The document outlines the calibration procedures for instruments and equipment. It states that a calibration schedule will be prepared annually listing the calibration frequency for all devices. Instruments should be calibrated in their location of use if possible. Any device found to be out of calibration must be labeled as such, have its history updated, and be reported so corrective actions can be taken until calibration is restored. Calibration reports must be prepared, reviewed, and documented. Re-calibration may also be needed after maintenance or relocation of a device.

Uploaded by

Yousif
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
50% found this document useful (2 votes)
1K views1 page

In House Calibration

The document outlines the calibration procedures for instruments and equipment. It states that a calibration schedule will be prepared annually listing the calibration frequency for all devices. Instruments should be calibrated in their location of use if possible. Any device found to be out of calibration must be labeled as such, have its history updated, and be reported so corrective actions can be taken until calibration is restored. Calibration reports must be prepared, reviewed, and documented. Re-calibration may also be needed after maintenance or relocation of a device.

Uploaded by

Yousif
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 1

In House Calibration

5.2.1      A calibration schedule shall be prepared in the beginning of each year for all instruments /

equipment which to be calibrated as per their individual frequencies.

5.2.2      Calibration of instruments shall be done as far as possible in the location of its use.

5.2.3      In case the instrument / equipment is found out of calibration, it shall be immediately labeled

“UNDER MAINTENANCE”, update the Instrument History Card as per Appendix II and inform to

Quality Assurance for appropriate action.

5.2.4      In case the instrument is out of calibration it gets promptly reported and steps taken to remove

it from service.

5.2.5  Instrument / equipment shall not be further used till complies the calibration requirements.

5.2.6      After successful calibration, the instrument shall be labelled with a Green calibration sticker

indicating the Instrument name, date of calibration, Tag No. / ID No., signature of the person who

carried out the calibration and next due date of calibration.

5.2.7      Calibration reports shall be prepared by the person carrying out the activity and shall

be reviewed by designated person and documented.

5.2.8      Maintenance or moving of an instrument can alter calibration status and accuracy of data.

Hence after any such activity with the instrument shall be assed and recalibrated (whenever required)

and document the calibration record.

5.2.9      Re-calibration shall be done in case of any maintenance or re-installation of instrument /

equipment (wherever applicable) and record shall be maintained.