B12 Cobas
B12 Cobas
Vitamin B12 II
                                                                                                               MODULAR ANALYTICS E170
                                                                                                               cobas e 411
 07212771 190            100
                                                                                                               cobas e 601
                                                                                                               cobas e 602
Vitamin B12 II
H314                 Causes severe skin burns and eye damage.                           Na‑heparin, Li‑heparin, K2‑EDTA and K3‑EDTA plasma. Li‑heparin plasma
                                                                                        tubes containing separating gel can be used.
H412                 Harmful to aquatic life with long lasting effects.                 Criterion: Slope 0.9‑1.1 + intercept within < ± 2x Limit of Blank + coefficient
Prevention:                                                                             of correlation ≥ 0.95.
                                                                                        Stable for 2 hours at 15‑25 °C, 48 hours at 2‑8 °C, 56 days at
P273                 Avoid release to the environment.                                  ‑20 °C (± 5 °C). Freeze once only.
                                                                                        Stability of serum obtained with separating tubes: 24 hours at 2‑8 °C (note
P280                 Wear protective gloves/ protective clothing/ eye protection/       the data provided by the tube manufacturer).
                     face protection.                                                   The sample types listed were tested with a selection of sample collection
Response:                                                                               tubes that were commercially available at the time of testing, i.e. not all
                                                                                        available tubes of all manufacturers were tested. Sample collection systems
P301 + P330 IF SWALLOWED: Rinse mouth. Do NOT induce vomiting.                          from various manufacturers may contain differing materials which could
+ P331                                                                                  affect the test results in some cases. When processing samples in primary
                                                                                        tubes (sample collection systems), follow the instructions of the tube
P303 + P361 IF ON SKIN (or hair): Take off immediately all contaminated                 manufacturer.
+ P353      clothing. Rinse skin with water/shower.                                     Centrifuge samples containing precipitates before performing the assay.
                                                                                        Do not use heat‑inactivated samples.
P304 + P340 IF INHALED: Remove person to fresh air and keep                             Avoid hemolysis.
+ P310      comfortable for breathing.
                                                                                        Do not use samples and controls stabilized with azide.
            Immediately call a POISON CENTER/doctor.
                                                                                        Vitamin B12 determinations should be performed on serum or plasma
P305 + P351 IF IN EYES: Rinse cautiously with water for several                         samples from fasting patients.
+ P338      minutes. Remove contact lenses, if present and easy to do.                  Note: Samples with extremely high total protein concentrations (e.g.
                                                                                        patients suffering from Waldenström's macroglobulinemia) are not suitable
+ P310      Continue rinsing. Immediately call a POISON CENTER/                         for use in this assay, since they may lead to the formation of protein gel in
            doctor.                                                                     the assay cup. Processing protein gel may cause a run abort. The critical
Product safety labeling follows EU GHS guidance.                                        protein concentration is dependent upon the individual sample composition.
                                                                                        The formation of protein gel was seen in samples (spiked with human IgG
Contact phone: all countries: +49-621-7590                                              or human serum albumin) having a total protein concentration > 160 g/L.
All human material should be considered potentially infectious. All products            Ensure the samples, calibrators and controls are at 20‑25 °C prior to
derived from human blood are prepared exclusively from the blood of                     measurement.
donors tested individually and shown to be free from HBsAg and antibodies
to HCV and HIV. The testing methods used assays approved by the FDA or                  Due to possible evaporation effects, samples, calibrators and controls on
cleared in compliance with the European Directive 98/79/EC, Annex II,                   the analyzers should be analyzed/measured within 2 hours.
List A.                                                                                 Materials provided
However, as no testing method can rule out the potential risk of infection              See “Reagents – working solutions” section for reagents.
with absolute certainty, the material should be handled with the same level
of care as a patient specimen. In the event of exposure, the directives of the          Materials required (but not provided)
responsible health authorities should be followed.8,9                                   ▪       07212780190, Vitamin B12 II CalSet, for 4 x 1.0 mL
Avoid foam formation in all reagents and sample types (specimens,                       ▪       05618860190, PreciControl Varia, for 4 x 3.0 mL
calibrators and controls).
