HPLC Column Protection Guide-1
HPLC Column Protection Guide-1
Version 0610
Includes:
• Mobile Phase Limitations
• Column Storage Tips
• Column Protection
Devices
column protection guide
TABLE OF CONTENTS
INTRODUCTION TO COLUMN CARE AND USAGE 1-4
Selecting the Right Tubing and Fittings
Column Installation
Figure 1
Tangent lines
are used in
determining
Peak width at peak width
at base (W) 10 % peak
half height (W½) W½ height
is actual width of
peak at half-height
inject
A B
( )
B t 2
As = N = 5.54 R1
A W½ W½
1
SELECTING THE RIGHT TUBING AND
FITTINGS
The tubing and fittings on an HPLC system contribute to system
dead volume. If not minimized, dead volume can lead to band
broadening and peak degradation. Please use the following
guideline to keep system dead volume to a minimum and to help
ensure optimum column performance.
TUBING
The choice of tubing material is based on its chemical
resistivity, application and HPLC system considerations
(i.e. flow rate, backpressure, etc). Please refer to Tables 1-3
for specifics.
TUBING DIAMETER
Table 1
High Pressure Tubing:
TUBING COMPATIBILITY
Table 2
Stainless Steel AVOID high concentrations
(Type 316) of acids or halogenated salts
TUBING APPLICATIONS
Table 3
Tubing ID Column IDs Typical Flow Rates
(Inch) (mm) (mL/min)
2
Fittings
All Phenomenex column end-fittings are female inverted (internal
type) with 10-32 type threading:
• The end-fitting can fit any 1/16 in. OD tubing (see page 2 for
tubing considerations)
• A 10-32 threaded male nut and ferrule or a polymeric
fingertight male nut* is used to swage or tighten the tubing
onto the fitting (see Figure 2)
Figure 2
Polymeric
Fingertight
Male Nut
32 threads per inch
No. 10 or 0.190 in. diameter
Ferrule
1/16 in. OD
Tubing Male Nut
(0.010 in. ID) /4 in. OD Standard Analytical
1
Seating depth of the nib
(250 x 4.6 mm ID)
is the critical dimension
10-32 Threaded Male Nut Female / Inverted (internal)
and Ferrule Column End-fitting
Installation
Figure 3
Considerations:
• The shape of the swaged
ferrule can differ between
manufacturers. For
Phenomenex columns, you
may use Phenomenex or
Valco type ferrules.
• VERY IMPORTANT:
The seating depth of
the nib (Figure 2) for
Phenomenex columns is
0.080 in. Tubing MUST be
seated all the way down
into the column end-fitting.
Failure to do so will result
in having a small mixing
chamber at the top or
bottom of the column.
This will lead to degraded
chromatography.
*Polymeric fingertight
fittings are easy to use.
They come in one piece,
require NO tools for
attachment and easily
conform to the shape of
the column end-fitting.
3
Column Installation
4
backpressure and flow rates
To maximize column life, flow rates should be adjusted to keep
pressures below 3500 psi.
Table 4
Particle Internal Typical Flow Typical Pressure(psi)
Size µm Diameter(mm) Rate(mL/min) 150 mm* 250 mm*
3 | 4.6 | 0.5 | 985 | 1640
5 1.0 0.1 1500 2500
5 2.0 0.2 750 1250
5 3.0 0.5 732 1226
5 4.6 1.0 710 1180
5 10.0 5.0 750 1250
10 4.6 2.0 355 590
10 21.2 20.0 170 280
* column length
Columns can be operated at any flow rate that is consistent with
the backpressure limitations described above. Flow rates should
be optimized to provide the best efficiency for your sample.
Column Storage
• Column storage conditions affect column lifetime
• Never store columns with buffers
• Flush with 5 column volumes of mobile phase without
buffer to remove any buffers or salts
Storage Conditions for Silica-Based HPLC Columns:
Table 6
Column Type Storage Solvent
Reversed Phase 65 % Acetonitrile/
C18, C12, C8, C4, C2, C1, 35 % Water
Phenyl, PFP
Normal Phase
Silica, CN, NH2, PAC, Isopropanol or Hexane
Diol Alumina
Ion-Exchange
Methanol*
SAX, SCX, WAX, WCX
Size-Exclusion 0.05 % NaN3 in water or
Diol 10 % methanol
HILIC 80 % Acetonitrile/
Luna HILIC 20 % Water
*Flush column with 50 mL HPLC grade water prior to storage solvent 5
Column Cleaning Procedures
The following conditions apply to Phenomenex silica-based
columns with the exception of chiral columns (see Parts IV
and V):
• Before starting any kind of cleaning procedure, make sure
your in-column solvent or mobile phase is miscible with the
recommended cleaning solvent(s).
• Flow rates should be 1/5 – 1/2 of the typical flow rate.
