Hematology Lab Testing Guide
Hematology Lab Testing Guide
- a branch of medicine concerning     - not applicable for ESR               conducts aperture current, and
the study of blood, the blood-        determination
                                                                             rinses instrument components
forming organs, and blood
                                      1. Beckman Coulter Ac.T 5 Diff         between analyses.
diseases.
                                      Analyzer
                                                                             used to lyse blood cells and to
- In the medical field, it includes
                                      > a fully automated hematology         determine
the treatment of blood disorders
and malignancies including types      analyzer providing a complete          hemoglobin concentration
of hemophilia, leukemia,
lymphoma and sickle cell anemia.      WBC five-part differential, which is   Reagent System
What specimen?                        > CBC: 48 hours @ room temp.           differentially stain granules of
                                                                             monocytes,
TEST                                  > Diff:
                                                                             neutrophils, and eosinophils
Complete Blood Count (CBC)            24 hours @ room temp.
                                                                             Used as a rinsing agent
Erythrocyte Sedimentation Rate        Reagent System
(ESR)                                                                        Ac.T 5 Diff Control Plus
                                      Reagent
Reticulocyte Count                                                           > A hematology quality control
                                      1. Diluent, 20 L                       material
Prothrombin Time
                                      2. Hgb Lyse,                           used to monitor the performance
Activated Partial Thromboplastin
                                      400 mL                                 of
Time
                                      EACMed                                 Ac.T 5 Diff analyzers in conjunction
D-dimer
                                      Function                               with the instrument's reagent
Clotting Time & Bleeding Time
                                                                             system
Malarial Smear                        used for counting and
                                      differentiating blood cells            >
Peripheral Blood Smear (PBS)
                                      dilutes whole-blood samples            Available in three levels (low,
Hematology Section Process Flow                                              normal,
                                      stabilizes cell membrane for
Turn around time                      accurate counting and                  and high)
                                                                                                                JMSE
>                                 • a quantitative, automated         Isotonic solution that dilutes the
                                                                      sample
Open stability: 15 days           hematology analyzer and
                                                                      Stabilizes cell membranes
Storage: 2-8°C                    WBC differential counter for
                                                                      V Conducts aperture current
Machine Linearity                 invitro diagnostic use in
                                                                      Carries and focuses the VCS
Parameter                         clinical laboratory.
                                                                      sample stream
Unit                              • Modes of operation:
                                                                      Rinses the system between
Linearity Range                   1. Primary mode                     samples
‡ 5 or + 5 % LYSE Ill lytic reagent Cleans and rinses the diluter parts
                                                                                                              JMSE
Calculated: (RBC x MCV) / 10    a. D-Dimer                            1.2. Run three levels of cell
                                                                      controls (low, normal and high).
Impedance with hydrotocus       b. von Willebrand Factor - Activity
                                                                      Results of
                                and Antigen
Derived from WBC scatterplot
                                                                      which must be within the range
                                C.
using VCS technology                                                  specified in the product insert.
                                Free Protein S
3. ACL ELITE PRO                                                      1.3. Enter the sample ID.
                                4.
a fully automated                                                     1.4. Press CBC/Diff to select the
                                Special tests                         desired analysis mode (CBC or
coagulation system for                                                CBC/Diff).
                                a.
specific clinical use in                                              The mode selected appears on the
                                LAC Screen and Confirm                screen.
hemostasis laboratory for
                                b.                                    1.5. Gently mix the sample by
clotting, chromogenic,
                                Silica Clotting Time Screen and       invertion for about 6-8 times.
special assay and                                                     Make sure the
                                Confirm
immunological testing.                                                sample is not clotted.
                                SAMPLE TRAY
ACL ELITE PRO                                                         1.6. Remove the cap of the sample
                                ~ 40 samples
used to perform the following                                         tub.
