offset softkey, then turn
                            the Entry knob to change the offset for the math waveform.
                       EXAMPLE OF MULTIPLY
Channel 1
Channel 2
1 * 2 waveform
Add or Subtract
                     When you select add or subtract, the Source 1 and Source 2 channel voltage values are
                     added or subtracted point by point, and the result is displayed.
                     You can use subtract to make a differential measurement or to compare two waveforms.
                     If your waveforms’ DC offsets are larger than the dynamic range of the oscilloscope's
                     input channels you will need to use a differential probe instead.
                     Example: Subtract channel 2 from channel 1
                        • Scale — lets you set your own vertical scale factors for subtract, expressed as
                          V/div (Volts/division) or A/div (Amps/division). Press the Scale softkey, then turn
                          the Entry knob to rescale. Units are set in the channel Probe Menu.
                        • Offset — lets you set your own offset for the math function. The offset value is in
                          Volts or Amps and is represented by the center horizontal grid line of the display.
                          Press the Offset softkey, then turn the Entry knob to change the offset of the math
                          waveform.
                        A scale unit of U (undefined) will be displayed for scale and offset if the two source
                        channels are set to dissimilar units. Units are set using the channel’s Probe Units
                        softkey.
EXAMPLE OF SUBTRACT
Channel 1
Channel 2
1 – 2 waveform
Differentiate
                     d/dt (differentiate) calculates the discrete time derivative of the selected source. You can
                     use differentiate to measure the instantaneous slope of a waveform. For example, the
                     slew rate of an operational amplifier may be measured using the differentiate function.
                     Because differentiation is very sensitive to noise, it is helpful to set acquisition mode to
                     Averaging in the Acquire Menu.
                     d/dt plots the derivative of the selected source using the “average slope estimate at 4
                     points” formula. The equation is:
                     di = yi+4 + 2yi+2 – 2yi-2 – yi-4
                                      8t
                     Where
                     d = differential waveform
                     y = channel 1, 2, 3, or 4, or g(t) (internal arithmetic operation) data points
                     i = data point index
                     t = point-to-point time difference
                     1 Press the [Math] key, press the Function softkey and select f(t), press the Operator
                       softkey and select d/dt. Press the Source, Scale, or Offset sofkeys if you want to
                       change the source, scaling, or offset for the differentiate function.
                        • Source — selects the source for d/dt. (See page 225 for information about using
                          g(t) as the source.)
                        • Scale — lets you set your own vertical scale factors for d/dt expressed in
                          units/second/division, where units can be V (Volts), A (Amps), or W (Watts). Units
                          are set in the channel Probe Menu. Press the Scale softkey, then turn the Entry
                          knob to rescale d/dt.
                        • Offset —lets you set your own offset for the dV/dt math function. The offset value
                          is in units/second where units can be V (Volts), A (Amps), or W (Watts) and is
                          represented by the center horizontal grid line of the display. Press the Offset
                          softkey, then turn the Entry knob to change the offset for d/dt.
                        A scale unit of U (undefined) will be displayed for scale and offset if the two source
                        channels are set to dissimilar units. Units are set using the channel’s Probe Units
                        softkey.
d/dt waveform
Channel 1
Integrate
                      dt (integrate) calculates the integral of the selected source. You can use integrate to
                     calculate the energy of a pulse in volt-seconds or measure the area under a waveform.
                      dt plots the integral of the source using the “Trapezoidal Rule”. The equation is:
                                              n
                      In = co + t   yi
                                              i=0
                     Where
                     I = integrated waveform
                     t = point-to-point time difference
                     y = channel 1, 2, 3, or 4, or g(t) (internal arithmetic operation)
                     co = arbitrary constant
                     i = data point index
                     1 Press the [Math] key, press the Function softkey and select f(t), press the Operator
                       softkey and select dt. Press the Source, Scale, or Offset sofkeys if you want to
                       change the source, scaling, or offset for the differentiate function.
                        • Source — selects the source for  dt. (See page 225 for information about using
                          g(t) as the source.)
                        • Scale — lets you set your own vertical scale factors for  dt expressed in
                          unit-seconds/division, where units can be V (Volts), A (Amps), or W (Watts). Units
                          are set in the channel Probe Menu. Press the Scale softkey, then turn the Entry
                          knob to rescale  dt.
                        • Offset — lets you set your own offset for the  Vdt math function. The offset value
                          is in unit-seconds where units can be V (Volts), A (Amps), or W (Watts) and is
                          represented by the center horizontal grid line of the display. Press the Offset
                          softkey, then turn the Entry knob to change the offset for  dt. The integrate
                          calculation is relative to the source signal’s offset. The following examples
                          illustrate the effects of signal offset.
Channel 1
0V
dt waveform
Channel 1
0V
dt waveform
Square Root
                     Square root () calculates the square root of the selected source.
                     1 Press the [Math] key, press the press the Function softkey and select f(t), press the
                       Operator softkey and select  (square root). Press the Source 1, Scale, or Offset
                       sofkeys if you want to change the source, scaling, or offset for the square root
                       function.
                        • Source 1 — selects the source for  (square root). (See page 225 for information
                          about using g(t) as the source.)
                        • Scale — lets you set your own vertical scale factors for  (square root) expressed
                          as V1/2/div (Volts-square-root/division), A1/2/div (Amps-square-root/division), or
                          W1/2/div (Watts-square-root/division or Volt-Amps-square-root/division). Units
                          are set in the channel Probe Menu. Press the Scale softkey, then turn the Entry
                          knob to rescale  (square root).
                        • Offset — lets you set your own offset for the multiply math function. The offset
                          value is in V1/2 (Volts-square-root), A1/2 (Amps-square-root), or W1/2
                          (Watts-square-root) and is represented by the center horizontal grid line of the
                          display. Press the Offset softkey, then turn the Entry knob to change the offset for 
                          (square root).
                        A scale unit of U (undefined) will be displayed for scale and offset if the two source
                        channels are set to dissimilar units. Units are set using the channel’s Probe Units
                        softkey.
Channel 1
 waveform
FFT Measurement
                     FFT is used to compute the fast Fourier transform using analog input channels or an
                     arithmetic operation g(t). FFT takes the digitized time record of the specified source and
                     transforms it to the frequency domain. When the FFT function is selected, the FFT
                     spectrum is plotted on the oscilloscope display as magnitude in dBV versus frequency.
                     The readout for the horizontal axis changes from time to frequency (Hertz) and the
                     vertical readout changes from volts to dB.
                     Use the FFT function to find crosstalk problems, to find distortion problems in analog
                     waveforms caused by amplifier non-linearity, or for adjusting analog filters.
                     FFT Units
                     0 dBV is the amplitude of a 1 Vrms sinusoid. When the FFT source is channel 1 or channel
                     2 (or channel 3 or 4 on 4-channel models), FFT units will be displayed in dBV when
                     channel units is set to Volts and channel impedance is set to 1 M.
                     FFT units will be displayed in dBm when channel units is set to Volts and channel
                     impedance is set to 50.
                     FFT units will be displayed as dB for all other FFT sources or when a source channel’s
                     units has been set to Amps.
                     DC Value
                     The FFT computation produces a DC value that is incorrect. It does not take the offset at
                     center screen into account. The DC value is not corrected in order to accurately represent
                     frequency components near DC.
                     Aliasing
                     When using FFTs, it is important to be aware of frequency aliasing. This requires that the
                     operator have some knowledge as to what the frequency domain should contain, and also
                     consider the sampling rate, frequency span, and oscilloscope vertical bandwidth when
                     making FFT measurements. The FFT sample rate is displayed directly above the softkeys
                     when the FFT Menu is displayed.
                     Aliasing happens when there are frequency components in the signal higher than half the
                     sample rate. Since the FFT spectrum is limited by this frequency, any higher components
                     are displayed at a lower (aliased) frequency.
                       The following figure illustrates aliasing. This is the spectrum of a 990 Hz square wave,
                       which has many harmonics. The FFT sample rate is set to 100 kSa/s, and the oscilloscope
                       displays the spectrum. The displayed waveform shows the components of the input
                       signal above the Nyquist frequency to be mirrored (aliased) on the display and reflected
                       off the right edge.
                       ALIASING
                       Since the frequency span goes from  0 to the Nyquist frequency, the best way to prevent
                       aliasing is to make sure that the frequency span is greater than the frequencies of
                       significant energy present in the input signal.
                       Spectral Leakage
                       The FFT operation assumes that the time record repeats. Unless there is an integral
                       number of cycles of the sampled waveform in the record, a discontinuity is created at the
                       end of the record. This is referred to as leakage. In order to minimize spectral leakage,
                       windows that approach zero smoothly at the beginning and end of the signal are
                       employed as filters to the FFT. The FFT Menu provides four windows: Hanning, flattop,
                       rectangular, and Blackman-Harris. For more information on leakage, see Agilent
                       Application Note 243, “The Fundamentals of Signal Analysis” at
                       http://cp.literature.agilent.com/litweb/pdf/5952-8898E.pdf.
                FFT Operation
                     1 Press the [Math] key, press the Function softkey and select f(t), press the Operator
                       softkey and select FFT.
                                                                                                    Sample
                                                                                                    rate
                        • Source 1 — selects the source for the FFT. (See page 225 for information about
                          using g(t) as the source.)
                        • Preset — sets the frequency Span and Center to values that will cause the entire
                          available spectrum to be displayed. The maximum available frequency is half the
                          effective FFT sample rate, which is a function of the time per division setting. The
                          current FFT sample rate is displayed above the softkeys.
                        • More FFT — displays the More FFT Settings Menu.
                     2 Press the More FFT softkey to display additional FFT settings.
                             • Rectangular — good frequency resolution and amplitude accuracy, but use only
                               where there will be no leakage effects. Use on self-windowing waveforms such
                               as pseudo-random noise, impulses, sine bursts, and decaying sinusoids.
                             • Blackman Harris — window reduces time resolution compared to a
                               rectangular window, but improves the capacity to detect smaller impulses due
                               to lower secondary lobes.
                          • Span — sets the overall width of the FFT spectrum that you see on the display (left
                            to right). Divide span by 10 to calculate the number of Hertz per division. It is
                            possible to set Span above the maximum available frequency, in which case the
                            displayed spectrum will not take up the whole screen. Press the Span softkey, then
                            turn the Entry knob to set the desired frequency span of the display.
                          • Center — sets the FFT spectrum frequency represented at the center vertical grid
                            line of the display. It is possible to set the Center to values below half the span or
                            above the maximum available frequency, in which case the displayed spectrum will
                            not take up the whole screen. Press the Center softkey, then turn the Entry knob to
                            set the desired center frequency of the display.
                          • Scale — lets you set your own vertical scale factors for FFT expressed in dB/div
                            (decibels/division). Press the Scale softkey, then turn the Entry knob to rescale
                            your math function.
                          • Offset — lets you set your own offset for the FFT. The offset value is in dB and is
                            represented by the center horizontal grid line of the display. Press the Offset
                            softkey, then turn the Entry knob to change the offset of your math function.
                       3 To make cursors measurements, press the [Cursors] key and set the Source softkey
                         to Math: f(t).
                          Use the X1 and X2 cursors to measure frequency values and difference between two
                          frequency values (X). Use the Y1 and Y2 cursors to measure amplitude in dB and
                          difference in amplitude (Y).
                       4 To make other measurements, press the [Quick Meas] key and set the Source softkey
                         to Math: f(t).
                       number of points in the FFT. The actual resolution of the display will not be this fine as the
                       shape of the window will be the actual limiting factor in the FFTs ability to resolve two
                       closely space frequencies. A good way to test the ability of the FFT to resolve two closely
                       spaced frequencies is to examine the sidebands of an amplitude modulated sine wave.
                       For the best vertical accuracy on peak measurements:
                       • Make sure the probe attenuation is set correctly. The probe attenuation is set from the
                         Channel Menu if the operand is a channel.
                       • Set the source sensitivity so that the input signal is near full screen, but not clipped.
                       • Use the Flat Top window.
                       • Set the FFT sensitivity to a sensitive range, such as 2 dB/division.
                       For best frequency accuracy on peaks:
                       • Use the Hanning window.
                       • Use Cursors to place an X cursor on the frequency of interest.
                       • Adjust frequency span for better cursor placement.
                       • Return to the Cursors Menu to fine tune the X cursor.
                       For more information on the use of FFTs please refer to Agilent Application Note 243,
                       ”The Fundamentals of Signal Analysis” at
                       http://cp.literature.agilent.com/litweb/pdf/5952-8898E.pdf. Additional information
                       can be obtained from Chapter 4 of the book “Spectrum and Network Measurements” by
                       Robert A. Witte.
                     Precision measurements and math are disabled in the same way, or you can disable them
                     by pressing [Save/Recall] > Default Setup.
7
Printing and Saving
Printing the Oscilloscope’s Display 244
   Quick Print 244
   To print the oscilloscope’s display 245
   Options 245
   Palette 246
Saving Oscilloscope Data 247
   Selecting a Destination for Your Saved Data 248
   Selecting a File Name 249
   Saving Waveform Trace and Oscilloscope Setup 250
   Display Image and Waveform Data File Formats 250
   Choosing Save Settings 251
   To save a waveform and/or setup to a USB mass storage device 255
   To save a waveform and/or setup to the oscilloscope’s internal
   memory 256
   To recall waveform trace and/or oscilloscope setup 256
   File Explorer 257
                                                              243
7     Printing and Saving
                  Quick Print
                       On earlier oscilloscopes the [Print] key was labeled [Quick Print] and the oscilloscope
                       attempted to print or save as soon as the key was pressed.
                       On newer oscilloscopes the key is labeled [Print], and by default, when it is pressed the
                       Print Configuration Menu is displayed.
                       Now you can select whether pressing the key displays the Print Configuration Menu or
                       performs a Quick Print.
                       To choose Print or Quick Print functionality:
                       1 Press [Save/Recall].
                       2 Press the Quick Print softkey to set Quick Print on or off.
                       When Quick Print is enabled, a Save function is performed immediately when the [Print]
                       (or [Quick Print]) key is pressed. The save parameters must be set up in the Save/Recall
                       Menu (see page 247) before pressing [Print].
                       The oscilloscope will print the last menu visited before you pressed the [Print] key.
