Industrial Digital Indicator Guide
Industrial Digital Indicator Guide
JIR-301-M
Your Industry.
Our Indicators.
96 11.5 104.5
106.2
Alarm Output (3 points) is Provided as Standard Standard Dust-proof/Drip-proof (for front panel only)
Alarm output (3 points) is available as standard. IP66 structure means the indicator can be used even in harsh
Alarm type and status Energized/De-energized can be easily environment exposed to dust and water splashes.
switched by keypad. (Factory default: No alarm action, Energized)
Specifications
■ Model ■ Rated Range
J I R -3 0 1 -M □, □□□ Series name: JIR-301-M (W96 x H48 x D100 mm) Input Type Input Range
Input M Multi-range (*1) -200 to 1370 ℃ -320 to 2500 ℉
K
100 to 240 V AC/DC -200.0 to 400.0 ℃ -200.0 to 750.0 ℉
Power supply voltage
1 24 V AC/DC (*2) J -200 to 1000 ℃ -320 to 1800 ℉
A4 Alarm 4 output (*3) R 0 to 1760 ℃ 0 to 3200 ℉
C5 Serial communication (RS-485) (*4) S 0 to 1760 ℃ 0 to 3200 ℉
P24 Insulated power output: 24 ± 3 V DC (*5), (*6) Thermocouple B 0 to 1820 ℃ 0 to 3300 ℉
P5 Insulated power output: 5 ± 0.5 V DC (*5), (*6) E -200 to 800 ℃ -320 to 1500 ℉
DSB Power for 2-wire transmitter (Current loop supply) (*6), (*7) T -200.0 to 400.0 ℃ -200.0 to 750.0 ℉
TA2(4-20) 4 to 20 mA DC N -200 to 1300 ℃ -320 to 2300 ℉
Direct current output
TA2(0-20) 0 to 20 mA DC PL-Ⅱ 0 to 1390 ℃ 0 to 2500 ℉
TV2(0-1) Transmission 0 to 1 V DC C (W/Re5-26) 0 to 2315 ℃ 0 to 4200 ℉
TV2(0-5) output 2 (*3) 0 to 5 V DC -200.0 to 850.0 ℃ -200.0 to 1000.0 ℉
Option DC voltage output Pt100
TV2(1-5) 1 to 5 V DC -200 to 850 ℃ -300 to 1500 ℉
RTD
TV2(0-10) 0 to 10 V DC -200.0 to 500.0 ℃ -200.0 to 900.0 ℉
JPt100
TA(0-20) Direct current output 0 to 20 mA DC -200 to 500 ℃ -300 to 900 ℉
TV(0-1) User specified 0 to 1 V DC 4 to 20 mA DC
TV(0-5) Transmission 0 to 5 V DC (Externally mounted 50 Ω -2000 to 10000 (*1), (*2)
DC voltage output shunt resistor)
TV(1-5) output (*8) 1 to 5 V DC
TV(0-10) 0 to 10 V DC 0 to 20 mA DC
BK Color: Black (Externally mounted 50 Ω -2000 to 10000 (*1), (*2)
Direct current shunt resistor)
TC Terminal cover
Three alarm outputs (A1, A2, A3 output) are standard. 4 to 20 mA DC
(Built-in 50 Ω shunt resistor)
-2000 to 10000 (*1), (*3), (*4)
Alarm types (4 types of A1, A2, and 5 types of A3 as well as No alarm action) and
Energized/De-energized can be selected. 0 to 20 mA DC
(*1) Thermocouple (10 types), RTD (2 types), Direct current (2 types) and DC voltage (4 types) (Built-in 50 Ω shunt resistor)
-2000~10000(*1), (*3)
input can be selected by keypad. 0 to 1 V DC -2000 to 10000 (*1)
(*2) Power supply voltage 100 to 240 V AC is standard. When ordering 24 V AC/DC, enter ‘1’ after 0 to 5 V DC -2000 to 10000 (*1)
the input code. DC voltage
1 to 5 V DC -2000 to 10000 (*1)
(*3) Alarm 4 output (A4 option) and Transmission output 2 (T □2 option) cannot be used together.