                                ~ 0.5 and 2.0 sample cups
test/s                                                                1.7. Present the sample to the
                                ~ full draw tubes                     probe and press the aspirate
1. Coagulometric tests
                                REAGENT POSITIONS                     switch. The
a. Prothrombin Time
                                Up to 22 total positions              LEDs flash during sample aspirat
1
                                1. COMPLETE BLOOD COUNT               1.8. When the red LED remains
b. Activated Partial                                                  illuminated. remove the tube from
                                (AUTOMATED)
Thromboplastin Time                                                   the
                                1. Ac.T 5 Diff Analyzer
C.                                                                    probe. When the green LED
                                1.1. Turn the instrument ON. The      remains illuminated, the
Thrombin Time
                                instrument performs the Startup       instrument is
2. Absorbance tests             routine.
                                                                      ready for the next analysis.
a.                              The Startup must be successful or
                                passed before sample analysis         1.9. The sample results appear on
Antithrombin                                                          the screen.
                                shall start.
b.                                                                    1.10. Retrieve the results using the
                                STARTUP ACCEPTABLE VALUE:             sample ID in the interface. Fill in
Protein C
                                                                      the
                                - WBC: 0.3 x 103/L
C.
                                                                      necessary data.
                                - RBC: 0.03 x 105/L
Plasminogen
                                                                      1.11. Validate the results
                                - Hemoglobin: 0.3 g/dL
d
                                                                      1.12. Print results.
                                - Platelet: 7.0 x 10¾/L
Fibrinogen
                                                                      2. HmX AL Analyzer
                                I. COMPLETE BLOOD COUNT
Immunological tests             (AUTOMATED)                           2.1. Machine start up
                                                                                                       JMSE
2.1.1. Turn on the machine by         22.1.9. Press the "Esc" key.         2.2.2.5. Press F3 (Secondary)
pressing the switch at the back of
                                      2.2.1.10. Load the cassette into     2.2.2.6. Mix the specimen gently
the
                                      load elevator of the machine.        but thoroughly.
analyzer.
                                      2.2.1.11. Click "Sample Analysis.    2.2.2.7. Remove the cap of the
2.1.2. Turn on the monitor for the                                         sample tube.
                                      2.2.1.12. Click "Run Samples"
Data Management System (DMS)
                                                                           2.2.2.8. Immerse the aspirator tip
by                                    2.2.1.13. Press "Enter"
                                                                           completely in the tube.
pressing the switch on the side of    2.2.1.14. Press F3 (Run).
                                                                           2.2.2.9. Press and release the
the instrument.
                                      2.2.1.15. Press F2 ( Start           sample bar.
2.1.3. Check status of reagents.
                                      2.2.2. Secondary Mode                2.2.2.10. Listen for the beep
2.1.4. Check the waste container.                                          before removing the tube.
                                      2.2.2.1. From the Main Menu,
2.1.5. On the Main Menu of the        select "Sample Analysis.             2.3. Retrieve the result using the
DMS, select "Diluter Functions"                                            sample ID in the interface. Fill in
then                                  2.2.2.2. Click "Run Samples"         the
"Start Up" to begin machine start     2.2.2.3. Type the sample ID and      necessary data.
up. It must be successful or          patient name then press "Enter"
                                                                           2.4. Validate the results.
passed before starting the analy      key.
                                                                           2.5. Print the results at the back of
2.2. Running of samples               2.2.2.4. Press F3 (Run)              the transaction slip.
2.2.1.1. Put the specimen tube        2.2.2.6. Mix the specimen gently     2.2.2.1. From the Main Menu,
into the cassette making sure any     but thoroughly.                      select "Sample Analysis.
bar code labels are visible through   2.2.2.7. Remove the cap of the       2.2.2.2. Click "Run Samples"
the top of the                        sample tube.
                                                                           2.2.2.3. Type the sample ID and
cassette.                             2.2.2.8. Immerse the aspirator tip   patient name then press "Enter"
                                      completely in the tube.
