User'S Manual Ehp-50F: Electric and Magnetic Field Probe - Analyzer From 1 HZ Up To 400 KHZ
User'S Manual Ehp-50F: Electric and Magnetic Field Probe - Analyzer From 1 HZ Up To 400 KHZ
User’s Manual
EHP-50F
ELECTRIC AND MAGNETIC
FIELD PROBE - ANALYZER
If the instrument is used in any other way than as described in this User’s Manual, it may become unsafe
Before using this product, the related documentation must be read with great care and fully understood to
familiarize with all the safety prescriptions.
To ensure the correct use and the maximum safety level, the User shall know all the instructions and
recommendations contained in this document.
This product is a Safety Class III instrument according to IEC classification and has been designed to meet
the requirements of EN61010-1 (Safety Requirements for Electrical Equipment for Measurement, Control and
Laboratory Use).
In accordance with the IEC classification, the battery charger of this product meets requirements Safety Class
II and Installation Category II (having double insulation and able to carry out mono-phase power supply
operations)..
It complies with the requirements of Pollution Class II (usually only non-conductive pollution). However,
occasionally it may become temporarily conductive due to condense on it.
Contents III
5 EHP-50F Stand-alone mode Page
5.1 Stand-alone mode description………….………………….… 5-1
5.2 EHP-50F Data Logger…………………………………………. 5-2
5.2.1 Run EHP-50F Stand-alone mode software………………. 5-2
5.2.2 Use EHP-50F Logger……………………………………….. 5-5
5.2.3 EHP-50F battery charging………………………………….. 5-5
IV Contents
Figures
Figure Page
Tables
Table Page
Contents V
SAFETY RECOMMENDATIONS AND INSTRUCTIONS
This product has been designed, produced and tested in Italy, and it left the factory in conditions fully
complying with the current safety standards. To maintain it in safe conditions and ensure correct use,
these general instructions must be fully understood and applied before the product is used.
• In case of devices permanently connected to the power supply, and lacking any fuses or other
devices of mains protection, the power line must be equipped with adequate protection
commensurate to the consumption of all the devices connected to it;
• Devices in Safety Class I, equipped with connection to the power mains by means of cord and plug,
can only be plugged into a socket equipped with a ground wire;
• Any interruption or loosening of the ground wire or of a connecting power cable, inside or outside the
device, will cause a potential risk for the safety of the personnel;
• Ground connections must not be interrupted;
• To prevent the possible danger of electrocution, do not remove any covers, panels or guards
installed on the device, and refer only to NARDA Service Centers if maintenance should be
necessary;
• To maintain adequate protection from fire hazards, replace fuses only with others of the same type
and rating;
• Follow the safety regulations and any additional instructions in this manual to prevent accidents and
damages.
VI Safety Consideration
Dichiarazione di Conformità
EC Declaration of Conformity
In accordo alla Decisione 768/2008/EC, conforme alle direttive EMC 2014/30/UE, Bassa Tensione 2014/35/UE e RoHS
2011/65/UE, ed anche alle norme ISO/IEC 17050-1 e 17050-2.
In accordance with the Decision 768/2008/EC, compliant to the Directives EMC 2014/30/EU, Low Voltage 2014/35/EU
and RoHS 2011/65/EU, also compliant to the ISO/IEC standard 17050-1 and 17050-2
Il costruttore
The manufacturer narda Safety Test Solutions S.r.l. Socio Unico
Indirizzo
Address Via Benessea, 29 / B
EMC - Emissioni:
EMC - Emission: EN 61326-1 (2013)
EMC - Immunità:
EMC - Immunity: EN 61326-1 (2013)
Sicurezza:
Safety: EN 61010-1 (2010)
Bassa Tensione
Low Voltage 2014/35/EU
Compatibiltà Elettromagnetica
EMC 2014/30/EU
RoHS
RoHS 2011/65/EU
Cisano sul Neva, 20 April 2016 Egon Stocca
General Manager
EC Conformity VII
This page has been left blank intentionally
1.2 Introduction This section provides a general overview of EHP-50F Electric and
Magnetic Field Analyzer.
EHP-50F is a low frequency electric and magnetic isotropic field probe-
analyzer. It provides an advanced technology solution for field analysis in
the 1 Hz to 400 kHz frequency range in an extremely high dynamic range.
It includes X, Y and Z axes simultaneous measurements with a powerful,
built in, spectrum analyzer.
EHP-50F includes a non volatile memory which stores frequency and level
calibration tables and an internal optical repeater which allows connection
to external devices through the optical fiber.