                                                                                        ▪       11732277122, Diluent Universal, 2 x 16 mL sample diluent or
Reagent handling                                                                                03183971122, Diluent Universal, 2 x 36 mL sample diluent
The reagents in the kit have been assembled into a ready‑for‑use unit that              ▪ General laboratory equipment
cannot be separated.
All information required for correct operation is read in from the respective           ▪ MODULAR ANALYTICS E170 or cobas e analyzer
reagent barcodes.                                                                       Accessories for cobas e 411 analyzer:
Storage and stability                                                                   ▪       11662988122, ProCell, 6 x 380 mL system buffer
Store at 2‑8 °C.                                                                        ▪       11662970122, CleanCell, 6 x 380 mL measuring cell cleaning
Do not freeze.                                                                             solution
Store the Elecsys reagent kit upright in order to ensure complete                       ▪       11930346122, Elecsys SysWash, 1 x 500 mL washwater additive
availability of the microparticles during automatic mixing prior to use.                ▪       11933159001, Adapter for SysClean
 Stability:                                                                             ▪       11706802001, AssayCup, 60 x 60 reaction cups
 unopened at 2‑8 °C                           up to the stated expiration date          ▪       11706799001, AssayTip, 30 x 120 pipette tips
 after opening at 2‑8 °C                      84 days (12 weeks)                        ▪       11800507001, Clean‑Liner
                                                                                        Accessories for MODULAR ANALYTICS E170, cobas e 601 and
 on the analyzers                             35 days (5 weeks) onboard                 cobas e 602 analyzers:
                                              or
                                                                                        ▪         04880340190, ProCell M, 2 x 2 L system buffer
                                              60 days when stored alternatively in
                                              the refrigerator and on the analyzer,     ▪         04880293190, CleanCell M, 2 x 2 L measuring cell cleaning
                                                                                            solution
                                              with the total time onboard on the
                                              analyzer not exceeding                    ▪         03023141001, PC/CC‑Cups, 12 cups to prewarm ProCell M and
                                                                                            CleanCell M before use
                                              10 x 8 hours
                                                                                        ▪         03005712190, ProbeWash M, 12 x 70 mL cleaning solution for run
Specimen collection and preparation                                                         finalization and rinsing during reagent change
Only the specimens listed below were tested and found acceptable.                       ▪         03004899190, PreClean M, 5 x 600 mL detection cleaning solution
Serum collected using standard sampling tubes or tubes containing                       ▪         12102137001, AssayTip/AssayCup, 48 magazines x 84 reaction
separating gel.                                                                             cups or pipette tips, waste bags
Vitamin B12 II
▪      03023150001, WasteLiner, waste bags                                       Criterion: Recovery within ± 10 % of initial value with samples > 200 pg/mL
▪      03027651001, SysClean Adapter M                                           and ≤ ± 20 pg/mL with samples ≤ 200 pg/mL.
Accessories for all analyzers:                                                   Samples should not be taken from patients receiving therapy with high
                                                                                 biotin doses (i.e. > 5 mg/day) until at least 8 hours following the last biotin
▪        11298500316, ISE Cleaning Solution/Elecsys SysClean,                    administration.
    5 x 100 mL system cleaning solution                                          No interference was observed from rheumatoid factors up to a
Assay                                                                            concentration of 1500 IU/mL.
For optimum performance of the assay follow the directions given in this         In vitro tests were performed on 16 commonly used pharmaceuticals. No
document for the analyzer concerned. Refer to the appropriate operator’s         interference with the assay was found.
manual for analyzer‑specific assay instructions.                                 In rare cases, interference due to extremely high titers of antibodies to
Resuspension of the microparticles takes place automatically prior to use.       analyte‑specific antibodies, streptavidin or ruthenium can occur. These
Read in the test‑specific parameters via the reagent barcode. If in              effects are minimized by suitable test design.