To estimate the column volume, use the following
equation: V= r 2L V = column volume in mL
r = column radius in cm
L = column length in cm
UNBONDED SILICA
Rinse with 10 Column Volumes each of:
• Hexane
• Methylene Chloride
• Isopropanol
• Methylene Chloride
• Mobile Phase
Water Removal Procedure:
Flush column with 30 mL 2.5 % 2,2-dimethoxy-propane and
2.5 % glacial acetic acid in Hexane
BONDED NORMAL PHASE (CN, NH2, Diol, PAC)
Rinse with 10 Column Volumes each of:
• Chloroform
• Isopropanol
• Methylene Chloride
• Mobile Phase
Exception: Luna Amino in reversed phase mode.
HILIC
Rinse with 10 Column Volumes each of:
• 95 % Water/5 % Acetonitrile (for buffer removal)
• 95 % 100 mM Ammonium Acetate, pH 5.8/5 % Acetonitrile
• 95 % Water/5 % Acetonitrile
• Mobile Phase
REVERSED PHASE
(C18, C12, C8, C4, C5, C2, C1, Phenyl, PFP, CN, NH2)
Rinse with 10 Column Volumes each of:
• 95 % Water/5 % Acetonitrile (for buffer removal)
• THF
• 95 % Acetonitrile/5 % Water
• Mobile Phase
REVERSED PHASE PROTEIN/PEPTIDE
(C18, C12, C8, C5, C4, Phenyl)
Rinse with 20 column volumes of mobile phase with buffer
removed. Run gradient (2x):
A) 0.1 % TFA in water
B) 0.1 % TFA in Acetonitrile/Isopropanol (1:2)
25 % B to 100 % B for 30 minutes
Equilibrate with 10 column volumes of mobile phase
Do not store column in TFA
ION EXCHANGE (SAX, SCX, NH2, WAX, WCX)
Rinse with 10 Column Volumes each of:
• 500 mM Phosphate Buffer pH 7
• 10 % Acetic Acid (Aq)
• 5 Column Volumes of Water
• 10 Column Volumes of Phosphate Buffer pH 7
• 5 Column Volumes of Water
• 10 Column Volumes of Methanol
• 10 Column Volumes of Water
For Protein Removal
Follow the above procedure with this exception:
Substitute 10 Column Volumes of Methanol with 10 Column
Volumes of 5 M Urea or 5 M Guanidine Thiocyanate
6
GFC/SEC (BioSep SEC*) *See Part VI for more details
Rinse with 5 column volumes of 0.1 M Phosphate Buffer
pH 3.0. For strongly retained proteins, run the following
gradient: 100 % Water to 100 % Acetonitrile to 100 % Water
over 60 minutes OR wash with 5 column volumes of SDS or 6 M
Guanidine Thiocyanate or 10 % DMSO
Sample Considerations
Always prefilter samples with Phenex™ syringe filters to avoid
particulate contaminants which may clog the column. Use of a
guard column is highly recommended to prolong the life of your
analytical to preparative column.
Equilibrating Column
SFC column stationary phases have a polar surface and may
be shipped under reversed phase or normal phase conditions.
Flush all columns with 10-30 column volumes of Methanol/
CO2 as intermediate solvent between CO2 and column shipping
conditions. Be aware of backpressure settings.
Equilibrate column to starting conditions with 10 column volumes
of mobile phase.
Cleaning Procedure
• Under extreme conditions the column can be flushed with
50/50 Acetonitrile/Isopropyl Alcohol followed by 100 %
Isopropyl Alcohol. Maintain backpressure below limits.
• Re- Equilibrate column to starting conditions with 10
column volumes of mobile phase
Column Storage
• Completely remove all buffers, acids, bases or other mobile
phase additives to prevent damage to media
• Flush with at least 10 column volumes of Methanol after
the last sample is purified
• Store column with end plugs firmly seated in endfittings to
ensure storage solvent does not evaporate
7
PART III - AXIA PACKED
PREPARATIVE COLUMN
Running Parameters
• Backpressure Limitation: 3500 psi
• Flow Rate Limitations: Determined by the viscosity
of mobile phase; flow rates to be controlled so that
backpressure limit of 3500 psi is not exceeded
• pH Limitations: Dictated by the media packed in the column
Cleaning Procedure
(for Axia reversed phase columns)
• Reverse flush the column with 10 column volumes of HPLC
grade water and then 10 column volumes of 100 % organic
solvent
• Under extreme conditions, the column can be flushed with
10 column volumes of 100 % THF (or IPA) followed by
100 % methylene chloride
• After cleaning, wash with 100 % THF (or IPA) and 50:50
Acetonitrile/Water, prior to equilibrating with the starting
mobile phase
COLUMN STORAGE
• Completely remove all buffers, acids, bases, or other
mobile phase additives to prevent physical damage to the
media
• Flush with at least 10 column volumes of 50:50
Acetonitrile/Water after the last sample is purified
• Store with column end plugs placed back in the end-fittings
to ensure that the packing media does not dry out
For additional information, consult the Care and Use of Axia Packed
Preparative HPLC Columns, included with each Axia column
purchased.
