Dosimass - Operatins Instructions
Dosimass - Operatins Instructions
Operating Instructions
Dosimass
Coriolis flowmeter
Dosimass
• Make sure the document is stored in a safe place such that it is always available when
working on or with the device.
• To avoid danger to individuals or the facility, read the "Basic safety instructions" section
carefully, as well as all other safety instructions in the document that are specific to
working procedures.
• The manufacturer reserves the right to modify technical data without prior notice. Your
Endress+Hauser Sales Center will supply you with current information and updates to
these instructions.
2 Endress+Hauser
Dosimass Table of contents
Table of contents
Endress+Hauser 3
Table of contents Dosimass
13 Maintenance . . . . . . . . . . . . . . . . . . . . . . 57
13.1 Maintenance tasks . . . . . . . . . . . . . . . . . . . . . 57
13.1.1 Exterior cleaning . . . . . . . . . . . . . . . . 57
13.1.2 Interior cleaning . . . . . . . . . . . . . . . . 57
13.2 Measuring and test equipment . . . . . . . . . . . . 57
13.3 Endress+Hauser services . . . . . . . . . . . . . . . . 57
14 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
14.1 General notes . . . . . . . . . . . . . . . . . . . . . . . . 58
14.2 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . 58
14.3 Endress+Hauser services . . . . . . . . . . . . . . . . 58
14.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
14.5 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
14.5.1 Removing the measuring device . . . . . 58
14.5.2 Disposing of the measuring device . . . 59
15 Accessories . . . . . . . . . . . . . . . . . . . . . . . 60
15.1 Communication-specific accessories . . . . . . . . 60
15.2 Service-specific accessories . . . . . . . . . . . . . . . 60
16 Technical data . . . . . . . . . . . . . . . . . . . . 62
16.1 Application . . . . . . . . . . . . . . . . . . . . . . . . . . 62
16.2 Function and system design . . . . . . . . . . . . . . 62
16.3 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
16.4 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
16.5 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . 64
16.6 Performance characteristics . . . . . . . . . . . . . . 65
16.7 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . 67
16.8 Environment . . . . . . . . . . . . . . . . . . . . . . . . . 67
16.9 Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
16.10 Mechanical construction . . . . . . . . . . . . . . . . 70
16.11 Operability . . . . . . . . . . . . . . . . . . . . . . . . . . 71
16.12 Certificates and approvals . . . . . . . . . . . . . . . 72
16.13 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . 73
16.14 Supplementary documentation . . . . . . . . . . . . 73
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
4 Endress+Hauser
Dosimass Document information
1 Document information
Symbol Meaning
DANGER!
DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
serious or fatal injury.
WARNING!
WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
serious or fatal injury.
CAUTION!
CAUTION This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
minor or medium injury.
NOTE!
NOTICE This symbol contains information on procedures and other facts which do not result in
personal injury.
Symbol Meaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Endress+Hauser 5
Document information Dosimass
Symbol Meaning
Tip
Indicates additional information.
Reference to documentation
Reference to page
Reference to graphic
, , … Series of steps
Result of a step
Visual inspection
1.3 Documentation
For an overview of the scope of the associated Technical Documentation, refer to the
following:
• The CD-ROM provided for the device (depending on the device version, the CD-ROM
might not be part of the delivery!)
• The W@M Device Viewer : Enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate
or scan the 2-D matrix code (QR code) on the nameplate.
For a detailed list of the individual documents along with the documentation code
Brief Operating Instructions Guide that takes you quickly to the 1st measured value
The Brief Operating Instructions contain all the essential information
from incoming acceptance to initial commissioning.
6 Endress+Hauser
Dosimass Document information
Endress+Hauser 7
Basic safety instructions Dosimass
Incorrect use
Non-designated use can compromise safety. The manufacturer is not liable for damage
caused by improper or non-designated use.
LWARNING
Danger of breakage of the measuring tube due to corrosive or abrasive fluids or from
environmental conditions.
Housing breakage due to mechanical overload possible!
‣ Verify the compatibility of the process fluid with the measuring tube material.
‣ Ensure the resistance of all fluid-wetted materials in the process.
‣ Keep within the specified pressure and temperature range.
8 Endress+Hauser
Dosimass Basic safety instructions
Residual risks
The external surface temperature of the housing can increase by max. 20 K due to the
power consumption of the electronic components. Hot process fluids passing through the
measuring device will further increase the surface temperature of the housing. The surface
of the sensor, in particular, can reach temperatures which are close to the fluid
temperature.
Possible burn hazard due to fluid temperatures!
‣ For elevated fluid temperature, ensure protection against contact to prevent burns.
Endress+Hauser 9
Basic safety instructions Dosimass
2.6 IT security
We only provide a warranty if the device is installed and used as described in the
Operating Instructions. The device is equipped with security mechanisms to protect it
against any inadvertent changes to the device settings.
IT security measures in line with operators' security standards and designed to provide
additional protection for the device and device data transfer must be implemented by the
operators themselves.
10 Endress+Hauser
Dosimass Product description
3 Product description
The device consists of a transmitter and a sensor.
The device is available as a compact version:
The transmitter and sensor form a mechanical unit.
2
A0026625
Endress+Hauser 11
Incoming acceptance and product identification Dosimass
• If one of the conditions is not satisfied, contact your Endress+Hauser Sales Center.
• Depending on the device version, the CD-ROM might not be part of the delivery!
The Technical Documentation is available via the Internet or via the Endress+Hauser
Operations App, see the "Product identification" section → 13.
12 Endress+Hauser
Dosimass Incoming acceptance and product identification
For an overview of the scope of the associated Technical Documentation, refer to the
following:
• The chapters "Additional standard documentation on the device" → 6 and
"Supplementary device-dependent documentation" → 7
• The W@M Device Viewer: Enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate or
scan the 2-D matrix code (QR code) on the nameplate.
