Calibration Labs: Mechanical Mass
Calibration Labs: Mechanical Mass
SPECIFIC CRITERIA
for CALIBRATION LABORATORIES
IN MECHANICAL DISCIPLINE :
MASS (Weights)
MASTER COPY
Reviewed by Approved by
10
1 General Requirements
1.1 Scope 3
1.2 Calibration Measurement Capability(CMC) 3
1.3 Personnel, Qualification and Training 3-4
1.4 Accommodation and Environmental Conditions 4-5
1.5 Special Requirements of Laboratory 5
1.6 Safety Precautions 5-6
1.7 Other Important Points 6
1.8 Proficiency Testing 6
2 Specific Requirements - Calibration of Weights
2.1 Scope 7
2.2 National/ International Standards, References and Guidelines 7
2.3 Metrological Requirements 7-8
2.4 Terms & Definitions 8-9
2.5 Selection of Reference Weights 9
2.6 Selection of Comparator/ Balance 9-10
2.7 Calibration Interval 10
2.8 Legal Aspects 11
2.9 Environmental Conditions 11-12
2.10 Calibration Methods 13
2.11 Determination of Air Density and its Uncertainty 13-15
Equations for Determination of Conventional Mass, True Mass
2.12 15-18
and their Uncertainties
2.13 Measurement Uncertainty 19
2.14 Reporting of Results 19
2.15 Evaluation of CMC 19-20
2.16 Sample Scope 20
2.17 Minimum Requirement for Accreditation 20
2.18 Maximum Permissible Error & Uncertainty Chart 21
Calibration of Newton Weights, Pressure Balance Weights and Non-
2.19 22
Metric Weights
2.20 Key Points 22
• The purpose of this document is to specify requirements with which a laboratory has to operate and
demonstrate its competency to carry out calibration in accordance with ISO/IEC 17025:2005.
• To achieve uniformity between the laboratories, assessors and assessment process in terms of
maximum permissible error, CMC, measurement uncertainty etc in line with National/International
standards.
1.1 Scope
This specific criteria lays down the specific requirements in calibration of weights under mechanical
discipline. This part of the document thus amplifies the specific requirements for calibration of weights
and supplements the requirements of ISO/IEC 17025:2005.
1.2.1 CMC is one of the parameters that is used by NABL to define the scope of an accredited
calibration laboratory, the others being parameter/quantity measured, standard/master used,
calibration method used and measurement range. The CMC is expressed as “the smallest
uncertainty that a laboratory can achieve when calibrating the best existing device”. It is an
expanded uncertainty estimated at a confidence level of approximately 95% corresponding to a
coverage factor k=2.
1.2.2 For evaluation of CMC laboratories should follow NABL 143 - Policy on Calibration and
Measurement Capability (CMC) and Uncertainty in Calibration.
The following are the specific requirements. However, qualification and experience
will not be the only criteria for the required activity. They have to prove their skill,
knowledge and competency in their specific field of calibration activity.
a) B.E / B.Tech or M.Sc. (having Physics as one of the subject) degree with 3
months experience in Basics of Mass Metrology.
b) B.Sc (with Physics as one of the subject) or Diploma with 6 months experience in
Basics of Mass Metrology.
The following are only guidelines. However, qualification and experience will not be
the only criteria for the required activity. They have to prove their skill, knowledge
and competency in analysis and interpretation of calibration results.
a) B.E / B.Tech or M.Sc. (with having Physics as one of the subject) degree with 6
months experience in Mass Metrology.
b) B.Sc. (with Physics as one of the subject) or Diploma with 1 year experience in
Mass Metrology.
Accommodation and environmental conditions adversely affect the results of calibration and
measurement accuracy unless they are controlled and monitored. Hence, they play a very important role.
The influencing parameters may be one or more of the following i. e. temperature, relative humidity,
atmospheric pressure, vibration, acoustic noise, dust, air currents/draft, illumination (wherever
applicable), voltage fluctuations, electrical earthing and direct sunlight etc., depending on the nature of
calibration services provided. The variables described above can play a major factor on calibration
results.