Table 1-1 Technical specifications of the EHP-50F Electric and Magnetic Field Analyzer
Electric Field Magnetic Field AUX input (MMCX Zin 1kΩ)
Frequency range 1 Hz ÷ 400 kHz
Measurement range (1) low range 5 mV/m ÷ 1 kV/m 0.3 nT ÷ 100 µT 30 nV ÷ 10 mV
high range 500mV/m ÷ 100 kV/m 30 nT ÷ 10 mT 3 uV ÷ 1 V
(146 dB) (150 dB) (150 dB)
Overload (high range) 200 kV/m (≤ 3 kHz) 20 mT (at 50 Hz) 2V
Dynamic range 106 dB 110 dB 110 dB
Resolution (2) 0.1 mV/m 0.1 nT 0.1 nV
1 mV/m Stand alone 1 nT Stand alone
Displayed average noise level (3)
Isotropic result 5 mV/m 0.3 nT
Single axis 3 mV/m 0.2 nT 30 nV
Flatness (@ 100 V/m, 3 µT, 1mV)
(5 Hz ÷ 40 Hz) ±0.35 dB ±0.50 dB ±0.35 dB
(40 Hz ÷ 300 kHz) ±0.35 dB ±0.35 dB ±0.35 dB
(300 kHz ÷ 400 kHz) ±0.35 dB ±0.50 dB ±0.35 dB
Anisotropicity (typ) 0.54 dB 0.12 dB ---
Linearity (referred to 100 V/m and 3 µT) 0.2 dB (1 V/m ÷ 1 kV/m) 0.2 dB (200 nT ÷ 10 mT) 0.2 dB (10 µV ÷ 1 V)
Internal memory Up to 24 hours regardeless the logging rate
Internal data logger 1 measurement every 30 or 60 seconds
Spectrum analysis method FFT
Time domain analysis method (4) ICNIRP 1998, 2010 and 2013/35/EU weighed peak analysis
Acquisition method Simultaneous three axis acquisition
SPAN 100 Hz, 200 Hz, 500 Hz, 1 kHz, 2 kHz, 10 kHz, 100 kHz, 400 kHz
(500Hz to 400kHz in Stand Alone mode)
Start frequency 1 Hz with SPAN 100 Hz; 1.2 % of the SPAN with wider SPAN
Stop frequency Equal to the SPAN
Immunity to E fields --- < 0.2 uT @ 20 kV/m ---
Immunity to H fields < 10 V/m @ 1 mT --- ---
Calibration (5) internal E2PROM
Typical temperature deviation
-8x10-3 dB/°C within -20 +23 °C
@ 55 Hz referred to 23°C -4x10-3 dB/°C within -20 +55 °C ---
+13x10-3 dB/°C within +23 +55 °C
(@ 50% of relative humidity when applicable)
Typical relative humidity deviation
+11x10-3 dB/% within 10 50 % -7x10-3 dB/% within 10 50 %
@ 55 Hz referred to 50% ---
+22x10-3 dB/% within 50 90 % +10x10-3 dB/% within 50 90 %
(@ 23 °C)
Dimensions 92 x 92 x 109 mm
Weight 550 g
Tripod support Threaded insert ¼”
Environmental protection IP42
Internal battery 3.7 V / 5.4 Ah Li-Ion, rechargeable
Operating time >9 hours in standard mode
24 hours in stand alone mode
Recharging time < 6 hours
External supply 10 ÷ 15 VDC, I = approx. 500 mA
Optical fiber connection Serial Optical Interface 38400 Baud - RP02 connector up to 40 m (USB-OC)
Firmware updating through the optical link by USB or RS232
Self test automatic at power on
Operating temperature -20 to +55 °C
Operating relative humidity (6) 0 to 95 %
Charging temperature 0 to +40°C
Storage temperature -30 to +75°C
(1) For each single axis. Ranges to be selected manually
(2) For the lower measurement range
(3) DANL is frequency and SPAN depending. The specified best performance is referred to f ≥ 50 Hz and SPAN ≤ 1 kHz
(4) The FPGA operates in the time domain, implementing the hardware filters exactly as if the analog signal were passing through a chain of RC filters, being able to reconfigure poles and
zeroes to adapt to legislations.
(5) Recommended re-calibration interval 24 month
(6) Without condensation
Specification are subject to change without notice
1.6.1 Measurements The total uncertainty derived from contributions of linearity, anisotropy,
Uncertainty of EHP-50F frequency response, temperature, relative humidity and with/without
(Spectrum mode) contribution of uncertainty of calibration.
(1) This uncertainty budget is for an ambient temperature of (23 +/- 4) °C, and relative humidity of (50 +/- 5) %
The expanded uncertainty for magnetic flux density for values close to 50 nT is calculated with negligible contribution
of noise level.
