0% found this document useful (0 votes)
405 views116 pages

Infinity Kappa

patient monitor drager

Uploaded by

Mohamed Abdulaal
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
405 views116 pages

Infinity Kappa

patient monitor drager

Uploaded by

Mohamed Abdulaal
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 116

Technical Documentation

Kappa
Monitorsystem

Revision 1.0
6013.052
9036308

Because you care


Contents

General

1 Symbols and Definitions 1

2 Notes 1

Function Description

1 Introduction 5

2 Computer Archictecture 5

2.1 Main Processor Bus ............................................................................................................... 5

2.2 Front End Bus ........................................................................................................................ 6

2.3 COMM Bus ............................................................................................................................. 6

2.4 Error Handling ........................................................................................................................ 6

2.5 Main Unit ................................................................................................................................ 7

2.6 Interfaces ................................................................................................................................ 8

2.7 Graphics Subsystem ............................................................................................................ 10

2.8 DSP System ......................................................................................................................... 11

2.9 POD COM Subsystem ......................................................................................................... 13

2.10 Power Conversion ............................................................................................................... 14

2.11 Front Bezel ........................................................................................................................... 17

2.12 MultiMed Front End .............................................................................................................. 17

2.13 Functional Description .......................................................................................................... 18

2.14 Communications ................................................................................................................... 23

2.15 NBP ...................................................................................................................................... 23

2.16 HemoMed Front End ............................................................................................................ 26

2.17 etCO2 Pod ............................................................................................................................ 27


Kappa_IEC_ SMIVZ.fm 02.08.06

2.18 HEMO 2/4 POD ................................................................................................................... 28


Copyright reserved.

2.19 Advanced Comm Option ...................................................................................................... 30

I
Contents

Schematics and Diagrams

1 Schematics and Diagrams 33

1.1 MultiMed Pod Cable Connector on Kappa ........................................................................... 33

1.2 HemoMed Pod Cable Connector on Kappa ......................................................................... 34

1.3 Aux Connector on Kappa ..................................................................................................... 35

1.4 Kappa Network Connector .................................................................................................. 35

1.5 Kappa Analog Output Connector ......................................................................................... 36

1.6 MultiMed 12 Pod .................................................................................................................. 36

1.7 Kappa RS-232, Keypad Input, Alarm Out Connector ........................................................... 37

1.8 Remote Alarm Cable ............................................................................................................ 37

1.9 Analog Cable ........................................................................................................................ 38

1.10 Cardiac Output Intermediate Cable Wiring Diagram ............................................................ 39

1.11 Input Connector on etCO2 Pod ............................................................................................ 39

1.12 HEMO Pod ........................................................................................................................... 40

1.13 Press Adapter Input ............................................................................................................. 41

Fault-Cause-Remedy

1 Introduction 45

2 Recommended Tools and Test Equipment 45

3 Power Problems 45

3.1 Power-On Problems ............................................................................................................. 45

3.2 Power On/Off Piezo Tone Fails to Sound. ............................................................................ 46

3.3 Power-Up Sequence Fails to Complete Properly ................................................................. 46

3.4 Monitor fails to operate on battery power ............................................................................. 47


Kappa_IEC_ SMIVZ.fm 02.08.06

Copyright reserved.

4 Rotary Knob Malfunction 47

5 Fan Malfunction. 47

II
Contents

6 VGA Display Malfunction. 47

7 Fixed Key Fails to Function 48

8 Isolating Cable Malfunctions 48

9 Visible or Audible Alarm Reporting Failure. 48

10 MultiMed POD - Parameter Signal Problems 48

11 NBP 50

11.1 NBP Error Messages ........................................................................................................... 50

11.2 NBP Troubleshooting ........................................................................................................... 50

12 etCO2 Malfunction. 51

13 HEMO2/4 Pod / HemoMed Pod 52

13.1 Readings Missing or Inaccurate ........................................................................................... 52

14 IBP Malfunctions. 52

15 No Printout from Recorder. 54

16 Patient-Related Data Not Retained, or Monitor Fails to Compute Trends 54

17 Software Loading Problems 54

18 Difficulty acquiring export protocol data 55

19 Total or partial loss of network communications 55

20 Problem Report 55

21 Diagnostic/Error Messages 56
Kappa_IEC_ SMIVZ.fm 02.08.06

21.1 Overview of Diagnostic Messages ....................................................................................... 56


Copyright reserved.

III
Kappa_IEC_ SMIVZ.fm 02.08.06

Copyright reserved.

Annex

Test List
Spare Parts
Contents

IV
General

1
2
Kappa General

1 Symbols and Defini-


tions WARNING
A WARNING statement provides important information about a poten-
tially hazardous situation which, if not avoided, could result in death
or serious injury.

CAUTION
A CAUTION statement provides important information about a potentially
hazardous situation which, if not avoided, may result in minor or moderate
injury to the user or patient or in damage to the equipment or other prop-
erty.

NOTE
A NOTE provides additional information intended to avoid inconvenience
during operation or servicing of the equipment.

Definitions according to German standard DIN 31051:


Inspection = examination of actual condition
Maintenance = measures to maintain specified condition
Repair = measures to restore specified condition
Servicing = inspection, maintenance, and repair

2 Notes This Technical Documentation conforms to the IEC 60601-1 standard.

Read each step in every procedure thoroughly before beginning any test.
Always use the proper tools and specified test equipment. If you deviate from
the instructions and/or recommendations in this Technical Documentation,
the equipment may operate improperly or unsafely, or the equipment could
be damaged.

Dräger recommends that only authentic Dräger repair parts be used for main-
tenance. Otherwise the correct functioning of the device maybe compro-
mised.

The maintenance procedures described in this Technical Documentation may


3.0_ Released_Printed on_02.08.06_General_Technical_Documentation.fm

be performed by qualified service personnel only. These maintenance proce-


dures do not replace inspections and servicing by the manufacturer.

The information in this Technical Documentation is confidential and may not


be disclosed to third parties without the prior written consent of the manufac-
turer.

This Technical Documentation is for the purpose of information only. Product


descriptions found in this Technical Documentation are in no way a substitute
for reading and studying the Instructions for Use/Operating Manual enclosed
with the product at the time of delivery.
Copyright reserved.

6013.052 1
General Kappa

NOTE
Unless otherwise stated, reference is made to laws, regulations or stan-
dards (as amended) applicable in the Federal Republic of Germany for
equipment used or serviced in Germany. Users or technicians in all other
countries must verify compliance with local laws or applicable international
standards.

3.0_ Released_Printed on_02.08.06_General_Technical_Documentation.fm


Copyright reserved.

2 6013.052
Function Description

3
4
Kappa Patient Monitor Function Description

1 Introduction The Kappa is a high-end single-board patient monitor. The board provides the
following parameters; 6 lead ECG, Respiration, two Temperatures, SpO2,
NBP, four IBPs, Cardiac Output, and two onboard 5 watt patient isolated ports
for additional parameters. It has connectors for external CRT, user interface,
audio, NBP pneumatic assembly, chart recorder, analog out, defib sync,
memory card, and Uarts. The board also contains the patient isolated front
ends.

2 Computer Archictec- Hardware architecture of the monitors is based on a dual processor design
ture using two Motorola MPC860s with onboard cache. The main processor is
responsible for graphics and communications, while the second processor is
dedicated to data acquisition and algorithm processing. A DSP subsystem
preprocesses the front end data.

There are three major bus structures within the system; MAIN processor bus,
FRONT END bus, and REMOTE COMM bus (see ). The buses operate at dif-
ferent speeds and efficiency. The FRONT END bus and REMOTE COMM
bus have multiple bus masters and common memory to allow exchange
between I/O devices.

The REMOTE COMM bus interfaces to the Advanced Comm Option. This
option includes the main circuit board from the IDS.
Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

Figure 1 Kappa Bus Structure

2.1 Main Processor Bus The Main processor bus is a 32 bit data bus connecting the MPC860 to its
main bank of 16 meg DRAM memory. The Program for the monitor is stored
in 8 meg Flash memory and uploaded to DRAM during initialization. The
Copyright reserved.

DRAM is optimized for multiple word transfers allowing efficient cache fills.
This bus has an optional daughter card connector allowing expansion of the

6013.052 5
Function Description Kappa Patient Monitor

main memory space. The graphics controller is connected to this bus to allow
high bandwidth access to video memory. The bus has a max bandwidth of 40
megbytes/sec.

This bus also has an I/O space implemented in an FPGA. These functions
include audio, chart recorder interface, keypad and rotary knob interface, and
EEPROM. The EEPROM contains serial #’s, calibration constants and con-
figurations. The I/O space also includes the Bridge to the FRONT END bus
and a port to the REMOTE COMM bus. The Bridge to the FRONT END bus is
unidirectional. This means that the Main processor may read and write to the
FRONT END bus, but the Front end processor can not access the MAIN bus.

2.2 Front End Bus The Front End bus is a 32 bit data bus connecting the second MPC860 to its
main bank of 4 meg DRAM memory. The program for this processor is down-
loaded from the main processor during initialization. The DRAM is optimized
for multiple word transfers allowing efficient cache fills. Both processors con-
tain 512K of battery-backed SRAM for trend and other patient data storage.
Data is exchanged through the common memory. This bus has multiple bus
masters that include the following:
• Front End 860
• Main 860
• DSP DMA
• POD Comm DMA (a POD is a configured front end)
• DRAM Refresh

2.3 COMM Bus The COMM bus interfaces to a network controller and other local serial buses
including MIB, lGraphics, Gas Monitoring, and other peripherals. The host is
stalled until completion of all read operations, but is released after a write is
latched to be serialized.

2.4 Error Handling The hardware provides several circuits for error detection, error recovery, and
safety. The main processor bus, front end processor bus, and COMM bus
both have timeouts implemented with the arbiter to prevent a lock up of the
system. The main 860 and the Front End 860 are both protected with watch-
dog timers. If a timer expires, the system initiates a reset and restarts the
monitor. Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

The power supply is also monitored with a piezo alarm that sounds during
power up (for test) and power down. This is to alert the user that the monitor
has turned off. The piezo is also sounded continuously if the monitor does not
reset properly after a watchdog timer has expired and the computer has
halted.

The main unit has been designed as a single board computer used in the
SC7000. In addition to the main board there is a connector board that adds
the functionality of the SC8000 which is different from the SC7000. These
functions include Analog Out buffering, Sync buffering, power for the ISD, fan
control, and circuitry to convert the cartridge connector of the SC7000 to a
pod connector. The other functions of this board are identical to the SC7000
Copyright reserved.

connector board which include EEProm and a cable harness. The main
board has three main sections (see ) -- processor, power conversion, and
front end. Each section is shielded from the others.

6 6013.052
Kappa Patient Monitor Function Description

Figure 2 Kappa Block Diagram

2.5 Main Unit

2.5.1 Processor Section The processor section contains all computer functions of the monitor. It con-
tains two MPC 860 processors, a VGA graphics controller, and a 2181 DSP.
All of these devices communicate using one large FPGA (field programmable
gate array) that is downloaded at initialization. The FPGA contains all custom
circuitry used in the computer, including the processor bridge, comm trans-
mitter, pod com DMA, DSP interface, NBP interface, and I/O interfaces for
Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

both processors. Configuration of the FPGA may be updated with the Soft-
ware through the memory card adapter.

2.5.2 Power Conversion Sec- The power conversion section operates on a DC input from +11 to +15 volts.
tion It switches between the power supply and the internal battery for the proper
power source, and generates all necessary dc voltages for the unit. It charges
and maintains the internal battery. This section also contains the patient isola-
tion for the two internal front ends as well as two general pod comm ports. It
also has the power control for the NBP pneumatics.

2.5.3 Front End Section The front end contains MultiMed and HemoMed circuitry. The MutiMed front
end provides the following parameters; 6-lead ECG, Respiration, Pulse Oxim-
etry, and Temperature. The NBP pressure transducer is also contained in this
Copyright reserved.

front end. The front end is based on a single 16 bit oversampling converter.
Oversampling allows for a reduction in anti-aliasing analog circuitry while
maintaining superior noise rejection. The HemoMed front end provides four
invasive pressures and Thermal Dilution Cardiac Output.

6013.052 7
Function Description Kappa Patient Monitor

2.5.4 Cooling System The cooling system for the main monitor uses a fan mounted on the rear of
the chassis. If the internal temperature of the circuit board exceeds 80° C the
monitor shuts down to prevent damage to the electronics. The monitor does
not restart until the temperature is below the shut off value.

2.5.5 Real Time Clock The Real Time Clock function is implemented with the EPSON-SEIKO
RTC4513 device, and is synchronized by the Central Station.

2.5.6 Non-volatile Memory The shared RAM and real time clock are provided with a lithium battery
Battery Backup and backup circuit to prevent corruption of this non-volatile memory during a
Power Reset power loss condition (both primary and battery power are lost). Note that the
battery used for non-volatile memory backup should not be confused with the
internal and external batteries used to provide power to the monitor base unit
when primary power is lost. Non-volatile memory lithium battery backup is
controlled by a power supervisory device that provides a power reset during
a power loss condition.

NOTE
No provisions have been made to recharge non-volatile memory backup
battery. Eventually (aprrox.10 years), battery must be replaced.

2.5.7 MPC 860 Communica- MPC 860 has an embedded communications processor capable of executing
tion Channels several protocols such as UART or Ethernet. The 860 communications chan-
nels are used as follows:

Main Processor

SCC1 Ethernet 10 Mbits/sec (future option)

SCC2 SC 9015 UART selectable baud

SCC3 MVP-1 UART selectable baud

SCC4 MVP-2 UART selectable baud

SMC1 main diag UART 19.2 Kbaud

SPI a/d (power monitor) Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

Front End Processor

SCC3 serial pod dataSCC4 serial pod data

SMC1 front end diag UART

An additional UART implemented in the FPGA contains a large FIFO and


interfaces to the chart recorder.

2.6 Interfaces
Copyright reserved.

2.6.1 Local Fixed Keys Inter- The monitor base unit has twelve fixed function keys and a fixed key dedi-
face cated as a power on/off switch. The power on/off switch is unique in that it is
not directly available via a status read command, but rather is input to the

8 6013.052
Kappa Patient Monitor Function Description

power supply subsystem interface, where the switch state is detected and
processed. Detection of a power off condition causes an interrupt to the host
processor.

2.6.2 Rotary Knob Interface The rotary knob is a 16 detent rotary knob. Each detent position indicates a
"click" clockwise or counter-clockwise. The change in detent position is
detected via a 2 bit quadrature code that changes value every time the rotary
knob is moved into a detent position. Also included in the rotary knob is a
push button switch that is operated by a press/release action. This switch is
used to select menu items on the screen.

2.6.3 Fast Analog Output The ANALOG OUT interface consists of two identical channels. Each ANA-
LOG OUT channel provides a 12 bit D/A function. The design uses a dual
DAC to produce the D/A conversion. The sampled analog data is then
passed through a 2 pole low pass filter. The analog output has a maximum
delay of 20 ms, and can be used for a defibrillator or balloon pump.

Separate Pacer Spike generation circuits for analog outputs 1 and 2 are pro-
vided.

2.6.4 HiFi Audible Alarm Inter- The Audible alarm interface consists of an FM synthesis and Audio DAC chip
face set. There is also power amplifier drive circuitry for the two speaker inter-
faces: the internal speaker located in the base unit and the speaker located
in the remote CRT. Circuitry has been included to provide a mechanism for
automatically generating an error tone when a watchdog failure occurs via
the piezo alarm ).

The chip is loaded with tone frequency, pitch, harmonics, and volume infor-
mation by the host processor, which controls the duration of the tone. The
audio DAC converts the received sampled tone data and produces a sampled
analog representation of the tone data.

The local speaker interface (also designated as main speaker interface) is


designed for an 8 ohm speaker load. This local speaker interface produces
1 watt of power into an 8 ohm load, and has thermal shutdown capability.

The remote speaker interface is designed to produce a 1 Vrms maximum sig-


nal into a 1 kohm load, and provides an ac coupled output.
Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

2.6.5 LED/Status Interface Five LEDs provide information in the present Kappa configuration. Two are
dedicated to the front end processor, to the DSP, and two to the main proces-
sor.

2.6.6 QRS Sync Out Interface A QRS sync output is provided. The QRS SYNC OUTPUT is an open collec-
tor type output driver that is pulled up to +12 volts (active HIGH). The output
is initialized to Gnd on reset or power on.

This QRS signal is available via an external connector mounted on the main
PC board. High level = +6 V min (10 KW load), +12.6 V (no load); Low level
(no QRS) = 1 V @ 5ma.

2.6.7 Local Alarm Out Inter- A Local Alarm output is provided. This Local Alarm Output is an open collec-
Copyright reserved.

face tor type output driver that is pulled up to +12 volts. The output is initialized to
ground (0 volts) on reset or power on (active HIGH).

Loopback status is available via a status read command.

6013.052 9
Function Description Kappa Patient Monitor

The Local Alarm Out signal is available via an external connector mounted on
the main PC board.

2.6.8 Recorder Interface The recorder interface provides all of the necessary control, data and power
supply signals required to drive an external recorder. The interface consists of
current limited DC power and a UART with handshake signals. The UART is
implemented in the main processor FPGA to allow for an extended FIFO.

2.6.9 Serial EEPROMS Four serial EEPROM devices, which contain the Monitor serial number, Eth-
ernet address, NBP pneumatic characterization and calibration constants,
and monitor setups, are located on the connector I/O board. If the main pro-
cessor board is replaced the monitor will keep its set ups from these serial
EEPROMs.

Two EEPROMs can be written only at the factory, and contain the Monitor
serial number and Ethernet ID address. The other devices are writable by the
main processor and are changed during service menu setups. These devices
are used for the monitor as well as network setups, device compatibility, and
software feature locks.

Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

Figure 3 Graphics Subsystem

2.7 Graphics Subsystem

2.7.1 Overview The Graphics Subsystem is based on a commercial VGA controller (see ),
and drives a CRT display from a local memory used to refresh the screen. It
uses a special video crystal which enables it to synchronize to most video
standards. The graphics chip is capable of running resolutions such as 800 x
600, when these displays are added to the monitor. The standard resolution
is set to 640 X 480.
Copyright reserved.

10 6013.052
Kappa Patient Monitor Function Description

2.7.2 Functional Description The VGA subsystem is designed to optimize the Bitblit operation, which
allows for quick updates of the screen. This is accomplished by writing
images to non-viewable areas of video memory before they are needed and
copying them to the screen on demand. The copy function is performed by
the VGA controller.

2.7.3 Video Output The Graphics Subsystem provides output to a standard VGA monitor. The
CRT interface uses three 8 bit DACs for its three color outputs. The front
bezel interface is digital and contains 6 bits for each color.
Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

Figure 4 DSP Subsystem


Copyright reserved.

