Rexmed Autoclave

This document provides specifications and operating instructions for the REXMED RAU-530D autoclave. It lists the sterilization time and temperature for a 1.5 liter water supply as 78 minutes total, broken down by phase. It also provides recommended sterilization exposure times at 121°C for various materials. Standard accessories include 2 wire baskets, 1 heating element, and 1 silicone door seal. Specifications include an AC 110/230V power supply, 2KW power, a 30x71cm stainless steel chamber, and a capacity of 50 liters.

Uploaded by

Shaif Zahrah
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
289 views2 pages

Rexmed Autoclave

This document provides specifications and operating instructions for the REXMED RAU-530D autoclave. It lists the sterilization time and temperature for a 1.5 liter water supply as 78 minutes total, broken down by phase. It also provides recommended sterilization exposure times at 121°C for various materials. Standard accessories include 2 wire baskets, 1 heating element, and 1 silicone door seal. Specifications include an AC 110/230V power supply, 2KW power, a 30x71cm stainless steel chamber, and a capacity of 50 liters.

Uploaded by

Shaif Zahrah
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

REXMED Industries Co., Ltd.

P.O.BOX 487, Kaohsiung, Taiwan


Tel: +886-7-3814010 Fax: +886-7-3814013
E-mail: info@rexmed.com, sales@rexmed.com
http://www.rexmed.com RAU-530D

STERILIZATION TIME AND TEMPERATURE


Water Supply Air Evacuation Pressure Build Sterilization Hot Water & Drying Time Necessary Time
Up Time Time Steam Exhaust

1.5 liters 19 min 8 min 20 min 1 min 60 min Total 78 min

RECOMMENDED PERIODS OF STERILIZE EXPOSURE 121°C


Material to be sterilized Time in minutes
Instruments in tray 20 min
Syringes & needles 20 min
Rubber gloves 20 min
Fabrics woven 30 min
Solutions 30 min
STANDARD ACCESSORIES
Wire basket 2pcs

Heating element 1pc

Silicone door seal 1pc

SPECIFICATIONS
Voltage AC 110/230V Auto add water Manual
Power supply 2 KW Chamber size 30 x 71 cm SS#304

Sterilizing pressure 1.2~2.1 kg/cm² Capacity 50 Liters

Sterilizing 121~134°C N.W. 70 KGS

Auto dry Yes Packing size L60 X W47 X H112 cm