0% found this document useful (0 votes)
330 views76 pages

Instructions For Use: Avanti J-26S XP

Uploaded by

saddam
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
330 views76 pages

Instructions For Use: Avanti J-26S XP

Uploaded by

saddam
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 76

Instructions For Use

Avanti J-26S XP

High-Performance Centrifuge

PN B10087AE
April 2019

Beckman Coulter, Inc.


250 S. Kraemer Blvd.
Brea, CA 92821 U.S.A.
Avanti J-26S XP
High-Performance Centrifuge
PN B10087AE (April 2019)

©2019 Beckman Coulter, Inc.


All rights reserved

Beckman Coulter, the stylized logo, and the Beckman


Coulter product and service marks mentioned herein are
trademarks or registered trademarks of Beckman Coulter,
Inc. in the United States and other countries.

Contact Us
If you have any questions, contact our Customer Support
Center.

• Worldwide, find us via our website at


www.beckman.com/support/technical.
• In the USA and Canada, call us at 1-800-369-0333.
• Outside of the USA and Canada, contact your local
Beckman Coulter Representative.

Find us on the World Wide Web at: www.beckman.com

Beckman Coulter Eurocenter S.A.


22, rue Juste-Olivier
Case Postale 1044
CH - 1260 Nyon 1, Switzerland
Tel: +41 (0) 22 365 36 11

Glossary of Symbols is available at


beckman.com/techdocs (PN C24689).

Original Instructions
Revision History
For labeling updates, go to www.beckman.com/techdocs and download the latest version of the manual or
system help for your instrument.

Issue AD, 05/2016


Updates were made to the following sections: Multi Compliance Label; CHAPTER 1, Control Features.

Issue AE, 04/2019


Updates were made to the following section: CHAPTER 1, Control Features

Note: Changes that are part of the most recent revision are indicated in text by a bar in the margin of the
amended page.

PN B10087AE iii
Revision History

iv PN B10087AE
Safety Notice
Read all product manuals and consult with Beckman Coulter-trained personnel before attempting
to operate the instrument. Do not attempt to perform any procedure before carefully reading all
instructions. Always follow product labeling and manufacturer’s recommendations. If in doubt as
to how to proceed in any situation, contact your Beckman Coulter Representative.

Alerts for Danger, Warning, Caution, Important, and Note

DANGER
DANGER indicates an imminently hazardous situation which, if not avoided,
will result in death or serious injury.

WARNING
WARNING indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.

CAUTION
CAUTION indicates a potentially hazardous situation, which, if not avoided,
may result in minor or moderate injury. It may also be used to alert against unsafe
practices.

IMPORTANT IMPORTANT is used for comments that add value to the step or procedure being performed.
Following the advice in the Important adds benefit to the performance of a piece of equipment or to a
process.

NOTE NOTE is used to call attention to notable information that should be followed during installation, use,
or servicing of this equipment.

Safety During Installation and/or Maintenance

This instrument is designed to be installed by a Beckman Coulter Field Service representative.


Installation by anyone other than authorized Beckman Coulter personnel invalidates any warranty
covering the instrument. Also, if the instrument needs to be moved, a Beckman Coulter Field
Service representative must reinstall and re-level the instrument in its new location.

Any servicing of this equipment that requires removal of any covers can expose parts which involve
the risk of electric shock or personal injury. Make sure that the power switch is turned off and the
instrument is disconnected from the main power source by removing the Mains (power) plug from
the outlet receptacle, and refer such servicing to qualified personnel.

PN B10087AE v
Safety Notice
Electrical Safety

Be sure to use the anchoring system to secure the centrifuge in place. The anchoring system is
designed to reduce the possibility of injury or damage that could result from instrument movement
in the event of a major rotor mishap.

Do not replace any centrifuge components with parts not specified for use on this instrument.

Electrical Safety

To reduce the risk of electrical shock, this instrument uses a three or five-wire electrical cord and
plug to connect this equipment to earth-ground. To preserve this safety feature:

• Make sure that the matching wall outlet receptacle is properly wired and earth-grounded.
Check that the line voltage agrees with the voltage listed on the name-rating plate affixed to
the centrifuge.
• Never use a three-to-two wire plug adapter.
• Never use a two-wire extension cord or a two-wire non-grounding type of multiple-outlet
receptacle strip.

Do not place containers holding liquid on or near the chamber door. If they spill, liquid may get into
the instrument and damage electrical or mechanical components.

Safety Against Risk of Fire

This instrument is not designed for use with materials capable of developing flammable or
explosive vapors. Do not centrifuge such materials (for example, chloroform or ethyl alcohol) in
this instrument nor handle or store them near the centrifuge.

Mechanical Safety

For safe operation of the equipment, observe the following:

• Use only the Beckman Coulter rotors and accessories designed for use in this instrument.
• Do not exceed the maximum rated speed of the rotor in use.
• NEVER attempt to slow or stop a rotor by hand.
• Do not lift or move the centrifuge while a rotor is spinning.
• NEVER attempt to override the door interlock system while the rotor is spinning.
• Do not lean on the centrifuge or place items on it while it is operating.

vi PN B10087AE
Safety Notice
Chemical and Biological Safety

Chemical and Biological Safety

Normal operation may involve the use of solutions and test samples that are pathogenic, toxic, or
radioactive. Such materials should not be used in this instrument unless all necessary safety
precautions are taken.
• Handle body fluids with care because they can transmit disease. No known test offers complete
assurance that they are free of micro-organisms. Some of the most virulent — Hepatitis (B and
C) and HIV (I–V) viruses, atypical mycobacteria, and certain systemic fungi — further emphasize
the need for aerosol protection. Handle other infectious samples according to good laboratory
procedures and methods to prevent spread of disease. Because spills may generate aerosols,
observe proper safety precautions for aerosol containment. Do not run toxic, pathogenic, or
radioactive materials in this centrifuge without taking appropriate safety precautions. Biosafe
containment should be used when Risk Group II materials (as identified in the World Health
Organization Laboratory Biosafety Manual) are handled; materials of a higher group require
more than one level of protection.
• Dispose of all waste solutions according to appropriate environmental health and safety
guidelines.
It is your responsibility to decontaminate the instrument and accessories before requesting service
by Beckman Coulter Field Service.

PN B10087AE vii
Safety Notice
RoHS Notice

RoHS Notice

These labels and materials declaration table (the Table of Hazardous Substance's Name and
Concentration) are to meet People’s Republic of China Electronic Industry Standard SJ/T11364-2006
“Marking for Control of Pollution Caused by Electronic Information Products” requirements.

China RoHS Caution Label


This label indicates that the electronic information product contains certain toxic or hazardous
substances. The center number is the Environmentally Friendly Useful Period (EFUP) date, and
indicates the number of calendar years the product can be in operation. Upon the expiration of the
EFUP, the product must be immediately recycled. The circling arrows indicate the product is
recyclable. The date code on the label or product indicates the date of manufacture.

China RoHS Environmental Label


This label indicates that the electronic information product does not contain any toxic or hazardous
substances. The center ‘e’ indicates the product is environmentally safe and does not have an
Environmentally Friendly Useful Period (EFUP) date. Therefore, it can safely be used indefinitely.
The circling arrows indicate the product is recyclable. The date code on the label or product
indicates the date of manufacture.

viii PN B10087AE
Contents

Revision History, iii


Safety Notice, v
Alerts for Danger, Warning, Caution, Important, and Note, v
Safety During Installation and/or Maintenance, v
Electrical Safety, vi
Safety Against Risk of Fire, vi
Mechanical Safety, vi
Chemical and Biological Safety, vii
RoHS Notice, viii
China RoHS Caution Label, viii
China RoHS Environmental Label, viii

Introduction, xv
Certification, xv
Scope of Manual, xv
Conventions, xv
Typographic Conventions, xvi
CFC-Free Centrifugation, xvi
Recycling Label, xvi
Multi Compliance Label, xvii

CHAPTER 1: Description, 1-1


Introduction, 1-1
Centrifuge Function and Safety Features, 1-1
Centrifuge Function, 1-1
Safety Features, 1-2
Centrifuge Chassis, 1-2
Housing and Door, 1-2
Rotor Chamber, 1-2
Drive, 1-2
Friction Reduction System (FRS), 1-3
Temperature Sensing and Control, 1-3
Overtemp System, 1-3
Name Rating Plate, 1-3

ix
Contents

Controls and Indicators, 1-4


Power Switch, 1-4
Control Panel, 1-4
Digital Display, 1-5
Knobs and Touch Keys, 1-5
System Keys, 1-7
Specifications, 1-7
Control Features, 1-7
Operational Features, 1-8
Physical Data, 1-8
Available Rotors, 1-9

CHAPTER 2: Operation, 2-1


Introduction, 2-1
Summary of Avanti J-26S XP Run Procedures, 2-1
Section and Loading, 2-2
Entering Run Parameters, 2-3
Selecting a Rotor, 2-3
Setting Run Speed, 2-4
Setting RPM, 2-4
Setting RCF, 2-5
Setting Run Time, 2-5
Timed Run, 2-5
Continuous (HOLD) Run, 2-6
Setting Run Temperature, 2-7
Setting Acceleration and Deceleration Rates, 2-7
Setting the Acceleration Rate, 2-8
Setting the Deceleration Rate, 2-9
Setting Up an Elutriation Run, 2-9
Starting a Run, 2-11
Changing Parameters During a Run, 2-11
Changing Time Settings, 2-12
Stopping a Run, 2-12
Unloading the Centrifuge, 2-13

CHAPTER 3: Troubleshooting, 3-1


Introduction, 3-1
User Messages, 3-1
Accessing the Rotor in Case of Power Failure, 3-5
JCF-Z Rotor Identification, 3-9

CHAPTER 4: Care and Maintenance, 4-1


Introduction, 4-1

x
Contents

Maintenance, 4-1
Cleaning, 4-2
Tube Breakage, 4-2
Decontamination, 4-3
Sterilization and Disinfection, 4-3
Replacing the Air Filter, 4-3
Circuit Breaker and Fuses, 4-4
Storage and Transport, 4-5
Storage, 4-5
Returning a Centrifuge, 4-5
Supply List, 4-6
Replacement Parts and Supplies, 4-6

APPENDIX A: Preinstallation Requirements, A-1


Preinstallation Requirements, A-1
Electrical Requirements, A-1
Single and Three-Phase Power Connections, A-2
Additional Requirements for Three-Phase Power
Connections, A-4
Space and Location Requirements, A-4
Securing the Centrifuge to the Floor, A-6
Bio-Safety Level 3 Installation, A-6
Non-Invasive Centrifuge Restraint System for Vinyl Floors, A-6
Using J2 Series Rotors in the Avanti J-26S XP, A-6
Checking for Rotor Drive Pins, A-6
Using the JA-18 Rotor, A-7
Using the JCF-Z Continuous Flow/Zonal Rotor, A-8
Zonal Bracket Kits, A-8
Ensuring Correct JCF-Z Continuous Flow/Zonal Rotor
Identification, A-8

APPENDIX B: Temperature Calibration Procedure, B-1


Introduction, B-1

Beckman Coulter, Inc.


