Water Meter Data
Water Meter Data
Abstract
Flow sensors based on the principle of electromagnetic induction were investigated as
alternatives to commonly used mechanical devices utilizing rotors and propellers. Proto-
type sensors were constructed showing considerable promise. Measurement accuracy in
excess of 1 cm see-l seems feasible with devices suited to long term battery operation. The
inertial effects and many of the reliability problems inherent in moving part devices would
be overcome by USC of an electromagnetic sensor.
Ocean currents have been routinely mca- field (E = V X B ) . The magnetic field pro-
sured by differences in potential caused by duced by a coil depends on the electrical
the clcctrically conductive seawater moving current and the number of turns. The
through the vertical component of the power supplied to the coil varies with 12.
Earth’s magnetic field (von Arx 1950). Thus, if the sensitivity of a coil is to bc in-
Blood flow in animals is also determined by creased lo-fold, the coil power must go up
electromagnetic induction, the magnetic 100 times. Large coils are efficient, but
field being produced by a coil placed next sensitivity increases with the square root of
to the blood vcsscl. The method is also used linear dimensions, while volume varies with
industrially for monitoring flow in pipes. the cube. Constructional simplicity is there-
But the use of a locally generated field has fore a useful criterion for choosing coil size.
found only occasional application in ocean- Coils of IO-cm diameter and l-cm2 cross
ography (Olson 1972; Tucker 1972; Bow- section are convenient. With a DC power
den and Fairbairn 1956). The continually of 1 W the field from such a coil reacts with
disappointing performance of mechanical a water flow of 1 m set-l to produce a po-
current meters has motivated us to take an tential difference of about 100 PV at the
independent look at the possibilities of elec- electrodes. If 10 mW of power are used,
tromagnetic flow sensors. Here WC describe the electrode output will still be 10 pV.
first the theory involved in the design of Such potentials are readily measured.
such a sensor. Then we show what results It is impossible to get two electrodes to
one can get in an actual device. Finally remain within a few microvolts of potential
we discuss potential uses, such as current difference because of chemical uncertainty
meters, to which this kind of flow sensor at the electrode surface. They can easily
can be applied. have a static offset 1,000 times this large.
If, however, the magnetic field is periodi-
Theory cally reversed, the polarity of the electrode
The flow sensor ( Fig. 1) contains a cop- potential due to water flow will change,
per coil to produce a magnetic Eicld. A set while the static offset remains constant. The
of electrodes measures the voltage gradient electrode voltages can be detected in syn-
across the face of the coil which is induced chrony with the field reversals. The magni-
in the water when it flows through the field. tude of the change is a function of the ve-
Faraday’s principle of induction states that locity of the flow. Electrode errors caused
the voltage field induced is the vector by electrochemical effects and input am-
product of the velocity and the magnetic plificr drift are almost entirely canceled.
---- It is ncccssary only that these offsets do not
1 Contribution No. 3448 from the Woods Hole change during one complete cycle of the
Oceanographic Institution. This work was SUP- field. Further, reversing the field means
ported by National Science Foundation grant
GA-31987. that we can use AC amplifiers; large gains
LIMNOLOGY AND OCEANOGRAPHY 174 MARCH 1975, V. 20(2)
Electromagnetic flow sensors 175
I
creases at lower frequencies, but remains
fairly constant above about 100 Hz. This,
as well as fast response, mandates a high
ELECTRODE,COVER
field reversal rate.
SENSOR ELECTRODES
Fig. 2. Block
El--
diagram
OSCILLATOR
4 X FIELD
FREQUENCY
of the flow
LOGIC
sensor electronics.
Fig. 3. Photograph of the complete flow sensor system with which the mcasurcments cited in this
articli were mad& -
vice is in use to monitor lateral as well as which make them worthy of study. These
fore and aft ship movements (Tucker et al. systems seem to 11s conceptually similar to
1970). e.m. devices. The choice between e.m. and
8. A determination of flow is completed acoustic flow sensors may rest on a balanc-
for each cycle of field reversal. It is possi- ing of engineering considerations not yet
ble to integrate the detector output in syn- apparent.
chrony with the field so as to eliminate any Errors-The device creates the magnetic
need to average large numbers of field cy- conditions in water which generate a flow
cles. The time of integration still sets the signal in the form of a square-wave voltage
bandwidth, however, so that time acuity at the electrodes. This is amplified and de-
and sensitivity must be balanced against tected to provide a DC level, the sign and
one another. For flows above 5 cm set-1 magnitude of which are a function of flow
one can see velocity detail with a time velocity. There are three kinds of errors
acuity down to at least 0.04 see; this may affecting the accuracy of this process: off-
make the device useful in turbulence stud- set errors or baseline drift, scale errors (i.e.
ies. sensitivity changes ) , and the catastrophic
Hydrodynamic properties-The flow prop- errors which involve dynamic limitation or
erties of the e.m. sensor may limit its use in device malfunction, and which usually re-
some applications. But a number of varia- sult in wildly inaccurate readings, or no
tions are possible to minimize such restric- readings. These errors are discussed below.