       NOTE            Therefore, if you have measurements (Amplitude, Frequency, etc.) showing on the display
                       before you select [Print], the measurements will be shown on the printout.
                       To print the display showing the Print Configuration Menu at the bottom, press the [Print]
                       key twice, then press the Press to Print softkey.
                 Options
                       Press the Options softkey to change the following options:
                       • Setup Information — Select to print oscilloscope setup information on your printout,
                         including vertical, horizontal, trigger, acquisition, math, and display settings.
                       • Invert Graticule Colors — The Invert Graticule Colors option can be used to reduce
                         the amount of black ink it takes to print oscilloscope images by changing the black
                         background to white. Invert Graticule Colors is the default mode.
                       • Form Feed — The Form Feed option can be selected to send a form feed command to
                         the printer after the waveform is printed and before the setup information is printed.
                         Switch Form Feed off if you want setup information printed on the same sheet of
                         paper with the waveform. This option only has an effect when the Setup Information
                         option is selected. Also, if the amount of setup information will not fit on the same
                         page with the waveform, it will be printed on a new page regardless of the Form Feed
                         setting.
                       • Landscape — Choose Landscape mode to print horizontally on the page instead of
                         vertically (portrait mode).
                  Palette
                       Press the Palette softkey to change the following options.
                       • Color — When Color printing is selected, the traces are printed in color.
                       • Grayscale — When Grayscale printing is selected, the traces are printed in shades of
                         gray rather than in color.
                       You can also save the oscilloscope’s display image using a web browser. See “Get
       NOTE            Image” on page 285 for details.
                       The File Explorer in the Save and Recall Menus can be used to save the following file
        NOTE           types: oscilloscope setup, waveform trace, display image and waveform data. It can be
                       used to recall oscilloscope setup and waveform trace files.
                       The File Explorer in the Utility Menu can be used for loading files into the oscilloscope or
                       deleting files.
                       If one or more USB mass storage devices are connected to the oscilloscope, you can
                       navigate to the USB device using the same method. The following display image shows
                       an image ready to be saved to a subfolder called “Rise time” on a USB mass storage
                       device. Push the Entry knob to save the image.
                       Overwriting a file
                       You can choose to overwrite an existing file name by browsing to that file and selecting it,
                       then choosing Press to Save.
                       4 You can press the Delete Character softkey to delete the current character and shift
                         the rest of the characters to the left.
                       5 When you select the Auto Increment option, the oscilloscope will add a numeric suffix
                         to your file name, and increment the number with each successive save. It will
                         truncate characters as necessary when the file name length is at maximum and more
                         digits are required for the numeric portion of the file name.
                       6 Press the Press to Save softkey to save the file.
                       • ASCII XY data file— This creates a file of comma-separated variables for each
                         displayed channel (including serial decode waveforms). If the oscilloscope acquisition
                         is stopped, then data records greater than 1000 points can be written. This format is
                         also suitable for spreadsheets.
                       • BIN data file — This creates a binary file with a header, and data in the form of time
                         and voltage pairs. This file is much smaller than the ASCII XY data file. If the
                         oscilloscope acquisition is stopped, then data records greater than 1000 points can be
                         written. (See also page 409.)
                       • ALB data file — This creates a file in an Agilent proprietary format which can be read
                         by Agilent logic analyzers. You can use Agilent’s B4610A Data import tool for offline
                         viewing and analysis.
                       • Mask test data file — This creates a mask file in an Agilent proprietary format which
                         can be read by Agilent InfiniiVision oscilloscopes. A mask data file includes certain
                         setup information, but not all setup information. To save all setup information
                         including the mask data file, choose “Trace and Setup” format instead. (See also
                         “Mask Test” on page 337.)
                       Setup Info
                       Select Setup Info if you want the oscilloscope setup information to be included.
                       Oscilloscope setup information includes vertical, horizontal, trigger, acquisition, math,
                       and display settings. The setup information will be sent to a separate file with a TXT
                       extension.
                       The Invert Graticule Colors option can be used to reduce the amount of black ink it takes
                       to print oscilloscope images by changing the black background to white. Invert Graticule
                       Colors is the default mode. Invert Graticule Colors is available when the BMP and PNG
                       formats are selected.
                       Palette
                       Press the Palette softkey to change the following options. The Palette options are
                       available when the BMP and PNG formats are selected.
                       • Color — When Color printing is selected, the traces are printed in color. Color printing
                         is not applicable to CSV format.
                       • Grayscale — When Grayscale printing is selected, the traces are printed in shades of
                         gray rather than in color. Grayscale printing is not applicable to CSV format.
                       Length Control
                       The Length softkey is available when the CSV, ASCII XY, BIN, or ALB formats are
                       selected. Length sets the number of data points that will be output to the file. Length can
                       be set to 100, 250, 500, or 1000 when the acquisition is running, or more when the
                       acquisition is stopped. (CSV length is limited to 1000 points.) It is important to save
                       enough points to represent the captured data accurately. Press the [Single] key to obtain
                       the maximum record length.
                       The maximum available record length depends on whether one or two channels are
                       active (turned on) per channel pair (channel pairs are channels 1 and 2 and channels 3
                       and 4), whether digital channels are on or off, and the horizontal timebase setting (how
                       much data is shown on screen).
                       Only displayed data points are output (except when you are using segmented memory
                       and the Save Segment control is set to All segments). Therefore, adjust the Horizontal
                       controls to display the data you want to save.
                       The length control has no effect on the amount of Serial Decode data saved; all serial
                       decode data will be output.
                       The Length control will perform a “1 of n” decimation of the data when necessary. For
                       example: if the Length is set to 1000, and you are displaying a record that is 5000 data
                       points in length, four of each five data points will be decimated, creating an output file
                       1000 data points in length. However, if serial decode is on, no decimation will occur.
                       For more information see “Minimum and Maximum Values in CSV Files” on page 417.
                       Save Segment
                       When using the optional segmented memory feature you can choose to save either the
                       currently displayed segment or all acquired segments in memory. (See also page 274.)
                       Press [Save/Recall]&Save&Format (choose CSV, ASCII XY, or BIN)&
                       Settings&Save Seg to make your selection. Set the Length control appropriately (see
                       Length Control, above).
RxRS232 Rx Ch2 D1
                       When saving display image files, the oscilloscope will save an image using the last menu
        NOTE           you visited before pressing the [Print] key. Therefore, if you have measurements
                       (Amplitude, Frequency, etc.) showing on the display before you select Save, the
                       measurements will be shown in the display image file.
                       To save an image of the display showing the Save/Recall Menu at the bottom, press the
                       [Save/Recall] key twice, then press the Press to Save softkey.
                 File Explorer
                       The File Explorer lets you navigate the file system. Using the File Explorer you can save,
                       recall, load, and delete files.
                       File system    The user-accessible file system consists of ten internal memory locations,
                       plus external USB mass storage devices you connect to the USB device ports on the
                       oscilloscope.
                       Internal memory The ten internal nonvolatile memory locations are located on the
                       oscilloscope at C:\setups and are named intern_0 through intern_9. You can save
                       waveform trace and oscilloscope setup in each of the ten internal memory locations, but
                       you cannot save display image files or waveform data files in the oscilloscope’s internal
                       memory. The latter must be stored on USB mass storage devices.
                       USB mass storage devices Most USB mass storage devices are compatible with the
                       oscilloscope. However, certain devices may be incompatible, and may not be able to be
                       read or written to.
                       When the USB mass storage device is connected to the oscilloscope’s front or rear USB
                       host port, a small four-color circle icon may be displayed briefly as the USB device is read.
                       You do not need to “eject” the USB mass storage device before removing it. Simply
                       ensure that any file operation you’ve initiated is done, and remove the USB drive from the
                       oscilloscope’s host port.
                       Do not connect USB devices that identify themselves as hardware type “CD” because
                       these devices are not compatible with the InfiniiVision 7000 Series oscilloscopes.
                       If two USB mass storage devices are connected to the oscilloscope, the first one is
                       designated “USB0” and the second one is designated “USB5,” not “USB1.” This
                       numbering method is normal; it is inherent in the USB driver.
                       A file that has been deleted from the USB mass storage device cannot be recovered by
                       the oscilloscope.
                       System software System software files can be loaded into the oscilloscope. These
                       software updatescan be downloaded from:
                       • www.agilent.com/find/5000sw
                       • www.agilent.com/find/6000sw
                       • www.agilent.com/find/7000sw
                       See “Software and Firmware Updates” on page 400 for more information.
           8
           Acquisition Modes
           To start and stop an acquisition 262
           To make a single acquisition 264
           Acquisition Modes 265
              Normal Mode 265
              Peak Detect Mode 266
              High Resolution Mode 266
              Averaging Mode 267
              Realtime Sampling Option 269
           Segmented Memory 271
                                                                    261
8     Acquisition Modes
                      Auto Single
                      Auto-trigger will generate a trigger for you if one is not found in the predetermined time
                      (about 40 ms) after you press [Single]. If you want to make a single acquisition and you
                      are not particularly interested in triggering the acquisition (for example, if you are probing
                      a DC level), set the trigger mode to Auto (see page 84) and press the [Single] key. If a
                      trigger condition occurs, it will be used; if a trigger doesn’t happen, an untriggered
                      acquisition will be taken.
Acquisition Modes
                       The InfiniiVision oscilloscopes have the following acquisition modes:
                       • Normal — for most waveforms (with normal decimating at slower sweep speeds, no
                         averaging).
                       • Peak Detect — for displaying narrow pulses that occur infrequently (at slower sweep
                         speeds).
                       • Averaging — for reducing noise and increasing resolution (at all sweep speeds,
                         without bandwidth or rise time degradation).
                       • High Resolution — for reducing random noise (at slower sweep speeds).
                       Realtime sampling (where the oscilloscope produces the waveform display from samples
                       collected during one trigger event) can be turned off or on in the Normal, Peak Detect,
                       and High Resolution modes.
                       (For information on the XY horizontal mode see “XY Horizontal Mode” on page 220.)
                 Normal Mode
                       In Normal mode at slower sweep speeds, extra samples are decimated (in other words,
                       some are thrown away). This mode yields the best display for most waveforms.
5 us/div 2 us/div 11
 20 us/div  10 us/div 12
                 Averaging Mode
                       The Averaging mode lets you average multiple acquisitions together to reduce noise and
                       increase vertical resolution (at all sweep speeds). Averaging requires a stable trigger.
                       The number of averages can be set from 2 to 65536 in powers-of-2 increments.
                       A higher number of averages reduces noise more and increases vertical resolution.
2 8
4 9
16 10
64 11
 256 12
                       The higher the number of averages, the slower the displayed waveform responds to
                       waveform changes. You must compromise between how quickly the waveform responds
                       to changes and how much you want to reduce the displayed noise on the signal.
                      To see the sample rate, press the [Acquire] key on the front panel. The sample rate is
                      displayed in the line just above the softkeys.
                                                                                                 Sample
                                                                                                 rate
Segmented Memory
                       You can purchase the oscilloscope with the segmented memory option factory-installed
                       (Option SGM) or you can easily install it yourself (order model number N5454A
                       “Segmented Memory”).
                       When capturing multiple infrequent trigger events it is advantageous to divide the
                       oscilloscope’s memory into segments. This lets you capture signal activity without
                       capturing long periods of signal “dead time.”
                       Each segment is complete with all analog channel, digital channel (on MSO models), and
                       serial decode data.
                       When using Segmented memory, use the Analyze Segments feature (see
                       “Measurements, statistics, and infinite persistence” on page 273) to show infinite
                       persistence across all acquired segments. See also “Infinite Persistence” on page 182
                       for details.
                      If the signal you are measuring has more than about 1 s of “dead time” (inactivity),
                      consider selecting Normal trigger mode to prevent AutoTriggering. Press the
                      [Mode/Coupling] key, then press the Mode softkey and select Normal.
Progress indicator
                                                                                     Sample
                                                                                     rate
                      Navigating segments
                      Press the Current Seg softkey and turn the Entry knob to display to the desired segment
                      along with a time tag indicating the time from the first trigger event.
                      You can also press the Knob softkey to toggle between Current Segment and Time. When
                      Current Segment is selected, the Horizontal position control knob moves the display to
                      the desired segment. When Time is selected, the Horizontal position control knob moves
                      the display horizontally just as it does when Segmented Memory mode is not selected.
                      The Knob softkey also appears in the Horizontal Menu so you can make the same
                      adjustments there.
                       Re-arm time
                       After each segment fills, the oscilloscope re-arms and is ready to trigger in about 8 s.
                       Remember though, for example: if the horizontal time per division control is set to
                       5 s/div, and the Time Reference is set to Center, it will take at least 50 s to fill all ten
                       divisions and re-arm. (That’s 25 s to capture pre-trigger data and 25 s to capture
                       post-trigger data.)
      Displayed segment
      number and number of
      segments acquired
9
Web Interface
Setting up the Oscilloscope’s LAN Connection 276
Accessing the Web Interface 280
Browser Web Control 281
Get Image 285
Identification Function 286
Instrument Utilities 287
Setting a Password 288
                                                                      275
9     Web Interface
                      Microsoft Internet Explorer is the recommended Web browser for communication and
                      control of the oscilloscope. Other Web browsers may work but are not guaranteed to
                      work with the oscilloscope. The Web browser must be Java-enabled with Sun
                      Microsystems™ Java Plug-in.
                      Before you can use the web interface, you must place the oscilloscope on the network
                      and set up its LAN connection.
                      Once you connect the oscilloscope to a LAN it is a good practice to limit access to the
        NOTE          oscilloscope by setting a password. By default, the oscilloscope is not password
                      protected. See page 288 to set a password.
                      Any time you modify the oscilloscope’s hostname it will break the connection between
        NOTE          the oscilloscope and the LAN. You will need to re-establish communication to the
                      oscilloscope using the new hostname.
                      4 Open a web browser and enter the oscilloscope’s hostname in the address field. The
                        oscilloscope’s Web page will be displayed.
                      Hostname
                      If you do not know the oscilloscope’s hostname you can reset the hostname as follows:
                      1 Press and release the Reset button (see page 354) on the front panel of the
                        6000L Series oscilloscope. This will reset the hostname to that which is printed on the
                        label near the oscilloscope’s power cord connection. The following LAN parameters
                        will be reset:
                         • Hostname
                         • IP address
                         • Domain name
                         • VISA address
                         • Password (see page 288.)