The laboratories are advised to follow the requirement of accommodation and environment depending
on the types of services provided as recommended
• As specified in the National/ International Standards or guidelines followed for the calibration.
The environmental monitoring equipments used should also meet the requirement of manufacturers’
recommendations and specifications as per the relevant standards followed.
If, accommodation and environmental conditions are not specified either by manufacturer or by
National/International standards / guidelines, the laboratory shall follow the below recommendations.
1.4.1 Vibration
The calibration area shall be free from vibrations generated by central air-conditioning plants,
vehicular traffic and other sources to ensure consistent and uniform operational conditions. The
laboratory shall take all special/ protective precautions like mounting of sensitive apparatus on
vibration free tables and pillars etc., isolated from the floor, if necessary.
Acoustic noise level in the laboratory shall be maintained to facilitate proper performance of
calibration work. Noise level shall be maintained less than 60 dBA, wherever it affects
adversely the required accuracy of measurement.
1.4.3 Illumination
The calibration area shall have adequate level of illumination. Where permissible, fluorescent
lighting is preferred to avoid localized heating and temperature drift. The recommended level
of illumination is 250-500 lux on the working table.
The environmental conditions for the activity of the laboratory shall be such as not to adversely
affect the required accuracy of measurement. Facilities shall be provided whenever necessary
for recording temperature, pressure and humidity values prevailing during calibration. The
atmospheric conditions maintained in the laboratory during calibration shall be reported in the
calibration report/ certificate.
1.5.1 The calibration laboratory shall make arrangements for regulated and uninterrupted power
supply of proper rating. The recommended voltage regulation level is ±2% or better, and
Frequency variation ± 2.5Hz or better on the calibration bench.
1.5.2 The reference standards shall be maintained at temperatures specified for their maintenance on
order to ensure their conformance to the required level of operation.
1.5.3 The laboratory shall take adequate measures against dust and external air pressure.
1.6.1 Relevant fire extinguishing equipment for possible fire hazards, shall be available in the
corridors or convenient places in the laboratory. Adequate safety measures against electrical,
chemical fire hazards must be available at the work place. Laboratory rooms/ areas where
1.6.2 Specification SP 31- 1986, a special publication in the form of a wall chart, giving the method
of treatment in case of electric shock, should be followed. The chart shall be placed near the
power supply switchgear and at other prominent places as prescribed under Indian Electricity
Rules 1956.
1.6.3 Effective mains earthing shall be provided in accordance with relevant specification IS: 3043.
This shall be periodically checked to ensure proper contact with earth rod.
1.7.1. Entry to the Calibration Area: As far as possible, only the staff engaged in the calibration
activity shall be permitted entry inside the calibration area.
1.7.2. Space in Calibration Area: The calibration Laboratory shall ensure adequate space for
calibration activity without adversely effecting the results.
To give further assurance to the accuracy or Uncertainty of measurements, a laboratory will be required
to participate, from time to time, in Proficiency Testing Program. The laboratory shall remain prepared
to participate in the Proficiency Testing Program through inter-laboratory, inter-comparison schemes
wherever it is technically feasible. (Ref. NABL 162, 163 and 164 for further details)
Note 1: Newton weights, non-metric weights can also be calibrated to accuracy class equivalent to
OIML R111 -1. However, the conventional mass values and its uncertainty should be given in
SI units.
Note 2: This technical requirement is based on the above mentioned guideline. Lab may follow
any relevant standard, however care shall be taken to follow the requirements in totality.
Note 3: ASTM standard weights can also be calibrated if the relevant standard is followed in total.
Note 4: Laboratory shall apply for calibration and not for verification of Weights. Verification may
require approval from Dept. of Legal Metrology, Regulatory Bodies, etc.
• ASTM E617 - 13 Standard Specifications for Laboratory Weights and Precision Mass Standards.
• OIML R 47 Edition 1079(E) - Standard weights for testing of high capacity weighing machines.