Avanti J Series Centrifuge Warranty
Related Documents

xi
Illustrations

Illustrations

1.1 The Power Switch, 1-4


1.2 The Control Panel, 1-4
2.1 Acceleration and Deceleration Settings Graph, 2-8
2.2 Elutriation Controls, 2-9
3.1 Emergency Door Release Latch Access, 3-5
3.2 Manual Door Release, 3-6
3.3 Venting the Chamber Vacuum, 3-8
3.4 Front Panel Retaining Strip, 3-9
A.1 Single-Phase Electrical Requirements, A-3
A.2 Three-Phase “Y” Electrical Requirements, A-3
A.3 Correct and Incorrect Three-Phase Power Configurations, A-4
A.4 Rear View and Dimensions, A-5
A.5 Checking the Rotor for Drive Pins, A-7

xii
Tables

Tables

2.1 Acceleration Settings, 2-7


2.2 Deceleration Settings, 2-8
3.1 Diagnostic Message Chart, 3-2
A.1 Nominal Supply Voltage Ratings for the Avanti J-26S XP, A-2
A.2 Required Wire Connections, A-3

xiii
Tables

xiv
Introduction

Certification

Beckman Coulter Avanti J-26S XP centrifuges are manufactured in a facility that maintains
certifications to both ISO 9001:2008 and ISO 13485:2003. They have been designed and tested to be
compliant (when used with Beckman Coulter rotors) with the laboratory equipment requirements
of applicable regulatory agencies. Declarations of conformity and certificates of compliance are
available at www.beckman.com.

Scope of Manual

This manual is designed to familiarize centrifuge users and site engineers with the Avanti J-26S XP
centrifuge, its functions, specifications, operation, and routine care and maintenance.
Beckman Coulter recommends that you read this entire manual, especially the Safety Notice and
all safety-related information, before operating the instrument or performing instrument
maintenance.

• CHAPTER 1, Description contains an overall description of the centrifuge, including a


description of system controls and indicators, and system specifications.
• CHAPTER 2, Operation summarizes procedures for operating the centrifuge.
• CHAPTER 3, Troubleshooting lists system diagnostic messages, together with probable causes
and user-performed corrective actions.
• CHAPTER 4, Care and Maintenance contains procedures for routine maintenance, as well as a
brief list of supplies and replacement parts.
• APPENDIX A, Preinstallation Requirements contains instructions for preparing your site for
centrifuge installation.
• APPENDIX B, Temperature Calibration Procedure contains a procedure to use when temperature
control within 1°C is required.

NOTE If the centrifuge is used in a manner other than specified in this manual, the safety and performance
of this equipment could be impaired. Further, the use of any equipment other than that recommended
by Beckman Coulter has not been evaluated for safety. Use of any equipment not specifically
recommended in this manual is the sole responsibility of the user.

Conventions

Certain symbols are used in the product labeling to call out safety related and other important
information. These international symbols may also be displayed on the centrifuge and are
reproduced on the inside back cover of the manual.

PN B10087AE xv
Introduction
CFC-Free Centrifugation

Typographic Conventions
Certain typographic conventions are used throughout this manual to distinguish names of user
interface components, such as keys and displays.

• Control knob names (for example, ROTOR and SPEED) appear in bold type.
• Key names (for example, START and STOP) appear in bold type.
• Display field names (for example, TEMP°C or SPEED) appear in bold type.

CFC-Free Centrifugation

To ensure minimal environmental impact, no CFCs are used in the manufacture or operation of the
Avanti J-26S XP centrifuge.

Recycling Label

A28219-AA

This symbol is required in accordance with the Waste Electrical and Electronic Equipment (WEEE)
Directive of the European Union. The presence of this marking on the product indicates:

1. The device was put on the European market after August 13, 2005 and
2. The device is not to be disposed via the municipal waste collection system of any member state
of the European Union.

It is very important that customers understand and follow all laws regarding the proper
decontamination and safe disposal of electrical equipment. For Beckman Coulter products bearing
this label please contact your dealer or local Beckman Coulter office for details on the take back
program that will facilitate the proper collection, treatment, recovery, recycling and safe disposal
of the device.

xvi PN B10087AE
Introduction
Multi Compliance Label

Multi Compliance Label

• The RCM mark is intended for use on products that comply with Austrailian Communications
Media Authority (ACMA) EMC Requirements.
• Recycling — Refer to the Recycling Label section in this document.
• 169502 — This label indicates recognition by a Nationally Recognized Testing Laboratory
(NRTL) that the instrument has met the relevant product safety standards.

NOTE 169502 is applicable to North American models only.

• — A “CE” mark indicates that a product has been assessed before being placed on the
market, and has been found to meet European Union safety, health, and/or environmental
protection requirements.

PN B10087AE xvii
Introduction
Multi Compliance Label

xviii PN B10087AE
CHAPTER 1
Description

Introduction

This section describes the Avanti J-26S XP centrifuge components and their functions. It also describes
system safety features and centrifuge controls and displays. Refer to the applicable rotor manual for rotor
descriptions.

Centrifuge Function and Safety Features

Centrifuge Function
The Avanti J-26S XP is a refrigerated centrifuge that generates centrifugal forces required for a wide
variety of applications. Together with the Beckman Coulter rotors designed for use in this
centrifuge, applications include:
• Routine processing such as sample preparations, pelleting, extractions, purifications,
concentrations, phase separations, and spin column and spin filter centrifugations.
• Rapid sedimentation of protein precipitates, large particles, and cell debris.
• Preparation of subcellular organelles such as mitochondria, nuclei, chloroplasts, and crude
microsomes.
• Separation of blood cells and cellular components.
• Pelleting of prokaryotic and eukaryotic cells.
• Gradient separation, for example, Ficoll-Hypaque and Percoll.
• Nucleic acid precipitation.
• Virus isolation.
• Bacteriophase isolation.

The Avanti J-26S XP centrifuge is microprocessor-controlled, providing interactive operation.* The


centrifuge features a brushless switched-reluctance drive motor,† an automatic rotor identification
system, FRS (friction reduction system), vacuum control circuitry, and a temperature control
system with automatic temperature compensation unique for each compatible rotor.

The user interface has a digital display, which can be toggled to show either actual centrifuge values
or user-entered setup values, knobs for parameter entry, and touch keys. Diagnostic messages on
the control panel will light to alert the operator to conditions that may need attention.

* Avanti J-26 XP software and firmware ©2005–2012 Beckman Coulter, Brea, CA, 92821.
† Manufactured under license from Switched Reluctance Drives Limited, Harrogate, U.K.

PN B10087AE 1-1
Description
Centrifuge Chassis

Safety Features
Avanti J-26S XP centrifuges have been designed and tested to operate safely indoors at altitudes up
to 2000 m (6562 ft).

Safety features include the following.

• An electromechanical door lock system prevents operator contact with spinning rotors and
prevents run initiation unless the door is shut and locked. The door locks when START is
pressed, or when the POWER switch is turned off.
• A steel casing surrounds the rotor chamber to provide operator protection in the unlikely event
of a rotor mishap.
• An automatic rotor identification system detects which rotor is installed and prevents the rotor
from running above its maximum rated speed.
• An imbalance detector monitors the system during operation, causing automatic shutdown if
rotor loads are severely out of balance.

Centrifuge Chassis

Housing and Door


The instrument control housing, door assembly, and cover panels are made from a highly-durable
unfinished plastic. The control panel is covered by a protective overlay made of coated
polycarbonate.

The door is opened by stepping on a foot pedal, which is located at the bottom right front of the
instrument. The door is hinged at the back left to open at a 60-degree angle to the centrifuge side
panels, providing clearance for loading and unloading of the centrifuge.

In the event of a power failure, the door can be manually unlocked for sample recovery (see
CHAPTER 3, Troubleshooting).

Rotor Chamber
The rotor chamber is made of stainless steel to resist corrosion. A rubber gasket around the
chamber opening ensures sealing. (Instrument gaskets have not been qualified as bioseals for
aerosol containment.)

Drive
The drive spindle is directly driven by a brushless, high-torque, switched-reluctance motor. The
instrument’s resilient suspension minimizes disturbance of the sample during acceleration and
deceleration, and reduces damage to the drive spindle if an imbalance occurs during centrifugation.

1-2 PN B10087AE
Description
Name Rating Plate 1

Friction Reduction System (FRS)


The friction reduction system (FRS) uses a mechanical rotary vane vacuum pump to reduce
chamber pressure to approximately one-quarter atmosphere (190 mm Hg). The pump turns on after
the run is started, before rotor friction reaches a high level. When the required vacuum level is
reached, the pump turns off. Vacuum in the chamber is vented during rotor deceleration.

Temperature Sensing and Control


The temperature control system is cooled by circulation of a non-CFC-based refrigerant. The
temperature control system is activated when the centrifuge power is on and when the door is
closed.

A thermistor in the rotor chamber continuously monitors the chamber temperature. The system
calculates the chamber temperature required to maintain the set rotor temperature,  2°C.*
Although the chamber temperature fluctuates during operation, the rotor’s large mass keeps the
sample temperature substantially constant. At the end of a run, the system continues controlling
the temperature to prevent freezing or overheating (more than 4°C over set temperature) of
the sample.

Overtemp System
An overtemp (over temperature) system provides sample protection and safety for the user.

• During a run, if the system temperature rises 4°C above the set temperature, the TEMP
diagnostic will light. The run will continue, unless stopped by the user. If the temperature rises
8°C above the set temperature, the system will shut down (using maximum brake).
• The system will always shut down (using maximum brake) if the system temperature reaches
50°C. The centrifuge cannot be restarted until the chamber has cooled.

Name Rating Plate

The name rating plate is affixed to the rear of the centrifuge. Check that the line voltage agrees with
the voltage listed on this name rating plate before connecting the centrifuge. Always mention the
serial number and the model number shown when corresponding with Beckman Coulter regarding
your centrifuge.

* Refer to the applicable rotor manual for specific temperature information.

PN B10087AE 1-3
Description
Controls and Indicators

Controls and Indicators

Power Switch
The power switch is located below the control panel (see Figure 1.1). This two-position rocker
switch (I, on; O, off) controls electrical power to the centrifuge.

Figure 1.1 The Power Switch

Control Panel
The control panel (Figure 1.2) is mounted at an angle on the centrifuge top rear for easy visibility
and access. It contains a digital display for actual and setup values, knobs and touch keys for
parameter entry, and additional touch keys for system control.

Figure 1.2 The Control Panel

ROTOR ID SPEED RPM/RCF TIME HR:MIN TEMP °c SET


ACTUAL
EXP
SET/
RPM rad 2 / s e c ACTUAL
AUTO RCG(xg) HOLD
1
ROTOR SPEED TIME TEMP
JA
˚C
JS JLA
JSP JV

JCF JE

RPM / TIME/
RCF HOLD

FRS SYS
ROTOR POWER
IMBAL DRIVE
MAX
SPEED REFR
MAX SLOW
DOOR TEMP
SLOW OFF

ACCEL DECEL CLEAR


START STOP

1. Digital Display

1-4 PN B10087AE
Description
Controls and Indicators 1

Digital Display
The display shows current status of the instrument whenever the power is turned on.

• The display shows both actual centrifuge values and user-set parameters. The SET/ACTUAL key
is used to toggle between the two views.
• During a run, actual values are displayed. If you turn a knob (other than the TIME knob) during
operation, the display will change to show the set values. If no parameters are entered or
changed for 5 seconds, the display will revert to showing actual values.
• During a timed run, the TIME display shows the remaining run time in hours and minutes. The
TIME display begins counting down when START is pressed and continues counting down to 0,
when deceleration begins. When the rotor decelerates to 0 RPM, End flashes in the TIME display.
• When HOLD is selected (no run time is specified), the TIME display shows the time elapsed since
START was pressed. After 3 hours (180 minutes) have elapsed, the timer will reset to 0 and
continue counting elapsed time until STOP is pressed.

Knobs and Touch Keys


The rotor name, speed, time, and temperature settings are entered by turning the appropriate knob
until the required value is displayed.