                       After the LAN connection is configured, you can find the oscilloscope’s IP address and
                       use it to access the instrument’s Web interface:
                       • On an oscilloscope that has a display, you can find the IP address assigned to the
                         oscilloscope in the [Utility]&I/O screen.
                       • On a 6000L Series oscilloscope without a display, you can use the Agilent Connection
                         Expert to find the oscilloscope’s IP address.
                       If Java is not installed on your PC, you will be prompted to install the Sun Microsystems
       NOTE            Java Plug-in. This plug-in must be installed on the controlling PC for the web interface’s
                       Remote Front Panel or Remote Programming operations.
                           Main Menu
                           Function Keys
                           Hint appears
                           when you
                           roll over with
                           mouse
                           Softkeys
                           (Left-click to
                           select, Right-
                           click for
                           Quick Help
                 Remote Programming
                       To send remote programming commands to the oscilloscope via the SCPI Commands
                       applet window:
                       1 Access the oscilloscope’s web interface (see “Accessing the Web Interface” on
                         page 280).
                       2 When the oscilloscope’s web interface is displayed, select Browser Web Control,
                         then select Remote Programming.
                       3 On the Remote Programming page, click SCPI Commands....
                          The SCPI Commands applet window appears.
                         If you get the “Unable to connect to the instrument. You may need to set the
                         instrument’s controller type to LAN.” message:
                         a On the oscilloscope’s front panel, press [Utility]&I/O.
                         b In the I/O Menu, press Control.
                         c Use the Entry knob to point to LAN; then, press Control again to enable the LAN
                           controller type.
                            Multiple controller types may be enabled.
Get Image
                       To save (or print) the oscilloscope’s display from the web interface:
                       1 Access the oscilloscope’s web interface (see “Accessing the Web Interface” on
                         page 280).
                       2 When the oscilloscope’s web interface is displayed, select the Get Image tab from the
                         left side of the Welcome screen. After a delay of several seconds, the oscilloscope’s
                         screen image will be displayed.
                       3 Right-click on the image and select “Save Picture As...” (or “Print Picture...”).
                       4 Select a storage location for the image file and click Save.
Identification Function
                      The Identification web interface feature is useful when trying to locate a specific
                      instrument in a rack of equipment.
                      1 Access the oscilloscope’s web interface (see “Accessing the Web Interface” on
                        page 280).
                      2 When the oscilloscope’s web interface Welcome Page is displayed, select the
                        Identification on radio button.
                         An “Identify” message is displayed on the oscilloscope; you can either select
                         Identification off or press the OK softkey on the oscilloscope to continue.
                         On 6000L Series oscilloscopes, the LAN indicator on the front of the instrument will
                         flash green to identify the oscilloscope until you select Identification off.
Identification Option
Instrument Utilities
                       The Instrument Utilities page of the web interface lets you:
                       • View installed options.
                       • View firmware versions.
                       • Install firmware upgrade files.
                       • View calibration status.
                       You can select these cababilities via a drop-down menu.
Setting a Password
                      Whenever you connect the oscilloscope to a LAN, it is good practice to set a password.
                      The password prevents someone from remotely accessing the oscilloscope via a Web
                      browser and changing parameters. Remote users can still view the Welcome screen,
                      view network status, etc., but they can’t operate the instrument or change its setup
                      without the password.
                      To set a password:
                      1 Access the oscilloscope’s web interface (see “Accessing the Web Interface” on
                        page 280).
                      2 When the oscilloscope’s web interface is displayed, select the Configure Network tab
                        from the instrument’s Welcome page.
                      3 Click the Modify Configuration button.
Step 2
Step 1
Step 3
                       When accessing the password protected oscilloscope, the user name is the IP address of
                       the oscilloscope.
10
Serial Decode/Lister
Serial Decode 292
Lister 293
CAN Serial Decode 295
FlexRay Serial Decode 302
I2C Serial Decode 306
I2S Serial Decode 310
LIN Serial Decode 314
MIL-STD 1553 Serial Decode 320
SPI Serial Decode 323
UART/RS232 Serial Decode 329
                                                                      291
10 Serial Decode/Lister
Serial Decode
                    Agilent’s hardware accelerated serial decode options can be installed when the
                    oscilloscope is manufactured, or added later. The following serial decode licenses are
                    available:
                    • The N5424A (Option AMS) license provides the ability to decode CAN (Controller Area
                      Network) and LIN (Local Interconnect Network) serial buses. Requires a 4-channel
                      oscilloscope.
                          Adding the Automotive Serial (AMS) serial decode license will add more CAN and LIN
                          trigger types to the trigger menu. See page 98 or page 128 for details.
                    • The N5432A (Option FRS) license provides the ability to trigger on and decode FlexRay
                      automotive serial buses. See page 108 for FlexRay triggering information. Requires a
                      4-channel MSO oscilloscope.
                    • The N5423A (Option LSS) license provides the ability to decode I2C (Inter-IC) and SPI
                      (Serial Peripheral Interface) serial buses. Requires a 4-channel oscilloscope.
                    • The N5468A (Option SND) license provides the ability to decode I2S (Inter-IC Sound or
                      Integrated Interchip Sound) serial buses. Requires a 4-channel oscilloscope.
                    • The N5457A (Option 232) license provides the ability to decode many UART (Universal
                      Asynchronous Receiver/Transmitter) protocols including RS232 (Recommended
                      Standard 232). Requires a 4-channel oscilloscope.
                    To determine whether these licenses are installed on your oscilloscope press the [Utility]
                    key, then press the Options softkey, the Licenses softkey, and the Show license
                    information softkey.
                    To order serial decode licenses, go to www.agilent.com and search for the product
                    number (for example, N5432A) or contact your local Agilent Technologies representative
                    (see www.agilent.com/find/contactus).
Lister
                       Lister is a powerful tool for investigating protocol failures. You can use Lister to view large
                       amounts of packet level serial data in a tabular format, including time tags and specific
                       decoded values. After pressing the [Single] key, you can rotate the Entry knob to select
                       an event and press the Zoom to Selection softkey to jump to the event.
                       To use the Lister:
                       1 Set up trigger and decode on the serial data signals to be analyzed. Instructions for
                         triggering on serial protocols are in “Triggering” on page 81. Instructions for setting
                         up serial decode are in this chapter (starting with “CAN Serial Decode” on page 295).
                       2 Press [Acquire]&Serial Decode&Lister.
                       3 Press Lister Display to enable the display.
                          Before you can select a row or navigate through the Lister data, oscilloscope
                          acquisitions must be stopped.
                    4 Press the [Single] key (in the Run Control group on the front panel) to stop the
                      acquisition.
                          Pressing [Single] instead of [Stop] fills the maximum memory depth.
                          When zoomed out and viewing a large number of packets, the Lister may not be able
                          to display information for all packets. However, when you press the [Single] key the
                          Lister will contain all on-screen serial decode information.
                    5 Press the Scroll Lister softkey and use the Entry knob to scroll through the data.
                          Time tags in the Time column indicate the event time relative to the trigger point. The
                          time tags of events that are shown in the waveform display area are displayed with a
                          dark background.
                    6 Press the Zoom to Selection softkey (or push the Entry knob on 7000 Series
                      oscilloscopes) to center the waveform display at the time associated with the selected
                      Lister row and automatically set the horizontal scale setting.
                    7 Press the Undo Zoom softkey to return to the horizontal scale and delay settings
                      before the last Zoom to Selection.
                    8 Press the Options softkey to open the Lister Options Menu. In this menu, you can:
                          • Enable or disable the Track Time option. When enabled, as you select different
                            Lister rows (using the Entry knob while acquisitions are stopped), the horizontal
                            delay changes to the Time of the selected row.
                    5 Press the Source softkey, and use the Entry knob to select the channel for the CAN
                      signal.
                          If you have connected one of the oscilloscope’s analog source channels to the CAN
                          signal: Adjust the trigger level for the selected analog channel to the middle of the
                          waveform by turning the Trigger Level knob.
                          If you have connected one of the oscilloscope’s digital source channels to the CAN
                          signal (this applies to MSO model oscilloscopes only): Press the [D15-D0] key and
                          select Thresholds to set an appropriate threshold level for digital channels.
                          The value of the trigger level or digital threshold is displayed in the upper-right corner
                          of the display.
                       6 Repeatedly press and release the Baud softkey to set the CAN signal baud rate to
                         match your CAN bus signal.
                          The CAN baud rate can be set to:
                          10 kb/s      50 kb/s         100 kb/s       500 kb/s
                          20 kb/s      62.5 kb/s       125 kb/s       800 kb/s
                          33.3 kb/s    83.3 kb/s       250 kb/s       1 Mb/s
                          User Defined
                                                
                                                  
                                   	
 
                                                     
                       Changing the Source settings in the Serial Decode setup also changes them in the Trigger
       NOTE            setup.
                    10 If the decode line does not appear on the display, press the Decode softkey to turn it
                       on.
                    11 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and decode data.
                    If the setup does not produce a stable trigger, the CAN signal may be slow enough that
          NOTE      the oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode
                    softkey to set the trigger mode from Auto to Normal.
You can use the Zoom horizontal sweep mode for easier navigation of the decoded data.
Active Error
Frame (red)
Missing
Acknowledge
(red)
CRC
(blue)
Data
(white)
Data Length
Code (blue)
Frame ID
(yellow)
                       • Data length code (DLC) appears in blue for data frames and green for remote frames.
                       • Data bytes appear in hex digits in white for data frames.
                       • Cyclic redundancy check (CRC) appears in hex digits in blue when valid, or in red to
                         indicate that the oscilloscope’s hardware decode calculated a CRC that is different
                         from the incoming CRC data stream.
                       • Angled waveforms show an active bus (inside a packet/frame).
                       • Mid-level blue lines show an idle bus.
                       • Decoded text is truncated at the end of the associated frame when there is insufficient
                         space within frame boundaries.
                       • Red dots in the decode line indicate that there is data that is not being displayed.
                         Scroll or expand the horizontal scale to view the information.
                       • Aliased bus values (undersampled or indeterminate) are drawn in red.
                       • Unknown bus values (undefined or error conditions) are drawn in red with a “?” label.
                       • Flagged error frames are drawn in red with an “ERR” label.
                 CAN Totalizer
                       The CAN totalizer provides a direct measure of bus quality and efficiency. The CAN
                       totalizer measures total CAN frames, flagged error frames, overload frames, and bus
                       utilization.
                       The totalizer is always running (counting frames and calculating percentages) and is
                       displayed whenever CAN decode is displayed. The totalizer counts even when the
                       oscilloscope is stopped (not acquiring data). Pressing the [Run/Stop] key does not affect
                       the totalizer. When an overflow condition occurs, the counter displays OVERFLOW. The
                       counters can be reset to zero by pressing the Reset CAN Counters softkey.
                       Types of Frames
                       • Active error frames are CAN frames in which a CAN node recognizes an error
                         condition during a data or remote frame and issues an active error flag.
                       • A partial frame occurs when the oscilloscope detects any error condition during a
                         frame that is not followed by an active error flag. Partial frames are not counted.
                     Counters
                     • The FRAMES counter gives the total number of completed remote, data, overload, and
                       active error frames.
                     • The OVLD counter gives the total number of completed overload frames and their
                       percentage of the total number of frames.
                     • The ERR counter gives the total number of completed active error frames and their
                       percentage of the total number of frames.
                     • The UTIL (bus load) indicator measures the percentage of time the bus is active. The
                       calculation is done on 330 ms periods, approximately every 400 ms.
                     Example: If a data frame contains an active error flag, both the FRAMES counter and the
                     ERR counter will be incremented. If a data frame contains an error that is not an active
                     error it is considered a partial frame and no counters are incremented.
                       In addition to the standard Time column, the CAN Lister contains these columns:
                       • ID — frame ID.
                       • Type — frame type (RMT remote frame, Data, etc.).
                       • DLC — data length code.
                       • Data — data bytes.
                       • CRC — cyclic redundancy check.
                       • Errors — highlighted in red. Errors can be Acknowledge (Ack, A), Form (Fo), or Frame
                         (Fr). Different kinds of errors can be combined like “Fo,Fr” in the above example.
                    While decoding FlexRay data, you can use FlexRay trigger or another trigger type, such as
      NOTE          duration, I2C, Nth edge burst, sequence, SPI, or USB.
                    When the oscilloscope is in FlexRay trigger mode, only FlexRay decode is available.
Payload
data
Cycle
Number
Header CRC
(blue)
Payload Length
(green)
Frame ID
(yellow)
Frame Type
(blue)
               FlexRay Totalizer
                    The FlexRay totalizer consists of counters that provide a direct measure of bus quality and
                    efficiency. The totalizer appears on screen whenever FlexRay Decode is ON in the Serial
                    Decode Menu.
                       In addition to the standard Time column, the FlexRay Lister contains these columns:
                       • FID — frame ID.
                       • Len — payload length.
                       • HCRC — header CRC.
                       • CYC — cycle number.
                       • Data.
                       • FCRC — frame CRC.
                       • Frames with errors are highlighted in red.
                    Because of oscilloscope hardware resource limitations, you cannot decode I2C data while
      NOTE          LIN triggering is selected.
                    If you have already set up an I2C trigger, the signals have already been specified, and you
      NOTE          have set the analog trigger level for each analog channel, you can continue on with
                    step 3.
                    Changing I2C signals in the Serial Decode setup also changes them in the Trigger setup.
                          b Press the SCL softkey, and use the Entry knob to select the channel probing the
                            clock signal.
                          c Press the SDA softkey, and use the Entry knob to select the channel probing the
                            data signal.
                       3 Set the trigger levels for the SCL and SDA signals to the middle of the signals:
                          • If your I2C signals are connected to analog channels, press the SCL softkey and
                            rotate the Trigger Level knob, then press the SDA softkey and rotate the Trigger
                            Level knob.
                          • If your I2C signals are connected to digital channels, press the [D15-D0] key and
                            the Thresholds softkey to access the threshold level setting softkeys and set the
                            thresholds to the approximate middle of the signals.