tions. For instance, to the extent that the Baseline clrift-Our device uses an AC
water next to the sensor is retarded by skin amplifier to magnify changing voltages to
friction, the indicated flow will be lower some arbitrarily large level; these relatively
than that in the bulk of the water. This can large changes are then demodulated to pro-
be overcome in part by making the sensor duce a DC level equal to the magnitude of
surface concave or by shaping the sensor as the change. Since demodulation must oc-
a toroid rather than a disk. The cutout cur at a finite level, the DC drift of the out-
center would allow free passage of the wa- put amplifier will affect zero stability. In
ter past the sensor. A toroidal shape would this instrument, a lm set-l current produced
probably be less sensitive to tilt and to a about 700 mV of DC output. The stability
biased indication of current flow resulting of the DC amplifier is about 5 PV deg-‘.
from any vertical pumping motion of the This implies that temperature effects from
mooring. this source are less than 1 X 10m5m see-l
Another sensor configuration would in- deg-l. Aging effects with DC amplifiers
volvc using two parallel field coils about a are such that drift is typically 15 PV
diameter apart. This would result in fairly month-l and this contributes further to
uniform distribution of magnetic field drift.
within the open “cylinder” defined by the A second source of zero drift exists in
two toroids. The struts connecting the long term unidirectional changes in the
toroids would support the electrodes. This electrode potentials. If the electrode differ-
would give a flow determination of the wa- ence voltage changes at sonic consistent
ter mass between the coils; very little dis- rate, an output will result that will look like
turbance by the sensor of this flow would a signal. For example, if the electrodes
be anticipated. Olson (1972) has reported change at a 1 PV set-l rate, the equivalent
good results with this double coil arrange- signal will be l/(2 x field frequency) times
ment. this. In the prototype, this would represent
Acoustic flow sensors-Flow sensors an error of 0.015 cm sccl. We find that the
based on various sound propagation tech- Ag-AgCl electrodes are within a millivolt
niques (pulse time-of-flight, continuous- of each other over long periods, so the ac-
wave phase-shift, doppler) have hydro- cumulated error from this source is appar-
dynamic properties and reliability potentials ently very small. If the amplifier is not
Electromagnetic flow sensors 179
continuously powered, howcvcr, as might DC level close to zero, Flows past the sen-
be the case in devices intended for inter- sor in excess of some limit will cause non-
mittent sampling, bias currents from the linear amplification of the electrode diffcr-
input operational amplifiers can cause sig- ence voltage, This will be indicated as a
nificant errors as the electrodes adjust to a steady DC output far in excess of nominal
new equilibrium. This effect does not ap- scale. In practice this is no problem.
pear to be significant after about 15 set for
stabilization, Potential uses of the e.m. sensor
A third source of baseline error arises
from electrical lcakagc between the field Moored current meter-The e.m. flow
sensor seems to bc an appropriate device
coil and the water. If this resistance is not
for use in a moored current meter. Several
extremely high (or perfectly balanced with
respect to the electrodes, which is very un- configurations are possible, some of which
likely) the field drive voltage will divide could bc adaptations of existing instru-
across this leakage resistance and the elec- mcnts. For example, an e.m. sensor with a
single electrode set could substitute for the
trode plus water path resistance, causing
mechanical rotor in any instrument that
a signal which produces offset. Changes
in this resistance are likely, so that the off- lines up with the current. An output con-
set cannot be permanently nullcd. A leak- sisting of magnitude and direction results.
age of 10” ohms will cause an offset on the The resultant polar form cannot be aver-
order of 1 cm sec- l, We insulated the coil aged directly, but the system would be suit-
carefully and did not identify leakage as an able for stable moorings and slowly chang-
error source in the sensor described here, ing currents.
although some earlier and less careful ef- Another approach would involve a two-
forts resulted in large errors. axes e.m. sensor. The orthogonal outputs of
Scale errors-Constant scale factor dc- such a device can be directly averaged.
pends on stable coil current and geometry Such a device could cope with mooring
and unchanging gain in the sense electron- mo,vemcnts on a time scale similar to the
ics and detector. Coil current is established averaging period.
with reference to a zener diode and has a If bearing information is available as sine
temperature coefficient of less than 0.02% and cosine of compass angle ( as it would be
per deg C. Most of the sources OFbaseline with a magnetomctcr compass), the orthog-
drift have some small dependence on field onal components from the e,m. sensor can
drive level, so that care in regard to, field be multiplied by the bearing information
stability may be important even though pre- and the result averaged and stored. This
cision in measurement of magnitude may might be the most straightforward and re-
not be. liable approach to a true vane. However,
The gain of the electrode amplifier and existing vector averaging electronics can be
detector also affects the scale factor. With utilized if the e.m. rectangular components
reasonable design it is easy to feel confident arc first converted to polar form. This
in a scale accuracy of better than 1% over simulates the format of the rotor/vane com-
time and temperature. bination. A very important advantage with
Catastrophic errors-Apart from the ob- both systems lies in the essentially instan-
vious failure of critical components, the taneous ability to resolve the directional
component of the water velocity vector.