(2) The expanded uncertainty of calibration U Cal used is 2,0%
Frequency from 5 Hz to 100 kHz 1 V/m to < 1000 V/m 8.8 10.3 (7)
Frequency from 100 to 400 kHz 1 V/m to < 1000 V/m 9.4 10.8 (7)
(5) This uncertainty budget is for an ambient temperature of (23 +/- 4) °C, and relative humidity of (50 +/- 5) %
(6) The expanded uncertainty of calibration U Cal used is 3.0%
U TOT = 1 . 96 ⋅ (u T EHP 50 F
) + (u )
2
Cal
2
where uT EHP 50 F
is the standard uncertainty
( uT EHP 50 F
= UT EHP 50 F
/ 1.96 )
uCal is the standard uncertainty
( uCal = U Cal / k )
where U Cal is the expanded uncertainty reported on certificate
and k is the coverage factor reported on certificate (normally
k=2 or 1.96)
Note: below 3 Hz and above 300 kHz the uncertainty increases to 0.85 dB (band limiting filter not considered).
S max
A = 20 ⋅ log 10 dB equation (1)
S max ⋅ S min
where S is the measured amplitude in field strength units.
Since the calibration systems are not infinite, these conditions can be
approximated but never reached, and so the evaluation of anisotropy for
our probes shows results worse than the reality. So, the real typical
anisotropy response is better than what specified.
EHP-50F is housed in a small cubic case. The bottom side panel includes
an optical fiber connector, extension rod screw, battery charger connector,
ON/OFF button and Status LED.
The geometric structure of the Narda probes, with sensors placed on the
peripheral faces of the cube, shows the same behavior as one with the
sensors accumulated in the center, when measuring far fields.
Moreover, this provides the opportunity to evaluate also very nearby fields,
albeit with all the limitations that this entails, knowing exactly where the
single sensor is, and allowing to minimize the contribution of the others.
References
[1] ISO/IEC “Guide to the expression of uncertainty in measurement” JCGM 100:2008 GUM 1995
with minor correction – International Organization for Standardization.
[2] CEI-IEC 61786-1 ed. 1.0 “Measurement of DC magnetic, AC magnetic and AC electric fields from
1 Hz to 100 kHz with regard to exposure of human beings – Part 1: Requirements for measuring
instruments.
[3] IEEE Std 1309™-2005 (revision of IEEE Std 1309-1996) “IEEE Standard for Calibration of
Electromagnetic Field Sensors and probes, Excluding Antennas, from 9 kHz to 40 GHz”.
[4] EA European co-operation for Accreditation – Public Reference EA-4/02 M:2013
Key:
1. Led
2. Battery charger connector
3. ON/OFF button
4. AUX input connector MMCX
male type
5. Fiber optic connector
2.3 Work environment The work environment of the Accessories, must come within the following
conditions:
• Temperature From -20°C to +55°C
• Humidity 0 to 95% relative
The meter must be stored in a clean and dry environment, free from acid
dusts and humidity.
The storage environment must come within the range of the following
conditions:
• Temperature From -30°C to + 75°C
• Humidity < 100% relative without condense
2.4 To return for repair When the meter needs to be returned to NARDA for repair, please
complete the questionnaire appended to this User’s Manual, filling in all the
data that will be useful for the service you have requested.
For reducing the period of time required for the repairs, it is necessary to
be as specific as possible in describing the problem. If the problem only
occurs in certain circumstances, please describe in detail how it happens.
If possible it is better to reuse the original packaging; making sure that the
apparatus is wrapped in thick paper or plastic.
Otherwise, use strong packaging by using a sufficient quantity of shock
absorbent material around all sides of the meter to ensure that it is
compact and does not move around inside the package.
In particular, take every precaution to protect the front panels.
Finish the package by sealing it up tightly.
Apply a FRAGILE label to the package to encourage greater care in its
handling.
2.5 To clean the meter Use a dry, clean and non-abrasive cloth for cleaning the meter.
Make a full charging cycle before using the Analyzer for longest battery
operation time.
The minimum voltage level for operation is about 3.25V. The batteries
must be recharged for lower voltages. Below such voltage the
analyzer will turn OFF automatically.
The time required for recharging the batteries is about 4-5 hours.
When the recharge is complete, this is indicated by the Led of the
EHP-50F, with a rapidly blinking Green light.
2.9 EHP-50F with EHP-50F is linked to NBM-550 via the fiber optic link.
NBM-550
2.10 Battery The Table summaries the battery autonomy of the Analyzer depending on
management its mode of operation.
The intensity of the measured field mainly depends on its voltage and
the geometry of the system under examination as well as the distance
between the conductors and the measurement points. In the vicinity
of cables, the reading of the field value may be very high and vary
with the location of the probe.
From the definition of the potential difference between to points:
r2
V 21
= − E dr
r1
It is evident that, keeping the potential difference constant as the
distance between the two points under examination decreases, the
intensity of the field necessarily increases.