                       4 If the decode line does not appear on the display, press the up-arrow softkey to return
                         to the previous menu, then press the Decode softkey to turn it on.
                       5 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and decode data.
                       If the setup does not produce a stable trigger, the I2C signal may be slow enough that the
       NOTE            oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode
                       softkey to set the trigger mode from Auto to Normal.
You can use the Zoom horizontal sweep mode for easier navigation of the acquired data.
                       In addition to the standard Time column, the I2C Lister contains these columns:
                       • Restart — indicated with an “X”.
                       • Address — colored blue for writes, yellow for reads.
                       • Data — data bytes.
                       • Missing Ack — highlighted in red, error if it has an “X”.
                    When triggering is set to CAN, LIN, Flexray, or UART/RS232 mode, decoding in I2S is not
      NOTE          allowed. When triggering in I2C or SPI modes, or any non-serial trigger mode, I2S is
                    available for decoding. When I2S is selected as the trigger mode, only I2S decode is
                    available.
                          a Press the Signals softkey to open the I2S Signals Menu. This menu is where you
                            select the oscilloscope channels that probe the serial clock, Word Select, and
                            serial data lines. It is the same menu you use when setting up I2S triggers. For more
                            information about the options in this menu, see step 6 on page 122.
                          b Press the Bus Config softkey to open the I2S Bus Configuration Menu. This menu
                            is where you identify the device under test’s I2S bus setup. It is the same menu you
                            use when setting up I2S triggers. For more information about the options in this
                            menu, see step 11 on page 123.
                          c Press the Base softkey to select the number base in which to display decoded data.
                       3 If the decode waveform does not appear on the display, press the up-arrow softkey to
                         return to the previous menu, then press the Decode softkey to turn it on.
                       4 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and decode data.
                       If the setup does not produce a stable trigger, the I2S signal may be slow enough that the
       NOTE            oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode
                       softkey to set the trigger mode from Auto to Normal.
You can use the Zoom horizontal sweep mode for easier navigation of the acquired data.
                       When the receiver word size is greater than the transmit word size, the decoder fills the
       NOTE            least significant bits with zeros and the decoded value does not match the trigger value.
                       In addition to the standard Time column, the I2S Lister contains these columns:
                       • Left Channel — displays the left channel data.
                       • Right Channel — displays the right channel data.
                       • Errors — highlighted in red and marked with an “X”.
           Decoded LIN
           data
                            b Press the Source softkey, and use the Entry knob to select the channel probing the
                              LIN signal.
                            c Press the Baud softkey and rotate the Entry knob to set the LIN signal baud rate to
                              match your LIN bus signal.
                               The LIN baud rate can be set to 2400 b/s, 9600 b/s, 10.4 kb/s, 19.2 kb/s, 115.2
                               kb/s, 625 kb/s, or User Defined. The default baud rate is 2400 b/s.
                             If the desired baud rate is not shown in the list, select User Defined. You can set
                             the LIN baud rate from 2.4 kb/s to 625 kb/s in increments of 100 b/s. Press the
                             User Baud softkey and rotate the Entry knob to make your selection.
                          d Repeatedly press and release the Smpl Pt softkey to select the sample point. This
                            controls the point within the bit’s time where the bit value is captured.
                                                   
                                                     
                                      	
 
                                                        
                          e Repeatedly press and release the Standard softkey to select the LIN standard (LIN
                            1.3 or LIN 2.0) of the measured signal. For LIN 1.2 signals, use the LIN 1.3 setting.
                            The LIN 1.3 setting assumes the signal follows the “Table of Valid ID Values” as
                            shown in section A.2 of the LIN Specification dated December 12, 2002. If your
                            signal does not comply with the table, use the LIN 2.0 setting.
                          f Repeatedly press and release the Sync Break softkey to define the minimum
                            number of clocks that define a Sync Break.
                    4 Set the trigger or threshold levels to the middle of the LIN signal.
                          If you have connected one of the oscilloscope’s analog source channels to the LIN
                          signal: Adjust the trigger level for the selected analog channel to the middle of the
                          waveform by turning the Trigger Level knob.
                          If you have connected one of the oscilloscope’s digital source channels to the LIN
                          signal (this applies to MSO model oscilloscopes only): Press the [D15-D0] key and
                          select Thresholds to set an appropriate threshold level for digital channels.
                          The value of the trigger level or digital threshold is displayed in the upper-right corner
                          of the display.
                    Changing the Trigger settings in the Serial Decode setup also changes them in the Trigger
      NOTE          setup.
                          g If the decode line does not appear on the display, press the up-arrow softkey to
                            return to the previous menu, then press the Decode softkey.
5 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and decode data.
                       If the setup does not produce a stable trigger, the LIN signal may be slow enough that the
       NOTE            oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode
                       softkey to set the trigger mode from Auto to Normal.
You can use the Zoom horizontal sweep mode for easier navigation of the decoded data.
                    • For LIN 1.3, the checksum appears in blue if correct, or red if incorrect. The checksum
                      always appears in white for LIN 2.0.
                    • Decoded text is truncated at the end of the associated frame when there is insufficient
                      space within frame boundaries.
                    • Red dots in the decode line indicate that there is data that is not being displayed.
                      Scroll or expand the horizontal scale to view the information.
                    • Unknown bus values (undefined or error conditions) are drawn in red.
                    • If there is an error in the synch field, SYNC will appear in red.
                    • If the header exceeds the length specified in the standard, THM will appear red.
                    • If the total frame count exceeds the length specified in the standard, TFM will appear
                      red (LIN 1.3 only).
                    • For LIN 1.3 a wakeup signal is indicated by WAKE in blue. If the wakeup signal is not
                      followed by a valid wakeup delimiter a wakeup error is detected and displayed as WUP
                      in red.
                       In addition to the standard Time column, the LIN Lister contains these columns:
                       • ID — frame ID.
                       • Data — (LIN 1.3 only) data bytes.
                       • Checksum — (LIN 1.3 only).
                       • Data and Checksum — (LIN 2.0 only).
                       • Errors — highlighted in red.
                    Before you can select serial decode of MIL-STD 1553 signals, the trigger type must be set
      NOTE          to MIL-STD-1553 or one of the non-serial triggering modes. See “MIL-STD 1553
                    Triggering” on page 133.
                    When MIL-STD-1553 is selected as the trigger mode, only MIL-STD-1553 decode is
                    available.
                    2 Press Settings.
                    3 In the MIL-STD-1553 Settings Menu, use the Base softkey to select between
                      hexadecimal and binary display of the decoded data.
                          The base setting is used for the display of the remote terminal address and the data, in
                          both the decode line and in the Lister.
                    You can use the Zoom horizontal sweep mode for easier navigation of the acquired data.
                    In addition to the standard Time column, the MIL-STD 1553 Lister contains these
                    columns:
                    • RTA – displays the Remote Terminal Address for Command/Status words, nothing for
                      Data words.
                    • Word Type – “Cmd/Status” for Command/Status words, “Data” for Data words. For
                      Command/Status words the background color is green to match the decode text color.
                    • Data – the 11 bits after the RTA for Command/Status words, or the 16 bits of a Data
                      word.
                    • Errors – “Sync”, “Parity”, or “Manchester” errors as appropriate. The background
                      color is red to indicate an error.
                    For more information on using the Lister display, see “Lister” on page 293.
                       Because of oscilloscope hardware resource limitations, you cannot decode SPI data
       NOTE            while CAN or LIN triggering is selected.
                    If you have already set up an SPI trigger, the signals and their trigger or threshold levels
      NOTE          have already been specified, and you can continue on with step 3.
                    Changing SPI signals in the Serial Decode setup also changes them in the Trigger setup.
 Graphic showing
 SPI signals setup
                             If you have connected one of the oscilloscope’s digital source channels to the
                             Clock signal (this applies to MSO model oscilloscopes only): Press the [D15-D0]
                             key and select Thresholds to set an appropriate threshold level for digital channels.
                          e Press the rising or falling edge softkey to specify the edge that the clock occurs on.
                          f Press the Data softkey, and use the Entry knob to select the channel probing the
                            data signal.
                          g If you have connected one of the oscilloscope’s analog source channels to the Data
                            signal: Adjust the trigger level for the selected analog channel to the middle of the
                            waveform by turning the Trigger Level knob.
                             If you have connected one of the oscilloscope’s digital source channels to the Data
                             signal (this applies to MSO model oscilloscopes only): Press the [D15-D0] key and
                             select Thresholds to set an appropriate threshold level for digital channels.
                          h Repeatedly press and release the Frame by softkey to select the method used to
                            identify frames:
                             • ~CS - Not Chip Select — an active low chip select signal.
                             • CS - Chip Select — an active high chip select signal.
                             • Clock Timeout — clock idle for a period of time.
                          i If you chose framing by chip select or not chip select, press the CS or ~CS softkey,
                            and use the Entry knob to select the channel probing the chip select signal.
                             If you have connected one of the oscilloscope’s analog source channels to the Chip
                             Select signal: Adjust the trigger level for the selected analog channel to the middle
                             of the waveform by turning the Trigger Level knob.
                             If you have connected one of the oscilloscope’s digital source channels to the Chip
                             Select signal (this applies to MSO model oscilloscopes only): Press the [D15-D0]
                             key and select Thresholds to set an appropriate threshold level for digital channels.
                          j If you chose framing by clock timeout, press the Timeout softkey and use the Entry
                            knob to specify the idle time.
                    3 Press the up-arrow softkey to return to the previous menu. Press the Word Size
                      softkey, and use the Entry knob to select the number of bits in a word.
                    4 If the decode line does not appear on the display, press the Decode softkey to turn it
                      on.
                    5 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and decode data.
                       If the setup does not produce a stable trigger, the SPI signal may be slow enough that the
       NOTE            oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode
                       softkey to set the trigger mode from Auto to Normal.
You can use the Zoom horizontal sweep mode for easier navigation of the acquired data.
                    • Red dots in the decode line indicate that there is data that is not being displayed.
                      Scroll or expand the horizontal scale to view the information.
                    • Aliased bus values (undersampled or indeterminate) are drawn in red.
                    • Unknown bus values (undefined or error conditions) are drawn in red.
                    In addition to the standard Time column, the SPI Lister contains these columns:
                    • Data — data bytes (MISO or MOSI).
                    Changing UART/RS232 signals in the Serial Decode setup also changes them in the
       NOTE         Trigger setup.
      Decoded
      data
                          b Press the Signals softkey, and select the oscilloscope channels that you have
                            connected to the Rx and Tx signals. Set the trigger levels. See “UART/RS232
                            Trigger” on page 166 for detailed instructions.
                       3 Press the “Return to previous menu” (up arrow) softkey.
                       Changing the Bus Configuration settings in the Serial Decode setup also changes them in
       NOTE            the Trigger setup.
                       In the decode display the most significant bit is always displayed on the left regardless of
       NOTE            how Bit Order is set.
                    8 Make sure the trigger or threshold levels are set to the middle of the signals:
                          • For analog channels, turn the Trigger Level knob.
                          • For digital channels, press the [D15-D0] key and the Thresholds softkey to access
                            the threshold level setting softkeys.
                    9 If the decode line does not appear on the display, press the up arrow softkey to return
                      to the previous menu, then press the Decode softkey to turn on serial decode.
                    10 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and decode data.
                    If the setup does not produce a stable trigger, the UART/RS232 signal may be slow
      NOTE          enough that the oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then
                    press the Mode softkey to set the trigger mode from Auto to Normal.
You can use the Zoom horizontal sweep mode for easier navigation of the acquired data.
                    • A parity error will cause the associated data word to be shown in red, which includes
                      the 5-8 data bits and the optional 9th bit.
               UART/RS232 Totalizer
                    The UART/RS232 totalizer consists of counters that provide a direct measure of bus
                    quality and efficiency. The totalizer appears on screen whenever UART/RS232 Decode is
                    ON in the Serial Decode Menu.
                    The totalizer is running, counting frames and calculating the percentage of error frames,
                    even when the oscilloscope is stopped (not acquiring data).
                    The ERR (error) counter is a count of Rx and Tx frames with parity errors. The TX FRAMES
                    and RX FRAMES counts include both normal frames and frames with parity errors. When
                    an overflow condition occurs, the counter displays OVERFLOW.
                    The counters can be reset to zero by pressing the Reset UART Counters softkey.
                       In addition to the standard Time column, the UART/RS232 Lister contains these columns:
                       • Rx — receive data.
                       • Tx — transmit data.
                       • Errors — highlighted in red, Parity Error or Unknown Error.
11
Mask Test
To create a mask from a “golden” waveform (Automask) 338
Setup Options 340
Mask Statistics 343
To manually modify a mask file 345
Building a Mask File 348
                                                                    337
11 Mask Test
                                                                                             Mask test
               4 Press the Automask softkey.
Automask
               5 Press the Source softkey and ensure the analog channel from which to create the
                 mask is selected.
               6 Adjust the mask’s horizontal tolerance (± Y) and vertical tolerance (± X). These are
                 adjustable in graticule divisions or in absolute units (volts or seconds), selectable
                 using the Units softkey.
                       7 Press the Create Mask softkey. The mask is created and testing begins. Whenever the
                         Create Mask softkey is pressed the old mask is erased and a new mask is created.
                       8 To clear the mask and switch off mask testing, press the up arrow softkey to return to
                         the Mask Test Menu, then press the Clear Mask softkey.
                       If infinite persistence display mode (see “Infinite Persistence” on page 182) is “on”
                       when mask test is enabled, it stays on. If infinite persistence is “off” when mask test is
                       enabled, it is switched on when mask test is switched on, then infinite persistence is
                       switched off when mask test is switched off.
Setup Options
               Run Until
                   The Run Until softkey lets you specify a condition on which to terminate testing.
                   Forever The oscilloscope runs continuously. However, if an error occurs the action
                   specified using the On Error softkey will occur.
                   Minimum # of Tests Choose this option and then use the # of Tests softkey to select
                   the number of times the oscilloscope will trigger, display the waveform(s), and compare
                   them to the mask. The oscilloscope will stop after the specified number of tests have
                   been completed. The specified minimum number of tests may be exceeded. If an error
                   occurs the action specified using the On Error softkey will occur. The actual number of
                   tests completed is displayed above the softkeys.