2.3.1 For Each weight, the expanded uncertainty, U, for k=2, of the conventional mass, shall be less
than or equal to one third of the maximum permissible error.
2.3.2 For each weight, the conventional mass, m c (determined with an expanded uncertainty, U,
according to 5.2 of OIML R-111-1) shall not differ from the nominal value of the weights, m0
by more than the maximum permissible error, δm minus the expanded uncertainty.
m 0 – (δm-U) ≤ m c ≤ m 0 + (δm-U)
2.3.3 For class E 1 and E 2 weights, which are always accompanied by certificates giving the
appropriate data, the deviation from the nominal value, m c -m 0 , shall be taken into account by
the user.
2.3.4 Calibration certificate shall state, as a minimum: the conventional mass of each weight, mc an
indication of whether a weight has been adjusted prior to calibration, its expanded uncertainty
U and the values of the coverage factor k.
2.3.6 The certificate for class E 2 weights shall state, as a minimum, the value of conventional mass
m c, of each weight, the expanded uncertainty, U, and the coverage factor k.
2.3.7 Altitude and corresponding changes in air density can affect the measurement error when
using the conventional mass of weight; therefore, the buoyancy correction shall be used,
which requires the density of the weight to be known. If class E weights are to be used above
330 m, the density of the weights shall be provided along with their associated uncertainty for
Class F1 the same is true above 800 m. Otherwise, the manufacturer shall take the lowered
buoyancy affect at higher altitude in to consideration in specifying the weights class for the
standards of conventional mass.
Accuracy Class
• Class designation of a Weight or Weight set which meets certain metrological requirements intended
to maintain the mass values within the specified limit.
Buoyancy Correction
• A buoyancy correction is the correction applied when Weights of different densities are compared
with each other during the calibration process, the buoyancy being a result of the upward force when
the weight is immersed in a fluid / air during the weighing process.
True Mass
• The true mass of a body relates to the amount of material it contains. The prefix true is added to the
word mass where it is important to make it clear that a particular mass being considered is not a
conventional mass value and it is important to avoid potential ambiguity. The International prototype
kilogram, on which the International mass scale is realized, is defined as a true mass of exactly 1
kilogram. Most high accuracy comparisons are performed on a true mass basis, and converted to
conventional mass when quoted on a certificate.
• The concept of true mass (mass in vacuum) might be appropriately described as a theoretical
comparison of a mass against a reference standard mass (with a known value) on a equal arm
balance inside a vacuum chamber. Although this scenario is impossible to achieve, one can easily
understand that such an arrangement, were it possible, would remove any influence on the weighing
process from the buoyant effect of air.
Discriminations / Sensitivity
• Sensitivity is the smallest change in mass that can be detected by the weighing Instrument. For
practical purpose discrimination is synonymous with readability. Discrimination is also synonymous
with the sensitivity of an analog weighing Instrument.
Resolution
• The readability expressed as a portion of the capacity. For example a weighing Instrument with a
capacity of 3000g and a readability of 0.1g has a resolution of 1 part in 30000.
Scale Interval
• The value expressed in units of Mass of in the case of analog indication, the difference between the
values that correspond to two scale marks. In the case of digital Indication, the difference between
two consecutively indicated values.
Weight
• A material measure of mass, regulated in regard to its physical and metrological characteristics:
Shape, Mass, material, surface quality, nominal value and maximum permissible error.
The reference weight shall generally be of a higher class of accuracy than the weight to be calibrated. In
the calibration of weights of class E 1 , the reference weight shall have similar or better metrological
characteristics (magnetic properties, surface roughness) than the weights to be calibrated [refer C.2.3 of
OIML R 111-1, 2004 for more details].
E1 √ √ √ √ √ √ √
E2 - - √ √ √ √ √
F1 - - - √ √ √ √
F2 - - - - √ √ √
Note: Standard weights of each class shall have the technical requirements such as shape,
construction, material, marking etc as per OIML R111. M1, M2, and M3 class weights are not
recommended to be used as a reference for the calibration of weights. However, M1 weights can
be used for above 20kg with coarser uncertainty.