For example: the intensity of the electric field between two armatures
of a parallel-plate capacitor situated at a distance of 0.1 m and having
a potential difference of 100 V is equal to:
100V
E= = 1 KV
0,1m m
It should be noted that a voltage of 100 V, in these conditions,
generates a field of 1000 V/m. It is, therefore, possible, in the vicinity
of 220 V conductors, that there may be a field, which is much higher
than 220 V/m.
Using the former optical/RS232 adapter, the selected COM port should be
assigned to the application software (see cap.5 EHP50-TS applications).
The new releases can be downloaded directly from the narda website.
http://www.narda-sts.it
1
EHP50-TS software is now installed in your PC, you can remove it, if
needed, simply running the “Uninstall EHP50-TS” application (see cap.8).
Dongle Utility: allows to enable the function ordered such Options using
the 40 Digit Activation Code received from NARDA Italy.
EHP50 – Stand Alone Mode: main task of this program is to set up EHP-
50F for standalone mode. The analyzer will take and log E or H field
measurements (according to the setting) for up to 24 hours, when switched
on without any connection to external devices (see operating manual for
details). Using EHP-50-Stand Alone mode program is then possible to
download measurement results and store them as a text file.
The User might have the need of administrator privileges to install and
run the software in Windows 7, Win8, Win8.1 and Win10; for further
information see the paragraphs in Chapter 3.
Do not pull the fiber optic by holding onto the cable but use the
connector so that the head does not get damaged.
Avoid dirt and other particles getting into the transducers of the fiber
optic.
4.2.1 Main menu Turn on EHP-50F by pushing the Red POWER button for one second
(ORANGE LED for about 1 second is shown on button press). After that,
the led turns off for 2 seconds, lights orange for about ½ second as an auto-
check and then blinks red.
Communication with PC is shown by the blinking GREEN LED.
Run EHP50-TS application; a welcome screenshot will appear for a few
seconds.
Pressing the POWER down button for more than 4 seconds forces the
hardware of the apparatus to shut down. Then, it is necessary to wait
several seconds before switching it on again.
4-2 Description
Otherwise the program EHP50-TS main window will be shown:
Description:
Span: to select a specific frequency span and, within the selected span, to
set zoom function
Description 4-3
4.3 Mode section Mode section allows setting different acquisition modes, including
Frequency and Time domais choice, as well as defining Electric, Magnetic
field, or both in the same graph, and preferred Unit.
Weighted Peak: Time Domain shows a continue running graph over the
specified period. This is a rolling peak. Once the first time period has been
completed, oldest data are discarded and replaced by the newest ones,
thus showing the values of the latest time period as specified.
Unit: to select unit according to the Input setting. Electric field can be
expressed in V/m and % of the limit only while Magnetic field can be
expressed in A/m, μT (magnetic induction), mG or % of selected limit. Dual
mode units are always V/m and A/m.
The unit “%” can be selected only if a Limit has been activated. Each
spectrum line will then represent the field strength expressed as
percent of the selected limit.
4-4 Description
4.4 Span section
Span: select one of the 8 available spans, keeping in mind that beside the
EHP-50F minimum operating frequency of 1 Hz, the minimum start
frequency of each Span is 1,2% of Span. For example, selecting 1 kHz
span the minimum start frequency should be 12 Hz (automatically adjusted
to the nearer step available of 12.5 Hz).This to avoid that the 0Hz signal,
common to every spectrum analyzer, is included in the measurement result.
Start frequency of 100 Hz and 200 Hz span is therefore 1 Hz (minimum
operating frequency).
Right click and drag on the graph window to define graphically the
Zoom frequency range.
Description 4-5
Fig.4-3 Zoom window
Also choosing the best scale range 120 dB or 140 dB can be helpful to
better distinguish signals and disturbances.
Sweep: the two radio buttons in this box allow the user to select the Linear
or Logarithmic frequency scale.
Logarithmic scale is used to emphathise and clearly display low
frequencies even thow high ones are included in the graph. High
frequencies are therefore compressed to the high side of displayed range.
4-6 Description
4.5 Standard Standard section allows the user to select standard ICNIRP, IEEE, BGV
and EC limits, already included by the software installation, as well as
user’s limits which can be created through the “Make Limit” function
provided by this section.
Safety Standard: This section can be used to load a default standard limit
saved into the memory at the factory. Default limit list shows limits that are
compatible to the actual setting only. If Electric field was selected in the
Mode section, electric field limits only are displayed. No default limit is
shown in case mG unit was selected.
User Defined: This section allows to select and activate a specific limit
among limits created and saved by the user. They can be Linear or Non
Linear (governed by a mathematic formula).
The Edit Limit button opens a window to edit linear limits and create new
ones:
Description 4-7
Type frequency and field strength for each point of the required limit and
click save to save it under the program directory. The limit file will be
created
as a linear interpolation between specified points.