exceptional cases the barcode cannot be read, enter the 15‑digit sequence        Because intrinsic factor is typically used as the binding protein in serum
of numbers (except for the cobas e 602 analyzer).                                vitamin B12 assays, anti-intrinsic factor antibodies (which are common in
MODULAR ANALYTICS E170, cobas e 601 and cobas e 602 analyzers:                   pernicious anemia) can lead to elevated vitamin B12 measurement
PreClean M solution is necessary.                                                values.2,11,12 The Elecsys Vitamin B12 II assay is designed to avoid
Bring the cooled reagents to approximately 20 °C and place on the reagent        interference due to anti-intrinsic factor antibodies.13
disk (20 °C) of the analyzer. Avoid foam formation. The system                   For diagnostic purposes, the results should always be assessed in
automatically regulates the temperature of the reagents and the                  conjunction with the patient’s medical history, clinical examination and other
opening/closing of the bottles.                                                  findings.
Calibration                                                                      Note: The presence of immunoglobulin-vitamin B12 complexes may cause
Traceability: This method has been standardized against the Vitamin B12          unexpectedly high values of vitamin B12.14,15
assay (      04745736190).                                                       Limits and ranges
Accuracy to WHO Standard 03/178: A study was performed to evaluate the           Measuring range
accuracy of the Elecsys Vitamin B12 II assay using the Vitamin B12 World         50.0‑2000 pg/mL or 36.9‑1476 pmol/L (defined by the Limit of Blank and the
Health Organization International Standard 03/178.10 Two reagent lots were       maximum of the master curve). Values below the Limit of Blank are
used on 16 instruments. The mean recovery of the target value of                 reported as < 50.0 pg/mL or < 36.9 pmol/L. Values above the measuring
WHO IS 03/178 (480 pg/mL) was 102 %.                                             range are reported as > 2000 pg/mL or > 1476 pmol/L.
Every Elecsys reagent set has a barcoded label containing specific               Lower limits of measurement
information for calibration of the particular reagent lot. The predefined
master curve is adapted to the analyzer using the relevant CalSet.               Limit of Blank, Limit of Detection and Limit of Quantitation
Calibration frequency: Calibration must be performed once per reagent lot        Limit of Blank = 50 pg/mL (36.9 pmol/L)
using fresh reagent (i.e. not more than 24 hours since the reagent kit was       Limit of Detection = 100 pg/mL (73.8 pmol/L)
registered on the analyzer).                                                     Limit of Quantitation = 150 pg/mL (111 pmol/L)
Calibration interval may be extended based on acceptable verification of         The Limit of Blank, Limit of Detection and Limit of Quantitation were
calibration by the laboratory.                                                   determined in accordance with the CLSI (Clinical and Laboratory Standards
Renewed calibration is recommended as follows:                                   Institute) EP17‑A requirements.
▪ after 1 month (28 days) when using the same reagent lot                        The Limit of Blank is the 95th percentile value from n ≥ 60 measurements of
                                                                                 analyte‑free samples over several independent series. The Limit of Blank
▪ after 7 days (when using the same reagent kit on the analyzer)                 corresponds to the concentration below which analyte‑free samples are
▪ as required: e.g. quality control findings outside the defined limits          found with a probability of 95 %.
Quality control                                                                  The Limit of Detection is determined based on the Limit of Blank and the
For quality control, use PreciControl Varia.                                     standard deviation of low concentration samples. The Limit of Detection
                                                                                 corresponds to the lowest analyte concentration which can be detected
In addition, other suitable control material can be used.                        (value above the Limit of Blank with a probability of 95 %).
Controls for the various concentration ranges should be run individually at      The Limit of Quantitation is defined as the lowest amount of analyte in a
least once every 24 hours when the test is in use, once per reagent kit, and     sample that can be accurately quantitated with a allowable imprecision of
following each calibration.                                                      ≤ 20 %.
The control intervals and limits should be adapted to each laboratory’s          It has been determined using low concentration vitamin B12 samples.
individual requirements. Values obtained should fall within the defined
limits. Each laboratory should establish corrective measures to be taken if      Dilution
values fall outside the defined limits.                                          Samples with vitamin B12 concentrations above the measuring range can
If necessary, repeat the measurement of the samples concerned.                   be manually diluted 1:2 with Diluent Universal. The concentration of the
                                                                                 diluted sample must be > 738 pmol/L or > 1000 pg/mL. After manual
Follow the applicable government regulations and local guidelines for            dilution, multiply the results by the dilution factor 2.
quality control.