On the basis of the accuracy class, a mass comparator is to be selected in such a way that its uncertainty
component is balanced in proportion to the overall uncertainty of the weighing result. The most
important uncertainty component of a mass comparator is calculated from its standard deviation. The
specification of the manufacturer can be selected as a first approximation for the value of a standard
deviation. It must be taken into account. However, that this indication is decisive for the smallest
nominal value. It should therefore not exceed an amount of 30% of the combined standard uncertainty
u1 (k=2).
Note: It is not recommended to calibrate a higher accuracy class weight with a lower accuracy class of
reference weight and comparator/balance with coarser resolution without the consent of the
customer.
2.6.2 Selection of comparator balance for calibration of weights depending on class of accuracy:
E1 E2 F1 F2 M1 M2 M3
Nominal Value
Standard Deviation of Repeatability in mg
5000 kg 2778 8889 27778 88889 277778
2000 kg 1111 3333 11111 33333 111111
1000 kg 178 556 1778 5556 17778 55556
500 kg 89 278 889 2778 8889 27778
200 kg 33 111 333 1111 3333 11111
100 kg 17.8 56 178 556 1778 5556
50 kg 2.78 8.9 28 89 278 889 2778
20 kg 1.11 3.3 11 33 111 333 1111
10 kg 0.56 1.78 6 18 56 178 556
5 kg 0.28 0.89 2.8 8.9 28 89 278
2 kg 0.11 0.33 1.11 3.3 11.1 33 111
1 kg 0.056 0.178 0.556 1.78 5.56 17.8 56
500 g 0.028 0.089 0.278 0.89 2.78 8.89 28
200 g 0.011 0.033 0.111 0.33 1.11 3.33 11.1
100 g 0.006 0.018 0.056 0.18 0.556 1.78 5.56
50 g 0.0033 0.011 0.034 0.11 0.333 1.11 3.33
20 g 0.0028 0.009 0.028 0.089 0.278 0.889 2.78
10 g 0.0022 0.007 0.022 0.067 0.222 0.667 2.22
5 g 0.0018 0.006 0.018 0.056 0.178 0.556 1.78
2 g 0.0013 0.004 0.013 0.044 0.133 0.444 1.33
1 g 0.0011 0.0034 0.011 0.033 0.111 0.333 1.11
500 mg 0.0009 0.0028 0.009 0.028 0.089 0.278 8.89
200 mg 0.0007 0.0022 0.007 0.022 0.067 0.222
100 mg 0.0006 0.0018 0.006 0.018 0.056 0.178
50 mg 0.0004 0.0013 0.004 0.013 0.044
20 mg 0.0003 0.0011 0.0034 0.011 0.034
10 mg 0.0003 0.0009 0.0028 0.009 0.028
5 mg 0.0003 0.0007 0.0022 0.006 0.022
2 mg 0.0003 0.0007 0.0022 0.006 0.022
1 mg 0.0003 0.0007 0.0022 0.006 0.022
For the reference Weights and comparators used in calibration of Weights at permanent lab facility
Note: Based on the historical data validity of reference weights may be extended upto 5 years for E 1 .
Calibration of weights done by any accredited laboratories is meant for scientific and industrial purpose
only. However, if used for commercial trading, additional recognition/ approval shall be complied as
required by Dept. of Legal metrology, Regulatory bodies, etc. This should be clearly mentioned in the
calibration certificate issued to the customer.
2.9.1.2 Accuracy or reliability of weighing results is closely connected with the place where,
mass comparators are installed, and also with the weights used, with the measuring
room conditions and operator's skill. The place of installation (measuring room) for
mass comparators shall be designed in such a way that, the disturbances of the
environment do not affect the result. Manufacturer’s recommendation shall be
considered.
2.9.2 The calibration of weights shall be performed at suitable conditions under ambient atmospheric
pressure at temperatures closer to room temperature (1) Typical recommended values are given
below:
Note (1): It is also important that the difference in temperature between the weights and the air
inside the mass comparator is as small as possible. Keeping the reference weight and
the test weight inside the mass comparator before and during the calibration to
reduce the temperature difference.