The Edit Non Linear Limit button opens a window to edit those limits that
are related to a mathematic formula, and create new ones:
Type in the Start and Stop frequencies in the appropriate boxes, and
choose the right measurement Unit.
Then choose the formula from the list and type the exponent value.
Click save to save it under the program directory. The limit file will be
created as a calculation of the mathematic expression.
The Max Hold function is not available when the ICNIRP or IEEE limit
is selected.
4-8 Description
4.6 Data section Even if EHP-50F takes measurement over the entire selected span, shown
results are related to the displayed spectrum only, allowing thus detailed
evaluation of user defineable frequency range through the zoom function.
The gray band on the left side of the graph highlights signals below the
minimum start frequency (1,2 % of Span) which are affected by residual
0 Hz peak. Calculation of Wideband result does not include any spectral
line within the gray band.
Marker: you can place a marker anywhere on the spectrum display with a
mouse click
Showed parameters:
Field strength at Marker frequency in the selected Unit (V/m in the above
picture)
Marker frequency (Hz)
% of contribution to the WideBand result
Description 4-9
Four additional frames, Acquisition, Marker, Save and Waterfall, are
included in the Data section:
RMS over: square average calculated over the specified time period (sec)
is shown. This is a moving average. Ones the first time period has been
completed, oldest data is discarded to be replaced by the newest one
showing thus the average value of the latest time period as specified.
Max Hold: the maximum field strength value of each frequency step is
retained and displayed since the Max Hold function has been activated.
Every spectral line is therefore updated only if the new value is grater than
the previously dislayed one showing thus the Maximum of each spectral
line since the Max hold function was activated.
Marker: in this area you can select on which trace to place the Marker:
specific axis (X,Y,Z) or Total result.
Three dedicated buttons, Highest Peak, Next Peak and Previous Peak,
allow Marker positioning over peaks shown by the graph.
If activated, Limit checkbox allows displaying the Limit value corresponding
to the Marker frequency.
Save: three buttons in this area to save spectrum as a picture file (.bmp),
as a text file (.txt) or to copy the spectrum picture to the Windows Clipboard.
4-10 Description
It is also possible to insert a comment.
The user can select the path where the files will be stored.
Description 4-11
Auto save text extract example
Round frequency: when this label is ticked, the frequency indication of the
Marker is corrected to the closest round value, thus avoiding approximation
errors.
The Max Hold function is not available when the ICNIRP or IEEE limit
is selected.
4-12 Description
4.6.1 Waterfall In addition to the spectrum view, another representation has been
introduced in the software, commonly called Waterfall.
The advantage of this view is that the disturbances are shown in a
tridimensional plot. Two dimensions are, as usual, frequency and level, and
the third is the time.
In this way it is possibile to detect and track time-varying signals.
The ranges for Frequency and Level are the same as the Spectrum view.
When pressing the Start Waterfall button, a pop-up message can appear
warning the operator if a file with the same name as the one in use, already
exists.
Press Save As to mantain the original file, or Overwrite to cancel the old
measurement with the starting new one.
In the Graph 2D, the horizontal axis represents Frequency, the vertical axis
is the Time and the colour is the Level.
The Color Scale referenced to levels and units, is indicated on the right
bottom of the screen.
The status-bar shows the number of tracks already acquired and the
maximum available.
On the right two sliders permit to set the preferred Reference Level and
Color Saturation (level dynamic range).
Description 4-13
When entering the Open Waterfall function a screen similar to the following
appears:
Like during the scan, on the left is the spectrogram of the measurements.
The column in the middle, called Events Horizon, reports the Marker with its
frequency and level and many parameters of the measurement setup and
of the probe used.
On the right, in addition to Ref. Level and Color Saturation, there are two
more sliders, useful to set the frequency resolution (Point of Track) and the
time interval (N° track).
It is also possible to Import or Export a full acquisition by pressing the
corresponding button in the Archive box. The name of the measurement
can be typed in the Name box and from the list it is possible to select an
already saved one.
In the Save box two or four buttons (depending on the marker activation)
can be pressed to save:
- a single trace (Levels vs Frequencies)
- a single Frequency (Levels vs Time)
- an image of the spectrogram
- the clipboard
4-14 Description
Fig.4-7 Waterfall - Graph 3D
In the Graph 3D, one axis (blue) represents Frequency, another the Level
(green) and the third the Time (red). So the Frequency can be on the
horizontal axis, the Level on the vertical axis and the Time in depth.
Description 4-15
Another option of the Graph 3D, the Time plot, is available when the
Marker is active.
A new window pops up and the plot represents the level at the marker
frequency versus time.
The horizontal axis of the grid is the time and the vertical axis is the field
level.
The Marker can be moved with the mouse to look for the desired
frequency.