4153 Reinach
1 Switzerland
Dosimass
2 Order Code: 10
3 Ser. no.:
TAG no.:
For installation and temp.
4 marking see control dwgs.
5 i
6 Materials:
7 Tm:
11
Ta+20°C/36°F
12
8 Ta:
13
9
A0007262
Order code
The measuring device is reordered using the order code.
Extended order code
• The device type (product root) and basic specifications (mandatory features) are
always listed.
• Of the optional specifications (optional features), only the safety and approval-
related specifications are listed (e.g. LA). If other optional specifications are also
ordered, these are indicated collectively using the # placeholder symbol (e.g. #LA#).
• If the ordered optional specifications do not include any safety and approval-related
specifications, they are indicated by the + placeholder symbol (e.g. XXXXXX-ABCDE
+).
Endress+Hauser 13
Incoming acceptance and product identification Dosimass
Symbol Meaning
WARNING!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious
or fatal injury.
Reference to documentation
Refers to the corresponding device documentation.
14 Endress+Hauser
Dosimass Storage and transport
Endress+Hauser 15
Installation Dosimass
6 Installation
Mounting location
A0023344
To prevent measuring errors arising from accumulation of gas bubbles in the measuring
tube, avoid the following mounting locations in the pipe:
• Highest point of a pipeline.
• Directly upstream of a free pipe outlet in a down pipe.
A0015596
16 Endress+Hauser
Dosimass Installation
8 ³⁄₈ 6 0.24
15 ½ 10 0.40
25 1 14 0.55
Orientation
The direction of the arrow on the sensor nameplate helps you to install the sensor
according to the flow direction (direction of medium flow through the piping).
Orientation Recommendation
A Vertical orientation
A0015591
1)
B Horizontal orientation, transmitter
head up Exception:
→ 4, 17
A0015589
2)
C Horizontal orientation, transmitter
head down Exception:
→ 4, 17
A0015590
1) Applications with low process temperatures may reduce the ambient temperature. To maintain the
minimum ambient temperature for the transmitter, this orientation is recommended.
2) Applications with high process temperatures may increase the ambient temperature. To maintain the
maximum ambient temperature for the transmitter, this orientation is recommended.
If a sensor is installed horizontally with a curved measuring tube, match the position of the
sensor to the fluid properties.
1 2
A0014057
Endress+Hauser 17
Installation Dosimass
Valves
Never install the sensor downstream from a filling valve. If the sensor is completely empty
this corrupts the measured value.
Correct measurement is only possible if the pipe is completely full. Perform sample
fillings before commencing filling in production.
1 2
2 1
3 3
A0003768
1 Measuring device
2 Filling valve
3 Container
Filling systems
The pipe system must be completely full to ensure optimum measurement.
1
2 1 2
2
1
3 3 3
A0003795
5 Filling system
1 Measuring device
2 Filling valve
3 Container
High temperatures
A0003830
To prevent the electronics from overheating in the event of strong heat formation
(e.g. CIP or SIP cleaning process), install the measuring device with the transmitter
part pointing downwards.
18 Endress+Hauser
Dosimass Installation
A0015597 A0015598
Installation dimensions
For the dimensions and installation lengths of the device, see the "Technical
Information" document, "Mechanical construction" section
Temperature tables
Observe the interdependencies between the permitted ambient and fluid
temperatures when operating the device in hazardous areas.
For detailed information on the temperature tables, see the separate document
entitled "Safety Instructions" (XA) for the device.
System pressure
It is important that cavitation does not occur, or that gases entrained in the liquids do not
outgas.
Cavitation is caused if the pressure drops below the vapor pressure:
• In liquids that have a low boiling point (e.g. hydrocarbons, solvents, liquefied gases)
• In suction lines
‣ Ensure the system pressure is sufficiently high to prevent cavitation and outgassing.
For this reason, the following mounting locations are recommended:
• At the lowest point in a vertical pipe
• Downstream from pumps (no danger of vacuum)
A0015594
Endress+Hauser 19
Installation Dosimass
Thermal insulation
In the case of some fluids, it is important that the heat radiated from the sensor to the
transmitter is kept to a minimum. A wide range of materials can be used for the required
insulation.
The connection adapter between the sensor and transmitter must always be kept free.
Note that a certain orientation might be required, depending on the medium temperature
→ 17.
For information on the permitted ambient temperature range
NOTICE
Electronics overheating on account of thermal insulation!
‣ Observe maximum permitted insulation height of the transmitter neck so that the
transmitter head is completely free.
Heating
NOTICE
Electronics can overheat due to elevated ambient temperature!
‣ Observe maximum permitted ambient temperature for the transmitter .
‣ Depending on the fluid temperature, take the device orientation requirements into
account .
NOTICE
Danger of overheating when heating
‣ Ensure that the temperature at the lower end of the transmitter housing does not
exceed 80 °C (176 °F)
‣ Ensure that convection takes place on a sufficiently large scale at the transmitter neck.
‣ Ensure that a sufficiently large area of the housing support remains exposed. The
uncovered part serves as a radiator and protects the electronics from overheating and
excessive cooling.
Heating options
If a fluid requires that no heat loss should occur at the sensor, users can avail of the
following heating options:
• Electrical heating, e.g. with electric band heaters
• Via pipes carrying hot water or steam
• Via heating jackets
Vibrations
The high oscillation frequency of the measuring tubes ensures that the correct operation of
the measuring system is not influenced by plant vibrations.
20 Endress+Hauser
Dosimass Installation
A0003761
1 Measuring device
2 Tank
3 Batching valve
4 Vessel
1 3
A0003762
1 Measuring device
2 Tank
3 Batching valve
4 Vessel
Endress+Hauser 21
Installation Dosimass
NOTICE
All Dosimass measuring devices are calibrated in accordance with state-of-the-art
technology. Calibration takes place under reference conditions .
Therefore, a zero point adjustment is generally not required for the Dosimass!
‣ Experience shows that a zero point adjustment is advisable only in special cases.