Column Storage
• Column storage for a longer period of time is recommended
in n-hexane/2-propanol (9:1, v/v).
• Columns used in reversed phase conditions should be first
flushed with water (whenever a buffer salt was used as
RP mobile phase additive) and then with methanol (or with
methanol only when no salt was used). The column can be
stored in methanol.
9
PART V - CHIREX Chiral
Columns
RUNNING PARAMETERS
• Temperature must not exceed 50 °C
• Column pressure must not exceed 3000 psi
• Maintain flow rate between 0.5-2.0 mL/min for 4.6 mm
ID columns
Sample Considerations
Always prefilter samples with Phenex™ syringe filters to avoid
particulate contaminants which may clog the column. Use of a
guard column is highly recommended to prolong the life of your
analytical to preparative column.
PART VI - BioSep-SEC-S
columns
RUNNING PARAMETERS
• Maximum flow rate is a function of pressure
• Column pressure must not exceed 1500 psi
• Maximum temperature: 50 °C for 316 stainless steel
Sample Considerations
Always prefilter samples with Phenex™ syringe filters to avoid
particulate contaminants which may clog the column. Use of a
guard column is highly recommended to prolong the life of your
analytical to preparative column.
Cleaning Procedure
• General protein removal: wash with 30 mL of 0.1 M
NaH2PO4, pH 3.0
• Hydrophobic protein removal: use Acetonitrile gradient
• Strongly adsorbed proteins: wash with 30 mL of 0.5 % SDS
or 6 M guanidine thiocyanate, or 10 % DMSO
Column Storage
• Overnight storage: run mobile phase at 0.2 mL/min
• Prolonged storage: use 0.05 % sodium azide in water or
10 % methanol in water
10
PART VII - REZEX
POLYMER-BASED COLUMN
Running Parameters
• Columns must be run at elevated temperatures
• (60-85 °C) except Rezex ROA and RHM for most
applications
• Column pressure for 8 % cross-linked material must not
exceed 1,000 psi; must not exceed 300 psi for 4 % cross-
linked material
• Clean and reverse flush column regularly with HPLC grade
water
Important: Never exceed maximum pressure limitations. This will
cause irreversible damage to the column.
Sample Considerations
Always prefilter samples with Phenex™ syringe filters to avoid
particulate contaminants which may clog the column. Use of a
guard column is highly recommended to prolong the life of your
analytical column.
Cleaning Procedure
Before utilizing any cleaning procedure outlined in the
Tables on pages 13 and 15, first try to clean your Rezex column
as follows:
Remove the guard column and reverse the direction of flow on
the analytical column. Run 100 % HPLC grade water through
the column as follows:
Table 7
Rezex Column Flow (mL/min) Temp. (°C)
Typical Pressure (psi @ Max Flow Rate) 400 200 200 400 400
Max. Pressure (psi @ Max Flow Rate) 1,000 300 300 1,000 1,000
Max. Flow Rate (mL/min)* 1.0 0.3 0.3 1.0 1.0
Max. Temperature (°C) 85 85 85 85 85
Typical Mobile Phase Water Water Water Water Water
pH Range Neutral Neutral Neutral Neutral Neutral
Guard Column Part No. 03B-0130-K0 03R-0133-N0 03R-0137-N0 03B-0136-K0 03B-0131-K0
SPECIFICATIONS AND OPERATING PARAMETERS
13
14
Table 9 RPM Monosaccharide RHM Monosaccharide ROA Organic Acid RFQ Fast Acid RCU Sugar Alcohols
Part Number 00H-0135-K0 00H-0132-K0 00H-0138-K0 00D-0223-K0 00G-0130-D0
Ionic Form Lead Hydrogen Hydrogen Hydrogen Calcium
Standard Dimensions 300 x 7.8 mm 300 x 7.8 mm 300 x 7.8 mm 100 x 7.8 mm 250 x 4.0 mm
Matrix Sulfonated Styrene Divinyl Benzene
Cross Linking 8% 8% 8% 8% 8%
Particle Size (µm) 8 8 8 8 8
PART VII - REZEX
Min. Efficiency (p/m) (based on last peak) 35,000 35,000 50,000 (Acetic Acid) 30,000 12,000
Typical Pressure (psi @ Max Flow Rate) 400 400 400 400 400
Max. Pressure (psi @ Max Flow Rate) 1,000 1,000 1,000 1,000 1,000
Max. Flow Rate (mL/min)* 1.0 1.0 1.0 1.0 0.5
Max. Temperature (°C) 85 85 85 85 85
Typical Mobile Phase Water Water 0.005N H2S04 0.005N H2S04 Water
pH Range Neutral 1–8 1–8 1–8 Neutral
Guard Column Part No. 03B-0135-K0 03B-0132-K0 03B-0138-K0 03B-0223-K0 03A-0130-D0
Specifications and OPERATING PARAMETERS
15
PART VIII - PolySep-GFC-p
columns
RUNNING PARAMETERS
• Column pressure must not exceed 650 psi
• Do not exceed 60 °C
PolySep Phase
1000 2000 3000 4000 5000 6000 Linear
Methanol 20 % 95 % 70 % 70 % 70 % 70 % 70 %
Acetonitrile 20 % 70 % 70 % 70 % 70 % 70 % 70 %
Cleaning Procedure
0.5 % SDS or 6 M guanidine thiocyanate. All PolySep columns
except for PolySep 1000 may also be cleaned with 50 %
acetonitrile. Make sure not to exceed a maximum pressure of
650 psi when cleaning.