In the 3D view also the dynamic can be selected between 120 and 140 dB,
to help for the best identification of disturbances.
4-16 Description
4.6.1.1 Data recording The subsequent spectra are automatically recorded and saved in a single
file *.WF5 localized in the folder Waterfall in the program root, inclusive of
all the analyzer settings.
The stored files can be recalled when in Data mode or at startup entering
the Open Waterfall window.
The filename is set by the operator prior to starting the test and it is
indicated at the upper right corner of the Waterfall window.
Description 4-17
4.7 Style section This unit concerns the appearance of the program.
Start and End Color buttons allow selection from a color palette
Tick the Disable colored bands box to avoid highlighting the bands for
which the software calculates the ICNIRP values.
4-18 Description
4.8 ICNIRP The International Commission on Non-Ionizing Radiation Protection
establishes guidelines for limiting EMF exposure that could affect human
health.
Standard section allows the user to select standard ICNIRP limits, already
included by the software installation.
The calculation of the ICNIRP value is performed if all three axes are
enabled
Description 4-19
In the screenshot above it is depicted an example how the software shows
the calculation of the ICNIRP value for the Electric Field measurement.
4-20 Description
4.9 IEEE One of the purposes of the Institute of Electrical and Electronic
Engineers, Inc. ("IEEE") is to establish exposure standards.
Standard section allows the user to select standard IEEE limits, already
included by the software installation.
When the IEEE limit is selected, the software automatically calculates the
corresponding total integration of the measured signals, and compares the
result with the threshold set by the standard, considering whether it is or is
not exceeded.
The calculation of the IEEE value is performed if all three axes are
enabled
Description 4-21
In the screenshot above it is depicted an example how the software shows
the calculation of the IEEE value for the Electric Field measurement.
4-22 Description
4.10 Weighted Peak The EHP-50F Weighted Peak (Time Domain) option performs test in the
Function range 1 Hz to 400 kHz to comply with the informative annex of the ICNIRP
(Time Domain) Guidelines 2010 (appendix “Determination of the weighted peak
exposure”) for coherent signal source.
The current legislation proposes limits for electromagnetic fields, the value
of which depends on the frequency.
When exposed to a sinusoidal field, it is relatively simple to relate the
measurement of the field to the reference limit, as the frequency is known
or easily measurable.
In case of multi-frequency fields, pulsed, non-sinusoidal waveforms, the
exposure assessment in relation to the limit becomes more complex.
In these cases, typical of industrial environments, it is not possible to relate
the amplitude of the field to the limit value at a specific frequency, as the
spectral content of the signal can be very complex (many frequencies are
present at the same time).
The ICNIRP guidelines propose a method for the evaluation of multi-
frequency fields, that consists of the sum of the field strength of the
individual spectral components related to the limit corresponding to the
frequency of each individual component. This method is appropriate in the
presence of incoherent components.
4.10.1 Weighted Peak The Weighted Peak function needs the following in the list order:
requirements
- In the Mode Section, select the Function “Weighted Peak (Time Domain)”
- In the Standard section set one of the available standards, such as, for
example, the ICNIRP 2010 limits for occupational or general public
exposure.
- In the Mode section select one of the ranges among the Electric 1 kV/m,
Electric 100 kV/m, Magnetic 100uT or Magnetic 10mT.
- In the Data section, enter the desired Hold Time.
Description 4-23
4.10.2 EHP50-TS Settings The Standard section allows the user to select one of the standard
ICNIRP or EU limits for occupational or general public exposure, already
included by the software installation.
In the Mode section you can select Electric 1 kV/m or Electric 100 kV/m
or Magnetic 100uT or Magnetic 10mT.
The Span section displays the frequency range, which is always full range
1 Hz to 400 kHz in Weighted Peak Mode.
The Data section allows the user to read the measurements for the highest
peak and Marker, to choose the axis and set the desired Hold Time.
4-24 Description
4.10.3 Weighted Peak Once the Weighted Peak has been enabled, the software shows the result
mode of the calculation and its graph over the time.
The Hold Time is the time resolution of the graph. During the Hold Time
the Weighted Peak results will be collected and the maximum value will be
shown in the graph.
Description 4-25
The minimum hold time is of 500 milliseconds.
The time span of the graph is 200 times the hold time. The rolling
measurement memory deletes all values that date back longer than
the set time period.
The limit selected is shown in the higher right corner of the graph.
The full scale is automatically adapted to the field and the limit selected.
In the centre, above the grid of the graph, the time indication is the actual
time, the instant in which the measurement is made, at the end of the hold
time step.
4-26 Description
4.10.3.1 Noise indication The User should make use the following information to choose the proper
setting for the desired measurement.
For every situation, the best choice is to adopt the lower Full Scale
available (for example 100 µT) until the meter indicates an over-range
warning; when it happens, the User should switch the meter to the higher
Full Scale setting (as for the example above, 10 mT).