                                                                                 Note: Sample-dependent non‑linearity upon dilution is seen with samples
Calculation                                                                      having analyte levels beyond the measuring range. As Diluent Universal
The analyzer automatically calculates the analyte concentration of each          may contain low levels of endogenous vitamin B12, it is recommended that
sample (either in pmol/L or pg/mL).                                              linearity studies be performed using a known low analyte-containing serum
                                                                                 pool. Samples outside the measuring range can be diluted 1:2 with Diluent
Conversion factors:                   pmol/L x 1.36 = pg/mL                      Universal; the effect of endogenous vitamin B12 concentration is
                                      pg/mL x 0.738 = pmol/L                     insignificant at these levels.
                                                                                 Expected values
Limitations - interference
                                                                                 Because differences may exist with respect to population and dietary
The assay is unaffected by icterus (bilirubin ≤ 1112 µmol/L or ≤ 65 mg/dL),      status, it is recommended that normal ranges be determined by each
hemolysis (Hb ≤ 0.025 mmol/L or ≤ 0.04 g/dL), lipemia (Intralipid                laboratory over a suitable period of time and in a statistically significant
≤ 17.1 mmol/L or ≤ 1500 mg/dL), biotin (≤ 205 nmol/L or ≤ 50 ng/mL), IgG         number of assays before clinical significance is attached to the results of
≤ 28 g/L, IgA ≤ 16 g/L and IgM ≤ 10 g/L.                                         these tests.
Vitamin B12 II
The values shown below were performed on samples from an apparently                    MODULAR ANALYTICS E170, cobas e 601 and cobas e 602 analyzers
healthy population, using the Elecsys Vitamin B12 II assay. The calculation
is based on 135 sera (68 men, 67 women). The age range was between 20                                                      Repeatability       Intermediate
and 78 years. Pregnant women were excluded. The reference population                                                                             precision
was selected according to normal homocysteine values.                                  Sample                  Mean         SD        CV         SD        CV
  N                  Median                Range (2.5th-97.5th percentile)                                    pg/mL        pg/mL       %       pg/mL       %
            pg/mL             pmol/L       pg/mL               pmol/L                  Human serum 4           1230         19.8      1.6       28.8       2.3
 135         425               314         197-771            145-569                  Human serum 5           1890         29.8      1.6       41.5       2.2
These values should only be used as guidelines.                                        PreciControl Varia1      448         7.16      1.6       15.3       3.4
Each laboratory should investigate the transferability of the expected values          PreciControl Varia2      917         12.0      1.3       27.8       3.0
to its own patient population and if necessary determine its own reference
ranges.                                                                                MODULAR ANALYTICS E170, cobas e 601 and cobas e 602 analyzers
Specific performance data                                                                                                  Repeatability       Intermediate
Representative performance data on the analyzers are given below.                                                                                precision
Results obtained in individual laboratories may differ.
                                                                                       Sample                  Mean         SD        CV         SD        CV
Precision
Precision was determined using Elecsys reagents, pooled human sera and                                        pmol/L       pmol/L      %       pmol/L      %
controls in a protocol (EP5‑A2) of the CLSI (Clinical and Laboratory                   Human serum 1            130         4.31      3.3       6.75       5.2
Standards Institute): 2 runs per day in duplicate each for 21 days (n = 84).