Column Storage
• Overnight storage: run buffer at low flow rate (0.2 mL/min
or less)
• Prolonged storage: store in 0.05 % sodium azide in water
or 10 % methanol in water
Sample Considerations
Always prefilter samples with Phenex™ syringe filters to avoid
particulate contaminants which may clog the column. Use of
a guard column is highly recommended to prolong the life of
your analytical to preparative column.
16
PART IX - PHENOGEL GPC
columns
Specifications
Matrix: Styrene-Divinyl Benzene Copolymer
Particle Size: 5, 10, 20 µm
Porosities: 50 Å to 106 Å, and mixed beds
Typical Pressure: 5 µm: 300 psi 10 µm: 200 psi
Maximum Pressure: 650 psi
Maximum Temperature: 140 °C (205 °C for UT)
Minimum Efficiency*: 5 µm: 45,000 P/m**
10 µm: 35,000 P/m**
Typical Flow Rates: 4.6 mm ID: 0.35 mL/min
7.8 mm ID: 1.0 mL/min
21.2 mm ID: 7.0 mL/min
End Fittings: Valco Compatible
*Tested in THF ** For 300 x 7.8 mm ID columns
Sample Considerations
Always prefilter samples with Phenex™ syringe filters to avoid
particulate contaminants which may clog the GPC column. Use
of a GUARD COLUMN is highly recommended to prolong the life
of your analytical or preparative column. For optimal results,
use the chart below to determine sample concentrations and
injection volumes.
Table 10
Molecular Weight Concentration (w/v) Max Injection Volume
<50 K 0.5 % 100 µL
50-600 K 0.25 % 100 µL
600-3000 K 0.05 % 100 µL
>3000 K 0.01 % 20 µL
Continued on p. 20
17
18
Polarity Refractive UV (nm) Boiling Viscosity Solubility in
Solvent Miscibility Table Solvent Index Index @ 20 °C Cutoff @ 1 AU Point (°C) (cPoise) Water (% w/w)
Acetic Acid 6.2 1.372 230 118 1.26 100
Table 11
Acetone 5.1 1.359 330 56 0.32 100
Solvent Polarity Chart Acetonitrile 5.8 1.344 190 82 0.37 100
Relative Compound Representative Benzene 2.7 1.501 280 80 0.65 0.18
Polarity Formula Group Solvent Compounds Butyl Acetate 4.0 1.394 254 125 0.73 0.43
NONPOLAR R - H | Alkanes | Petroleum ethers, ligroin, hexanes n-Butanol 3.9 1.399 215 118 2.98 7.81
Ar - H Aromatics Toluene, benzene
R-O-R Ethers Diethyl ether Carbon tetrachloride 1.6 1.466 263 77 0.97 0.08
R-X Alkyl halides Tetrachloromethane, chloroform Chloroform 4.1 1.446 245 61 0.57 0.815
R - COOR Esters Ethyl acetate Cyclohexane 0.2 1.426 200 81 1.00 0.01
R - CO - R Aldehydes Acetone, methyl ethyl ketone 1,2-Dichloroethane1 3.5 1.444 225 84 0.79 0.81
and ketones (MEK) Dichloromethane2 3.1 1.424 235 41 0.44 1.6
R - NH2 Amines Pyridine, triethylamine Dimethylformamide 6.4 1.431 268 155 0.92 100
R - OH Alcohols Methanol, ethanol, Dimethyl sulfoxide3 7.2 1.478 268 189 2,00 100
isopropanol, butanol Dioxane 4.8 1.422 215 101 1.54 100
Increasing Polarity
R - COHN2 Amides Dimethylformamide Ethyl Acetate 4.4 1.372 260 77 0.45 8.7
R - COOH Carboxylic Ethanoic acid Ethanol 5.2 1.360 210 78 1.20 100
acids di-Ethyl Ether 2.8 1.353 220 35 0.32 6.89
POLAR H - OH Water Water Heptane 0.0 1.387 200 98 0.39 0.0003
Hexane 0.0 1.375 200 69 0.33 0.001
Methanol 5.1 1.329 205 65 0.60 100
Methyl-t-Butyl Ether4 2.5 1.369 210 55 0.27 4.8
Methyl Ethyl Ketone5 4.7 1.379 329 80 0.45 24
Pentane 0.0 1.358 200 36 0.23 0.004
n-Propanol 4.0 1.384 210 97 2.27 100
iso-Propanol6 3.9 1.377 210 82 2.30 100
di-iso-Propyl Ether 2.2 1.368 220 68 0.37
Tetrahydrofuran 4.0 1.407 215 65 0.55 100
Toluene 2.4 1.496 285 111 0.59 0.051
Trichloroethylene 1.0 1.477 273 87 0.57 0.11
Water 9.0 1.333 200 100 1.00 100
Xylene 2.5 1.500 290 139 0.61 0.018
& Miscible
SYNONYM TABLE
Immiscible* 1 4
Ethylene Chloride tert-Butyl Methyl Ether
2 5
Methylene Chloride 2-Butanone
3 6
*Immiscible means that in some proportions two phases will be produced. Methyl Sulfoxide 2-Propanol