Touch keys allow the user to start and stop the centrifuge, and to select speed mode (RPM or RCF),
time mode (specific run length or hold for continuous operation), and acceleration and deceleration
rates, and to toggle the display between set and actual values.

ROTOR Knob Used to select the rotor in use. As the knob is turned, the name of each
Beckman Coulter rotor that can be run in the centrifuge appears in succession in
ROTOR the ROTOR ID display, and the LED by each rotor type abbreviation
JS
JA
JLA
(JCF: continuous flow rotors; JSP: special type rotors; JS: swinging bucket rotors;
JSP JV JA: fixed angle rotors; JLA: lightweight fixed angle rotors; JV: vertical tube rotors;
JCF JE JE: elutriator rotors) lights when appropriate.

SPEED Selection The RPM/RCF key is pressed to select the speed mode (RPM or RCF). After the
mode is selected, the SPEED knob functions as follows.
• In RPM mode, each SPEED knob increment is 100 RPM at speeds up to
10,000 RPM. Above 10,000 RPM, each knob increment is 500 RPM.
• In RCF mode, at settings less than or equal to 3000  g, each SPEED knob
increment is equivalent to incrementing/decrementing the speed by 100  g.
At settings greater than 3000  g, each SPEED knob increment is equivalent to
incrementing/decrementing the speed by 250  g.
• Speed increments for elutriator rotors are either 10 RPM or 10  g.

PN B10087AE 1-5
Description
Controls and Indicators

TIME Selection Two time modes are available. The HR:MIN mode is used for runs of specified
lengths. The HOLD mode is used for continuous runs of unspecified lengths. The
HOLD key is used to toggle between the two modes.
In HR:MIN mode:
• The TIME knob is used to set run time, which can be set up to 3 hours and
0 minutes (180 minutes).
• Each TIME knob increment is 1 minute. A full turn of the knob equals
60 minutes.
• The time setting can be changed during a run.
• During a run, the time remaining in the run is displayed. When 0 is reached, the
run ends and the rotor decelerates to a stop.
In HOLD mode:
• When HOLD mode is selected, the time display shows the word HOLD.
• During a run, the time elapsed is displayed. When the display reaches 3 hours
and 0 minutes, the system automatically resets the display to 0 and begins
counting the elapsed time again. The run will continue until STOP is pressed.
• If you switch from HOLD mode to HR:MIN mode during a run, the system
remembers the previous time setting. Turning the TIME knob increments or
decrements the set time from that point.

TEMPERATURE Used to select the rotor temperature, from –10 to +40°C.


Knob • The minimum and maximum allowable set temperatures depend on the set
speed and the rotor in use. If a temperature is entered that cannot be achieved
by the installed rotor at the set speed, the TEMP°C field will flash.
• The maximum achievable rotor temperature depends on the frictional heat
generated inside the chamber during operation. At low run speeds or low
ambient temperature, the centrifuge may not be able to achieve some higher
temperatures.

ACCELERATION Used to select either maximum or slow acceleration. An indicator light above the
Key ACCEL key displays the selection. Acceleration rates are described in Table 2.1).

MAX
SLOW
ACCEL

DECELERATION Used to select maximum or slow deceleration, or off (no brake). An indicator light
Key above the DECEL key displays the selection. Deceleration rates are described in
Table 2.2.
MAX
SLOW
OFF
DECEL

1-6 PN B10087AE
Description
Specifications 1

System Keys

START Pressed to begin a run. When START is pressed, the display immediately shows the
actual centrifuge values. The green START light flashes during acceleration until set
speed is reached, and then remains on continuously until deceleration begins.

STOP Pressed to end a run; the centrifuge decelerates to a complete stop according to the
deceleration setting. The red STOP light blinks during deceleration. Deceleration can be
terminated and the centrifuge restarted by pressing START.

CLEAR Pressed to stop the flashing of a diagnostic indicator light. See CHAPTER 3,
Troubleshooting, for information on diagnostic messages.

SET/ACTUAL Pressed to toggle the display view between set and actual values. The indicator for the
selected view will be lit.

Specifications

Only values with tolerances or limits are guaranteed data. Values without tolerances are
informative data, without guarantee.

Control Features

Specifications Description

Speed • Setting range:


— 100 to 26,000 RPM (in 100-RPM increments below 10,000 RPM,
500-RPM increments above 10,000 RPM), or equivalent RCF (in
100  g increments)
• Elutriation speed setting range:
— 0 to 5000 RPM (in 10-RPM or 10  g increments)
• Speed display:
— from 0 to 10,000 RPM, display indicates actual rotor speed  10 RPM;
from 10,000 to 26,000 RPM, indicates actual rotor speed  0.1%, or
equivalent RCF

Speed Control • Low speed (2000 RPM) control = ±10 RPM


• High speed (Max/Rotor) control = ±0.1% RPM

Time • Setting range:


— to 180 minutes or continuous (hold)
• Time display:
— Indicates time remaining (timed run) or time elapsed (hold run)

PN B10087AE 1-7
Description
Specifications

Specifications Description

Temperature • Setting range:


— –10 to +40°C (in 1°C increments)
• Accuracy:
— rotor temperature controlled to within  2°C of set temperature (after
equilibration)a
• Ambient temperature range:
— 16 to 38°C (60 to 100°F)b
• Cooling fluid:
— refrigerant R452A (HFO)

Acceleration maximum or slow

Deceleration maximum, slow, or off


a. During transient conditions, such as acceleration and deceleration, rotor temperature may be outside this range. Refer to
applicable rotor manuals for specific rotor operating range information.
b. To reach temperatures above ambient, the centrifuge is dependent on the frictional heat generated inside the chamber
during operation. At low run speeds or low ambient temperatures, the centrifuge may not be able to achieve some higher
temperatures.

Operational Features

Specifications Description

Door 6.1-cm (2.4-in.) thick plastic resin with steel plate

Rotor Chamber Diameter 51.3 cm (20 in.)

Friction Reduction 190 mm (7.5 in.) Hg


System (FRS)

Physical Data

Specification Description

Width 71 cm (28 in.)


Depth • 86 cm (34 in.)
• including air diverter extending from back panel
— 102 cm (40.25 in.)
Height • with door closed
— 86 cm (34 in.)
• to top of control head
— 116 cm (45.5 in.)
• to top of open door
— 149 cm (58.5 in.)

1-8 PN B10087AE
Description
Available Rotors 1

Specification Description

Weight 290 kg (640 lb)


Clearances • sides
(for adequate — 7.6 cm (3 in.)
ventilation)
• back (place air diverter against the wall)
— 16 cm (6.25 in.)
Surface Finish polyester control panel with polycarbonate coating overlay; uncoated plastic
on door and covers; acrylic baking enamel on metal surfaces
Electrical Requirements • 200/208/240-V, single-phase instrument
— 180–264 VAC, 30 A, 50/60 Hz
• 230-V, single-phase instrument
— 180–264 VAC, 30 A, 50 Hz
• 220/380-V plus neutral, three-phasea instrument
— 313–457 VAC plus neutral, 16 A, 50 Hz
Electrical Supply Class I
Maximum Heat 2 kW (~6900 BTu/h)
dissipation into Room
Under Steady-State
Conditions
Humidity restrictions <95% (noncondensing)
Noise Level 0.91 m (3 ft) 57 dBa
in front of instrument at
26,000 RPM
Installation II
(overvoltage) Category
Pollution Degree 2b
a. Unbalanced three-phase. Split for single-phase operation internally.
b. Normally, only nonconductive pollution occurs; occasionally, however, a temporary conductivity caused by condensation
must be expected.

Available Rotors

Refer to the applicable rotor manual for complete rotor descriptions.

PN B10087AE 1-9
Description
Available Rotors

Rotor Profile Description Rotor Max Max Max Capacity Rotor


Code RPMa RCF Manual
( g) Number

JA-30.50 Ti b 30.50 26,000 c 81,800 8  50 mL J-TB-070


Fixed Angle, 34°
(8 place)
rmax = 108 mm

JA-25.50 Ti 25.50 25,000 75,600 8  50 mL J-TB-056


Fixed Angle, 34°
(8 place)d
rmax = 108 mm

JA-25.15 Ti 25.15 25,000 24 15 mL J-TB-057


Fixed Angle, 25°
(24 place)d
rmax = 106 mm 74,200
(outer row)
rmax = 86 mm 60,200
(inner row)

JA-21 21 21,000 50,400 18  10 mL J-TB-002


Fixed Angle, 40°
(18 place)
rmax = 102 mm

JA-20.1 20.1 20,000 32  15 mL J-TB-022


Fixed Angle, 23°
(32 Place)
rmax = 115 mm 51,500
(outer row)
rmax = 98 mm 43,900
(inner row)

JA-20 20 20,000 48,400 8  50 mL J-TB-003


Fixed Angle, 34°
(8 place)
rmax = 108 mm

JA-14.50 14.50 14,000 35,000 16  50 mL B32164


Fixed Angle, 35°
(16 place)e
rmax = 160 mm

1-10 PN B10087AE
Description
Available Rotors 1

Rotor Profile Description Rotor Max Max Max Capacity Rotor


Code RPMa RCF Manual
( g) Number

JA-18.1 18.1 J-TB-037


Fixed Angle
(24 place)
45° adapter 18,000 42,100 24  1.8 mL
rmax = 116 mm
25° adapter 17,000 36,300 24  1.8 mL
rmax = 112 mm

JA-18 18 18,000 47,900 10  100 mL J-TB-035


Fixed Angle, 23°
(10 place)
rmax = 132 mm

JA-17 17 17,000 39,800 14  50 mL J-TB-017


Fixed Angle, 25°
(14 place)
rmax = 132 mm

JLA-16.250 16.250 16,000 38,400 6  250 mL J-TB-072


Fixed Angle, 25°
(6 place)
rmax = 134 mm

JA-14 14 14,000 30,100 6  250 mL J-TB-004


Fixed Angle, 25°
(6 place)
rmax = 137 mm

F14BCI-14x50cy F50C 14,000 33,500 14  50 mL —


Fixed Angle, 34°
(14 place)
rmax = 153 mm

F14BCI-6x250y F250 14,000 30,000 6  250 mL —


Fixed Angle, 23°
(6 place)
rmax = 134 mm

JA-12 12 12,000 23,200 12  50 mL J-TB-051


Fixed Angle, 35°
(12 place)
rmax = 144 mm

PN B10087AE 1-11
Description
Available Rotors

Rotor Profile Description Rotor Max Max Max Capacity Rotor


Code RPMa RCF Manual
( g) Number

JA-10 10 10,000 17,700 6  500 mL J-TB-006


Fixed Angle, 25°
(6 place)
rmax = 158 mm

JLA-10.500 10.500 10,000 18,600 6  500 mL J-TB-048


Fixed Angle, 20°
(6 place)
rmax = 166 mm

F10BCI-6x500y F500 10,000 17,696 6  500 mL —


Fixed Angle, 23° (previ-
(6 place) ously
rmax = 158 mm 10.1)