2.2.1.2. From the Main Menu,                                               key.
select "Sample Analysis." Click       2.2.2.9. Press and release the
                                      sample bar.                          2.2.2.4. Press F3 (Run)
"Worklist."
                                      2.2.2.10. Listen for the beep        2.2.2.5. Press F3 (Secondary)
2.2.1.3. Press "F12" (Seq)            before removing the tube.
                                                                           2.2.2.6. Mix the specimen gently
2.2.1.4. Press the "Enter" key        2.2.2. Secondary Mode                but thoroughly.
twice
                                      2.2.2.1. From the Main Menu,         2.2.2.7. Remove the cap of the
2.2.1.5. Input the cassette number    select "Sample Analysis.             sample tube.
in use.
                                      2.2.2.2. Click "Run Samples"         2.2.2.8. Immerse the aspirator tip
2.2.1.6. Press "F12" (Seg)                                                 completely in the tube.
                                      2.2.2.3. Type the sample ID and
2.2.1.7. Press the "Ent               patient name then press "Enter"      2.2.2.9. Press and release the
                                                                           sample bar.
2.2.1.8. Input appropriate patient    key-
identifiers (sample ID and                                                 2.2.2.10. Listen for the beep
                                      2.2.2.4. Press F3 (Run)              before removing the tube.
patient name).
                                                                                                            JMSE
2.3. Retrieve the result using the      1.7. When the red LED remains           2.3.5.2. Press and release the
sample ID in the interface. Fill in     illuminated, remove the tube from       sample bar.
the                                     the
                                                                                2.3.5.3. Listen for the beep or
necessary data.                         probe. When the green LED               verify the status line message
                                        remains illuminated, the
2.4. Validate the results.                                                      changes from "Aspirating" to
                                        instrument is
                                                                                "Diluting" before removing
2.5. Print the results at the back of
                                        ready for the next analysis.
the transaction slip.                                                           the bottle.
                                        1.8. Repeat steps 4 through 8 until
• RUNNING OF CELL CONTROL                                                       2.3.6. After the analysis, review
                                        you have run all three levels of cell
                                                                                the primer results at the DMS.
1. Ac.T 5 Diff Analyzer
                                        control.
                                                                                2.3.6.1. If count is unacceptable,
1.1. Turn the instruments ON. The
                                        1.9. Review the control results to      rerun the Latron primer.
instrument performs the Startup
                                        ensure they are within the
routine                                                                         2.3.6.2. If count is acceptable, run
                                        acceptable
                                                                                Latron Control.
The Startup must be successful or
                                        ranges.
passed before sample analysis                                                   2.4. Running of Latron Control
shall                                   1.10. Print quality control result.
                                                                                2.4.1. Press the "Esc" ley to close
start.                                  1.11. Log the results in the            the primer Run Window.
                                        designated log she
1.2. Enter the cell control level                                               2.4.2. Press "F3" (Run).
(LOW. NORMAL OR HIGH) as the            2. Hm AL Analyzer
                                                                                2.4.3. Press "F3" [Control(Seconda
sample
                                        2.1. Ensure that start up is
                                                                                2.4.4. Gently invert the Latron
ID.                                     complete.
                                                                                Control bottle 5 to 8 times,
1.3. Press CBC/Diff to select the       2.2. Bring Latron Primer and
                                                                                2.4.5. Run the Latron Control in
desired analysis mode (CBC or           Control at room temperature.
                                                                                the Secondary mode.
CBC/Diff).
                                        2.3. Running of Latron Primer
                                                                                2.4.5.1. Immerse the aspirator tip
The mode selected appears on the
                                        2.3.1. From the Main menu of the        completely in the Primer. (Do
screen.
                                        DMS software. select "Controls."
                                                                                not place the tip against the side
1.4. Gently mix the sample by
                                        2.3.2. Select "Control Run"             of the bottle.)
inversion for about 6-8 times.