Xylene
Water
Trichloroethylene
Toluene
Tetrahydrofuran
di-iso-Propyl Ether
iso-Propanol6
n-Propanol
Pentane
Methyl Ethyl Ketone5
Methyl-t-Butyl Ether4
Methanol
Hexane
Heptane
di-Ethyl Ether
Ethyl Acetate
Ethanol
Dioxane
Dimethyl Sulfoxide3
Dimethylformamide
Dichloromethane2
1,2-Dichloroethane1
Cyclohexane
Chloroform
Carbon tetrachloride
n-Butanol
Butyl Acetate
Benzene
Acetonitrile
Acetone
Acetic Acid
19
Column Storage
Solvents such as THF (stabilized THF only), Chloroform, Methylene
Chloride, and Toluene are commonly used for column storage.
Be sure to follow solvent switching instructions (see below) if
using solvents other than THF. Storage solvents that remain
liquefied at ambient temperatures and are not oxidizing can be
used for storage.
BE SURE THAT ANY COLUMN THAT IS NOT USED IS CAPPED
TIGHTLY WITH END-PLUGS TO AVOID EVAPORATION OF
SOLVENTS FROM COLUMN. COLUMN DESICCATION IS THE
MOST COMMON SOURCE OF COLUMN FAILURE.
SOLVENT SWITCHING
CONSIDERATIONS FOR
NON-AQUEOUS GPC COLUMNS
Phenogel columns are rugged and exhibit wide solvent
compatibility. Different solvents, however, produce different
swell characteristics (Table 12). Improper solvent switches can
result in a void. For this reason, we recommend that you dedicate
columns to specific solvents.
If you need to switch solvents, it is VERY IMPORTANT to take the
following into consideration:
1. Reduce flow rate to 0.2 mL/min.
2. Backpressure must NEVER exceed 650 psi.
3. Always check solvent miscibility in a beaker or follow the
solvent miscibility table on page 18-19 before proceeding
with ANY solvent switch.
4. Compare the swell characteristics of solvent 1 (old solvent)
to solvent 2 (new solvent) and use the following guideline:
• If the solvent 1 and solvent 2 belong to the same swell
category (Table 12), check the solvent miscibility and
proceed with the switch.
• If solvent 1 and solvent 2 belong to successive swell
categories as indicated by the arrows on Table 12, check
the miscibility and proceed with the switch.
• If solvent 1 and solvent 2 DO NOT belong to the same OR
successive swell categories, switch to an intermediate
solvent FIRST, as indicated by the arrows on Table 12.
Table 12
A – 70-80 % SWELL
Toluene TCB
THF p-Xylene
Chloroform Cyclopentane
Benzene Pyridine
o-Dichlorobenzene
B – 60 % SWELL
Ethyl Ether Cycloheptane
Methylene Chloride Ethyl Acetate
Methyl Ethyl Ketone
C – 50 % SWELL
Acetone Dimethyl Acetamide
m-Cresol N-Methyl Pyrrolidone
o-Chlorophenol DMSO
Dimethyl Formamide Dioxane
D – 30 % SWELL
Acetonitrile Freon 113 n-Butyl Alcohol
Cyclohexane HFIP Methanol
20 n-Hexane Iso-Propanol
PHENOGEL PORE SIZE Suggested PHENOGEL PORE SIZE Suggested
Linear Operating Linear Operating
Mobile Phase Solvent 50 Å 100 Å 500 Å 103 Å 104 Å 105 Å 106 Å & Mixed Temperature 3 4 5 6 Temperature
Mobile Phase Solvent 50 Å 100 Å 500 Å 10 Å 10 Å 10 Å 10 Å & Mixed
Acetone Y Y Y Y Y Y Y Y Hexane Y Y Y Y Y Y Y Y
Benzene Y Y Y Y Y Y Y Y m-Cresol Y* Y Y Y Y Y Y Y 100 °C
Carbon Tetrachloride Y Y Y Y Y Y Y Y Methyl Ethyl Ketone Y Y Y Y Y Y Y Y
Chloroform Y Y Y Y Y Y Y Y Methylene Chloride Y Y Y Y Y Y Y Y
30 % HFIP/chloroform Y Y Y Y Y Y Y Y o-Chlorophenol Y* Y Y Y Y Y Y Y 100 °C
Diethyl Ether Y Y Y Y Y Y Y Y o-Dichlorobenzene Y* Y Y Y Y Y Y Y 135 °C
Table 13
Xylene Y Y Y Y Y Y Y Y
SOLVENT COMPATIBILITY CHART
21
PART X - Polymerx rp
columns
Specifications
• Matrix: Polystyrene Divinylbenzene (PSDVB)
• Particle Size: 3, 5, 7, 10 µm
• Pore Size: 100 Å
RUNNING PARAMETERS
• Maximum temperature: 60 °C
• Maximum pressure: 2500 psi
Cleaning Procedure
• 100 % Water to 100 % Acetonitrile. Repeat 3 times.
