Select Analyst: Operating Manual
Select Analyst: Operating Manual
SELECT ANALYST
OPERATING MANUAL
ISSUE - A
1/9/09
CONTENTS PAGE
1 DOCUMENT HISTORY 5
2 CONTACT DETAILS 5
7 TROUBLE SHOOTING 25
8 TECHNICAL SPECIFICATION 27
8.1 Product outputs 27
8.2 Treated water specification 27
8.3 Water storage 27
8.4 Fuse rating/ type 27
9 INSTALLATION DETAILS
9.1 Media cartridge packs 29
9.2 Feedwater supply requirements 29
9.3 Electrical Supply 30
9.4 Unit positioning 30
9.5 Drain connection 32
9.6 Installation of media cartridge packs 32
10 APPENDIX 35
10.1 CE Declaration 35
10.2 WEEE Declaration 36
10.3 Process Flow Diagram 37
1 DOCUMENT HISTORY
2 CONTACT DETAILS
We trust the system will meet all your expectations. However, should you
have any additional questions concerning the equipment please do not
hesitate to contact Purite Ltd at:
Purite Ltd
Bandet Way
Thame e
Oxfordshire
OX9 3SJ
United Kingdom
I This symbol points out important information for working with the
system in the proper manner. Failure to observe these references
may result in malfunctions in the system or impact on the
environment.
The units must only be operated with water supplied in accordance with
the quality described and the operating parameters specified in
“installation details”.
The units must not be operated unless in proper working order. Any
malfunctions must be rectified immediately.
Only persons who have read and understood this Operating Manual
should be permitted to operate the unit. When operating the units, it is
important to observe all safety information.
The operator must take care to ensure that all maintenance, inspection
and assembly tasks are performed by authorised and qualified
professionals who have been sufficiently informed regarding the task at
hand (via use of Operating Manual).
The system must be shut down and be protected from further use until all
repair and maintenance tasks have been completed. It is essential to
observe the shutdown procedure as described in this Operating Manual.
The Purite Select Analyst purifies potable water. The water is purified by a
combination of processes including; ion-exchange, activated carbon and
reverse osmosis.
The purified water produced by the unit complies with latest standards for
laboratory grade purified water.
Each of the above variants is available with 4 different output flow rates:
Please note that the above outputs are based on an 8 hour working day at
60psi (4 bar) inlet pressure, 25oC and maximum of 400 ppm total dissolved
solids.
The unit type is detailed on its serial plate which can be found on the
reverse side of the front cover.
The Select Analyst is provided with an installation kit for quick installation.
Both media packs are provided and packaged separately. Refer to Section
9.1 and 9.6.
water a bib tap has been provided at the front of the tank.
Note: Generation units are NOT supplied with a water storage tank
The front cover is easily removed and provides access to the cartridge
media packs.
Your select Analyst is operated via the touch screen display. Buttons are
provided on the display to access both the start-up and set-up menus.
MENU
WELCOME
START
ANALYST
Start
Stop
Menu
→ Language
→ Time and date
→ Set time
→ Set date
→ set format
→ Maintenance
→ Pre-treatment cartridge pack date
→ Puripac cartridge pack date
→ Product Information
→ Engineer Menu
→ Auto restart
→ Enable
→ Disable
The unit runs an initialisation for 10 seconds before the unit can be
activated. To switch the unit on press the START button after the
initialisation is compete
When start is pressed the following screen is displayed to show the unit
is in operation and processing water.
QUALITY 10.0 MΩ
TEMPERATURE 20.0 °C
STOP
Outlet water quality and outlet water temperature are displayed on the
screen for your information.
A single red button is featured, STOP. When pressed the display reverts to
the Power On Screen and all functions are terminated.
Pressing the MENU button takes you to the Settings Menu where system
parameters can be set up.
MENU
WELCOME
ANALYST START
All menu
nu screens feature four buttons BACK, UP, DOWN, ENTER.
UP and DOWN are used to move through the available options and/or to
increase/decrease numerical vales.
ENTER selects to the option highlighted.
BACK returns to the previous screen.
SETTINGS MENU
BACK
Language
Time and Date
Maintenance
Product information
Engineer menu
Auto Restart
ENTER
5.2.1 Language
LANGUAGE
BACK
ENGLISH
FRENCH
ENTER
Highlight the appropriate tick box by pressing on the relevant screen area
and press enter.
ENTER
Set time.
PROCESSING
BACK
12:54
HH:MM
NEXT
Set date
SET DATE
BACK
04/01/09
DD/MM/YY
NEXT
SET DATE
BACK
DD MM YY
DD MM YYYY
MM DD YY
MM DD YYY ENTER
The maintenance menu provides the capability to set defined dates for
replacement of media cartridge packs in the device. These dates, when
expired, will trigger an alarm which will be displayed on the menu bar.
MAINTENANACE
BACK
Pretreatment Pack Date
Puripac Pack PP 8 Date
ENTER
Enter the menu and press reset. Failure to reset and replace media
packs will lead to poor water quality.