The following results were obtained:                                                   Human serum 2            300         6.08      2.0       9.37       3.1
                                                                                       Human serum 3            745         9.74      1.3       15.6       2.1
                               cobas e 411 analyzer
                                                                                       Human serum 4            908         14.6      1.6       21.3       2.3
                                          Repeatability       Intermediate
                                                                precision              Human serum 5           1395         22.0      1.6       30.6       2.2
 Sample                        Mean         SD        CV       SD        CV            PreciControl Varia1      331         5.28      1.6       11.3       3.4
                               pg/mL      pg/mL       %      pg/mL           %         PreciControl Varia2      677         8.86      1.3       20.5       3.0
 Human serum 1                  176        8.86       5.0     12.7       7.2       Method comparison
 Human serum 2                  405        13.0       3.2     17.5       4.3       a) A comparison of the Elecsys Vitamin B12 assay (calibrated with
                                                                                   Vitamin B12 CalSet II; x) and the Elecsys Vitamin B12 II assay (calibrated
 Human serum 3                  960        19.7       2.1     31.0       3.2       with Vitamin B12 II CalSet; y) using clinical samples gave the following
 Human serum 4                  1230       27.4       2.2     46.4       3.8       correlations (pg/mL):
                                                                                   Number of samples measured: 100
 Human serum 5                  1940       40.9       2.1     72.6       3.7
 PreciControl Varia1            447        12.2       2.7     18.6       4.2       Passing/Bablok16                      Linear regression
 PreciControl Varia2            934        20.2       2.2     38.4       4.1       y = 0.952x + 15.1                     y = 0.957x + 11.6
                                                                                   τ = 0.977                             r = 0.999
                               cobas e 411 analyzer
                                                                                   The sample concentrations were between 69 and 1890 pg/mL (51 and
                                          Repeatability       Intermediate         1395 pmol/L).
                                                                precision          b) A comparison of the Elecsys Vitamin B12 II assay (y) and a commercially
 Sample                        Mean         SD        CV       SD        CV        available method (x) using clinical samples gave the following correlations
                                                                                   (pg/mL):
                               pmol/L     pmol/L      %      pmol/L          %     Number of samples measured: 106
 Human serum 1                  130        6.54       5.0     9.37       7.2
                                                                                   Passing/Bablok16                      Linear regression
 Human serum 2                  299        9.59       3.2     12.9       4.3
                                                                                   y = 0.923x + 4.90                     y = 0.881x + 27.6
 Human serum 3                  708        14.5       2.1     22.9       3.2
                                                                                   τ = 0.952                             r = 0.993
 Human serum 4                  908        20.2       2.2     34.2       3.8
                                                                                   The sample concentrations were between 182 and 1797 pg/mL (134 and
 Human serum 5                  1432       30.2       2.1     53.6       3.7       1326 pmol/L).
 PreciControl Varia1            330        9.00       2.7     13.7       4.2       c) A comparison of the Elecsys Vitamin B12 II assay on the cobas e 601
                                                                                   analyzer (y) and the Elecsys Vitamin B12 II assay on the cobas e 411
 PreciControl Varia2            689        14.9       2.2     28.3       4.1       analyzer (x) using clinical samples gave the following correlations (pg/mL):
  MODULAR ANALYTICS E170, cobas e 601 and cobas e 602 analyzers                    Number of samples measured: 117
                                          Repeatability       Intermediate         Passing/Bablok16                      Linear regression
                                                                precision          y = 1.01x - 2.77                      y = 1.01x + 3.22
 Sample                        Mean         SD        CV       SD        CV        τ = 0.933                             r = 0.995
                               pg/mL      pg/mL       %      pg/mL           %     The sample concentrations were between 56 and 1887 pg/mL (41 and
 Human serum 1                  176        5.84       3.3     9.14       5.2       1393 pmol/L).
 Human serum 2                  407        8.24       2.0     12.7       3.1       Analytical specificity
                                                                                   The following cross-reactivities were found, tested with vitamin B12
 Human serum 3                  1010       13.2       1.3     21.1       2.1       concentrations of 129 pg/mL and 550 pg/mL.
Vitamin B12 II
 Cross-reactant                  Maximum                Cross-reactivity                                     Volume after reconstitution or mixing
                             concentration tested             %
                                                                                   GTIN                      Global Trade Item Number
                                   ng/mL
 Cobinamide dicyanide                  210                    0.003              COBAS, COBAS E, ELECSYS and PRECICONTROL are trademarks of Roche. INTRALIPID is a trademark of
                                                                                 Fresenius Kabi AB.