Inspect the                             2.3.3. Press the "F2" key (File).       2.4.5.2. Press and release the
                                                                                sample bar.
vial's contents to ensure that all      2.3.4. Select the current Latron
cells are uniformly distributed.        File (active lot number)                2.4.5.3 .Listen for the beep or
                                                                                verify the status line message
1.5. Remove the cap of the sample       235. Press "F3" (Run).
tube.                                                                           changes from "Aspirating" to
                                        2.3.4. Press F4 (Primer).
                                                                                "Diluting" before removing
1.6. Present the sample to the
probe and press the aspirate            2.3.5. Run the Latron Primer in the
                                                                                the bottle.
switch. The                             Secondary mode.
                                                                                2.4.5.4. After the analysis, review
LEDs flash during sample                2.3.5.1. Immerse the aspirator tip
                                                                                the the results in the Control
aspiration.                             completely in the Primer. (Do
                                                                                Run Screen.
                                        not place the tip against the side
                                        of the botti                            2.5. Preparation of 5C cell control
                                                                                                                  JMSE
2.5.1. Warm control vials at          - Reconstitute with 8.5 mL PT        1.1. PT Fibrinogen HS Plus
ambient temperature for 10-15         Buffer (5 days upon
                                                                           - Reconstitute with 8.5 mL PT
minutes                               reconstitution).
                                                                           Buffer (5 days upon
EAGMed before mixing.                 1.2. APTT Reagent Synthasil          reconstitution).
2.5.2. Mix gently by hand using the   - Liquid ready reagent and has 30    1.2. APTT Reagent Synthasil
&x8x8 method twice                    days stability upon opening.
                                                                           - Liquid ready reagent and has 30
2.5.3. After the second mixing,       1.3. Normal Control Assayed          days stability upon opening.
check the sides and bottom of the     (Normal Control)
                                                                           1.3. Normal Control Assayed
tube
                                      - Reconstitute with 1 mL distilled   (Normal Control)
to verify the control is mixed        water (24 hour stability upon
                                                                           - Reconstitute with 1 mL distilled
completely.
                                      reconstitution).                     water (24 hour stability upon
2.6. Running of 5C Cell Control
                                      1.4. Calibration Plasma (Normal      reconstitution).
2.6.1. From the Main menu of the      pool)
                                                                           1.4. Calibration Plasma (Normal
DMS software. select "Controls.
                                      - Reconstitute with 1 mL distilled   pool)
2.6.2. Select "Control Run"           water (24 hour stability upon
                                                                           - Reconstitute with 1 mL distilled
2.6.3. Press the "F2" key (File).     Reagent preparation                  water (24 hour stability upon
(24 hour stability upon                i.e., the end that has not come        VI. LEUKOCYTE NUMBER
reconstitution).                       into                                   CONCENTRATION
2. Check status of reagents.           contact with the blood, with soft      1. Place a dampened gauze in the
                                       wax or plastic modeling clay.          bottom of petri dish with two
3. Running of control.                 Check                                  halves of
3.1. On the main screen, click         that it is completely plugged to a     wooden stick placed on the gauze
"Analysis"                             depth of about 2mm.                    to serve as stand for the
3.2. Click "Multi-test session" or     3.                                     hemocytometer.
"Single test session",
                                                                                                               JMSE
2. Place a whole blood in a rotor     remove and clean the coverslip;       1. Collect a drop of blood of about
for at least 1 to 2 minutes or hand   clean the counting chamber; and       4mm in diameter on one end of
invert                                refill                                the
3. To make a 1:20 dilution, attach    11. Leave the counting chamber        2. Hold the slides with one hand.
a suction device to the pipette and   on the bench for about 3 minutes      Using the other hand, place the
draw a                                to allow                              edge of
sample of whole blood to exactly      the cells to settle. Be careful it    the spreader just in front of the
0.5 mark on pipette stem. Do not      does not dry up                       drop of blood.
allow
                                      12. Place the chamber on the          3. Draw the spreader back until it
air bubbles to enter the pipette.     stage of the microscope. Use the x    touches the drop of blood.