2.8 DSP System The monitor uses a DSP for preprocessing of oversampled data (see ). The
DSP is a specialized microprocessor that executes high speed repetitive
functions such as digital filters. The DSP acquires data from the incoming

6013.052 11
Function Description Kappa Patient Monitor

serial pod comm data streams. The data sent to the DSP is selected by the
control words in the pod com memory buffer. Typically only high acquisition
rate data is sent to the DSP.

The DSP has two other communication ports both of which can access the
internal 32Kword memory. The IDMA port is used to DMA data to and from
the common memory. Bus sizing logic converts the DSP 16 bit port to the 32
bit FRONT END bus. During initialization this path is used to download code
to the DSP. The main processor takes control of the DMA port during this
time. Once the system is operational the DSP takes control of the DMA con-
troller by using its I/O port. The I/O port is a dedicated 8 bit path into the main
FPGA, which allows the DSP access to the DSP DMA controller as well as
other internal FPGA registers, including analog out and QRS sync.

Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm


Copyright reserved.

Figure 5 Pod Communications

12 6013.052
Kappa Patient Monitor Function Description

2.9 POD COM Sub- A pod is a front end device that acquires data for a particular set of parame-
system ters. A pod may contain a processor and return preprocessed data or it may
provide raw A/D samples.

Refer to Figure 5.

2.9.1 Overview The pod com subsystem has four channels. Channel 1 is dedicated to the
two internal front ends; the MultiMed and HemoMed. Channel 2 is dedicated
to the slot on front for etCO2. This connector does not require patient isola-
tion and has higher power than the pod com connectors. Channels 3 and 4
are both used to communicate with external pods. They have full patient iso-
lation for both power and data.

Data acquisition of the monitor is controlled by several DMA controllers that


operate on circular buffers residing in common memory on the FRONT END
bus. There are four channels, each allocated a 16 bit transmit buffer and a 16
bit receive buffer. It takes four 32 bit transfers to update one location in every
buffer, since each access consists of high and low data from different chan-
nels. The transmit buffer tells the pod either what sample to take or to change
a control setting. The receive buffer contains a/d samples and status informa-
tion from the pod. A control register in the FPGA sets a mux to the DSP’s
communication port and connects the selected pod com channel.

2.9.2 Outputs

2.9.3 Error Handling The pod com channels provide error detection by performing CRC checks on
data in both directions. CRC errors are reported to the front end processor
through interrupts.
Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

Copyright reserved.

6013.052 13
Function Description Kappa Patient Monitor

Figure 6 Power Conversion

2.10 Power Conversion Refer to Figure 6.

2.10.1 Power Control Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

Power Buss Most monitor loads are powered from a DC power buss, called VBUSS,
within the monitor. VBUSS powers the +3.3 VDC, +5.0 VDC, ±12 VDC,
+40 VDC and charger power converters. VBUSS also powers the external
pods, cartridge, strip recorder and backlight. The NBP pump and valves as
well as the internal multimed and hemomed front ends are powered from the
regulated +12 V supply.

Control and Load The switching of the VBUSS power inputs and the power converters is man-
Sequencing aged by the power supply gate array. This gate array controls the power on
and power off of the monitor, and the battery charging process. It also pro-
vides a safety timer for the NBP pneumatics, which are controlled by the main
processor FPGA.
Copyright reserved.

Logic circuits on the main gate array sequence the power to the Pods, Car-
tridges, and Recorder to reduce power on load transients.

14 6013.052
Kappa Patient Monitor Function Description

Power On / Off The monitor is normally switched on by the user pushing the On/Off button for
at least 1 second. (The monitor may switch on when the switch is pushed for
as short a time as 50 msec.)

The power down sequence may be initiated either by the user pushing the
on/off switch for at least 1 second or when the batteries are depleted. When
the power down sequence is initiated, the power conversion board control
logic generates an interrupt for the processor. 100 ms later, the power supply
shuts down. An immediate shutdown is initiated if a power fault occurs (such
as overvoltage).

Power Source Control Power for the monitor is provided by the internal power supply or internal bat-
tery.

This input is monitored by a voltage comparator to determine that adequate


voltage is present for internal power supply operation. The main battery also
has a voltage comparator indicating that its voltage is high enough to provide
power.

Based on the information provided by the comparators, a power source is


connected to VBUSS in the priority of main power supply and then battery.

Battery Charging The battery charger is a two-level constant voltage charger with a fixed cur-
rent limit and temperature compensated voltage levels. When the main power
comes on, the battery is fast charged at the high voltage until the current
drops below a specific threshold. Then the charger voltage drops to the lower
“float” voltage.

Indicator LEDs Two green LED indicators on the front bezel of the monitor indicate power
and charger status, as given in Table 1.

Table 1 Power and Charger LED Indicators

LED CONDITION LED STATE


Power Processor power on on
Processor power off off
Charger Main power on on
Main power off off
Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

*Battery or power fault off


* The charger LED is off if the battery temperature is excessive or if there is
a power fault.

2.10.2 Power Mode Indication The source of power is indicated to the processor via the power mode bits, as
given in Table 2.

Table 2 Power Mode Table

MODE1 MODE0 INDICATION


1 X operating on main power
Copyright reserved.

0 1 operating on battery

6013.052 15
Function Description Kappa Patient Monitor

2.10.3 Piezo Alarm The piezo alarm activates at power up, power down, and if a software watch-
dog is activated. At turn on, the software shuts the piezo off after two sec-
onds. The piezo functions are as indicated in Table 3.

Table 3 Piezo Alarm

Cause Duration
Turn on 2 seconds
Turn off >4 seconds, <10 seconds
Processor watchdog Continous (until sucessful reboot)
+5 V Undervoltage >4 seconds, <10 seconds
+3.3 V Undervoltage >4 seconds, < 10 seconds

2.10.4 Fault Protection Reverse polarity protection for the battery and the main power input are pro-
vided by shunt diodes and fuses on the connector board. There is also a fuse
in series with the battery harness. +5 V, +3.3 V, and +12 V supplies are pro-
vided with overvoltage protection.

The battery has a temperature sensor on the Connector I/O PC board that is
used to disable charge or discharge of the battery if the temperature is exces-
sive.

A temperature sensor in the power supply section of the main board shuts
down the power system if the board temperature is excessive.

All power converters are fused to limit fault currents.

2.10.5 External Pod Overload External pod current limit circuits are implemented as follows:
Protection
When an overload occurs, the load is switched off after the 0.2 second over-
load timeout. A retry occurs after 5 seconds.

2.10.6 Electrical Specifications The following specifications indicate the design limits of the power system
and do not relate to a present design configuration of the Kappa.

2.10.7 Power Supply Power Supply Input Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

100 Vac @ 2.5A; 240 Vac @ 1.3A; 50/60 Hz

Power Supply Output

11.0 to 15 volts DC @ 6.0 Amps Max.Buss Fault Detection <8.97 V±1%

Battery Source (in DC Power Supply): 12 V Lead Acid, 9.8 to 15 VDC @


6.0 Amps Max.

2.10.8 Battery Specifications Voltage 12 V


(in DC Power Supply)
Discharge Time: 20 minutes
Copyright reserved.

16 6013.052
Kappa Patient Monitor Function Description

2.11 Front Bezel

2.11.1 Introduction The Front Bezel provides an interface to the various operator related func-
tions. The interface consists of circuitry and connectors that allow the main
processor to access all of the operator related functions.

2.11.2 Functional Description The front panel interface section of the Connector I/O PC board provides an
interface between the main board and the front bezel components. It is a
cable harness for these components and is unique in its construction. The
board provides for unit rotary knob input and keypad interface.

2.11.3 Local Rotary Knob/fixed The Kappa base unit has twelve fixed keys. An additional key is dedicated as
Keys Interface the power on/standby switch. The rotary knob interface provides a 2-bit
encoder output and also a rotary knob push button signal output. All of the
key/rotary knob signals are filtered. All of the keypad switches have pull-
downs except the power on/standby switch. Thus, the power switch signal
output from the front bezel is pulled up by the power switch interface located
in the power supply section.

2.11.4 Battery/power LED Inter- The battery LED is turned on or off via the associated LED control signal from
face the main board. The power LED is connected to +5 V. Both the battery and
power led's are green when turned on. The power and battery LED's have
been integrated into the membrane switch interface used for the fixed key
and power on/standby switch. The LED on/off control signals are provided by
the power supply.

2.12 MultiMed Front End The MutiMed front end section of the main board combines 6-lead ECG, 2-
lead respiration, temperature, and saturated oxygen data gathered by the
MultiMed Pod from transducers at the patient and converts them to digital
form for transmission through isolators to the computer section of the main
board. This section also houses the NBP pressure transducer which uses the
same acquisition system.

The hardware design uses a single oversampling 16 bit converter to measure


all of the parameters. This allows bulky analog filters to be replaced by soft-
ware filters. Careful shielding and filters protect against very high frequency
interference from upsetting measurements.
Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

2.12.1 Safety • Patient isolation withstands 5 kV during defib.


• Leakage currents are limited to safe values normally and during single
fault conditions.
• Patient is protected against electrosurgical burns at the electrodes.
• Defibrillation protection does not drain excessive current away from the
patient.
Copyright reserved.

6013.052 17
Function Description Kappa Patient Monitor

Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

Figure 7 MultiMed Front End

Specially shielded connectors and cables are used to provide excellent


immunity up to 1000 MHz and can not be touched by patient even when dis-
connected.

Single cable from MultiMed Pod to Kappa reduces clutter between bed and
monitor.

2.13 Functional Descrip- Transducers gather physiological data at the patient and feed them into the
Copyright reserved.

tion small MultiMed Pod at the bed. The MultiMed Pod in turn is connected via a
3-meter cable to the MultiMed front end in the main unit where analog ECG,
Respiration, Temperature, and SpO2 signals are converted to digital form and
sent through isolators for processing.

18 6013.052
Kappa Patient Monitor Function Description

2.13.1 ECG/Resp The MultiMed Pod located close to the patient accepts a set of 3, 5 or 6
shielded ECG electrode leads, an SpO2 (Nellcor) cable adapter, and a tem-
perature sensor. The ECG section contains RF filters, and overvoltage
clamps that include 1k series resistors to limit shunting of defibrillator current.
The SpO2 and temperature sections also contain RF filters. Impedance respi-
ration is sensed through the ECG electodes. Void-free potting and internal
shielding enable compact containment of high voltage defibrillator and elec-
trosurgery pulses. The small interconnecting cable to the main assembly is
captive at the MultiMed POD but plugs into the MultiMed front end via a spe-
cially shielded connector.

The front end accepts physiological signals from the MultiMed POD connec-
tor and feeds temperature, respiration, and ECG signals via RF filters, config-
uration multiplexers, and pre-amplifiers to a high-speed multiplexer driving a
16-bit analog-to-digital (A/D) converter. The data stream is sent to the Main
Processor board via an opto-isolator. Control commands from the Processor
are sent out to the front end on a similar isolating link. Isolated DC power is
also provided.

The ECG signals are conductively coupled to the isolated circuits via current-
limiting series resistors, whereas the SpO2 signals are optically isolated at
the transducer. Temperature signals are doubly insulated at the patient by
disposable boots on the sensors. AC (40 kHz) excitation currents for respira-
tion monotoring are dc-isolated by high-voltage ceramic capacitors.

The A/D samples the following parameters:

Table 4 Parameter Sampling Table

Parameter # of Channels
ECG 4
Pace 2
SpO2 Red 1
SpO2 IR 1
NBP 1
Resp 1
Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

Temp 2

The pace signal samples are used directly by the DSP to detect pace pulses.
All other signals are decimated and filtered using digital signal processing to
the above specifications. Additional filtering is user selectable and invokes
additional digital signal processing in the computer section of the board. The
high oversampling rate is required to minimize the requirements (and size) of
the analog anti alias filters. Superior rejection to ESU and other types of inter-
ference is achieved with this type of design.

ECG • Pacer pulses may be detectable by software on two lead-pairs.


• Bandwidth is set flexibly by software filters.
Copyright reserved.

• Reconfigurable neutral selector can drive any electrode.


• Lead-on detection functions with even poor electrodes.
• Calibration voltages can be superimposed on patient wave-forms or onto
flat baselines.

6013.052 19
Function Description Kappa Patient Monitor

See Figure 8. Composite electrocardiographic (ECG) signals generated by


the heart and by a pacemaker are filtered to reduce RF interference from
impedance respiration and electrosurgery and then injected with dc lead-off
detection currents. Over-voltage clamps protect the semiconductors from the
surges passing the sparkgaps in the MultiMed Pod and also reduce the dc
current applied to the patient due to a component fault.

Figure 8 Lead Forming Network

The Wilson point, "W", the average of the LA, RA, and LL electrode poten-
tials, serves as the negative reference potential for the V and V' lead-pairs
and is also used as a measure of the common-mode potential of the patient.
By driving the isolated common of the front end at the same potential as the
Wilson point, the common-mode voltage across the electrodes is reduced
Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm
nearly to zero and the effective common-mode rejection is improved. As most
of the common-mode current is now forced through the neutral electrode, it
becomes noisier and hence is not used as part of another signal path.
Switches are provided to select other electrodes to be neutral if the RL elec-
trode is off or missing. If the V' electrode is present, then it can be selected to
be neutral so that the three Einthoven and the V lead pairs can still be used.
However, the V' lead-pair will be corrupted due to neutral current noise. Simi-
larly the V electrode can be selected to be neutral. Now that the RL is discon-
nected from the neutral driver, its potential can be monitored to determine
whether it has been reconnected to the patient and thus is able to be rese-
lected to be neutral.

If only the three Einthoven (LA, RA, and LL) electrodes are connected, one is
selected as neutral leaving the remaining two electrodes to form one valid
Copyright reserved.

lead-pair. The "W" now contains the neutral drive signal which bypasses the
neutral electrode and reduces the gain of the neutral driver amplifier. To
improve the resulting poor common-mode rejection, a Wilson Grounding
"WG" switch is activated to selectively disable the offending input to the "W".

20 6013.052
Kappa Patient Monitor Function Description

Figure 9 Respiration Functional Block Diagram

Respiration Refer to Figure 9.


• Respiration is both ac- and dc-coupled in hardware. DC is used for high Z
sensing; ac is used for signal acquisition.
• Respiration may be monitored on leads I and II.
• Detection sensitivity has low dependence on base resistance or electrode
unbalance.

Impedance respiration is monitored by injecting a nominally 40 kHz square


wave of current into one ECG electrode and removing it at another ECG elec-
trode. The resulting 40 kHz voltage drop across those electrodes is propor-
tional to the impedance. Specially balanced true current sources do not load
the ECG electrodes or distort the ECG morphology. The waveform of the cur-
rent is preemphasized to reduce bypassing effects of cable capacitance. The
returning 40 kHz differential voltage is amplified, synchronously demodu-
lated, and low-pass filtered. The resulting dc-coupled waveform is converted
to single-ended form, further low-pass filtered, and passed to the A/D multi-
plexer. An ac-coupled stage with an "autobloc" dc-restorer feeding a separate
input to the A/D multiplexer also provides additional gain.
Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

Figure 10 Temperature Functional Block Diagram

2.13.2 Temperature Refer to Figure 10.


• Designed to meet the stringent German PTB requirements including
Copyright reserved.

detection of marginal accuracy due to degradation of a single component.


• A second temperature channel is also available.

6013.052 21
Function Description Kappa Patient Monitor

Temperature is sensed at the patient by a non-linear negative-temperature-


coefficient thermistor. This is linearized with a precision resistor network and
excited by the same reference as the A/D converter to a produce ratiometric
digital output. An input multiplexer (MUX) selects among the external signal
and internal reference dividers simulating -5 and +50 °C. The dc amplifier
matches the dynamic range of the A/D by combining, amplifying, and pre-
cisely offsetting the small signal from the multiplexer. Power supplies whose
failure would invalidate temperature measurements are also monitored and
compared against the A/D reference.

Figure 11 SpO2 Functional Block Diagram

2.13.3 SpO2 Determination of the concentration of oxygen in the blood depends on the
principle that the absorption of red (R) light depends on the degree of oxy-
genation of the blood, whereas the absorption of infrared (IR) radiation is
independent of oxygenation and causes only constant attenuation. Refer to
Figure 11. In the SpO2 sensor, R and IR emitting leds are alternately pulsed
on at a 25% duty cycle. The intensity of light (including ambient) transmitted
through or scattered by the blood is converted to a current by a photodiode in
the sensor. The current that appears when both leds are off depends mainly
on the ambient light. This ambient contribution is later subtracted to leave Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

only the R or IR signal levels. The large dynamic range of the light intensities
requires constant automatic monitoring and adjustment.

The intensities of the R and IR sources are independently controlled by two


digital-analog converters attenuating the 2.5 V reference.

Attenuated radiation falling on the photodiode in the sensor is converted to a


current which passes through an RF filter balun in the HVPOD and enters the
current-to-voltage converters in the MultiMed front end. The resulting unipolar
stream of pulses is then ac-coupled to a controllable-gain differential ampli-
fier. The signal is then synchronously demodulated into Red and IRed signals
with ambient light subtracted. Additional gain control, filtering, and signal off-
set are provided for each signal prior to A/D conversion.
Copyright reserved.

The calibration of each sensor is coded into the value of a precision resistor
built into the sensor. The value of this resistor is sensed by forming a voltage
divider. The value of the resistor ratio is read by a separate A/D input, and out
of range values are interpreted as “sensor unplugged.”

22 6013.052
Kappa Patient Monitor Function Description

2.14 Communications The multiplexers and A/D are controlled by the Main Processor via a
Manchester-encoded serial communications channel (Pod Com) optically
coupled to the isolated front end. Most of the digital logic is contained in the
MultiMed FPGA. Outputs from the A/D are Manchester-encoded in the Mul-
tiMed FPGA and fed to the opto-coupled data flow to the Main Processor.

A power-on monitor resets the FPGA until both ±5 V have risen to normal
range. The isolated dc-dc converters are synchronized to the data acquisition
sequence via the Main Processor FPGA. The A/D converter is automatically
calibrated after the power-on reset is cleared.

Figure 12 NBP Functional Block Diagram


Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

2.15 NBP Refer to Figure 12.

2.15.1 Introduction The NBP design measures blood pressure non-invasively using an inflatable
cuff and the oscillometric method. The NBP algorithms are performed in the
front end processor. The NBP circuit contains two pressure transducers
which measure the hose pressure. The second redundant pressure sensor is
used to measure overpressure for safety. This pressure transducer is
mounted in the power section while the other pressure transducer is mounted
in the MultiMed front end. A plastic manifold connects the two transducers
together and to the pneumatic assembly in the rear case. The MultiMed front
end A/D samples the pressure transducer.
Copyright reserved.