JLA-9.1000 9.1000 9,000 16,800 4  1000 mL J-TB-073


Fixed Angle, 20°
(4 place)
rmax = 185 mm

JLA-8.1000 8.1000 8,000 15,970 6  1000 mL J-TB-073


Fixed Angle, 20°
(6 place)
rmax = 222.8 mm

JS-13.1 13.1 13,000 26,500 6  50 mL J-TB-036


Swinging Bucket
(6 place)
rmax = 140 mm

JS-24.38 24.38 10,000f 18,000 6  38.5 mL J-TB-058


Swinging Bucket
(6 place)
rmax = 161 mm

JS-24.15 24.15 10,000f 19,200 6  15 mL J-TB-058


Swinging Bucket
(6 place)
rmax = 171.3 mm

1-12 PN B10087AE
Description
Available Rotors 1

Rotor Profile Description Rotor Max Max Max Capacity Rotor


Code RPMa RCF Manual
( g) Number

JS-7.5 S 7.5 7,500 10,400 4  250 mL J-TB-007


Swinging Bucket
(4 place)
rmax = 165 mm

JS-5.3 5.3 5,300 6,130 24 microplates J-TB-089


Swinging Bucket 8 deep-well
(4 place) plates
rmax = 194.8 mm 4 square-well
plates

JS-4.3 4.3 4,300 4,220 4  750 mL J-TB-050


Swinging Bucket
(4 place)
rmax = 204 mm

JS-4.0 4.0 4,000 4,050 4  1 liter J6-TB-006


Swinging Bucket 4 blood bags
(4 place) 12 microplates
rmax = 226 mm 148 RIA tubes

JCF-Z JCFZ 20,000 39,900 660 mL JCFZ-IM


Continuous Flow/ (standard core)
Zonal Rotorg 1250 mL
(large core)
240 mL
(small core)

JE-5.0 5.0 5,000 JE5-IM


Elutriation Rotor
rmax
large 4,700 40 mL
chamber = 168 mm
standard 3,500 4.0 mL
chamber = 125 mm
Sanderson 4,230 5.5 mL
chamber = 126 mm
a. Maximum speeds are based on a solution density of 1.2 g/mL with the following exceptions: the JA-18.1 rotor is rated for a
density of 1.4 g/mL; the JCF-Z rotor is rated for a density of 1.45 g/mL; and the JE-5.0 rotor is rated for a density of 3 g/mL.
b. Temperature performance for the JA-30.50 Ti rotor in the Avanti J-26S XP is as follows: 15°C minimum at 26,000 RPM (at 30°C
ambient); 4°C minimum at 21,500 RPM (30°C ambient).
c. Maximum speed for the JA-30.50 Ti rotor is 26,000 RPM in the Avanti J-26S XP centrifuge, and 30,000 RPM in the Avanti J-30 I
centrifuge.
d. Temperature performance for the JA-25.50 and JA-25.15 rotors in the Avanti J-26S XP is as follows: 10°C minimum at
25,000 RPM (at 30°C ambient); 4°C minimum at 21,500 RPM (30°C ambient).
e. Temperature performance for the JA-14.50 rotor is as follows: 4°C minimum at 14,000 RPM (at 38°C ambient).
f. Maximum speed for the JS-24.38 and JS-24.15 rotor is 10,000 RPM in the Avanti J-26S XP centrifuge, and 24,000 RPM in the
Avanti J-30 I centrifuge.
g. In the Avanti J-26S XP, the JCF-Z Continuous Flow/Zonal Rotor can be used in the continuous flow and reorienting gradient
zonal configurations only.

PN B10087AE 1-13
Description
Available Rotors

1-14 PN B10087AE
CHAPTER 2
Operation

Introduction

This section contains detailed centrifuge operating procedures. A summary is provided on page 2-1.
If you are an experienced user of this centrifuge, you can turn to the summary for a quick review of
operating steps.

WARNING
Normal operation may involve the use of solutions and test samples that are
pathogenic, toxic, or radioactive. Handle body fluids with care because they can
transmit disease. No known test offers complete assurance that they are free of
micro-organisms. Some of the most virulent — Hepatitis (B and C) and HIV (I–V)
viruses, atypical mycobacteria, and certain systemic fungi — further emphasize
the need for aerosol protection. Handle other infectious samples according to
good laboratory procedures and methods to prevent spread of disease. Because
spills may generate aerosols, observe proper precautions for aerosol
containment.
Do not run toxic, pathogenic, or radioactive materials in this centrifuge without
taking appropriate safety precautions. Biosafe containment should be used when
Risk Group II materials (as identified in the World Health Organization Laboratory
Biosafety Manual) are handled; materials of a higher group require more than one
level of protection.

WARNING
Do not use the centrifuge in the vicinity of flammable liquids or vapors, and do not
run such materials in the centrifuge. Do not lean on the centrifuge or place items
on it while it is operating.

Summary of Avanti J-26S XP Run Procedures

For runs at temperatures other than ambient, precool or prewarm the rotor to the required temperature
before the run.

1 Press the POWER switch to on (I).


a. Open the centrifuge door.

2 Install the rotor.

PN B10087AE 2-1
Operation
Section and Loading

a. Make sure that the rotor load is balanced and that the rotor is securely fastened to the
centrifuge drive hub.
b. Close the centrifuge door.

3 Turn the rotor knob until the installed rotor name appears in the ROTOR ID display.

4 Turn the SPEED knob until the required run speed (RPM) appears in the SPEED display.
a. Or press RPM/RCF to select RCF mode and turn the SPEED knob until the required RCF
appears in the SPEED display.

5 Turn the TIME knob until the required run length appears in the TIME display.
a. Or press TIME/HOLD for continuous operation.

6 Turn the TEMP knob until the required temperature appears in the TEMP°C display.

7 Press ACCEL repeatedly to select MAX (maximum) or SLOW acceleration.

8 Press DECEL repeatedly to select MAX (maximum) or SLOW deceleration, or select OFF
(no brake).

9 Check that all parameters are correct and that the door is closed.
a. Press START.

10 Wait for the set time to count down to zero, or end the run by pressing STOP.

11 When the run is complete, open the door and unload the rotor.

Section and Loading

For fast temperature equilibration, precool or prewarm the rotor to the required temperature before
the run.

1 Turn the power switch on (I).


Power is applied to the system, and the display illuminates.

2-2 PN B10087AE
Operation
Entering Run Parameters 2

2 Step on the foot pedal to open the door.


The door opens.

3 Install the rotor according to the instructions in the rotor manual.


a. Always run the rotor with a balanced load.
b. (If you are using a swinging bucket rotor, fill all positions on the yoke with buckets.)
c. (If you are using a J-Lite rotor with cannisters, place a cannister in each rotor body cavity.)

CAUTION
Do not drop the rotor onto the drive hub. The drive shaft can be bent if the rotor
is forced sideways or dropped onto the hub. Install the rotor by centering it over
the hub and carefully lowering it straight down.

4 Securely attach the rotor lid knob, or tie-down knob in rotors without lids, to the drive shaft by
turning it to the right (clockwise).

NOTE If the knob turns loosely and you do not feel threads engaging, the rotor drive hole pins may not
be properly seated on the centrifuge hub. Lift the rotor up above the hub, rotate it slightly, and lower
it onto the hub again. Tighten the knob.

5 Close the centrifuge door firmly.

CAUTION
If you leave the rotor in the centrifuge between runs, make sure the rotor is seated
on the drive hub and the tie-down knob is tight before each run. (Remove the rotor
from the centrifuge if you anticipate a long period between runs.)

Entering Run Parameters

Selecting a Rotor

1 Turn the ROTOR knob until the rotor in use appears in the display.

PN B10087AE 2-3
Operation
Entering Run Parameters

Small green LEDs by each rotor type abbreviation light one at a time, and the rotors of that type
appear consecutively in the display.

ROTOR
JA
JS JLA
JSP JV
JCF JE

Setting Run Speed


Enter a run speed up to 26,000 RPM, or up to the maximum speed of the rotor in use. Or, enter a
relative centrifugal field (RCF) value up to the maximum achievable RCF of the rotor in use.

Setting RPM

1 Check to see if the red RPM light is lit on the display. (RPM is the default speed mode.)
a. If it is lit, go to step 3.
b. If the centrifuge is already in the RPM mode, skip step 2 and go on to step 3.

2 Press the RPM/RCF key to select the RPM mode.


On the display, the red RPM LED lights.

RPM

3 Turn the SPEED knob until the required RPM appears in the SPEED display.
When the run is started, the centrifuge operates at the set speed.
The corresponding RCF is automatically calculated.
a. Press RPM/RCF to display the RCF value.

2-4 PN B10087AE
Operation
Entering Run Parameters 2

Setting RCF

1 Check to see if the red RCF light is lit on the display.


a. If it is lit, go to step 3.

2 Press the RPM/RCF key to select the RCF mode.


On the display, the red RCF ( g) LED lights.

3 Turn the SPEED knob until the required RCF value appears in the SPEED display.
The centrifuge operates at a speed calculated to produce the set RCF value.
a. Press RPM/RCF to display the RPM value.

Setting Run Time


Pressing the TIME/HOLD key toggles the system between HR:MIN mode and HOLD mode. HR:MIN
mode is the default.

Timed Run

1 Turn the TIME knob until the required run time appears in the TIME display.

PN B10087AE 2-5
Operation
Entering Run Parameters

If you enter more than 59 minutes in the minutes field, the system automatically converts the
entry to hours and minutes after ENTER is pressed.

2 When all run parameters are entered, press START.


• The rotor begins to accelerate and the display begins showing the remaining run time.
• The run will end when the time counts down to 0 or when STOP is pressed.

Continuous (HOLD) Run

1 Press the TIME/HOLD key to select the HOLD mode.


The red HOLD indicator light comes on in the TIME display.

2 When all run parameters are entered, press START.


• The rotor begins to accelerate and the display begins showing the elapsed time.

• The time will count up to 3 hours, 0 minutes, and then will reset to 0 and continue counting.

3 Press STOP to end the run.


The rotor decelerates to a stop.

2-6 PN B10087AE
Operation
Entering Run Parameters 2

Setting Run Temperature


Select a run temperature between –10 and +40°C, in 1°C increments.

NOTE For runs at temperatures other than ambient, always refrigerate or warm the rotor to the required run
temperature before the run. For low-temperature runs, precool the system by running a precooled rotor
at 2000 RPM at the required temperature for at least 30 minutes.

1 Turn the TEMP°C knob until the required temperature appears in the display.
• After you release the knob, the display shows the selected temperature.
• After 5 seconds, the display reverts to the actual current chamber temperature.
• When the run begins, the display shows the rotor temperature 2°C (after a period of
temperature equilibration).
NOTE The system controls the chamber temperature to within 2°C of the set temperature.* To control
temperature within 1°C of set, perform the temperature calibration procedure provided in
APPENDIX B, Temperature Calibration Procedure.

Setting Acceleration and Deceleration Rates


Table 2.1 and Table 2.2 describe the acceleration and deceleration settings. Figure 2.1 provides a
graphic representation of these settings.

Table 2.1 Acceleration Settings

Maximum Acceleration Slow Acceleration


(not available with the
JS-24.38 and JS-24.15 rotors)

Full torque is used from 0 RPM Reduced torque is used from 0 to 500 RPM, lasting approximately
until set speed is reached 2 minutes. (The time varies depending on the mass of the rotor in use.
Larger rotors take longer.). Above 500 RPM, full torque is used until set
speed is reached.

Application: pelleting Application: gradients

* During transient conditions, such as acceleration and deceleration, rotor temperature may be outside this range.

PN B10087AE 2-7
Operation
Entering Run Parameters

Table 2.2 Deceleration Settings

Maximum Deceleration Slow Deceleration Off

Full brake is used from set Reduced torque is used from set speed to No brake is used. Rotor
speed to near 0 RPM, to 500 RPM, causing deceleration to 500 RPM to coasts to 0 RPM. This
bring the rotor to a stop in take twice as long as maximum deceleration. process can take up to
the shortest possible time. (The exact rate depends on the mass of the 1 hour, depending on the
Reduced brake is used rotor in use. Larger rotors will take longer.) rotor in use and the
during the last few RPM Deceleration time from 500 to 0 Rpm is speed at the start of
until 0 is reached, to approximately 2 minutes (larger rotors take deceleration.
minimize disturbance of longer).
the sample.