(*4) If Serial communication (RS-485) (C5 option) is ordered, the Event input function cannot be used. 0 to 10 V DC -2000 to 10000 (*1)
(*5) Insulated power output (P24 option) and Insulated power output (P5 option) cannot be used (*1) Input range and decimal point place can be selected.
together. If Insulated power output (P24 or P5 option) is ordered, A2 output cannot be used. (*2) Connect a 50 Ω shunt resistor (sold separately) between input terminals.
(*6) Insulated power output (P24 or P5 option) cannot be used with the Power for 2-wire transmitter (*3) This input type has a built-in 50 Ω shunt resistor.
(DSB option). (*4) If Power for 2-wire transmitter (DSB option) is ordered, only 4 to 20 mA DC input
(*7) If Power for 2-wire transmitter (DSB option) is ordered, only 4 to 20 mA DC input (Built-in 50 Ω (Built-in 50 Ω shunt resistor) can be used.
shunt resistor) can be used.
(*8) TA (4-20 mA DC) is a standard feature.
UP key FAST key
HOLD indicator
TX/RX indicator
A1 to A4 value indicator
■ Terminal Arrangement
GND Ground terminal
PWR Power supply voltage
TRANSMIT OUTPUT1 Transmission output 1
A1 A1 output
A2 A2 output
A3 A3 output
EVENT INPUT Event input
TC Thermocouple input
RTD RTD input
Direct current input, DC voltage input
DC For direct current input (externally mounted 50 Ω shunt resistor),
connect a 50 Ω shunt resistor (sold separately) between input terminals.
P24 Insulated power output 24 V (P24 option)
P5 Insulated power output 5 V (P5 option)
RS-485 Serial communication (RS-485) (C5 option)
TRANSMIT OUTPUT2 Transmission output 2 (T □2 option)
A4 A4 output (A4 option)
A Direct current input (DSB option)
24V Power for 2-wire transmitter (DSB option)
Alarm unit 2
Recorder
44.5
48
Power supply
96 11.5 98.5
2(*)
10 9 8 7 6 5 4 3 2 1
104.5(*)
106.2
(*) When terminal cover is used
20 19 18 17 16 15 14 13 12 11
Heater Thermocouple
45
■ Solderless Terminal
n×48-3
5.8mm以下
92 0.8
0
Caution: If vertical close mounting is used for the
instrument, IP66 (Dust-proof/Drip-proof) may be
compromised, and all warranties will be invalidated.
Caution • The terminal block of this instrument is designed to be wired from the upper side.
• If Insulated power output (P24 or P5 option) is ordered, A2 output cannot be used.
• If Serial communication (C5 option) is ordered, the Event input function cannot be used.
● To ensure safe and correct use, thoroughly read and understand the manual before using this instrument. Caution with respect to
● This instrument is intended to be used for industrial machinery, machine tools and measuring equipment. Export Trade Control ordinance
Verify correct usage after purpose-of-use consultation with our agency or main office. (Never use this instrument for
medical purposes with which human lives are involved.) To avoid this instrument from being used as a
component in, or asbeing utilized in the
● External protection devices such as protective equipment against excessive temperature rise, etc. must be installed,
manufacture of weapons of mass destruction
as malfunction of this product could result in serious damage to the system or injury to personnel. (i.e. military applications, military equipment, etc.),
SAFETY Proper periodic maintenance is also required. please investigate the end users and the final
PRECAUTIONS ● This instrument must be used under the conditions and environment described in the manual. Shinko Technos use of this instrument.
Co., Ltd. does not accept liability for any injury, loss of life or damage occurring due to the instrument being used In the case of resale, ensure that this instrument
under conditions not otherwise stated in the manual. is not illegally exported.
• This catalog is as of April 2021 and its contents are subject to change without notice.
• Photos used in this catalog do not show unit in operating status.
• If you have any inquiries, please consult us or our agency.