2.15.2 Pneumatic Subassembly The pneumatic subassembly consists of two modulating solenoid valves (V1,
V2), a pump (P1), a filter, and a manifold. The manifold provides the inter-
connection of the air passages between the individual components and pro-

6013.052 23
Function Description Kappa Patient Monitor

vides for their mechanical mounting. It also provides an acoustic attenuation


of the valve and pump noise. The filters prevent contamination from entering
the pneumatic system from the cuff hose or ambient air.

P1 provides the pressurized air to inflate the blood pressure cuff. V1 and V2
are used to control the air flow during the de-flation phase of a blood pressure
measurement. V1 is a normally closed exhaust valve with a relatively small
orifice. V2 is a normally open exhaust valve with a comparatively large orifice.

When a blood pressure measurement is initiated V2 is closed, P1 is turned


on and the rising cuff pressure is monitored via pressure transducers. When
the pressure has reached the target inflation pressure, P1 is turned off. Neo-
nate inflation cycles are identical except that a speed control circuit is used to
reduce the pump output to approximately 15% of the adult mode.

After the inflation, there is a short delay after the pump stops to allow thermal
transients to settle. Either V1 or V2 is now modulated to control the deflation
rate. The choice of V1 or V2 and the initial pulse width is made based on the
inflation cycle. The chosen valve is modulated and the pulse width (open
time) is continuously adjusted to provide a constant deflation rate. If initial
deflation was started with V1 the software may determine that it needs to
switch to V2 to maintain proper deflation. In any case when the measurement
cycle is complete, V2 is opened fully (de-energized) to allow for rapid defla-
tion.

2.15.3 Transducers The measurement pressure transducer is DC coupled to a 16 bit A/D con-
verter so that cuff pressure is measured with adequate resolution to detect
blood pressure pulses.

The overpressure transducer has two threshold settings. The adult setting is
300 ±30 mmHg and the nominal neonatal setting is 158 ±7 mmHg. Both
transducers share a common manifold and are mounted on the main PC
board.

2.15.4 Pneumatic Controls The P1 control provides 3 functions.


• It limits current to the pump when the pump starts to prevent power sup-
ply overload.
• It rapidly decelerates the pump when the pump is shut off, by applying a Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm
low resistance across the motor.
• It provides a closed loop speed control for low speed neonatal operation.

A relatively high pulse voltage is used to drive V1 and V2 to get quick


response. This pulse lasts for approximately 2 milliseconds after which time
the valve voltage is lowered to a holding value. At the end of the valve "on"
time period, the valve voltage is allowed to reverse and the energy stored in
the solenoid inductance is rapidly released into a relatively high voltage
clamp circuit.

P1 and V2 are supplied by a redundant power switch so that, under fault con-
ditions, they can be de-energized.
Copyright reserved.

2.15.5 Safety timer The software limits measurement time to 119 secs for adult mode, 89 secs for
neonatal mode and 59 secs for French neonatal mode. A safety timer circuit
monitors current in P1 and V2, and if due to some failure (hardware or soft-
ware), P1 or V2 remain activated for more than 120±1 seconds in adult

24 6013.052
Kappa Patient Monitor Function Description

mode, 90±1 seconds for neonatal mode or 60±1 seconds in French neonatal
mode, the circuit latches on, causing the redundant power switch to P1 and
V2 to switch off. When the safety timer latch has been set, V1 is opened as
an additional safety feature. Only recycling the monitor resets the safety timer
latch. The safety timer circuit is functionally independent of the logic gate
array.

When the unit is powered up, the safety timer is de-activated until the pump is
started the first time. This feature allows service calibration without triggering
the safety timer. Once the pump has been activated the timer circuit becomes
functional.

2.15.6 Logic gate array The main FPGA provides the following control functions for the pneumatics
and the communications.
• Clock generation for safety timer
• 12 bit 20 Hz PWM and pulse control for V1 and V2
• Pump control
• Neonatal mode switching of pump and overpressure
• Safety logic

2.15.7 Non-volatile memory A EEPROM stores pneumatic component flow factors. During calibration at
production system test and in the field, a 0.5 liter canister is connected to the
NBP input on the monitor. The monitor automatically measures the pump and
valve flow rates and determines their flow factors for the use in the flow con-
trol algorithm.

2.15.8 Hose detection An electromagnetic coil located at the hose connector detects the metal in the
hose connector when the connector is present.

2.15.9 Watchdog Timer A watchdog timer is implemented in the power conversion FPGA to monitor
the safety timer clock input from the main FPGA in case the main FPGA or its
crystal become damaged.
Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

Copyright reserved.

6013.052 25
Function Description Kappa Patient Monitor

Figure 13 HemoMed Front End

2.16 HemoMed Front End

2.16.1 Introduction Refer to Figure 13. The HemoMed front end section of the monitor’s main
board takes invasive blood pressure, and thermal dilution cardiac output data
gathered by the HemoMed Pod from transducers at the patient and converts
them to digital form for transmission through isolators to the computer section
of the main board. The HemoMed front end may also be used with a single or
dual pressure cable instead of using the HemoMed.

Pressure The pressure data acquisition front end is designed to operate with resistive
strain gage pressure transducers having an output impedance of less than
3000 Ohms and an input impedance between 3000 and 200 Ohms. Excita-
tion voltage is applied in pairs. Press 1 and 3 share a driver as well as Press Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

2 and 4. The output signals generated from the pressure sensors are passed
through filter and clamp networks which limit and filter RF noise. The pres-
sure excitations are monitored for fault detection.

Cardiac Output The two thermistor signals are connected to a precision resistor network to
linearize voltage vs temp curve of the thermistor. The thermistor signals are
filtered and clamped before amplification. Two calibration voltages are also
sampled by the A/D converter to correct amplifier offset and gain errors. The
catheter also has a reference resistor which is read for calibration.

The front bezel switches on the HemoMed are converted to unique voltages
when pressed, allowing a voltage to be sent to the A/D converter, which can
be decoded by the front end processor into the corresponding switch closure.
Copyright reserved.

26 6013.052
Kappa Patient Monitor Function Description

Figure 14 etCO2 Sensing Process Functional Block Diagram

2.17 etCO2 Pod The etCO2 pod non-invasively monitors end-tidal CO2 using a technique that
relies on the selective absorption properties of the CO2 to specific frequen-
cies of infra-red radiation. See Figure 14.

In the sensor a thick film infra-red source is pulsed at a rate of approximately


87 Hz, generating a broad- band spectrum of IR. Selective filtering separates
this into two narrow regions, one inside and one outside the band of CO2
absorption. The detector associated with the filter outside the band of CO2
absorption records the maximum level of the source energy since the signal it
receives is not affected by CO2. It provides a baseline which serves as a Ref-
erence for the level of CO2 in the airway. The other detector senses a filtered
energy level modified by the presence of CO2. As the level of CO2 increases,
the CO2 gas molecules in the airway absorb more of the light energy and less
signal reaches the detector. This signal, converted by the detector, is referred
to as the Data signal. Current through the thick-film source is bidirectional to
offset the tendency of particles within the source to migrate when exposed to
a strong unidirectional electric field caused by current flow only in one direc-
tion. This keeps the structure of the source uniform and enhances system
Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

integrity and life of the product.

To acquire a precise level of CO2, both channels are simultaneously sampled


and the level of CO2 is determined from the ratio of the Data and the Refer-
ence channels. The ratio is compared to a look-up table in memory to estab-
lish the correct value in units of mmHg.

The pod then sends the results to the host system for further processing and
display.

2.17.1 System Memory The system has three types of memory:


– PROM: Programmable Read Only Memory
– SRAM: Static Random Access Memory
Copyright reserved.

– EEPROM: Electrically Erasable Read Only Memory

6013.052 27
Function Description Kappa Patient Monitor

PROM stores the pod's program. Its contents remain intact even when power
is removed from the pod. It has been socketed to allow for future program
updates, if required. Besides containing the pod's program, it also contains
various look-up tables for calculating CO2 parameters and the Interrupt Vec-
tor Table.

The system's Static RAM functions as a scratch pad to temporarily hold vari-
ous system variables until they are either no longer needed by the system
and are overwritten with new information, or power is removed from the pod
and the RAM contents are lost.

The EEPROM holds system parameter information that must be retained


when power is removed, but must also be modifiable by the processor. The
device contains multiple copies of system information such as calibration fac-
tors, sensor serial number, and span cell number, to ensure data integrity.

A Supervisor chip performs various monitoring tasks to ensure that the micro-
processor and system run properly.

2.17.2 User Interface The user interface provides capability for airway and adapter calibration, and
also compensation for effects of N2O and O2. When calibrating the acces-
sory assembly, switches inside the sensor, one for the Zero Cell and one for
the Span Cell, tell the processor when the assembly has been placed on the
proper cell for system calibration.

2.18 HEMO 2/4 POD

Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm


Copyright reserved.

Figure 15 HEMO 2/4 POD Functional Block Diagram

28 6013.052
Kappa Patient Monitor Function Description

2.18.1 Functional Description HEMO 2/4 PODs have provisions for monitoring either 2 or 4 invasive blood
pressures, 2 temperatures and cardiac output. See Figure 15.

2.18.2 Pressure The pressure data acquisition front end is designed to operate with resistive
strain gage pressure transducers having an output impedance of less than
3000 Ohms and an input impedance between 3000 and 200 Ohms. Excita-
tion voltage is applied, one at a time, to each resistive strain gauge pressure
transducers by a single, current limited voltage reference circuit which is time-
multiplexed across four pressure sensors. The differential output signals gen-
erated by the pressure sensors are passed through filter and clamp networks
which limit the differential and common mode voltage swings and filter out RF
noise.

Next, the signals enter a functional block that converts the differential signals
into single ended signals which are then presented one at a time in a time-
multiplexed fashion to a fixed gain single ended amplifier. Calibration volt-
ages for zero and 200 mmHg are periodically switched into the amplifier input
to correct errors in amplifier offset and gain respectively.

An A/D converter samples the resulting output voltage. Timing is coordinated


by the logic gate array.

Figure 16 IBP Functional Block Diagram


Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

2.18.3 IBP Each of two thermistors is connected to a functional block that consists of a
precision resistor network to partially linearize the voltage verses temperature
transfer curve of the thermistor. This functional block also consists of means
for filtering RF noise and limiting the voltage swing. A 4.5 Volt reference is
connected to power each linearization network.

A multiplexer selects one of the two temperatures or one of the two calibra-
tion points and connects the voltage to the input of a fixed gain amplifier. The
two calibration points are used to correct gain and offset errors in the ampli-
fier circuits.

An offset is added to center the signal within the dynamic range of the A/D
Copyright reserved.

converter. The signal is then further multiplexed with two power supply volt-
age monitors and Cardiac Output. A fixed gain of two is finally applied to
match the signal range to the full scale range of the A/D converter.

6013.052 29
Function Description Kappa Patient Monitor

2.18.4 Cardiac Output Cardiac output operates in much the same way as temperature. The ther-
mistor signals are filtered and clamped then multiplexed to the input of a fixed
gain amplifier. Two calibration voltages are also multiplexed in to correct
amplifier offset and gain errors.

Next, an offset is added to the signal to center it to the dynamic range of the
A/D converter. The amplified signal is then multiplexed with temperature, then
through a fixed gain of two and finally to the A/D converter.

2.18.5 EEPROM Storage Two EEPROM's are used for non-volatile information storage. One EEPROM
is used for reading and writing data that changes during the operation of the
POD, such as pressure offsets, the other stores more permanent information
such as POD serial number and is therefore write protected. A state machine
inside the logic gate array supports communications between the Host and
the two EEPROM's. A mechanism is provided which allows service personnel
to disable the write protection of the otherwise write protected EEPROM.

2.18.6 LCD and Push Buttons A total of 16 LCD characters are provided for use as pressure labels. Each
pressure channel is allocated 4 LCD characters. The Logic Gate Array sup-
ports communication of controll between the Host and the LCD's.

Up to three push buttons are provided for user interface. There is one for
pressure zero, one for Cardiac Output Start and one spare. The interface of
the buttons to the Host is handled by the gate array.

2.18.7 Current Limiting the Volt- In the event a defective pressure sensor presents a short circuit to the excita-
age Reference tion voltage source, the voltage source goes into current limit during the bad
transducer's time slot.

2.19 Advanced Comm The Kappa has been designed to function in standalone mode or in an Infinity
Option Network. It is not compatible with SIRENET.

The Comm Option PC board supplies power and communications interface


for peripheral devices associated with the monitor.

2.19.1 Comm Option Board The major circuits include a high speed serial link to the Patient Monitor, con-
Hardware trol and status registers to the 68302 processor, and miscellaneous functions.
The serial link functions as a bus master on the local bus. The 68302 per-
forms bus arbitration. The registers and miscellaneous functions are slave Revision 1_ Released_Printed on_02.08.06_F6013052_Kappa_Function_Description.fm

devices on the bus and completely accessible to the 68302.

2.19.2 EEPROMs The Comm Option PC board contains SERIAL EEPROMs which are pro-
grammed in the factory with various configuration parameters. One contains
factory-programmed field service data and is read-only in the field. A second
EEPROM is field-programmable and contains various configuration parame-
ters.
Copyright reserved.

30 6013.052
Schematics and Diagrams

31
32
Kappa Patient Monitor Schematics and Diagrams

1 Schematics and Dia-


grams

Figure 1 Kappa Monitor Front Panel Connectors

Figure 2 Kappa Monitor Rear Panel Connectors


Revision 1_ Released_Printed on_02.08.06_S6013052_Kappa_Schematics_and_Diagrams.fm

1.1 MultiMed Pod Cable


Connector on Kappa

Figure 3 MultiMed Pod Cable Connector Pinouts


Copyright reserved.

Table 1 MultiMed Pod Cable Connector Pinouts

Pin No. Signal


1 RCALIB

6013.052 33
Schematics and Diagrams Kappa Patient Monitor

Pin No. Signal


2 REDK
3 HVPODIDR
4 VEE_1
5 LFT_LEG
6 DET_A
7 DET_K
8 TB
9 TCOM
10 RT_LEG
11 LFT_ARM
12 (RCALRTN)
13 IRK
14 TA
15 VEE_2
16 RT_ARM

1.2 HemoMed Pod Cable


Connector on Kappa

Figure 4 HemoMed Pod Cable Connector

Revision 1_ Released_Printed on_02.08.06_S6013052_Kappa_Schematics_and_Diagrams.fm


Table 2 HemoMed Pod Cable Connector Pinouts

Pin No. Signal


1 -IBP1
2 -IBP2
3 -IBP3
4 -IBP4
5 P1SWITCHES
6 +IBP1
7 +IBP2
Copyright reserved.

8 +IBP3
9 +IBP4
10 TINJ

34 6013.052
Kappa Patient Monitor Schematics and Diagrams

Pin No. Signal


11 CO_SW_RET
12 PRES_RET
13 PRES1_3+
14 PRES2_4+
15 7R37
16 TBLD

1.3 Aux Connector on


Kappa

Figure 5 Auxillary Connector on Kappa

Table 3 Aux Connectors, Pinouts

Pin No. Signal


1 TX+
2
3 TX-
4 RX+
5 Gnd
6 PWR
7 RX-
Revision 1_ Released_Printed on_02.08.06_S6013052_Kappa_Schematics_and_Diagrams.fm

1.4 Kappa Network Con-


nector

Figure 6 RJ 45 Network Connector

Table 4 RJ45 Network Connector

Pin No. Signal


1 TX+
Copyright reserved.

2 TX-
3 RX+

6013.052 35
Schematics and Diagrams Kappa Patient Monitor

Pin No. Signal


6 RX-

1.5 Kappa Analog Out-


put Connector

Figure 7 Analog Output Connectors, Front/Rear

Table 5 Analog Output Connectors Pinouts

Pin No. Signal


1 GND
2 Hd Audio Out
3 Hd Audio Rtn
4 NC
5 Diag TxD
6 Chan2 Rtn
7 Chan2 Analg
8 Hd Audio Det
9 AGND
10 AGND
11

Revision 1_ Released_Printed on_02.08.06_S6013052_Kappa_Schematics_and_Diagrams.fm


12 Chan1 Anlg
13 Chan1 Rtn
14 Diag RxD

1.6 MultiMed 12 Pod


Copyright reserved.

Figure 8 MultiMed 12 Pod

36 6013.052
Kappa Patient Monitor Schematics and Diagrams

Table 6 MultiMed 12 Pod Connector Pinouts

Aux. SPO2 ECG


Pin No. Signal Pin No. Signal Pins 1, 3, 5, 7, 9, 11, 13, 15, 17, 19, 21 = SHGND
1 1 DETA Pin No. IEC-1/IEC/2 Signals Pin.No. IEC-1/IEC-2 Signals
2 2 DETK SH 2 L LA 14 C6 V6
3 3 NC 4 F LL 16 C5 V5
4 4 REDK 6 R RA 18 C4 V4
5 5 RCALRTN 8 C/C1 VC/V1 20 C3 V3
6 6 RCALIB 10 N RL 22 C2 V2
7 7 IRK 12 C+ V+

1.7 Kappa RS-232, Key-


pad Input, Alarm Out
Connector

Figure 9 RS-232, Keypad Input, Alarm Out Connector

Table 7 RS-232, Keypad Input, Alarm Out Connector Pinouts

Pin No. Signal Pin No. Signal


1 GND 8 COM 1 RXD
2 COM 1 TXD 9 ALRM OUT L
3 REM PWR 10 GND
Revision 1_ Released_Printed on_02.08.06_S6013052_Kappa_Schematics_and_Diagrams.fm

4 MC800TL 11 COM2 RXD


5 COM2 TXD 12 SER. D IN
6 SER. D OUT 13 SER. CLK
7 SER. FSL 14 GND

1.8 Remote Alarm Cable


Copyright reserved.

Figure 10 Remote Alarm Cable - Unterminated

6013.052 37
Schematics and Diagrams Kappa Patient Monitor

Table 8 Remote Alarm Cable Connector Pinouts and Wire Color Code

Connector Pin No. Relay Input Wire Color SPDT Relay Output Circuit Status
1 TAN Brown RTN
2 - 8, 10 - 14 NC Green Inactive Open
9 ORANGE White Inactive Closed

1.9 Analog Cable

Figure 11 Analog Cable (Unterminated)

Table 9 Analog Cable Color Code

Connector Pin Wire Color Function Connector Pin Wire Color Function
No. No.
1 TAN NC 8 BROWN NC
2 WHITE NC 9 ORANGE NC
3 BLACK NC 10 GREY NC
4 RED NC 11 VIOLET NC
5 GREEN NC 12 PINK CHAN 1 ANLG
6 YELLOW CHAN 2 RTN 13 LT BLUE CHAN 1 RTN
7 BLUE CHAN 2 ANLG 14 LT GREEN NC

Revision 1_ Released_Printed on_02.08.06_S6013052_Kappa_Schematics_and_Diagrams.fm


Copyright reserved.