Application: pelleting Application: gradients, pelleting Application: gradients

Figure 2.1 Acceleration and Deceleration Settings Graph

1 5
3

7
500
8
4

0
2

1. Rotor Speed (RPM) 5. Max Decel


2. Time 6. Run Speed
3. Max Accel 7. Off (no brake)
4. Slow Accel 8. Slow Decel

Setting the Acceleration Rate


NOTE The default setting for both acceleration and deceleration is maximum (MAX). If the MAX indicator
lights are already lit, no input is required. To choose a different setting, follow the steps below.

1 Press the ACCEL key to toggle between the MAX and SLOW acceleration settings.

2-8 PN B10087AE
Operation
Entering Run Parameters 2

The selected setting lights.

MAX
SLOW
ACCEL

Setting the Deceleration Rate

1 Press the DECEL key to toggle between the MAX, SLOW, and OFF deceleration settings.
The selected setting lights.

MAX
SLOW
OFF
DECEL

Setting Up an Elutriation Run


Centrifuges equipped for elutriation have elutriation controls as shown in Figure 2.2 and an
elutriation power box mounted on the back of the control panel. See the JE-5.0 elutriator rotor
manual (JE5-IM) for complete rotor setup and operating instructions.

Figure 2.2 Elutriation Controls

DELAY

ELUTRIATOR

DELAY

ELUTRIATO
R

1 Install the rotor.

2 Set up sample and buffer reservoirs and tubing lines as described in the rotor manual.
a. Route the tubing lines through the elutriator port holes at the left side of the door seal.
b. Close the centrifuge door.

PN B10087AE 2-9
Operation
Entering Run Parameters

3 Turn the ROTOR knob to the JE position.


5.0 appears in the ROTOR ID display.

ROTOR
JA
JS JLA
JSP JV
JCF JE

4 Enter the remaining run parameters.


• The run is ready to begin.
• Refer to the JE-5.0 rotor manual for rotor setup and operating instructions.

5 Press START.
The rotor accelerates to set speed.

6 When the rotor reaches set speed, turn on the strobe controls by pressing the strobe
power switch.
The power switch indicator lights.

1 1. Strobe Power Switch

DELAY

ELUTRIATOR

7 When the rotor speed has stabilized, look through the port in the centrifuge door and turn the
DELAY knob until the elutriation chamber in the rotor is synchronized with the strobe.
• (The chamber will appear to be motionless.)
• Sample can now be injected into the lines.
• See the rotor instruction manual for detailed instructions.

2-10 PN B10087AE
Operation
Changing Parameters During a Run 2

1. Delay Control Knob

DELAY
1
ELUTRIATOR

NOTE At the end of the run, after disassembling the liquid and power lines, be sure to replace the
black rubber stoppers in the ports at the left side of the centrifuge door. Failure to seal these
openings will prevent proper operation of the FRS system, which will cause a diagnostic shut
down during the next run.

Starting a Run

1 Check that all parameters are correct and the door is firmly closed.

2 Press START.
• The display will start showing actual centrifuge values within 5 seconds.
• The LED on the START key flashes during acceleration.
• When set speed is reached, the light stays on continuously until the centrifuge begins
decelerating.
• As the run begins, the system identifies the installed rotor.
• The system verifies that the user-selected rotor and the installed rotor are the same.
• If the system detects that the installed rotor is not the rotor that the user selected with
the ROTOR knob, and the speed selected by the user is greater than the installed rotor’s
maximum permitted speed, the system will reduce the run speed to the maximum allowed
for the installed rotor.

Changing Parameters During a Run

While a run is in progress, the speed, time, temperature, and acceleration or deceleration
setting can be changed without stopping the run. The rotor name parameter cannot be changed
during a run.

PN B10087AE 2-11
Operation
Stopping a Run

Changing Time Settings

1 To increase or decrease the remaining run time, turn the knob to the new run time.
• The system will add the new remaining time to (or subtract from) the time already elapsed,
and the display will show the new remaining run time.
• If you turn the TIME knob to a value less than the length of time the centrifuge has already
been running, the run will end and the rotor will begin to decelerate immediately.
— For example, if the centrifuge has been running for 20 minutes and you turn the knob
until 10 shows in the display, the run will end.

2 To set a completely new run time while a timed run is in progress, turn the TIME knob to the
new time setting and press START.
The system will ignore any elapsed time.

3 To change from HR:MIN mode to HOLD mode, press the TIME/HOLD key.
The display will show the total accumulated run time and will begin counting the time elapsed
from that point.

4 To change from HOLD mode to HR:MIN mode, press the TIME/HOLD key and turn the TIME knob
until the run length appears in the display.
a. Press START.
The system will ignore any elapsed time and will begin displaying the time remaining in
the run.

Stopping a Run

A timed run ends automatically when the TIME display counts down to zero. “End” will flash in the
TIME display. To end a HOLD run or a run in progress for any reason, press STOP.

NOTE If a diagnostic condition causes the run to stop, with or without use of the brake, the appropriate
diagnostic message will light and the light on the STOP key will flash until the rotor comes to a
complete stop.

1 Press STOP to initiate deceleration.


The light on the STOP key flashes until the rotor comes to a stop.

2-12 PN B10087AE
Operation
Unloading the Centrifuge 2

2 After the rotor stops spinning, step on the foot pedal located on the front of the centrifuge, to
open the door.

NOTE During runs at very cold temperatures, nominally –10°C or below, ice may form around the door
opening, causing the door to stick shut. If this happens, press firmly on the door in several locations
around the door opening and depress the foot pedal again.
To minimize icing, remove moisture from inside the chamber, from the chamber gasket, and from
the inner door surface before each run by wiping these surfaces with a clean cloth. Also, keep the
door closed as much as possible.

Unloading the Centrifuge

At the end of a run, unload the rotor following the instructions in the applicable rotor manual.

CAUTION
If disassembly reveals evidence of leakage, you should assume that some fluid
escaped the rotor. Apply appropriate decontamination procedures to the
centrifuge and accessories as required.

PN B10087AE 2-13
Operation
Unloading the Centrifuge

2-14 PN B10087AE
CHAPTER 3
Troubleshooting

Introduction

This chapter lists possible malfunctions, along with probable causes and corrective actions. Maintenance
procedures are given in CHAPTER 4, Care and Maintenance. For any problems not covered here, contact
Beckman Coulter Field Service (1-800-742-2345 in the United States; outside the U.S. contact your local
Beckman Coulter office or visit us at www.beckman.com) for assistance.
NOTE It is your responsibility to decontaminate the instrument, as well as any rotors and/or accessories,
before requesting service by Beckman Coulter Field Service.

User Messages

FRS SYS
ROTOR POWER
IMBAL DRIVE
SPEED REFR
DOOR TEMP
CLEAR

If a problem occurs during operation, the appropriate diagnostic message will light and a beep will
sound to alert you to the condition. See Table 3.1 for a list of the diagnostic messages, the probable
causes, and the recommended actions.

Two kinds of diagnostic conditions can occur.

• Cautionary diagnostic messages alert you to a condition that may need attention when the run is
complete. Cautionary diagnostic messages provide information only; they do not shut down the
run in progress. Press CLEAR to stop the light from flashing. If the condition still exists, the
diagnostic message will begin flashing again, even though the run continues.
• Shutdown diagnostic messages occur when a problem exists that prevents the run from
continuing. The diagnostic message will flash and the red STOP key light will also flash,
indicating that the rotor is decelerating to a stop. You can press CLEAR to stop the diagnostic
message from flashing. However, if the diagnostic condition still exists (for example, if the rotor
is out of balance), another beep will sound and the message will start flashing again.

Table 3.1 contains information on causes and recommended actions for each diagnostic condition.
If you are unable to correct the problem, call Beckman Coulter Field Service (in the United States,
call 1-800-742-2345; outside of the U.S., call your local Beckman Coulter office or visit us at

PN B10087AE 3-1
Troubleshooting
User Messages

www.beckman.com). To help the Field Service representative diagnose and correct the problem,
provide as much information about the situation as you can, including:
• the diagnostic message,
• the operating situation when the diagnostic condition occurred (such as, rotor in use, speed, or
load type), and
• any unusual environmental and/or operating conditions (such as ambient temperature, or
voltage fluctuations).

NOTE The information in Table 3.1 is provided as a user guide, and is not a comprehensive checklist.

Table 3.1 Diagnostic Message Chart

Diagnostic Message Problem Result Recommendation

FRS FRS integrity is not Run shuts down with • Check and clean door sealing
(friction reduction established within maximum brake area and door gasket.
system) 10 seconds after the pump • Wipe excess moisture from
is turned on chamber.
• Make sure that elutriator port
stoppers are in place at left side
of door seal.

Required FRS level Run shuts down with Call Beckman Coulter Field
not reached within maximum brake Service.a
90 seconds

FRS level stays too high for Run continues Press CLEAR to stop LED from
45 seconds flashing.

FRS level stays too high for Run shuts down with Call Beckman Coulter Field
60 seconds maximum brake Service.a

IMBAL Rotor load is imbalanced Run shuts down with • Make sure that the tubes are
(imbalance) maximum brake loaded symmetrically in the
rotor.
• With swinging bucket rotors,
lubricate the buckets according
to the rotor manual
instructions. Unlubricated pivot
pins can prevent the buckets
from reaching horizontal
position, which can cause
imbalance.

3-2 PN B10087AE
Troubleshooting
User Messages 3

Table 3.1 Diagnostic Message Chart (Continued)

Diagnostic Message Problem Result Recommendation

ROTOR System-identified rotor System lowers run speed Press CLEAR to stop LED from
does not match user-set to the maximum of the flashing.
rotor; set speed above rotor rotor identified by system
maximum

System cannot identify Run shuts down with Check rotor and restart run.
rotor (rotor ID system maximum brake
malfunction)

No rotor in chamber Run starts; within seconds Install rotor and restart run.
the system determines
that no rotor is installed
and the run shuts down
with maximum brake

SPEED Speed control system is Rotor spinning above set Call Beckman Coulter Field
malfunctioning speed; run shuts down Service.a
with maximum brake

DOOR Door is not latched properly DOOR LED lights after Open the door and close it firmly;
or interlock is out of position START is pressed and press CLEAR to stop LED from
system shuts down flashing. Restart the run.

Door stays locked after Sample cannot be Restart the centrifuge and perform
rotor stops spinning retrieved a brief run. If door will still not open
after deceleration, call
Beckman Coulter Field Service.a

Door latches do not close Run cannot start Check the latch hole (at left of door)
and lock for debris. Gently clean the area
with a lintless cloth or tissue. Be
careful not to damage sensitive
electronics in the area.

SYS There is a problem with the System shuts down Call Beckman Coulter Field
(system) system control software, Service.a
EPROM, or RAM

POWER Momentary power failure; Run continues when Press CLEAR to stop LED from
rotor does not come to a power resumes; elapsed flashing.
complete stop time continues counting
from current value

Power failure; rotor comes Run restarts when power Press CLEAR to stop LED from
to a complete stop resumes; elapsed time flashing.
starts counting from 0

PN B10087AE 3-3
Troubleshooting
User Messages

Table 3.1 Diagnostic Message Chart (Continued)

Diagnostic Message Problem Result Recommendation

DRIVE Erroneous signal from Erroneous display of Call Beckman Coulter Field
tachometer board to 0 RPM immediately after Service.a Wait for 4 hours before
display board or drive failure, rotor still spinning. trying to open the door.
system problem. Door
lockout time may not be
sufficient to prevent access
to spinning rotor.