                   Minimum Time Choose this option and then use the Test Time softkey to select how
                   long the oscilloscope will run. When the selected time has passed the oscilloscope will
                   stop. The specified time may be exceeded. If an error occurs the action specified using
                   the On Error softkey will occur. The actual test time is displayed above the softkeys.
                   Minimum Sigma Choose this option and then use the Sigma softkey to select a
                   minimum sigma. The mask test runs until enough waveforms are tested to achieve a
                   minimum test sigma. (If an error occurs the oscilloscope will perform the action specified
                   by the On Error softkey.) Note that this is a test sigma (the max achievable process
                   sigma, assuming no defects, for a certain number of tested waveforms) as opposed to a
                   process sigma (which is tied to the amount of failures per test). The sigma value may
                   exceed the selected value when a small sigma value is chosen. The actual sigma is
                   displayed.
               On Error
                   The On Error setting specifies the action to take when the input waveform does not
                   conform to the mask. This setting supersedes the Run Until setting.
                   Stop The oscilloscope will stop when the first error is detected (on the first waveform
                   that does not conform to the mask). This setting supersedes the Minimum # of Tests and
                   Minimum Time settings.
                       Save The oscilloscope will save the screen image based on the settings you have
                       chosen. See “Saving Oscilloscope Data” on page 247.
                       Print   The oscilloscope will print the waveform display when the first error is detected.
                       This option is only available when a printer is connected as described in “To print the
                       oscilloscope’s display” on page 245.
                 Source Lock
                       When you turn on Source Lock using the Source Lock softkey, the mask is redrawn to
                       match the source whenever you move the waveform. For example, if you change the
                       horizontal timebase or the vertical gain the mask is redrawn with the new settings.
                       When you turn off Source Lock, the mask is not redrawn when horizontal or vertical
                       settings are changed.
                       If you change the Source channel, the mask is not erased. It is re-scaled to the vertical
                       gain and offset settings of the channel to which it is assigned. To create a new mask for
                       the selected source channel, press the up arrow softkey, then press Automask, then
                       press Create Mask.
                       The Source softkey in the Mask Setup Menu is the same as the Source softkey in the
                       Automask Menu.
               The Pass/fail status is evaluated periodically. When the evaluation of the testing period
               results in a pulse output, the trigger output pulses high (+ 5 V). Otherwise, the trigger
               output remains at low (0 V).
Mask Statistics
                       From the Mask Test Menu, press the Statistics softkey to enter the Mask Statistics
                       Menu.
               Reset Statistics
                   Press this softkey to reset the statistics counters to zero.
                   Note that statistics are also reset when:
                   • Mask Test is switched on after being switched off.
                   • Clear Mask softkey is pressed.
                   • An Automask is created.
                   Additionally, the accumulated time counter is reset whenever the oscilloscope is run after
                   the acquisition was stopped.
               Transparent
                   Enable Transparent mode to write measurement values and statistics on the screen
                   without a background. Disable the Transparent mode to show them with a light blue
                   background. This controls the display of statistics for both Quick Meas and the optional
                   Mask test feature.
                       Mask Title
                       The Mask Title is a string of ASCII characters.
                       Example: autoMask CH1 OCT 03 09:40:26 2008
                       When a mask file contains the keyword “autoMask” in the title, the edge of the mask is
                       passing by definition. Otherwise, the edge of the mask is defined as a failure.
Region 1
Region 2
               Up to 8 regions can be defined for a mask. They can be numbered 1-8. They can appear in
               any order in the .msk file. The numbering of the regions must go from top to bottom, left
               to right. The lowest numbered region
               An Automask file contains two special regions: the region “glued” to the top of the
               display, and the region that is “glued” to the bottom. The top region is indicated by
               y-values of “MAX” for the first and last points. The bottom region is indicated by y-values
               of “MIN” for the first and last points.
               The top region must be the lowest numbered region in the file. The bottom region must be
               the highest numbered region in the file.
               Region number 1 is the top mask region. The vertices in Region 1 describe points along a
               line; that line is the bottom edge of the top portion of the mask.
               Similarly, the vertices in Region 2 describe the line that forms the top of the bottom part of
               the mask.
               The vertices in a mask file are normalized. There are four parameters that define how
               values are normalized:
               • X1
                       • X
                       • Y1
                       • Y2
                       These four parameters are defined in the Oscilloscope Setup portion of the mask file.
                       The Y-values (normally voltage) are normalized in the file using the following equation:
                       Ynorm = (Y - Y1)/Y
                       where Y = Y2 - Y1
                       To convert the normalized Y-values in the mask file to voltage:
                       Y = (Ynorm * Y) + Y1
                       where Y = Y2 - Y1
                       The X-values (normally time) are normalized in the file using the following equation:
                       Xnorm = (X - X1)/X
                       To convert the normalized X-values to time:
                       X = (Xnorm * X) + X1
"All Regions"
               /* Region Number */ 1
               /* Number of vertices */ 4
                 -12.50, MAX
                 -10.00, 1.750
                  10.00, 1.750
                  12.50, MAX
               /* Region Number */ 2
               /* Number of vertices */ 5
                 -10.00, 1.000
                 -12.50, 0.500
                 -15.00, 0.500
                 -15.00, 1.500
-12.50, 1.500
                       /* Region Number */ 3
                       /* Number of vertices */ 6
                         -05.00, 1.000
                         -02.50, 0.500
                          02.50, 0.500
                          05.00, 1.000
                          02.50, 1.500
                         -02.50, 1.500
                       /* Region Number */ 4
                       /* Number of vertices */ 5
                          10.00, 1.000
                          12.50, 0.500
                          15.00, 0.500
                          15.00, 1.500
                          12.50, 1.500
                       /* Region Number */ 5
                       /* Number of vertices */ 5
                         -10.00, -1.000
                         -12.50, -0.500
                         -15.00, -0.500
                         -15.00, -1.500
                         -12.50, -1.500
                       /* Region Number */ 6
                       /* Number of vertices */ 6
                         -05.00, -1.000
                         -02.50, -0.500
                          02.50, -0.500
                          05.00, -1.000
                          02.50, -1.500
                         -02.50, -1.500
                       /* Region Number */ 7
                       /* Number of vertices */ 5
                          10.00, -1.000
                          12.50, -0.500
                          15.00, -0.500
                          15.00, -1.500
                          12.50, -1.500
/* Region Number */ 8
               /* Number of vertices */ 4
                 -12.50, MIN
                 -10.00, -1.750
                  10.00, -1.750
                  12.50, MIN
               setup
               :MTES:ENAB 1
               :CHAN1:RANG +4.00E+00;OFFS +0.0E+00;COUP DC;IMP ONEM;DISP 1;BWL 0;INV 0
               :CHAN1:LAB "1";UNIT VOLT;PROB +1.0E+00;PROB:SKEW +0.0E+00;STYP SING
               :CHAN2:RANG +16.0E+00;OFFS +1.62400E+00;COUP DC;IMP FIFT;DISP 0;BWL 0;INV 0
               :CHAN2:LAB "2";UNIT VOLT;PROB +1.0E+00;PROB:SKEW +0.0E+00;STYP SING
               :CHAN3:RANG +40.0E+00;OFFS +0.0E+00;COUP DC;IMP ONEM;DISP 0;BWL 0;INV 0
               :CHAN3:LAB "3";UNIT VOLT;PROB +1.0E+00;PROB:SKEW +0.0E+00;STYP SING
               :CHAN4:RANG +40.0E+00;OFFS +0.0E+00;COUP DC;IMP ONEM;DISP 0;BWL 0;INV 0
               :CHAN4:LAB "4";UNIT VOLT;PROB +1.0E+00;PROB:SKEW +0.0E+00;STYP SING
               :EXT:BWL 0;IMP ONEM;RANG +5E+00;UNIT VOLT;PROB +1.0E+00;PROB:STYP SING
               :TIM:MODE MAIN;REF CENT;MAIN:RANG +50.00E-09;POS +0.0E+00
               :TRIG:MODE EDGE;SWE AUTO;NREJ 0;HFR 0;HOLD +60E-09
               :TRIG:EDGE:SOUR CHAN1;LEV -75.00E-03;SLOP POS;REJ OFF;COUP DC
               :ACQ:MODE RTIM;TYPE NORM;COMP 100;COUNT 8;SEGM:COUN 2
               :DISP:LAB 0;CONN 1;PERS MIN;SOUR PMEM1
               :HARD:APR "";AREA SCR;FACT 0;FFE 0;INKS 1;PAL NONE;LAY PORT
               :SAVE:FIL "mask_0"
               :SAVE:IMAG:AREA GRAT;FACT 0;FORM NONE;INKS 0;PAL COL
               :SAVE:WAV:FORM NONE
               :MTES:SOUR CHAN1;ENAB 1;LOCK 1
               :MTES:AMAS:SOUR CHAN1;UNIT DIV;XDEL +3.00000000E-001;YDEL +2.00000000E-001
               :MTES:SCAL:BIND 0;X1 +0.0E+00;XDEL +1.0000E-09;Y1 +0.0E+00;Y2 +1.00000E+00
               :MTES:RMOD FOR;RMOD:TIME +1E+00;WAV 1000;SIGM +6.0E+00
               :MTES:RMOD:FACT:STOP 0;PRIN 0;SAVE 0
               end_setup
12
Controls and Connectors
Front Panel 352
   5000/6000 Series Front Panel (4-channel) 352
   5000/6000 Series Front Panel (2-Channel, differences only) 353
   6000L Series Front and Rear Panel 354
   7000 Series Front Panel (4-Channel) 355
   7000 Series Front Panel (2-Channel, differences only) 356
   Front Panel Control and Connector Descriptions 357
                                                                  351
12 Controls and Connectors
Front Panel
                                                                                                        17.
                                                                                                      Trigger
                                                                                                      Controls
                           21.                                                                          14.
                          Entry                                                                       [Utility]
                          Knob                                                                          Key
                         20.                                                                           15. File
                     [AutoScale]                                                                        Keys
                         Key
                                                                                                          7.
                          23.                                                                          [Label]
                        Softkeys                                                                         Key
                                                                                                      4. Channel
                                                                                                      Input BNC
                                                                                                      Connector
                                                                                                          5.
                                                                                                      AutoProbe
                                                                                                      Interface
        1.     12.    2. USB  3. Probe  13. Digital 8. Vertical   9.   6. [Math] 10. Vertical
      Power Intensity Host Compensation Channel      Position Channel      Key   Sensitivity
      Switch Control Port    Terminals   Controls    Control    On/Off
                                                                 Key
                                                                                                         External
                                                                                                        Trigger
                                                                                                        Input
                       The differences between the front panel of the 4-channel oscilloscopes and the
                       2-channel oscilloscopes are:
                       • The 2-channel oscilloscope has two sets of channel controls
                       • The 2-channel oscilloscope’s external trigger input is on the front panel instead of the
                         rear panel. Some trigger features differ. See “External Trigger Input” on page 92.
FRONT PANEL
  1. Power      25.       26. LAN    27. Reset       12.        2. USB    3. Probe     5.    4. Channel
   Switch     Power        Status     Switch      Intensity    Host Port Compen- AutoProbe Input BNC
             Indicator   Indicator                 Control                 sation  Interface Connector
                                                                         Terminals
REAR PANEL
                                                                                                     13. Digital
                              20.                                                                     Channel
                          [AutoScale]                                                                 Controls
                              Key
                                21.                                                                      12.
                               Entry                                                                  Intensity
                               Knob                                                                    Control
                               22.                                                                     11.
                             Measure                                                                 Waveform
                              Keys                                                                    Keys
                               23.
                             Softkeys                                                                 10. Vertical
                                                                                                      Sensitivity
                                                                                                         9.
                                                                                                      Channel
                                                                                                      On/Off
                                                                                                        Key
                                                                                                    External
                                                                                                    Trigger
                                                                                                    Input
                    The differences between the front panel of the 4-channel oscilloscopes and the
                    2-channel oscilloscopes are:
                    • The 2-channel oscilloscope has two sets of channel controls
                    • The 2-channel oscilloscope’s external trigger input is on the front panel instead of the
                      rear panel. Some trigger features differ. See “External Trigger Input” on page 92.
                       1. Power Switch     Press once to switch power on; press again to switch power off. See
                       page 49.
                       2. USB Host Port     This port is for connecting USB mass storage devices or printers to
                       the oscilloscope.
                       Connect a USB compliant mass storage device (flash drive, disk drive, etc.) to store or
                       recall oscilloscope setup files or waveforms.
                       To print, connect a USB compliant printer. For more information about printing see “To
                       print the oscilloscope’s display” on page 245.
                       You can also use the USB port to update the oscilloscope’s system software when
                       updates are available.
                       You do not need to take special precautions before removing the USB mass storage
                       device from the oscilloscope (you do not need to “eject” it). Simply unplug the USB mass
                       storage device from the oscilloscope when the file operation is complete.
                       For more information about saving to a USB mass storage device see “Saving
                       Oscilloscope Data” on page 247.
                       Do not connect a host computer to the oscilloscope’s USB host port: use the device
     CAUTION
                       port. A host computer sees the oscilloscope as a device, so connect the host
                       computer to the oscilloscope’s device port (on the rear panel). See “Web
                       Interface” on page 275.
                       3. Probe Compensation Terminals Use the signal at this terminal to match a probe’s
                       input capacitance to the oscilloscope channel to which it is connected. See page 77.
                       4. Channel Input BNC Connector Attach the oscilloscope probe or BNC cable to the
                       BNC connector. This is the channel’s input connector.
                       5. AutoProbe Interface When you connect a probe to the oscilloscope, the AutoProbe
                       Interface attempts to determine the type of probe and set its parameters in the Probe
                       Menu accordingly. On all models except 100 MHz 6000 Series. See “AutoProbe
                       Interface” on page 72.
                    6. [Math] Key The [Math] key provides access to add, subtract, multiply, differentiate,
                    integrate, FFT (Fast Fourier Transform) and square root functions. See “Measurements
                    and Math Functions” on page 191.
                    7. [Label] Key    Press this key to access the Label Menu, which lets you enter labels to
                    identify each trace on the oscilloscope display. See “Using Labels” on page 183.