16/12/2009
6 MONTH EXPIRY
RESET
Puripac date
BACK
16/12/2009
6 MONTH EXPIRY
RESET
Product Information
BACK
Type: Analyst
Serial number: 123456
Unit ID: 00
Unit Version: R.1.5
Pod Version : R.1.6
Commissioning Date: 01/01/01
Memory Checksum: OxFE66
Address 02
If selected will return the unit into operation automatically in the event of a
power cut.
AUTO RESTART
BACK
Disabled
Enabled
ENTER
6 MAINTENANCE
Leave the unit to stand for 30 seconds to relieve any internal pressure.
Remove the front door cover by pulling it gently from the unit using the
hand holds provided to expose the two media cartridge packs.
Press down the lever over the corresponding cartridge to be removed and
pull it away from its retainer.
Discard the exhausted media pack. Remove the new pack from its bag
removing the two black plugs from the inlet/outlet ports.
Insert the media pack into the unit locating it securely into its retaining
block, push it home until it clicks into position and the retaining lever clicks
up.
Press “START” and follow the instructions on screen for setting new pack
replacement date (see previous section)
After resetting the pack date the unit can be started from the power on
screen.
The inlet screen filter is located in the right hand side of the unit.
o The filter is located in the exposed panel on the right hand side
of the unit.
To obtain access to the filter screen unscrew the black knurled top anti-
clockwise. The screen will then be exposed. Pull the screen away
from the housing and clean away any debris.
Once cleaned, replace the screen and black top. Replace the side
panel and cover using the reverse procedure and switch on the water
supply. Start the unit from the power on screen
Sanitisation of the unit should be carried out at least once every 6 months.
Open the front door cover to reveal the media cartridge packs.
To sanitise the unit insert a cleaning pack adaptor (Purite part number
L998549) into each of the media pack locations.
Allow the unit to fill the tank. When the tank is full the unit will stop.
Allow the unit to stand with a full tank for two hours. Drain the tank
using the bib tap located on the front of the tank.
The unit will automatically restart and begin to fill the tank. When the
tank is full drain it and allow it to fill once more.
When full switch the unit off, drain the tank and fit new media packs.
Restart the unit and allow it to fill the tank. When filled, switch the unit
off and drain the tank again.
7 TROUBLE SHOOTING
The touch screen will display various alarms and/or error messages if a fault
is identified with the unit. The guide below can be used to rectify common
alarms.
Line cell error Faulty cable – open circuit Call service department
8 TECHNICAL SPECIFICATION
Unit Size
Product Outputs
40 80 160 320
@ 10 °C 3.6 7.2 14.4 30
@ 25 °C 6 12 24 48
Inorganics 1-10MΩ.cm
pH1 Neutral
Bacteria >99% rejection
Organics (TOC) <50 ppb
1
pH of stored water may decrease due to absorption of free carbon
Pure water storage Optional 20 litre tank (external 50 and 100 litre
tanks available)
9 INSTALLATION KIT
Contents:
3 metres ½” Blue hose,
7metres 8mm nylon tubing,.
1 off Jubilee clip.
3 metres ,12mm nylon tubing,.
1 off 12mm stem elbow.
3 off 8mm stem elbows.
Your Select Analyst is supplied with two different media cartridge packs
which are available in two sizes depending on your unit capacity.
The pretreatment pack must be fitted in the left hand position and the
Puripac cartridge in the right hand position when viewed from the front.
The unit has been designed to operate only on potable tap water and
should comply with the following limits.
Your Select Anaylst will require a single phase electrical supply that meets
the requirements below
Ensure that the mains supply is provided via a convenient switched outlet
and is protected with a 5 amp fuse.
Secure the bracket to the wall. Remove the feet of the Analyst by
unscrewing them. Place the Analyst on the wall bracket and fix to the
bracket using screws through the feet mounting points.
Fit 8mm tubing to the water inlet connector found on the rear of the
unit and the other end to your water supply.
Attach a suitable length of 8mm tubing to the drain connection so that the
free end may be inserted into an appropriate drain. The tubing must not be
kinked or the flow restricted in anyway.
If the drain is to be run over 5 meters and/or rises above the level of the
unit then contact our Service Department for advice.
Attach a suitable length 12mm hose to the overflow connection on the unit
and run to a suitable drain point. If the drain is to be run over 5 meters
and/or rises above the level of the unit then contact our Services Dept for
advice.
Remove the door of the unit by tugging gently on the hand hold provided in
the door.
Insert the Pretreatment (PT8 or PT16) in the left hand retaining block
pushing it home until it clicks into position and retaining mechanism clicks
up.
Insert the Puripac PP8 or PP16 pack in the right hand retaining block
pushing it home until it clicks into position and retaining mechanism clicks
up.
Replace the door by locating the top of the door under the retaining
mouldings and lifting the bottom of the door until it slides home.
Switch on water supply to the unit.
Switch on the electrical supply to the unit.
This touch screen display will illuminate and display the start screen.
MENU
WELCOME
ANALYST START
10 Appendix
10.1 CE declaration
The boost pump increases the feed water pressure to a level to enable
the reverse osmosis (RO) membranes to operate efficiently.
Flow diagram