Drive system Run stops, usually with Call Beckman Coulter Field
malfunctioning (for no brake. Door may not Service.a Before trying to open the
example, over current, unlock for up to an hour. door, listen carefully and make
over/under voltage, sure that no sound is coming from
components over the chamber (indicating a spinning
temperature) rotor). Follow the directions under
Accessing the Rotor in Case of
Power Failure, below.

REFR Refrigeration system Run shuts down with Call Beckman Coulter Field
(refrigeration malfunctioning maximum brake Service.a
system)

TEMP The refrigeration system Run shuts down with Call Beckman Coulter Field
is not cooling and the maximum brake Service.a
chamber temperature has
exceeded 50°C.

— During low-temperature Door will not open at the Press firmly in several locations
runs (near –10°C), ice forms end of a run around the door opening, then
around the door opening depress the foot pedal again.
To minimize icing, wipe moisture
from the chamber, the chamber
gasket and the inner door surface
before each run. Keep the door
closed as much as possible.
a. In the United States, call 1-800-742-2345. Outside the U.S., contact your local Beckman Coulter office or visit us at www.beckman.com.

3-4 PN B10087AE
Troubleshooting
Accessing the Rotor in Case of Power Failure 3

Accessing the Rotor in Case of Power Failure

WARNING
Any maintenance procedure requiring removal of a panel exposes the operator to
the possibility of electrical shock and/or mechanical injury. Therefore, turn the
power off (O) and disconnect the instrument from the main power source by
removing the Mains (power) plug from the outlet receptacle, and refer such
maintenance to qualified service personnel.

WARNING
Before performing this procedure, verify that the rotor is not spinning by listening
carefully for any noise coming from the chamber. Proceed only if the instrument
is quiet. Never attempt to override the door interlock system while the rotor is
spinning.

If the facility power fails only momentarily, the centrifuge will resume operation when the power
is restored and the rotor will return to the set speed. In the event of an extended power failure, it
may be necessary to override the door-locking mechanism manually to remove the rotor and
retrieve your sample.

To access the door-locking mechanism, you must remove the centrifuge front panel. Two latches
secure the front panel in place; these latches are accessible through two holes at the upper right
and left of the panel (see Figure 3.1).

Figure 3.1 Emergency Door Release Latch Access

Avanti
Cent
J-26S XP
rifug
e

PN B10087AE 3-5
Troubleshooting
Accessing the Rotor in Case of Power Failure

1 Turn the power switch to off (O) and unplug the power cord from the main source by removing
the Mains (power) plug from the outlet receptacle.
• The Mains (power) plug is the disconnect device and must remain easily accessible.
— Position the centrifuge so that it is easy to remove the Mains (power) plug from the
outlet receptacle.

2 Insert a 4-mm or 5/32-in. Allen wrench straight through one of the holes (either one) and turn
the wrench to the left (counterclockwise) until the latch disengages.

3 Repeat step 2, inserting the Allen wrench through the other hole.
• After the second latch disengages, the front panel will fall forward about an inch from
the top.
• The bottom of the front panel is held in place by three grooved brackets, attached to the
front panel, that seat over a metal retaining strip on the centrifuge bottom panel.

4 Lift the front panel up and off the centrifuge; set it aside.
• You will see an inner front panel that extends about halfway down from the centrifuge top
(see Figure 3.2).
• Below this inner panel is a black manual door release interlock lever (see Figure 3.2).

Figure 3.2 Manual Door Release

1. Inner Front Panel 2. Retaining Strip 3. Manual Door Release Interlock Lever

3-6 PN B10087AE
Troubleshooting
Accessing the Rotor in Case of Power Failure 3

5 Pull the interlock lever out and to the left (at about a 45-degree angle), and while holding it out,
step on the foot pedal.
Depending on the level of vacuum in the chamber at the time of the power failure, the door may
or may not open.
a. If it opens, go to step 6.
b. If it does not open, you will need to manually vent the vacuum. Go to step 7.

6 After the door opens, first release the foot pedal and then release the interlock lever.
Sample can be removed.
a. Go to step 10.

7 To vent the chamber vacuum, grasp the red rubber hose and pull it up until it comes off of the
pump fitting (see Figure 3.3).
a. Use a back-and-forth motion as you pull; this takes quite a bit of force.
The vacuum will be completely vented several seconds after the hose is detached.

NOTE Several tubing lines are tied to the red hose. As you grasp the hose, place your hand over
these lines. Make sure that they stay connected as you pull.

CAUTION
Do not twist the hose to the left (counterclockwise) as you pull the hose, or you
will unscrew the hose fitting.

8 Replace the hose by pushing it down over the fitting as far as it can go.

9 Pull the interlock level out and left, and while holding it out, step on the foot pedal.

10 Remove your sample.

PN B10087AE 3-7
Troubleshooting
Accessing the Rotor in Case of Power Failure

DANGER
Never try to slow or stop the rotor by hand.

Figure 3.3 Venting the Chamber Vacuum

1. Step 1 5. Grasp red hose and pull up Do not disconnect the tubing lines tied to the
2. Step 2 red hose.
3. Step 3 6. Red hose removed from fitting
4. Vacuum Pump 7. Replace red hose

To replace the panel on the centrifuge,

3-8 PN B10087AE
Troubleshooting
JCF-Z Rotor Identification 3

1 Holding the panel at the top (with the instrument label facing out), insert the three grooved
brackets on the bottom inside of the door over the front panel retaining strip (see Figure 3.4).
a. Push the top of the front panel into place.

Figure 3.4 Front Panel Retaining Strip

1. Retaining Strip

2 Hold the left top edge in place and insert the Allen wrench through the hole.
a. Turn to the right (clockwise) until the latch engages.

3 Repeat step 2 on the right side.


The second latch engages.

JCF-Z Rotor Identification

The rotor identification system can, under two specific conditions, misidentify the JCF-Z
continuous flow/zonal rotor. Misidentification has occurred if a rotor name other than JCF-Z
appears in the ROTOR display during the run, when rotor identification is complete. If the rotor is
misidentified, stop the run and follow one of the steps below.

NOTE Misidentification can be prevented by returning your JCF-Z rotor to the Beckman Coulter factory to
have magnets added to the rotor body. The magnets ensure correct identification. Call your local
Beckman Coulter Field Service representative* for more information.

JCF-Z rotors manufactured after March, 1997, have factory-installed magnets.

* In the United States, call 1-800-742-2345. Outside the U.S., contact your local Beckman Coulter office or visit us at
www.beckman.com.

PN B10087AE 3-9
Troubleshooting
JCF-Z Rotor Identification

1 On the first few uses of a new JCF-Z rotor, or in an older JCF-Z rotor with newly replaced bearings, run
the rotor from 0 to 5000 RPM and back to 0 RPM three times.
This procedure will distribute lubricant around the bearings in the rotating seal assembly,
reducing drag.

2 In older JCF-Z rotors with bearings that have not been replaced recently, follow the same procedure
as for a new JCF-Z rotor (perform three runs from 0 to 5000 RPM and back to 0 RPM) to ensure
that the bearings are properly lubricated.
If misidentification recurs after the third run, when the bearings are lubricated, this indicates
that the bearings are worn and need to be replaced.
a. Replace the bearings following the instructions in the JCF-Z manual (JCFZ-IM).

NOTE If the JCF-Z rotor is misidentified, the run speed will be limited to the maximum speed for the
identified rotor. (The maximum speed for the JCF-Z rotor is 20,000 RPM.)

3-10 PN B10087AE
CHAPTER 4
Care and Maintenance

Introduction

This section describes routine care and maintenance procedures that you should perform regularly or
as required. For maintenance not covered in this manual, contact Beckman Field Service.* Refer to the
applicable rotor manual and Rotors and Tubes for J Series Centrifuges (publication JR-IM) for
instructions on the care of rotors and accessories.

NOTE It is your responsibility to decontaminate the instrument, as well as any rotors and/or accessories,
before requesting service by Beckman Coulter Field Service.

WARNING
Any maintenance procedure or servicing of this equipment that requires removal
of any covers can expose parts which involve the risk of electric shock or personal
injury. Make sure that the power switch is off (O) and the centrifuge is
disconnected from the main power source by removing the Mains (power) plug
from the outlet receptacle, and refer such servicing to qualified service personnel.

WARNING
Do not use alcohol or other flammable substances in or near operating
centrifuges.

Maintenance

Perform the following procedures regularly to ensure continued performance and long service life
of the centrifuge.

1 Inspect the centrifuge chamber for accumulations of sample, dust, or glass particles from
broken sample tubes.
a. Clean as required (see Cleaning below).

2 Check the air filter on the back panel for obstructions.


a. Keep vents clear and clean.

* In the United States, call 1-800-742-2345. Outside the U.S., contact your local Beckman Coulter office or visit us at
www.beckman.com.

PN B10087AE 4-1
Care and Maintenance
Maintenance

3 Wipe condensation out of the chamber between runs with a sponge or clean cloth to prevent
chamber icing.

4 If chamber icing occurs, defrost the system and wipe moisture out of the chamber before use.
a. To defrost the system, set the temperature to 30°C for 20 minutes.
(These are suggested settings that may be adjusted as appropriate for your laboratory
conditions.)

NOTE Before using any cleaning or decontamination methods except those recommended by the
manufacturer, users should check with the manufacturer that the proposed method will not damage
the equipment.

Cleaning
Clean the centrifuge frequently. Always clean up spills when they occur to prevent corrosives or
contaminants from drying on component surfaces.

1 To prevent accumulations of sample, dust, and/or glass particles from broken sample tubes,
keep the chamber clean and dry by frequent wiping with a cloth or paper towel.
a. For thorough cleaning, wash the chamber using a mild detergent such as Solution 555.
b. Dilute the detergent with water (10 parts water to 1 part detergent).
c. Rinse thoroughly and dry completely.

2 Clean the centrifuge exterior surfaces by wiping with a cloth dampened with Solution 555.
Dilute the detergent with water (10 parts water to 1 part detergent).
Do not use acetone.

3 Clean the drive hub regularly using Solution 555 and a soft brush.
a. Dilute the detergent with water (10 parts water to 1 part detergent).
b. Rinse thoroughly and dry completely.

Tube Breakage
If a glass tube breaks, and all the glass is not contained in the bucket or rotor, be sure to thoroughly
clean the chamber.

4-2 PN B10087AE
Care and Maintenance
Maintenance 4

WARNING
Be careful when examining or cleaning the chamber and chamber gasket, as sharp
glass fragments may be embedded in their surfaces.

1 Examine the chamber gasket to make sure that no glass particles are retained in it.
a. Carefully remove any glass particles that may remain.

2 Carefully wipe away any glass particles that remain in the chamber.

Decontamination
If the instrument and/or accessories are contaminated with radioactive or pathogenic solutions,
perform all appropriate safety and decontamination procedures. Refer to Chemical Resistances
(publication IN-175) to be sure the decontamination method will not damage any part of the
instrument

Sterilization and Disinfection


Ethanol (70%) may be used to clean any exterior surface of the centrifuge. See Chemical Resistances
(publication IN-175) for chemical resistances of centrifuge and accessory materials. For this
centrifuge refer to the PET column in the chemical resistances chart for the plastic surfaces and the
polyurethane paint column for the painted metal surfaces.

CAUTION
Ethanol is a flammability hazard. Do not use it in or near operating centrifuges.