Note (2): This is the change in the temperature of the laboratory. Thermal stabilization of
balances and weights also requires an appropriate temperature stability of laboratory
for 24 hours before calibration.
Note (3): The upper limit is mainly important when storing the weight.For E1 and E2 class
weights, the temperature should be within ˚C 18
to 27˚C. The environmental
conditions shall be within the specifications of the weighing instruments.
However, laboratory shall evaluate the requirement of accuracy, resolution and uncertainty
depending on the CMC aimed at.
• Prior to performing any calibration, the weights need to be acclimated to the ambient
conditions of the laboratory. In particular, weights of classes E 1 , E 2 and F 1 shall be
close to the temperature in the weighing area.
There are two methods for determination of conventional mass of weights in a weight set.
Note: Method used for calibration should be clearly mentioned in the calibration certificate
issued to the customer.
2.11.1 In 2008 CIPM recommended that the following equation be used to determine ρa the density
of Air
The mole fraction of water vapor, xv that is a function of the relative humidity, rh
or dew-point temperature tr, an enhancement factor, f and the moist air saturation
vapor pressure, psv is given as follows:
2.11.1.2 The moist air saturation vapor pressure, psv can be calculated using the following
equation:
f = α + βp + γt2
2.11.1.5 Table: Recommended Values for Constants of Psv (A, B, C, D), f (α, β, γ) and
Z (a 0 , a 1 , a 2 , b 0 , b 1 , c 0 , c 1 , d, e)
Psv f Z
1991 1991 1991
Constant Recommended Unit Constant Recommended Unit Constant Recommended Unit
value value value
A 1.2378847 × 10-5 K-2 α 1.00062 a0 1.58123 × 10-6 KPa-1
B -1.9121316 × 10-2 K-1 β 3.14 × 10-8 Pa-1 a1 -2.9331 × 10-8 Pa-1
C 33.93711047 γ 5.60 × 10 -7
K -2
a2 1.1043 × 10-10 K-1Pa-1
D -6.3431645 × 103 K b0 5.707 × 10-6 KPa-1
b1 -2.051 × 10-8 Pa-1
c0 1.9898 × 10-4 KPa-1
c1 -2.376 × 10-6 Pa-1
d 1.83 × 10-11 K2Pa-2
e -7.65 × 10-8 K2Pa-2
2.11.2 Approximation Formula as per OIML R-111-1: 2004 (Page No. 76)
ρa = 0.34848p -0.009*h*exp(0.061*t)
273.15+t (E-3.1 OIML)
2.11.3 For class E1 weights, the density of air should always be determined based o corresponding
measurements. However, the following approximation equation is a way to estimate air
density at laboratories that have no means of determining the air density at the site. The height
above sea level is always known. Therefore, if the air density is not measured, it should be
calculated as a mean value for the laboratory site as follows:
-ρ 0
ρa = ρ 0 * exp *gh
P0
Where p 0 = 101325 Pa
ρ 0 = 1.2 kg/m3
g = 9.81 m/s 2
h = height above sea level expressed in meter.
2.12 Equations for Determination of Conventional Mass, True mass and their uncertainties
2.12.1 Determination of Conventional Mass with reference to volume of the reference and test
Weights
m ct =[m cr + (V t - V r )(ρ a -ρ 0 )+ ∆m w ]
Where,
V t = Volume of the test weight
V r = Volume of the reference weight
m ct = Conventional mass of test weight
m cr = Conventional mass of reference weight
ρ a = Density of Air during calibration in kg/m3
ρ 0 = Density of Air (Conventional) 1.2 kg/m3
∆m w = difference in mass observed using ABBA weighing cycle.
Note: Vt and Vr can be calculated from density and the nominal mass of the
weights.