‣ When maximum accuracy is required and flow rates are very low.
‣ Under extreme process or operating conditions (e.g. very high process temperatures or
very high-viscosity fluids).
A0008558
7
22 Endress+Hauser
Dosimass Installation
Does the arrow on the sensor nameplate match the direction of flow of the fluid through the
piping → 17?
Are the measuring point identification and labeling correct (visual inspection)?
Endress+Hauser 23
Electrical connection Dosimass
7 Electrical connection
The measuring device does not have an internal circuit breaker. For this reason,
assign the measuring device a switch or power-circuit breaker so that the power
supply line can be easily disconnected from the mains.
Electrical safety
In accordance with applicable federal/national regulations.
Signal cable
Cables are not included in the scope of delivery; they can be ordered as an accessory
→ 60.
Pulse/frequency/switch output
Standard installation cable is sufficient.
24 Endress+Hauser
Dosimass Electrical connection
RSE 8
M12x1
A B
2 2
8 3 3 8
1 4 4 1
7 5 5 7
6 6
A B
1
2
3
4
5
+ 6
E 7
-
8
A0023237
8 Connection to device
A Coupling: Supply voltage, pulse/freq./switch output
B Connector: Supply voltage, pulse/freq./switch output
E PELV or SELV power supply
1 to Pin assignment
8
Pin assignment
Connection: Coupling (A) – Connector (B)
Pin Assignment
1 L+ Supply voltage
2 + Service interface RX
3 + Service interface TX
4 L- Supply voltage
5 + Pulse/frequency/switch output
6 – Pulse/frequency/switch output 1
7 – Pulse/frequency/switch output 2
Endress+Hauser 25
Electrical connection Dosimass
Supply voltage
DC 24 V (nominal voltage: DC 20 to 30 V)
• The power unit must be tested to ensure that it meets safety requirements (e.g.
PELV, SELV).
• The supply voltage must not exceed a maximum short-circuit current of 50 A.
A B
A0023690
A Coupling
B Plug
Grounding
Grounding is by means of a cable socket.
A0007235
To guarantee IP67 degree of protection, Type 4X enclosure, carry out the following steps
after the electrical connection:
26 Endress+Hauser
Dosimass Electrical connection
Does the supply voltage in the system match the specifications on the device's nameplate?
Are the maximum values for voltage and current at the pulse and status output being observed?
→ 63
Endress+Hauser 27
Operation options Dosimass
8 Operation options
1 2
A0017760
28 Endress+Hauser
Dosimass Operation options
2 5
+
1 4
-
A0007228
1 Supply voltage 24 V DC
2 Service adapter
3 Dosimass
4 Commubox FXA291
5 Computer with "FieldCare" or "DeviceCare" operating tool
The service adapter, cable and Commubox FXA291 are not included in the delivery.
These components can be ordered as accessories → 60.
8.2.2 FieldCare
Function scope
FDT-based plant asset management tool from Endress+Hauser. It can configure all smart
field devices in a system and helps you manage them. By using the status information, it is
also a simple but effective way of checking their status and condition.
Access is via:
Service adapter and Commubox FXA291 → 28
Typical functions:
• Configuring parameters of transmitters
• Loading and saving device data (upload/download)
• Documentation of the measuring point
• Visualization of the measured value memory (line recorder) and event logbook
For details, see Operating Instructions BA00027S and BA00059S
Establishing a connection
Service adapter, Commubox FXA291 and "FieldCare" operating tool
1. Start FieldCare and launch the project.
2. In the network: Add a device.
The Add device window opens.
3. Select the CDI Communication FXA291 option from the list and press OK to
confirm.
4. Right-click CDI Communication FXA291 and select the Add device option in the
context menu that opens.
Endress+Hauser 29
Operation options Dosimass
5. Select the desired device from the list and press OK to confirm.
6. Establish the online connection to the device.
User interface
2 3 4 5 6 7
Xxxxxx/…/…/
Device name: Xxxxxxx Mass flow: 12.34 kg/h
Device tag: Xxxxxxx
1 Volume flow: 12.34 m³/h
Status: Good
10 11
A0021051-EN
1 Header
2 Picture of device
3 Device name
4 Tag name
5 Status area with status signal
6 Display area for current measured values
7 Edit toolbar with additional functions such as save/restore, event list and create documentation
8 Navigation area with operating menu structure
9 Working area
10 Range of action
11 Status area
8.2.3 DeviceCare
Function scope
Tool to connect and configure Endress+Hauser field devices.
The fastest way to configure Endress+Hauser field devices is with the dedicated
"DeviceCare" tool. Together with the device type managers (DTMs) it presents a convenient,
comprehensive solution.
For details, see Innovation Brochure IN01047S
30 Endress+Hauser
Dosimass System integration
9 System integration
Endress+Hauser 31
Commissioning Dosimass
10 Commissioning
If the device does not start up successfully, depending on the cause, a diagnostic
message is displayed in the system asset management tool "FieldCare" .