While Beckman Coulter has tested ethanol (70%) and found that it does not damage the centrifuge,
no guarantee of sterility or disinfection is expressed or implied. When sterilization or disinfection
is a concern, consult your laboratory safety officer regarding proper methods to use.

Replacing the Air Filter

1 Check the air filter regularly and replace it about once a year, or more often if it looks dirty.
The air filter is not fastened to the centrifuge, so no tools are required for removal or
installation.

PN B10087AE 4-3
Care and Maintenance
Circuit Breaker and Fuses

1. Frame Edge
2. Air Filter
3. Retaining Strip
4. Frame Edge

2
4

2 To remove the air filter, hold the side edges and lift the filter straight up until the bottom edge
is above the centrifuge bottom retaining strip.
a. Pull the filter out, bottom edge first, and discard it.

3 Install a new filter (885218).


a. The filter has a directional arrow on one of its edges; install the filter with this arrow
pointing toward the centrifuge.
b. Holding the filter by the side edges, insert the top half behind the frame edge and lift up
until the bottom half clears the retaining strip.
c. Then set the bottom edge down.

Circuit Breaker and Fuses

There are no user-replaceable fuses in the Avanti J-26S XP centrifuge.

If the centrifuge circuit breaker trips for any reason, the power switch will move to the OFF (O)
position. Reset the circuit breaker by turning the power switch back to the ON (I) position. If it trips
again immediately, do not reset it. Call Beckman Coulter Field Service.

4-4 PN B10087AE
Care and Maintenance
Storage and Transport 4

CAUTION
Repeated attempts to reset the centrifuge circuit breaker can cause substantial
damage to electrical and electronic components.

Storage and Transport

Storage
To ensure that the centrifuge does not get damaged, contact Beckman Coulter Field Service* for
specific instructions and/or assistance in preparing the equipment for transport or long-term
storage. Temperature and humidity conditions for storage should meet the environmental
requirements described under Specifications in CHAPTER 1.

Returning a Centrifuge
Before returning a centrifuge or accessory for any reason, prior permission must be obtained from
Beckman Coulter. Contact your local Beckman Coulter office* to obtain the required form and
instructions for packaging and shipping.

To protect our personnel, it is the customer’s responsibility to ensure that all parts are free from
pathogens and/or radioactivity. Sterilization and decontamination must be done before returning
the parts.

All parts must be accompanied by a note, plainly visible on the outside of the box or bag, stating that they
are safe to handle and that they are not contaminated with pathogens or radioactivity. Failure to attach
this notification will result in return or disposal of the items without review of the reported problem.

* In the United States, call 1-800-742-2345. Outside the U.S., contact your local Beckman Coulter office or visit us at
www.beckman.com.

PN B10087AE 4-5
Care and Maintenance
Supply List

Supply List

Contact Beckman Coulter Sales* or visit www.beckman.com for information about ordering parts,
supplies, or publications. Refer to the High Performance, High Speed, High Capacity Rotors, Tubes &
Accessories catalog (BR-8102, available at www.beckman.com). For your convenience, a partial list of
centrifuge supplies is given below.

Refer to the applicable rotor manual for materials and supplies needed for rotors.

Replacement Parts and Supplies


NOTE For MSDS information, go to the Beckman Coulter website at www.beckman.com.

Description Part
Number

Air filter 885218

Solution 555 (1 qt) 339555

Zonal Bracket Kit for Cole-Parmer size 16 tubing


(6.4-mm [1/4-in.] O.D.) 366431
Zonal Bracket (qty 1) 366430
Thumbscrew, stainless steel, M4 (threads) x 19 mm (qty 2) 893412
Thumbscrew, stainless steel, M4 (threads) x 12 mm (qty 2) 893411
Cable Clamp, nylon, 6.4-mm (1/4-in.) I.D. (qty 10) 000499

Zonal Tubing Adapter for Cole-Parmer size 14 tubing


(4.8-mm [3/16-in.] (O.D.) 363844
Note: kit 366431 is also required.
Tubing Adapter, stainless steel, for size 14 tubing (qty 2) 363830
Cable Clamp, nylon, 4.8-mm (3/16-in.) I.D. (qty 10) 003343

Zonal Tubing Adapter for Cole-Parmer size 15 tubing


(9.6-mm [3/8-in.] O.D.) 363845
Note: kit 366431 is also required.
Tubing Adapter, stainless steel, for size 15 tubing (qty 2) 363831
Cable Clamp, nylon, 9.6-mm (3/8-in.) I.D. (qty 10) 000596

Elutriation upgrade kit 366562

* In the United States, call 1-800-742-2345. Outside the U.S., contact your local Beckman Coulter office or visit us at
www.beckman.com.

4-6 PN B10087AE
APPENDIX A
Preinstallation Requirements

Preinstallation Requirements

Do not attempt to install this instrument. Its purchase price includes installation by Beckman Coulter
personnel. Installation by anyone other than an authorized Beckman Coulter representative invalidates
any warranty covering the instrument.

Preinstallation requirements have been sent prior to shipment of the instrument. The following
information is provided in case the centrifuge must be relocated.

The centrifuge will be installed upon initial purchase by a Beckman Coulter Field Service
representative after preinstallation requirements for power and site preparation have been met.
The following equipment is required for preinstallation:
• Voltmeter
• For single phase centrifuges: two 30-ampere circuit breakers
• For three phase centrifuges: three 16-ampere circuit breakers
• Power receptacle (see Figure A.1 or Figure A.2)
• Drill for drilling holes in the floor for installation of anchoring kit bolts (see Securing the
Centrifuge to the Floor, later in this section). A 9.5 mm (3/8-inch) drill is required for concrete
floors. A 6.4 mm (1/4-inch) drill is required for wood floors.

Electrical Requirements
Power to the centrifuge should originate directly from a main power line transformer at a power
source known to be clear of erratic loads, spikes, and electromagnetic interference. Make sure that
there are properly rated thermal circuit breakers at the service panel to protect the centrifuge
circuit. If fuses must be used instead of the specified circuit breakers, the fuses may require a rating
of greater than 30 amperes (for single-phase centrifuges) or greater than 16 amperes (for three-
phase centrifuges).

Terminate the open end of the harmonized cord with a certified single- or three-phase connector
suitable for the power supplied in the country of intended use (see Table A.1). Install only one
centrifuge per circuit.

PN B10087AE A-1
Preinstallation Requirements
Preinstallation Requirements

Table A.1 Nominal Supply Voltage Ratings for the Avanti J-26S XP

Nominal Instrument Instrument Nominal Power Cord and Plug


Instrument Part Part Supply Description
Voltage Rating Number Number Frequency
(Elutriation
System)

Single-phase, B14535 B14541 180–264 V, permanently attached three-wire


200/208/240 V 50/60 Hz, 30 A UL/CSA-approved cord with NEMA
type 6-30P plug

Single-phase, 230 V B14536 B14542 180–264 V, permanently attached three-wire


50 Hz, 30 A CENELEC harmonized cord
without plug

Three-phase, B14537 B14543 313–457 V, permanently attached five-wire


220/380 V + plus neutral, CENELEC harmonized cord
Neutrala 50 Hz, 16 A without plug

a. Unbalanced three-phase. Split for single-phase operation internally.

To ensure optimal safety, the centrifuge should be wired to a remote emergency switch (preferably
outside the room where the centrifuge is housed, or adjacent to the exit from that room). In case of
a malfunction, the centrifuge can be disconnected from the main power source by removing the
Mains (power) plug from the outlet receptacle.

WARNING
To reduce the risk of electrical shock, this equipment uses a three-wire or five-
wire electrical cord and plug to connect the centrifuge to earth-ground. To
preserve this safety feature, make sure that the matching wall outlet receptacle
is properly wired and earth-grounded.

Prior to instrument purchase, the power configuration should be determined and the appropriate
instrument ordered.

Single and Three-Phase Power Connections


Figure A.1 shows the power connection for single-phase centrifuges, including earth-ground and
two power leads with 30-ampere circuit breakers. Figure A.2 shows the power connection for three-
phase centrifuges, including earth-ground and three power leads with minimum 16-ampere circuit
breakers and common neutral connection. Table A.2 contains wiring information.

A-2 PN B10087AE
Preinstallation Requirements
Preinstallation Requirements A

Figure A.1 Single-Phase Electrical Requirements


1

4
3

1. 30-ampere Circuit Breaker 3. Earth-Ground


2. Wall Outlet: Hubell 9930, 4. Measured Line Voltage
Bryant 96-30-FR, or Equivalent
(NEMA 6-30 R)

Figure A.2 Three-Phase “Y” Electrical Requirements


1

2
3

1. 16-ampere Circuit Breaker 3. Measured Line Voltage Phase to Phase


2. Neutral 4. Earth-Ground

Table A.2 Required Wire Connections

Wire Insulation Color Terminal Symbol

Harmonized North American

Green/Yellow Earth ground

Light Blue Neutral N L


Brown Live or Line L L
Black (three-phase Live or Line (qty 2) L —
connections only)

PN B10087AE A-3
Preinstallation Requirements
Preinstallation Requirements

Additional Requirements for Three-Phase Power Connections


For three-phase power service to the centrifuge, use the “Y” connected configuration shown in
Figure A.3. Also note the following information:

• The steady state current draw of the centrifuge can be as high as 12 amperes per phase,
depending on the voltage.
• Inrush current to the centrifuge is 100 amperes for up to 2 seconds during the refrigeration
start period. Circuit breakers, whether thermal or magnetic actuating, must be a “motor start”
delay type.
• The centrifuge can tolerate a drop of 15 volts during the start period at low line voltage (see
Figure A.3 for the acceptable voltage ranges). The mains power for the centrifuge must be of
sufficient wire gauge to provide this condition. The wire gauge required at a particular site
must be determined by a facilities engineer at that site.

Figure A.3 Correct and Incorrect Three-Phase Power Configurations


1 2

3
312–456

180–264
4

1. Correct Configuration “Y” 3. Safety Ground


2. Incorrect Configuration “Delta” 4. Circuit Common (Neutral)

Space and Location Requirements

1 Locate the instrument on a clean, level floor.


• The Mains (power) plug is the disconnect device and must remain easily accessible.
— Position the centrifuge so that it is easy to remove the Mains (power) plug from the
outlet receptacle.

2 Select a location away from heat-producing laboratory equipment.


If ambient temperature exceeds 38°C (100°F), premature component failure may result.

3 In addition to space for the centrifuge, allow a 7.7-cm (3-in.) clearance on each side of the
centrifuge and a 16-cm (6.25-in.) clearance behind the centrifuge for air circulation.
The centrifuge must have adequate air ventilation to ensure compliance to local requirements
for vapors produced during centrifuge operation.

A-4 PN B10087AE
Preinstallation Requirements
Preinstallation Requirements A

4 Position the centrifuge so that the air diverter, shown in Figure A.4, touches the wall behind the
centrifuge.
a. Place the power cord to one side of the air diverter.
b. To avoid damaging the power cord when installing or moving the centrifuge, be sure to
move the cord out of the way before pushing the centrifuge towards the wall.

WARNING
Do not place the centrifuge near areas containing flammable reagents or
combustible fluids. Vapors from these materials could enter the instrument’s
air system and be ignited.

Figure A.4 Rear View and Dimensions

cm
in.
86
34

1
149
2 58.5
1.8 m 116
6.0 ft. 45.5
3 86
34
4 16.0
6.25
5

71
28

1. Air Exit 4. Depth


2. Power Cable 5. Air Intake
3. Air Diverter

PN B10087AE A-5
Preinstallation Requirements
Using J2 Series Rotors in the Avanti J-26S XP

Securing the Centrifuge to the Floor


Avanti J series centrifuges are certified to meet the requirements of the European CE mark. To meet
these requirements, the centrifuge must be secured to the floor using the anchoring hardware
shipped with the instrument. This will prevent the centrifuge from moving in the unlikely event of
a rotor mishap.