38 6013.052
Kappa Patient Monitor Schematics and Diagrams

1.10 Cardiac Output


Intermediate Cable
Wiring Diagram

Figure 12 Cardiac Output Intermediate Cable Wiring Diagram

1.11 Input Connector on


etCO2 Pod
Revision 1_ Released_Printed on_02.08.06_S6013052_Kappa_Schematics_and_Diagrams.fm

Figure 13 Input Connector on etCO2 Pod

Table 10 etCO2 Pod Input Connector Pins

Pin No. Signal


Copyright reserved.

1 SRC-
2 SRC+
3 EE CS

6013.052 39
Schematics and Diagrams Kappa Patient Monitor

Pin No. Signal


4 REF IN
5 CASE HTR
6 DATA IN
7 EE SK
8 HTR RTN
9 EE DOUT
10 +5V
11 EE DIN
12 SPAN SW
13 SRC SHLD
14 ZERO SW
15 CASE THRM
16 DET HTR
17 DET THRM
18 -12V
19 +12V
20 AGND

1.12 HEMO Pod

Revision 1_ Released_Printed on_02.08.06_S6013052_Kappa_Schematics_and_Diagrams.fm


Figure 14 HEMO Pod Connectors

Table 11 HEMO Pod Connector Signals

Pin No. PRESSURE Temp A C.O./Temp B IBP Kappa


FRONT END
INTERFACE
1 +SIG (P1) TEMP 1 TEMP 2 CNTRLH
2 GND GND
3 GND GND GND CNTRLL
Copyright reserved.

4 +EXCIT (P1) GND .7R37 DATH


5 +EXCIT (P2) GND TBLD VDCRTN
6 GND GND TINJ VDC

40 6013.052
Kappa Patient Monitor Schematics and Diagrams

Pin No. PRESSURE Temp A C.O./Temp B IBP Kappa


FRONT END
INTERFACE
7 GND GND GND DATL
8 +SIG (P2)
9 -SIG (P1)
10 GND
11 -EXCIT (P1)
12 GND
13 -EXCIT (P2)
14 GND
15 -SIG (P2)

1.13 Press Adapter Input

Figure 15 PRESS Input Connector

Table 12 Press Adapter Input

Pin No. A, B, C, D Pin No. A, B, C, D


1 +EXCIT 6 +SIGNAL
2 7
Revision 1_ Released_Printed on_02.08.06_S6013052_Kappa_Schematics_and_Diagrams.fm

3 -EXCIT 8 GND
4 Cable In 9
5 -SIGNAL 0
Copyright reserved.

6013.052 41
42
Schematics and Diagrams
Kappa Patient Monitor

6013.052
Copyright reserved.
Revision 1_ Released_Printed on_02.08.06_S6013052_Kappa_Schematics_and_Diagrams.fm
Fault-Cause-Remedy

43
44
Kappa Patient Monitor Fault-Cause-Remedy

1 Introduction Troubleshooting Kappa monitors is either a two-step or three-step process


depending on whether the problem is internal or external to the Monitor.

1. Determine whether the trouble is external, such as a malfunctioning pod,


connector or cable, or internal.
2. Do either a or b as appropriate.
a) If external, replace malfunctioning item.
b) If internal, go on to step 3.
3. Usual remedial action is to replace malfunctioning component or subas-
sembly. See “Spare Parts” for complete listing of replaceable system
components and subassemblies. After replacing an internal subassembly,
functionally verify all monitor functions before returning monitor to clinical
service.

Kappa monitoring system devices consist of the Base Unit, and peripheral
devices including the R50 Recorder, etCO2 Pod, MultiMed or MultiMed 12
Pod, HEMO or HemoMed Pod, and VGA Display.

2 Recommended Use the specified, recommended tools and test equipment, or a known equiv-
Tools and Test alent, when troubleshooting Kappa monitors and peripheral equipment. Sub-
Equipment stitutions are approved only if an equivalent is listed. Use of other test
equipment and/or accessories could result in inconclusive tests or damage to
system components

Table 1 Recommended Tools and Test Equipment

Tools & Test Equipment Description


Digital Multimeter (DMM), 4.5 digit Fluke model 8050A (or equivalent)
Patient ECG/RESP, Temp, IBP simulator DNI Medsim 300B (or equivalent)
Leads: Three-lead grabber set, or IEC color code 1 33 75 230
IEC color code 2 33 75 248
Five-lead grabber set IEC color code 1 33 75 255
IEC color code 2 33 75 263
Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm

Patient SpO2 simulator, or Nellcor PT2500 (or equivalent)


Reusable SpO2 sensor: Durasensor, Adult 45 34 475 EH50U
NBP simulator (calibrated) or DNI CuffLink‚
Mercury manometer with hand bulb, or Baumanometer, 0-300 mmHg
Electronic pressure indicator with handbulb Veri-Cal Pressure Transducer
Tester (or equivalent)
NBP Calibration Assembly 28 77 855
NBP connection hose 3.7 m 12 75 275

3 Power Problems Before troubleshooting power problems, keep in mind that fuses are used to
protect delicate circuits from potentially harmful currents. Replacement of a
Copyright reserved.

fuse may provide only a temporary solution and may not remove the source
of the fault. Never replace a fuse with other than what is specified.

3.1 Power-On Problems No Response When POWER ON/OFF Key Pressed

6013.052 45
Fault-Cause-Remedy Kappa Patient Monitor

Table 2 Power-On Problems

Conditions Possible Cause(s) Troubleshooting and Remedial Action


Monitor plugged into clini- Clinical site power source not 1. Assure that clinical site power source is
cal site power source. AC active, or defective power cable. active, and known-good power cable is
power LED not illuminated. plugged into Kappa power supply.
Defective power supply or power 2. If problem persists, check for +12Vdc across
harness red and black wires at output connector of
power supply. If voltage not OK, go to step 5.
Connector I/O PCB malfunction
3. If voltage OK, replace power cable.
4. If problem persists, go to step 5.
5. If problem persists, contact your Dräger Med-
ical service representative.
Monitor plugged into clini- Power Supply Check for +12Vdc across red and black wires at
cal site power source. AC output connector of power supply. If voltage not
power LED illuminated. Connector I/O PCB malfunction OK, contact your Dräger Medical service repre-
sentative.
Main Processor PCB malfunction

3.2 Power On/Off Piezo


Tone Fails to Sound.
Table 3 Power-off Alarm Malfunction

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


Piezo tone fails to sound Main Processor PCB malfunction Contact your Dräger Medical service representa-
when monitor powered tive.
on, monitor loses power,
or on power-off.

3.3 Power-Up Sequence


Fails to Complete
Properly

Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm


Table 4 Power-up on AC Power Process Malfunction

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


Monitor fails to complete Software program corrupted 1. If power ON LED illuminates but monitor fails
power-up sequence, e.g., to complete power-up sequence, software
powers itself down before Main Processor PCB malfunction may have become corrupted. Try reinstalling
MAIN screen displays, software.
sounds continuous tone, 2. If problem persists, contact your Dräger Medi-
repeatedly resets. cal service representative.
Copyright reserved.

46 6013.052
Kappa Patient Monitor Fault-Cause-Remedy

3.4 Monitor fails to oper-


ate on battery power
Table 5 Power-up on Battery Power Process Malfunction

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


Battery fails to maintain Battery malfunctionBattery charger 1. Check that battery properly connected to
operation of monitor dur- malfunction power harness. If connected, continue.
ing power failure, for 2. If monitor powered from clinical site for ≥3.5
time duration specified in hours but fails to maintain operation for time
User guide. duration specified in User guide when power
removed, replace battery.
3. If problem persists, contact your Dräger Medi-
cal service representative.

4 Rotary Knob Mal- Table 6 Rotary Knob Malfunction


function

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


Rotary knob fails to prop- Rotary knob subassembly mal- Contact your Dräger Medical service representa-
erly select fields, or press- function tive.
ing the knob in fails to
activate a menu or select Connector I/O PCB malfunction
a default.
Main Processor PCB malfunction

5 Fan Malfunction. Table 7 Fan Malfunction

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


Monitor completes normal Fan or temp sensor malfunction 1. Replace Fan subassembly.
power-up sequence but 2. If problem persists, contact your Dräger Medi-
fan fails to start. Connector I/O PCB malfunction cal service representative.
Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm

6 VGA Display Mal- Table 8 VGA Display Malfunction


function.

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


Display powers on, but VGA cable malfunction 1. Replace VGA cable with known-good cable.
waveforms fail to appear 2. If problem persists, replace VGA Display with
on screen Display malfunction known-good Display.

Connector I/O PCB malfunction 3. If problem persists, contact your Dräger Medi-
cal service representative.
Main Processor PCB malfunction
Areas of display missing or Connector I/O PCB malfunction Contact your Dräger Medical service representa-
color contaminated tive.
Copyright reserved.

Graphics Processor on Main Pro-


cessor PCB malfunction

6013.052 47
Fault-Cause-Remedy Kappa Patient Monitor

7 Fixed Key Fails to Table 9 Fixed Key Malfunction


Function

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


A Fixed Key fails to initiate Front Bezel malfunction Contact your Dräger Medical service represen-
change tative.
Connector I/O PCB malfunction

Main Processor PCB malfunction

8 Isolating Cable Mal- In general, the troubleshooting and repair approach for cable malfunctions is
functions to use a known input signal for any given parameter, and then replace a cable
or sensor found to be malfunctioning. Cable malfunctions, including those
associated with connectors on the cables, generally fall into one of three
cagegories -- Open circuits, Short circuits, and Intermittent conditions

Open circuits and short circuits manifest themselves as a loss of signal. Soft-
ware in the Monitor senses the loss, and generates an error message such
as “ECG Leads Off” and “SpO2 Transparent.” Typically, short circuits result in
software resets.

An intermittent condition manifests itself as noise on the signal. The source of


the noise can often be isolated by removing the signal and shorting the
input(s) to the cable. Then flex along the cable, particularly at connectors,
while watching for noise indications on the monitor display.

9 Visible or Audible
Alarm Reporting
Failure.
Table 10 Alarm Malfunctions

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


Audible Alarm O.K., but Visi- Software problem 1. Try reinstalling software.

Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm


ble Alarm Fails. 2. If problem persists, contact your Dräger Medi-
cal service representative.

10 MultiMed POD -
Parameter Signal
Problems
Table 11 Parameter Signal Problems

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


Failure to report lead-off MultiMed POD malfunction 1. Check MultiMed Pod contact for possible con-
condition tamination. Clean if required. If problem per-
Main Processor PCB malfunction sists, replace Pod.
Copyright reserved.

2. If problem persists, contact your Dräger Medi-


cal service representative.

48 6013.052
Kappa Patient Monitor Fault-Cause-Remedy

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


ECG/Resp waveform Incorrect setting of line frequency 1. Access the Service Menu under Monitor
noise for customer site Setup →Biomed →Service in the Main menu.
2. Assure that the setting is proper for the fre-
Poor connection or intermittent quency of the power source at the customer
break site.

MultiMed POD malfunction 3. If problem persists, check cables, connec-


tions, and MultiMed POD for intermittent
Main Processor PCB malfunction breaks.
4. Connect grabber connectors of an ECG lead
set to a common conductor, such as the
shank of a screwdriver blade, to produce a
flatline ECG waveform.
5. Watch for a distinct change in noise level,
indicating the source of the problem, as you
flex each lead and cable (particularly at con-
nectors).
6. Replace a defective ECG lead, cable, or Mul-
tiMed POD
7. If problem persists, contact your Dräger Medi-
cal service representative.
SpO2 parameter box fails Sensor malfunction 1. Check illumination of red LED in sensor.
to appear when sensor 2. If LED not illuminated, replace sensor and
plugged into MultiMed Open cable or connector continue to step 3.
POD, or SpO2 readings
missing MultiMed POD malfunction 3. If LED illuminated, place sensor on your fin-
ger and go on to step 4.
Main Processor PCB malfunction 4. If problem persists, replace intermediate
cable between sensor and MultiMed POD.
5. If problem persists, replace MultiMed POD.
6. If problem persists, contact your Dräger Medi-
cal service representative.
SpO2 waveform noise Poor connection or intermittent 1. With SpO2 sensor on your finger, observe
Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm

break in cable waveform while you flex lead and cable, par-
ticularly at sensor, at connectors, and at Mul-
MultiMed POD malfunction tiMed POD.
2. Watch for distinct change in waveform (dis-
Main Processor PCB malfunction tinctly different from artifact) as you flex along
length of cables as well as at sensor and con-
nectors.
3. Replace defective sensor, cable or MultiMed
POD.
4. If problem persists, contact your Dräger Medi-
cal service representative.
Copyright reserved.

6013.052 49
Fault-Cause-Remedy Kappa Patient Monitor

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


Temp parameter box fails Sensor malfunction 1. Replace sensor.
to appear when sensor 2. If problem persists, replace MultiMed POD.
plugged into MultiMed Open cable or connector
POD, Temp readings 3. If problem persists, contact your Dräger Medi-
missing or inaccurate POD malfunction cal service representative.

Main Processor Board malfunc-


tion

11 NBP

11.1 NBP Error Messages


NBP Open Line Results if the NBP measurement circuit does not sense pressure changes
after initiating inflation. Possible causes include an open line out to the
patient, an open or occluded line on the pressure measurement line running
between the NBP assembly and the pressure transducer on the main proces-
sor board, a leaky cuff or cuff connector, or a defective valve on the NBP
pump assembly.

NBP Blocked Line Result of the NBP measurement circuit detecting an occlusion on the line to
the patient, or a neonatal cuff with monitor in adult mode.

NBP Overpressure Result attributable to hardware or software detecting overpressure.

NBP Cuff Deflation Error Result of the two minute NBP measurement timer expiring. It is typically trig-
gered when an NBP measurement had been taken prior to placing the unit in
calibration mode. (When calibrating, power-cycle monitor and then don’t run
pump until after calibration.)

NBP Artifact Result of erratic pressure values being sensed and could be related to an
application problem or could caused by an intermittent connection to the
sense line.

11.2 NBP Troubleshoot- Before attempting to troubleshoot NBP malfunctions, do the following:
ing
Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm
1. Set up NBP Calibration assembly.
2. Perform leakage test on fixed volume and hose, and on system.
3. Check calibration.

NOTE
Refer to procedures in Noniinvasive Blood Pressure Section in Functional
Verification for procedures to set up NBP Calibration assembly, perform
leakage test, and check NBP calibration.

Table 12 NBP Malfunctions

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


Copyright reserved.

NBP fails to zero prop- NBP pneumatic system malfunc- Contact your Dräger Medical service representa-
erly, or fails calibration tion tive.
check
Main Processor PCB malfunction

50 6013.052
Kappa Patient Monitor Fault-Cause-Remedy

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


NBP pump fails to Front Bezel malfunction 1. If monitor reporting NBP in fault mode, or
start/stop when NBP key error message displays, power-cycle monitor.
on front panel is pressed Connector I/O PCB malfunction 2. If problem persists, contact your Dräger Medi-
cal service representative.
NBP pump subassembly mal-
function

Main Processor PCB malfunction


NBP pump starts, but cuff Cuff assembly malfunction 1. Recheck cuff assembly and installation, and
fails to inflate/deflate prop- replace cuff assembly if defective.
erly NBP pneumatic system malfunc- 2. If problem persists, contact your Dräger Medi-
tion cal service representative.
NBP parameter box fails MAIN screen display mode set 1. Assure MAIN screen display mode set to
to appear when cuff hose for Manual automatic.
plugged into NBP module 2. If problem persists, contact your Dräger Medi-
Cuff sensor defective or not cal service representative.
plugged into Connector I/O PC
Board.

Main Processor PCB malfunction

12 etCO2 Malfunction.
Table 13 etCO2 Malfunctions

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


Parameter box fails to Sensor or cable malfunction 1. Be sure etCO2 cable plugged into AUX con-
appear when sensor nector on front of Kappa. If problem persists,
plugged into pod etCO2 pod malfunction continue.
2. Clean calibration windows.
Main Processor PCB malfunction
3. Replace etCO2 Sensor.
4. If problem persists, replace etCO2 pod.
Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm

5. If problem persists, contact your Dräger Medi-


cal service representative.
Sensor fails calibration Sensor or cable malfunction 1. Replace etCO2 Sensor.
2. If problem persists, replace etCO2 pod.
etCO2 pod malfunction
3. If problem persists, contact your Dräger Medi-
Main Processor PCB malfunction cal service representative.

Persistent Adapter Fail- Airway adapter or sensor window 1. If adapter or sensor window occluded, clean
ure message occluded window.
2. If problem persists, replace airway adapter.
Airway adapter malfunction
3. If problem persists, replace sensor.
Sensor malfunction 4. If problem persists, replace etCO2 pod.
Copyright reserved.

6013.052 51
Fault-Cause-Remedy Kappa Patient Monitor

13 HEMO2/4 Pod /
HemoMed Pod

13.1 Readings Missing or Disruption in communications to an external pod is caused by the pod itself,
Inaccurate by a cable problem, or by a communication problem on the main processor
board. Power for an external pod is supplied from the main processor board
when the board senses that a load is present. If the monitor is not reporting
connection of a Hemo2/4 pod, 12-lead pod or tcpO2/CO2 pod, use the follow-
ing test to determine first whether or not the pod is being detected.

NOTE
Only the Hemo2/4 pod “communicates” with the monitor. The Hemo2/4
Pod is a digital device; the HemoMed Pod is analog.

Connect a 150W ¼W resistor between pins 1 and 3 on the suspect port, and
check monitor display for a pod connection error “DEVICE FAILURE ON
HEMO CONNECTOR x.”
• If no error message displays, the problem is on the main processor
board.
• If error message displays, problem may be in cable, pod, or main proces-
sor board. Replace defective cable or pod, as required. If problem per-
sists, contact your Dräger Medical service representative.
• If only one input or output on a HEMO2/4 or HemoMed Pod fails to func-
tion, try replacing the sensor or cable. If problem persists, replace the
Pod.

14 IBP Malfunctions. Table 14 IBP Malfunctions

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


Zero All key, C.O.Start key Pod malfunction 1. Try replacing Pod / monitor cable.
or Wedge key on Pod fails 2. If problem persists, replace Pod.
to initiate function Cable malfunction
3. If problem persists, contact your Dräger Medi-

Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm


Main Processor PCB malfunction cal service representative. Copyright reserved.