Navigation
"Setup" menu
Setup
Device tag → 33
‣ System units → 33
‣ Pulse/frequency/switch output 1 → 35
to 2
‣ Advanced setup → 41
32 Endress+Hauser
Dosimass Commissioning
Navigation
"Setup" menu → Device tag
Device tag Enter the name for the measuring point. A maximum of 32 characters Dosimass
such as letters, numbers or
special characters (e.g. @, %, /)
Navigation
"Setup" menu → System units
‣ System units
Mass unit
Volume unit
Density unit
Temperature unit
Mass flow unit Select mass flow unit. Unit choose list Country-specific:
• g/s
Result
• oz/s
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
Endress+Hauser 33
Commissioning Dosimass
Volume flow unit Select volume flow unit. Unit choose list Country-specific:
• ml/s
Result
• fl oz/s (us)
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
34 Endress+Hauser
Dosimass Commissioning
Navigation
"Setup" menu → Pulse/frequency/switch output 1 to 2
‣ Pulse/frequency/switch output 1
to 2
Operating mode
Channel 2
Pulse width
Failure mode
Channel 2 The Pulse option is selected in Select impulse with or without • Off Off
the Operating mode time offset. • Redundant 0°
parameter. • Redundant 90°
• Redundant 180°
Assign pulse output One of the following options is Select process variable for • Off Off
selected in the Operating pulse output. • Mass flow
mode parameter: • Volume flow
• Pulse
• Automatic pulse
Endress+Hauser 35
Commissioning Dosimass
Value per pulse One of the following options is Enter measured value at which Signed floating-point Depends on the
selected in the Operating a pulse is output. number nominal diameter:
mode parameter: • DN 8 (³/₈"):
• Pulse 0.02 g (0.001 oz)
• Automatic pulse • DN 15 (½"):
0.1 g (0.004 oz)
One of the following options is
• DN 25 (1"):
selected in the Assign pulse
0.2 g (0.007 oz)
output parameter (→ 35):
• Mass flow
• Volume flow
Pulse width In the Operating mode Define time width of the 0.05 to 3.75 ms 0.05 ms
parameter, the Pulse option is output pulse.
selected and one of the
following options is selected in
the Assign pulse output
parameter (→ 35):
• Mass flow
• Volume flow
Failure mode One of the following options is Define output behavior in • Actual value Actual value
selected in the Operating alarm condition. • No pulses
mode parameter:
• Pulse
• Automatic pulse
One of the following options is
selected in the Assign pulse
output parameter (→ 35):
• Mass flow
• Volume flow
Navigation
"Setup" menu → Pulse/frequency/switch output 1 to 2
‣ Pulse/frequency/switch output 1
to 2
Operating mode
36 Endress+Hauser
Dosimass Commissioning
Failure mode
Failure frequency
Assign frequency output The Frequency option is Select process variable for • Off Off
selected in the Operating frequency output. • Mass flow
mode parameter (→ 35). • Volume flow
• Density
• Temperature
Minimum frequency value In the Operating mode Enter minimum frequency. 0.0 to 10 000.0 Hz 0.0 Hz
parameter, the Frequency
option is selected and one of
the following options is
selected in the Assign
frequency output parameter
(→ 37):
• Mass flow
• Volume flow
• Density
• Temperature
Maximum frequency value In the Operating mode Enter maximum frequency. 0.0 to 10 000.0 Hz 10 000.0 Hz
parameter, the Frequency
option is selected and one of
the following options is
selected in the Assign
frequency output parameter
(→ 37):
• Mass flow
• Volume flow
• Density
• Temperature
Measuring value at maximum In the Operating mode Enter measured value for Signed floating-point Depends on country
frequency parameter, the Frequency maximum frequency. number and nominal
option is selected and one of diameter
the following options is
selected in the Assign
frequency output parameter
(→ 37):
• Mass flow
• Volume flow
• Density
• Temperature
Endress+Hauser 37
Commissioning Dosimass
Failure mode In the Operating mode Define output behavior in • Actual value 0 Hz
parameter (→ 35), the alarm condition. • Defined value
Frequency option is selected • 0 Hz
and one of the following
options is selected in the
Assign frequency output
parameter (→ 37):
• Mass flow
• Volume flow
• Density
• Temperature
Failure frequency In the Operating mode Enter frequency output value 0.0 to 10 000.0 Hz 0.0 Hz
parameter (→ 35), the in alarm condition.
Frequency option is selected
and one of the following
options is selected in the
Assign frequency output
parameter (→ 37):
• Mass flow
• Volume flow
• Density
• Temperature
Navigation
"Setup" menu → Pulse/frequency/switch output 1 to 2
‣ Pulse/frequency/switch output 1
to 2
Operating mode
Assign limit
Assign status
Switch-on value
Switch-off value
38 Endress+Hauser
Dosimass Commissioning
Failure mode
Switch output function The Switch option is selected Select function for switch • Off Off
in the Operating mode output. • On
parameter. • Diagnostic
behavior
• Limit
• Flow direction
check
• Status
Assign diagnostic behavior • The Switch option is Select diagnostic behavior for • Alarm Alarm
selected in the Operating switch output. • Alarm or warning
mode parameter. • Warning
• The Diagnostic behavior
option is selected in the
Switch output function
parameter.
Assign limit • The Switch option is Select process variable for limit • Mass flow Mass flow
selected in the Operating function. • Volume flow
mode parameter. • Density
• The Limit option is selected • Temperature
in the Switch output
function parameter.
Assign flow direction check • The Switch option is Select process variable for flow • Off Mass flow
selected in the Operating direction monitoring. • Volume flow
mode parameter. • Mass flow
• The Flow direction check
option is selected in the
Switch output function
parameter.
Assign status • The Switch option is Select device status for switch • Partially filled pipe Partially filled pipe
selected in the Operating output. detection detection
mode parameter. • Low flow cut off
• The Status option is
selected in the Switch
output function parameter.
Switch-on value • The Switch option is Enter measured value for the Signed floating-point Country-specific:
selected in the Operating switch-on point. number • 0 g/s
mode parameter. • 0 oz/s
• The Limit option is selected
in the Switch output
function parameter.
Endress+Hauser 39
Commissioning Dosimass
Switch-off value • The Switch option is Enter measured value for the Signed floating-point Country-specific:
selected in the Operating switch-off point. number • 0 g/s
mode parameter. • 0 oz/s
• The Limit option is selected
in the Switch output
function parameter.
Navigation
"Setup" menu → Low flow cut off
Assign process variable – Select process variable for low • Off Mass flow
flow cut off. • Mass flow
• Volume flow
On value low flow cutoff One of the following options is Enter on value for low flow cut Positive floating- For liquids: depends
selected in the Assign process off. point number on country and
variable parameter (→ 40): nominal diameter
• Mass flow
• Volume flow
Off value low flow cutoff One of the following options is Enter off value for low flow cut 0 to 100.0 % 50 %
selected in the Assign process off.
variable parameter (→ 40):
• Mass flow
• Volume flow
Pressure shock suppression One of the following options is Enter time frame for signal 0 to 100 s 0s
selected in the Assign process suppression (= active pressure
variable parameter (→ 40): shock suppression).