The table above shows the noise for each standard and setting, for Electric
field measurements.
The table above shows the noise for each standard and setting, for
magnetic field measurements.
Description 4-27
For convenience, a noise threshold has been established, under
which the field indication becomes “LOW”, as in the example below.
4-28 Description
5 - EHP-50F Stand alone mode
5.1 Stand alone mode EHP-50F has been designed to be used also in stand-alone mode. This
description mode is useful because no external devices is necessary to the analyzer
during measurements. Once the measurement parameters have been
programmed through a PC, the EHP-50F analyzer can start its acquisition
by storing the data over 24 hours in a stand-alone mode. It is necessary to
position it over the TR-02A tripod and to activate the start. After 24 hours it
will stop automatically and later it would be possible to download all the
data to the PC by using the provided EHP50-TS PC software. From PC it
is possible to select to measure the electric or magnetic field, to select the
range, the Highest or Wideband mode, the desired SPAN and the
sampling rate of one minute or 30 seconds.
Some typical applications are:
- Magnetic fields near high, medium and low voltage transformers
- Measurements in proximity of power line
- Safety measurement at worker’s site
- Measurements close to machines, air conditioning systems, home
appliances etc.
- Development of new products
The acquisition can be terminated earlier than 24 hours. Just turn off
the EHP-50F and run the application to download the data to the PC.
The Read Logger will not be available unless the communication between
the PC and the analyzer has been established. Click this button to read the
actual EHP-50F setup.
The Set Logger window will not be available unless you change some
parameters. The Set Logger will be active and available for the user to
transfer all measurement parameters (setup) to the EHP-50F internal
memory.
Answering YES, the date & time of your PC will be transferred inside the
analyzer.
Chronicle: the data are ordered by increasing time like in the following
example:
Median: the data are ordered from the lowest to the highest WIDE T value
like in the following example:
Push the button Download to transfer all data from the analyzer to the PC
To save the data into your own PC directory, it is necessary to activate the
button Save.
Data will be saved in TXT format according to the following window:
5.2.3 EHP-50F battery If the battery charger is plugged to the EHP-50F while the software is
charging running, the analyzer will be disconnected and the software will display:
This section provides all the information required for firmware updating.
6.2 To run the Turn off the EHP-50F and connect it to a free USB or RS232 port of the
update software PC.
Run EHP50F Update Firmware to start the update program.
The User should check that the battery level is at least 3.9 V before
starting the upgrade.
6.3 To transfer data Main window displayed after the updating program EHP-50F Update
Firmware has been run:
Turn the EHP-50F OFF (it could seem to be already OFF but it is not) and
turn it ON again.
The EHP-50F is now updated with the new version of the internal firmware
or FPGA.
It is now possible to disconnect the cable connected to the PC, with the
EHP-50F meter either switched on or off.
Subsequently, when the meter is switched on again, the new version will be
displayed in the NBM-550, EHP50-TS or EHP-50F Stand Alone Mode
application software.
Before removing any shared system file, the uninstaller will ask for a
confirmation.
Answer “NO” in case you are not sure whether the showed system file is
required for other applications.
Click “finish” to exit the uninstaller, USB driver is now removed from your
system.
9.1.2 Work environment Unless otherwise specified, the work environment of the Accessories, must
come within the following conditions:
The Accessories must be stored in a clean and dry environment, free from
dust, acids and humidity.
The storage environment must come within the range of the following
conditions:
9.1.3 Return for repair When the Accessories need to be returned to NARDA for repair, please
complete the questionnaire appended to this User’s Manual, filling in all the
data that will be useful for the service you have requested.
For reducing the period of time required for the repairs, it is necessary to
be as specific as possible in describing the problem. If the problem only
occurs in certain circumstances, please describe in detail how it happens.
If possible it is better to reuse the original packaging; making sure that the
apparatus is wrapped in thick paper or plastic.
Otherwise, use strong packaging by using a sufficient quantity of shock
absorbent material around all sides of the product to ensure that it is
compact and does not move around inside the package.
In particular, take every precaution to protect the front panels.
Finish the package by sealing it up tightly.
Apply a FRAGILE label to the package to encourage greater care in its
handling.