52 6013.052
Kappa Patient Monitor Fault-Cause-Remedy

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


IBP fails to zero properly Cable malfunction 1. Unplug all patient parameter inputs to the
or fails calibration check monitor.
Pod malfunction 2. Set Patient simulator for an IBP static pres-
sure = 0 mmHg, and plug simulator into Pod.
Main Processor PCB malfunction
3. Plug Pod output cable into monitor.
4. Check that “Zero Required” appears in mes-
sage field and that IBP parameter box
appears on display.
5. If either message or parameter box fails to
appear, replace Pod and cable with known-
good Pod and cable.
6. If problem persists, contact your Dräger Medi-
cal service representative. Otherwise, con-
Aux/Hemo2S or tinue. Select IBP parameter field on MAIN
Aux/Hemo3 screen, and assure that Cal Factor is set to
100.
7. Select Zero in menu, and press in on rotary
knob.If “Zero Accepted” appears in message
field, continue.
8. If “--- Did Not Zero” appears in message field,
contact your Dräger Medical service repre-
sentative.
9. Increase simulator pressure to 100 mmHg.
10. If monitor reading = 100 ±1 mmHg, return
monitor to clinical service.
11. If monitor reading ≠ 100 ±1 mmHg, contact
your Dräger Medical service representative
IBP or Temp (HEMO2/4 Defective sensor or cable 1. Replace sensor and cable with known-good
Pod only) parameter box sensor and cable. If problem persists, con-
fails to appear when sen- Pod malfunction tinue.
sor plugged into Pod 2.
Pod / Monitor cable malfunction
Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm

• With HEMO2/4 connected to monitor,


Main Processor PCB malfunction assure that parameter labels are visible in
LED windows on HEMO Pod.
• If HemoMed Pod, go on to step 4.
3.
• If labels are visible, replace sensor. If
problem persists, go on to step 4.
• If labels NOT visible, try replacing HEMO
POD / Monitor interconnecting cable. If
problem persists, continue.
4. Replace Pod.
5. If problem persists, contact your Dräger Medi-
cal service representative.
Copyright reserved.

6013.052 53
Fault-Cause-Remedy Kappa Patient Monitor

15 No Printout from Table 15 Recorder Problems


Recorder.

Symptom(s) Possible Cause(s) Troubleshooting and Remedial Action


Recorder connected Recorder malfunction 1. Detach Interface Plate from Recorder and
directly to Monitor check voltage between pins 1 and 2 on Inter-
Cabling malfunction face Plate docking connector. See illustration
Power LED on Recorder below.
NOT illuminated Interface Plate (if installed) mal- 2. If voltage OK, replace Recorder.
function
3. If voltage not OK, check for +12VDC between
Connector I/O PCB malfunction pins 2 and 15 of X13 on monitor.
• If voltage OK replace cable.
Main Processor PCB malfunction • If voltage not OK contact your Dräger
Medical service representative.
4. If problem persists, contact your Dräger Medi-
cal service representative.
Recorder connected Recorder malfunction 1. With an ECG waveform on Monitor display,
directly to Monitor provided by patient simulator, press Record
Interconnecting cable or connec- key.
Recorder power LED illu- tion malfunction 2.
minated
Connector I/O PCB malfunction • If “Recorder Not Connected” message
appears in the message field, go to step
Main Processor PCB malfunction 3.
• Press Zoom key and select Event Recall.
After ≈ 20 sec, BED TIMED strip should
appear on display. If BED TIMED strip
fails to appear, contact your Dräger Medi-
cal service representative. Otherwise,
continue.
3. If problem persists, and Recorder Cable
4318130 is installed, replace Recorder cable
and go to step 5.
4. If problem persists, replace Recorder Inter-

Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm


face Plate.
5. If problem persists, replace Recorder.
6. If problem persists, contact your Dräger Medi-
cal service representative.

16 Patient-Related Data Contact your Dräger Medical service representative.


Not Retained, or
Monitor Fails to
Compute Trends
Copyright reserved.

17 Software Loading For problems concerning software downloads, refer to software installation
Problems instructions for version of software you are attempting to install.

54 6013.052
Kappa Patient Monitor Fault-Cause-Remedy

18 Difficulty acquiring This problem is usually caused by an incorrect connection or communications


export protocol data protocol. Refer to Infinity RS-232 Export Protocol Reference Booklet, 64 93
212.

19 Total or partial loss Temporary network communication losses are caused by either an inter-rup-
of network commu- tion with the Ethernet port on the main processor or in the network line.
nications
1. Verify that network configuration data in monitor are correct.
2. Contact your Dräger Medical service representative, if problem isolated to
monitor.

20 Problem Report SC Series Patient Monitoring

Enter all applicable data in the spaces provided, and include a copy of this
form when faxing a request for technical assistance.

Name of contact

Telephone

Fax

Email Address (If available)

Monitoring Site: Country:

Region / State / Province:

Hospital or Clinical Site:

Device Type:

Device Serial Number:

Device Operating Software:


Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm

Care Unit Type:

Parameters being monitored at time of fault:

Network / Stand-alone Use

Brief Description of Fault:

_____________________________________________________________

_____________________________________________________________

_____________________________________________________________

_____________________________________________________________
Copyright reserved.

Can the problem be reproduced or is the problem intermittent?

Has TSS been previously contacted concerning this problem?

6013.052 55
Fault-Cause-Remedy Kappa Patient Monitor

Has a customer complaint on this product been filed?

Complaint Reference Number (If applicable)

Fax inquiry to: Draeger Medical Systems,


Danvers, MA 01923 U.S.A. 978 750-7655

21 Diagnostic/Error
Messages

21.1 Overview of Diag- The diagnostic codes given in the following Tables may appear in a Kappa
nostic Messages Diagnostic Log, and help in troubleshooting a malfunctioning Monitor. Both
possible cause and suggested remedial action are listed for the field-signifi-
cant codes.

Only codes for field-related procedures are given in the Table. There are also
a number of other diagnostic codes, not listed in the Table, that have signifi-
cance for engineering in helping to improve product performance. If
observed, report these codes to the factory. Download the Diagnostic Log to
a DataCard.

To access the Diagnostic Log, on the Main Menu select BIOMED, then DIAG.
LOG.

Diagnostic messages caused by an error condition can be generated by four


sources:
• Replaceable hardware malfunction, such as an etCO2 or HEMO Pod
• Software error
• Intermittent condition
• Hardware error

Notes for the following Tables --


• When action says replace..., do so only if problem can be reproduced.

Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm


• Convention of A, B, C etc. in the messages given in the Description col-
umn is an abbreviation of actual messages, which will have numbers in
place of the A, B, C, etc., that help engineers uncover the actual cause.
• When action says “upgrade to Vxx”, problem can also be solved by
upgrading to any SW version released after Vxx.
Copyright reserved.

56 6013.052
Kappa Patient Monitor Fault-Cause-Remedy

Table 16 Diagnostic Log Codes

Code Description Cause Action

xxx = any
alphanumeric
characters
80007xxx Try upgrading SW to currently released version.

80008xxx If error persists, contact your Dräger Medical


service representative.
80009xxx

8000axxx
8000bxxx Software did not Try disabling items connected or disabling Auto
allocate enough Dual View.
8000cxxx memory for user’s
current setup. If error persists, contact your Dräger Medical
service representative.
800c9009 atten OOR = A cal fac- Speaker too quiet Replace speaker
tor = B
800c9022 cal factor=A Speaker broken Replace speaker
max_sample_value=B
800cd007 Loopback Active = A Hardware malfunc- Check “Nurse Call” circuit.
Expect Active =B tion in “nurse call”
circuit.
800d3xxx Cardiac Output- Try disconnecting/reconnecting cable.
related
800d4xxx If problem persists, swap cable with known-good
cable.
800d5xxx
If problem persists try swapping pod.
800d6xxx
If problem persists, contact your Dräger Medical
Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm

service representative.
800d7xxx Temperature - related Try disconnecting/reconnecting cable.
(MultiMed Pod)
800d8xxx If problem persists, swap cable with known-good
cable.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.
800d9xxx Pod communica- Try disconnecting/reconnecting cable.
tions problem.
800daxxx If problem persists, swap cable with known-good
Pod cable, Pod, or cable.
CPU board failed
Copyright reserved.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.

6013.052 57
Fault-Cause-Remedy Kappa Patient Monitor

Code Description Cause Action

xxx = any
alphanumeric
characters
800dbxxx etCO2 Pod - related Try swapping out airway adaptor with known-
good components.
800dcxxx
Try disconnecting/reconnecting cable.

If problem persists, swap cable with known-good


cable.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.
800ddxxx HEMOPOD / Try disconnecting/reconnecting cable.
HEMOMED - related
800dexxx If problem persists, swap cable with known-good
cable.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.
800dfxxx Pod communica- Try disconnecting/reconnecting cable.
tions problem.
800e0xxx If problem persists, swap cable with known-good
Pod cable, Pod, or cable.
CPU board failed
If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.

Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm


800e1xxx NIBP - related Try disconnecting/reconnecting pneumatic tub-
ing.
800e2xxx
If problem persists, contact your Dräger Medical
service representative.
800e5xxx RESP-related Try disconnecting/reconnecting cable.

800e6xxx If problem persists, swap cable with known-good


cable.

If problem persists try swapping pod.

If problem persists, replace Main Processor


Board Subassembly
Copyright reserved.

58 6013.052
Kappa Patient Monitor Fault-Cause-Remedy

Code Description Cause Action

xxx = any
alphanumeric
characters
800e7xxx SPO2 - related Try disconnecting/reconnecting cable.

If problem persists, try replacing sensor.

If problem persists, swap cable with known-good


cable.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.
800e9xxx Temperature - related Try disconnecting/reconnecting cable.
(MultiMed Pod)
800eaxxx If problem persists, swap cable with known-good
cable.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.
800f1xxx Temperature - related Try disconnecting/reconnecting cable.
(MultiMed Pod)
800f2xxx If problem persists, swap cable with known-good
cable.

If problem persists try swapping pod.

If problem persists, replace Main Processor


Board Subassembly
800f5xxx HEMOPOD / Try disconnecting/reconnecting cable.
Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm

HEMOMED - related
800f6xxx If problem persists, swap cable with known-good
cable.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.
800fbxxx NIBP - related Try disconnecting/reconnecting pneumatic tub-
ing.
800fcxxx
If problem persists, contact your Dräger Medical
service representative.
80101xxx Analog Out Contact your Dräger Medical service representa-
Copyright reserved.

tive.
80102xxx

6013.052 59
Fault-Cause-Remedy Kappa Patient Monitor

Code Description Cause Action

xxx = any
alphanumeric
characters
80107xxx RESP-related Try disconnecting/reconnecting cable.

80108xxx If problem persists, swap cable with known-good


cable.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.
8010dxxx SPO2 - related Try disconnecting/reconnecting cable.

8010exxx If problem persists, try replacing sensor.

If problem persists, swap cable with known-good


cable.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.
80117xxx Serial Interface to Try disconnecting/reconnecting keypad.
Remote Keypad -
80118xxx related If problem persists, swap cable with known-good
keypad.

If problem persists, contact your Dräger Medical


service representative.
80119xxx Temperature - related Try disconnecting/reconnecting cable.
(MultiMed Pod)
8011axxx If problem persists, swap cable with known-good

Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm


cable.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.
80127xxx bkg_operational deter- Background entity Try upgrading SW to currently released version.
mined MONITORING determined that the
80128xxx SW is corrupt. Main Software was If error persists, contact your Dräger Medical
corrupted (BAD service representative.
CRC)
8012901e boot_validate_downlo Software memory Re-download monitor software. If problem per-
ad_card could not card may have sists, replace memory card.
open/close one of the been removed dur-
Copyright reserved.

boot files. ing a download or


the card may have
become corrupted

60 6013.052
Kappa Patient Monitor Fault-Cause-Remedy

Code Description Cause Action

xxx = any
alphanumeric
characters
8014bxxx RECORDINGS - Try disconnecting/reconnecting cable.
related
8014cxxx If problem persists, swap cable with known-good
cable.

If problem persists try swapping Recorder.

If problem persists, contact your Dräger Medical


service representative.
8016a00a Image checksum test Bad mem- Restart monitor. If error persists replace moni-
failed due to TAXI ory/access/fetch tor.
error!
80179xxx Pod communica- Try disconnecting/reconnecting cable.
tions problem.
8017axxx If problem persists, swap cable with known-good
Pod cable, Pod, or cable.
CPU board failed
If problem persists try swapping pod.

If problem persists, replace Main Processor


Board Subassembly
8017fxxx Temperature -related Try disconnecting/reconnecting cable.
(MultiMed Pod)
80180xxx If problem persists, swap cable with known-good
cable.

If problem persists try swapping pod.

If problem persists, replace Main Processor


Board Subassembly
Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm

80187020 Error Log was cleared! Manual clear of None


error log via system
console menu
8018bxxx etCO2 Pod - related Try swapping out airway adaptor with known-
good components.
8018cxxx
Try disconnecting/reconnecting cable.

If problem persists, swap cable with known-good


cable.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


Copyright reserved.

service representative.
8018d005 sn A, status B etCO2 Hardware Replace etCO2 Pod.
Fail

6013.052 61
Fault-Cause-Remedy Kappa Patient Monitor

Code Description Cause Action

xxx = any
alphanumeric
characters
8018fxxx Try upgrading SW to currently released version.

80190xxx If error persists, contact your Dräger Medical


service representative.

80196006 Chip timed out at: A Flash memory Re-download software. If problem persists then
wrote: B, read: C failed tests when replace Main Processor Board Subassembly .
doing a download.
Bit one is always
on.
8019601d Flash program write Monitor flash failed Re-download software. If problem persists, con-
verify failed at A. tests when doing a tact your Dräger Medical service representative.
wrote: B, read: C download
8019700c time=A Heart blip not Ignore
showing; most
likely due to rapid
screen re-draws
80199xxx PCMCIA Interface - Retry download. If error persists, return card.
related
8019axxx
801ac00b Pod Type A Conn B, Pod would not Check/replace in this order pod/cartrige, cable,
S/N=C, event D state power on or off. Main Processor Board Subassembly.
E Connector number
in description string
are: 1,2,3-hemo
pod 4-reserved 5-
etCO2
801ac00c Pod Type A, Conn B, Comm error, CRC If problem persists, check connector or replace
S/N=C, event D state error pod

Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm


E
801adxxx Try upgrading SW to currently released version.

801aexxx If error persists, contact your Dräger Medical


service representative.

801b7000 ERROR: load_34010 34010 Failure Contact your Dräger Medical service representa-
tive.
801b7001 ERROR: load_34010 34010 Failure Contact your Dräger Medical service representa-
tive.
Copyright reserved.

62 6013.052
Kappa Patient Monitor Fault-Cause-Remedy

Code Description Cause Action

xxx = any
alphanumeric
characters
801c9xxx RECORDINGS - Try disconnecting/reconnecting cable.
related
801caxxx If problem persists, swap cable with known-good
cable.

If problem persists try swapping Recorder.

If problem persists, contact your Dräger Medical


service representative.
801e1xxx HEMOPOD / Try disconnecting/reconnecting cable.
HEMOMED - related
801e2xxx If problem persists, swap cable with known-good
cable.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.
801e7xxx Try upgrading SW to currently released version.

801e8xxx If error persists, contact your Dräger Medical


service representative.

8020fxxx Software did not Try disabling items connected or disabling Auto
allocate enough Dual View.
80210xxx memory for user’s
current setup. If error persists, try upgrading SW to currently
released version.

If error persists, contact your Dräger Medical


Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm

service representative.
80231005 LCOM_BAD_SWITCH Comm error This indicates a SW error and should be
_COUNT_ERROR reported to Customer service via DataCard.
8023100a LCOM_CHECKSUM_ Comm error This indicates a SW error and should be
DIAG_ERROR reported to Customer service via DataCard.
8023102b LCOM_RAM_DIAG_E Net comm failure This indicates a SW error and should be
RROR reported to Customer service via DataCard.
8023102c LCOM_ROM_DIAG_E Net comm failure This indicates a SW error and should be
RROR reported to Customer service via DataCard.
8023102e LCOM_SHRAM_CHE Net comm failure This indicates a SW error and should be
CKSUM_ERROR reported to Customer service via DataCard.
80231030 LCOM_SHRAM_DIA Net comm failure This indicates a SW error and should be
Copyright reserved.

G_ERROR reported to Customer service via DataCard.


80231031 LCOM_SHRAM_PAC Net comm failure This indicates a SW error and should be
KET_ERROR reported to Customer service via DataCard.

6013.052 63
Fault-Cause-Remedy Kappa Patient Monitor

Code Description Cause Action

xxx = any
alphanumeric
characters
80231032 LCOM_SHRAM_PRL net comm failure This indicates a SW error and should be
_ERROR reported to Customer service via DataCard.
8023dxxx PCMCIA Interface - Retry download. If error persists, return card.
related
8023exxx

80249xxx

8024axxx
80255xxx Try upgrading SW to currently released version.

80256xxx If error persists, contact your Dräger Medical


service representative.

8025bxxx RS-232 Output - Ignore.


related
8025cxxx If problem persists, contact your Dräger Medical
service representative.
80263xxx

80264xxx
80264001 Invalid status after Recorder ASIC Power cycle unit. If error persists, contact your
RX_RESET! failed hardware test Dräger Medical service representative.
rx_data=A
80264002 TX Empty should be Recorder ASIC Power cycle unit. If error persists, contact your
set after TX_RESET! failed cold start test Dräger Medical service representative.
tx_data=A
8026dxxx Network Communica- Try disconnecting/reconnecting cable between
tions - related Monitor and network wallbox.

Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm


8026exxx
Verify proper SW version(s) installed in all
related devices, in accordance with SW compat-
ibility chart for Monitor SW.

If problem persists, contact your Dräger Medical


service representative.
80271xxx Network Communica- Try disconnecting/reconnecting cable between
tions - related Monitor and network wallbox.
80272xxx
Verify proper SW version(s) installed in all
related devices, in accordance with SW compat-
ibility chart for Monitor SW.
Copyright reserved.

If problem persists, contact your Dräger Medical


service representative.

64 6013.052
Kappa Patient Monitor Fault-Cause-Remedy

Code Description Cause Action

xxx = any
alphanumeric
characters
80273xxx NIBP - related Try disconnecting/reconnecting pneumatic tub-
ing.
80274xxx
If problem persists, contact your Dräger Medical
service representative.
8027dxxx Network Communica- Try disconnecting/reconnecting cable between
tions - related Monitor and network wallbox.
8027exxx
Verify proper SW version(s) installed in all
related devices, in accordance with SW compat-
ibility chart for Monitor SW.