• Mass flow
• Volume flow
40 Endress+Hauser
Dosimass Commissioning
Navigation
"Setup" menu → Partially filled pipe detection
Low value partial filled pipe The Density option is selected Enter lower limit value for Signed floating-point Depending on
detection in the Assign process variable deactivating partialy filled pipe number country:
parameter. detection. • 200 kg/m3
• 12.5 lb/ft3
High value partial filled pipe The Density option is selected Enter upper limit value for Signed floating-point Depending on
detection in the Assign process variable deactivating partialy filled pipe number country:
parameter. detection. • 6 000 kg/m3
• 374.6 lb/ft3
Response time part. filled pipe In the Assign process variable Enter time before diagnostic 0 to 100 s 1s
detect. parameter, the Density option message is displayed for
is selected. partially filled pipe detection.
Navigation
"Setup" menu → Advanced setup
‣ Advanced setup
‣ Sensor adjustment → 42
Endress+Hauser 41
Commissioning Dosimass
‣ Totalizer 1 to 3 → 42
‣ Administration → 54
Navigation
"Setup" menu → Advanced setup → Sensor adjustment
‣ Sensor adjustment
Installation direction
Progress
Installation direction Set sign of flow direction to match the • Flow in arrow direction Flow in arrow direction
direction of the arrow on the sensor. • Flow against arrow direction
Zero point adjustment control Start zero point adjustment. • Cancel Cancel
• Busy
• Zero point adjust failure
• Start
Navigation
"Setup" menu → Advanced setup → Totalizer 1 to 3
‣ Totalizer 1 to 3
Mass unit
Volume unit
42 Endress+Hauser
Dosimass Commissioning
Failure mode
Assign process variable – Select process variable for • Off Mass flow
totalizer. • Volume flow
• Mass flow
Mass unit The Mass flow option is Select mass unit. Unit choose list Depending on
selected in the Assign process country:
variable parameter (→ 43) • g
of the Totalizer 1 to 3 • oz
submenu.
Volume unit The Volume flow option is Select volume unit. Unit choose list Depending on
selected in the Assign process country:
variable parameter (→ 43) • ml
of the Totalizer 1 to 3 • fl oz (us)
submenu.
Totalizer operation mode One of the following options is Select totalizer calculation • Net flow total Net flow total
selected in the Assign process mode. • Forward flow total
variable parameter (→ 43) • Reverse flow total
of the Totalizer 1 to 3
submenu:
• Volume flow
• Mass flow
Failure mode One of the following options is Define totalizer behavior in • Stop Stop
selected in the Assign process alarm condition. • Actual value
variable parameter (→ 43) • Last valid value
of the Totalizer 1 to 3
submenu:
• Volume flow
• Mass flow
10.6 Simulation
The Simulation submenu enables you to simulate, without a real flow situation, various
process variables in the process and the device alarm mode and to verify downstream
signal chains (switching valves or closed-control loops).
Navigation
"Diagnostics" menu → Simulation
‣ Simulation
Endress+Hauser 43
Commissioning Dosimass
Assign simulation process variable – Select a process variable for • Off Off
the simulation process that is • Mass flow
activated. • Volume flow
• Density
• Temperature
Value process variable One of the following options is Enter the simulation value for Depends on the 0
selected in the Assign the selected process variable. process variable
simulation process variable selected
parameter (→ 44):
• Mass flow
• Volume flow
• Density
• Temperature
Simulation device alarm – Switch the device alarm on and • Off Off
off. • On
44 Endress+Hauser
Dosimass Operation
11 Operation
Navigation
"Operation" menu → Locking status
Temporarily locked Write access to the parameters is temporarily lock due to device-internal
processing (e.g. data upload/download, reset). Once the internal processing has
been completed, the parameters can be changed once again.
Locking status Indicates the write protection with the Temporarily locked Temporarily locked
highest priority that is currently active.
Navigation
"Operation" menu → Access status tooling
Access status tooling Shows the access authorization to the • Operator Maintenance
parameters via the operating tool. • Maintenance
Endress+Hauser 45
Operation Dosimass
Navigation
"Diagnostics" menu → Measured values → Process variables
‣ Process variables
Mass flow
Volume flow
Density
Temperature
Mass flow Displays the mass flow currently measured. Signed floating-point number
Dependency
The unit is taken from the Mass flow unit parameter
Volume flow Displays the volume flow currently calculated. Signed floating-point number
Dependency
The unit is taken from the Volume flow unit parameter
Temperature Shows the medium temperature currently measured. Signed floating-point number
Dependency
The unit is taken from: Temperature unit parameter
11.3.2 Totalizer
The Totalizer submenu contains all the parameters needed to display the current
measured values for every totalizer.
Navigation
"Diagnostics" menu → Measured values → Totalizer
‣ Totalizer
Totalizer value 1 to 3
Totalizer overflow 1 to 3
46 Endress+Hauser
Dosimass Operation
Totalizer value One of the following options is selected Displays the current totalizer counter Signed floating-point
in the Assign process variable value. number
parameter (→ 43) of the Totalizer 1
to 3 submenu:
• Volume flow
• Mass flow
Totalizer overflow One of the following options is selected Displays the current totalizer overflow. Integer with sign
in the Assign process variable
parameter (→ 43) of the Totalizer 1
to 3 submenu:
• Volume flow
• Mass flow
Navigation
"Diagnostics" menu → Measured values → Output values
‣ Output values
Pulse output
Output frequency
Switch status
Pulse output One of the following options is selected Displays the pulse frequency currently Positive floating-point
in the Operating mode parameter: output. number
• Pulse
• Automatic pulse
Output frequency The Frequency option is selected in the Displays the value currently measured 0.0 to 10 000.0 Hz
Operating mode parameter. for the frequency output.