Column Storage
• 75:25 Acetonitrile / Water
phenex™
Syringe filters
• Increase column lifetime (save money!)
• Ensure more accurate, consistent results
• Eliminate damaging microparticlates
Particulates can damage expensive equipment, valves, columns
and pumps. They can also lead to erratic analytical results. Pre-
filtering samples prior to analysis is critical in preventing column
and frit blockage, undue wear on valve seals, and abnormally
high operating pressures.
Table 14
Sample or Mobile Filter Membrane
Phase Volume (mL) (diameter, mm) Format
≤2 4 Syringe filter
2 to 10 15-17 Syringe filter
10 to 100 25-28 Syringe filter
> 100 47 Membrane disk
> 1000 90 Membrane disk
ORDERING INFORMATION
Part No. Pore Size (µm) Phenex Membrane Housing
4 mm Diameter (500/pk)
AF0-3103-52 0.45 RC PP
AF0-3102-52 0.45 PTFE6 PP
AF3-3107-52 0.45 NY PP
AF0-3203-52 0.20 RC PP
AF0-3202-52 0.20 PTFE6 PP
AF3-3207-52 0.20 NY PP
15–17 mm Diameter (500/pk)
AF0-2103-52 0.45 RC PP
AF0-2102-52 0.45 PTFE6 PP
AF0-2107-52 0.45 NY PP
AF0-2203-52 0.20 RC PP
AF0-2202-52 0.20 PTFE6 PP
AF0-2207-52 0.20 NY PP
25–28 mm Diameter (500/pk)
AF0-8103-525 0.45 RC PP
AF0-8108-527 0.45 PES3 PP
AF0-1102-52 0.45 PTFE6 PP
AF0-1107-52 0.45 NY PP
AF0-8B09-527 0.45 GF/CA2,3,4 MBS
AF0-8203-52 5
0.20 RC PP
AF0-8208-52 7
0.20 PES3 PP
AF0-1202-52 0.20 PTFE6 PP
AF0-1207-52 0.20 NY PP
AF0-8A09-527 0.20 GF/CA2,3,4,7 MBS
AF0-8515-52 7
1.20 GF2,3 MBS
Housing is made of medical-grade polypropylene (PP), unless otherwise indicated. Above
syringe filters are non-sterile.
1. 17 mm diameter. 4. Cellulose acetate is surfactant-free.
2. Glass fiber filters are 28 mm diameter 5. 26 mm diameter.
and made of borosilicate. They will 6. Hydrophobic membrane. Can be made
remove 90 % of all particles >1.2 µm. hydrophilic by pre-wetting with IPA.
3. Housing material is methacrylate 7. 28 mm diameter.
butadiene styrene (MBS) polymerisate.
Also known as Cryolite™.
23
PHENEX™ DISPOSABLE CENTRIFUGAL
FILTER UNITS
• Convenient filtration of multiple HPLC and GC samples
• High recovery for small samples
• Nylon, Cellulose Acetate, and PTFE (Teflon®)
membrane materials
ORDERING INFORMATION
Pore Volumes (mL) Membrane
Part No. Size (μm) Sample/Receiver Non-Sterile Unit
AF0-0438 0.2 2.0 / 5.0 Nylon 25/pk
AF0-0439 0.45 2.0 / 5.0 Nylon 25/pk
AF0-0440 0.2 2.0 / 5.0 PTFE 25/pk
AF0-0441 0.45 2.0 / 5.0 PTFE 25/pk
AF0-8353 0.2 2.0 / 5.0 CA 25/pk
AF0-8354 0.45 2.0 / 5.0 CA 25/pk
Above centrifugal filters are non-sterile.
24
Guard Cartridge
System
Clean Dirty
CONVENIENCE
Knowing when to replace your guard is no longer a mystery!