                                      10
4. Tilt pipette so that the stem is                                         4. Let the blood run along the
slightly above horizontal and wipe    objective.                            edge of the spreader.
stem
                                      13. Count the leukocyte in area of    5. Push the spreader to the end of
carefully with slightly dampened      4mm3 of the chamber. Include in       the slides with a smooth
gauze, moving from pipette bulb       the count                             movement (all
toward
                                      the leukocytes seen on the lines of   the blood should be used).
tip                                   two sides of each square counted,
                                                                            6. Air dry smear. Adequate drying
5Gently rotating pipette between      using the letter "L" or "." as your   is essential to preserve the quality
forefinger and thumb, draw            guide. This represents one of the     of the
diluent                               four
                                                                            smear.
steadily into pipettes exactly to     squares counted. Repeat the
                                                                            7. Write with a lead pencil on the
the 11 mark.                          procedure in the three other.
                                                                            thick part of the smear not used
6. Immediately cover tip of pipette   14. Calculate the number of           for
with a finger and remove the          leukocytes in 1 liter of blood by
                                                                            examination patient's name and
suction                               multiplying the
                                                                            date of examination.
device.                               number of leukocytes counted in
                                                                            8. Fix the thin smear with
                                      the four squares by 0.05. Report
7. Shake pipette to mix by hand.                                            methanol for 2-3 minutes.
                                      the
8.                                                                          9. Stain according to the
                                      result as "number" by 109
                                                                            manufacturer's instructions.
Place a coverslip on the
                                      Example:
hemocytometer.                                                              10. Wash the stain off in a stream
                                      Number of leukocytes counted in       of buffered water. Do not tip the
9. Discard the first four drops.
                                      4 squares = 188                       stain off as
10. Dispense the mixture into the
                                      Number of leukocytes per liter =      thin will leave a deposit of stain on
counting chamber. If the liquid
                                      (188 x 0.05) x 109                    the smear.
overflows
                                      Result Reported: 9.4 x 109            11. Leave clean water on the slide
into the channel between the two
                                                                            for 2-3 minutes to differentiate
chambers, you must start again:       VI. PREPARATION AND STAINING
                                                                            the smear.
                                      OF
                                      BLOOD SMEAR
                                                                                                             JMSE
12. Tip off the water and place the   and never substitute other types     1. Using at least 30 specimens
slide in a draining rack to dry.      of lubricants for immersion oil,     from healthy subjects, perform
                                      since                                machine
13.Examine the color of the
smear. It should appear mauve-        substitutes may damage optical       platelet count in a hematology
neither too blue                      lenses.                              analyzer. Record results.
14.3. The smear must not be too       1. Count the number of               number per single oil-immersion
long and too thick.                   identifiable platelets in 10 oil     field
                                      immersion fields by
14.4. The smear must not contain                                           Platelet Estimation Factor
holes because a greasy slide          using only the area where 50% of
                                                                           5. For each specimen, divide the
                                      the red cells are overlapping in
has been used                                                              corresponding machine count by
                                      doublets or triplets.                the
VIII.LEUKOCYTE DIFFERENTIAL
COUNT                                 2. Divide the total number of        average number of platelets per
                                      platelets counted by 10 to obtain    oil-immersion field to obtain
1. Check slide identification.                                             individual
                                      the average
2. Perform low power scan to                                               platelet estimation factors (PEF)
                                      number per field.
review blood smear adequacy.
