Control Unit
9.3 Control Unit
The most important functions of the Dornier Compact Delta II, Delta II are executed
via the control unit. The control unit is integrated in a housing and connected to the
Dornier Compact Delta II, Delta II via a helix cable.
Figure 9-1 Control Unit of the Dornier Compact Delta II, Delta II with the most important
Elements
1 Display (H1) 8 Coupling pressure (S6, S7)
2 Reset (S1) 9 Movement buttons (S8 to S13)
3 Display (H2) 10 Zero position (S16)
4 Pulse frequency (S2, S3) 11 Shock-wave release button (S17)
5 Display (H3) 12 C-arm CC+ (S14)
6 Shock-wave intensity (S4, S5) 13 C-arm CC- (S15)
7 Display (H4)
CS3.SM_GB_A 9.3
Control Unit
Table 9-1 Elements of Control Unit
Symbol Designation Description
- Display of shock-wave number or error message
H1
Display 1 - Display of the Service Menu
H2 - Display of the type of triggering / fixed trigger
Display 2 frequency
- Display of the Default values of the single
Service Menus
H3 Display of shock-wave intensity
Display 3
H4 Display of coupling pressure
Display 4
S1 - Pressing simultaneously with S4,
Reset Reset of shock-wave counter to zero
- Pressing simultaneously with S16,
switching between isocentric mode, the mode for
image oriented movement and the mode for
normal movement
- Acknowledge of error message
- Switch-over Front/Side Mode
S3 - Pulse frequency (+) switches step-by-step to the
Pulse frequency (+) next higher value,
S2 - Pulse frequency (-) switches step-by-step to the
Pulse frequency (-) next lower value.
- Selection between ECG triggering and fixed
frequency triggering.
Sequence: ECG 60, 70, 80, 90, 100, 110, 120, 150
and 180 SW/min
Press repeatedly to switch from one value to the
next. To switch to next value after 180, press pulse
frequency (-). In ECG mode, press pulse frequency
(+) to select next value.
S5 Shock-wave intensity (+)
Shock-wave intensity (+) increases shock-wave intensity
S4 Shock-wave intensity (-)
Shock-wave intensity (-) reduces shock-wave intensity
S7 Coupling pressure (+) increases the coupling
Coupling pressure (+) pressure.
S6 Coupling pressure (-) reduces coupling pressure.
Coupling pressure (-)
Coupling pressure is displayed with 0 as lowest
value to 5 as highest value on display H4.
9.4 CS3.SM_GB_A
Control Unit
Symbol Designation Description
S15 X-ray C-bow moves in preset position for CC-
Position for CC- locating locating as long as key is pressed. Intermediate
position is PA position. Press key once more to
move to CC- end position.
Release key during movement to enable
intermediate positions.
S14 X-ray C-bow moves in preset position for CC+
+
Position for CC locating locating as long as key is pressed. Intermediate
position is PA position. Press key once more to
move to CC+ end position.
Release key during movement to enable
intermediate positions.
S10 Move therapy head and X-ray C-bow in Z axis.
Z axis (+) Z axis (+) raises,
Z axis (-) lowers therapy head and X-ray C-bow.
With image oriented movement:
movement in direction T+ and T-
S11
Z axis (-)
S12 Move therapy head and X-ray C-bow in Y axis.
Y axis (+) Y axis (+) moves therapy head and X-ray C-bow
towards the unit,
Y axis (-) moves therapy head and X-ray C-bow
towards patient table.
When isocentric mode is activated:
S13 Movement of TKCF axis
Y axis (-)
With image oriented movement:
movement in direction V+ and V-
Move therapy head and X-ray C-bow in X axis.
X axis (-) moves therapy head and X-ray C-bow to
the left (view from the front onto the unit),
S9 S8 X axis (+) moves therapy head and X-ray C-bow
X axis (-) X axis (+) to the right (view from the front onto the unit).
When isocentric mode is activated:
Movement of TKCD axis
With image oriented movement:
movement in direction H+ and H-
CS3.SM_GB_A 9.5
Control Unit
Symbol Designation Description
Move therapy head and X-ray C-bow in any axis
up to set zero position (setup position). Hold key
pressed until zero position (setup position) is
S16 reached.
Zero position,
setup position, (only if a In combination with reset key:
table Dornier Relax + is Switching to the isocentric mode, the mode for
available) image oriented movement and the mode for
normal movement
Releases shock-waves
S17 (shock-wave release continues as long as button is
Shock-wave release pressed)
button
Key reset in combination with key
S1 + S16 zero position :
Switching between isocentric mode, the mode for
image oriented movement and the mode for
normal movement
With isocentric mode:
S9 Turning therapy head in CD axis to 190°.
TKCD (-)
S8 Turning therapy head in CD axis to +10°.
TKCD (+)
S12 With isocentric mode:
TKCF (+)
Move therapy head in CF axis up to
max. +45°.
Move therapy head in CF axis down to
S13 max. -8°.
TKCF (-)
With isocentric mode:
S10 blocked
S11 blocked
9.6 CS3.SM_GB_A
Control Unit
Symbol Designation Description
With image oriented movement:
S9 H- (moving the concrement to the left on the
ultrasound image)
S8 H+ (moving the concrement to the right on the
ultrasound image)
With image oriented movement:
S12 V+ (moving the concrement up on the
ultrasound image)
S13 V- (moving the concrement down on the
ultrasound image)
With image oriented movement:
S10 T+ (moving the concrement along the depth inside
on the ultrasound image)
S11 T- (moving the concrement along the depth
outside on the ultrasound image)
All movement and rotating axes are moved to a
defined position in order to bring the positioning
S16 + S17 into the transport position.
Transport position
H6 Status display
Status LED Off: System O.K.
Green flashing: System O.K. and
isocentric mode activated
Green: System O.K. and
image oriented movement
activated
Orange: Message
Orange flashing: Message,, acknowledge with
Reset
Red: Error Message
Red flashing: Error message, acknowledge
with Reset
CS3.SM_GB_A 9.7
Control Unit
Figure 9-2 Component Placement on PCB M370A
9.8 CS3.SM_GB_A
Control Unit
Figure 9-3 Circuit Diagram M370A
CS3.SM_GB_A 9.9
Control Unit
Figure 9-4 Circuit Diagram M370A
9.10 CS3.SM_GB_A
Control Unit
Figure 9-5 Circuit Diagram M370A
CS3.SM_GB_A 9.11
Control Unit
9.4 Data Interface
The data interface is a serial interface according to RS 422 standard (CCITT/X27 resp.
DIN 66259, part 3). The receiving lines are terminated in a 68 Ohm impedance to
ground.
The interface parameters are the following:
Transmission: asynchronous, full duplex
Transfer rate: 9600 Baud
Character length: 8 bit
Frame: 1 start bit, 1 stop bit, parity even
Handshake: software handshake XON/XOFF
The XOFF signal is sent by the controller as soon as the receiving buffer is almost full.
The receiving buffer has a total depth of 20 bytes.
Table 9-2 Pin Assignment on 1X2 and at the Control Unit
PIN Assignment at Function PIN Assignment at the Control
1X2 Unit
A +24 VDC 1
B GND 2
C T(A) 8
D T(B) 9
E R(A) 10
F R(B) 11
G MovEn 12
H screen screen + green/black
J HandTrig 2 3
K HandTrig 2 4
L HandTrig 1 5
M HandTrig 1 6
9.12 CS3.SM_GB_A