Complete instructions for installing the anchoring kit are packaged with the hardware, which is
shipped with the centrifuge. The instructions (publication B10122) include a full-size template to be
used as a guide for drilling holes in the floor. Refer to this document for additional installation
instructions.

NOTE Beckman Coulter representatives are not equipped to drill holes in your floor. The holes must be drilled
before your scheduled installation.

Bio-Safety Level 3 Installation


For laboratories with epoxy aggregate (resinous) floors, such as BSL-3 labs, a non-invasive
installation kit (PN 393316) is available. The kit which consists of an adhesive-backed mounting
plate, is CSA certified for use on epoxy aggregate floors only.

Non-Invasive Centrifuge Restraint System for Vinyl Floors


For laboratories with vinyl tile floors or vinyl seamless floors, a non-invasive installation kit
(PN A84005) is available. The kit which consists of an adhesive-backed mounting plate, is CSA
certified for use on vinyl floors only.

Using J2 Series Rotors in the Avanti J-26S XP

CAUTION
Do not use the Beckman Coulter JA-10, JS-7.5, JA-14, or JS-13 rotors in the
Avanti J-26S XP centrifuge before reading this information.

Checking for Rotor Drive Pins


Rotors used in Avanti J series centrifuges must have drive pins in the rotor drive hole (see
Figure A.5). These drive pins engage with the centrifuge spindle hub to ensure that the rotor does
not slip during acceleration. Some Beckman Coulter rotors, including the JA-10, the JS-7.5, the
JA-14, and the JS-13, have been manufactured without drive pins because pins were not needed
when these rotors were used in J2 series centrifuges.

A-6 PN B10087AE
Preinstallation Requirements
Using J2 Series Rotors in the Avanti J-26S XP A

Figure A.5 Checking the Rotor for Drive Pins


KMA N
EC

B
3
4

KM
EC A
N
B

1 2

1. Top View 4. Drive Pins (Angled pins shown; pins can also be vertical or horizontal.)
2. Side View 5. Drive Spindle Assembly
3. Rotor Drive Hole

The rotor pins are positioned parallel to the Beckman name engraved at the center of the rotor body
(see Figure A.5). Knowing the pin orientation before you install the rotor will help to ensure that
you position the rotor properly on the hub, minimizing the chance of hub damage.

Check all J2 series rotors for drive pins before using them in an Avanti J series centrifuge and do not use
rotors without drive pins in the Avanti J. To check for drive pins, hold the rotor up or turn it on its side
and look into the drive hole. If you do not see two metal pins in the hole, do not use the rotor in the
Avanti J. Call your local Beckman Coulter office* for information on returning the rotor to the
factory for upgrading.

Using the JA-18 Rotor


In Avanti J series centrifuges, the JA-18 rotor must be run with the lid attached. See the JA-18 rotor
manual (publication J-TB-035) for complete rotor usage instructions.

* In the United States, call 1-800-742-2345. Outside the US, contact your local Beckman Coulter office or visit us at
www.beckman.com.

PN B10087AE A-7
Preinstallation Requirements
Using J2 Series Rotors in the Avanti J-26S XP

Using the JCF-Z Continuous Flow/Zonal Rotor

Zonal Bracket Kits


A special bracket and mounting hardware are required when the JCF-Z Continuous Flow/Zonal
Rotor is used. These parts are contained in the Zonal Bracket Kit (part number 366431). The Zonal
Bracket Kit includes a bracket, mounting hardware, and assembly instructions. This kit is for use
with Cole-Parmer tubing size 16 (6.4 mm [1/4-in.] O.D.), the tubing size most commonly used with
the JCF-Z rotor.

Two additional kits are available, one which contains adapters for size 14 tubing (4.8 mm
[3/16-in.] O.D.) and one which contains adapters for size 15 tubing (9.6 mm [3/8-in.] O.D.). If you
wish to use either of these sizes of tubing, you will need to purchase one of these kits in addition to
the basic kit (366431).

See Replacement Parts and Supplies in CHAPTER 4 for a complete list of Zonal Bracket Kit part
numbers and kit contents.

Ensuring Correct JCF-Z Continuous Flow/Zonal Rotor Identification


The centrifuge’s rotor identification system can, under two specific conditions, misidentify the
JCF-Z continuous flow/zonal rotor. These conditions and corrective actions are listed below.

NOTE JCF-Z rotors manufactured after March, 1997, contain special magnets that ensure correct rotor
identification in Avanti J series centrifuges. We recommend that older JCF-Z rotors be returned to the
Beckman Coulter factory for addition of the magnets before use in the Avanti J-26S XP. Call
Beckman Coulter Field Service* for more information.

The following information pertains to JCF-Z rotors manufactured before March, 1997, and to JCF-Z
rotors that have not been modified at the factory.

• When the bearings in the JCF-Z rotating seal assembly get old or worn, the resulting “drag” on
the bearings changes the rotor’s dynamic properties, making them similar to those of several
Beckman Coulter fixed angle rotors. If rotor misidentification occurs when your JCF-Z rotor is
used, first perform three runs from 0 to 5000 RPM and back to 0 RPM. If rotor misidentification
recurs, replace the bearings. See the JCF-Z rotor instruction manual (publication JCFZ-IM) for
bearing replacement instructions. Be sure to follow instructions in the next paragraph below
for “wearing in” the bearings.
• On the first few uses of a new JCF-Z rotor, or when the bearings have been replaced in an older
JCF-Z rotor, lubrication has not yet been thoroughly distributed around the bearings in the
bearing housing. The excess lubrication produces drag on the bearings, which can change the
rotor’s dynamic properties enough to cause rotor misidentification. To “wear in” the bearings
and distribute the lubrication, perform three runs from 0 to 5000 RPM and back to 0 RPM.

NOTE If the JCF-Z rotor is misidentified, the run speed will be limited to the maximum speed for the identified
rotor. (The maximum speed for the JCF-Z rotor is 20,000 RPM.)

* In the United States, call 1-800-742-2345. Outside the US, contact your local Beckman Coulter office or visit us at
www.beckman.com.

A-8 PN B10087AE
APPENDIX B
Temperature Calibration Procedure

Introduction

The Avanti J-26S XP centrifuge specification for temperature control is 2°C of the set temperature.
That means that your sample will stay within 2°C of set temperature at all times, after rotor and
system equilibration, described below. (During transient conditions, such as acceleration and
deceleration, the rotor temperature may be outside this range.) The following procedure is
provided for those cases in which temperature control within 1°C is required.

In this procedure, you will measure the temperature of your sample after equilibrating the rotor
temperature for a minimum of 1 hour. Based on the test results, you can then adjust the centrifuge
temperature setting to bring your sample as close as possible to the required temperature.

1 For runs at other than ambient temperature, precool or prewarm the rotor and test samples to
the required temperature.
The length of precooling/prewarming time depends on how different the rotor starting
temperature is from the required run temperature.

2 Install the rotor with adapters, if applicable, and tubes or bottles filled with sample buffer or
water (if above 2°C).

3 Set the speed to 2000 rpm, the time to HOLD, and the temperature to the required run
temperature.
a. Start the run.

4 After 30 minutes, set the required run speed, and allow the system to run for at least 1 hour.

NOTE The system must run for at least 1 hour before the temperature of the buffer or water is
measured. The length of time required for equilibration depends on the rotor and centrifuge starting
temperatures and the ambient room temperature.

5 After at least 1 hour, measure the temperature of the buffer or water using a thermometer or
thermocouple.

6 If the measured and set temperatures are different, note how many degrees different they are
and adjust the set temperature up or down that number of degrees.

PN B10087AE B-1
Temperature Calibration Procedure
Introduction

For example:

If the required sample And the measured Set the temperature to


temperature is buffer/water temperature
is

4°C 6°C 2°C

5°C 4°C 6°C

B-2 PN B10087AE
Beckman Coulter, Inc.
Avanti J Series Centrifuge Warranty
Subject to the exceptions and upon the conditions specified below Beckman Coulter, Inc. agrees to
correct either by repair, or, at its election, by replacement, any defects of material or workmanship
which develop within one (1) year (3 years for the drive motor) after delivery of the Avanti J Series
Centrifuge (the product), to the original buyer by Beckman Coulter or by an authorized
representative, provided that investigation and factory inspection by Beckman Coulter discloses
that such defect developed under normal and proper use.

Some components and accessories by their nature are not intended to and will not function for as
long as one (1) year. If any such component or accessory fails to give reasonable service for a
reasonable period of time, Beckman Coulter will repair or, at its election, replace such component
or accessory. What constitutes either reasonable service and a reasonable period of time shall be
determined solely by Beckman Coulter.

Replacement
Any product claimed to be defective must, if requested by Beckman Coulter, be returned to the
factory, transportation charges prepaid, and will be returned to Buyer with the transportation
charges collect unless the product is found to be defective, in which case Beckman Coulter will pay
all transportation charges.
Beckman Coulter makes no warranty concerning products or accessories not manufactured by it.
In the event of failure of any such product or accessory, Beckman Coulter will give reasonable
assistance to the Buyer in obtaining from the respective manufacturer whatever adjustment is
reasonable in light of the manufacturer’s own warranty.

Conditions
Beckman Coulter shall be released from all obligations under all warranties, either expressed or
implied, if the product covered hereby is repaired or modified by persons other than its own
authorized service personnel, unless such repair by others is made with the written consent of
Beckman Coulter, or unless such repair in the sole opinion of Beckman Coulter is minor, or unless
such modifications is merely the installation of a new Beckman Coulter plug-in component for such
product.

Disclaimer
IT IS EXPRESSLY AGREED THAT THE ABOVE WARRANTY SHALL BE IN LIEU OF ALL WARRANTIES OF
FITNESS AND OF THE WARRANTY OF MERCHANTABILITY AND THAT BECKMAN COULTER, INC.
SHALL HAVE NO LIABILITY FOR SPECIAL OR CONSEQUENTIAL DAMAGES OF ANY KIND
WHATSOEVER ARISING OUT OF THE MANUFACTURE, USE, SALE, HANDLING, REPAIR,
MAINTENANCE, OR REPLACEMENT OF THE PRODUCT.

PN B10087AE Warranty-1
Beckman Coulter, Inc. Avanti J Series Centrifuge Warranty

Warranty-2 PN B10087AE
Related Documents
Rotors and Tubes for Beckman Coulter J2, The JE-5.0 Elutriation System JA-18 Fixed-
J6,and Avanti J Series Centrifuges Angle Rotor
PN JR-IM-10 PN J-TB-035
• Rotors Available in hard copy or electronic pdf by
• Tubes, Bottles, and Accessories request.
• Using Tubes and Accessories
• Using Fixed-Angle Rotors High-Performance and High-Capacity
• Using Swinging-Bucket Rotors Centrifuges Catalog
• Using Vertical-Tube and Rack Type Rotors PN BR-8102
• Care and Maintenance Available at www.beckman.com
• Chemical Resistances
• Temperature Compensation Tables
• Gradient Materials
• Blood Component Separation
Available in electronic pdf or CD-ROM by
request.

Chemical Resistances for Beckman Coulter


Centrifugation Products
PN IN-175
Available in hard copy or electronic pdf by
request.
Available at www.beckman.com

www.beckman.com

© 2019 Beckman Coulter, Inc.


All Rights Reserved

You might also like