                                      3. Multiply the average number by    6. Add all individual PEF from 30
2.1. Check feather edge for fibrin                                         specimens and divide by 30 to
                                      the platelet estimation factor.
clots.                                                                     obtain the
                                      4
2.2. Examine smear edges for                                               average PEF. Round-off the
possible aggregation of               The number obtained is reported      number to the nearest digit and
leukocytes.                           as Platelet Estimate:                use this
2.3. Verify stain quality             x 109                                number as the PEF.
2.4. Note also erythrocyte            Platelet Estimation Factor           Formula:
distribution pattern, size and
                                      > derived by correlating a machine   MC/SC = iPEF
shapes.
                                      count and a smear count, as
3. Perform oil-immersion              follows:                             PEF = Total iPEF ÷ 30
examination of the blood smear.
                                                                           Legend:
Use oil sparingly
                                                                                                          JMSE
MC = Machine Count                    oxalated vacuum tube. Mix well            6. At interval of 30 seconds, blot
                                      but gently.                               the blood which has oozed out
SC = Smear Count
                                                                                with
                                      2.
¡PEF = Individual PEF
                                                                                absorbent or filter paper.
                                      Uncap the tube and insert gently
X. RETICULOCYTE COUNT
                                      the commercially available                7. Record the time until the
1. Place three drops each of          Westergren tube                           bleeding stop
reagent and blood sample in a test
                                      unto the tube with venous tube.           XIII.CLOTTING TIME
tube.
                                      3.                                        1. Obtain four drops of blood by
2. Mix blood and reagent by
                                                                                prick method and place them
swirling the tube.                    Check that the Westergren tube is
                                                                                separately on
                                      filled with blood up to the zero
3. Allow the blood and reagent
                                      mark.                                     a clean slide. Time simultaneously.
mixture to stand for 15-20
minutes.                              4.                                        2. At intervals of one minute, draw
                                                                                the tip of the needle to each of the
4.                                    Allow the tube to stand in a
                                                                                drops
                                      vertical position in the Westergern
With the use of a pasteur pipette
                                      rack exactly one                          successively. The appearance of
(or any of equivalent), place 1
                                                                                fibin at the point of the needle is
drop of                               hour at room temperature
                                                                                the
                                      without vibration or exposure to
the mixture on one side of clean
                                      direct sunlight.                          coaqulation time
slide.
                                      5.                                        3. Record the results.
5. Prepare a thin blood dye film in
usual manner and allow to dry.        After exactly 60 minutes or one           CBC Reference Range
                                      hour, the distance from the O
6. Using oil immersion lens, count                                              Parameter
                                      mark to the top of
the number of reticulocytes per
                                                                                Unit
1000                                  the column of red cells is recorded
                                      in milliliters as the ESR value. If the   Screenshot (74).png
erythrocytes counted and record.
                                      demarcation between plasma and            Reference Range
7. Calculate the percent
                                      the red cell column is hazy, the
erythrocytes using the following                                                2
                                      level is taken
formula:
                                      where the full density is firts appar     NATIONAL EXTERNAL QUALITY
Number of reticulocyte counted y                                                ASSESSMENT
1000 = Reticulocyte ct. in %          XI1. BLEEDING TIME
                                                                                SCHEME (NEQAS) ENROLLMENT
1000 RBC                              1. Clean the ring finger of the
                                      patient with cotton and alcohol.          > aims to help laboratories attain
XI. ERYTHROCYTE                                                                 accurate and reliable analytical
SEDIMENTATION RATE                    2. Let it dry.
                                                                                results and achieve inter-
(ESR)                                 3. Make a deep puncture on the            laboratory comparability as well.
                                      tip of the ring finger using sterile
1.                                                                              > one of the requirements for the
                                      lancet.
                                                                                renewal of the License to
Collect venous blood into an EDTA
                                      4. Discard the first drop
(Ethylene Diamine Tetraacetic                                                   Operate (LTO) of Clinical
Acid) tube or                         5. Start timer.                           Laboratories of all category.
                                                                                                                JMSE
> Reference Laboratory for
Hematology: National Kidney and
Transplant Institut
your Laboratory.
JMSE