Switch status The Switch option is selected in the Displays the current switch output • Open
Operating mode parameter. status. • Closed
Reset + hold The totaling process is stopped and the totalizer is reset to 0.
Preset + hold The totaling process is stopped and the totalizer is set to its defined start value
from the Preset value parameter.
Endress+Hauser 47
Operation Dosimass
Options Description
Reset + totalize The totalizer is reset to 0 and the totaling process is restarted.
Preset + totalize The totalizer is set to the defined start value in the Preset value parameter and the
totaling process is restarted.
Reset + totalize Resets all totalizers to 0 and restarts the totaling process. This deletes all the flow
values previously totalized.
Navigation
"Operation" menu → Totalizer handling
‣ Totalizer handling
Control Totalizer 1 to 3
Preset value 1 to 3
Control Totalizer One of the following options is Control totalizer value. • Totalize Totalize
selected in the Assign process • Reset + hold
variable parameter (→ 43) • Preset + hold
of the Totalizer 1 to 3 • Reset + totalize
submenu: • Preset + totalize
• Volume flow • Hold
• Mass flow
Preset value One of the following options is Specify start value for totalizer. Signed floating-point Country-specific:
selected in the Assign process number • 0 kg
Dependency
variable parameter (→ 43) • 0 lb
of the Totalizer 1 to 3
submenu:
The unit of the selected
process variable is
• Volume flow specified for the totalizer
• Mass flow in the Assign process
variable parameter. If
the following is selected
in the Assign process
variable parameter:
• Volume flow option:
Volume flow unit
parameter
• Mass flow option:
Mass flow unit
parameter
48 Endress+Hauser
Dosimass Diagnostics and troubleshooting
No write access to parameters Current user role has limited access Check access authorization status .
authorization
No connection via service adapter Incorrect configuration of USB Observe the documentation for the
interface on PC or driver not Commubox.
installed correctly.
FXA291: Document
"Technical Information"
TI00405C
Xxxxxx/…/…/
Device name: Xxxxxxx Mass flow: 12.34 kg/h
Device tag: Xxxxxxx Volume flow: 12.34 m³/h
Status signal: Function check (C)
Xxxxxx
P Diagnostics 1: C485 Simu...
P Remedy information: Deactivate...
P Access status tooling: Mainenance Failure (F)
A0021799-EN
Furthermore, diagnostic events that have occurred can be viewed in the Diagnostics
menu:
• Via parameter
• Via submenu → 53
Endress+Hauser 49
Diagnostics and troubleshooting Dosimass
Diagnostic information
The fault can be identified using the diagnostic information. The short text helps you by
providing information about the fault.
Diagnostic information
Diagnostic code
↓ ↓ ↓
3-digit number
Options Description
Alarm Measurement is interrupted. Signal outputs and totalizers assume the defined alarm
condition. A diagnostic message is generated.
Warning Measurement is resumed. The signal outputs and totalizers are not affected. A
diagnostic message is generated.
Logbook entry only The device continues to measure. The diagnostic message is entered in the Event
logbook (events list) submenu only and is not displayed in alternation with the measured
value display.
Off The diagnostic event is ignored, and no diagnostic message is generated or entered.
50 Endress+Hauser
Dosimass Diagnostics and troubleshooting
Diagnostic of sensor
Diagnostic of electronic
Diagnostic of configuration
Endress+Hauser 51
Diagnostics and troubleshooting Dosimass
Diagnostic of process
Navigation
"Diagnostics" menu
52 Endress+Hauser
Dosimass Diagnostics and troubleshooting
Previous diagnostics
Actual diagnostics A diagnostic event has occurred. Shows the current occured diagnostic Symbol for diagnostic
event along with its diagnostic behavior, diagnostic code
information. and short message.
Ifsimultaneously,
two or more messages occur
the message with
the highest priority is shown on
the display.
Previous diagnostics Two diagnostic events have already Shows the diagnostic event that Symbol for diagnostic
occurred. occurred prior to the current diagnostic behavior, diagnostic code
event along with its diagnostic and short message.
information.
Endress+Hauser 53
Diagnostics and troubleshooting Dosimass
I1089 Power on
Navigation
"Expert" menu → System → Administration
‣ Administration
Device reset
Navigation
"Diagnostics" menu → Device information
‣ Device information
Device tag
Serial number
54 Endress+Hauser
Dosimass Diagnostics and troubleshooting
Firmware version
Device name
Order code
ENP version
Device tag Enter the name for the measuring point. A maximum of 32 characters Dosimass
such as letters, numbers or
special characters (e.g. @, %, /)
Serial number Shows the serial number of the measuring A maximum of 11-digit –
device. character string comprising
letters and numbers.
Firmware version Shows the device firmware version installed. Character string with the 03.00
following format:
xx.yy.zz
Order code Shows the device order code. Character string composed of –
letters, numbers and certain
It can be found in the "Order code" field
on the nameplate. punctuation marks (e.g. /).
Extended order code 1 Shows the 1st part of the extended order Character string –
code.
Itfield
can be found in the "Ext. ord. cd."
on the nameplate.
Extended order code 2 Shows the 2nd part of the extended order Character string –
code.
Extended order code 3 Shows the 3rd part of the extended order Character string –
code.
Endress+Hauser 55
Diagnostics and troubleshooting Dosimass
For the compatibility of the firmware version with the previous version, the installed
device description files and operating tools, observe the information about the device
in the "Manufacturer's information" document.
The manufacturer's information is available:
• In the Downloads area of the Endress+Hauser web site: www.endress.com →
Downloads
• Specify the following details:
– Product root: e.g. 8RE
– Text search: Manufacturer's information
– Media type: Documentation – Technical Documentation
56 Endress+Hauser
Dosimass Maintenance
13 Maintenance
For a list of some of the measuring and test equipment, refer to the "Accessories"
chapter of the "Technical Information" document for the device.