                    8. Vertical Position Control     Use this knob to change the channel’s vertical position on
                    the display. There is one Vertical Position control for each channel. See “Using the
                    Analog Channels” on page 59.
                    9. Channel On/Off Key Use this key to switch the channel on or off, or to access the
                    channel’s menu in the softkeys. There is one Channel On/Off key for each channel. See
                    “Using the Analog Channels” on page 59.
                    10. Vertical Sensitivity There is one knob marked           for each channel. Use these
                    knobs to change the vertical sensitivity (gain) of each channel. See “Using the Analog
                    Channels” on page 59.
                    11. Waveform Keys The [Acquire] key lets you set the oscilloscope to acquire in
                    Normal, Peak Detect, Averaging, or High Resolution modes (see “Acquisition Modes” on
                    page 265), and lets you turn Realtime sampling off or on (see page 269). The [Display]
                    key lets you access the menu where you can select infinite persistence (see page 182),
                    switch vectors on or off (see page 187), and adjust the display grid (graticule) intensity
                    (see page 181).
                    13. Digital Channel Controls These controls switch the digital channels on and off,
                    and can be used to select a digital channel for repositioning on the display.
                    Press the [D15-D0] key to turn on the digital channels (the [D15-D0] key will illuminate).
                    There are two indicators above the Digital Channel Control Knob: Select and Pos
                    (Position). When Select is illuminated, rotate the knob to select a digital trace. The
                    selected trace will change from blue to red. Push the Digital Channel Control Knob to
                    toggle from Select to Pos. Rotate the knob and press it again to re-position the digital
                    trace.
                       If a trace is repositioned over an existing trace the indicator at the left edge of the trace
                       will change from Dnn designation (where nn is a one or two digit channel number from 0
                       to 15) to D*. The “*” indicates that two channels are overlaid.
                       You can rotate the knob to select an overlaid channel, then press the knob to position it
                       just as you would any other channel.
                       For more information on digital channels see Chapter 13, “Digital Channels,” starting on
                       page 371.
                       14. [Utility] Key   Press this key to access the Utility Menu, which lets you configure
                       the oscilloscope’s I/O settings, use the file explorer, set preferences, access the service
                       menu, and choose other options.
                       15. File Keys   Press the [Save/Recall] key to save or recall a waveform or setup, or to
                       access the Default Setup function, which restores many settings to their factory
                       configuration (details on “To restore the oscilloscope to its default configuration” on
                       page 392). The [Print] key opens the Print Configuration Menu so you can print the
                       displayed waveforms. See “Saving Oscilloscope Data” on page 247 and “Printing the
                       Oscilloscope’s Display” on page 244.
                       16. Run Controls Press [Run/Stop] to make the oscilloscope begin looking for a
                       trigger. The [Run/Stop] key will illuminate in green. If the trigger mode is set to
                       “Normal,” the display will not update until a trigger is found. If the trigger mode is set to
                       “Auto,” the oscilloscope looks for a trigger, and if none is found, it will automatically
                       trigger, and the display will immediately show the input signals. In this case, the Auto
                       indicator at the top of the display will change to Auto? and the background will flash,
                       indicating that the oscilloscope is forcing triggers.
                       Press [Run/Stop] again to stop acquiring data. The key will illuminate in red. Now you
                       can pan across and zoom-in on the acquired data.
                       Press [Single] to make a single acquisition of data. The key will illuminate in yellow until
                       the oscilloscope triggers. See “To start and stop an acquisition” on page 262.
                       17. Trigger Controls These controls determine how the oscilloscope triggers to
                       capture data. See Chapter 4, “Triggering,” starting on page 81.
                    Horizontal Sweep Speed Control Turn the knob in the Horizontal section that is
                    marked          to adjust the sweep speed. The symbols under the knob indicate that this
                    control has the effect of spreading out or zooming in on the waveform using the
                    horizontal scale. The horizontal sweep speed control changes the time per horizontal
                    division on the display. For more information see “To set up the Horizontal time base” on
                    page 64.
                    Horizontal Position Control Turn the knob marked           to pan through the waveform
                    data horizontally. You can see the captured waveform before the trigger (turn the knob
                    clockwise) or after the trigger (turn the knob counterclockwise). If you pan through the
                    waveform when the oscilloscope is stopped (not in Run mode) then you are looking at the
                    waveform data from the last acquisition taken. See “To set up the Horizontal time
                    base” on page 64.
                    19. Display The display shows captured waveforms using a different color for each
                    channel. For more information about display modes see Chapter 5, “Displaying,” starting
                    on page 173. Signal detail is displayed using 256 levels of intensity. For more information
                    about viewing signal detail see “Waveform Intensity and Signal Detail” on page 180.
                    20. [AutoScale] Key When you press the [AutoScale] key the oscilloscope will
                    quickly determine which channels have activity, and it will turn these channels on and
                    scale them to display the input signals. See “AutoScale” on page 177
                    21. Entry Knob The entry knob is used to select items from menus and to change
                    values. Its function changes based upon which menu is displayed. Note that the curved
                    arrow symbol        above the entry knob illuminates whenever the entry knob can be used
                    to select a value. Rotate and push the entry knob to select among the choices that are
                    shown on the softkeys. More information about the Entry knob is on page 53.
                       22. Measure Keys Press the [Cursors] key to turn on cursors that you can use for
                       making measurements. Press the [Quick Meas] key to access a set of predefined
                       measurements. See Chapter 6, “Measurements and Math Functions,” starting on page
                       191.
                       23. Softkeys     The functions of these keys change based upon the menus shown on the
                       display directly above the keys.
                       24. Battery Status Indicators These indicators appear only on 6000A models with
                       option BAT (battery power). See “Power-on the Battery-Powered 6000A Series
                       Oscilloscope” on page 49 for important operation and safety information.
25. Power Indicator (6000L models only.) Illuminates green when power is on.
                       26. LAN Status Indicator (6000L models only.) This indicator illuminates green when
                       a LAN connection has been detected and is connected as configured. The LAN status
                       indicator illuminates red when the following LAN faults occur:
                       • failure to acquire a valid IP address
                       • detection of a duplicate IP address
                       • failure to renew an already acquired DHCP lease
                       The LAN status indicator flashes green when the identify function is activated (see
                       “Identification Function” on page 286).
                       27. Reset Switch   (6000L models only.) When power is on, press and release this
                       recessed pushbutton to default LAN parameters.
Rear Panel
0RODUCT 3ERIAL .UMBER ,ABEL
>8:H$CB7"%%&
                                                   L6GC>C</B6>CI6>C                                                                                    40
                                                   <GDJC9ID6KD>9                                                                         >HB<GE&"6
                                                                                                               :A:8IG>86A
                                                   :A:8IG>8H=D8@                                              :A:8IGDC>8
                                                                                                               B:6HJG>C<                     C&%&).
                                                                                                               I:HI>C<
                                                                                                               :FJ>EB:CI      8     JH
                                                   s A>C:&'%LB6M                                                                                           86A>7G6I>DCJCEGDI:8I:9
                                                   &%%"&'%K!*%$+%$)%%=o
                                                   &%%"')%K!*%$+%=o                                                 :MIIG><             IG><                86A>7G6I>DCEGDI:8I:9
                                                                                                                        >C                DJI
                                                                                                                                   &*K
                                                                               JH7                         
                                                                                                           CDIJH:9
                                                                                                           ;DG'"8=              B6M
E
  Kensington         Video       USB Device       USB Host Port                   LAN Port                                        External                    Trigger            GPIB
     Lock            Output        Port for       for saving data                                                                 Trigger                     Output           Connector
                                  controlling           from                                                                       Input                       BNC
                                 oscilloscope      oscilloscope                                                                     BNC
                                                    and loading
                                                      software
                                                      updates
0RODUCT 3ERIAL .UMBER ,ABEL
>8:H$CB7"%%&
                                                  L6GC>C</B6>CI6>C                                                                                  40
                                                  <GDJC9ID6KD>9                                                                       >HB<GE&"6
                                                                                                            :A:8IG>86A
                                                  :A:8IG>8H=D8@                                            :A:8IGDC>8
                                                                                                            B:6HJG>C<                     C&%&).
                                                                                                            I:HI>C<
                                                                                                            :FJ>EB:CI      8     JH
                                                  s A>C:&'%LB6M                                                                                         86A>7G6I>DCJCEGDI:8I:9
                                                  &%%"&'%K!*%$+%$)%%=o
                                                  &%%"')%K!*%$+%=o                                               :MIIG><             IG><                86A>7G6I>DCEGDI:8I:9
                                                                                                                     >C                DJI
                                                                                                                                &*K
                                                                                                           ;DG'"8= 
                                                                                                          CDIJH:9
                                                                              JH7                                              B6M
E
  Kensington         Video       USB Device      USB Host Port                   LAN Port                                      External                    Trigger            GPIB
     Lock            Output        Port for      for saving data                                                               Trigger                     Output           Connector
                                  controlling          from                                                                     Input                       BNC
                                 oscilloscope     oscilloscope                                                                   BNC
                                                   and loading
                                                     software
                                                     updates
                                          86A>7G6I>DCJCEGDI:8I:9                                                               s A>C:&'%LB6M
                            :MI                                                                                                  &%%"&'%K!*%$+%$)%%=o
                                          86A>7G6I>DCEGDI:8I:9                                                                  &%%"')%K!*%$+%=o
                            IG><                                                                         JH7
                             >C                                                                                             L6GC>C</B6>CI6>C<GDJC9
                             &*K   IG><   &%B=o            M<6                                                              ID6KD>9:A:8IG>8H=D8@
                            B6M
                                   DJI     G:;     &*K    K>9:D                                       =DHI   9:K>8:
                                                  B6M      DJI               9><>I6A9&*"9%                          A6CCdc"6jid"B9>M     <E>7
JH:G:8DBB:C9:9867A:DCAN
54684e92
                                  86A>7G6I>DCJCEGDI:8I:9                                                          s A>C:&'%LB6M
                                                                                                                    &%%"&'%K!*%$+%$)%%=o
                                  86A>7G6I>DCEGDI:8I:9                                                             &%%"')%K!*%$+%=o
                                                                                            JH7
                                                                                                               L6GC>C</B6>CI6>C<GDJC9
                           IG><   &%B=o            M<6                                                         ID6KD>9:A:8IG>8H=D8@
                           DJI     G:;     &*K    K>9:D                                  =DHI   9:K>8:
                                          B6M      DJI          9><>I6A9&*"9%                          A6CCdc"6jid"B9>M     <E>7
JH:G:8DBB:C9:9867A:DCAN
         4. Calibration 6. XGA Video                                  5. Digital                     11. Power                             13. Ground                    10. Kensington
             Protect        Output                                      Input                        Connector                              Terminal                          Lock
             Switch                                                   Connector
                                         86A>7G6I>DCJCEGDI:8I:9
                           :MI           86A>7G6I>DCEGDI:8I:9                                                                  L6GC>C<
                           IG><                                                                         JH7
                                                                                                                      6AL6NH8DCC:8II=:G:6GE6C:A<GDJC9
                            >C
                            &*K   IG><   &%B=o            M<6                                                         I:GB>C6A7A68@ID:6GI=<GDJC9L=:C
                                                                                                                      DE:G6I>C<I=:>CHIGJB:CI;GDB76II:GNEDL:G
                           B6M
                                  DJI     G:;     &*K    K>9:D                                       =DHI   9:K>8:
                                                 B6M      DJI               9><>I6A9&*"9%                          A6C
                                                                                                                                                 <E>7
JH:G:8DBB:C9:9867A:DCAN
54684e94
                                                                                                                                                                                                                                9. LAN
                              :MIIG><   IG><     &%B=o                                                                                                    &-&$53*$"-
                                                                                                                                                           &-&$530/*$               -3$
                                                                                                                                                                                                 *$&4/.#
                                                                                                                                                                                                                                 Port
                                 >C      DJI       G:;                                           M<6K>9:D                                                 .&"463*/(
                               &*K                &*K                    9><>I6A9&*"9%                                                                    5&45*/(
                                                                                                                                                                           $   64
40
                                          86A>7G6I>DCJCEGDI:8I:9
                                          86A>7G6I>DCEGDI:8I:9
                                                                                                                                           L6GC>C</B6>CI6>C<GDJC9
                                                                                                                                           ID6KD>9:A:8IG>8H=D8@
                                                                                                                                           sA>C:&'%LB6M
                                                                                                                                           .+"&))K!)-"))%=o
                                                                                                                                           &.'"'--K!)-"++=o
*$&4/.#
40
                          86A>7G6I>DCJCEGDI:8I:9
                          86A>7G6I>DCEGDI:8I:9
                                                                                                                           L6GC>C</B6>CI6>C<GDJC9
                                                                                                                           ID6KD>9:A:8IG>8H=D8@
                                                                                                                           sA>C:&'%LB6M
                                                                                                                           .+"&))K!)-"))%=o
                                                                                                                           &.'"'--K!)-"++=o
F
                       1. External Trigger Input   See “External Trigger Input” on page 92 for an explanation
                       of this feature.
                       5. Digital Input Connector Connect the digital probe cable to this connector (MSO
                       models only).
                       6. XGA Video Output This video output is always active. You can connect an external
                       monitor to this connector.
                       7. USB Host Port This port functions identically to the USB host port on the front
                       panel. USB Host Port is used for saving data from the oscilloscope and loading software
                       updates. See also page 357.
                       8. USB Device Port This port is for connecting the oscilloscope to a host PC. You can
                       issue remote commands from a host PC to the oscilloscope via the USB device port. See
                       “Remote Programming with Agilent IO Libraries” on page 284.
                       9. LAN Port    You can communicate with the oscilloscope and use the Remote Front
                       Panel feature using the LAN port. See “Web Interface” on page 275 and “Accessing the
                       Web Interface” on page 280.
                       10. Kensington Lock     This is where you can attach a Kensington lock for securing the
                       instrument.
                       11. Power Connector Attach the power cord here, or on 6000A Option BAT models
                       attach power connector from 0950-4866 power supply.
                       12. GPIB Connector This port is for controlling the oscilloscope via GPIB. You can
                       issue remote commands from a host computer and retrieve data from the oscilloscope.