From the equation of the Clause 8.12.1.1, the term air buoyancy correction (B.C.)
is:
2.12.1.3 Equation for Determining the Uncertainty due to Air Buoyancy Correction
u'b2 = (V t - V r )2*uρ a 2+ (uv t 2 + uv r 2)*(ρ a-
ρ 0 )2
Note: If, the volume is determined from known density and mass values and
their uncertainties, uncertainty of volume (uV t ) can be calculated by the
relation uV t = uρ t /ρ τ ∗V t for test weight and similarly for the reference
weight.
2.12.2 Determination of Conventional Mass with reference to Density of the Reference and
Test Weights
m ct = m cr (1+C)+∆m w
Where,
m ct = Conventional mass of test weight
m cr = Conventional mass of reference weight
C= Buoyancy correction factor
∆m w = Difference in weight by ABBA method
Where,
ρr = Density of Reference Weight in kg/m3
ρt = Density of Test Weight in kg/m3
ρa = Density of Air during Calibration in kg/m3
ρ0 = Density of Air (Conventional) 1.2 kg/m3
Where,
u' b = Uncertainty of air buoyancy correction of conventional mass
m cr = Conventional mass of reference weight in kg
ρr = Density of reference weight in kg/m3
ρt = Density of test weight in kg/m3
National Accreditation Board for Testing and Calibration Laboratories
Doc. No: NABL 122-02 Specific Criteria for Calibration Laboratories in Mechanical Discipline – Mass (Weights)
Issue No: 05 Issue Date: 12.08.2014 Amend No: 00 Amend Date: - Page No: 16 of 22
ρa = Density of Air during calibration in kg/m3
ρ a1 = Density of Air during earlier calibration of reference wt in kg/m3
uρ r = Uncertainty in Density of reference weight in kg/m3
uρ t = Uncertainty in Density of test weight in kg/m3
uρ a = Uncertainty in Density of air in kg/m3
ρ0 = Density of air (conventional) 1.2 kg/m3
- If the air density is not measured and the average air density for the site is
used, than the uncertainty for the air density is to be estimated as:
u(pa)= 0.12/√3.
- At sea level the density of air shall be assumed to be 1.2.kg per m3.
- For class E weights, the density of air should be determined. Its uncertainty is
usually estimated from the uncertainties for temperature, pressure, and air
humidity. For class E 1 , the CIPM formula or an approximation can be used for
the calculation of air density.
Note: Where the air density is 1.2 kg/m3 within ±10%, conventional mass value
shall be used in calculation and true mass shall be calculated from
conventional mass (OIML R- 111 C.2.1.2).
2.12.3 True Mass with reference to Volume of the Reference and Test Weights
m tt = [m cr + (V t -V r )(ρ a )+ ∆m]
2.12.3.2 Equation for Determining the Uncertainty due to Air Buoyancy Correction
Note: If, the volume is determined from known density and mass values and
their uncertainties, uncertainty of volume (uV t ) can be calculated by
the relation uV t = uρ t /ρ τ ∗V t for test weight and similarly for the
reference weight.
2.12.4 True Mass with reference to Density of the Reference and Test Weights
M tt = m tr (1+B)+∆m w
2.12.4.2 Buoyancy Correction Factor for Calibration of Mass on True Mass basis
B = [(ρ a )*(1/ρ t )-(1/ρ r )]
Where,
ρr = Density of reference weight in kg/m3
ρt = Density of test weight in kg/m3
ρa = Density of air in kg/m3
Where,
ub= Uncertainty in buoyancy correction in true mass
mr = Mass of reference weight in kg
ρr = Density of reference weight in kg/m3
ρt = Density of test weight in kg/m3
ρa = Density of air in kg/m3
uρ r = Uncertainty in density of reference weight in kg/m3
uρ t = Uncertainty in density of test weight in kg/m3
uρ a = Uncertainty in density of air in kg/m3
Note 1: As per the standard if the air density deviates from 1.2 kg/m3 by more
than 10%, true mass values shall be used for calculations and the
conventional mass shall be calculated from the true mass.
Note 2: True mass is required for the realization of force, pressure and torque
etc by using dead weights.