If problem persists, contact your Dräger Medical


service representative.
8027fxxx Software did not Try disabling items connected or disabling Auto
allocate enough Dual View.
80280xxx memory for user’s
current setup. If error persists, contact your Dräger Medical
service representative.
80281xxx HEMOPOD / Try disconnecting/reconnecting cable.
HEMOMED - related
80282xxx If problem persists, swap cable with known-good
cable.
80283xxx
If problem persists try swapping pod.
80284xxx
If problem persists, contact your Dräger Medical
service representative.
80289xxx Pod communica- Try disconnecting/reconnecting cable.
tions problem.
Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm

8028axxx If problem persists, swap cable with known-good


Pod cable, Pod, or cable.
80291xxx CPU board failed
If problem persists try swapping pod.
80292xxx
If problem persists, contact your Dräger Medical
service representative.
8029fxxx Power -related Replace Battery

802a0xxx If problem persists, contact your Dräger Medical


service representative.
802a1xxx

802a2xxx
Copyright reserved.

6013.052 65
Fault-Cause-Remedy Kappa Patient Monitor

Code Description Cause Action

xxx = any
alphanumeric
characters
802a3xxx QRS Sync pulse - Try disconnecting/reconnecting cable.
related
802a4xxx If problem persists, swap cable with known-good
cable.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.
802abxxx RECORDINGS - Try disconnecting/reconnecting cable.
related
802acxxx If problem persists, swap cable with known-good
cable.
802adxxx
If problem persists try swapping Recorder.
802aexxx
If problem persists, contact your Dräger Medical
service representative.
802b9xxx RESP-related Try disconnecting/reconnecting cable.

802baxxx If problem persists, swap cable with known-good


cable.

If problem persists try swapping pod.

If problem persists, replace Main Processor


Board Subassembly
802bbxxx Try upgrading SW to currently released version.

802bcxxx If error persists, contact your Dräger Medical

Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm


service representative.

802bf008 memory partition 2 is Memory utilization Ignore


91 percent used. note
802d4002 shutdown not due to Monitor was shut If runing monitor on internal battery, ignore this
user. Could be low down without using message. Otherwise report condition to cus-
battery. the user’s front tomer service
panel switch. Usu-
ally this means that
the monitor was
running on internal
batteries and ran
out of power.
Copyright reserved.

66 6013.052
Kappa Patient Monitor Fault-Cause-Remedy

Code Description Cause Action

xxx = any
alphanumeric
characters
802dfxxx SPO2 - related Try disconnecting/reconnecting cable.

802e0xxx If problem persists, try replacing sensor.

If problem persists, swap cable with known-good


cable.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.
802e701c Powering up the Sys- Power was turned Ignore
tem... on.
802e701d Indicates that new No action nessessary
software was
installed
802ef011 trends completely Software warning Ignore
cleared, new sofware
loaded
802f1xxx Network Communica- Try disconnecting/reconnecting cable between
tions - related Monitor and network wallbox.
802f2xxx
Verify proper SW version(s) installed in all
related devices, in accordance with SW compat-
ibility chart for Monitor SW.

If problem persists, contact your Dräger Medical


service representative.
8030500d b A a=B w=C d=D 0 SW error Contact your Dräger Medical service representa-
Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm

e=E 0 s=F rx=G ok=H tive.


flags=I df=J,K
80309xxx Try upgrading SW to currently released version.

8030axxx If error persists, contact your Dräger Medical


service representative.

80311xxx Network Communica- Try disconnecting/reconnecting cable between


tions - related Monitor and network wallbox.
80312xxx
Verify proper SW version(s) installed in all
related devices, in accordance with SW compat-
ibility chart for Monitor SW.
Copyright reserved.

If problem persists, contact your Dräger Medical


service representative.

6013.052 67
Fault-Cause-Remedy Kappa Patient Monitor

Code Description Cause Action

xxx = any
alphanumeric
characters
80325001 Speaker Out of Range Failed speaker, If sound is heard replace Connector I/O Board.
value: A CPU board or front Otherwise, replace Main Processor Board Sub-
panel board assembly
80359xxx Network Communica- Try disconnecting/reconnecting cable between
tions - related Monitor and network wallbox.
8035axxx
Verify proper SW version(s) installed in all
related devices, in accordance with SW compat-
ibility chart for Monitor SW.

If problem persists, contact your Dräger Medical


service representative.
8035bxxx Serial Interface to Try disconnecting/reconnecting keypad.
Remote Keypad -
8035cxxx related If problem persists, swap cable with known-good
keypad.

If problem persists, contact your Dräger Medical


service representative.
8035dxxx Network Communica- Try disconnecting/reconnecting cable between
tions - related Monitor and network wallbox.
8035exxx
Verify proper SW version(s) installed in all
8036fxxx related devices, in accordance with SW compat-
ibility chart for Monitor SW.
80370xxx
If problem persists, contact your Dräger Medical
service representative.
80371xxx MGM - related Verify proper SW version(s) installed in all
related devices, in accordance with SW compat-

Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm


80372xxx ibility chart for Monitor SW.

If problem persists, try disconnecting/reconnect-


ing cable(s) between Monitor and MGM.

If problem persists,try replacing cable(s)


between Monitor and MGM.

If problem persists, try swapping MGM with


known-good device.

If problem persists try replacing MIB 1&2 Option.

If problem persists, contact your Dräger Medical


service representative.
Copyright reserved.

68 6013.052
Kappa Patient Monitor Fault-Cause-Remedy

Code Description Cause Action

xxx = any
alphanumeric
characters
80373xxx MIB - related Verify proper SW version(s) installed in all
related devices, in accordance with SW compat-
80374xxx ibility chart for Monitor SW.

If problem persists, try disconnecting/reconnect-


ing cable(s) between Monitor and MIB converter.

If problem persists,try replacing cable(s)


between Monitor and MIB converter.

If problem persists, try swapping MIB converter


with known-good device.

If problem persists, try replacing MIB 1&2


Option.

If problem persists, contact your Dräger Medical


service representative.
80375xxx ISD (SDC) - related Verify proper SW version(s) installed in all
related devices, in accordance with SW compat-
80376xxx ibility chart for Monitor SW.

If problem persists, try disconnecting/reconnect-


ing cable(s) between Monitor and ISD (SDC).

If problem persists,try replacing cable(s)


between Monitor and ISD (SDC).

If problem persists, try swapping ISD (SDC) with


known-good device.
Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm

If problem persists, contact your Dräger Medical


service representative.
803a1xxx 12 LEAD POD - Try disconnecting/reconnecting cable.
related
803a2xxx If problem persists, swap cable with known-good
cable.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.
803a3xxx Serial Interface to Try disconnecting/reconnecting keypad.
Remote Keypad -
803a4xxx related If problem persists, swap cable with known-good
Copyright reserved.

keypad.

If problem persists, contact your Dräger Medical


service representative.

6013.052 69
Fault-Cause-Remedy Kappa Patient Monitor

Code Description Cause Action

xxx = any
alphanumeric
characters
803e5xxx SPO2 - related Try disconnecting/reconnecting cable.

803e6xxx If problem persists, try replacing sensor.

803e7xxx If problem persists, swap cable with known-good


cable.
803e8xxx
If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.
80361xxx 12 LEAD POD - Try disconnecting/reconnecting cable.
related
80362xxx If problem persists, swap cable with known-good
cable.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical


service representative.
803f1xxx etCO2 Pod - related Try swapping out airway adaptor with known-
good components.
803f2xxx
Try disconnecting/reconnecting cable.
803f3xxx
If problem persists, swap cable with known-good
803f4xxx cable.

If problem persists try swapping pod.

If problem persists, contact your Dräger Medical

Revision 1_ Released_Printed on_02.08.06_R6013052_Kappa_IEC_Fault_Cause_Remedy.fm


service representative.
ffffffff boot_program_images Flash program- Retry download.
failed with status: A on ming error when
image: B performing a down- If problem persists, contact your Dräger Medical
load service representative.
Copyright reserved.

70 6013.052
Annex

Spare Parts

Test List
Kappa Test List
Kappa Patient Monitor
Monitor System

Revision 1.0
6013.052

Because you care


Contents

1 Introduction 1

2 Recommended Tools and Test Equipment 1

3 Power Circuits and Start-up 3

3.1 Power ON/OFF ....................................................................................................................... 3

3.2 Power-Up Sequence .............................................................................................................. 3

4 Rotary Knob 3

5 Fixed Keys 3

5.1 ON/OFF Key ........................................................................................................................... 3

5.2 Main Screen Key .................................................................................................................... 4

5.3 Alarm Silence Key .................................................................................................................. 4

5.4 Alarm Limits Key .................................................................................................................... 4

5.5 All Alarms Off Key .................................................................................................................. 4

5.6 Code Key ................................................................................................................................ 4

5.7 Record Key ............................................................................................................................. 4

5.8 Print Screen Key .................................................................................................................... 5

5.9 NBP Start/Stop Key ................................................................................................................ 5

5.10 Zoom Key ............................................................................................................................... 5

5.11 Help Key ................................................................................................................................. 5

5.12 Mark Key ................................................................................................................................ 5

6 ECG/RESP Functions 5

6.1 Test Setup .............................................................................................................................. 5

6.2 Waveforms/Digital Readouts/Tones ....................................................................................... 6

6.3 Pacer Detection ...................................................................................................................... 6

6.4 Lead-Off Indicators ................................................................................................................. 6


Kappa_IEC_Test_ListIVZ.fm 02.08.06

6.5 Alarm Function ....................................................................................................................... 6


Copyright reserved.

6.6 Asystole .................................................................................................................................. 7

I
Contents

7 SpO2 Function 7

7.1 SpO2 Test Setup .................................................................................................................... 7

7.2 Waveforms/Digital Readouts/Tones ....................................................................................... 7

7.3 Pulse Tone Generator ............................................................................................................ 7

7.4 SpO2 Limits Alarms ................................................................................................................ 8

8 Temperature Function 8

8.1 Temperature Test Setup ......................................................................................................... 8

8.2 Digital Readout ....................................................................................................................... 8

8.3 Temperature Calibration Check .............................................................................................. 8

9 Non-Invasive Blood Pressure Function 9

9.1 System Setup and Pneumatics Leakage Test ........................................................................ 9

9.2 Calibration Check ................................................................................................................. 10

9.3 Hardware Overpressure ....................................................................................................... 10

9.4 Pump .................................................................................................................................... 10

9.5 Interval Mode ........................................................................................................................ 10

9.6 Safety Timer ......................................................................................................................... 11

10 etCO2 Function 11

11 HemoMed Pod 12

11.1 IBP Function ......................................................................................................................... 12

11.2 Cardiac Output Function ...................................................................................................... 13

12 HEMO POD 2/4 14

12.1 IBP Function ......................................................................................................................... 14

12.2 Memory Backup .................................................................................................................. 16

12.3 MIB and CAN Options (if installed) ....................................................................................... 16


Kappa_IEC_Test_ListIVZ.fm 02.08.06

12.4 Adv Comm Option (if installed) ............................................................................................. 16


Copyright reserved.

12.5 Configuration ........................................................................................................................ 16

12.6 Recorder Function ................................................................................................................ 17

II
Contents

13 Leakage Tests 17

13.1 Resistance Test .................................................................................................................... 17

13.2 Leakage Current Tests ......................................................................................................... 18

Clinical Site Report


Kappa_IEC_Test_ListIVZ.fm 02.08.06

Copyright reserved.

III
IV
Contents

Copyright reserved.
Kappa_IEC_Test_ListIVZ.fm 02.08.06
Kappa Patient Monitor Test List

Functional Verification
Tests and Calibration
Checks

1 Introduction Calibration required for proper operation of Kappa Patient Monitors has been
programmed into the software for the monitor. Specific calibration and adjust-
ment procedures needed in routine operations of the monitor are detailed in
the User Guide for the installed software version, and are required to be per-
formed only when directed to do so by a message in the message field on the
monitor display.

If specific reading of a patient parameter is suspect, Dräger Medical recom-


mends functional verification for that function. Dräger Medical recommends a
full functional verification for all patient parameters, including an NBP calibra-
tion check, a pneumatic leakage check, a Temperature calibration check and
a leakage current check be performed in accordance with local regulations or
at least annually.

NOTE
Calibration checks for NBP and Temperature should be performed by only
qualified personnel with the necessary equipment. Regulations in some
national jurisdictions may place specific requirements on who may perform
these functions.

The following describes functional verification and calibration check proce-


dures for monitors in the field. The procedures assure proper operation of the
monitor, using industry-standard patient simulators and test equipment, to
verify that the monitor properly processes and reports patient physiological
parameter signal inputs. Record results of functional verification testing in a
copy of the Test List.

Software and certain of the monitor’s hardware components are continually


checked, first during power-up and later during on-line self-tests. All monitor
functions that are tested are performing in accordance with design specifica-
tions if no error messages exhibit.

Perform all functional verification procedures with the monitor off-line. Discon-
nect all patient parameter signal inputs from the monitor, and plug in signal
inputs from patient simulators and industry-standard test equipment only as
Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

instructed to do so. To assure that the monitor is off-line, unplug network


cable from monitor.

If the monitor fails to perform as indicated for any specific function during
functional verification procedures, troubleshoot the monitor using procedures
in Fault and Remedy document and remedy the malfunction before attempt-
ing to complete functional verification.

2 Recommended Use specified recommended tools and test equipment given in Table 1 below,
Tools and Test or a known equivalent, when performing functional verification tests. Substitu-
Equipment tions are approved only if an equivalent is listed. Use of other test equipment
Copyright reserved.

and/or accessories could result in inconclusive tests or damage to system


components.

6013.052 1
Test List Kappa Patient Monitor

Table 1 Recommended Tools and Test Equipment

Tools & Test Equipment Description


ECG/Resp:
Patient ECG/Resp simulator DNI Medsim 300B or equivalent
Leads: Three-lead grabber set, or IEC color code 1 33 75 230

IEC color code 2 33 75 248


Five-lead grabber set IEC color code 1 33 75 255

IEC color code 2 33 75 263


SpO2
Patient SpO2 simulator, or Nellcor‚ PT2500 or equivalent
Reusable SpO2 sensor: Durasensor Adult 45 34 475
Non-Invasive Blood Pressure:
*NBP simulator (calibrated) or Dynatech CuffLink, Veri-Cal Pressure
Transducer Tester, or equivalent
*Mercury manometer with hand bulb Baumanometer‚ 0-300 mmHg
NBP Calibration Assembly 28 77 855
NBP connection hose, 3.7 m 12 75 275
etCO2
5.00 ±0.03% CO2, balance N2, Calibration Gas Cylinder 28 68 532
Calibration Gas Regulator, with elbow fitting 28 68 540
Mercury barometer with mm grade
Invasive blood pressure:
IBP simulator w/ test cable DNI Medsim 300B or equivalent
Cardiac Output
SHP ACC C.O. INTERMEDIATE CABEL 33 68 458
Temperature Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

Temperature simulator w/ test cable (2 cables req’d) DNI Medsim 300B or equivalent
Temperature Y Cable 55 92 154
SHP ACC TEMP ADPT CBL 1/4” JACK (2 req’d) 51 98 333
Leakage
Leakage Tester Bender µP-Safety Tester 601/751, or
equivalent
Output Cables:
Recorder ÷ Interface Plate or CPS/IDS 47 21 770
IDS (or equivalent CPS), unless Monitor used only in DirectNet mode 52 60 110
Copyright reserved.

*For functional verification and calibration check. An equivalent must be accurate to ±0.3 mmHg from 0 to
330 mmHg.

2 6013.052
Kappa Patient Monitor Test List

3 Power Circuits and The following procedures check the monitor’s power circuits, power-up
Start-up sequence, and power off indicator. Begin this procedure with the monitor
turned off and plugged into a hospital-grade power source.

3.1 Power ON/OFF 1. Press ON/OFF switch on front panel and verify that power LED in
ON/OFF key turns on.
2. Press and hold ON/OFF key for approximately two seconds.
3. Verify that high pitched piezo tone sounds briefly and power LED in
ON/OFF key turns off.

3.2 Power-Up Sequence 1. Press ON/OFF key and verify following sequence of events:
2. Power LED in ON/OFF key turns on.
3. Display illuminates briefly, monitor beeps, high-pitched piezo tone sounds
briefly, and Battery charger LED illuminates briefly.
4. After a few moments, during which display is NOT illuminated, monitor
sounds a chime.
5. After a few more moments, New Patient prompt displays.
6. Select NO, and press rotary knob in to clear prompt.

NOTE
With all patient inputs removed from the monitor, only “adult” and “ALL
Alarms OFF” appear on MAIN screen.

7. Battery charge level bar graph, date and time report in message area at
bottom of display.

4 Rotary Knob The Rotary Knob in the lower right corner of the front panel controls an optical
encoder for pointing to and selecting display fields and functions.

1. Press Menu key, and verify that selected menu item changes as knob is
rotated one complete revolution in either direction.
2. Rotate knob until ¨ in upper left corner of Main Menu is selected.
3. Press knob in and verify that MAIN screen displays.

5 Fixed Keys The following tests verify that membrane switches on the front panel are func-
Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

tioning properly, and that the signal from the key is processed by the Front
Panel Control PCB. Functions controlled by the fixed keys are individually
verified elsewhere in this Chapter as required.

NOTE
Before beginning Key tests access Main menu. Select Monitor Setup →
Monitor Options →Speaker Volumes, and assure that Attention Tone Vol-
ume is set to other than OFF.

5.1 ON/OFF Key The ON/OFF key initiates the power-on sequence if the monitor is powered
Copyright reserved.

off, and powers-off the monitor, initiating a brief power-off piezo alarm, if the
monitor is powered-on.

6013.052 3
Test List Kappa Patient Monitor

NOTE
This test can be omitted if the procedure of section has already been per-
formed.

1. Press and momentarily hold ON/OFF key.


2. Verify that powered state of monitor changes.
3. Set monitor to powered-on state, if monitor powered off.

5.2 Main Screen Key The Main Screen key sets the display to the MAIN screen.

1. Press Menu key to display Main menu.


2. Press Main Screen key, and verify that Main menu extinguishes, and dis-
play returns to MAIN screen.

5.3 Alarm Silence Key The Alarm Silence key silences an alarm tone for one minute.

1. Assure that HR alarm is enabled, and without any input applied to Mul-
tiMed POD, plug MultiMed or MultiMed 12 cable into monitor. Monitor
should Alarm.
2. Press Alarm Silence key and verify that alarm ceases.
3. Turn off HR alarm in Alarm Limits Table before proceeding.

5.4 Alarm Limits Key The Alarm Limits fixed key calls up a setup table on which upper and lower
alarm limits for physiologic parameters can be assigned, and alarms and
alarm recordings can be enabled or disabled.