                       See “Remote Programming with Agilent IO Libraries” on page 284.
                    13. Ground Terminal Option BAT oscilloscopes: Connect the oscilloscope to earth
                    ground using this terminal. Maintain oscilloscope ground connection. Do not negate the
                    protective action of the ground connection to the oscilloscope. If the device under test
                    has voltages greater than 30 Vrms, the oscilloscope must be grounded through its ground
                    terminal to prevent electric shock to the user or damage to the oscilloscope.
13
Digital Channels
To connect the digital probes to the device under test 372
Acquiring waveforms using the digital channels 375
To display digital channels using AutoScale 376
Interpreting the digital waveform display 378
To switch all digital channels on or off 379
To switch groups of channels on or off 379
To switch a single channel on or off 379
To change the displayed size of the digital channels 379
To reposition a digital channel 380
To change the logic threshold for digital channels 380
To display digital channels as a bus 381
Digital channel signal fidelity: Probe impedance and grounding 385
To replace digital probe leads 389
                                                                    371
13 Digital Channels
Off
                      2 Connect the digital probe cable to the D15 - D0 connector on the rear panel of the
                        mixed-signal oscilloscope. The digital probe cable is indexed so you can connect it
                        only one way. You do not need to power-off the oscilloscope.
                      Probe cable for digital channels: Use only the Agilent part number 54620-68701
      CAUTION
                      digital probe kit supplied with the mixed-signal oscilloscope.
3 Route the cable under the oscilloscope and bring it out in front.
                       4 Connect the ground lead on each set of channels (each pod), using a probe grabber.
                         The ground lead improves signal fidelity to the oscilloscope, ensuring accurate
                         measurements.
                          Channel
                          Pod Ground
                                                     Circuit
                                                     Ground
                       5 Connect a grabber to one of the probe leads. (Other probe leads are omitted from the
                         figure for clarity.)
Grabber
                      7 For high-speed signals, connect a ground lead to the probe lead, connect a grabber to
                        the ground lead, and attach the grabber to ground in the device under test.
Signal Lead
Ground Lead
Grabber
                       8 Repeat steps 3 through 6 until you have connected all points of interest.
                                                 Signals
Ground
               Example
                      Install probe clips on channels 0 and 1 of the digital probe cable. Connect the probes for
                      digital channels 0 and 1 to the Probe Comp output on the front panel of the instrument.
                      Be sure to connect the ground lead to the ground lug beside the Probe Comp output. Set
                       the instrument to the factory default configuration by pressing the [Save/Recall] key,
                       then the Default Setup softkey. Then press the [AutoScale] key. You should see a display
                       similar to the following.
                                                                                                       Threshold
                                                                                                       level
 Digital channel
 identifier
                                                                                                    Activity
                                                                                                    indicator
                      Activity indicator
                      When any digital channels are turned on, an activity indicator is displayed in the status
                      line at the bottom of the display. A digital channel can be always high ( ), always low
                      ( ), or actively toggling logic states ( ). Any channel that is turned off will be grayed out
                      in the activity indicator.
TTL +1.4 V
CMOS +2.5 V
ECL –1.3 V
                      The threshold you set applies to all channels within the selected D15 Thru D8 or
                      D7 Thru D0 group. Each of the two channel groups can be set to a different threshold if
                      desired.
                      Values greater than the set threshold are high (1) and values less than the set threshold
                      are low (0).
                      If the Thresholds softkey is set to User, press the User softkey for the channel group,
                      then turn the Entry knob to set the logic threshold. There is one User softkey for each
                      group of channels.
                      On 7000 Series oscilloscopes:          To select a waveform, push the knob in the Digital
                      controls until the Select indicator is lit; then, turn the knob. The selected waveform is
                      highlighted in red. To position the selected waveform, push the knob until the Pos
                      indicator is lit; then, turn the knob.
                       If a trace is repositioned over an existing trace the indicator at the left edge of the trace
                       will change from Dnn designation (where nn is a one or two digit channel number from 0
                       to 15) to D*. The “*” indicates that two channels are overlaid.
Bus softkey
                       Next, select a bus. Rotate the Entry knob, then press the Entry knob or the Bus1/Bus2
                       softkey to switch it on.
                       Use the second softkey from the left and the Entry knob to select individual channels to
                       be included in the bus. You can rotate the Entry knob and push the softkey to select
                       channels. You can also press the Select/Deselect D15-D8 and Select/Deselect D7-D0
                       softkeys to include or exclude groups of eight channels in each bus.
                       If the bus display is blank, completely white, or if the display includes “...”, you need to
                       expand the horizontal scale to allow space for the data to be shown, or use the cursors to
                       display the values (see page 382).
                       The Base softkey lets you choose to display the bus values in hex or binary.
      Bus values
      can be
      shown in
      hex or
      binary
                       4 Use the Entry knob and the X1 and X2 softkeys to position the cursors where you want
                         to read the bus values.
X1 cursor
X2 cursor
Bus values
    Bus values at
    cursors shown
    here
                       When you press the [D15-D0] key to display the Digital Channel Menu, the digital activity
                       indicator is shown where the cursor values were and the bus values at the cursors are
                       displayed in the graticule.
      Digital channel
      values at cursor
      Analog channel
      values at cursor
      Trigger pattern
      definition
                                                                                         Bus values
                                                                                         displayed
                 Input Impedance
                       The logic probes are passive probes, which offer high input impedance and high
                       bandwidths. They usually provide some attenuation of the signal to the oscilloscope,
                       typically 20 dB.
                       Passive probe input impedance is generally specified in terms of a parallel capacitance
                       and resistance. The resistance is the sum of the tip resistor value and the input
                       resistance of the test instrument (see the following figure). The capacitance is the series
                       combination of the tip compensating capacitor and the cable, plus instrument
                       capacitance in parallel with the stray tip capacitance to ground. While this results in an
                       input impedance specification that is an accurate model for DC and low frequencies, the
                       high-frequency model of the probe input is more useful (see the following figure). This
                       high-frequency model takes into account pure tip capacitance to ground as well as series
                       tip resistance, and the cable’s characteristic impedance (Zo).
                       DC AND LOW-FREQUENCY PROBE EQUIVALENT CIRCUIT
8.5 pF 100 kW
1 pF 100 kW 150 W
                      The impedance plots for the two models are shown in these figures. By comparing the
                      two plots, you can see that both the series tip resistor and the cable’s characteristic
                      impedance extend the input impedance significantly. The stray tip capacitance, which is
                      generally small (1 pF), sets the final break point on the impedance chart.
                      IMPEDANCE VERSUS FREQUENCY FOR BOTH PROBE CIRCUIT MODELS
                                  100k 
                                  10k                                            High
                                                                                  Frequency
                                                                                  Model
                                   1k 
                      Impedance
                                                               Typical
                                  100                         Model
10 
                                    1
                                     10kHz   100kHz    1MHz     10MHz    100MHz    1GHz
                                                       Frequency
                      The logic probes are represented by the high-frequency circuit model shown above. They
                      are designed to provide as much series tip resistance as possible. Stray tip capacitance to
                      ground is minimized by the proper mechanical design of the probe tip assembly. This
                      provides the maximum input impedance at high frequencies.
                 Probe Grounding
                       A probe ground is the low-impedance path for current to return to the source from the
                       probe. Increased length in this path will, at high frequencies, create large common mode
                       voltages at the probe input. The voltage generated behaves as if this path were an
                       inductor according to the equation:
                                                                         di
                                                                V = L -----
                                                                         dt
                       Increasing the ground inductance (L), increasing the current (di) or decreasing the
                       transition time (dt), will all result in increasing the voltage (V). When this voltage exceeds
                       the threshold voltage defined in the oscilloscope, a false data measurement will occur.
                       Sharing one probe ground with many probes forces all the current that flows into each
                       probe to return through the same common ground inductance of the probe whose ground
                       return is used. The result is increased current (di) in the above equation, and, depending
                       on the transition time (dt), the common mode voltage may increase to a level that causes
                       false data generation.
                       Common Mode Input Voltage Model
                       Probe 1                   Zin
                                                                    i1
                                              L (GND)
                        Probe                                            Vn (Common Mode)
                        Ground                           i1+i2+in
i2 +in
Probe 2 Zin
in
Probe N Zin
                       In addition to the common mode voltage, longer ground returns also degrade the pulse
                       fidelity of the probe system. Rise time is increased, and ringing, due to the undamped LC
                       circuit at the input of the probe, is also increased. Because the digital channels display
                      reconstructed waveforms, they do not show ringing and perturbations. You will not find
                      ground problems through examination of the waveform display. In fact, it is likely you will
                      discover the problem through random glitches or inconsistent data measurements. Use
                      the analog channels to view ringing and perturbations.
For other replacement parts, consult the InfiniiVision Oscilloscope Service Guides.
14
Utilities
To restore the oscilloscope to its default configuration 392
To perform service functions 393
User Calibration 394
Self Test 397
                                                             391
14 Utilities
Horizontal normal mode, 100 s/div scale, 0 s delay, center time reference.
                Trigger Edge trigger, Auto sweep mode, 0 V level, channel 1 source, DC coupling, rising
                edge slope, 60 ns holdoff time.
Other Acquire mode normal, [Run/Stop] to Run, cursors and measurements off.
                Labels All custom labels that you have created in the Label Library are preserved (not
                erased), but all channel labels will be set to their original names.
User Calibration
               Perform user-calibration:
               • Each year or after 2000 hours of operation.
               • If the ambient temperature is >10° C from the calibration temperature.
               • If you want to maximize the measurement accuracy.
               The amount of use, environmental conditions, and experience with other instruments
               help determine if you need shorter User Cal intervals.
               User Cal performs an internal self-alignment routine to optimize the signal path in the
               oscilloscope. The routine uses internally generated signals to optimize circuits that affect
               channel sensitivity, offset, and trigger parameters. Disconnect all inputs and allow the
               oscilloscope to warm up before performing this procedure.
               Performing User Cal will invalidate your Certificate of Calibration. If NIST (National
               Institute of Standards and Technology) traceability is required, perform the Performance
               Verification procedure in the Agilent InfiniiVision Oscilloscopes Service Guide using
               traceable sources.
                       For a 2-channel oscilloscope, connect a BNC tee to the equal length cables. Then connect
                       a BNC(f)-to-BNC(f) (also called a barrel connector) to the tee as shown below.
                       USER CALIBRATION CABLE FOR 2-CHANNEL OSCILLOSCOPE
                                                                                       Longer cable
                        To Channel 1                                                   to TRIG OUT
To Channel 2
                       For a 4-channel oscilloscope, connect BNC tees to the equal-length cables as shown
                       below. Then connect a BNC(f)-to-BNC(f) (barrel connector) to the tee as shown below.
                       USER CALIBRATION CABLE FOR 4-CHANNEL OSCILLOSCOPE
                                To Channel 1
To Channel 2
To Channel 3
                                                                                          Longer cable
                                                                                          to TRIG OUT
                                                                To Channel 4
               3 Connect a BNC cable (40 inches maximum) from the TRIG OUT connector on the rear
                 panel to the BNC barrel connector.
               4 Press the [Utility] key, then press the Service softkey.
               5 Begin the Self Cal by pressing the Start User Cal softkey.
               6 When the User Cal is completed, set the rear-panel CALIBRATION switch to
                 PROTECTED.
Self Test
                       Pressing [Utility]&Service&Start Self Test performs a series of internal procedures to
                       verify that the oscilloscope is operating properly.
                       It is recommended you run Self Test:
                       • after experiencing abnormal operation.
                       • for additional information to better describe an oscilloscope failure.
                       • to verify proper operation after the oscilloscope has been repaired.
                       Successfully passing Self Test does not guarantee 100% of the oscilloscope's
                       functionality. Self Test is designed to provide an 80% confidence level that the
                       oscilloscope is operating properly.
15
Reference
About Oscilloscope 400
Software and Firmware Updates 400
Installed Licenses 400
Upgrading to an MSO 401
Secure Environment Mode Option 402
Measurement Category 403
Environmental Conditions 405
Using the 10 MHz reference clock 406
To supply a sample clock to the oscilloscope 406
To synchronize the timebase of two or more instruments 408
To clean the oscilloscope 408
Binary Data (.bin) Format 409
CSV and ASCII XY files 417
Acknowledgements 418
To check warranty and extended services status 418
To return the instrument 419
Contact us 420
                                                           399
15 Reference
About Oscilloscope
               Press [Utility]&Service&About Oscilloscope to display information about your
               oscilloscope model number, serial number, software version, boot version, graphics
               version, and installed licenses.
Installed Licenses
               This line in the About This Oscilloscope dialog contains information about the licenses
               that have been installed on the oscilloscope. For example, it can show:
               • 232 — UART (Universal Asynchronous Receiver/Transmitter) and RS232
                 (Recommended Standard 232) trigger and decode. This license provides trigger and
                 decode capability for many UART protocols including RS232.
               • 553 — MIL-STD 1553 trigger and decode. This license provides trigger and serial
                 decode capability for MIL-STD 1553 buses.
               • ALT — Altera FPGA Dynamic Probe. This license provides internal FPGA visibility and
                 quick instrument setup.
               • AMS — Automotive Serial Decode. This license provides serial decode of CAN and
                 LIN buses.
               • FPG — Xilinx FPGA Dynamic Probe. This license provides internal FPGA visibility and
                 quick instrument setup.
               • FLX — FlexRay trigger and decode (for 4 channel or 4+16 channel models only).
                 Includes mask limit test option (LMT), segmented memory option (SGM), and FlexRay
                 physical layer conformance test application option (FRC).
               • FRC — FlexRay Physical Layer Conformance Test Application.
               • LMT — Mask Limit Test. This license lets you create a mask and test waveforms to
                 determine whether they comply to the mask.
                       • LSS — Low Speed Serial Decode. This license provides serial decode of I2C and SPI
                         buses.
                       • mem8M — Memory Upgrade. This license is always provided with 7000 Series
                         oscilloscopes. It shows the total memory depth (8 Mpts interleaved).
                       • MSO — Mixed Signal Oscilloscope. This license adds 16 digital channels. You do not
                         have to install any hardware. A digital probe cable is provided for connection to your
                         device under test.
                       • PWR— This license lets your oscilloscope communicate with software running on a
                         PC for characterizing power devices.