Endress+Hauser 57
Repair Dosimass
14 Repair
14.4 Return
The measuring device must be returned if it is need of repair or a factory calibration, or if
the wrong measuring device has been delivered or ordered. Legal specifications require
Endress+Hauser, as an ISO-certified company, to follow certain procedures when handling
products that are in contact with the medium.
To ensure safe, swift and professional device returns, please refer to the procedure and
conditions for returning devices provided on the Endress+Hauser website at
http://www.endress.com/support/return-material
14.5 Disposal
58 Endress+Hauser
Dosimass Repair
Endress+Hauser 59
Accessories Dosimass
15 Accessories
Various accessories, which can be ordered with the device or subsequently from Endress
+Hauser, are available for the device. Detailed information on the order code in question is
available from your local Endress+Hauser sales center or on the product page of the
Endress+Hauser website: www.endress.com.
60 Endress+Hauser
Dosimass Accessories
Endress+Hauser 61
Technical data Dosimass
16 Technical data
16.1 Application
The measuring device is suitable for flow measurement of liquids and gases only.
Depending on the version ordered, the measuring device can also measure potentially
explosive, flammable, poisonous and oxidizing media.
To ensure that the device remains in proper operating condition for its service life, use the
measuring device only for media against which the process-wetted materials are
adequately resistant.
Measuring principle Mass flow measurement based on the Coriolis measuring principle
16.3 Input
[mm] [kg/h]
8 0 to 2 000
15 0 to 6 500
25 0 to 18 000
62 Endress+Hauser
Dosimass Technical data
[in] [lb/min]
³⁄₈ 0 to 73.50
½ 0 to 238.9
1 0 to 661.5
To calculate the measuring range, use the Applicator product selection tool → 60
16.4 Output
Pulse output
Frequency output
Endress+Hauser 63
Technical data Dosimass
Switch output
Pulse/frequency/switch output
Pulse output
Failure mode Choose from:
• Actual value
• No pulses
Frequency output
Failure mode Choose from:
• Actual value
• 0 Hz
• Defined value: 0 to 10 000 Hz
Switch output
Failure mode Choose from:
• Current status
• Open
• Closed
Low flow cut off The switch points for low flow cut off are user-selectable.
Terminal assignment → 24
64 Endress+Hauser
Dosimass Technical data
Electrical connection → 26
Cable specification → 24
Base accuracy
Mass flow and volume flow (liquids)
• ±0.15 % o.r.
• ±0.30 % ± [(zero point stability : measured value) · 100] % o.r.
• ±5 % ± [(zero point stability : measured value) · 100] % o.r.
Endress+Hauser 65
Technical data Dosimass
Density (liquids)
• Reference operating conditions: ±0.0005 g/cm³
• Field density calibration: ±0.0005 g/cm³
(after field density calibration under process conditions)
• Standard density calibration: ±0.02 g/cm³
(valid over the entire temperature range and density range )
Temperature
±0.5 °C ± 0.005 · T °C (±0.9 °F ± 0.003 · (T – 32) °F)
15 ½ 0.65 0.024
25 1 1.80 0.066
Flow values
Flow values as turndown parameter depending on nominal diameter.
SI units
US units
Accuracy of outputs
The outputs have the following base accuracy specifications.
Pulse/frequency output
o.r. = of reading
Accuracy Max. ±50 ppm o.r. (across the entire ambient temperature range)
66 Endress+Hauser
Dosimass Technical data
≥ 0.75 0.2
≥ 1.5 0.1
≥ 3.0 0.05
Density (liquids)
±0.00025 g/cm3
Temperature
±0.25 °C ± 0.0025 · T °C (±0.45 °F ± 0.0015 · (T–32) °F)
Influence of medium A difference between the calibration pressure and process pressure does not affect
pressure accuracy.
16.7 Installation
"Mounting requirements"
16.8 Environment
Ambient temperature
range Temperature tables
Observe the interdependencies between the permitted ambient and fluid
temperatures when operating the device in hazardous areas.
For detailed information on the temperature tables, see the separate document
entitled "Safety Instructions" (XA) for the device.
Endress+Hauser 67
Technical data Dosimass
16.9 Process
Medium pressure range max. 40 bar (580 psi), depending on process connection
(nominal pressure)
Secondary containment The housing does not have pressure vessel classification.
pressure rating For the dimensions and installation lengths of the device, see the "Technical
Information" document, "Mechanical construction" section
Flow limit Select the nominal diameter by optimizing between the required flow range and
permissible pressure loss.
For an overview of the full scale values for the measuring range, see the "Measuring
range" section → 62
• The minimum recommended full scale value is approx. 1/20 of the maximum full scale
value
• For the most common applications, 20 to 50 % of the maximum full scale value can be
considered ideal
• Select a low full scale value for abrasive media (e.g. liquids with entrained solids): Flow
velocity < 1 m/s (< 3 ft/s).
Pressure loss To calculate the pressure loss, use the Applicator sizing tool → 73
68 Endress+Hauser
Dosimass Technical data
Heating NOTICE
Electronics can overheat due to elevated ambient temperature!
‣ Observe maximum permitted ambient temperature for the transmitter .
‣ Depending on the fluid temperature, take the device orientation requirements into
account .
NOTICE
Danger of overheating when heating
‣ Ensure that the temperature at the lower end of the transmitter housing does not
exceed 80 °C (176 °F)
‣ Ensure that convection takes place on a sufficiently large scale at the transmitter neck.
‣ Ensure that a sufficiently large area of the housing support remains exposed. The
uncovered part serves as a radiator and protects the electronics from overheating and
excessive cooling.
Heating options
If a fluid requires that no heat loss should occur at the sensor, users can avail of the
following heating options:
• Electrical heating, e.g. with electric band heaters
• Via pipes carrying hot water or steam
• Via heating jackets
Vibrations The high oscillation frequency of the measuring tubes ensures that the correct operation of
the measuring system is not influenced by plant vibrations.