Note 3: Calibration of weights in terms of Newton, bar, pascal etc (local "g"
value shall be known to be sufficient accuracy).
1-ρ 0 /ρ c
mt = mc x
1-ρ 0 /ρ
Where,
mt = True mass
mc= Conventional mass
ρ0 = Density of conventional Air (1.2) in kg/m3
ρc = Conventional density of Mass (8000) in kg/m3
ρ= Density of mass in kg/m3
The estimation of the uncertainty measurement for the weight calibrated by a laboratory shall
consider at least the following contributions.
u c = √ (u 1 2+ u 2 2+ u 3 2+ u 4 2+ u 5 2)
Where,
u 1 = standard uncertainty associated with standard deviation of weighing
u 2 = Standard uncertainty associated with air buoyancy correction
u 3 = standard uncertainty associated with drift of the reference standard weight
u 4 = standard uncertainty associated with resolution of the comparator/balance
u 5 = Standard uncertainty associated with reference standard weight
Expanded uncertainty U = k x u c
The calibration certificates issued to the customer shall be in accordance with clause 5.10 of
ISO/IEC/17025:2005. Apart from that it shall also include the following:
2.15.2 CMC value is not the same as expanded uncertainty reported in the calibration
Certificate/Report. CMC values exclude the uncertainties which are attributed to the DUC
(Device under calibration).
National Accreditation Board for Testing and Calibration Laboratories
Doc. No: NABL 122-02 Specific Criteria for Calibration Laboratories in Mechanical Discipline – Mass (Weights)
Issue No: 05 Issue Date: 12.08.2014 Amend No: 00 Amend Date: - Page No: 19 of 22
2.15.3 For the purpose of CMC evaluation the following components should be considered:
Discipline: Mechanical
* Only for Electro-technical discipline; scope shall be recommended parameter wise (where applicable) and the
ranges may be mentioned frequency wise.
** NABL 143 shall be referred for the recommendation of CMC
+
Remarks shall also include whether the same scope is applicable for site calibration as well. NABL 130 shall be
referred while recommending the scope for site calibration.
Signature, Date Signature, Date & Name of Assessor(s) Signature, Date & Name of Lead Assessor
& Name of Lab
Representative
2.17.1 Laboratory shall have minimum F2 class of reference Weights for the calibration below 20
kg along with appropriate comparator / balance. However, M1 class reference weights may
be used for calibration above 20 kg.
2.17.2 Demonstration of any CMC values doesn't automatically qualify for granting accreditation
until the lab satisfies the stipulated requirement given above.
2.18.1 Permissible error for classification weights as per OIML R-111-1, 2004.
Note: For each weight expanded uncertainty U, for k=2 the conventional mass, shall be as per the standard less than or equal to one-
third of the maximum permissible error (U = 1/3 δm)
Class E1 E2 F1 F2 M1 M2 M3
Nominal Error Uncertainty Error Uncertainty Error Uncertainty Error Uncertainty Error Uncertainty Error Uncertainty Error Uncertainty
Value ± mg ± mg ± mg ± mg ± mg ± mg ± mg
5000 kg 25000 8333 80000 26667 250000 83333 800000 266667 2500000 833333
2000 kg 10000 3333 30000 10000 100000 33333 300000 100000 1000000 333333
1000 kg 1600 533 5000 1667 16000 5333 50000 16667 160000 53333 500000 166667
500 kg 800 267 2500 833 8000 2667 25000 8333 80000 26667 250000 83333