SecurityGuard’s direct-view feature lets you inspect the packing
material for visual contaminants and indicates when it’s time
to replace the cartridge. No other guard cartridge has this
convenient feature.
EXTRA PROTECTION
SecurityGuard offers the option of stacking two cartridges in
the same holder, using the simple stacking ring provided. Extra
length provides extra protection. When the first cartridge becomes
exhausted, contaminants are retained by the second cartridge.
VERSATILITY
One direct-connect holder conveniently finger-tightens into
virtually any brand of HPLC column worldwide. How can one
holder be direct-connect and universal at the same time when
end-fittings have different depths? Answer- the length of the
stainless steel nib at the end of the holder automatically adjusts
to the precise depth of a column’s endfitting. SecurityGuard’s
fingertight connection will withstand pressures up to 5000 psi and
it features a completely inert and biocompatible flowpath.
ACCURACY
The cartridges can be used with virtually any matching phase of
virtually any brand of column without affecting efficiency, retention
time or backpressure. There are 34 different phases to choose
from, including cartridges for general purpose, pharmaceutical,
protein and polypeptide, aqueous size exclusion, carbohydrate
and organic acid applications. SecurityGuard phases can be
used with columns containing 3, 3.5, 4, 5, 10, 15 µm or larger
diameter particle sizes.
Continued on p. 26
25
Security Guard
ORDERING INFORMATION
Analytical Holder Assembly Kit
Part No. Description Unit
KJ0-4282 Guard Cartridge Kit ea
Semi-Preparative Preparative
AJ0-8420 Oligo-WAX 15 x 30 ea
(WA, Weak Anion Exchanger)
27
Security Guard ORDERING
INFORMATION (continued)
Cartridges for Protein/Polypeptide Reversed Phase
For use with all silica columns for separation of proteins and peptides, such as Jupiter®
(Phenomenex); Vydac® 218TP, 214TP (Alltech Associates, Inc.); SynChropak® 300 C18,
C4 (Eprogen, Inc.); Nucleosil® 300 Å C18, C4; HYPERSIL® 300 Å and all other widepore
or 300 Å brands.
pH Dimensions
Part No. Material Description Stability L x ID(mm) Unit
AJ0-4320 Widepore C18 (ODS) 1.5-10 4 x 2.0 10/pk
AJ0-4321 Widepore C18 (ODS) 1.5-10 4 x 3.0 10/pk
AJ0-7224 Widepore C18 (ODS) 1.5-10 10 x 10 3/pk
AJ0-7230 Widepore C18 (ODS) 1.5-10 15 x 21.2 ea
AJ0-8313 Widepore C18 (ODS) 1.5-10 15 x 30 ea
AJ0-4326 Widepore C5 (Pentyl) 1.5-10 4 x 2.0 10/pk
AJ0-4327 Widepore C5 (Pentyl) 1.5-10 4 x 3.0 10/pk
AJ0-7371 Widepore C5 (Pentyl) 1.5-10 10 x 10 ea
AJ0-4329 Widepore C4 (Butyl) 1.5-10 4 x 2.0 10/pk
AJ0-4330 Widepore C4 (Butyl) 1.5-10 4 x 3.0 10/pk
AJ0-7225 Widepore C4 (Butyl) 1.5-10 10 x 10 3/pk
AJ0-7231 Widepore C4 (Butyl) 1.5-10 15 x 21.2 ea
AJ0-8314 Widepore C4 (Butyl) 1.5-10 15 x 30 ea
28
Cartridges for Chiral (cont’d)
Dimensions
Part No. Material Description* pH Stability L x ID(mm) Unit
AJ0-8473 Lux Amylose-2 2-9 15 x 21.2 ea
AJ0-8474 Lux Amylose-2 2-9 15 x 30 ea
AJ0-8621 Lux Cellulose-3 2-9 4 x 2.0 10/pk
AJ0-8622 Lux Cellulose-3 2-9 4 x 3.0 10/pk
AJ0-8623 Lux Cellulose-3 2-9 10 x 10.0 3/pk
AJ0-8624 Lux Cellulose-3 2-9 15 x 21.2 ea
AJ0-8625 Lux Cellulose-3 2-9 15 x 30.0 ea
AJ0-8626 Lux Cellulose-4 2-9 4 x 2.0 10/pk
AJ0-8627 Lux Cellulose-4 2-9 4 x 3.0 10/pk
AJ0-8628 Lux Cellulose-4 2-9 10 x 10.0 3/pk
AJ0-8629 Lux Cellulose-4 2-9 15 x 21.2 ea
AJ0-8630 Lux Cellulose-4 2-9 15 x 30.0 ea
* Lux Cellulose-1 is cellulose tris(3,5-dimethylphenylcarbamate)
Lux Cellulose-2 is cellulose tris(3-chloro-4-methylphenylcarbamate)
Lux Amylose-2 is amylose tris(5-chloro-2-methylphenylcarbamate)