                       • SEC — Secure Environment mode. This license prevents oscilloscope setup and trace
                         information from being stored in the oscilloscope’s internal memory.
                       • SGM — Segmented memory. This license lets you capture infrequent or burst signals
                         with great resolution by eliminating the capture of your signal’s “dead time.”
                       • SND — I2S serial decode option (for 4 channel or 4+16 channel models only).
Upgrading to an MSO
                       A license can be installed to activate the digital channels of an oscilloscope that was not
                       originally ordered as a mixed-signal oscilloscope (MSO). A mixed signal oscilloscope has
                       analog channels plus 16 time-correlated digital timing channels.
                       To view the currently installed licenses press [Utility]&Options&Licenses&Show
                       license information.
                       For information about upgrading your oscilloscope through licensing, contact your local
                       Agilent Technologies representative or see www.agilent.com/find/7000.
Measurement Category
                 Measurement Category
                       The InfiniiVision oscilloscopes are intended to be used for measurements in
                       Measurement Category I.
                       Use this instrument only for measurements within its specified measurement
    WA R N I N G
                       category.
                   Maximum input voltage at analog inputs and 2-channel external trigger input:
      CAUTION
                   CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
                   CAT II 100 Vrms, 400 Vpk
                   50 input: 5 Vrms
                   Input protection is enabled in 50 mode and the 50 load will disconnect if greater
                   than 5 Vrms is detected. However the inputs could still be damaged, depending on
                   the time constant of the signal. The 50 input protection only functions when the
                   oscilloscope is powered on.
                   1 M ohm input, 100 MHz models:
                   For steady-state sinusoidal waveforms derate at 20 dB/decade above 200 kHz to a
                   minimum of 10 Vpk
                   1 M ohm input, 350 MHz, 500 MHz, 1 GHz models:
                   For steady-state sinusoidal waveforms derate at 20 dB/decade above 57 kHz to a
                   minimum of 5 Vpk
                   With N2863A 10:1 probe: CAT I 600 V, CAT II 300 V (DC + peak AC)
                   With 10073C or 10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
                   Maximum input voltage at digital channels: ±40 V peak CAT I; transient overvoltage
      CAUTION
                   800 Vpk
Environmental Conditions
                   Maximum input voltage at 10 MHz REF connector: Do not apply more than ±15 V at
      CAUTION
                   the 10 MHz REF BNC connector on the rear panel or damage to the instrument may
                   occur.
A black locked padlock icon will appear at the top of the display.
                       If the externally supplied sample clock varies greater than 0.5% from 10 MHz a soft
                       unlock will occur. The oscilloscope will continue to acquire data but the lock symbol in
                       the upper right part of the display will become a red unlocked padlock icon.
                       If the externally supplied sample clock signal is lost, a hard unlock will occur. The lock
                       symbol in the upper right part of the display will become a red unlocked padlock icon and
                       the oscilloscope will stop acquiring data. The oscilloscope will resume sampling when
                       the externally supplied sample clock becomes stable again.
                  Agilent provides the .m files, which need to be copied into the work directory for
                  MATLAB. The default work directory is C:\MATLAB7\work.
                  File Header
                  There is only one file header in a binary file. The file header consists of the following
                  information.
                  Cookie      Two byte characters, AG, that indicate the file is in the Agilent Binary Data file
                  format.
File Size A 32-bit integer that is the number of bytes that are in the file.
                  Number of Waveforms           A 32-bit integer that is the number of waveforms that are
                  stored in the file.
                  Waveform Header
                  It is possible to store more than one waveform in the file, and each waveform stored will
                  have a waveform header. When using segmented memory, each segment is treated as a
                  separate waveform. The waveform header contains information about the type of
                  waveform data that is stored following the waveform data header.
Header Size A 32-bit integer that is the number of bytes in the header.
                  Waveform Type         A 32-bit integer that is the type of waveform stored in the file:
                  • 0 = Unknown.
                  • 1 = Normal.
                  • 2 = Peak Detect.
                  • 3 = Average.
                  • 4 = Not used in InfiniiVision oscilloscopes.
                  • 5 = Not used in InfiniiVision oscilloscopes.
                  • 6 = Logic.
                       Number of Waveform Buffers          A 32-bit integer that is the number of waveform buffers
                       required to read the data.
Points A 32-bit integer that is the number of waveform points in the data.
                       Count A 32-bit integer that is the number of hits at each time bucket in the waveform
                       record when the waveform was created using an acquisition mode like averaging. For
                       example, when averaging, a count of four would mean every waveform data point in the
                       waveform record has been averaged at least four times. The default value is 0.
                       X Display Range      A 32-bit float that is the X-axis duration of the waveform that is
                       displayed. For time domain waveforms, it is the duration of time across the display. If the
                       value is zero then no data has been acquired.
                       X Display Origin A 64-bit double that is the X-axis value at the left edge of the display.
                       For time domain waveforms, it is the time at the start of the display. This value is treated
                       as a double precision 64-bit floating point number. If the value is zero then no data has
                       been acquired.
                       X Increment A 64-bit double that is the duration between data points on the X axis. For
                       time domain waveforms, this is the time between points. If the value is zero then no data
                       has been acquired.
                       X Origin     A 64-bit double that is the X-axis value of the first data point in the data
                       record. For time domain waveforms, it is the time of the first point. This value is treated as
                       a double precision 64-bit floating point number. If the value is zero then no data has been
                       acquired.
                       X Units     A 32-bit integer that identifies the unit of measure for X values in the acquired
                       data:
                       • 0 = Unknown.
                       • 1 = Volts.
                       • 2 = Seconds.
                       • 3 = Constant.
                       • 4 = Amps.
                       • 5 = dB.
                       • 6 = Hz.
               Y Units A 32-bit integer that identifies the unit of measure for Y values in the acquired
               data. The possible values are listed above under “X Units”.
               Frame     A 24 byte character array that is the model number and serial number of the
               oscilloscope in the format of: MODEL#:SERIAL#.
               Waveform Label A 16 byte character array that contains the label assigned to the
               waveform.
               Time Tags  A 64-bit double, only used when saving multiple segments (requires
               segmented memory option). This is the time (in seconds) since the first trigger.
               Segment Index A 32-bit unsigned integer. This is the segment number. Only used
               when saving multiple segments.
               Waveform Data Header Size           A 32-bit integer that is the size of the waveform data
               header.
               Buffer Type A 16-bit short that is the type of waveform data stored in the file:
               • 0 = Unknown data.
               • 1 = Normal 32-bit float data.
               • 2 = Maximum float data.
               • 3 = Minimum float data.
               • 4 = Not used in InfiniiVision oscilloscopes.
               • 5 = Not used in InfiniiVision oscilloscopes.
               • 6 = Digital unsigned 8-bit char data (for digital channels).
Bytes Per Point A 16-bit short that is the number of bytes per data point.
                       Buffer Size A 32-bit integer that is the size of the buffer required to hold the data
                       points.
                        File Header
                                        Number of Waveforms = N
                          12 bytes
                      Waveform Data
                                        Buffer Type = 1 (floating point)
                        Header 1
                                        Bytes per Point = 4
                        12 bytes
                       Voltage Data 1
                         buffer size
                      Waveform Data
                                        Buffer Type = 1 (floating point)
                        Header 2
                                        Bytes per Point = 4
                        12 bytes
                       Voltage Data 2
                         buffer size
                      Waveform Data
                                        Buffer Type = 1 (floating point)
                        Header N
                                        Bytes per Point = 4
                        12 bytes
                      Voltage Data N
                        buffer size
                            File Header
                                             Number of Waveforms = 2
                              12 bytes
                          Waveform Data
                                             Buffer Type = 6 (unsigned char)
                            Header 1
                                             Bytes per Point = 1
                            12 bytes
                          Waveform Data
                                             Buffer Type = 6 (unsigned char)
                            Header 2
                                             Bytes per Point = 1
                            12 bytes
                  File Header
                                   Number of Waveforms = N = Number of Segments
                    12 bytes
                Waveform Data
                                   Buffer Type = 1 (floating point)
                  Header 1
                                   Bytes per Point = 4
                  12 bytes
                 Voltage Data 1
                   buffer size
                Waveform Data
                                   Buffer Type = 1 (floating point)
                  Header 2
                                   Bytes per Point = 4
                  12 bytes
                 Voltage Data 2
                   buffer size
                Waveform Data
                                   Buffer Type = 1 (floating point)
                  Header N
                                   Bytes per Point = 4
                  12 bytes
                 Voltage Data N
                   buffer size
Acknowledgements
               RealVNC is licensed under the GNU General Public License. Copyright (C) 2002-2005
               RealVNC Ltd. All Rights Reserved.
               This is free software; you can redistribute it and/or modify it under the terms of the GNU
               General Public License as published by the Free Software Foundation; either version 2 of
               the License, or (at your option) any later version.
               This software is distributed in the hope that it will be useful, but WITHOUT ANY
               WARRANTY; without even the implied warranty of MERCHANTABILITY or FITNESS FOR
               A PARTICULAR PURPOSE. See the GNU General Public License for more details. The
               license is located on the Agilent InfiniiVision Oscilloscopes Documentation CD-ROM.
               RealVNC source code can be obtained from RealVNC or by contacting Agilent. Agilent
               will charge for the cost of physically performing the source distribution.
Contact us
               Americas
               Canada (877) 894-4414
               Latin America 305 269 7500
               United States (800) 829-4444
               Asia Pacific
               Australia 1 800 629 485
               China 800 810 0189
               Hong Kong 800 938 693
               India 1 800 112 929
               Japan 81 426 56 7832
               Korea 080 769 0800
               Malaysia 1 800 888 848
               Singapore 1 800 375 8100
               Taiwan 0800 047 866
               Thailand 1 800 226 008
               Europe
               Austria 0820 87 44 11
               Belgium 32 (0) 2 404 93 40
               Denmark 45 70 13 15 15
               Finland 358 (0) 10 855 2100
               France 0825 010 700
               Germany 01805 24 6333*
               *0.14€/minute
               Ireland 1890 924 204
               Italy 39 02 92 60 8484
               Netherlands 31 (0) 20 547 2111
               Spain 34 (91) 631 3300
               Sweden 0200-88 22 55
               Switzerland (French)
               44 (21) 8113811 (Opt 2)
               Switzerland (German)
               0800 80 53 53 (Opt 1)
               United Kingdom 44 (0) 7004 666666
               Other European countries:
               www.agilent.com/find/contactus
EEPROM data read, I2C trigger, 117   firmware versions, 287                   hex bus trigger, 139
energy of a pulse, 232               flash drive, 357                         Hex softkey, 139
Entry knob, 53, 176, 360             Flat top window, 238                     HF Reject, 87, 88
event table, 293                     FlexRay frame counter, 304               high-frequency noise rejection, 88
expand about, 60, 187                FlexRay serial decode, 302               high-resolution mode, 265, 266
expand about center, 187             FlexRay totalizer, 304                   holdoff, 90
expand about ground, 187             FlexRay trigger, 108                     horizontal controls, 64
exporting waveform, 247              FLX Option, 400                          horizontal fine adjustment, 65
external memory device, 357          FPG Option, 24, 400                      horizontal Menu/Zoom key, 360
external timebase, 406               FR2 Option, 400                          horizontal position control, 360
external trigger, 92                 frame mode                               horizontal sweep speed control, 360
    input impedance, 93, 94              segmented memory, 271                horizontal time/div control, 360
    probe attenuation, 92            frame trigger, I2C, 118                  host name, 275, 277
    probe settings, 92               Frequency measurement, 200
    probe units, 93                  front panel                              I
Extinction Ratio measurement, 207        5000/6000 Series, 352, 353
eye diagrams, 269                                                             I2C serial decode, 302, 306
                                         6000L Series, 354
                                                                              I2C Signals Menu, 307
                                         7000 Series, 355, 356
                                                                              I2C trigger, 115
F                                        language overlay, 40
                                     front panel keys, 53                     I2S serial decode, 310, 320
f(t), 225                                                                     I2S trigger, 121, 133
                                     FRS Option, 24
factory default configuration, 392                                            idle serial bus, 299, 308, 312, 317, 327,
Fall time measurement, 201                                                          333
feet, 36
                                     G                                        Imped softkey, 61
FFT measurements, 236                g(t), 225                                impedance
FFT window, 238                      gateway IP, 277                              digital probes, 385
file                                 glitch capture, 112                          external trigger, 94
     save, recall, load, 257         glitch trigger, 140                      increment statistics, 196
file browser, save to, 248           golden waveform test, 337                indeterminate state, 216
file explorer, 248, 257              grabber, 373, 374                        indicator
     using, 258                      graphic symbols in menus, 176                LAN status, 361
file format                          graphical user interface languages, 56   indicators, battery-powered
     ASCII, 251                      graphics version, 400                          oscilloscopes, 49
     BIN, 251                        graticule intensity, 181                 infinite persistence, 112, 182, 262
     BMP, 250                        grid intensity, 181                      infrequent triggers, 269
     CSV, 250                        ground level, 60                         input impedance
     PNG, 250                        ground post, 50                              channel input, 61
file keys, 359                       ground wire, 50                              external trigger, 94
file name, new, 249                                                           installed licenses, 400
fine adjustment, channel, 60         H                                        installed options, 287
fine adjustment, sweep speed, 65                                              instantaneous slope of a waveform, 230
firmware updates, 400                Hanning window, 238                      Instrument Utilities web page, 287
firmware upgrade files, 287          HDTV trigger, 156                        Integrate math function, 232
firmware version information, 275    help system, built-in, 55                intensity control, 180, 358
waveform
  cursor tracking, 215
  intensity, 180
  keys, 358
  printing, 244
  recall, 256
  reference point, 187
  save options, 251
  save to scope memory, 256
  save to USB, 255
  saving and exporting, 247
Width - measurement, 202
Width + measurement, 202
Window, FFT, 238
www, 258
X
X at Max Y measurement, 202
X at Max Y on FFT, 199
X at Min Y measurement, 202
X at Min Y on FFT, 199
XY mode, 69, 220
Z
Z-axis blanking, 69, 222
zoom and pan, 179
Zoom mode, 66
zoomed time base measurement
    window, 68