1. With MAIN screen displayed, press Alarm Limits fixed key.


2. Verify that Alarms Setup Table displays.

5.5 All Alarms Off Key The All Alarms Off key silences all alarms for a period of 2 minutes.

1. Press All Alarms Off key.


2. Verify that 2 minute countdown field appears at top center of display.
Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

5.6 Code Key The Code key can perform any of several functions, depending on other
parameter settings entered into the monitor. Refer to operating instructions in
the User Guide for more detailed information. The following procedure veri-
fies only that the monitor responds to a key press.

1. Press the key and verify that a change occurs in the display.
2. Double-click key to return display to former status.

5.7 Record Key The Record key starts and stops a recording of limited duration when the
monitor is mounted on an IDS (or CPS).
Copyright reserved.

1. With monitor mounted on IDS (or CPS docking station) press Record key.
2. Verify that message “Recording Not Accepted” appears in message field
at bottom right-hand side of display.

4 6013.052
Kappa Patient Monitor Test List

5.8 Print Screen Key Print Screen key initiates printing a copy of the display, on network printer.

1. Press Print Screen key twice.


2. Verify that message “Printer Unavailable, Waiting” appears in message
field at bottom right-hand side of display.

5.9 NBP Start/Stop Key The NBP Start/Stop key initiates or terminates the inflation cycle for the non-
invasive blood pressure monitor function.

1. Press Menu key. Access Monitor Setup →Monitor Options →Speaker Vol-
umes, and set Attention Tone Volume to 50%.
2. Press NBP Start/Stop key.
3. Verify that monitor sounds a tone. (Cuff must not be plugged into cuff con-
nector.)

5.10 Zoom Key When a patient is being monitored, the Zoom key calls up a 1-hour trend dis-
play for a quick overview of patient status.

1. Press Zoom key, and verify that “Fast Access Menu” displays.
2. Press Main Screen key to return.

5.11 Help Key 1. Press Help key and observe that Main Help Menu appears on display.
2. Press Main Screen key to clear display.

5.12 Mark Key The Mark key inserts parameter values with time and date stamp in the tabu-
lar trends. Verify that Mark membrane switch is functional as follows:

1. Press Menu key and select Review →Trend Table.


2. Press Mark key and observe that time stamp on HR readout changes
from white to green and Mark symbol appears above stamp.

6 ECG/RESP Func- With the cable plugged into the monitor connect either a 3-lead, 5-lead, or 6-
tions lead ECG cable from the Patient Simulator into the MultiMed POD.

6.1 Test Setup


Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

ECG/RESP Test Setup 1. Select HR parameter box and press rotary knob in to bring up ECG
menu.
2. Set all ECG Lead settings at default values and remaining parameters as
follows:
• ARR Monitoring: Basic
• RESP Monitoring: ON
• Pacer Detection: ON
• QRS Sync Marker: ON
• Pulse Tone Source: ECG
Copyright reserved.

• Pulse Tone Volume: 10%

Simulator Setup Set simulator as follows:


• ECG = Normal Sinus

6013.052 5
Test List Kappa Patient Monitor

• HR = 80 beats per minute (bpm)


• amplitude = 1.0 mV
• RESPIRATION = Normal Rest.
• rate = 20 breaths per minute (BPM)
• ohms = 1.0
• LEAD SELECT = II/RA-LL
• BASELINE IMPEDANCE = 500

6.2 Waveforms/Digital Verify the following:


Readouts/Tones • Waveform and HR correspond to data provided by simulator.
• Heart symbol (♥) blinks and pulse tone sounds for each QRS complex.
• White spike present at each QRS complex.
• RESP and HR digital readout correspond to settings of simulator.
• Vary Tone Volume setting and verify that pulse tone volume changes.
• Set Tone Volume to OFF, and verify that pulse tone stops.

6.3 Pacer Detection 1. Apply paced signal from simulator.


2. Verify that small “P” accompanies heart symbol (P♥) for every detected,
paced beat, blue spike appears for each paced signal, and HR digital
value agrees with pacer bpm setting.
3. Generate asystole condition in simulator, with pacer output still active.
4. Verify that ASY appears in ARR parameter box, an asystole alarm
sounds, and waveform is flatline with pacer pulses.
5. Disable pacer signal, and return simulator to setup above (Section 6.1).

6.4 Lead-Off Indicators 1. One at a time, disconnect each ECG lead from simulator, and verify the
following:
• For LL and RA electrodes, verify “Lead-Off” and “ECG Leads Invalid”
messages appear in message area, pulse tone ceases, *** replaces
digital heart rate in HR field.
• For V and RL electrodes, verify “Lead-Off” message, loss of V trace.
Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

• For LA electrode, verify “Lead-Off” message only.


2. Reconnect all leads to simulator.

6.5 Alarm Function This procedure also tests that the alarm function of the monitor, as applicable
to all other patient parameters, is operational in the monitor.

1. In Alarm Limits Table, set HR alarm parameters as follows:


• Upper limit = 110 bpm
• Lower limit = 40 bpm
• Alarm = ON
Copyright reserved.

2. Set emulator to HR = 120 bpm.


3. Verify that monitor responds with following Serious Alarm indications:
• HR in parameter field = 120

6 6013.052
Kappa Patient Monitor Test List

• HR parameter field blinks and color changes.


• Serious Alarm tone sounds.
• Message HR > 110 appears in message area at bottom of display.
4. Reset simulator to HR = 80 bpm.
5. Verify the following:
• HR parameter field returns to normal color
• HR returns to 80
• Message area continues to report most recent alarm, HR >110.
6. Press Alarm Silence fixed key.
7. Verify that “HR > 110” ceases to be reported.

6.6 Asystole 1. Switch power to simulator OFF.


2. Verify that HR parameter field reports ASY, “Asystole” appears in mes-
sage area at bottom of display, and monitor responds with Life-Threaten-
ing alarm.
3. Switch power to simulator ON.

7 SpO2 Function The SC 8000 monitors oxygen saturation (SpO2) and pulse rate using the
spectrophotometric method. SpO2 software is checked on monitor power-up
and also periodically while the monitor is in operation.

7.1 SpO2 Test Setup The SpO2 parameter box appears when an SpO2 input is applied to the mon-
itor through the MultiMed POD.

1. Access SpO2 parameter box menu. Set parameters as follows:


• Pulse Tone Source - SpO2
• Pulse Tone Volume -10%
• Waveform Size - 10%
• Averaging - Normal
2. Do either of the following as appropriate:
• If using a variable SpO2 simulator, set SpO2 level to 98% and pulse
rate to 70 bpm, and plug simulator into SpO2 input adapter cable to
MultiMed POD.
Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

• If using a Nellcor PT-2500 pocket tester or equivalent, plug tester into


SpO2 input adapter cable to MultiMed POD.

7.2 Waveforms/Digital 1. Verify the following:


Readouts/Tones • Simulated SpO2 waveform appears, and digital SpO2 and pulse rate
(PLS) values correspond to simulator settings.
• © symbol blinks in SpO2 field, and pulse tone sounds synchronous
with appearance of © symbol.

7.3 Pulse Tone Genera- 1. Select Pulse Tone Volume in SpO2 menu.
tor 2. Vary volume setting and verify that pulse tone volume changes.
Copyright reserved.

3. Set Pulse Tone Volume to OFF, and verify that tone stops. Then reset
pulse tone to ON.
4. Do either of the following, as appropriate:

6013.052 7
Test List Kappa Patient Monitor

• If using variable simulator, change oxygen saturation value and verify


that pulse tone frequency (pitch) increases as SpO2 level increases,
and decreases as SpO2 level decreases.
• If using pocket tester or equivalent, replace tester with adult finger
sensor and place sensor on your finger. Your SpO2 reading should be
> reading obtained from tester, and frequency (pitch) of pulse tone
should increase.

7.4 SpO2 Limits Alarms 1. In Alarm Limits Table, set SpO2 upper limit to 100%, lower limit to 80%,
and assure that alarm is ON.
2. Set simulator to an SpO2 value outside of set limits.
3. Verify that monitor responds with serious Alarm indication.
4. Reset simulator to value within alarm range.
5. Verify that alarm ceases

8 Temperature Func-
tion

8.1 Temperature Test 1. Using Temperature Y Cable input to MultiMed Pod, set up patient simula-
Setup tor to supply temperature input to Temp A.
2. Set simulator for standard 37°C.

8.2 Digital Readout 1. Verify that monitor indicates temperature of 37±0.1 °C.
2. Change simulator to temperature above and then below 37 °C.
3. Verify that monitor readout agrees with simulator settings ±0.1 °C.
4. Repeat procedure for Temp B input to Temperature Y Cable.

8.3 Temperature Cali- Use the following procedure to check temperature calibration. Record and
bration Check retain results in a copy of Table 2. Also record results in Acceptance Test
Report. Monitor must be returned to Dräger Medical for repair if calibration
check fails.

Table 2 Resistance Value vs Temperature

Resistance Set Temper- Reading On Tolerance Pass/Fail


Setting ature Monitor
Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

9045 -4.0 -4.1 to -3.9


3539 15.0 14.9 to 15.1
1355 37.0 36.9 to 37.1
843.2 49.0 48.9 to 49.1

8.3.1 Recommended Equip- • Decade Resistor, ±0.1% accuracy (or fixed resistors with same accuracy)
ment
• Siemens Temp Adapter Cable, 51 98 333 (Optional)
• Siemens “Y” Adapter, 55 92 154
Copyright reserved.

8.3.2 Procedure 1. Connect MultiMed cable to input of patient monitor.


2. Connect temp “Y” cable to temp input of MultiMed Pod.
3. Connect decade resistor to Temp A temperature input.

8 6013.052
Kappa Patient Monitor Test List

4. For each resistance value in Table 2, verify that monitor reports “Set Tem-
perature” value ±0.1°C.
5. Repeat procedure for Temp B input to Temperature Y Cable.

9 Non-Invasive Blood The SC 8000 Monitor measures non-invasive blood pressure (NBP) accord-
Pressure Function ing to the oscillometric method. Perform NBP characterization after replacing
the NBP Subassembly and also after replacing the Rear Housing Subassem-
bly (which includes the Connector I/O PCB). Use tools listed in Table 1 to
check calibration of the pressure transducer. Note pressure indicator required
for NBP calibration.

Figure 1 NBP Calibration Check / Calibration Test Setup

9.1 System Setup and 1. Set up NBP Calibration assembly (28 77 855) as illustrated in Figure 1.
Pneumatics Leak- 2. Assure that patient category is set to Adult.
age Test 3. Set following in NBP parameter field menu:
• Interval Time - OFF
• Continuous Mode - OFF
Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

• Calibrate Mode - ON. (Observe “NBP Cal. = 0 mmHg” appears.)


4. Clamp pneumatic hose (with hemostat or clamp) between T-connector
and monitor (1 in Figure 1), and using pressure bulb, increase pressure to
250 ±5 mmHg. Then clamp hose at inflation bulb (2 in Figure 1), and let
pressure stabilize for 1 minute. Do NOT run pump.
5. Observe pressure drop for an additional 5 minutes. Drop should be
<2 mmHg in 5 minutes. If not, tighten all connections and fittings and
retest equipment for leakage. When leakage test OK, go on to. With both
clamps removed, reinflate to 250±5 mmHg, if necessary, and then re-
clamp hose at inflation bulb.Observe pressure drop for 1 minute. Drop
should be <4 mmHg. If not, service monitor’s internal pneumatics system
and retest system for leakage. When leakage test OK, remove clamp at
Copyright reserved.

inflation bulb and go on to step 7.


6. Assure that pneumatic leakage is within specifications before continuing
to Calibration Check.

6013.052 9
Test List Kappa Patient Monitor

9.2 Calibration Check


NOTE
You may have only 2 min. to perform this test if the pump has run since the
monitor was powered on, before the H/W deflation error initiates. If this
happens, you will need to reset the monitor. If the pump has not run, how-
ever, there is no time limit.

1. Return to Main menu. Using hand bulb, increase pressure to 250


±5 mmHg, if necessary, and allow it to stabilize for 1 minute.
2. Verify that pressure values displayed on monitor and pressure indicator
are within ±3 mmHg of each other.
3. Slowly release pressure in 50 mmHg increments. At pressures of 200,
150, 100, and 50 mmHg, verify that pressure values on monitor and pres-
sure indicator are within ±3 mmHg of each other at each level.
4. If NBP function fails calibration check, contact your Dräger Medical ser-
vice representative. Otherwise, if calibration check OK, go on to 9.3 Hard-
ware Overpressure.

9.3 Hardware Overpres- 1. With monitor still in calibration mode, slowly increase pressure. while
sure observing pressure rise on monitor’s screen.

NOTE
Pressure indicator accuracy of ±0.3 mmHg is not required for hardware
overpressure and subsequent NBP tests.

2. Verify that pressure suddenly drops at 300 ±30 mmHg.


3. Release all remaining pressure before proceeding.Set patient category to
neonatal and repeat steps 1 through 3, using pressure limit of
157±8 mmHg for step 2.

9.4 Pump 1. Set Calibrate Mode in NBP parameter field menu to OFF.
2. Press NBP Start/Stop key.
3. Verify the following:
• Pressure increases and then decreases.
Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

• Inflation and deflation pressures reported in message field.


• Message “NBP No Pulsation” exhibits at end of deflation phase.
• All digital readouts in NBP field are ***.

9.5 Interval Mode 1. With NBP Calibrate Mode set to OFF, set Interval Time to 1 min.
2. Verify the following:
• One-minute countdown bar graph appears at bottom of NBP parame-
ter box.
• NBP pump starts immediately when rotary switch is pressed in.
• NBP sequences through an inflation/deflation cycle and produces
Copyright reserved.

“NBP No Pulsation” message at bottom of display.


• NBP pump starts again when one-minute countdown bar resets to ini-
tiate another cycle.
3. Press NBP Start/Stop key to stop cycle, and reset Interval Time to OFF.

10 6013.052
Kappa Patient Monitor Test List

9.6 Safety Timer 1. Assure that NBP Calibrate Mode is set to OFF.
2. Press NBP Start/Stop fixed key to start pump.
3. Press same key again to stop measurement.
4. Set NBP Calibrate Mode to ON, and press MAIN SCREEN key.
5. Press CODE key to start screen stopwatch. Observe monitor screen.
6. Press CODE key again to stop watch when “NBP Cuff Deflation Error”
message displays, indicating that safety timer has activated.
7. Verify that elapsed time is as follows:
• Adult: 120 ±5 seconds.
• Neonatal: 90 ±5 seconds
• French Homologation: 60 ±5 seconds
8. Press NBP Start/Stop fixed key.Verify that tone sounds and pump fails to
start.Power-cycle monitor to clear fault condition.

10 etCO2 Function With ≥VE0 software installed in the Kappa Monitor, the etCO2 Pod enables
the Monitor to non-invasively monitor end-tidal CO2 (etCO2) using a tech-
nique that relies on the selective absorption properties of CO2 to specific fre-
quencies of infrared radiation. The pod automatically compensates for
variations in ambient barometric pressure if Monitor set to automatic mode.
Before beginning this procedure, use a mercury column barometer or equiva-
lent other device to determine local atmospheric pressure. Record this value.

1. Plug etCO2 Pod into Aux. connector (between Hemo Med 1 connector
and NBP port) on front of Monitor.

NOTE
Observe that the parameter box appears on the monitor display screen,
and that “etCO2 Sensor Warming Up” followed by “etCO2 Place Sensor on
Zero Cell” appears in the message field.

2. While sensor is warming up, select etCO2 parameter box.


3. Assure that Atmospheric Pressure is set to “Manual”, and is set to value
indicated by mercury column barometer.
4. Place sensor on Zero Cell.
Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

5. Note that “etCO2 Calibrating Sensor” appears in message field, followed


by “etCO2 Place Sensor on Ref Cell”.
6. Place sensor on Reference Cell.
7. Verify that”etCO2 Verifying Sensor Cal” followed by “etCO2 Sensor Cal
Verified” appears in message area, and reading in etCO2 parameter box
= 38 ±2mmHg.
8. Attach adult airway adapter to calibration gas cylinder. Do NOT open
valve on cylinder.

NOTE
Copyright reserved.

As CO2 is heavier than room air, set up the airway adapter such that the
point where the gas exits from the adapter is higher in elevation than the
point where it enters.

9. Place sensor on adult airway adapter and note reading = 0 ±1 mmHg.

6013.052 11
Test List Kappa Patient Monitor

10. Turn valve on the cylinder until it is fully open.


11. Wait for 30 seconds and record displayed value.
12. Close valve and remove sensor from airway.Verify that measured value is
in range (0.05 x local pressure) ±3, rounded to nearest integer.
13. Select etCO2 parameter box and set Atmospheric Pressure to AUTO.
14. Repeat steps 4 through 12.
15. Verify that measured value = previously measured value ±3, rounded to
nearest integer.

11 HemoMed Pod

Figure 2 IBP Functional Verification Test Setup for HemoMed Pod

Refer to Figure 2 for test setup. Invasive blood pressure and cardiac output
functions of HEMO2/4 PODs are incorporated in the HemoMed Pod. A single
cable connects the Pod to the HemoMed 1 input on the front of the Kappa
Monitor. Pressure labels for HemoMed Pod channels are set in the Monitor.

11.1 IBP Function


Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

11.1.1 IBP Test setup 1. With MultiMed cable and all other patient inputs unplugged from monitor,
power-cycle monitor. Select NO to clear display.
2. Plug cable from HemoMed Pod output into HemoMed 1 connector on
front of Monitor.

Channel A 1. With MAIN screen displayed, connect BP output from simulator to first
input, channel A, on HEMO POD adapter.Set IBP simulator for a static
pressure = 0 mmHg.
Copyright reserved.

NOTE
“Zero Required” message, appears on display.

Monitor Zero Function 2. Press Zero All key on HemoMed Pod.

12 6013.052
Kappa Patient Monitor Test List

NOTE
All four pressures are zeroed simultaneously in the monitor, even if only
channels A and B are to be used in the HemoMed Pod.

3. Verify that “Zero Accepted” message that changes to “Static Pressure”


appear in message field.
4. Select a pulsatile pressure on patient simulator.
5. Assure that Cal Factor and Manometer Cal are set to 100.
6. Verify that pressure reading on monitor agrees with values generated by
pressure signal from simulator.
7. If verifying only Kappa, omit remaining steps in this Section and go to
Section 11.2. If also verifying HEMO POD, go on to step 9.

Channel B 8. Unplug BP adapter cable from channel A on HEMO POD and plug it into
Channel B.
9. Verify that pressure reading on monitor agrees with values generated by
pressure signal from simulator.
10. Do either step a or step b as appropriate.
• If HemoMed Pod has four channels, unplug the BP adapter cable
from channel B and plug it into channel C. Then go on to step 12.
• If HemoMed Pod has only two channels, bypass remaining steps of
this section and go to Section 11.2.