Endress+Hauser 69
Technical data Dosimass
Design, dimensions For the dimensions and installation lengths of the device, see the "Technical
Information" document, "Mechanical construction" section
8 3.5
15 4.0
25 4.5
Weight in US units
³⁄₈ 7.7
½ 8.8
1 9.9
Device plugs
Sensor housing
• Acid and alkali-resistant outer surface
• Stainless steel 1.4301 (304)
Measuring tubes
Stainless steel, 1.4539 (904L)
Process connections
• Flanges according to EN (DIN):
Stainless steel, 1.4404 (316/316L)
• Flanges according to DIN 32676:
Stainless steel, 1.4435 (316L)
• All other process connections:
Stainless steel, 1.4404 (316L)
List of all available process connections → 71
70 Endress+Hauser
Dosimass Technical data
Seals
Welded process connections without internal seals
16.11 Operability
Local operation This device cannot be operated locally using a display or operating elements.
Endress+Hauser 71
Technical data Dosimass
2 5
+
1 4
-
A0007228
1 Supply voltage 24 V DC
2 Service adapter
3 Dosimass
4 Commubox FXA291
5 Computer with "FieldCare" or "DeviceCare" operating tool
The service adapter, cable and Commubox FXA291 are not included in the delivery.
These components can be ordered as accessories → 60.
CE mark The measuring system is in conformity with the statutory requirements of the applicable
EC Directives. These are listed in the corresponding EC Declaration of Conformity along
with the standards applied.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications
and Media Authority (ACMA)".
Ex approval The devices are certified for use in hazardous areas and the relevant safety instructions are
provided in the separate "Safety Instructions" (XA) document. Reference is made to this
document on the nameplate.
Pressure Equipment • With the PED/G1/x (x = category) marking on the sensor nameplate, Endress+Hauser
Directive confirms compliance with the "Essential Safety Requirements" specified in Annex I of the
Pressure Equipment Directive 97/23/EC.
• Devices not bearing this marking (PED) are designed and manufactured according to
good engineering practice. They meet the requirements of Art.3 Section 3 of the
Pressure Equipment Directive 97/23/EC. The range of application is indicated in tables 6
to 9 in Annex II of the Pressure Equipment Directive.
72 Endress+Hauser
Dosimass Technical data
16.13 Accessories
Overview of accessories available for order → 60
Dosimass KA00043D
Dosimass GP01050D
Technical Information
Dosimass TI00065D
Endress+Hauser 73
Technical data Dosimass
ATEX Ex nA XA00079D
cCSAus FES0232
74 Endress+Hauser
Dosimass Index
Index
A Previous diagnostic event . . . . . . . . . . . . . . . . . . . 52
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 Display values
Adapting the diagnostic behavior . . . . . . . . . . . . . . . . 50 For locking status . . . . . . . . . . . . . . . . . . . . . . . . . 45
Ambient temperature Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Influence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67 Document
Ambient temperature range . . . . . . . . . . . . . . . . . . . 19 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8, 62 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 Document function . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Down pipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
C
C-Tick symbol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 E
CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9, 72 Electrical connection
Certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 Degree of protection . . . . . . . . . . . . . . . . . . . . . . . 26
Check Measuring device . . . . . . . . . . . . . . . . . . . . . . . . . 24
Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Electromagnetic compatibility . . . . . . . . . . . . . . . . . . 68
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Endress+Hauser services
Checklist Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Post-connection check . . . . . . . . . . . . . . . . . . . . . 27 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Post-installation check . . . . . . . . . . . . . . . . . . . . . 23 Environment
Cleaning Storage temperature . . . . . . . . . . . . . . . . . . . . . . . 67
Cleaning in place (CIP) . . . . . . . . . . . . . . . . . . . . . 57 Error messages
Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . 57 see Diagnostic messages
Interior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . 57 Event history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Sterilization in place (SIP) . . . . . . . . . . . . . . . . . . . 57 Events list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Cleaning in place (CIP) . . . . . . . . . . . . . . . . . . . . . . . . 68 Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Extended order code
Connecting cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Connecting the measuring device . . . . . . . . . . . . . . . . 26 Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Connecting the transmitter . . . . . . . . . . . . . . . . . . . . 26 F
Connection Field of application
see Electrical connection Residual risks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Connection conditions . . . . . . . . . . . . . . . . . . . . . . . . 24 FieldCare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Current consumption . . . . . . . . . . . . . . . . . . . . . . . . . 65 Device description file . . . . . . . . . . . . . . . . . . . . . . 31
Establishing a connection . . . . . . . . . . . . . . . . . . . 29
D Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . 9 User interface . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Degree of protection . . . . . . . . . . . . . . . . . . . . . . 26, 67 Filtering the event logbook . . . . . . . . . . . . . . . . . . . . 53
Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 Firmware
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Release date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Device components . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Device description files . . . . . . . . . . . . . . . . . . . . . . . 31 Firmware history . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Device documentation Flow direction . . . . . . . . . . . . . . . . . . . . . . . . . . . 17, 23
Supplementary documentation . . . . . . . . . . . . . . . . 7 Flow limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Device locking, status . . . . . . . . . . . . . . . . . . . . . . . . 45 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Device name Functions
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 see Parameter
DeviceCare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Diagnostic information G
Design, description . . . . . . . . . . . . . . . . . . . . . . . . 50 Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
FieldCare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
H
Hygienic compatibility . . . . . . . . . . . . . . . . . . . . . . . . 72
Remedy information . . . . . . . . . . . . . . . . . . . . . . . 50
Diagnostic list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
I
Display
Identifying the measuring device . . . . . . . . . . . . . . . . 12
Current diagnostic event . . . . . . . . . . . . . . . . . . . . 52
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . 12
Endress+Hauser 75
Index Dosimass
76 Endress+Hauser
Dosimass Index
Endress+Hauser 77
www.addresses.endress.com