limits imposed by dynamic range of clec-
This occurs because a vane, with its atten-
trical devices are significant. The internal dant inertia and friction, is not required.
circuitry that we used can compensate elec- Magnetometer compass-Conventional
trode differentials up to about 12 mV. For compasses suffer some of the disadvantages
electrode offset potentials above this, the inherent to electromechanical devices.
device is completely inoperative. The out- They have fragile bearings, uncertain dy-
put will be some arbitrary and unchanging namic propertics, and are difficult to read
180 Kanwisher and Lawson
Fig. 4. Continuous 4%hr recording of the flow through the Eel Pond channel.
electrically. They are listed as the greatest ford 1970) show currents with a finely
source of trouble in a recent upwelling ex- stratified, rapidly changing structure. In
periment on the Oregon coast (Pillsbury et this sense the contribution of moored cur-
al. 1974). These considerations may point rent meters may be to confirm this com-
to the use of magnetometers using high pcr- plexity.
meability toroidal cores for directional in- Other applications-We have tested our
formation. The Earth’s magnetic field pro- sensor by monitoring the flow through the
ducts an imbalance in the magnetic core channel leading into Eel Pond here in
which can be electrically sensed: the re- Woods Hole. We had expected to see a
sult is an output signal which varies as the simple tidal movement. The record in Fig.
sine of the magnetic heading. This informa- 4 shows otherwise, The tidal flow is ob-
tion allows the orthogonal e.m. sensor out- scured by a short period (about 16 min)
puts to be resolved on a time scale of 1 set oscillation which is constantly present and
or less. The entire instrument case can ro- accounts for 10 times as much flow as the
tate randomly without afEecting the valid- tide. It varies 3-fold in amplitude on a
ity of the current data. schedule which we cannot correlate. The
A magnetometer of the type described record shown is full scale 2 25 cm see-l.
can operate with sufficient accuracy ( -t- 3 The cross section of the channel is well
deg or better) at low power levels com- defined (2.5 X 11 m). The integrated flow
patible with long term measurement prob- during the 8 min of a half cycle can be ex-
lems, i.e. 10 mW or less. We have a work- pressed as volume OFwater moving through
ing prototype which we are combining the channel. This volume distributed over
with the e.m. sensor to produce an all elec- the known area of the pond ( 49,000 m2)
tronic current meter with no moving parts. tells the change in height. Since the time
It may turn out that when oceanography constant OF the pond is short compared to
finally gets a working current meter, the 16 min, the same change in height must
real ocean will appear more complicated occur outside the channel. This periodic
than we can treat conceptually. Certainly movement up and down represents a long
the initial results of in situ potentials as an slow wave driving the water back and forth
index of water movement (Drevcr and San- through the channel. The waves shown in
Electromagnetic flow sensors 181
channel=2.5 x lim=27.5m2
volume= 14.6m3
Many physical oceanographers presently NEUMANN, G., AND W. J. PIU~SON. 1966. Prin-
involved in the construction and use of cur- ciples of physical oceanography. Prentice-
Hall.
rent meters must receive a part of the credit OLSON, J. R. 1972. Two-component electromag-
for the work described here. Their con- netic flowmeter. Mar. Technol. Sot. J. 6:
tinued lack of interest in other than Sa- 19-24.
vonius rotor current meters has finally PILLSBURY, R. D., J, S. BOTTERO, R. E. STILL, AND
stimulated us to the present effort. We hope W.E. GILBERT. 1974. A compilation of ob-
scrvations from moored current meters.
their current agonizing reappraisal, growing School of Oceanogr. Data Rep. 57. Oregon
out of the disappointing MODE results, State Univ. Ref. 74-2.
will make them more receptive to alternate TUCKER, M. J. 1972. Electromagnetic current
methods such as e.m. sensors and magne- meters. Proc. Sot. Underwater Technol. 2:
53-58.
tome tcrs. N. D. SMITH, F. E. PIERCE, AND E. P.
GdLLINS. 1970. A two-component electro-
References magnetic ships’ log. J. Inst. Navig. 23: 302-
BOWDEN, K. F., AND L. A. FAIRBAIRN. 1956. 316.
Measurement of turbulent fluctuations and VON ARX, W. S. 1950. An electromagnetic
Reynolds stresses in a tidal current. method for measuring the velocities of ocean
Proc. R.
Sot. Land. Ser. A 237: 442-438. currents from a ship underway. Pap. Phys.
DREVER. R. G., AND T. SANFORD. 1970. A free- Oceanogr. Mctcorol. 11: l-60.
fall electromagnetic current meter. IEEE
( Inst. Elec. Electron. Eng. ) Conf. Proc. 19 : Submitted: 11 March 1974
353-370. Accepted: 12 November 1974