Lux Cellulose-3 is cellulose tris(4-methylbenzoate)
Lux Cellulose-4 is cellulose tris(4-chloro-3-methylphenylcarbamate)
Replacement Parts
Part No. Description Unit
AJ0-4283 PEEK Ferrules 3/pk
AJ0-4285 Stacking Rings 2/pk
AQ0-1389 PEEK Fingertight Male Nuts 10/pk
AJ0-4284 Security Guard Wrenches 2/pk
AQ0-8374 PREP Coupler, SS w/ PEEK Ferrule Inserts ea
10-32 Threads, 1/16 in. OD x 0.020 in. ID
AQ0-8375 Replacement Ferrule Inserts, for PREP 10/pk
Coupler, PEEK, 0.020 in. ID
AQ0-8222 PREP Replacement O-Rings, Kalrez® 2/pk
For 15 x 21.2 mm SG Holder, Size 2-021
AQ0-8318 PREP Replacement O-Rings, Kalrez® 2/pk
For 15 x 30 mm SG Holder, Size 2-025
29
HPLC SYSTEM TEST KIT
ORDERING INFORMATION
Part No. Description Unit
CH0-1684 | HPLC System Test Kit, Reversed | ea
Phase, includes: C18 column,
isocratic and gradient test mixes
CH0-1685 Isocratic Test Mix 5/pk
CH0-1686 Gradient Test Mix 5/pk
COLUMN PERFORMANCE
CHECK STANDARDS
• Convenient way to check column performance
• Affordable and easy to use
30
COLUMN PERFORMANCE CHECK
STANDARDS (continued)
31
COLUMN PERFORMANCE CHECK
STANDARDS (continued)
32
COLUMN PERFORMANCE CHECK
STANDARDS (continued)
33
PART XII - SOLID PHASE
EXTRACTION (SPE)
Increase column and instrument life by injecting samples
cleaned-up with Strata®.
• Optimal Flow
• Lot-to-Lot Reproducibility
• Wide Range of Selectivity
• Available chemistries include:
C18-E, C18-U, C18-T, C8,
Phenyl, SDB-L, CN, Si-1, WCX, FI-PR, NH2, SAX, SCX,
Melamine
equipment
• Column Chiller-Heater
• Column Heater
• Column Selectors
• Degasser
• Fluid Processors
• Mobile Phase Recycler
• Temperature Controllers
For ordering and additional information, please contact your Phenomenex
Technical Consultant.
ORDERING INFORMATION
ThermaSphere™ TS-130
Part No. Description
EH0-7057 ThermaSphere TS-130 HPLC Column Heater 25-90 °C,
95 to 265 VAC, 50/60 Hz
EH0-7058 Stand for ThermaSphere TS-130 HPLC Column Heater
More Accessories available. See Phenomenex Catalog for details.
35
PHENOMENEX WARRANTY
Phenomenex products are warranted to meet the stated
performance and quality and to be free of defects in
material and workmanship. They are not warranted, nor
does Phenomenex assume liability, if misused. NO OTHER
WARRANTY OR REPRESENTATION IS IMPLIED OR EXPRESSED
BY PHENOMENEX FOR ITS PRODUCTS WITH RESPECT TO
MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE,
OR ANY OTHER MATTER. PHENOMENEX SHALL NOT UNDER
ANY CIRCUMSTANCES BE LIABLE FOR ANY INCIDENTAL,
CONSEQUENTIAL, OR COMPENSATORY DAMAGE ARISING
FROM THE USE OF, OR IN CONJUNCTION WITH, ITS PRODUCTS.
The maximum liability which can be assumed by Phenomenex for
breach of warranty shall be the invoice price of the product.
36
column protection guide
Column: ____________________________________
Dimensions: _________________________________
37
www.phenomenex.com
Phenomenex products are available worldwide. For the distributor in your country,
contact Phenomenex USA, International Department by telephone, fax or email: international@phenomenex.com.
Australia Austria Canada Denmark France Germany
mail: PO Box 4084 Zeppelinstr. 5 411 Madrid Ave. Gydevang 39-41 Parc des Grillons, Bat.3 Zeppelinstr. 5
www.phenomenex.com
Lane Cove, NSW 2066
Australia
63741 Aschaffenburg
Germany
Torrance, CA
90501-1430
3450 Allerød 60support@phenomenex.com
route de Sartrouville
78232 Le Pecq Cedex
63741 Aschaffenburg
Germany
GU54810610
Denmark
USA France
tel.: 02-9428-6444 01-319-1301 (800) 543-3681 4824 8048 01 30 09 21 10 06021-58830-0
fax: 02-9428-6445 01-319-1300 (310) 328-7768 4810 6265 01 30 09 21 11 06021-58830-11
email: info@ anfrage@ info@ dkinfo@ franceinfo@ anfrage@