References                                                                       All other product names and trademarks are the property of their respective owners.
                                                                                 Additions, deletions or changes are indicated by a change bar in the margin.
1 Thomas L. Clinical Laboratory Diagnostics: Use and Assessment of
     Clinical Laboratory Results; 1st Edition, Frankfurt/Main: TH-Books-         © 2017, Roche Diagnostics
     Verl.-Ges.,1998:424-431.
2 Stabler SP. Vitamin B12 deficiency. N Engl J Med 2013;368:149-160.
3 Allen LH. Vitamin B-12. Adv Nutr 2012;3(1):54-55. doi:                                    Roche Diagnostics GmbH, Sandhofer Strasse 116, D-68305 Mannheim
     10.3945/an.111.001370. Epub 2012 Jan 5.                                                www.roche.com
4 Allen LH. How common is vitamin B-12 deficiency? Am J Clin Nutr
     2009;89(2):693S-696S. Epub 2008 Dec 30.
5 Chatthanawaree W. Biomarkers of cobalamin (vitamin B12) deficiency
     and its application. J Nutr Health Aging 2011 Mar;15(3):227-313.
6 Yetley EA, Pfeiffer CM, Phinney KW, et al., Biomarkers of vitamin B-12
     status in NHANES: a roundtable summary. Am J Clin Nutr 2011
     Jul;94(1):313S-321S.
7 Hvas AM, Nexo E. Diagnosis and treatment of vitamin B12 deficiency -
     an update. Haematologica 2006;91(11):1506-1512.
8 Occupational Safety and Health Standards: Bloodborne pathogens.
     (29 CFR Part 1910.1030). Fed. Register.
9 Directive 2000/54/EC of the European Parliament and Council of
     18 September 2000 on the protection of workers from risks related to
     exposure to biological agents at work.
10 Thorpe SJ, Heath A, Blackmore S, et al. International Standard for
     serum vitamin B12 and serum folate: international collaborative study to
     evaluate a batch of lyophilised serum for B12 and folate content. Clin
     Chem Lab Med 2007;45(3):380-386.
11 Yang DT, Cook RJ. Spurious elevations of vitamin B12 with pernicious
     anemia. N Engl J Med 2012;366:1742-1743.
12 Carmel R, Agrawal YP. Failures of cobalamin assays in pernicious
     anemia. N Engl J Med 2012;367:385-386. [Erratum, N Engl J Med
     2012;367:976.]
13 Schilling KA, Wiesgigl M. The Elecsys® Vitamin B12 assay is not
     affected by anti-intrinsic factor antibodies. Clin Chem Lab Med 2013
     Jun 29;51(11):e251-e252.
14 Jeffery J, Millar H, MacKenzie P, et al. An IgG complexed form of
     vitamin B12 is a common cause of elevated serum concentrations. Clin
     Biochem 2010 Jan;43(1-2):82-88. doi:
     10.1016/j.clinbiochem.2009.08.022. Epub 2009 Sep 8.
15 Bowen RA, Drake SK, Vanjani R, et al. Markedly increased vitamin B12
     concentrations attributable to IgG-IgM-vitamin B12 immune complexes.
     Clin Chem 2006;52(11):2107-2114.
16 Bablok W, Passing H, Bender R, et al. A general regression procedure
     for method transformation. Application of linear regression procedures
     for method comparison studies in clinical chemistry, Part III.
     J Clin Chem Clin Biochem 1988 Nov;26(11):783-790.
For further information, please refer to the appropriate operator’s manual for
the analyzer concerned, the respective application sheets, the product
information and the Method Sheets of all necessary components (if
available in your country).
A point (period/stop) is always used in this Method Sheet as the decimal
separator to mark the border between the integral and the fractional parts of
a decimal numeral. Separators for thousands are not used.
Symbols
Roche Diagnostics uses the following symbols and signs in addition to
those listed in the ISO 15223‑1 standard (for USA: see
https://usdiagnostics.roche.com for definition of symbols used):
                     Contents of kit
                     Analyzers/Instruments on which reagents can be used
                     Reagent
                     Calibrator