200 kg 300 100 1000 333 3000 1000 10000 3333 30000 10000 100000 33333
100 kg 160 53 500 167 1600 533 5000 1667 16000 5333 50000 16667
50 kg 25 8.3 80 27 250 83 800 267 2500 833 8000 2667 25000 8333
20 kg 10 3.3 30 10 100 33 300 100 1000 333 3000 1000 10000 3333
10 kg 5.0 1.7 16 5 50 17 160 53 500 167 1600 533 5000 1667
5 kg 2.5 0.8 8 3 25 8 80 27 250 83 800 267 2500 833
2 kg 1.0 0.3 3 1.0 10 3 30 10 100 33 300 100 1000 333
1 kg 0.5 0.2 1.6 0.5 5 2 16 5 50 17 160 53 500 167
500 g 0.25 0.08 0.8 0.3 2.5 1 8.0 3 25 8 80 27 250 83
200 g 0.10 0.03 0.3 0.10 1.0 0.3 3.0 1 10 3 30 10 100 33
100 g 0.05 0.02 0.16 0.05 0.5 0.2 1.6 0.5 5 2 16 5 50 17
50 g 0.03 0.01 0.10 0.03 0.3 0.1 1.0 0.3 3 1 10 3 30 10
20 g 0.025 0.01 0.08 0.03 0.25 0.1 0.8 0.3 2.5 1 8 3 25 8
10 g 0.020 0.01 0.06 0.02 0.20 0.1 0.6 0.2 2 1 6 2 20 7
5 g 0.016 0.01 0.05 0.02 0.16 0.1 0.5 0.2 1.6 1 5 2 16 5
2 g 0.012 0.004 0.04 0.01 0.12 0.04 0.4 0.1 1.2 0.4 4 1 12 4
1 g 0.010 0.003 0.03 0.01 0.10 0.03 0.3 0.1 1.0 0.3 3 1 10 3
500 g 0.008 0.003 0.025 0.008 0.08 0.03 0.25 0.1 0.8 0.3 2.5 1
200 g 0.006 0.002 0.020 0.007 0.06 0.02 0.20 0.1 0.6 0.2 2.0 1
100 mg 0.005 0.002 0.016 0.005 0.05 0.02 0.16 0.1 0.5 0.2 1.6 1
50 mg 0.004 0.001 0.012 0.004 0.04 0.01 0.12 0.04 0.4 0.1
20 mg 0.003 0.001 0.010 0.003 0.03 0.01 0.10 0.03 0.3 0.1
10 mg 0.003 0.001 0.008 0.003 0.025 0.01 0.08 0.03 0.25 0.1
5 mg 0.003 0.001 0.006 0.002 0.020 0.01 0.06 0.02 0.20 0.1
2 mg 0.003 0.001 0.006 0.002 0.020 0.01 0.06 0.02 0.20 0.1
1 mg 0.003 0.001 0.006 0.002 0.020 0.01 0.06 0.02 0.20 0.1
2.19.1 Weights used for realization of Pressure in Dead weight pressure balance or weights used for
realization of force in Newton are to be calibrated on true mass basis.
2.19.2 If, weights are calibrated on conventional mass basis equation for conversion from
conventional mass to true mass to be mentioned to enable the user to apply appropriate
buoyancy correction.
2.19.3 Newton or Force weights are typically of a slotted design or with a centre hole and are
typically marked with a nominal Force in Newton. Force is calculated with respect to Local
gravity ‘g L ’ during calibration using the formula given below:
Where
F = Force in Newton
m = True Mass in Kg
ρ a = air density in kg/m3
ρ m = density of weights in kg/m3
g L = Local gravity in m/sec2 (value of ‘g’ at the customer’s site )
The force values can be converted either to the standard ‘g’ value or to the customer’s ‘g’
value using the formula given below:
2.19.4 When the customer requires the force weight with respect to his local ‘g’ value he has to
provide the same with uncertainty. Then the force value shall be calculated using the local
‘g’ value and declare in the certificate in terms mass value along with the calculated value in
Newton. ‘g’ value of the calibration laboratory shall also be known to sufficient accuracy.
2.19.5 The Laboratory may calibrate weights of non-metric units (e.g. Pound or Ounce etc.)
However, the results shall be reported in SI units like kg, g, mg, etc. along with the
calculated equivalent value in the non-metric unit or mention the conversion factor to be
used.
Demonstration of any CMC values doesn't automatically qualify for granting accreditation until the
lab satisfies the stipulated requirement given in this document.