Channel C 11. Unplug BP adapter cable from Channel B on HemoMed Pod and plug it
into Channel C.
12. Verify that pressure reading on monitor agrees with values generated by
pressure signal from simulator.

Channel D 13. Unplug BP adapter cable from Channel C on HemoMed Pod and plug it
into Channel D.
14. Verify that pressure reading on monitor agrees with values generated by
pressure signal from simulator.
15. Unplug simulator input adapter cable from HemoMed Pod, and go on to
Section 11.2.

11.2 Cardiac Output 1. Plug C.O. output from simulator into C.O. test adapter cable, and plug
Function adapter cable into C.O input on HemoMed Pod.
Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

2. When READY appears in CO parameter field verify that blood tempera-


ture indication is 37±0.15 °C.
3. Press C.O. START key on simulator, and verify an Injectate Temperature
indication of 1±0.2 °C.
Copyright reserved.

6013.052 13
Test List Kappa Patient Monitor

12 HEMO POD 2/4

Figure 3 IBP Functional Verification Test Setup for HEMO2/4 PODs

Refer to Figure 3 for test setup.

12.1 IBP Function IBP parameter boxes for up to two pressures for HEMO 2 POD or up to four
pressures for HEMO 4 POD display automatically when IBP signal from POD
is plugged into monitor, when Display Mode is set to automatic.

12.1.1 IBP Test setup 1. With MultiMed cable and all other patient inputs unplugged from monitor,
power-cycle monitor. Select NO to clear display.
2. Plug cable from HEMO POD IBP SC 9000 output into Aux./Hemo3 input
on front of Kappa Monitor.

Channel A 3. With MAIN screen displayed, connect BP output from simulator to first
input, Channel A, on HEMO POD adapter.
4. Set IBP simulator for static pressure = 0 mmHg.
Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

NOTE
“Zero Required” message, identified by same pressure label is shown in
window for channel A on front of HEMO POD appears on display.

5. Monitor Zero Function


6. Press Zero All key on HEMO POD.
7. Verify that “Zero Accepted” message that changes to “Static Pressure”,
both identified by same pressure label, appear in message field.
8. Select a pulsatile pressure on patient simulator.
9. Assure that Cal Factor and Manometer Cal are set to 100.Verify that
pressure reading on monitor is in agreement with values generated by
Copyright reserved.

pressure signal from simulator.


10. Plug cable from HEMO POD IBP SC 9000 output into Aux./Hemo2 input
and repeat steps 3 through 9. Otherwise, continue.

14 6013.052
Kappa Patient Monitor Test List

11. If functionally verifying only the Kappa Monitor, omit remaining steps in
this section and go to Section 12.1.2. Otherwise, if also functionally verify-
ing HEMO POD, go on to step 12.

Channel B 12. Unplug BP adapter cable from Channel A on HEMO POD and plug it into
Channel B.
13. Observe that pressure label changes to same pressure label as is shown
in window for Channel B on front of HEMO POD.
14. Verify that pressure reading on monitor is in agreement with values gen-
erated by pressure signal from simulator.
15. Do either step a or step b as appropriate.
a) If the HEMO POD has four channels, unplug the BP adapter cable
from channel B and plug it into channel C. Then go on to step 16.
b) If the HEMO POD has only two channels, bypass remaining steps of
this section and go to Section 12.1.2.

Channel C 16. Unplug BP adapter cable from Channel B on HEMO POD and plug it into
Channel C.
17. Observe that pressure label changes to same pressure label as is shown
in window for Channel C on front of HEMO POD.
18. Verify that pressure reading on monitor is in agreement with values gen-
erated by pressure signal from simulator.

Channel D 19. Unplug BP adapter cable from Channel C on HEMO POD and plug it into
Channel D.
20. Observe that pressure label changes to same pressure label as is shown
in window for Channel D on front of HEMO POD.
21. Verify that pressure reading on monitor is in agreement with values gen-
erated by pressure signal from simulator.
22. Unplug simulator input adapter cable from HEMO POD, and go on to
Section 12.1.2.

12.1.2 Temperature Function 1. To perform temperature calibration check, refer to procedure of Section
8.3. Connect decade resistor directly across temp input to Hemo 2/4 Pod
instead of using Y cable.
2. Plug fixed temperature (37 °C) output from simulator into TEMP A con-
nector HEMO POD.
Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

3. Plug selectable temperature output from simulator into TEMP B connec-


tor HEMO POD, and set temperature for other than 37 °C.
4. Verify the following:
• T field appears on MAIN screen
• T1a temperature = 37 ±0.1 °C
• T1b temperature = simulator setting ±0.1 °C.
5. Select Temperature field, and access TEMP1 menu.
6. Select TEMP Display and then select temperature difference key.
7. Verify that T1b changes and reports temperature difference between T1a
and T1b ±0.2 °C.
Copyright reserved.

8. Reset TEMP Display in TEMP1 menu to T1b.Press Main Screen key to


clear table and return to MAIN Screen.

6013.052 15
Test List Kappa Patient Monitor

12.1.3 Cardiac Output Function 1. Plug C.O. output from simulator into C.O. test adapter cable, and plug
adapter cable into C.O./Temp B input on HEMO POD.
2. When READY appears in CO parameter field verify that blood tempera-
ture indication is 37±0.15 °C.
3. Press C.O. START key on simulator, and verify an Injectate Temperature
indication of 1±0.2 °C.

12.2 Memory Backup The monitor retains patient-related data, such as alarm limits, trends, and
stored alarm recordings when it is powered off.

1. With monitor powered-up and no patient inputs applied, press Alarms


Limits fixed key and change limit for any parameter.

NOTE
Note time on Real Time Clock.

2. Power monitor off for approximately 2 minutes, and then power it back on
again.
3. After MAIN screen displays, press Alarm Limits fixed key to call up Alarm
Limits Table.
4. Verify that new limit(s) you set in Step 1 have been retained, and that
clock has advanced the proper amount of time.

12.3 MIB and CAN Omit steps 1 and 2 if MGM not used with this monitor.
Options (if installed)
1. With MAIN screen displayed on the Kappa Monitor, connect MGM to
Comm connector input on Monitor and to a hospital-grade power source.
2. Switch MGM power on, and verify that Agent parameter box displays on
Kappa screen.
3. Plug MIB - External Vendor Device cable into any MIB input and into
External Vendor Device port as given in Software Compatibility Chart for
installed monitor software.
4. Refer to Operating Instructions for External Vendor Device to set up
device as required in the Software Compatibility Chart for the installed
software version.
5. Power External Vendor Device ON, and verify that vendor parameters
Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

appear on Monitor Display in accordance with MIB User Guide.

NOTE
If test mode not possible on External Vendor Device, only parameter and
waveform templates may display on Patient Monitor. Refer to User Guide
for installed software version for instructions on using MIB option.

12.4 Adv Comm Option (if Functional verification of proper operation of the Adv Comm Option has been
installed) programmed into the software. To verify proper operation, connect an SDC
(Surgical Display Controller) to the Adv Comm connector. The Monitor auto-
Copyright reserved.

matically performs a functional verification and reports successful completion


on the Display.

12.5 Configuration This checks for proper configuration and for monitor communication via the
Infinity Network.

16 6013.052
Kappa Patient Monitor Test List

1. Connect monitor to Infinity Network using ethernet connector on back of


monitor.
2. Setup monitor for simulated patient using external patient simulator.
3. Verify that waveforms display on MVWS.

NOTE
If Infinity Network not equipped with MVWS, use remote view function of
another monitor on network to view waveforms, skip step 4 and go on to
Section 12.6.

4. Access net Info on MVWS and verify that monitor labels are unique.

12.6 Recorder Function The R50 Recorder connects to the Kappa Monitor through X13 on back of the
monitor. The following procedure verifies that the monitor is communicating
with the Recorder.

1. Connect known good R50 Recorder to monitor.


2. Press Menu key, and access Monitor Setup →Biomed →Service.
3. Enter Service password (4712) and select Accept.
4. Select Bedside Setup, and set Waveform Simulator to ON.
5. After monitor cycles, bring up MAIN screen.
6. Press Record fixed key.
7. Verify that recorder begins to dispense waveform printout.
8. Press Stop key on recorder to stop paper flow.
9. Press Menu key, and access Monitor Setup →Biomed →Service.
10. Enter Service password (4712) and select Accept.
11. Select Bedside Setup, and set Waveform Simulator to OFF.

13 Leakage Tests Kappa Monitors are AC voltage operated devices. Leakage tests assure that
under both normal and fault conditions, any leakage current does not exceed
values given below. Record all values in copy of Acceptance Test Report.

13.1 Resistance Test 1. Using DMM (Fluke, model 8050A or equivalent) measure resistance
between ground stud on rear panel and earth pin on the AC inlet (i.e.
Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm

where the cord plugs in).

NOTE
Since the value of resistance is very small, you are reminded that it is nec-
essary to account for lead resistance in the measurement. Prior to measur-
ing the resistance between the ground stud on the rear panel and the earth
pin on the AC inlet, connect both meter leads to the ground stud and mea-
sure the meter lead resistance. Subtract this lead resistance value from the
DVM reading obtained in the ground stud / earth pin resistance measure-
ment.
Copyright reserved.

6013.052 17
Test List Kappa Patient Monitor

Figure 4 Leakage Current Test Setup

2. Verify that resistance between ground stud and earth pin <0.5Ω. Record
reading in space provided on Test List.

13.2 Leakage Current 1. Perform leakage current tests with monitor plugged into leakage tester.
Tests See Figure 4.
2. Follow leakage tester manufacturer’s instructions to measure each of
leakage currents given below:

a) Enclosure (Case) Leakage Current:


Normal: Limit <10 µa
Open Ground: Limit <500 µa at 240 vac
or
<300 µa at 120 vac
Open Ground and Reverse Mains: Limit <500 µa at 240 vac
or
<300 µa at 120 vac
b) Patient Combined Leakage Current:
Normal: Limit <10 µa
Open Ground: Limit <50 µa
c) Leakage with Mains on Patient Leads:
Limit <50 µa Revision 1_ Released_Printed on_02.08.06_P6013052_Kappa_IEC_Test_List.fm
Copyright reserved.

18 6013.052
Test List Report Kappa Patient Monitor

Test List Report X = Test Passed

Power Circuits and Startup _______

Power ON/OFF key _______

Power LED _______

Piezo tone _______

Power-Up Sequence _______

Monitor Powered By External Source _______

Rotary Knob _______

Pointing Function _______

Selecting Function _______

Fixed Keys _______

ON/OFF _______

Main Screen _______

Alarm Silence _______

Alarm Limits _______

All Alarms Off _______

Code _______

Record _______

Print Screen _______

NBP Start/Stop _______

Zoom _______
All rights reserved. Copyright reserved.

Help _______
Kappa_IEC_Test_List_Report.fm 02.08.06

Mark _______

ECG/Resp Functions _______

Waveforms/Digital Readouts/Tones _______

Pacer Detection _______

Lead-Off Indicators _______

Draeger Medical Systems, Inc. 6013.052 Revision 1 Released 19


Kappa Patient Monitor Test List Report

Alarm Function _______

Asystole _______

SpO2 Function _______

Waveforms/Digital Readouts/Tones _______

Pulse Tone Generator _______

Limits Alarm _______

Temperature Function (MultiMed POD) _______

Digital Readout Temp A_______

Temp B_______

Temperature Calibration _______

Non-Invasive Blood Pressure _______

Calibration OK _______

Hardware Overpressure _______

Pump _______

Interval Mode _______

Safety Timer _______

etCO2 _______

Sensor Cal. _______

Sensor Zero _______

Measured Value w/ Manual Atm. Press._______

Measured Value w/ AUTO Atm. Press. _______


All rights reserved. Copyright reserved.

HemoMed Pod _______

Invasive Blood Pressure Function _______


Kappa_IEC_Test_List_Report.fm 02.08.06

HEMO Channel A Monitor Zero Function _______


(Aux./Hemo3 input)
Monitor Pressure Reading _______

HEMO Channel A Monitor Zero Function _______


(Aux./Hemo2 input)
Monitor Pressure Reading _______

20 Draeger Medical Systems, Inc. 6013.052 Revision 1 Released


Test List Report Kappa Patient Monitor

HEMO POD Channel B _______

HEMO POD Channel C _______

HEMO POD Channel D _______

Cardiac Output Function _______

Blood temperature ______

Injectate temp ______


HEMO POD2/4 _______

Invasive Blood Pressure Function _______

HEMO Channel A Monitor Zero Function_______


(Aux./Hemo3 input)
Monitor Pressure Reading_______

HEMO Channel A Monitor Zero Function_______


(Aux./Hemo2 input)
Monitor Pressure Reading_______

HEMO POD Channel B _______

HEMO POD Channel C _______

HEMO POD Channel D _______

Temperature Function _______

T1a_______

T1b_______

T1a - T1b_______

Cardiac Output Function _______

Blood temperature ______

Injectate temp ______

Memory Backup _______


All rights reserved. Copyright reserved.
Kappa_IEC_Test_List_Report.fm 02.08.06

MIB and CAN Options (if installed) _______

CAN Option_______

MIB Option_______

Adv Comm Option (if installed) _______

Draeger Medical Systems, Inc. 6013.052 Revision 1 Released 21


Kappa Patient Monitor Test List Report

Configuration_______

Battery Charger Circuit _______

Recorder Function _______

Leakage Tests _______

Resistance Test Resistance Reading (Ω_________

Leakage Current Tests

Enclosure (Case) Normal: _________


Leakage Current:
Open Ground:

@ 240 vac_________

or @ 120 vac_________

Open Ground and Reverse Mains:

@ 240 vac_________

or @ 120 vac_________

Patient Combined Normal: _________


Leakage Current:
Open Ground:_________

Leakage with Mains _________


on Patient Leads:

All rights reserved. Copyright reserved.


Kappa_IEC_Test_List_Report.fm 02.08.06

22 Draeger Medical Systems, Inc. 6013.052 Revision 1 Released


Test List Report Kappa Patient Monitor

1 Clinical Site Report


Clinical Site Name: Date:

Address:

_______________________________________Clinical Site Facility Manager:

_______________________________________Clinical Site Contact Person:

Tel. No: FAX No.

Int’l Code Number Ext:

Monitoring Unit _______________________________

Care Unit _________________

Monitor Serial No. _________________________

Monitor has passed all required tests.

Dräger Medical Systems Service Product Representative:

___________________________________ __________________________________ ________________


All rights reserved. Copyright reserved.

Name (Print) Signature Date


Kappa_IEC_Test_List_Report.fm 02.08.06

Clinical Site Representative:

___________________________________ __________________________________ ________________

Name (Print) Signature Date

Draeger Medical Systems, Inc. 6013.052 Revision 1 Released 23


Kappa Patient Monitor Test List Report

All rights reserved. Copyright reserved.


Kappa_IEC_Test_List_Report.fm 02.08.06

24 Draeger Medical Systems, Inc. 6013.052 Revision 1 Released


Kappa Patient Monitor Spare Parts

Spare Parts Refer to the Figures and corresponding Tables for a listing of replacement
subassemblies and components. The lists contain all information available as
of the publication date of this Manual. Field experience and technological
development, however, may require future modifications. Subsequent
changes may be published as a supplement, as well as posted on the Dräger
Medical Intranet site.
Revision 1_ Released_Printed on_02.08.06_E6013052_Kappa_IEC_Spare_Parts.fm

Figure 1 Basic Kappa Monitor (Top Cover Removed)

Table 1 Basic Kappa Patient Monitor

Item No. Description Part Number


1 Battery 59 43 563
2 Connector I/O PCB 57 37 577
Copyright reserved.

3 NBP Pump Subassembly 59 43 571


4 NBP Air Intake Filter 28 66 726
5 Speaker Subassembly 51 94 365

6013.052 1
Spare Parts Kappa Patient Monitor

Item No. Description Part Number


6 Power Supply 57 38 450
7 Power Supply Harness 57 38 013
8 Power Supply Entry Module / Harness 57 38 468
9 R393 Access Port Cover 59 56 979
10 Fan 57 40 936
11 Main Processor PCB Subassembly 57 40 050
12 Rear Bezel w/o installed Adv. Comm Option 57 36 470
13 Top Cover 57 36 397

Figure 2 Front Bezel Subassembly and Top Cover

Table 2 Front Bezel Subassembly and Top Cover

Item No. Description Part Number


1 Front Bezel Subassembly 59 43 597
2 Label Kit for Front Panel 59 43 589
3 Optical Encoder Subassembly 43 11 622 Revision 1_ Released_Printed on_03.08.06_E6013052_Kappa_IEC_Spare_Parts.fm

4 Rotary Knob 43 16 662


5 R393 Access Port Cover 59 56 979
6 Top Cover 57 36 397
Copyright reserved.

2 6013.052
Kappa Patient Monitor Spare Parts

Figure 3 Adv Comm and MIB 1&2 Options


Revision 1_ Released_Printed on_03.08.06_E6013052_Kappa_IEC_Spare_Parts.fm

Table 3 Adv Comm and MIB 1&2 Options

Item No. Description Part Number


1 MIB 1&2 PCB 50 90 794
2 IDS Adv Comm PCB 59 48 356
3 14-P USD Adv Comm Cable Asy 59 49 693
4 CAN Adv Com Cable Asy 59 49 685
5 Rear Bezel w/ Adv Comm Option 59 47 630
6 Ethernet Interconnect Cable 57 40 985
Copyright reserved.

6013.052 3
4
Spare Parts
Kappa Patient Monitor

6013.052
Copyright reserved.
Revision 1_ Released_Printed on_03.08.06_E6013052_Kappa_IEC_Spare_Parts.fm
Manufacturer: EC Representative:
Draeger Medical Systems, Inc. Dräger Medical AG & Co. KG
6 Technology Drive Moislinger Allee 53 – 55
Andover, MA 01810 23542 Lübeck
USA Germany

Phone: (978) 379-8000 Phone: (+49) (0) 1805-3723437


Fax: (978) 379-8330 Fax: (+49) 451/882 - 3779
Web: http://www.draeger.com Web: http://www.draeger.com

Subject to change without notice.


Will not be replaced in the event of modifications.

© Copyright by Dräger Medical AG & Co. KG, 2006, Lübeck, Germany.


The reproduction, distribution and utilization of this document as well as the communication of its contents to
others without explicit authorization is prohibited. Offenders will be held liable for the payment of damages. All
rights reserved in the event of the grant of a patent, utility model or design.

The warranty and liability conditions of the general terms and conditions for business transactions of
Dräger Medical AG & Co. KG are not extended by this Technical Documentation.

You might also like