9.1.4 Cleaning Use a dry, clean and non-abrasive cloth for cleaning the instruments.
Battery charger:
output: DC, 10 - 15 V, ~ 500 mA
- +
Connector:
9-2 Accessories
9.2 NBM-550 Controller and Display Unit for EHP-50F
EHP-50F can be easily operated using the Broadband Field Meter NBM-
550 as a controller and display unit. In order not to influence the field to be
measured, communication between the EHP-50F Analyzer and the NBM
Unit is through a rugged optical fiber cable. The NBM-550 supports the
following settings and measurement modes:
• Electric or magnetic field selection (1 kV/m; 100 kV/m; 100 µT; 10 mT)
• Span selection (100 Hz 400 kHz)
• Wideband or highest peak modes
• Spectrum Mode
• Monitor Mode (Actual; Max; Avg; Min)
• XYZ Mode (wideband or highest peak)
• Weighted Peak (WPM) ranging from 1 Hz to 400 kHz
Accessories 9-3
NBM-550 screenshots of EHP-50F results:
Fig. 9-7 Weighted Peak Mode Fig. 9-8 Weighted Peak Marker
9-4 Accessories
9.3 USB-OC Optical USB Converter
9.3.1 Introduction USB-OC is standard accessory of the EHP-50F Analyzer.
It converts the signals of some of the system’s accessories, which are only
connected via fiber optic, into USB-compatible signals. It, therefore, makes
it possible to link the following items up to the USB port of any Personal
Computer to operate them in conjunction with specific application software
and for firmware updating:
• EHP-50F Electric and Magnetic Field Analyzers, 1 Hz – 400 kHz
• EHP-200A Electric and Magnetic Field Analyzers, 9 kHz – 30 MHz
• NBM-550 Broadband meter
9.3.2 Installation Insert USB-OC in the connector of a free USB port of the PC, connect the
fiber optic coming from the probe or other Accessories treating the locating
key with care.
Considering the very low consumption of the device, the power required by
USB-OC is taken directly from the USB port of the PC. This means no
maintenance is needed.
The link between USB-OC and a HUB USB device or USB cable
extension could not work properly. Connect the USB-OC to the PC
directly.
Front view
Key:
Rear view
Key:
Power supply USB-OC is powered directly from the USB port of the PC.
Accessories 9-5
This page has been intentionally left blank
9-6 Accessories
NARDA
Sales & Support: Manufacturing Plant:
Safety Via Leonardo da Vinci, 21/23 Via Benessea, 29/B
Test 20090 Segrate (MI) - ITALY 17035 Cisano sul Neva (SV)
Solutions Tel.: +39 02 2699871 Tel.: +39 0182 58641
S.r.l. Socio Unico Fax: +39 02 26998700 Fax: +39 0182 586400
http://www.narda-sts.it
Mod. 18-1
Caro cliente
grazie per aver acquistato un prodotto NARDA! Sei in possesso di uno strumento che per molti anni ti garantirà un’alta qualità di servizio. NARDA riconosce
l'importanza del Cliente come ragione di esistenza; ciascun commento e suggerimento, sottoposto all'attenzione della nostra organizzazione, è tenuto in grande
considerazione. La nostra qualità è alla ricerca del miglioramento continuo. Se uno dei Suoi strumenti NARDA necessita di riparazione o calibrazione, può aiutarci
a servirla più efficacemente compilando questa scheda e accludendola all’apparecchio.
Tuttavia, anche questo prodotto diventerà obsoleto. In questo caso, ti ricordiamo che lo smaltimento dell'apparecchiatura deve essere fatto in conformità con i
regolamenti locali. Questo prodotto è conforme alle direttive WEEE dell’Unione Europea (2002/96/EC) ed appartiene alla categoria 9 (strumenti di controllo). Lo
smaltimento, in un ambiente adeguato, può avvenire anche attraverso la restituzione del prodotto alla NARDA senza sostenere alcuna spesa. Può ottenere ulteriori
informazioni contattando i venditori NARDA o visitando il nostro sito Web www.narda-sts.it.
Dear Customer
thank you for purchasing a NARDA product! You now own a high-quality instrument that will give you many years of reliable service. NARDA recognizes the
importance of the Customer as reason of existence; in this view, any comment and suggestion you would like to submit to the attention of our service organization
is kept in great consideration. Moreover, we are continuously improving our quality, but we know this is a never ending process. We would be glad if our present
efforts are pleasing you. Should one of your pieces of NARDA equipment need servicing you can help us serve you more effectively filling out this card and
enclosing it with the product.
Nevertheless, even this product will become obsolete. When that time comes, please remember that electronic equipment must be disposed of in accordance with
local regulations. This product conforms to the WEEE Directive of the European Union
(2002/96/EC) and belongs to Category 9 (Monitoring and Control Instruments). You can return the instrument to us free of charge for proper environment friendly
disposal. You can obtain further information from your local NARDA Sales Partner or by visiting our website at www.narda-sts.it.
Servizio richiesto: Service needed:
Solo taratura Riparazione Riparazione & Taratura Taratura SIT Altro:
Calibration only Repair Repair & Calibration Certified Calibration Other:
Ditta:
Company:
Indirizzo:
Address:
Persona da contattare: Telefono:
Technical contact person: Phone n.
Modello: Numero di serie:
Equipment model: Serial n.